Abstract
We present broadband measurements of the optical properties of tissue-mimicking solid phantoms using a single integrating sphere to measure the hemispherical reflectance and transmittance under a direct illumination at the normal incident angle. These measurements are traceable to reflectance and transmittance scales. An inversion routine using the output of the adding-doubling algorithm restricted to the reflectance and transmittance under a direct illumination was developed to produce the optical parameters of the sample along with an uncertainty budget at each wavelength. The results for two types of phantoms are compared to measurements by time-resolved approaches. The results between our method and these independent measurements agree within the estimated measurement uncertainties.
OCIS codes: (300.0300) Spectroscopy, (120.3150) Integrating spheres, (170.7050) Turbid media, (290.5820) Scattering measurements, (300.1030) Absorption
1. Introduction
Well-characterized tissue-mimicking phantoms are essential to validate the performance and to calibrate measurement results in the development and clinical applications of biomedical instruments from the bench and to the bedside. The phantoms are fabricated from turbid materials to simulate tissue optical properties with desired absorption coefficient μa, scattering coefficient μs and anisotropy of scattering g.
A variety of measurement techniques with the aids of physics-based photon-transport models have been developed to measure these parameters. Measurements of the optical properties of turbid media in the time domain are based on the estimation of the temporal spreading of a light pulse subjected to scattering and absorption events as it travels through the sample. Analysis using the diffusion approximation of the radiative transfer equation (RTE) or Monte Carlo (MC)-based model were used to obtain μa and = μs(1 − g) (reduced scattering coefficient) of liquid [1,2] and solid phantoms [3]. In the scope of the diffusion approximation, frequency domain measurements techniques were used to measure the optical properties of turbid liquids [4–6]. Spatial domain measurement techniques either fiber-based [7, 8] or non-contact [9–11] were developed as well. Integrating sphere measurement techniques associated with a MC model or the adding-doubling (AD) algorithm [12] were also used to measure the optical properties of biomedical phantoms and tissues [13–15].
In the scope of the current effort to implement a national scale for the optical properties of turbid media at the National Institute of Standards and Technology (NIST), in previous studies [16,17] we have established an inversion routine of the AD algorithm that allowed for the computation of the absorption coefficient μa and the reduced scattering coefficient along with the total uncertainty budget at each measurement wavelength. The measurements were made following a procedure described in the inverse adding-doubling (IAD) manual [19] using a double integrating sphere setup with identical spheres and the substitution method. The details on how to make these measurements can be found elsewhere [13,18]. Contrary to the comparison method, in the substitution method the sphere efficiencies change when the sample is substituted for the reflectance standard used as reference in the measurements [20]. In his Phd. thesis [18], Moffitt presented the equation describing the sphere efficiency of a single sphere in terms of the geometrical parameters of the spheres, the reflectance of the sphere wall, the reflectance of the detector and and further expressed the measured reflectance RMeas and transmittance TMeas for measurements with one integrating sphere and with double integrating spheres (for both measurements techniques the corresponding formulas are also presented in the documentation related to IAD source code that’s part of the IAD package). For double integrating sphere systems, cross talk between spheres lead to complex relationships between (RMeas, TMeas), the sample hemispherical reflectance and transmittance under a direct illumination, and , and the hemispherical reflectance and transmittance under a diffuse illumination, and .
Reflectance and transmittance measurements should be traceable to reflectance and transmittance scales. Under the substitution method, it’s easier using a single integrating sphere since the measurement equations are simpler. Measurements of diffuse samples using a single integrating sphere have been underway for at least a decade and has been coupled with IAD [13, 21] or inverse Monte Carlo routine [22, 23] to obtain the optical properties of the samples. Within the scope of IAD, whereas the measurements are made with one or two spheres, an important experimental step (so-called sphere calibration step) in the measurement procedure is to estimate the reflectance of the sphere wall for each sphere by aiming the incident beam toward the sphere wall and measuring the signal for (i) an empty sample port and (ii) a sample port blocked by a reference standard. Since the illumination is diffuse in that case, it’s possible to measure by adding a measurement step where the sample is set at the sample port of the sphere, provided that the sphere model is valid. By measuring a set of samples with known reflectance under a diffuse illumination, this measurement step can then be used for the traceability to the reflectance scale and also provide a means of testing the IAD model.
In this study we measure for a set of calibrated quasi-Lambertian reflectance standards in the 5 % to 99 % reflectance range. A discrepancy between the reflectance results and their expected values led us to modify our measurement procedure and move from measurements made with a double integrating sphere system to using a single integrating sphere to measure the hemispherical reflectance and transmittance under a direct illumination. The inversion algorithm of the AD routine is modified accordingly and the optical properties are accompanied by an uncertainty budget. For the first time, using a single integrating sphere, the hemispherical reflectance and transmittance under a direct illumination used as input to the AD inversion routine are traceable to reflectance and transmittance scales. The results include measurements of a commonly produced and a custom phantom that were also measured independently by a time-resolved method.
