Finite element analysis of dental implants with validation: to what extent can we expect the model to predict biological phenomena? A literature review and proposal for classification of a validation process

Yuanhan Chang 1,, Abhijit Anil Tambe 2, Yoshinobu Maeda 1, Masahiro Wada 1, Tomoya Gonda 1
PMCID: PMC5842167  PMID: 29516219

Abstract

A literature review of finite element analysis (FEA) studies of dental implants with their model validation process was performed to establish the criteria for evaluating validation methods with respect to their similarity to biological behavior. An electronic literature search of PubMed was conducted up to January 2017 using the Medical Subject Headings “dental implants” and “finite element analysis.” After accessing the full texts, the context of each article was searched using the words “valid” and “validation” and articles in which these words appeared were read to determine whether they met the inclusion criteria for the review. Of 601 articles published from 1997 to 2016, 48 that met the eligibility criteria were selected. The articles were categorized according to their validation method as follows: in vivo experiments in humans (n = 1) and other animals (n = 3), model experiments (n = 32), others’ clinical data and past literature (n = 9), and other software (n = 2). Validation techniques with a high level of sufficiency and efficiency are still rare in FEA studies of dental implants. High-level validation, especially using in vivo experiments tied to an accurate finite element method, needs to become an established part of FEA studies. The recognition of a validation process should be considered when judging the practicality of an FEA study.

Keywords: Finite element analysis, Dental implant, Validation, Verification

Review

Background

Finite element analysis (FEA) has been applied to investigate dental implant designs, the structure and material of the superstructure, and the stability of the surrounding bone [1, 2]. According to PubMed, only 10 FEA studies of dental implants were published in 1990, while 102 papers were published in 2014.

FEA has become an increasingly useful tool in the past few decades. In the medical field, the behavior of any structure or tissue under a particular stimulation can be evaluated using FEA, and biomechanical changes in the tissues can be analyzed. Additionally, FEA allows for measurement of the stress distribution inside of the bone and various dental implant designs during mastication; such measurements are impossible to perform in vivo [1, 2, 3].

A large number of FEA regarding dental implant and bone were published in these decades; however, the precision and accuracy of those studies in silico are still questionable. In 2009, Dumont et al. [4] indicated that FEA studies of biological structures should be validated experimentally whenever possible. Hannam [5] stated that the minimum requirements of FEA studies should include comparisons with data from other work or any data that can be gleaned from the living subjects being modeled.

According to the American Society of Mechanical Engineers Committee on verification and validation in computational solid mechanics, verification is defined as “the process of determining that a computational model accurately represents the underlying mathematical model and its solution,” while validation is defined as “the process of determining the degree to which a model is an accurate representation of the real world from the perspective of the intended uses of the model.” In simple terms, verification is the process of “solving the equations right,” whereas validation is the process of “solving the right equations” [69]. Validation is a process by which computational predictions are compared with experimental data in an effort to assess the modeling error [69]. The sole purpose of these “experiments” is to produce data for comparison with model predictions rather than to address specific scientific hypotheses.

FEA studies with validation have recently become more common in the biomechanical field. FEA validations can be divided into two types: (1) direct validation, which involves experiments on the quantities of interest (from basic material characterizations to hierarchical system analysis such as model experiments and in vitro experiments), and (2) indirect validation, which involves the use of literature or the results of previous clinical studies. Indirect validation is clearly less favored than direct validation because of its uncertain experimental quality, sources of error, and high degree of variability. However, indirect validation may be unavoidable in FEA because no concrete biological outcome can be directly attributed to most FEA studies of force distribution; thus, it is difficult to generate outcome data for comparison with experimental data. Therefore, FEA studies should include a validation method to prove the close similarity of the results to the actual clinical situation. Validation is the process of “solving the right equations” [69] and comparing computational predictions with experimental data (the “gold standard”) in an effort to assess the modeling error.

The purpose of this literature review of FEA studies was to examine their model validation process and establish the criteria for evaluating validation methods with respect to their similarity to biological behavior.

Materials and methods

All studies included in this review (eligibility criteria) were FEA studies of the stress distribution of dental implants and surrounding bone using any type of validation method, and all were published in English. The exclusion criteria were publication in a language other than English, appearance of the word “validation” only in the references, no mention of the validation method for numerical FEA analysis, and mentioning of the requirement for validation without conduction of the actual validation.

An electronic literature search of PubMed was conducted up to January 2017 using the Medical Subject Headings “dental implants” and “finite element analysis.” After accessing the full text, the full context was searched using the words “valid” and “validation,” and all articles containing these words were read to determine whether they met the inclusion criteria. The selected articles were then read and summarized, and the validation techniques used in each article were assessed and categorized in a hierarchy (Fig. 1).

Fig. 1.

Fig. 1

Flowchart of literature review. An electronic literature search of PubMed was conducted up to January 2017 using the Medical Subject Headings “dental implants” and “finite element analysis.” After accessing the full texts, the context of each article was searched using the words “valid” and “validation” and articles in which these words appeared were read to determine whether they met the inclusion criteria for the review

Results

In total, 601 articles were obtained from the PubMed electronic search using the Medical Subject Headings “dental implants” and “finite element analysis.” After excluding articles for which the full text could not be accessed (n = 69) and that were not written in English (n = 10), 522 articles remained. These articles were searched using the terms “validation,” “validity,” and “valid” to determine whether the studies had performed a validation; after this process, 122 articles remained. These 122 articles were read, and 47 that met the eligibility criteria were selected and are summarized in Table 1. These articles were all FEA studies published from 1997 to 2016. The articles were categorized according to the method of validation as follows: in vivo experiments, model experiments, use of others’ clinical data or literature, and other software (Figs. 2 and 3).

Table 1.

