The Magnetohydrodynamic Effect and its Associated Material Designs for Biomedical Applications: A State-of-the-Art Review

T Stan Gregory 1, Rui Cheng 1, Guoyi Tang 2, Leidong Mao 1,+, Zion Tsz Ho Tse 1,+
PMCID: PMC5844576  NIHMSID: NIHMS837114  PMID: 29527149

Abstract

The presented article discusses recent advances in biomedical applications of classical Magnetohydrodynamics (MHD), with a focus on operating principles and associated material considerations. These applications address novel approaches to common biomedical problems from micro-particle sorting for lab-on-a-chip devices to advanced physiological monitoring techniques. 100 papers in the field of MHDs were reviewed with a focus on studies with direct biomedical applications. The body of literature was categorized into three primary areas of research including Material Considerations for MHD Applications, MHD Actuation Devices, and MHD Sensing Techniques. The state of the art in the field was examined and research topics were connected to provide a wide view of the field of biomedical MHDs. As this field develops, the need for advanced simulation and material design will continue to increase in importance in order to further expand its reach to maturity. As the field of biomedical MHDs continues to grow, advances towards micro-scale transitions will continue to be made, maintaining its clinically driven nature and moving towards real-world applications.

Keywords: Magnetohydrodynamics, electrocardiogram, micropump, mixing, drug delivery

1. Introduction

The Magnetohydrodynamic (MHD) Effect is a physical phenomenon describing the motion of a conducting fluid flowing under the influence of an external magnetic field. Its applications have been studied extensively across multiple disciples ranging from the study of solar winds[1, 2] to MHD-derived biomedical sensors[3] and actuators[47]. Despite this research, alternative uses of the MHD effect are emerging topics, particularly those of clinical significance or addressing biomedical problems.

The field of MHDs with a focus on studies with direct biomedical applications was surveyed, and literature that helps to advance the current state of the art was reviewed and presented in this body of work. The research was introduced through a theoretical presentation of the underlying physics followed by three primary research areas: (1) Macro- and Micro-Scale Theoretical Magnetohydrodynamics; (2) Material Considerations for MHD Applications; (3) MHD Actuation Devices; and (4) MHD Sensing Techniques. These research topics focus on macro- and micro-scale treatments, with applications being considered in the field of Magnetic Resonance Imaging (MRI), Microfluidics and Analytical Chemistry, and Guided Drug Delivery among others.

It is the aim of this manuscript to present the current state of the literature with regards to clinical applications of the MHD effect, and outline predictions for the further growth and development of the field.

2. Theory

The study of the MHD effect involves the coupling of electromagnetism, specifically regarding induction, to dynamics. This coupling is necessary when the material in question can behave as a conductor and through this have the ability to modify and be modified by the external magnetic fields when there is relative motion; creating a dominant physical force in conductive fluids with high volumetric flow rates and in large current carriers. The magnitude and direction of the fluid flow (v⇀) and the magnetic field ( Bo) both directly affect the magnitude and resultant direction of the MHD effect, creating a variance in induced Hall Effect voltages [8]. The following is a presentation of MHD theory through the derivation of Maxwell’s equations coupled with fluid flow. Table 1 lists a table of terms used in this derivation.

Table 1.

Terms used in derivation of MHD underlying physics

Term Description Units
v⇀ Magnitude and direction of fluid flow mL/s
L Length of fluid flow field m
Bo
Magnitude and direction of the applied magnetic field T, G
E Electric Field N/C, V/m
ρe Charge density C/m2
J Current density A/m2
ε0 Permittivity of free space F/m
μ0 Permeability of free space H/m, N/A2
μ Magnetic Permeability H/m, N/A2
c Magnetic Susceptibility --
σ Material conductivity S/m
η Magnetic Diffusivity mA2W/N
Rem Magnetic Reynold’s number --
Q Flow Rate m3/s
ρfluid Fluid density kg/m3
μV Coefficient of fluid viscosity m2/s
p Fluid Pressure N/m2
fEM External applied forces N
fL Lorentz force N
fVB Magnetophoretic force N
fE Electrostatic force N

2.1. Electromagnetism

The basis of the MHD effect must first be examined in terms of the primary magnetic field ( Bo), and Maxwell’s equations (Eqns. 14) expressed in differential form to predict the interactions between magnetic fields and electric fields, currents, and charges.

·E=ρeεo (1)
·B=0 (2)
×E=-Bt (3)
×B=μo(J+εoEt) (4)

Assuming that the medium is highly conductive (such as in blood flow [9]), the displacement current becomes negligible as the propagation of electric fields is inhibited. This allows Eqn. 4 to be further simplified to remove the displacement current (second parenthetical term). In order to define this phenomenon in terms of the external magnetic field, Ohm’s law must be employed (Eqn. 5) and further modified to include induction (Eqn. 6).

J=σE (5)
J=σ(E+v×B) (6)

In such a situation where a current is allowed to flow across the medium, an additional force is present (the Lorentz force) due to the interaction between the external magnetic field and that of the conductor (Eqn. 7). Eqn. 3 can be substituted into Eqn. 7 to yield a more relevant version of the Lorentz Force (Eqn. 8).

F=J×B (7)
F=(1μo)(×B)×B (8)

Through the derivation of the behavior of interacting magnetic fields, Maxwell’s equations can then be applied to fluid flow and MHDs.

2.2. Application to Fluid Flow

In order to fully describe the behavior of a flowing fluid in the presence of an external magnetic field, first consider a fluid channel under normal conditions and then subject it to an external applied force. The Navier-Stokes equation for an incompressible Newtonian fluid flowing through a pipe (Eqn. 9) provides a simple solution for the system under applied external electromagnetic body forces in this case.

ρ(vt+v·v)=-p+μV2v+fEM (9)

When considering the flowing molten metal or blood within the human body under the exposure to an applied magnetic field, micro-scale particulate interactions must be considered due to variances in compositions. As these solutions are typically considered well-mixed, homogeneous solutions, the bulk of flow can be considered nearly electrically neutral, and the total electromagnetic force can be expressed as a summation of Lorentz, magnetophoretic (occurring in ferrous materials), and electrostatic forces [1013] (Eqn. 10).

fEM=fL+fB+fE (10)

During the consideration of most clinical applications of the MHD effect, the applied magnetic field is uniform and static, especially during MRI, and magnetophoretic and electrostatic forces become null in Eqn. 10 [11]. This simplification in this special case, allows for the only remaining force, the Lorentz Force (Eqn. 8) to be substituted into the Navier-Stokes equation (Eqn. 11). This describes the flow of a conductive fluid when subjected to an external magnetic field.

ρ(vt+v·v)=-p+μV2v+(1μo)(×B)×B (11)

2.3. Scaling Effect and Dimensionless Number

Scaling effects in MHDs are important consideration in biomedical applications. MHD affects flows in channels that can vary from 10−6 to 10−2 meters in width [14]. Nguyen et. al. summarized basic dimensionless numbers for scaling the traditional MHD model down to this range [15]. For example, the magnetic Reynolds number characterizes magnetic advection to the magnetic diffusion (Eqn. 12) [16].

