Patient-Specific Organ and Effective Dose Estimates in Pediatric Oncology Computed Tomography

Yiming Gao a,*, Brian Quinn a, Neeta Pandit-Taskar b, Gerald Behr b, Usman Mahmood a, Daniel Long a, X George Xu c, Jean St Germain a, Lawrence T Dauer a,b
PMCID: PMC5828028  NIHMSID: NIHMS930531  PMID: 29472080

Abstract

Purpose

Estimate organ and effective doses from computed tomography scans of pediatric oncologic patients using patient-specific information.

Materials and Methods

With IRB approval patient-specific scan parameters and patient size obtained from DICOM images and vendor-provided dose monitoring application were obtained for a cross-sectional study of 1250 pediatric patients from 0 through 20 y-olds who underwent head, chest, abdomen-pelvis, or chest-abdomen-pelvis CT scans. Patients were categorized by age. Organ doses and effective doses were estimated using VirtualDose CT based on patient-specific information, tube current modulation (TCM), and age-specific realistic phantoms. CTDIvol, DLP, and dose results were compared with those reported in the literature.

Results

CTDIvol and DLP varied widely as patient size varied. The 75th percentiles of CTDIvol and DLP were no greater than in the literature with the exception of head scans of 16–20 y-olds and of abdomen-pelvis scans of larger patients. Eye lens dose from a head scan was up to 69 mGy. Mean organ doses agreed with other studies at maximal difference of 38% for chest and 41% for abdomen-pelvis scans. Mean effective dose was generally higher for older patients. The highest effective doses were estimated for the 16–20 y-olds as: head 3.3 mSv, chest 4.1 mSv, abdomen-pelvis 10.0 mSv, chest-abdomen-pelvis 14.0 mSv.

Conclusion

Patient-specific organ and effective doses have been estimated for pediatric oncologic patients from <1 through 20 y-olds. The effect of TCM was successfully accounted for in the estimates. Output parameters varied with patient size. CTDIvol and DLP results are useful for future protocol optimization.

Keywords: Pediatric, Computed Tomography, Diagnostic Reference Level, Organ Dose

Introduction

Computed Tomography (CT) is an important diagnostic imaging tool that helps save lives, but it does expose patients to ionizing radiation in addition to natural background radiation. On average, CT scans account for half of all medical radiation exposure in the United States (US) [1]. One in four persons in the US (per capita) had a CT scan in 2015 [2]. Concerns about the high scan number and the large contributions to medical exposure have led to campaigns and guidelines for justification and optimization of image quality and dose in the radiological community [36].

The advent of multi-detector CT (MDCT) has enabled rapid image acquisition without loss of spatial or contrast resolution. This is particularly useful in the pediatric oncologic setting, where cooperation with breath holds and motion avoidance are especially challenging for the young patients, many of whom have previously undergone numerous medical tests. Modern MDCT decreases reliance on and utilization of anesthesia [7]. However, MDCT technology itself does not directly decrease radiation exposure [8]. Recent successful decreases in pediatric exposure have resulted largely from national campaigns, which have raised awareness of the issue of CT parameter optimization within pediatric radiology practices [5]. This issue is of the utmost importance in the care of pediatric oncologic patients, as these patients tend to require numerous follow-up exams and subsequently receive relatively high cumulative dose.

In oncologic imaging, both pediatric and adult patients frequently undergo CT scans. While it is always important to justify and optimize CT radiation dose, it is particularly important to control the CT radiation dose to children. Compared to adults, pediatric patients are considered to have a higher risk of developing radiation-induced diseases such as leukemia, thyroid, breast, brain, or skin cancer in their lifetime because of long life expectation as well as rapid cell replication and division in the body [912]. To accurately evaluate associated radiation-induced risks, knowledge of the doses to specific organs is recommended [12].

Volumetric CT dose index (CTDIvol), dose length product (DLP) and, more recently in vendor provided application DoseWatch (General Electric Inc., Milwaukee, USA), size-specific dose estimate (SSDE) are reported at the end of CT examinations [3, 13], but typically without any organ dose information. Diagnostic reference level (DRL) is a tool recommended by the radiological protection community for the management and evaluation of CT dose quantities and for identification of unusually high doses [4, 14]. DRLs are not limitations of diagnostic radiation dose or references of organ dose, but rather quantities for comparing protocols and for promoting protocol optimization and avoidance of unnecessary dose [4, 14]. National or regional DRLs of CTDIvol, DLP and SSDE are published and available for institutional dose management [1517].

Extra efforts are required to obtain specific organ dose by measurements or computational methods. One can utilize a humanoid phantom of tissue equivalent material and a number of detectors and perform CT scans on the phantom to obtain dose readings in locations designated as various organs [1820]. On the other hand, one can also use sophisticated computational software to estimate organ dose for patients of various sizes and ages [2123] or employ the use of size-based computational methods and transverse patient size to estimate organ dose [2426]. These methods have advantages and disadvantages and can be selected based on available resources and demands [27, 28].

Organ dose calculation for a large number of pediatric patients is complicated by the variations in patient age and patient body size and the various CT scan techniques, as well as the use of tube current modulation. Although studies have shown that individual patient-specific computational phantoms could be generated and Monte Carlo simulations performed to estimate organ dose [25, 26], obtaining estimates in this manner currently takes a significant amount of time and is unrealistic for assessing doses in large numbers of patients. Moreover, the widely-used tube current modulation (TCM) technique changes the tube current angularly and longitudinally during CT scans to achieve an acceptable level of image quality with a lower average tube current, which redistributes the dose to organs and lowers the dose to some of them [26, 29, 30]. The complete TCM profile of a scan is usually proprietary and hard to obtain [25, 29], while a feasible way of obtaining longitudinal TCM is to extract the tube current of each image from the DICOM header [26]. In comparing several available organ dose estimation methods that generated patient-size based empirical functions by Monte Carlo simulations or phantom measurements [28], a reasonable and innovative approach would be to use modified computational software [21] that is capable of dose calculation for pediatric patients from <1 year (y) olds through 20 y-olds and can extract and directly apply the modulated tube current from DICOM images. The software was previously validated and compared to several other available tools as discussed in the literature [21, 28].

The aim of the present study was to estimate the organ and effective doses from computed tomography scans of oncologic pediatric patients with patient-specific information and to assess the reliability of these estimates by comparing them to estimates calculated with other methods from the literature [20, 2426].

Methods

Patient population

Institutional review board approval with a waiver of patient informed consent was obtained for this HIPAA-compliant, two-year retrospective study of CT scans of pediatric patients at a large academic cancer center performed during the period from September 2014 through September 2016. All DICOM images were anonymized prior to use. A total of 1250 cases were randomly selected to cover pediatric patients in five age categories: <1 y-olds, 1 to 5 y-olds, 6 to 10 y-olds, 11 to 15 y-olds, and 16 to 20 y-olds. In each age group, routine CT scans with or without imaging contrast of four body regions were selected: head (brain), chest, abdomen-pelvis (AP), and chest-abdomen-pelvis (CAP).

CT examination techniques

Records of pediatric CT scans were gathered by specifying time span via the vendor-provided dose monitoring application DoseWatch (General Electric Inc., Milwaukee, USA), and routine scans of the four aforementioned body regions were selected with in-house queries. Archived DICOM images of these CT scans were gathered through the HERMES GOLD (Hermes Medical Solutions Inc., Stockholm, Sweden) system. All scans were performed on HD750 CT scanners (General Electric Inc., Milwaukee, USA). For organ dose and effective dose calculation, parameters were extracted from the DICOM headers of images; these included patient age (y), gender, body part examined, tube voltage (kVp), tube current (mA) of each image, revolution time (second), beam collimator width (mm), pitch, bowtie filter type, CTDIvol (mGy), DLP (mGy*cm), and SSDE (mGy).

Percentiles calculation for Diagnostic Reference Level

DRLs for specific CT protocols in the broad patient cohort were estimated at the 75th percentiles of the distributions of radiation dose quantities (e.g., CTDIvol, DLP and SSDE) [4, 17]. The distributions of the CTDIvol, DLP and SSDE of the examinations were analyzed and the 75th percentiles were calculated and compared to DRLs from several national or regional surveys inside and outside the U.S. [1517].

Patient radiation dose assessment

Patient organ dose was calculated using a modified version of the Monte Carlo-based CT dose calculation software VirtualDose CT (VDCT) (Virtualphantoms Inc., Albany, USA). The modification integrated VDCT with in-house computer programs to extract the tube current of each image and other parameters from DICOM headers. VDCT was also modified to enable automatic transfer of the parameters and automatic estimation of organ and effective doses.