All studies in the literature that considered with an actual validation of FEA

Ranking Authors Year FE model FEM geometry reference Material properties of tissues around implants Validation Material of validation model Comparison items
A Heckmann et al. [10] 2006 Implants embedded in a bone block CAD: bone block with a cortical layer and cancellous layer was constructed by CAD Homogeneous isotropic linear elasticity: cortical and trabecular bone (1) In vitro testing: strain gauge on implant support bridge in resin bone model
(2) In vivo testing: strain gauge on pontic of a 3-unit bridge in humans
(1) Epoxy resin
(2) In vivo: in a patient’s mouth
Surface strain of resin/resin
B Hou et al. [12] 2009 Implants embedded in bone (rat’s mouth) CT: CT data of the implant in a rat model Not mentioned In vivo experiment: implants placed in rat, and histologic findings compared after loading In vivo: rat Histologic findings
B Natali et al. [11] 1997 Implants embedded in a bone block CAD: bone section constructed by CAD Homogeneous isotropic linear elasticity: cortical and trabecular bone In vivo experiment: implant insertion in animal (dog) for loading and creation of sections of bone and implants In vivo: dog Visualization of change in bone and stress analysis by FEA
B Cha et al. [13] 2015 Implant engaged in bone disc (model was used to calculate torque) CAD: based on histology of the bone–implant interface Homogeneous isotropic linear elasticity In vivo experiment: implant insertion in animal (mice) with different insertion torques In vivo: mice Histomorphometric analyses
C Nagasao et al. [16] 2009 Implants embedded in normal mandibles and reconstructed mandible (with fibulae or ribs) and under mastication movement (dynamic condition) CT: dry mandibles, ribs, and fibulae Homogeneous isotropic linear elasticity: cortical and cancellous bone of every part of mandible, fibula, and rib Mechanical testing: implant embedded in 3 full mandibles and surface strain under loading measured by strain gauge Dry mandible as mandible Surface strain of bone under same conditions in FEA/experiment
C Nagasao et al. [17] 2010 Implants embedded in normal mandibles and reconstructed mandible (with fibulae or ribs) under mastication movement (dynamic condition) CT: dry mandibles, ribs, and fibulae Homogeneous isotropic linear elasticity: cortical and cancellous bone of every part of mandible, fibula, and rib Mechanical testing: implant embedded in 2 full mandibles and surface strain under loading measured by strain gauge Dry mandible as mandible Surface strain of bone under same conditions in FEA/experiment
C Eser et al. [18] 2009 Four implants embedded in the maxilla with bar superstructure CAD: model of nonanatomic maxilla, individualized arch form according to implant alignment Homogeneous isotropic linear elasticity: cortical bone, cancellous bone, Ti, Alloy, bar-superstructure Ex vivo strain gauge measurement of cadaver’s maxilla (with implants) Cadaver Surface strain of bone (maxilla)
C Nagasao et al. [19] 2006 Implants embedded in normal maxilla and cleft maxilla CT and CAD: normal maxilla: CT from a dry skull; palatal cleft, alveolar cleft, and complete cleft were designed by computer Homogeneous isotropic linear elasticity: cortical and trabecular bone Mechanical testing: strain measurement by strain gauge and implant embedded in actual skull model Dry skull Surface strain of bone
D Bardyn et al. [20] 2010 Implants embedded in bone (polyurethane foam and sheep bone) CT: polyurethane foam block and sheep bone Nonhomogeneous: calculated from CT data Mechanical testing in both polyurethane foam and sheep bone: measurement of removal torque of the implant Polyurethane foam and sheep bone Removal torque of implants
D Olsen et al. [21] 2005 Implants embedded in porcine mandibles from CT data and application of loading on the implant of FEM CT: porcine mandibles Nonhomogeneous: calculated from CT data Mechanical testing: comparison of displacement with actual measurements under the same testing load Block of porcine mandible Implant displacement under loading
D Huang et al. [22] 2002 Implant embedded in bone block CAD: bone block model constructed by CAD Homogeneous isotropic linear elasticity: cortical and trabecular bone In vitro model testing experiments: implant in bone cubic and measurement of resonance frequencies Bone section from lumbar vertebrae of hogs Value of resonance frequency
D Hasan et al. [23] 2012 Implant (implant and abutment together) embedded in bovine bone CT: scan of the models used for the experiment (implant embedded in bovine rib) Homogeneous isotropic linear elasticity: bovine cortical bone, bovine cancellous bone Mechanical tests: implant displacement and rotation under loading were measured using a biomechanical measurement system (laser pinhole and camera) Bovine rib section as mandible bone Displacement of the abutment
D Chatzigianni et al. [24] 2011 Mini-implant embedded in bone CT: scan of the specimen used for the experiment (implant embedded in bovine rib) Homogeneous isotropic linear elasticity: bovine cortical bone, bovine cancellous bone Mechanical tests: implant displacement and rotation under loading were measured using a 3D mobility measurement system (laser beams and camera) Bovine rib section as mandible bone Displacement of the abutment
E1 Tiossi et al. [14] 2013 Implants and tooth (acrylic) embedded in resin block model, crowns (splint and non-splint) CAD: epoxy model block Nil (in this FEM, there was a resin block only and no living tissue simulation) Digital image correlation (DIC): images of deforming body captured and strain calculated. Mechanical testing with implants embedded in resin block Resin block as mandible bone Calculated surface strain by DIC and FEA
E2 Ozçelik et al. [25] 2007 Three-unit bridge fixed prosthesis (with rigid connector and non-rigid) supported by an implant and a natural tooth, with an adjacent tooth and surrounding bone CAD: a bone section (2D) was constructed by CAD with a cortical layer and spongious bone and PDL Homogeneous isotropic linear elasticity: enamel, dentin, pulp, cortical bone, cancellous bone Photoelastic stress analysis methods (PSAM): implants placed in photoelastic resin, then force loaded and photograph taken Photoelastic resin as bone Stress distribution in bone/resin
E3 Chou et al. [26] 2014 A section of mandible and implant CT data and 2D FE model used in previous study Homogeneous isotropic linear elasticity Mechanical testing: implant embedded in resin bone and strain measured by strain gauge 3D printer to build acrylic-based polymer Surface strain of bone
E3 Mobilio et al. [29] 2013 Implant embedded in a bone block CAD: bone block built by CAD with a cortical (1.5 mm) and trabecular (28.5 mm) layer Homogenous anisotropic linear elasticity cortical bone: orthotopic linearly elastic material; trabecular bone: transversely isotropic linearly elastic material Mechanical testing: implant embedded in resin block and strain measured by strain gauge Resin block as mandible bone Load and strain relationship
E3 Chang et al. [30] 2012 Short implants with crowns embedded in left posterior segment of maxilla CT: CT scan of a dry human male skull Homogeneous isotropic linear elasticity: cortical bone, cancellous bone (high and low density) Mechanical testing: strain measured by strain gauge and implant embedded in resin block under loading ABS plastic bone as maxillary bone Surface strain of bone/resin
E3 Tu et al. [31] 2010 Implant embedded in resin block CAD: a resin block with a cortical layer and cancellous layer was constructed by CAD Nil (in this FEM, there was a resin block only and no living tissue simulation) Mechanical testing: strain measured by strain gauge and implant embedded in resin mandible section Resin bone as mandible bone Surface strain of bone/resin
E3 Lin et al. [32] 2010 Implant embedded in the left maxilla with crown CT of intact healthy male patient Homogeneous isotropic linear elasticity: cortical bone, cancellous bone Mechanical testing: strain measured by strain gauge and implant embedded in resin mandible section ABS plastic bone as maxillary bone Surface strain of bone/resin
E3 Qian et al. [33] 2009 Implant embedded in bone block CAD: a bone cubic with cortical layer and cancellous layer was constructed by CAD Homogeneous isotropic linear elasticity: cortical bone, cancellous bone (1) In vitro experiment: mechanical testing with resin bone and digital image correlation to calculate displacement of implant and strain on bone
(2) Literature data: strain gauge measurement in model experiment
Resin block as mandible bone (1) Displacement of implant and strain on bone
(2) Surface strain of bone
E3 Karl et al. [34] 2009 Implant embedded in base made by 3 materials CAD: acrylic, G10 epoxy resin, aluminum Homogeneous isotropic linear elasticity: acrylic, G10 epoxy resin, aluminum Mechanical testing: strain gauge. FEA-calculated strain was compared with strain gauge results Acrylic resin, glass-filled epoxy, aluminum Surface strain of acrylic resin, glass-filled epoxy, aluminum
