Abstract
Objectives
The aim of this pictorial review is to illustrate the use of CBCT in a broad spectrum of musculoskeletal disorders and to compare its diagnostic merit with other imaging modalities, such as conventional radiography (CR), Multidetector Computed Tomography (MDCT) and Magnetic Resonance Imaging.
Background
Cone Beam Computed Tomography (CBCT) has been widely used for dental imaging for over two decades.
Discussion
Current CBCT equipment allows use for imaging of various musculoskeletal applications. Because of its low cost and relatively low irradiation, CBCT may have an emergent role in making a more precise diagnosis, assessment of local extent and follow-up of fractures and dislocations of small bones and joints. Due to its exquisite high spatial resolution, CBCT in combination with arthrography may be the preferred technique for detection and local staging of cartilage lesions in small joints. Evaluation of degenerative joint disorders may be facilitated by CBCT compared to CR, particularly in those anatomical areas in which there is much superposition of adjacent bony structures. The use of CBCT in evaluation of osteomyelitis is restricted to detection of sequestrum formation in chronic osteomyelitis. Miscellaneous applications include assessment of (symptomatic) variants, detection and characterization of tumour and tumour-like conditions of bone.
Teaching Points
• Review the spectrum of MSK disorders in which CBCT may be complementary to other imaging techniques.
• Compare the advantages and drawbacks of CBCT compared to other imaging techniques.
• Define the present and future role of CBCT in musculoskeletal imaging.
Keywords: Cone beam computed tomography, Multidetector computed tomography, Conventional radiography, Magnetic resonance imaging, Musculoskeletal imaging
Introduction
Although initially used for dental imaging, Cone Beam Computed Tomography (CBCT) is currently installed in many radiology departments as an integral part of the imaging armamentarium. CBCT uses a conical x-ray beam which falls on a flat panel detector unlike conventional Multidetector Computed Tomography (MDCT), where a fan shaped beam and linear detectors are used (Fig. 1). In CBCT, the X-ray tube and the detector synchronously rotate 360° around the patient. At certain degree intervals, single projection images or “basis” images, are acquired. Software programs incorporating sophisticated algorithms including back-filtered projection are applied to these projection data to generate a volumetric data set, which can be used for reconstruction images in three orthogonal planes [1]. In our department, we use a CBCT with a gantry of 58 cm patient aperture and a movable table allowing horizontal positioning and multifunctional use (NewTom 5 G, QR systems, Verona, Italy). The specific purpose of this paper is to present a pictorial overview of the clinical usefulness of the CBCT of evaluation in a broad spectrum of musculoskeletal disorders and to compare its diagnostic merit with other imaging modalities, such as conventional radiography (CR), Multidetector Computed Tomography (MDCT) and Magnetic Resonance Imaging (MRI).
Advantages and disadvantages
A major advantage of CBCT is its high spatial resolution resulting in exquisite detail of bone microarchitecture (Fig. 2a) [2, 3]. CBCT after intra-articular contrast injection (CBCT arthrography; CBCT-A) offers high resolution images of the articular cartilage surface. (Fig. 2b) [4]. With our equipment, the spatial resolution ranges between 300 μm for a standard scan to 75 μm for high resolution images. Recent studies, performed on phantoms and also in a cohort of paediatric patients, confirm the significant lower dose of CBCT, compared to MDCT [5–11]. Effective dose for paranasal sinuses imaging in CBCT is approximately 40% lower than standard MDCT and 30% lower than low-dose sinus CT scans [12]. Studies on phantoms in the ankle region showed 21.4 μSv of effective dose for MDCT, for CBCT it was reported ranging from 1.9 μSv to14.3 μSv [11]. Lower radiation dose results from single rotation of the gantry required for acquisition of the whole scan volume, smaller field of view, pulsed X-ray beams instead of the constant radiation stream and the use of a large high quality flat panel detector [13].
Our NewTom 5 G equipment uses a fixed tube voltage of 110 kVp, but is fitted with the SafeBeam technology™, allowing reduction of the radiation dose.
Initial optimization of the tube current (mA) occurs on estimation of the patient size based on the attenuation information derived from an anteroposterior and lateral scout views [11].
Angular tube current modulation further equalizes in real time the photon flux to the detector as the X-ray tube rotates about the patient among the anteroposterior and lateral position according to the measured attenuation from the previous projection, allowing further adaptation of the dose for the anatomy of the patient [14, 15].
