Microfluidic methods for aptamer selection and characterization

Sean K Dembowski a, Michael T Bowser a,*
PMCID: PMC5819361  NIHMSID: NIHMS917402  PMID: 29094731

Abstract

Nucleic acid aptamers have tremendous potential as molecular recognition elements in biomedical targeting, analytical arrays, and self-signaling sensors. However, practical limitations and inefficiencies in the process of selecting novel aptamers (SELEX) have hampered widespread adoption of aptamer technologies. Many factors have recently contributed to more effective aptamer screening, but no influence has done more to increase the efficiency, scale, and automation of aptamer selection than that of new microfluidic SELEX techniques. This review introduces aptamers as a powerful chemical and biological tool, briefly highlights traditional SELEX techniques and their limitations, covers in detail the recent advancements in microfluidic methods of aptamer selection and characterization, and suggests possible future directions of the field.

Graphical abstract

graphic file with name nihms917402u1.jpg

Introduction

Aptamers

Aptamers are synthetic sequences of RNA or single-stranded DNA that fold into a particular three-dimensional conformation that allows them to bind a molecular target with high affinity and specificity. Because nucleic acids possess an innate flexibility and adopt a wide variety of structural motifs, aptamers have been developed for virtually every type of target imaginable: metal ions,13 small molecules,47 lipids,8, 9 proteins,1014 viruses,15 and whole cells.16, 17 Though often likened to antibodies, aptamers have a number of distinct advantages over their protein counterparts. Most notably, specific aptamers are generated by in vitro solid-phase synthesis, an inexpensive, scalable, and easily automated technique with excellent batch-to-batch reproducibility. This avoids the use of model animals required of antibody production and the associated time and expertise required. Additionally, aptamers are highly stable over wide ranges of pH, temperature and ionic strength, and they refold spontaneously after denaturation. Finally, aptamers are easy to chemically link or modify during the synthesis process at virtually any position for customized applications.

Aptamers are utilized in a variety of laboratory affinity assays and separations, especially those previously dominated by antibodies such as biomolecule purification,18, 19 chiral separation,20 and enzyme-linked immunosorbent assays (ELISA).2123 Because nucleic acid aptamers can undergo conformational changes upon binding, they are well suited to the development FRET- or quenching-based biosensors for their target, while antibodies are too inflexible for such behavior.2, 24, 25 Aptamers are also gaining favor in biomedical applications as both diagnostic and therapeutic tools. Aptamers have been used extensively to target imaging agents,2629 siRNA,30, 31 and drugs3234 both in vitro and in vivo. While targeting is the most common medical use, modified aptamers have also had success as primary therapeutic agents, most notably Macugen for the treatment of age-related macular degeneration (AMD). While Macugen is thus far the only FDA approved aptamer-based drug, several aptamer therapeutics are in clinical trials and numerous others are in preclinical stages.3537

Given the tremendous applicability and the advantages of aptamers, it is not surprising that the number of aptamer-related articles published annually has increased more than 10-fold since 2001.35 However, the selection of aptamers for novel targets is less well represented in the literature, primarily because of the time, resources, and expertise required to combine the disparate steps in aptamer development. In addition, successful selection of useful and applicable aptamers is never a guarantee, even in the most careful and well-intentioned cases.

One innovation that has dramatically lessened the time- and labor-intensiveness of aptamer technologies is the incorporation of novel microfluidic strategies; no single advancement has led to as much improvement in speed, throughput, and automation in aptamer selection and characterization. (Microfluidics have also led to notable advances in aptamer-based sensing (aptasensors), especially involving electrochemical detection3841 and low-cost printed or paper-based devices;4244 however, microfluidic applications of aptamers will not be covered here. Aptasensors have been recently reviewed elsewhere.4548)

The current review provides a detailed overview of the different types of microfluidic aptamer selection schemes, highlighting key examples of these families and the major contributions of the work. Current work in aptamer microarrays for selection and characterization are also covered, and the authors offer possible future directions of the field.

The SELEX method

New aptamers are selected by an iterative combinatorial process called Systematic Evolution of Ligands by Exponential Enrichment (SELEX),49, 50 which is outlined in Figure 1. Briefly, a library of randomized RNA or DNA oligonucleotide sequences (oligos) is incubated with the target of interest. The large number of sequences creates sufficient structural diversity for numerous sequences to compete for binding with a limited amount of target. The in vitro nature of this incubation allows parameters such as temperature and ionic strength to be adjusted for optimized binding under the specified conditions. Next, unbound oligos are separated by one of a variety of partitioning techniques and discarded while those sequences complexed with the target are amplified by PCR. (In the case of RNA aptamers, sequences are reverse transcribed prior to PCR and transcribed back to RNA afterward.) Finally, the PCR products are purified and complementary strands are removed. This results in a new single-stranded pool that has been enriched with sequences that display affinity for the target. At the end of each cycle, the average affinity of the pool for the target of interest is measured, and the cycle is repeated until optimal binding affinity is reached. Finally, the nucleotide pool is sequenced and analyzed, and a subset of those sequences is synthesized and characterized in terms of their affinity, selectivity, and physical properties.

Figure 1.

Figure 1

Schematic of SELEX process for isolating aptamers. The nucleic acid library undergoes incubation with the target of interest, separation to remove sequences not bound to the target, amplification of binding sequences, and purification to remove complementary nucleotide strands. RNA selections require additional transcription and reverse transcription steps (not shown).

Conventional SELEX

Nitrocellulose filtration was the first major partitioning technique employed because of its ability to pass nucleic acid strands while retaining proteins.49 While the technique is simple in that it requires no immobilization of the target and little adjustment from protein to protein, filtration is relatively inefficient and time-consuming, requiring a dozen rounds of SELEX or more and weeks to months to achieve suitable affinity. Nitrocellulose filtration is also only effective for proteins, making it unsuitable for many proposed targets. Affinity chromatography-based techniques, which rely on immobilizing the target of interest on a solid support, were introduced to address challenges presented by nitrocellulose SELEX and are now the technique of choice for many. The most frequently used supports are microbeads,5153 which are easily retained by a microcolumn filter during selection, but other surfaces such as microplates are also used.54, 55 In both cases, the target can be attached covalently with linker chemistry,56 adsorbed non-specifically,54 or linked with biotin to a streptavidin-coated surface.57 While this technique enhances flexibility and the number of allowable targets, macroscale affinity chromatography on the whole does not substantially increase the efficiency of the selection and often uses relatively large amounts of material. Additionally, the solid supports, and any linker or capping molecules used, present potential surfaces with which aptamers may develop unwanted, non-specific affinity, requiring negative selections to obtain target-specific aptamers.

Microfluidic SELEX techniques

While different SELEX techniques offer certain pros and cons, there are a number of advantages that hold true of all microfluidic SELEX methods. First, because most separation techniques typically only operate at a surface interface, shrinking the system to yield a larger surface-to-volume ratio dramatically increases separation efficiency while reducing material consumption. In the case of affinity chromatography, this decrease in size also reduces the potential for non-specific and off-target binding, in many cases allowing negative selections to be removed altogether. Second, it has been shown that reducing the amount of target (thereby increasing the stringency of the selection) results in faster convergence to the aptamer pool to a small number of sequences.58 Microfluidic methods excel at handling small amounts of reagents, both increasing stringency and cutting material costs. Finally, a major drawback to SELEX in general is that each phase of the cycle can be labor-intensive, time-consuming, and require special expertise. Microfluidic techniques have a much higher potential for automation and multiplexing of this process, which will reduce the time and energy invested into each new aptamer and streamline the SELEX process. This has led to the development of many microfluidic SELEX separation techniques, including those based on capillary electrophoresis, microfluidic bead-based separations, and sol-gel encapsulation.

