Electrocardiogram Delineation in a Wistar Rat Experimental Model

Pedro David Arini 1,2, Sergio Liberczuk 2,3,, Javier Gustavo Mendieta 2,4, Martín Santa María 5, Guillermo Claudio Bertrán 5
PMCID: PMC5822908  PMID: 29593828

Abstract

Background and Objectives

The extensive use of electrocardiogram (ECG) recordings during experimental protocols using small rodents requires an automatic delineation technique in the ECG with high performance. It has been shown that the wavelet transform (WT) based ECG delineator is a suitable tool to delineate electrocardiographic waveforms. The aim of this work is to implement and evaluate the ECG waves delineation in Wistar rats applying WT. We also describe the ECG signal of the Wistar rats giving the characteristics of its spectrum among other useful information.

Methods

We evaluated a delineator based on WT in a Wistar rat electrocardiograms database which was annotated manually by experienced observers.

Results

The delineation showed an “overall performance” such as sensitivity and a positive predictive value of 99.2% and 83.9% for P-wave, 100% and 99.9% for QRS complex, and 100% and 99.8% for T-wave, respectively. We also compared temporal analysis based ECG delineator with the WT based ECG delineator in RR interval, QRS duration, QT interval, and T-wave peak-to-end duration. The results showed that WT outperforms the temporal delineation technique in all parameters analyzed.

Conclusions

Finally, we propose a WT based ECG delineator as a methodology to implement in a wide diversity of experimental ECG analyses using Wistar rats.

1. Introduction

Electrocardiograms (ECGs) are broadly used for diagnosing many cardiac pathologies as well as for researching on experimental animals. Since physiological information is present in ECG intervals and amplitudes, it is essential to develop delineation algorithms to obtain automatic characteristic wave peaks and boundary ECG points. Indeed, the performance of the ECG analysis depends on the accuracy of the technique employed in the electrocardiographic signal delineation. Moreover, ECG data emerging from long-term recordings, for example, Holter in humans or experimental protocols in animals, demand automatic ECG delineation with a good performance. ECG signal presents patterns with different frequency contents that repeat cyclically beat-to-beat. Typically, in the ECG delineation process, the algorithm firstly detects the QRS complex [1], which is the most pronounced wave of the heartbeat. Lastly, the delineator locates the onset, the peak, and the end of the P-wave, the QRS complex, and the T-wave.

There are several completely automatic methods for the delineation of ECG waveforms used in the literature, which are based on the phasor transform [2], mathematical models [3, 4], Hilbert transform [5], correlation analysis [6], template matching [7], K-nearest neighbor [8], artificial neural networks [911], hidden Markov model, hybrid hidden Markov models combined with the wavelet transform [12, 13], wavelet transform (WT) [14, 15], genetic algorithms, fuzzy logic methods, support vector machines, self-organizing maps, and Bayesian techniques [16], with each approach exhibiting its own advantages and disadvantages. Although various techniques have been considered, most of the studies presented in the literature use the wavelet transform technique, and researchers claim that this is the best method for extracting features from the ECG signal [11, 17, 18]. Moreover, analyzing several techniques mentioned previously [28], it can be observed that the WT technique presents the best results in terms of robustness and noise immunity [14]. This is because the WT method allows information extraction from both frequency and time domains, different from what is usually achieved by the traditional Fourier transform which permits the analysis of only the frequency domain. This way, the ECG can be described at different scales of temporal and frequency resolution, and thus high frequency waves (such as the QRS complex) can be distinguished from low frequency waves (such as P- and T-waves). Moreover, the WT can be easily implemented with a cascade of Finite-Impulse Response (FIR) filters [19]. In summary, the wavelet transform based ECG delineator (WTD) is a suitable tool to detect waveforms (such as mono- or biphasic P-wave, QRS complex, and mono- or biphasic T-wave) and is useful in different applications. Numerous methods, based on the wavelet transform ECG delineator technique proposed by Martínez et al., have been proposed [1424].

As part of several research topics, a wide diversity of works utilized ECG analysis in small rodents. The rat and mouse models for research offer advantages with respect to larger animals such as lower cost, less physiological variability, and the possibility of using transgenic models. Normann et al. [25] analyzed ECG changes in rats during myocardial necrosis and infarction, both cases induced experimentally. Sgoifo et al. [26] showed ECG responses to different acute stressors in healthy rats, computing the heart rate variability (HRV) scores. Previously, they have implemented a software package developed to measure RR intervals and to interactively detect rhythm disturbances [27]. Also, telemetry systems have been widely used in rat experiments to analyze the relationship between ECG characteristics and behavioral responses during and after environmental challenges [28, 29]. Kuwahara et al. studied power spectral analysis of HRV in rats to assess autonomic nervous activity [30]. They developed an offline ECG analysis in order to detect R-waves and calculate the HRV. On the other hand, one investigation revealed that the stability of HRV estimators, for both time and frequency domain analysis in unrestrained rats, was highly dependent on the average heart rate (related to physical rat activity) as well as the length of the ECG records [31]. Moreover, Opitz et al. showed evidence that reperfusion can improve arrhythmia-related mortality by the prevention of ventricular fibrillation episodes [32]. They analyzed ECG offline in a semiautomatic fashion, as previously reported [33]. A recent work, making use of ECG biomarkers such as RR interval, QRS complex, and QT interval, showed that concomitant administration of trigonelline and sitagliptin produced a cardioprotective effect in contrast with monotherapy in diabetic rats [34]. Also, several toxicological studies showed the usefulness of ECG analysis in small rodents [35].

