Skip to main content
The Journal of the Acoustical Society of America logoLink to The Journal of the Acoustical Society of America
. 2018 Mar 16;143(3):1491–1503. doi: 10.1121/1.5026796

Quantifying undesired parallel components in Thévenin-equivalent acoustic source parameters

Kren Rahbek Nørgaard 1,a),, Stephen T Neely 2, Daniel M Rasetshwane 2
PMCID: PMC5856597  PMID: 29604709

Abstract

The calibration of an ear probe to determine its Thévenin-equivalent acoustic source parameters facilitates the measurement of ear-canal impedance and reflectance. Existing calibration error metrics, used to evaluate the quality of a calibration, are unable to reveal undesired parallel components in the source parameters. Such parallel components can result from, e.g., a leak in the ear tip or improperly accounting for evanescent modes, and introduce errors into subsequent measurements of impedance and reflectance. This paper proposes a set of additional error metrics that are capable of detecting such parallel components by examining the causality of the source admittance in the frequency domain and estimating the source pressure in the time domain. The proposed and existing error metrics are applied to four different calibrations using two existing calibration methods, representing typical use cases and introducing deliberate parallel components. The results demonstrate the capability of the proposed error metrics in identifying various undesired components in the source parameters that might otherwise go undetected.

I. INTRODUCTION

Measurements of ear-canal acoustic impedance and reflectance have been proposed for diagnoses of conductive hearing disorders (Piskorski et al., 1999; Keefe et al., 2000; Feeney and Keefe, 2001; Keefe et al., 2012; Ellison et al., 2012; Merchant et al., 2014) and for calibrating stimulus levels in situ (Scheperle et al., 2008; McCreery et al., 2009; Lewis et al., 2009; Scheperle et al., 2011; Withnell et al., 2009; Souza et al., 2014). To facilitate such probe-based measurements, the acoustic source parameters of an ear probe, typically the Thévenin-equivalent source parameters, must be determined in a preliminary calibration procedure. The quality, or accuracy, of the source parameters is traditionally assessed by examining the calibration error, defined as the ability of the source parameters to estimate the known reference impedances of the waveguides or the measured probe pressures. Evanescent modes occur in the discontinuity between the probe tube and load waveguide and affect probe-based measurements in the ear canal (Brass and Locke, 1997) as higher-order, non-propagating modes are excited. However, they also affect the calibration procedure, which typically constitutes measurements using the probe in a set of reference loads with known acoustic input impedances. Evanescent modes can be approximated as an acoustic inertance in series with the plane-wave acoustic input impedance and essentially translates the position of impedance minima in frequency (Keefe and Benade, 1981; Fletcher et al., 2005).

Siegel and Neely (2017) demonstrated that the presence of evanescent modes during a widely used calibration methodology, proposed by Allen (1985), introduces an apparent non-causality into the source impedance. Siegel and Neely (2017) also demonstrated how a physical modification to the probe tube reduced the evanescent modes in front of the probe, and removed a large part of this undesired non-causal behavior in the source impedance.

Nørgaard et al. (2017b) further quantified the effect of evanescent modes on the source parameters and showed that it can be approximated by a phantom acoustic compliance in parallel with the source parameters. The compliance is phantom in the sense that it is not physically present during the calibration procedure, but is rather introduced as a consequence of indirectly estimating the waveguide model lengths from probe pressures that are affected by evanescent modes. They also proposed a method to account for evanescent modes during the calibration procedure and indirectly demonstrated how this approach appeared to eliminate the parallel compliance from the source parameters. However, they noted that the calibration error, typically used to assess the outcome of an acoustic Thévenin calibration, is incapable of identifying such undesired parallel component as such component is inherited into the source parameters.

Thus, there is a need for alternative error metrics that can reveal any undesired parallel components in the source parameters. These error metrics should provide a direct, absolute, and objective assessment of the physical validity of the source parameters produced by such acoustic Thévenin calibration procedure. This paper proposes a set of error metrics that can be utilized in combination with existing and during development of new calibration procedures to assess the physical validity of the source parameters and quantify parallel components. The proposed error metrics are based partly on the study of causality in the source admittance using the Hilbert transform, and a time-limited estimation of the source pressure from the source volume flow under in a resistive load. In addition, this paper will further explain the concept of parallel components in the source parameters, limitations in representing transfer functions in the time domain, and the seemingly non-causality associated with lumped acoustic elements that arises as a result of finite frequency spectra.

II. BACKGROUND

A. Existing error metrics

In the existing, widely used methodology (e.g., Allen, 1985; Keefe et al., 1992; Voss and Allen, 1994), the probe pressures Pi in N short, rigidly terminated waveguides are measured. By modeling the input impedances Zi of each waveguide as an acoustic transmission line, including thermo-viscous wall losses (Keefe, 1984), the Thévenin-equivalent source parameters (the source pressure Ps and source impedance Zs) are obtained by solving an overdetermined system of N equations using the least-squares method,

[Z1P1Z2P2ZNPN][PsZs]=[P1Z1P2Z2PNZN]. (1)

The lengths li of the calibration waveguides, used to calculate the analytical reference impedances Zi, are estimated, and the final evaluation of the accuracy of the obtained source parameters can be assessed by the relative error in estimating the analytically modeled reference impedances,

ϵi=|1Z^iZi|, (2)

where the estimated impedances

Z^i=ZsPiPsPi. (3)