Patients were matched to the available computational anthropomorphic phantoms in VDCT according to their age: <1 y-olds simulated by newborn phantoms, 1–5 y-olds simulated by 1 y-old phantoms, 6–10 y-olds simulated by 5 y-old phantoms, 11–15 y-olds simulated by 10 y-old phantoms, and 16–20 y-olds simulated by 15 y-old phantoms. The body part examined was extracted and matched to the body region of the phantoms, with the start and end of scans defined as from top of head through the base of the skull for head scans, from the clavicles through the base of the lungs for chest scans, from the top of the liver through the pubic symphysis for AP scans, and from the clavicles through the pubic symphysis for CAP scans. Overscan of half of the beam width was assumed and added to each end of a simulated scan, since at least half of an extra rotation was needed for image reconstruction of helical CT scans [31]. Exam technique parameters were used to estimate organ dose as described previously [21], where tube current and time were applied in a slice-by-slice manner to address potential TCM technique. Results of organ doses and effective doses using tissue weighting factors from ICRP Publication No. 103 [32] were saved to a local database.

To account for the patient size variation within each age group, the ratio of patient-specific size-specific dose estimate (SSDE) to phantom SSDE was applied to all organ and effective dose results, as shown in the following equation:

Dpatient=Dphantom×SSDEpatient/SSDEphantom, (1)

Where Dpatient is the patient organ dose after SSDE correction, Dphantom is the organ dose obtained from the VirtualDose software using the corresponding phantom, SSDEpatient is the SSDE of the patient scan obtained from DoseWatch, and SSDEphantom is the SSDE of the corresponding phantom generated by multiplying the CT scan CTDIvol with the previously calculated f-factor (as described in AAPM report No.220) for the scanned phantom region [33].

Effective dose was also roughly calculated using the DLP of the scans and the DLP-to-effective-dose conversion factors (known as “k factors”) [13]. For chest scans and AP scans, organ doses estimated in this study were compared to doses estimated with methods in the literature, which employed Monte Carlo method or measurements and generated patient-size based organ dose estimation equations [20, 2426]. Specifically, CTDIvol, SSDE and patient effective diameter of the scanned region were applied to these methods to estimate organ dose.

Statistical analysis

Statistical analyses were performed using the Matlab 2016b (Natick, Massachusetts, USA) software. Quantities calculated included minimum, 25th percentile, mean, 75th percentile, and maximum. Two-way ANOVA analyses (rows: different methods, columns: doses to different organs) were performed for comparisons between the results of this study and doses estimated with different methods in the literature. The null hypothesis was that the two compared sets of data were drawn from the same population with the same mean. The statistical significance level was defined as p < 0.05.

Results

Patient population

Patient demographics are shown in Table 1. Of the 1250 CT scans evaluated in this study, 699 (59%) were performed on male patients and 551 (41%) on female patients (Table 1). For each of the age groups except the <1 y-olds, the number of scans was over 200. The mean body weight of the patient age groups varied from 6.9 kg to 67.1 kg. Pediatric patients with large body weight were found in the 11–15 y-olds (maximum of 142 kg) and the 16–20 y-olds (maximum of 149 kg).

Table 1.

Patient demographics

Age Group No. of Examinations Age (y) Weight (kg)

Head Chest Abdomen and Pelvis Chest, Abdomen, and Pelvis Subtotal Mean (Min-Max) Mean (Min-Max)
<1 year 2 13 1 3 19 0.0 (0.0–0.0) 6.9 (3.0–11.0)
1–5 years 41 55 17 98 211 2.8 (1.0–4.0) 14.4 (7.0–30.0)
6–10 years 26 94 21 128 269 6.8 (5.0–9.0) 22.3 (13.0–44.0)
11–15 years 20 130 26 79 255 12.3 (10.0–14.0) 49.0 (23.0–142.0)
16–20 years 53 222 55 166 496 17.6 (15.0–20.0) 67.1 (27.0–149.0)

Note.—The mean and the minimum-to-maximum range in parenthesis are provided for age and body weight of patients in each age group.

CT examination techniques

Table 2 lists mean values of CT technique parameters. The tube potential of the study cohort was 100 kVp or 120 kVp. The tube current was modulated in 55% of the 1250 cases. Head scans had the highest effective mAs (mean, 224 mAs), followed by AP (116 mAs), CAP (87 mAs), and chest scans (77 mAs). The variation (standard deviation or SD over mean) in the effective mAs was large, ranging from 27% for head scans to 42% for CAP scans. The bowtie filter type was medium filter for all head scans and body filter for all other scans.

Table 2.

CT technique summary

Scanning region Tube potential (kVp) Tube current (mA) Revolution time (s) Collimation (mm) Pitch CTDIvol (mGy) DLP (mGy·cm) SSDE (mGy)
Head 100 or 120 226.1 (120.0–300.0) 1.0 (1.0–1.0) 17.3 (10.0–40.0) 1.01 (0.98–1.07) 45.3 (26–65.3) 841.4 (339.7–2036.3) NA*
Chest 100 or 120 111.6 (40.0–380.0) 0.6 (0.5–0.8) 33.2 (20.0–40.0) 1.26 (0.98–1.38) 5.5 (1.0–15.0) 202.3 (20.9–676.6) 8.7 (2.2–16.4)
Abdomen and pelvis 100 or 120 201.4 (60.0–435.0) 0.7 (0.4–0.8) 39.2 (20.0–40.0) 1.22 (0.98–1.38) 8.4 (1.7–21.3) 441.2 (43.5–1218.9) 12.7 (3.2–28.4)
Chest, abdomen and pelvis 100 or 120 178.7 (50.0–380.0) 0.6 (0.5–0.8) 40.5 (20.0–160.0) 1.31 (0.98–1.38) 6.4 (1.4–21.2) 444.2 (45.7–1750.1) 10.1 (2.6–24.6)

Note.— The mean and the minimum-to-maximum range in parenthesis for all ages are provided. CTDIvol = Volumetric Computed Tomography Dose Index, DLP = Dose Length Product, SSDE = Size Specific Dose Estimate, NA = Not Applicable.

*

SSDE for head scans were not used in this study.

Percentiles calculation for Diagnostic Reference Level

The mean and 75th percentile of CTDIvol and DLP of various age groups from this study were compared to data from British and German national surveys (Table 3) [15, 16]. The data from Shrimpton et al [15] were for single-slice CT scanners, and their latest survey of multi-detector CT scanners [34] showed higher CTDIvol and DLP than their single-slice CT survey for all ages and protocols investigated. For CTDIvol the 75th percentiles in our study were no greater than those in the literature [15, 16] for any age- or region-specific groups except for AP scans of 1–5 y-olds. For DLP, 75th percentiles for head scans of 16–20 y-olds in our study were slightly higher (2%) than those in the German survey. The 75th percentiles of DLP for chest scans of all age groups in our study were lower than the literature values. The 75th percentile DLP values of AP scans in our study were higher than those in the German survey for 1 y-olds through 15 y-olds, but the 75th percentile CTDIvol values of these scans were very close to the literature values and lower than the 20 mGy DRL for 5 y-olds suggested by NCRP [14].

Table 3.

Mean and 75th percentile comparisons with international data for various pediatric age groups

Exam Age CTDIvol DLP

This study Galanski et al (2005)* Shrimpton et al (2003) This study Galanski et al (2005)* Shrimpton et al (2003)
Head <1 y 26 (26) 26 (34) 25 (28) 340 (340) 302 (393) 230 (270)
1–5 y 34 (35) 36 (49) 34 (43) 594 (558) 452 (611) 383 (465)
6–10 y 36 (38) 44 (58) 44 (51) 599 (607) 582 (711) 508 (619)
11–15 y 49 (48) 53 (65) 56 (64) 810 (815) 764 (920) 694 (787)
16–20 y 58 (60) 61 (60) 56 (64) 1183 (1119) 881 (1100) 694 (787)
Chest <1 y 1 (1) 3 (4) 11 (12) 28 (29) 40 (49) 159 (204)
1–5 y 3 (4) 3 (4) 11 (13) 86 (101) 61 (73) 198 (228)
6–10 y 4 (5) 5 (6) 14 (17) 124 (153) 102 (128) 303 (368)
11–15 y 5 (7) 6 (8) 10 (12) 198 (230) 180 (244) 402 (488)
16–20 y 7 (8) 11 (10) 10 (12) 277 (318) 368 (345) 402 (488)
Abdomen and pelvis <1 y 2 (2) 3 (4) NA 43 (43) 79 (82) NA
1–5 y 4 (6) 4 (5) NA 162 (220) 115 (147) NA
6–10 y 6 (7) 6 (7) NA 247 (293) 180 (227) NA
11–15 y 8 (10) 8 (10) 11 (13) 438 (502) 328 (402) 473 (534)
16–20 y 11 (12) 13 (15) 11 (13) 611 (719) 568 (980) 473 (534)

Note.—Dose data are mean (75th percentile). CTDIvol = Volumetric Computed Tomography Dose Index, DLP = Dose Length Product, NA = Not Applicable.