Rem=vlη,η=1μσ (12)

In traditional MHD systems, such as the sun, its flow field has strong interaction with the magnetic field, as a result, the magnetic Reynolds number is very large [17]. However, for a typical scale (10−6~10−2 m) and a velocity (10−4~10−1 m/s) in biomedical flows, the upper limit of magnetic Reynolds number is on the order of 10−2 [18]. (Here, the magnetic diffusivity is selected as 1 m2/s. [15]). Therefore, advection effect is not important in biomedical flows. Another two dimensionless numbers comparing the Lorenz force against the fraction force and the inertial force are the Hartmann number (Ha) [16]:

Ha=LorenzforceFrictionforce=BlσμV (13)

and the interaction number N [16],

N=LorenzforceInertialforce=σB2lρfluidv (14)

Both numbers decrease linearly as the dimension of the system decreases, which suggests Lorenz force does not benefit from miniaturization, and a larger magnetic field flux density may be needed for pronounced MHD effects. Finally, the governing equation for MHD (Eqn. 11) at microscale becomes,

0=-p+μV2v+J×B (15)

3. Material Considerations for MHD Applications

In order to harness the MHD phenomena in real-world applications, such as chemical mixing and physiological monitoring, several basic materials considerations must be examined. These applications become of a particular importance when dealing with very strong external magnetic fields, such as the 1.5T or 3T magnetic field employed by the majority of MRI scanners, and when interacting with biological tissue. These considerations are primarily twofold and involve the selection of materials to maintain: (1) a desired level of functionality within a strong external magnetic field; and (2) a level of bio-compatibility in which negative interactions with biological tissues do not occur over time. An overview of these considerations in presented in Table 2.

Table 2.

Overview of Biocompatible MHD Material Considerations

Material Magnetic Susceptibility (c) Electrical Conductivity (S/cm) Typical Applications Ref
Nylon Diamagnetic
c<0
10−13 Drug Delivery Devices, Skin/Cartilage [87, 88]
PTFE 10−19 to 10−18
Wood 10−14 to 10−4 N/A
Copper 6 × 107 Biosensors, Electrodes [89]
Dacron 10−25 to 10−23 Cardiovascular Prostheses [9092]
Teflon 10−25 to 10−23
Silicone (Hard Rubber) 10−14 Artificial Organ Replacement [9395]
Cellulose 10−13
Aluminum (Al2O3) Paramagnetic
0<c<1
2.5 × 106 Dental Implants
Platinum 9.4 × 106 Biosensors, Electrodes [9698]
Magnesium (WE43) 6.7 × 106 Cardiovascular Prosthseses [99, 100]
Titanium 5–7 × 106 Skeletal or Joint implants [28, 29, 33]
Cp-Ti 5–7 × 106
Ti-5Al-2.5Fe 5–7 × 106
Ti-6Al-7Nb 5–7 × 106
Ti-6Al-4V 5–7 × 106 Skeletal implants
Stainless Steel Ferromagnetic c>1 1.45 × 106 Skeletal implants [28, 29, 33]

3.1 External Magnetic Field Exposure

As previously mentioned, several external electromagnetic forces may interact with constituent system materials when reduced to the micro-scale. In the case of ferrous materials, an external magnetophoretic force becomes significant and can adversely affect predicted results and proper chemical mixing in MHD driven systems. As the strength of the magnetic field increases, it may produce high applied forces and torques on ferromagnetic or highly paramagnetic materials introduced into the field, offsetting applied MHD forces [1921]. Furthermore, in the case of micro-scale applications, any inhomogeneity in the external magnetic field may reduce method accuracy, which drives the need for the removal of ferromagnetic and highly paramagnetic materials in the design and construction of MHD actuators.

When selecting materials in these environments, first and foremost, magnetic susceptibility (c) must be examined, the measure of a material’s tendency to interact with an external magnetic field and offset its desired value [22]. The choice of material magnetic susceptibility in these applications, particularly in high-field MRI applications, must be either diamagnetic (c < 0), or slightly paramagnetic (0 < c < 0.1) [22]. Highly desirable materials to prevent interactions with the MHD effect, and produce solely an applied Lorentz force, are typically nylon, PTFE, wood, copper, aluminum (Al), titanium (Ti), ceramics, and the majority of engineering plastics [23]. Higher paramagnetic metals such as Al and Ti may produce an image artifact in MRI applications, but are generally safe for use in commercial MHD actuators and mixers [22].

During pulsed or rapid magnetic field gradient ramps, material conductivity is also an important factor to consider, especially in human-device interactions, due to induced heating and eddy currents [22]. Conventionally, human-device interactions for bio-sensing are performed using non-ferrous conducting electrodes, and may be capacitively coupled to patient tissue through the layering of conductive and non-conductive materials [24]. In either case, they are typically isolated using RF filters designed to divert RF energy from the test electrode, and are externally interfaced to with carbon lead clips and high-impedance cabling [24, 25].

As the local magnetic field strength decreases, these potential material selection issues become less of a concern, but must be accounted for during the design stages. Furthermore, the careful control of the applied magnetic field can yield the ability to place ferromagnetic materials in magnetic fringe fields, such as in the case of ferrous conducting rods or electrodes designed to produce an electric field across a magnetic field.

3.2 Bio-compatible Materials and Implants

In the case of medical applications of the MHD effect, several issues in bio-compatibility must be addressed, and only nonviable materials should be chosen for implantation or interaction with biological tissue [26]. Common biomaterials selected for implantable medical devices are dependent on the human body application, and may involve: (1) Ti or Ti alloys for skeletal or joint replacements; (2) Ti alloys, Dacron, Teflon, or similar engineering plastics for cardiovascular or blood vessel prostheses; (3) polyurethane, silicone rubber, silicone, or cellulose for artificial organ replacement; and (4) hydrogels, collagen, or platinum electrodes for sensing and ocular devices [27].

As Ti and Ti alloys maintain a relatively low magnetic susceptibility, serving as a mid-level paramagnetic material, they are often suitable for both implantable medical devices and operation within an external magnetic field. Ti alloys have conventionally been used in biomedical implants since the early 1970s, with their accepted forms and material specifications described in ASTM and ISO 5832 standards [28]. The use of Ti alloys as biomaterials increased due to their reduced elastic modulus, superior biocompatibility, high strength to weight ratio, and enhanced corrosion resistance when compared to conventional stainless steel and Co-Cr alloys [29]. The CP-Ti and Ti-6Al-4V alloys are the most common in medical applications, with CP-Ti being favored due to its surface properties which result in the spontaneous build-up of a stable and inert oxide layer [30, 31].

Due to its poor wear resistance, the Ti-6Al-4V is not suitable for bearing surface applications such as hip heads and femoral knees, without a coating or surface treatment, and furthermore Vanadium (V) and Al elements have been shown to be toxic to the human body, resulting in the development of β alloys free of V such as Ti-6Al-7Nb and Ti-5Al-2.5Fe, which yields better biocompatibility as well as additional advantages over Ti-6Al-4V such as higher fatigue strength and lower elastic modulus [3234].

