Revised. Amendments from Version 1

We hereby present a revised version of our review, based on the comments made by the referees and new developments since version 1 was published. In our description of the physical DNA sequencing process, part 1 of our review, minor revisions were made; the R9.5 pore is now briefly introduced under “structure and charge of the nanopore”, mock data used in the example of the MinION raw signal in Figure 3 has been replaced by actual data, and the flowcell grid layout cartoon in Figure 5 has been adapted to show more diverse defects that may occur in a well. Mentions of the R9 pore where we intended to refer to the series designated with prefix R9 have been adapted to clarify this. Part 2 has undergone more significant revisions. The section on basecallers now is focused on raw-signal processing and transducer-based basecalling. To that end, the text on Metrichor basecallers has been revised and the descriptions of older basecallers, Nanocall and DeepNano, have been omitted, while two newer tools, Chiron and BasecRAWller, have been added. In the section on long read assemblers, notes on the quality of MinION-only assemblies versus that of hybrid assemblies and a comment on the role of and comparison with SMRT technology has been added. Descriptions that were found too detailed in retro-perspective have been simplified so that the text now describes the tools in more uniform depth and a description of SMARTdenovo has been added. Short descriptions of tools, along with their performance in published benchmark efforts are noted in Table 1. Under post-assembly correction, the recent updates to Nanopolish allowing much faster processing speeds and methylation-aware polishing have been taken into account. We thank all involved referees for their thorough assessments, which we believe have helped us to improve the manuscript.

Abstract

Nanopore technology provides a novel approach to DNA sequencing that yields long, label-free reads of constant quality. The first commercial implementation of this approach, the MinION, has shown promise in various sequencing applications. This review gives an up-to-date overview of the MinION's utility as a de novo sequencing device. It is argued that the MinION may allow for portable and affordable de novo sequencing of even complex genomes in the near future, despite the currently error-prone nature of its reads. Through continuous updates to the MinION hardware and the development of new assembly pipelines, both sequencing accuracy and assembly quality have already risen rapidly. However, this fast pace of development has also lead to a lack of overview of the expanding landscape of analysis tools, as performance evaluations are outdated quickly. As the MinION is approaching a state of maturity, its user community would benefit from a thorough comparative benchmarking effort of de novo assembly pipelines in the near future. An earlier version of this article can be found on  bioRxiv.

Keywords: nanopore sequencing, MinION, \textit{de novo} sequencing, review

Introduction

The development of novel genome sequencing methods has been a major driving force behind the rapid advancements in genomics of the last decades. Notably, the advent of second generation sequencing (SGS) provided researchers with the required throughput and costefficiency to sequence many more genomes than was previously deemed feasible. Recent years saw the dawn of what can be considered a third generation; one that allows amplification-free reading of single DNA molecules in long consecutive stretches 1. Currently, this new generation is dominated by two methods: nanopore sequencing and single-molecule real time (SMRT) sequencing, championed by Oxford Nanopore Technologies (ONT) and Pacific Biosciences (PacBio), respectively.

Conceptually, nanopore sequencing is easier to explain than most other sequencing methods. An electrical potential is applied across an insulating membrane in which a single small pore is inserted. A DNA strand is pulled through the pore and the sequence is inferred from the characteristic way in which the passing base combinations influence the current. In 1989, David Deamer roughly sketched this concept as it is applied today, although it took more than two decades of key innovations to bring the concept to fruition 2. Since the introduction of the first commercially available nanopore sequencing device, ONT’s MinION, and the start of the MinION Access program (MAP) in 2014, the field of nanopore sequencing has been advancing at a rapid pace; both new applications and improvements to existing ones are published on a regular basis.

The advantages of the MinION over other sequencing devices are numerous. Both its size, roughly that of a cellphone, and its initial investment cost, a thousand dollars for a starter kit, are a mere fraction of that of competitors. Running the MinION is also reasonably time- and cost-effective; a 48-hour sequencing run currently costs around 800 dollars 1 and yields up to 5 Gbases of raw sequenced data 3. Furthermore, the technique does not rely on any labeling techniques to recognize different bases, while Sanger, second generation and SMRT sequencing methods do require some form of labeling of nucleotides. Amplification by PCR is optional for the MinION, while this step is mandatory for Sanger and SGSmethods. Not only does omitting these steps simplify sample preparation for MinION samples, it also helps to avoid errors and biases (e.g. the CG-bias for PCR) and allows detection of modified bases 4. Finally, the maximum read length produced by the MinION is many times greater than that of both second-generation and Sanger sequencing and only paralleled by SMRT sequencing, which is highly advantageous in resolving repeat sequences.

The most prominent disadvantages of the MinION, with respect to its competitors, are the lower signal-to-noise ratio, stochasticity introduced by its biological components, and the resulting high error rate of basecalling. Indeed, the MinION is a product in development and the used materials (i.e. membranes, nanopores and buffers) are still being optimized. Furthermore, it is thought that significant improvements are still possible in the software pipelines that translate current signal to DNA sequence. In this review, an up-to-date overview of de novo nanopore sequencing and assembly is provided. First, the physical sequencing process as it takes place inside the MinION is outlined. Then, the general structure of analysis pipelines is described, along with currently available software implemented in these pipelines and their respective strengths and weaknesses. It should be noted that nanopore sequencing is a rapidly advancing field. While some work discussed in this paper is considered cutting-edge at the moment of writing, the reader is advised to keep the publication date of said work in mind.

1 Physical basis of DNA sequencing using nanopores

The underlying principle of nanopore sequencing can be explained as follows: a microscopic opening wide enough to allow single-stranded DNA to pass the nanopore is introduced in an insulating membrane between two compartments filled with saline solution and an electric potential is applied across it. DNA strands are then added to one compartment and allowed to diffuse toward the nanopore, where they are captured by the electric field and threaded through the pore. While a strand is passed through, the characteristic way in which the bases influence the electric current through the nanopore is measured. These measurements can then be decoded to retrieve the sequence of the DNA strand ( Figure 1).

Figure 1. Sequencing of a DNA strand using nanopores.

Figure 1.

From left to right, double-stranded DNA with attached motor protein attaches to a pore protein in an insulating membrane. The applied potential pulls one strand through the pore, while the motor protein unzips the DNA in a step-wise fashion. After the DNA has been unzipped completely and one strand has passed through, the complex detaches from the pore entrance and the pore is ready to receive another strand. Image courtesy of Oxford Nanopore Technologies Ltd.

In recent years, several key discoveries rapidly transformed nanopore sequencing into a usable DNA analysis method. In a step-by-step exploration of the sequencing process, these discoveries will be discussed next.

Choice of pore: Biological versus solid-state Nanopore sequencing efforts are sub-categorized in two groups based on the choice of nanopore. Most current efforts implement biological nanopores, which are protein multimers derived from naturally occurring counterparts. Through genetic engineering, biological nanopores are modifiable in terms of dimensions and placement of electrical charge. These properties are also highly reproducible from one pore to the next. Functionality can be further modified by attaching compatible enzymes to the pore opening. Like their naturally occurring counterparts however, they need to be embedded in a lipid membrane, which is generally prone to disruption, particularly when exposed to varying electrical potentials. In the MinION, this was partly solved by constructing membranes out of a more stable single layer of polymers, rather than the traditional bilayer. Solid-state nanopores on the other hand, are made by burning openings in a synthetic membrane using a focused electron or ion beam 5. Contrary to biological nanopores, solid-state nanopores are compatible with a wide range of strong and chemically stable materials with equally diverse properties. Pores are also more easily parallelized and integrated in electrical readout circuits. A major disadvantage at the moment is the irreproducibility of the pore dimensions. They also do not combine as easily with modifying enzymes. As a result, solid-state nanopores currently produce noisier and less easily interpretable signals than biological nanopores. In the following, the focus will lie on biological nanopore sequencing and the term nanopore will refer to the biological kind.

Structure and charge of the nanopore One important structural property that makes a biological pore suitable for DNA sequencing is a constriction site at which the passing strand exerts the most influence on the electrical current. The length of the constricting passage largely determines how many bases simultaneously influence the electrical current and thus the number of bases that is “read” simultaneously at a given time. This number should be kept low enough to allow recognition of a signature current for each different combination of bases and high enough to allow for some overlap between subsequent base combinations, as this benefits basecalling accuracy by allowing every base to be read multiple times. Modified versions of both pore proteins that have seen application in the MinION, MspA (denoted by ONT with series numbers prefix “R7”) and the currently used CsgG 6 (denoted with prefix “R9”, Figure 2), have a constricted passage that allows detection of a manageable number of bases. For the 10Å-long constriction of the CsgG pore, basecalling models previously relied on the assumption that five nucleotides sufficiently influence the current at any given time to discern all different nucleotide combinations, and thus 5-mers were assigned to stretches of signal ( Figure 3). Although this worked reasonably well, it was found that this assumption does not always hold, e.g. due to specific base sequences and the secondary structure of the molecule influencing the current differently. Newer basecalling models therefore no longer make this assumption and assign a variable number of bases (see also section 2.1).

Figure 2. Protein structure of the CsgG pore protein complex, a variant of which is used in current generation MinION flow cells.

Figure 2.

Positive and negative residues are colored blue and red, respectively. Image generated by the authors using PyMOL v1.7.0.0. PDB ID: 4UV3 6.

Figure 3. Example of a MinION DNA read as raw data (grey line) and the event data (red lines) extracted from it, corresponding to discrete sets of bases.

Figure 3.

For the sake of illustration it is assumed that five bases influence the current at a given time, although in reality this assumption may not always hold. Data used in this figure was obtained from the Nanopore WGS consortium (third release) 3.

For sequencing to commence, a DNA strand first needs to diffuse towards one side of the pore, referred to as the cis-side, where it is captured by the electric field resulting from the applied potential. It is then threaded through the pore and extruded at the other end, called the trans-side.

Two forces should be considered. First and most importantly, the electrophoretic force induced by a positive electric potential applied at the trans-side attracts the negatively charged DNA and pulls it in. As negative particles leave the cis-side and positive particles simultaneously move in the opposite direction, a positively charged zone forms around the cis entrance of the pore, strengthening attraction of DNA strands. Secondly, strand translocation is influenced by the electro-osmotic flow (EOF), the force induced by the net water and ion flow through the pore. While a DNA strand is in the pore, the EOF normally opposes the direction of the electrophoretic force and thus of translocation; however, this effect is relatively minor.

Through iterative optimization of internal architecture, it was found that positive internal surface charges are important for efficient DNA capture 7, 8, while base recognition was found to improve with bulky or hydrophobic amino acid side chains placed at the constriction site, as these direct ion flow toward the DNA strand 9. Although the structures of the modified pores used in MinION flow cells have not been publicly released by ONT, modifications to these properties have likely been made. Currently, ONT maintains two types of flow cells containing different modified CsgG pores 6, designated R9.4 and R9.5. Reportedly, alterations between R9.4 and its successor R9.5 were solely made to facilitate a novel sequencing mode (dubbed 1D 2, see below) and should not influence sequencing accuracy in any other way. These alterations thus likely pertain to different properties of the pore.

Processive control It should be noted that the processive speed of the strand without any further modifications is too high for the sensor to accurately detect changes in electrical current (between 2 · 10 6 and 10 · 10 6 bases/s in wild-type MspA) 7. Currently, the most successful way to exert control over the speed has proven to be the addition of a motor protein, such as phi29 DNA polymerase 10 or a helicase 11. In a preparatory step, poly-T or “leader” adapters are attached to the doublestranded DNA. Motor proteins attach to these adapters, but due to specialized bases in the adapter sequence (possibly acridine residues as used by 12, but left unspecified by ONT 11), they cannot unzip it at this stage. Once one end of the complex is adjacent to the cis-side of the pore, the leader adapter previously blocking the motor protein is released, presumably due to the force exerted on the strand as demonstrated by 13 and described in 14. The DNA is then fed base-by-base through the pore by the motor protein as it processes the strand, where it can now be read at a regular pace. A modified helicase is currently used as motor protein in the MinION 11. The latest release of this motor protein at the time of writing (dubbed E8) maintains an average throughput speed of 450 bases/s (as noted in e.g. 3).

Reading the DNA strand During a MinION sequencing run, the potential over the membrane is kept stable, while the electrical current (in the pA-range) is sampled at a frequency in the kHz range ( Figure 3). This signal is characteristic for the subsequent bases moving through the pore and will ultimately serve as the basis for basecalling. As the amount of electrolyte is increasingly depleted during the run, the applied potential (typically starting at -180mV) is further decreased by 5mV per two hours of runtime and increased by 5mV when the MinION switches to another set of wells filled with fresher buffer (see next section).

While the MinION can read the first strand of a dsDNAstretch that is threaded through the pore - by definition, the template strand - and discard the complementary strand, it is possible to instead read the complementary strand immediately after the template, thus performing a second read of the same stretch in reverse complement ( Figure 4). Combining reads of both strands has been shown to increase sequencing accuracy significantly 15.

Figure 4. The three categories of DNA reading chemistries for the MinION.

Figure 4.

( A) When using 1D chemistry, only the template strand (blue) is threaded by its motor protein (green) and read. The complement strand (red) is discarded at the cis side of the pore. The tethers (dark-green) allow for selection of properly ligated complexes during sample preparation and attach to the membrane to increase the availability of strands near pores during sequencing. ( B) The now-deprecated 2D chemistry connected template and complement strand using a hairpin, thus allowing sequencing of the complement strand immediately after the template strand. An additional tether that attached to the hairpin allowed for selection of correctly ligated strands during sample preparation. ( C) 1D 2 chemistry, the successor of 2D, also allows sequencing of both strands, but rather than attaching the two, the complement strand is tethered to the membrane while the template is sequenced. After the template strand is threaded through, the complement strand is drawn in and the tether is pulled loose. Based on 17 by permission from Macmillan Publishers Ltd: Nature Methods, copyright(2015), the ONT kit content description, and ONT's technical update of March 2017.

The currently implemented method for doing so is referred to as 1D 2 sequencing (versus 1D sequencing if only the template strand is read). The 1D 2 chemistry provided by ONT includes different adapters that allow the complement strand to attach to the membrane while the template strand is read. Shortly after the template strand has completely left the pore, the complement strand is pulled in and sequenced. The mirrored reads are then decoded jointly so that any sequencing errors may be corrected. A previously offered method with the same aim, referred to as 2D-sequencing, involved covalently connecting the 3’-end of the template and the 5’-end of its complement using an abasic hairpin adapter, thus allowing the complement strand to be pulled in automatically after the template strand. However, due to several issues, including the hairpin’s tendency to ligate different strands into chimeric reads 16 and a lower read quality and sequencing speed for the complement strand 15 reportedly caused by secondary structure changes in the strand while rezipping after sequencing, this approach was deprecated in favor of 1D 2-sequencing in May of 2017.

Channel parallelization Lastly, throughput can be greatly increased by reading the signal from multiple pores in parallel. The current generation of the MinION’s disposable cartridges, called flow cells, can read the signal of up to 512 pores in parallel ( Figure 5). The flow cell is equipped with 2048 wells, which are connected in groups of four to multiplexers (MUXs), the switches that control which of the four cells per group is controlled and read out by the circuits. During the initial platform quality check, DNA strands (of unreleased source and sequence), present in the buffer with which the flow cells are shipped, are sequenced to discern wells suitable for sequencing (i.e. containing an intact membrane and precisely one correctly inserted, properly functioning pore) from wells in which correct pore insertion has failed ( see ONT platform quality check explanation). The latter scenario may occur, as the insertion of pores is a stochastic process. In a second quality check, the MUX scan, each MUX chooses up to three wells in order of signal quality and begins readout in the best-quality well. As well quality is expected to decline during the run, the standard protocol switches to the second-best quality pore after eight hours, and the third-best quality after another eight hours. This way, the best and most output is expected in the first part of the run. While a run using a group of wells is in progress, the circuits connected to the MUXs regulate the current in each selected well individually. This also allows expelling of eventual blockades from a pore, by temporarily reversing the current in the affected well while the rest of the wells continue to function normally.

Figure 5. Layout of a MinION flowcell grid.

Figure 5.

Large circles denote wells in the grid, small black circles denote inserted nanopores. In reality, the pore diameter (12 nm) is much smaller with respect to the well diameter (about 10 µm). Each group of four wells is controlled by a multiplexer (MUX). During an initial quality check, wells that are unusable e.g. due to erroneous pore insertion, membrane defects or pore blockades are marked as unusable (hatch pattern). Right before sequencing, the wells are tested a second time and three wells per MUX are ranked on signal quality (if possible). Sequencing of the sample will then commence, starting read-out from the best-performing well (green) and switching to second and third best (yellow) after eight hours each. The white wells are usable for sequencing, but are left unused unless the user designates otherwise in the MinION protocol.

2 Currently available software for MinION basecalling and de novo assembly

Following the process in section 1, a current signal is obtained that is subsequently translated into the underlying DNA sequence by a so-called basecaller. Next, the read sequences may be de novo assembled using assembly tools that can make use of the long read length while mitigating the error-prone nature of the reads. This is often followed by a last error correction or ’polishing’ step, in which a better consensus between the assembly and the raw reads is sought. In this section, these steps are detailed and a selection of available software tools to fulfill each step is explored.

2.1 Basecallers

Before basecalling takes place, some preparatory steps may be required. First, if the (now deprecated) 2D chemistry was used, the signal derived from the template strand should be separated from that of the hairpin and the complement strand. This process is commonly referred to as segmentation. Furthermore, older basecallers require the signal to be subdivided into discrete averaged stretches, or events, each corresponding to a particular set of k bases. Both segmentation and event detection can be performed by MinKNOW, the MinION control software provided by ONT. For event detection, MinKNOW was reported to calculate a simple t-statistic between sliding adjacent windows of set size. Peaks in the t-statistic above a certain threshold are then assumed to signify the borders between adjacent events.

Initially, basecallers were designed to find the most likely set of k bases for each event detected in this manner 18, 19. As it became clear that the number of bases per event is too variable for this approach, newer tools generally infer the events and the underlying sequence simultaneously from the raw signal (e.g. Albacore v 2.0.1, Chiron, BasecRAWller) 20, 21.

To assess the quality of basecalling performance, a 3.6 kbase calibration strand derived from the Lambda genome may be added to the sample 15, 22. MinKNOW automatically detects reads derived from the Lambda genome and separates those from the sample reads. Software tools may also use these strands for parameter optimization (e.g. as PoreSeq does to adjust its basecall correction algorithm 22).

Several dedicated basecalling tools are available to MinION users. In this section, the underlying principles and implementation of these tools are explored, along with their reported strengths and weaknesses. Unfortunately, most basecallers do not support calling 1D 2 reads, thus performance measures will focus on 1D calling. Wick et al. have provided a benchmark for basecallers on a 1D, R9.4 Kleibsiella pneumoniae dataset generated with a SQK-LSK108 chemistry kit. To the author’s knowledge, this is currently the only comprehensive and up-to-date benchmarking effort. Comparisons made in this section are based on their analysis and reports made by the authors of the open-source basecallers in their publications. In the latter case, the used read type, pore and chemistry kit is listed between brackets each time, e.g. for the Wick et al. study: (1D, R9.4, SQK-LSK108).

Metrichor basecallers Metrichor, a spin-off company of ONT and its main developer of proprietary analysis software, maintains a range of basecallers that have remained the go-to option for most MinION users. Currently, four Metrichor basecallers are available to users: Albacore, the MinKNOW integrated basecaller, Nanonet and Scrappie. A cloud-based version was previously integrated in the EPI2ME platform, but this service has been discontinued. Both Nanonet and Scrappie are unsupported development basecallers, while Albacore and the MinKNOW version are stable tools intended for regular MinION users.

Initially, the Metrichor basecallers relied on hidden Markov models (HMMs) to assign k-mers of set size k to event-called data. As of early 2016, the HMM model was replaced by a more accurate recurrent neural network (RNN)-implementation. This approach was first introduced in Nanonet (source code publicly available), a basecaller written in Python and using the CURRENNT library 23 to implement its RNN. It is able to perform all steps from raw MinION signal to base sequence (i.e. segmentation, event-calling and basecalling). The next major advancement was the addition of a transducer after the RNN in April of 2017 which, rather than assigning a k-mer to each event, uses the newly input signal and the bases it previously emitted to determine whether to output none, one or multiple bases for the next event. Importantly, this allowed the detection of homopolymer sequences longer than a given k-mer size 3. This was previously impossible, as the sliding window t-test used in event detection could not discern individual events in homopolymer stretches, effectively merging them into a single event which would then be assigned a single k-mer 22, 24, 25.

From June of 2017, event-based calling was abandoned all together in favor of a more accurate raw signal-based approach. Both the transducer and raw signal-based calling were first introduced as options in Scrappie, a newer developer basecaller written in C (source code publicly available), and were later implemented in Albacore (transducer as of v1.0.1, raw signal interpretation as of v2.0.1). To date, Albacore also remains the only basecaller able to make use of 1D 2 reads. The MinKNOW basecaller lags slightly behind Albacore but is otherwise identical. The source code of Albacore and the MinKNOW basecaller is currently only open to developer users.

Metrichor’s up-to-date basecaller implementations (i.e. Albacore, MinKNOW and Scrappie) first center and scale the raw signal using the median signal over the entire read (as first described in 26) and then consecutively feed it through a strided convolutional filter and unidirectional RNN layers of gated recurrent units (GRUs) which receive their memory from alternating directions. The stacked unidirectional layers and use of GRUs allows the RNN to interpret the convolved signal in a long-range context from both sides, while remaining computationally efficient in use. The output of the RNN is fed into a transducer, which assigns a number of bases to each raw data point as described above 2. Lastly, the translocation speed of the strand is estimated using found non-homopolymeric events, which is then used to detect and correct probable collapsed homopolymer sequences.

The processing speed 21 and accuracy of Albacore, MinKNOW and Scrappie is currently considered to be the highest of all available basecallers. Wick et al. estimated median identity with the reference genome of a transducer-based raw signal-processing Albacore version (v2.0.2) at 87.6%. The introduction of raw data interpretation lead to some increase in accuracy; Albacore v2.0.1 scored 87.6% identity versus 86.5% for v1.2.6 (the last version without raw calling included by Wick et al.) and a similar difference was seen between Scrappie v1.1.1 processing event-called (85.8% identity) and raw data (88.1%). The effect of the introduction of the transducer at v1.0.1 can be seen in the read length, which is closer to the reference read length, and the higher corrected assembly identity, which indicates that fewer systematic errors are made. Both observations can be explained by the fact that the transducer allows for more accurate calls in homopolymer regions in particular, as was also shown by 3. As expected, the outdated Nanonet (v2.0.0) does not perform as well as Scrappie and Albacore (85.6% identity). Albacore’s median identity rate on 1D 2 reads has been reported by ONT at around 97%, however this has yet to be confirmed by thorough independent studies.

Chiron Chiron 21 is a third-party basecaller that shows high similarity with current Metrichor basecallers. It was written in Python and its neural network is implemented using the TensorFlow library 27.

Chiron first centers the raw signal around the mean and scales it over the standard deviation, after which the signal is divided up in partly overlapping batches to allow parallel processing. Much like current Metrichor basecallers, it then feeds the signal through a convolutional filter, several RNN-layers and a transducer which outputs probabilities for each base (or the absence of a base) for each raw data point. Finally, the returned base sequences for the split signal are fused into a single sequence for the entire read by finding the largest overlap.

Although Chiron’s overall structure is similar to that of Metrichor basecallers, its multiple convolutional layers, the usage of the more elaborate long short-term memory (LSTM) cells instead of GRUs and the more conventional bidirectional RNN architecture make Chiron more complex. Indeed, the benchmark published by Chiron’s authors shows that it performs slightly slower than Albacore v1.1.1 but similarly in terms of accuracy; on reads of lambda phage DNA, E. coli and Mycobacterium tubercolosis (all 1D, R9.4, SQK-LSK108), the difference between sequence identities of Albacore and Chiron did not rise above 1.2%. Albacore did do slightly better than Chiron on a human dataset generated with the same chemistry; Chiron’s authors hypothesize that this could be because Chiron was not trained on human data. These results are largely in line with the benchmark by Wick et al.; indeed Chiron (v0.2) performs similarly to Albacore v1.1.2, but the raw data-based Albacore v2.0.2 performs notably better. In terms of sequencing speed, Chiron’s authors showed that Albacore (2975 bases per second on a CPU) easily outperformed Chiron (21 bases per second on a CPU, 1652 on a GPU).

BasecRAWller While other basecallers prioritize accuracy, BasecRAWller’s 20 primary goal is to allow “streaming basecalling”, i.e. basecalling during sequencing directly from the raw signal. As its authors note, streaming basecalling may prove highly advantageous in selected applications, such as rejection of strands from the pore during sequencing if, based on the retrieved base sequence, it is decided that the strand is not of interest to the user. BasecRAWller is written in Python and uses the TensorFlow library 27 for its neural network implementation.

Like Metrichor basecallers, BasecRAWller uses a medianbased normalization method 26 to pre-process the raw signal. However, as the median of the signal of the entire strand (as used by Metrichor) is not available in streaming basecalling, it is approximated by using the median unoccupied pore signal, as these values were found to correlate sufficiently. The normalized signal is then consecutively fed into a unidirectional LSTM-RNN and a fully connected feed-forward network, which assigns a 4-mer to each measurement and a probability that the measurement should be recognized as the start of a new event. This information is ultimately passed on to another unidrectional LSTM-RNN which assigns zero, one or multiple bases to each event. Although bidirectional RNNs have the advantage of utilizing both past and future measurements to place a prediction in a proper context, the choice for a unidirectional network was consciously made to retain the ability to basecall in a streaming fashion.

As its authors state in their own assessment of BasecRAWller’s performance, some accuracy was surrendered to allow for streaming basecalling; Metrichor basecallers reached significantly higher accuracy on both an E. coli dataset (1D, R9, SQK-NSK007) and a human dataset (2D, R9.4, SQK-LSK108) (89.4% and 76% respectively, versus 82.9% and 72.5% for BasecRAWller). It should be noted that Albacore was able to take advantage of the 2D chemistry used for the human dataset, while BasecRAWller could not. Similarly, Wick et al. found a median identity of 74.0% for BaseCrawller (v0.1) versus Albacore’s (v2.0.2) 87.6%. An assessment by Teng et al. found slightly higher identity rates for BasecRAWller (v0.1) of around 82% on Lambda phage E. coli, M. tuberculosis and human datasets (all 1D, R9.4, SQK-LSK108), which were still 2% lower than that of Albacore (v1.1.1) on human data and around 8% lower for the other datasets 21. BasecRAWller’s authors indicate a processing speed of up to 900 bases per second using the current MinION throughput speed and sampling frequency, while Teng et al. indicated a maximum sequencing speed of 81 bases per second 21. The cause of this large difference is unclear, but important to investigate further, as a speed below 450 bases per second (the current average throughput speed of the MinION) would indicate that BasecRAWller is currently not able to function as a true streaming basecaller.

2.2 Assemblers

Once nanopore reads have been basecalled, they may serve several purposes. If SGS reads are available, one of several approaches to hybrid assembly (i.e. combining long error-prone and short accurate reads) may be chosen; short reads may be mapped to the nanopore reads to correct sequencing errors pre-assembly 31 or to create large low-error contigs. The latter goal may be achieved by using nanopore reads to close gaps and resolve repeat regions in SGS assemblies 32, by using them as scaffolds to properly align short reads 3335, by correcting a long read-only assembly using short reads 36, 37 (referred to as “polishing”, see also next section), or by creating short accurate seed regions from short reads, which are then bridged by nanopore reads 25. All described approaches were shown to result in accurate and highly contiguous de novo assemblies and in identification of repeats that were collapsed in SGS-only assemblies 25, 31. If no SGS reads are available, nanopore-only assembly pipelines can be used. It has been shown that using these pipelines, a cheap and highly contiguous MinION-only de novo draft genome can already be sequenced and assembled within one week (e.g. as was done for the 54 Mbase fungal genome of Rhizoctonia solani 38). If speed, cost or only the general structure of the genome are of major importance, a MinION-only approach may thus already be adequate. However, it should be noted that MinION-only assemblies are still generally inferior to those of hybrid methods in terms of accuracy, due to the error-prone nature of the reads 39, 40. If the goal is the construction of a highly accurate and contiguous assembly and SGS reads can be obtained, hybrid assemblies should be preferred. This accuracy gap is expected to diminish in the future due to the steadily increasing quality of MinION reads. With this and the cost- and time-effectiveness of the MinION in mind, the focus of this review lies on tools that can be used in de novo MinION-only sequencing.

As PacBio sequencers were available before nanopore sequencing had come to fruition, most assemblers able to work with MinION reads were either initially intended as PacBio tools or were written with both technologies in mind. Some tools offer specific parameter settings to account for differences in read properties between the two technologies, most importantly the differing error distributions. Giordano et al. showed that, on datasets of comparable size and read length distribution, assemblers consistently constructed more accurate assemblies with SMRT reads than with MinION reads (although the latter were generated with older chemistries and basecallers, see also Table 1) 39. While the difference in accuracy is in large part attributable to the higher number and less random distribution of sequencing errors, it does seem that those adapted for use with MinION reads are better able to mitigate its sequencing errors.

Table 1. Summary of comparisons between long read assemblers.

( A) Selected metrics for three benchmarking efforts on MinION reads, including chemistries used in the respective studies. Bold values denote the best score per metric. ( B) Short descriptions and reference papers for all assemblers discussed in this paper. 1: reads were corrected by Canu prior to assembly.

A Judge et al. 41 Istace et al. 40 Giordano et al. 39
subs/
kbase
indels/
kbase
N50
(Mbase)
subs/
kbase
indels/
kbase
N50
(Mbase)
subs/
kbase
indels/
kbase
N50
(Mbase)
PBcR 1.0 12.2 1.20 0.2 17 0.616
Canu 0.3 7.8 2.80 0.105 10.0 0.610 0.1 17 0.698
SMARTdenovo 0.580 11.1 0.783 0.3 14 0.625
Minimap &
miniasm
6.7 18.6 6.60  0.207 1  13.5 1  0.736 1 34 67 0.739
ABruijn 0.130 10.1 0.816 0.1 15 0.769
Chemistry MAP006 MAP005/MAP006 MAP006/007
Read type 2D 2D 2D
Pore R7.3 R7.3 R7.3/R9
Basecaller EPI2ME EPI2ME EPI2ME
Organism Enterobacter
kobei
S. cerevisiae S. cerevisiae
B Description Ref.
PBcR Celera OLC assembler adapted for long error-prone reads. 42
Canu The more accurate successor of PBcR. 43
SMARTdenovo Fast and reasonably accurate assembler without prior error correction step. Github
Minimap &
miniasm
Fast assembly pipeline without error correction and consensus steps. 44
ABruijn DBG assembler that fuses unique strings prior to assembly, produces highly contiguous
assemblies.
45
TULIP uses seed extension principle to efficiently assemble large genomes. 25
HINGE Assesses coverage of low complexity regions prior to assembly and processes them more
efficiently.
46

Assembly of MinION and SMRT reads requires a different approach than that of SGS reads; as the reads are longer, finding a correct overlap should be easier, yet they are more error-prone, which increases the uncertainty of overlaps. Because of these differences, a return of interest in overlap-layout-consensus (OLC) algorithms - which were at the peak of their popularity in the era of Sanger sequencing - is seen. Traditional De-Bruijn graph (DBG) assemblers, the more popular choice for SGS reads, were reported to return lower quality assemblies of MinION reads than OLC-based methods, but proved faster in some cases 47. A selection of available long read OLC and DBG assemblers is discussed in this section.

Software using traditional greedy extension algorithms (e.g. SSAKE) is rarely used in MinION read assembly as it was found to perform decidedly less well in a de novo assembly setting, both in terms of assembly quality and required computational resources 47, and is therefore not further discussed here. Furthermore, only tools that provide a full solution to their respective step in the assembly pipeline are reported here. As current assemblers either include their own error correction module 43 or work with uncorrected reads 25, 4446, stand-alone pre-assembly error correction tools are excluded as well. A short summary of each assembler’s characteristics and the limited number of available benchmarks is given in Table 1, although it should be noted that a proper evaluation is difficult due to the different and outdated chemistries and basecallers used. Thus, while performances noted here may provide an initial orientation in the available choice in long read assemblers, results are likely to differ when using current technology.

PBcR & Canu Originally developed for the first human genome draft, the Celera assembly pipeline 48 and its extensions 43, 49, 50 have remained a popular choice in a growing landscape of OLC assemblers. Briefly, the Celera assembler uses read overlaps to find contigs of which the structure can unambiguously be derived from overlap information, referred to as unitigs. It then separates unitigs that were found to occur multiple times from unique ones and attempts to orient the unique unitigs with respect to eachother. Where possible, gaps between unique unitigs are filled with non-unique unitigs. As a high read error rate is detrimental to the quality of the assembly 51, two different modifications to the pipeline are available. The PacBio corrected Reads (PBcR) algorithm, originally developed for the correction of PacBio reads suffering from similar error rates, uses accurate short reads mapped with high confidence to the long reads to correct errors. The assembly then proceeds as usual by Celera 42. Celera’s successor, Canu 43, provides a more accurate solution that does not require short accurate reads. Like PBcR, Canu was shown to succesfully assemble both MinION and PacBio reads 39. The pipeline includes three stages; correction, trimming and assembly. Overlaps are found using the efficient minhash alignment process (MHAP) 52, which hashes k-mers using different hash functions and for each hash function stores the smallest integer to which a k-mer of the sequence is hashed. Comparing the hashed k-mers per read results in initial overlap hits, which are then used to perform error correction by consensus seeking. By selecting overlaps for correction on quality, but limiting the number of overlaps a read can contribute to, Canu attempts to prevent masking of true repeat variants. Shorter reads are used at this stage to improve accuracy of longer reads. In the trimming step, overlaps are recalculated to locate and filter out regions of low coverage and high error. Reads are overlapped two more times to correct specific types of errors (i.e. missed hairpin sections for 2D reads, adapters, chimeric reads) and to adjust the error rate per overlap, before the actual assembly phase starts. With adjustments to account for erroneous alignments and residual errors, assembly essentially follows the same procedure as CABOG, another Celera-based pipeline 49.

Due to its thorough yet relatively efficient correction steps, Canu is significantly more accurate than both its predecessor Celera/PBcR and most other tested assemblers. In benchmarks on Enterobacter kobei and S. cerevisiae reads, it often produced an assembly with fewer indels and mismatches than others, often with higher contiguity 3941. These results are in line with the author’s own assessment 43.

SMARTdenovo SMARTdenovo is a long read OLCassembly pipeline that was originally intended to work with PacBio reads, but has been shown to produce assemblies of reasonably high continuity from MinION reads as well 39. Surprisingly, it does so without an error correction step prior to assembly, making SMARTdenovo a faster alternative to Canu.

As detailed on its Github page, SMARTdenovo first attempts to find read overlaps for each read in three steps at increasing accuracy by first searching hits in sorted k-mer tables twice and then using a banded Smith-Waterman algorithm. To find overlaps that were missed in this process, it subsequently repeats the process for pairs of reads that should overlap, given the extent to which they are overlapped by other reads. Next, low quality or chimeric read ends are identified by their decreased coverage by other reads and removed. Finally, SMARTdenovo borrows PacBio’s directed alignment graph consensus (DAGCon) algorithm 53 to produce the consensus assembly.

As expected, SMARTdenovo was shown to outperform Canu in terms of computing efficiency 39, 54. However, benchmarks on S. cerevisiae reads demonstrated that assemblies by Canu generally show higher identity with the reference sequence 39, 40. This is possibly due to the fact that the HGAP algorithm leveraged for error correction was originally intended to work with PacBio reads, which have a different error distribution. Notably, Schmidt et al. showed that SMARTdenovo produced an assembly of higher contiguity for the large tomato ( Solanum pennellii) genome and, when preceded by Canu’s pre-assembly error correction module, obtained an even more contiguous assembly with fewer predicted errors than either Canu or SMARTdenovo could, while still remaining faster than Canu alone 54.

Minimap & Miniasm In terms of speed and computational efficiency, the OLC-based pipeline consisting of Minimap and Miniasm 44 has a definite advantage over other existing tools 3941. This efficiency was reached through the omission of the consensus step and the use of minimizers. Much like the k-mer hash table used by Canu’s MHAP 43, a minimizer is a memory-efficient hashed representation of a sequence. Minimap computes the set of minimizers of a sequence, the “sketch”, by finding the k-mers represented by the smallest hash value within a certain window size of each position of the sequence. The complement of each k-mer is also considered. Decreasing the window size will increase the returned number of minimizers and allow for more accurate alignment, at the cost of increased computational requirements. Minimap then performs all-versus-all mapping by identifying hits between minimizers of different sequences. The found overlaps are passed on to Miniasm, which constructs an assembly graph. First, potential artefacts are removed from each read by identifying the longest stretch with a coverage of three or more other reads, and then clipping off the ends that fall outside this region. Then reads contained within other reads are removed and small bubbles, less than 50 kb in length, are popped (i.e. a consensus is taken in cases where paths split and later join up again). Finally, sequences can be extracted from stretches of the graph without multi-edges to form unitigs. The error rate at this point is practically the same as that in the raw reads, emphasizing that correct basecalling is essential for the eventual quality of the assembly. The graphical fragment assembly (GFA) output format of Miniasm conveniently allows both graphing of the uncorrected assembly and addition of consensus error correction tools, such as Nanopolish or Racon, to the pipeline.

In March of 2016, the authors of Minimap and Miniasm reported assembly of MinION reads of an E. coli genome in a single contig. In May of the same year, Judge et al. assembled an Enterobacter kobei genome in 16 contigs with an N50 of 662 kbase in two minutes, while the next fastest assembler (Canu) took two hours, however their benchmark showed that the omission of an error correction step caused the eventual assembly quality of E. kobei to be too low to properly assess by the QUAST analysis tool 41.

ABruijn While more traditional DBG assemblers performed worse than OLC assemblers on assembling long error-prone reads 47, the approach taken by the ABruijn assembler has shown more promise 45. To account for the high error rate, ABruijn filters all k-mers occurring in the reads by their frequency; if a k-mer occurs few times for given dataset and genome sizes, it is assumed that it contains basecalling errors and it is removed. Then k-mers are fused into so-called “solid strings”, sequences that contain no other occurring sequences as substring. The ABruijn graph is then drawn by representing solid strings as vertices and connecting them where connections exist in the reads. The edges are weighted by the number of positions between the first bases of the connected solid strings. The assembler consults the weights in this graph to quickly identify overlaps between reads, allowing to select on a minimum overlap length and maximum overhang length. The assembly graph is constructed by starting with the graph for an arbitrary read and iteratively extending it by overlapping it with other reads. ABruijn also includes an error correction routine, during which a best consensus between reads is found by identifying low-error stretches and, in between those stretches, choosing the consensus sequence that maximizes the likelihood of the read sequences.

In two independent benchmarking efforts (2D, R7.3, MAP005/006 and 2D, R7.3/R9, MAP006/007), ABruijn assembled an S. cerevisiae genome with higher contiguity than other included assemblers (Canu, Minimap/Miniasm, SMARTdenovo and PBcR) 39, 40 ( Table 1). However, ABruijn was also the only assembler to produce chimeric contigs. Furthermore, Canu’s assemblies showed higher identity with the reference genome. Thus ABruijn’s assembly routine tends to return longer contigs, while Canu is less error-prone.

TULIP As more reads are required to cover larger genomes, and as the time required for all-vs-all overlapping increases quadratically with an increasing number of reads, it follows that the overlap step of OLC assemblers may take unfeasibly long for very large genomes. To tackle this issue, The Uncorrected Long read Integration Process (TULIP) takes a different approach to read overlapping 25. Instead of all-vs-all alignment, short seed sequences are selected, which the assembler then attempts to align with long reads. This drastically cuts down the overlapping complexity and makes efficient use of long reads to cover long stretches of the genome between the seed regions. The resulting graph represents seeds as vertices and the connecting reads as edges. In a graph cleaning step, vertices with multiple in- or outgoing edges are revisited. Spurious and superfluous edges are removed aggressively, thus producing a linear graph. Note that, as the name implies, TULIP does not perform basecalling error correction.

The success of assembly using TULIP highly depends on proper seed selection. To avoid spurious connections between reads, the seeds need to be sufficiently unique in the genome and contain few sequencing errors. If available, SGS reads may be used to construct seeds, although with the increasing accuracy of MinION reads, the ends of long reads may be used as well. Apart from cutting out the need for SGS methods, the latter approach has the added advantage that pairs of seeds are connected by at least one long read. Furthermore, as TULIP is not able to assemble regions in which the gap between seeds is larger than the read length, a proper seed density over the entire genome is required. If a marker map is available for the genome, this information can be used to control the distribution of seeds in the selection process.

As a first demonstration of TULIP’s efficiency, Jansen et al. assembled the genome of the European eel Anguilla anguilla (approximately 850Mbp) with 18x coverage in three hours (excluding sequence polishing), requiring only 4.4GB of RAM and four threads 25. The resulting assembly was more continuous than the SGS-based reference genome. As was the case with Minimap/Miniasm however, the current quality of MinION reads combined with the lack of an error correction step necessitates post-assembly correction. The authors further showed that missed seed alignments were the most commonly encountered issue during graph simplification, followed by tangled alignments due to repetitive seeds and spurious alignments. The seeds, constructed from short SGS reads, only underwent selection by uniqueness, which did not lead to an equal distribution over the entire genome; however, density remained high enough for successful assembly. The authors noted that assembly using the tips of MinION reads as seeds proved successful for Escherichia coli genomes, but this has not been attempted for larger genomes yet (personal communication, May 1, 2017).

HINGE Although long reads provide a definite edge when attempting to resolve repeat regions, issues may still occur if not all individual repeats are spanned by at least one whole read. In such cases, HINGE may provide a solution. Rather than attempting to resolve frayed rope structures in the assembly graph afterwards, HINGE preprocesses the reads to separate repeat regions that are entirely spanned by a read (and are thus more easily resolvable) from those that are not, and collapses the latter beforehand 46.

First, HINGE attempts to identify reads that wholly or partly overlap a repeat region. It does so by performing all-vs-all alignment and then selecting those reads of which a stretch aligns to a proportionally larger number of other reads than the rest of the read. The intuition behind this is that reads from all copies of a repeat region existing in the genome align to each other, thus causing a characteristic abrupt increase in alignments for reads that overlap these repeat regions. Repeat regions covered entirely by at least one read can be easily resolved and are omitted from the following procedure. Of the reads lining the same repeat region, the reads that extend furthest into the repeat region (regardless of the location of the actual copy), are designated “hinges”. In the subsequent greedy extension of the hinges, the contigs will split at the hinge regions. Like Miniasm, HINGE outputs its assembly in the form of a graph. As its authors show, this is particularly useful for circular genomes.

HINGE provides an elegant solution to long repeat resolution, by separating resolvable regions from unresolvable ones beforehand. Its authors compared HINGE to Miniasm on PacBio reads of 997 circular bacterial genomes and found that overall, HINGE produced a completed genome in more cases than Miniasm could 46. Whether the precaution taken by HINGE is necessary is dependent on the genome under consideration and the used reads; if the genome is known to contain repeats longer than most of the reads, the described approach would be justified.

2.3 Post-assembly correction tools

A number of tools attempt to improve, or “polish”, assemblies by remapping long reads to the assembly and adapting the assembly to increase local resemblance to the reads. These polishing tools may be essential to use after assembly pipelines that do not include a consensus step themselves, such as Minimap/Miniasm, but have also frequently been used to polish assemblies produced by assemblers that do include this step. In this section, a selection of polishing tools is described. Notably, ONT recently published the source code for their own neural network-based polisher, Medaka. Although this tool may become a valuable addition to assembly pipelines in the future, it is currently in an early stage of development.

Nanopolish Nanopolish attempts to find an optimal consensus between an assembly and the raw current signal output by the MinION, by iteratively proposing and evaluating small adaptations to the assembly based on the original reads 24. The proposal mechanism for adaptations works in two steps. First, reads are aligned to the assembly and the resulting multiple alignment is divided in 50 bp subsequences of the assembly. For each read aligning partly or fully to a subsequence, sections in which events perfectly align to the assembly are detected. The consensus sequence between each pair of aligning sections is replaced by the aligned read subsequence, creating an initial set of alternative candidate sequences. In the second step, this set is further extended by proposing every possible one-base deletion, insertion and substitution in the previously generated candidate sequences. Of this set, the sequence maximizing the likelihood of observing the raw signal is picked. This process allows Nanopolish to explore a decent number of likely modifications, while remaining computationally tractable. As of v0.8.4, available information on methylation sites can be used to improve the quality of those sites even further. As epigenetic modifications were shown to influence the current signal 26, this may result in a significant improvement.

Nanopolish was found to improve assembly quality, regardless of the assembly tool used. One study on E. coli sequencing data reported that identity to the reference genome rose from 89% to 99% when Nanopolish (v0.4.0) was applied after Minimap/Miniasm, while improvement after Canu was more modest (98.2% to 99.6%) 55. Notably, the previously mentioned Wick et al. benchmark showed that methylation-aware polishing brought the identity of reference-based assemblies up significantly to 99.9% versus 99.7% after polishing without methylationawareness. An assessment on a de novo assembly has yet to be made.

Despite its efficient searching heuristic of block replacement and mutation, running Nanopolish remains a time-consuming step; in two separate benchmarking efforts, one on an E. kobei assembly produced by Minimap/Miniasm and one on a S. cerevisiae assembly by Canu, running Nanopolish (v0.4.0 and v0.5.0 respectively) required more than a month of extra CPU time 39, 41. Later versions of Nanopolish (especially v0.7.0 and up) were reported by its authors to work much faster.

Racon Racon 56 corrects MinION assemblies by finding a consensus sequence between reads and the assembly through the construction of partial order alignment (POA) graphs. After alignment of the reads by a mapper of choice (e.g. Minimap or Graphmap), Racon segments the sequence and finds the best alignment between a POA graph of the reads and the assembly. By default, the alignment is performed using the Needleman-Wunsch algorithm, which can align sequence and POA graph with little adaptation. The alignment process is sped up by parallelization. Racon was reported by its authors to be two orders of magnitude faster than the popular (yet currently deprecated) Nanocorrect 24 after assembly of an E. coli genome by Miniasm, albeit not quite as good at diminishing the error rate (to 1.31% versus 0.62% for Nanocorrect). Compared to consensus steps in Falcon 57 and Canu 43 on that same assembly, Racon remains an order of magnitude faster while producing similar error rates. A closer look at the remaining errors reveals that the majority consists of indels. As indel basecalling has drastically improved in newer basecallers (versus the pre-transducer basecallers used by Racon’s authors), these would likely allow Racon to reach even lower error rates. Finally, the total genome size estimate following application of Racon was closer to the reference genome size than the estimates of Canu, Falcon and Nanocorrect.

3 Discussion

Nanopore sequencing is a promising new venue in biology research. Inexpensive, small, capable of producing long reads and freed from the need for nucleotide labeling or amplification, it is conceivable that the MinION will make cost-effective, fast and portable de novo whole genome sequencing of even complex genomes possible in the future. In this review, an attempt was made to give an updated overview of the progress in this field, focusing in particular on de novo whole genome sequencing.

Available basecaller tools have been improving rapidly in accuracy. Notable recent improvements include the move toward raw signal-based calling and the inclusion of a transducer. For the next step in a typical sequencing routine, assembly, OLC-assemblers are currently considered the best option for accurate de novo nanopore-based assembly. The choice of assembler should be adapted to the characteristics of the genome and the priorities of the user. Canu is a complete and accurate solution, although SMARTdenovo was shown to be much faster against slightly diminished accuracy. The best of both methods may be obtained by combining Canu’s error correction module with SMARTdenovo. Minimap/Miniasm is by far the fastest option available, but as it lacks any form of error correction, cannot produce a usable genome draft without any post-assembly correction. For large, complex genomes, TULIP may be the more tractable alternative. Lastly, stand-alone post-assembly consensus error correction tools Nanopolish and Racon are a worthwhile addition in de novo sequencing pipelines and a necessity in combination with assemblers that do not contain a sequencing error correction step of their own.

Currently, the most prominent obstacle for de novo sequencing using the MinION is the high error rate of the reads. Improving basecalling accuracy would not only improve assembly quality in a direct manner, but may also allow more computationally efficient assembly.

The active research community surrounding the MinION has booked great progress in both the development of new applications and improvements on accuracy of existing ones. ONT also continuously works on improvements for both its hardware and software platforms, and regularly updates its users on this. Although these updates often entail welcome new features or some form of accuracy improvement, it should be noted that this policy has also lead to some difficulties. Developers may not be able to keep pace with ONT when evaluating, updating or calibrating their tools, and users may not always know which tool is suited best to their data and needs. As a result, most published studies, including tool benchmarking efforts, were conducted using older or multiple chemistries. Although such growing pains are to be expected for a novel fast-developing field of research, the MinION’s current state of development may allow for some increase in stability, thus giving the user community the time for proper evaluation.

Notes

1Estimate based on a purchase of 24 flowcells and a 1 D/1D2 sequencing kit, 13th of October 2017

2A thorough discussion of neural network architectures and their respective properties is outside the scope of this article. Interested readers are referred to 28 and 29 for introductions to RNNs and convolutional networks respectively, and 30 for more information on transducers.

Acknowledgements

Giovanni Maglia (University of Groningen, Groningen, The Netherlands) provided helpful advice related to the physical basis of nanopore sequencing.

Funding Statement

The author(s) declared that no grants were involved in supporting this work.

[version 2; referees: 2 approved]

References

  • 1. van Dijk EL, Auger H, Jaszczyszyn Y, et al. : Ten years of next-generation sequencing technology. Trends Genet. 2014;30(9):418–426. 10.1016/j.tig.2014.07.001 [DOI] [PubMed] [Google Scholar]
  • 2. Deamer D, Akeson M, Branton D: Three decades of nanopore sequencing. Nat Biotechnol. 2016;34(5):518–524. 10.1038/nbt.3423 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Jain M, Koren S, Quick J, et al. : Nanopore sequencing and assembly of a human genome with ultra-long reads. bioRxiv. 2017. 10.1101/128835 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Simpson JT, Workman R, Zuzarte PC, et al. : Detecting DNA methylation using the oxford nanopore technologies minion sequencer. bioRxiv. 2016;047142 10.1101/047142 [DOI] [Google Scholar]
  • 5. Miles BN, Ivanov AP, Wilson KA, et al. : Single molecule sensing with solid-state nanopores: novel materials, methods, and applications. Chem Soc Rev. 2013;42(1):15–28. 10.1039/c2cs35286a [DOI] [PubMed] [Google Scholar]
  • 6. Goyal P, Krasteva PV, Van Gerven N, et al. : Structural and mechanistic insights into the bacterial amyloid secretion channel CsgG. Nature. 2014;516(7530):250–253. 10.1038/nature13768 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Butler TZ, Pavlenok M, Derrington IM, et al. : Single-molecule DNA detection with an engineered MspA protein nanopore. Proc Natl Acad Sci U S A. 2008;105(52):20647–20652. 10.1073/pnas.0807514106 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Maglia G, Restrepo MR, Mikhailova E, et al. : Enhanced translocation of single DNA molecules through alpha-hemolysin nanopores by manipulation of internal charge. Proc Natl Acad Sci U S A. 2008;105(50):19720–19725. 10.1073/pnas.0808296105 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Stoddart D, Heron AJ, Klingelhoefer J, et al. : Nucleobase recognition in ssDNA at the central constriction of the alpha-hemolysin pore. Nano Lett. 2010;10(9):3633–3637. 10.1021/nl101955a [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Lieberman KR, Cherf GM, Doody MJ, et al. : Processive replication of single DNA molecules in a nanopore catalyzed by phi29 DNA polymerase. J Am Chem Soc. 2010;132(50):17961–17972. 10.1021/ja1087612 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Bowen RV, Brown CG, Bruce M, et al. : Method for controlling the movement of a polynucleotide through a transmembrane pore. US Patent App. 15/113,174,2017. Reference Source [Google Scholar]
  • 12. Cherf GM, Lieberman KR, Rashid H, et al. : Automated forward and reverse ratcheting of DNA in a nanopore at 5-Å precision. Nat Biotechnol. 2012;30(4):344–8. 10.1038/nbt.2147 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Manrao EA, Derrington IM, Laszlo AH, et al. : Reading DNA at single-nucleotide resolution with a mutant MspA nanopore and phi29 DNA polymerase. Nat Biotechnol. 2012;30(4):349–353. 10.1038/nbt.2171 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Heron AJ, Alves DA, Clarke JA, et al. : Enzyme stalling method. US Patent App. 14/773,164,2016. Reference Source [Google Scholar]
  • 15. Ip CLC, Loose M, Tyson JR, et al. : MinION Analysis and Reference Consortium: Phase 1 data release and analysis [version 1; referees: 2 approved]. F1000Res. 2015;4:1075. 10.12688/f1000research.7201.1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. White R, Pellefigues C, Ronchese F, et al. : Investigation of chimeric reads using the MinION [version 2; referees: 2 approved]. F1000Res. 2017;6:631. 10.12688/f1000research.11547.2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Jain M, Fiddes IT, Miga KH, et al. : Improved data analysis for the MinION nanopore sequencer. Nat Methods. 2015;12(4):351–356. 10.1038/nmeth.3290 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. David M, Dursi LJ, Yao D, et al. : Nanocall: an open source basecaller for Oxford Nanopore sequencing data. Bioinformatics. 2017;33(1):49–55. 10.1093/bioinformatics/btw569 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Boža V, Brejová B, Vinař T: Deepnano: deep recurrent neural networks for base calling in minion nanopore reads. arXiv preprint:1603.09195. 2016. Reference Source [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Stoiber M, Brown J: BasecRAWller: Streaming nanopore basecalling directly from raw signal. bioRxiv. 2017;133058 10.1101/133058 [DOI] [Google Scholar]
  • 21. Teng H, Cao MD, Hall MB, et al. : Chiron: Translating nanopore raw signal directly into nucleotide sequence using deep learning. bioRxiv. 2017;179531 10.1101/179531 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Szalay T, Golovchenko JA: De novo sequencing and variant calling with nanopores using PoreSeq. Nat Biotechnol. 2015;33(10):1087–1091. 10.1038/nbt.3360 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Weninger F, Bergmann J, Schuller BW: Introducing currennt: the munich open-source cuda recurrent neural network toolkit. J Mach Learn Res. 2015;16(3):547–551. Reference Source [Google Scholar]
  • 24. Loman NJ, Quick J, Simpson JT: A complete bacterial genome assembled de novo using only nanopore sequencing data. Nat Methods. 2015;12(8):733–735. 10.1038/nmeth.3444 [DOI] [PubMed] [Google Scholar]
  • 25. Jansen HJ, Liem M, Jong-Raadsen SA, et al. : Rapid de novo assembly of the European eel genome from nanopore sequencing reads. Sci Rep. 2017;7(1):7213. 10.1038/s41598-017-07650-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Stoiber MH, Quick J, Egan R, et al. : De novo identification of DNA modifications enabled by genome-guided nanopore signal processing. bioRxiv. 2016;094672 10.1101/094672 [DOI] [Google Scholar]
  • 27. Abadi M, Agarwal A, Barham P, et al. : Tensorflow: Large-scale machine learning on heterogeneous distributed systems. arXiv preprint:1603.04467. 2016. Reference Source [Google Scholar]
  • 28. Lipton ZC, Berkowitz J, Elkan C: A critical review of recurrent neural networks for sequence learning. arXiv preprint: 1506.00019. 2015. Reference Source [Google Scholar]
  • 29. Zeiler MD, Fergus R: Visualizing and understanding convolutional networks. In European conference on computer vision Springer,2014;8689:818–833. 10.1007/978-3-319-10590-1_53 [DOI] [Google Scholar]
  • 30. Graves A: Sequence transduction with recurrent neural networks. arXiv preprint: 1211.3711, 2012. Reference Source [Google Scholar]
  • 31. Goodwin S, Gurtowski J, Ethe-Sayers S, et al. : Oxford Nanopore sequencing, hybrid error correction, and de novo assembly of a eukaryotic genome. Genome Res. 2015;25(11):1750–1756. 10.1101/gr.191395.115 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Antipov D, Korobeynikov A, McLean JS, et al. : hybridSPAdes: an algorithm for hybrid assembly of short and long reads. Bioinformatics. 2016;32(7):1009–1015. 10.1093/bioinformatics/btv688 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Madoui MA, Engelen S, Cruaud C, et al. : Genome assembly using Nanopore-guided long and error-free DNA reads. BMC Genomics. 2015;16(1):327. 10.1186/s12864-015-1519-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Cao MD, Nguyen SH, Ganesamoorthy D, et al. : Scaffolding and completing genome assemblies in real-time with nanopore sequencing. bioRxiv. 2016;054783 10.1101/054783 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Ye C, Hill CM, Wu S, et al. : DBG2OLC: Efficient Assembly of Large Genomes Using Long Erroneous Reads of the Third Generation Sequencing Technologies. Sci Rep. 2016;6:31900. 10.1038/srep31900 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Walker BJ, Abeel T, Shea T, et al. : Pilon: an integrated tool for comprehensive microbial variant detection and genome assembly improvement. PLoS One. 2014;9(11):e112963. 10.1371/journal.pone.0112963 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Tyson JR, O’Neil NJ, Jain M, et al. : Whole genome sequencing and assembly of a Caenorhabditis elegans genome with complex genomic rearrangements using the MinION sequencing device. bioRxiv. 2017;099143 10.1101/099143 [DOI] [Google Scholar]
  • 38. Datema E, Hulzink RJM, Blommers L, et al. : The megabase-sized fungal genome of Rhizoctonia solani assembled from nanopore reads only. bioRxiv. 2016;084772 10.1101/084772 [DOI] [Google Scholar]
  • 39. Giordano F, Aigrain L, Quail MA, et al. : De novo yeast genome assemblies from MinION, PacBio and MiSeq platforms. Sci Rep. 2017;7(1):3935. 10.1038/s41598-017-03996-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Istace B, Friedrich A, d'Agata L, et al. : de novo assembly and population genomic survey of natural yeast isolates with the Oxford Nanopore MinION sequencer. bioRxiv. 2016;066613 10.1101/066613 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Judge K, Hunt M, Reuter S, et al. : Comparison of bacterial genome assembly software for MinION data and their applicability to medical microbiology. Microb Genom. 2016;2(9):e000085. 10.1099/mgen.0.000085 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Koren S, Schatz MC, Walenz BP, et al. : Hybrid error correction and de novo assembly of single-molecule sequencing reads. Nat Biotechnol. 2012;30(7):693–700. 10.1038/nbt.2280 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Koren S, Walenz BP, Berlin K, et al. : Canu: scalable and accurate long-read assembly via adaptive k-mer weighting and repeat separation. Genome Res. 2017;27(5):722–736. 10.1101/gr.215087.116 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Li H: Minimap and miniasm: fast mapping and de novo assembly for noisy long sequences. Bioinformatics. 2016;32(14):2103–10. 10.1093/bioinformatics/btw152 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Lin Y, Yuan J, Kolmogorov M, et al. : Assembly of long error-prone reads using de Bruijn graphs. Proc Natl Acad Sci U S A. 2016;113(52):E8396–E8405. 10.1073/pnas.1604560113 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Kamath GM, Shomorony I, Xia F, et al. : HINGE: long-read assembly achieves optimal repeat resolution. Genome Res. 2017;27(5):747–756. 10.1101/gr.216465.116 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Cherukuri Y, Janga SC: Benchmarking of de novo assembly algorithms for Nanopore data reveals optimal performance of OLC approaches. BMC Genomics. 2016;17(Suppl 7):507. 10.1186/s12864-016-2895-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Myers EW, Sutton GG, Delcher AL, et al. : A whole-genome assembly of Drosophila. Science. 2000;287(5461):2196–2204. 10.1126/science.287.5461.2196 [DOI] [PubMed] [Google Scholar]
  • 49. Miller JR, Delcher AL, Koren S, et al. : Aggressive assembly of pyrosequencing reads with mates. Bioinformatics. 2008;24(24):2818–2824. 10.1093/bioinformatics/btn548 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50. Zimin A: MSR-CA–efficient De Novo genome assembler for long and short read data.In Plant and Animal Genome XXI Conference.Plant and Animal Genome.2013. Reference Source [Google Scholar]
  • 51. Salzberg SL, Phillippy AM, Zimin A, et al. : GAGE: A critical evaluation of genome assemblies and assembly algorithms. Genome Res. 2012;22(3):557–567. 10.1101/gr.131383.111 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Berlin K, Koren S, Chin CS, et al. : Assembling large genomes with single-molecule sequencing and locality-sensitive hashing. Nat Biotechnol. 2015;33(6):623–630. 10.1038/nbt.3238 [DOI] [PubMed] [Google Scholar]
  • 53. Chin CS, Alexander DH, Marks P, et al. : Nonhybrid, finished microbial genome assemblies from long-read SMRT sequencing data. Nat Methods. 2013;10(6):563–569. 10.1038/nmeth.2474 [DOI] [PubMed] [Google Scholar]
  • 54. Schmidt MHW, Vogel A, Denton A, et al. : Reconstructing the gigabase plant genome of Solanum pennellii using nanopore sequencing. bioRxiv. 2017;129148 10.1101/129148 [DOI] [Google Scholar]
  • 55. Lu H, Giordano F, Ning Z: Oxford Nanopore MinION Sequencing and Genome Assembly. Genomics Proteomics Bioinformatics. 2016;14(5):265–279. 10.1016/j.gpb.2016.05.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56. Vaser R, Sović I, Nagarajan N, et al. : Fast and accurate de novo genome assembly from long uncorrected reads. Genome Res. 2017;27(5):737–746. 10.1101/gr.214270.116 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57. Chin CS, Peluso P, Sedlazeck FJ, et al. : Phased diploid genome assembly with single-molecule real-time sequencing. Nat Methods. 2016;13(12):1050–1054. 10.1038/nmeth.4035 [DOI] [PMC free article] [PubMed] [Google Scholar]
F1000Res. 2018 Jan 16. doi: 10.5256/f1000research.14564.r28964

Referee response for version 2

Christiaan V Henkel 1,2, Michael Liem 1

The authors have significantly improved this paper, and addressed all our previous issues and concerns. The paper is now an excellent entry point into the fast-moving and therefore sometimes confusing world of nanopore data analysis.

We have detected no new issues that need further amendment. There are of course many further discussion points possible on the future of nanopore sequencing, but the review nicely succeeds in capturing the current state-of-the-art.

We have read this submission. We believe that we have an appropriate level of expertise to confirm that it is of an acceptable scientific standard.

F1000Res. 2017 Dec 13. doi: 10.5256/f1000research.14564.r28965

Referee response for version 2

David A Eccles 1

Carlos de Lannoy, Dick de Ridder, and Judith Risse have put in a huge amount of effort in improving their review of nanopore sequencing. I'm satisfied that the manuscript is now sufficiently close to describing the current state of nanopore sequencing technology, particularly when considering the level of detail present in the discussions of different aspects of the technology. While I think it's important to make reviews as current as is possible, I understand that it's also important to make sure that the information is accurate; trying to predict where ONT technology will be at the time of publication can be an exercise in futility.

In light of this, I provide here some minor corrections / suggestions for the text. These represent my attempt at a comprehensive nit-picking of what has been discussed in the manuscript, rather than a selection of demonstrative comments as in my previous review.

"its size, roughly that of a cellphone"

  • Nick Loman has used "office stapler" previously, which I think is a closer comparison

"yields up to 5 Gbases of raw sequenced data"

"second generation and SMRT sequencing methods do require some form of labeling of nucleotides"

  • Be careful with this description. The IonTorrent sequencing process doesn't use labelled nucleotides, but still depends on a specific nucleotide model due to the process of nucleotide flow during sequencing. I prefer to say that nanopore sequencing is an observational process that has a minimal dependence on models for signal capture.

"[Amplification by PCR] is mandatory for Sanger and SGSmethods"

"the maximum read length produced by the MinION is... only paralleled by SMRT sequencing"

"Solid-state nanopores... are made by burning openings in a synthetic membrane"

  • Somewhat related: see Kerstin Göpfrich's PhD thesis, which discusses nanopores constructed from DNA, embedded into lipid membranes. It's sort of a half-way point between fully biological nanopores and solid-state nanopores.

Figure 5 is much improved; thanks.

"much smaller with respect to the well diameter (about 10 µm)"

"Chiron (v0.2) performs similarly to Albacore v1.1.2"

  • Note: Ryan Wick has updated his analysis to include Chiron v0.3, which produces much more accurate consensus assemblies.

"The source code... is currently only open to developer users."

  • clarification: "users who have signed a restricted-distribution developer agreement with ONT"

"due to the error-prone nature of the reads"

  • For now, I prefer to assume that the errors are mostly in the basecallers, rather than the reads themselves. I acknowledge that I could be wrong on this, and leave the distinction as a moot point.

"after polishing without methylationawareness"

  •  -> "after polishing without methylation-awareness"

I have read this submission. I believe that I have an appropriate level of expertise to confirm that it is of an acceptable scientific standard.

F1000Res. 2017 Jul 25. doi: 10.5256/f1000research.12992.r24095

Referee response for version 1

Christiaan V Henkel 1,2, Michael Liem 1

This paper by Carlos de Lannoy et al. is a timely review of the choices available for genome sequencing based on Oxford Nanopore technology. This is a fast-moving field, and potential entrants will be well served by this exposition of the many options available. However, precisely because it is a fast-moving field, several of these options are already obsolete (yet relevant in providing context for recent developments). In addition, the many options in bioinformatics are confounded with a multiplicity of genome types (small prokaryotic to large repetitive) and nanopore chemistries.

Many readers would benefit from additional support in navigating these multiple menageries. For this reason, we would like to ask the authors to address the following issues in more detail:

  1. All methods discussed have been tested using nanopore data, but never using comparable datasets. Therefore, evaluating for example the percentages of sequence correctness is not straightforward. A table simply listing the methods with some of their key characteristics/assembly statistics/nanopore chemistries/genomes assembled would be helpful.

    Similar confusion is always looming because of the many nanopore chemistries that have been available. For example, on page 7, on the improvement of the Metrichor basecallers: the improvement here, from 68% to 95%, also coincides with the shift from R7.3 to R9. While this is mentioned as ‘numerous improvements in chemistry and hardware’, it would be clearer if the nanopore chemistry generation is explicitly mentioned whenever such percentages are listed.

  2. The in-depth discussion of assembly methodology is clearly delimited to methods focusing on nanopore data only. But is this actually already a feasible option, and if not, what is needed to make this happen? All of the assembly pipelines discussed still fall short of reference genome quality. At the moment, sequence correction using short-read data (Pilon) remains necessary. Why/when should one choose for nanopore-only assembly, as opposed to hybrid?

  3. Both basecalling and polishing are nanopore-specific bioinformatics tasks, however the de novo assemblers are in principle (and often in practice) suitable for any long-read technology. Therefore, a brief discussion comparing PacBio and nanopore for genome assembly is appropriate. PacBio-specific assemblers are briefly mentioned (SMARTdenovo, Falcon), but how do they relate to e.g. Canu?

 

A few minor issues:

  • p2, p6: please change ‘chapter’ to ‘paper’.

  • p3, second column, modifications of the R9/CsgG pore: this is a good location to briefly introduce R9.4/R9.5

  • p5: ‘multiple reads may be linked together by hairpins into chimeric reads’ -> multiple molecules

  • Figure 5: explain that the white wells are ignored.

  • p8, Canu: ‘Where possible, gaps between unique unitigs are filled with repetitive elements.’ This may need some rephrasing, ‘repetitive element’ is a specific term – not necessarily identical with ‘non-unique genome content’.

  • p8: percentages for DeepNano: ‘more accurate … (approximately 70% versus 77%) and … (approximately 88% versus 87%)’. Which is which?

  • Figure 7, 3rd kmer: the reverse complement of CTC isn’t TCT.

  • p11: a few extra days of CPU time spent on polishing is acceptable for large genomes, but a major bottleneck for bacteria – polishing then takes longer than sequencing. Do these timings refer to prokaryotic genomes?

We have read this submission. We believe that we have an appropriate level of expertise to confirm that it is of an acceptable scientific standard, however we have significant reservations, as outlined above.

F1000Res. 2017 Jul 10. doi: 10.5256/f1000research.12992.r24090

Referee response for version 1

David A Eccles 1

Carlos de Lannoy, Dick de Ridder, and Judith Risse have written a technical review paper of MinION sequencing. This is a challenging task due to the high rate of MinION consumable obsolescence, and the breadth of use cases. Unfortunately I feel that this challenge has not been met in their paper, which appears to be a scattered mixture of discussions of differing recency with no cohesive story threading them together. I think it is especially important that a review paper of MinION technology is as current as possible, because it has the potential to be used extensively by others as an indication of the quality of sequencing at a particular point in time.

I should perhaps point out that the main thing I found the most frustrating to read was a reference to R9 as current technology, and my other comments are likely coloured by me spotting this in my first scan through the article.

Here are some additional comments about the content of the paper. This is not a comprehensive review, and should only be used as a guide to the areas of concern I have with the content:

  • Title: "Coming of age" -- overlaps with Sara Goodwin's Nature Reviews Genetics paper [doi:10.1038/nrg.2016.49], which is also a review paper on sequencing technology (including nanopore). See, for example, this image:

    [https://www.nature.com/nrg/journal/v17/n6/fig_tab/nrg.2016.49_F5.html]

    A search for "coming of age" and nanopore returns another paper:

    [http://pubs.rsc.org/en/content/articlelanding/2007/mb/b702845h]

  • The paper talks about "current" nanopore technology as R9, which was discontinued last year in preference to R9.4. Shortly before this, the paper discusses solid state nanopores for a review paper, when there is no commercially-available product. This is despite mentioning the introduction of 1D² technology in May 2017, which was introduced after R9.4 pores.

  • The paper references internal nanopore community technical documentation, which is unlikely to be available to general readers of the article (and particularly not those who are considering its use).

  • "The exact start and end of the hairpin is not always clear from the signal" -- I disagree with this. While current software may have difficulty interpreting the start/end, the location is obvious when looking at the raw signal. It is surprising to see this shortly before referencing our chimeric reads paper, where we give visual examples of the raw signal where the characteristic harpins are annotated.

  • Figure 5: I like the idea of this figure. It's uncommon for people to give a visual indication of multiple pores and the MUX switching. However, some changes (and/or different annotation) are warranted. Using colours for both the mux type and the QC error is distracting; I would recommend using a hatch pattern to indicate failed pores. The pores are much larger than their actual size relative to the well size; this should be stated in the legend. Pores can also fail QC due to the membrane breaking, due to blocked pores, and presumably other reasons; the QC check is not only counting pores.

  • Figure 6: There is enough event-annotated nanopore data available that mock data should not be necessary. Spikes and within-event signal variation that I am used to seeing in raw signal data are not present in this figure.

  • "Any comparison in this chapter" -- might want to change that 'chapter' to a 'paper'.

  • Canu is now published in Genome Research:

    [dx.doi.org/10.1101/gr.215087.116]

  • The nanopolish algorithm has now been improved such that it takes substantially less time for v0.7:

    [http://simpsonlab.github.io/2017/06/30/nanopolish-v0.7.0/]

  • The paper makes many statements around the accuracy of base-calling and assembly solutions which [mostly] ignores post-transducer base-calling. I would like to see a review paper that (if possible) concentrates on post-transducer software, which would probably mean discarding discussion on HMM-based callers.

I have read this submission. I believe that I have an appropriate level of expertise to state that I do not consider it to be of an acceptable scientific standard, for reasons outlined above.

References

  • 1. Goodwin S, McPherson J, McCombie W: Coming of age: ten years of next-generation sequencing technologies. Nature Reviews Genetics.2016;17(6) : 10.1038/nrg.2016.49 333-351 10.1038/nrg.2016.49 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Ying L: Single molecule biology: coming of age. Mol Biosyst.2007;3(6) : 10.1039/b702845h 377-80 10.1039/b702845h [DOI] [PubMed] [Google Scholar]
  • 3. Koren S, Walenz BP, Berlin K, Miller JR, Bergman NH, Phillippy AM: Canu: scalable and accurate long-read assembly via adaptive k-mer weighting and repeat separation. Genome Res.2017;27(5) : 10.1101/gr.215087.116 722-736 10.1101/gr.215087.116 [DOI] [PMC free article] [PubMed] [Google Scholar]
F1000Res. 2017 Jul 12.
Carlos de Lannoy 1

Thank you for this thorough review. It clearly raises a number of valid points, which we will address in a revised version after other reviews have become available. We would like to point out at this time that we did intend our review to cover the most recent work, and that by “R9” we actually meant “R9.x” (R9.0-R9.5), as opposed to R7. This also explains why indeed we discuss more recent developments than R9.0, such as ID^2 sequencing. We hope this takes away some of the issues.

F1000Res. 2017 Jul 12.
David Eccles 1

"we did intend our review to cover the most recent work"

Note that the author guidelines for Review articles state that "Reviews should provide a balanced and comprehensive overview of the latest discoveries in a particular field." See here. This is what I had in my mind in my decision for Approve/Reject.

It is particularly important for nanopore reviews that this is the case because ONT advances their technology and software at a very rapid rate. Academics who decide on funding for research frequently latch onto the latest review articles, as they expect those articles to be a comprehensive representation of the current state of the technology.


Articles from F1000Research are provided here courtesy of F1000 Research Ltd

RESOURCES