E3 Hsu et al. [35] 2009 Implant embedded in resin block (with resin’s parameter for consistence with experiment) Nil Homogeneous isotropic linear elasticity: Resin (epoxy and Tempron) Mechanical testing: implant embedded in resin bone section and surface strain under loading was measured by strain gauge Resin block as mandible bone Surface strain of resin/resin
E3 Nagasawa et al. [36] 2008 Implant embedded in a bone block (only compact bone) CAD: a bone block (compact bone) was constructed by CAD Homogeneous isotropic linear elasticity: compact bone Mechanical loading test for implant, sectioned longitudinally Nil Implant deformation; no scientific values
E3 Huang et al. [37] 2005 Splinted or non-splinted 2-unit crowns supported by 2 or 3 implants embedded in bone CT: CT of posterior portion of a cadaver mandible Homogeneous anisotropic linear elasticity cortical bone: orthotopic linearly elastic material; trabecular bone: transversely isotropic linearly elastic material Mechanical test: strain measured by strain gauge on model Acrylic resin as mandible bone Surface strain of resin/bone
E3 Iplikçioğlu et al. [38] 2003 Implant embedded in bone block CAD: a resin block model was constructed by CAD Nil (in this FEM, there was a resin model only and no living tissue simulation) Mechanical test: measurement of stress on the implant, abutment, and resin Resin block as bone Stress distribution in resin and implants
E3 Chang et al. [27] 2016 Ball attachment overdenture (mandible, implant and attachment, mucosa, denture) CT: from a single human mandible (edentulous 65-year-old woman) Homogeneous isotropic linear elasticity Mechanical test: strain measured by strain gauge on surface of bone model Rapid prototype ABS plastic bone model, and a 3-mm layer of silicone to simulate mucosa Surface strain of resin/bone
E3 Rezende et al. [28] 2015 Bone section with embedded implant and prosthesis (metal coping and porcelain), screws CT: in vitro model (resin bone) Homogeneous isotropic linear elasticity Mechanical test: strain measured by strain gauge on surface of bone model Polyurethane resin Surface strain of resin/bone
E3 Chang et al. [39] 2012 Implants embedded in maxilla section with imperfect and perfect osseointegration under force loading CT: data of maxillary first molar area Inhomogeneous anisotropic linear elasticity cortical bone: anisotropic Trabecular bone: transversely isotropic linearly elastic material Mechanical testing: strain measured by strain gauge and implant embedded in resin block Resin block as maxillary bone Surface strain of bone/resin block
E3 Chang et al. [40] 2012 Implants and crowns in a section of the maxilla CAD: a bone block with a cortical layer and cancellous layer was constructed by CAD Homogeneous anisotropic linear elasticity compact bone, cancellous bone Mechanical testing: strain measured by strain gauge and implant embedded in resin block ABS resin block as mandible bone Surface strain of bone/resin block
E4 Zhiyong et al. [41] 2004 (1) Single tooth in bone block
(2) Single implant in bone block
(3) Various FPD supported by tooth and implant
CAD: a bone block model was constructed by CAD Homogeneous isotropic linear elasticity: cortical and trabecular bone, dentin, PDL Mechanical testing: comparison of displacement with actual measurements under the same tested loading conditions Not mentioned Implant displacement under loading
E4 Chang et al. [42] 2012 Implant embedded in a bone block CAD: a bone block with a cortical layer and cancellous layer was constructed by CAD Homogeneous isotropic linear elasticity: compact bone, cancellous bone Mechanical testing: pullout testing of mini-implant inserted in synthetic bone material Synthetic bone material as mandible bone Pullout strength of mini-implant
E5 Inglam et al. [43] 2013 Implant embedded in a bone block CAD: a bone block with a cortical layer and cancellous layer was constructed by CAD Homogeneous anisotropic linear elasticity cortical bone: orthotopic isotropic
Trabecular bone: transversely isotropic linearly elastic material
Mechanical testing: strain measured by strain gauge and implant embedded in resin block Resin block as mandible bone Surface strain of bone/resin block
E5 Necchi et al. [44] 2003 Implant (fixture, abutment, and connecting screw) Nil Nil Mechanical failure tests: preloading and functional loading conditions Not mentioned Maximum breaking force
E5 Genna et al. [45] 2003 Implant embedded in bone block CAD: a resin block model was constructed by CAD Nil (in this FEM, there was a resin model only and no living tissue simulation) Cyclic mechanical fatigue testing: implant placed in epoxy resin and section of specimen examined under microscope Epoxy resin block as bone Comparison of locations of stress focus
E5 Perriard et al. [46] 2002 Different types of implant bodies and abutments embedded in resin CAD: epoxy resin as bone Nil (in this FEM, there was a resin model only and no living tissue simulation) Mechanical fatigue testing of implant model: until half of samples still survived under loading Resin block Comparison of locations of stress concentrations
F1 Bruno Salles Sotto-Maior et al. [47] 2016 A bone model of mandibular right posterior region CT: from a patient’s mandible Homogeneous isotropic linear elasticity: cortical and trabecular bone Clinical findings of bone loss at 1-year follow-up In vivo: radiographic films of patients Mechanoregulatory tissue model was employed to monitor the morphological changes in bone subjected to biomechanical loading
F1 Wang et al. [48] 2013 A 3D model of maxillary bone CT image of maxillary bone section missing both central incisors Homogeneous isotropic linear elasticity Radiographs qualitatively compared regarding resemblance between computational remodeling results and clinical data In vivo: radiographic films of patients Comparison of variations in maxillary bone densities
F1 Choi et al. [50] 2012 Implants embedded in anterior maxilla CT: anterior maxillary bone Homogeneous isotropic linear elasticity compact bone, cancellous bone Comparison of model implant displacement under the same load with clinical outcomes in literature Literature Model implant displacement
F1 Shen et al. [51] 2010 Implant embedded in mandibular right first molar area CAD: a bone block with a cortical layer and cancellous layer was constructed by CAD Homogeneous anisotropic linear elasticity: cortical bone, cancellous bone Clinical data; comparison of implant displacement value under 20-N loading from clinical data Clinical results Implant displacement under loading
F1 Lin et al. [52] 2010 Implant embedded in mandible (cortical and cancellous bone), crown, teeth CT: in vivo CT of a segment of mandible Inhomogeneous anisotropic linear elasticity: cortical bone, cancellous bone (properties varied with density) Clinical data: comparison of bone density with other clinical follow-up X-ray images X-ray images of human X-ray images
F2 MacGinnis et al. [49] 2014 3D skull model with masked sutures CT: from 42-year-old man, 3D skull image excluding the mandible Homogeneous isotropic linear elasticity Comparison with past literature Comparison with conclusions of past literature
F2 Fanuscu et al. [53] 2004 Unilateral edentulous posterior maxilla with grafted sinus CAD: unilateral edentulous posterior maxilla with grafted sinus was constructed by computer Homogeneous isotropic linear elasticity: cortical and trabecular bone Validation with previous study by one of the authors in which photoelastic modeling with similar geometry was used Location of stress
F2 Mellal et al. [54] 2004 Cylindrical implant, bone consisting of a cancellous core coated with cortical envelope CAD: a bone section model was constructed by CAD Homogeneous isotropic linear elasticity: cortical and trabecular bone Literature: systematic search of the literature was conducted to relate the numerical predictions to existing in vivo data
F2 Zarone et al. [55] 2003 Mandible with 6 implants and prosthetic superstructure Laser: a man’s total mandible by laser digitizer Homogeneous isotropic linear elasticity: cortical and trabecular bone Data from previous experiments: comparisons of range of medial convergence during opening and protrusive movements
G Bulaqi et al. [56] 2015 Implants embedded in a bone block CT data: mandible Homogeneous isotropic linear elasticity Comparison with theoretically predicted values (calculated with the equations) values of conical to wretch torque ratio
G Vayron et al. [57] 2015 Implants embedded in a bone block CAD: cortical bone, newly formed bone, and trabecular bone Homogeneous isotropic mechanical properties Comparison with results using a 2D finite difference numerical model

