Abstract
Purpose
To develop and evaluate a cardiac phase-resolved myocardial T1 mapping sequence.
Methods
The proposed method for TempOrally-resolved Parametric Assessment of Z-magnetization recovery (TOPAZ) is based on contiguous FLASH imaging readout after magnetization inversion from the pulsed steady-state. Thereby, segmented k-space data is acquired over multiple heart-beats, before reaching steady-state. This results in sampling of the inversion recovery curve for each heart-phase at multiple points separated by an R-R interval. Joint T1 and B1+ estimation is performed for reconstruction of cardiac phase-resolved T1 and B1+ maps. Sequence parameters are optimized using numerical simulations. Phantom and in-vivo imaging is performed to compare the proposed sequence to a spin-echo reference and SAPPHIRE T1 mapping sequence in terms of accuracy and precision.
Results
In phantom, TOPAZ T1 values with integrated B1+ correction are in good agreement with spin-echo T1 values (NRMSE=4.2%) and consistent across the cardiac cycle (coefficient-of-variation=1.4±0.78%), and different heart-rates (coefficient-of-variation=1.2±1.9%). In-vivo imaging shows no significant difference in TOPAZ T1 times between the cardiac phases (ANOVA:p=0.14, coefficient-of-variation=3.2±0.8%) but underestimation compared with SAPPHIRE (T1-time±precision: 1431±56 ms vs. 1569±65ms). In-vivo precision is comparable to SAPPHIRE T1 mapping until mid-diastole (p>0.07) but deteriorates in the later phases.
Conclusion
The proposed sequence allows cardiac phase-resolved T1 mapping with integrated B1+ assessment at a temporal resolution of 40ms.
Keywords: quantitative myocardial tissue characterization, T1 mapping, B1+ mapping, cine, cardiac imaging
Introduction
Cardiovascular MRI (CMR) has a major role in the diagnosis, staging and monitoring of numerous ischemic and non-ischemic cardiomyopathies (1–3). Recently, parameter mapping in the heart has expanded the CMR toolbox, offering the chance to detect diffuse pathologies and to perform quantitative diagnosis. Among quantitative imaging technologies myocardial T1 mapping has shown promising diagnostic and prognostic value in a wide range of diseases (4–7).
Early techniques for quantification of the longitudinal relaxation time (T1) in the heart have employed continuous imaging using equidistant FLASH excitations, following an initial inversion pulse, as originally proposed by Look and Locker (8). While such approaches did not provide a pixel-wise T1 map due to the absence of ECG triggering, they allowed for a regional estimation of myocardial T1 using a ROI analysis (9). For voxel-wise quantification, the Modified Look-Locker Inversion recovery sequence (MOLLI) was introduced, performing single-shot imaging, triggered to the end-diastolic quiescence in a Look-Locker type inversion-recovery experiment (10). This enabled spatially-resolved quantification of the T1 time as a parameter map (T1 mapping) and established widespread use of quantitative CMR. Other imaging sequences, based on inversion (11–13) or saturation (14,15) recovery, or a combination of both (16,17) have been subsequently introduced for myocardial T1 mapping, each offering a distinct profile of advantages and disadvantages (18–20).
Myocardial T1 maps are conventionally acquired at a single end-diastolic phase. Recently, quantitative imaging during systole has been introduced, promising reduced effects of partial voluming (21), as well as increased resilience to heart-rate variability (22). Also, imaging throughout the cardiac cycle using variable flip-angle steady-state acquisitions, has been explored for quantitative mapping in a preclinical mouse study (23). However, due to limitations of the B1+ correction in the sequence, quantitative T1 times were only derived at the end-diastolic phase.
Quantitative analysis of the T1 time in the myocardium conventionally entails manually delineating the myocardium from the surrounding blood pools. Hence, consistent contouring of the myocardium is paramount for the reproducibility of cardiac T1 measurements (19,24). However, sub-epicardial fat, fatty infiltrations, myocardial crypts, trabeculae and other structures have previously been shown to hamper the identification of the blood myocardial interface in single-phase cardiac images (25). Furthermore, wall motion abnormalities including asynchrony patterns caused by bundle branch blocks are challenging in diastolic imaging and are commonly observed in cardiac diseases. Superior temporal resolution or the ability to view the changes in these structures over the cardiac phase may facilitate differentiation.
In this study, we sought to develop a method that allows for temporally and spatially resolved mapping of the native myocardial T1 times. The proposed sequence for TempOrally-resolved Parametric Assesment of Z-magnetization recovery (TOPAZ) is based on a steady-state look-locker inversion recovery, using segmented spoiled gradient echo imaging of multiple cardiac phases. B1+ correction is performed in an integrated manner using the inversion-efficiency coefficient of a rectangular inversion pulse. Acquisition parameters are chosen to yield optimal T1 mapping precision, based on numerical simulations. Accuracy and consistency of the T1 time quantification across the cardiac phases are assessed using phantom imaging. In-vivo imaging in healthy subjects is performed in a single breath-hold acquisition, and compared to a conventional myocardial T1 mapping method.
Methods
Sequence
The sequence diagram of the proposed TOPAZ acquisition is depicted in Figure 1. Initially, FLASH imaging pulses are used to drive the magnetization to steady-state. An inversion pulse is then played following the detection of the subsequent R-wave of the ECG signal. Following the inversion, contiguous imaging throughout multiple heart-beats is performed using FLASH pulses in a segmented fashion, until steady-state is reached again. This way a k-space segment is acquired at each cardiac phase for several sample times on the inversion recovery curve, where the inversion times for a given cardiac phase are separated by the duration of the R-R interval. After multiple heart-beats the magnetization is driven to steady-state and the acquisition of the next k-space segments starts with another inversion pulse played at the next R-wave. To avoid deviation from the pulsed steady-state recovery curve in the presence of R-R interval variations, dummy pulses, with no corresponding signal readout, are played after the acquisition of a pre-determined number of cardiac phases until the detection of the subsequent R-wave.
Figure 1.
Sequence diagram of the proposed TOPAZ T1 mapping acquisition: The signal is first driven to steady-state using continuous FLASH excitations. Following, an initial magnetization inversion, FLASH pulses are continuously played to allow for the sampling of an apparent T1 relaxation curve. k-space readouts are segmented across several heart-phases for sufficient temporal resolution. Multiple inversion times are sampled on the inversion recovery curve until pulsed steady-state is reached and another magnetization inversion is performed, after 5 R-R intervals in the diagram, to sample the next k-space segment. This is repeated several times to fill the k-space for each cardiac phase at five different inversion-times.
Flip-angle and repetition time were chosen based on numerical simulations to optimize TOPAZ T1 mapping sensitivity. The number of segments was selected between 8 and 12, depending on the heart-rate to yield scan time between 17–23 s.
Joint Reconstruction of Temporally-Resolved T1 and B1+ Maps
The magnetization signal during contiguous FLASH pulses can be described by a three parameter inversion recovery model (26)
| [1] |
where describes an apparent longitudinal relaxation time, which is a function of the flip-angle (FA) and the repetition time (TR):
| [2] |
The magnetization inversion is performed from the steady-state signal using a rectangular hard pulse. Hence, for an inversion flip angle of αINV, the initial signal can be described as
| [3] |
leading to the magnetization
| [4] |
Thus parameter quantification can be performed using a three-parameter least square fit:
| [5] |
for j=1,…,C, where C is the number of cardiac phases, denotes the inversion time of the kth T1-weighted image of the jth cardiac phase, and the corresponding signal for a given pixel location.