It is possible to investigate such calibration error, absolute or relative, in terms of any variable in the system Zi, Pi, Ps, Zs, or other derived quantities. What all these possible different errors have in common is that they, in the same way, merely characterize the capability of the linear model in representing the data points used to derive the model, though with differences in scaling. These errors can reveal certain inaccuracies in the calibration, but do not provide a complete objective assessment of the accuracy of the source parameters. Scheperle et al. (2011) used, similar to Eq. (3), the source parameters to estimate the probe pressures

P^i=PsZiZs+Zi, (4)

and compared them to the measured probe pressures Pi using the relative averaged error across frequency ω and waveguides i,

ϵPi=iω|PiP^i|2iω|Pi|2. (5)

One single number is thus obtained to represent the quality of the calibration. Such averaging has the disadvantage that it cannot identify large errors confined within small frequency ranges. Notice that, due to the order of the operations (the summations precede the division) in Eq. (5), this summed probe-pressure error ϵPi is sensitive to the frequency contents of the stimulus. Other authors have also employed similar single-number error metrics (e.g., Keefe et al., 1992).

B. Parallel components

Nørgaard et al. (2017b) noted that existing error metrics are incapable of identifying undesired parallel components in the source parameters. Such parallel component can be introduced due to, for example, not accounting for evanescent modes when waveguide lengths are estimated or a leak in the ear tip. Figure 1(a) shows a Thévenin-equivalent circuit of a probe and its source parameters Ps and Zs, the measured ith probe pressure Pi in a given reference load impedance Zi, and an additional impedance Zpar in parallel with the load. The distinction between physical and phantom parallel components is important. If a parallel component is physically present during the calibration procedure, such as a leak in the ear tip, and not accounted for in the modeled reference impedances Zi, its effect will be represented in the measured probe pressures Pi and thus inherited into the source parameters. Conversely, a phantom parallel component is not physically present during the measurement, and the probe pressures Pi consequently unaffected by it, but rather, unintentionally, included in the modeled reference impedances Zi. It thus changes its sign of operation when it is inherited into the source parameters.1 An alternative set of source parameters Ps and Zs can be defined that include such parallel component, physical or phantom. The alternative source pressure Ps is given by a voltage division of the true source pressure Ps between the true source impedance Zs and the impedance of the parallel component Zpar,

Ps=PsZparZs+Zpar. (6)

Similarly, the alternative source impedance Zs is given by the parallel coupling of the true source impedance Zs and the impedance of the parallel component Zpar,

Zs=ZsZparZs+Zpar. (7)

Interestingly, these alternative Thévenin source parameters are, on a relative scale, equivalently related to their true counterparts, as they are multiplied by the same complex factor Zpar/(Zs+Zpar). Figure 1(b) shows the Norton-equivalent circuit of the acoustical system under consideration. The source volume flow Us now replaces the source pressure Ps and the source impedance Zs appears as a parallel component. Thus,

Us=PsZs=PsZs, (8)

and the source volume flow Us remains unaffected by the parallel component Zpar, while Eq. (7) still applies for the alternative source impedance Zs.

FIG. 1.

FIG. 1.

(a) Thévenin- and (b) Norton-equivalent circuit diagrams of the probe and load, including the source parameters (source pressure Ps or source volume flow Us, and source impedance Zs), an undesired parallel impedance Zpar, the ideal impedance of the ith calibration waveguide Zi, and probe pressure Pi.

C. Limitations in discrete-time representation of acoustic transfer functions

Acoustic response functions, such as the probe pressure, are typically calculated from a time-domain waveform using the Fourier transform. Conversely, acoustic transfer functions, such as impedance and reflectance, are well-defined, measured, and modeled in the frequency domain. As the term “transfer function” suggests, such functions consist of the ratio of two frequency-domain response functions. Under certain assumptions, it is possible to transform such transfer functions into the time domain using the inverse Fourier transform. The continuous-time impedance z(t) can be thought of as the resulting sound pressure in response to a Dirac-delta volume-flow impulse, and the continuous-time reflectance r(t) as the pressure response following an incident Dirac-delta pressure impulse. However, such transformation imposes certain constraints and assumptions onto the quantities in the practical case of a finite frequency spectrum and, thus, discrete-time representation. Without further modification, the band-limited, Hermitian-symmetric, discrete spectrum supplied to the inverse Fourier transform is implicitly multiplied by an ideal, zero-phase, low-pass filter. Such filter has a non-causal impulse response of a sinc function, which the time-domain pressure and volume-flow response functions are implicitly convolved by. With the continuous-time volume-flow Dirac-delta impulse occurring at t = 0, its discrete-time representation becomes

u(n)=sincπn, (9)

where n is the integer sample number. The volume-flow response function thus yields u(0)=1,u(n0)=0, and does as such not pose a limitation. However, the pressure-response function of an arbitrary load does not necessarily arrive back to the input plane at integer samples. The discrete-time representation of any pressure pulse arriving back after time τ can be represented by

p(n)=sincπ(nτfs), (10)

where fs is the sampling frequency. If τ is not an integer multiple of the sampling interval 1/fs, it results in p(n0)0 and this severely limits the interpretation of time-domain transfer functions. Similar considerations apply to the Hilbert transform, as it consists of intermediate steps of the inverse and forward Fourier transforms, and the residuals from the sinc function introduces non-zero components into negative time.

