Skip to main content
Iranian Journal of Pharmaceutical Research : IJPR logoLink to Iranian Journal of Pharmaceutical Research : IJPR
. 2017 Autumn;16(4):1415–1423.

Preclinical Evaluation of 68Ga-MAA from Commercial Available 99mTc-MAA Kit

Saeed Shanehsazzadeh a,*, Amir Reza Jalilian a, Afsaneh Lahooti b, Parham Geramifar c, Davood Beiki c, Hassan Yousefnia a, Amir Rabiee a, Mohammad Mazidi a, Seyedeh Fatemeh Mirshojaei a, Stephan Maus d
PMCID: PMC5843303  PMID: 29552050

Abstract

99mTc-Macroaggregated Albumin (99mTc-MAA) has been used as a perfusion agent. This study described development of the 68Ga-MAA via commercially available kits from Pars-Isotopes Company as a 99mTc-MAA kit. 68Ge/68Ga generator was eluted with suprapure HCl (0.6 M, 6 mL) in 0.5 mL fractions. The two fractions with the highest 68GaCl3 activity were generally used for labeling purposes. After labeling, the final product was centrifuged 2 times to purify the solution. Five rats were sacrificed at each exact time interval (from 15 min to 2 h post injection) and the percentage of injected dose per gram (%ID/g) of each organ was measured by direct counting from 11 harvested organs of rats. The RTLC showed that labeling yields before centrifuges were 90% and 95% for Pars-Isotopes and GE kits, respectively and after centrifuges, they became 100%. The microscopic size examination showed a shift in the particle sizes post centrifuges and the biodistribution data revealed the efficiency and benefits of centrifuges in terms of preventing the of liver and bone marrow uptakes especially for Pars-Isotopes kits. Our results showed that after centrifuges of the final product, the lung uptakes increased from 89% to more than 97% of %ID/g after 5 min post injections. The whole procedure took less than 25 min from elution to the final product. Since 99mTc-MAA remained longer than 68Ga-MAA in the lung and 68Ga-MAA showed better image qualities, using 68Ga-MAA is recommended.

Key Words: Biodistribution, Macroaggregated Albumin, 68Ga, Imaging, pulmonary ventilation/perfusion scan

Introduction

A pulmonary ventilation/perfusion scan (V/Q scan) involves two nuclear scan tests to measure breathing (ventilation) and circulation (perfusion) in all areas of the lungs. Nowadays, during the perfusion scan, the intravenous injection (i.v.) of radioactive technetium macro aggregated albumin (99mTc-MAA). A gamma camera acquires the images for both phases of the study. A SPECT image can also be taken following an injection of commercially used particles are MAA, which are labeled with 99mTc (1, 2). These micro particles are 15–100 μm in size and lodge in the pulmonary capillaries and in the pre-capillary arterioles. The particle distribution accurately defines regional lung perfusion. When performing the study, an important factor is the number of particles injected and the size distribution of the particles (1). A minimum of 60,000 particles are required to obtain uniform distribution of activity reflecting regional perfusion (3). Normally, about 400,000 labeled particles are injected. Bearing in mind that there are over 280 billion pulmonary capillaries and 300 million pre-capillary arterioles, the administration of up to 400,000 particles will result in obstruction of only a very small fraction of pulmonary vessels. A reduction in the number of particles administered to between 100,000 and 200,000 is recommended in patients with known pulmonary hypertension, the right to left heart shunt or after single lung transplantation. In infants and children, the number of particles may be further reduced in accordance with patient’s weight (4).

Due to some physical limitations for SPECT such as spatial resolution, it would be better to perform the scan with positron emission tomography (PET) rather than conventional gamma camera scintigraphy (5, 6). This has been performed with Gallium-68 labeled carbon nanoparticles (Galligas) using a conventional Technegas machine for ventilation images, and with Gallium-68 labeled MAA (68Ga-MAA) for perfusion images. PET has multiple potential benefits including superior resolution, speed and quantification (5).