As the B parameter describes the inversion efficiency, phase-resolved B1+ information can be estimated from the three-parameter fit. When a nominal inversion flip angle of is utilized, results in the temporally-resolved pixel-wise B1+ flip-angle scaling factor and can be used as a B1+ map. Accordingly, the actual excitation flip-angle can be derived from the three fit parameters as
| [6] |
Here αexc describes the nominal excitation flip-angle after compensation for slice-profile effects. Details of the slice-profile correction are provided in Appendix A and Supporting Information S1–S3. The T1 time can then be derived from the three fit parameters as
| [7] |
In this study, the three-parameter fitting in equation [5] was performed using a bounded parameter range for the inversion efficiency ([1.4 2.4]), to avoid detrimental effects of outliers in the B1+ maps of late heart-phases.
Numerical Simulations
The sensitivity of the parameter estimation is a function of the T1, the flip-angle and the repetition time in TOPAZ T1 mapping. Numerical Monte-Carlo simulations were performed to optimize the T1 time precision in dependence of the sequence parameters. The signal throughout the proposed sequence was simulated using Bloch-simulations. Rician-noise characteristics were achieved by adding complex noise with independently normal distributed real and imaginary part and using the absolute value for further analysis. Simulations were performed with a heart-rate of 60 beats-per-minute (bpm), no B1+ variations and a tissue T1 time of 1550 ms to resemble healthy myocardial tissue at 3T (19). Noise was simulated at an SNR of 30 relative to the steady-state magnetization for parameters TR = 4.5 and flip-angle = 4.4° (Ernst-angle). T1 time accuracy was obtained as the difference between the mean of the estimated T1 across 1000 noise patterns and the simulated ground-truth T1 time. T1 time precision was calculated as the standard deviation across the noise patterns.
To ensure sufficient recovery before the reinversion, the number of heart-beats between the inversion pulses in the simulated sequence, was adapted to allow for recovery to at least 95% of the steady-state signal, for each (TR, FA) pair and the given T1 value. This leads to different scan times for different (TR, FA) settings and restricts the applicable parameter range in order to avoid long breath-hold durations. The simulated parameter range spanned 1° to 20° degrees flip-angle in steps of 1°, and 3.5 to 15.0 ms TR in steps of 0.5 ms.
Phantom Experiments
All imaging was performed at a 3T Siemens Magnetom Prisma (Siemens Healthcare, Erlangen, Germany) system using a 30-channel receiver coil-array.
The proposed method was compared to inversion-recovery spin-echo T1 mapping (16) in a cylindrical phantom containing several compartments filled with gadolinium doped agarose-gel to achieve T1 and T2 times in the in-vivo range (T1/T2 = 390/70, 696/84, 1233/106, 1665/442, 2147/~2000ms (27)). Imaging was performed with a simulated ECG signal at 60 bpm. All T1 maps were acquired with 10 repetitions. TOPAZ T1 times were assessed for all cardiac phases. Accuracy was defined based on manually drawn ROIs in comparison to SAPPHIRE T1 times. Spatial-variability was assessed as a surrogate for precision by calculating the standard deviation of the T1 time within the ROI drawn in the homogenous phantom compartment. Consistency in the T1 time accuracy and precision are assessed as the coefficient of variation (CoV).
The dependence of TOPAZ T1 mapping on the heart-rate was studied by acquiring images in the same phantom with various simulated ECGs. The RR-intervals were varied between 600 ms (100 bpm) and 1200 ms (50 bpm). For each manually drawn ROI the T1 time and spatial variation within the ROI were averaged across all phases. Dependence on the RR-interval of the simulated ECG was assessed using CoV and correlation analysis.
Imaging sequence parameters for TOPAZ T1 mapping were as follows, in the present study: TR/TE/FA=5/2.5ms/3°, bandwidth=350Hz/Px, FOV=300×225mm2, resolution=1.9×1.9mm2, matrix-size=160×120, slice thickness=10mm, partial-Fourier=6/8, GRAPPA-factor=2, reference-lines=24 (in-plane), #phase-encode lines=55, time between inversion pules=5–6 RR-intervals, inversion-pulse: 0.5ms rectangular pulse (160° flip-angle, to ensure the acos is well defined, which allows for estimation of the B1+ amplitude using equation [6]), excitation-pulse: symmetrical sinc-pulse with bandwidth-time-product=2.0, temporal resolution=40–60ms, breath-hold duration=17–23s.
Reference T1 times were acquired using an inversion-recovery spin-echo sequence, with TR/TE/FA=10s/12ms/90°, bandwidth=300Hz/Px, FOV=240×240mm2, resolution=1.9×1.9mm2, slice thickness=10mm, inversion-times=50,100,200,…,6400ms and one image without inversion-pulse, scan time= 3 h 12 min.
In-vivo experiments
The imaging protocol was approved by the local institutional review board, and written informed consent was obtained from all participants prior to each examination for this HIPAA-compliant study. Cardiac phase-resolved T1 maps were obtained from 9 healthy subjects (4 male, 32 ± 15 y/o) using the proposed TOPAZ T1 sequence and compared to SAPPHIRE T1 mapping (14). TOPAZ was acquired with the same imaging parameters as described for the phantom experiment. SAPPHIRE T1 mapping was acquired for reference with FLASH imaging readout and the following parameters: TR/TE/FA=4.0/2.0ms/10°, bandwidth=505Hz/Px, FOV=320×320mm2, resolution=2.0×2.1mm2, matrix-size = 160×152, slice thickness=10mm, slice gap=10mm, partial-Fourier=6/8, GRAPPA-factor=2, reference-lines=40 (in-plane), #phase-encode lines=77, linear k-space ordering, number of images = 15, inversion-pulse: 2.56ms tan/tanh adiabatic full passage (28), saturation-pulse: 4-compartment WET (19).
All images were acquired in a single mid-ventricular slice. Myocardial T1 times were assessed throughout the entire cardiac cycle using manually drawn ROIs in the septal wall. In-vivo precision was defined as the standard-deviation within the septal ROI. Due to a different number of cardiac phases acquired with TOPAZ in volunteers with different heart-rates, the vectors of septal T1 times acquired with TOPAZ throughout the cardiac cycle had different lengths for each volunteer. To facilitate quantitative comparison among subjects, each vector of septal T1 times was interpolated to 20 elements. T1 times and T1 precision between the methods were statistically compared using paired Student’s T-Test and Wilcoxon signed rank tests, respectively. T1 times at different cardiac phases were compared for statistical significant differences using one-way ANOVA. Differences in the precision across the cardiac phases were studied using Kruskal-Wallis group analysis. All statistical tests were compared at a significance level of p < 0.05.
Analysis of the estimated B1+ variation across cardiac regions and cardiac phases was performed by manually drawing profiles across the heart from the apex through the mid-septum to the lateral wall in the B1+ maps, in all cardiac phases. To allow for inter-subject comparison B1+ values were interpolated to 20 virtual cardiac phases as described for T1 time above and mean and standard deviation across all subjects were calculated as a function of cardiac region and cardiac phase. B1+ maps were quantitatively analyzed with respect to the B1+ gradient along the line from the apex through the mid-septum to the lateral free wall. Linear trends were compared among the cardiac phases using Pearson’s correlation coefficient, and statistically compared using ANOVA following Fisher’s z-transformation. Furthermore, the slope of the spatial gradient was calculated and statistically compared among subjects.
Results
Numerical Simulations
Figure 2 2 depicts the precision, accuracy and duration of the proposed technique as functions of TR and flip-angle. High flip-angles and short TRs cause rapid recovery to the steady-state magnetization. This prevents accurate T1 estimation and hampers the estimation precision. On the other hand, low flip-angles, well below the Ernst angle, result in compromised T1 mapping precision due to insufficient imaging SNR. Long TRs increase the duration of the recovery and the steady-state signal, leading to improved precision, although at the cost of prolonged scan time. TR = 5 ms and FA = 3° were chosen as a trade-off between T1 mapping precision and scan time, in the remainder of the study. This setting requires 5 RR-intervals pulsed recovery before re-inversion of the magnetization at a heart rate of 60 bpm.