The aim of this work is to develop, validate, and provide a methodology for implementing WTD for Wistar rat experimental protocols. We offer here a robust and noise-immune tool to delineate all the characteristic points of the ECG and not only the one corresponding to the R-wave. Complete delineation of the ECG allows evaluating not only the HRV but also the PR interval and P-wave duration in order to analyze atria activation and QRS complex with the aim of evaluating ventricular depolarization and QT interval for measuring ventricular depolarization and repolarization jointly and T-wave peak-to-end interval to quantify transmural ventricular repolarization. Also, amplitudes of the P-, R-, S-, and T-waves are in general all altered in heart diseases.

In this sense, we have focused on the following: (1) study the ECG signal in Wistar rats and its corresponding power spectral density, (2) implement and evaluate WTD in an experimental model widely used in research, specifically Wistar rats, and (3) make a comparison between two ECG delineation systems in Wistar rats, such as WTD versus temporal analysis based ECG delineator (TAD). Finally, despite the existence of software that detects and analyzes ECG parameters for different species such as pigs, dogs, rabbits, guinea pigs, mice, and rats [34, 3638], to our knowledge, WTD applied to Wistar rats has not previously been described.

2. Materials and Methods

2.1. Wistar Rats ECG Database

In order to analyze the WTD algorithm in small rodent models, a Wistar rat database (WRDB) was developed. This database was acquired at the Institute of Medical Research “Alfredo Lanari” that belongs to the University of Buenos Aires, Argentina. Its contains a group of 57 Wistar rats, 29 males and 28 females, which were included in the study. When the measurements were carried out, this group contained 30 adult animals (424.5 ± 54.5 g) which were above 4 months of age and 27 young animals (245.5 ± 43.5 g) which were below 1 month of age. All these animals were in sinus rhythm at the time of ECG recordings.

After anesthesia with ketamine (75 mg/kg) and Rompun (0,75 mg/kg xylazine) administered subcutaneously and once the animal had been stabilized, the ECG signals were continuously recorded for at least 15 minutes in I-II, V1, V3, and V6 leads.

The ECG was acquired using EcoSur S.A. equipment (Buenos Aires, Argentina), digitized at 12-bit resolution with a sampling frequency of 1 kHz. These recordings were stored on a computer hard disk with custom-made software for posterior analysis.

In this work, we have only analyzed the delineation technique using lead II of the ECG, because it has shown robust P-, R-, S-, and T-waves. Lead II can be achieved in the Wistar rat by placement of the negative electrode near the right shoulder and the positive electrode to the left of the xyphoid space, in the same way as the Einthoven triangle (right arm position in the negative electrode and left leg position in the positive electrode).

Analysis of the ECG variables was carried out manually by two experienced observers (Ob#1 and Ob#2) using a computer calibrated cursor. Each ECG recording (57 in total) was annotated for each expert in 20 beats (1140 beats annotated), that is, a total of 2280 cardiac beats in order to validate WTD.

This process required manual measurement of 9120 characteristic ECG points by each expert observer in rat ECG measurement, that is, a total of 18240 electrocardiographic marks. The background noise level was 20 microV. To verify the reproducibility of the measurement methodology, 500 randomly selected recordings were analyzed by a third experienced observer (Ob#3). Animals were cared for according to Argentina's National Drug, Food and Medical Technology Administration Standards (Regulation 6344/96) and the National Institutes of Health Guide for the Care and Use of Laboratory Animals (NIH pub. number 85-23, revised 1996).

2.2. Temporal and Spectral Characteristics of the Wistar Rat ECG Signal

It is important to highlight that Q-wave does not exist in lead II WRECG [35], and hence we assumed the QRSon temporal position as the start of the ventricular depolarization process. We also observed some special electrocardiographic characteristics in WRECG, for example, as short QT interval, the absence of Q-wave in most ECG leads (in lead II, it does not exist), and the lack of ST segment. So, due to the absence of the ST segment in WRECG, we observed that the beginning of the T-wave was located in the same position as the end of the QRS complex.

Also, normal ventricular repolarization was characterized by a J-wave which was identified as the down-sloping portion of the ST segment and could be attributed to the heterogeneity of the ventricular wall, as was described by Antzelevitch and Fish [39], who explained this phenomenon by IKto-mediated current present in the epicardium, but not endocardium.

We calculated the power spectrum of WRECG signals using 300 beats from six animals. This diagram represented the QRS complex and P- and T-waves power spectrum separately (Figure 2(b)). The ECG lead II has been used in order to plot the ECG power spectrum in Wistar rats. As far as we know, no results exist regarding ECG spectrum in Wistar rats. While, in HECG, frequency components are approximately between 0.05 and 40 Hz, the range of WRECG goes from 4-5 to 120 Hz, as we can observe in Figure 2.

Figure 2.

Figure 2

Representative power spectrum of ECG in human beings (a) and ECG in Wistar rats (b). The location of spectral contents in the P-wave, the QRS complex, and the T-wave can be observed.