Capillary electrophoresis SELEX

The first microfluidic technique introduced to enhance SELEX partitioning was capillary electrophoresis (CE). CE separates components of a mixture based on their electrophoretic mobility. Unbound DNA or RNA migrate together while oligo-target complexes, because of their altered size and charge, separate from the unbound sequences and are collected (Figure 2). Because DNA and RNA have a higher negative charge density than virtually any target molecule of interest, complexation results in a significant and predictable decrease in mobility. This, together with the dramatic increase in the performance and prevalence of CE technology during the 1990s in response to the Human Genome Project, made CE the natural choice for SELEX’s stringent nucleic acid separation requirements.

Figure 2.

Figure 2

Example electropherogram of CE-SELEX separation. Target and unbound oligos have very different electrophoretic mobilities and separate during CE. Target-oligo complexes migrate between the two component peaks. Target and complexes are collected for amplification while unbound oligos are not.

CE-based aptamer selection (CE-SELEX) was first implemented by Mendonsa and Bowser seeking a DNA aptamer for human immunoglobulin E (IgE).59 Because of the high partitioning efficiency of CE, only four rounds of SELEX were required to obtain multiple aptamer sequences with dissociation constants (Kd) below 30 nM, drastically shortening the overall selection time. Additionally, the entire incubation and separation process can be performed in free solution, which allows the process to be done without tagging or immobilizing either molecule. Similar CE-SELEX techniques have been used to select aptamers for a variety of proteins, including human immunodeficiency virus-reverse transcriptase (HIV-RT),60 human vascular endothelial growth factor 165 (hVEGF165),61 α-fetoprotein,62 and human epididymis protein 4 (HE4).63 Although the separation between bound and unbound sequences is more substantial with larger targets, CE-SELEX has also been used successfully on smaller targets including the 4.3-kDa neuropeptide-Y64 and the 580-Da N-methyl mesoporphyrin (NMM),65 demonstrating the flexibility and robust nature of the technique.

An additional advantage of CE-SELEX is that CE provides an excellent technique for determining the binding affinity of both the mixed selection pool and individual aptamer sequences. However, this is typically done in an additional post-selection step. The Krylov group developed Non-Equilibrium Capillary Electrophoresis of Equilibrium Mixtures (NECEEM), which applied a novel analysis to extract kinetic and thermodynamic parameters of the nucleic acid pool.66 By supplying target-free buffer at the capillary inlet, the migrating complexes slowly unbound depending on the aptamer dissociation equilibrium and rate constants (Kd and koff, respectively), which were analyzed in the separation electropherogram. Krylov further reduced the required time by removing the intermediate amplification and purification steps altogether. This “non-SELEX” method facilitated three successive NECEEM separation steps in one hour and improved the affinity of the DNA pool for the h-RAS protein by over four orders of magnitude.67 While the selection stringency of this technique may be limited by the need to collect sufficient DNA for subsequent partitioning steps, the utility of non-SELEX has been further demonstrated in identifying aptamers for three signal transduction proteins, each with at least three orders of magnitude improvement in affinity.68

One major concern over CE-SELEX is the small sample volume that CE can handle. Injections are often less than 10 nL,61 so even when very high DNA concentrations are used, only about 1012-1013 oligos are sampled per run. One way around this has been utilizing bulk separation techniques like nitrocellulose filtration prior to CE to remove a large portion of non-binding sequences before CE-SELEX.69 This effectively samples more relevant sequences by removing a large fraction of non-binders before the CE step, although it does so at the cost of additional time and complexity.

Another approach to increase the number of sequences assessed is to move from discrete separations to continuous partitioning techniques. One example of this is micro-free flow electrophoresis (µFFE). In this technique, buffer and sample are pressure-driven through a planar channel while an electric field is applied across the channel, perpendicular to the direction of flow (Figure 3). Because species are deflected laterally based on their mobility, sample can be streamed continuously through the device rather than injected in discrete plugs. Binding sequences are collected through a separate exit port than non-binders. This continuous approach eliminates the complicated manual timing required in CE to obtain reproducible sample collections. A µFFE-SELEX technique was demonstrated that sampled nearly 2×1014 DNA sequences in a 30-minute separation, a 300-fold improvement over CE-SELEX.70 Aptamers for human IgE with a Kd of 20–30 nM were selected in only two rounds, indicating no loss in performance versus CE-based techniques.

Figure 3.

Figure 3

Schematic of µFFE-SELEX device. Buffer pressure at the inlet drives sample along the channel at a constant rate. A voltage is applied to the electrodes to create an electric field across the channel, deflecting non-binding sequences to waste while oligo-target complexes are collected from a separate outlet.

Bead–based microfluidic SELEX

Magnetic methods

Implementing microbeads as a solid support allows even small volume systems to benefit from a large surface area, increasing the physical selection space. SELEX on microparticles was first implemented using magnetic beads almost 20 years ago and is still widely used today.71, 72 Briefly, the target molecule is covalently attached to the surface of a magnetic (usually superparamagnetic) microparticle or nanoparticle and incubated with the nucleic acid library. Binding oligos complex with the target and become bound to the magnetic particles as well. A strong external magnet or magnetic field is then applied, which either isolates the particles from the suspension or holds them immobile while they are rinsed. While the theory is simple, the efficiency and practicality of bulk magnetic separation are limited compared to other techniques. However, combining a microfluidic approach with traditional microbead separation enhances the effectiveness of the technique dramatically.

The Soh group was the first to implement magnetic microparticle SELEX on the microfluidic scale. Their first device, a continuous-flow magnetic activated chip separation device (CMACS), used a multi-stream laminar flow architecture to keep the bulk sample at the edges of the channel and a fresh buffer stream along the center (Figure 4). This layout had less lateral diffusion across the interface than in a single-stream flow, which enhanced separation efficiency.73 Tiny ferromagnetic structures created strong, precise magnetic fields to guide the beads into the central buffer stream for collection. This microfluidic addition enhanced recovery of beads to over 99.5% under optimal conditions and achieved a bulk Kd of 33 nM in just one round.74 Despite the excellent results, CMACS suffered from a number of difficulties in practice. Ideal flow conditions were only reached with careful adjustment under microscopic monitoring, undermining the robustness of the technique. Microbubbles and blockages also occurred frequently, which disrupted the flow streams and had a detrimental impact on aptamer separation.

Figure 4.

Figure 4

Schematic of the CMACS device. Laminar flow architecture keeps the sample stream near the channel walls and flows to waste. Angled magnetized Ni strips guide the magnetic beads containing bound oligos into the central buffer stream which is collected through a separate outlet. Optical micrographs show the device during operation at A) the channel wall, B) the buffer stream, and C) the outlet. The large distance m prevents diffusion of unwanted sequences into the central stream. Adapted from Lou et al73 Copyright 2009 National Academy of Sciences.