68Ga is an excellent positron emitting radioisotope suitable for clinical PET scan applications (7). The physical characteristics of this radioisotope (positron emission (89%), low abundance of 1,077 keV photon emission (3.22%) and the relatively short half-life (= 67.71 min) (8). permit PET applications of the 68Ga-radiopharmacuticals, while maintaining an acceptable radiation dose to the patient. Besides to these specifications, the cyclotron independent availability of 68Ga from a 68Ge/68Ga generator at a reasonable cost makes it an attractive and realistic option for countries with limited or no cyclotron facilities.

Although there are several studies that compare the results of 99mTc-MAA with 68Ga-MAA but there is still lack of data about the distribution of 68Ga-MAA in the body especially in the first minutes post injections and the clearance of 68Ga-MAA from the blood (9-11).

In continues to our previous studies and with the aim of developing new diagnostic agents in our country (12-14), in this study, we tried to develop and optimize the production of 68Ga-MAA via commercially available kits from GE and Pars-Isotopes kits for 99mTc-MAA kits.

Experimental

68Ge/68Ga generator was eluted with suprapure HCl (0.6 M, 6 mL) in 0.5 mL fractions. The two fractions with the highest 68GaCl3 (700-900 MBq) activity were generally used for labeling purposes. The MAA kits were obtained from Pars-Isotope Company (TCK-Pars-1800) which contain 3mg HAS (human serum albumin). All other chemical reagents were purchased from Sigma-Aldrich Chemical Co. U.K. Whatman paper No. 2 was obtained from Whatman (UK). Radio-chromatography was performed by Whatman paper using a thin layer chromatography scanner, Bioscan AR 2000, Paris, France. The activity of the samples were measured by a p-type coaxial high-purity germanium (HPGe) detector (model: EGPC 80-200R) coupled with a multichannel analyzer system. Calculations were based on the 511 keV peak for 68Ga. All values were expressed as mean ± standard deviation (Mean ± SD). Animal studies were performed in accordance with the United Kingdom biological council´s guidelines on the use of living animals in scientific investigations, second edition (15).

Size measurements

The particle size distributions were measured via sedimentation in Andreasen cylinders and via laser diffraction (using the Fraunhofer approximation) on a Fritsch Particle Sizer Analysette 22 Nanotec (Fritsch Laborgerätebau GmbH, Germany). The size of the particles was measured pre and post labeling.

Preparation of MAA from MAA kits

The general procedure for the preparation of 99mTc labeled commercial human serum MAA kits was based on and similar to the process described by previous studies (11, 16, 17). Each kit was suspended in 10 mL of 0.9% sodium chloride and vigorously mixed for 15 min to make the particles free of additional compounds such as stannous chloride dihydrates. After centrifuges (at 4000 rpm for 5 min), the supernatants were wasted and the retained particles were reconstituted in 0.3–0.5 mL sterile water and ready for radioactive labeling.

Labeling of MAA with 68 Ga

The MAA particles were mixed with 1.5 mL of 68Ge/68Ga generator eluted (contains 600MBq) and the pH adjusted to 4 with 500 mg of HEPES buffer. The mixture was allowed to react and stirred gently for 8 min at 75 °C. The labeled compound then quenched via 10 mL of sterile water. In order to purify the 68Ga-MAA and get rid of free 68Ga the sample were centrifuged again for 5 min (at 4000 rpm) and supernatant were squandered.

Determination of labeling yields

In order to determine the labeling yields, we used different methods of radio thin layer chromatography (RTLC) to determine the radioactive labeling yields. We used RTLC on silica gel impregnated glass fiber sheets as the stationary phase and 0.9% NaCl solution as the mobile phase. The 68Ga labeled with MAA remained at the start position and the free 68Ga migrated with the solvent front. The percentages of each fraction were determined relative to the total activity of the chromatogram (Figure 1). As represented in Figure 1, the labeling yield was about 95% therefore we considered to centrifuge the compound again for 5 min (at 4000 rpm). The RTLC results showed that there is no free activity.