The spectral characteristic of a normal P-wave in humans is usually low frequency, below 10–15 Hz. Instead, most of the energy of the P-wave in rats is in the band between 20 and 60 Hz (see Figure 2). The most important frequency content of the QRS complex in humans is approximately between 5 and 25 Hz, while in WRECG, this energy is situated in the band between 30 and 80 Hz. Finally, the T-wave in humans is similar to a P-wave below 15 Hz approximately, but with higher amplitude than P-wave, while in Wistar rats the T-wave ranges from 5 to 80 Hz.

2.3. Wavelet Transform

The WT provides an alternative to the short time Fourier transform for the analysis of nonstationary signals, utilizing short analysis windows at high frequencies and long analysis windows at low frequencies. This can be thought of as the decomposition of a signal in a set of basis functions, obtained by means of dilation (a) and translation (b) of a single prototype function wavelet called ψ(t). The WT of a signal x(t) is defined as Waxb=1/a-+xtψt-b/adt, with a > 0. As the scale factor a increases, the wavelet becomes wider, giving information about lower frequency components of the signal, and upside down. The parameters a and b can be discretized according to a dyadic grid, being a = 2k and b = 2kl. The transform is then called dyadic wavelet transform, with basis functions ψk,l(t) = 2k/2ψ(2ktl) with k, lZ+.

All details of WTD have been extensively described in [14]; therefore, discrete time implementation is not explained in this work. The frequency responses of the first five scales were computed and obtained in the same way as Martínez et al. did [14].

However, it is important to highlight that the ECG delineation algorithm performs the detection of all characteristic points of the ECG waves using a quadratic spline WT, which, at scale 2k, is proportional to the derivative of the filtered version of the input ECG signal with a smoothing function at the same scale 2k.

In this sense, the zero crossings of the WT correspond to the minimum or maximum of the smoothed ECG signal at different k scales, and the maximum absolute values of the WT are related to maximum slopes in the smoothed cardiac electric signal. The local maximum, minimum, and zero crossings were used at different scales (21 to 25) in order to identify the characteristic points of the ECG signals.

2.4. WT Based Delineation Technique

In this section, we explain the delineation algorithms applied to Wistar rat electrocardiograms (WRECG).

Because a priori we did not know the WRECG spectrum characteristics, we took the greatest amount of information that we could get, using a sampling rate of 1 kHz. Our system has been based on [14, 15]. There, the sampling frequency of HECG signals was 250 Hz. So, as Martínez et al. indicated in [14], we would normally have to redesign the filters according to the new sampling frequency and rescale the filter banks. But as we have mentioned before, WRECG spectrum bandwidth changed with respect to HECG spectrum and this balanced the cutoff frequencies in the WT scales, so that we did not need to change filter bank cutoff frequencies as described below.

The WT scales were compressed almost four times when the sample rate increased in the same order, producing a relative increase in the cutoff frequencies of each filter. That is, each curve belonging to a filter or a WT scale remained fixed (in the absolute sense of the equivalent frequency response) but changed its cutoff frequency in a relative way due to the change produced in the sampling frequency.

On the other hand, the input signal changed its bandwidth as it has been mentioned before. The WRECG signal did not have the same spectrum as HECG signal and it changed its maximum frequency, which increased nearly 3 times from 40 Hz to 120 Hz, compensating this way its frequency situation regarding the filters or wavelets scales which remained constant (in absolute terms). The compensatory effect kept the entire signal spectrum covered by the same number of filters (or scales) designed previously for validation when the ECG human sampling rate was four times lower. Finally, the maximum frequency spectrum of the signal in WRECG (120 Hz) coincided approximately with the cutoff frequency of a particular filter bank or WT scale (k = 2).

2.4.1. QRS-Complex Detection

We built a vector containing maximum and minimum values above a certain threshold of scale 2 in WRECG and afterward a window of 15 ms in steps of 2 ms was shifted. Whenever a maximum following a lower maximum in that window is found, we call npre to the first maximum and npost to the second maximum. Then (in the modulus of scale 2) the minimum or lowest value between npre and npost was considered as the R-wave position in WRECG (see Figure 1(a)).

Figure 1.

Figure 1

ECG beats delineation from WRDB. In (a), we can observe the QRS complex with its WT at scales 22 and 23 and the peak and QRS boundaries obtained by the algorithm proposed. The npre and post positions were located in scale 22, while nfirst was located in scale 23. In (b), we can see the P-wave and T-wave with their WT scales 24 and 25 and marks of peaks, onset, and end of characteristics ECG points. The npost, last, and minT positions were located in scale 24, while nlast was located in scale 25.

2.4.2. QRS-Complex Delineation

The algorithm started from the R-wave previously detected, which had to be flanked by a pair of maximum moduli at |W22x(n)| in WRECG that were called npre and npost (see Figure 1(a)).

From the position of maximum value npost, we have searched to the right within a window size of 15 ms until a maximum was found in |W23x(n)| (see Figure 1(a)); this maximum corresponds to the end of the S-wave which also corresponds to the end of the QRS complex, so-called QRSend.