Through the presented advantages of these Ti alloys as a biomaterial, and the paramagnetic characterization of some of these Ti alloys, Ti presents itself as a highly favorable material for the development of biocompatible medical devices, and in the case of medical devices and applications driven by the MHD effect in which strength is a key design factor.

4. MHD Actuation Devices

Since the discovery of the MHD effect, efforts have been made towards scaling it down to microscale and applying it in a variety of areas. Among them, MHD based micro-pumps, micro-mixers, particle and cell manipulation were developed. We will review these applications.

4.1 MHD Micro-Pumps

MHD micro-pumps were developed with a goal of applying them in applications such as drug delivery and lab-on-a-chip polymerase chain reaction (PCR) [4]. In a typical setup, MHD micro-pump drives conductive liquid through an applied Lorenz force, which can be exerted onto the liquid without any mechanical moving parts. As a result, MHD micro-pumps possess a much simpler structure and fabrication process when compared to their mechanical counterparts. Both direct current (DC) and alternate current (AC) MHD micro-pumps have been developed. Figure 1a shows a typical drawing of a MHD micro-pump, including an electrode pair and permanent magnets. The flowing direction of the liquid can be controlled via the direction of the current and magnetic field. Continuous flow devices using circular micro-channels and driven by AC MHD (Figure 1b) were also developed for thermal cycling PCR [35].

Figure 1.

Figure 1

MHD based microflow manipulations: DC operated micropumps in a straight channel (a) and a circular channel (b); AC operated micromixer in straight channels (c,e) and circular chambers(d,f); MHD micropump based sorting device (g). Images are reproduced from reference [14], [2425], [27] and [3031].

In order to use MHD micro-pumps for drug delivery, stability of the flow rate is critical. Flow rate of MHD micro-pumps depends on the current density and the magnetic flux density[4]. Since the flux density from a permanent magnet is much larger than that of an electromagnet, a DC MHD pump can achieve higher flow rates than its AC counterpart, however, the problem with DC pumps is that hydrolysis often occurs during their operation, which generates bubbles and leads to electrode corrosion. Bubble generation is a serious issue, as bubbles formed on the electrode surface can isolate the electrodes from conductive solutions, which may deteriorate the pump or even result in total failure[11]. Bubbles can also obstruct flows and create instabilities for drug delivery [36].

There are several approaches in solving the bubble problem in MHD pumps. First, bubble generation is less of a problem in AC MHD micro-pumps [37]. Second, one can create additional chambers for bubble releasing and trapping, in order to reduce the bubble generation and still take advantages of the high current density associated with DC pumps [38, 39]. Third, redox electrolyte pairs and inert electrodes can be used to minimize bubble generation [40].

4.2 MHD Micro-Mixers

Rapid and efficient mixing is critical to most microfluidic systems because many biological processes – such as DNA hybridization, cell activation, enzyme reactions, and protein synthesis – require rapid reactions that inevitably require efficient mixing of these reagents [4143]. Typically, biological reagents have relatively low diffusion coefficients; for a diffusion-based device, the mixing process can take on the order of tens of seconds or minutes. This is particularly true when the solution contains macromolecules (e.g., DNA and proteins) or large particles (e.g., bacteria and blood cells) that have diffusion coefficients orders of magnitude lower than that of most small molecules. For this purpose, MHD micro-mixers were developed.

Typically, MHD mixers use Lorenz force exerted on electrolyte solutions in order to generate complex flow and achieve mixing. When electrodes are configured to induce recirculation in the electrolyte, complex flows can be created to enhance mixing in the channel. The patterns of complex flow depend on the electrode configuration; two of such designs are shown in Figure 1c and Figure 1e [44]. In other designs, chaotic advections were induced in circular cavities, as shown in Figure 1d and Figure 1f [45]. Such complex flow and chaotic advection greatly enhance the mass transport of reagents. MHD mixers have also been integrated with an axial flow so that the mixing process was accomplished along with fluid transport [46]. MHD pumps and mixers could be integrated within one device [47], or to form complex fluidic networks[10].

4.3 Other Microscale MHD Devices and Applications

A MHD based microfluidic switcher was developed and shown in Figure 1g [48]. In this design, two AC MHD micro-pumps were integrated into the arms of a Y-shape channel. As a result, flow can be switched from one arm to the other very quickly. This opens the door to device design that can control on-chip assays. One of such assays is cell sorting, which was also demonstrated via mouse neural stem cells and neuroblastoma cells by the same group [49].

4.4. Remote Drug Delivery

The development of the MHD effect as a propulsive force, further allows for the design of advanced surgical micro-robots, taking advantage of the strong magnetic field of the MRI and inherent RF capabilities; further allowing for remote navigation and control during imaging using wireless power transfer schemas (Fig. 2a) [50]. This device was shown to generate a peak force of 0.31mN (Fig. 2b), derived based on the micro-scale considerations presented and the unique potential for MRI-compatible devices to harness the strong magnetic field of the MRI scanner, which is typically 1.5 to 3T in magnitude, allowing for large increases in the Lorentz body force applied to the capsule as compared to the much lower field strengths in similar medical applications of the MHD effect.

Figure 2.

Figure 2

Development of a MRI-compatible Magnetohydrodynamically driven endocapsule. a) Translation of the MHD effect into a source of propulsion (top), and the design of a MHD thruster for use in a MRI scanner (bottom). b) Quantification of system dynamic performance; step input response.

Material electrodes must be constructed from either redox pairs, such as FeCl2/FeCl3, or corrosion resistant titanium, which maintains a high level of biocompatibility and is easily sterilized. The outer capsule must serve to isolate the capsule from the body, without impairing body function. Polycarbonate materials are often utilized for this, due to their accepted usage in the development of medical instruments [51, 52], and their ease of use in modern rapid-prototyping technologies [53, 54]. This application of the Lorentz force as a propulsor in the environment of a MRI scanner, illustrates comparable body forces to similarly designed devices with alternate methods of propulsion, and demonstrating the device ability to operate remotely and its potential for remote drug delivery [5557].

5. MHD Sensing Techniques

In addition to macro- and micro-scale actuation techniques that have been presented, a field of MHD sensing has also emerged, using an external magnetic field to quantify fluid flow for the applications of mechanical rate sensing [3, 5860], hemodynamic monitoring[61], and advanced feedback control[62].

5.1. Mechanical Rate and Fluid Flow Sensing

Due to the physical scale of modern hardware employed in MHD-enabled actuators, the ability to develop compact sensing platforms based on this technology became an apparent goal. Early work included the development of MHD angular rate sensors to quantify angular velocity, based on a small circular channel filled with a conductive fluid [3, 59]. An external electromagnetic winding or a static magnet is configured perpendicular to the flow direction, and responsible for induced MHD interactions. In response to the rotation of the device within the magnetic field, a circumferential fluid velocity is produced and an external signal is generated which can be measured by complementary electrode pairs or a pick-up coil [6365]. The resultant signal produces a measure of angular velocity, and can be employed easily with a static magnet without an external power source. These sensors are commonly applied when the need to correct/quantify low level vibrations [66] or high angular rates [63].

