Abstract
A multi-country outbreak of Mycobacterium chimaera infection associated with contaminated heater–cooler devices (HCDs) has been reported, with more than 70 cases in Europe and the United States and two cases in Canada to date. The epidemiological and microbiological characteristics of this outbreak provide evidence for common-source transmission of M. chimaera from the exhaust air of intrinsically contaminated HCDs to patients during cardiac surgery. To date, all reported cases have been associated with Stöckert 3T HCDs manufactured at one plant by LivaNova prior to September 2014. Implantation of prosthetic material increases the risk of infection. Infections usually present as prosthetic valve endocarditis, vascular graft infection or disseminated infection. Reported mortality rates have varied, but were often over 40%.
Several measures are recommended to facilitate case-finding and mitigate risk of exposure. The feasibility of some risk mitigation measures and their effectiveness in reducing the risk of exposure are yet to be determined. Until HCDs are redesigned in a manner that prevents water contamination and aerosolization, separating the HCD exhaust air from the operating room air during surgery may be the most effective risk mitigation strategy. However, possible unintended consequences of this approach should be considered. This overview summarizes findings from peer-reviewed and other relevant national documents on key features of the outbreak, including the source, identified risk factors for infection, signs and symptoms of infection, burden of disease, risk mitigation measures, management challenges and knowledge gaps.
Introduction
Health care–associated infections related to medical device contamination and biofilm formation have been documented in the literature (1). Recently, heater–cooler devices (HCDs) used during cardiopulmonary bypass (CPB) for cardiac surgeries and during extracorporeal membrane oxygenation (ECMO) have come under scrutiny due to infections linked to contaminated devices (2,3).
Heater-cooler devices have water tanks that pump temperature-controlled water through closed circuits to external heat exchangers that regulate patient body temperature by convection (4). The device is equipped with a radiator and fan to facilitate constant dissipation of excess heat through grid openings and the stirring of water in the tank results in aerosolization via the exhaust air (4,5). HCDs are subject to biofilm formation. A biofilm is an aggregate of microorganisms embedded within an extracellular matrix that adhere to each other and to internal surfaces, such as the interior of HCDs.
Several types of microorganisms have been isolated from contaminated HCDs, including nontuberculous mycobacteria (NTM), which are ubiquitous in soil and water and have been linked to health care-associated infections (6-9). Investigations of NTM infection clusters following cardiac surgery detected Mycobacterium chimaera as the causative microorganism. M. chimaera is a slow-growing NTM included in the Mycobacterium avium complex (3,9-12). It is less susceptible to disinfection procedures due to its cell wall constituents and its ability to form biofilms. Isolation and identification of M. chimaera from clinical specimens requires specialized microbiological techniques (3). Transmission was associated with a single model of HCD manufactured by Sorin (now LivaNova) (3,13). Cultures from HCD water tanks, water circuits and air samples taken while HCDs were in use have grown M. chimaera (5,8,9,11).
Although M. chimaera contamination of ECMO devices has been reported, contamination did not spread to the air in the room while the devices were running and no ECMO-associated M. chimaera infections were reported (2,14). Nonetheless, the need to assess potential patient exposure from ECMO has been recognized since patients treated with ECMO, who are often critically ill and highly immunocompromised, may be exposed to the device for an extended period of time (2). National guidance documents and safety communications describing risk mitigation measures and testing recommendations in Canada, United Kingdom (UK), United States (US) and Australia have been published (13,15-20).
The objective of this overview is to summarize relevant literature on the current multi-country outbreak of M. chimaera infection. The source of exposure, risk factors for infection, signs and symptoms of infection, disease burden, risk mitigation measures, challenges and gaps are summarized. This overview may be a helpful resource for Canadian health care facilities and providers who use HCDs. It may also support informed decision-making by authorities responsible for implementing infection prevention and control measures.
Scope
A worldwide literature search was undertaken by the Health Library (Health Canada) using Ovid MEDLINE, EMBASE and Global Health databases for studies published from January 1, 2007 to March 8, 2017. The search strategy was developed using database-specific thesauri for “Mycobacterium chimaera”, “heater-cooler devices”, and “cardiac surgery”. The search was limited to studies in English and French with no filters applied to limit retrieval by study design. A grey literature search was also conducted by the Health Library to identify relevant national guidance documents and safety communications. The reference lists of relevant guidance documents were hand searched for additional relevant studies.
