Skip to main content
International Journal of Cardiology. Heart & Vessels logoLink to International Journal of Cardiology. Heart & Vessels
. 2014 Jul 6;4:90–96. doi: 10.1016/j.ijchv.2014.06.008

Regional variations in baseline characteristics of cardiac rhythm device recipients: The PANORAMA observational cohort study

Fawziah Al Kandari a,1, Andrejs Erglis b,1, Raed Sweidan c,1, Ingrid Dannheimer d,1, Milan Sepsi e,1, Juan Bénézet f,1, Michal Padour g,1, Ajay Naik h,1, Jaime Escudero i,1, Teena West j,1, Reece Holbrook k,⁎,1, Faizel Lorgat l,1; On behalf of the PANORAMA investigators
PMCID: PMC5801449  PMID: 29450186

Abstract

Background

The PANORAMA study was designed to collect concurrent data on subjects from different worldwide regions implanted with CRM devices.

Methods

In this prospective, multi-center study, we analyzed baseline data on 8586 subjects implanted with CRM devices with no additional selection criteria (66% pacemaker (IPG), 16% implantable cardiac defibrillators (ICD), 17% cardiac resynchronization therapy (CRT) and < 1% Internal Loop Recorder) from 156 hospitals across 6 geographical regions between 2005 and 2011.

Results

Regardless of the device implanted, subjects from the Middle East and India often had more diabetes than other regions. Eastern and Western Europe had higher rates of atrial fibrillation reported, and men were more likely to smoke than women (46% vs 11%, p < 0.001). Within the CRT cohort there was significant variation in the proportion of males receiving a device, ranging from 55% in India to 83% in Eastern Europe.

Conclusions

We provide comprehensive descriptive data on patients receiving CRM devices from a range of geographies that are not typically reported in literature. We found significant variations in clinical characteristics and implant practices. Long term follow-up data will help evaluate if these variations require adjustments to outcome expectations.

Abbreviations: ICD, implantable cardioverter-defibrillator; IPG, implantable pulse generator; CRT, cardiac resynchronization therapy; CRM, cardiac rhythm management; AV, atrio-ventricular; DFT, defibrillation threshold testing

Keywords: Cardiac pacing, Defibrillators, Cardiac resynchronization, Bradyarrhythmias, Tachyarrhythmias, Heart failure

1. Introduction

Device therapies for treating cardiac rhythm disorders include pacemakers (IPG) and implantable cardiac defibrillators (ICD), both with and without cardiac resynchronization therapy (CRT). Randomized trials have established the effectiveness of device therapies for cardiac rhythm and disease management 1, 2, 3, 4, 5. These trials have played an important role in establishing guidelines for the application of these therapies.

Translating evidence from randomized trials into global clinical practice guidelines involves extrapolating results from a study cohort to the population of interest in the guideline. However it is not a priori certain that study results from a specific patient population in a specific region can be extrapolated to a less well-defined patient population or to other geographies. Regional variations in disease incidence 6, 7, 8, patient demographics and comorbidities [9], genetics [10], healthcare systems and reimbursement conditions [11], cultural attitudes to disease and implant practices exist which affect the choice of therapy and may influence the expected therapy outcomes [12]. Risk factors can be considered and treated differently across geographies [13] and approaches to diagnostic testing can also differ. Understanding the patients, the practice patterns and the healthcare settings of a geography are important for setting expectations and interpreting patient outcomes.

While randomized trials are indispensable in understanding the benefit of therapies in strictly controlled settings, observational studies are designed to assess the relevance and credibility of clinical trial outcomes in real-life settings [14]. Our current understanding of the real world application of cardiac rhythm management (CRM) therapies comes largely from registries conducted in North America and Europe 15, 16. To date there is little evidence available to shed light on regional differences in practice patterns, particularly to understand demographics, comorbidities and treatment patterns in emerging geographies.

This is the first report of the world wide PANORAMA study, a long term, multi-center, prospective, non-randomized observational study. The primary purpose of the study was to construct a computerized database of national profiles and epidemiological data on patients wearing Medtronic implantable pacemakers, cardioverter defibrillators (both with and without cardiac resynchronization therapy), and implantable loop recorders. The study was conducted in 34 countries across 6 geographical regions. Minimal selection criteria ensured that the study population included elderly patients, patients with comorbidities and patients presenting for a replacement device in an effort to ensure that participants were representative of patients receiving therapy for cardiac rhythm disorders. The objective of this analysis is to describe the patients and implant procedures and to provide information on clinical characteristics in regions previously underrepresented.

2. Methods

2.1. Study population

Consenting patients included those who were implanted (either de novo device or replacement) with a Medtronic market-released CRM device (IPG or ICD with or without CRT capability); no other selection criteria were applied. The protocol specified that enrollment should take place within 30 days of the planned/performed implant procedure.

PANORAMA (ClinicalTrials.gov Identifier: NCT00382525) was designed and conducted in compliance with the local ethical considerations and according to the principles outlined in the ‘World Medical Association Declaration of Helsinki’ (October 2000) and the laws and regulations in the countries in which the study was conducted. The study was submitted to locally appointed ethics committees and informed consent was obtained from the subjects (or their guardians).

Patients were enrolled from 6 geographies: Latin America (Argentina, Bahamas, Brazil, Colombia, Dominican Republic, Ecuador, Mexico, Puerto Rico, Uruguay, Venezuela, and Virgin Islands), Western Europe (Austria, Belgium, Denmark, Spain, Germany, Greece, Luxemburg, Netherlands, and United Kingdom), Eastern Europe (Belarus, Czech Republic, Latvia, Lithuania, Poland, Romania, Russian Federation, Serbia, Slovakia, and Turkey), Middle East (Kuwait and Saudi Arabia), South Africa and India.