By using the pulsed emission technology, exposure is restricted to intermittent bursts of radiation for each degree instead of using a constant stream of radiation during the 360 ° rotation. This results in a considerable decrease of effective exposure time (e.g., if a 360 ° rotation lasts for 18 to 36 s, the effective exposure time during the 360° rotation is 2.4 to 7.3 s) [16]. This does not affect the overall image quality as the dataset obtained from 360 projections (one for each degree during 360° rotation) may be used for qualitative image reconstruction.
For small joints, the spectrum of the effective dose (ED) in CBCT ranges between 1 to 15.3 μSv applying a conversion factor of 0.01 mSv/Gy x cm2 for peripheral joints, which is significantly lower than values reported for MDCT. Among those imaging methods using radiation, Conventional Radiography (CR) still remains the one with the lowest ED, between 0.07 to 5 μSv) [17].
However, image quality of CT and CBCT images may be altered by implanted metal elements reducing the contrast, obscuring structures and impairing the detection of areas of interest. This image degradation may be reduced by using dedicated algorithms and software [18, 19]. The overall cost of the equipment is far less than MDCT [20], thus it is suited for private practices or small medical centres or as additional CT equipment in large institutions. Table 1 provides a short overview of the parameters that may influence the economic rentability of installing a CBCT. However, as the referral pattern for certain examinations may differ among different health care centres and reimbursement can significantly differ depending on the health policy of each country, it is not possible to provide general recommendations.
Table 1.
CBCT (high-end) | MDCT (mid-end) | |
---|---|---|
Purchase price equipment* | 21–30% (200,000 Euro) | (665,000–968,000 Euro) |
Annual maintenance service* | 10% (10,000 Euro) | (100,000 Euro) |
Required area for placement of equipment + operating space for medical staff (minimum versus our hospital) | 16.5 m2–33 m2 | 36 m2–52 m2 |
Investment cost for room preparation | 100,000 Euro | 150,000 Euro |
Operating cost (electricity, other utilities) | 50% | 100% |
Cost medical staff | Similar (10 min acquisition and reconstruction time) | Similar (10 min acquisition and reconstruction time) |
Honorarium examination | Similar (e.g. 50.90 Euro for MSK examination)** | Similar (e.g. 50.90 Euro for MSK) |
Other applications than small bones/joints | Dental, petrous bone, sinuses | Brain, Spine, Abdomen, Chest, Bone and Joints (including large joints), CT-angiography, … |
*Based on list price provided by the manufacturer of our high-end CBCT equipment versus a MDCT mid-end equipment (range of different manufacturers)
**Currently pending approval of the National Institute for Sickness and Invalidity Insurance of our country
Abbreviations: m = meter; m2 = square meter; MSK = musculoskeletal; % = percentage
A disadvantage of CBCT is the limited field of view (FOV), which ranges from 6 × 6 cm to maximum 18 × 16 cm with our equipment. Therefore, CBCT is not suitable for imaging of large joints. Another drawback is its limitation to assess soft tissue pathology due to lack of contrast resolution and Hounsfield Units (HU) measurements. Furthermore, CBCT is more time consuming (18 to 36 s of acquisition time) resulting in higher susceptibility to motion artifacts. Table 2 summarizes the main advantages and disadvantages of musculoskeletal CBCT.
Table 2.
ADVANTAGES | DISADVANTAGES |
---|---|
Lower radiation dose than MDCT | Radiation exposure higher than CR |
More comfortable positioning than MRI for patients suffering from claustrophobia | Prone to motion artifacts (patient with tremor, children) |
Suitable for postoperative follow-up in patient with metallic implants using appropriate metal artifact algorithms | Limited field of view |
High spatial resolution images of bone architecture | Limited evaluation of soft tissue pathology |
High spatial resolution images of cartilage surface after intra-articular contrast injection | Mildly more time consuming than CR, comparable examination time to MDCT due to easier positioning |
Relative low cost of equipment | |
Joint imaging in weight-bearing position with some CBCT equipment |
Evaluation of fractures, dislocations and their follow-up
CR is the first-line imaging technique in case of clinical suspicion of fractures, although being less sensitive than cross-sectional imaging. CBCT imaging shows higher sensitivity in detection of small bone and joint trauma than CR and may visualize fractures being occult on CR (Fig. 3 and Fig. 4) or confirm doubtful fractures [17, 21, 22]. In most cases, this has an impact on the treatment strategy [5]. Comparison of MDCT as the gold standard with CBCT for finger fractures showed similar results in depicting the fracture and assessment of articular involvement [7]. In case of high clinical suspicion of carpal fractures (especially the scaphoid bone), but negative CR, subsequent MRI is often recommended for exclusion of occult nondisplaced fractures and bone marrow contusion. In this scenario, MRI remains more sensitive than CBCT [21]. Nonetheless, it is not always possible to perform MRI immediately following trauma and MRI cannot be performed in every patient, due to potential contraindications or lesser accessibility. Therefore, CBCT should be considered as a second line imaging modality in assessing complex anatomical sites with multiple overlapping bones, such as the wrist (Fig. 5) and foot, in case of negative CR but with high clinical suspicion of a fracture. Prompt, accurate evaluation of fractures may obviate the need of MRI at a later date [21] .