During modern metallurgic processes, such as the manufacture of high quality steels, high accuracy velocity and flow measurements of molten metal are necessary metrics for effective quality control [67]. In the case of flowing molten metal, magnetic probes are often used to induce MHD signals, and directly quantify flow at each stage in the manufacturing process [6770].

5.2. Human Body Hemodynamic Sensing

As the study of MHD sensing applications continues to grow, additional efforts in flow sensing were applied to the human body, with early observations of the MHD effect being made in the strong magnetic field of the MRI scanner [7173]. In the environment of an MRI scanner, a MHD-derived voltage overlay has been observed on traditional Electrocardiograms (ECG), measures of cardiac electrical activity, known to be induced by interactions between arterial blood flow during early systole and the MRI main magnetic field (B0) [7476]. As a large component of human blood consists of erythrocytes and a blood plasma solution of sugars, fat, proteins, and salts, human blood flow tends to be conductive [9], causing positively and negatively charged blood particles which are flowing transversely to the strong magnetic field of the MRI scanner to be deflected by the Lorentz force, in turn inducing electrical voltages across vessel walls [74].

These induced MHD voltages (VMHD) can often be recorded on the surface ECG inside the MRI bore, theorizing the relationship between VMHD and arterial blood flow, which has generated preliminary studies which have been able to show correlation in timing and onset between VMHD recorded on intra-MRI ECGs and MRI-derived aortic blood (Fig. 3) [77, 78].

Figure 3.

Figure 3

Correlation between induced VMHD in intra-MRI ECG recordings and MRI-derived aortic blood flow; illustrating temporal alignment. a) Extracted VMHD with measurements of duration of MHD negative deflection and delay from QRS complex. b) Extracted MHD signal (solid line) and MRI flow (dashed line) with the subject in the supine position.

The development of early models of the MHD effect in arterial vasculature, led to the partial validation of the hypothesis, proving that VMHD of a comparable magnitude and direction could be induced by aortic blood flow, with the primary contribution to the net signal being attributed to the aortic arch, a large diameter vasculature in the human body transverse to the MRI magnetic field and with a high volumetric flow rate [74, 79]. Similarly, a phantom in-vitro model was developed to record bipolar VMHD signals (Fig. 4a) and shown to be comparable to a simplified analytical model in an MRI environment (Fig. 4b), predicting a linear dependence on field strength [80]. The phantom model was constructed using rigid plastic tubing, a fluid reservoir, and a Ventricular Assist Device, allowing the device to circulate a volume of conducting fluid while maintaining MRI-compatibility.

Figure 4.

Figure 4

Development of a MHD surrogate model for the validation of induced VMHD as a blood flow potential. a) Experimental design for the recording of induced VMHD using an EP catheter located within a flow phantom. b) Comparison of MRI flow derived VMHD (U10cycles,MRI) and measured VMHD (U10cycles) in the flow phantom over 10 cycles, illustrating flow dependence.

Dependence has been verified experimentally in human subjects by comparing directionality of induced VMHD extracted from 12-lead ECGs with the orientation of the aortic arch, illustrating an error of <3° in 4 human subjects [81]. In this environment, Ag/AgCl ECG electrodes are commonly used in conjunction with carbon clips and non-ferrous cabling and shielding to reduce the chance for the generation of RF burns on subjects during the gradient ramps which occur during MRI scanning [25]. Similarly, a single subject was rotated on the MRI scanner table in 90° increments to illustrate the dependence of extracted VMHD on the direction of blood flow (Fig. 5) [81]. Due to the distribution of vasculature in the human body, a study sought to determine contributions to the net recorded VMHD from vasculature networks. Using MRI validation, a 1-D VMHD distribution across the human body was demonstrated, showing increases in VMHD corresponding to major vasculature branches and solid organs[82].

Figure 5.

Figure 5

Assessment of dependence of induced VMHD on the direction of blood flow, achieved through subject rotation. a) Initial position of subject in the Feet-In position on the scanner table. b) Extracted VMHD vectors achieved though signal time-integration, illustrating near 90 degree rotation.

Despite the theorized correlation to blood flow, VMHD overlays onto surface ECGs has created difficulties in synchronizing heart motion through its electrical activity with cardiac MRI imaging, inducing motion based artifacts onto acquisitions [83]. To increase the ability to perform cardiac synchronization during MRI imaging, customized gating and synchronization algorithms have been developed to identify the timing of ventricular contraction as marked by the QRS complex of the ECG [83, 84]. Early correlation to blood flow, and therefore cardiac activity, has led to the development of VMHD driven MRI synchronization algorithms, based on the cyclic rhythm of the true QRS complex generated from the sinoatrial node and induced MHD from aortic blood flow, which allow for fine-tuning of image acquisition to better capture flow [8386].

5.3. Hemodynamic Measurements

Upon the examination of induced VMHD as a meaningful signal rather than a noise source which reduces accuracy in MRI synchronization [78], and the ability to extract induced VMHD from 12-lead ECGs obtained intra-MRI [25], came the drive to determine the clinical significance of induced VMHD. Early studies have shown the correlation of extracted VMHD with Stroke Volume (SV) and aortic blood flow (Fig. 6a), illustrating a < 4% error in SV estimation at rest, and a <10% error during a series of exercise stress testing (Fig. 6b) [61]. The usefulness of this technique is limited to inside the MRI scanner, with an applied magnetic field of 3 Tesla (T), however there are efforts to translate this technology into a stand-alone device for VMHD monitoring and potentially stroke volume or perfusion estimation.

Figure 6.

Figure 6

Aortic blood flow estimation using VMHD-derived methodology. a) MRI and VMHD derived flow (blue), compared to extracted VMHD components in the Vectorcardiogram frame of reference (green). b) Comparison of beat-to-beat aortic blood flow using VMHD (blue) and MRI (red) methods.

These devices utilize conventional physiological monitoring hardware with an external neodymium magnet (grade N52), including high-impedance inputs and an embedded microcontroller for signal processing, with the major difference being in the usage of stainless steel dry electrodes rather than conventional wet Ag-AgCl electrodes (Fig. 7a). Usefulness of the device has only been demonstrated preliminarily, and requires further validation and development to accurately predict SV (Fig. 7b).

Figure 7.

Figure 7

Development of a portable device for quantifying induced MHD voltages and estimated variations in blood flow during exercise stress testing. a) Typical usage of prototype portable flow monitoring device, illustrated during ECG acquisition. b) Time-integrated SV metrics during Exercise Stress Testing in a normal health subject (left) and in an athlete (right).