To combat these issues, a new design was developed: the micromagnetic separation chip (MMS). The MMS relied on grids of patterned titanium and nickel, which were magnetized externally, to catch target-coated magnetic beads as they flowed through the channel. Once enough sample had passed through the chip, the channel was rinsed thoroughly to remove non-specifically adhered oligos. The external magnets were then removed, releasing the microbeads and target-binding sequences into solution for collection (Figure 5).74 The MMS chip demonstrated recovery and partition efficiency as good as or better than those of CMACS (roughly 99.5% and 106, respectively) while offering significantly more robust and reproducible operation. MMS was first demonstrated using the protein streptavidin as the target, and later applied to platelet-derived growth factor-BB to obtain three aptamers with Kd’s below 3 nM.75 The process has also been incorporated into special selections, such as obtaining orthogonal aptamer pairs for sandwich assays,76 aptamers with a long binding lifetime,77 and self-reporting aptamer biosensors.78

Figure 5.

Figure 5

Schematic of the MMS device. A) Top view and B) side view. Sequential buffer and sample inlets flow sample towards the outlet. The patterned Ti/Ni grids become magnetized by external rare-earth magnets and traps magnetic beads containing bound oligos. Adapted from Qian et al74 with permission from the American Chemical Society.

Magnetic microbead SELEX continues to be perhaps the most popular microfluidic SELEX mechanism, with constant tweaking and adjustment pushing the boundaries of the technique. Recently, a device by Hong et al.79 incorporated both positive and negative selection chambers into the chip separated by a microvalve, reducing the time and hassle of performing a negative selection. A quantitative fluorescence binding analysis was performed on a separate microfluidic chip using magnetic target immobilization, demonstrating a trend towards miniaturization of additional steps of the aptamer selection process.

Non-magnetic methods

Non-magnetic beads have also transitioned from bulk affinity chromatography-based separations to microfluidic SELEX. These beads, generally made of silica derivatives or organic polymers like agarose and acrylate, supply the same increased surface area with lower cost and without the challenges associated with magnetic separation. Recently, Zhao et al. demonstrated a simple custom chip that used two sequential pinches to trap acrylate beads in the channel, the first beads containing multiple non-target proteins (negative selection) and the second beads functionalized with myoglobin (positive selection).80 The chip is shown schematically in Figure 6. DNA was introduced continually at the top of the channel and flowed first through the negative selection zone to remove unwanted non-specific binders and then through the positive selection zone to retain sequences with affinity for myoglobin. The second pinch was then rinsed to remove weak binders before aptamers were eluted with DDT. This technique exemplifies the ease of performing simultaneous positive and negative selection that microfluidic SELEX offers. A peak Kd of 5 nM was obtained with exquisite specificity.

Figure 6.

Figure 6

Schematic of dual positive-negative selection chip. Beads in the first pinch are coated with multiple non-target proteins for negative selection, while beads in the second pinch are coated with the target of interest. The oligos are flowed through inlet 1, and only sequences with affinity for the target but not for any of the negative proteins will bind in the second pinch. These oligos are then eluted by the addition of DDT to inlet 2. Adapted from Wang et al80 with permission from the American Chemical Society.

Since these beads do not move throughout the separation, similar strategies have been employed that forego the microbeads altogether, immobilizing the target directly on the device surface with similar effect. Liu et al. constructed a circuitous microfluidic path over a protein microarray so that oligos passed over four different negative selection proteins before reaching the region displaying the target, lactoferrin.81 Only those sequences bound to lactoferrin were eluted, achieving multiple sequences with Kd’s below 10 nM.

In some cases though, the microbeads don’t just provide a static surface but are actually a dynamic part of the device as with many magnetic systems. Recently, a continuous separation device was developed using migration by acoustic waves (acoustophoresis) for aptamer selection.82 Similarly to the CMACS device, the incubated sample of target-coated beads and DNA was introduced at the edges of the channel with a central buffer stream holding the sample stream against the channel walls. As sample passed down the channel, an acoustic standing wave field focused the beads to the node in the central buffer stream for collection while leaving smaller species such as unbound DNA in the waste-bound edge streams. An aptamer was obtained with sub-nanomolar affinity for prostate-specific antigen (PSA) after eight rounds of SELEX. As with other techniques not requiring a separate elution step (µFFE, CMACS), this device can overcome a common weakness of microfluidic SELEX, the small number of sequences sampled, by simply extending the time of continuous operation.

Sol-gel–based microfluidic SELEX

While SELEX techniques that rely on immobilizing the target have undoubtedly been successful in many cases, there has always been some concern over whether immobilization affects target conformation, occludes possible binding sites, or has other unintended consequences on the selection. One approach to circumvent covalent immobilization has been the use of sol-gel nanocomposites to trap proteins or other biomolecules.83, 84 Briefly, monomers or nanoparticles of the gel precursor (often one or more silica derivatives) are dissolved to create an aqueous sol. Chemical additives, along with the target, are then mixed into the sol, causing the silica to coalesce and solidify into a porous framework, trapping the target in the growing gel. Gels with pores of two different sizes are favored; the smaller pores trap the macromolecule in place while the larger pores allow easy diffusion of aptamers or other ligands throughout the network. It has been shown that enzymes in the gel not only maintain their function but can be more stable against chemical and thermal denaturation than in free solution.85, 86

Sol-gel encapsulation was first demonstrated as a SELEX technique by Park et al.87 The microfluidic device consisted of a narrow channel connecting five identical chambers, each constructed over an aluminium microheater. A droplet of the protein-containing sol-gel was spotted into each chamber and allowed to gel overnight, then a polydimethylsiloxane (PDMS) cover sealed the chip. Incubated oligos were observed to bind with high efficiency to the trapped target but not significantly to the gel matrix. Applying current through the heater electrodes yielded enough heat to denature and elute the oligo sequences. A strength of the technique is high target density, with each 300-µm droplet holding about 0.6 ng of protein. The device isolated RNA aptamers with affinity for a yeast transcription factor87 and TATA binding protein88 with high affinity and in fewer rounds than with conventional SELEX. The high target density and heat-triggered release lend themselves well to a multiplexed selection, but the first attempts at simultaneous selection were fraught with issues of contamination. However, the design was improved by independent sample and elution lines as well as pneumatic valves between the reaction chambers (Figure 7). The new device performed selections against five targets simultaneously in less time than the previous system without contamination.89 The capabilities and applicability of sol-gel–based SELEX have continued to expand, as the technique has selected aptamers against the small molecule xanthine, even when traditional SELEX could not,90 and against Bisphenol-A, which is too insoluble for other SELEX methods.91

Figure 7.

Figure 7

Multiplex sol-gel–based SELEX chip, showing (a) a schematic of the individual layers and (b) a picture of the assembled chip. Electrode heaters (b, left inset) are patterned onto the bottom glass layer and spotted with the sol-gel. The pneumatic layer controls the valves between reaction chambers to prevent cross-contamination during elution, while the remaining layers contain the main fluidic network. With valves open, sample is flowed through the reaction chambers to incubate with each sol-gel droplet, followed by washing to remove non-specific binders. Valves are closed and aptamers are eluted by running current through the electrode heaters and collecting at each elution outlet. Reproduced from Lee et al89 with permission from Springer.

Integrated microfluidic SELEX systems

As discussed previously, the SELEX cycle consists of four main steps: incubation of oligos with the target to compete for binding; partitioning of free oligos from those bound to the target; amplification of the oligo pool by PCR; and purification, whereby the complementary strand is removed to achieve a single-stranded pool again. (Additional transcription and reverse transcription steps are necessary when RNA aptamers are sought.) The techniques discussed thus far have focused on incorporating the partitioning and sometimes the incubation steps into to microfluidic environment. However, one of the main draws of a microfluidic SELEX technique, the enhanced potential for automation, is significantly limited until all parts of this cycle can be linked in a single integrated system. For this reason, many groups have pursued not only the microfluidic partitioning methods described above but also miniaturization of the amplification, purification, and intermediate steps as well.