In case of suspicion of joint instability, weight-bearing CBCT provides information about joint alignment [5, 8] although this can only be performed on dedicated CBCT equipment [20].
For follow-up of bone healing and callus formation, CR may be difficult, especially in the presence of overlying splints or casts. As CBCT can provide more detailed information on bone architecture in comparison to CR, it can also help in evaluation of the healing process, which can be over- or underestimated on CR [6]. As a cross-sectional technique and the possibility for multidirectional reformations and 3-D reconstructions, CBCT is superior to CR in assessment of callus formation, osseous bridging and evaluation of residual fracture lines. In case of postoperative follow-up after placement of metallic hardware, incomplete healing as well as early detection of hardware loosening may be facilitated by CBCT. For hardware fractures, however, CR still remains the preferred method because of potential metallic streak artifacts on CBCT. The possible explanation for this potential discrepancy is related to the size of the metal objects. Indeed, at the bone–screw interface, higher contrast and spatial resolution of CBCT dominates the effect of the beam hardening owing to the relatively small size of metallic screws. Conversely, the beam hardening artifact surrounding large side plates in CBCT images compared with plain radiograph dominates the effect of better contrast resolution on CBCT [23]. Despite the use of current Metal Artifact Reduction sequences, the overall usefulness of MRI after screw fixation is limited due to metal artifacts.
Bone tumours and tumour-like lesions
The value of CBCT in the assessment of tumour and tumour-like conditions of the jaw bones has been reported previously in the dental literature [24–26]. Compared to the limited 2-dimensional information of conventional panoramic view, CBCT provides more precise information on location, morphology, intra-osseous extent, cortical breakthrough periosteal reaction and local effect on adjacent structures and teeth roots. Although definitive characterization of these lesions is often difficult or even impossible due to overlapping imaging characteristics, analysis of the matrix and intralesional calcifications and relationship with dentition are very helpful parameters in identification of odontogenic and non- odontogenic tumour and tumour-like conditions of the jaws. A disadvantage of CBCT is its limited assessment of the potential soft tissue component of the lesions. For bone tumours located in extremities, the use of CBCT is less documented. Similar to lesions in the jaws, CBCT allows for a more accurate evaluation of lesion location (either in the longitudinal or transverse axis of the bone), cortical breakthrough and periosteal reaction than CR (Fig. 6). Although these semiological features may help in lesion characterisation, MDCT and especially MRI are better suited for evaluation of the soft tissue component. Therefore, CBCT cannot be regarded as preferred technique for assessment of bone tumours.
Osteomyelitis
Although the preferred imaging modalities in evaluation of osteomyelitis are CR as baseline examination for follow-up and MRI for early detection and staging, CT is the best technique for assessment of a sequestrum in chronic osteomyelitis (Fig.7). In the jaw bones and appendicular skeleton, CBCT may have an equal diagnostic performance as MDCT. On MRI, a sequestrum is difficult to distinguish from sclerotic but viable bone. CBCT can also provide more detailed visualization of osteolytic changes caused by infection in the presence of metallic hardware.
Degenerative joint disease
In comparison to CR, CBCT shows more precisely the degree and extent of degenerative joint changes. It can depict pathology in small joints not visible due to overlying of bony structures, for example, in sesamoid bones at the level of metatarsal head (Fig.8) or in the presence of a metallic screw (Fig.9). Small osteophytes, joint space narrowing and subtle areas of subchondral sclerosis can be detected. Evaluation of this subtle cartilage loss in small joints on MRI is far more challenging because of poor spatial resolution. However, bone marrow oedema indicating disease activity is only detected on MRI.