Fig. 2.

Fig. 2

Hierarchy of validations based on their similarity to real biomechanical behaviors. The articles (n = 47) were categorized according to their validation method as follows: in vivo experiments in humans (n = 1) and other animals (n = 3), model experiments (n = 32), others’ clinical data and past literature (n = 9), and other software (n = 2)

Fig. 3.

Fig. 3

Proportion of dental implant FEA articles with a validation. (Left) Among totally 522 FEA articles of dental implants which we were able to access English full text up to January 2017, there are only 47 articles with a validation. (Right) The articles with a validation were categorized according to their validation method as follows levels: A, in vivo (human bodies); B, performed in vivo (heterogeneous animals); C, model experiment performed using part of a cadaver; D, model experiment performed using heterogeneous bone; E, model experiment performed using artificial materials; F, comparison with past literature; and G, performed with other software (n = 2)

Based on the validation methods described in the articles, the top portion of the validation hierarchy comprised studies that used a customized clinical method in a human for validation [10]. The next level of the hierarchy comprised studies that used models for validation, including animal models [1113] and mechanical experiments. Mechanical experiments were divided according to the material used for bone models and the techniques used for testing those models. The materials were divided into homogenous bone, heterogeneous bone, and artificial materials such as acrylic, polyurethane, plastic bone material, and others. Various validation methods were used in studies that employed mechanical testing of bone models using these specific artificial materials, such as digital image correction [11], photo-elastic stress analysis [15], and use of a strain gauge test attached to a model (this was the most commonly used method, described in 15 of 48 articles). These techniques also involved measurement of the implant displacement and fatigue testing of an implant body. The next level of the hierarchy comprised studies that used literature or clinical data from other articles to compare with results of FEA. The final level comprised studies that used other computer software for support but did not perform an actual experiment.

We classified all validation processes based on their similarity to real biomechanical behaviors into the following hierarchy (levels A to G) (Fig. 2):

  • Level A: performed in vivo (human bodies) (n = 1) [10]

The top level of the hierarchy, level A, includes in vivo methods of FEA validation conducted in humans. In 2006, Heckmann et al. [10] quantified the degree of stress that occurs in the bone around the implants as a result of fixation of cemented and screw-retained fixed partial dentures. They used a computer-aided design (CAD) model of an implant embedded in a bone block for FEA, and strain gauge experiments were performed under the same loading conditions with the use of a resin bone model and a human being for validation.

  • Level B: performed in vivo (heterogeneous animals) (n = 3) [1113]

Three studies conducted animal experiments for FEA validation. In 2009, Hou et al. [12] conducted an FEA validation study involving rats to assess the histological change in the mechanical environment surrounding loaded and unloaded implants. In 1997, Natali et al. [11] performed a validation study in which they compared the influence of axial and nonaxial forces on the bone tissue surrounding oral implants placed in dogs. Both research groups used computed tomography data and CAD techniques to create an FEA model. Similarly, in 2015, Cha et al. [13] used murine femurs to place implants with low and high insertion torques for FEA validation.

  • Level C: model experiment performed using part of a cadaver (n = 4) [1619]

  • Level D: model experiment performed using heterogeneous bone (n = 5) [2024]

The next two levels in the hierarchy comprised in vivo model experiments on a section of a cadaver (level C) and the bone of heterogeneous animals (level D). Most of these studies involved mechanical testing, such as recording strain by a strain gauge attached to a dry skull or a section of bovine, porcine, or sheep bone. Bardyn et al. [20] compared the FEA-predicted removal torque with that measured using sheep bone and polyurethane foam as a validation technique. Olsen et al. [21] scanned a porcine mandible to create an FEA model and compared the FEA-predicted implant displacement with that measured on the same porcine mandible as a validation technique. Additionally, in 2002, Huang et al. [22] determined the vibrating behavior of a dental implant under various surrounding bone conditions using bone sections from hogs and FEA. The resonance frequency was compared between the two techniques, but in this case, FEA seemed more likely to serve as a validation technique to support the results of the model experiment.

  • Level E: model experiment performed using artificial materials (n = 23) [14, 25, 2646]

Artificial materials such as acrylic resin, polyurethane, or plastic bone models were commonly used as embedded “bone” implants in validation experiments. Level E includes the use of special materials and specific methods to measure the force distribution and photoelastic resin as well as a technique called digital image correlation described by Tiossi et al. [14] in 2013. Comparisons of these artificial materials is difficult because it is challenging to determine how much more accurate one technique is over another technique. Even after subcategorizing the techniques from E1 to E5, we found that no one technique was superior to any other.

  • Level F: performed by comparison with past literature (n = 9) [4755]

Validations in this level involve comparison of FEA with clinical data (F1) or other literature (F2). Most such studies compared FEA with “similar” conditions in patients, but either the comparisons were not customized or indirect and ill-defined clinical results (e.g., bone resorption volume in length or radiographic X-ray images) were compared with force in FEA. Level F2 includes validation using past literature with similar results or conclusions that were mostly summarized in few words in the “Discussion” section of an article.

  • Level G: performed by comparison with other software (n = 2) [56, 57]

The last level, level G, includes validation performed by another type of computer software such as two-dimensional FEA, i.e., an FEA model built in a computer is validated by another computer simulation or calculated values.

Discussion

The use of FEA for dental implants and surrounding bone has increased during the past few decades. Our PubMed search using the terms “dental implants” and “finite element analysis” revealed about 450 papers published in the past 10 years. However, FEA studies of implants using validation experiments are comparatively rare. While prior studies had effectively outlined the importance of validation in biomechanical FEA, no reviews of studies that applied validation to computational biomechanics of dental implants have been performed.

Table 1 shows all studies in the literature that considered the need for validation of FEAs. According to these studies, we established a hierarchy based on the evidence level of the validations (A to G, i.e., high to low) (Fig. 2).

Level A: validation using living humans

Level B: validation using living heterogeneous animals

Levels C and D: validation using homogenous and heterogeneous bone

Level E: validation using artificial bone materials

Level F: validation using past literature

Level G: validation using other software

FEA using model verification cannot be considered to be a validation method for entire study. Model verification should be performed to ensure accurate FEA; however, finite element models verified with clinical data such as a patient’s computed tomography findings are categorized in a low level of validation or without validation. For this reason, studies that used only model verification (some studies may called it by "model validation") were not included in our review [5862].