Full texts of all relevant studies were screened to identify those reporting on HCD-associated M. chimaera infection in postoperative cardiac surgery patients and any risk mitigation measures described. A narrative synthesis of the relevant peer-reviewed publications, national guidance documents and/or safety communications was done.
Findings
A total of 95 articles were retrieved from peer-reviewed and grey literature searches, including a reference list search of identified documents. Information from 38 relevant documents was included in this overview. Fifty-seven articles were excluded for one of several reasons including studies that reported on case(s) already described in detail elsewhere; studies that focused on NTM in general (not specifically M. chimaera); studies that did not discuss patient exposure or transmission; and national guidance documents or safety communications that did not provide additional information to that obtained from similar documents from Canada, the US, Australia and Europe.
Source of exposure
To date, all cases of M. chimaera infection reported internationally have been associated with Stöckert 3T HCDs manufactured in Germany by LivaNova before September 2014 (3,9,13,15,21-23). Phylogenetic analysis by whole genome sequencing and other means showed that isolates from infected patients and from water and exhaust air of used and new Stöckert 3T HCDs were closely related, suggesting global distribution of contaminated HCDs and a hospital-independent, common source for the current outbreak (5,9,12,22,24-26). LivaNova implemented changes to their disinfection processes in an attempt to reduce the risk of M. chimaera contamination of 3T HCDs manufactured after September 2014 (13,15,27,28). Tests conducted on HCDs manufactured by a different company detected M. chimaera in the water but not in air samples, and the isolate obtained was genetically distinct from isolates obtained from Stöckert 3T HCDs (12,25,29).
During surgeries, the HCD is often positioned adjacent to the cardiopulmonary bypass machine and the patient. Recently, one of the considerations related to minimizing patient exposure to exhaust air from the HCD has to do with the feasibility of positioning the HCD immediately beside the floor-level exhaust in the operating room.
Risk factors for infection
Cases of M. chimaera infection following exposure to HCDs during cardiopulmonary bypass have been reported in patients who had undergone surgery in Europe (UK, France, Switzerland, Netherlands, Germany, Ireland and Spain) as well as in the US, Australia, Canada and Hong Kong Special Administrative Region (18). Patients undergoing cardiac surgery involving cardiopulmonary bypass where body temperature is regulated by HCDs are at risk of exposure and infection (8). Patients undergoing cardiopulmonary bypass for over two hours had higher odds of NTM infection (odds ratio: 16.5; 95% CI: 3.2–84) (8). In hospitals where at least one HCD-associated M. chimaera infection was identified, the risk of a patient getting an infection was approximately 0.1–1% (11,30,31). Of 115,664 surgical procedures in England involving repair or replacement of cardiac valves (between 2007 and 2014), the risk of NTM infections increased from less than 0.2/10,000 person-years before 2010 to 1.65/10,000 person-years in 2013 (29).
Implantation of prosthetic material (e.g., heart valve, vascular graft, left ventricular assist device) increased the risk of infection (3,11,13,29). Limited data suggest that heart transplants may also increase the risk of infection (3,32).
No case has occurred in operating room personnel exposed to aerosolization from HCDs.
Signs and symptoms of infection
Signs and/or symptoms of invasive M. chimaera infection following exposure to aerosols from an HCD may not occur for months or years after exposure, with a mean time between exposure and diagnosis of 1.6 years (range: 0.1–6.3 years) (3,10,14,23,32). The infection usually presents as prosthetic valve endocarditis, vascular graft infection or disseminated infection although a variety of extracardiac sites may also be infected (Table 1) (9-11,13,18,29,33). Clinical manifestations of infection are diverse and symptoms may be nonspecific (12,23). In some cases, extracardiac manifestations preceded cardiovascular disease (11). A description of a compatible syndrome for NTM infection published by the Canadian Public Health Laboratory Network (CPHLN) is shown in Table 1 (16).