*

Adult data is used for 16–20 years group.

Adult data is used for both 11–15 years group and 16–20 years group.

The median and 75th percentiles of CTDIvol, DLP and SSDE for 16–20 y-olds were compared with the data of adult CT scans from regional surveys inside the U.S. by Kanal et al [17] (Table 4). For head scans both median and 75th percentiles were slightly higher (22% and 7% respectively) than in the survey, but the 75th percentile of CTDIvol was still below the 75 mGy DRL for adults suggested by NCRP [14]. For chest scans and CAP scans, the 50th and 75th percentiles were lower than in the survey. AP scans from our study had higher median DLP and higher 75th percentiles for SSDE than those in the U.S. survey. The size-specific data of the survey confirmed the trend of increasing CTDIvol, DLP and SSDE with increasing patient body size for body scans [17].

Table 4.

Median and 75th percentile comparisons with U.S. data for 16–20 years group

Examination CTDIvol (mGy) DLP (mGycm) SSDE (mGy)

This study Kanal et al (2017)* This study Kanal et al (2017)* This study Kanal et al (2017)*
Head without contrast 60 (60) 49 (57) 958 (1078) 849 (1011) 58 (60) NA
Chest without contrast 6 (8) 9 (12) 259 (338) 334 (443) 9 (11) 11 (15)
Chest with contrast 7 (9) 10 (13) 281 (359) 353 (469) 9 (12) 11 (15)
Abdomen and pelvis with contrast 12 (12) 12 (15) 648 (708) 608 (755) 17 (19) 15 (18)
Chest, abdomen, and pelvis with contrast 9 (11) 12 (15) 637 (878) 779 (947) 13 (16) 14 (18)

Note.—Dose data are 50th percentile (75th percentile). CTDIvol = Volumetric Computed Tomography Dose Index, DLP = Dose Length Product, SSDE = Size Specific Dose Estimate, NA = Not Applicable.

*

Adult data is used for comparison.

Patient radiation dose

Tables 58 show the minimum, 25th percentiles, mean, 75th percentiles, and maximum of effective doses as well as the organ doses. In addition to the five age groups, the dose results are further broken down by gender. For 1007 of the 1250 scans that reported patient SSDE, organ dose was estimated and the SSDE ratio was applied to the dose results.

Table 5.

Radiation dose to pediatric patients by head CT scans

Patient age <1 y 1–5 y 6–10 y 11–15 y 16–20 y

Sex* M (n=2) F (n=0) M (n=25) F (n=16) M (n=23) F (n=3) M (n=15) F (n=5) M (n=39) F (n=14)
Effective Dose (mSv)
Min 1.9 NA 0.3 2.3 2.1 3.1 3.3 3.3 2.7 2.8
25th percentile 1.9 NA 2.3 2.4 3.1 3.1 3.4 3.3 3.3 3.4
Mean 1.9 NA 2.3 2.6 3.1 3.2 3.5 3.4 3.3 3.3
75th percentile 1.9 NA 2.5 2.5 3.2 3.2 3.5 3.5 3.5 3.4
Max 1.9 NA 3.0 4.5 3.2 3.2 4.4 3.5 3.5 3.4
Sex-averaged mean 1.9 2.4 3.1 3.5 3.3
By k-factor 3.7 4.0 2.4 2.6 2.5

Organ Dose (mGy)
Red bone marrow 7.7 NA 9.0 10.2 11.2 11.6 15.5 14.9 15.3 15.6
(7.7,7.7) NA (1.1,11.4) (9.3,17.2) (7.4,11.7) (11.5,11.8) (14.3,20.3) (14.4,15.2) (12.5,17.2) (12.9,16.1)
Breast 0.5 NA 0.5 0.6 0.5 0.6 0.3 0.3 0.3 0.1
(0.5,0.5) NA (0.1,0.6) (0.5,1.0) (0.3,0.6) (0.6,0.6) (0.3,0.4) (0.3,0.3) (0.3,0.3) (0.1,0.1)
Colon 0.1 NA 0.1 0.1 0.0 0.0 0.0 0.0 0.0 0.0
(0.1,0.1) NA (0.0,0.1) (0.1,0.1) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0)
Lung 0.8 NA 0.8 1.0 1.3 1.4 0.7 0.7 0.5 0.6
(0.8,0.8) NA (0.1,1.1) (0.9,1.7) (0.8,1.4) (1.4,1.4) (0.7,0.9) (0.7,0.8) (0.4,0.5) (0.5,0.6)
Stomach 0.2 NA 0.2 0.3 0.2 0.2 0.1 0.1 0.0 0.1
(0.2,0.2) NA (0.0,0.3) (0.2,0.5) (0.1,0.2) (0.2,0.2) (0.1,0.1) (0.1,0.1) (0.0,0.1) (0.1,0.1)
Bladder 0.1 NA 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0
(0.1,0.1) NA (0.0,0.0) (0.0,0.1) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0)
Esophagus 1.1 NA 2.0 2.3 2.9 3.1 1.5 1.5 0.8 0.8
(1.1,1.1) NA (0.2,2.7) (2.1,4.1) (1.2,3.0) (3.0,3.1) (1.4,1.9) (1.4,1.5) (0.7,0.9) (0.7,0.8)
Gonads 0.0 NA 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0
(0.0,0.0) NA (0.0,0.0) (0.0,0.1) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0) (0.0,0.0)
Liver 0.3 NA 0.2 0.3 0.2 0.3 0.1 0.1 0.1 0.1
(0.3,0.3) NA (0.0,0.3) (0.3,0.5) (0.2,0.3) (0.3,0.3) (0.1,0.2) (0.1,0.2) (0.1,0.1) (0.1,0.1)
Thyroid 3.2 NA 2.9 3.4 5.2 5.6 4.1 4.1 2.1 2.6
(3.2,3.2) NA (0.2,3.9) (3.0,6.0) (2.3,5.5) (5.5,5.7) (3.9,5.2) (4.0,4.2) (1.6,2.3) (2.1,2.7)
Bone surface 10.3 NA 12.0 13.6 7.6 7.9 10.4 10.0 13.5 13.7
(10.3,10.3) NA (1.5,15.1) (12.3,22.7) (5.0,8.0) (7.8,8.0) (9.6,13.6) (9.7,10.2) (11.1,15.1) (11.4,14.2)
Brain 29.8 NA 32.0 36.1 30.9 32.0 43.0 41.2 46.4 48.7
(29.8,29.8) NA (6.5,39.6) (32.7,59.5) (21.0,32.4) (31.6,32.4) (39.7,57.1) (39.9,42.1) (38.0,52.9) (40.3,50.4)
Salivary Glands 5.2 NA 8.8 9.8 30.8 31.5 30.1 28.9 31.1 34.3
(5.2,5.2) NA (0.5,13.5) (8.5,19.7) (13.6,32.9) (31.0,31.9) (26.7,37.9) (28.0,29.5) (18.2,32.8) (28.4,35.6)
Skin 5.8 NA 6.2 7.0 5.3 5.5 4.9 4.7 4.0 4.1
(5.8,5.8) NA (0.9,7.8) (6.3,11.8) (3.8,5.6) (5.5,5.6) (4.5,6.2) (4.6,4.8) (3.3,4.3) (3.4,4.3)
Lens of eye 30.9 NA 35.1 39.6 40.1 41.7 54.9 52.9 66.4 68.6
(30.9,30.9) NA (1.1,38.7) (36.5,57.7) (28.8,42.1) (41.1,42.2) (51.0,69.0) (51.3,54.1) (54.5,72.3) (56.8,71.0)

Note.—Organ dose is displayed as mean (minimum, maximum). NA = not applicable, M = male, F = female.

*

Sex-specific examination number is given in parenthesis.

Mean effective doses were estimated using k-factors from AAPM report No. 96.