6. Conclusion

A review of the current state of the field of biomedical Magnetohydrodynamics was presented, and organized to illustrate the current state of the art in three primary research areas: (1) Material Considerations for MHD Applications; (2) MHD Actuation Devices; and (3) MHD Sensing Techniques. As Magnetohydrodynamics continues to be large field of study primarily focused on macro-scale interactions, increasing effort into micro-scale considerations and adaptions of this phenomenon is being produced, with a large attention to the growing medical field and the potential for clinical applications to expedite and increase the potential associated with current medical procedures and fields of study. The study and translation of these effects to clinical applications requires careful design considerations to be made as an effort to encourage the development of stand-alone clinical devices which can harness the MHD effect for micro-fluidics, particle mixing and sorting, and advanced physiological monitoring and imaging.

In order to further support the rapidly maturing field of biomedical MHDs, the growth of collaborative and interdisciplinary efforts must continue in order to meet increasing design and material constraints present in the harsh magnetic environment that may be present in MHD driven devices. These areas will revolve around sophisticated electromagnetic simulation and MEMS fabrication for advanced micro-scale design, and further collaborative efforts as these devices aim to grow in portability and translate into clinical tools.

The efficacy of these clinical and diagnostic tools must also be examined to observe how the MHD effect could predict hemodynamics as well as how its application in medical devices may affect the human body, particularly in high-risk subjects or when the subject is placed in a large external magnetic field, which may be required for some MHD technologies. These studies may subsequently influence US Food and Drug Administration regulations regarding human magnetic field exposure and its associated effects.

Furthermore, considerations into the development of MRI-compatible and diamagnetic/paramagnetic biomaterials must be made to enable the interfacing of MHD driven medical devices with the human body, as the majority of such research has been a byproduct of conventional biomaterials research and is only recently growing into a field of research. The need for high conductivity biomaterials and composite materials, designed to capacitively couple with the human body, with limited magnetic susceptibility also presents a growing challenge to limit the external forces present on MHD actuators, drug delivery vehicles, and chemical mixers.

Ultimately, while these applications of the MHD effect are technically feasible, advancement of MHD technologies should rely on a clinical need, rather than a technological push, as are the biomedical applications presented in this manuscript. This need will allow for the benefits of current and newly proposed MHD technologies to have a broad and real impact in scientific community and the field of medicine.

Acknowledgments

This material is based upon work supported by the National Science Foundation under Grants No. 1150042 (Rui Cheng and Leidong Mao). Research reported in this publication is also supported by the National Institute of General Medical Sciences of the National Institutes of Health under Award Numbers R21GM104528 (Leidong Mao). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.

Biographies

graphic file with name nihms837114b1.gif

Leidong Mao is an Associate Professor in the College of Engineering at the University of Georgia. He received his B.S. degree in Materials Science from Fudan University, Shanghai, China, in 2001. He received his Master of Science degree, and Doctor of Philosophy degree in Electrical Engineering from Yale University, New Haven, CT, in 2002 and 2007, respectively. His current research interests include magnetic liquids and its application in biomedical applications such as cell sorting and single cell study. He received the NSF Career Award in 2012, and a Young Scientist Award from International Conference of Magnetic Fluids in 2013.

graphic file with name nihms837114b2.gif

Zion Tsz Ho Tse received his Ph. D in Mechatronics in Medicine from Imperial College London. Currently he is an assistant professor in the College of Engineering and the Principal Investigator of the Medical Robotics Lab at the University of Georgia, Athens, GA, USA. Before joining UGA, he was a research fellow at Harvard Medical School and Brigham and Women’s Hospital. Dr. Tse has been involved in designing and prototyping a broad range of novel analog-digital electronic devices, most of which have been applied in numerous clinical and industrial environments.

