Abstract
The objective of this study was to assess the recommended DVH parameter (e.g., D2 cc) addition method used for combining EBRT and HDR plans, against a reference dataset generated from an EQD2‐based DVH addition method. A revised DVH parameter addition method using EBRT DVH parameters derived from each patient's plan was proposed and also compared with the reference dataset. Thirty‐one biopsy‐proven cervical cancer patients who received EBRT and HDR brachytherapy were retrospectively analyzed. A parametrial and/or paraaortic EBRT boost were clinically performed on 13 patients. Ten IMRT and 21 3DCRT plans were determined. Two different HDR techniques for each HDR plan were analyzed. Overall D2 cc and D0.1 cc OAR doses in EQD2 were statistically analyzed for three different DVH parameter addition methods: a currently recommended method, a proposed revised method, and a reference DVH addition method. The overall D2 ccEQD 2 values for all rectum, bladder, and sigmoid for a conformal, volume optimization HDR plan generated using the current DVH parameter addition method were significantly underestimated on average −5 to −8% when compared to the values obtained from the reference DVH addition technique (P < 0.01). The revised DVH parameter addition method did not present statistical differences with the reference technique (P > 0.099). When PM boosts were considered, there was an even greater average underestimation of −8~−10% for overall OAR doses of conformal HDR plans when using the current DVH parameter addition technique as compared to the revised DVH parameter addition. No statistically significant differences were found between the 3DCRT and IMRT techniques (P > 0.3148). It is recommended that the overall D2 cc EBRT doses are obtained from each patient's EBRT plan.
Keywords: generalized equivalent uniform dose, HDR plan evaluation, high dose rate brachytherapy, OAR doses, overall D2 cc dose
1. INTRODUCTION
Integration of concomitant chemotherapy, external beam radiotherapy (EBRT), and intracavitary brachytherapy (BT) is the standard of care in the curative management of locally advanced cervical cancer.1 Using a BT boost is linked with improved pelvic control2 and overall survival.2, 3 The first use of BT for the treatment of cervical cancer dates back to 1903.4 The use of three‐dimensional (3D) imaging techniques, such as computerized tomography (CT) and magnetic resonance imaging (MRI), have been rapidly replacing planar x‐ray imaging in BT treatment planning. This follows the recommendations of the Groupe Européen de Curiethérapie‐European Society for Therapeutic Radiology and Oncology (GEC‐ESTRO),5, 6, 7 the American Brachytherapy Society (ABS),8, 9 EMBRACE (An intErnational study on MRI‐guided BRachytherapy in locally Advanced CErvical cancer) protocol,10 and a recent International Commission on Radiation Units and Measurements (ICRU) Report #89.11 Volumetric dose parameters for targets and organs‐at‐risk (OARs) were introduced and used, allowing clinicians to customize isodose lines with the goal of achieving maximal coverage of the high‐risk clinical target volume (HR‐CTV) while irradiating OARs as little as possible. These adaptive, conformal BT approaches have resulted in significantly improved clinical outcomes.12 Volumetric OAR dose constraints, such as the minimal dose of the 2 cc of normal tissue with the highest dose (D2cc) or D0.1 cc, have been investigated13, 14, 15 as an alternative to conventional rectum and bladder point doses. These alternatives originated from the ICRU Report #38,16 and are mainly applicable to Point A‐based BT planning techniques. In order to integrate overall volumetric OAR doses (D2 cc and D0.1 cc) from EBRT and BT, it was recommended that the EBRT and BT doses be added even though the location of given hot‐spots (D2 cc or D0.1 cc regions) may not be identical for each of the plans. This was initially called a “worst case assumption” in the GEC‐ESTRO recommendation,7 but a worse case would occur due to intra‐fraction organ or applicator motion. The adopted EMBRACE protocol phrase for this is “DVH parameter addition”.10 In this DVH parameter addition technique, the EBRT component dose distributions (at least for the volumetric OAR parameters (D2 cc and D0.1 cc)), are assumed to be completely uniform EBRT prescription doses following the recommendations of the EMBRACE protocol.5, 7, 10
There have been efforts to accurately estimate overall doses from EBRT and HDR BT plans17, 18, 19 but he previous studies were performed using either a phantom study17 or a dosimetric planning study with no statistical analysis for six or fewer patients,18, 19 and they did not present a practical approach on how to estimate the overall OAR doses (e.g., D2 ccEQD2) without exporting and processing dose DICOM files (dose distribution (DVHEQD2) addition) or using DIR‐based DVH analysis. In this study, we present a practical revised DVH parameter addition method where the volumetric OAR parameters (e.g., D2 cc) are simply obtained from each patient's EBRT plan, instead of assuming a completely uniform EBRT prescription dose. The proposed, revised DVH parameter addition method was compared with the current DVH parameter addition method that has been used in the overall dose integration framework of GEC‐ESTRO guidelines7 and the EMBRACE protocol,10 and it assumes the completely uniform EBRT prescription dose. A dose distribution (DVHEQD2) addition was used as a reference dataset to compare those two approaches. Two different BT planning techniques (a) conventional Point A HDR plans and (b) 3 Tesla MRI‐guided, conformal, and adaptive volume optimization HDR plans were examined. In addition, the integrated, single EQD2‐based DVH, generalized equivalent uniform doses (gEUDEQD2) for the rectum, bladder, and sigmoid are presented and compared with conventional D2cc values as a potential, additional plan evaluation metric for OAR.
2. MATERIALS AND METHODS
2.A. EBRT and 3T MRI‐guided, adaptive/conformal volume optimization HDR plans
Following the approval from our institutional review board (IRB), a retrospective study was performed with 31 biopsy‐proven cervical cancer patients whose International Federation of Gynecology and Obstetrics (FIGO) stages varied from Ib to IV. All patients received EBRT and HDR BT treatments. IMRT (10 patients) or 3D‐conformal EBRT (21 patients) plans were clinically generated in Pinnacle3 v9.8 (Philips Healthcare, Inc., Thornton, CO, USA). The type of EBRT planning techniques was clinically determined and the EBRT plans were retrospectively analyzed. All EBRT plans had HDR boosts in which an HDR plan was created for each fraction using BrachyVision v11.0 (Varian Medical System, Inc. Palo Alto, CA, USA). Patients were prescribed 45 Gy in 25 fractions, and 12 patients each received a 5.4–10 Gy EBRT boost in 3–5 fractions due to the involvement of PM (13 patients) or paraaortic (PA) (5 patients) regions. Two patients received both PM and PA EBRT boosts. Standard Point A‐based HDR plans for use with a Fletcher‐Suit‐Declos tandem‐and‐ovoids (T&O) applicator (Varian Medical System, Inc.) were clinically generated according to the ABS consensus guidelines.8, 9 The prescription dose was 33–36 Gy in 5–7 fractions, typically 5.5 Gy × 5 fractions or 7 Gy × 4 fractions). The clinical Point A plans were generated on 3 Tesla T2‐ and T1‐weighted MRI data sets20 (MAGNETOM TrioTM, Siemens Medical System Inc., Erlangen, Germany). A staff physician contoured the bladder, rectum, and sigmoid structures using T2‐weighted MR images.11 A T&O applicator was reconstructed (digitized) on T1‐weighted MR images.6 The details of the HDR workflow, imaging, and planning have been previously described.21, 22, 23, 24
An adaptive/conformal volume optimization HDR plan was retrospectively created for each clinical Point A plan through a hybrid‐inverse optimization process that includes a combination of an inverse optimization and manual forward planning. The hybrid‐inverse optimization process21, 22, 25, 26 includes three main steps: (a) generate a conventional Point A plan, (b) set dose‐volume objective constraints for inverse optimization based upon the resulting DVH parameters, and (c) perform final dose shaping using graphical optimization based upon DVH parameters and isodose lines. As a last step, a physician reviews the isodose lines for each slice on coronal, sagittal, and axial views. A patient's initial EBRT plan isodose lines, PA boost, and PM boost are depicted in Fig. 1, along with two different HDR plan techniques: a Point A plan and conformal volume‐optimization plan.