Table 8.

Radiation dose to pediatric patients by Chest-Abdomen-Pelvis CT scans

Patient age <1 y 1–5 y 6–10 y 11–15 y 16–20 y

Sex* M (n=3) F (n=0) M (n=58) F (n=39) M (n=73) F (n=54) M (n=47) F (n=30) M (n=88) F (n=63)
Effective Dose (mSv)
Min 2.7 NA 2.6 3.1 3.1 3.2 5.8 5.0 5.3 6.7
25th 2.9 NA 4.7 7.4 6.9 6.6 8.4 8.3 10.6 10.4
Mean 4.3 NA 6.9 7.5 7.8 7.5 9.9 12.2 14.9 12.8
75th 5.9 NA 8.2 8.2 8.9 8.4 11.3 14.0 18.6 13.6
Max 6.8 NA 10.3 9.1 12.9 12.3 15.5 24.6 25.7 23.7
Sex-averaged mean 4.3 7.2 7.7 10.8 14.0
By k-factor 3.5 4.4 4.3 7.1 11.4

Organ Dose (mGy)
Red bone marrow 2.4 NA 3.7 4.0 4.5 4.3 5.8 7.1 6.4 6.1
(1.5,3.7) NA (1.4,5.9) (1.6,4.8) (1.7,7.5) (1.8,7.2) (3.2,8.8) (3.1,13.6) (2.4,10.7) (3.1,10.9)
Breast 4.7 NA 6.4 7.1 8.0 7.8 8.4 11.2 11.9 11.9
(3.0,7.5) NA (2.4,11.7) (2.9,8.6) (3.2,13.6) (3.0,13.3) (5.2,13.9) (4.6,24.0) (5.7,20.9) (5.7,25.9)
Colon 5.2 NA 7.9 8.8 9.3 9.0 12.4 15.8 18.3 18.8
(3.3,8.2) NA (3.0,11.3) (3.6,10.6) (3.9,16.4) (3.9,15.5) (7.2,22.5) (6.0,32.8) (5.7,33.5) (8.3,40.0)
Lung 5.0 NA 8.6 9.3 9.2 8.9 11.3 13.4 15.2 13.0
(3.2,7.8) NA (3.2,15.0) (3.8,11.2) (3.5,14.8) (3.7,14.3) (6.9,15.5) (6.2,24.1) (6.2,24.6) (6.8,23.1)
Stomach 5.1 NA 7.9 8.7 9.0 8.7 12.2 15.0 18.1 15.7
(3.3,8.1) NA (3.0,12.0) (3.6,10.5) (3.6,16.5) (4.0,14.9) (6.8,19.5) (5.8,30.8) (5.5,33.1) (7.5,33.1)
Bladder 5.1 NA 7.8 8.6 9.3 9.6 12.1 16.7 17.9 17.6
(3.2,8.0) NA (3.0,12.4) (3.5,10.4) (3.8,15.2) (4.1,17.0) (6.7,22.8) (6.1,35.7) (6.2,32.0) (7.5,32.5)
Esophagus 4.1 NA 6.5 7.1 6.9 6.7 8.6 10.0 12.3 10.5
(2.6,6.5) NA (2.5,12.4) (2.9,8.5) (2.6,11.6) (2.9,10.9) (5.0,12.0) (4.9,17.4) (5.0,19.4) (5.4,18.3)
Gonads 3.2 NA 7.6 8.5 9.9 8.5 14.0 14.8 22.0 16.9
(2.1,5.1) NA (2.9,13.4) (3.5,10.2) (3.9,16.5) (3.5,15.0) (7.2,24.8) (5.4,30.7) (7.4,38.4) (7.3,31.6)
Liver 5.3 NA 8.2 9.0 9.1 8.8 12.1 14.9 18.9 15.5
(3.4,8.3) NA (3.1,12.4) (3.7,10.8) (3.6,16.7) (4.1,14.9) (6.9,19.5) (5.8,30.6) (5.7,34.3) (7.4,32.7)
Thyroid 2.9 NA 7.4 7.9 3.1 3.1 4.2 4.5 9.4 5.2
(1.9,4.6) NA (2.8,15.2) (3.2,9.6) (1.1,5.4) (1.3,4.7) (2.2,6.2) (2.2,6.6) (4.2,15.9) (2.8,8.2)
Bone surface 2.2 NA 3.5 3.8 4.9 4.7 6.4 7.8 9.9 9.3
(1.4,3.5) NA (1.3,5.6) (1.5,4.5) (1.9,8.2) (2.0,7.9) (3.5,9.8) (3.4,14.9) (3.7,16.4) (4.8,16.6)
Brain 0.2 NA 0.2 0.2 0.2 0.2 0.2 0.3 0.2 0.2
(0.1,0.3) NA (0.1,0.4) (0.1,0.3) (0.1,0.3) (0.1,0.3) (0.1,0.3) (0.1,0.4) (0.1,0.4) (0.1,0.4)
Salivary Glands 0.7 NA 1.2 1.3 0.8 0.7 0.9 0.9 1.1 0.9
(0.5,1.1) NA (0.5,2.4) (0.5,1.5) (0.3,1.3) (0.3,1.1) (0.5,1.3) (0.5,1.4) (0.5,1.8) (0.5,1.4)
Skin 2.9 NA 4.2 4.6 4.2 4.1 5.4 6.7 7.8 6.8
(1.9,4.6) NA (1.6,6.3) (1.9,5.6) (1.7,6.8) (1.8,6.7) (3.2,8.6) (2.7,13.2) (2.9,13.3) (3.5,12.5)
Lens of eye 0.2 NA 0.1 0.2 0.1 0.1 0.1 0.2 0.3 0.1
(0.1,0.3) NA (0.1,0.3) (0.1,0.2) (0.0,0.2) (0.0,0.2) (0.1,0.2) (0.1,0.2) (0.1,0.4) (0.1,0.2)

Note.—Organ dose is displayed as mean (minimum, maximum). NA = not applicable, M = male, F = female.

*

Sex-specific examination number is given in parenthesis.

Mean effective dose were estimated using k-factors from AAPM report No. 96.

The mean effective dose estimated by VDCT showed an increasing trend as patients’ ages increased, while effective dose roughly estimated with k factors showed similar trend except for head scans. The differences between VDCT estimated dose and k factors estimated dose were up to twofold for head scans, 70% for chest scans, 50% for AP scans, and 80% for CAP scans.

The five organs with the highest doses were: for head scans lens of eye, brain, salivary glands, red bone marrow and bone surface; for chest scans lung, breast, esophagus, liver, and thyroid; for AP scans gonads, colon, bladder, liver, and stomach; for CAP scans gonads, liver, colon, stomach, and bladder.

Comparison to Dose Estimates by methods from the Literature

For chest scans the average doses to eight organs were compared between this study and those obtained with the methods of Moore et al (M method), Tian et al (Ti method), and Papadakis et al (Pa method), as shown in Figure 1 [20, 25, 26]. ANOVA analysis showed the doses estimated from this study and those estimated with the M method differed significantly (p < 0.05). Multiple comparison analysis showed the mean organ dose of this study was lower than that estimated with the M method [20], with the greatest difference being −3.5 mGy (−38%, 5.8 mGy versus 9.3 mGy) for 11–15 y-olds (where the negative sign indicates that doses in this study were lower). Estimated doses of the eight organs did not differ significantly between our method and Ti method (p > 0.05) for any of the age groups compared. Dose estimates for five of the eight organs were compared between our method and Pa method [26], and significant differences in mean organ dose were found (p < 0.05) of −1.1 mGy (−11%, 8.7 mGy versus 9.8 mGy, 16–20 y old) and 0.8 mGy (13%, 6.9 mGy versus 6.1 mGy, 1–5 y old), where a negative sign again indicates doses of this study were lower.

Figure 1.

Figure 1

Graph shows comparison of dose to 8 organs by chest CT scans with method of this study and methods from literature (references): A. comparison for 1–5 years patients, B. comparison for 6–10 years patients, C. comparison for 11–15 years patients, D. comparison for 16–20 years patients. The whiskers represent one standard deviation from the mean. The 0 y-olds are not compared due to the limited number of examinations obtained in this study.