Figure 2.
Simulation results depicting phantom accuracy, precision and sequence duration as a function of repetition time (TR) and flip angle, as assessed using noisy Bloch-simulations. The white line indicates the Ernst-angle for each TR/flip-angle pair, yielding optimal signal strength in a conventional, continuous FLASH acquisition. High flip-angles and short TRs impair sensitivity to the T1 time and substantially compromise fit accuracy and precision. Long TRs and low flip-angles on the other hand result in long recovery periods leading to increased scan time. Based on the simulation results, TR = 5 ms, and flip-angle = 3° (white circle) were selected as a compromise resulting in 1559±151 ms (T1 time±precision, simulated T1: 1550 ms) and scan duration of 25 HBs.
Phantom Experiments
Figure 3 shows the results of the T1 time measurement in phantom. The effect of the proposed B1+ correction is depicted in Figure 3a, where severe underestimation of the T1 time compared to spin-echo measurements is observed without any correction for excitation flip angles (y = 0.77 x + 91 ms). Corrected T1 times are in good agreement with spin-echo measurements, showing only slight deviations (−1.1 − 7.5%, y = 1.04 x − 29 ms, NRMSE = 4.2%). Figure 3b shows the estimated T1 times and the precision as a function of the cardiac phase. TOPAZ T1 times show high consistency throughout the cardiac cycle (Coefficient of Variation: CoV = 1.4 ± 0.78%), although slight reduction in precision is observed in the later cardiac phases.
Figure 3.
a) Phantom TOPAZ T1 values averaged across phases compared to spin-echo T1 values show substantial underestimation without B1+ correction. Good agreement is obtained with the corrected method. b) Phantom accuracy and precision of the proposed TOPAZ T1 mapping method (represented by circles and intervals respectively) compared across all cardiac phases. The estimated T1 times are consistent and accurate across cardiac cycles and T1 values, although slight deviation in precision is observed in later cardiac phases for higher T1 values.
The evaluation of TOPAZ T1 times at different heart-rates is shown in Figure 4. Good consistency of the average T1 across a wide range of RR-intervals is observed, resulting in a CoV of 1.2 ± 1.9%. No correlation with the RR-interval is observed across the phantom T1 times (R = 0.0026). A slight trend of increased spatial variability is observed at longer RR intervals (CoV = 34.4 ± 23.3%, R = 0.54), where more cardiac phases are sampled at the trade-off against fewer time points on the inversion recovery curve. This trend is particularly pronounced for the shortest T1 time (CoV = 73.6%, R = 0.95), which leads to the fastest recovery to the pulsed steady-state, and requires a denser sampling of the beginning of the inversion-recovery curve.
Figure 4.

T1 time and spatial variability, averaged across all cardiac phases, acquired with TOPAZ T1 mapping at different simulated RR-intervals. Good consistency of the T1 time is obtained across a wide range of simulated heart-rates. A slight trend of increased spatial variability is observed at long RR intervals, especially for very short T1 times, due to insufficient sampling of the early part of the inversion recovery curve.
In-vivo experiments
Representative base-line images of an exemplary TOPAZ scan with 50ms temporal resolution are shown in Figure 5 along with the corresponding T1 maps (All six inversion times and 12 cardiac phases for this subject are shown in Supplementary Figure S4). Raw images show strong inversion-recovery contrast, with differential signal changes of the myocardium and the blood pool signal intensity. Zero-crossing of the myocardial signal happens around one RR interval after the inversion pulse.
Figure 5.
Representative baseline images showing four inversion times at six cardiac phases acquired in a healthy subject with the proposed TOPAZ T1 mapping sequence. All images are displayed with the same window-level. Visually high contrast is depicted between the myocardium and the blood pools in the earlier images before the zero crossing, which happens around the time of one RR interval after the inversion pulse. Longer inversion times depict higher overall signal and weaker contrast, displaying increased proton density weighting. The corresponding T1 maps show comparable T1 times across the cardiac cycle. Visually decreased homogeneity is observed in the later phases, which is in accordance with the decreased dynamic signal range. All six inversion times at 12 cardiac phases with 50 ms temporal resolution are provided in Supplementary Figure S4.
Figure 6 shows example TOPAZ T1 and B1+ maps of selected cardiac phases acquired in a healthy subject at a temporal resolution of 40 ms (all cardiac phases are shown in Supplementary Figure S5). The T1 maps show homogenous T1 quantification throughout the myocardium at all cardiac phases. Sharp delineation toward the blood pools suggests sufficient temporal resolution to minimize temporal blurring of the myocardial contours. B1+ maps appear noise resilient and largely T1 insensitive at the earlier phases. However, increased inhomogeneity and compromised B1+ precision is observed at the later cardiac phases. Additional examples of phase-resolved T1 maps are provided as quantitative movies in Supporting Information SV1 – SV3.
Figure 6.
In-vivo T1 and B1+ maps jointly acquired in a healthy subject with TOPAZ T1 mapping, at representative cardiac phases acquired with a temporal resolution of 40 ms. Visually high T1 map quality with homogeneous signal in the myocardium is observed. Good B1+ quality is observed at the earlier heart-phases, but is degraded towards the end of the cardiac cycle. T1 and B1+ maps of all cardiac phases for this subject are provided in Supplementary Figure S5.
Figure 7 depicts diastolic and systolic phases of the TOPAZ T1 maps, as well as the comparison SAPPHIRE T1 maps acquired as a single temporal snap-shot during diastole. As with phantom imaging, uncorrected TOPAZ T1 maps, show severe underestimation compared to SAPPHIRE T1 maps. This is mitigated using the proposed joint estimation of B1+ maps for correction of the apparent T1 time, yielding T1 maps that are comparable to SAPPHIRE maps in terms of visual quality.
Figure 7.
Representative diastolic T1 maps acquired with SAPPHIRE compared with a systolic and diastolic phase acquired with the proposed TOPAZ T1 mapping method. TOPAZ T1 maps are shown before and after B1+ correction along with the B1+ maps used in the correction. As in phantom imaging, the corrected TOPAZ T1 maps are consistent with conventional SAPPHIRE T1 maps. The B1+ maps obtained from the joint estimation do not show substantial spatial variation for these cardiac phases.
Average myocardial T1 times and T1 time precision was quantified throughout the cardiac cycle across all subjects as shown in Figure 8 along with T1 maps of a single subject at 5 representative cardiac phases. Additionally, the average and standard deviation of in-vivo B1+ corrected and uncorrected T1 times as well as precision and B1+ amplitude are listed for all cardiac phases in Table 1. At peak systole and peak diastole the T1 times acquired using the TOPAZ sequence were 1475±66ms and 1431±56ms. This means a 5.9% and 8.6% decrease compared with SAPPHIRE T1 mapping (1569±65ms, p < 0.01). Slightly elevated T1 time was found during end systole, but with no significance in the difference between the cardiac phases (ANOVA p = 0.14, CoV = 3.2 ± 0.8 %). In-vivo precision was comparable to SAPPHIRE T1 mapping at early heart-phases (p > 0.07) but showed a significant increase toward the end of the cardiac cycle (Kruskal-Wallis p < 10−3, compared to SAPPHIRE, p < 0.004).
Figure 8.
Myocardial T1 times and in-vivo precision of TOPAZ T1 mapping, shown at 20 interpolated cardiac phases. The top row displays the mean T1 time per subject averaged over all healthy volunteers. The middle row depicts the standard deviation within the septal ROI of a subject averaged across the cohort. Shadings indicate the standard deviation across the nine subject. In the bottom row 5 myocardial T1 maps that are representative of the trend throughout the cardiac cycle are presented. Consistent T1 times with no major variation are achieved across the cardiac cycle. Early phases show good in-vivo precision, comparable, to SAPPHIRE T1 mapping. A trend of increased in-vivo variability is observed in the later phases of the cardiac cycle.
Table 1.
Cardiac phase-resolved quantification of longitudinal relaxation time across nine healthy subjects. T1 time depicts the B1+ corrected T1 time assessed in manually drawn ROIs in the septum of the myocardium. Precision shows the spatial variation within the ROI. T1* time provides the relaxation time without B1+ correction. Relative B1+ gives the measured B1+ amplitude relative to the nominal B1+ amplitude. All datasets were up-sampled to 20 cardiac phases for quantitative analysis. All reported values are mean and standard deviations per up-sampled cardiac phases across healthy subjects.
| Cardiac Phase | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 | 15 | 16 | 17 | 18 | 19 | 20 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| T1 Time (ms) | 1391.6±43.2 | 1403.3±56.4 | 1403.6±68.9 | 1407.3±68.9 | 1411.4±67.8 | 1420.7±50.2 | 1427.4±39.5 | 1426.2±64.6 | 1440.1±66.3 | 1474.6±66.4 | 1477.2±68.5 | 1469.8±71.9 | 1472.5±69.1 | 1467.2±90.2 | 1444.3±87.4 | 1463.5±82.9 | 1454.4±69.1 | 1431.3±56.2 | 1430.4±59.3 | 1436.3±78.8 |
| Precision (ms) | 94.0±26.7 | 101.4±28.8 | 96.9±31.5 | 92.4±30.3 | 91.5±30.6 | 93.8±28. | 98.5±36.5 | 94.5±28.2 | 94.7±32.7 | 105.3±39.8 | 108.8±36.9 | 111.4±26.2 | 124.9±39. | 127.2±40.8 | 123.2±27.6 | 137.6±40.5 | 141.0±37.9 | 122.9±34.2 | 129.0±20.7 | 156.5±28.6 |
| T1* Time (ms) | 1212.0±39.5 | 1225.6±47.4 | 1227.1±56.8 | 1231.2±60.8 | 1236.0±59.0 | 1244.6±40.8 | 1249.3±31.2 | 1246.4±51.8 | 1256.6±51.3 | 1287.3±53.7 | 1289.6±56.2 | 1282.9±58.2 | 1284.4±58.5 | 1281.3±79.6 | 1262.0±74.0 | 1280.5±70.2 | 1273.9±59.5 | 1250.4±53.7 | 1247.7±56.2 | 1251.9±73.6 |
| Relative B1+ (%) | 87.4±5.6 | 86.0±4.2 | 85.6±3.8 | 85.3±4.3 | 84.9±4.1 | 84.4±3.1 | 84.6±3.2 | 85.2±3.2 | 85.3±2.9 | 84.2±2.9 | 84.2±2.9 | 84.5±2.7 | 84.8±3.1 | 84.5±3.8 | 85.0±3.5 | 84.1±3.4 | 84.1±3.4 | 85.7±4.1 | 86.5±4.5 | 86.7±6.3 |
Figure 9 depicts the results of the B1+ variation analysis as a function of cardiac regions and cardiac phases. The profiles exhibit dependence on cardiac regions, with a trend of increased B1+ amplitude from the apex to the lateral wall. They also show an increase in B1+ inhomogeneity in later cardiac phases, in accordance with the unfavorable conditioning for the joint estimation in these cardiac phases. Correlation-coefficient between B1+ amplitude and the spatial location as well as the spatial gradient slope of the B1+ amplitude are fairly consistent over the earlier phases but detriment towards the end of the cardiac cycle (ANOVA: Correlation coefficient: p<10−9, Slope<10−4).
Figure 9.
Quantitative analysis of the B1+ profiles obtained with the proposed TOPAZ sequence. a) The left panel depicts B1+ trends across the heart at five representative cardiac phases, as mean and standard-deviation across all healthy subjects. The right panel shows B1+ maps of a single healthy subject, including the spatial location of the B1+ profile. A trend of increased B1+ amplitude from apex to the lateral free wall is consistently observed across all cardiac phases and across all subjects. B1+ maps at later phases have a tendency of increased variability, especially in apical locations. b) Slope of spatial B1+ gradient from apex to lateral wall as average and standard-deviation over all subjects. A comparable slope is maintained over the beginning of the cardiac cycle. A decreased B1+ gradient is observed in the later phases. c) Correlation statistics of the B1+ amplitude to the spatial location. A strong linear trend indicated by high correlations is shown consistently in the earlier phases, with the trend mitigating towards the end of the cardiac cycle.
Discussion
In this study, we developed a sequence to acquire cardiac phase-resolved native T1 maps of the myocardium, with integrated B1+ quantification at a temporal resolution of up to 40 ms in a single breath-hold. The sequence parameters for the proposed TOPAZ T1 mapping sequence were optimized numerically. High accuracy and good consistency of the T1 times across the cardiac phases and at different heart-rates were achieved in phantom measurements, while in-vivo maps showed slight underestimation of the myocardial T1 times.
Native myocardial T1 mapping is an emerging tool, establishing itself in the CMR sequence portfolio for diagnosis, risk stratification and prognosis of numerous cardiomyopathies. However, large-scale clinical evaluation has commenced only recently, and limited exposure of clinicians to this technique likely hampers its diagnostic value. Signal originating from different structures, such as the myocardial muscle, the blood pools or extra-cardiac fat, exhibit severely distinct T1 characteristics and need to be carefully delineated in order to maximize the diagnostic quality of T1 mapping. T1 mapping during systole was proposed to increase the readily evaluable myocardial area and to facilitate manual contouring. Recently, blood-suppressed myocardial T1 mapping was also introduced, to eliminate contamination of the myocardial signal by the neighboring blood pool (29). Both approaches yield increased resilience to ROI placement and decreased observer variability. The proposed TOPAZ technique might further increase the diagnostic confidence in myocardial T1 maps, by presenting quantification at multiple phases throughout the heart-beat. This might ease the delineation of blood, sub-epicardial fat and other structures and altered T1 times observed in a single phase can be substantiated by cross-comparison throughout the cardiac cycle. Thus, the technique might be particularly useful in diseases, which form complicated focal/diffuse scar patterns such as HCM or amylodosis. A recent study has demonstrated, that in healthy volunteers, who usually have thin myocardial walls, quantitative tissue characterization is sensitive to slice orientation, myocardium thickness and wall motion (30) – a problem that cannot be easily alleviated by improved ROI placement. T1 mapping throughout the myocardial cycle, at various states of myocardial contraction, could improve robustness in this cohort, and therefore, minimize false positive outliers. Accordingly, patients exhibiting pathologically decreased myocardial wall-thickness, for example when suffering from dilated cardiomyopathy, might benefit from the proposed technique.
Conventional myocardial T1 mapping methods utilize single-shot imaging over a window of 150 – 250 ms to allow the acquisition of a single T1-weighted snap-shot of the heart per heart-beat. This temporal resolution is sufficient for imaging the left-ventricular myocardium during the end-diastolic quiescence in most patients and healthy volunteers. However, the achieved temporal resolution is often insufficient to quantify highly mobile extra myocardial structures, such as cardiac tumors, including myxomas, or thrombi. Also, the long acquisition windows of single-shot sequences have been shown to impair quantitative cardiac mapping, even in healthy volunteers, when residual motion is abundant in mid-diastole (30). Lastly, patients suffering from wall motion abnormalities, including bundle branch blocks, commonly exhibit low image quality in diastolic imaging. The proposed method enables imaging at a temporal resolution of up to 40 ms. Conventional cine imaging at this temporal resolution has been proven valuable in depicting even highly mobile structures (31). Hence, the proposed method might be used for improved characterization of subjects with residual diastolic motion or motion abnormalities. Furthermore, it might extend the applicability of T1 quantification to extra-myocardial structures, potentially easing non-invasive characterization or classification of cardiac masses, such as myxomas or other tumors. Further evaluation of the proposed technique in this patient cohort is warranted to assess diagnostic sensitivity.
High accuracy and minimal deviation from the reference method were shown for the proposed technique in phantom scans, over a wide range of T1 times. However, in-vivo quantification of the myocardial T1 times show underestimation of about 10% compared to SAPPHIRE T1 times, which are well in line with previous studies (19,32). Due to the use of a B1+ adjusted correction of the apparent T1* time in the present method, this level of accuracy still compares favorably to previously reported accuracy obtained with commonly used inversion-recovery based myocardial T1 mapping methods at 3T and saturation-recovery methods as reference (19,20).
Accurate T1 quantification in the proposed sequence requires imaging of the magnetization during the approach to the pulsed steady-state. Cardiac motion will cause displacement of the heart within the inhomogeneous B1+ field. This induces cyclic changes in the effective flip-angle and the apparent relaxation time. The accuracy of an estimation using a constant relaxation time, might therefore be impaired in-vivo. However, our data and previous cardiac B1+ papers (33,34) indicate the slope of the B1+ amplitude gradient being low. Given the current sequence parameters this leads to less than 0.1° flip-angle variation throughout the cardiac cycle, which can be expected to impair the estimation only marginally. More importantly, due to non-rigid cardiac motion, myocardial tissue that is not subjected to the previous excitation pulses, may move into the imaging plane and corrupt the signal. While we have used a 10mm slice thickness, further increase in slice thickness, as well as a flat slice profile, reduce sensitivity at the edges of the slice excitation to through-plane motion and may be employed at the cost of reduced spatial resolution. The effect of through-plane motion may be more pronounced in other views, such as 2-chamber of horizontal long-axis. Such effects were not explored in this introductory study, and warrants further investigation. Three dimensional imaging might be used to ensure excitation of the entire cardiac volume, mitigating effects of through-plane motion. However, the corresponding increase in acquisition time requires other means of respiratory motion compensation, such as navigator gating and slice-tracking as previously proposed for myocardial T1 mapping (13,35,36). Development of three-dimensional free-breathing extensions of our technique requires further investigation.
In the proposed sequence, the transmit field strength was quantified based on the inversion efficiency of a rectangular inversion pulse. In order to ensure a unique solution to the inverse problem from the efficiency to the inversion flip-angle, a flip-angle smaller than 180 degrees was chosen. However, T2* recovery, magnetization transfer effects and B0 inhomogneities (off-resonances), compromise inversion efficiency and might corrupt the accuracy of the B1+ maps (37). Furthermore, B1+ sensitization is only performed at one time point throughout the cardiac cycle, at the time of inversion. Hence, flow and motion effects will affect the B1+ profiles at later cardiac phases. Nonetheless, the proposed B1+ mapping shows high visual quality and similar trends compared to B1+ maps obtained with dedicated methods in previous studies (33,34,38), with a consistent increase of B1+ amplitude from the apex to the lateral wall.
Sequential-flipping (10) was used in this study for polarity restoration in the magnitude images. In phantom imaging, for the vial with T1 = 1233 ms and at a single cardiac phase (700 ms after R-wave), this leads to points near zero-crossing, eventually causing increased standard deviation in the T1 measurements. While this effect was not observed for myocardial T1 values in the in-vivo range, it can be further mitigated by the use of phase-sensitive reconstruction (39).
Increased variability is observed toward the end of the heart-beat in both B1+ and T1 estimates. This is likely explained by the increased minimal inversion times at these phases, where the earlier parts of the inversion recovery curve were not sampled. This hampers the differentiation between T1 and B1+ effects in the fit. To mitigate the effects of outliers in the B1+ map at later phases on the T1 time precision, bounded fitting was performed, with restricting the inversion efficiency to an expected range based on literature values (33). However, previously proposed methods for cardiac transmit field mapping, did only allow for the generation of a single B1+ map. This prevents the application of quantitative methods throughout the entire cardiac cycle. The proposed method, on the other hand, provides B1+ estimation at 40 ms temporal resolution in an integrated manner with T1 mapping, at the cost of increased B1+ variability in parts of the cardiac cycle.
Tissue characterization following injection of contrast agents has proven clinical value in form of LGE imaging (40) and quantification of extra-cellular volume (ECV) quantification (41). For TOPAZ the interrelation between quantification sensitivity, sequence parameters and T1 time might cause suboptimal performance in the post-contrast T1 range with the presented scheme might achieve. Faster T1 recovery requires fewer heart-beats to reach steady-state magnetization, and enables the acquisition of multiple inversion-recovery experiments during a single breath-hold. Furthermore, optimal precision is likely to be achieved at different flip-angle and TR values, and separate optimization for the post-contrast T1 regime is warranted. Phase-resolved ECV mapping can be enabled by the acquisition of phase-resolved native and post-contrast T1 maps. This additionally requires co-registration of the native and post-contrast maps, as previously proposed (42,43). High temporal resolution in the dynamic T1 maps can be employed to ensure adequate synchronization of the cardiac phases in the separate scans. Therefore, residual motion between matched cardiac phases in the two T1 mapping acquisitions can be expected to be largely due to variations in the end-expiratory breath-hold position.
Limitations
This study and the proposed method have several limitations. T1 mapping was performed of the native myocardium only and post-contrast application remains subject of future studies. Slice-thickness of 10 mm was used as trade-off between through-plane motion sensitivity, SNR and spatial resolution. This slice-thickness is employed in several recent studies (44–48). However, reduced slice-thickness is recommended for increased sharpness of the blood-myocardial interface (49).
The breath-hold duration of 17–23 seconds might not be accomplishable by some patients, although this is in line with some clinical CINE protocols. Alternative parameter choices allow the trade-off between scan-time, temporal and spatial resolution to be tailored to the actual target population in future studies. Furthermore, advanced under-sampling and acceleration techniques, such as compressed sensing with temporal regularization might be applied to drastically shorten the acquisition time and are currently being investigated (50). Furthermore, a rectangular FOV with a relatively low extent in the y-dimension was used in this study. For larger patients, an increase in FOV would cause either a coarser resolution or longer scan time. While this was not an issue in the subject cohort for this study, such trade-offs may be necessary in clinical settings, and also requires further exploration into higher acceleration approaches. Lastly, no pathologies were evaluated in this proof-of-concept study, and evaluation of the diagnostic benefit over single-phase T1 mapping is warranted in a separate study.
Conclusion
This study demonstrates the feasibility of cardiac phase-resolved T1 mapping at a temporal resolution of 40 ms. The proposed TOPAZ T1 mapping technique also allows for integrated acquisition of temporally resolved T1 and B1+ maps. High-quality in-vivo quantitative maps are obtained throughout the entire cardiac cycle. This bears promise to improve diagnostic confidence in myocardial T1 mapping, by easing the delineation of surrounding structures.
Supplementary Material
Cardiac phase-resolved myocardial T1 maps acquired in a healthy subject using TOPAZ T1 mapping. The dynamic T1 maps are depicted in a cine view in form of a video. Average T1 time and standard-deviation within a manually drawn septal ROI are indicated in the colorbar.
Cardiac phase-resolved myocardial T1 maps acquired in a healthy subject using TOPAZ T1 mapping. The dynamic T1 maps are depicted in a cine view in form of a video. Average T1 time and standard-deviation within a manually drawn septal ROI are indicated in the colorbar.
Cardiac phase-resolved myocardial T1 maps acquired in a healthy subject using TOPAZ T1 mapping. The dynamic T1 maps are depicted in a cine view in form of a video. Average T1 time and standard-deviation within a manually drawn septal ROI are indicated in the colorbar.
Supporting Figure S1: Illustration of the second order approximation to the inversion-recovery signal model, as described in the Appendix A Equation [A10]. a) In the top row the signal is plotted against the flip-angle for several inversion-times t. Highly non-linear trends can be observed with non-zero curvature even for small flip-angles. In the lower row the correlation coefficient between the signal and the flip-angle is listed. Consistently low values indicate poor-approximation of the signal function for small flip-angles. b) In the top row the same signal is plotted against the flip-angle, where the flip-angle axis is now scaled quadratically instead of linear. Good approximation of the signal curve in the small flip-angle regime can now be obtained by a linear function as indicated with the dashed lines. The lower row lists the correlation coefficients between the signal and the square of the flip-angle, confirming good approximation with the model proposed in Appendix A Equation [A10].
Supporting Figure S2: Numerical Bloch-simulation of a TOPAZ sequence were performed to study the accuracy of the proposed slice-profile correction as explained in the methods section, except no noise was added in this simulation. The simulated T1 deviation from the ground-truth with and without using the proposed slice-profile correction, at various flip-angles and slice-profiles (bandwidth-time product) are shown above. TR was fixed to 5 ms for this simulation and the simulated ground-truth T1 time was 1550 ms. The red circles indicates the actual acquisition parameters from the paper, resulting in a T1 deviation of 4.3 ms using the proposed correction.
Supporting Figure S3: Difference between slice profile correction using ground truth T1 time in the calculation of MSS(α) and using an iteratively estimated T1, with a varying number of iterations (see Appendix A). Simulations were performed with TR=5 ms, α=3°, bandwidth-time products (BWT) as listed in the legend. The simulated T1=1550 ms.
Supporting Figure S4: All baseline images as acquired in a healthy subject with the proposed TOPAZ T1 mapping sequence, providing five different inversion times and 12 cardiac phases at 50 ms temporal resolution. All images are displayed with the same window-level. Visually high contrast is depicted between the myocardium and the blood pools in the earlier images before the zero crossing, which happens around the time point of one RR interval after the inversion pulse. Longer inversion times depict higher overall signal and weaker contrast, displaying increased proton density weighting. The corresponding T1 maps show comparable T1 times across the cardiac cycle. Visually decreased homogeneity is observed in the later phases, which is in accordance with the decreased dynamic signal range.
Supporting Figure S5: In-vivo T1 and B1+ maps jointly acquired in a healthy subject with TOPAZ T1 mapping, at a temporal resolution of 40 ms. Visually high T1 map quality with homogeneous signal in the myocardium is observed. Good B1+ quality is observed at the earlier heart-phases, but is degraded towards the end of the cardiac cycle.
Acknowledgments
The authors thank Ehsan Sedaghat-Nejad for his valuable help with designing the Figures.
Funding:
NIH; Grant numbers: R00HL111410, P41EB015894; NSF; Grant number: CCF-1651825
Appendix A
The signal at time t with flip angle α is given by:
| [A1] |
where
| [A2] |
For notational brevity we use . We expand R(t, α) as a Taylor series with respect to α:
| [A3] |
| [A4] |
| [A5] |
for α=0, this yields
| [A6] |
| [A7] |
| [A8] |
| [A9] |
leading to
| [A10] |
This approximation is illustrated in Supporting Figure 1.
The signal along a non-constant slice profile, α(x) is obtained as
| [A11] |
Given the Taylor series expansion above we can approximate this as
| [A12] |
where the last line follows by the Taylor series approximation.
Given a three-parameter model fit, with :
| [A11] |
and assuming a global minimum exists as per convention, indicating the objective function is locally convex around it, we conclude
| [A13] |
Numerical simulations were performed to study the accuracy of this slice-profile correction. Details and results are provided in Supporting Figure S2.
Note that [A13] requires the knowledge of T1, to calculate MSS(α) as given in [A2]. To overcome this circularity, we iteratively calculate T1 using [A13], initialized with MSS(α), corresponding to the T1 obtained without slice-profile correction. Numerical simulations were performed to illustrate the convergence of this iterative process and are provided in Supporting Figure S3.
References
- 1.Kramer CM. Role of Cardiac MR Imaging in Cardiomyopathies. J Nucl Med. 2015;56(Suppl 4):39S–45S. doi: 10.2967/jnumed.114.142729. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.von Knobelsdorff-Brenkenhoff F, Schulz-Menger J. Cardiovascular magnetic resonance imaging in ischemic heart disease. J Magn Reson Imaging. 2012;36(1):20–38. doi: 10.1002/jmri.23580. [DOI] [PubMed] [Google Scholar]
- 3.van der Wall EE, Vliegen HW, de Roos A, Bruschke AV. Magnetic resonance imaging in coronary artery disease. Circulation. 1995;92(9):2723–2739. doi: 10.1161/01.cir.92.9.2723. [DOI] [PubMed] [Google Scholar]
- 4.Schelbert EB, Messroghli DR. State of the Art: Clinical Applications of Cardiac T1 Mapping. Radiology. 2016;278(3):658–676. doi: 10.1148/radiol.2016141802. [DOI] [PubMed] [Google Scholar]
- 5.Jellis CL, Kwon DH. Myocardial T1 mapping: modalities and clinical applications. Cardiovasc Diagn Ther. 2014;4(2):126–137. doi: 10.3978/j.issn.2223-3652.2013.09.03. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Ambale-Venkatesh B, Lima JA. Cardiac MRI: a central prognostic tool in myocardial fibrosis. Nat Rev Cardiol. 2015;12(1):18–29. doi: 10.1038/nrcardio.2014.159. [DOI] [PubMed] [Google Scholar]
- 7.Radenkovic D, Weingartner S, Ricketts L, Moon JC, Captur G. T1 mapping in cardiac MRI. Heart Fail Rev. 2017;22(4):415–430. doi: 10.1007/s10741-017-9627-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Look DC, Locker DR. Time Saving in Measurement of NMR and EPR Relaxation Times. Review of Scientific Instruments. 1970;41:250–251. [Google Scholar]
- 9.Nacif MS, Turkbey EB, Gai N, Nazarian S, van der Geest RJ, Noureldin RA, Sibley CT, Ugander M, Liu S, Arai AE, Lima JA, Bluemke DA. Myocardial T1 mapping with MRI: comparison of look-locker and MOLLI sequences. J Magn Reson Imaging. 2011;34(6):1367–1373. doi: 10.1002/jmri.22753. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52(1):141–146. doi: 10.1002/mrm.20110. [DOI] [PubMed] [Google Scholar]
- 11.Kvernby S, Warntjes MJ, Haraldsson H, Carlhall CJ, Engvall J, Ebbers T. Simultaneous three-dimensional myocardial T1 and T2 mapping in one breath hold with 3D-QALAS. J Cardiovasc Magn Reson. 2014;16:102. doi: 10.1186/s12968-014-0102-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Mehta BB, Chen X, Bilchick KC, Salerno M, Epstein FH. Accelerated and navigator-gated look-locker imaging for cardiac T1 estimation (ANGIE): Development and application to T1 mapping of the right ventricle. Magn Reson Med. 2015;73(1):150–160. doi: 10.1002/mrm.25100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Weingartner S, Roujol S, Akcakaya M, Basha TA, Nezafat R. Free-breathing multislice native myocardial T1 mapping using the slice-interleaved T1 (STONE) sequence. Magn Reson Med. 2014 doi: 10.1002/mrm.25387. [DOI] [PubMed] [Google Scholar]
- 14.Chow K, Flewitt JA, Green JD, Pagano JJ, Friedrich MG, Thompson RB. Saturation recovery single-shot acquisition (SASHA) for myocardial T(1) mapping. Magn Reson Med. 2014;71(6):2082–2095. doi: 10.1002/mrm.24878. [DOI] [PubMed] [Google Scholar]
- 15.Higgins DM, Ridgway JP, Radjenovic A, Sivananthan UM, Smith MA. T1 measurement using a short acquisition period for quantitative cardiac applications. Med Phys. 2005;32(6):1738–1746. doi: 10.1118/1.1921668. [DOI] [PubMed] [Google Scholar]
- 16.Weingartner S, Akcakaya M, Basha T, Kissinger KV, Goddu B, Berg S, Manning WJ, Nezafat R. Combined saturation/inversion recovery sequences for improved evaluation of scar and diffuse fibrosis in patients with arrhythmia or heart rate variability. Magn Reson Med. 2014;71(3):1024–1034. doi: 10.1002/mrm.24761. [DOI] [PubMed] [Google Scholar]
- 17.Weingartner S, Akcakaya M, Roujol S, Basha T, Tschabrunn C, Berg S, Anter E, Nezafat R. Free-breathing combined three-dimensional phase sensitive late gadolinium enhancement and T1 mapping for myocardial tissue characterization. Magn Reson Med. 2015;74(4):1032–1041. doi: 10.1002/mrm.25495. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Roujol S, Weingärtner S, Foppa M, Chow K, Kawaji K, Ngo LH, Kellman P, Manning WJ, Thompson RB, Nezafat R. Accuracy, precision, and reproducibility of four T1 mapping sequences: a head-to-head comparison of MOLLI, ShMOLLI, SASHA, and SAPPHIRE. Radiology. 2014;272(3):683–689. doi: 10.1148/radiol.14140296. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Weingartner S, Messner NM, Budjan J, Lossnitzer D, Mattler U, Papavassiliu T, Zollner FG, Schad LR. Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI. J Cardiovasc Magn Reson. 2016;18(1):84. doi: 10.1186/s12968-016-0302-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Teixeira T, Hafyane T, Stikov N, Akdeniz C, Greiser A, Friedrich MG. Comparison of different cardiovascular magnetic resonance sequences for native myocardial T1 mapping at 3T. J Cardiovasc Magn Reson. 2016;18(1):65. doi: 10.1186/s12968-016-0286-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Reiter U, Reiter G, Dorr K, Greiser A, Maderthaner R, Fuchsjager M. Normal diastolic and systolic myocardial T1 values at 1. 5-T MR imaging: correlations and blood normalization. Radiology. 2014;271(2):365–372. doi: 10.1148/radiol.13131225. [DOI] [PubMed] [Google Scholar]
- 22.Ferreira VM, Wijesurendra RS, Liu A, Greiser A, Casadei B, Robson MD, Neubauer S, Piechnik SK. Systolic ShMOLLI myocardial T1-mapping for improved robustness to partial-volume effects and applications in tachyarrhythmias. J Cardiovasc Magn Reson. 2015;17:77. doi: 10.1186/s12968-015-0182-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Coolen BF, Geelen T, Paulis LE, Nauerth A, Nicolay K, Strijkers GJ. Three-dimensional T1 mapping of the mouse heart using variable flip angle steady-state MR imaging. NMR Biomed. 2011;24(2):154–162. doi: 10.1002/nbm.1566. [DOI] [PubMed] [Google Scholar]
- 24.Kellman P, Hansen MS. T1-mapping in the heart: accuracy and precision. J Cardiovasc Magn Reson. 2014;16:2. doi: 10.1186/1532-429X-16-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Kellman P, Bandettini WP, Mancini C, Hammer-Hansen S, Hansen MS, Arai AE. Characterization of myocardial T1-mapping bias caused by intramyocardial fat in inversion recovery and saturation recovery techniques. J Cardiovasc Magn Reson. 2015;17:33. doi: 10.1186/s12968-015-0136-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Deichmann R, Haase A. Quantification of T1 Values by SNAPSHOT-FLASH NMR Imaging. Journal of Magnetic Resonance. 1992;96:608–612. [Google Scholar]
- 27.Lakkadi N, Rosen M, Bolan PJ. Design of a Phantom for Multiparametric Quantitative MR Imaging and Spectroscopy. Proc Intl Soc Mag Reson Med; Toronto, Canada. 2015. p. 4285. [Google Scholar]
- 28.Kellman P, Herzka DA, Hansen MS. Adiabatic inversion pulses for myocardial T1 mapping. Magn Reson Med. 2014;71(4):1428–1434. doi: 10.1002/mrm.24793. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Weingartner S, Messner NM, Zollner FG, Akcakaya M, Schad LR. Black-blood native T1 mapping: Blood signal suppression for reduced partial voluming in the myocardium. Magn Reson Med. 2016 doi: 10.1002/mrm.26378. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Wassmuth R, Prothmann M, Utz W, Dieringer M, von Knobelsdorff-Brenkenhoff F, Greiser A, Schulz-Menger J. Variability and homogeneity of cardiovascular magnetic resonance myocardial T2-mapping in volunteers compared to patients with edema. J Cardiovasc Magn Reson. 2013;15:27. doi: 10.1186/1532-429X-15-27. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Krishnamurthy R, Cheong B, Muthupillai R. Tools for cardiovascular magnetic resonance imaging. Cardiovasc Diagn Ther. 2014;4(2):104–125. doi: 10.3978/j.issn.2223-3652.2014.03.06. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Weingartner S, Moeller S, Schmitter S, Auerbach E, Kellman P, Shenoy C, Akcakaya M. Simultaneous multislice imaging for native myocardial T1 mapping: Improved spatial coverage in a single breath-hold. Magn Reson Med. 2017 doi: 10.1002/mrm.26770. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Weingärtner S, Zimmer F, Metzger GJ, Ugurbil K, Van de Moortele PF, Akcakaya M. Motion-robust cardiac B1+ mapping at 3T using interleaved bloch-siegert shifts. Magn Reson Med. 2016 doi: 10.1002/mrm.26395. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Cunningham CH, Pauly JM, Nayak KS. Saturated double-angle method for rapid B1+ mapping. Magn Reson Med. 2006;55(6):1326–1333. doi: 10.1002/mrm.20896. [DOI] [PubMed] [Google Scholar]
- 35.Weingartner S, Akcakaya M, Roujol S, Basha T, Stehning C, Kissinger KV, Goddu B, Berg S, Manning WJ, Nezafat R. Free-breathing post-contrast three-dimensional T1 mapping: Volumetric assessment of myocardial T1 values. Magn Reson Med. 2015;73(1):214–222. doi: 10.1002/mrm.25124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Chow K, Yang Y, Shaw P, Kramer CM, Salerno M. Robust free-breathing SASHA T1 mapping with high-contrast image registration. J Cardiovasc Magn Reson. 2016;18(1):47. doi: 10.1186/s12968-016-0267-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Schweser F, Deistung A, Reichenbach JR. Foundations of MRI phase imaging and processing for Quantitative Susceptibility Mapping (QSM) Z Med Phys. 2016;26(1):6–34. doi: 10.1016/j.zemedi.2015.10.002. [DOI] [PubMed] [Google Scholar]
- 38.Schar M, Vonken EJ, Stuber M. Simultaneous B(0)- and B(1)+-map acquisition for fast localized shim, frequency, and RF power determination in the heart at 3 T. Magn Reson Med. 2010;63(2):419–426. doi: 10.1002/mrm.22234. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Xue H, Greiser A, Zuehlsdorff S, Jolly MP, Guehring J, Arai AE, Kellman P. Phase-sensitive inversion recovery for myocardial T1 mapping with motion correction and parametric fitting. Magn Reson Med. 2013;69(5):1408–1420. doi: 10.1002/mrm.24385. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Kim RJ, Wu E, Rafael A, Chen EL, Parker MA, Simonetti O, Klocke FJ, Bonow RO, Judd RM. The use of contrast-enhanced magnetic resonance imaging to identify reversible myocardial dysfunction. New England Journal of Medicine. 2000;343(20):1445–1453. doi: 10.1056/NEJM200011163432003. [DOI] [PubMed] [Google Scholar]
- 41.Kellman P, Wilson JR, Xue H, Ugander M, Arai AE. Extracellular volume fraction mapping in the myocardium, part 1: evaluation of an automated method. J Cardiovasc Magn Reson. 2012;14:63. doi: 10.1186/1532-429X-14-63. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Xue H, Shah S, Greiser A, Guetter C, Littmann A, Jolly MP, Arai AE, Zuehlsdorff S, Guehring J, Kellman P. Motion correction for myocardial T1 mapping using image registration with synthetic image estimation. Magn Reson Med. 2012;67(6):1644–1655. doi: 10.1002/mrm.23153. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Roujol S, Foppa M, Weingartner S, Manning WJ, Nezafat R. Adaptive registration of varying contrast-weighted images for improved tissue characterization (ARCTIC): application to T1 mapping. Magn Reson Med. 2015;73(4):1469–1482. doi: 10.1002/mrm.25270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Swoboda PP, McDiarmid AK, Erhayiem B, Law GR, Garg P, Broadbent DA, Ripley DP, Musa TA, Dobson LE, Foley JR, Fent GJ, Page SP, Greenwood JP, Plein S. Effect of cellular and extracellular pathology assessed by T1 mapping on regional contractile function in hypertrophic cardiomyopathy. J Cardiovasc Magn Reson. 2017;19(1):16. doi: 10.1186/s12968-017-0334-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Nakamori S, Dohi K, Ishida M, Goto Y, Imanaka-Yoshida K, Omori T, Goto I, Kumagai N, Fujimoto N, Ichikawa Y, Kitagawa K, Yamada N, Sakuma H, Ito M. Native T1 Mapping and Extracellular Volume Mapping for the Assessment of Diffuse Myocardial Fibrosis in Dilated Cardiomyopathy. JACC Cardiovasc Imaging. 2017 doi: 10.1016/j.jcmg.2017.04.006. [DOI] [PubMed] [Google Scholar]
- 46.Reinstadler SJ, Stiermaier T, Liebetrau J, Fuernau G, Eitel C, de Waha S, Desch S, Reil JC, Poss J, Metzler B, Lucke C, Gutberlet M, Schuler G, Thiele H, Eitel I. Prognostic Significance of Remote Myocardium Alterations Assessed by Quantitative Noncontrast T1 Mapping in ST-Segment Elevation Myocardial Infarction. JACC Cardiovasc Imaging. 2017 doi: 10.1016/j.jcmg.2017.03.015. [DOI] [PubMed] [Google Scholar]
- 47.Vasanji Z, Sigal RJ, Eves ND, Isaac DL, Friedrich MG, Chow K, Thompson RB. Increased Left Ventricular Extracellular Volume and Enhanced Twist Function in Type 1 Diabetic Individuals. J Appl Physiol. 1985;2017 doi: 10.1152/japplphysiol.00012.2017. jap 00012 02017. [DOI] [PubMed] [Google Scholar]
- 48.Wu LM, An DL, Yao QY, Ou YZ, Lu Q, Jiang M, Xu JR. Hypertrophic cardiomyopathy and left ventricular hypertrophy in hypertensive heart disease with mildly reduced or preserved ejection fraction: insight from altered mechanics and native T1 mapping. Clin Radiol. 2017 doi: 10.1016/j.crad.2017.04.019. [DOI] [PubMed] [Google Scholar]
- 49.Cameron D, Vassiliou VS, Higgins DM, Gatehouse PD. Towards accurate and precise T 1 and extracellular volume mapping in the myocardium: a guide to current pitfalls and their solutions. MAGMA. 2017 doi: 10.1007/s10334-017-0631-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Steen M, Weingärtner S, Akcakaya M. Multi-Scale Locally Low-Rank Noise Reduction for High-Resolution Dynamic Quantitative Cardiac MRI. Annual International Conference of the IEEE Engineering in Medicine and Biology Society; JeJu Island, Korea. 2017. p. 837. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Cardiac phase-resolved myocardial T1 maps acquired in a healthy subject using TOPAZ T1 mapping. The dynamic T1 maps are depicted in a cine view in form of a video. Average T1 time and standard-deviation within a manually drawn septal ROI are indicated in the colorbar.
Cardiac phase-resolved myocardial T1 maps acquired in a healthy subject using TOPAZ T1 mapping. The dynamic T1 maps are depicted in a cine view in form of a video. Average T1 time and standard-deviation within a manually drawn septal ROI are indicated in the colorbar.
Cardiac phase-resolved myocardial T1 maps acquired in a healthy subject using TOPAZ T1 mapping. The dynamic T1 maps are depicted in a cine view in form of a video. Average T1 time and standard-deviation within a manually drawn septal ROI are indicated in the colorbar.
Supporting Figure S1: Illustration of the second order approximation to the inversion-recovery signal model, as described in the Appendix A Equation [A10]. a) In the top row the signal is plotted against the flip-angle for several inversion-times t. Highly non-linear trends can be observed with non-zero curvature even for small flip-angles. In the lower row the correlation coefficient between the signal and the flip-angle is listed. Consistently low values indicate poor-approximation of the signal function for small flip-angles. b) In the top row the same signal is plotted against the flip-angle, where the flip-angle axis is now scaled quadratically instead of linear. Good approximation of the signal curve in the small flip-angle regime can now be obtained by a linear function as indicated with the dashed lines. The lower row lists the correlation coefficients between the signal and the square of the flip-angle, confirming good approximation with the model proposed in Appendix A Equation [A10].
Supporting Figure S2: Numerical Bloch-simulation of a TOPAZ sequence were performed to study the accuracy of the proposed slice-profile correction as explained in the methods section, except no noise was added in this simulation. The simulated T1 deviation from the ground-truth with and without using the proposed slice-profile correction, at various flip-angles and slice-profiles (bandwidth-time product) are shown above. TR was fixed to 5 ms for this simulation and the simulated ground-truth T1 time was 1550 ms. The red circles indicates the actual acquisition parameters from the paper, resulting in a T1 deviation of 4.3 ms using the proposed correction.
Supporting Figure S3: Difference between slice profile correction using ground truth T1 time in the calculation of MSS(α) and using an iteratively estimated T1, with a varying number of iterations (see Appendix A). Simulations were performed with TR=5 ms, α=3°, bandwidth-time products (BWT) as listed in the legend. The simulated T1=1550 ms.
Supporting Figure S4: All baseline images as acquired in a healthy subject with the proposed TOPAZ T1 mapping sequence, providing five different inversion times and 12 cardiac phases at 50 ms temporal resolution. All images are displayed with the same window-level. Visually high contrast is depicted between the myocardium and the blood pools in the earlier images before the zero crossing, which happens around the time point of one RR interval after the inversion pulse. Longer inversion times depict higher overall signal and weaker contrast, displaying increased proton density weighting. The corresponding T1 maps show comparable T1 times across the cardiac cycle. Visually decreased homogeneity is observed in the later phases, which is in accordance with the decreased dynamic signal range.
Supporting Figure S5: In-vivo T1 and B1+ maps jointly acquired in a healthy subject with TOPAZ T1 mapping, at a temporal resolution of 40 ms. Visually high T1 map quality with homogeneous signal in the myocardium is observed. Good B1+ quality is observed at the earlier heart-phases, but is degraded towards the end of the cardiac cycle.