References

  • 1.Priest ER. Solar magnetohydrodynamics. Vol. 1. Cambridge, UK: Cambridge University Press; 2003. [Google Scholar]
  • 2.Freidberg JP. Ideal magnetohydrodynamics. New York, NY: Plenum Press; 1987. [Google Scholar]
  • 3.Laughlin DR. A magnetohydrodynamic angular motion sensor for anthropomorphic test device instrumentation. SAE Technical Paper. 1989 [Google Scholar]
  • 4.Jang J, Lee SS. Theoretical and experimental study of MHD (magnetohydrodynamic) micropump. Sensors and Actuators A: Physical. 2000;80:84–89. [Google Scholar]
  • 5.Lemoff AV, Lee AP. An AC magnetohydrodynamic micropump. Sensors and Actuators B: Chemical. 2000;63:178–185. [Google Scholar]
  • 6.Louis J, Lothrop J, Brogan T. Fluid dynamic studies with a magnetohydrodynamic generator. Physics of Fluids (1958–1988) 1964;7:362–374. [Google Scholar]
  • 7.Rosa RJ. Physical principles of magnetohydrodynamic power generation. Physics of Fluids (1958–1988) 1961;4:182–194. [Google Scholar]
  • 8.Blandford R, Thorne K. Applications of Classical Physics. CA: CalTech; 2004. Magnetohydrodynamics. [Google Scholar]
  • 9.Hirsch FG, TEXTER EC, WOOD LA, BALLARD WC, Horan FE, Wright IS, et al. The Electrical Conductivity of Blood In Relationship to Erthrocyte Concentration. Blood. 1950;5:1017–1035. [PubMed] [Google Scholar]
  • 10.Bau HH, Zhu J, Qian S, Xiang Y. A magneto-hydrodynamically controlled fluidic network. Sensors and Actuators B: Chemical. 2003;88:205–216. [Google Scholar]
  • 11.Qian S, Bau HH. Magneto-hydrodynamics based microfluidics. Mechanics Research Communications. 2009;36:10–21. doi: 10.1016/j.mechrescom.2008.06.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Qin M, Bau HH. When MHD-based microfluidics is equivalent to pressure-driven flow. Microfluidics and nanofluidics. 2011;10:287–300. [Google Scholar]
  • 13.Xiang Y, Bau HH. Complex magnetohydrodynamic low-Reynolds-number flows. Physical Review E. 2003;68:016312. doi: 10.1103/PhysRevE.68.016312. [DOI] [PubMed] [Google Scholar]
  • 14.Hetsroni G, Mosyak A, Pogrebnyak E, Yarin L. Fluid flow in micro-channels. International Journal of Heat and Mass Transfer. 2005;48:1982–1998. [Google Scholar]
  • 15.Nguyen NT. Micro-magnetofluidics: interactions between magnetism and fluid flow on the microscale. Microfluidics and Nanofluidics. 2012 Jan;12:1–16. [Google Scholar]
  • 16.Davidson PA. An introduction to magnetohydrodynamics. Cambridge ; New York: Cambridge University Press; 2001. [Google Scholar]
  • 17.Léorat J, Pouquet A, Frisch U. Fully developed MHD turbulence near critical magnetic Reynolds number. Journal of Fluid Mechanics. 1981;104:419–443. [Google Scholar]
  • 18.Wang PJ, Chang CY, Chang ML. Simulation of two-dimensional fully developed laminar flow for a magneto-hydrodynamic (MHD) pump. Biosensors and Bioelectronics. 2004;20:115–121. doi: 10.1016/j.bios.2003.10.018. [DOI] [PubMed] [Google Scholar]
  • 19.Elhawary H, Zivanovic A, Rea M, Davies BL, Besant C, Young I, et al. A modular approach to MRI-compatible robotics. Engineering in Medicine and Biology Magazine, IEEE. 2008;27:35–41. doi: 10.1109/EMB.2007.910260. [DOI] [PubMed] [Google Scholar]
  • 20.Gassert R, Moser R, Burdet E, Bleuler H. MRI/fMRI-compatible robotic system with force feedback for interaction with human motion. Mechatronics, IEEE/ASME Transactions on. 2006;11:216–224. [Google Scholar]
  • 21.Shellock FG. Magnetic resonance safety update 2002: implants and devices. Journal of magnetic resonance imaging. 2002;16:485–496. doi: 10.1002/jmri.10196. [DOI] [PubMed] [Google Scholar]
  • 22.Stoianovici D. Multi-imager compatible actuation principles in surgical robotics. The International Journal of Medical Robotics and Computer Assisted Surgery. 2005;1:86–100. doi: 10.1002/rcs.19. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Gassert R, Dovat L, Lambercy O, Ruffieux Y, Chapuis D, Ganesh G, et al. A 2-DOF fMRI compatible haptic interface to investigate the neural control of arm movements. Robotics and Automation, 2006. ICRA 2006. Proceedings 2006 IEEE International Conference on; 2006. pp. 3825–3831. [Google Scholar]
  • 24.Wahlstrand CD, Hoegh TB, Hrdlicka GA, Cross TE, Jr, Olsen JM. Google Patents. 2007. Lead electrode for use in an MRI-safe implantable medical device. [Google Scholar]
  • 25.Tse Z, Dumoulin CL, Clifford GD, Schweitzer J, Qin L, Oster J, et al. A 1.5T MRI-conditional 12-lead electrocardiogram for MRI and intra-MR intervention. Magn Reson Med. 2013 Apr 11;71:1336–1347. doi: 10.1002/mrm.24744. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Williams DF. Definitions in biomaterials: proceedings of a consensus conference of the European Society for Biomaterials; Chester, England. March 3–5, 1986; Elsevier Science Limited; 1987. [Google Scholar]
  • 27.Ratner BD, Hoffman AS, Schoen FJ, Lemons JE. Biomaterials science: an introduction to materials in medicine. Academic press; 2004. [Google Scholar]
  • 28.Breme J. Titanium and titanium alloys, biomaterials of preference. Sixth World Conference on Titanium. I. 1988:57–68. [Google Scholar]
  • 29.Long M, Rack H. Titanium alloys in total joint replacement—a materials science perspective. Biomaterials. 1998;19:1621–1639. doi: 10.1016/s0142-9612(97)00146-4. [DOI] [PubMed] [Google Scholar]
  • 30.Dearnley P, Dahm K, Çimenoǧlu H. The corrosion–wear behaviour of thermally oxidised CP-Ti and Ti–6Al–4V. Wear. 2004;256:469–479. [Google Scholar]
  • 31.Velten D, Biehl V, Aubertin F, Valeske B, Possart W, Breme J. Preparation of TiO2 layers on cp-Ti and Ti6Al4V by thermal and anodic oxidation and by sol-gel coating techniques and their characterization. Journal of biomedical materials research. 2002;59:18–28. doi: 10.1002/jbm.1212. [DOI] [PubMed] [Google Scholar]
  • 32.Khan M, Williams R, Williams D. The corrosion behaviour of Ti–6Al–4V, Ti–6Al–7Nb and Ti–13Nb–13Zr in protein solutions. Biomaterials. 1999;20:631–637. doi: 10.1016/s0142-9612(98)00217-8. [DOI] [PubMed] [Google Scholar]
  • 33.Niinomi M. Mechanical properties of biomedical titanium alloys. Materials Science and Engineering: A. 1998;243:231–236. [Google Scholar]
  • 34.Semlitsch MF, Weber H, Streicher RM, Schön R. Joint replacement components made of hot-forged and surface-treated Ti-6Al-7Nb alloy. Biomaterials. 1992;13:781–788. doi: 10.1016/0142-9612(92)90018-j. [DOI] [PubMed] [Google Scholar]
  • 35.West J, Karamata B, Lillis B, Gleeson JP, Alderman J, Collins JK, et al. Application of magnetohydrodynamic actuation to continuous flow chemistry. Lab on a Chip. 2002;2:224–230. doi: 10.1039/b206756k. [DOI] [PubMed] [Google Scholar]