Many recent papers [1012, 14, 15, 2531, 33, 35, 36, 39, 4145, 47, 48, 5052, 54, 55, 5874] have described the use of FEA to evaluate the stress distribution of implant fixtures and surrounding bone, with a particular focus on different fixture lengths, shapes, connection designs, and prostheses. However, most such studies [15, 5874] were performed without validation executed under the same conditions with the FEA. The following questions are worthy of consideration by oral scientists and clinicians: Can a finite element model really create a virtual condition simulating the biomechanical behavior of the craniomandibular system? To what extent can we predict biological activities with finite element models [9]?

The complexity of living organisms and internal biological phenomena is impossible to fully and precisely duplicate with individual-level specificity using a computer. However, we can evaluate the limitations of current technology and build a model with the highest level of evidence possible.

Because of the limitations of computer technology, most FEA models [7579] simplify the skeletal muscle architecture in terms of a uniform fiber length, pennation angle, and line of action and represent the architecture using a Hill-based muscle model. However, how well the modeling of skeletal muscles as one-dimensional strings represents the behavior of the full three-dimensional muscles remains unknown. Reducing the complexity of the muscles to strings entirely neglects the variations in muscle density (deformation) and structure during the complex movement of real muscle specimens, which is difficult to acquire.

This review focused on validation of FEA and established a hierarchy of validation techniques from high to low as a reference for further FEA studies. However, due to the limitations of this study, the boundary conditions and finite element method (FEM) settings were not considered. For example, some research may have involved high-level validation performed in vivo, but the original FEM model was built by CAD using only a simple flat two-layer bone and without any model verification. Some other studies used a simulated bone (computed tomography data from homogeneous, heterogeneous, or artificial materials) as an FEM geometry reference and performed the validation on that material only, without seeking to perform validation using a more realistic material. Both the use of a detailed, accurate model that closely resembles the real condition and the performance of validation to prove its accuracy are important. As computer technology has progressed, model verification has become more sophisticated and complicated; however, validation still should not be ignored.

While conducting this review, we also considered future efforts. There are two types of FEA studies: time-dependent studies, which have a validity period within which the process must take place, and time-independent studies, which have no validity period but only analyze the stress distribution at a single point in time. To date, several biomechanical studies have been published with time-independent analysis [1012, 14, 15, 2531, 33, 35, 36, 39, 4145, 47, 48, 5052, 54, 55, 5874, 80] (e.g., examination of bone resorption underneath the denture base, analysis of the instant stress distribution of a dental implant, and the bones or components of an artificial knee joint). Maeda and Wood [80] simulated a bone-dependent bone resorption process using an FEM model and shape-optimization algorithm.

To explain or analyze the mechanical properties involved in biological phenomena such as motor tasks (mastication, walking, or heart contraction), a time-dependent finite element model may provide a more realistic view. However, if time-dependent performance criteria are considered (the most common is to clarify the influence of musculoskeletal structure on function or the performance of a motor task), dynamic optimization and boundary conditions are required. This means that a much more complex model including many parameters and properties must be built, despite some of these real-world physiological data being unknown. This difficulty may explain why time-dependent models of mastication for FEA are rare.

Conclusions

  1. High-level validation of FEA using in vivo experiments is still rare in the dental implant field.

  2. It is necessary to clearly indicate the validation process of the model when a study using FEA is presented.

  3. The hierarchy proposed in this study based on the evidence level of the validations can be applied to evaluate the clinical significance of studies using FEA.

Authors’ contributions

YC and YM studied the concept and design of the study. YC, AAT, and TG contributed to the acquisition of the data. YC, AAT, and MW analyzed and interpreted the data. YC drafted the manuscript. YC and YM critically revised the article. All authors read and approved the final manuscript.

Competing interests

Yuanhan Chang, Abhijit Anil Tambe, Yoshinobu Maeda, Masahiro Wada, and Tomoya Gonda declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Contributor Information

Yuanhan Chang, Email: tyou@dent.osaka-u.ac.jp.

Abhijit Anil Tambe, Email: drabhijittambe24@gmail.com.

Yoshinobu Maeda, Email: ymaeda@dent.osaka-u.ac.jp.

Masahiro Wada, Email: masahiro@dent.osaka-u.ac.jp.

Tomoya Gonda, Email: tgonda@dent.osaka-u.ac.jp.

