Abstract
The impact of the microbiota on our health is rapidly gaining interest. While several bacteria have been associated with disease, and others being indicated as having a probiotic effect, the individual biomolecules behind these alterations are often not known. A major problem in the study of these factors in vivo is their low abundance in complex environments. We recently identified the first secreted bacterial antioxidant protein, RoxP, from the skin commensal Propionibacterium acnes, suggesting its relevance for maintaining the redox homeostasis on the skin. In order to study the effect, and prevalence, of RoxP in vivo, a capacitive biosensor with a recognition surface based on molecular imprinting was used to detect RoxP on skin in vivo. In vitro analyses demonstrated the ability to detect and quantify RoxP in a concentration range of 1 x 10−13 M to 1 x 10−8 M from human skin swabs; with a limit of detection of 2.5 x 10−19 M in buffer systems. Further, the biosensor was highly selective, not responding to any other secreted protein from P. acnes. Thus, it was possible to demonstrate the presence, and quantity, of RoxP on human skin. Therefore, the developed biosensor is a very promising tool for the detection of RoxP from clinical samples, offering a rapid, cost-effective and sensitive means of detecting low-abundant bacterial proteins in vivo in complex milieus.
Introduction
Propionibacterium acnes is one of the most common bacteria on human skin [1], and is involved in several skin disorders including acne vulgaris [2], and infections of prosthetic devices [3,4] due to biofilm formation on orthopedic implants [5]. Further, during microbial dysbiosis, as seen in the oxidative stress driven disease psoriasis [6], the prevalence of P. acnes is significantly decreased [7]. Due to the general high abundance of P. acnes on our skin, and its association with acne vulgaris, it has gained much scientific interest. Recently, in the omics-era, the secretome of P. acnes was published [8], demonstrating the presence of a highly secreted protein in all studied strains of P. acnes during anaerobic culturing in complex media in vitro. Using mass spectrometry Bek-Thomsen et al further demonstrated the presence of RoxP (e.g. PPA1939) from sebaceous follicles in vivo, though the abundance was lower than in vitro [9]. This protein, further on denoted as RoxP (Radical oxygenase of Propionibacterium acnes) was later demonstrated by us to be a unique secreted bacterial antioxidant, playing a critical role for the colonization of P. acnes on the skin, allowing the facultative anaerobe to tolerate the oxic environment [10].
The human skin is constantly being exposed to free radicals and oxidative stress, from among others UV-irradiation, activated immune cells, and regular metabolic activities [11–13]. The human body is well adapted to cope with this stress through intracellular antioxidants, like heme-oxygenase 1 that is up-regulated during UV stimulation [14]. Similarly, most bacteria, including P. acnes, have intracellular defense mechanisms against oxygen radicals, including catalases, superoxide dismutases and NADH oxidases. These antioxidants play a critical role in maintaining our health, protecting from oxidative stress and malignancies [15]. Therefore, it was suggested that the presence of the bacterial antioxidant RoxP on the skin could act as a benevolence factor, benefiting our health by protecting the skin from excess oxidative stress [10]; thus being part of maintaining our redox homeostasis. The prevalence of RoxP on skin could therefore serve as a biomarker for presence or proneness of developing oxidative diseases in individuals. However, so far no one has been able to detect and quantitate this bacterial protein from the complex skin environment.
Capacitive biosensors, a type of impedance biosensors [16], measure changes in dielectric properties, or the thickness of a dielectric layer at an electrolyte-electrode interface during the interaction between an analyte and its biorecognition element on the sensor surface [16]. Capacitive biosensors based on the use of biorecognition molecules have been used for detection of several targets, including peptides [17], toxins [18], nucleic acids [19], antibodies [20], antigens [21], and proteins [22]. To further improve on the selectivity, sensitivity, and operational stability of the biosensor surface, the use of molecularly imprinted polymers (MIPs) have been used to create biorecognition cavities on capacitive gold electrodes [23,24]. Molecular imprinting is a technique by which artificial recognition sites can be created in a polymer matrix which are complementary to the template in terms of size, shape, and chemical functionality [25,26]. MIPs are cost-effective, stable, and robust which make them superior compared to antibodies [27].