2. Material and methods
2.1. Experimental setup
Figure 1 describes the experimental setup. It’s an evolution of what was already described in Ref. [16] with the notable difference that it only includes a single integrating sphere (UMBK-190, Gigahetrz Optik, Türkenfeld, Germany; 196 mm internal diameter coated with ODM98 synthetic material, 12 mm thick; all dimensions are nominal unless noted otherwise) instead of two for reasons explained in Section 2.4. The integrating sphere has an entrance port (diameter 25.1 mm), a sample port (diameter 63.5 mm reduced to 38.1 mm by an internally conical port reducer to get a knife-edge contact with the sample), a fiber port (diameter 12.7 mm), a photodiode port (diameter 12.7 mm) inside a cylindrical baffle (diameter 30 mm, height 18 mm, with a 10 mm diameter hole on the side of the cylinder) that prevents direct reflections from the sample to the photodiode. The other notable difference from our previous setup is that the illumination sources are now of two types: one laser-based from one of three options (HeNe laser: λ = 543 nm, JDS Uniphase Corporation, Milpitas, CA, USA; λ = 632 nm, Research Electro-Optics Inc., Boulder, CO, USA; laser diode: λ = 785 nm, Spectra-Physics, Santa Clara, CA, USA) and the other using a laser driven light source (LDLS, EQ 1500, Energetiq Technology Inc., MA, USA; broadband from 170 nm to 2100 nm). The choice of illumination source is made by blocking the unwanted illumination using a mechanical light shutter LS4 (blocks the laser illumination) or LS5 (blocks the broadband illumination) and by flipping the flipping mirror FM down (laser illumination) or up (broadband illumination).
The selection of the laser wavelength is made by blocking unwanted laser beams using the manual light shutters LS1, LS2 or LS3. The beam is directed toward the sample by a set of mirrors (M1, M2, M3) and dichroic mirrors (DM1, DM2). The polarization of the laser beam is controlled by a linear polarizer P placed in front of a beam splitter BS1 that diverts a portion of the beam for detection by a photodiode D1 to simultaneously monitor power fluctuations. The remaining portion of the beam is normally incident to the front face of the sample set at the sample port of the integrating sphere (Fig. 1: reflectance measurement situation; the sphere is rotated 180° around the vertical axis perpendicular to the optical table for transmittance measurements). The signal of the corresponding diffuse hemispherical reflectance or hemispherical transmittance is detected by a second photodiode D2. The photocurrents are amplified by two current-voltages amplifiers CV1 and CV2 and the corresponding voltages are acquired by a data acquisition board (DAQ).
The output of the LDLS source is collimated by C1 (biconvex lens, f = 30 mm) and the beam is converged to the tip of an optical fiber OF1 by a parabolic mirror PM. At the output of OF1, the beam is collimated by C2 (biconvex lens, f = 25 mm) and the beam diameter is adjusted by an iris I. A band pass filter BPF can be inserted in the the beam path for wavelength selection. The beam is directed by FM to a beam splitter BS2 which diverts a portion of the beam to the tip of a fiber OF2 connected to a spectrometer SM1 for reference. The remaining part of the beam is directed to the sample and the reflectance (transmittance) signal is detected by a fiber OF3 connected to a spectrometer SM2 (detection range of the spectrometers: 350 nm to 1050 nm). In this paper, the measurements are made using the broadband source.
2.2. Samples
Two types of solid phantoms are measured in this study. The first type is a set of three polyurethane phantoms from the Institut National d’Optique (INO, Quebec, Canada) with titanium dioxide (TiO2) as a scatterer and carbon black as an absorber (batch B0430). The nominal optical properties of these phantoms are: μa = 0.01 mm−1 and = 1 mm−1 at λ = 805 nm. The samples were cut to three different thicknesses t = 5 mm, 7 mm and 10 mm (nominal) and have the same lateral dimensions (100 mm × 100 mm). The surface aspect of their faces is smooth from machining by INO and present a diffuse reflectance. We used a dial gage micrometer to measure the actual thicknesses of the samples at nine locations set on a 3×3 grid with a center point near the center of the sample. The combined uncertainty associated with thickness, was computed combining the variance of the measured thickness, , with the variance attributable to the precision of the dial micrometer, . The measured thickness values are: t = 4.94 ± 0.04 mm, 6.95 ± 0.02 mm and 9.83 ± 0.04 mm (coverage factor k = 2, used in the rest of the paper unless noted). The second type of solid phantoms were fabricated at NIST [25]. They are made of polydimethylsiloxane (PDMS) and have different concentrations of TiO2 (0.2 %, 0.1 % and 0.05 % in w/w of TiO2/PDMS) but no absorber added. They were molded using Petri dishes (diameter 87 mm) and their measured thicknesses are: t = 4.85 ± 0.18 mm (TiO2 = 0.2%), t = 4.39 ± 0.14 mm (TiO2 = 0.1%), t = 4.32 ± 0.13 mm (TiO2 = 0.05%). For comparison with time-resolved measurements at the Physikalisch-Technische Bundesanstalt (PTB, Berlin, Germany), samples of different dimensions (diameter 51.5 mm, nominal thickness 12 mm) were prepared from the same PDMS batch. The samples faces are specular, one from molding, the other from the air-PDMS interface.
2.3. Measurement of the index of refraction of the specular samples
The index of refraction of the base material of the specular samples was obtained by fitting the measurements of the Fresnel reflectance at an 8° incident angle with a Cauchy dispersion law over the measurement wavelengths. The instrument we used was a bench-top spectrophotometer equipped with an integrating sphere detector and a specular port (Lambda 1050, Perkin-Elmer, Waltham, MA, USA). The Fresnel reflectance was estimated by subtracting the diffuse hemispherical reflectance measured with the specular port open from the total hemispherical reflectance measured with the specular port blocked for λ = 400 nm to 1000 nm in steps of 10 nm. The index of refraction of PDMS was estimated from the mean value of the Fresnel reflectance of the three samples (uncertainty 0.015 at k = 1). A set of three specular phantoms from INO was measured to estimate the index of refraction of the polyurethane material [26]. We assume that the B0430 samples are made of an identical base material (uncertainty 0.015 at k = 1).
2.4. Test of the validity of the IAD model of the integrating spheres
IAD assumes an “ideal” integrating sphere with an entrance port, a sample port and a detector port, and neglects the area of the baffle but not its effect [19]. Since our spheres have a more complex internal geometry, we assumed in one of our previous papers [16] that we could use the IAD “ideal” sphere model by adjusting the diameter to basically get the same sphere gain as the real ones. This hypothesis put our system in conformity with the model used in IAD and not surprisingly, the results we obtained with our inversion of the adding-doubling algorithm were consistent with what was obtained with IAD with the additional benefit of getting the total uncertainty budget on the optical parameters μa and .
Under the scope of the IAD sphere model (see Appendix), the diffuse hemispherical reflectance of the sample under a diffuse illumination (beam aimed at the sphere wall) (diffuse/diffuse geometry [24]) can be estimated using Eq. (A5)
(1) |
by successively measuring the voltages (i) , with no sample, (ii) , with a reference standard RStdRef and (iii) ,with the sample, respectively set at the sample port of the sphere (Fig. 2(a)). This approach is known as the so-called first Taylor method [27,28].
Using a NIST-traceable 99 % reflectance standard as a reference, we measured a set of calibrated quasi-Lambertian reflectance standards in the range [2 %, 5 %, 10 %, 20 %, 40 %, 60 %, 80 %, 99 %] (Avian Technologies LLC, Sunapee, NH, USA) and compared the values to the measurements from 400 nm to 1000 nm of the directional-hemispherical reflectance with specular component included at a 6° illumination angle, R(6 : di) [24], made using the integrating sphere instrument of NIST’s Spectral Tri-function Automated Reference Reflectometer (STARR) [29]. Figure 3(a) presents the results of this comparison from 450 nm to 850 nm. For both measurement techniques, the uncertainties are computed by propagation of a Gaussian distribution of uncertainties with a coverage factor k = 2. There is a discrepancy larger than the error bars between the measured reflectance R(d : d) and the reference reflectance R(6 : di) for the 20 % to 60 % reflectance standards. This shouldn’t be the case for these quasi-Lambertian reflectance standards and the results cast doubts on the validity of the IAD sphere model for the integrating sphere used in our experimental apparatus.
Moreover, using Eq. (A3)
(2) |
where A1 and B1 are dependent of the geometrical and reflectance constants of the sphere, we can fit the measured voltage ratios for each wavelength as a function of the reference reflectance of the samples . Figure 3(b) presents the results at λ = 794 nm (wavelength of the smallest root-mean-square error (rmse), 2.62 × 10−4). The IAD model doesn’t fit the measurements well, specifically in the 40 % to 80 % reflectance range. The IAD model of the integrating sphere is not adequate for our application since it leads to substantial errors in . That’s also the case for the IAD model for a double integrating sphere system at a 0° incident angle using the substitution method since it is based on the model for a single integrating sphere. This invalidates the measurements made with our original double integrating sphere system [16,17]. However, it is crucial to perform this measurements step with a single integrating sphere to get the values of , and for the estimation of and .
2.5. Measurements of
The measurement of the diffuse hemispherical reflectance of the sample under a direct illumination at a 0° incident angle with specular component excluded, [24], can be estimated using Eq. (A7)
(3) |
by successively measuring the voltages (i) , with a reference standard RStdRef and (ii) , with the sample, respectively set at the sample port of the sphere (Fig. 2(b)). One should note that the voltages ratio formed from the measurements presented in Section 2.4 is essential to compute .
We measured for the set of calibrated reflectance standards previously used and compared the results to their R(6 : di) values (Fig. 4(a)). The uncertainty on are computed propagation of a Gaussian distribution of uncertainties with with a coverage factor k = 2. The measured values by our setup agree within the error bars with the calibrated values of the reflectance standards. Figure 4(b) presents a linear fit of the data at λ = 456 nm (wavelength of the largest rmse value 3.87 × 10−4). One should note that for a Lambertian reflectance standard, and Eq. (3) becomes , i.e. the measurement procedure is identical to the comparison method with the sphere wall used as a reference (measurement steps (a) and (b) of Fig. 2 with the sample set at the sphere sample port).
2.6. Measurements of
The measurement of the diffuse hemispherical transmittance of the sample under a direct illumination at a 0° incident angle with specular component excluded [24], , can be estimated using Eq. (A11)
(4) |
by successively measuring the voltages (i) , no sample with the beam directed toward the sphere wall and (ii) , with the sample set at the sample port of the sphere and the beam at a 0° incident angle (Fig. 2(c)). One should note that the voltages ratio formed from the measurements presented in Section 2.4 is essential to compute .