Figure 1.

Figure 1.

The RTLC results (on the right sides) and the Fritsch Particle sizer results (left side) of the MAA a) after labeling with 68Ga b) after first centrifuge and c) after 2 times centrifuge post labeling. The X and Y axises for the Fritsch Particle sizer were the mean size diameter and the intensity, respectively

Animal Models

For biodistribution evaluation of the 68Ga-MAA, normal female wild-type rats were used (purchased from Razi Institute, Karaj, Iran). During the entire study, autoclaved food and drinking water were available ad libitum

Biodistribution and Imaging in wild-type rats

The distribution of radiolabeled complex (15, 30, 45, 60 and 120 min) among tissues was determined in wild-type rats. The total amount of radioactivity injected into each rat was measured by counting the syringe before and after injection in a dose calibrator with fixed geometry. The rats were sacrificed using the animal care protocols at selected times after injection. Blood samples were rapidly taken from the rats’ aorta after scarification. The tissues (heart, lung, intestine, skin, stomach, kidneys, spleen, liver, muscle, urine, carcass and bone) were weighed and rinsed with normal saline and their specific activities were determined with an HPGe detector (count 3 times) and the percentage of injected dose per gram (%ID/g) of each organ was measured (18, 19).

Measurement of activity

The activity in the syringes was measured before and after administration of the radiopharmaceutical with well-type ionization chamber (CRC-15R, Capintec, USA N.J.). All samples were background subtracted, the decay correction was not performed and then similar samples were averaged together (20).

For each of these measurements, three samples (from each organ) were weighed and then counted by HPGe to determine the percentage of injected dose per gram (which was equivalent to the percentage of injected activity per gram %IA/g≡%ID/g); all the organ activity measurements were normalized to injected activity. Uncertainties in the determinations were minimal because each assay collected at least 10,000 counts, which results the standard deviation (SD) of less than 1%. In all the measurements, we tried to keep same geometry and volume in order to prevent overestimation and underestimation in dose measurements. All samples became background subtracted and decay correction was not considered for all measurements and the similar samples were averaged together (18, 21).

The HPGe detector gave us the counts, therefore we used the below formula to convert the counts into the activity (22):

A(Bq)Tissue=Areat*Eff*Br Equ 1

Where t is the time of counts and Eff is the efficiency of the detector for the selected energy and Br is the decay yield of selected energy (512 keV) for the 68Ga.

The 68Ga activity concentration at time t, %ID/g (t) was then calculated as the percentage of injected activity per gram of tissue (%IA/g).

%ID/g=ATissueMTissueATotal*100 Epu 2

Where Atissue is the 68Ga activity in the sample, Mtissue is the mass of the sample and Atotal is the total activity of 68Ga injected into the rat (23, 24).

PET scan imaging

The rats were anesthetized with ketamine 50 mg/kg and xylazine 5 mg/kg; the 68Ga-MAA was injected into the rats’ tail vein (25). Static PET/CT imaging of the first rat was performed 50 min after 6.66 MBq injection of 68Ga- MAA on the SIEMENS Biograph 6 clinical PET/CT scanner (26). The decay corrected acquisition protocol was set to 15 min per bed position. Figure 3 shows the distribution of 68Ga-MAA in the first rat via PET/CT and PET MIP images.

Figure 3.

Figure 3

Static PET/CT fused image (left) and PET MIP image (right) of 68Ga-MAA in the first rat 50 min after injection show significant accumulation of 68Ga-MAA in lungs. The injected dose was 6.66 MBq

Dynamic PET/CT images were acquired for 45 min in list mode (32 bit) format a few sec after injection (Figures 4 and 5). In all of the acquisitions, the rats were placed in supine position and CT scans performed for anatomical reference and attenuation correction (spatial resolution 1.25 mm, 80 kV, 150 mAs) with a total CT scanning time of 20 sec.

Figure 4.

