The Proximal Optimisation Technique for Intervention of Coronary Bifurcations

Angela Hoye 1
PMCID: PMC5808474  PMID: 29588738

Abstract

The proximal optimisation technique (POT) has been proposed as a strategy to improve the results of stent scaffolding of bifurcation lesions. It is a straightforward technique whereby a short, appropriately-sized balloon is inflated in the main vessel just proximal to the carina. The technique has several advantages: it reduces the risk of side branch compromise related to shifting of the carina, it improves stent apposition in the proximal main vessel, and it facilitates side branch access after main vessel stent implantation. When treating bifurcations, final kissing balloon dilation (KBD) has traditionally been used routinely to optimise angiographic results. However, recent clinical data have questioned this philosophy and bench models have demonstrated several shortcomings of KBD. Instead, the optimal strategy may centre on performing POT, followed by side branch dilation, and completed with a final (re)-POT. The following review article describes how to perform POT and presents the evidence to support its’ routine use.

Keywords: Bench test, bifurcation, mal-apposition, optimisation, side branch, stent


The coronary tree is comprised of arteries which divide into ever smaller branches to supply the myocardium. This means that the diameter of the vessel proximal to a bifurcation is always larger than the diameter of the main vessel distal to the bifurcation. The proximal optimisation technique (POT) was proposed by Dr Olivier Darremont as a technique to compensate for this difference in diameters when bifurcation lesions are stented. The technique is fully supported by the European Bifurcation Club and is a key part of the consensus statement produced by this group of experts.[1] A short balloon is expanded in the proximal main vessel just up to the carina. This enables full expansion and complete apposition of the stent in the proximal main vessel. In addition, the POT also facilitates re-crossing into the side branch in order to optimise the final result if required. The following review article describes how to perform the technique and presents the evidence to support its routine use.

Murray’s Law

The coronary tree is composed of vessels which branch into vessels of decreasing diameters. This follows Murray’s Law which states that the radii of daughter branches are related to the radius of the parent branch. Therefore, at each branch point within the coronary tree, the diameter of the vessels can be predicted. Professor Gérard Finet simplified Murray’s Law and showed that the proximal main vessel diameter closely approximates to the sum of the daughter vessel diameters (distal main plus side) multiplied by 0.67 (Figure 1).

Figure 1: Finet’s formula of bifurcation vessel size.

Figure 1:

The proximal main vessel lumen diameter (D1) = 0.678 x (distal main vessel diameter (D2) + side branch diameter (D3)).

Therefore, for all bifurcations, the diameter of the distal main vessel will be consistently smaller than the diameter of the proximal main vessel. The greater the diameter of the side branch, the greater is the disparity between the diameter of the proximal main vessel and the vessel distal to the bifurcation. This difference in diameters is important when selecting a stent of appropriate diameter. If a stent is chosen based on the diameter of the proximal main vessel, this will be ‘over-sized’ for the distal main vessel. This may be associated with compromise of flow into the side branch as it can lead to shifting of the carina (Figure 2A). However, if the stent is chosen based on the diameter of the distal main vessel, the proximal part of the stent will be small relative to the vessel size (Figure 2B). This mal-apposition has implications particularly if a new guidewire is advanced through the stent as it may inadvertently pass behind the stent struts. Any further balloon dilatation will then crush the stent potentially leading to major complications including stent thrombosis.[2] Importantly, this may not necessarily be evident to the operator on angiography or by ‘tactile feedback’ of the smoothness of wire passage.

Figure 2: Bifurcation lesion.