Herein we have developed a highly specific and selective capacitive biosensor for the in vivo detection and quantification of the skin bacterial antioxidant RoxP, using molecular imprinting. In buffer, RoxP could be detected in the attomolar range with a limit of detection of 2.5 x 10−19 M but could also be detected in a complex skin environment down to picomolar levels. Thus, the developed tool may be used as a highly sensitive method for detection and quantification of bacterial factors in vivo as exemplified by RoxP in this study.
Materials and methods
Materials
Acrylamide, N-hydroxymethylacrylamide (NHMA), N-isopropylacrylamide (NIPAm), N,N-methylenebisacrylamide (MBAAm), ammonium persulphate (APS), N,N,N’,N’-tetramethylethyldiamine (TEMED), Tris-HCl, Tween-20, Immunoglobulin A (IgA), Collagen, Bovine serum albumin (BSA), 3-aminopropyl-triethoxysilane (APTES), Glutaraldehyde (50%, w/v), Acryloylchloride, Triethylamine, 1-dodecanethiol, and Tyramine (99%) were all purchased from Sigma Aldrich. The capacitive biosensor and gold electrodes were supplied by CapSenze Biosystems AB, Lund, Sweden. All buffers were prepared with a Milli-Q system from Millipore. Prior to use, all buffers were filtered through a Millipore filter (pore size: 0.22 μm) and degassed for 1 h.
Preparation of RoxP-imprinted capacitive biosensor electrodes
Glass cover slips (24 x 50 mm) which were to be used as RoxP stamps were cleaned and modified according to the protocol of Ertürk et al. [28]. Following sequential modification by treatment with 10% (v/v) APTES and 5% (v/v) glutaraldehyde [23], the glass cover slips were immersed in a 0.1 mg.mL-1 RoxP solution in 10 mM phosphate buffer (pH 7.4) at 4°C overnight, resulting in a RoxP immobilization on the surface of the glass cover slips.
In the second step, capacitive gold electrodes were prepared. Gold electrodes were cleaned before tyramine electropolymerization was carried out to introduce free primary amino groups on the surface as described elsewhere [21,23]. Following tyramine modification, electrodes were immersed in a solution of 30 mM acryloyl chloride and 30 mM trimethylamine in toluene overnight at room temperature to introduce vinyl groups on the surface.
In the last step, hydrogel based molecularly imprinted polymers were prepared according to the protocols described by Reddy et al [29–31]. Acrylamide (54 mg), NHMA (140 μL), and NIPAm (85.6 mg) were used as functional monomers and methylenebisacrylamide (MBAAm) (9.5 mg) was used as a cross-linker [30]. Functional monomers and cross-linker were dissolved in 820 μL of Milli-Q water. TEMED (20 μL, 5% (v/v)) was added to the polymer matrix and purged with nitrogen gas for 5 min. Freshly prepared APS (20 μL, 10% (w/v)) was added to the mixture, and 1.5 μL of this solution was pipetted onto the modified gold electrode surface before the RoxP stamp was brought into contact with the monomer treated electrode. Polymerization continued for 5 h at room temperature after which the RoxP stamp was gently removed from the surface, and the electrode was rinsed with 10 mM phosphate (pH 7.4) buffer. In order to cover pinholes in the insulating layer on the gold surface, electrodes were immersed in 1-dodecanethiol (10 mM in ethanol) for 20 min. After rinsing the electrodes with water and drying them with nitrogen, a RoxP-MIP electrode was inserted into the electrochemical flow cell of the capacitive system for analysis (CapSenze HB, patent no. US9304096 B2, 2016).
A schematic presentation of the preparation of RoxP imprinted biosensor is shown in Fig 1. A non-imprinted (NIP) electrode was also prepared as a control in the presence of exactly the same monomer composition but in the absence of RoxP. A glass slide which was not containing any protein on the surface was used as the stamp for the preparation of the NIP electrode.