The first major step to this end was a prototype by Hybarger et al. that incorporated transcription, incubation, selection, reverse transcription, and amplification into a system of microlines, all controlled by valves actuated in LabView.92 The main achievement of the prototype was not just a seamless transition between the steps but also the use of pressurized reagent manifolds to fill, rinse, and clear sections of the system accordingly with no manual intervention. The full prototype still had a large footprint and its own share of practical difficulties, but it nonetheless set a high bar for future integrated SELEX systems.

The first integrated microfluidic SELEX chip, developed in the lab of Gwo-Bin Lee, was a three-layer, four-chamber PDMS and glass chip.93 The processes of incubation, separation, and amplification were performed in a central sample chamber connected to reagent and waste chambers by pneumatic microvalves (Figure 8). The system used typical magnetic bead SELEX, first demonstrated with C-reactive protein (CRP) as the target conjugated to superparamagnetic polystyrene beads. Efficient mixing of the sample chamber was achieved within three seconds by deflection of the PDMS membrane layer with pressurized air. After incubation, the microbeads and any bound aptamers were held by an external magnetic field and rinsed thoroughly from the wash buffer chamber. PCR reagents were then introduced and the temperature was cycled in the sample chamber using microheaters and microsensors patterned on the glass substrate. The inherent automation coupled with the faster temperature cycling allowed by on-chip PCR achieved a round of incubation, selection and amplification in under 60 minutes.

Figure 8.

Figure 8

A schematic of the integrated SELEX chip by Huang et al. (A) The bottom glass layer is patterned with microheaters and sensors (platinum) with associated electrodes (gold) for temperature control during PCR. The thin-film PDMS membrane and thick PDMS block facilitate all liquid handling, including pumping and mixing by pneumatic membrane deflection. (B) The assembled device contains a main incubation/PCR chamber connected to chambers for wash buffer, PCR reagents, and waste by low-volume microchannels to prevent dead space and bubble formation. Adapted from Huang et al93 with permission from Elsevier.

The system was further enhanced to improve mixing and transport, and a competitive assay chip was designed to accompany the SELEX chip and further improve automation and speed of the overall process.94 Though the competitive assay chip was not a direct in-line addition and required intermediate manual processing, the increased efficiency of a chip-based assay provides a glimpse into the possible future of combined SELEX-aptamer characterization chips. The system has been continuously adapted and expanded to improve function and robustness. Example improvements include additional reagent and sample chambers, vacuum-driven pumps to accommodate live cells without lysis, and temperature-controlled regions of the device to speed cycling while avoiding reagent degradation.9597

Purification of the PCR products to remove reagents and complementary nucleic acid strands has proven the most difficult piece to incorporate into microfluidic systems, either performed off-chip (reducing speed and automation) or absent altogether (reducing efficiency) in the previous devices. A new design by the Lin group utilized a bead-based PCR system to isolate the active aptamer from its complement generated during PCR.98 The device consisted of two chambers, one for incubation and selection, the other for amplification and purification (Figure 9). The chambers were connected by a gel-filled channel to facilitate electrophoretic transfer of the DNA from one chamber to the other and back. Each chamber was surrounded by a weir to keep the beads in the appropriate chamber while allowing reagents and buffers to come and go freely. The key to purification was the use of reverse PCR primers immobilized onto microbeads in the amplification chamber. By performing PCR with immobilized reverse primers, only aptamer strands were transferred back to the selection chamber while the complementary strands remained bound to the PCR beads. The device performed three complete rounds of selection in under 10 hours with no off-chip activity, obtaining aptamers for both protein (IgE) and small molecule (bisboronic acid) targets. While this strategy provided an elegant solution to the purification problem, the electrophoretic transfer was not without practical issues, which led to a new design that incorporated both electrophoresis and pressure-driven flow into the system.99

Figure 9.

Figure 9

Schematic of the integrated SELEX chip by Kim et al. (A) A top view of the device, showing the selection and amplification chambers connected by a gel-filled channel. (B) A side view of each chamber, showing weirs to prevent bead movement and patterned microheater for elution or PCR. Adapted from Kim et al98

Microfluidic aptamer characterization

Sequencing

The majority of microfluidic influence in SELEX has focused on optimizing the separation and partitioning methods used in the selection itself, with the goal of obtaining sequence pools with the highest average affinity and the most high-affinity binders in the least amount of time. However, even given highly efficient selection and enrichment, the ability to identify the best aptamers from the pool depends heavily on the nucleotide sequencing and binding characterization methods used.

Traditionally, aptamer pools were cloned into bacterial vectors and sequenced by the Sanger method, which demanded labor-intensive handling of individual bacterial colonies and usually returned a few dozen to a hundred sequences at most, a tiny percent of the selected pool. A small number of those sequences were then synthesized and had their affinity tested individually. Though this technique is still commonplace in the field, finding the best aptamers by this method is largely governed by blind luck in picking the colonies. The introduction of high-throughput next-generation sequencing (NGS) techniques has allowed researchers to view a much larger fraction of the pool and to rationally choose sequences for further characterization.75, 100 While sequence-based criteria (usually the copy number or enrichment-fold of a motif) are not always the best indicator of affinity,61, 75 this strategy of rational sequence selection by NGS is arguably more effective than picking colonies randomly. As commercial NGS services become quicker and cheaper, choosing aptamer candidates via NGS analysis is rapidly becoming the method of choice.

Affinity measurement

Once individual sequences are chosen, some number of them, usually up to a few dozen, are synthesized in high purity, and the affinities of the individual sequences are determined for the target of interest as well as possibly for similar targets to determine selectivity. While NGS sequence analysis may hint at the best aptamers, it is no substitute for experimentally determining the affinity of a large number of sequences, as the true trends in affinity always deviate from those predicted by the informatics data. Even when using the microfluidic selection techniques above, affinity characterization has almost universally been done sequentially on large instrumentation. Microplate fluorescence assays, affinity capillary electrophoresis, and surface plasmon resonance are among the most common techniques for finding target affinity, in each case determining the amount of aptamer bound at many different concentrations of target to construct a binding curve and determine the aptamer dissociation constant. Because instrumental characterization techniques are robust and highly automated, it is unlikely that microfluidic characterization methods outside microarrays will see widespread adoption until they display significant advances in performance, automation, or multiplexing.

Custom chip-based methods

DNA microarrays have largely cornered the market on microfluidic affinity assays, but custom characterization chips do still appear in the literature from time to time. For example, the magnetic microbead SELEX by Hong et al.79 was presented with a custom chip that measured the fluorescence intensity of each magnetically bound sample post-partitioning, performing a microfluidic version of the normal microplate fluorescence binding assay. There is the potential for future integration of this type of technique directly into the selection, particularly when the partitioning already makes use of magnetic beads as well. Rather than integration, the Afi-Chip developed by Song et al.101 presents a universal, low-cost design for potential commercialization. The quantitative gas generation in a catalase-linked assay moves a dye bar for each sample, meaning little expertise and no expensive equipment is required to interpret the length-based results. These types of chip-based characterization platforms provide an interesting glimpse into the possible future of the microfluidic SELEX workflow; however, they don’t yet represent an improvement in speed or performance compared to instrumental techniques, which is the main reason most researchers still rely on workhorse analytical techniques or DNA microarrays for characterization.