For the onset of the QRS complex, we search in |W23x(n)| from the R-wave previously detected but to the left within a window of 15 ms long until a maximum was found; this point was called nfirst. Note that the Q-wave is not present in the rat signal, so from that position we searched a new point that conforms to a lower value with respect to the following threshold:

ξQRSon=0.35W23xnfirst. (1)

2.4.3. T-Wave Delineation

For each R-wave position previously detected, we applied an analogous procedure to that applied in the QRS-complex boundaries detection. This way, we found the maximum after that corresponding R-wave, but this time in the scale |W24x(n)| we call it npost (see Figure 1(b)). Note that, due to the absence of the ST segment in WRECG, we observed that the beginning of the T-wave was located in the same position as the end of the QRS complex. Starting from this npost, we searched to the right for a maximum within a window size of 100 ms. Once this maximum was detected, we searched to the right in another window size of 100 ms to find a minimum; we call it MinT (see Figure 1(b)). The peak location of the T-wave had to be 2 ms before MinT. From the location of the aforementioned minimum, we searched the following maximum so-called nlast. Finally, having found that point, nlast, we searched in a new window of 100 ms until we found a point in |W25x(n)| (see Figure 1(b)) that was lower than the following threshold:

ξTend=0.45W25xnlast. (2)

This value corresponds to the T-wave end position.

2.4.4. P-Wave Delineation

The algorithm of P-wave delineation was analogous to the T-wave algorithm previously detailed (see Figure 1(b)). In this case, we searched to the left of the R-wave. The search windows to detect maximums and minimums were 40 ms of maximum length and the thresholds to detect the onset and offset of the P-wave were

ξPon=0.6W24xnfirst,ξPend=0.9W24xnlast. (3)

2.5. TA Based Delineation Technique in WRECG

In order to extend the WRDB analysis, we compared the WTD technique versus the TAD technique [40] using clinical ECG parameters such as RR interval, QRS duration, QT interval duration, and T-wave peak-to-end duration. Below, we briefly explained the TAD technique as was described in the software LabChart 7 Pro v7.3.1, ADInstruments, Sydney, Australia [40].

The R-wave was identified as the most positive value in the neighborhood of the cardiac beat selected. QRSon and QRSend were determined by searches on each side of the R-wave for regions where the slope, computed as dV/dt, decreased to sufficiently low values. Moreover, the isoelectric level was computed as the median of all data values preceding QRSon. In this sense, Ppeak was the point of the greatest absolute deviation from the isoelectric level, in an interval from pre-P-wave to just before QRSon. The P-wave was detected only if the peak exceeded a measure of uncertainty (the mean absolute deviation) in the isoelectric value. Pon was determined from a straight line fitted by least squares to points preceding Ppeak (those points in the range 15–60% of the peak were used). The intersection of this line with the isoelectric level was Pon; the search Pend was analogous to Pon.

With respect to the T-wave delineation, searching was carried out for the first significant peak of either sign, starting from a point after QRSend. When T-wave was selected, the starting point was close to QRSend. If it was not selected, the starting point was further to the right. If a suitable peak was found, a straight line was fitted by least squares to the tail of the wave, over a data range 70–30% of Tpeak. Finally, Tend was found as the intersection of this line with the isoelectric level.

2.6. Validation and Evaluation of WTD

The evaluation in WRDB was done between the reference ECG annotations provided by two experienced observers (gold standard) and WTD implemented. We calculated the sensitivity as Se = Tp/(Tp + Fn) and the positive predictive value as P+ = Tp/(Tp + Fp). Tp (true positive) detection was considered when the ECG characteristic point was annotated and the algorithm detects its presence within a window < 25 ms in WRDB. Fn (false negative) detection was computed when the delineation algorithm located an ECG characteristic point which was not annotated by the expert. Lastly, a Fp (false positive) was calculated as the number of misdetections (the algorithm fails to locate the annotated characteristic ECG point within the abovementioned tolerance).

When there was no annotation of the ECG characteristic point, it was not possible to know whether the expert decided that there were no ECG characteristic points to mark or simply was not confident to mark it [14, 23]. Whatever the reason was, the P+ value was computed under the assumption that each absent mark meant that there was no waveform. Therefore, the computed P+ was interpreted as a lower limit so-called Pmin+ (see Table 1).

Table 1.

Performance of wavelet transform based ECG delineator applied to Wistar rats ECG database. We can observe the results of sensitivity and positive predictive value of this work in comparison with both experienced observers (Ob#1 and Ob#2). The analysis of the ECG variables was carried out manually by two experienced observers using a computer calibrated cursor. Also, interobserver (Ob#1 versus Ob#2) annotation differences mean and variability are presented. Moreover, delineation of WRDB showed an overall performance that was computed as a mean value between the characteristic points for each electrocardiographic wave, such as Se and P+ of 99.2% and 83.9% for P-wave, 100% and 99.9% for QRS complex, and 100% and 99.8% for T-wave, respectively.

Param. Pon Ppeak Pend QRSon   Rpeak QRSend Tpeak Tend
This work versus Ob#1 Se (%) 99.3 99.3 99.3 100 100 100 100 100
P + (%) 82.1 82.1 82.1 100 100 100 100 100
Me ± SD (ms) 2.59 ± 3.99 0.33 ± 3.93 0.47 ± 3.86 0.66 ± 1.72 −0.04 ± 1.4 0.03 ± 2.15 1.72 ± 3.05 2.40 ± 11

This work versus Ob#2 Se (%) 99.1 99.1 99.07 100 100 100 100 100
P + (%) 86.65 87.06 87.27 99.8 100 99.6 99 100
Me ± SD (ms) 3.4 ± 5.58 0.17 ± 4.81 0.21 ± 4.61 0.03 ± 1.73 −0.21 ± 0.63 0.39 ± 1.86 1.59 ± 2.9 −1.75 ± 9.2

Me ± SD (Ob#1 versus Ob#2) (ms) 0.72 ± 2.38 −0.2 ± 1.31 −0.38 ± 1.93 −0.63 ± 2.16 −0.17 ± 1.45 0.36 ± 1.87 −0.14 ± 2.14 −4.1 ± 9.58

Overall Se (%) 99.2 100 100
Performance P + (%) 83.9 99.9 99.8

The mean valueMe of the errors was computed as the temporal position difference between reference annotation and automatic delineation, and the average standard deviationSD of the error was computed by averaging the intra-ECG recording standard deviation (see Table 1).