Surface characterization of RoxP-MIP capacitive electrodes with scanning electron microscopy (SEM)
For SEM, specimens were mounted on aluminum holders with adhesive carbon tape and sputtered with palladium/gold to create a 10 nm thick layer. The samples were examined in a DELPHI correlative light and electron microscope (Phenom-World) at the IQ Biotechnology Platform, Lund University.
Real time capacitive measurements with the RoxP-MIP capacitive system
Capacitance measurements were performed based on the current pulse method introduced by Erlandsson et al [20]. The capacitance measurement was performed by using a current pulse method which is based on the principle of an electrical double layer and the electrode-solution interface. This current pulse technique offers some advantages over the commonly used potential pulse technique. The use of microampere constant current helps the electrodes to be more resistant against external electronic disturbances which can cause to inaccurate measurement and poor baseline stability. The calculation of capacitance is done in a way that the potential response is linear to the pulse of current which makes it simple and more accurate. In this system, four values of capacitance are calculated from one cycle of current supply. One analysis cycle was started with the injection of regeneration buffer (25 mM glycine-HCl, pH 2.5, including 0.05 M Tween-20) to regenerate the surface, followed by equilibration in running buffer (50 mM Tris, pH 7.4) for 25 min in total before introduction of the sample to be analyzed. RoxP (1.0 x 10−18 M—1.0 x 10−13 M) dissolved in running buffer was injected sequentially into the system. All samples were measured in triplicates. After injecting the samples into the capacitive system, an average of the last five readings was calculated automatically by the CapSenze software and the calibration graphs were obtained by plotting the change in capacitance (-pF.cm-2) vs the log concentration of the analyte. The binding of RoxP to the surface resulted in a decrease in the total capacitance of the system according to the equation:
where total capacitance is equal to the capacitance contributions of insulating, biosensing and diffuse layers, respectively.
Selectivity of the RoxP-MIP biosensor
Selectivity experiments were carried out in order to show the selectivity of the system for RoxP versus other competing proteins. For selectivity, IgA, collagen, and BSA were used as competing proteins; and a NIP electrode as a negative control. Further, culture media from a ΔroxP P. acnes strain [32] was compared to wildtype culture media. The changes in capacitance (ΔC) after the injection of competing proteins were compared with the change due to RoxP injection, as well as the responses on RoxP-MIP and NIP electrodes. The measured values were used to calculate selectivity (k) and relative selectivity coefficients (k’). The selectivity coefficients (k) were calculated as ΔCRoxP/ΔCcompetitor. Relative selectivity coefficient was calculated as kRoxP-MIP/kNIP for IgA, collagen and BSA. All measurements were performed in triplicates.
Collection of skin swab samples
A skin surface area (cheek, forehead) of 4x4 cm was gently rubbed with a cotton swab (ESwab, Copan Diagnostics) for 20 seconds on one middle-aged (30–35 years) man and woman without any current skin disorders. The swab was put back into its container, gently shaken, and left at 4°C. Cells and debris were collected by centrifugation (4000 g, 15 min), and the supernatant frozen until analysis on the capacitive biosensor. All samples were collected with an informed written consensus of the individuals, according to the Declaration of Helsinki. An ethical approval to collect the samples was granted by the Lund University ethical committee (Dnr 2016/465).
Detection of RoxP on skin
A calibration graph of RoxP in Copan ESwab storage solution was performed in the concentration range of 1.0 x 10−13–1.0 x 10−9 M to take into account the background signal from the media itself. Extracted skin samples were diluted 1/10–1/105 in 50 mM Tris-HCl (pH 7.4), and injected into the biosensor system. All measurements were performed in triplicates.
Results
Surface characterization of bare and RoxP-MIP gold electrodes
In order to be able to detect and quantify the bacterial benevolence factor RoxP, a gold electrode was covered with a polymer matrix having cavities based on RoxP. In comparison to a bare gold electrode (Fig 2A), the RoxP imprints on the RoxP-MIP electrode could be visualized as small non-distinctive cavities by SEM (Fig 2B–2D).