Different approaches can be employed to circumvent the effects of these constraints. Rasetshwane and Neely (2011) and Rasetshwane et al. (2012) multiplied the frequency-domain reflectance by a zero-phase frequency-domain Blackman window prior to calculating the inverse Fourier transform. The time-domain transfer functions are thus convolved by the impulse response of this window rather than the sinc function and are smeared in time, leading to a possible overlap between effects that were otherwise limited in time. Nørgaard et al. (2017a) used a frequency-domain truncation methodology that discarded part of the high-frequency spectrum to restore differentiability at the Nyquist frequency between the original and complex-conjugate part of the replicated spectrum. This approach effectively changed the sampling rate of the synthesized time-domain transfer function, ensuring that response components occur at integer sample numbers. While this approach circumvents most of the limitations, it is still dependent on the capability of restoring such differentiability. Zero-phase, band-limited filters are inherently non-causal and thus, using either approach, it is not possible to study the time-domain behavior of an acoustic transfer function quantity without applying a non-causal filter to the transfer function. Applying instead a causal filter shifts the response functions in time and thus alters the real and imaginary parts of the transfer function, which complicates the assessment of causality.

D. Non-causality and discrete-time representation of acoustic lumped components

Acoustic impedance is inherently a causal quantity. A pressure response cannot occur prior to a volume-flow stimulation. However, Siegel and Neely (2017) observed that the source impedance, obtained from an acoustic Thévenin calibration procedure, was non-causal when evanescent modes were improperly accounted for. This non-causality can likely be accredited to the presence of the phantom parallel compliance in the source impedance (Nørgaard et al., 2017b). Similarly, Rasetshwane and Neely (2011) and Nørgaard et al. (2017a) noted that probe-based measurements of acoustic impedance and reflectance were non-causal when affected by evanescent modes, and thus an approximate series inertance. The continuous-time relationships between the sound pressure p(t) and volume flow u(t) for an acoustic inertance L and compliance C are given by

p(t)=Ldu(t)dt, (11)
u(t)=Cdp(t)dt, (12)

respectively, and are both causal components. In order to interpret p(t) as the time-domain impedance of the inertance zL(t) in Eq. (11), the Fourier transform of the volume flow must satisfy F[u(t)]=1 for all frequencies. Similarly for u(t) and the time-domain admittance of the compliance yC(t) in Eq. (12), the Fourier transform of the pressure must satisfy F[p(t)]=1. For the continuous-time representation, this condition is only satisfied with the Dirac-delta function F[δ(t)]=1, and the time-domain impedance of an acoustic inertance zL(t) and admittance of an acoustic compliance yC(t) are given by the derivative of the Dirac-delta function,

zL(t)=Ldδ(t)dt, (13)
yC(t)=Cdδ(t)dt. (14)

When these components are investigated using the inverse Fourier transform of a band-limited spectrum, a signal non-causality is introduced due to the incapability of properly representing the Dirac-delta function derivative of the physical system in the discrete-time representation. With the term “signal non-causality” is understood a degree of non-causality resulting from a signal-processing artifact that occurs as a result of the underlying assumptions in the inverse Fourier transform. It is thus not the physical system under consideration that is non-causal. By truncating the continuous-time spectrum, the Dirac-delta function is represented by the sinc function, which has the derivative

ddtsincπt=cosπttsinπtπt2. (15)

The respective discrete-time representations of zL(n) and yC(n), derived from the inverse Fourier transform, thus result in

zL(n)=fsL(cosπnnsinπnπn2), (16)
yC(n)=fsC(cosπnnsinπnπn2). (17)

While the integer samples n coincided with the zeros in Eq. (9) for n0, they now occur close to the extrema in the sinc-function derivative. It is interesting to note that, even though zL(n) and yC(n) are physically causal, the signal non-causality appears due to these remaining residuals of the sinc-function derivative in the discrete-time representation. Figure 2 shows the continuous- and discrete-time behavior of the sinc function and its derivative. Similar considerations hold for the inverse functions, the time-domain admittance of an inertance yL(t) and impedance of a compliance zC(t), which are both represented by a Heaviside step function in continuous time.

FIG. 2.

FIG. 2.

(Color online) Continuous-time t and discrete-time n behavior of the sinc function and the sinc-function derivative sinc′. The discrete-time functions are plotted for integer n.

Although inertances and compliances are inherently physically causal components, it is useful to consider the hypothetical nonphysical case of negative L and C. This is a relevant consideration as the evanescent-modes inertance L is not a physically present acoustic inertance, but rather an approximation of the low-frequency behavior of higher-order modes between the speaker and microphone apertures in the probe tube. Although this inertance tends to be positive, it can also be negative in waveguides that are narrow relative to the probe-tube geometry (Fletcher et al., 2005). In addition, a negative parallel compliance is inherited into the source impedance in the presence of a positive evanescent-modes inertance during calibration (Nørgaard et al., 2017b). In general, inverting the sign of operation of the imaginary part of a transfer function effectively reverses its time-domain representation. The effect of such negative inertance or compliance thus does not change the physical causality of Eqs. (13) and (14), as the polarity of the Dirac-delta function derivative is simply reversed. However, the Heaviside step function, representing the inverse functions yL(t) and zC(t), is reversed in time in the case of negative quantities. This is an anti-causal, non-physical case where the behavior, for a negative compliance, can be interpreted as the sound pressure stored in the compliance discharging in reverse time, prior to the volume-flow excitation. For this case of a negative inertance or compliance, zL(t) and yC(t) are thus causal, but not minimum phase, since their inverse functions yL(t) and zC(t) are anti-causal (Oppenheim and Schafer, 1975).