2.7. Statistical Analysis in Comparison with Temporal Technique

Firstly, D'Agostino-Pearson's normality test was applied with the aim of quantifying the discrepancy between the ECG indices (RR interval, QRS duration, QT interval, and TPE duration) and ideal Gaussian distributions for each case. Secondly, because distribution variables were non-Gaussian, we applied the two-sided Mann–Whitney U test in unpaired samples and the Wilcoxon signed rank test in paired samples. Finally, in order to determine the statistical significance of ECG indices between temporal analysis and wavelet transform based ECG delineators, nonparametric tests were applied. When the p value was <0.05, differences were considered statistically significant.

3. Results

Figure 1 shows representative ECG beats obtained from WRDB. In (a), we can observe one beat with its QRSon, Rpeak, and QRSend points detected, and their WT at scales 22 and 23 used for that purpose. Likewise, Figure 1(b) shows one beat with its Pon, Ppeak, Pend, Tpeak, and Tend characteristic points detected, and their WT at scales 24 and 25 utilized during the delineation process.

Figure 2 presents the HECG (a) and WRECG (b) power spectrum characteristics. This diagram serves as a rough guide of where to locate the spectral components despite variations between beats of different leads, origins, and subjects.

The results obtained in WRDB are given in Table 1; we observe the “overall performance” of WTD, which has been calculated as a mean representative value of Se and P+ in each electrocardiographic wave, that is, one unique value for P-wave, T-wave, and QRS-complex, respectively (see Table 1). Also, in this table, we show interobserver (Ob#1 versus Ob#2) annotation differences expressed as mean and SD, which were calculated over the same 1140 cardiac beats selected for manual measurement.

The purpose of Figure 3 is to compare two delineation techniques, namely, temporal analysis based ECG delineator (TAD) and wavelet transform based ECG delineator (WTD), for measuring ECG parameters with clinical utility, such as RR interval, QRS duration, QT interval duration, and T-wave peak-to-end duration. Box and whiskers plots of the mean value and the standard deviation of both WTD (Section 2.4) and TAD (Section 2.5) techniques are reported.

Figure 3.

Figure 3

Box and whisker plots showing the mean (a) and standard deviation (b) values for different ECG parameters in Wistar rats. The ECG parameters measured were the RR interval (RR), QRS duration (QRS), QT interval duration (QT), and T-wave peak-to-end duration (TPE). The temporal analysis based ECG delineator (TAD) was represented with clear boxes and the wavelet transform based ECG delineator (WTD) was represented with dark boxes. p < 0.05, p < 0.001, and p < 0.0001 indicate statistically significant differences of ECG parameters between TAD and WTD. NS: nonstatistically significant differences between both groups.

4. Discussion

The ECG waves delineation in Wistar rats provides fundamental features such as amplitude and duration of each cardiac beat. This information can be utilized in order to evaluate pathologies, behavior, and/or drug effects [35]. In this sense, it is necessary to achieve very good performance to detect characteristic waves, peaks, and boundaries of ECG points.

4.1. The WTD Performance in Wistar Rats

We have implemented WTD in signals obtained from WRDB, as we can see in Table 1. We observed very good sensitivity and positive predictive values with respect to both experienced observers (Ob#1 and Ob#2) in P-wave, QRS complex, T-wave delineation, and QRS detection. Taking into account the delineation of the onset, the peak, and the end, we observed that Se values were higher than 99% in the P-wave, equal to 100% in the QRS complex and T-wave, and likewise 100% in the QRS detection. Regarding the positive predictive value, it ranged from 82% up to 87% in the P-wave delineation and was approximately 100% in the QRS-complex and T-wave delineation. Also, in the QRS detection, it was approximately 100%. We did not observe statistically significant differences between the results presented by both observers (p = NS). In conclusion, we observed a high performance to delineate WRECG signals regardless of the observers, that is, Ob#1 or Ob#2. Also, in order to verify the reproducibility of the measurement methodology, 500 randomly selected ECG recordings were analyzed by a third observer (Ob#3). The interobserver differences between both Ob#1 versus Ob#3 and Ob#2 versus Ob#3 were not significant (NS).

The difference in performance between sensitivity and positive predicted value in the data is presented only in the detection of the P-wave, but not in the rest of the waves, and is due to different factors, such as the absence, low amplitude, M-shape, and/or biphasic morphology of the P-wave. This leads to having more false positives (Fp) than false negatives (Fn), which finally explains the difference in performance.

4.2. Comparison with Temporal Techniques

In order to analyze the delineation performance in WRECG, we compared WTD proposed in this work with the commercial software LabChart 7 Pro version 7.3.1 (ADInstruments, Sydney, Australia, [40]); the latter was based on  TAD technique. Electrocardiographic parameters such as RR interval, QRS duration, QT interval duration, and T-wave peak-to-end duration were computed from both methods named above using frontal ECG lead II.