CBCT arthrography
To evaluate chondral lesions, arthroscopy is a reference standard procedure offering also simultaneous treatment. Considering the operative risk and invasive nature of arthroscopy, appropriate preoperative imaging is preferable for diagnosis and local staging of cartilage lesions. For some cartilage lesions MR-arthrography (MRA) is widely used, particularly at the wrist. CBCT arthrography has -however- a better spatial resolution, the ability for thin multiplanar reformats allowing more accurate staging of articular cartilage lesions (Fig.10). This method can be notably useful in case of orthopaedic implants located near the area of interest [9, 27], in which MRI is less feasible due to susceptibility artifacts [27]. Intra-articular loose bodies (Fig. 11) or synovial tumour and tumour-like conditions such as Pigmented Villonodular Synovitis (PVNS) can be also better visualized after intra-articular contrast injection (Fig. 12).
Miscellaneous
Another useful application of CBCT is the imaging of various anatomic variants, bony coalitions, osseous defects simulating cartilage lesions on other imaging modalities (Fig. 13) or accessory or bifid bones (Fig.14).
Interventional radiology and future developments
Some modern flat panel detector C-arm units combine fluoroscopy with CBCT imaging offering guidance for interventional radiology procedures [28]. Spinal interventions including nucleoplasty, vertebroplasty or bone biopsies are among the most frequently performed percutaneous interventional musculoskeletal procedures. More and more ablation or palliative procedures are performed under imaging guidance, including CBCT [29]. When precise biopsy needle positioning is difficult to achieve, CBCT application can help in choosing the best approach, precise tracking and detection of errors in the operating room. The reduction of cumulative dose to patient and staff can be achieved by reduction of fluoroscopy time due to the usage of CBCT guidance [28, 30]. Fusion with MRI for precise lesion targeting has been reported as well [31]. The overall duration of those procedures under MDCT and CBCT guidance are similar with less dose applied to both the patient and performing physician for CBCT [32]. Especially in young patients, lower radiation dose is of utmost importance. Other systems on the market are designed for weight-bearing extremities examinations, allowing evaluation of joint stability [20]. Given the specific design (the size of the gantry, presence of guide lights for needle position) for each application, every business case for purchasing CBCT equipment should be tailored to the special needs in each department.
Conclusions
CBCT is a promising method that may be very useful for evaluation of trauma of small joints and bones, particularly when CR is negative or doubtful despite high clinical suspicion for fractures. In combination with arthrography, CBCT offers high anatomical detail of articular cartilage, which may be an advantage to routine MRI in evaluation of osteochondral lesions in joints with thin cartilage such as the ankle joint. Furthermore, the technique may serve as an alternative method to MRI for a variety of musculoskeletal diseases in patients with claustrophobia or other contraindications of MRI. In all these scenarios, an improved diagnosis may result in a more timely and appropriate treatment regime. Although MDCT equipment remains the preferred CT technology for multifunctional purposes in most imaging departments, due to its low cost in purchase and maintenance, CBCT may be a useful tool in private practices with a high turn-over of musculoskeletal procedures or as an additional imaging tool to MDCT in large hospitals. Awareness of the advantages and disadvantages of the technique is a prerequisite for its dedicated use.
Abbreviations
- CBCT
Cone Beam Computed Tomography
- CBCT-A
Cone Beam Computed Arthrography
- CR
Conventional Radiography
- FOV
Field of View
- HU
Hounsfield Units
- kVp
KiloVoltage power
- mAs
Milliamperage seconds
- MDCT
Multidetector Computed Tomography
- MRI
Magnetic Resonance Imaging
Compliance with ethical standards
Competing interests
The authors have no competing interests.