Real-time detection of RoxP with RoxP-MIP capacitive biosensors
To verify that the molecular imprints of RoxP in the polymer could interact with free RoxP, a dilution series of RoxP (1.0 x 10−18–1.0 x 10−13 M) was sequentially injected into the capacitive biosensor flow system. The ΔC (change in capacitance; Fig 3A) increased linearly with increasing concentrations of RoxP with a good fitting to a linear regression model (R2 = 0.99) (Fig 3B). The limit of detection was calculated as 2.5 x 10−19 M based on IUPAC guidelines [33].
The RoxP-MIP capacitive biosensor is highly selective for RoxP
When analyzing environmental or patient samples, the complexity of the material is often contributing to either blocking the signal or creating false positives; mainly due to the low abundance of the analyte and high abundance of interfering material. Therefore, in order to detect low-abundant molecules in the presence of high-abundant molecules, the biosensor must have high selectivity and specificity. Due to the presence of RoxP on skin, we challenged the system with two other common skin proteins IgA and collagen. Further, BSA was included as a general control protein for selectivity purposes. Despite a concentration difference of approximately 104, favoring the competing proteins (IgA: 6.2 x 10−9 M; collagen: 3.7 x 10−9 M; BSA: 1.45 x 10−8 M), RoxP still generated higher response levels (Fig 4A). No difference in affinity between proteins could be observed using a NIP electrode (Fig 4B). Based on the data the selective coefficient (k) as well as the relative selectivity coefficient (k’) was calculated (Table 1), indicating the preference of RoxP-MIP to interact with RoxP (kMIP 3.09–5.14) while non-specific electrodes (NIP) did not show any preference for RoxP interaction (kNIP 1.02–1.06).
Table 1. Selectivity coefficients of RoxP-MIP capacitive biosensor.
Protein | Concentration (mg/mL) |
ΔMIP | ΔNIP | kMIP | kNIP | k’ |
---|---|---|---|---|---|---|
RoxP | 10−8 | 7744 | 5696 | - | - | - |
Collagen | 10−3 | 2152 | 5491 | 3.60 | 1.04 | 3.47 |
IgA | 10−3 | 2505 | 5352 | 3.09 | 1.06 | 2.90 |
BSA | 10−3 | 1507 | 5580 | 5.14 | 1.02 | 5.03 |
ΔMIP: Capacitance change of the MIP electrode for the proteins; ΔNIP: Capacitance change of the NIP electrode for the proteins; kMIP: selectivity coefficient of the MIP electrode versus competing proteins; kNIP: selectivity coefficient of the NIP electrode versus competing proteins; k’: relative selectivity coefficient of the MIP electrode versus the NIP electrode.
To further verify the selectivity of the RoxP-MIP, culture media from a wildtype P. acnes and its isogenic roxP mutant was analyzed on the capacitive biosensor. The wildtype P. acnes displayed a well-fitted linear regression (R2 = 0.998) with a high k-value, indicating an efficient response to RoxP (Fig 4D). However, the isogenic roxP mutant, lacking RoxP in the culture media, did not display a linear increase between the dilutions (R2 = 0.344), but rather a horizontal base line (Fig 4C).
RoxP can be detected and quantified from skin swabs
Since RoxP can be found on skin, it is imperative to investigate the ability of the capacitive biosensor to detect and quantify RoxP in such a complex sample. A mock skin swab was spiked with a dilution series of RoxP to generate a calibration curve in the specific media (Fig 5A). As a proof of principle that the biosensor can detect RoxP from skin in vivo, a swab from the forehead of a healthy middle-aged man was measured in a dilution series to verify the dose-dependency, confirming that the signal was not due to binding of interfering material (Fig 5B). Extrapolating the change in capacitance to a RoxP concentration demonstrated the presence of 10−9 M (9.4 nM ± 15.9 nM) RoxP in the skin swab sample, equivalent to 8.25 ng RoxP/cm2 skin (± 15.5 ng). The presence of other components in the sample matrix sometimes prevents the accurate and very sensitive measurement of the target. This can be one of the reasons why the LOD value obtained from skin swab is higher compared to the LOD value obtained from the buffer. Further studies will be able to investigate gender, age, and disease specific abundancies of RoxP on skin using this novel developed highly sensitive, specific, and selective biosensor.