  • 36.Tsai NC, Sue CY. Review of MEMS-based drug delivery and dosing systems. Sensors and Actuators a-Physical. 2007 Mar 15;134:555–564. [Google Scholar]
  • 37.Woias P. Micropumps - past, progress and future prospects. Sensors and Actuators B-Chemical. 2005 Feb 14;105:28–38. [Google Scholar]
  • 38.Homsy A, Koster S, Eijkel JCT, van den Berg A, Lucklum F, Verpoorte E, et al. A high current density DC magnetohydrodynamic (MHD) micropump. Lab on a Chip. 2005;5:466–471. doi: 10.1039/b417892k. [DOI] [PubMed] [Google Scholar]
  • 39.Nguyen B, Kassegne SK. High-current density DC magenetohydrodynamics micropump with bubble isolation and release system. Microfluidics and Nanofluidics. 2008 Sep;5:383–393. [Google Scholar]
  • 40.Qian S, Bau HH. Magnetohydrodynamic flow of RedOx electrolyte. Physics of Fluids. 2005 Jun;17 [Google Scholar]
  • 41.Nguyen NT, Wu ZG. Micromixers - a review. Journal of Micromechanics and Microengineering. 2005 Feb;15:R1–R16. [Google Scholar]
  • 42.Lee CY, Chang CL, Wang YN, Fu LM. Microfluidic Mixing: A Review. International Journal of Molecular Sciences. 2011 May;12:3263–3287. doi: 10.3390/ijms12053263. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Capretto L, Cheng W, Hill M, Zhang XL. Micromixing Within Microfluidic Devices. Microfluidics: Technologies and Applications. 2011;304:27–68. doi: 10.1007/128_2011_150. [DOI] [PubMed] [Google Scholar]
  • 44.Bau HH, Zhong JH, Yi MQ. A minute magneto hydro dynamic (MHD) mixer. Sensors and Actuators B-Chemical. 2001 Oct 15;79:207–215. [Google Scholar]
  • 45.Yi MQ, Qian SZ, Bau HH. A magnetohydrodynamic chaotic stirrer. Journal of Fluid Mechanics. 2002 Oct 10;468:153–177. [Google Scholar]
  • 46.Affanni A, Chiorboli G. Development of an enhanced MHD micromixer based on axial flow modulation. Sensors and Actuators B-Chemical. 2010 Jun 3;147:748–754. [Google Scholar]
  • 47.Qian S, Bau HH. Magneto-hydrodynamic stirrer for stationary and moving fluids. Sensors and Actuators B-Chemical. 2005 May 13;106:859–870. [Google Scholar]
  • 48.Lemoff AV, Lee AP. An AC magnetohydrodynamic microfluidic switch for micro total analysis systems. Biomedical Microdevices. 2003 Mar;5:55–60. [Google Scholar]
  • 49.Wang LS, Flanagan L, Monuki E, Jeon NL, Lee AP. A magnetohydrodynamic (MHD) microlfuidic platform for cell switching. 2005 3rd IEEE/EMBS Special Topic Conference on Microtechnology in Medicine and Biology; 2005. pp. 276–279. [Google Scholar]
  • 50.Gregory TS, Wu KJ, Yu J, Box JB, Cheng R, Mao L, et al. Magnetohydrodynamic-Driven Design of Microscopic Endocapsules in MRI. IEEE/ASME Trans Mechatronics. 2015 [Google Scholar]
  • 51.Anderson JM. Biological responses to materials. Annual Review of Materials Research. 2001;31:81–110. [Google Scholar]
  • 52.Wallin R, Arscott E. A practical guide to ISO 10993-5: Cytotoxicity. Medical Device and Diagnostic Industry. 1998;20:96–98. [Google Scholar]
  • 53.Speakman S. Google Patents. 2000. 3D printing and forming of structures. [Google Scholar]
  • 54.Gross BC, Erkal JL, Lockwood SY, Chen C, Spence DM. Evaluation of 3D printing and its potential impact on biotechnology and the chemical sciences. Analytical chemistry. 2014;86:3240–3253. doi: 10.1021/ac403397r. [DOI] [PubMed] [Google Scholar]
  • 55.Burke M. Bidirectional Propulsion of Devices Along the Gastrointestinal Tract Using Electrostimulation. UCL (University College London); 2013. [Google Scholar]
  • 56.Kósa G, Jakab P, Jólesz F, Hata N. Swimming capsule endoscope using static and RF magnetic field of MRI for propulsion. Robotics and Automation, 2008. ICRA 2008. IEEE International Conference on; 2008. pp. 2922–2927. [Google Scholar]
  • 57.Simi M, Valdastri P, Quaglia C, Menciassi A, Dario P. Design, fabrication, and testing of a capsule with hybrid locomotion for gastrointestinal tract exploration. Mechatronics, IEEE/ASME Transactions on. 2010;15:170–180. [Google Scholar]
  • 58.Bialek J, Boozer AH, Mauel ME, Navratil GA. Modeling of active control of external magnetohydrodynamic instabilities. Physics of Plasmas (1994–present) 2001;8:2170–2180. [Google Scholar]
  • 59.Laughlin DR. Google Patents. 1997. Active magnetohydrodynamic rate sensor. [Google Scholar]
  • 60.Lee AP, Lemoff AV. Google Patents. 2000. Micromachined magnetohydrodynamic actuators and sensors. [Google Scholar]
  • 61.Gregory TS, Schmidt EJ, Zhang SH, Kwong RY, Stevenson WG, Oshinski J, et al. Rapid quantification of stroke volume using magnetohydrodynamic voltages in 3T MRI: a feasibility study. Journal of Cardiovascular Magnetic Resonance. 2015;17:P32. doi: 10.1161/CIRCIMAGING.115.003282. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Lurie K. Applied optimal control theory of distributed systems. Vol. 43. Springer Science & Business Media; 2013. [Google Scholar]
  • 63.Laughlin DR. Google Patents. 1991. Magnetohydrodynamic angular rate sensor for measuring large angular rates. [Google Scholar]
  • 64.Priede J, Buchenau D, Gerbeth G. Force-free and contactless sensor for electromagnetic flowrate measurements. Magnetohydrodynamics. 2009;45:451–458. [Google Scholar]
  • 65.Titterton D, Weston JL. Strapdown inertial navigation technology. Vol. 17. IET; 2004. [Google Scholar]
  • 66.Eckelkamp-Baker D, Sebesta HR, Burkhard K. Magnetohydrodynamic inertial reference system. AeroSense 2000. 2000:99–110. [Google Scholar]
  • 67.Ricou R, Vives C. Local velocity and mass transfer measurements in molten metals using an incorporated magnet probe. International Journal Of Heat And Mass Transfer. 1982;25:1579–1588. [Google Scholar]
  • 68.Eckert S, Cramer A, Gerbeth G. Magnetohydrodynamics. Springer; 2007. Velocity measurement techniques for liquid metal flows; pp. 275–294. [Google Scholar]
  • 69.Fujisaki K, Ueyama T, Toh T, Uehara M, Kobayashi S. Magnetohydrodynamic calculation for electromagnetic stirring of molten metal. Magnetics, IEEE Transactions on. 1998;34:2120–2122. [Google Scholar]
  • 70.Reed C, Picologlou B, Dauzvardis P, Bailey J. Techniques for measurement of velocity in liquid-metal MHD flows. Fusion Science and Technology. 1986;10:813–821. [Google Scholar]
  • 71.Abi-Abdallah D, Robin V, Drochon A, Fokapu O. Alterations in human ECG due to the MagnetoHydroDynamic effect: a method for accurate R peak detection in the presence of high MHD artifacts. Conf Proc IEEE Eng Med Biol Soc. 2007:1842–5. doi: 10.1109/IEMBS.2007.4352673. [DOI] [PubMed] [Google Scholar]
  • 72.Tenforde T, Gaffey C, Moyer B, Budlnger T. Cardiovascular alterations in Macaca monkeys exposed to stationary magnetic fields: experimental observations and theoretical analysis. Bioelectromagnetics. 1983;4:1–9. doi: 10.1002/bem.2250040102. [DOI] [PubMed] [Google Scholar]