References

  • 1.Van Staden RC, Guan H, Loo YC. Application of the finite element method in dental implant research. Comput Methods Biomech Biomed Engin. 2006;9:257–270. doi: 10.1080/10255840600837074. [DOI] [PubMed] [Google Scholar]
  • 2.Maminskas J, Puisys A, Kuoppala R, Raustia A, Juodzbalys G. The prosthetic influence and biomechanics on peri-implant strain: a systematic literature review of finite element studies. J Oral Maxillofac Res. 2016;7:e4. doi: 10.5037/jomr.2016.7304. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Gass SI. Decision-adding models: validation, assessment and related issues for policy analysis. Oper Res. 1983;31:603–631. doi: 10.1287/opre.31.4.603. [DOI] [Google Scholar]
  • 4.Dumont ER, Grosse IR, Slater GJ. Requirements for comparing the performance of finite element models of biological structures. J Theor Biol. 2009;256:96–103. doi: 10.1016/j.jtbi.2008.08.017. [DOI] [PubMed] [Google Scholar]
  • 5.Hannam AG. Current computational modelling trends in craniomandibular biomechanics and their clinical implications. J Oral Rehabil. 2011;38:217–234. doi: 10.1111/j.1365-2842.2010.02149.x. [DOI] [PubMed] [Google Scholar]
  • 6.Boehm BW. Software engineering economics. Englewood Cliffs: Prentice Hall; 1981. [Google Scholar]
  • 7.Blottner FG. Accurate Navier-Stokes results for the hypersonic flow over a spherical nosetip. AIAA J Spacecr Rocket 1990;27:113-122.
  • 8.Anderson AE, Ellis BJ, Weiss JA. Verification, validation and sensitivity studies in computational biomechanics. Comput Methods Biomech Biomed Engin. 2007;10:171–184. doi: 10.1080/10255840601160484. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Viceconti M, Olsen S, Nolte LP, Burton K. Extracting clinically relevant data from finite element simulations. Clin Biomech (Bristol, Avon) 2005;20:451–454. doi: 10.1016/j.clinbiomech.2005.01.010. [DOI] [PubMed] [Google Scholar]
  • 10.Heckmann SM, Karl M, Wichmann MG, Winter W, Graef F, Taylor TD. Loading of bone surrounding implants through three-unit fixed partial denture fixation: a finite-element analysis based on in vitro and in vivo strain measurements. Clin Oral Implants Res. 2006;17:345–350. doi: 10.1111/j.1600-0501.2005.01177.x. [DOI] [PubMed] [Google Scholar]
  • 11.Natali AN, Meroi EA, Williams KR, Calabrese L. Investigation of the integration process of dental implants by means of a numerical analysis of dynamic response. Dent Mater. 1997;13:325–332. doi: 10.1016/S0109-5641(97)80103-0. [DOI] [PubMed] [Google Scholar]
  • 12.Hou X, Weiler MA, Winger JN, Morris JR, Borke JL. Rat model for studying tissue changes induced by the mechanical environment surrounding loaded titanium implants. Int J Oral Maxillofac Implants. 2009;24:800–807. [PubMed] [Google Scholar]
  • 13.Cha JY, Pereira MD, Smith AA, Houschyar KS, Yin X, Mouraret S, et al. Multiscale analyses of the bone-implant interface. J Dent Res. 2015;94:482–490. doi: 10.1177/0022034514566029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Tiossi R, Vasco MA, Lin L, Conrad HJ, Bezzon OL, Ribeiro RF, et al. Validation of finite element models for strain analysis of implant-supported prostheses using digital image correlation. Dent Mater. 2013;29:788–796. doi: 10.1016/j.dental.2013.04.010. [DOI] [PubMed] [Google Scholar]
  • 15.Alrbata RH, Yu W, Kyung HM. Biomechanical effectiveness of cortical bone thickness on orthodontic microimplant stability: an evaluation based on the load share between cortical and cancellous bone. Am J Orthod Dentofac Orthop. 2014;146:175–182. doi: 10.1016/j.ajodo.2014.04.018. [DOI] [PubMed] [Google Scholar]
  • 16.Nagasao T, Miyamoto J, Kawana H. Biomechanical evaluation of implant placement in the reconstructed mandible. Int J Oral Maxillofac Implants. 2009;24:999–1005. [PubMed] [Google Scholar]
  • 17.Nagasao T, Miyamoto J, Tamaki T, Kawana H. A comparison of stresses in implantation for grafted and plate-and-screw mandible reconstruction. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;109:346–356. doi: 10.1016/j.tripleo.2009.10.009. [DOI] [PubMed] [Google Scholar]
  • 18.Eser A, Akça K, Eckert S, Cehreli MC. Nonlinear finite element analysis versus ex vivo strain gauge measurements on immediately loaded implants. Int J Oral Maxillofac Implants. 2009;24:439–446. [PubMed] [Google Scholar]
  • 19.Nagasao T, Miyamoto J, Jin H, Tamaki T, Isshiki Y, Kaneko T, et al. The dynamics in implantation for patients with clefts. Cleft Palate Craniofac J. 2006;43:84–91. doi: 10.1597/04-091R1.1. [DOI] [PubMed] [Google Scholar]
  • 20.Bardyn T, Gédet P, Hallermann W, Büchler P. Prediction of dental implant torque with a fast and automatic finite element analysis: a pilot study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;109:594–603. doi: 10.1016/j.tripleo.2009.11.010. [DOI] [PubMed] [Google Scholar]
  • 21.Olsen S, Ferguson SJ, Sigrist C, Fritz WR, Nolte LP, Hallermann W, et al. A novel computational method for real-time preoperative assessment of primary dental implant stability. Clin Oral Implants Res. 2005;16:53–59. doi: 10.1111/j.1600-0501.2004.01071.x. [DOI] [PubMed] [Google Scholar]
  • 22.Huang HM, Lee SY, Yeh CY, Lin CT. Resonance frequency assessment of dental implant stability with various bone qualities: a numerical approach. Clin Oral Implants Res. 2002;13:65–74. doi: 10.1034/j.1600-0501.2002.130108.x. [DOI] [PubMed] [Google Scholar]
  • 23.Hasan I, Röger B, Heinemann F, Keilig L, Bourauel C. Influence of abutment design on the success of immediately loaded dental implants: experimental and numerical studies. Med Eng Phys. 2012;34:817–825. doi: 10.1016/j.medengphy.2011.09.023. [DOI] [PubMed] [Google Scholar]
  • 24.Chatzigianni A, Keilig L, Duschner H, Götz H, Eliades T, Bourauel C. Comparative analysis of numerical and experimental data of orthodontic mini-implants. Eur J Orthod. 2011;33:468–475. doi: 10.1093/ejo/cjr097. [DOI] [PubMed] [Google Scholar]
  • 25.Ozçelik T, Ersoy AE. An investigation of tooth/implant-supported fixed prosthesis designs with two different stress analysis methods: an in vitro study. J Prosthodont. 2007;16:107–116. doi: 10.1111/j.1532-849X.2007.00176.x. [DOI] [PubMed] [Google Scholar]
  • 26.Chou IC, Lee SY, Jiang CP. Effects of implant neck design on primary stability and overload in a type IV mandibular bone. Int J Numer Method Biomed Eng. 2014;30:1223–1237. doi: 10.1002/cnm.2653. [DOI] [PubMed] [Google Scholar]
  • 27.Chang SH, Huang SR, Huang SF, Lin CL. Mechanical response comparison in an implant overdenture retained by ball attachments on conventional regular and mini dental implants: a finite element analysis. Comput Methods Biomech Biomed Engin. 2016;19:911–921. doi: 10.1080/10255842.2015.1067687. [DOI] [PubMed] [Google Scholar]
  • 28.Rezende CE, Chase-Diaz M, Costa MD, Albarracin ML, Paschoeto G, Sousa EA, et al. Stress distribution in single dental implant system: three-dimensional finite element analysis based on an in vitro experimental model. J Craniofac Surg. 2015;26:2196–2200. doi: 10.1097/SCS.0000000000001977. [DOI] [PubMed] [Google Scholar]