For AP scans the average doses to seventeen organs were compared between our method and M method and that of Turner et al (Tu method), as shown in Figure 2 [20, 24]. Mean organ dose differed significantly between our method and M method [20] (p < 0.05), with differences ranging between 0.8 mGy (10%, 8.5 mGy versus 7.7 mGy, 1–5 y-olds) and 4.1 mGy (41%, 14.1 mGy versus 10.0 mGy, 16–20 y-olds). No significant difference was found between our method and Tu method [24] (p > 0.05) except for 16–20 y-old, where the average difference was −1.7 mGy (−10%, 15.0 mGy versus 16.7 mGy).

Figure 2.

Figure 2

Graph shows comparison of dose to 17 organs by abdomen and pelvis CT scans with method of this study and methods from literature (references): A. comparison for 1–5 years patients, B. comparison for 6–10 years patients, C. comparison for 11–15 years patients, D. comparison for 16–20 years patients. The whiskers represent one standard deviation from the mean.

In summary the organ dose comparisons showed the dose estimation in our study was reasonably accurate: the maximal difference between VDCT doses and doses estimated by methods from the literature [20, 2426] was 38% for mean organ dose in chest scans and 41% for mean organ dose in AP scans.

Discussion

To the best of our knowledge, this is the first study that estimates organ and effective dose for over a thousand pediatric patients receiving oncologic CT scans with TCM. The use of patient-specific scan parameters from CT images, individual patient size information from a vendor-provided dose monitoring application, and the sophisticated and validated Monte Carlo based dose estimator with realistic pediatric phantoms facilitated patient-specific organ and effective dose estimation. Regarding patient-specific organ dose estimation, our study did not only obtain the longitudinal TCM from DICOM headers of patient images as described in the literature, but also modified the computer code to automatically choose a phantom as close to a patient as possible based on age and sex, used the SSDE correction for size differences between the chosen phantom and the corresponding patient, and was able to process hundreds of CT scans in several minutes once all required information was obtained.

Except for the <1 y-olds, the number of exams was more than (or close to) 20 for each specific age group and body region, thus enabling the collection of representative organ dose results and the sufficient resultant data for valid statistical analyses.

The 11–15 y-olds and 16–20 y-olds included patients of a very large size for their age (with body weight of 142 kg or more), for whom the use of an average-size human phantom was overly conservative. Therefore, for scans of the body regions, organ dose results were adjusted by the ratio of the SSDE of the patient to the SSDE of the phantom.

In our study, the CTDIvol and DLP were no greater than the DRLs from all literature surveys [1517] except for head scans of 16–20 y-olds, and the SSDE was lower than the DRL from the U.S. survey [17] except for AP scans with contrast. The calculation and analysis of the 75th percentiles of CTDIvol and DLP could be used to establish institutional DRLs to compare with reference levels and guide protocol optimization efforts that seek to reduce dose without compromising image quality [35]. The head scan dose percentiles suggest the opportunity for further optimization of protocols, and the AP scan dose 75th percentiles were affected by protocol adjustment for large patient size (up to 149 kg for 16–20 y-olds). A review of the data shows that for some of the 16–20 y-olds, an adult head/brain protocol was utilized; this, coupled with some patients with extended scan length above the head, contributed to additional dose from head scans in this age group. Improvement on patient categorization by age and protocol selection is an ongoing mission at our institution.

Large variations in the scanner output were found to accommodate the large variations in patient sizes: For example, the variation in CTDIvol (one SD over mean) was 26% for head scans, 45% for chest scans, 46% for AP scans, and 56% for CAP scans. In our study, the large variations in scan parameters accommodating variations in patient weight and size had led to variations in organ dose (one SD over mean) ranging from 13% through 50%.

For head scans, comparison was not performed due to the lack of organ dose data from the literature that was based on anatomically realistic phantoms and could account for TCM. The doses to organs in the head were believed reasonably estimated because they were within 20% difference from CTDIvol. The dose to lens of eye reached as high as 69 mGy for one brain CT scan for the 16–20 y-olds, and ten such scans during oncologic treatment fractions could lead to a total dose well above the 500-mGy dose threshold of the radiation-induced occurrence of cataract [36], suggesting that further optimization of lens of eye doses should perhaps be pursued [37].

For chest scans, doses estimated by the M method [20] were higher because their method could not account for the dose reduction effect of tube current modulation and their SSDE-to-dose factors (1.0 to 1.5 for in-range organs) led to estimated organ doses consistently greater than SSDE, whereas estimated organ doses in our study varied around SSDE [38]. For Ti method [25] and Pa method [26], extra steps were taken to obtain organ dose estimates. Tian et al [25] and Papadakis et al [26] generated their equations based on a 16-cm CTDI phantom, while in our study, the scan protocols in body regions were all based on a 32-cm CTDI phantom regardless of patients’ ages. Additional protocol-correction factors for Ti method [25] and Pa method [26] were generated in our comparisons, and these factors were the ratios of the CTDIvol of the specific scanner for 16-cm phantom they used in their study to the CTDIvol of GE HD750 scanner for 32-cm phantom in our study. Moreover, the water-equivalent diameter was required as input by Pa method [26] but unavailable in either DICOM headers or DoseWatch, and approximated by dividing the effective diameter with a factor of 1.22, which was the ratio of the two kinds of parameters of small patients for chest scans in the literature [33]. The extra efforts required by these methods may prevent clinical professionals from promptly and correctly estimating organ doses.

For AP scans, protocol-correction factor was not needed for Tu method [24], as their equations were based on a 32-cm CTDI phantom. However, Turner et al only estimated doses to seven organs [24]. For 16–20 y-olds we used a pair of male and female 15 y phantoms for dose estimation while Moore et al used one adult female phantom [20]. The size and anatomical difference between 15 y phantom and adult female phantom could have caused the large dose differences, and relatively small dose conversion factors were reported by Moore et al using the adult phantom [20]. In addition the AP scan range of this study covered testes by considering overscan, while Moore et al avoided direct irradiation of testes [20], and this range difference led to the large testes dose discrepancy as discussed in the literature [28].

Our study had several limitations. First, five pairs of pediatric phantoms were used to represent patients who were categorized into five age groups. Although patient size variation within each group was considered by applying the SSDE ratio of patient to phantom, ideally computational phantoms should be generated for each patient and Monte Carlo simulations performed on each one. However, this would require significant advances in computational speed. Second, only the longitudinal TCM was accounted for and the angular TCM was ignored because of lack of proprietary data. The angular TCM, especially the organ-based TCM could potentially decrease dose to sensitive organs such as lens of eye in head scans or female breasts in chest scans by up to 38%, while at the same time increasing dose to other organs such as bone surface and bone marrow by about 10% [3942]. However, female breast dose reduction was lower (only up to 19%) in clinical settings for patients than the reduction in experiments because of the lateral shift of breasts in supine position especially for elder patients [43]. Organ-based TCM was not used in any of the patient CT scans discussed in this study and based on literature we speculate using angular TCM the doses would be decreased by about 20% for superficial organs such as female breasts and the doses for bone surface and bone marrow would be increased by about 10% [4346]. Finally, our study provided very limited data on <1 y-olds and could not provide definitive findings toward these patients.

Conclusions

Organ and effective doses were estimated for a large number of pediatric patients with CT scan-specific and patient-specific parameters, taking TCM into account. The 75th percentiles of scan parameters such as CTDIvol and DLP were calculated and found to be lower than DRLs in the literature except in head scans.

These percentiles could be used to establish institutional DRLs and assist in protocol optimization in the future. As some adolescents in the study had large body sizes for their ages, SSDE ratios were used to account for patient size variations. Average organ dose estimates were found to be reasonably consistent (maximal difference of 38% for chest scans and 41% for AP scans) with estimates derived with other methods in the literature. Due to the variations in scan parameters, the relative standard deviation of organ dose was found to be as high as 50% of the mean dose. The large degree of variation in scan parameters called for patient-specific dose estimations, as performed in this study. Dose estimation methods utilizing CTDIvol with a multiplication factor should be utilized carefully to take into account the size of the CTDI phantom on which they are based. Dose to the lens of eye could be as high as 69 mGy for one brain CT scan and several scans could lead to lens doses greater than the threshold of radiation-induced cataract [36, 47]. As the method of dose estimation employed in this study allows consideration of the effects of TCM and accommodates a wide range of patients scanned with different parameters and is easy to use, it should be an appealing option until Monte Carlo simulations can be rapidly performed on computational phantoms generated for individual patients.

Table 6.