Figure 2:

(A) The stent is sized to the proximal reference diameter of the main vessel. Stent deployment will displace the carina thereby compromising flow into the side branch. (B) The stent is sized to the distal reference diameter of the main vessel. Following stent deployment, the proximal part of the stent is mal-apposed. (C) Illustration of the proximal optimisation technique. A short balloon sized to the proximal main vessel diameter is positioned just up to the carina and inflated. (D) Following POT, there is good apposition of the stent throughout its length and without shifting of the carina. POT= Proximal Optimisation Technique

Technique

The aim of the POT is to fully appose the stent proximal to the bifurcation; in addition, POT expands the stent cells overlying the side branch ostium. A short balloon is taken and must be accurately positioned with the distal marker of the balloon placed at the carina, level with the side branch ostium (Figure 2C). This is best done by confirming a good position on orthogonal views. If positioned too distal, this could cause carinal shift, if too proximal then this will not appropriately expand the stent overlying the side branch ostium. If required, an additional balloon inflation is made more proximally to ensure that the inflow of the stent is fully apposed. To help accurately position the balloon, it is useful to use radiographic optimisation with techniques such as stent boost imaging. It is important to choose a correctly-sized balloon (non-compliant or semi-compliant) that will adequately expand the proximal part of the stent.

Evidence in Favour of POT from Bench Models

Bench models have been key to developing the concept of the POT and demonstrate the associated excellent anatomical / scaffolding result.[36] In 2011, Foin et al.[3] evaluated the concept in a bench model of commercially available stents. They demonstrated that a final POT was able to significantly improve the anatomical result following single stent implantation for a bifurcation, confirming the reduction in mal-apposition of the stent in the proximal main vessel. Furthermore, they also showed the potential downside of final kissing balloon dilation (KBD) which has traditionally been recommended in clinical practice. KBD caused asymmetrical stent (over)expansion in the region where both balloons were expanded together, as well as development of mal-apposition of the proximal part of the main vessel stent (Figure 3).

Figure 3: Effect of kissing balloon dilation inflation on strut distribution during provisional bifurcation stenting and Impact of Proximal Optimisation Technique on stent apposition and side branch access.

Figure 3:

(A) Effect of kissing balloon dilation (KBD) inflation on strut distribution during provisional bifurcation stenting Strut mal-apposition and stent under-expansion during stenting of the main vessel (A.i) and dilation of the side branch (A.ii). Longitudinal and cross-sectional views show the asymmetrical distortion of the stent in the proximal main vessel after KBD (A.iii). Post-dilation of the proximal part of the main vessel (proximal optimisation technique) corrects the stent distortion while ensuring complete apposition of stent struts (A.iv). (B) Impact of proximal optimsation technique on stent apposition and side branch access. Comparison of the stent before (deployment) and after POT. View of the strut configuration from the side branch (right panel) shows how POT facilitates re-crossing by creating a ‘funnel effect’ through enlargement of the cells located in front of the side branch, reducing the number of possible cells available for crossing. Source: Foin et al. 2011[3] and Foin et al. 2013.[13] Reproduced with permission from Europa Digital & Publishing, © 2011, 2013.

Computational simulations have been useful in assessing the negative impact of KBD. Studies have shown that in addition to creating elliptical deformation of the stent in the proximal main vessel, KBD also causes altered strut configuration, damage to the stent coating, as well as arterial injury to the side branch ostium.[79] KBD results in higher stresses in the arterial wall[10] and stent ‘over-stretch’ has been shown to increase the inflammatory response to stent implantation. Otake et al. demonstrated, by Optical Coherence Tomography (OCT), that stent asymmetry was correlated with stent thrombosis and an increased rate of major adverse clinical events (MACE) in patients treated with drug-eluting stents.[11]

Foin and colleagues went on to carry out a further bench study and proposed a step-wise post-dilation strategy in preference to KBD.[4] In this second study, following stent implantation, they performed POT followed by side branch balloon-inflation. The final step was either KBD or a final POT. As before, KBD was shown to have unfavourable features causing elliptical over-expansion of the stent proximal to the carina, as well as distortion of the proximal part of the main vessel stent with development of mal-apposition (even though POT had been done prior to the KBD). However, the step-wise approach (POT, side branch dilation, then re-POT) led to a more favourable result with significantly less strut mal-apposition at the proximal edge of the stent. In addition, POT expanded the stent cell overlying the side branch ostium. This is advantageous as it has the potential to make it easier to re-cross a guidewire into the side branch and may also facilitate ‘optimal re-cross’. When a provisional strategy is used, re-cross should ideally be performed through a distal stent cell.[12] This ensures that balloon inflation will produce better clearance of struts from the side branch ostium.