Figure 4

Dynamic PET/CT fused images of 68Ga-MAA in the second rat during the first minute after injection show quick absorption of 68Ga-MAA in lungs and both kidneys. The injected dose was 1.85 MBq

Figure 5.

Figure 5

Dynamic PET/CT fused images of 68Ga-MAA in the second rat after 5, 15, 30 and 40 min after injection. The activity was extracted by the kidneys to the bladder

Reconstruction was performed using the TrueX algorithm (including resolution recovery) with attenuation correction. The reconstruction settings were 2 iterations and 21 subsets to a 336×336 matrix. Images were filtered with Gaussian filter of 5mm full-width half maximum (FWHM). Transmission data were reconstructed into a matrix of equal size by means of filtered back-projection, yielding a co-registered image set (26).

Results

The size distribution results were depicted in Figure 1. As shown in Figure 1, after centrifuges the size became more uniform and after two times the uniformity became better. The average sizes of the MAA particles were 24.31 µm, 25.79 and 26.44 before and after 1 time and 2 times centrifuge, respectively.

The biodistribution of 68Ga-MAA (with one and two times centrifuge post labeling) was represented in Figure 2. As shown in Figure 2 the lung took up the majority of the activities. Our results showed that centrifuging the final product, increased lung uptakes from 89% to 97% of the total injecting dose per gram (ID/g) after 5 min post injections.

Figure 2.

Figure 2

Biodistribution of 68Ga-MAA after 1 and 2 times centrifuge post labeling (the data were decay corrected) at various time points (from 15 to 120 min) post injection

68 Ga-MAA Preclinical PET/CT Imaging

The PET/CT of 68Ga-MAA after first centrifuge was represented in Figures 3, 4 and 5

Discussion

Previous studies stated that size is an important issue on biodistribution of the particles (19, 25, 27, 28). If the particles are larger than 0.08 µm and less than 10 µm their target organs trough the i.v. injection were in the reticuloendothelial system (RES) and bones instead of Lungs (19, 27, 28). As represented in Figure 1, particle size measurements show that after first centrifuge about 1.31 % of MAA particles were less than 2 µm and about 5.2% were between 2-10 µm, which results the unwanted uptakes in liver and kidneys and about 0.6% undesirable bone marrow uptakes. The uninvited uptakes problem can be solved by centrifuge the 68Ga-MAA again; however, this purification procedure is takes 10 min more and losing the activities with the magnitude of 15-20%. As demonstrated in Figure 2 b after more purification the lung uptake will rise from 89% into 97% of %ID/g, which is in accordance with the previous studies on 68Ga-MAA (9-11), though none of the previous studies were not assess the uptakes of 68Ga-MAA before 1 h post injection. Therefore they could have some unwanted uptakes in liver or kidneys at 15 min post injection.

Although this study was in continues to the previous studies which discuss the advantages of the using 68Ga-MAA, but there is still a lack of data in terms of the kinetics of MAA in blood and lung also none of the previous studies discuss the undesirable uptakes in RES and in bone marrow. Therefore, the main advantages of this study are showing the uptakes and kinetics of 68Ga-MAA from the first minutes post injection and explain how to prevent these phenomena. As shown in Figure 3, the PET/CT and PET MIP (Maximum Intensity Projection) images of the first injected rat clearly show the significant activity concentration of 68Ga-MAA in lungs 50 min after injection. The low levels of uptake in other organs were clearly seen. During the first minute of 68Ga-MAA injection, as shown in Figure 4, quick uptake of lungs, and both kidneys are noticeable. Most of the injected activity was detected in the lungs and kidneys after tail vein injection (Figure 4) while after a few minutes, the main activity was extracted by the kidneys to the bladder as represented in Figure 5. For quantitative uptake behavior, maximum activity concentrations (Bq/mL) were measured for lungs, kidneys and bladder at different time points (as represented in Table 1).

Table 1.