Radiation dose to pediatric patients by chest CT scans

Patient age <1 y 1–5 y 6–10 y 11–15 y 16–20 y

Sex* M (n=3) F (n=10) M (n=16) F (n=24) M (n=23) F (n=27) M (n=33) F (n=52) M (n=87) F (n=47)
Effective Dose (mSv)
Min 1.4 1.1 1.3 1.2 2.3 1.6 2.8 1.4 2.8 2.3
25th 1.4 1.1 3.7 3.1 3.4 3.6 3.3 3.2 3.4 3.2
Mean 1.4 1.3 3.6 3.4 4.1 3.9 3.7 4.0 4.2 3.9
75th 1.5 1.4 4.1 4.0 5.0 4.1 4.1 4.7 4.9 4.7
Max 1.5 1.6 5.5 5.3 6.2 6.0 5.4 5.9 7.5 6.5
Sex-averaged mean 1.3 3.5 4.0 3.9 4.1
By k-factor 1.1 2.2 2.2 2.6 3.9

Organ Dose (mGy)
Red bone marrow 0.9 0.8 2.3 2.2 2.7 2.4 2.4 2.6 2.0 1.8
(0.9,1.0) (0.7,1.1) (0.8,3.7) (0.8,3.6) (1.5,4.0) (1.0,3.7) (1.8,3.4) (0.9,3.7) (1.3,3.3) (1.1,2.9)
Breast 2.9 2.6 6.2 6.0 7.7 7.6 6.9 7.5 7.7 7.6
(2.8,3.0) (2.2,3.2) (2.2,9.3) (2.2,9.6) (4.2,12.2) (3.1,11.8) (4.9,10.5) (2.9,11.7) (5.1,15.0) (4.2,12.8)
Colon 0.4 0.3 0.4 0.4 0.4 0.4 0.4 0.4 0.3 0.3
(0.4,0.4) (0.3,0.4) (0.1,0.7) (0.1,0.7) (0.2,0.7) (0.2,0.8) (0.3,0.6) (0.1,0.6) (0.2,0.6) (0.2,0.5)
Lung 2.9 2.6 7.9 7.5 8.9 8.3 8.8 9.5 9.3 8.8
(2.8,3.1) (2.2,3.3) (2.8,11.8) (2.7,11.7) (4.8,13.5) (3.4,12.4) (6.6,12.4) (3.5,14.0) (6.2,16.4) (5.3,14.5)
Stomach 1.3 1.2 3.5 3.3 5.1 4.8 3.9 4.0 4.3 4.9
(1.3,1.4) (1.0,1.5) (1.2,5.4) (1.2,5.3) (2.9,8.0) (2.0,9.1) (2.6,5.9) (1.5,6.2) (2.7,7.6) (2.7,8.2)
Bladder 0.1 0.1 0.1 0.1 0.1 0.1 0.0 0.0 0.0 0.0
(0.1,0.1) (0.0,0.1) (0.0,0.2) (0.0,0.2) (0.0,0.1) (0.0,0.1) (0.0,0.0) (0.0,0.0) (0.0,0.1) (0.0,0.0)
Esophagus 2.4 2.1 6.2 5.8 6.6 6.2 6.6 7.1 7.4 6.8
(2.3,2.5) (1.8,2.7) (2.1,9.6) (2.1,9.2) (3.7,10.0) (2.4,9.3) (4.8,9.2) (2.5,10.1) (4.8,12.6) (4.0,11.0)
Gonads 0.0 0.1 0.0 0.1 0.0 0.1 0.0 0.0 0.0 0.0
(0.0,0.0) (0.1,0.1) (0.0,0.1) (0.0,0.2) (0.0,0.0) (0.0,0.2) (0.0,0.0) (0.0,0.1) (0.0,0.0) (0.0,0.1)
Liver 2.1 1.8 4.3 4.0 6.1 5.7 5.7 5.9 6.1 6.2
(2.0,2.2) (1.6,2.3) (1.5,6.5) (1.4,6.4) (3.4,9.3) (2.3,10.7) (3.9,8.7) (2.2,9.0) (3.8,10.8) (3.4,10.4)
Thyroid 1.8 1.6 7.3 6.8 3.1 2.9 3.5 3.8 6.7 3.7
(1.7,1.9) (1.4,2.0) (2.5,12.0) (2.4,10.6) (1.7,4.8) (1.1,4.1) (2.3,4.9) (1.2,5.2) (3.9,12.2) (2.2,5.3)
Bone surface 0.9 0.8 2.2 2.1 2.7 2.5 2.5 2.7 3.1 2.9
(0.9,1.0) (0.7,1.0) (0.8,3.6) (0.8,3.4) (1.5,4.1) (1.0,3.8) (1.8,3.4) (0.9,3.7) (2.0,5.2) (1.7,4.6)
Brain 0.1 0.1 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2
(0.1,0.1) (0.1,0.1) (0.1,0.3) (0.1,0.3) (0.1,0.3) (0.1,0.2) (0.1,0.2) (0.1,0.3) (0.1,0.3) (0.1,0.2)
Salivary Glands 0.4 0.4 1.2 1.0 0.8 0.7 0.7 0.8 0.7 0.6
(0.4,0.4) (0.3,0.5) (0.4,1.9) (0.4,1.6) (0.4,1.2) (0.3,1.0) (0.5,1.0) (0.2,1.1) (0.5,1.3) (0.4,0.9)
Skin 0.8 0.7 1.7 1.6 1.8 1.7 1.7 1.8 2.0 1.7
(0.8,0.8) (0.6,0.9) (0.6,2.6) (0.6,2.6) (1.0,2.6) (0.7,2.5) (1.2,2.3) (0.6,2.5) (1.3,3.4) (1.0,2.8)
Lens of eye 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.2 0.1
(0.1,0.1) (0.1,0.1) (0.0,0.2) (0.0,0.2) (0.1,0.2) (0.0,0.2) (0.1,0.2) (0.0,0.2) (0.1,0.3) (0.1,0.1)

Note.—Organ dose is displayed as mean (minimum, maximum). NA = not applicable, M = male, F = female.

*

Sex-specific examination number is given in parenthesis.

Mean effective dose were estimated using k-factors from AAPM report No. 96.

Table 7.

Radiation dose to pediatric patients by Abdomen-Pelvis CT scans

Patient age <1 y 1–5 y 6–10 y 11–15 y 16–20 y

Sex* M (n=0) F (n=1) M (n=11) F (n=6) M (n=3) F (n=17) M (n=8) F (n=18) M (n=24) F (n=13)
Effective Dose (mSv)
Min NA 2.8 2.3 5.6 5.3 4.1 5.2 5.6 5.9 6.0
25th NA 2.8 3.5 5.9 5.8 5.4 6.3 6.9 8.7 6.5
Mean NA 2.8 5.5 7.5 6.8 6.8 8.5 7.9 10.6 8.8
75th NA 2.8 7.7 8.6 7.6 7.6 9.3 8.9 12.1 11.9
Max NA 2.8 8.4 10.3 7.7 10.5 17.9 11.9 15.4 14.8
Sex-averaged mean 2.8 6.2 6.8 8.1 10.0
By k-factor 2.1 4.9 4.9 6.6 9.2