References
- 1.Scarfe WC, Farman AG, Sukovic P. Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc. 2006;72(1):75–80. [PubMed] [Google Scholar]
- 2.Klintstrom E, Smedby O, Moreno R, Brismar TB. Trabecular bone structure parameters from 3D image processing of clinical multi-slice and cone-beam computed tomography data. Skelet Radiol. 2014;43(2):197–204. doi: 10.1007/s00256-013-1766-5. [DOI] [PubMed] [Google Scholar]
- 3.Demehri S, Muhit A, Zbijewski W, et al. Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system. Eur Radiol. 2015;25(6):1742–1751. doi: 10.1007/s00330-014-3546-6. [DOI] [PubMed] [Google Scholar]
- 4.Ramdhian-Wihlm R, Le Minor JM, Schmittbuhl M, et al. Cone-beam computed tomography arthrography: an innovative modality for the evaluation of wrist ligament and cartilage injuries. Skelet Radiol. 2012;41(8):963–969. doi: 10.1007/s00256-011-1305-1. [DOI] [PubMed] [Google Scholar]
- 5.Pugmire BS, Shailam R, Sagar P, et al. Initial clinical experience with extremity cone-beam CT of the foot and ankle in pediatric patients. AJR Am J Roentgenol. 2016;206(2):431–435. doi: 10.2214/AJR.15.15099. [DOI] [PubMed] [Google Scholar]
- 6.Huang AJ, Chang CY, Thomas BJ, MacMahon PJ, Palmer WE. Using cone-beam CT as a low-dose 3D imaging technique for the extremities: initial experience in 50 subjects. Skelet Radiol. 2015;44(6):797–809. doi: 10.1007/s00256-015-2105-9. [DOI] [PubMed] [Google Scholar]
- 7.Faccioli N, Foti G, Barillari M, Atzei A, Mucelli RP. Finger fractures imaging: accuracy of cone-beam computed tomography and multislice computed tomography. Skelet Radiol. 2010;39(11):1087–1095. doi: 10.1007/s00256-010-0911-7. [DOI] [PubMed] [Google Scholar]
- 8.Carrino JA, Al Muhit A, Zbijewski W, et al. Dedicated cone-beam CT system for extremity imaging. Radiology. 2014;270(3):816–824. doi: 10.1148/radiol.13130225. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Koskinen SK, Haapamaki VV, Salo J, et al. CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT) Skelet Radiol. 2013;42(5):649–657. doi: 10.1007/s00256-012-1516-0. [DOI] [PubMed] [Google Scholar]
- 10.Loubele M, Bogaerts R, Van Dijck E, et al. Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications. Eur J Radiol. 2009;71(3):461–468. doi: 10.1016/j.ejrad.2008.06.002. [DOI] [PubMed] [Google Scholar]
- 11.Koivisto J, Kiljunen T, Kadesjo N, Shi XQ, Wolff J. Effective radiation dose of a MSCT, two CBCT and one conventional radiography device in the ankle region. J Foot Ankle Res. 2015;8:8. doi: 10.1186/s13047-015-0067-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Al Abduwani J, ZilinSkiene L, Colley S, Ahmed S. Cone beam CT paranasal sinuses versus standard multidetector and low dose multidetector CT studies. Am J Otolaryngol. 2016;37(1):59–64. doi: 10.1016/j.amjoto.2015.08.002. [DOI] [PubMed] [Google Scholar]
- 13.Haridas H, Mohan A, Papisetti S, Ealla KK. Computed tomography: will the slices reveal the truth. J Int Soc Prev Community Dent. 2016;6(Suppl 2):S85–S92. doi: 10.4103/2231-0762.189734. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.McCollough CH, Bruesewitz MR, Kofler JM., Jr CT dose reduction and dose management tools: overview of available options. Radiographics : Rev Publ Radiol Soc North Am Inc. 2006;26(2):503–512. doi: 10.1148/rg.262055138. [DOI] [PubMed] [Google Scholar]
- 15.Kalra MK, Maher MM, Toth TL, et al. Techniques and applications of automatic tube current modulation for CT. Radiology. 2004;233(3):649–657. doi: 10.1148/radiol.2333031150. [DOI] [PubMed] [Google Scholar]
- 16.Scarfe WC, Farman AG. What is cone-beam CT and how does it work? Dent Clin N Am. 2008;52(4):707–730. doi: 10.1016/j.cden.2008.05.005. [DOI] [PubMed] [Google Scholar]
- 17.De Smet E, De Praeter G, Verstraete KL, Wouters K, De Beuckeleer L, Vanhoenacker FM. Direct comparison of conventional radiography and cone-beam CT in small bone and joint trauma. Skelet Radiol. 2015;44(8):1111–1117. doi: 10.1007/s00256-015-2127-3. [DOI] [PubMed] [Google Scholar]