In Figure 3, we can observe box and whisker plots of the mean (a) and standard deviation (b) of the mean values of 57 Wistar rats (40 beats processed for each animal) for relevant ECG parameters.

When we compared the WTD technique with the TAD technique, we observed that our algorithm outperforms clearly the temporal technique when the standard deviation values were computed. Moreover, statistically significant differences between both delineation techniques can be observed.

4.3. ECG Variables in Normal and Anesthetized WRs

Recently, the ECG time intervals (registered in lead II) have been presented in unanesthetized WRs [41]. Representative values, the mean, and the range from the minimum to the maximum values included between square brackets were shown in [41] as seen below: HR 460 [228–600] bpm, P 13 [10–16] ms, PR 40 [33–50] ms, QRS 17 [12–26], and QT 66 [38–80] ms.

In order to compare our results with [41], we have computed the average values mean ± SD of the animals used in the present work, which are detailed below: HR 251 ± 22 bpm, P-wave 24.5 ± 2.0 ms, PR interval 54.7 ± 5.3 ms, QRS complex 17.9 ± 1.6 ms, and QT interval 83.3 ± 4.0 ms. It can be observed that the major discrepancy was in the P-wave definition. Moreover, representative amplitude values, mean, and range were reported in [41] as seen below: P 110 [20–200] uV, Q 0 uV, R 1060 [220–1500] uV, S 200 [0–500] uV, and T 150 [50–300] uV. Also, we have calculated the average values mean ± SD, which are presented below: P-wave 93.8 ± 40.6 uV, Q-wave 0 uV, R-wave 610.8 ± 140.1 uV, S-wave −385.2 ± 152.3 uV, and T-wave 163.8 ± 58.4 uV. The values obtained in this work do not present discrepancy with unanesthetized WRs presented in [41]. We emphasize that normal variants presented in [41] cannot be considered abnormal. Finally, it has been concluded that the values of the ECG parameters in anesthetized WRs did not change rhythm or morphology in any way or affected the results of the delineation.

5. Conclusions

There are a lot of experimental works which have shown the utility of ECG in small rodents in order to discern the mechanisms of action of drugs, infer disturbances of the cardiac rhythm and electrical conduction, detect the presence of ischemic injury, and analyze behavioral responses, among others. In spite of being a useful tool, it is not sufficient as a standalone methodology to describe abnormal electrophysiological mechanisms. However, when combined with other electrophysiological techniques that can only be carried out in animal models, the ECG analysis may be used to obtain a more complete understanding of the electrical cardiac mechanism. Also, it is important to highlight that replacement of printed ECG with digitized ECG, as well as delineation algorithms with high performance values, is collectively streamlining ECG assessments in experimental protocols.

In the present work, we have shown and validated the capability to perform QRS detection and location of ECG wave boundaries using WTD in ECG Wistar rats. Also, the methodology implemented has provided reliable and accurate delineation in the ECG in comparison with the TA technique delineation. Finally, due to the importance of having a high performance in the ECG delineated stage, we propose the use of WTD during experimental protocols which use Wistar rats.

Acronyms

ECG:

Electrocardiogram

HECG:

Human being electrocardiograms

P+:

Positive predictive value

Se:

Sensitivity

TA:

Temporal analysis

TAD:

Temporal analysis based ECG delineator

WRDB:

Wistar rats database

WRECG:

Wistar rats electrocardiogram

WT:

Wavelet transform

WTD:

Wavelet transform based ECG delineator.

Conflicts of Interest

The authors declare that they have no conflicts of interest regarding the publication of this paper.