Figure 1.

A case of EBRT (45 Gy) (a) with paraortic boost (PA: 7.2 Gy) (b) and parametria boost (PM: 9 Gy) (c) received with a Point A HDR plan (e) that retrospectively regenerated as a volume optimization, adaptive HDR plan (d).
2.B. Two different techniques to access overall OAR DVH parameters (D2 cc and D0.1 cc)
Two different approaches to assess overall OAR DVH parameters (e.g., D2 cc and D0.1 cc) from EBRT and each HDR plan were tested. The first is the recommended GEC‐ESTRO7 and EMBRACE protocol10 DVH parameter addition technique where a completely uniform EBRT prescription dose is assumed. EBRT DVH parameters (e.g., D2 cc) were assumed to receive the full EBRT prescription dose from the initial EBRT plan and PA boost but receive no additional doses from the PM boost due to its central block. A 4 cm central block was used for all EBRT PM boost plans. The second method is a revised DVH parameter addition technique where EBRT DVH parameters (e.g., D2 cc) are obtained from each patient's EBRT plan. In both approaches, their physical D2 cc and D0.1 cc parameters were converted into EQD2‐based values (D2ccEQD2 and D0.1ccEQD2) according to a linear‐quadratic cell survival model following GEC‐ESTRO guidelines5, 7 and the ICRU Report #89.11 The α/β ratio of 3 and repair halftime of 1.5 hr were used.11
| (1) |
Here , d, and g represent a fraction number, a dose per fraction, and an incomplete repair function that is 1 for HDR. Afterward, the DVH parameters (e.g., D2 cc) in EQD2 were added for each EBRT and HDR plan. Both approaches can be simply done using an Excel spreadsheet (Microsoft Corporation, Redmond, WA, USA) available as a template on the American Brachytherapy Society website (www.americanbrachytherapy.org).
2.C. Integrated, single EQD2‐based DVH as a reference dataset
In order to test these two different approaches, an integrated single EQD2‐based, differential DVH was generated as a reference dataset. This was done through three steps: (a) each physical dose map (i.e., dose DICOM file) was converted into an EQD2 dose map to account for the different fractionation schemes between the EBRT and HDR BT plans, (b) a differential DVH was generated from each EQD2 dose map that is EQD2‐based, differential DVH (DVHEQD2), and (c) all differential DVHEQD2 were combined to create a single, integrated differential DVHEQD2. The radiobiological plan evaluation tool, RadioBioEval, was developed in‐house as a stand‐alone software application in order to convert physical dose maps in DICOM format from EBRT treatment planning system (TPS) (Pinnacle, Philips Healthcare, Inc.,) and HDR TPS (BrachyVision, Varian Medical System, Inc.) into EQD2 dose maps, to generate a single, differential DVHEQD2 from EQD2 dose maps of EBRT and HDR plans and to evaluate an overall D2 ccEQD2, D0.1 ccEQD2, and gEUDEQD2 (see Fig. 2). A user put radiobiological parameters into “RadioBiological Parameters” of the RadioBioEval software to generate generalized EUD (gEUD) or other radiobiological metrics evaluated in the previous studies such as equivalent uniform dose (EUD) and tumor control probability (TCP)27 for tumor and normal tissue complication probability (NTCP).28 Afterward, DICOM dose files are imported through “Open Files” (seen in the left upper corner of the software, Fig. 2), then the imported file structures are presented in the bottom of Fig. 2. When “Target” is clicked, the α/β ratio of 10 is used. Otherwise, an α/β ratio of 3 is used. In this study, the α/β ratio of 3 was used for all OARs. In this demonstration case, a composite EBRT plan with EBRT boost and three HDR plans of fraction #1–#3 were imported. The DVH Graph of Fig. 2 presented the solid line DVHs of physical doses and the dashed line DVHs of EQD2 doses. When clicking “DVH Metrics” tab, the overall DVH parameters in EQD2 doses (e.g., D2 cc or D0.1 cc), referred to as the dose distribution addition (DVHEQD2), were obtained from the integrated, single, differential DVHEQD2. On “RadBio Metrics” tab in Fig. 2, gEUD and NTCP values for OARs and EUD and TCP values for tumors are presented.
Figure 2.

In‐house radiobiological evaluation tool (RadioBioEval) for integrated EBRT and each HDR brachytherapy plans through which physical EBRT and HDR DICOM dose map plans (solid lines on DVH Graph) are converted into EQD2‐dose maps and DVHs (dashed lines). In this demonstration case, a composite EBRT plan with EBRT boost and three HDR plans of fraction #1–#3 were imported.
2.D. EQD2‐based, generalized EUD (gEUDEQD2) as an additional plan evaluation metric
Currently, the GEC‐ESTRO working group,5, 7 EMBRACE protocol,10 and the ICRU report #8911 all recommend assuming maximal dose–volume parameters. D2cc values sufficiently represent each OAR's dose distribution including whole DVH. As an additional plan evaluation metric, the use of an EQD2‐based, generalized equivalent uniform dose (gEUDEQD2) was proposed. The gEUDEQD2 was obtained from the integrated, differential DVHEQD2 that was described in the previous section. The gEUDEQD2 is determined by solving the following eq. (2):
| (2) |
where n is a volume effect parameter, and EQD2 i is the differential dose bin obtained from a single integrated, differential DVHEQD2. The n values of the rectum, bladder, and sigmoid were 0.23,29 0.5,30 and 0.17,30 respectively, and were based upon the available literature. The values of gEUDEQD2 were compared with the values of current OAR plan evaluation metric, D2ccEQD2.