Discussion
The impact of bacteria in both health and disease has been the focus for many researchers, and several factors contributing to disease (e.g. virulence factors) have been characterized. As is often the case with scientific questions, we need to simplify the experimental settings in order to examine the function of the target protein/factor. The beauty of the simplicity is the more conclusive answers it generates; the drawback being the loss of the thousands of natural factors also affecting the system. In many instances, the lack of human in vivo data can be attributed to the difficulty of detecting and quantitating the analyte. Factors to be studied in blood, saliva, skin, tissues, or other complex samples often need a high degree of processing before any analysis can take place, with high-abundant proteins/molecules often negatively affecting the readout [34]. Here we developed a highly sensitive, selective, and specific biosensor based on molecular imprinting of an antioxidant (RoxP) from the skin commensal P. acnes, able to detect and quantitate attomolar quantities of RoxP under optimized conditions, Importantly, we were able to detect RoxP in vivo on skin.
The biological function(s) of RoxP is still under characterization. We recently showed the importance of RoxP for colonization of P. acnes on skin, as well as survival in oxic environments in vitro [10]. However, the actual oxic conditions in the pilosebaceous follicles, the main habitat of P. acnes, and the necessity of RoxP to colonize this region of the skin, has not been investigated. The lowered abundance of RoxP in the sebaceous follicles compared to in vitro suggests that RoxP may partake in spreading of P. acnes (e.g. exposure to oxygen) rather than actual colonization in the sebaceous follicles. Due to its high affinity to heme, it has also been argued that the secreted porphyrins of P. acnes [35] may be the actual substrate of RoxP [10], reducing inflammatory UV-oxidized porphyrins. Further molecular investigations within this area will be needed to shed light upon this interaction.
In order to achieve the high sensitivity, several parameters have to be taken into account, the running buffer being but one of them. Earlier studies by El-Sharif et al, using a similar monomer composition as in our study, demonstrated the high selectivity of their system in Tris-HCl, while PBS reduced the selectivity [30]. In slightly basic conditions, Tris buffer will present three non-dissociated hydroxyl groups that can form hydrogen bonds with the MIP and the template protein, providing ideal conditions for a native protein interaction with the cavities within the MIP. While the buffer conditions may partly influence the selectivity, the monomer composition of the MIP is critical. The usage of hydrogel based molecular imprinted polymers (Hydro-MIPs) have shown high selectivity for template proteins as compared to NIP polymers [29,31,36]. Thus, even though highly selective, an even higher selectivity can likely be achieved for RoxP by optimizing the ratio of the monomers used for the polymer.
The issue of high background is prevalent in many methods. Due to the high sensitivity of the MIP biosensor, and thus the ability to dilute the sample several 10-folds, the problem with the complexity of the sample is reduced. Still, components of the sample may interact with the MIP, as seen with the culture media from the isogenic roxP mutant P. acnes strain. This interaction, likely due to short peptides present in the broth, is however of a much lower affinity than that of the analyte-MIP interaction, as evident by the significant fit of a linear regression model to a spiked culture media sample.
In this study we have focused on the secreted bacterial protein RoxP, an antioxidant that has been speculated to partake in the protection and redox homeostasis of the skin [10]. Since the skin is constantly exposed to oxidative stress through UV-irradiation, activated immune cells, and regular metabolic processes [6], the presence of an efficient antioxidant on the skin may prove beneficial. The involvement of P. acnes in several diseases has so far mainly been indicative rather than conclusive, much owing to its high prevalence on the skin. However, it is commonly reported to be underrepresented in skin dysbiosis (e.g. psoriasis) [7], diseases with a known element of oxidative stress [6]. Therefore, the ability to measure the presence of P. acnes, and specific proteins (RoxP) in real-time in vivo would generate valuable data allowing us to further evaluate the importance of this skin commensal and its role in health and disease. Earlier investigations have demonstrated the presence of RoxP in vitro and in vivo [8,9], but none of the methods employed (e.g. SDS-PAGE and MS/MS) are suitable for absolute quantification, nor relative quantification, due to the differences in molecular mass and ability of peptides to fly in the mass spectrometer.