Table 1. Clinical symptoms of patients with Mycobacterium chimaera infection.
| Type of symptoms | Clinical symptoms |
|---|---|
| Constitutional | Recurrent or prolonged fever, fatigue, shortness of breath, weight loss, night sweats |
| Cardiac | Prosthetic valve endocarditis and/or prosthetic vascular graft infection |
| Extracardiac | Bone infection, sternotomy surgical wound infection, mediastinitis, hepatitis, bloodstream infection, ocular infection (panuveitis, multifocal choroiditis, chorioretinitis) |
| Immunologic/embolic | Splenomegaly, cytopenia |
| Infants | Febrile episodes and failure to thrive |
Reproduced with permission from the Canadian Public Health Laboratory Network (16)
Disease burden
M. chimaera infection requires aggressive medical treatment with combination antimycobacterial therapy and sometimes repeat surgical intervention. The infection generally results in substantial morbidity with prolonged hospitalization, adverse effects of medical and surgical treatment, and/or treatment failure (3,11,18,29). In Europe, at least 52 cases have been reported as of January 2017 (12,18). Three cases have been identified in Australia, 24 in the US and two in Canada (20,23,32,34). Individual patient information was not always reported. From the data available, most cases were in older adults although patient age ranged from one to 81 years old, including two pediatric patients. Approximately 83% of the patients were male. Most studies reported a mortality rate over 40% (see Table 2) (3,11,12,29,32) and mortality was high when significant delays in diagnosis occurred and patients were severely ill when appropriate antimycobacterial treatment was implemented. It remains unclear whether increased awareness and earlier diagnosis will reduce the mortality associated with M. chimaera infection.
Table 2. Reported mortality from Mycobacterium chimaera infection associated with heater–cooler devices.
| Reference (country / region) | Number of patients diagnosed | Number of deaths (mortality) % |
|---|---|---|
| Kohler et al., 2015 (Europe) (11) | 101 | 4 (40%)2 |
| Chand et al., 2016 (Europe) (29) | 183 | 9 (50%) |
| Appenheimer et al., 2016 (US) (32) | 24 | NR (46%)4 |
| European Centre for Disease Prevention and Control, 2016 (Europe) (18) | 52 5 | 10 (<19%)6 |
| Haller et al., 2016 (Germany) (9) | 5 | 1 (20%)7 |
| Tan et al., 2016 (US) (33) | 3 | 2 (67%)6,7 |
| Public Health England, 2017 (Europe) (12) | 26 | 15 (58%) |
| Australian Commission on Safety and Quality in Health Care, 2017 (Australia) (20) | 3 | 0 (0%) |
Abbreviations: NR, not reported; US, United States; %, percentage 1 Nine cases were confirmed and one was probable 2 An additional death was not linked to M. chimaera infection 3 All cases were probable 4 Number of deaths were not reported 5 Some of these cases have been reported in other publications 6 Cause of death not described or not all deaths attributed to the infection 7 Percent mortality not reported in the study, was calculated for inclusion in this table
Risk mitigation measures
Key measures identified to facilitate case-finding and mitigate future exposure to M. chimaera are summarized in Table 3.
Table 3. Recommended measures to facilitate case-finding and mitigate future risk of Mycobacterium chimaera exposure.
| Risk mitigation measure | Additional context and/or limitation |
|---|---|
| Health care provider notification and education (11,12,28,32) |
|
| Patient notification (8,12,28,32) |
|
| Enhanced prospective NTM surveillance (9,21) |
|
| Ensure traceability of HCDs in use (12) |
|
| Remove potentially contaminated HCDs from service (12,15,27) |
|
| Replace contaminated HCDs, plus accessories, tubing and connectors, to prevent recontamination (13,15,27,35) |
|
| Use manufacturer’s operation protocol including updated cleaning and disinfection procedures (3,9,12,15,27,28,35) |
|
| Routine microbiological testing of HCDs in use (12,15,17,25,27,36) |
|
| Apply engineering solutions to enable reliable separation of HCD exhaust air from operating room air (4,5,12,13,15,18,25,26,37) | Options include:
|
Abbreviations: CPB, cardiopulmonary bypass; ECDC, European Centre for Disease Prevention and Control; HCD, heater–cooler device; NTM, nontuberculous mycobacteria; US FDA, United States Food and Drug Administration
Challenges and gaps
Table 4 summarizes the challenges and gaps in evidence informing the clinical management of M. chimaera infection.