Decay corrected organ maximum activity concentration (Bq/mL) of 68Ga-MAA during the first 45 min

Organ 68 Ga- MAA
1 min 5 min 10 min 15 min 20 min 25 min 35 min 40 min 45 min
Bladder 11990 45092 55371 61601 67875 54429 94294 81037 22838
R. Kidney 32354 22703 15795 14808 13282 12653 11905 10603 9561
L. Kidney 28374 20763 16618 14189 11945 12693 11362 12145 8703
Lungs 145166 145371 147475 131815 142757 122164 110887 107585 97605

As shown in table 1, from bladder to kidney maximum activity concentration ratios were increased from nearly zero in the first minute to 5.7 and 6.25 for right and left kidneys in the 45 min time point, respectively. Furthermore, the lungs to right kidney and lungs to left kidney maximum activity concentration ratios were increased from 4.5 and 5.1 in the first minute to 10.2 and 11.2 in the 45 min post injection. While initial kidneys uptake was decreasing via excretion into the bladder, table 1 clearly shows the preferential allocation of 68Ga-MAA in the lungs over the period studied.

There is another merit of 68Ga-MAA compares to the 99mTc-MAA in terms of lung uptakes were other studies showed that the lung uptake for 99mTc were about 87 and 79 after 2 and 4 h post injection, respectively this value for 68Ga-MAA is more than 90% (10, 11). This advantage let the physician prescribe the lower injected dose for 68Ga-MAA due to its higher uptakes and better physical characteristics of the PET compare to the SPECT (29).

Conclusion

As RTLC showed, the centrifuges increase labeling yield from 90-95% for Pars-Isotopes and GE kits respectively to 100%. In addition, the microscopic size examination demonstrated a shift in the particles size post centrifuge. Moreover, the biodistribution data revealed the efficiency and benefits of the centrifuge in terms of preventing liver and bone marrow uptakes especially for Pars-Isotopes kits. In conclusion, the whole procedure took less than 25 min from elution to the final product. Using 68Ga-MAA is recommended, because it remained shorter than 99mTc-MAA in the lung and showed better image qualities.