2.2. Study design

Patients were assessed at study entry and during follow-up visits for at least 1 year after implant. Patients were followed according to the standard follow-up visit scheme of the participating centers, and did not require any procedures beyond regular practice. All treatment decisions were at the discretion of the treating physician. PANORAMA was designed to enroll 10,000 patients.

2.3. Data collection and measures

Clinical data were collected by the investigators using an electronic case report form designed specifically for the study, and stored in a centralized database. The data collected at baseline included demographics and clinical characteristics, medical history, and cardiovascular pharmacological therapy. At implant, data were collected on the implantation procedure and techniques, adverse experiences and device programming.

The IPG cohort includes patients implanted with a single or dual chamber pacemaker. Indications were defined as: AV block (any form of atrioventricular conduction disorder), sinus node disease (any form of atrial based bradyarrhythmia) or other (neither of the previous two). The ICD cohort includes patients implanted with a single or dual chamber ICD. Indications were defined as: secondary prevention (survivor of prior sustained ventricular tachyarrhythmia) or primary prevention (risk factors for sudden cardiac arrest without prior episode). The CRT cohort includes patients implanted with a CRT-D or a CRT-P.

2.4. Statistics

Data are reported as mean + standard deviation (SD), median (interquartile range (IQR)) or as n (percentage). For continuous variables, comparisons across the regional groups were made using analysis of variance (ANOVA) with pairwise comparisons performed using Tukey's studentized range test. For categorical variables a chi-square test was used. All statistical analyses were performed using SAS software version 9.3 (SAS Institute). P values less than 0.05 were considered statistically significant. Stratification of the analysis was specified a priori by region, pathology, indication, and device type.

3. Results

A total of 10,064 subjects were enrolled in the PANORAMA study between 2005 and 2011. From the total study population 1478 patients were excluded from the database due to the lack of evidence of a signed Patient Informed Consent or Patient Data Release Form (1428), or because the same patient was, by mistake, created twice in the electronic database (50). This analysis includes data of the remaining 8586 subjects, all implanted with a CRM device, and enrolled by 156 centers across 34 countries. Two thirds of the study population was implanted with an IPG, 16% with ICD, 17% with CRT, and 1% were implanted with an implantable loop recorder or missing information about the device type. One third of the subjects were enrolled from Eastern Europe (EE), 17% from Western Europe (WE), 17% from South Africa (SA), 17% from the Middle East (ME), 13% from Latin America (LA) and 7% from India (IN) (Table 1).

Table 1.

Overall baseline demographics and device characteristics by region.

Characteristics Total
Eastern Europe
India
Latin America
Middle East
South Africa
Western Europe
n = 8586 n = 2496 n = 571 n = 1127 n = 1463 n = 1432 n = 1497
Number of countries 34 10 1 11 2 1 9
Number of sites 156 34 9 34 5 13 61
Device (%)
IPG 5592 (65) 1769 (71) 368 (64) 840 (75) 782 (53) 827 (58) 1006 (67)
ICD 1407 (16) 322 (13) 94 (16) 129 (11) 400 (27) 61 (4) 401 (27)
CRT 1455 (17) 400 (16) 108 (19) 116 (10) 276 (19) 469 (33) 86 (6)
Othera 132 (2) 5 (< 1) 1 (< 1) 42 (4) 5 (< 1) 75 (5) 46 (1)

IPG = pacemaker; ICD = implantable cardiac defibrillator; CRT = cardiac resynchronization therapy; IQR = interquartile range.

a

Other includes implantable loop recorder (n = 76), missing (n = 56).

Regardless of the type of device implanted, several cardiovascular risk factor patterns were noted to be similar. In particular, subjects from the Middle East had substantially more diabetes present than other regions (47% vs 20% EE, 19% LA, 15% SA, 25% WE, 33% IN, p < 0.001 for ME vs all); subjects from India were significantly lighter in weight (66 ± 13 kg vs 81 ± 16 EE, 71 ± 15 LA, 82 ± 20 SA, 79 ± 16 WE, 76 ± 19 ME, p < 0.001 for IN vs all) and in all regions men were more likely to smoke than females (46% vs 11%, p < 0.001).

There were also distinct differences in the amount of atrial fibrillation reported across the regions with Eastern and Western Europe reporting more atrial fibrillation than other regions (42% EE and 36% WE vs 15% LA, 15% ME, 26% SA, 9% IN, p < 0.001 for EE vs all and p < 0.05 for WE vs all).

The use of general anesthesia for all types of device implants was higher in Latin America than in other regions (26% vs 2% EE, 1% IN, 2% ME, 12% SA, 14% WE, p < 0.001 for LA vs all).

In the IPG cohort 46% were female and 41% were aged over or equal to 75 years of age. Table 2 reports the baseline and initial treatment characteristics of patients who received IPG therapy stratified by region. There were significant differences in median age across the regions ranging from 65 years in the Middle East and India to 76 years in Western Europe (p < 0.001), which remained even after adjusting for replacement device. AV block as a primary indication for implant was more common in India (77%), Latin America (62%) and the Middle East (72%) and least frequent in South Africa (26%) (p < 0.001). 22% of the IPG cohort was implanted with a single chamber device with the highest numbers in Latin America (25%), the Middle East (30%) and South Africa (31%) (p < 0.001). In 62% of the patients implanted with a single chamber device, persistent or permanent AF had not been reported.

Table 2.

Baseline demographics, device and implant characteristics for the IPG cohort by region.