  • 29.Mobilio N, Stefanoni F, Contiero P, Mollica F, Catapano S. Experimental and numeric stress analysis of titanium and zirconia one-piece dental implants. Int J Oral Maxillofac Implants. 2013;28:e135–e142. doi: 10.11607/jomi.2335. [DOI] [PubMed] [Google Scholar]
  • 30.Chang SH, Lin CL, Hsue SS, Lin YS, Huang SR. Biomechanical analysis of the effects of implant diameter and bone quality in short implants placed in the atrophic posterior maxilla. Med Eng Phys. 2015;2015. Article ID 943839. p. 16. [DOI] [PubMed]
  • 31.Tu MG, Hsu JT, Fuh LJ, Lin DJ, Huang HL. Effects of cortical bone thickness and implant length on bone strain and interfacial micromotion in an immediately loaded implant. Int J Oral Maxillofac Implants. 2010;25:706–714. [PubMed] [Google Scholar]
  • 32.Lin CL, Lin YH, Chang SH. Multi-factorial analysis of variables influencing the bone loss of an implant placed in the maxilla: prediction using FEA and SED bone remodeling algorithm. J Biomech. 2010;43:644–651. doi: 10.1016/j.jbiomech.2009.10.030. [DOI] [PubMed] [Google Scholar]
  • 33.Qian L, Todo M, Matsushita Y, Koyano K. Effects of implant diameter, insertion depth, and loading angle on stress/strain fields in implant/jawbone systems: finite element analysis. Int J Oral Maxillofac Implants. 2009;24:877–886. [PubMed] [Google Scholar]
  • 34.Karl M, Kelly JR. Influence of loading frequency on implant failure under cyclic fatigue conditions. Dent Mater. 2009;25:1426–1432. doi: 10.1016/j.dental.2009.06.015. [DOI] [PubMed] [Google Scholar]
  • 35.Hsu JT, Fuh LJ, Lin DJ, Shen YW, Huang HL. Bone strain and interfacial sliding analyses of platform switching and implant diameter on an immediately loaded implant: experimental and three-dimensional finite element analyses. J Periodontol. 2009;80:1125–1132. doi: 10.1902/jop.2009.090013. [DOI] [PubMed] [Google Scholar]
  • 36.Nagasawa S, Hayano K, Niino T, Yamakura K, Yoshida T, Mizoguchi T, et al. Nonlinear stress analysis of titanium implants by finite element method. Dent Mater J. 2008;27:633–639. doi: 10.4012/dmj.27.633. [DOI] [PubMed] [Google Scholar]
  • 37.Huang HL, Huang JS, Ko CC, Hsu JT, Chang CH, Chen MY. Effects of splinted prosthesis supported a wide implant or two implants: a three-dimensional finite element analysis. Clin Oral Implants Res. 2005;16:466–472. doi: 10.1111/j.1600-0501.2005.01124.x. [DOI] [PubMed] [Google Scholar]
  • 38.Iplikçioğlu H, Akça K, Cehreli MC, Sahin S. Comparison of non-linear finite element stress analysis with in vitro strain gauge measurements on a Morse taper implant. Int J Oral Maxillofac Implants. 2003;18:258–265. [PubMed] [Google Scholar]
  • 39.Chang CL, Chen CS, Yeung TC, Hsu ML. Biomechanical effect of a zirconia dental implant-crown system: a three-dimensional finite element analysis. Int J Oral Maxillofac Implants. 2012;27:e49–e57. [PubMed] [Google Scholar]
  • 40.Chang CL, Chen CS, Huang CH, Hsu ML. Finite element analysis of the dental implant using a topology optimization method. Med Eng Phys. 2012;34:999–1008. doi: 10.1016/j.medengphy.2012.06.004. [DOI] [PubMed] [Google Scholar]
  • 41.Zhiyong L, Arataki T, Shimamura I, Kishi M. The influence of prosthesis designs and loading conditions on the stress distribution of tooth-implant supported prostheses. Bull Tokyo Dent Coll. 2004;45:213–221. doi: 10.2209/tdcpublication.45.213. [DOI] [PubMed] [Google Scholar]
  • 42.Chang JZ, Chen YJ, Tung YY, Chiang YY, Lai EH, Chen WP, et al. Effects of thread depth, taper shape, and taper length on the mechanical properties of mini-implants. Am J Orthod Dentofac Orthop. 2012;141:279–288. doi: 10.1016/j.ajodo.2011.09.008. [DOI] [PubMed] [Google Scholar]
  • 43.Inglam S, Chantarapanich N, Suebnukarn S, Vatanapatimakul N, Sucharitpwatskul S, Sitthiseripratip K. Biomechanical evaluation of a novel porous-structure implant: finite element study. Int J Oral Maxillofac Implants. 2013;28:e48–e56. doi: 10.11607/jomi.1666. [DOI] [PubMed] [Google Scholar]
  • 44.Necchi S, Migliavacca F, Gastaldi D, Pizzagalli M, Del Fabbro M, Weinstein R, et al. The effect of fixture neck design in a realistic model of dental implant: a finite element approach. Comput Methods Biomech Biomed Engin. 2003;6:289–297. doi: 10.1080/10255840310001646301. [DOI] [PubMed] [Google Scholar]
  • 45.Genna F. On the effects of cyclic transversal forces on osseointegrated dental implants: experimental and finite element shakedown analyses. Comput Methods Biomech Biomed Engin. 2003;6:141–152. doi: 10.1080/1025584031000091696. [DOI] [PubMed] [Google Scholar]
  • 46.Perriard J, Wiskott WA, Mellal A, Scherrer SS, Botsis J, Belser UC. Fatigue resistance of ITI implant-abutment connectors—a comparison of the standard cone with a novel internally keyed design. Clin Oral Implants Res. 2002;13:542–549. doi: 10.1034/j.1600-0501.2002.130515.x. [DOI] [PubMed] [Google Scholar]
  • 47.Sotto-Maior BS, Mercuri EG, Senna PM, Assis NM, Francischone CE, Del Bel Cury AA. Evaluation of bone remodeling around single dental implants of different lengths: a mechanobiological numerical simulation and validation using clinical data. Comput Methods Biomech Biomed Engin. 2016;19:699–706. doi: 10.1080/10255842.2015.1052418. [DOI] [PubMed] [Google Scholar]
  • 48.Wang C, Li Q, McClean C, Fan Y. Numerical simulation of dental bone remodeling induced by implant-supported fixed partial denture with or without cantilever extension. Int J Numer Method Biomed Eng. 2013;29:1134–1147. doi: 10.1002/cnm.2579. [DOI] [PubMed] [Google Scholar]
  • 49.MacGinnis M, Chu H, Youssef G, Wu KW, Machado AW, Moon W. The effects of micro-implant assisted rapid palatal expansion (MARPE) on the nasomaxillary complex—a finite element method (FEM) analysis. Prog Orthod. 2014;15:52. doi: 10.1186/s40510-014-0052-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Choi KS, Park SH, Lee JH, Jeon YC, Yun MJ, Jeong CM. Stress distribution on scalloped implants with different microthread and connection configurations using three-dimensional finite element analysis. Int J Oral Maxillofac Implants. 2012;27:e29–e38. [PubMed] [Google Scholar]
  • 51.Shen WL, Chen CS, Hsu ML. Influence of implant collar design on stress and strain distribution in the crestal compact bone: a three-dimensional finite element analysis. Int J Oral Maxillofac Implants. 2010;25:901–910. [PubMed] [Google Scholar]
  • 52.Lin D, Li Q, Li W, Duckmanton N, Swain M. Mandibular bone remodeling induced by dental implant. J Biomech. 2010;43:287–293. doi: 10.1016/j.jbiomech.2009.08.024. [DOI] [PubMed] [Google Scholar]
  • 53.Fanuscu MI, Vu HV, Poncelet B. Implant biomechanics in grafted sinus: a finite element analysis. J Oral Implantol. 2004;30:59–68. doi: 10.1563/0.674.1. [DOI] [PubMed] [Google Scholar]
  • 54.Mellal A, Wiskott HW, Botsis J, Scherrer SS, Belser UC. Stimulating effect of implant loading on surrounding bone. Comparison of three numerical models and validation by in vivo data. Clin Oral Implants Res. 2004;15:239–248. doi: 10.1111/j.1600-0501.2004.01000.x. [DOI] [PubMed] [Google Scholar]