DNA microarrays

DNA microarrays are the only method of microfluidic characterization that has been widely adopted. A DNA microarray consists of thousands or millions of “spots” on a 2-D surface, with each spot containing many copies of a known DNA sequence. DNA can either be synthesized in situ or generated first and deposited onto the surface. A sample containing the fluorescently labelled target(s) of interest is then flowed over the array, and fluorescence intensity (and color, in the case of multiple targets) is scanned across the array to indicate binding. The basics of DNA microarrays and their usage in microfluidics have been reviewed elsewhere.102, 103

For characterization of aptamers, arrays were first used to examine and improve existing sequences. For example, aptamers against IgE have been of particular interest because a long consensus region is highly conserved through many known aptamers. Microarrays with thousands of sequences systematically mutated from a few IgE aptamers have been used to interrogate the roles of specific bases.104, 105 Mutations of certain bases or motifs completely eliminated binding behavior while other changes had no effect, allowing the original sequence to be truncated further.104 Similar work has been done using microarrays with over 10,000 mutants to analyze and improve existing aptamer sequences binding to green fluorescent protein (GFP) and Drosophila negative elongation factor E (NELF-E).106 Ever improving technology has continued to increase both the number of sequences that can be tested and the amount of sample space that can be interrogated by more complex rational algorithms to control mutation. Work by Kinghorn et al.107 tested nearly 200,000 mutants of previously selected aptamers against a malarial antigen. The mutants were determined using the public tool Resample, which not only recognized and generated preliminary mutants around motif families but also filtered out candidates based on predicted disadvantageous pairing and structural motifs. The process produced a truncated version of an existing sequences, as well as three novel sequences, all with Kds below 10 nM, substantially better than the starting aptamers.

Arrays are most commonly applied after SELEX in this manner, but their use as a part of the SELEX process is an interesting shortcut to condense the workflow and reduce backtracking. This was demonstrated by Cho et al. by combining the MMS device, NGS, and commercial microarrays into the quantitative parallel aptamer selection system (QPASS).108 Briefly, four rounds of SELEX were performed on the MMS against the protein angiopoietin-2 (Ang2). A small aliquot from each pool underwent NGS and returned ~107 sequences, and the sequences were ranked by copy number. Eight identical arrays were then generated, each containing the 235 most common sequences from each pool, including different types of linkers and control sequences, with each spot in triplicate per array. By incubating each array with a different concentration of fluorescent target, an 8-point binding isotherm (and the associated Kd) was determined simultaneously for each of the over 200 sequences. Characterizing that many sequences sequentially by standard methods would be totally unthinkable, demonstrating the power of array-based SELEX.

The researchers have since used microarrays for rapid discovery of aptamer pairs that bind to different epitopes of the same target.109 Such pairs are necessary for some highly selective analytical assays or for development of bidentate aptamer binding schemes. Isolation of these aptamer pairs was difficult and relied largely on luck before this technique. By first generating an array of the Ang2 aptamers found previously, then incubating the target protein with the array, Ang2 was now presented for potential binding but with the binding epitope of the first aptamer occluded. A new library was then flowed over this protein-bound array to identify new sequences that bound to an independent epitope. Screening of this huge number of sequences allowed identification of multiple pairs with the strongest original aptamer sequence. By connecting these two aptamer sequences with a poly-T linker, the resulting aptamer exhibited an astounding 62 pM Kd, 300-fold stronger affinity than either sequence individually. There is still tremendous room for growth, as these examples can be scaled up to current commercially-available arrays of 1–2 million features with almost no increase in time or labor. As the scale, complexity, and economics of microfluidic DNA arrays continue to improve, it seems inevitable that microarrays will become inextricably linked with the SELEX process itself.

Conclusions

Despite many advantages of aptamers over other affinity reagents such as antibodies, conventional SELEX methods for isolating new aptamers are labor-intensive and fickle at best, which has bottlenecked progress in the field. However, incorporating new microfluidic SELEX techniques, either by scaling down existing procedures or by inventing novel strategies only feasible at the microscale, has produced tremendous growth in all parts of the selection and characterization process. While it is likely that no single SELEX technique will emerge as the optimal method in all cases (given the variety of targets, selection requirements, and intended aptamer uses), these abundant new microfluidic SELEX methods all have improved efficiency and speed over the older bulk techniques they have begun to replace.

Based on the current state of the field, we predict two significant future directions. First, extensive work has already been done on specific selection techniques, and we anticipate a continued shift toward integrated microfluidic methods combining selection with purification, characterization, negative selection, and/or multiplexing to take full advantage of the superb automation that microfluidics provides. Second, the trend has thus far been towards increased complexity in prototype design to maximize speed and elegance, creating a divide between method-focused groups designing new selection techniques and those groups selecting aptamers for use in their research. We predict a future movement toward simpler designs that are more cost-effective and accessible to these application-focused groups with less microfluidics expertise, such as 3D printed devices or commercially available disposable devices. In both cases, we eagerly await exciting future microfluidic developments to meet the growing aptamer screening needs in bioanalytical and medical fields.

Acknowledgments

The authors gratefully acknowledge funding from the National Institutes of Health (R01 GM063533).

Biographies

graphic file with name nihms917402b1.gif

Sean Dembowski

Sean Dembowski is a Ph.D. candidate in the Department of Chemistry at the University of Minnesota. He graduated Phi Beta Kappa from Oberlin College with a B.A. in Chemistry. Sean’s current research focuses on selection of aptamers for membrane proteins using capillary electrophoresis and applications of aptamers for in vivo transport and delivery.

graphic file with name nihms917402b2.gif

Michael Bowser

Michael Bowser is a professor of Chemistry at the University of Minnesota. He received his Ph.D. in Chemistry from the University of British Columbia and was a postdoctoral fellow at the University of Florida under Prof. Robert Kennedy. Prof. Bowser’s group has pioneered aptamer selection by capillary electrophoresis (CE-SELEX) and microscale free-flow electrophoresis (µFFE). Applications of interest include adipocyte signaling in obesity, nucleic acid modulation of heart function, aptamers to facilitate cell delivery, high-speed microdialysis assays, and multidimensional separations.