Table 4. Challenges and gaps in evidence informing the management of Mycobacterium chimaera infection.
| Challenge / gap | Additional context |
|---|---|
| The magnitude of the risk of M. chimaera infection and the extent of the outbreak is unknown (12,14,29) |
|
| Delay in symptom onset and diagnosis of infection (3,10,14,23,32) |
|
| Effectiveness and adverse effects of therapy (3,11,29,32) |
|
| Development of new HCD designs is pending (5,15) |
|
| Extent of HCD-associated infections caused by other microorganisms such as Legionella species is unknown (12,29) |
|
Abbreviation: HCD, heater–cooler device; HEPA, high-efficiency particulate air; UK, United Kingdom
Discussion
Findings from this overview indicate a low but increased risk of M. chimaera infection with use of common source-contaminated HCDs during CPB (29). Given the long latency period, additional cases are expected. The true magnitude of risk following exposure is uncertain; current estimates are based on very limited data. Nonetheless, the risk of delaying cardiac surgery is generally considered far greater than the risk posed by this infection, even when the infection risk has not been entirely mitigated (28).
Future patient exposure may be prevented by implementing risk mitigation measures, including the use of uncontaminated HCDs or replacement of contaminated HCDs as soon as possible. Case finding may be expedited by the development of polymerase chain reaction (PCR)-based assays for rapid and reliable detection of M. chimaera in clinical or environmental samples.
Improved HCD designs that facilitate reliable decontamination and prevent aerosols from reaching the operative field are urgently needed (5,11). These developments may require collaborative discussions between medical device manufacturers, engineers and infection prevention and control experts.
This overview is limited by insufficient data to estimate the true magnitude of risk of infection and absence of data on efficacy and feasibility of risk mitigation measures.
Conclusion
The epidemiological and microbiological characteristics of this outbreak provide evidence for transmission of M. chimaera from the exhaust air of contaminated Stöckert 3T HCDs to patients during CPB, resulting in endocarditis, surgical site infections and/or disseminated infections. The true magnitude of risk following exposure is uncertain; it is currently estimated based on very limited data.
Strategies that separate the HCD exhaust air from the operating room air during surgery may be the most effective risk mitigation measures. The feasibility of implementing currently recommended risk mitigation measures is yet to be determined and studies are needed to determine if there are any unintended consequences from implementation of these measures. Development of HCDs with new designs that are airtight and/or not susceptible to biofilm formation may help address this problem.
Authors’ statement
TO – project administration, conceptualization, methodology, research, data abstracting and writing (original draft, review and editing); GT, LJ, MM – clinical expertise, intellectual content, review, editing and writing; KA - clinical expertise, intellectual content, review and editing; AC, K Defalco – review of abstracted data, research, review and editing; K Dunn – conceptualization, supervision, review and editing; JJ, SS, JE, BH, JS – clinical expertise, intellectual content, review and editing.
Acknowledgements
The authors would like to thank the Canadian Public Health Laboratory Network (CPHLN) for providing information presented in Table 1 of this manuscript.
The authors would also like to thank Lynda Gamble (Health Library, Health Canada) for conducting the literature search for this overview; Caroline M. Desjardins (Public Health Agency of Canada) for citing and generating references for this manuscript; and Dr. Margaret Gale-Rowe (Public Health Agency of Canada) for reviewing the final version of the manuscript.
Footnotes
Conflict of interest: None.
Funding: This work was supported by the Public Health Agency of Canada.