To validate this part of the experimental procedure, we measured for a set of BK7-based neutral density filters with optical density OD = [0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 1.0] and compared the results to the transmittance coefficients T(0 : 0) (normal/normal geometry [24]) measured using NIST’s Reference Transmittance Spectrophotometer (RTS) between 400 nm and 1000 nm [30]. The filters are non-scattering and for a 0° incident angle the beam exits the sphere through the open exit port (see Fig. 2(c)) which cannot be blocked in order to maintain the sphere efficiency as measured in Fig. 2(a). In this situation, no signal is detected. Near a 0° incident angle, the differences in Fresnel transmittance values of BK7 are small. Hence, the angle of incidence is changed to 8° by tilting the sphere around the vertical axis of the setup so that the transmitted beam is reflected by the sphere wall and the transmitted signal is detected. One should note that for regular diffusive samples, the angle of incidence is maintained at 0°. Figure 5(a) shows that the results with our system are consistent with the results from RTS within the error bars for λ = 450 to 850 nm. For both measurement techniques, the uncertainty are computed propagation of a Gaussian distribution of uncertainties with with a coverage factor k = 2. Figure 5(b) presents a linear fit of the data at λ = 452 nm (wavelength of the largest rmse value 5.90 × 10−4). One should note that and are identical and that Eq. (4) becomes (measurement steps (a) and (c) of Fig. 2 with sample only).
and are measurable quantities using one integrating sphere with a beam incident at 0°. The measured values are traceable to reflectance and transmittance scales. Compared to our previous study [16], we modified our measurement routine following the measurements steps described in Fig. 2 and also the inversion algorithm we implemented to obtain μa and .
3. Adding-doubling and modification of the inversion routine
Adding-doubling computes the total hemispherical reflectance and transmittance under a direct and a diffuse illumination using the incident angle θ, the thickness of the sample d, its index of refraction n, the anisotropy coefficient g and the optical parameters of the sample μa and μs as input parameters. It solves the RTE for samples with homogeneous optical properties, infinite plane-parallel slab geometry and smooth boundaries for a distribution of light independent of time and without considering polarization effects. The method gets its name from the fact that it uses a formal solution of the RTE for a thin layer of material in the single scattering approximation and computes the total reflectance and transmittance for a direct and a diffuse illumination by successively adding-doubling the values until the sample thickness is reached [31]. The inversion routine, either IAD or our previous work, computes an estimation of the measured reflectance RMeas and transmittance TMeas using the model of the spheres and the values of , and (i) (measurements with one sphere) or (ii) and (measurements with two spheres) and further uses d, n, g, RMeas and TMeas as input parameters to estimate μa and by iteration.
The modified version of our inversion algorithm restricts the use of the output of the AD algorithm to and since one only needs these two values to infer the two optical parameters μa and . Hence
(5) |
From Eq. (3) and Eq. (4), the input parameters of the inversion routine are the measurement voltages , , , , , , , the reflectance standard RStdRef, d, n, g and the incident angle θ.
One should note that in the case of the measurement of a specular sample at a 0° incident angle, is not the total hemispherical reflectance since the Fresnel specular reflectance is rejected through the entrance port of the integrating sphere. This can be corrected by
(6) |
In the case of smooth non specular samples, is assumed to be equivalent to the total hemispherical reflectance needed by the inversion algorithm as an input even though the adding-doubling algorithm does not take the roughness of the interfaces into account.
As described in Ref. [16], we consider the uncertainty on the input parameters and the measured signals to compute the total uncertainty budget of (μa; ) by propagation of a Gaussian distribution of the input uncertainties. The type A uncertainty uA and type B uncertainty uB are estimated to produce the combined uncertainty uC with , following the Guide to the Estimation of Uncertainty in Measurement (GUM) [32]. The expanded uncertainty is estimated with a coverage factor k = 2.
4. Results
A NIST-traceable 99 % polytetrafluoroethylene (PTFE)-based reflectance standard RStdRef = 0.99 ± 0.002 (k = 1) was used as a reference for the measurements. Figure 6 presents the measurements of μa and for the three B0430 samples from λ = 450 nm to 850 nm. The sample faces are smooth so no correction accounting for the Fresnel reflectance was considered in the analysis of the data. The values are compared to results by INO at λ = (475, 540, 543, 630, 632, 780, 805 and 850) nm on a 2 cm thick reference sample with smooth faces. INO uses a time-domain transmittance measurement [3] coupled to an analysis based on simulations using a modified version of the Monte Carlo modeling of light transport in Multi-Layered tissues (MCML) algorithm [33] that accounts for reflection by lateral boundaries. However, one should note that like MCML and AD their model does not take the roughness of the interfaces into account. INO does not provide an estimation of the uncertainties of their results so we used the uncertainty values computed for different samples at λ = 660 nm as presented in their reference paper (Ref. [3]). We also used their estimation of g = 0.621 ± 0.015 (k = 1) that was measured by non-scattered transmission of scattering-only thin wedges by retrieving the attenuation coefficient in the single scattering regime. With these considerations, our results for all three thicknesses are in good agreement with the ones from INO over the wavelength range (Fig. 6(a) and Fig. 6(b)). Typical uncertainty values for the integrating sphere instrument are about 4 % to 8 % on μa and about 12 % on compared to 11 % and 7 % for the INO results, respectively. The uniformity in the error bars is explained by the dominance of type B uncertainties. An example of uncertainty budget at λ = 632 nm for the 7 mm thick sample presented in Table 1 shows that n and g are dominant parameters for the uncertainties on μa and . Table 2 presents our results and the ones obtained by INO at λ = [475, 540, 543, 630, 632, 780, 805, 850] nm. A comparison between the results obtained with our single integrating sphere setup, our previous double integrating sphere setup and the INO measurements is shown in Fig. 7 for one of the B0430 samples (nominal thickness t = 5 mm). The results for μa for both integrating sphere measurements techniques do not overlap for most wavelengths whereas there is an overlap over all wavelengths for (uncertainty at k = 2). The μa results from INO overlap with both integrating sphere measurements techniques due to higher uncertainty values. Less so for the results from INO that marginally overlap for higher wavelengths.
Table 1.
Absolute uncertainties | Relative uncertainties | ||||
---|---|---|---|---|---|
| |||||
σ | μa (mm−1) | (mm−1) | μa(%) | (%) | |
Type B parameters | |||||
| |||||
RStdRef | 0.002 | 6.6 × 10−5 | 0.0034 | 0.58 | 0.33 |
d | 0.02 | 9.3 × 10−5 | 0.0086 | 0.81 | 0.81 |
n | 0.015 | 2.1 × 10−4 | 0.012 | 1.8 | 1.1 |
g | 0.015 | 1.6 × 10−6 | 0.059 | 0.014 | 5.6 |
θ | 0.0005 | 2.0 × 10−15 | 5.7 × 10−10 | 1.7 × 10−11 | 5.4 × 10−8 |
| |||||
Total Type B uB | 2.4 × 10−4 | 0.061 | 2.1 | 5.8 | |
Repeatability | 5.9 × 10−5 | 0.0033 | 0.51 | 0.31 | |
Reproducibility sR | 5.4 × 10−5 | 0.0028 | 0.47 | 0.27 | |
Total type A uA | 8.0 × 10−5 | 0.0043 | 0.69 | 0.41 | |
Combined uc | 2.5 × 10−4 | 0.061 | 2.2 | 5.8 | |
Expanded U(k = 2) | 5.0 × 10−4 | 0.12 | 4.4 | 12 |
Table 2.
λ(nm) | μa × 10−3 (mm−1) | (mm−1) | ||
---|---|---|---|---|
| ||||
INO | NIST | INO | NIST | |
475 | 15.0 ± 1.7 | 14.6 ± 0.6 | 1.04 ± 0.07 | 1.09 ± 0.13 |
540 | 12.0 ± 1.5 | 12.5 ± 0.5 | 1.02 ± 0.07 | 1.09 ± 0.13 |
543 | 11.9 ± 1.3 | 12.5 ± 0.5 | 1.00 ± 0.07 | 1.08 ± 0.13 |
630 | 12.2 ± 1.4 | 11.5 ± 0.5 | 1.01 ± 0.07 | 1.06 ± 0.12 |
632 | 12.0 ± 1.4 | 11.5 ± 0.5 | 0.997 ± 0.070 | 1.05 ± 0.12 |
780 | 9.5 ± 1.1 | 9.5 ± 0.5 | 0.954 ± 0.067 | 1.00 ± 0.12 |
805 | 9.3 ± 1.1 | 9.2 ± 0.5 | 0.948 ± 0.066 | 0.98 ± 0.11 |
850 | 9.0 ± 1.0 | 8.9 ± 0.6 | 0.928 ± 0.065 | 0.97 ± 0.11 |
Figure 8 presents the measurements of for the three PDMS samples from λ = 450 nm to 850 nm. The faces of the samples are specular so the reflectance values were corrected for the Fresnel reflectance rejected through the entrance port of the sphere. The anisotropy factor g was set to 0.5 in the data analysis. The values of are compared to measurements on samples from the same batch at λ = (500, 600, 700 and 800) nm performed at PTB by means of time-resolved diffuse optical spectroscopy [34,35]. In this case the optical properties were derived from measurements of time-resolved diffuse transmittance and reflectance that were conducted independently using two systems based on fast detectors and time-correlated single photon counting. One system used a Ti:Sapphire laser, free-space optics and a microchannel plate photomultiplier. The second system employed a supercontinuum laser with acousto-optic filter, optical fibers to guide the light to and from the sample and a hybrid photodetector. The time resolution of these systems was about 35 ps and 125 ps, respectively. The fitting procedure for obtaining and μa was based on the Monte Carlo method for a slab-like turbid medium, with a database of photon time-of-flight distributions created for multiple values. The results presented here were obtained combining the data from both measurement systems. Instrumentation and analysis are described in more detail in Ref. [25]. It should be noted that the PDMS phantoms did not contain an added absorber and that μa was too small to be determined reliably. There is an agreement between both measurement techniques at the wavelengths of interest (Fig. 8). Table 3 presents the results by PTB and NIST results at λ = (500, 600, 700 and 800) nm.
Table 3.
λ (nm) | (mm−1) | ||
---|---|---|---|
| |||
PTB | NIST | ||
TiO2 = 0.05% | |||
| |||
500 | 0.590 ± 0.030 | 0.582 ± 0.052 | |
600 | 0.518 ± 0.030 | 0.517 ± 0.045 | |
700 | 0.444 ± 0.030 | 0.449 ± 0.039 | |
800 | 0.384 ± 0.030 | 0.391 ± 0.035 | |
TiO2 = 0.1% | |||
| |||
500 | 1.21 ± 0.06 | 1.19 ± 0.11 | |
600 | 1.04 ± 0.05 | 1.05 ± 0.10 | |
700 | 0.886 ± 0.044 | 0.904 ± 0.083 | |
800 | 0.773 ± 0.039 | 0.776 ± 0.071 | |
TiO2 = 0.2% | |||
| |||
500 | 2.30 ± 0.12 | 2.46 ± 0.24 | |
600 | 2.07 ± 0.10 | 2.17 ± 0.21 | |
700 | 1.77 ± 0.09 | 1.86 ± 0.18 | |
800 | 1.54 ± 0.08 | 1.57 ± 0.15 |
5. Conclusion
We present the broadband measurements of the optical parameters of two types of tissue-mimicking solid phantoms, one polyurethane-based from INO and the other made at NIST using a PDMS base material. The optical parameters come with an uncertainty budget obtained by propagation of a Gaussian distribution of input uncertainties at each measurement wavelength. Our instrument measures the hemispherical reflectance and transmittance of the sample at a 0° incident by means of an integrating sphere and the substitution method. The data is analyzed using an inversion routine of the adding-doubling algorithm.
The non-validity of the IAD integrating sphere model is asserted by measuring the hemispherical reflectance under a diffuse illumination of a set of quasi-Lambertian reflectance standards and comparing the results to measurement made with the integrating sphere instrument of NIST STARR. We show that measuring the hemispherical reflectance and the hemispherical transmittance under a direct illumination (0° incident angle) is however possible with our setup. This is accomplished by measuring the same set of reflectance standards (reflectance under a direct illumination, comparison to results by NIST STARR facility) and a set of neutral density filters (transmittance under a direct illumination, comparison to results by NIST RTS). The measured reflectance are then traceable to reflectance and transmittance scales. We define our measurement procedure and restrict the use of the output of the adding-doubling algorithm to the hemispherical reflectance and transmittance under a direct illumination in our inversion routine. The optical properties of the measured sample are accompanied by an uncertainty budget.
A correction based on an independent measurement of the index of refraction is implemented to deal with the rejection of the Fresnel reflectance through the entrance port of the reflectance sphere while measuring specular samples. The optical parameters of the smooth and non specular samples are obtained with expanded uncertainties (k = 2) of about 5 % on μa and 12 % on . They are compared to independent measurements by INO (k = 2 uncertainties: 12 % on μa, 7 % on ). The results for the PDMS samples show similar uncertainties and the values are compared to independent measurements by PTB (k = 2 uncertainties: 8 %).
Our future work plan toward the establishment of a reference scale is to develop an independent measurement of the anisotropy factor g and to define the range of optical properties values measurable by our system. A set of liquid phantoms with different concentration of intralipid and nigrosin are envisioned. A liquid cell has been fabricated for this purpose.
Acknowledgments
Certain commercial materials and equipment are identified in order to adequately specify the experimental procedure. Such identification does not imply recommendation by the National Institute of Standards and Technology. This work was supported by the NIST Innovation in Measurement Science (IMS) program. Authors thank William Guthrie for supporting this work.
Appendix: IAD model of the integrating sphere
The mathematical description of this model can be found in Moffitt’s Ph.D. dissertation [18] and is further expanded here. The sphere has an entrance port (area Ae), a sample port (area As) and a detector port (area Ad). The area of the sphere wall is A and the fractional areas of interest are: , and . For a single sphere, the reflected optical power is
(A1) |
with the sphere gain
(A2) |
where A1 = 1 − awrw − (1 − ae)rwadrd, B1 = (1 − ae)rwas, P is the incident optical power, rw is the reflectance of the sphere wall, rd is the reflectance of the detector, f = 0 for a direct incident illumination and f = 1 for a diffuse incident illumination.
Under a diffuse illumination, i.e. with the incident beam directed toward the sphere wall, two measurements can be made: (i) with the sample port blocked by a reference standard RStdRef and (ii) with the sample set at the sample port of the sphere. The ratio of the reflected optical powers is
(A3) |
The measured voltages and are proportional to the optical powers P(0, ) and P(0, RStdRef), respectively. With an additional measurement of were no sample is set at the sample port of the sphere ,
(A4) |
Combining Eq.(A3) and Eq.(A4) and solving for gives
(A5) |
which defines the diffuse hemispherical reflectance under a diffuse illumination from measurable quantities under the IAD model of the integrating sphere (see Fig. 2(a)).
Under a direct illumination, i.e. with the incident beam directed toward the sample, two measurements can be made: (i) with the sample set at the sample port of the sphere and (ii) with the sample port blocked by a reference standard RStdRef (see Fig. 2(b)). The optical power is then proportional to the reflectance of the sample times the sphere gain for case(i) (RStdRef times for case (ii)). Forming the ratio of the reflected optical powers, the proportionality factor cancels out and we have
(A6) |
where the ratio of sphere gains is obtained as a ratio of voltages from the measurement of , . So
(A7) |
For a single sphere, the measurement of the transmittance under a direct illumination requires to rotate the sphere 180° around a vertical axis perpendicular to the plane of the optical bench, as compared to the previous experimental situation. The transmitted optical power is
(A8) |
In that situation, two measurements can be made: (i) with an empty sample port, i.e. unit transmittance and (ii) with the sample set at the sample port of the sphere. Forming the ratio of the optical powers, the proportionality factor cancels out and
(A9) |
where (the exit port of the sphere is open so that the sphere gain does not change compared to the measurement situation). So
(A10) |
But is not measurable since the beam exits through the exit port of the sphere. It is then assumed that the signal for a unit transmittance is equivalent to the measured signal for an empty sample port when the beam hits the sphere wall. This experimental situation is obtained by tilting the sphere as shown in Fig. 2(c). Then
(A11) |
Disclosures
The authors declare that there are no conflicts of interest related to this article.
References and links
- 1.Spinelli L., Martelli F., Farina A., Pifferi A., Torricelli A., Cubeddu R., Zaccanti G., “Calibration of scattering and absorption properties of a liquid diffusive medium at NIR wavelengths. time-resolved method,” Opt. Express 15, 6589–6604 (2007). 10.1364/OE.15.006589 [DOI] [PubMed] [Google Scholar]
- 2.Liebert A., Wabnitz H., Grosenick D., Möller M., Macdonald R., Rinneberg H., “Evaluation of optical properties of highly scattering media by moments of distributions of times of flight of photons,” Appl. Opt. 42, 5785–5792 (2003). 10.1364/AO.42.005785 [DOI] [PubMed] [Google Scholar]
- 3.Bouchard J.-P., Veilleux I., Jedidi R., Noiseux I., Fortin M., Mermut O., “Reference optical phantoms for diffuse optical spectroscopy. part 1–error analysis of a time resolved transmittance characterization method,” Opt. Express 18, 11495–11507 (2010). 10.1364/OE.18.011495 [DOI] [PubMed] [Google Scholar]
- 4.Pham T. H., Coquoz O., Fishkin J. B., Anderson E., Tromberg B. J., “Broad bandwidth frequency domain instrument for quantitative tissue optical spectroscopy,” Rev. Sci. Inst. 71, 2500–2513 (2000). 10.1063/1.1150665 [DOI] [Google Scholar]
- 5.Bevilacqua F., Berger A. J., Cerussi A. E., Jakubowski D., Tromberg B. J., “Broadband absorption spectroscopy in turbid media by combined frequency-domain and steady-state methods,” Appl. Opt. 39, 6498–6507 (2000). 10.1364/AO.39.006498 [DOI] [PubMed] [Google Scholar]
- 6.Cletus B., Künnemeyer R., Martinsen P., McGlone A., Jordan R., “Characterizing liquid turbid media by frequency-domain photon-migration spectroscopy,” J. Biomed. Opt. 14, 024041 (2009). 10.1117/1.3119282 [DOI] [PubMed] [Google Scholar]
- 7.Martelli F., Zaccanti G., “Calibration of scattering and absorption properties of a liquid diffusive medium at NIR wavelengths. cw method,” Opt. Express 15, 486–500 (2007). 10.1364/OE.15.000486 [DOI] [PubMed] [Google Scholar]
- 8.Di Ninni P., Martelli F., Zaccanti G., “Intralipid: towards a diffusive reference standard for optical tissue phantoms,” Phys. Med. Biol. 56, N21 (2010). 10.1088/0031-9155/56/2/N01 [DOI] [PubMed] [Google Scholar]
- 9.Foschum F., Jäger M., Kienle A., “Fully automated spatially resolved reflectance spectrometer for the determination of the absorption and scattering in turbid media,” Rev. Sci. Inst. 82, 103104 (2011). 10.1063/1.3648120 [DOI] [PubMed] [Google Scholar]
- 10.Andree S., Reble C., Helfmann J., Gersonde I., Illing G., “Evaluation of a novel noncontact spectrally and spatially resolved reflectance setup with continuously variable source-detector separation using silicone phantoms,” J. Biomed. Opt. 15, 067009 (2010). 10.1117/1.3526367 [DOI] [PubMed] [Google Scholar]
- 11.Cuccia D. J., Bevilacqua F., Durkin A. J., Ayers F. R., Tromberg B. J., “Quantitation and mapping of tissue optical properties using modulated imaging,” J. Biomed. Opt. 14, 024012 (2009). 10.1117/1.3088140 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Prahl S. A., van Gemert M. J., Welch A. J., “Determining the optical properties of turbid media by using the adding–doubling method,” Appl. Opt. 32, 559–568 (1993). 10.1364/AO.32.000559 [DOI] [PubMed] [Google Scholar]
- 13.Moffitt T., Chen Y.-C., Prahl S. A., “Preparation and characterization of polyurethane optical phantoms,” J. Biomed. Opt. 11, 041103 (2006). 10.1117/1.2240972 [DOI] [PubMed] [Google Scholar]
- 14.Aernouts B., Zamora-Rojas E., Van Beers R., Watté R., Wang L., Tsuta M., Lammertyn J., Saeys W., “Supercontinuum laser based optical characterization of intralipid phantoms in the 500–2250 nm range,” Opt. Express 21, 32450–32467 (2013). 10.1364/OE.21.032450 [DOI] [PubMed] [Google Scholar]
- 15.Azimipour M., Baumgartner R., Liu Y., Jacques S. L., Eliceiri K., Pashaie R., “Extraction of optical properties and prediction of light distribution in rat brain tissue,” J. Biomed. Opt. 19, 075001 (2014). 10.1117/1.JBO.19.7.075001 [DOI] [PubMed] [Google Scholar]
- 16.Lemaillet P., Bouchard J.-P., Allen D. W., “Development of traceable measurement of the diffuse optical properties of solid reference standards for biomedical optics at National Institute of Standards and Technology,” Appl. Opt. 54, 6118–6127 (2015). 10.1364/AO.54.006118 [DOI] [PubMed] [Google Scholar]
- 17.Lemaillet P., Bouchard J.-P., Hwang J., Allen D. W., “Double-integrating-sphere system at the National Institute of Standards and Technology in support of measurement standards for the determination of optical properties of tissue-mimicking phantoms,” J. Biomed. Opt. 20, 121310 (2015). 10.1117/1.JBO.20.12.121310 [DOI] [PubMed] [Google Scholar]
- 18.Moffitt T. P., “Light transport in polymers for optical sensing and photopolymerization,” Ph.D. thesis, Oregon Health & Science University; (2005). [Google Scholar]
- 19.Prahl S. A., “Inverse adding-doubling,” http://omlc.org/software/iad/.