We took a first step towards enabling absolute quantification of RoxP by developing a RoxP-MIP biosensor, optimizing it for detection and quantification of RoxP in the complex skin environment, detecting low ng quantities on the skin; a concentration comparing well to that of extrapolated in vitro measurements. For comparison, though the skin hardly can be described as a flat organ, a monolayer of proteins on a flat surface will reach quantities close to 300 ng/cm2 indicating that RoxP is highly abundant in the skin of healthy individuals. It should however be noted that we only took a skin swab, and that skin samples from different compartments (e.g. sebaceous follicles, scrub-wash fractions, etc) would generate different numbers; in particular since it is still unknown whether it is only the surface associated P. acnes expressing RoxP, or if also the follicular P. acnes express this antioxidant. Further research, taking advantage of this highly sensitive and selective biosensor, will shed light upon the in vivo role of RoxP in skin health and disease.
Conclusion
Herein, we have developed a highly sensitive, selective, and specific biosensor for the detection and in real-time quantification of the bacterial benevolence factor RoxP. Not only can this method be used as a research tool, or diagnostics to measure the presence of RoxP on skin. Rather, this method can be generalized to target any protein/biomolecule of interest within a complex milieu for a highly sensitive and selective detection and quantification, as exemplified with the quantification of the putative oxidative stress biomarker RoxP on the skin in vivo
Acknowledgments
We thank Maria Baumgarten (IQ Biotechnology Platform, Infection Medicine, Lund University) for performing and providing scanning electron micrographs. Tilde Andersson and Angelica Bengtsson are acknowledged for their skillful help with the project.
Data Availability
All relevant data are within the paper.
Funding Statement
This work was supported by grants from the Crafoord Foundation (20170669) to TR, and the Foundations Tornspiran, Alfred Österlund, Magnus Bergvall, Åke Wiberg, Längmanska Kulturfonden, Zoégas, Lars Hierta, Sigurd and Elsa Golje’s Memory, the Royal PhysiographicSociety Lund, and the Swedish Medical Society to RL. Bo Mattiasson and Martin Hedström are associated with the company CapSenze, on whose instrument the analyses were performed, providing support in the form of instrument leasing, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section. The other funders had no role in study design, data collection and interpretation, or the decision to submit the work for publication.
References
- 1.McGinley KJ, Webster GF, Leyden JJ. Regional variations of cutaneous propionibacteria. Appl Environ Microbiol. 1978. January;35(1):62–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Webster GF. Acne vulgaris. BMJ. 2002. August;325(7362):475–479. [PMC free article] [PubMed] [Google Scholar]
- 3.Tunney MM, Patrick S, Gorman SP, Nixon JR, Anderson N, Davis RI, et al. Improved detection of infection in hip replacements. A currently underestimated problem. J Bone Joint Surg Br. 1998. July;80(4):568–72. [DOI] [PubMed] [Google Scholar]
- 4.Tunney MM, Patrick S, Curran MD, Ramage G, Hanna D, Nixon JR, et al. Detection of prosthetic hip infection at revision arthroplasty by immunofluorescence microscopy and PCR amplification of the bacterial 16S rRNA gene. J Clin Microbiol. 1999. October;37(10):3281–90. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Holmberg A, Lood R, Mörgelin M, Söderquist B, Holst E, Collin M, et al. Biofilm formation by Propionibacterium acnes is a characteristic of invasive isolates. Clin Microbiol Infect. 2009. August;15(8):787–95. doi: 10.1111/j.1469-0691.2009.02747.x [DOI] [PubMed] [Google Scholar]