2.E. Statistical correlation analysis
The percent differences of the two different DVH parameter addition techniques were statistically analyzed in comparison with the reference dataset. The statistical differences between the two were also measured. The impact of EBRT techniques such as 3D conformal radiotherapy (3DCRT) and intensity‐modulated radiotherapy (IMRT) were examined for the EBRT cases without PA or PM boosts. The consequence of PM boost contributions were evaluated by comparing a DVH parameter addition with no PM boost assumption method and the revised DVH parameter addition method. All analysis was separately performed for Point A and conformal volume optimization HDR plans, along with separate statistical analysis on each rectum, bladder, and sigmoid. All P values were calculated based upon paired, two‐sample Student t‐tests.
3. RESULTS AND DISCUSSION
The overall D2ccEQD2 parameters for the rectum, bladder, and sigmoid that were obtained from a revised DVH parameter addition technique presented no statistical differences (P > 0.0981) with the reference dataset values regardless of conformal, volume optimization, and Point A HDR plans (see Table 1). The overall D0.1ccEQD2 parameters for the rectum and sigmoid also did not present statistically significant differences between the revised DVH parameter addition and the reference dataset (P > 0.3831). Only the overall D0.1ccEQD2 values of the bladder were found to be significantly higher (P < 0.001) in the revised DVH parameter addition for either Point A or conformal HDR plan with a 5.9%–7.2% (7.5–7.8 GyEQD2) average difference from the reference datasets. It is recommended that D0.1ccEQD2 values be recorded and monitored but with no dose limits.5, 7, 10, 11 When the overall D2ccEQD2 and D0.1ccEQD2 values were obtained from the current DVH parameter addition technique, all OAR values were significantly underestimated over the values of the reference dataset, regardless of HDR planning techniques (P < 0.01) (see Table 1). On average significantly lower D2ccEQD2 for all OAR were recorded than the values of the reference dataset in both conformal HDR plans (−4.7 to −8.3% (−2.8 to −5.6 GyEQD2)) and Point A plans (−2.5 to −5.7% (−1.7 to −3.8 GyEQD2)). The overall D2ccEQD2 values were underestimated over the reference dataset for the rectum, bladder, and sigmoid. The values reached as low as −11.7% (−6.9 GyEQD2), −18.5% (−13.1 GyEQD2), and −18.2% (−9.2 GyEQD2), respectively. The overall D2ccEQD2 values were also significantly underestimated over the revised DVH addition technique (P < 0.004) for conformal and Point A HDR plans, reaching −4.2 to −6.0% (−2.8 to −4.6 GyEQD2) and −4.1 to −5.8% (−3.1 to −4.8 GyEQD2), respectively (see Table 2). The current DVH parameter addition technique resulted in significantly underestimated DVH parameter values (D2ccEQD2 and D0.1ccEQD2) for all OARs regardless of Point A or conformal HDR planning technique when compared to either the revised DVH parameter addition or the dose distribution (DVHEQD2) addition technique.
Table 1.
The differences of DVH parameters (i.e., D2 cc and D0.1 cc) between either current DVH parameter addition or the proposed, revised DVH parameter addition and the reference dataset (i.e., either current DVH parameter addition or the revised addition — the reference dataset). The rectum, bladder, and sigmoid D2 ccEQD2 values of the revised DVH parameter addition did not present significant difference with the reference dataset (P > 0.097) regardless of HDR planning techniques; for both conformal volume optimization and Point A plans
| ∆ D2ccEQD2 [% (GyEQD2)] | P‐value | ∆ D0.1 ccEQD2 [% (GyEQD2)] | P‐value | |||
|---|---|---|---|---|---|---|
| Mean* | Std Dev | Mean* | Std Dev | |||
| Current DVH parameter addition with Uniform EBRT Rx Dose Assumption | ||||||
| Conformal HDR Plan | ||||||
| Rectum | −4.7 (−2.8) | 4.8 (3.1) | <0.001 | −5.0 (−3.4) | 5.1 (4.0) | <0.001 |
| Bladder | −7.4 (−5.6) | 5.9 (4.2) | <0.001 | −13.0 (−12.9) | 8.9 (8.5) | <0.001 |
| Sigmoid | −8.3 (−4.9) | 4.9 (2.6) | <0.001 | −7.9 (−5.4) | 5.2 (3.3) | <0.001 |
| Point A HDR Plan | ||||||
| Rectum | −2.5 (−1.7) | 5.2 (3.4) | 0.009 | −4.0 (−3.3) | 7.3 (6.3) | 0.007 |
| Bladder | −4.1 (−3.2) | 6.0 (5.4) | 0.002 | −11.3 (−12.8) | 6.1 (8.3) | <0.001 |
| Sigmoid | −5.7 (−3.8) | 6.2 (4.0) | <0.001 | −5.8 (−5.1) | 7.2 (7.3) | <0.001 |
| Revised DVH parameter addition | ||||||
| Conformal HDR Plan | ||||||
| Rectum | −0.1 (+0.02) | 5.6 (3.4) | 0.980 | +0.2 (+0.3) | 5.2 (3.8) | 0.720 |
| Bladder | −1.2 (−1.0) | 6.2 (4.9) | 0.332 | −7.2 (−7.8) | 9.1 (9.3) | <0.001 |
| Sigmoid | −1.6 (−0.8) | 6.6 (3.8) | 0.297 | −1.4 (−0.8) | 6.3 (4.4) | 0.383 |
| Point A HDR Plan | ||||||
| Rectum | +1.8 (+1.4) | 6.9 (4.6) | 0.098 | +0.7 (+0.6) | 7.9 (7.0) | 0.621 |
| Bladder | +1.5 (+1.6) | 6.4 (6.0) | 0.152 | −5.9 (−7.5) | 6.9 (8.9) | <0.001 |
| Sigmoid | +0.5 (+0.5) | 7.1 (5.0) | 0.551 | +0.2 (−0.2) | 7.5 (7.7) | 0.872 |
Negative sign refers the values of current/revised DVH parameter addition underestimate when compared to the reference datasets.
Table 2.