- 18.Bechara BB, Moore WS, McMahan CA, Noujeim M. Metal artefact reduction with cone beam CT: an in vitro study. Dentomaxillo fac Radiol. 2012;41(3):248–253. doi: 10.1259/dmfr/80899839. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Meilinger M, Schmidgunst C, Schutz O, Lang EW. Metal artifact reduction in cone beam computed tomography using forward projected reconstruction information. Z Med Phys. 2011;21(3):174–182. doi: 10.1016/j.zemedi.2011.03.002. [DOI] [PubMed] [Google Scholar]
- 20.Richter M, Seidl B, Zech S, Hahn S. PedCAT for 3D-imaging in standing position allows for more accurate bone position (angle) measurement than radiographs or CT. Foot Ankle Surg : Off J Eur Soc Foot and Ankle Surg. 2014;20(3):201–207. doi: 10.1016/j.fas.2014.04.004. [DOI] [PubMed] [Google Scholar]
- 21.Edlund R, Skorpil M, Lapidus G, Backlund J. Cone-beam CT in diagnosis of scaphoid fractures. Skelet Radiol. 2016;45(2):197–204. doi: 10.1007/s00256-015-2290-6. [DOI] [PubMed] [Google Scholar]
- 22.Vanhoenacker FM, Desimpel J. Cone beam CT of trauma of small bones and joints. Semin Musculoskelet Radiol. 2017;21(3):290–302. doi: 10.1055/s-0037-1602405. [DOI] [PubMed] [Google Scholar]
- 23.Osgood GM, Thawait GK, Hafezi-Nejad N, et al. (2017) Image Quality of Cone Beam Computed Tomography for Evaluation of Extremity Fractures in the Presence of Metal Hardware: Visual Grading Characteristics Analysis. Br J Radiol :20160539. doi: 10.1259/bjr.20160539 [DOI] [PMC free article] [PubMed]
- 24.Dabbaghi A, Nikkerdar N, Bayati S, Golshah A. Rare appearance of an odontogenic myxoma in cone-beam computed tomography: a case report. J Dent Res Dent Clin Dent Prospects. 2016;10(1):65–68. doi: 10.15171/joddd.2016.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Hunter AK, Muller S, Kalathingal SM, Burnham MA, Moore WS. Evaluation of an ameloblastic fibro-odontoma with cone beam computed tomography. Tex Dent J. 2012;129(6):619–624. [PubMed] [Google Scholar]
- 26.Araki M, Kameoka S, Matsumoto N, Komiyama K. Usefulness of cone beam computed tomography for odontogenic myxoma. Dentomaxillofacial Radiol. 2007;36(7):423–427. doi: 10.1259/dmfr/85361774. [DOI] [PubMed] [Google Scholar]
- 27.Suojarvi N, Haapamaki V, Lindfors N, Koskinen SK (2017) Radiocarpal injuries: cone beam computed tomography arthrography, magnetic resonance arthrography, and arthroscopic correlation among 21 patients. Scand J Surg 106(2):173–179. 10.1177/1457496916659226 [DOI] [PubMed]
- 28.Schafer S, Nithiananthan S, Mirota DJ, et al. Mobile C-arm cone-beam CT for guidance of spine surgery: image quality, radiation dose, and integration with interventional guidance. Med Phys. 2011;38(8):4563–4574. doi: 10.1118/1.3597566. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Cazzato RL, Garnon J, Tsoumakidou G, et al. Percutaneous image-guided screws meditated osteosynthesis of impeding and pathological/insufficiency fractures of the femoral neck in non-surgical cancer patients. Eur J Radiol. 2017;90:1–5. doi: 10.1016/j.ejrad.2017.02.022. [DOI] [PubMed] [Google Scholar]
- 30.Kroes MW, Busser WM, Hoogeveen YL, de Lange F, Schultze Kool LJ. Laser guidance in C-arm cone-beam CT-guided radiofrequency ablation of osteoid osteoma reduces fluoroscopy time. Cardiovasc Intervent Radiol. 2017;40(5):728–734. doi: 10.1007/s00270-016-1533-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Thakor AS, Patel PA, Gu R, Rea V, Amaral J, Connolly BL. MR cone-beam CT fusion image overlay for fluoroscopically guided percutaneous biopsies in pediatric patients. Pediatr Radiol. 2016;46(3):407–412. doi: 10.1007/s00247-015-3479-5. [DOI] [PubMed] [Google Scholar]
- 32.Tselikas L, Joskin J, Roquet F, et al. Percutaneous bone biopsies: comparison between flat-panel cone-beam CT and CT-scan guidance. Cardiovasc Intervent Radiol. 2015;38(1):167–176. doi: 10.1007/s00270-014-0870-9. [DOI] [PubMed] [Google Scholar]