Footnotes

Conflicts of Interest

There are no conflicts of interest to declare

References

  • 1.Qu H, Csordas AT, Wang JP, Oh SS, Eisenstein MS, Soh HT. ACS Nano. 2016;10:7558–7565. doi: 10.1021/acsnano.6b02558. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Rajendran M, Ellington AD. Analytical and Bioanalytical Chemistry. 2008;390:1067–1075. doi: 10.1007/s00216-007-1735-8. [DOI] [PubMed] [Google Scholar]
  • 3.Cho YS, Lee EJ, Lee GH, Hah SS. Bioorganic & Medicinal Chemistry Letters. 2015;25:5536–5539. doi: 10.1016/j.bmcl.2015.10.056. [DOI] [PubMed] [Google Scholar]
  • 4.Mann D, Reinemann C, Stoltenburg R, Strehlitz B. Biochemical and Biophysical Research Communications. 2005;338:1928–1934. doi: 10.1016/j.bbrc.2005.10.172. [DOI] [PubMed] [Google Scholar]
  • 5.Ma XY, Wang WF, Chen XJ, Xia Y, Duan N, Wu SJ, Wang ZP. Food Control. 2015;47:545–551. [Google Scholar]
  • 6.Chen XJ, Huang YK, Duan N, Wu SJ, Ma XY, Xia Y, Zhu CQ, Jiang Y, Wang ZP. Analytical and Bioanalytical Chemistry. 2013;405:6573–6581. doi: 10.1007/s00216-013-7085-9. [DOI] [PubMed] [Google Scholar]
  • 7.Stoltenburg R, Nikolaus N, Strehlitz B. Journal of Analytical Methods in Chemistry. 2012:14. doi: 10.1155/2012/415697. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Khvorova A, Kwak YG, Tamkun M, Majerfeld I, Yarus M. Proceedings of the National Academy of Sciences of the United States of America. 1999;96:10649–10654. doi: 10.1073/pnas.96.19.10649. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Vlassov A, Khvorova A, Yarus M. Proceedings of the National Academy of Sciences of the United States of America. 2001;98:7706–7711. doi: 10.1073/pnas.141041098. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Toscano-Garibay JD, Benitez-Hess ML, Alvarez-Salas LM. In: Cervical Cancer: Methods and Protocols. Keppler D, Lin AW, editors. Vol. 1249. Humana Press Inc; Totowa: 2015. pp. 221–239. [DOI] [PubMed] [Google Scholar]
  • 11.Bock LC, Griffin LC, Latham JA, Vermaas EH, Toole JJ. Nature. 1992;355:564–566. doi: 10.1038/355564a0. [DOI] [PubMed] [Google Scholar]
  • 12.Nonaka Y, Sode K, Ikebukuro K. Molecules. 2010;15:215–225. doi: 10.3390/molecules15010215. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Lupold SE, Hicke BJ, Lin Y, Coffey DS. Cancer Research. 2002;62:4029–4033. [PubMed] [Google Scholar]
  • 14.Shigdar S, Lin J, Yu Y, Pastuovic M, Wei M, Duan W. Cancer Science. 2011;102:991–998. doi: 10.1111/j.1349-7006.2011.01897.x. [DOI] [PubMed] [Google Scholar]
  • 15.Kwon HM, Lee KH, Han BW, Han MR, Kim DH, Kim DE. Plos One. 2014;9:9. doi: 10.1371/journal.pone.0097574. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Lu M, Zhou L, Zheng XH, Quan Y, Wang XL, Zhou XN, Ren J. Cancer Biomarkers. 2015;15:163–170. doi: 10.3233/CBM-140450. [DOI] [PubMed] [Google Scholar]
  • 17.Ji KL, Lim WS, Li SFY, Bhakoo K. Analytical and Bioanalytical Chemistry. 2013;405:6853–6861. doi: 10.1007/s00216-013-7155-z. [DOI] [PubMed] [Google Scholar]
  • 18.Javaherian S, Musheev MU, Kanoatov M, Berezovski MV, Krylov SN. Nucleic Acids Research. 2009:37. doi: 10.1093/nar/gkp176. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Zhao Q, Li XF, Shao YH, Le XC. Analytical Chemistry. 2008;80:7586–7593. doi: 10.1021/ac801206s. [DOI] [PubMed] [Google Scholar]
  • 20.Ruta J, Ravelet C, Desire J, Decout J-L, Peyrin E. Analytical and Bioanalytical Chemistry. 2008;390:1051–1057. doi: 10.1007/s00216-007-1552-0. [DOI] [PubMed] [Google Scholar]
  • 21.Toh SY, Citartan M, Gopinath SCB, Tang TH. Biosensors & Bioelectronics. 2015;64:392–403. doi: 10.1016/j.bios.2014.09.026. [DOI] [PubMed] [Google Scholar]
  • 22.Drolet DW, MoonMcDermott L, Romig TS. Nature Biotechnology. 1996;14:1021–1025. doi: 10.1038/nbt0896-1021. [DOI] [PubMed] [Google Scholar]
  • 23.Nie J, Deng Y, Deng QP, Zhang DW, Zhou YL, Zhang XX. Talanta. 2013;106:309–314. doi: 10.1016/j.talanta.2012.11.018. [DOI] [PubMed] [Google Scholar]
  • 24.Nutiu R, Li YF. Journal of the American Chemical Society. 2003;125:4771–4778. doi: 10.1021/ja028962o. [DOI] [PubMed] [Google Scholar]
  • 25.Li D, Song S, Fan C. Accounts of Chemical Research. 2010;43:631–641. doi: 10.1021/ar900245u. [DOI] [PubMed] [Google Scholar]
  • 26.Dougherty CA, Cai WB, Hong H. Current Topics in Medicinal Chemistry. 2015;15:1138–1152. doi: 10.2174/1568026615666150413153400. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Hong H, Goel S, Zhang Y, Cai W. Current Medicinal Chemistry. 2011;18:4195–4205. doi: 10.2174/092986711797189691. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Chu TC, Shieh F, Lavery LA, Levy M, Richards-Kortum R, Korgel BA, Ellington AD. Biosensors & Bioelectronics. 2006;21:1859–1866. doi: 10.1016/j.bios.2005.12.015. [DOI] [PubMed] [Google Scholar]
  • 29.Cui ZQ, Ren Q, Wei HP, Chen Z, Deng JY, Zhang ZP, Zhang XE. Nanoscale. 2011;3:2454–2457. doi: 10.1039/c1nr10218d. [DOI] [PubMed] [Google Scholar]
  • 30.McNamara JO, Andrechek ER, Wang Y, Viles KD, Rempel RE, Gilboa E, Sullenger BA, Giangrande PH. Nature Biotechnology. 2006;24:1005–1015. doi: 10.1038/nbt1223. [DOI] [PubMed] [Google Scholar]
  • 31.Tan Y, Zhang XM. Progress in Biochemistry and Biophysics. 2012;39:410–415. [Google Scholar]
  • 32.Chen ZJ, Tai ZG, Gu FF, Hu CL, Zhu QG, Gao S. European Journal of Pharmaceutics and Biopharmaceutics. 2016;107:130–141. doi: 10.1016/j.ejpb.2016.07.007. [DOI] [PubMed] [Google Scholar]
  • 33.Liu LS, Lu XL, Zhao YX. Journal of Nanoscience and Nanotechnology. 2016;16:6611–6621. [Google Scholar]
  • 34.Wang ZH, Xia JF, Cai F, Zhang FF, Yang M, Bi S, Gui RJ, Li YH, Xia YZ. Colloids and Surfaces B-Biointerfaces. 2015;134:40–46. doi: 10.1016/j.colsurfb.2015.06.031. [DOI] [PubMed] [Google Scholar]
  • 35.Mattice CMC, DeRosa MC. Biodrugs. 2015;29:151–165. doi: 10.1007/s40259-015-0126-5. [DOI] [PubMed] [Google Scholar]