References

  • 1.Grüning T, Drake BE, Farrell SL, Nokes T. Three-year clinical experience with VQ SPECT for diagnosing pulmonary embolism diagnostic performance. Clin. Imaging . 2014;38:831–5. doi: 10.1016/j.clinimag.2014.04.003. [DOI] [PubMed] [Google Scholar]
  • 2.Mason RJ, Broaddus VC, Martin T, Gotway MB, King Jr TE, Schraufnagel D, Murray JF, Nadel JA. Murray and Nadel´s textbook of respiratory medicine: 2-volume set. Elsevier Health Sciences; 2010. [Google Scholar]
  • 3.Heck LL, Duley Jr JW. Statistical Considerations in Lung Imaging with 99mTc Albumin Particles 1. Radiology . 1974;113:675–9. doi: 10.1148/113.3.675. [DOI] [PubMed] [Google Scholar]
  • 4.Suga K, Yasuhiko K, Iwanaga H, Tokuda O, Matsunaga N. Relation between lung perfusion defects and intravascular clots in acute pulmonary embolism: assessment with breath-hold SPECT-CT pulmonary angiography images. Clin. Imaging . 2008;33:164. doi: 10.1016/j.ejrad.2007.08.023. [DOI] [PubMed] [Google Scholar]
  • 5.Hofman MS, Beauregard J-M, Barber TW, Neels OC, Eu P, Hicks RJ. 68Ga PET/CT Ventilation–Perfusion Imaging for Pulmonary Embolism: A Pilot Study with Comparison to Conventional Scintigraphy. J. Nucl. Med. . 2011;52:1513–9. doi: 10.2967/jnumed.111.093344. [DOI] [PubMed] [Google Scholar]
  • 6.Jalilian AR, Shanesazzadeh S, Rowshanfarzad P, Bolourinovin F, Majdabadi A. Biodistribution study of [61Cu] pyruvaldehyde-bis (N-4-methylthiosemicarbazone) in normal rats as a PET tracer. Nucl. Sci. Tech. . 2008;19:159–64. [Google Scholar]
  • 7.Roesch F, J Riss P. The renaissance of the 68Ge/68Ga radionuclide generator initiates new developments in 68Ga radiopharmaceutical chemistry. Curr. Top. Med. Chem. . 2010;10:1633–68. doi: 10.2174/156802610793176738. [DOI] [PubMed] [Google Scholar]
  • 8.Fazaeli Y, Jalilian AR, Mohammadpour Amini M, Majdabadi A, Rahiminejad A, Bolourinovin F, Pouladi M. Development of Ga-67 Maltolate complex as an imaging agent. Iran. J. Pharm. Res. . 2012;11:755–62. [PMC free article] [PubMed] [Google Scholar]
  • 9.Ament S, Maus S, Reber H, Buchholz H, Bausbacher N, Brochhausen C, Graf F, Miederer M, Schreckenberger M. PET lung ventilation/perfusion imaging using 68Ga aerosol (Galligas) and 68Ga-labeled macroaggregated albumin. Theranostics, Gallium-68, and Other Radionuclides. Springer; 2013. pp. 395–423. [DOI] [PubMed] [Google Scholar]
  • 10.Amor-Coarasa A, Milera A, Carvajal DA, Mcgoron AJ. 99mTc-MAA vs. 68Ga-MAA as Perfusion Agents. Biomedical Engineering Conference (SBEC), 2013 29th Southern; IEEE; 2013. pp. 131–2. [Google Scholar]
  • 11.Maus S, Buchholz H-G, Ament S, Brochhausen C, Bausbacher N, Schreckenberger M. Labelling of commercially available human serum albumin kits with 68 Ga as surrogates for 99 m Tc-MAA microspheres. Appl. Radiat. Isotopes. . 2011;69:171–5. doi: 10.1016/j.apradiso.2010.09.008. [DOI] [PubMed] [Google Scholar]
  • 12.Sadeghzadeh M, Shanehsazzadeh S, Lahooti A. Assessment of the effective absorbed dose of 4-benzyl-1-(3-[125I]-iodobenzylsulfonyl) piperidine in humans on the basis of biodistribution data of rats. Nucl. Med. Commun. . 2015;36:90–4. doi: 10.1097/MNM.0000000000000210. [DOI] [PubMed] [Google Scholar]
  • 13.Jalilian A, Shanehsazzadeh S, Akhlaghi M, Kamali-Dehghan M, Moradkhani S. Development of [111In]-DTPA-buserelin for GnRH receptor studies. Radiochim. Acta . 2010;98:113–9. [Google Scholar]
  • 14.Jalilian A, Shanehsazzadeh S, Akhlaghi M, Garoosi J, Rajabifar S, Tavakoli M. Preparation and evaluation of [67Ga]-DTPA-β-1–24-corticotrophin in normal rats. Radiochim. Acta . 2008;96:435–9. [Google Scholar]
  • 15.Council B. Guidelines on the use of living animals in scientific investigations. Biological Council; 1987. [Google Scholar]