E. Lumped waveguide extension

Nørgaard et al. (2017b) showed that, if evanescent modes are not accounted for during calibration, an impedance measurement in a waveguide of radius identical to the calibration waveguides results in an apparent plane-wave impedance. This is a result of an evanescent-modes inertance and parallel compliance effectively translating the position of impedance minima and maxima equivalently. They also noted the approximate relationship between the evanescent-modes inertance L, the parallel compliance C, and the characteristic impedance of the calibration waveguides Z0,cal,

CLZ0,cal2. (18)

Figure 3 shows the schematic of such evanescent-modes inertance and parallel compliance in front of a transmission line with the characteristic impedance Z0. The combination of an infinitesimally small inertance and compliance satisfying Eq. (18) represents an elemental length of a lossless transmission line and thus an extension of the transmission line in Fig. 3. This transmission-line extension can thus also be characterized by the time delay τ,

τ=CZ0=LZ0. (19)

However, the transmission-line extension is lumped since it only consists of one non-infinitesimal L-C coupling. This is the reason for Eq. (18) being approximate, and the transmission-line extension is only well approximated when the wavelength λcτ, where c is the speed of sound. represents the effective length of the extension. It is interesting to consider the consequence of such inertance and compliance on the signal causality of impedance. As long as the approximation is valid, the impedance of a transmission line with such extension appears completely causal as it simply represents a delay in the system. This potentially limits the methodology of Nørgaard et al. (2017a) in estimating the evanescent modes inertance in presence of a parallel compliance. Notice, that for a negative L, the corresponding C and τ would also become negative and thereby result in an effective decrease in length of the transmission-line. This could violate the physical causality of the system.

FIG. 3.

FIG. 3.

Coupling of a series inertance L and a parallel compliance C to a uniform transmission line with characteristic impedance Z0.

III. METHODS

Two error metrics are proposed that can be used to assess the quality of the source parameters and reveal undesired physical and phantom parallel components. For the first metric, a Hilbert-transform-based approach will be used to quantify the signal causality of the source admittance. The second metric employs the source pressure and volume flow, whereby it is not subject to the limitations and assumptions of the discrete inverse Fourier transform and Hilbert transform of transfer functions.

A. Source-admittance error metric

The source impedance Zs can be physically interpreted as the input impedance of the probe tube and is therefore an inherent minimum-phase quantity. Minimum phase is equivalent to a quantity being causal, stable, and having a causal inverse (Oppenheim and Schafer, 1975). With the primary motivation for proposing additional error metrics in revealing parallel components, it is convenient to investigate the behavior in the inverse function of the source impedance Zs, namely, the source admittance Ys = 1/Zs, as this quantity must also conform to causality. The parallel component in Eq. (7) now simply results in the alternative source admittance

Ys=Ys+Ypar, (20)

which is advantageous in characterizing this parallel component.

Nørgaard et al. (2017a) employed the Hilbert transform to identify and compensate for the signal non-causality introduced into waveguide acoustic impedance measurements as a result of an evanescent-modes inertance. Here, a similar approach will be employed to assess the causality of the source admittance that can reveal and quantify physical and phantom parallel components, such as a compliance, that introduce a signal or physical non-causality. Considering the assumed causal function, the time-domain source admittance ys(t < 0) = 0, the Hilbert transform relates the real and imaginary parts of the Fourier transform of that function, the frequency-domain admittance Ys, and the behavior in limωYs(ω)=Y0 (Papoulis, 1962; Hsu, 1967),

Re{Ys}=Y0+H[Im{Ys}], (21)
Im{Ys}=H1[Re{Ys}]. (22)

H[·] is the Hilbert-transform operator, Re{·} and Im{·} denote the real and imaginary parts, respectively, and Y0=A/(ρc) is the characteristic admittance of the probe tube with geometrical cross-sectional area A, air density ρ, and speed of sound c. The source admittance can now be estimated,

Y^s=Y0+H[Im{Ys}]+jH1[Re{Ys}], (23)

and the source-admittance estimation error defined,

ϵYs=YsY^s. (24)

Similar to Nørgaard et al. (2017a), the non-causality will be quantified from the real and imaginary source-admittance estimation errors

Re{ϵYs}=Re{Ys}Y0H[Im{Ys}], (25)
Im{ϵYs}=Im{Ys}H1[Re{Ys}]. (26)

For a source admittance containing a purely non-causal parallel component such as Eq. (20), the real and imaginary source admittance estimation errors yield

Re{ϵYs}=Re{Ypar}+H[Im{Ypar}], (27)
Im{ϵYs}=Im{Ypar}+H1[Re{Ypar}]. (28)

Nørgaard et al. (2017a) evaluated the Hilbert transform on the reflectance due to the non-smooth behavior of the impedance and admittance of a rigidly terminated waveguide, and transformed the reflectance estimation error ϵR into an impedance estimation error ϵZ. For the investigated cases in this study, the source admittance was generally a smooth function with frequency, rendering such transformation unnecessary. Frequency truncation is employed where possible in order to restore differentiability of the spectrum by adjusting the truncation frequency to coincide with a local extreme of the real part and an ordinate zero-crossing of the imaginary part. For a band-limited transfer function, the source-admittance estimation error characterizes the signal non-causality, and thus not the non-causality of the physical system. The Hilbert transform therefore reveals any lumped acoustic components, such as a compliance, in the source admittance, although they do not introduce any physical non-causality into the system.

B. Source-pressure error metric

Equations (6)–(8) showed that any parallel component, regardless of causality, scales the source impedance and source pressure equivalently, while leaving the source volume flow unchanged. It might therefore be unnecessary to assess the causality properties of the source pressure. However, the source pressure and volume flow differ from the source admittance in that they are response functions rather than transfer functions, and are unaffected by the inherent limitations regarding the inverse Fourier transform and Hilbert transform of transfer functions, as described in Sec. II C. They can therefore readily be transformed into the time domain without applying any windowing, truncation, or further precautions. An alternative metric to assess the causality of the source pressure is proposed that is based only on the inverse Fourier transform of response functions.