The studies by Foin have been subsequently supported by a further bench model study[6] that evaluated six strategies: (i) KBD alone, (ii) POT then KBD, (iii) POT then asymmetrical KBD with the main vessel balloon inflated and maintained at 12 atm and the side branch balloon inflation reduced from 12 atm to 4 atm, (iv) POT then asymmetrical KBD with side branch balloon inflated and maintained at 12 atm and the main vessel balloon inflation reduced from 12 atm to 4 atm, (v) POT then side branch inflation (SBI), (vi) POT then SBI at 12 atm then re-POT.[6] Each strategy was evaluated in two different stent designs. The results demonstrated that the best geometric result was obtained from POT – SBI –re-POT. Compared to techniques utilising KBD, the POT – SBI – re-POT strategy reduced the proximal area overstretch and restored circularity, reduced the side branch ostium stent-strut obstruction, and significantly reduced global strut mal-apposition (from 40 ± 6.2 % to 2.6 ± 1.4 %).

This reduction in strut mal-apposition may have important clinical implications as strut mal-apposition disturbs blood flow causing high shear stress. This activates platelets and has been shown to delay neointimal coverage of struts, both potentially increasing thrombogenicity and the potential for late stent thrombosis.[14,15] Furthermore, disturbed flow patterns caused by the geometrical complexity of bifurcation lesions may explain the relative increase in the risk of developing restenosis.[16]

Recently, Murasato et al.[5] used a bench model to suggest that a compliant balloon may be preferable to the use of a non-compliant balloon to optimise strut apposition during POT. High pressure is not a pre-requisite as the balloon merely needs to dilate the unopposed stent; however, an adequately large balloon diameter is obligatory. For some larger vessels, the expansion range of non-compliant balloons will limit their ability to fully appose the struts whereas the greater expansion range of a semi-compliant balloon could be advantageous. For example, if a vessel is 4.9 mm diameter, a 4.5 mm non-compliant balloon may only reach a maximal diameter of 4.8 mm whereas a semi-compliant balloon will easily expand to 5.0 mm.

All of these bench studies provide evidence to support the potential benefits of POT (Table 1) and emphasise the importance of always completing the procedure with a final POT. This can correct both the mal-apposition as well as improve the elliptical shape back to a circular lumen which will favourably influence the flow dynamics. This information forms the basis by which experts including the European Bifurcation Club have adopted and strongly support the routine use of POT.

Table 1: Potential Benefits of the Proximal Optimisation Technique.

Reduce of strut mal-apposition especially in the main vessel proximal to the bifurcation
Reduce the risk of inadvertent abluminal re-wiring of the main vessel stent
Open the stent cells overlying the side branch ostium
Facilitate re-wiring into the side branch
Optimise the final stent geometry and flow dynamics

Implications for single stent bifurcation strategy

It is important to appreciate that the clinical trials evaluating various bifurcation strategies (e.g. one stent versus two-stent strategies) were performed prior to the widespread adoption of POT. However, there seems no reason why the important information obtained from bench model studies should not be incorporated into clinical practice.

Randomised controlled studies have demonstrated that, for the majority of bifurcations, a single stent strategy is preferred over a two-stent strategy; with recent data suggesting improved 5-year survival for the simpler technique.[17] The recommendation from the EBC is that both vessels are wired, with pre-dilation of the main vessel if required.[1] Pre-dilatation of the side branch is to be avoided if possible as, if this causes dissection, the side branch will be more difficult to re-wire once the main vessel has been stented. The main vessel is stented using a stent diameter based on the reference diameter of the main vessel distal to the bifurcation (Figure 4). The stent must be long enough to cover sufficient length of the proximal main vessel to allow POT to be performed, bearing in mind that the shortest available balloon may be 6 mm or 8 mm in length. After stenting, POT is performed to optimise stent expansion. In addition, this balloon dilatation will also open the stent cells overlying the side branch ostium and may partially scaffold the side branch ostium. In the majority of situations, the procedure can be terminated at this point. However, if there is concern regarding flow into the side branch, POT facilitates successful re-wiring of the side branch to enable further balloon dilatation +/- stent implantation. Based on information from the bench models, the procedure should always end with a final POT.