  • 16.Even GA, Green MA. Gallium-68-labeled macroaggregated human serum albumin, 68 Ga-MAA. Int. J. Rad. Appl. Instrum. B . 1989;16:319–21. doi: 10.1016/0883-2897(89)90014-7. [DOI] [PubMed] [Google Scholar]
  • 17.Mathias CJ, Green MA. A convenient route to [68 Ga] Ga-MAA for use as a particulate PET perfusion tracer. Appl. Radiat. Isotopes. . 2008;66:1910–2. doi: 10.1016/j.apradiso.2008.06.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Lahooti A, Shanehsazzadeh S, Jalilian AR, Tavakoli MB. Assessment of effective absorbed dose of 111In-DTPA-Buserelin in human on the basis of biodistribution rat data. Radiat. Prot. Dosim. . 2013;154:1–8. doi: 10.1093/rpd/ncs137. [DOI] [PubMed] [Google Scholar]
  • 19.Shanehsazzadeh S, Lahooti A. Biodistribution of 80 nm iron oxide nanoparticles labeled with 99 mTc in Balb/c mice. Nucl. Med. Biol. . 2014;41:625. [Google Scholar]
  • 20.Shanehsazzadeh S, Lahooti A, Sadeghi HR, Jalilian AR. Estimation of human effective absorbed dose of 67Ga–cDTPA–gonadorelin based on biodistribution rat data. Nucl. Med. Commun. . 2011;32:37–43. doi: 10.1097/MNM.0b013e328340b916. [DOI] [PubMed] [Google Scholar]
  • 21.Lahooti A, Shanehsazzadeh S, Oghabian MA, Allen BJ. Assessment of human effective absorbed dose of Tc-99m-USPIO based on biodistribution rat data. J Label Compd Rad. 111 RIVER ST, HOBOKEN 07030-5774, NJ USA: WILEY-BLACKWELL; 2013. pp. S258–S. [Google Scholar]
  • 22.Jalilian AR, Sardari D, Kia L, Rowshanfarzad P, Garousi J, Akhlaghi M, Shanehsazzadeh S, Mirzaii M. Preparation, quality control and biodistribution studies of two [111In]-rituximab immunoconjugates. Sci. Pharm. 2008;76:151–70. [Google Scholar]
  • 23.Jalilian AR, Shanehsazzadeh S, Akhlaghi M, Kamali-dehghan M, Moradkhani S. Development of [In-111]-DTPA-buserelin for GnRH receptor studies. Radiochim. Acta . 2010;98:113–9. [Google Scholar]
  • 24.Jalilian AR, Shanehsazzadeh S, Akhlaghi M, Garousi J, Rajabifar S, Tavakoli MB. Preparation and biodistribution of [Ga-67]-DTPA-gonadorelin in normal rats. J. Radioanal. Nucl. Ch. . 2008;278:123–9. [Google Scholar]
  • 25.Shanehsazzadeh S, Gruettner C, Lahooti A, Mahmoudi M, Allen BJ, Ghavami M, Daha FJ, Oghabian MA. Monoclonal antibody conjugated magnetic nanoparticles could target MUC-1-positive cells in vitro but not in-vivo. Contrast. Media. Mol. Imaging . 2015;10:225–36. doi: 10.1002/cmmi.1627. [DOI] [PubMed] [Google Scholar]
  • 26.Mirzaei A, Jalilian A, Aghanejad A, Mazidi M, Yousefnia H, Shabani G, Ardaneh K, Geramifar P, Beiki D. Preparation and Evaluation of 68Ga-ECC as a PET Renal Imaging Agent. Nucl. Med. Mol. Imaging . 2015;49:208–16. doi: 10.1007/s13139-015-0323-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Shanehsazzadeh S, Oghabian MA, Lahooti A, Abdollahi M, Haeri SA, Amanlou M, Daha FJ, Allen BJ. Estimated background doses of [67Ga]-DTPA-USPIO in normal Balb/c mice as a potential therapeutic agent for liver and spleen cancers. Nucl. Med. Commun. . 2013;34:915–25. doi: 10.1097/MNM.0b013e328362d2fb. [DOI] [PubMed] [Google Scholar]
  • 28.Shanehsazzadeh S, Oghabian MA, Daha FJ, Amanlou M, Allen BJ. Biodistribution of ultra small superparamagnetic iron oxide nanoparticles in BALB mice. J. Radioanal. Nucl. Ch. . 2013;295:1517–23. [Google Scholar]
  • 29.Shanehsazzadeh S, Lahooti A, Shirmardi SP, Erfani M. Comparison of estimated human effective dose of 67Ga-and 99 mTc-labeled bombesin based on distribution data in mice. J. Radioanal. Nucl. Ch. . 2015;305:513–20. [Google Scholar]

Articles from Iranian Journal of Pharmaceutical Research : IJPR are provided here courtesy of Brieflands

RESOURCES