Consider the incident source pressure Ps,0 resulting from the source volume flow Us into the probe-tube characteristic impedance Z0. In the frequency domain, this quantity is inherently different from the source pressure Ps due to the non-resistive components in the source impedance Zs,

Ps,0=UsZ0Ps. (29)

Evaluating the inverse Fourier transform, the convolution us(n)z0(n)=us(n)Z0, as Z0 is a constant. With regards to the typical length of a probe tube (1–2 cm), the time-domain source impedance zs(n) can be assumed to behave purely resistive with no reflections until a certain point in time τZ0=nZ0/fs, and therefore zs(n<nZ0)=z0(n), if the characteristic impedance of the probe tube remains constant along the probe tube. Thus,

ps,0(n)=us(n)Z0=ps(n),forn<nZ0. (30)

While the time-domain source pressure ps(n) is affected by a parallel component, the source volume flow us(n) is unaffected [Eq. (8)]. The probe-tube characteristic impedance Z0=ρc/A does not contain this parallel component, as it is calculated from the known cross-sectional area A of the probe tube. Besides visually assessing the similarity of the functions ps(n) and ps,0(n) until the sample number nZ0, the following normalized source-pressure error metric is proposed:

ϵps=n=M/2+1nZ0|ps(n)ps,0(n)|2n=M/2+1nZ0|ps(n)|2. (31)

M is the fast Fourier transform (FFT)-length and nZ0 must be chosen such that only the resistive behavior in zs(n) is represented in the error. In this study, nZ0 was chosen to simply coincide with the first peak in us(n), and it might thus not contain the entire resistive part of the source impedance. In other cases it might be necessary to adopt different ways to determine nZ0.

C. Predicted incident sound power

Although it does not immediately serve as an error metric, it is useful to observe how physical and phantom parallel components contribute to the predicted sound power delivered into a non-reflecting resistive load, corresponding to the probe-tube characteristic impedance Z0. To do this, the predicted incident pressure P0 and volume flow U0 into this load are first calculated,

P0=PsZ0Zs+Z0, (32)
U0=PsZs+Z0. (33)

Using the inverse Fourier transform, the corresponding time-domain quantities p0(n) and u0(n) are obtained and the instantaneous incident sound power w0(n) can be calculated,

w0(n)=p0(n)u0(n). (34)

By summing the sound power, the predicted total-incident dissipated acoustic energy E0 can be calculated,

E0=1fsn=M/2+1M/2w0(n). (35)

D. Measurements and equipment

The measurements reported in this study were carried out using a FireFace UC sound card (RME Audio, Haimhausen, Germany) controlled through custom-written MATLAB (The MathWorks, Inc., Natick, MA) software and the third-party utility Playrec.2 Both an ER-10X ear probe (Etymōtic Research, Inc., Elk Grove Village, IL) and a Titan-based ear probe (Interacoustics A/S, Middelfart, Denmark), but modified to improve the high-frequency performance via internal cross-talk reduction, were used.

Probe pressures were obtained by applying frequency-equalized wideband chirps to the probes so as to provide a flat incident probe pressure in a non-reflecting load of radius a = 4 mm, similar to an adult ear canal. The chirps were played back in the probes in 128 phase-locked blocks with a length of M = 2048 samples at a sampling rate of fs = 44.1 kHz, which were each recorded using the probe microphones and averaged to reduce the noise in the measurements. The derived source pressures and source volume flows were divided by the complex frequency response of the electrically supplied chirp to transform them into their respective impulse responses.

The hilbert function, as part of the MATLAB Signal Processing Toolbox, was used to calculate discrete Hilbert transforms from the Hermitian-symmetric source-admittance spectra. When possible, the source-admittance spectra were truncated to restore differentiability (Nørgaard et al., 2017a). Frequency-domain impedances and admittances were multiplied by a frequency-domain Blackman window (Rasetshwane and Neely, 2011; Rasetshwane et al., 2012) before calculating their time-domain equivalents using the ifft function. Time-domain source pressures and volume flows were calculated from the entire spectrum using only the ifft function. Due to the anti-aliasing filters in the sound card, the maximum considered calibration frequency was 20 kHz. Thus, the obtained source impedances and admittances beyond this frequency and up to the Nyquist frequency fs/2 are neither considered valid nor included in calculating the inverse Fourier transform and Hilbert transform (Nørgaard et al., 2017a).

Two existing methodologies were used for calibrating each probe type and four calibrations were conducted in total:

  • (1)

    Calibration of the ER-10X ear probe using the original methodology of Allen (1985), Keefe et al. (1992), and Voss and Allen (1994) of estimating waveguide lengths and not accounting for evanescent modes. The lengths were estimated by minimizing the relative impedance calibration error ϵi [Eq. (2)] in a region averaged around the first probe-pressure minima. The ER-10X built-in calibration waveguide of radius a = 4 mm was used with software-selected lengths li = 1.4, 1.7, 2, and 2.3 cm. The probe was inserted into the waveguide using a standard, mushroom-shaped, green, 9 mm, rubber ear tip (Sanibel Supply, Middelfart, Denmark).

  • (2)

    Identical to calibration (1), but cutting off an annular sector of the mushroom part of the ear tip, thus, deliberately introducing a leak into the ear tip, representing a physically present, undesired parallel component.