Organ Dose (mGy)
Red bone marrow NA 1.2 2.4 3.4 3.5 3.5 4.5 4.2 4.0 3.8
NA (1.2,1.2) (1.0,3.7) (2.5,4.8) (2.6,4.0) (2.1,5.5) (2.8,9.4) (3.0,6.4) (2.3,5.9) (2.5,6.3)
Breast NA 3.3 2.1 3.0 0.8 0.8 2.3 2.2 0.6 9.1
NA (3.3,3.3) (0.9,3.4) (2.2,4.1) (0.6,0.9) (0.5,1.3) (1.4,5.0) (1.6,3.5) (0.3,0.9) (6.3,16.4)
Colon NA 3.9 8.9 12.4 13.2 13.2 15.2 14.1 17.7 17.5
NA (3.9,3.9) (3.7,13.8) (9.2,16.8) (10.2,14.8) (8.0,20.5) (9.2,31.1) (10.0,22.3) (9.9,25.9) (12.2,30.2)
Lung NA 2.4 4.2 5.7 1.7 1.8 3.8 3.5 3.4 3.0
NA (2.4,2.4) (1.7,6.6) (4.2,7.7) (1.3,2.0) (1.1,2.7) (2.3,8.1) (2.6,5.5) (1.9,5.0) (2.1,5.4)
Stomach NA 3.8 8.3 11.5 10.9 11.0 13.9 12.7 15.8 14.6
NA (3.8,3.8) (3.4,13.1) (8.5,15.5) (8.3,12.5) (6.8,16.9) (8.1,29.1) (9.3,19.7) (8.8,23.0) (10.2,26.3)
Bladder NA 3.9 8.7 12.4 13.0 13.5 14.1 14.4 17.6 15.0
NA (3.9,3.9) (3.6,13.5) (9.1,17.6) (10.2,14.6) (8.0,20.9) (9.5,30.2) (9.8,21.0) (9.9,24.9) (8.8,21.7)
Esophagus NA 1.8 2.1 2.9 2.1 2.1 3.2 2.9 3.2 3.5
NA (1.8,1.8) (0.9,3.3) (2.1,3.9) (1.6,2.4) (1.3,3.3) (1.9,6.7) (2.1,4.6) (1.8,4.7) (2.4,6.3)
Gonads NA 3.5 8.5 12.1 13.3 12.2 15.7 12.8 21.4 14.5
NA (3.5,3.5) (3.6,13.2) (9.0,17.0) (10.5,14.9) (7.2,19.1) (10.6,33.1) (8.7,19.2) (11.8,29.9) (8.5,21.7)
Liver NA 3.8 8.6 11.8 10.6 10.6 13.3 12.1 16.1 14.1
NA (3.8,3.8) (3.5,13.5) (8.7,15.9) (8.1,12.1) (6.7,16.4) (7.8,27.9) (8.9,18.8) (8.9,23.4) (9.9,25.4)
Thyroid NA 0.3 0.6 0.8 0.3 0.3 0.3 0.3 0.2 0.3
NA (0.3,0.3) (0.2,0.9) (0.6,1.1) (0.2,0.3) (0.2,0.4) (0.2,0.6) (0.2,0.4) (0.1,0.4) (0.2,0.5)
Bone surface NA 1.2 2.3 3.2 4.0 4.0 5.0 4.7 6.1 5.7
NA (1.2,1.2) (0.9,3.5) (2.3,4.5) (3.0,4.6) (2.4,6.4) (3.1,10.4) (3.3,7.2) (3.4,8.8) (3.7,9.3)
Brain NA 0.1 0.1 0.1 0.0 0.0 0.0 0.0 0.0 0.0
NA (0.1,0.1) (0.0,0.1) (0.1,0.1) (0.0,0.0) (0.0,0.1) (0.0,0.1) (0.0,0.1) (0.0,0.0) (0.0,0.0)
Salivary Glands NA 0.2 0.2 0.3 0.1 0.1 0.1 0.1 0.1 0.1
NA (0.2,0.2) (0.1,0.3) (0.2,0.4) (0.1,0.1) (0.1,0.2) (0.1,0.2) (0.1,0.2) (0.0,0.1) (0.0,0.1)
Skin NA 1.8 3.6 5.0 4.1 4.1 4.8 4.5 5.5 4.8
NA (1.8,1.8) (1.5,5.5) (3.7,6.8) (3.2,4.6) (2.5,6.4) (3.0,10.0) (3.2,6.7) (3.1,8.0) (3.1,7.8)
Lens of eye NA 0.1 0.1 0.1 0.0 0.0 0.0 0.0 0.0 0.0
NA (0.1,0.1) (0.0,0.1) (0.1,0.1) (0.0,0.0) (0.0,0.1) (0.0,0.1) (0.0,0.0) (0.0,0.1) (0.0,0.0)

Note.—Organ dose is displayed as mean (minimum, maximum). NA = not applicable, M = male, F = female.

*

Sex-specific examination number is given in parenthesis.

Mean effective dose were estimated using k-factors from AAPM report No. 96.

Highlights.

  • Organ and effective doses are estimated for over a thousand patients.

  • Tube current modulation and patient body size variations are considered.

  • Eye lens dose can be higher than nominal effect threshold in few repeated scans.

  • Local CTDIvol and DLP 75th percentiles are mostly lower than national DRLs.

Acknowledgments

The authors would like to acknowledge the help of Dr. Hedvig Hricak, Mrs. Ada Muellner, and Dr. Robert Prins during the course of manuscript development. This research was funded in part through the NIH/NCI Cancer Center Support Grant P30 CA008748.

Footnotes

Authors’ contributions

YG collected and analyzed the data, drafted the manuscript, and participated in designing the study. BQ participated in designing the study and provided technical insights. NPT, GB, UM, DL, and JSG provided practical and clinical input and helped draft the manuscript. XGX provided the VirtualDose estimator and helped draft the manuscript. LTD conceived of and designed the study, provided technical oversight for dose calculation methods and results. All authors read and approved the manuscript.