  • 73.Tenforde TS, Gaffey CT, Liburdy RP, Levy L. Biological effects of magnetic fields. BIOLOGY AND MEDICINE DIVISION ANNUAL REPORT 1986. 1985:66. [Google Scholar]
  • 74.Gupta A, Weeks AR, Richie SM. Simulation of elevated T-waves of an ECG inside a static magnetic field (MRI) IEEE Trans Biomed Eng. 2008 Jul;55:1890–6. doi: 10.1109/TBME.2008.919868. [DOI] [PubMed] [Google Scholar]
  • 75.Abi Abdallah D, Drochon A, Robin V, Fokapu O. Effects of static magnetic field exposure on blood flow. The European Physical Journal Applied Physics. 2009;45:11301. [Google Scholar]
  • 76.Bhatt B, Reddy M. Theoretical Analysis of Induced Potentials due to Blood Flow under the Static Magnetic Field of MRI. Indian Journal of Biomechanics. 2009:51–55. [Google Scholar]
  • 77.Nijm G, Swiryn S, Larson A, Sahakian A. Characterization of the magnetohydrodynamic effect as a signal from the surface electrocardiogram during cardiac magnetic resonance imaging. IEEE; 2006. [Google Scholar]
  • 78.Nijm G, Swiryn S, Larson A, Sahakian A. Extraction of the magnetohydrodynamic blood flow potential from the surface electrocardiogram in magnetic resonance imaging. Med Biol Eng Comput. 2008;46:729–733. doi: 10.1007/s11517-008-0307-1. [DOI] [PubMed] [Google Scholar]
  • 79.Oster J, Llinares R, Payne S, Tse ZTH, Schmidt EJ, Clifford GD. Comparison of three artificial models of the magnetohydrodynamic effect on the electrocardiogram. Computer methods in biomechanics and biomedical engineering. 2015;18:1400–1417. doi: 10.1080/10255842.2014.909090. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Buchenberg WB, Mader W, Hoppe G, Lorenz R, Menza M, Büchert M, et al. In vitro study to simulate the intracardiac magnetohydrodynamic effect. Magnetic Resonance in Medicine. 2014 doi: 10.1002/mrm.25456. pp. n/a-n/a. [DOI] [PubMed] [Google Scholar]
  • 81.Gregory TS, Schmidt EJ, Zhang SH, Kwong RY, Stevenson WG, Murrow JR, et al. Left-Ventricular Mechanical Activation and Aortic-Arch Orientation Recovered from Magneto-Hydrodynamic Voltages Observed in 12-Lead ECGs Obtained Inside MRIs: A Feasibility Study. Annals of Biomedical Engineering. 2014;42:2480–9. doi: 10.1007/s10439-014-1109-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82.Gregory TS, Schmidt EJ, Zhang SH, Murrow JR, Zion TT. Magnetohydrodynamic Quantification of the Peripheral Perfusion Index in High-field MRI. Circulation. 2014 [Google Scholar]
  • 83.Gregory TS, Schmidt EJ, Zhang SH, Tse ZTH. 3DQRS: A method to obtain reliable QRS complex detection within high field MRI using 12-lead electrocardiogram traces. Magnetic Resonance in Medicine. 2014;71:1374–1380. doi: 10.1002/mrm.25078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Fischer SE, Wickline SA, Lorenz CH. Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions. Magn Reson Med. 1999 Aug;42:361–70. doi: 10.1002/(sici)1522-2594(199908)42:2<361::aid-mrm18>3.0.co;2-9. [DOI] [PubMed] [Google Scholar]
  • 85.Frauenrath T, Dieringer M, Patel N, Ozerdem C, Hentschel U, Renz W, et al. From Artifact to Merit: Cardiac Gated MRI at 7T and 3T Using Magneto-Hydrodynamic Effects for Synchronization. Proc Intl Soc Mag Reson Med. 2011;19 [Google Scholar]
  • 86.Frauenrath T, Fuchs K, Dieringer M, Ozerdem C, Patel N, Renz W, et al. Detailing the Use of Magnetohydrodynamic Effects for Synchronization of MRI With the Cardiac Cycle: A Feasibility Study. Journal of Magnetic Resonance Imaging. 2012;36:364–372. doi: 10.1002/jmri.23634. [DOI] [PubMed] [Google Scholar]
  • 87.Langer R, Peppas N. Present and future applications of biomaterials in controlled drug delivery systems. Biomaterials. 1981;2:201–214. doi: 10.1016/0142-9612(81)90059-4. [DOI] [PubMed] [Google Scholar]
  • 88.Nair LS, Laurencin CT. Biodegradable polymers as biomaterials. Progress in polymer science. 2007;32:762–798. [Google Scholar]
  • 89.Kang X, Mai Z, Zou X, Cai P, Mo J. A sensitive nonenzymatic glucose sensor in alkaline media with a copper nanocluster/multiwall carbon nanotube-modified glassy carbon electrode. Analytical biochemistry. 2007;363:143–150. doi: 10.1016/j.ab.2007.01.003. [DOI] [PubMed] [Google Scholar]
  • 90.Chandy T, Das GS, Wilson RF, Rao GH. Use of plasma glow for surface-engineering biomolecules to enhance bloodcompatibility of Dacron and PTFE vascular prosthesis. Biomaterials. 2000;21:699–712. doi: 10.1016/s0142-9612(99)00231-8. [DOI] [PubMed] [Google Scholar]
  • 91.Ratner BD. The catastrophe revisited: blood compatibility in the 21st century. Biomaterials. 2007;28:5144–5147. doi: 10.1016/j.biomaterials.2007.07.035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 92.White RA. The effect of porosity and biomaterial on the healing and long-term mechanical properties of vascular prostheses. ASAIO Journal. 1988;34:95–100. doi: 10.1097/00002480-198804000-00004. [DOI] [PubMed] [Google Scholar]
  • 93.Czaja W, Krystynowicz A, Bielecki S, Brown RM. Microbial cellulose—the natural power to heal wounds. Biomaterials. 2006;27:145–151. doi: 10.1016/j.biomaterials.2005.07.035. [DOI] [PubMed] [Google Scholar]
  • 94.Hou A, Zhou M, Wang X. Preparation and characterization of durable antibacterial cellulose biomaterials modified with triazine derivatives. Carbohydrate Polymers. 2009;75:328–332. [Google Scholar]
  • 95.Refojo MF. Current status of biomaterials in ophthalmology. Survey of ophthalmology. 1982;26:257–265. doi: 10.1016/0039-6257(82)90161-8. [DOI] [PubMed] [Google Scholar]
  • 96.Kotwal A, Schmidt CE. Electrical stimulation alters protein adsorption and nerve cell interactions with electrically conducting biomaterials. Biomaterials. 2001;22:1055–1064. doi: 10.1016/s0142-9612(00)00344-6. [DOI] [PubMed] [Google Scholar]
  • 97.Robblee L, McHardy J, Marston J, Brummer S. Electrical stimulation with Pt electrodes. V. The effect of protein on Pt dissolution. Biomaterials. 1980;1:135–139. doi: 10.1016/0142-9612(80)90035-6. [DOI] [PubMed] [Google Scholar]
  • 98.Yang M, Qu F, Lu Y, He Y, Shen G, Yu R. Platinum nanowire nanoelectrode array for the fabrication of biosensors. Biomaterials. 2006;27:5944–5950. doi: 10.1016/j.biomaterials.2006.08.014. [DOI] [PubMed] [Google Scholar]
  • 99.Staiger MP, Pietak AM, Huadmai J, Dias G. Magnesium and its alloys as orthopedic biomaterials: a review. Biomaterials. 2006;27:1728–1734. doi: 10.1016/j.biomaterials.2005.10.003. [DOI] [PubMed] [Google Scholar]
  • 100.Witte F, Hort N, Vogt C, Cohen S, Kainer KU, Willumeit R, et al. Degradable biomaterials based on magnesium corrosion. Current opinion in solid state and materials science. 2008;12:63–72. [Google Scholar]

RESOURCES