  • (3)

    Identical to calibration (1), but using the beveled probe tube proposed by Siegel and Neely (2017) to reduce the effect of evanescent modes, and thus proposedly eliminating the phantom parallel compliance from the source parameters.

  • (4)

    Calibration of the Titan probe using the waveguides and methodology proposed by Nørgaard et al. (2017b) of accounting for evanescent modes and flow losses during calibration, and thus proposedly eliminating the phantom parallel compliance from the source parameters. These calibration waveguides were of lengths li = 1.2, 1.45, 1.75, and 2 cm and radius a = 2 mm. The probe was inserted into the waveguides using their same utilized hard-rubber part.

IV. EXPERIMENTAL RESULTS

A. Calibration error and source admittance

Figures 4(a)–4(d) show the relative impedance calibration errors ϵi [Eq. (2)] of calibrations (1)–(4), respectively, and Figs. 4(e) and 4(f) show the real and imaginary parts of the obtained source admittances, respectively, normalized by the probe-tube characteristic admittances, Ys/Y0 of calibrations (1)–(4). The summed probe-pressure errors ϵPi [Eq. (5)] (0.25–18 kHz, Scheperle et al., 2011) are listed in Table I. Since frequency-equalized chirps were used for each ear probe, ϵPi should be comparable.

FIG. 4.

FIG. 4.

(Color online) The resulting relative impedance calibration errors ϵi [Eq. (2)] for each waveguide i of calibrations (a) (1) ER-10X, (b) (2) ER-10X leak, (c) (3) ER-10X bevel, and (d) (4) Titan (see Sec. III D for details). The resulting (e) real and (f) imaginary parts of the source admittances, normalized by the characteristic admittance of the probe tubes, Ys /Y0 for each calibration (1)–(4).

TABLE I.

The obtained probe-pressure summation errors ϵPi [Eq. (5)], source-pressure error metrics ϵps [Eq. (31)], estimated time limits τZ0 for the resistive behavior in the source impedance, and predicted incident acoustic energy E0 [Eq. (35)] for calibrations (1)–(4) (see Sec. III D for details).

Calibration (1) ER-10X (2) ER-10X leak (3) ER-10X bevel (4) Titan
ϵPi 1.87×104 2.55×104 3.14×104 1.87×106
ϵps 1.38 6.45 0.40 0.016
τZ0 (ms) 0.136 0.136 0.136 0.113
E0 (nJ) 0.054 0.020 0.092 0.12

Comparing first calibrations (1) and (2), it is worth noticing that ϵi in Figs. 4(a) and 4(b) remain almost unchanged across the frequency spectrum, despite the introduction of a leak into the ear tip, although ϵPi appears to have increased slightly. Despite these consistent ϵi, Re{Ys} and Im{Ys} in Figs. 4(e) and 4(f) undergo quite a significant change below 2 kHz. The leak appears to be similar to the combination of a pressure-release condition, a resistance, and an inertance. Such inertance has an admittance YL=1/(jωL), which is indeed the behavior observed between Im{Ys} below 2 kHz in Fig. 4(f). In addition, it should be noted that, while Re{Ys} are positive, Im{Ys} have a negative gradient, representing the phantom parallel compliance, and are thus not minimum phase in its literal interpretation.

Comparing now calibrations (1) and (3) the observation is made, that ϵi appear to increase slightly between Figs. 4(a) and 4(c), which is further indicated by the slightly larger ϵPi. However, the parallel compliance in Im{Ys} in Fig. 4(f) has been reduced to roughly one-third of its value with the standard probe tube, though not entirely removed. Re{Ys} in Fig. 4(e) have changed slightly, however, this is to be expected as the beveled probe tube constitutes a physical modification and thus a potential change in resonances, etc. It is not immediately possible to observe this effect in Im{Ys} due to the dominating parallel compliance. Calibrations (1)–(3) clearly demonstrate the incapability of the existing error metrics in identifying undesired parallel components.

Finally, calibration (4), which was conducted using a different probe and methodology that incorporates evanescent modes into the model rather than attempting to physically reduce them, shows much lower ϵi and ϵPi than the previous calibrations (1)–(3). This can presumably be accredited to the methodology also accounting for flow losses in the transition between probe tube and calibration waveguide. It has the effect that ϵi are further reduced in impedance minima, which are the points that are most distinct in the other calibrations (1)–(3). This approach could be potentially risky as the methodology cannot distinguish between including evanescent modes and flow losses, or maybe an entirely different undesired phenomenon in the model, such as cross talk. Despite this concern, Im{Ys} appears to be centered around the abscissa indicating that there is no, or at least only a small, parallel compliance and that Ys could be minimum phase.

B. Hilbert-transform assessment of source-admittance causality

Figures 5(a)–5(d) show the normalized time-domain source impedances zs(t)/Z0 and admittances ys(t)/Y0 of calibrations (1)–(4), respectively, and Figs. 5(e) and 5(f) show the real and imaginary parts of the source-admittance estimation errors, respectively, normalized by the probe-tube characteristic admittances, ϵYs/Y0 [Eq. (24)] of calibrations (1)–(4). Due to the large magnitude of the parallel compliances, it is only possible to apply the frequency-truncation process (Nørgaard et al., 2017a) to calibration (4), and thus the Blackman window was applied to all time-domain impedance and admittance measures for better comparison.

FIG. 5.

FIG. 5.

(Color online) Time-domain impedances zs(t) and admittances ys(t) of calibrations (a) (1) ER-10X, (b) (2) ER-10X leak, (c) (3) ER-10X bevel, and (d) (4) Titan (see Sec. III D for details). The resulting (e) real and (f) imaginary parts of the source-admittance estimation errors ϵYs [Eq. (24)] for each calibration (1)–(4). Quantities are normalized by either the characteristic impedance Z0 or admittance Y0 of the probe tubes.

Comparing first calibrations (1) and (2), zs(t) and ys(t) in Figs. 5(a) and 5(b) are non-causal. While ys(t) mainly show the double-peak effect originating from the sinc-function derivative discussed in Sec. II D, zs(t) extend much further into negative time, especially evident for calibration (1). This is presumably due to the non-physical anti-causal property of the parallel compliance being negative and thus discharging into the resistive part of the source impedance in reverse time. In calibration (2), zs(t) seems to be of much smaller magnitude, which can presumably be accredited to the pressure-release properties of the ear-tip leak. Im{ϵYs} in Fig. 5(f) appear to have accurately identified the parallel compliance from the source impedance in calibration (1). On the other hand, calibration (2) is affected by the additional parallel inertance, seen from the deviation from frequency proportionality below 2 kHz. This also introduces instability into the Hilbert transform at low frequencies due to the increased noise and diverging admittance resulting from the pressure-release condition. As the non-causality is mostly represented by an imaginary component in calibration (1), Re{ϵYs} in Fig. 5(e) essentially represent the Hilbert transform of the imaginary part of the parallel component H[Im{Ypar}] [see Eq. (27)]. However, this also means that Im{ϵYs} is largely unaffected by any H1[Re{Ypar}].

Comparing calibrations (1) and (3), the non-causality in zs(t) and ys(t) has been reduced between Figs. 5(a) and 5(c), which is evident from the decreased extension into negative time of zs(t) and reduced double peak of ys(t). Both obtained zs(t) align fairly well with those presented by Siegel and Neely (2017), and Re{ϵYs} and Im{ϵYs} in Figs. 5(e) and 5(f), respectively, indicate that the parallel compliance has indeed been much reduced as a consequence of the beveling of the probe tube.

Finally, calibration (4) indicates causal zs(t) and ys(t) in Fig. 5(d), aside from the remaining smearing of the Blackman window. In addition, it is worth noticing that the contribution at t = 0 represents the characteristic resistive behavior in both quantities, which they are both normalized by, and are thus identical. Finally, Re{ϵYs} and Im{ϵYs} have been reduced to a very low level in Figs. 5(e) and 5(f), respectively, also indicating causality.

C. Source-pressure estimation from source volume flow

Figures 6(a)–6(d) show for calibrations (1)–(4), respectively, the time-domain source pressures ps(t) and the estimated incident source pressures ps,0(t) [Eq. (30)]. Table I further lists the source-pressure error metrics ϵps [Eq. (31)] and time limits τZ0, obtained for calibrations (1)–(4).

FIG. 6.

FIG. 6.

(Color online) Time-domain source pressures ps(t) and the predicted incident source pressures ps,0(t) [Eq. (30)] of calibrations (a) (1) ER-10X, (b) (2) ER-10X leak, (c) (3) ER-10X bevel, and (d) (4) Titan (see Sec. III D for details).

Comparing calibrations (1) and (2) in Figs. 6(a) and 6(b), respectively, a similar behavior and extension into negative time of ps(t) is observed as for the source impedances zs(t) in Figs. 5(a) and 5(b). This is in accordance with Eqs. (6) and (7), that Ps and Zs are scaled by the same factor. In addition, ps,0(t) are very similar between the two calibrations and do in either case not coincide with ps(t) over any intervals prior to τZ0. ϵps in Table I indicate that the calibrations are quite poor.

Comparing calibrations (1) and (3) in Figs. 6(a) and 6(c), respectively, a behavior is also observed between ps(t) that is similar to zs(t) in Figs. 5(a) and 5(c). ps,0(t) are still very similar, and for calibration (3), ps(t) and ps,0(t) are to a larger degree aligned, but not completely. Also, ϵps in Table I is reduced accordingly.

Finally, calibration (4) in Fig. 6(d) shows well-aligned ps(t) and ps,0(t) until approximately 0.2 ms, representing the actual time-limit of the resistive behavior of the probe tube. This indicates that the source pressure is indeed unaffected by any non-causal, non-resistive parallel components. Finally, ϵps in Table I has been reduced to a very low value.

D. Incident sound power

Figures 7(a)–7(d) show, for calibrations (1)–(4), respectively, the predicted incident sound power w0(t) [Eq. (34)], and Table I further lists the predicted incident acoustic energy E0 [Eq. (35)]. Comparing calibrations (1) and (2), it is evident that the leak in the ear tip dramatically decreases w0(t) and E0, mainly due to the much lower source pressure. Comparing calibrations (1) and (3), the reduction of the parallel compliance in the source parameters increases w0(t) and E0. It is difficult to interpret the results of calibration (4) as there is nothing to compare with, but it is still reported for reference.

FIG. 7.

FIG. 7.

(Color online) Predicted incident sound power w0(t) [Eq. (34)] of calibrations (a) (1) ER-10X, (b) (2) ER-10X leak, (c) (3) ER-10X bevel, and (d) (4) Titan (see Sec. III D for details).