References

  • 1.Pan J., Tompkins W. J. A real-time QRS detection algorithm. IEEE Transactions on Biomedical Engineering. 1985;32(3):230–236. doi: 10.1109/TBME.1985.325532. [DOI] [PubMed] [Google Scholar]
  • 2.Martínez A., Alcaraz R., Rieta J. J. A new method for automatic delineation of ECG fiducial points based on the Phasor Transform. Proceedings of the 2010 32nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC'10; September 2010; Argentina. pp. 4586–4589. [DOI] [PubMed] [Google Scholar]
  • 3.Sörnmo L. A model-based approach to QRS delineation. Computers and Biomedical Research. 1987;20(6):526–542. doi: 10.1016/0010-4809(87)90024-3. [DOI] [PubMed] [Google Scholar]
  • 4.Sayadi O., Shamsollahi M. B. A model-based Bayesian framework for ECG beat segmentation. Physiological Measurement. 2009;30(3):335–352. doi: 10.1088/0967-3334/30/3/008. [DOI] [PubMed] [Google Scholar]
  • 5.Madeiro J. P., Cortez P. C., Marques J. A., et al. An innovative approach of qrs segmentation based on first-derivative, hilbert and wavelet transforms. Medical Engineering & Physics. 2012;34(9):1236–1246. doi: 10.1016/j.medengphy.2011.12.011. [DOI] [PubMed] [Google Scholar]
  • 6.Homaeinezhad M. R., ErfanianMoshiri-Nejad M., Naseri H. A correlation analysis-based detection and delineation of ECG characteristic events using template waveforms extracted by ensemble averaging of clustered heart cycles. Computers in Biology and Medicine. 2014;44(1):66–75. doi: 10.1016/j.compbiomed.2013.10.024. [DOI] [PubMed] [Google Scholar]
  • 7.Krasteva V., Jekova I. QRS template matching for recognition of ventricular ectopic beats. Annals of Biomedical Engineering. 2007;35(12):2065–2076. doi: 10.1007/s10439-007-9368-9. [DOI] [PubMed] [Google Scholar]
  • 8.Saini I., Singh D., Khosla A. QRS detection using K-Nearest Neighbor algorithm (KNN) and evaluation on standard ECG databases. Journal of Advanced Research. 2013;4(4):331–344. doi: 10.1016/j.jare.2012.05.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Dokur Z., Ölmez T., Yazgan E., Ersoy O. K. Detection of ECG waveforms by neural networks. Medical Engineering & Physics. 1997;19(8):738–741. doi: 10.1016/S1350-4533(97)00029-5. [DOI] [PubMed] [Google Scholar]
  • 10.Silipo R., Marches C. Artificial neural networks for automatic ECG analysis. IEEE Transactions on Signal Processing. 1998;46(5):1417–1425. doi: 10.1109/78.668803. [DOI] [Google Scholar]
  • 11.Güler I., Übeyli E. D. ECG beat classifier designed by combined neural network model. Pattern Recognition. 2005;38(2):199–208. doi: 10.1016/j.patcog.2004.06.009. [DOI] [Google Scholar]
  • 12.De Lannoy G., Frenay B., Verleysen M., Delbeke J. Supervised ECG delineation using the wavelet transform and hidden markov models. Proceedings of the 4th European Conference of the International Federation for Medical and Biological Engineering, ECIFMBE 2008; November 2008; Belgium. pp. 22–25. [DOI] [Google Scholar]
  • 13.Andreäo R. V., Boudy J. Combining wavelet transform and hidden Markov models for ECG segmentation. EURASIP Journal on Advances in Signal Processing. 2007;2007 doi: 10.1155/2007/56215.56215 [DOI] [Google Scholar]
  • 14.Martínez J. P., Almeida R., Olmos S., Rocha A. P., Laguna P. A wavelet-based ECG delineator: evaluation on standard databases. IEEE Transactions on Biomedical Engineering. 2004;51(4):570–581. doi: 10.1109/TBME.2003.821031. [DOI] [PubMed] [Google Scholar]
  • 15.Li C., Zheng C., Tai C. Detection of ECG Characteristic Points Using Wavelet Transforms. IEEE Transactions on Biomedical Engineering. 1995;42(1):21–28. doi: 10.1109/10.362922. [DOI] [PubMed] [Google Scholar]
  • 16.Karpagachelvi S., Arthanari M., Sivakumar M. Ecg feature extraction techniques-a survey approach. International Journal of Computer Science and Information Security. 2010;8(1) [Google Scholar]
  • 17.Kutlu Y., Kuntalp D. Feature extraction for ECG heartbeats using higher order statistics of WPD coefficients. Computer Methods and Programs in Biomedicine. 2012;105(3):257–267. doi: 10.1016/j.cmpb.2011.10.002. [DOI] [PubMed] [Google Scholar]
  • 18.Lin C.-H., Du Y.-C., Chen T. Adaptive wavelet network for multiple cardiac arrhythmias recognition. Expert Systems with Applications. 2008;34(4):2601–2611. doi: 10.1016/j.eswa.2007.05.008. [DOI] [Google Scholar]
  • 19.Ghaffari A., Homaeinezhad M. R., Khazraee M., Daevaeiha M. M. Segmentation of holter ECG waves via analysis of a discrete wavelet-derived multiple skewness-kurtosis based metric. Annals of Biomedical Engineering. 2010;38(4):1497–1510. doi: 10.1007/s10439-010-9919-3. [DOI] [PubMed] [Google Scholar]
  • 20.Bahoura M., Hassani M., Hubin M. DSP implementation of wavelet transform for real time ECG wave forms detection and heart rate analysis. Computer Methods and Programs in Biomedicine. 1997;52(1):35–44. doi: 10.1016/S0169-2607(97)01780-X. [DOI] [PubMed] [Google Scholar]
  • 21.Sivannarayana N., Reddy D. C. Biorthogonal wavelet transforms for ECG parameters estimation. Medical Engineering & Physics. 1999;21(3):167–174. doi: 10.1016/S1350-4533(99)00040-5. [DOI] [PubMed] [Google Scholar]