  • 36.Hanout M, Ferraz D, Ansari M, Maqsood N, Kherani S, Sepah YJ, Rajagopalan N, Ibrahim M, Do DV, Quan Dong N. Biomed Research International. 2013 doi: 10.1155/2013/830837. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Parashar A. Journal of Clinical and Diagnostic Research. 2016;10:BE1–BE6. doi: 10.7860/JCDR/2016/18712.7922. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Limi T, Lee SY, Yang J, Hwang SY, Ahn Y. Biochip Journal. 2017;11:109–115. [Google Scholar]
  • 39.Moon JM, Kim DM, Kim MH, Han JY, Jung DK, Shim YB. Biosensors & Bioelectronics. 2017;91:128–135. doi: 10.1016/j.bios.2016.12.038. [DOI] [PubMed] [Google Scholar]
  • 40.Jin SR, Ye ZZ, Wang YX, Ying YB. Scientific Reports. 2017;7:8. doi: 10.1038/srep43175. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Matharu Z, Patel D, Gao YD, Hague A, Zhou Q, Revzin A. Analytical Chemistry. 2014;86:8865–8872. doi: 10.1021/ac502383e. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Fu HY, Yang JH, Guo L, Nie JF, Yin QB, Zhang L, Zhang Y. Biosensors & Bioelectronics. 2017;96:194–200. doi: 10.1016/j.bios.2017.05.012. [DOI] [PubMed] [Google Scholar]
  • 43.Jiang P, He MY, Shen L, Shi AN, Liu ZH. Sensors and Actuators B-Chemical. 2017;239:319–324. [Google Scholar]
  • 44.Sathish S, Ricoult SG, Toda-Peters K, Shen AQ. Analyst. 2017;142:1772–1781. doi: 10.1039/c7an00273d. [DOI] [PubMed] [Google Scholar]
  • 45.Dhiman A, Kaira P, Bansal V, Bruno JG, Sharma TK. Sensors and Actuators B-Chemical. 2017;246:535–553. [Google Scholar]
  • 46.Duan N, Wu SJ, Dai SL, Gu HJ, Hao LL, Ye H, Wang ZP. Analyst. 2016;141:3942–3961. doi: 10.1039/c6an00952b. [DOI] [PubMed] [Google Scholar]
  • 47.Hasanzadeh M, Shadjou N, de la Guardia M. Trac-Trends in Analytical Chemistry. 2017;89:119–132. doi: 10.1016/j.trac.2017.07.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Kim YS, Raston NHA, Gu MB. Biosensors & Bioelectronics. 2016;76:2–19. doi: 10.1016/j.bios.2015.06.040. [DOI] [PubMed] [Google Scholar]
  • 49.Tuerk C, Gold L. Science. 1990;249:505–510. doi: 10.1126/science.2200121. [DOI] [PubMed] [Google Scholar]
  • 50.Ellington AD, Szostak JW. Nature. 1990;346:818–822. doi: 10.1038/346818a0. [DOI] [PubMed] [Google Scholar]
  • 51.Bruno JG, Carrillo MP, Phillips T, King B. In Vitro Cellular & Developmental Biology-Animal. 2008;44:63–72. doi: 10.1007/s11626-008-9086-0. [DOI] [PubMed] [Google Scholar]
  • 52.Kowalska E, Bartnicki F, Pels K, Strzalka W. Analytical and Bioanalytical Chemistry. 2014;406:5495–5499. doi: 10.1007/s00216-014-7937-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Latulippe DR, Szeto K, Ozer A, Duarte FM, Kelly CV, Pagano JM, White BS, Shalloway D, Lis JT, Craighead HG. Analytical Chemistry. 2013;85:3417–3424. doi: 10.1021/ac400105e. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Kim CH, Lee LP, Min JR, Lim MW, Jeong SH. Biosensors & Bioelectronics. 2014;51:426–430. doi: 10.1016/j.bios.2013.08.003. [DOI] [PubMed] [Google Scholar]
  • 55.Mudili V, Makam SS, Sundararaj N, Siddaiah C, Gupta VK, Rao PVL. Scientific Reports. 2015;5:12. doi: 10.1038/srep15151. [DOI] [PMC free article] [PubMed] [Google Scholar] [Retracted]
  • 56.Yang DK, Chen LC, Lee MY, Hsu CH, Chen CS. Biosensors & Bioelectronics. 2014;62:106–112. doi: 10.1016/j.bios.2014.06.027. [DOI] [PubMed] [Google Scholar]
  • 57.Grozio A, Gonzalez VM, Millo E, Sturla L, Vigliarolo T, Bagnasco L, Guida L, D’Arrigo C, De Flora A, Salis A, Martin EM, Bellotti M, Zocchi E. Nucleic Acid Therapeutics. 2013;23:322–331. doi: 10.1089/nat.2013.0418. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Levine HA, Nilsen-Hamilton M. Computational Biology and Chemistry. 2007;31:11–35. doi: 10.1016/j.compbiolchem.2006.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Mendonsa SD, Bowser MT. Journal of the American Chemical Society. 2004;126:20–21. doi: 10.1021/ja037832s. [DOI] [PubMed] [Google Scholar]
  • 60.Mosing RK, Mendonsa SD, Bowser MT. Analytical Chemistry. 2005;77:6107–6112. doi: 10.1021/ac050836q. [DOI] [PubMed] [Google Scholar]
  • 61.Jing M, Bowser MT. Analytical Chemistry. 2013;85:10761–10770. doi: 10.1021/ac401875h. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Dong LL, Tan QW, Ye W, Liu DL, Chen HF, Hu HW, Wen D, Liu Y, Cao Y, Kang JW, Fan J, Guo W, Wu WZ. Scientific Reports. 2015;5:10. doi: 10.1038/srep15552. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Eaton RM, Shallcross JA, Mael LE, Mears KS, Minkoff L, Scoville DJ, Whelan RJ. Analytical and Bioanalytical Chemistry. 2015;407:6965–6973. doi: 10.1007/s00216-015-8665-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Mendonsa SD, Bowser MT. Journal of the American Chemical Society. 2005;127:9382–9383. doi: 10.1021/ja052406n. [DOI] [PubMed] [Google Scholar]
  • 65.Yang J, Bowser MT. Analytical Chemistry. 2013;85:1525–1530. doi: 10.1021/ac302721j. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Berezovski M, Drabovich A, Krylova SM, Musheev M, Okhonin V, Petrov A, Krylov SN. Journal of the American Chemical Society. 2005;127:3165–3171. doi: 10.1021/ja042394q. [DOI] [PubMed] [Google Scholar]
  • 67.Berezovski M, Musheev M, Drabovich A, Krylov SN. Journal of the American Chemical Society. 2006;128:1410–1411. doi: 10.1021/ja056943j. [DOI] [PubMed] [Google Scholar]
  • 68.Tok J, Lai J, Leung T, Li SFY. Electrophoresis. 2010;31:2055–2062. doi: 10.1002/elps.200900543. [DOI] [PubMed] [Google Scholar]
  • 69.Ashley J, Ji KL, Li SFY. Electrophoresis. 2015;36:2616–2621. doi: 10.1002/elps.201400540. [DOI] [PubMed] [Google Scholar]
  • 70.Jing M, Bowser MT. Lab on a Chip. 2011;11:3703–3709. doi: 10.1039/c1lc20461k. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Bruno JG. Biochemical and Biophysical Research Communications. 1997;234:117–120. doi: 10.1006/bbrc.1997.6517. [DOI] [PubMed] [Google Scholar]
  • 72.Xi ZJ, Huang RR, Li ZY, He NY, Wang T, Su EB, Deng Y. ACS Applied Materials & Interfaces. 2015;7:11215–11223. doi: 10.1021/acsami.5b01180. [DOI] [PubMed] [Google Scholar]