  • 55.Zarone F, Apicella A, Nicolais L, Aversa R, Sorrentino R. Mandibular flexure and stress build-up in mandibular full-arch fixed prostheses supported by osseointegrated implants. Clin Oral Implants Res. 2003;14:103–114. doi: 10.1034/j.1600-0501.2003.140114.x. [DOI] [PubMed] [Google Scholar]
  • 56.Bulaqi HA, Mousavi Mashhadi M, Geramipanah F, Safari H, Paknejad M. Effect of the coefficient of friction and tightening speed on the preload induced at the dental implant complex with the finite element method. J Prosthet Dent. 2015;113:405–411. doi: 10.1016/j.prosdent.2014.09.021. [DOI] [PubMed] [Google Scholar]
  • 57.Vayron R, Nguyen VH, Bosc R, Naili S, Haïat G. Finite element simulation of ultrasonic wave propagation in a dental implant for biomechanical stability assessment. Biomech Model Mechanobiol. 2015;14:1021–1032. doi: 10.1007/s10237-015-0651-7. [DOI] [PubMed] [Google Scholar]
  • 58.Cicciù M, Cervino G, Bramanti E, Lauritano F, Lo Gudice G, Scappaticci L, et al. FEM analysis of mandibular prosthetic overdenture supported by dental implants: evaluation of different retention methods. Comput Math Methods Med. 2015;2015:16. http://dx.doi.org/10.1155/2015/943839. [DOI] [PMC free article] [PubMed]
  • 59.Topkaya T, Solmaz MY. The effect of implant number and position on the stress behavior of mandibular implant retained overdentures: a three-dimensional finite element analysis. J Biomech. 2015;48:2102–2109. doi: 10.1016/j.jbiomech.2015.03.006. [DOI] [PubMed] [Google Scholar]
  • 60.Costa C, Peixinho N, Silva JP, Carvalho S. Study and characterization of the crest module design: a 3D finite element analysis. J Prosthet Dent. 2015;113:541–547. doi: 10.1016/j.prosdent.2014.12.008. [DOI] [PubMed] [Google Scholar]
  • 61.Arat Bilhan S, Baykasoglu C, Bilhan H, Kutay O, Mugan A. Effect of attachment types and number of implants supporting mandibular overdentures on stress distribution: a computed tomography-based 3D finite element analysis. J Biomech. 2015;48:130–137. doi: 10.1016/j.jbiomech.2014.10.022. [DOI] [PubMed] [Google Scholar]
  • 62.Abou-Emara M, Spintig T, Lackmann J, Müller WD. Finite element analysis of the biomechanical effects of PEEK dental implants on the peri-implant bone. J Biomech. 2015;48:1–7. doi: 10.1016/j.jbiomech.2014.11.017. [DOI] [PubMed] [Google Scholar]
  • 63.Capek L, Simunek A, Henys P, Dzan L. The role of implant’s surface treatment to its preload. Comput Methods Biomech Biomed Engin. 2014;17(Suppl 1):8–9. doi: 10.1080/10255842.2014.931051. [DOI] [PubMed] [Google Scholar]
  • 64.Brunski JB. Biomechanical aspects of the optimal number of implants to carry a cross-arch full restoration. Eur J Oral Implantol. 2014;7(Suppl 2):S111–S131. [PubMed] [Google Scholar]
  • 65.Harirforoush R, Arzanpour S, Chehroudi B. The effects of implant angulation on the resonance frequency of a dental implant. Med Eng Phys. 2014;36:1024–1032. doi: 10.1016/j.medengphy.2014.05.007. [DOI] [PubMed] [Google Scholar]
  • 66.Xiao W, Li Z, Shen S, Chen S, Wang Y, Wang J. Theoretical role of adjunctive implant positional support in stress distribution of distal-extension mandibular removable partial dentures. Int J Prosthodont. 2014;27:579–581. doi: 10.11607/ijp.3866. [DOI] [PubMed] [Google Scholar]
  • 67.Kim S, Kim S, Choi H, Woo D, Park YB, Shim JS, et al. A three-dimensional finite element analysis of short dental implants in the posterior maxilla. Int J Oral Maxillofac Implants. 2014;29:e155–e164. doi: 10.11607/jomi.3210. [DOI] [PubMed] [Google Scholar]
  • 68.van Staden RC, Li X, Guan H, Johnson NW, Reher P, Loo YC. A finite element study of short dental implants in the posterior maxilla. Int J Oral Maxillofac Implants. 2014;29:e147–e154. doi: 10.11607/jomi.3234. [DOI] [PubMed] [Google Scholar]
  • 69.Piotrowski B, Baptista AA, Patoor E, Bravetti P, Eberhardt A, Laheurte P. Interaction of bone-dental implant with new ultra low modulus alloy using a numerical approach. Mater Sci Eng C Mater Biol Appl. 2014;38:151–160. doi: 10.1016/j.msec.2014.01.048. [DOI] [PubMed] [Google Scholar]
  • 70.dos Santos MB, Bacchi A, Correr-Sobrinho L, Consani RL. The influence of clip material and cross sections of the bar framework associated with vertical misfit on stress distribution in implant-retained overdentures. Int J Prosthodont. 2014;27:26–32. doi: 10.11607/ijp.3627. [DOI] [PubMed] [Google Scholar]
  • 71.Santiago Junior JF, Pellizzer EP, Verri FR, de Carvalho PS. Stress analysis in bone tissue around single implants with different diameters and veneering materials: a 3-D finite element study. Mater Sci Eng C Mater Biol Appl. 2013;33:4700–4714. doi: 10.1016/j.msec.2013.07.027. [DOI] [PubMed] [Google Scholar]
  • 72.Ding X, Liao SH, Zhu XH, Wang HM. Influence of orthotropy on biomechanics of peri-implant bone in complete mandible model with full dentition. Biomed Res Int. 2014;2014:709398. doi: 10.1155/2014/709398. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Gačnik F, Ren Z, Hren NI. Modified bone density-dependent orthotropic material model of human mandibular bone. Med Eng Phys. 2014;36:1684–1692. doi: 10.1016/j.medengphy.2014.09.013. [DOI] [PubMed] [Google Scholar]
  • 74.Kang N, Wu YY, Gong P, Yue L, Ou GM. A study of force distribution of loading stresses on implant-bone interface on short implant length using 3-dimensional finite element analysis. Oral Surg Oral Med Oral Pathol Oral Radiol. 2014;118:519–523. doi: 10.1016/j.oooo.2014.05.021. [DOI] [PubMed] [Google Scholar]
  • 75.Chi SW, Hodgson J, Chen JS, Reggie Edgerton V, Shin DD, Roiz RA, et al. Finite element modeling reveals complex strain mechanics in the aponeuroses of contracting skeletal muscle. J Biomech. 2010;43:1243–1250. doi: 10.1016/j.jbiomech.2010.01.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Yucesoy CA, Koopman BH, Huijing PA, Grootenboer HJ. Three-dimensional finite element modeling of skeletal muscle using a two-domain approach: linked fiber-matrix mesh model. J Biomech. 2002;35:1253–1262. doi: 10.1016/S0021-9290(02)00069-6. [DOI] [PubMed] [Google Scholar]
  • 77.Koolstra JH, van Eijden TM. A method to predict muscle control in the kinematically and mechanically indeterminate human masticatory system. J Biomech. 2001;34:1179–1188. doi: 10.1016/S0021-9290(01)00053-7. [DOI] [PubMed] [Google Scholar]
  • 78.Koolstra JH, van Eijden TM. Combined finite-element and rigid-body analysis of human jaw joint dynamics. J Biomech. 2005;38:2431–2439. doi: 10.1016/j.jbiomech.2004.10.014. [DOI] [PubMed] [Google Scholar]
  • 79.Houdijk H, Bobbert MF, de Haan A. Evaluation of a Hill based muscle model for the energy cost and efficiency of muscular contraction. J Biomech. 2006;39:536–543. doi: 10.1016/j.jbiomech.2004.11.033. [DOI] [PubMed] [Google Scholar]
  • 80.Maeda Y, Wood WW. Finite element method simulation of bone resorption beneath a complete denture. J Dent Res. 1989;68:1370–1373. doi: 10.1177/00220345890680091601. [DOI] [PubMed] [Google Scholar]

Articles from International Journal of Implant Dentistry are provided here courtesy of Springer

RESOURCES