Characteristics Total
Eastern Europe
India
Latin America
Middle East
South Africa
Western Europe
n = 5592 n (%) n = 1769 n (%) n = 368 n (%) n = 840 n (%) n = 782 n (%) n = 827 n (%) n = 1006 n (%)
Median age, years (IQR) 72 (63,79) 73 (66,79) 65 (56,74) 74 (64,80) 65 (51,74) 72 (62,80) 76 (70,81)
Male 3030 (54) 884 (50) 236 (64) 431 (51) 406 (52) 461 (56) 612 (61)
Indication
Sinus node dysfunction 2421 (43) 920 (52) 75 (20) 266 (32) 187 (24) 512 (62) 461 (46)
AV block 2733 (49) 686 (39) 283 (77) 529 (62) 561 (72) 218 (26) 465 (46)
Othera 431 (8) 161 (9) 10 (3) 54 (6) 34 (4) 95 (11) 77 (8)
Body mass index (kg/m2)
Mean (SD) 27 (5.3) 28 (4.7) 25 (4.4) 26 (4.4) 29 (6.9) 27 (5.8) 27 (4.5)
Overweight/obeseb 2914 (52%) 1033 (58%) 170 (46%) 416 (50%) 522 (67%) 363 (44%) 410 (41%)
Previous or current smoker 1558 (28) 478 (27) 55 (15) 299 (36) 172 (22) 160 (19) 394 (39)
Hypertension 3418 (61) 1321 (75) 164 (45) 531 (63) 425 (54) 376 (46) 601 (60)
Diabetes 1258 (23) 379 (21) 118 (32) 166 (20) 292 (37) 109 (13) 194 (19)
Coronary artery disease 1256 (23) 441 (25) 133 (36) 103 (12) 156 (20) 206 (25) 217 (22)
Prior myocardial infarction 477 (9) 237 (13) 17 (5) 42 (5) 44 (6) 54 (7) 83 (8)
Coronary artery bypass grafting 340 (6) 93 (5) 30 (8) 25 (3) 48 (6) 91 (11) 53 (5)
Mean ejection fraction, % (SD) 57 (11.9) 55 (10.4) 54 (11.4) 59 (11.8) 54 (11.6) 59 (12.6) 61 (12.8)
NYHA
I 742 (13) 369 (21) 26 (7) 122 (15) 50 (6) 35 (4) 140 (14)
II 1273 (23) 768 (43) 109 (30) 147 (18) 45 (6) 54 (7) 150 (15)
III 340 (6) 168 (10) 52 (14) 43 (5) 9 (1) 19 (2) 49 (5)
IV 44 (1) 11 (1) 3 (1) 4 (1) 3 (< 1) 18 (2) 5 (1)
No heart failure 3175 (57) 451 (26) 178 (48) 523 (62) 674 (86) 700 (85) 650 (65)
Atrial fibrillation
Paroxysmal 1093 (20) 586 (33) 33 (9) 69 (8) 57 (7) 83 (10) 265 (26)
Persistent/permanent 847 (11) 226 (13) 5 (1) 75 (9) 73 (9) 133 (16) 115 (11)
CHADS2 score 2 or more 2960 (53) 1192 (67) 170 (46) 438 (52) 326 (42) 287 (35) 547 (54)
Single chamber device 1236 (22) 363 (21) 57 (16) 210 (25) 233 (30) 252 (31) 121 (12)
Specialization of implanting physicianc
Electrophysiologist 2122 (38) 436 (25) 363 (99) 385 (46) 741 (95) 124 (15) 73 (7)
Cardiologist 2516 (45) 765 (43) 3 (1) 237 (28) 9 (1) 702 (85) 800 (80)
Surgeon 919 (16) 565 (32) 2 (1) 211 (25) 15 (2) 0 (0) 126 (13)
EP or cardiac catheter lab implant setting 4212 (75) 1199 (68) 364 (99) 617 (74) 763 (98) 812 (98) 457 (45)
General anesthesia 372 (7) 7 (< 1) 4 (1) 178 (21) 24 (3) 59 (7) 100 (10)

IPG = implantable pulse generator; EP = electrophysiology; IQR = interquartile range; SD = standard deviation.

a

Other includes syncope, (bradycardia due to) atrial fibrillation and supraventricular arrhythmia.

b

Overweight defined as BMI (kg/m2) between 25 and 29 and obese defined as BMI > 30.

c

Specialization of physician was self-reported.

Table 3 reports the baseline and initial treatment characteristics of patients who received ICD therapy stratified by region. The proportion of males was higher in the ICD cohort compared to the IPG cohort (82% vs 54%, p < 0.001) and ranged from 77% in South Africa and Latin America to 89% in India. Primary prevention as an indication for implant was highest in the Middle East (66%) and Western Europe (70%) and lowest in South Africa (25%) (p < 0.001). In all regions there was a small proportion of patients that had NYHA class II and QRS > 150 ms or NYHA III/IV and QRS > 120 ms with LVEF < 35% (Table 3). The use of a single chamber device was lowest in the Middle East (25%) and highest in Western Europe (92%) (p < 0.001). The use of defibrillation threshold testing (DFT) at implant varied between regions from 35% in the Middle East to 89% in South Africa. There was no difference in age or gender when comparing patients who received defibrillation testing to those who did not receive defibrillation testing. A decline in the proportion of patients receiving defibrillation testing over time (2005–2007 66% vs 2008–2009 62% vs 2010–2011 48%, p for trend = 0.001) was seen.

Table 3.

Baseline demographics, device and implant characteristics for ICD cohort by region.

Characteristics
Total
Eastern Europe
India
Latin America
Middle East
South Africa
Western Europe
n = 1407 n (%) n = 322 n (%) n = 94 n (%) n = 129 n (%) n = 400 n (%) n = 61 n (%) n = 401 n (%)
Median age, years (IQR) 61 (52,70) 60 (52,68) 56 (48,66) 64 (54,73) 58 (48,67) 58 (44,66) 67 (58,73)
Male 1149 (82) 261 (81) 84 (90) 99 (77) 330 (83) 47 (77) 328 (82)
Indication for ICD
Secondary preventiona 643 (46) 216 (67) 64 (68) 60 (47) 137 (34) 46 (75) 120 (30)
Primary prevention 764 (54) 106 (33) 30 (32) 69 (54) 263 (66) 15 (25) 281 (70)
Body mass index (kg/m2)
Mean (SD) 28 (5.6) 28 (5.2) 25 (4.0) 27 (5.0) 29 (6.3) 28 (6.0) 28 (5.2)
Overweight/obeseb 865 (61%) 196 (61%) 32 (34%) 79 (61%) 295 (74%) 30 (49%) 233 (58%)
Previous or current smoker 699 (50) 168 (52) 18 (19) 53 (41) 220 (55) 24 (39) 216 (54)
Hypertension 774 (55) 195 (61) 21 (22) 63 (49) 236 (59) 25 (41) 234 (58)
Diabetes 481 (34) 49 (15) 29 (31) 26 (20) 227 (57) 3 (5) 147 (37)
Coronary artery disease 944 (67) 208 (65) 57 (61) 72 (56) 284 (71) 34 (56) 289 (72)
Prior myocardial infarction 609 (43) 154 (48) 25 (27) 42 (33) 166 (42) 18 (30) 204 (51)
Coronary artery bypass grafting 273 (19) 54 (17) 23 (25) 14 (11) 64 (16) 12 (20) 106 (26)
Mean ejection fraction, % (SD) 34 (13.3) 40 (14.3) 34 (14.8) 34 (15.0) 31 (10.9) 43 (15.7) 31 (11.2)
NYHA
I 232 (17) 48 (15) 9 (10) 25 (20) 111 (28) 8 (13) 31 (8)
II 594 (42) 138 (43) 46 (49) 45 (35) 164 (41) 9 (15) 192 (48)
III 313 (22) 97 (30) 14 (15) 27 (21) 35 (9) 7 (12) 133 (33)
IV 27 (2) 3 (1) 1 (1) 3 (2) 2 (1) 4 (7) 14 (4)
No heart failure 236 (17) 35 (11) 24 (26) 29 (23) 88 (22) 33 (54) 27 (7)
QRS duration > 120 ms 285 (20) 93 (29) 17 (18) 31 (24) 66 (17) 12 (20) 66 (17)
CRT indicationc 95 (7) 33 (10) 5 (5) 8 (6) 13 (3) 4 (7) 32 (8)
Atrial fibrillation
Paroxysmal 117 (8) 51 (16) 3 (3) 5 (4) 20 (5) 4 (7) 34 (9)
Persistent/permanent 162 (12) 49 (15) 2 (2) 5 (4) 16 (4) 1 (2) 89 (22)
CHADS2 score 2 or more 830 (59) 190 (59) 33 (35) 66 (51) 242 (61) 15 (25) 284 (71)
Single chamber device 804 (57) 179 (56) 73 (78) 40 (31) 98 (25) 44 (72) 370 (92)
Specialization of implanting physiciand
Electrophysiologist 761 (54) 129 (40) 94 (100) 67 (52) 376 (94) 40 (66) 55 (14)
Cardiologist 417 (30) 42 (13) 0 (0) 38 (30) 0 (0) 21 (34) 316 (79)
Surgeon 206 (15) 150 (47) 0 (0) 24 (19) 2 (1) 0 (0) 30 (8)
EP or cardiac catheter lab implant setting 975 (69) 154 (48) 94 (100) 97 (75) 395 (99) 61 (100) 174 (43)
Anesthesia
General anesthesia 224 (16) 38 (12) 0 (0) 60 (47) 4 (1) 32 (53) 90 (22)
Defibrillation testing performed 806 (57) 193 (60) 67 (71) 97 (75) 139 (35) 54 (89) 256 (64)

ICD = implantable cardioverter defibrillator; EP = electrophysiology; IQR = interquartile range; SD = standard deviation.

a

Secondary prevention indication includes: cardiac arrest (presumably) due to VT or VF, spontaneous sustained VT or VF with structural heart disease (symptomatic and non-symptomatic), unexplained syncope with structural heart disease, unexplained syncope with (depressed EF and) inducible VT/VF.

b

Overweight defined as BMI (kg/m2) between 25 and 29 and obese defined as BMI > 30.

c

CRT indication defined as LVEF < 35% with NYHA class II and QRS > 150 ms or NYHA III/IV and QRS > 120 ms.

d

Specialization of physician was self-reported.

The CRT cohort was of similar age with the ICD cohort and younger than the IPG cohort (63 years vs 61 (ICD) and 72 (IPG)). Table 4 reports the baseline and initial treatment characteristics of patients who received CRT therapy stratified by region. Notable differences between regions in subject demographics include male gender ranging from 55% in India to 83% in Eastern Europe, and median age ranging from 58 years in India to 73 years in Western Europe. Regional variation in the use of CRT-D vs CRT-P was considerable with 85% of the CRT devices implanted in the Middle East having defibrillator functionality compared to only 12% in South Africa. 20% of the CRT-D cohort was female compared to 33% of the CRT-P cohort, p < 0.001. In those that had QRS data available, less than half of the CRT recipients in South Africa, had a QRS > 120 ms. 50% of the patients had the LV lead implanted in the posterolateral location, but in a significant percentage of patients in Eastern Europe and Latin America, the LV lead was implanted in the middle lateral vein location. The use of defibrillation threshold testing (DFT) for CRT-D device implants varied between regions from 7% in South Africa to 30% in Latin America.

Table 4.

Baseline demographics, device and implant characteristics for CRT cohort by region.