The differences of DVH parameters (i.e., D2 cc and D0.1 cc) when current DVH parameter addition and the proposed, revised DVH parameter addition were directly compared (i.e., current DVH parameter addition — the revised addition). The rectum, bladder, and sigmoid D2 ccEQD2 and D0.1 ccEQD2 values of the current DVH parameter addition were statistically significantly underestimated when compared to the revised DVH parameter addition (P < 0.0027) regardless of HDR planning techniques; for both conformal volume optimization and Point A plans
| ∆ D2 ccEQD2 [% (GyEQD2)] | P‐value | ∆ D0.1 ccEQD2 [% (GyEQD2)] | P‐value | |||
|---|---|---|---|---|---|---|
| Mean* | Std Dev | Mean* | Std Dev | |||
| Conformal HDR Plan | ||||||
| Rectum | −4.2 (−2.8) | 6.5 (4.1) | 0.003 | −4.7 (−3.7) | 6.0 (4.3) | <0.001 |
| Bladder | −5.6 (−4.6) | 5.4 (4.5) | <0.001 | −5.1 (−5.1) | 4.5 (4.5) | <0.001 |
| Sigmoid | −6.0 (−4.0) | 7.2 (4.2) | <0.001 | −6.0 (−4.6) | 6.2 (4.1) | <0.001 |
| Point A HDR Plan | ||||||
| Rectum | −4.1 (−3.1) | 5.5 (3.7) | <0.001 | −4.4 (−3.9) | 5.3 (4.0) | <0.001 |
| Bladder | −5.3 (−4.8) | 5.1 (4.3) | <0.001 | −4.8 (−5.3) | 4.4 (4.3) | <0.001 |
| Sigmoid | −5.8 (−4.4) | 6.2 (4.0) | <0.001 | −5.6 (−4.9) | 5.2 (4.0) | <0.001 |
Negative sign refers the values of current DVH parameter addition underestimate when compared to the revised DVH parameter addition.
No statistically significant differences in D2ccEQD2 values were recorded between two different EBRT techniques: 3D conformal radiotherapy (3DCRT) and intensity‐modulated radiotherapy (IMRT) cases for both conformal and Point A HDR plans, regardless of the rectum, bladder, or sigmoid (P > 0.3148) (see Table 3). For all PM boost cases, the current DVH parameter addition method assumes no contributions from 4 cm central block PM boosts. When directly compared to the revised DVH parameter addition method using values obtained from each patient's EBRT plan, D2 ccEQD2 and D0.1ccEQD2 values for all rectum, bladder, and sigmoid, regardless of conformal or Point A HDR planning technique, were statistically significantly underestimated (P < 0.008) (Table 4). On average, −8 to −10% (−5.4 to −8.0 GyEQD2) underestimation occurred overall for OAR doses for the conformal HDR plans.
Table 3.
The differences of D2 cc values between the reference dataset and either original DVH parameter addition or a revised addition method when two different EBRT techniques, 3DCRT and IMRT were compared. EBRT cases with PA or PM boosts were excluded for this analysis. No statistically significant differences were recorded (p > 0.3148)
| Organ | 3DCRT [% (GyEQD2)]a | IMRT [% (GyEQD2)]a | Difference Between 3DCRT and IMRT [% (GyEQD2)]b | |||
|---|---|---|---|---|---|---|
| ∆D2ccEQD2 Mean | ∆D2ccEQD2 Std Dev | ∆D2ccEQD2 Mean | ∆D2ccEQD2 Std Dev | ∆D2ccEQD2 Mean | P‐value | |
| Current DVH parameter addition with Uniform EBRT Rx Dose Assumption | ||||||
| Conformal HDR Plan | ||||||
| Rectum | −3.8 (−2.3) | 1.8 (1.2) | −3.9 (−1.9) | 8.4 (5.3) | 0.1 (−0.4) | 0.9053 |
| Bladder | −5.0 (−3.7) | 9.2 (6.3) | −6.2 (−4.8) | 2.9 (2.3) | 1.2 (1.1) | 0.6913 |
| Sigmoid | −8.2 (−4.7) | 5.4 (2.8) | −7.1 (−4.5) | 2.5 (1.5) | −1.1 (−0.2) | 0.4407 |
| Point A HDR Plan | ||||||
| Rectum | −2.1 (−1.4) | 3.1 (2.2) | −3.6 (−2.3) | 6.7 (4.6) | 1.5 (0.9) | 0.8845 |
| Bladder | −3.9 (−3.1) | 5.4 (4.9) | −1.8 (−1.0) | 7.6 (7.7) | −2.1 (−2.1) | 0.7509 |
| Sigmoid | −7.3 (−4.7) | 5.9 (2.7) | −4.9 (−3.5) | 6.8 (5.0) | −2.4 (−1.2) | 0.8157 |
| Revised DVH parameter addition | ||||||
| Conformal HDR Plan | ||||||
| Rectum | −3.4 (−2.0) | 8.6 (4.9) | −1.9 (−1.1) | 1.2 (0.7) | −1.5 (−0.9) | 0.9949 |
| Bladder | −2.2 (−1.9) | 10.0 (7.6) | −2.7 (−2.1) | 1.6 (1.4) | 0.5 (0.3) | 0.5071 |
| Sigmoid | −5.7 (−3.2) | 9.4 (4.5) | −2.6 (−1.8) | 1.8 (1.3) | −3.1 (−1.4) | 0.3149 |
| Point A HDR Plan | ||||||
| Rectum | +1.6 (+1.4) | 4.6 (3.5) | −1.3 (−0.8) | 1.4 (1.0) | 2.9 (2.2) | 0.7352 |
| Bladder | +0.5 (+0.6) | 5.0 (5.0) | +1.2 (+1.8) | 7.2 (7.8) | −0.8 (−1.2) | 0.8419 |
| Sigmoid | −0.4 (−0.4) | 3.2 (2.4) | −0.7 (−0.5) | 6.7 (5.5) | 0.3 (0.1) | 0.9156 |
Negative sign refers the values of current/revised DVH parameter addition underestimate when compared to the reference datasets.
The values = 3DCRT – IMRT. Thus, negative signs refers IMRT values are bigger than 3DCRT values.
Table 4.
The differences of DVH parameters (i.e., D2cc and D0.1cc) only for PM boost cases between a DVH parameter addition with the assumption of no contributions from PM boost and a revised DVH parameter addition in which D2cc and D0.1cc parameters were obtained from each patient's EBRT plan (i.e., the revised DVH parameter addition — the current DVH parameter addition). The rectum, bladder, and sigmoid D2ccEQD2 and D0.1ccEQD2 values of the current DVH parameter addition were statistically significantly underestimated when compared to the revised DVH parameter addition (p < 0.0008) regardless of HDR planning techniques; for both conformal volume‐optimization and Point A plans
| ∆ D2ccEQD2 [% (GyEQD2)] | P‐value | ∆ D0.1ccEQD2 [% (GyEQD2)] | P‐value | |||
|---|---|---|---|---|---|---|
| Mean* | Std Dev | Mean* | Std Dev | |||
| Conformal HDR Plan | ||||||
| Rectum | −7.9 (−5.4) | 4.2 (2.7) | 0.0003 | −8.5 (−6.8) | 3.1 (2.2) | <0.0001 |
| Bladder | −9.3 (−8.0) | 3.0 (2.7) | <0.0001 | −7.8 (−8.3) | 3.0 (2.8) | <0.0001 |
| Sigmoid | −9.9 (−6.9) | 3.4 (2.5) | <0.0001 | −9.4 (−7.4) | 2.6 (2.2) | <0.0001 |
| Point A HDR Plan | ||||||
| Rectum | −5.2 (−3.9) | 6.8 (4.4) | 0.007 | −5.7 (−5.0) | 6.7 (5.1) | 0.004 |
| Bladder | −7.1 (−6.5) | 7.1 (6.0) | 0.002 | −6.3 (−6.8) | 6.3 (6.1) | 0.002 |
| Sigmoid | −6.7 (−5.4) | 8.7 (5.7) | 0.005 | −6.5 (6.0) | 7.3 (5.6) | 0.002 |
* Negative sign refers the values of current DVH parameter addition underestimate when compared to the revised DVH parameter addition.
The gEUDEQD2 values were statistically different from the current evaluation dose–volume parameters of D2 ccEQD2 values for all rectum, bladder, and sigmoid, regardless of which HDR planning technique was used (conformal or Point A HDR plans (P < 0.0001)) (see Table 5). All D2ccEQD2 values for the conformal HDR plan cases were measured as statistically significantly lower than those of Point A HDR plan cases (P < 0.02). For the rectum and sigmoid, the conformal HDR cases presented statistically significantly lower values than the Point A cases (P < 0.0496) but not for the bladder (P = 0.06423). The absolute D2ccEQD2 values were significantly higher than gEUDEQD2 values (P < 0.0001), since D2ccEQD2 values are maximal doses while gEUDEQD2 values are radiobiological mean doses accounting for the volume effect characteristics of each organ.
Table 5.
The EQD2‐based gEUD values (gEUDEQD2) of integrated EBRT and HDR are compared with overall D2cc for each volume optimization, adaptive/conformal HDR plans, and Point A HDR plans
| D2ccEQD2 [GyEQD2] | gEUDEQD2 [GyEQD2] | P‐value | |||||
|---|---|---|---|---|---|---|---|
| Mean | Std Dev | Max | Mean | Std Dev | Max | ||
| Conformal HDR Plan | |||||||
| Rectum | 64.3 | 7.5 | 80.9 | 52.2 | 5.7 | 64.1 | <0.0001 |
| Bladder | 82.9 | 9.2 | 95.3 | 53.5 | 5.3 | 67.5 | <0.0001 |
| Sigmoid | 65.7 | 5.7 | 75.9 | 56.1 | 4.6 | 68.5 | <0.0001 |
| Point A HDR Plan | |||||||
| Rectum | 70.3 | 10.8 | 93.0 | 55.3 | 5.9 | 72.0 | <0.0001 |
| Bladder | 90.2 | 12.2 | 115.1 | 56.0 | 4.3 | 64.8 | <0.0001 |
| Sigmoid | 74.3 | 14.2 | 122.3 | 60.8 | 6.7 | 79.1 | <0.0001 |
In this study, we found that the overall D2ccEQD2 values for all rectum, bladder, and sigmoid for either conformal, volume optimization or Point A HDR plan that were generated through the current DVH parameter addition method, were significantly underestimated, when compared to the values obtained from the reference technique (P < 0.01) and the proposed, revised DVH parameter method (P < 0.004). The proposed, revised DVH parameter addition method, where the EBRT DVH parameters (e.g., D2cc) were simply obtained from each patient's EBRT plan instead of assuming a completely uniform prescription dose, did not present statistical differences with the reference dataset values (P > 0.099). In order to avoid significant underestimation of overall D2ccEQD2 values in clinical practice, it is recommended to simply obtain each patient's EBRT D2cc values of the rectum, bladder, and sigmoid from his or her EBRT plan instead of assuming a completely uniform prescription dose (the current DVH parameter addition) or exporting and processing dose DICOM files for each HDR plan (dose distribution (DVHEQD2) addition), which requires additional software. To more accurately estimate overall OAR doses by accounting for their locations, deformable image registration (DIR) techniques have been investigated.31, 32, 33, 34, 35
Van de Kamer et al. first tested the current DVH parameter addition method and found that it still yielded a sufficient approximation without an additional EBRT parametrial (PM) or paraaortic (PA) boost when compared to a dose distribution addition method (i.e., adding the EBRT and BT DVHs). However, they found that the overall D90 HR‐CTV with an EBRT PM (PA) boost for HDR was underestimated by 9.9 ± 6.2% (2.8 ± 1.4%) when a current DVH parameter addition method was used.19 They concluded that a “dose distribution (i.e., DVH) addition” method should be considered when an EBRT boost is given. Tamaki et al. highlighted the dosimetric impact of PM boosts with 3–4 cm central block17 through a simulated phantom setup. They found the contributions from central block of PM boost plan were on average 9% (5%–6%) and 28%–32% (11%–16%), for the D2ccEQD2 values of the rectum and bladder, respectively, when a 3 cm (4 cm) central block was used.17 Fenkell et al.18 also reported that an EBRT PM boost caused D2 ccEQD2 dose to increase by more than 50% over the boost prescription dose in four out of six patients and that the central PM boost shield does not predictably protect the high dose regions (D2 cc) of OARs. They proposed the use of intracavitary plus interstitial applicator treatments to properly cover the parametrial region instead of using the EBRT PM boost technique. Andersen et al. reported DIR‐based DVH parameter addition for the bladder was possible in 42 out of 77 cases (54%) and found mean deviations of 1.5 ± 1.8% in D2 cc EQD2 bladder values. They observed D2ccEQD2 dose deviations greater than 5% occurred in only 2% of the subjects, and they concluded that the recommended “current DVH parameter addition” provides a good estimate for the overall bladder D2ccEQD2 values. The Andersen et al. study31 included pathological node EBRT boosts but did not include PM boosts due to the use of tandem‐and‐ring (T&R) + interstitial applicators. It is worth noting that Andersen et al. could only test the overall D2cc values of the bladder by using a surface mesh DIR technique,32 but due to the considerable uncertainties of the DIR technique could not test the rectum, sigmoid, and small bowel. ICRU report #8911 states that “currently no deformable registration program is capable of tracking the location and dose‐exposure history of relevant biological structures within the target volumes and OAR”. Using DIR techniques is challenging mainly due to three factors: (a) 3D images of BT include vaginal packing with an applicator in situ and are not presented in EBRT image datasets; DIR algorithms assume there are paired pixels on two image datasets, (b) the significant deformation of organs such as the uterus and the sigmoid between EBRT and BT, and (c) the technical difficulties of reconciling DIR between MRI datasets which are the optimal 3D image guidance technique for BT, and CT datasets which are the most popular primary images for EBRT.
ICRU report #8911 addresses the assumption of a completely uniform EBRT dose that may not be valid for IMRT or volumetric‐modulated arc therapy (VMAT), and it identifies that fact that a special analysis is required when an EBRT PA or PM boost is used. When an EBRT PM boost is used, the evaluation of overall doses is especially challenging,11 since the central shield PM boosts do not predictably protect the OARs (D2ccEQD2). However, ICRU report #89 could not present a practical DVH parameter addition approach without processing the dose DICOM files for each EBRT and HDR plan.
One way to eliminate the complexity of dose integration of EBRT and BT is using EBRT boosts as an alternative to BT boost. Pioneering studies have investigated the feasibility of using IMRT or intensity‐modulated proton therapy (IMPT) boosts.36 Georg et al. concluded that 3D image‐guided, volume optimization, conformal BT techniques need to be used, since both advanced IMRT and IMPT boosts seem to be inferior to advanced BT boosts. For patients unable to undergo BT, stereotactic‐body radiotherapy (SBRT) techniques have been explored with different fractionation schemes such as 20 Gy in five fractions (23 Gy10 in EQD2),37, 38 14 Gy in two fractions (20 Gy10 in EQD2),38, 39 25 Gy in five fractions (31 Gy10 in EQD2),40 5 Gy in three fractions (31 Gy10 in EQD2),41 or 6 Gy in five fractions (40 Gy10 in EQD2).42 3D conformal radiotherapy (3D‐CRT) or IMRT techniques38, 39, 41, 43 along with CyberKnife (Accuracy Inc., Sunnyvale, CA, USA)37, 40, 42 have also been tested. However, the majority of clinical SBRT results37, 38, 40, 42 are still in the early stage with only small number of patients (<20 patients) and short‐term follow‐up (<26 months).
As preliminary data as a precursor to a large‐scale clinical outcome analysis, a radiobiological volumetric dose parameter (gEUDEQD2) which accounts for full DVH, and each organ's volume effect characteristic, such as serial or parallel organ, are presented here and demonstrated as significantly different from the current dose–volume plan evaluation metric (D2 ccEQD2). Clinical outcome (acute and late side effect) correlation analysis of using D2ccEQD2 alone, gEUDEQD2 alone, or a combination of D2ccEQD2 and gEUDEQD2 for OARs will be followed. Kim et al.28 presented the finding that overall EBRT gEUDEQD2 and each HDR plan has considerably higher statistical correlation with predicted normal tissue complication probability (NTCP) than the current overall D2ccEQD2. Shaw et al.44 proposed driving gEUDEQD2 criteria, resulting in conformal, volume optimization HDR plans per GEC‐ESTRO recommendations.5, 7 Yao et al. explored the use of gEUD as an inverse optimization objective function.45 The D2ccEQD2 OAR values are used as OAR plan evaluation metrics based upon the recommendations of the GEC‐ESTRO working group,5, 7 the EMBRACE protocol,10 and ICRU report #8911 that assumes that D2ccEQD2 HDR values or pulse dose rate (PDR) BT plans represent the full dose distribution (DVH) of each OAR reasonably well. As cautioned by ICRU report #89,11 the overall DVH pattern of OARs can change considerably with EBRT PA or PM boost, especially when IMRT or VMAT are used. Similar OAR DVHs pattern changes have occurred with BT treatments following the introduction of hybrid intracavitary and interstitial applicators such as the Vienna46, 47 or Urecht applicator.48 It is for these aforementioned reasons that similar overall D2ccEQD2 values, but significantly different volumetric doses (DVHs) are possible. As Koom et al. described, reporting organ doses in terms of a single point, or a maximal small volume dose assessment is not appropriate considering an inhomogeneous dose distribution, even in the event of a steep gradient over the high‐risk clinical target volume (HR‐CTV) and OARs near the source.15 In this study, all physical HDR plan doses (D2cc and D0.1 cc) were calculated based upon the AAPM TG 43 formula49, 50, 51 without heterogeneity‐corrections. The uncertainties of TG 43 dose calculations have been validated through a model‐based dose calculation algorithm that accounts for tissue and applicator heterogeneity.52 The uncertainties of OAR D2 cc in HDR BT plans for cervical cancer have been reported as on average 1%~3% for plastic tandem‐and‐ring (T&R) applicators calculated using a Grid‐Based Boltzmann Solver (GBBS) algorithm (Acuros, Varian Medical System, Inc.)53 or by a collapsed‐cone convolution algorithm (ACE, Elekta Ltd., Stockholm, Sweden).54 This is also true of titanium T&O applicators.26
4. CONCLUSIONS
The overall D2 ccEQD2 values for all rectum, bladder, and sigmoid for conformal, volume optimization HDR plans generated using the current DVH parameter addition method, (which assumes a completely uniform prescription dose of EBRT), were significantly underestimated by on average −4.7%~−8.3%, when compared to the reference technique of dose distribution (DVH) addition values (P < 0.01). The revised DVH parameter addition method where the EBRT DVH parameters (e.g., D2 cc) were simply obtained from each patient's EBRT plan did not present statistical differences with the reference technique's values (P > 0.099). When PM boosts were used, there was an average −8%~−10% underestimation of overall OAR doses occurred for conformal HDR plans when using the current DVH parameter addition technique when compared to the revised DVH parameter addition. No significant differences between 3DCRT and IMRT techniques were found when using different approaches to estimate the overall OAR DVH parameters.
CONFLICT OF INTEREST
The authors declare that there is no conflict of interest.
REFERENCES
- 1. Eifel PJ. Chemoradiotherapy in the treatment of cervical cancer. Semin Radiat Oncol. 2006;16:177–185. [DOI] [PubMed] [Google Scholar]
- 2. Lanciano RM, Martz K, Coia LR, Hanks GE. Tumor and treatment factors improving outcome in stage III‐B cervix cancer. Int J Radiat Oncol Biol Phys. 1991;20:95–100. [DOI] [PubMed] [Google Scholar]
- 3. Chafe S, Moughan J, McCormick B, et al. Late toxicity and patient self‐assessment of breast appearance/satisfaction on RTOG 0319: a phase 2 trial of 3‐dimensional conformal radiation therapy‐accelerated partial breast irradiation following lumpectomy for stages I and II breast cancer. Int J Radiat Oncol Biol Phys. 2013;86:854–859. [DOI] [PubMed] [Google Scholar]
- 4. O'Brien FW. The radium treatment of cancer of the cervix; a historical review. Am J Roentgenol Radium Ther. 1947;57:281–297. [PubMed] [Google Scholar]
- 5. Haie‐Meder C, Potter R, van Limbergen E, et al. G.E.C.E.W.G. Gynaecological, “Recommendations from Gynaecological (GYN) GEC‐ESTRO Working Group (I): concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV”. Radiother Oncol. 2005;74:235–245. [DOI] [PubMed] [Google Scholar]
- 6. Hellebust TP, Kirisits C, Berger D, et al. G.E.C.E.W.G. Gynaecological, “Recommendations from Gynaecological (GYN) GEC‐ESTRO Working Group: considerations and pitfalls in commissioning and applicator reconstruction in 3D image‐based treatment planning of cervix cancer brachytherapy”. Radiother Oncol. 2010;96:153–160. [DOI] [PubMed] [Google Scholar]
- 7. Potter R, Haie‐Meder C, Van Limbergen E, et al. G.E.W. Group, “Recommendations from gynaecological (GYN) GEC ESTRO working group (II): concepts and terms in 3D image‐based treatment planning in cervix cancer brachytherapy‐3D dose volume parameters and aspects of 3D image‐based anatomy, radiation physics, radiobiology. Radiother Oncol. 2006;78, 67–77. [DOI] [PubMed] [Google Scholar]
- 8. Beriwal S, Demanes DJ, Erickson B, et al. S. American Brachytherapy, “American Brachytherapy Society consensus guidelines for interstitial brachytherapy for vaginal cancer”. Brachytherapy. 2012;11:68–75. [DOI] [PubMed] [Google Scholar]
- 9. Viswanathan AN, Thomadsen B & American Brachytherapy Society Cervical Cancer Recommendations Committee, American Brachytherapy Society . American Brachytherapy Society consensus guidelines for locally advanced carcinoma of the cervix. Part I: general principles. Brachytherapy. 2012;11:33–46. [DOI] [PubMed] [Google Scholar]
- 10. EMBRACE , “www.embracestudy.dk.”
- 11. ICRU‐report89 , Prescribing, Recording, and Reporting Brachytherapy for Cancer of the Cervix. (Oxford University Press, 2013). [DOI] [PubMed]
- 12. Potter R, Georg P, Dimopoulos JC, et al. Clinical outcome of protocol based image (MRI) guided adaptive brachytherapy combined with 3D conformal radiotherapy with or without chemotherapy in patients with locally advanced cervical cancer. Radiother Oncol. 2011;100:116–123. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Fidarova EF, Berger D, Schussler S, et al. Dose volume parameter D2cc does not correlate with vaginal side effects in individual patients with cervical cancer treated within a defined treatment protocol with very high brachytherapy doses. Radiother Oncol. 2010;97:76–79. [DOI] [PubMed] [Google Scholar]
- 14. Jamema SV, Mahantshetty U, Tanderup K, et al. Inter‐application variation of dose and spatial location of D(2 cm(3)) volumes of OARs during MR image based cervix brachytherapy. Radiother Oncol. 2013;107:58–62. [DOI] [PubMed] [Google Scholar]
- 15. Koom WS, Sohn DK, Kim JY, et al. Computed tomography‐based high‐dose‐rate intracavitary brachytherapy for uterine cervical cancer: preliminary demonstration of correlation between dose‐volume parameters and rectal mucosal changes observed by flexible sigmoidoscopy. Int J Radiat Oncol Biol Phys. 2007;68:1446–1454. [DOI] [PubMed] [Google Scholar]
- 16. International Commission on Radiation Units & Measurements Inc . Dose and Volume Specification for Reporting Intracavitary Therapy in Gynecology (Report 38). (ICRU, Bethesda, MD, 1985). [Google Scholar]
- 17. Tamaki T, Noda SE, Ohno T, Kumazaki Y, Kato S, Nakano T. Dose‐volume histogram analysis of composite EQD2 dose distributions using the central shielding technique in cervical cancer radiotherapy. Brachytherapy. 2016;15:598–606. [DOI] [PubMed] [Google Scholar]
- 18. Fenkell L, Assenholt M, Nielsen SK, et al. Parametrial boost using midline shielding results in an unpredictable dose to tumor and organs at risk in combined external beam radiotherapy and brachytherapy for locally advanced cervical cancer. Int J Radiat Oncol Biol Phys. 2011;79:1572–1579. [DOI] [PubMed] [Google Scholar]
- 19. van de Kamer JB, De Leeuw AA, Moerland MA, Jurgenliemk‐Schulz IM. Determining DVH parameters for combined external beam and brachytherapy treatment: 3D biological dose adding for patients with cervical cancer. Radiother Oncol. 2010;94:248–253. [DOI] [PubMed] [Google Scholar]
- 20. Kim Y, Muruganandham M, Modrick JM, Bayouth JE. Evaluation of artifacts and distortions of titanium applicators on 3.0‐Tesla MRI: feasibility of titanium applicators in MRI‐guided brachytherapy for gynecological cancer. Int J Radiat Oncol Biol Phys. 2011;80:947–955. [DOI] [PubMed] [Google Scholar]
- 21. Anderson J, Huang Y, Kim Y. Dosimetric impact of point A definition on high‐dose‐rate brachytherapy for cervical cancer: evaluations on conventional point A and MRI‐guided, conformal plans. J Contemp Brachytherapy. 2012;4:241–246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Anderson JW, Xia J, Flynn RT, et al. High resolution (3 Tesla) MRI‐guided conformal brachytherapy for cervical cancer: consequences of different high‐risk CTV sizes. J Contemp Brachytherapy. 2013;5:101–109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Schindel J, Zhang W, Bhatia SK, Sun W, Kim Y. Dosimetric impacts of applicator displacements and applicator reconstruction‐uncertainties on 3D image‐guided brachytherapy for cervical cancer. J Contemp Brachytherapy. 2013;5:250–257. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Sun W, Bhatia SK, Jacobson GM, Flynn RT, Kim Y. Target volume changes through high‐dose‐rate brachytherapy for cervical cancer when evaluated on high resolution (3.0 Tesla) magnetic resonance imaging. Pract Radiat Oncol. 2012;2:e101–e106. [DOI] [PubMed] [Google Scholar]
- 25. Kim Y, Modrick JM, Pennington EC, Kim Y. Commissioning of a 3D image‐based treatment planning system for high‐dose‐rate brachytherapy of cervical cancer. J Appl Clin Med Phys. 2016;17:5818. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Hyer DE, Sheybani A, Jacobson GM, Kim Y. The dosimetric impact of heterogeneity corrections in high‐dose‐rate (1)(9)(2)Ir brachytherapy for cervical cancer: investigation of both conventional Point‐A and volume‐optimized plans. Brachytherapy. 2012;11:515–520. [DOI] [PubMed] [Google Scholar]
- 27. Kim Y, Basak CS, Mart CJ, S.M. M , Sun W, Bhatia SK Presented at the IUPESM World Congress on Medical Physics & Biomedical Engineering, Toronto, Canada, 2015. (unpublished). [Google Scholar]
- 28. Kim Y, Basak C, Cheung KH, Mart C, Sun W, Bhatia SK. Generalized equivalent uniform dose to evaluate overall OAR Doses of EBRT and 3D Image‐Guided Adaptive HDR brachytherapy for cervical cancer. Brachytherapy. 2014;13:S24–S25. [Google Scholar]
- 29. Rancati T, Fiorino C, Gagliardi G, et al. Fitting late rectal bleeding data using different NTCP models: results from an Italian multi‐centric study (AIROPROS0101). Radiother Oncol. 2004;73:21–32. [DOI] [PubMed] [Google Scholar]
- 30. Burman C, Kutcher GJ, Emami B, Goitein M. Fitting of normal tissue tolerance data to an analytic function. Int J Radiat Oncol Biol Phys. 1991;21:123–135. [DOI] [PubMed] [Google Scholar]
- 31. Andersen ES, Noe KO, Sorensen TS, et al. Simple DVH parameter addition as compared to deformable registration for bladder dose accumulation in cervix cancer brachytherapy. Radiother Oncol. 2013;107:52–57. [DOI] [PubMed] [Google Scholar]
- 32. Noe KO, Tanderup K, Sorensen TS Surface membrane based bladder registration for evaluation of accumulated dose during brachytherapy in cervical cancer. IEEE international symposium on biomedical imaging (ISBI), 1253–1256 (2011).
- 33. Sabater S, Andres I, Sevillano M, Berenguer R, Machin‐Hamalainen S, Arenas M. Dose accumulation during vaginal cuff brachytherapy based on rigid/deformable registration vs. single plan addition. Brachytherapy. 2014;13:343–351. [DOI] [PubMed] [Google Scholar]
- 34. Xiong L, Viswanathan A, Stewart AJ, et al. Deformable structure registration of bladder through surface mapping. Med Phys. 2006;33:1848–1856. [DOI] [PubMed] [Google Scholar]
- 35. Christensen GE, Carlson B, Chao KS, et al. Image‐based dose planning of intracavitary brachytherapy: registration of serial‐imaging studies using deformable anatomic templates. Int J Radiat Oncol Biol Phys. 2001;51:227–243. [DOI] [PubMed] [Google Scholar]
- 36. Georg D, Kirisits C, Hillbrand M, Dimopoulos J, Potter R. Image‐guided radiotherapy for cervix cancer: high‐tech external beam therapy versus high‐tech brachytherapy. Int J Radiat Oncol Biol Phys. 2008;71:1272–1278. [DOI] [PubMed] [Google Scholar]
- 37. Haas JA, Witten MR, Clancey O, Episcopia K, Accordino D, Chalas E. CyberKnife boost for patients with cervical cancer unable to undergo brachytherapy. Front Oncol. 2012;2:25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Molla M, Escude L, Nouet P, et al. Fractionated stereotactic radiotherapy boost for gynecologic tumors: an alternative to brachytherapy? Int J Radiat Oncol Biol Phys. 2005;62:118–124. [DOI] [PubMed] [Google Scholar]
- 39. Jorcano S, Molla M, Escude L, et al. Hypofractionated extracranial stereotactic radiotherapy boost for gynecologic tumors: a promising alternative to high‐dose rate brachytherapy. Technol Cancer Res Treat. 2010;9:509–514. [DOI] [PubMed] [Google Scholar]
- 40. Kubicek GJ, Xue J, Xu Q, et al. Stereotactic body radiotherapy as an alternative to brachytherapy in gynecologic cancer. Biomed Res Int. 2013;2013:898953. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Guckenberger M, Bachmann J, Wulf J, et al. Stereotactic body radiotherapy for local boost irradiation in unfavourable locally recurrent gynaecological cancer. Radiother Oncol. 2010;94:53–59. [DOI] [PubMed] [Google Scholar]
- 42. Marnitz S, Kohler C, Budach V, et al. Brachytherapy‐emulating robotic radiosurgery in patients with cervical carcinoma. Radiat Oncol. 2013;8:109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Hsieh CH, Tien HJ, Hsiao SM, et al. Stereotactic body radiation therapy via helical tomotherapy to replace brachytherapy for brachytherapy‐unsuitable cervical cancer patients – A preliminary result. Onco Targets Ther. 2013;6:59–66. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Shaw W, Rae WI, Alber ML. Equivalence of Gyn GEC‐ESTRO guidelines for image guided cervical brachytherapy with EUD‐based dose prescription. Radiat Oncol. 2013;8:266. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Yao R, Templeton AK, Liao Y, Turian JV, Kiel KD, Chu JC. Optimization for high‐dose‐rate brachytherapy of cervical cancer with adaptive simulated annealing and gradient descent. Brachytherapy. 2014;13:352–360. [DOI] [PubMed] [Google Scholar]
- 46. Dimopoulos JC, Kirisits C, Petric P, et al. The Vienna applicator for combined intracavitary and interstitial brachytherapy of cervical cancer: clinical feasibility and preliminary results. Int J Radiat Oncol Biol Phys. 2006;66:83–90. [DOI] [PubMed] [Google Scholar]
- 47. Kirisits C, Lang S, Dimopoulos J, Berger D, Georg D, Potter R. The Vienna applicator for combined intracavitary and interstitial brachytherapy of cervical cancer: design, application, treatment planning, and dosimetric results. Int J Radiat Oncol Biol Phys. 2006;65:624–630. [DOI] [PubMed] [Google Scholar]
- 48. Jurgenliemk‐Schulz IM, Tersteeg RJ, Roesink JM, et al. MRI‐guided treatment‐planning optimisation in intracavitary or combined intracavitary/interstitial PDR brachytherapy using tandem ovoid applicators in locally advanced cervical cancer. Radiother Oncol. 2009;93:322–330. [DOI] [PubMed] [Google Scholar]
- 49. Nath R, Anderson LL, Luxton G, Weaver KA, Williamson JF, Meigooni AS. Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine. Med Phys. 1995;22:209–234. [DOI] [PubMed] [Google Scholar]
- 50. Rivard MJ, Coursey BM, DeWerd LA, et al. Update of AAPM Task Group No. 43 Report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31:633–674. [DOI] [PubMed] [Google Scholar]
- 51. Rivard MJ, Ballester F, Butler WM, et al. Supplement 2 for the 2004 update of the AAPM Task Group No. 43 Report: joint recommendations by the AAPM and GEC‐ESTRO. Med Phys. 2017;44:e297–e338. [DOI] [PubMed] [Google Scholar]
- 52. Beaulieu L, Carlsson Tedgren A, Carrier JF, et al. Report of the Task Group 186 on model‐based dose calculation methods in brachytherapy beyond the TG‐43 formalism: current status and recommendations for clinical implementation. Med Phys. 2012;39:6208–6236. [DOI] [PubMed] [Google Scholar]
- 53. Hofbauer J, Kirisits C, Resch A, et al. Impact of heterogeneity‐corrected dose calculation using a grid‐based Boltzmann solver on breast and cervix cancer brachytherapy. J Contemp Brachytherapy. 2016;8:143–149. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Jacob D, Lamberto M, DeSouza Lawrence L, Mourtada F. Clinical transition to model‐based dose calculation algorithm: a retrospective analysis of high‐dose‐rate tandem and ring brachytherapy of the cervix. Brachytherapy. 2017;16:624–629. [DOI] [PubMed] [Google Scholar]