Figure 4: Case example of Proximal Optimisation Technique during provisional stenting in a 67-year-old patient admitted to hospital with an acute coronary syndrome.

Figure 4:

There is a tight stenosis at the ostium of the left anterior descending artery seen on angiography (A) as well as by intravascular ultrasound (B). There was no significant plaque at the ostium of the left circumflex artery (C) so a single stent provisional strategy was chosen. The reference diameter of the left anterior descending artery was 3.0 mm (D) so the lesion was treated with a 3.0 x 23 mm Xience stent (Abbott Vascular) implanted at 16 atm. There is significant mal-apposition of the proximal part of the stent in the left main stem (E). A short 4.5 mm balloon was inflated in the left main stem to perform POT (F). Excellent final result both on angiography (G) as well as on intravascular ultrasound (H).

Although there is a lack of robust clinical data to demonstrate the advantages of POT, there is evidence from a randomised study to show that that there is no benefit (or perhaps a disadvantage) to undertaking KBD routinely.[18,19] The NORDIC III trial randomised 477 patients with bifurcation lesions to final KBD versus no KBD after main vessel stenting.[19] At 6 months, there was no significant difference in the primary endpoint of cardiac death, non-procedural myocardial infarction, target lesion revascularisation (TLR) or stent thrombosis (2.1 % versus 2.5 %, p=1.0). However, the KBD procedures were longer and utilised higher contrast use. The 5-year clinical outcomes of this trial were presented in 2015.[20] There were no differences in the rates of myocardial infarction, TLR or stent thrombosis. Interestingly however, cardiac mortality was significantly higher in the group treated with routine KBD (4.2 % versus 0.8 %, p=0.02). Though this may be a chance finding, it could also be hypothesized that in the absence of routine POT, some of the late events might be due to inadvertent abluminal wiring occurring in the pursuit of performing KBD.[2]

The efficacy of KBD from registry data is contradictory. The first COBIS registry[21] evaluated 1,065 patients of which 736 (69 %) were treated with KBD. At 22 months’ follow-up, there was a significantly higher rate of MACE in those treated with KBD (10.0 % versus 4.9 %, p<0.05), driven by a higher rate of TLR in the KBD group (9.1 % versus 3.4 %, p<0.05). The subsequent COBIS II registry demonstrated quite the opposite and suggested that KBD might in fact be favourable.[22] This study evaluated 1,901 patients with a bifurcation lesion treated by single drug-eluting stent implantation. KBD was performed at the discretion of the operator in 620 patients (32.6 %), notably a lower proportion than that seen in the first COBIS registry. At a median follow-up of 36 months, those treated with KBD had a significantly lower rate of MACE (adjusted hazard ratio [HR]: 0.68, 95 % CI: 0.46 to 0.99; p=0.048). This benefit remained apparent after propensity score matching (545 pairs, adjusted HR: 0.50, 95 % CI: 0.30 to 0.85; p=0.01); with a significantly lower rate of TLR in the KBD group.

Ongoing in vivo work is being undertaken to evaluate POT in clinical practice and will potentially provide clinical evidence to back-up the evidence from bench studies. Studies have been designed that incorporate OCT imaging as a tool to identify stent strut mal-apposition, which is proposed as a surrogate marker for a higher risk of adverse clinical events.