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.NCRP. Ionizing Radiation Exposure of the Population of the United States. NCRP Publications; Bethesda, MD: 2009. [Google Scholar]
  • 2.The Organisation for Economic Co-operation and Development (OECD) Computed tomography (CT) exams (indicator) 2016. [Google Scholar]
  • 3.AAPM. Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations. AAPM Publications; 2011. AAPM Report No. 204. Online. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.ICRP. Radiation Protection in Medicine. ICRP Publication 105. Annals of the ICRP. 2007;37:1–63. doi: 10.1016/j.icrp.2008.08.001. [DOI] [PubMed] [Google Scholar]
  • 5.Goske MJ, Applegate KE, Boylan J, Butler PF, Callahan MJ, Coley BD, et al. The Image Gently campaign: working together to change practice. AJR American journal of roentgenology. 2008;190:273–4. doi: 10.2214/AJR.07.3526. [DOI] [PubMed] [Google Scholar]
  • 6.Applegate KE, Frush DP. Image Gently: A Decade of International Collaborations to Promote Appropriate Imaging for Children. Journal of the American College of Radiology: JACR. 2017;14:956–7. doi: 10.1016/j.jacr.2017.04.039. [DOI] [PubMed] [Google Scholar]
  • 7.Pappas JN, Donnelly LF, Frush DP. Reduced frequency of sedation of young children with multisection helical CT. Radiology. 2000;215:897–9. doi: 10.1148/radiology.215.3.r00jn34897. [DOI] [PubMed] [Google Scholar]
  • 8.Valentin J ICRP. Managing patient dose in multi-detector computed tomography(MDCT). ICRP Publication 102. Annals of the ICRP. 2007;37:1–79. doi: 10.1016/j.icrp.2007.09.001. [DOI] [PubMed] [Google Scholar]
  • 9.Brenner DJ, Hall EJ. Computed tomography--an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–84. doi: 10.1056/NEJMra072149. [DOI] [PubMed] [Google Scholar]
  • 10.Frush DP, Donnelly LF, Rosen NS. Computed tomography and radiation risks: What pediatric health care providers should know. Pediatrics. 2003;112:951–7. doi: 10.1542/peds.112.4.951. [DOI] [PubMed] [Google Scholar]
  • 11.Kalender WA. Dose in x-ray computed tomography. Physics in medicine and biology. 2014;59:R129–50. doi: 10.1088/0031-9155/59/3/R129. [DOI] [PubMed] [Google Scholar]
  • 12.UNSCEAR. UNSCEAR Publications. New York, NY: United Nations Scientific Committee on the Effects of Atomic Radiation; 2013. SOURCES, EFFECTS AND RISKS OF IONIZING RADIATION. [Google Scholar]
  • 13.American Association of Physicists in Medicine (AAPM) The Measurement, Reporting, and Management of Radiation Dose in CT. College Park, MD: AAPM; 2008. AAPM Report No. 96 ed. [Google Scholar]
  • 14.National Council on Radiation Protection and Measurements (NCRP) Reference Levels and Achievable Doses in Medical and Dental Imaging: Recommendations for the United States. Bethesda, MD: NCRP; 2012. Report No. 172 ed. [Google Scholar]
  • 15.Shrimpton P, Hillier M, Lewis M, Dunn M. Doses from computed tomography (CT) examinations in the UK-2003 review. Chilton, UK: National Radiological Protection Board (NRPB); 2005. NRPB-W67 ed. [Google Scholar]
  • 16.Galanski M, Nagel H, Stamm G. Paediatric CT exposure practice in the Federal Republic of Germany. Results of a nation-wide survey in. 2005;6:2006. [Google Scholar]
  • 17.Kanal KM, Butler PF, Sengupta D, Bhargavan-Chatfield M, Coombs LP, Morin RL. U.S. Diagnostic Reference Levels and Achievable Doses for 10 Adult CT Examinations. Radiology. 2017:161911. doi: 10.1148/radiol.2017161911. [DOI] [PubMed] [Google Scholar]
  • 18.Brisse HJ, Robilliard M, Savignoni A, Pierrat N, Gaboriaud G, De Rycke Y, et al. Assessment of organ absorbed doses and estimation of effective doses from pediatric anthropomorphic phantom measurements for multi-detector row CT with and without automatic exposure control. Health physics. 2009;97:303–14. doi: 10.1097/HP.0b013e3181ad00ed. [DOI] [PubMed] [Google Scholar]
  • 19.Zhang D, Li X, Gao Y, Xu XG, Liu B. A method to acquire CT organ dose map using OSL dosimeters and ATOM anthropomorphic phantoms. Medical physics. 2013;40:081918. doi: 10.1118/1.4816299. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Moore BM, Brady SL, Mirro AE, Kaufman RA. Size-specific dose estimate (SSDE) provides a simple method to calculate organ dose for pediatric CT examinations. Medical physics. 2014;41:071917. doi: 10.1118/1.4884227. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Ding A, Gao Y, Liu H, Caracappa PF, Long DJ, Bolch WE, et al. VirtualDose: a software for reporting organ doses from CT for adult and pediatric patients. Physics in medicine and biology. 2015;60:5601–25. doi: 10.1088/0031-9155/60/14/5601. [DOI] [PubMed] [Google Scholar]
  • 22.ImPACT. ImPACT CT dosimetry calculator, version 1.0.4. London: St. George’s Healthcare NHS Trust; 2011. [Google Scholar]
  • 23.Stamm G, Nagel HD. CT-Expo V2.0.1: a tool for dose evaluation in computed tomography. Hannover: 2014. [Google Scholar]
  • 24.Turner AC, Zhang D, Khatonabadi M, Zankl M, DeMarco JJ, Cagnon CH, et al. The feasibility of patient size-corrected, scanner-independent organ dose estimates for abdominal CT exams. Medical physics. 2011;38:820–9. doi: 10.1118/1.3533897. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Tian X, Li X, Segars WP, Paulson EK, Frush DP, Samei E. Pediatric chest and abdominopelvic CT: organ dose estimation based on 42 patient models. Radiology. 2014;270:535–47. doi: 10.1148/radiol.13122617. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Papadakis AE, Perisinakis K, Damilakis J. Development of a method to estimate organ doses for pediatric CT examinations. Medical physics. 2016;43:2108. doi: 10.1118/1.4944867. [DOI] [PubMed] [Google Scholar]
  • 27.Brady Z, Cain TM, Johnston PN. Comparison of organ dosimetry methods and effective dose calculation methods for paediatric CT. Australasian physical & engineering sciences in medicine. 2012;35:117–34. doi: 10.1007/s13246-012-0134-4. [DOI] [PubMed] [Google Scholar]
  • 28.Gao Y, Quinn B, Mahmood U, Long D, Erdi Y, St Germain J, et al. A comparison of pediatric and adult CT organ dose estimation methods. BMC medical imaging. 2017;17:28. doi: 10.1186/s12880-017-0199-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Khatonabadi M, Kim HJ, Lu P, McMillan KL, Cagnon CH, DeMarco JJ, et al. The feasibility of a regional CTDIvol to estimate organ dose from tube current modulated CT exams. Medical physics. 2013;40:051903. doi: 10.1118/1.4798561. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Schlattl H, Zankl M, Becker J, Hoeschen C. Dose conversion coefficients for paediatric CT examinations with automatic tube current modulation. Physics in medicine and biology. 2012;57:6309–26. doi: 10.1088/0031-9155/57/20/6309. [DOI] [PubMed] [Google Scholar]
  • 31.Schilham A, van der Molen AJ, Prokop M, de Jong HW. Overranging at Multisection CT: An Underestimated Source of Excess Radiation Exposure. Radiographics. 2010;30:1057–67. doi: 10.1148/rg.304095167. [DOI] [PubMed] [Google Scholar]
  • 32.ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Annals of the ICRP. 2007;37:1–332. doi: 10.1016/j.icrp.2007.10.003. [DOI] [PubMed] [Google Scholar]
  • 33.American Association of Physicists in Medicine (AAPM) Use of Water Equivalent Diameter for Calculating Patient Size and Size-Specific Dose Estimates (SSDE) in CT. College Park, MD: AAPM; 2014. AAPM Report No. 220 ed. [PMC free article] [PubMed] [Google Scholar]
  • 34.Shrimpton P, Hillier M, Meeson S, Golding S. Doses from computed tomography (CT) examinations in the UK-2011 review. Chilton, UK: Public Health England; 2014. PHE-CRCE-013 ed. [Google Scholar]
  • 35.Tonkopi E, Duffy S, Abdolell M, Manos D. Diagnostic Reference Levels and Monitoring Practice Can Help Reduce Patient Dose From CT Examinations. Am J Roentgenol. 2017;208:1073–81. doi: 10.2214/AJR.16.16361. [DOI] [PubMed] [Google Scholar]
  • 36.ICRP. ICRP Statement on Tissue Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs – Threshold Doses for Tissue Reactions in a Radiation Protection Context. ICRP Publication 118. Annals of the ICRP. 2012:41. doi: 10.1016/j.icrp.2012.02.001. [DOI] [PubMed] [Google Scholar]
  • 37.Vano E, Miller D, Dauer L. Implications in medical imaging of the new ICRP thresholds for tissue reactions. Annals of the ICRP. 2015;44:118–28. doi: 10.1177/0146645314562322. [DOI] [PubMed] [Google Scholar]
  • 38.Schlattl H, Zankl M, Becker J, Hoeschen C. Dose conversion coefficients for CT examinations of adults with automatic tube current modulation. Physics in medicine and biology. 2010;55:6243–61. doi: 10.1088/0031-9155/55/20/013. [DOI] [PubMed] [Google Scholar]
  • 39.Duan X, Wang J, Christner JA, Leng S, Grant KL, McCollough CH. Dose reduction to anterior surfaces with organ-based tube-current modulation: evaluation of performance in a phantom study. AJR American journal of roentgenology. 2011;197:689–95. doi: 10.2214/AJR.10.6061. [DOI] [PubMed] [Google Scholar]
  • 40.Wang J, Duan X, Christner JA, Leng S, Yu L, McCollough CH. Radiation dose reduction to the breast in thoracic CT: comparison of bismuth shielding, organ-based tube current modulation, and use of a globally decreased tube current. Medical physics. 2011;38:6084–92. doi: 10.1118/1.3651489. [DOI] [PubMed] [Google Scholar]
  • 41.Matsubara K, Sugai M, Toyoda A, Koshida H, Sakuta K, Takata T, et al. Assessment of an organ-based tube current modulation in thoracic computed tomography. J Appl Clin Med Phys. 2012;13:3731. doi: 10.1120/jacmp.v13i2.3731. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Gandhi D, Crotty DJ, Stevens GM, Schmidt TG. Technical Note: Phantom study to evaluate the dose and image quality effects of a computed tomography organ-based tube current modulation technique. Medical physics. 2015;42:6572–8. doi: 10.1118/1.4933197. [DOI] [PubMed] [Google Scholar]
  • 43.Euler A, Szucs-Farkas Z, Falkowski AL, Kawel-Bohm N, D’Errico L, Kopp S, et al. Organ-based tube current modulation in a clinical context: Dose reduction may be largely overestimated in breast tissue. European radiology. 2016;26:2656–62. doi: 10.1007/s00330-015-4085-5. [DOI] [PubMed] [Google Scholar]
  • 44.Fu W, Tian X, Sturgeon GM, Agasthya G, Segars WP, Goodsitt MM, et al. CT breast dose reduction with the use of breast positioning and organ-based tube current modulation. Medical physics. 2017;44:665–78. doi: 10.1002/mp.12076. [DOI] [PubMed] [Google Scholar]
  • 45.Fujii K, Nomura K, Muramatsu Y, Obara S, Akahane K, Kusumoto M. Organ Dose Evaluations Based on Monte Carlo Simulation for Ct Examinations Using Tube Current Modulation. Radiation protection dosimetry. 2017;174:387–94. doi: 10.1093/rpd/ncw144. [DOI] [PubMed] [Google Scholar]
  • 46.Kim JS, Kwon SM, Kim JM, Yoon SW. New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study. Radiol Med. 2017;122:601–8. doi: 10.1007/s11547-017-0755-5. [DOI] [PubMed] [Google Scholar]
  • 47.National Council on Radiation Protection and Measurements (NCRP) Guidance on Radiation Dose Limits for the Lens of the Eye. Bethesda, MD: NCRP; 2016. Commentary No. 26 ed. [Google Scholar]