- 20.Goebel D., “Generalized Integrating-Sphere Theory,” Appl. Opt. 6, 125–128 (1967). 10.1364/AO.6.000125 [DOI] [PubMed] [Google Scholar]
- 21.Zhang Y., Chen Y., Yu Y., Xue X, Tuchin V. V., Zhu D., “Visible and near-infrared spectroscopy for distinguishing malignant tumor tissue from benign tumor and normal breast tissues in vitro,” J. Biomed. Opt. 18, 077003 (2013). 10.1117/1.JBO.18.7.077003 [DOI] [PubMed] [Google Scholar]
- 22.Salomatina E. V., Jiang B., Novak J., Yaroslavsky A. N., “Optical properties of normal and cancerous human skin in the visible and near-infrared spectral range,” J. Biomed. Opt. 11, 064026 (2006). 10.1117/1.2398928 [DOI] [PubMed] [Google Scholar]
- 23.Friebel M., Roggan A., Müller G. J., Meinke M. C., “Determination of optical properties of human blood in the spectral range 250 to 1100 nm using Monte Carlo simulations with hematocrit-dependent effective scattering phase functions,” J. Biomed. Opt. 11, 034021 (2006). 10.1117/1.2203659 [DOI] [PubMed] [Google Scholar]
- 24.“Colorymetry,” Commission Internationale de l’Eclairage Technical Report, 15:2004 (2004). [Google Scholar]
- 25.Hwang J., Kim H.-J., Lemaillet P., Wabnitz H., Grosenick D., Yang L., Gladytz T., McClatchy D., Allen D., Briggman K., Pogue B., “Polydimethylsiloxane tissue-mimicking phantoms for quantitative optical medical imaging standards,” Proc. SPIE 10056, Design and Quality for Biomedical Technologies X, 1005603 (2017). [Google Scholar]
- 26.Levine Z. H., Streater R. H., Lieberson A.-M. R., Pintar A. L., Cooksey C. C., Lemaillet P., “Algorithm for rapid determination of Optical Scattering parameters,” Opt. Express 25, 26728–26746 (2017). 10.1364/OE.25.026728 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Taylor A. H., “Measurement of Diffuse Reflection Factors: And a New Absolute Reflectometer” (US Government Printing Office, 1920). [Google Scholar]
- 28.Germer T. A., Zwinkels J. C., Tsai B. K., “Spectrophotometry: Accurate measurement of optical properties of materials”, vol. 46 (Elsevier, 2014). [Google Scholar]
- 29.Barnes P., Early E. A., Parr A., “NIST Special Publication 250-48,” US Dept. of Commerce; (1998). [Google Scholar]
- 30.Allen D. W., Early E. A., Tsai B. K., Cooksey C. C., “NIST Special Publication 250-69,” US Dept. of Commerce; (2011). [Google Scholar]
- 31.Pickering J. W., Prahl S. A., Van Wieringen N., Beek J. F., Sterenborg H. J., Van Gemert M. J., “Double-integrating-sphere system for measuring the optical properties of tissue,” Appl. Opt. 32, 399–410 (1993). 10.1364/AO.32.000399 [DOI] [PubMed] [Google Scholar]
- 32.“Jcgm 100: Evaluation of measurement data - guide to the expression of uncertainty in measurement,” Tech. Rep., Joint Committee for Guides in Metrology; (2008). [Google Scholar]
- 33.Wang L., Jacques S. L., Zheng L., “MCML-Monte Carlo modeling of light transport in multi-layered tissues,” Comp. Meth. and Prog. Biomed. 47, 131–146 (1995). 10.1016/0169-2607(95)01640-F [DOI] [PubMed] [Google Scholar]
- 34.Spinelli L., Botwicz M., Zolek N., Kacprzak M., Milej D., Sawosz P., Liebert A., Weigel U., Durduran T., Foschum F.,“Determination of reference values for optical properties of liquid phantoms based on intralipid and india ink,” Biomed. Opt. Express 5, 2037–2053 (2014). 10.1364/BOE.5.002037 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Rothfischer R., Grosenick D., Macdonald R., “Time-resolved transmittance: a comparison of the diffusion model approach with Monte Carlo simulations,” Proc. SPIE 9538, 95381H (2015). 10.1117/12.2183762 [DOI] [Google Scholar]