Characteristic
Total
Eastern Europe
India
Latin America
Middle East
South Africa
Western Europe
n = 1455 n (%) n = 400 n (%) n = 108 n (%) n = 116 n (%) n = 276 n (%) n = 469 n (%) n = 86 n (%)
Median age, years (IQR) 63 (54,71) 59 (51,67) 58 (50,66) 63 (53,71) 64 (56,72) 64 (56,72) 73 (67,77)
Male 1052 (72) 332 (83) 59 (55) 85 (73) 199 (72) 317 (68) 60 (70)
Body mass index (kg/m2)
Mean (SD) 29 (6.0) 29 (5.1) 25 (4.1) 26 (5.0) 30 (6.6) 30 (6.6) 27 (4.7)
Overweight/obesea 886 (61%) 260 (65%) 44 (41%) 52 (45%) 206 (75%) 277 (59%) 47 (55%)
Previous or current smoker 568 (39) 206 (52) 17 (16) 38 (33) 131 (47) 141 (30) 35 (41)
Hypertension 785 (54) 232 (58) 26 (24) 59 (51) 183 (66) 241 (51) 44 (51)
Diabetes 433 (30) 70 (18) 43 (40) 24 (21) 172 (62) 97 (21) 27 (31)
Coronary artery disease 616 (42) 199 (50) 46 (43) 33 (28) 139 (50) 143 (31) 56 (65)
Prior myocardial infarction 316 (22) 159 (40) 10 (9) 23 (20) 58 (21) 35 (8) 31 (36)
Coronary artery bypass grafting 245 (17) 74 (19) 17 (16) 8 (7) 47 (17) 71 (15) 28 (33)
Mean ejection fraction, % (SD) 29 (12.2) 28 (7.8) 23 (9.0) 23 (7.8) 25 (9.4) 35 (16.6) 30 (7.2)
NYHA
I 89 (6) 2 (1) 5 (5) 5 (4) 39 (14) 35 (8) 3 (4)
II 246 (17) 42 (11) 9 (8) 23 (20) 69 (25) 90 (19) 13 (15)
III 689 (47) 273 (68) 71 (66) 55 (47) 141 (51) 107 (23) 42 (49)
IV 318 (22) 82 (21) 23 (21) 25 (22) 22 (8) 163 (35) 3 (4)
No heart failure 93 (6) 1 (< 1) 0 (0) 8 (7) 5 (2) 74 (16) 5 (6)
QRS duration, ms
Not reported 271 (19) 56 (14) 62 (57) 37 (32) 18 (7) 59 (13) 39 (45)
< 120 385 (27) 42 (11) 4 (4) 12 (10) 45 (16) 275 (59) 7 (8)
120 to 150 378 (26) 120 (30) 9 (8) 26 (22) 128 (46) 79 (17) 16 (19)
> 150 421 (29) 182 (46) 33 (31) 41 (35) 85 (31) 56 (12) 24 (28)
Left bundle branch block 712 (49) 264 (66) 82 (76) 49 (42) 173 (63) 98 (21) 46 (54)
Right bundle branch block 81 (6) 18 (5) 9 (8) 8 (7) 29 (11) 15 (3) 2 (2)
Atrial fibrillation
Paroxysmal 147 (10) 45 (11) 3 (3) 2 (2) 27 (10) 59 (13) 11 (13)
Persistent/permanent 223 (15) 94 (24) 3 (2) 12 (10) 22 (8) 79 (17) 13 (15)
CHADS2 score 2 or more 931 (64) 265 (66) 53 (49) 65 (56) 211 (76) 285 (61) 52 (61)
CRT defibrillator 682 (50) 212 (53) 31 (29) 87 (75) 234 (85) 58 (12) 60 (70)
Specialization of implanting physicianb
Electrophysiologist 775 (53) 175 (44) 108 (100) 74 (64) 273 (99) 116 (25) 29 (34)
Cardiologist 439 (30) 10 (3) 0 (0) 22 (19) 0 (0) 352 (75) 55 (64)
Surgeon 238 (16) 215 (54) 0 (0) 20 (17) 0 (0) 1 (< 1) 2 (2)
EP or cardiac catheter lab implant setting 1246 (86) 229 (57) 108 (100) 105 (91) 275 (100) 461 (98) 68 (79)
General anesthesia 154 (11) 15 (4) 0 (0) 45 (39) 5 (2) 78 (17) 11 (13)
LV lead location
Postero lateral 545 (50) 107 (36) 42 (46) 28 (33) 86 (48) 269 (69) 13 (24)
Middle lateral vein 218 (20) 90 (30) 7 (8) 25 (30) 19 (11) 69 (18) 8 (15)
Basal posterior vein 66 (6) 15 (5) 1 (1) 2 (2) 36 (20) 8 (2) 4 (7)
Middle anterior vein 66 (6) 29 (10) 5 (6) 9 (11) 2 (1) 19 (5) 2 (4)
Middle posterior vein 38 (4) 14 (5) 0 (0) 7 (8) 14 (8) 1 (< 1) 2 (4)
Other 113 (11) 34 (11) 30 (33) 8 (10) 21 (12) 20 (6) 0 (0)
Defibrillation testing performedc 215 (15) 93 (23) 10 (9) 35 (30) 36 (13) 31 (7) 10 (12)

CRT = cardiac resynchronization therapy; EP = electrophysiology; LV = left ventricular; IQR = interquartile range; SD = standard deviation.

a

Overweight defined as BMI (kg/m2) between 25 and 29 and obese defined as BMI > 30.

b

Specialization of physician was self-reported.

c

Counts and percentages based on CRT-defibrillator devices only. Percentages do not always add to 100, due to missing values.

4. Discussion

The PANORAMA database provides unprecedented observational data collected from more than 8500 subjects being treated for cardiac rhythm disturbances from diverse geographies, facilitating comparisons and examination of patient demographics, disease comorbidities, and implant practices. Despite some difficulties associated with conducting research outside of the usual western settings, the importance of including such subjects should not be underestimated; majority of RCT data for cardiac rhythm disturbances are based on North American (US) patients. However, when looking to the worldwide implanted CRM devices, the US population receives only about 20% of the total number of pacemakers and 40% of the total implanted ICDs [17].

The subjects enrolled in PANORAMA showed important differences across the regions. Subjects from India and the Middle East were distinct from those in other regional groups in terms of the high prevalence of diabetes and coronary artery disease. This confirms previous publications, and likely reflects genetic and environmental differences and dietary preferences 9, 18, 19. Differences across the regions were most pronounced when looking at the indications and type of device, for example in India, Latin America and the Middle East, there was more use of IPG therapy in AV block (a more stark presentation) than sinus node disease and in South Africa very few ICDs were implanted for primary prevention. It is likely that these differences reflect the economic situation and reimbursement and insurance structures within a country, however, there also seems to be an evolution in pacing indication, from primarily AV block in the early regional pacing era to more sinus node disease when cardiac pacing has become more common [20]. Of the 34 participating countries, 71% are considered to be emerging or developing economies, based on International Monetary Fund (IMF) guidelines [21]. This represents 82% of the study population.

The diagnosis of AF appeared to be comparatively low in Latin America, India and the Middle East. Recent evidence is indicating that even small amounts of AF regardless of symptoms may be correlated with increased stroke risk [22].

Also seen were expected patterns such as the higher use of general anesthesia which reflects local preferences. Other observations were more difficult to explain such as DFT at implants for ICD devices with and without CRT capability. This may reflect changing attitudes about the clinical need for DFT at implant, for which the evidence base is growing 23, 24, 25.

In general, the PANORAMA IPG cohort overall was similar to previously reported cohorts. Compared to the DINAMIT and MADIT-II trials, the PANORAMA ICD cohort had slightly more comorbidities present at implant 26, 27 but when compared with previous secondary prevention studies our cohort was similar [28]. We saw a higher prevalence of single chamber ICD use, predominantly in Western Europe, India and South Africa as compared to other regions. This may not only be reflective of economic realities, but may also indicate differing attitudes regarding the level of evidence available for patients without a pacemaker indication [29]. The PANORAMA study included CRT patients with a fairly typical patient profile. When compared to the European CRT survey [16] our cohort was of similar age but had more CRT-P patients and slightly more patients with NYHA II. There were a number of patients receiving CRT therapy despite the lack of documentation of a QRS duration greater than 120 ms. Although it might be possible that the QRS duration has been less documented for those patients with a QRS duration greater than 120 ms, it might also be that the PANORAMA reflects real clinical practice of implanting CRT devices for other (clinical) reasons, irrespective of the QRS duration. There is evidence that CRT devices are electively implanted in patients with a QRS duration below 150 ms after total AV node ablation for AF 30, 31. However, this cannot be confirmed by the PANORAMA data since prior AV nodal ablation information had not been collected.

Regardless of the reasons for heterogeneity in subjects and therapy practice these differences are likely to impact the outcomes for patients. A smoking history was present in more than 50% of the male ICD and CRT patients, and from the male patients receiving a first ICD up to 24% (Middle East) was smoking at the time of device implant. From the male patients receiving a first CRT the percentage of current smokers was up to 18% (South Africa and Eastern Europe). As failure to stop smoking will likely have a negative impact on survival of these patients, especially those with ischemic heart disease, the PANORAMA data may indicate a need to strongly emphasize smoking cessation possibilities to patients before receiving a medical device.

5. Study limitations

These data need to be interpreted in the context of the limitations of the study. This was a single-manufacturer sponsored study where specific device characteristics might limit applicability across all device manufacturers; however, an attempt was made to compare general therapy parameters that would apply regardless of the manufacturer. In this global analysis, some regions of the world were not represented, but this is the most regionally diverse dataset reported to date. The number of centers and countries represented in the data may not be sufficient to accurately characterize the full regional practice and some of the differences observed could be center specific. Some countries limited the type of device enrolled, so broad conclusions about the proportion of IPG/ICD/CRT use are not possible. To limit overinterpretation, in cases where the number of patients was limited (specific device types), general observations were avoided in this report.

Missing data could have introduced biases and where substantial data was missing it has been reported and interpreted with caution. Some of the measures of patient conditions were collected without specific definition (e.g. hypertension), potentially introducing variability into the response; however, these measures are interpreted in a broad, non-quantitative manner. No adjustments were made for multiple comparisons; however, results are reported with p values so the reader can interpret the clinical and statistical significance of the results.

6. Conclusions

We provide comprehensive descriptive data on patients receiving cardiac rhythm management devices from a range of geographies that is not typically reported in literature; this may aid in the interpretation and application of findings from other studies. We found significant variations in age, cardiovascular diseases, rhythm disturbances, type of device implant, and implant practices across the regions. Some of these may be reflections of larger socioeconomic variations, while others may be due to regional guidelines and practice preferences.

Sources of funding

This work was supported by the Bakken Research Center, Maastricht, Netherlands, a wholly owned division of Medtronic, Inc.

Conflicts of interest

Fawziah Al Kandari has received consultant honoraries from Medtronic. Juan Bénézet has received consultant honoraries and speaker fees from Medtronic. Ajay Naik has received consultant honoraries from Medtronic, St. Jude Medical, Boston Scientific, Biotronik, Sanofi and Astra Zeneca. Teena West and Reece Holbrook are Medtronic employees.

Acknowledgments

We would like to thank all PANORAMA investigators and the PANORAMA team for their continued efforts during this long term study.

Footnotes

(ClinicalTrials.gov number: NCT00382525)

Available online 6 July 2014

Appendix A

Supplementary data to this article can be found online at http://dx.doi.org/10.1016/j.ijchv.2014.06.008.

Appendix A. Supplementary data

Supplementary material.

mmc1.docx (39.2KB, docx)

References

  • 1.Bardy G.H., Lee K.L., Mark D.B., Poole J.E., Packer D.L., Boineau R. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352(3):225–237. doi: 10.1056/NEJMoa043399. [DOI] [PubMed] [Google Scholar]
  • 2.Young J.B., Abraham W.T., Smith A.L., Leon A.R., Lieberman R., Wilkoff B. Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD trial. JAMA. 2003;289(20):2685–2694. doi: 10.1001/jama.289.20.2685. [DOI] [PubMed] [Google Scholar]
  • 3.Linde C., Abraham W.T., Gold M.R., Sutton St John M., Ghio S., Daubert C. Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms. J Am Coll Cardiol. 2008;52(23):1834–1843. doi: 10.1016/j.jacc.2008.08.027. [DOI] [PubMed] [Google Scholar]
  • 4.Moss A.J., Hall W.J., Cannom D.S., Klein H., Brown M.W., Daubert J.P. Cardiac-resynchronization therapy for the prevention of heart-failure events. N Engl J Med. 2009;361(14):1329–1338. doi: 10.1056/NEJMoa0906431. [DOI] [PubMed] [Google Scholar]
  • 5.Tang A.S., Wells G.A., Talajic M., Arnold M.O., Sheldon R., Connolly S. Cardiac-resynchronization therapy for mild-to-moderate heart failure. N Engl J Med. 2010;363(25):2385–2395. doi: 10.1056/NEJMoa1009540. [DOI] [PubMed] [Google Scholar]
  • 6.Anjana R.M., Pradeepa R., Deepa M., Datta M., Sudha V., Unnikrishnan R. Prevalence of diabetes and prediabetes (impaired fasting glucose and/or impaired glucose tolerance) in urban and rural India: phase I results of the Indian Council of Medical Research-INdia DIABetes (ICMR-INDIAB) study. Diabetologia. 2011;54(12):3022–3027. doi: 10.1007/s00125-011-2291-5. [DOI] [PubMed] [Google Scholar]
  • 7.Yusuf S., Reddy S., Ounpuu S., Anand S. Global burden of cardiovascular diseases: part II: variations in cardiovascular disease by specific ethnic groups and geographic regions and prevention strategies. Circulation. 2001;104(23):2855–2864. doi: 10.1161/hc4701.099488. [DOI] [PubMed] [Google Scholar]
  • 8.Yusuf S., Reddy S., Ounpuu S., Anand S. Global burden of cardiovascular diseases: part I: general considerations, the epidemiologic transition, risk factors, and impact of urbanization. Circulation. 2001;104(22):2746–2753. doi: 10.1161/hc4601.099487. [DOI] [PubMed] [Google Scholar]
  • 9.Saadi H., Carruthers S.G., Nagelkerke N., Al-Maskari F., Afandi B., Reed R. Prevalence of diabetes mellitus and its complications in a population-based sample in Al Ain, United Arab Emirates. Diabetes Res Clin Pract. 2007;78(3):369–377. doi: 10.1016/j.diabres.2007.04.008. [DOI] [PubMed] [Google Scholar]
  • 10.Prins B.P., Lagou V., Asselbergs F.W., Snieder H., Fu J. Genetics of coronary artery disease: genome-wide association studies and beyond. Atherosclerosis. 2012;225(1):1–10. doi: 10.1016/j.atherosclerosis.2012.05.015. [DOI] [PubMed] [Google Scholar]
  • 11.Bhojani U., Beerenahalli T.S., Devadasan R., Munegowda C.M., Devadasan N., Criel B. No longer diseases of the wealthy: prevalence and health-seeking for self-reported chronic conditions among urban poor in Southern India. BMC Health Serv Res. 2013;13:306. doi: 10.1186/1472-6963-13-306. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Laslett L.J., Alagona P., Clark B.A., Drozda J.P., Saldivar F., Wilson S.R. The worldwide environment of cardiovascular disease: prevalence, diagnosis, therapy, and policy issues: a report from the American College of Cardiology. J Am Coll Cardiol. 2012;60(25 Suppl.):S1–S49. doi: 10.1016/j.jacc.2012.11.002. [DOI] [PubMed] [Google Scholar]
  • 13.Shehab A., Al-Dabbagh B., Almahmeed W., Bustani N., Agrawal A., Yusufali A. Characteristics, management, and in-hospital outcomes of diabetic patients with acute coronary syndrome in the United Arab Emirates. Sci World J. 2012;2012:698597. doi: 10.1100/2012/698597. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Nallamothu B.K., Hayward R.A., Bates E.R. Beyond the randomized clinical trial: the role of effectiveness studies in evaluating cardiovascular therapies. Circulation. 2008;118(12):1294–1303. doi: 10.1161/CIRCULATIONAHA.107.703579. [DOI] [PubMed] [Google Scholar]
  • 15.Hammill S.C., Kremers M.S., Stevenson L.W., Heidenreich P.A., Lang C.M., Curtis J.P. Review of the registry's fourth year, incorporating lead data and pediatric ICD procedures, and use as a national performance measure. Heart Rhythm. 2010;7(9):1340–1345. doi: 10.1016/j.hrthm.2010.07.015. [DOI] [PubMed] [Google Scholar]