- 6.Kadam DP, Suryakar AN, Ankush RD, Kadam CY, Deshpande KH. Role of oxidative stress in various stages of psoriasis. Indian J Clin Biochem. 2010. October;25(4):388–92. doi: 10.1007/s12291-010-0043-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Gao Z, Tseng C, Strober BE, Pei Z, Blaser MJ. Substantial alterations of the cutaneous bacterial biota in psoriatic lesions. PLoS One. 2008. July 23;3(7):e2719 doi: 10.1371/journal.pone.0002719 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Holland C, Mak TN, Zimny-Arndt U, Schmid M, Meyer TF, Jungblut PR, et al. Proteomic identification of secreted proteins of Propionibacterium acnes. BMC Microbiol. 2010. August 27;10:230 doi: 10.1186/1471-2180-10-230 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Bek-Thomsen M, Lomholt HB, Scavenius C, Enghild JJ, Brüggemann H. Proteome analysis of human sebaceous follicle infundibula extracted from healthy and acne-affected skin. PLoS One. 2014. September 19;9(9):e107908 doi: 10.1371/journal.pone.0107908 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Allhorn M, Arve S, Brüggemann H, Lood R. A novel enzyme with antioxidant capacity produced by the ubiquitous skin colonizer Propionibacterium acnes. Sci Rep. 2016. November 2;6(1):36412. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Bossi O, Gartsbein M, Leitges M, Kuroki T, Grossman S, Tennenbaum T. UV irradiation increases ROS production via PKCdelta signaling in primary murine fibroblasts. J Cell Biochem. 2008. September 1;105(1):194–207. doi: 10.1002/jcb.21817 [DOI] [PubMed] [Google Scholar]
- 12.Babior BM, Kipnes RS, Curnutte JT. Biological defense mechanisms. The production by leukocytes of superoxide, a potential bactericidal agent. J Clin Invest. 1973. March;52(3):741–744. doi: 10.1172/JCI107236 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Grange PA, Chéreau C, Raingeaud J, Nicco C, Weill B, Dupin N, et al. Production of superoxide anions by keratinocytes initiates P. acnes-induced inflammation of the skin. PLoS Pathog. 2009. July 24;5(7):e1000527 doi: 10.1371/journal.ppat.1000527 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Ferrándiz ML, Devesa I. Inducers of heme oxygenase-1. Curr Pharm Des. 2008;14(5):473–86. [DOI] [PubMed] [Google Scholar]
- 15.Birben E, Sahiner UM, Sackesen C, Erzurum S, Kalayci O. Oxidative stress and antioxidant defense. World Allergy Organiz J. 2012. January 13;5(1):9–19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Berggren C, Bjarnason B, Johansson G. Capacitive Biosensors. Electroanalysis. 2001. March 1; [Google Scholar]
- 17.Lebogang L, Hedström M, Mattiasson B. Development of a real-time capacitive biosensor for cyclic cyanotoxic peptides based on Adda-specific antibodies. Anal Chim Acta. 2014. May 15;826:69–76. doi: 10.1016/j.aca.2014.03.028 [DOI] [PubMed] [Google Scholar]
- 18.Gutierrez R AV, Hedström M, Mattiasson B. Screening of self-assembled monolayer for aflatoxin B1 detection using immune-capacitive sensor. Biotechnol Rep (Amst). 2015. December;8:144–51. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Mahadhy A, Ståhl-Wernersson E, Mattiasson B, Hedström M. Use of a capacitive affinity biosensor for sensitive and selective detection and quantification of DNA-A model study. Biotechnol Rep (Amst). 2014. September;3:42–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Erlandsson D, Teeparuksapun K, Mattiasson B, Hedström M. Automated flow-injection immunosensor based on current pulse capacitive measurements. Sensors and Actuators B: Chemical. 2014. January;190:295–304. [Google Scholar]
- 21.Ertürk G, Hedström M, Tümer MA, Denizli A, Mattiasson B. Real-time prostate-specific antigen detection with prostate-specific antigen imprinted capacitive biosensors. Anal Chim Acta. 2015. September 3;891:120–9. doi: 10.1016/j.aca.2015.07.055 [DOI] [PubMed] [Google Scholar]