Implications for a 2-stent strategy

Whatever stenting technique is used, POT is an integral part of two-stent strategies to appose the stent in the proximal main vessel as well as to open up the stent cells to facilitate easier guidewire crossing through the side of the stent. In contrast to a single stent strategy, there is strong evidence in favour of kissing balloon dilation for two-stent strategies.[23,24] Failure to perform successful KBD is associated with both a greater rate of stent thrombosis as well as a significantly higher rate of restenosis, TLR and MACE. Indeed, some of the best clinical results have been obtained from the double kissing strategy of the DK-crush technique.[2527] Figure 5 illustrates the principle of incorporating POT during two-stent bifurcation procedures. The first POT is performed after the first stent has been implanted, to appose the stent in the proximal main vessel and open the struts to facilitate optimal guidewire re-crossing. The second stent is implanted according to the preferred strategy, with further POT used as required to facilitate guidewire re-crossing. Final KBD is mandated, however, as demonstrated in Figures 5A and 6B, such KBD causes elliptical deformation of the stented proximal main vessel. This has the potential to impact negatively on the flow dynamics, but can be improved following a final POT which restores the circular morphology.[28]

Figure 5: Case example of POT during a two-stent strategy in a 79-year-old patient with an acute coronary syndrome and on-going symptoms of chest discomfort.

Figure 5:

(A) There is a filling defect in the ostium of the left circumflex artery (LCx) extending back into the left main stem and compromising flow into the left anterior descending artery (LAD). There was no improvement in the angiographic appearances after pre-dilation so a 2-stent strategy was performed (Culotte technique). The first stent was implanted in the LCx back to the ostium of the LMS (B). POT was performed to fully appose the proximal part of the stent in the LMS (C) and facilitate re-wiring of the LAD. After further balloon dilation, the LAD was stented to cover the ostium of the LMS (D). The result was optimised with kissing balloon inflation (E) with a good angiographic result (F). LAD= left anterior descending artery; LCx=left circumflex artery; POT= Proximal Optimisation Technique

The importance of POT to facilitate optimal guidewire re-crossing during two-stent bifurcation procedures is illustrated by a case report published by Wurtz and colleagues.[2] They reported a patient who presented with stent thrombosis six years after undergoing Culotte bifurcation stenting. They showed, by OCT, that there had been inadvertent abluminal wiring during the original procedure. Instead of seeing a circular double layer of stent struts in the proximal main vessel, there was evidence that the side branch stent had been completely crushed in the proximal main vessel. This was not evident on angiography at the time of the index procedure. Importantly, this could have been avoidable if POT had been performed prior to re-wiring; however, this technique was not part of clinical practice at the time the original procedure was undertaken.

Figure 6: Improved strut apposition and restoration of circularity when POT was performed.

Figure 6:

Despite a good result on angiography (Ai), IVUS demonstrated that there was still significant mal-apposition at the proximal part of the LMS as well as elliptical deformation of the stented vessel (Aii). A final balloon inflation (POT) was performed with a 4.5 mm balloon (Bi). IVUS demonstrated good strut apposition as well as restoration of circularity (Bii).

Bioresorbable scaffolds

Bioresorbable scaffolds (BRS) have some expansion limitations beyond which there is a risk of strut fracture.[29] Therefore, when selecting the most appropriate size, it is important to appreciate the dilation limitations. The expansion capacity of the Absorb scaffold (Abbott Vascular, Santa Clara, California) allows post-dilation up to 0.5 mm above the nominal diameter. An ‘aggressive’ POT should be avoided for BRS so it may be necessary to choose a diameter that is a little larger than the distal main vessel reference diameter and closer to the proximal main vessel reference diameter. The scaffold is deployed at nominal pressure, and then optimised using both an appropriately-sized balloon for the distal part of the stent as well as a larger balloon for the POT. If the SB appears compromised, the recommended technique is POT, followed by balloon inflation in the side branch, followed by re-POT.[30] Bench studies suggest that for a 3.0 mm Absorb scaffold implanted in the main vessel, the side branch can be dilated with (up to) a 3.0 mm balloon inflated to a maximum of 10 atm.[31] Importantly, because of the risk of scaffold fracture with overstretch, KBD with overlapping balloons is strongly discouraged.[1]

Conclusion

The proximal optimisation technique is a key part of treating large bifurcation lesions and will optimise results of both single and two-stent strategies. Technique is important – an appropriately sized balloon should be positioned and inflated just up to the carina. When performed well, the enhanced lesion scaffolding, reduced strut mal-apposition and improved flow dynamics are likely to translate into improved clinical results.

References

  • 1.Lassen JF, Holm NR, Banning A et al. Percutaneous coronary intervention for coronary bifurcation disease: 11th consensus document from the European Bifurcation Club. EuroIntervention. 2016;12:38–46. doi: 10.4244/EIJV12I1A7. DOI: [DOI] [PubMed] [Google Scholar]
  • 2.Wurtz M, Christiansen EH, Kristensen SD, Holm NR. Accidentally crushed stent during complex bifurcation treatment. A potential cause of very late stent thrombosis. Int J Cardiol. 2015;197:113–5. doi: 10.1016/j.ijcard.2015.06.045. DOI: [DOI] [PubMed] [Google Scholar]
  • 3.Foin N, Secco GG, Ghilencea L et al. Final proximal post-dilatation is necessary after kissing balloon in bifurcation stenting. EuroIntervention. 2011;7:597–604. doi: 10.4244/EIJV7I5A96. DOI: [DOI] [PubMed] [Google Scholar]
  • 4.Foin N, Torii R, Mortier P et al. Kissing balloon or sequential dilation of the side branch and main vessel for provisional stenting of bifurcations: lessons from micro-computed tomography and computational simulations. JACC Cardiovasc Interv. 2012;5:47–56. doi: 10.1016/j.jcin.2011.08.019. DOI: [DOI] [PubMed] [Google Scholar]
  • 5.Murasato Y, Finet G, Foin N. Final kissing balloon inflation: the whole story. EuroIntervention. 2015;11:81–5. doi: 10.4244/EIJV11SVA18. Suppl V:V. DOI: [DOI] [PubMed] [Google Scholar]
  • 6.Finet G, Derimay F, Motreff P et al. Comparative Analysis of Sequential Proximal Optimizing Technique Versus Kissing Balloon Inflation Technique in Provisional Bifurcation Stenting: Fractal Coronary Bifurcation Bench Test. JACC Cardiovasc Interv. 2015;8:1308–17. doi: 10.1016/j.jcin.2015.05.016. DOI: [DOI] [PubMed] [Google Scholar]
  • 7.Guerin P, Pilet P, Finet G et al. Drug-eluting stents in bifurcations: bench study of strut deformation and coating lesions. Circ Cardiovasc Interv. 2010;3:120–6. doi: 10.1161/CIRCINTERVENTIONS.108.846089. DOI: [DOI] [PubMed] [Google Scholar]
  • 8.Basalus MW, Tandjung K, van Apeldoorn AA et al. Effect of oversized partial postdilatation on coatings of contemporary durable polymer-based drug-eluting stents: a scanning electron microscopy study. J Interv Cardiol. 2011;24:149–61. doi: 10.1111/j.1540-8183.2010.00617.x. DOI: [DOI] [PubMed] [Google Scholar]
  • 9.Basalus MW, van Houwelingen KG, Ankone MJ et al. Micro-computed tomographic assessment following extremely oversized partial postdilatation of drug-eluting stents. EuroIntervention. 2010;6:141–8. DOI: 10.4244/ [PubMed] [Google Scholar]
  • 10.Gastaldi D, Morlacchi S, Nichetti R et al. Modelling of the provisional side-branch stenting approach for the treatment of atherosclerotic coronary bifurcations: effects of stent positioning. Biomech Model Mechanobiol. 2010;9:551–61. doi: 10.1007/s10237-010-0196-8. DOI: [DOI] [PubMed] [Google Scholar]
  • 11.Otake H, Shite J, Ako J et al. Local determinants of thrombus formation following sirolimus-eluting stent implantation assessed by optical coherence tomography. JACC Cardiovasc Interv. 2009;2:459–66. doi: 10.1016/j.jcin.2009.03.003. DOI: [DOI] [PubMed] [Google Scholar]
  • 12.Ormiston J, Darremont O, Iwasaki K et al. Lessons from the real bench: non-BRS. EuroIntervention. 2015;11:27–30. doi: 10.4244/EIJV11SVA6. Suppl V:V. DOI: [DOI] [PubMed] [Google Scholar]
  • 13.Foin N, Mattesini A, Ghione M et al. Tools & techniques clinical: optimising stenting strategy in bifurcation lesions with insights from in vitro bifurcation models. EuroIntervention. 2013;9:885–7. doi: 10.4244/EIJV9I7A144. doi: [DOI] [PubMed] [Google Scholar]
  • 14.Foin N, Gutierrez-Chico JL, Nakatani S et al. Incomplete stent apposition causes high shear flow disturbances and delay in neointimal coverage as a function of strut to wall detachment distance: implications for the management of incomplete stent apposition. Circ Cardiovasc Interv. 2014;7:180–9. doi: 10.1161/CIRCINTERVENTIONS.113.000931. DOI: [DOI] [PubMed] [Google Scholar]
  • 15.Lee R, Foin N, Ng J et al. Early coverage of drug-eluting stents analysed by optical coherence tomography: evidence of the impact of stent apposition and strut characteristics on the neointimal healing process. EuroIntervention. 2016;12:e605–14. doi: 10.4244/EIJV12I5A100. DOI: [DOI] [PubMed] [Google Scholar]
  • 16.Giannoglou GD, Antoniadis AP, Koskinas KC, Chatzizisis YS. Flow and atherosclerosis in coronary bifurcations. EuroIntervention. 2010;6:16–23. doi: 10.4244/EIJV6SUPJA4. Suppl J:J. DOI: [DOI] [PubMed] [Google Scholar]
  • 17.Behan MW, Holm NR, de Belder AJ et al. Coronary bifurcation lesions treated with simple or complex stenting: 5-year survival from patient-level pooled analysis of the Nordic Bifurcation Study and the British Bifurcation Coronary Study. Eur Heart J. 2016;37:1923–8. doi: 10.1093/eurheartj/ehw170. DOI: [DOI] [PubMed] [Google Scholar]
  • 18.Korn HV, Yu J, Ohlow MA et al. Interventional therapy of bifurcation lesions: a TIMI flow-guided concept to treat side branches in bifurcation lesions--a prospective randomized clinical study (Thueringer bifurcation study, THUEBIS study as pilot trial). Circ Cardiovasc Interv. 2009;2:535–42. doi: 10.1161/CIRCINTERVENTIONS.108.833046. DOI: [DOI] [PubMed] [Google Scholar]
  • 19.Niemela M, Kervinen K, Erglis A et al. Randomized comparison of final kissing balloon dilatation versus no final kissing balloon dilatation in patients with coronary bifurcation lesions treated with main vessel stenting: the Nordic-Baltic Bifurcation Study III. Circulation. 2011;123:79–86. doi: 10.1161/CIRCULATIONAHA.110.966879. DOI: [DOI] [PubMed] [Google Scholar]
  • 20.Niemela MJ, Holms NR, Kervinen K et al. Randomized Comparison of Final Kissing Balloon Dilatation Versus No Final Kissing Balloon Dilatation in Patients With Coronary Bifurcation Lesions Treated With Main Vessel Stenting. Five Year Clinical Outcome in The Nordic-Baltic Bifurcation Study III. J Am Coll Cardiol. 2015;66(15, suppl):B11–B11. doi: 10.1161/CIRCULATIONAHA.110.966879. [DOI] [PubMed] [Google Scholar]
  • 21.Gwon HC, Hahn JY, Koo BK et al. Final kissing ballooning and long-term clinical outcomes in coronary bifurcation lesions treated with 1-stent technique: results from the COBIS registry. Heart. 2012;98:225–31. doi: 10.1136/heartjnl-2011-300322. DOI: [DOI] [PubMed] [Google Scholar]
  • 22.Yu CW, Yang JH, Song YB et al. Long-Term Clinical Outcomes of Final Kissing Ballooning in Coronary Bifurcation Lesions Treated With the 1-Stent Technique: Results From the COBIS II Registry (Korean Coronary Bifurcation Stenting Registry). JACC Cardiovasc Interv. 2015;8:1297–307. doi: 10.1016/j.jcin.2015.04.015. DOI: [DOI] [PubMed] [Google Scholar]