  • 22.Boichat N., Khaled N., Rinconz F., Atienza D. Wavelet-based ECG delineation on a wearable embedded sensor platform. Proceedings of the 2009 6th International Workshop on Wearable and Implantable Body Sensor Networks, BSN 2009; June 2009; USA. pp. 256–261. [DOI] [Google Scholar]
  • 23.Di Marco L. Y., Chiari L. A wavelet-based ECG delineation algorithm for 32-bit integer online processing. Biomedical Engineering Online. 2011;10, article no. 23 doi: 10.1186/1475-925X-10-23. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Ghaffari A., Homaeinezhad M. R., Akraminia M., Atarod M., Daevaeiha M. A robust wavelet-based multi-lead electrocardiogram delineation algorithm. Medical Engineering & Physics. 2009;31(10):1219–1227. doi: 10.1016/j.medengphy.2009.07.017. [DOI] [PubMed] [Google Scholar]
  • 25.Normann S. J., Priest R. E., Benditt E. P. Electrocardiogram in the normal rat and its alteration with experimental coronary occlusion. Circulation Research. 1961;9:282–287. doi: 10.1161/01.RES.9.2.282. [DOI] [PubMed] [Google Scholar]
  • 26.Sgoifo A., Stilli D., De Boer S. F., Koolhaas J. M., Musso E. Acute social stress and cardiac electrical activity in rats. Aggressive Behavior. 1998;24(4):287–296. doi: 10.1002/(SICI)1098-2337(1998)24:4&#x0003c;287::AID-AB4&#x0003e;3.0.CO;2-G. [DOI] [Google Scholar]
  • 27.Sgoifo A., Stilli D., Aimi B., Parmigiani S., Manghi M., Musso E. Behavioral and electrocardiographic responses to social stress in male rats. Physiology & Behavior. 1994;55(2):209–216. doi: 10.1016/0031-9384(94)90125-2. [DOI] [PubMed] [Google Scholar]
  • 28.Bohus B. Telemetered heart rate responses of the rat during free and learned behavior. Biotelemetry. 1974;1(4):193–201. [PubMed] [Google Scholar]
  • 29.Sgoifo A., Stilli D., Medici D., Gallo P., Aimi B., Musso E. Electrode positioning for reliable telemetry ECG recordings during social stress in unrestrained rats. Physiology & Behavior. 1996;60(6):1397–1401. doi: 10.1016/S0031-9384(96)00228-4. [DOI] [PubMed] [Google Scholar]
  • 30.Kuwahara M., Yayou K.-I., Ishii K., Hashimoto S.-I., Tsubone H., Sugano S. Power spectral analysis of heart rate variability as a new method for assessing autonomic activity in the rat. Journal of Electrocardiology. 1994;27(4):333–337. doi: 10.1016/S0022-0736(05)80272-9. [DOI] [PubMed] [Google Scholar]
  • 31.Couderc J.-P., Elder A., Cox C., Zareba W., Oberdorster G. Limitations of power-spectrum and time-domain analysis of heart rate variability in short-term ECG recorded using telemetry in unrestrained rats. Proceedings of the Computers in Cardiology 2002; September 2002; usa. pp. 589–592. [Google Scholar]
  • 32.Opitz C. F., Finn P. V., Pfeffer M. A., Mitchell G. F., Pfeffer J. M. Effects of reperfusion on arrhythmias and death after coronary artery occlusion in the rat: Increased electrical stability independent of myocardial salvage. Journal of the American College of Cardiology. 1998;32(1):261–267. doi: 10.1016/S0735-1097(98)00173-9. [DOI] [PubMed] [Google Scholar]
  • 33.Opitz C. F., Mitchell G. F., Pfeffer M. A., Pfeffer J. M. Arrhythmias and death after coronary artery occlusion in the rat: Continuous telemetric ECG monitoring in conscious, untethered rats. Circulation. 1995;92(2):253–261. doi: 10.1161/01.CIR.92.2.253. [DOI] [PubMed] [Google Scholar]
  • 34.Kamble H. V., Bodhankar S. L. Cardioprotective effect of concomitant administration of trigonelline and sitagliptin on cardiac biomarkers, lipid levels, electrocardiographic and heamodynamic modulation on cardiomyopathy in diabetic Wistar rats. Biomedicine & Aging Pathology. 2014;4(4):335–342. doi: 10.1016/j.biomag.2014.07.009. [DOI] [Google Scholar]
  • 35.Farraj A. K., Hazari M. S., Cascio W. E. The utility of the small rodent electrocardiogram in toxicology. Toxicological Sciences. 2011;121(1):11–30. doi: 10.1093/toxsci/kfr021. [DOI] [PubMed] [Google Scholar]
  • 36.Yakushev S., Band M., Tissot van Patot M. C., Gassmann M., Avivi A., Bogdanova A. Cross talk between S-nitrosylation and S-glutathionylation in control of the Na,K-ATPase regulation in hypoxic heart. American Journal of Physiology-Heart and Circulatory Physiology. 2012;303(11):H1332–H1343. doi: 10.1152/ajpheart.00145.2012. [DOI] [PubMed] [Google Scholar]
  • 37.Tomat A. L., Juriol L. V., Gobetto M. N., et al. Morphological and functional effects on cardiac tissue induced by moderate zinc deficiency during prenatal and postnatal life in male and female rats. American Journal of Physiology-Heart and Circulatory Physiology. 2013;305(11):H1574–H1583. doi: 10.1152/ajpheart.00578.2013. [DOI] [PubMed] [Google Scholar]
  • 38.Zajaczkowski S., Smolinska M., Badtke P., et al. Time-domain and spectral analysis of heart rate variability in rats challenged with hypoxia. Proceedings of the Computing in Cardiology Conference (CinC), 2014; 2014; Cambridge, Mass, USA. pp. 785–788. [Google Scholar]
  • 39.Antzelevitch C., Fish J. Electrical heterogeneity within the ventricular wall. Basic Research in Cardiology. 2001;96(6):517–527. doi: 10.1007/s003950170002. [DOI] [PubMed] [Google Scholar]
  • 40.LabChart 7 Pro. v7.3.1. 2011, http://www.ADInstruments.com/
  • 41.Macfarlane P. W., van Oosterom A. Comprehensive electrocardiography. 2011.