- 22.Teeparuksapun K, Hedström M, Wong EY, Tang S, Hewlett IK, Mattiasson B. Ultrasensitive detection of HIV-1 p24 antigen using nanofunctionalized surfaces in a capacitive immunosensor. Anal Chem. 2010. October 15;82(20):8406–11. doi: 10.1021/ac102144a [DOI] [PubMed] [Google Scholar]
- 23.Ertürk G, Berillo D, Hedström M, Mattiasson B. Microcontact-BSA imprinted capacitive biosensor for real-time, sensitive and selective detection of BSA. Biotechnology Reports. 2014. September;3:65–72. doi: 10.1016/j.btre.2014.06.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Haupt K. Peer reviewed: molecularly imprinted polymers: the next generation. Anal Chem. 2003. September;75(17):376 A–383 A. [DOI] [PubMed] [Google Scholar]
- 25.Lenain P, De Saeger S, Mattiasson B, Hedström M. Affinity sensor based on immobilized molecular imprinted synthetic recognition elements. Biosens Bioelectron. 2015. July 15;69:34–9. doi: 10.1016/j.bios.2015.02.016 [DOI] [PubMed] [Google Scholar]
- 26.Ertürk G, Mattiasson B. From imprinting to microcontact imprinting-A new tool to increase selectivity in analytical devices. J Chromatogr B Analyt Technol Biomed Life Sci. 2016. May 15;1021:30–44. doi: 10.1016/j.jchromb.2015.12.025 [DOI] [PubMed] [Google Scholar]
- 27.Ertürk G, Mattiasson B—Advanced Molecularly Imprinting, 2017, et al. Molecular Imprinting: The Creation of Biorecognition Imprints on Biosensor Surfaces. [Google Scholar]
- 28.Ertürk G, Hedström M, Mattiasson B. A sensitive and real-time assay of trypsin by using molecular imprinting-based capacitive biosensor. Biosens Bioelectron. 2016. December 15;86:557–65. doi: 10.1016/j.bios.2016.07.046 [DOI] [PubMed] [Google Scholar]
- 29.Hawkins DM, Stevenson D, Reddy SM. Investigation of protein imprinting in hydrogel-based molecularly imprinted polymers (HydroMIPs). Anal Chim Acta. 2005. June;542(1):61–5. [Google Scholar]
- 30.El-Sharif HF, Phan QT, Reddy SM. Enhanced selectivity of hydrogel-based molecularly imprinted polymers (HydroMIPs) following buffer conditioning. Anal Chim Acta. 2014. January 27;809:155–61. doi: 10.1016/j.aca.2013.11.052 [DOI] [PubMed] [Google Scholar]
- 31.Reddy SM, Phan QT, El-Sharif H, Govada L, Stevenson D, Chayen NE. Protein crystallization and biosensor applications of hydrogel-based molecularly imprinted polymers. Biomacromolecules. 2012. December 10;13(12):3959–65. doi: 10.1021/bm301189f [DOI] [PubMed] [Google Scholar]
- 32.Sorensen M, Mak TN, Hurwitz R, Ogilvie LA, Mollenkopf HJ, Meyer TF, et al. Mutagenesis of Propionibacterium acnes and analysis of two CAMP factor knock-out mutants. J Microbiol Methods. 2010. November;83(2):211–216. doi: 10.1016/j.mimet.2010.09.008 [DOI] [PubMed] [Google Scholar]
- 33.McNaught AD, Wilkinson A. Compendium of Chemical Terminology—IUPAC Recommendations (IUPAC Chemical Data). 2nd ed Wiley; 1997. [Google Scholar]
- 34.Wilson R. Sensitivity and specificity: twin goals of proteomics assays. Can they be combined? Expert Rev Proteomics. 2013. April;10(2):135–49. doi: 10.1586/epr.13.7 [DOI] [PubMed] [Google Scholar]
- 35.Shu M, Kuo S, Wang Y, Jiang Y, Liu YT, Gallo RL, et al. Porphyrin metabolisms in human skin commensal Propionibacterium acnes bacteria: potential application to monitor human radiation risk. Curr Med Chem. 2013;20(4):562–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Hawkins DM, Trache A, Ellis EA, Stevenson D, Holzenburg A, Meininger GA, et al. Quantification and confocal imaging of protein specific molecularly imprinted polymers. Biomacromolecules. 2006. September;7(9):2560–4. doi: 10.1021/bm060494d [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All relevant data are within the paper.