References
- 1.Percival SL, Suleman L, Vuotto C, Donelli G. Healthcare-associated infections, medical devices and biofilms: risk, tolerance and control. J Med Microbiol 2015. Apr;64(Pt 4):323–34. 10.1099/jmm.0.000032 [DOI] [PubMed] [Google Scholar]
- 2.Trudzinski FC, Schlotthauer U, Kamp A, Hennemann K, Muellenbach RM, Reischl U et al. Clinical implications of Mycobacterium chimaera detection in thermoregulatory devices used for extracorporeal membrane oxygenation (ECMO), Germany, 2015 to 2016. Euro Surveill 2016. Nov;21(46):1–8. 10.2807/1560-7917.ES.2016.21.46.30398 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Sommerstein R, Schreiber PW, Diekema DJ, Edmond MB, Hasse B, Marschall J et al. Mycobacterium chimaera outbreak associated with heater-cooler devices: piecing the puzzle together. Infect Control Hosp Epidemiol 2017. Jan;38(1):103–8. 10.1017/ice.2016.283 [DOI] [PubMed] [Google Scholar]
- 4.Sommerstein R, Rüegg C, Kohler P, Bloemberg G, Kuster SP, Sax H. Transmission of Mycobacterium chimaera from heater-cooler units during cardiac surgery despite an ultraclean air ventilation system. Emerg Infect Dis 2016. Jun;22(6):1008–13. 10.3201/eid2206.160045 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Sax H, Bloemberg G, Hasse B, Sommerstein R, Kohler P, Achermann Y et al. Prolonged outbreak of Mycobacterium chimaera infection after open-chest heart surgery. Clin Infect Dis 2015. Jul;61(1):67–75. 10.1093/cid/civ198 [DOI] [PubMed] [Google Scholar]
- 6.Behr M, Jarand J, Marras TK. Canadian Tuberculosis Standards 7th edition [Internet]. Ottawa (ON): Public Health Agency of Canada; 2014 [cited 2017 Mar 16]. Available from: http://www.phac-aspc.gc.ca/tbpc-latb/pubs/tb-canada-7/tb-standards-tb-normes-ch11-eng.php [Google Scholar]
- 7.Galassi L, Donato R, Tortoli E, Burrini D, Santianni D, Dei R. Nontuberculous mycobacteria in hospital water systems: application of HPLC for identification of environmental mycobacteria. J Water Health 2003. Sep;1(3):133–9. [PubMed] [Google Scholar]
- 8.Lyman M, Grigg C, Kinsey CB, Keckler MS, Moulton-Meissner H, Noble-Wang J et al. Invasive nontuberculous mycobacteria infections among cardiothoracic surgery patients - Hospital A, Pennsylvania, 2010-2015. 65th Annual Epidemic Intelligence Service (EIS) Conference. 2016 [cited 2017 Mar 16]:94. Available from: http://www.cdc.gov/eis/downloads/eis-conference-2016.pdf
- 9.Haller S, Höller C, Jacobshagen A, Hamouda O, Abu Sin M, Monnet DL et al. Contamination during production of heater-cooler units by Mycobacterium chimaera potential cause for invasive cardiovascular infections: results of an outbreak investigation in Germany, April 2015 to February 2016. Euro Surveill 2016. Apr;21(17). 10.2807/1560-7917.ES.2016.21.17.30215 [DOI] [PubMed] [Google Scholar]
- 10.Achermann Y, Rössle M, Hoffmann M, Deggim V, Kuster S, Zimmermann DR et al. Prosthetic valve endocarditis and bloodstream infection due to Mycobacterium chimaera. J Clin Microbiol 2013. Jun;51(6):1769–73. 10.1128/JCM.00435-13 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Kohler P, Kuster SP, Bloemberg G, Schulthess B, Frank M, Tanner FC et al. Healthcare-associated prosthetic heart valve, aortic vascular graft, and disseminated Mycobacterium chimaera infections subsequent to open heart surgery. Eur Heart J 2015. Oct;36(40):2745–53. 10.1093/eurheartj/ehv342 [DOI] [PubMed] [Google Scholar]
- 12.Public Health England. Infections associated with heater cooler units used in cardiopulmonary bypass and ecmo: information for healthcare providers in the UK, version 2 London: Public Health England; 2017 [cited 2017 Mar 16]. Available from: http://www.gov.uk/government/publications/infections-associated-with-heater-cooler-units-used-in-cardiopulmonary-bypass-and-ecmo
- 13.UPDATE. Mycobacterium chimaera infections associated with LivaNova PLC (formerly Sorin Group Deutschland GmbH) Stöckert 3T Heater-Cooler System: FDA Safety Communication [Internet]. Silver Spring (MD): U.S.: Food and Drug Administration [updated 2016 Oct. 13; cited 2017 Mar 16]. Available from: http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm520191.htm
- 14.Struelens MJ, Plachouras D. Mycobacterium chimaera infections associated with heater-cooler units (HCU): closing another loophole in patient safety. Euro Surveill 2016. Nov;21(46):1–3. 10.2807/1560-7917.ES.2016.21.46.30397 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.LivaNova Stöckert 3T Heater-Cooler Device - Risk of Mycobacterium chimaera Infections [Internet]. Ottawa (ON): Health Canada; 2017 [cited 2017 Mar 16]. Available from: http://healthycanadians.gc.ca/recall-alert-rappel-avis/hc-sc/2017/62252a-eng.php
- 16.Antonation K, Patel S, Trumble Waddell J, Poliquin G, Alexander DC, Hoang L et al. Interim laboratory testing guidance for the detection of non-tuberculous Mycobacterium (NTM) infections in post-operative patients exposed to heater-cooler units. Can Commun Dis Rep 2017;43(1):25–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Canadian Public Health Laboratory Network (CPHLN) Testing Working Group. Device testing recommendations regarding non-tuberculous Mycobacteria (NTM) contamination in heater-cooler units [Internet]. Ottawa (ON): Government of Canada [updated 2017 Feb. 9; cited 2017 Mar 16]. Available from: www.canada.ca/en/public-health/services/publications/diseases-conditions/device-testing-non-tuberculous-mycobacteria-contamination-heater-cooler.html
- 18.European Centre for Disease Prevention and Control (ECDC). Rapid risk assessment: Invasive cardiovascular infection by Mycobacterium chimaera associated with the 3T heater-cooler system used during open-heart surgery [Internet]. Stockholm (SE): ECDC; 2016 [cited 2017 Mar 16]. Available from: http://www.ecdc.europa.eu/en/publications/Publications/RRA-mycobacterium-chimaera-November-2016.pdf
- 19.Centers for Disease Control and Prevention. Non-tuberculous Mycobacterium (NTM) infections and heater-cooler devices interim practical guidance: Updated October 27, 2015 [Internet]. Atlanta (GA): CDC [updated 2015 Oct 27; cited 2017 Mar 16]. Available from: www.cdc.gov/HAI/pdfs/outbreaks/CDC-Notice-Heater-Cooler-Units-final-clean.pdf
- 20.National infection control guidance for non-tuberculous Mycobacterium associated with heater-cooler devices [Internet]. Sydney (AUS): Australian Commission on Safety and Quality in Health Care; 2017 [cited 2017 Mar 16]. Available from: https://www.safetyandquality.gov.au/publications/national-infection-control-guidance-for-non-tuberculous-mycobacterium-associated-with-heater-cooler-devices/
- 21.European Centre for Disease Prevention and Control (ECDC). EU protocol for case detection, laboratory diagnosis and environmental testing of Mycobacterium chimaera infections potentially associated with heater cooler units: case definition and environmental testing methodology [Internet]. Stockholm (SE): ECDC; 2015 [cited 2017 Mar 16]. Available from: http://ecdc.europa.eu/en/publications/Publications/EU-protocol-for-M-chimaera.pdf
- 22.Williamson D, Howden B, Stinear T. Mycobacterium chimaera spread from heating and cooling units in heart surgery. N Engl J Med 2017. Feb;376(6):600–2. Available from: http://www.nejm.org/doi/full/10.1056/NEJMc1612023 10.1056/NEJMc1612023 [DOI] [PubMed] [Google Scholar]
- 23.Bursle E, Playford EG, Coulter C, Griffin P. First Australian case of disseminated Mycobacterium chimaera infection post-cardiothoracic surgery. Infect Dis Health. Forthcoming 2017; Available from: http://www.sciencedirect.com/science/article/pii/S2468045117300020 10.1016/j.idh.2017.01.003 [DOI] [Google Scholar]
- 24.Perkins KM, Lawsin A, Hasan NA, Strong M, Halpin AL, Rodger RR et al. Notes from the Field: mycobacterium chimaera contamination of heater-cooler devices used in cardiac surgery - United States. MMWR Morb Mortal Wkly Rep 2016. Oct;65(40):1117–8. 10.15585/mmwr.mm6540a6 [DOI] [PubMed] [Google Scholar]
- 25.Svensson E, Jensen ET, Rasmussen EM, Folkvardsen DB, Norman A, Lillebaek T. Mycobacterium chimaera in heater-cooler units in Denmark related to isolates from the United States and United Kingdom. Emerg Infect Dis 2017. Mar;23(3):507–9. Available from: https://wwwnc.cdc.gov/eid/article/23/3/16-1941_article 10.3201/eid2303.161941 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Schreiber PW, Kuster SP, Hasse B, Bayard C, Rüegg C, Kohler P et al. Reemergence of Mycobacterium chimaera in heater-cooler units despite intensified cleaning and disinfection protocol. Emerg Infect Dis 2016. Oct;22(10):1830–3. Available from: http://wwwnc.cdc.gov/eid/article/22/10/16-0925_article 10.3201/eid2210.160925 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Sorin Group. Sorin 3T Heater-Cooler System [Internet]. [updated 2016; cited 2017 Mar 16]. Available from: http://www.livanova.sorin.com/products/cardiac-surgery/perfusion/hlm/3t
- 28.Kanamori H, Weber DJ, Rutala WA. Healthcare-associated Mycobacterium chimaera transmission and infection prevention challenges: role of heater-cooler units as a water source in cardiac surgery. Clin Infect Dis 2017. Feb;64(3):343–6. Available from: http://cid.oxfordjournals.org/content/early/2016/12/23/cid.ciw755.full.pdf+html 10.1093/cid/ciw755 [DOI] [PubMed] [Google Scholar]
- 29.Chand M, Lamagni T, Kranzer K, Hedge J, Moore G, Parks S et al. Insidious risk of severe Mycobacterium chimaera infection in cardiac surgery patients. Clin Infect Dis 2017. Feb;64(3):335–42. Available from: http://cid.oxfordjournals.org/content/early/2016/12/05/cid.ciw754.full.pdf+html 10.1093/cid/ciw754 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Centers for Disease Control and Prevention. Contaminated devices putting open-heart surgery patients at risk [Internet]. Atlanta (GA): CDC [updated 2016 Oct 13; cited 2017 Mar 16]. Available from: http://www.cdc.gov/media/releases/2016/p1013-contaminated-devices-.html
- 31.Miller MM, Baer L, Freer C, Hnatuck P, Julian KG, Mincemoyer S et al. Mycobacterium chimaera infections following open chest cardiac surgery Pennsylvania. Am J Infect Control 2016;44(6 Suppl):S26–7. 10.1016/j.ajic.2016.04.204 [DOI] [Google Scholar]
- 32.Appenheimer AB, Diekema DJ, Berriel-Cass D, Crook T, Daley CL, Dobbie D et al. Mycobacterium chimaera outbreak response: experience from four United States healthcare systems [Abstract] [Internet] Open Forum Infect Dis 2016;3 Supp 1:S2392. [Google Scholar]
- 33.Tan N, Sampath R, Abu Saleh OM, Tweet MS, Jevremovic D, Alniemi S et al. Disseminated Mycobacterium chimaera infection after cardiothoracic surgery. Open Forum Infect Dis 2016. Jun;3(3):ofw131. Available from: http://ofid.oxfordjournals.org/content/3/3/ofw131.full 10.1093/ofid/ofw131 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.L'ICM avise ses patients opérés à coeur ouvert depuis 2012 d'un risque d'infection [Internet]. Montreal (QC): Institut de Cardiologie de Montréal; 2016 Oct. 24 [cited 2017 Mar 16]. Available from: http://www.icm-mhi.org/fr/salle-presse/nouvelles/licm-avise-ses-patients-operes-coeur-ouvert-depuis-2012-dun-risque-dinfection
- 35.Garvey MI, Ashford R, Bradley CW, Bradley CR, Martin TA, Walker J et al. Decontamination of heater-cooler units associated with contamination by atypical mycobacteria. J Hosp Infect 2016. Jul;93(3):229–34. Available from: http://www.sciencedirect.com/science/article/pii/S0195670116001274 10.1016/j.jhin.2016.02.007 [DOI] [PubMed] [Google Scholar]
- 36.Hnatuck P, Brumbach A, Stolberg DA, Snyder R, Baer L, Mincemoyer S et al. A case study: discordant results in mycobacterial culturing of cardiac bypass heater-cooler machines—seek and you may find. Am J Infect Control 2016;44(6 Suppl):S50 10.1016/j.ajic.2016.04.042 [DOI] [Google Scholar]
- 37.Gotting T, Jonas D, Scheibert D, Klassen S, Ebner W. Infection risk in cardiothoracic surgery due to contaminated heater cooler units. From 3rd International Conference on Prevention and Infection Control (ICPIC 2015). BioMed Central [Internet]. 2015 4(Suppl 1):73. [Google Scholar]