  • 16.Bogale N., Priori S., Cleland J.G., Brugada J., Linde C., Auricchio A. The European CRT survey: 1 year (9–15 months) follow-up results. Eur J Heart Fail. 2012;14(1):61–73. doi: 10.1093/eurjhf/hfr158. [DOI] [PubMed] [Google Scholar]
  • 17.Mond H.G., Proclemer A. The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009–a World Society of Arrhythmia's project. Pacing Clin Electrophysiol. 2011;34(8):1013–1027. doi: 10.1111/j.1540-8159.2011.03150.x. [DOI] [PubMed] [Google Scholar]
  • 18.Malik M., Bakir A., Saab B.A., Roglic G., King H. Glucose intolerance and associated factors in the multi-ethnic population of the United Arab Emirates: results of a national survey. Diabetes Res Clin Pract. 2005;69(2):188–195. doi: 10.1016/j.diabres.2004.12.005. [DOI] [PubMed] [Google Scholar]
  • 19.Pradeepa R., Prabhakaran D., Mohan V. Emerging economies and diabetes and cardiovascular disease. Diabetes Technol Ther. 2012;14(Suppl. 1):S59–S67. doi: 10.1089/dia.2012.0065. [DOI] [PubMed] [Google Scholar]
  • 20.Ector H., Rickards A.F., Kappenberger L., Linde C., Vardas P., Oto A. The world survey of cardiac pacing and implantable cardioverter defibrillators: calendar year 1997—Europe. Pacing Clin Electrophysiol. 2001;24(5):863–868. doi: 10.1046/j.1460-9592.2001.00863.x. [DOI] [PubMed] [Google Scholar]
  • 21.International Monetary Fund; 2008. World Economic Outlook Housing and the Business Cycle April 2008. http://www.imf.org/external/pubs/ft/weo/2008/01/ [[cited 2013 8th February]. Statistical Appendix, Table F] [Google Scholar]
  • 22.Lip G.Y., Brechin C.M., Lane D.A. The global burden of atrial fibrillation and stroke: a systematic review of the epidemiology of atrial fibrillation in regions outside North America and Europe. Chest. 2012;142(6):1489–1498. doi: 10.1378/chest.11-2888. [DOI] [PubMed] [Google Scholar]
  • 23.Russo A.M., Sauer W., Gerstenfeld E.P., Hsia H.H., Lin D., Cooper J.M. Defibrillation threshold testing: is it really necessary at the time of implantable cardioverter-defibrillator insertion? Heart Rhythm. 2005;2(5):456–461. doi: 10.1016/j.hrthm.2005.01.015. [DOI] [PubMed] [Google Scholar]
  • 24.Gula L.J., Massel D., Krahn A.D., Yee R., Skanes A.C., Klein G.J. Is defibrillation testing still necessary? A decision analysis and Markov model. J Cardiovasc Electrophysiol. 2008;19(4):400–405. doi: 10.1111/j.1540-8167.2007.01095.x. [DOI] [PubMed] [Google Scholar]
  • 25.Kolb C., Tzeis S., Zrenner B. Defibrillation threshold testing: tradition or necessity? Pacing Clin Electrophysiol. 2009;32(5):570–572. doi: 10.1111/j.1540-8159.2009.02328.x. [discussion 2] [DOI] [PubMed] [Google Scholar]
  • 26.Steinberg J.S., Fischer A., Wang P., Schuger C., Daubert J., McNitt S. The clinical implications of cumulative right ventricular pacing in the multicenter automatic defibrillator trial II. J Cardiovasc Electrophysiol. 2005;16(4):359–365. doi: 10.1046/j.1540-8167.2005.50038.x. [DOI] [PubMed] [Google Scholar]
  • 27.Dorian P., Hohnloser S.H., Thorpe K.E., Roberts R.S., Kuck K.H., Gent M. Mechanisms underlying the lack of effect of implantable cardioverter-defibrillator therapy on mortality in high-risk patients with recent myocardial infarction: insights from the defibrillation in acute myocardial infarction trial (DINAMIT) Circulation. 2010;122(25):2645–2652. doi: 10.1161/CIRCULATIONAHA.109.924225. [DOI] [PubMed] [Google Scholar]
  • 28.Brodsky M.A., McAnulty J., Zipes D.P., Baessler C., Hallstrom A.P., Investigators A. A history of heart failure predicts arrhythmia treatment efficacy: data from the antiarrythmics versus implantable defibrillators (avid) study. Am Heart J. 2006;152(4):724–730. doi: 10.1016/j.ahj.2006.04.021. [DOI] [PubMed] [Google Scholar]
  • 29.Almendral J., Arribas F., Wolpert C., Ricci R., Adragao P., Cobo E. Dual-chamber defibrillators reduce clinically significant adverse events compared with single-chamber devices: results from the DATAS (dual chamber and atrial tachyarrhythmias adverse events study) trial. Europace. 2008;10(5):528–535. doi: 10.1093/europace/eun072. [DOI] [PubMed] [Google Scholar]
  • 30.Brignole M., Botto G., Mont L., Iacopino S., De Marchi G., Oddone D. Cardiac resynchronization therapy in patients undergoing atrioventricular junction ablation for permanent atrial fibrillation: a randomized trial. Eur Heart J. 2011;32(19):2420–2429. doi: 10.1093/eurheartj/ehr162. [DOI] [PubMed] [Google Scholar]
  • 31.Stavrakis S., Garabelli P., Reynolds D.W. Cardiac resynchronization therapy after atrioventricular junction ablation for symptomatic atrial fibrillation: a meta-analysis. Europace. 2012;14(10):1490–1497. doi: 10.1093/europace/eus193. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary material.

mmc1.docx (39.2KB, docx)

Articles from International Journal of Cardiology. Heart & Vessels are provided here courtesy of Elsevier

RESOURCES