  • 23.Hoye A, Iakovou I, Ge L et al. Long-term outcomes after stenting of bifurcation lesions with the “crush” technique: predictors of an adverse outcome. J Am Coll Cardiol. 2006;47:1949–58. doi: 10.1016/j.jacc.2005.11.083. DOI: [DOI] [PubMed] [Google Scholar]
  • 24.Sgueglia GA, Chevalier B. Kissing balloon inflation in percutaneous coronary interventions. JACC Cardiovasc Interv. 2012;5:803–11. doi: 10.1016/j.jcin.2012.06.005. DOI: [DOI] [PubMed] [Google Scholar]
  • 25.Chen SL, Zhang JJ, Ye F et al. Study comparing the double kissing (DK) crush with classical crush for the treatment of coronary bifurcation lesions: the DKCRUSH-1 Bifurcation Study with drug-eluting stents. Eur J Clin Invest. 2008;38:361–71. doi: 10.1111/j.1365-2362.2008.01949.x. DOI: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Chen SL, Xu B, Han YL et al. Clinical Outcome After DK Crush Versus Culotte Stenting of Distal Left Main Bifurcation Lesions: The 3-Year Follow-Up Results of the DKCRUSH-III Study. JACC Cardiovasc Interv. 2015;8:1335–42. doi: 10.1016/j.jcin.2015.05.017. DOI: [DOI] [PubMed] [Google Scholar]
  • 27.Chen SL, Santoso T, Zhang JJ et al. Clinical Outcome of Double Kissing Crush Versus Provisional Stenting of Coronary Artery Bifurcation Lesions: The 5-Year Follow-Up Results From a Randomized and Multicenter DKCRUSH-II Study (Randomized Study on Double Kissing Crush Technique Versus Provisional Stenting Technique for Coronary Artery Bifurcation Lesions). Circ Cardiovasc Interv. 2017;10(2):pii–pii. doi: 10.1161/CIRCINTERVENTIONS.116.004497. DOI: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Rigatelli G, Dell’Avvocata F, Zuin M et al. Complex coronary bifurcation revascularization by means of very minimal crushing and ultrathin biodegradable polymer DES: Feasibility and 1-year outcomes of the “Nano-crush” technique. Cardiovasc Revasc Med. 2017;18:22–7. doi: 10.1016/j.carrev.2016.07.003. DOI: [DOI] [PubMed] [Google Scholar]
  • 29.Ormiston JA, Webber B, Ubod B et al. An independent bench comparison of two bioresorbable drug-eluting coronary scaffolds (Absorb and DESolve) with a durable metallic drug-eluting stent (ML8/Xpedition). EuroIntervention. 2015;11:60–7. doi: 10.4244/EIJY15M02_03. DOI: [DOI] [PubMed] [Google Scholar]
  • 30.Derimay F, Souteyrand G, Motreff P et al. Sequential Proximal Optimizing Technique in Provisional Bifurcation Stenting With Everolimus-Eluting Bioresorbable Vascular Scaffold:Fractal Coronary Bifurcation Bench for Comparative Test Between Absorb and XIENCE Xpedition. JACC Cardiovasc Interv. 2016;9:1397–406. doi: 10.1016/j.jcin.2016.04.021. DOI: [DOI] [PubMed] [Google Scholar]
  • 31.Ormiston JA, Webber B, Ubod B et al. Absorb everolimus-eluting bioresorbable scaffolds in coronary bifurcations: a bench study of deployment, side branch dilatation and post-dilatation strategies. EuroIntervention. 2015;10:1169–77. doi: 10.4244/EIJY14M05_08. DOI: [DOI] [PubMed] [Google Scholar]

Articles from Interventional Cardiology Review are provided here courtesy of Radcliffe Cardiology