  • 73.Lou XH, Qian JR, Xiao Y, Viel L, Gerdon AE, Lagally ET, Atzberger P, Tarasow TM, Heeger AJ, Soh HT. Proceedings of the National Academy of Sciences of the United States of America. 2009;106:2989–2994. doi: 10.1073/pnas.0813135106. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74.Qian JR, Lou XH, Zhang YT, Xiao Y, Soh HT. Analytical Chemistry. 2009;81:5490–5495. doi: 10.1021/ac900759k. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75.Cho M, Xiao Y, Nie J, Stewart R, Csordas AT, Oh SS, Thomson JA, Soh HT. Proceedings of the National Academy of Sciences of the United States of America. 2010;107:15373–15378. doi: 10.1073/pnas.1009331107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Csordas AT, Jorgensen A, Wang JY, Gruber E, Gong Q, Bagley ER, Nakamoto MA, Eisenstein M, Soh HT. Analytical Chemistry. 2016;88:10842–10847. doi: 10.1021/acs.analchem.6b03450. [DOI] [PubMed] [Google Scholar]
  • 77.Oh SS, Ahmad KM, Cho M, Kim S, Xiao Y, Soh HT. Analytical Chemistry. 2011;83:6883–6889. doi: 10.1021/ac201269f. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 78.Oh SS, Plakos K, Lou XH, Xiao Y, Soh HT. Proceedings of the National Academy of Sciences of the United States of America. 2010;107:14053–14058. doi: 10.1073/pnas.1009172107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79.Hong SL, Wan YT, Tang M, Pang DW, Zhang ZL. Analytical Chemistry. 2017;89:6535–6542. doi: 10.1021/acs.analchem.7b00684. [DOI] [PubMed] [Google Scholar]
  • 80.Wang Q, Liu W, Xing YQ, Yang XH, Wang KM, Jiang R, Wang P, Zhao Q. Analytical Chemistry. 2014;86:6572–6579. doi: 10.1021/ac501088q. [DOI] [PubMed] [Google Scholar]
  • 81.Liu XH, Li H, Jia WC, Chen Z, Xu DK. Lab on a Chip. 2017;17:178–185. doi: 10.1039/c6lc01208f. [DOI] [PubMed] [Google Scholar]
  • 82.Park JW, Lee SJ, Ren S, Lee S, Kim S, Laurell T. Scientific Reports. 2016;6:9. doi: 10.1038/srep27121. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83.Avnir D, Coradin T, Lev O, Livage J. Journal of Materials Chemistry. 2006;16:1013–1030. [Google Scholar]
  • 84.Bhatia RB, Brinker CJ, Gupta AK, Singh AK. Chemistry of Materials. 2000;12:2434–2441. [Google Scholar]
  • 85.Nguyen DT, Smit M, Dunn B, Zink JI. Chemistry of Materials. 2002;14:4300–4306. [Google Scholar]
  • 86.Jin W, Brennan JD. Analytica Chimica Acta. 2002;461:1–36. [Google Scholar]
  • 87.Park SM, Ahn JY, Jo M, Lee DK, Lis JT, Craighead HG, Kim S. Lab on a Chip. 2009;9:1206–1212. doi: 10.1039/b814993c. [DOI] [PubMed] [Google Scholar]
  • 88.Ahn JY, Jo M, Dua P, Lee DK, Kim S. Oligonucleotides. 2011;21:93–100. doi: 10.1089/oli.2010.0263. [DOI] [PubMed] [Google Scholar]
  • 89.Lee S, Kang J, Ren S, Laurell T, Kim S, Jeong OC. Biochip Journal. 2013;7:38–45. [Google Scholar]
  • 90.Bae H, Ren S, Kang J, Kim M, Jiang Y, Jin MM, Min IM, Kim S. Nucleic Acid Therapeutics. 2013;23:443–449. doi: 10.1089/nat.2013.0437. [DOI] [PubMed] [Google Scholar]
  • 91.Ahn JY, Lee S, Jo M, Kang J, Kim E, Jeong OC, Laurell T, Kim S. Analytical Chemistry. 2012;84:2647–2653. doi: 10.1021/ac202559w. [DOI] [PubMed] [Google Scholar]
  • 92.Hybarger G, Bynum J, Williams RF, Valdes JJ, Chambers JP. Analytical and Bioanalytical Chemistry. 2006;384:191–198. doi: 10.1007/s00216-005-0089-3. [DOI] [PubMed] [Google Scholar]
  • 93.Huang CJ, Lin HI, Shiesh SC, Lee GB. Biosensors & Bioelectronics. 2010;25:1761–1766. doi: 10.1016/j.bios.2009.12.029. [DOI] [PubMed] [Google Scholar]
  • 94.Huang CJ, Lin HI, Shiesh SC, Lee GB. Biosensors & Bioelectronics. 2012;35:50–55. doi: 10.1016/j.bios.2012.02.024. [DOI] [PubMed] [Google Scholar]
  • 95.Lai HC, Wang CH, Liou TM, Lee GB. Lab on a Chip. 2014;14:2002–2013. doi: 10.1039/c4lc00187g. [DOI] [PubMed] [Google Scholar]
  • 96.Hung LY, Wang CH, Hsu KF, Chou CY, Lee GB. Lab on a Chip. 2014;14:4017–4028. doi: 10.1039/c4lc00587b. [DOI] [PubMed] [Google Scholar]
  • 97.Weng CH, Hsieh IS, Hung LY, Lin HI, Shiesh SC, Chen YL, Lee GB. Microfluidics and Nanofluidics. 2013;14:753–765. [Google Scholar]
  • 98.Kim J, Olsen TR, Zhu J, Hilton JP, Yang KA, Pei R, Stojanovic MN, Lin Q. Scientific Reports. 2016;6:10. doi: 10.1038/srep26139. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 99.Olsen T, Zhu J, Kim J, Pei RJ, Stojanovic MN, Lin Q. Slas Technology. 2017;22:63–72. doi: 10.1177/2211068216659255. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 100.Berezhnoy A, Stewart CA, McNamara JO, Thiel W, Giangrande P, Trinchieri G, Gilboa E. Molecular Therapy. 2012;20:1242–1250. doi: 10.1038/mt.2012.18. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 101.Song YL, Shi YZ, Li XR, Ma YL, Gao MX, Liu D, Mao Y, Zhu Z, Lin H, Yang CY. Analytical Chemistry. 2016;88:8294–8301. doi: 10.1021/acs.analchem.6b02140. [DOI] [PubMed] [Google Scholar]
  • 102.Bumgarner R. Curr Protoc Mol Biol. 2013;22 doi: 10.1002/0471142727.mb2201s101. 22.21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 103.Wang L, Li PCH. Analytica Chimica Acta. 2011;687:12–27. doi: 10.1016/j.aca.2010.11.056. [DOI] [PubMed] [Google Scholar]
  • 104.Katilius E, Flores C, Woodbury NW. Nucleic Acids Research. 2007;35:7626–7635. doi: 10.1093/nar/gkm922. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 105.Fischer NO, Tok JBH, Tarasow TM. Plos One. 2008;3:9. doi: 10.1371/journal.pone.0002720. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 106.Tome JM, Ozer A, Pagano JM, Gheba D, Schroth GP, Lis JT. Nature Methods. 2014;11:683. doi: 10.1038/nmeth.2970. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 107.Kinghorn AB, Dirkzwager RM, Liang SL, Cheung YW, Fraser LA, Shiu SCC, Tang MSL, Tanner JA. Analytical Chemistry. 2016;88:6981–6985. doi: 10.1021/acs.analchem.6b01635. [DOI] [PubMed] [Google Scholar]
  • 108.Cho M, Oh SS, Nie J, Stewart R, Eisenstein M, Chambers J, Marth JD, Walker F, Thomson JA, Soh HT. Proceedings of the National Academy of Sciences of the United States of America. 2013;110:18460–18465. doi: 10.1073/pnas.1315866110. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 109.Cho M, Oh SS, Nie J, Stewart R, Radeke MJ, Eisenstein M, Coffey PJ, Thomson JA, Soh HT. Analytical Chemistry. 2015;87:821–828. doi: 10.1021/ac504076k. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES