The Advantages of Viscous Dissipation Rate over Simplified Power Loss as a Fontan Hemodynamic Metric

Zhenglun Alan Wei †,#, Michael Tree *, Phillip M Trusty , Wenjun Wu , Shelly Singh-Gryzbon , Ajit Yoganathan
PMCID: PMC5811362  NIHMSID: NIHMS917297  PMID: 29094292

Abstract

Flow efficiency through the Fontan connection is an important factor related to patient outcomes. It can be quantified using either a simplified power loss or a viscous dissipation rate metric. Though practically equivalent in simplified Fontan circulation models, these metrics are not identical. Investigation is needed to evaluate the advantages and disadvantages of these metrics for their use in in vivo or more physiologically-accurate Fontan modeling. Thus, simplified power loss and viscous dissipation rate are compared theoretically, computationally, and statistically in this study. Theoretical analysis was employed to assess the assumptions made for each metric and its clinical calculability. Computational simulations were then performed to obtain these two metrics. The results showed that apparent simplified power loss was always greater than the viscous dissipation rate for each patient. This discrepancy can be attributed to the assumptions derived in theoretical analysis. Their effects were also deliberately quantified in this study.

Furthermore, statistical analysis was conducted to assess the correlation between the two metrics. Viscous dissipation rate and its indexed quantity show significant, strong, linear correlation to simplified power loss and its indexed quantity (p < 0.001, r > 0.99) under certain assumptions. In conclusion, viscous dissipation rate was found to be more advantageous than simplified power loss as a hemodynamic metric because of its lack of limiting assumptions and calculability in the clinic. Moreover, in addition to providing a time-averaged bulk measurement like simplified power loss, viscous dissipation rate has spatial distribution contours and time-resolved values that may provide additional clinical insight. Finally, viscous dissipation rate could maintain the relationship between Fontan connection flow efficiency and patient outcomes found in previous studies. Consequently, future Fontan hemodynamic studies should calculate both simplified power loss and viscous dissipation rate to maintain ties to previous studies, but also provide the most accurate measure of flow efficiency. Additional attention should be paid to the assumptions required for each metric.

INTRODUCTION

Fontan connection power loss is a hemodynamic metric used to quantify fluid flow efficiency through a palliative surgery for single ventricle congenital heart disease patients. Fluid flow efficiency through the Fontan connection is important because the Fontan circulation functions on a single ventricle, leaving the pulmonary circulation without ventricular support. Lack of pulmonary ventricular support increases the importance of fluid flow efficiency from the systemic venous return through the Fontan connection, and into the single atrium. Decreased Fontan flow efficiency decreases single ventricle filling14 and patient exercise performance17,34, negatively impacting the entire circulation.

The importance of flow efficiency through the Fontan connection, and the studies conducted to understand it, has created different efficiency-related hemodynamic metrics. One of the first studies analyzing Fontan connection flow efficiency was by Low, et al20. This study introduced a total energy loss coefficient to quantify Fontan connection flow efficiency. The current power loss metric was first computed by Dubini, et al, though it was termed the “hydraulic dissipated power”8. Dubini’s power loss employed a simplified control volume energy analysis easily calculated by measuring only inlet and outlet volumetric flow rate, cross sectional area, and static pressure. This simplified power loss is the power loss metric that remains in the literature today, gaining recent popularity because of its relationship to patient outcomes13,17,34. The realization that flow efficiency losses through the Fontan connection are almost entirely due to mechanical energy converted to heat led to the introduction of viscous dissipation rate as a Fontan connection hemodynamic metric6,15,21,26.

Unfortunately, simplified power loss and viscous dissipation rate are not equivalent quantities11,21. Though practically equivalent for some Fontan hemodynamic studies that assume simplified physiological models5,9,18,34, we will show the inequality of the two metrics in this study.

The purpose of this study is to compare simplified power loss and viscous dissipation rate as hemodynamic metrics, and show the benefits of using viscous dissipation rate for future Fontan studies. Any hemodynamic metric fit for future, more physiologically-accurate Fontan studies must fulfill three requirements: (1) accurately describe Fontan physiological flow efficiency; (2) be easily calculable from clinical measurements; and (3) maintain the relationship between Fontan connection flow efficiency and patient outcomes from previous studies. These criteria were selected to focus the hemodynamic metric on clinical utilization.

This study hypothesizes that viscous dissipation rate fulfills all three requirements as a Fontan hemodynamic metric and is fit for future studies, and that viscous dissipation rate has advantages over simplified power loss as a hemodynamic efficiency metric. This hypothesis is tested using theoretical, computational, and statistical methods.

MATERIALS AND METHODS

Theoretical Methods

Theoretical methods were used to derive simplified power loss and viscous dissipation rate. Simplified power loss was derived using a control volume energy conservation analysis. Viscous dissipation rate was derived from a differential volume energy conservation analysis. The simplified power loss derivation identifies assumptions separating it from true power loss and viscous dissipation rate values (which are equal). The individual terms required to calculate viscous dissipation rate were considered for clinical calculability against known clinical measurement devices.

Patient Information

Computational fluid dynamics (CFD) and statistical analyses were conducted on a patient cohort (n=10) extracted from the Georgia Tech Cardiovascular Fluid Mechanics Fontan Database. The ten patients selected for this study were sampled to evenly span the range of total mean inflow (cardiac output) values found in the database. A large range of total inflows was desired because total inflow is associated with total energy at the Fontan connection inlets, and, therefore, different amounts of potential energy loss. The minimum and maximum total time-averaged volumetric inflow rate (Q) for the simulations was 2.18 L/min and 6.30 L/min, respectively. Figure 1 shows the Fontan connection anatomies used for the CFD analysis ordered by increasing total mean inflow rate.

Figure 1.

Figure 1

Ten patient-specific 3D anatomy models used for CFD analysis. The inferior vena cava (IVC) and superior vena cava (SVC) inflows are located at the bottom and top of each anatomy, respectively. The left pulmonary artery (LPA) outlet is toward the right side and the right pulmonary artery (RPA) and right upper pulmonary artery (RUPA) outlets are toward the left side of each model shown.

CFD Model Solution

Unsteady simulations were employed using ANSYS Fluent 17.0 (ANSYS, Inc., Canonsburg, PA) to acquire simplified power loss and viscous dissipation rate values. The simulation domain and boundary conditions were defined according to magnetic resonance imaging data, further detailed by Wei et al32. The Fourier transform in MATLAB (The MathWorks, Inc., Natick, MA) was used to convert time-varying clinical measured flow rates to the frequency domain. User-defined functions were utilized to apply reverse Fourier transform to reconstruct the time-varying flow rate at the inlets and time-varying flow ratio at the outlets. This procedure ensured better flow periodicity between cycles than directly applying clinical measurements at the boundaries. Figure 2 shows a good agreement (R2 > 0.999) between simulated flow and clinical measurements.

Figure 2.

Figure 2

The comparison of the flow rate and flow ratio between clinical measurement (markers) and the corresponding values from simulation (corresponding solid lines) for a representative case. Purple circular marker: IVC flow rate; Green rectangular marker: SVC flow rate; Orange triangular marker: flow ratio of LPA over RPA.

The coupled flow model was employed to solve the continuity and momentum equations simultaneously. Temporal and spatial discretizations employed second-order schemes; the momentum equation employed a third order scheme. The warped-face gradient correction was enabled to improve gradient calculation accuracy. Flow extensions equal to 10 times the vessel diameter were added to all inlets and outlets. Time step size was set at 0.001 seconds to ensure the Flow Courant number was ~ 0.1. Convergence criteria for continuity and momentum were set to 10−4. The vessel walls were assumed as rigid, and the fluid was assumed to be Newtonian with density equal to 1060 kg-m−3 and dynamic viscosity equal to 3.4×10−3 Pa-s.

Mesh Generation

The simulation was conducted on polyhedral meshes made in ANSYS Fluent 17.0 (ANSYS, Inc., Canonsburg, PA), converted from an unstructured tetrahedral mesh generated in ANSYS Workbench Meshing (ANSYS, Inc., Canonsburg, PA). A mesh independence study indicated that the adequate mesh has an edge size equal to Davg/50, where Davg is the average diameter of all TCPC inlets and outlets. Relevant results are discussed in Results section below. Moreover, boundary layer zones were created to improve numerical accuracy near the walls. A prismatic boundary mesh was generated by using the smooth transition approach with 10 layers and a geometric growth ratio of 1.05.

Hemodynamic Metrics

The CFD simulations resulted in time-resolved velocity vector and pressure scalar fields from which 4 metrics were calculated: simplified power loss (Ėsim), indexed simplified power loss7 (sim), viscous dissipation rate (Φ), and indexed viscous dissipation rate (). The indexed versions of the simplified power loss and viscous dissipation rate values were added, as indexed simplified power loss is correlated to Fontan patient exercise capacity17, and employed in multiple other studies23,24,27. Equations (1), (2), (3), and (4) define each of these metrics, respectively:

E˙sim=inlets(p+12ρ[QA]2)Qoutlets(p+12ρ[QA]2)Q (1)
iE˙sim=E˙sim·BSA2ρQ3 (2)
Φ=μV2[(uxx)2+(uyy)2+(uzz)2]+(uxy+uyx)2+(uyz+uzy)2+(uzx+uxz)2dV (3)
iΦ=Φ·BSA2ρQ3 (4)

where p is the fluid static pressure, ρ is the fluid density, A is the inlet or outlet cross-sectional area, BSA is the patient-specific body surface area, u is the directional velocity component (either x, y, or z direction), μ is the fluid dynamic viscosity, and V is the Fontan connection volume. The single value reported for statistical analysis was the time-averaged value of each metric.

Statistical Analysis

The statistical analyses comparing the simplified power loss and viscous dissipation rate metrics were performed using IBM SPSS Statistics (IBM, Inc., Aramark, NY). The metrics were first analyzed for normality using the Shapiro-Wilk test. Depending on normality results, bivariate correlations were analyzed using either Spearmen’s or Pearson’s correlation test. P < 0.05 was considered a statistically significant correlation.

RESULTS

Control Volume Energy Conservation

The simplified power loss term, Ėsim, comes from a control volume energy conservation analysis of the TCPC with several assumptions. True power loss, Ėloss, is derived first to highlight the assumptions made to arrive at Ėsim. The derivation that follows closely aligns with Munson, Young, and Okiishi22.

The first law of thermodynamics for a system states that the time rate of increase of the total stored energy of the system is equal to the sum of the net time rate of energy addition by heat transfer into the system and the net time rate of energy addition by work transfer into the system. This is represented symbolically in Equation (5):

tVρe dV+CSρeU·n dA=Q·in+W·in (5)

where e is the intensive total energy, U⃑ is the fluid velocity vector, n⃑ is a unit vector normal (outward) to the control surface (cs), in is the extensive heat rate added to the control volume, and in is the extensive work rate added to the control volume. The intensive total energy, e, is related to the intensive internal energy, û, the intensive kinetic energy, U2/2, and the intensive potential energy, gz, by Equation (6):

e=u^+gz+U22 (6)

The work term, in, is broken into several forms of work by Equation (7):

W˙in=W˙in,shaft+W˙in,press+W˙in,viscous (7)

where in,shaft is the net shaft work rate into the system (i.e. via turbine), in,press is the net pressure work into the system acting on all the control surfaces, and in,viscous is the net viscous work rate into the system due to stress tangential to the control surfaces. Both in,press and in,viscous definitions are provided in Equation (8) and Equation (9), respectively:

W˙in,press=CSp(U·n) dA (8)
W˙in,viscous=CSτ·U dA (9)

where p is the hydrostatic fluid pressure and τ⃑ is the shear stress vector on the control surface.

Substituting Equations (6), (7), (8), and (9) into (5) results in Equation (10):

tVρ(u^+gz+U22) dV+CS(u^+gz+U22)ρU·n dA=Q˙in+CSτ·U dACSp(U·n) dA+W˙in,shaft (10)

The true power (energy rate) lost (Ėloss) during the TCPC flow process is equal to the sum of the time rate of internal energy change and the heat rate out of the control volume. Isolating these terms on the left-hand side of the energy balance results in Equation (11):

E˙loss=W˙in,shaft+CSτ·U dACS(gz+U22+Pρ)ρU·n dAtVρ(gz+U22) dV (11)

Equation (11) is the most general form of power loss, and is applicable to all Newtonian fluids. To calculate simplified power loss, Ėsim, 5 assumptions are first applied to the general power loss equation (Equation (11)):

  1. No shaft work is present in the system:
    W˙in,shaft=0
  2. No viscous work is present on the system control surface:
    W˙τ=CSτ·U dA0
  3. The control volume does not change in time:
    W˙q=tVρ(gz+U22) dV0

    It is worth noticing that q also involves the effect of time-derivative of dynamic pressure, which will be discussed later. Nevertheless, the time-averaged effect of dynamic pressure should be negligible because of conservation of mass.

  4. There is no change in elevation between control surfaces:
    CS(gz+U22+Pρ)ρU·n dA=CS(U22+Pρ)ρU·n dA
  5. The normal velocity profile at the control surfaces is constant
    Un=U·nQA

Applying all of these assumptions to Equation (11) results in the simplified power loss most prevalent in the Fontan hemodynamic literature, as shown in Equation (1).

Differential Energy Conservation

The viscous dissipation rate metric comes from a differential energy conservation analysis. The first law of thermodynamics, restated for convenience, is that the time rate of increase of the total stored energy of the system is equal to the sum of the net time rate of energy addition by heat transfer into the system and the net time rate of energy addition by work transfer into the system. The differential work rate transferred into the system is broken into work done by pressure forces normal to the surface (Equation (12)), work done by shear forces tangential to the surface (Equation(13)) and body forces (Equation (14)).

dw˙p=·(pI¯·U) dV (12)
dw˙s=·(τ¯·U) dV (13)
dw˙b=ρf·U dV (14)

where p is the static pressure, is the identity tensor, U⃑ is the velocity vecotry, τ̲ is the symmetric shear stress tensor, and f⃑ is a body force (gravity) per unit mass.

The differential heat rate transferred into the system is broken into volumetric heating (Equation (15)) and thermal conduction heating (Equation (16)) components.

dQ˙v=ρq˙v dV (15)
dQ˙c=(·q˙c)dV (16)

where v is a term representing heat added to the entire volume per unit mass (i.e. due to radiation or a chemical reaction), and c is the conductive heat vector.

The differential total energy rate increase of the system is defined in Equation (17).

de˙=ρDDt(u^+U22) dV (17)

where ė is the total energy rate, û is the intensive internal energy, and U2/2 is the intensive kinetic energy. The potential energy contribution for this analysis is considered as part of the body forces term (Equation (14)).

Constructing Equations (12) through (14) to meet the first law of thermodynamics results in Equation (18):

ρDDt(u^+U22)=ρq˙v(·q˙c)·(pI¯·U)+·(τ¯·U)+ρf·U (18)

Similar to the control volume analysis, the sum of the change in internal energy and the heat rate leaving the system are equal to the power lost. These terms are isolated on the left-hand side of Equation (18) by subtracting the differential kinetic energy rate (Equation (19)) from both sides.

ρDDt(U22)=U·p+U·(·τ¯)+ρf·U (19)

Equation (20) results, and defines the differential power lost in the Fontan connection.

e˙loss=p(·U)+τ¯:U (20)

Assuming the fluid in question is Newtonian allows for a relationship between the shear stress vector and the velocity, which can then be substituted into Equation (20).

e˙loss=p(uxx+uyy+uzz)+μ[2(uxx)2+2(uyy)2+2(uzz)2+(uxy+uyx)2+(uxz+uzx)2+(uyz+uzy)2]23μ(uxx+uyy+uzz)2 (21)

Equation (21) is simplified by assuming incompressible flow, which leaves Equation (22):

VDR=μ[2(uxx)2+2(uyy)2+2(uzz)2+(uxy+uyx)2+(uxz+uzx)2+(uyz+uzy)2] (22)

A differential analysis is evaluated over a differential volume, so in order to truly equate viscous dissipation rate to the general power loss from the control volume analysis, Φ, (including equating their units of measure) the differential viscous dissipation rate must be integrated over the entire domain volume. The final relationship is found in Equation (23).

Φ=μV[2(uxx)2+2(uyy)2+2(uzz)2+(uxy+uyx)2+(uxz+uzx)2+(uyz+uzy)2] dV (23)

Mesh-independence Study

The mesh-independence study was conducted based on the case with the highest flow rate (model #10). The study started with a Davg/20 edge size, which was used in previous studies31. Every refinement reduced the edge size by 20%, resulting in ~50% reduction in mesh cell volume. Consequently, edge sizes were 1/20, 1/25, 1/31.25, 1/40, 1/50, and 1/62.5 of Davg. It was found that the Davg/20 mesh produced mesh-independent power loss as its resultant value has less than 2% discrepancy compared to the Davg/25 mesh. However, the Davg/50 mesh was needed for the calculated viscous dissipation rate to show less than a 5% discrepancy between the further refined mesh (Davg/62.5). Therefore, the Davg/50 mesh was chosen for all simulations, resulting in about 5 million polyhedral cells per mesh. The estimated numerical errors were 0.1% and 5% for power loss and viscous dissipation rate, respectively.

Viscous Dissipation Rate and Power Loss Comparison

The results of the CFD simulations were time-resolved three-dimensional velocity fields and pressure values for each of the 10 patient-specific models. Figure 3 shows viscous dissipation rate contours on a coronal slice and outlet velocity contours for two different models.

Figure 3.

Figure 3

Coronal slices of viscous dissipation rate (VDR) contours and colored vectors representing velocity values.

Similar results exist for all 10 patient-specific geometries. Each result shows increased viscous dissipation rate values near the wall and stagnation points, and non-uniform velocity profiles at the outlets. The time-averaged values of the viscous dissipation rate and power loss terms from the 10 patient-specific CFD simulations are summarized in Table 1. Table 1 also quantifies the effect of three out of five assumptions made by the simplified power loss and viscous dissipation rate in comparison to the true power loss metric. The effect of the remaining two assumptions will be elaborated in the discussion section.

Table 1.

Time-averaged viscous dissipation rate, power loss, and analyses of assumptions 2,3, and 5 (see appendix for more details).

Ėsim (mW)*1 ĖsimU⃑n (mW)*2 τ(μW)*3 q(μW)*4 Ėloss(mW)*5 Φ(mW)*6 ΔĖsimǂ1 Δ(ĖsimU⃑n)ǂ2 Δτǂ3 Δqǂ4 ΔΦǂ5
AVG§ 3.9 3.3 2.8 4.98 3.33 3.22 −17.70% −17.70% 0.1% 0.1% 2.7%
SD§§ 3.2 2.9 1.45 7.68 2.92 2.79 9.00% 8.80% 0.1% 0.3% 1.3%
§AVG average value
§§SD standard deviation
*1Ėsim simplified power loss
*2ĖsimU⃑n power loss with correctly calculated velocity profile (U⃑n)
*3τ = ∫CS τ⃑ · U⃑ dA viscous work present on the control surface
*4
W˙q=tVρ(U22) dV
effect of time-derivatives of dynamic pressure and volume
*5Ėloss = (ĖsimU⃑n) + τq true power loss
*6Φ viscous dissipation rate
ǂ1ΔĖsim = (ĖlossĖsim)/Ėloss discrepancy between simplified and true power losses
ǂ2Δ(ĖsimU⃑n) = 1 ∸ Esim/(ĖsimU⃑n) percentage discrepancy between power loss obtained using an accurate spatial velocity profile (assumption #5) and true power loss.
ǂ3 ΔẆτ = τ/Ėloss effect of viscous work on the control surface, assumption #2
ǂ4ΔẆq = q/Ėloss effect of control volume and dynamic pressure, assumption #3
ǂ5ΔΦ = (ĖlossΦ)/Ėloss discrepancy between viscous dissipation rate and true power loss

Figure 4 depicts the time-resolved difference between viscous dissipation rate and different forms of power loss as shown in Table 1. The instantaneous Ėloss looks identical to instantaneous viscous dissipation rate, and the time-resolved effect of viscous work on the control surface (assumption 2) is minimal. Additionally, the time-resolved discrepancy between Ėsim and ĖsimU⃑n are on the same order of the time-averaged discrepancy between them. However, the instantaneous difference between ĖsimU⃑n and Ėloss are remarkable, while the time-averaged difference between them are very small (combined effect of Δτ and Δq, less than 0.1 ± 0.3% as shown in Table 1).

Figure 4.

Figure 4

Time-resolved differences between different forms of power loss and viscous dissipation rate over a cardiac cycle for model #8. Ėsim: simplified power loss; ĖsimU⃑n: power loss with correctly calculated velocity profile (U⃑n); (ĖsimU⃑n) + τ: true power loss without the effect of time-derivatives of dynamic pressure and control volume; Ėloss: true power loss; Φ: viscous dissipation rate.

Both the comparison between Φ and Ėsim, and the comparison between and sim showed statistically significant relationships. Bland-Altman plots2 are illustrated in Figure 5. The 95% limits of agreement are indicated by the 2SD lines. Ėsim and sim are generally higher than Φ and , respectively, as indicated by the positive average differences,

Figure 5.

Figure 5

Bland-Altman plots for (a) simplified power loss and viscous dissipation rate values and (b) indexed simplified power loss and indexed viscous dissipation rate values resulting from the CFD simulation. Ėsim: simplified power loss; Φ: viscous dissipation rate;. loss: indexed simplified power loss; : indexed viscous dissipation rate.

Moreover, neither Ėsim, nor Φ, were found to be normal distributions, according to the Shapiro-Wilk test, so a Spearman’s correlation test was performed to compare these two metrics. The Spearman’s correlation test found that Ėsim is significantly (R2 = 0.992, p < 0.001) correlated with Φ, as shown in Figure 6. Both sim, and , were normally distributed according to a Shapiro-Wilk test, so a Pearson’s correlation test was performed to determine their relationship. Similar to the non-indexed terms, a very strong (r = 0.991) statistically significant (p < 0.001) correlation was found between sim and , as shown in Figure 6.

Figure 6.

Figure 6

Scatter plot comparing (a) simplified power loss and viscous dissipation rate values and (b) indexed simplified power loss and indexed viscous dissipation rate values resulting from the CFD simulation. The purple-dotted line stands for the linear curve-fitting line. Ėsim: simplified power loss; Φ: viscous dissipation rate;. sim: indexed simplified power loss; : indexed viscous dissipation rate.

DISCUSSION

This study shows advantages of viscous dissipation rate over simplified power loss as a Fontan connection hemodynamic metric. The focus, in particular, is on future Fontan in vitro and in silico studies with increased physiological modelling accuracy, or in vivo studies directly measuring true physiology.

The derivation of simplified power loss from true power loss via a control volume analysis highlighted the use of five assumptions. Of these five assumptions, two may be valid for physiologically accurate studies. Assuming no shaft work is present in the system (assumption 1) is always valid, unless the control volume included a mechanical energy source (i.e. a ventricular assist device). Assuming no change in elevation between the control surfaces (assumption 4) is also valid because most clinical measurements are made while the Fontan patient is in the supine position. The three remaining assumptions may not be completely valid in physiological settings.

First, involving properly measured velocity profiles (assumption 5) is important for all previous and future in vitro, in silico, and in vivo studies. Figure 2 showed two examples that can easily refute the validity of assumption 5. Table 1 quantitatively demonstrated that the effect of a constant velocity profile at the control surfaces, Δ(ĖsimU⃑n) = −17.7±8.8%, is the major contribution to the discrepancy between the simplified power loss and the true power loss, ΔĖsim= −17.7±9.0%. Highlighting the lack of proper velocity treatment for the simplified power loss calculation is not new. Grigioni et al. noted it as a critical issue when studying flow through the Fontan connection11. However, this study did not identify all of the other assumptions made to arrive at a simplified power loss calculation, neither did they systematically quantify the effect of these assumptions.

Second, accounting for control volume changes in time (assumption 3) is important for in vivo studies, and is becoming important for in vitro and in silico studies as researchers add flexible Fontan models to experiments25,30 and fluid-structure interaction models to computations19. Even with an unchanged control volume, the comparison between (ĖsimU⃑n) + τ and Ėloss in Figure 4 illustrates that q (assumption 3) plays a very important role to time-resolved discrepancies between the simplified power loss and viscous dissipation rate. Assumption 3 involves the time-derivative of the velocity field. The integration of the time-derivative over a cardiac cycle is negligible according to Table 1Ėτ = 0.1±0.1%), but its instantaneous values could be very large as the flow in the Fontan connection may be very unsteady.

Third, the presence of viscous work on control surfaces (assumption 2) is dependent on control volume selection and velocity measurement location. In addition, the velocity profile at the outlet control surfaces is rarely known prior to any (in vitro, in silico, or in vivo) Fontan hemodynamic study. Unless the inflow and outflow control surfaces are everywhere orthogonal to the velocity field, viscous work on the control surfaces may exist but could be neglected, ΔĖq = 0.1±0.3% as demonstrated in Table 1.

The results of the differential energy conservation derivation showed the assumptions made to arrive at the viscous dissipation rate metric. The viscous dissipation rate metric assumes that the fluid is Newtonian and that flow is incompressible. Incompressibility is commonly held for blood for cardiovascular systems. The Newtonian assumption is valid as long as flow within the Fontan connection exceeds a shear rate of 50/s3,10. Even with a non-Newtonian effect, one can easily use non-Newtonian models to adjust Equation (23), e.g. Cheng et al.4.

Utilizing all of the results, one can compare simplified power loss and viscous dissipation rate according to the three requirements outlined for a future Fontan hemodynamic efficiency metric. The first requirement states that the metric must describe Fontan flow efficiency accurately while maintaining physical meaning. The derivation of both simplified power loss and viscous dissipation shows that simplified power loss assumes more potentially invalid conditions. Additionally, simplified power loss is only capable of representing time-averaged hemodynamics in the Fontan connection while the viscous dissipation rate is useful for both the time-resolved and time-averaged analyses. The time-resolved values may assist studies investigating Fontan hemodynamics under free-breathing or exercise conditions because of the significant hemodynamic variation between cardiac cycles under these conditions33. The second requirement states that the metric should be easily calculable from clinical measurement devices. Simplified power loss, again, is the less ideal solution, as it requires invasive catheterization measurements to obtain static pressure data. Viscous dissipation rate requires a time-resolved three dimensional velocity field – data achievable via 4D MRI technology12. It is also worth noticing that power loss is only a bulk measurement, while the viscous dissipation rate also provides distribution contours, like in Figure 2. The distribution contours help visualize energetic differences between patients and identify regions of high viscous dissipation rate. This information can be used for making better clinical strategies. The final requirement is that the Fontan hemodynamic efficiency metric of the future maintain the relationship between Fontan connection flow efficiency and patient outcomes from previous studies. Simplified power loss was the metric used for studies relating to patient outcomes, so it automatically fulfills this final requirement14,17,34. In addition, the Bland-Altman plots in Figure 5 show the viscous dissipation rate is statistically equivalent to the simplified power loss. Strong correlations were also observed between viscous dissipation rate and simplified power loss for the CFD simulation results. The CFD methods employed are similar to those used in previous studies relating Fontan flow efficiency to patient outcomes, so this study concludes that viscous dissipation rate fulfills the final requirement as well. Ultimately, viscous dissipation rate fulfills all three requirements while simplified power loss only completely fulfills one. This shows the advantages of viscous dissipation rate over simplified power loss as a Fontan hemodynamic metric.

Nonetheless, a much higher resolution velocity field is required to obtain viscous dissipation rate in comparison to power loss. Therefore, the ability of the viscous dissipation rate metric to be entirely calculable in the clinic is dependent on 4D MRI technology. Work by Cibis et al explored the effects of 4D MRI spatial resolution on viscous dissipation rate calculations. Cibis’ study concluded that a CFD-based (high resolution) viscous dissipation rate was significantly greater than those based on a down-sampled CFD mesh, a down-sampled CFD mesh with simulated random noise, and a 4D MRI flow measurement. The study by Cibis showed lower viscous dissipation rate values for coarser resolution data fields across all patients. So, even though 4D MRI has a lower viscous dissipation rate value due to inadequate spatial resolution, comparisons between patients should still be valid. Nevertheless, advanced post-processing techniques are in development to improve the accuracy of MRI-derived hemodynamic metrics by merging CFD and MRI1. However, the study by Cibis et al. calculated viscous dissipation rate values on grids as fine as 0.1 mm, but never showed the minimum grid size at which changes in viscous dissipation rate no longer occurred. Determining this sort of resolution requirement to achieve accurate viscous dissipation rate measurements should be the subject of future research. Additionally, the noise in 4D MRI may be amplified by the calculation of velocity derivatives needed to obtain the viscous dissipation rate6,29. Future post-processing techniques shall also minimize the effect of the 4D MRI noise or quantify the budget of uncertainty for the impact of the noise on hemodynamic metrics, especially those derived from velocity derivatives. After all, 4D MRI technology must be thoroughly validated before clinical use for in vivo evaluation.

Closer examination of the CFD simulation methods show why the statistical correlation between the simplified power loss and viscous dissipation rate terms was so strong. The CFD simulations were conducted for patient-specific, rigid-walled Fontan connections under pulsatile flow conditions. This pre-established assumption drew the simplified power loss values closer to the viscous dissipation rate values, and undoubtedly played a role in the correlation strength and statistical significance. For example, with the vessel wall deformation, assumption 3 may not only influence the time-resolved discrepancy between simplified power loss and viscous dissipation rate but also affect the time-averaged one. However, examining the statistical relationship under the rigid wall assumption is permissible for the current study because similar pre-established assumptions were widely employed in previous studies that related simplified power loss to Fontan outcomes17,28,34. The strength and significance of the relationship between simplified power loss and viscous dissipation rate are expected to decrease under more physiologically accurate conditions, i.e. the introduction of flexible vessel wall. Hence, viscous dissipation rate would be better than simplified power loss at representing true physical meaning under such conditions.

Moreover, the CFD simulations did not include any additional forms of work within the control volume (in,shaft), nor did they consider gravity effect. It is hypothesized that the presence of gravity would affect both simplified power loss and viscous dissipation rate in the same way. Therefore, even though gravity effect will affect the absolute values, any change in the relationship between these two metrics could be negligible.

Additionally, the CFD simulations used the Newtonian fluid, thereby ignoring the non-Newtonian effect. A non-Newtonian important factor (IL = μeff, where μeff is the effective viscosity) was introduced to determine the non-Newtonian effect, and the cut-off value was suggested to be 1.7516. In order to estimate μeff, this study plugged the shear rate from the simulation to a Carreau model which curve-fitted experimental data from Cheng et al4. One case, in this study, exhibited an IL slightly above the cut-off line, i.e. 1.78. For this case, the differences between the Newtonian and non-Newtonian results were considered negligible for two reasons. Firstly, the non-Newtonian simulation indicates that both simplified power loss and viscous dissipation rate were increased by 4% compared to the Newtonian simulation. This marginally affected the difference between simplified power loss and viscous dissipation rate (18.3% with Newtonian properties vs. 19.0% with non-Newtonian properties). Secondly, including the non-Newtonian value for this patient produced the same statistical results in Bland-Altman plots and bivariate correlations. Therefore, to maintain consistency across patients for the sake of this study, the Newtonian results were used for this patient. This decision had no effect on our results.

Furthermore, energy equation was not involved in simulations of all patients. The effect of including energy equation was examined for the case with the maximum volumetric inflow rate. The simulation with the conservation of energy (converged to 10−7) resulted in less than 0.2% discrepancy in comparison with the simulation not including the energy equation. Therefore, it is reasonable to exclude the computation of conservation of energy for saving computational time because the impact is negligible.

Finally, the selection of 10 patient-specific cases that cover the full range of possible cardiac outputs is not the only way the cohort could have been determined. Similar CFD Fontan studies have selected cohorts according to Fontan connection type or Fontan connection anatomy characteristics while maintaining similar flow rates. However, it is worth noting that Fontan type and anatomy could affect power loss and viscous dissipation values differently. For example, the geometry would have impact on the normal velocity profile involved in assumption 5. Therefore, different type of Fontan anatomy could change the value of the discrepancy between simplified power loss and viscous dissipation rate but may not change the fact that assumption 5 dominates this discrepancy.

Overall, this study highlighted advantages of viscous dissipation rate over simplified power loss as a Fontan hemodynamic metric. Future studies of Fontan connection hemodynamics should include both simplified power loss and viscous dissipation rate metrics, with special attention paid to viscous dissipation rate’s relationship to patient outcomes for more complex and physiologically accurate studies.

Acknowledgments

This study was supported by the National Heart, Lung, and Blood Institute Grants HL67622 and HL098252. The authors acknowledge the use of ANSYS software which was provided through an Academic Partnership between ANSYS, Inc. and the Cardiovascular Fluid Mechanics Lab at the Georgia Institute of Technology. Additionally, the authors would like to acknowledge Luyu Zhang from Emory University for assistance in statistical analysis.

APPENDIX

Table II.

Details for Table 1: Time-averaged viscous dissipation rate, power loss, and analyses of assumptions 2,3, and 5

Model 1 2 3 4 5 6 7 8 9 10 AVG§ SD§§
Ėsim (mW)*1 0.56 1.87 2.68 3.88 3.69 5.97 4.16 3.56 0.87 11.7 3.89 3.19
ĖsimU⃑n(mW)*2 0.52 1.67 2.53 3.28 3.05 4.65 3.24 2.84 0.71 10.9 3.34 2.93
τ(μW)*3 1.47 3.34 2.11 1.67 1.33 5.95 4.07 3.04 2.08 3.7 2.88 1.45
q(μW)*4 −1.58 2.36 −1.09 2.29 1.72 10.03 1.16 22.9 0.39 11.6 4.98 7.68
Ėloss(mW)*5 0.52 1.67 2.53 3.28 3.05 4.65 3.24 2.82 0.71 10.9 3.33 2.92
Φ(mW)*6 0.52 1.64 2.42 3.17 2.93 4.56 3.14 2.76 0.7 10.4 3.22 2.79

ΔĖsim ǂ1 −7% −12% −6% −18% −21% −28% −28% −26% −22% −8% −17.7% 9.0%
Δ(ĖsimU⃑n) ǂ2 −8% −12% −6% −19% −21% −28% −28% −25% −23% −8% −17.7% 8.8%
Δτ ǂ3 0.3% 0.2% 0.1% 0.1% 0.0% 0.1% 0.1% 0.1% 0.3% 0.0% 0.1% 0.1%
Δq ǂ4 −0.3% 0.1% 0.0% 0.1% 0.1% 0.2% 0.0% 0.8% 0.1% 0.1% 0.1% 0.3%
ΔΦ ǂ5 0.6% 1.9% 4.5% 3.3% 3.9% 1.8% 3.2% 2.1% 1.6% 4.5% 2.7% 1.3%

§Avg = average value §§SD = standard deviation
*1Ėsim simplified power loss
*2ĖsimU⃑n power loss with correctly calculated velocity profile (U⃑n)
*3τ = ∫CS τ⃑ · U⃑ dA viscous work present on the control surface
*4
W˙q=tVρ(U22) dV
effect of time-derivatives of dynamic pressure and volume
*5Ėloss = (ĖsimU⃑n) + τq true power loss
*6Φ viscous dissipation rate
ǂ1ΔĖsim = (ĖlossĖsim)/Ėloss discrepancy between simplified and true power losses
ǂ2Δ(ĖsimU⃑n) = 1 ∸ Ėsim/(ĖsimU⃑n) time-averaged percentage discrepancy between power loss obtained using an accurate spatial velocity profile (assumption #5) and true power loss.
ǂ3 Δτ = τloss time-averaged effect of viscous work on the control surface, assumption #2
ǂ4 Δq = qloss time-averaged effect of control volume and dynamic pressure, assumption #3
ǂ5ΔΦ = (Ėloss − Φ)/Ėloss time-averaged discrepancy between viscous dissipation rate and true power loss

Footnotes

The authors report no conflicts of interest.

References

  • 1.Bakhshinejad A, Baghaie A, Vali A, Saloner D, Rayz VL, D’Souza RM. Merging Computational Fluid Dynamics and 4D Flow MRI Using Proper Orthogonal Decomposition and Ridge Regression. J. Biomech. 2017 doi: 10.1016/j.jbiomech.2017.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat. Methods Med. Res. 1999;8:135–160. doi: 10.1177/096228029900800204. [DOI] [PubMed] [Google Scholar]
  • 3.Chandran KB, Rittgers SE, Ajit P Yoganathan AP. Biofluid mechanics : the human circulation. CRC Press, Taylor & Francis Group; 2012. p. 431. at < https://www.crcpress.com/Biofluid-Mechanics-The-Human-Circulation-Second-Edition/Chandran-Rittgers-Yoganathan/p/book/9781439845165>. [Google Scholar]
  • 4.Cheng AL, Takao CM, Wenby RB, Meiselman HJ, Wood JC, Detterich JA. Elevated Low-Shear Blood Viscosity is Associated with Decreased Pulmonary Blood Flow in Children with Univentricular Heart Defects. Pediatr. Cardiol. 2016;37:789–801. doi: 10.1007/s00246-016-1352-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Chopski SG, Rangus OM, Moskowitz WB, Throckmorton AL. Experimental measurements of energy augmentation for mechanical circulatory assistance in a patient-specific fontan model. Artif. Organs. 2014;38:791–9. doi: 10.1111/aor.12255. [DOI] [PubMed] [Google Scholar]
  • 6.Cibis M, Jarvis K, Markl M, Rose M, Rigsby C, Barker AJ, Wentzel JJ. The effect of resolution on viscous dissipation measured with 4D flow MRI in patients with Fontan circulation: Evaluation using computational fluid dynamics. J. Biomech. 2015;48:2984–2989. doi: 10.1016/j.jbiomech.2015.07.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Dasi LP, Pekkan K, Katajima HD, Yoganathan AP. Functional analysis of Fontan energy dissipation. J. Biomech. 2008;41:2246–2252. doi: 10.1016/j.jbiomech.2008.04.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Dubini G, de Leval MR, Pietrabissa R, Montevecchi FM, Fumero R. A Numerical Fluid Mechanical Study of Repaired Congenital Heart Defects. Application to the Total Cavopulmonary Connection. J. Biomech. 1996;29:111–121. doi: 10.1016/0021-9290(95)00021-6. [DOI] [PubMed] [Google Scholar]
  • 9.Ensley AE, Lynch P, Chatzimavroudis GP, Lucas C, Sharma S, Yoganathan AP. Toward designing the optimal total cavopulmonary connection: an in vitro study. Ann. Thorac. Surg. 1999;68:1384–1390. doi: 10.1016/s0003-4975(99)00560-3. [DOI] [PubMed] [Google Scholar]
  • 10.Fung YC. Biomechanics: Circulation. Springer; 1997. [Google Scholar]
  • 11.Grigioni M, Avenio GD, Del Gaudio C, Morbiducci U. Critical issues in studies of flow through the Fontan circuit after 10 years of investigation. Cardiol. Young. 2005;15:68–73. doi: 10.1017/S1047951105001678. [DOI] [PubMed] [Google Scholar]
  • 12.Ha H, Kim GB, Kweon J, Lee SJ, Kim Y-H, Lee DH, Yang DH, Kim N. Hemodynamic Measurement Using Four-Dimensional Phase-Contrast MRI: Quantification of Hemodynamic Parameters and Clinical Applications. Korean J. Radiol. 2016;17:445. doi: 10.3348/kjr.2016.17.4.445. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Haggerty CM, Restrepo M, Tang E, de Zélicourt DA, Sundareswaran KS, Mirabella L, Bethel J, Whitehead KK, Fogel MA, Yoganathan AP. Fontan hemodynamics from 100 patient-specific cardiac magnetic resonance studies: A computational fluid dynamics analysis. J. Thorac. Cardiovasc. Surg. 2014;148:1481–1489. doi: 10.1016/j.jtcvs.2013.11.060. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Haggerty CM, Whitehead KK, Bethel J, Fogel MA, Yoganathan AP. Relationship of Single Ventricle Filling and Preload to Total Cavopulmonary Connection Hemodynamics. Ann. Thorac. Surg. 2015;99:911–917. doi: 10.1016/j.athoracsur.2014.10.043. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Healy TM, Lucas C, Yoganathan AP. Noninvasive Fluid Dynamic Power Loss Assessments for Total Cavopulmonary Connections Using the Viscous Dissipation Function: A Feasibility Study. J. Biomech. Eng. 2001;123:317. doi: 10.1115/1.1384875. [DOI] [PubMed] [Google Scholar]
  • 16.Johnston BM, Johnston PR, Corney S, Kilpatrick D. Non-Newtonian blood flow in human right coronary arteries: Steady state simulations. J. Biomech. 2004;37:709–720. doi: 10.1016/j.jbiomech.2003.09.016. [DOI] [PubMed] [Google Scholar]
  • 17.Khiabani RH, Whitehead KK, Han D, Restrepo M, Tang E, Bethel J, Paridon SM, Fogel Ma, Yoganathan AP. Exercise capacity in single-ventricle patients after Fontan correlates with haemodynamic energy loss in TCPC. Heart. 2015;101:139–143. doi: 10.1136/heartjnl-2014-306337. [DOI] [PubMed] [Google Scholar]
  • 18.Lardo AC, Webber SA, Friehs I, Del Nido PJ, Cape EG. Fluid dynamic comparison of intra-atrial and extracardiac total cavopulmonary connections. J. Thorac. Cardiovasc. Surg. 1999;117:697–704. doi: 10.1016/S0022-5223(99)70289-8. [DOI] [PubMed] [Google Scholar]
  • 19.Long CC, Hsu M-CCM, Bazilevs Y, Feinstein JA, Marsden AL. Fluid – structure interaction simulations of the Fontan procedure using variable wall properties. Int. j. numer. method. biomed. eng. 2012;28:513–527. doi: 10.1002/cnm.1485. [DOI] [PubMed] [Google Scholar]
  • 20.Low HTT, Chew YTT, Lee CNN. Flow studies on atriopulmonary and cavopulmonary connections of the Fontan operations for congenital heart defects. J. Biomed. Eng. 1993;15:303–7. doi: 10.1016/0141-5425(93)90006-k. [DOI] [PubMed] [Google Scholar]
  • 21.Moyle K, Mallinson G, Cowan B. Volumetric methods for evaluating irreversible energy losses and entropy production with application to bioengineering flows. Int. J. Numer. Methods Fluids. 2006;50:1357–1368. [Google Scholar]
  • 22.Munson BR, Young DF, Okiishi TH. Fundamentals of Fluid Mechanics. John Wiley & Sons; 2006. [Google Scholar]
  • 23.Restrepo M, Luffel M, Sebring J, Kanter KR, Del Nido PJ, Veneziani A, Rossignac J, Yoganathan AP. Surgical Planning of the Total Cavopulmonary Connection: Robustness Analysis. Ann. Biomed. Eng. 2014 doi: 10.1007/s10439-014-1149-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Restrepo M, Tang E, Haggerty CM, Khiabani RH, Mirabella L, Bethel J, Valente AM, Whitehead KK, McElhinney DB, Fogel MA, Yoganathan AP. Energetic Implications of Vessel Growth and Flow Changes Over Time in Fontan Patients. Ann. Thorac. Surg. 2015;99:163–170. doi: 10.1016/j.athoracsur.2014.08.046. [DOI] [PubMed] [Google Scholar]
  • 25.Santhanakrishnan A, Maher KO, Tang E, Khiabani RH, Johnson J, Yoganathan AP. Hemodynamic effects of implanting a unidirectional valve in the inferior vena cava of the Fontan circulation pathway: an in vitro investigation. Am. J. Physiol. - Hear. Circ. Physiol. 2013;305:H1538–H1547. doi: 10.1152/ajpheart.00351.2013. [DOI] [PubMed] [Google Scholar]
  • 26.Soerensen DD, Pekkan K, Sundareswaran KS, Yoganathan AP. New power loss optimized Fontan connection evaluated by calculation of power loss using high resolution PC-MRI and CFD. Conf. Proc. IEEE Eng. Med. Biol. Soc. 2004;2:1144–7. doi: 10.1109/IEMBS.2004.1403367. [DOI] [PubMed] [Google Scholar]
  • 27.Tang E, Restrepo M, Haggerty CM, Mirabella L, Bethel J, Whitehead KK, Fogel MA, Yoganathan AP. Geometric Characterization of Patient-Specific Total Cavopulmonary Connections and its Relationship to Hemodynamics. JACC Cardiovasc. Imaging. 2014;7:215–224. doi: 10.1016/j.jcmg.2013.12.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Tang E, Alan Wei Z, Whitehead KK, Khiabani RH, Restrepo M, Mirabella L, Bethel J, Paridon SM, Marino BS, Fogel MA, Yoganathan AP. Effect of Fontan geometry on exercise haemodynamics and its potential implications. Heart. doi: 10.1136/heartjnl-2016-310855. heartjnl-2016-310855. [DOI] [PubMed] [Google Scholar]
  • 29.Volonghi P, Tresoldi D, Cadioli M, Usuelli AM, Ponzini R, Morbiducci U, Esposito A, Rizzo G. Automatic Extraction of Three-Dimensional Thoracic Aorta Geometric Model from Phase Contrast MRI for Morphometric and Hemodynamic Characterization. 2016;882:873–882. doi: 10.1002/mrm.25630. [DOI] [PubMed] [Google Scholar]
  • 30.Vukicevic M, Conover TA, Jaeggli M, Zhou J, Pennati G, Hsia T-YT, Figliola RS. Control of respiration-driven retrograde flow in the subdiaphragmatic venous return of the Fontan circulation. ASAIO J. 2014;60:21–23. doi: 10.1097/MAT.0000000000000093. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Wang C, Pekkan K, de Zélicourt D, Horner M, Parihar A, Kulkarni A, Yoganathan AP. Progress in the CFD Modeling of Flow Instabilities in Anatomical Total Cavopulmonary Connections. Ann. Biomed. Eng. 2007;35:1840–1856. doi: 10.1007/s10439-007-9356-0. [DOI] [PubMed] [Google Scholar]
  • 32.Wei ZA, Trusty PM, Tree M, Haggerty CM, Tang E, Fogel M, Yoganathan AP. Can time-averaged flow boundary conditions be used to meet the clinical timeline for Fontan surgical planning? J. Biomech. 2017;50:172–179. doi: 10.1016/j.jbiomech.2016.11.025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Wei Z, Whitehead KK, Khiabani RH, Tree M, Tang E, Paridon SM, Fogel MA, Yoganathan AP. Respiratory Effects on Fontan Circulation During Rest and Exercise Using Real-Time Cardiac Magnetic Resonance Imaging. Ann. Thorac. Surg. 2016;101:1818–1825. doi: 10.1016/j.athoracsur.2015.11.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Whitehead KK, Pekkan K, Kitajima HD, Paridon SM, Yoganathan AP, Fogel Ma. Nonlinear Power Loss During Exercise in Single-Ventricle Patients After the Fontan: Insights From Computational Fluid Dynamics. Circulation. 2007;116:I-165–I-171. doi: 10.1161/CIRCULATIONAHA.106.680827. [DOI] [PubMed] [Google Scholar]