V. DISCUSSION

The incapability of identifying parallel components in the source parameters has been a significant limitation for the most widely used methodologies for acoustic Thévenin calibration of ear probes. A parallel component, such as an acoustic compliance resulting from improperly accounting for evanescent modes, can cause large errors in subsequent measurements of impedance and reflectance (Nørgaard et al., 2017b). This paper has proposed a set of error metrics, in addition to the calibration error, that can effectively identify such parallel components in the source parameters. Nørgaard et al. (2017a) demonstrated that the Hilbert transform is an efficient method to characterize the signal non-causality in the measured waveguide reflectance. The results presented in this paper further support the viability of the methodology of Nørgaard et al. (2017a). An important property of the Hilbert transform, and thus the Hilbert-transform-based estimation error, of a band-limited spectrum, is that they quantify the signal causality of a transfer function. Any signal non-causality of a transfer function represents the superimposed effects of actual physical non-causality and non-causality resulting from signal-processing artifacts when observing the physical system through the time-domain transfer function using the inverse Fourier transform. A non-zero Hilbert-transform estimation error can therefore not be interpreted as a physically non-causal system. However, a low Hilbert-transform estimation error does give an indication that the physical system is causal.

One of the main advantages of the Hilbert transform is that it is able to provide information on inaccuracies in a calibration that are confined to small frequency ranges. Such effect is visible around 4 kHz in Figs. 5(e) and 5(f) for calibrations (1)–(3) where small variations in the source admittance estimation error can be observed. Proposedly, this is caused by a combination of the source parameters mainly being determined from impedance minima in this region and not accounting for flow losses. Despite the effectiveness of the Hilbert transform, it might not be possible to utilize it in certain cases due to potential unknown problems encountered with frequency truncation, besides those resulting from a component in parallel with the source impedance, and noise. It should be noted that the source-admittance estimation error ϵYs relies on that no physical lumped elements are actually present in the true source parameters, e.g., that the effect of the material properties of the ear tip are negligible in the source admittance and that the probe tube is uniform at the tip of the probe. While the assessment of the probe-pressure estimation error ϵps might be less direct than quantifying the non-causality of the source admittance, it is a more robust methodology in its basic formulation because it does not rely on the inverse Fourier transform of a transfer function. The limitations of this probe-pressure estimation error are mainly constituted in determining a representative time limitation τZ0. This quantity will vary depending on the length of the probe tube, and the source-pressure error might not be immediately comparable between different probe types. An additional fundamental difference between the two proposed error metrics is worth emphasizing. While the Hilbert-transform methodology relies on the undesired parallel component comprising either a physical or signal non-causality, the probe-pressure estimation error only relies on the time-limited resistive behavior of the probe tube. In addition, neither of the proposed error metrics are applicable in the case of a protruded microphone tube in the probe (e.g., Keefe et al., 1992; Voss and Allen, 1994), as this protrusion results in an actual lumped compliance in parallel with the source parameters. In principle, this condition of resistive behavior in the source admittance for both error metrics is not satisfied with the beveling of the probe tube on the ER-10X probe either. However, the phantom parallel compliance appears to be much more dominant than potential errors due to this effect.

A fundamental prerequisite in the methodology proposed by Nørgaard et al. (2017b) and utilized for calibration (4) is that the waveguide lengths are known quantities, which was also the case for the waveguides used for this calibration. The same assumption cannot be made for the calibration in the built-in ER-10X calibration waveguide due to the uncertainty associated with the insertion of the probe using an ear tip and the movable piston termination. The results of this study indicate that properly accounting for evanescent modes during calibration is crucial for obtaining a set of causal source parameters. The proposed error metrics could potentially enable such calibration procedure without preliminary knowledge of the waveguide lengths, though this has not been further investigated. However, it is interesting to note that introducing a signal non-causal component into the reference impedances, in the form of an inertance to model evanescent modes, during calibration results in causal source parameters. Conversely, failing to account for evanescent modes by assuming causal, plane-wave reference impedances results in non-causal source parameters.

Although the presented results are each based on one single calibration, the standard calibrations of the ER-10X and Titan probes, i.e., calibrations (1) and (4), respectively, yielded reproducible results with negligible deviations over the course of several calibrations, utilizing different probe tubes and ear tips. However, the physical modifications performed on the probes in calibrations (2) and (3) are less reproducible. The ear-tip modification in calibration (2) involved manually cutting away an annular sector of the mushroom part of the ear tip. This introduces a different leak when performing a similar modification to another ear tip. In addition, the leak in the ear tip is shaped slightly different with different insertions into the calibration waveguide. The beveling of the probe tube in calibration (3) also involves a manual modification of the probe tube, resulting in an inevitable variation between different beveled probe tubes, unless the beveling can be incorporated as part of the manufacturing process. In the light of these inherent variations in deliberately induced parallel components, calibrations (2) and (3) should merely be considered as examples of such errors and not an assessment of, for example, the general accuracy in the methodology in beveling the probe tube of the ER-10X probe.

The present study has proposed a set of error metrics that may be used in combination with the existing calibration error to further assess the validity of an acoustic Thévenin calibration procedure. However, no error thresholds have been defined that the source parameters should satisfy to be considered valid, although the presented examples in calibrations (1)–(4) can be used as a set of guidelines for what could be considered acceptable and poor calibrations. Defining such thresholds would not only require additional data from different probes, but should also include a systematic assessment of acceptable errors in subsequent measurements under similar conditions to the expected use case, e.g., adult or infant ears. The error metrics considered in this study were normalized by probe-tube characteristic impedance or admittance, mainly to yield similar values for the source impedance and admittance and facilitate the comparison of data between the two utilized probes. To assess the potential influence of a given parallel component on a subsequent measurement, normalization should be carried out with respect to the characteristic impedance of the considered load. While the proposed error metrics can be a valuable tool in improving acoustic Thévenin-equivalent calibration methods, they could also be utilized in routine calibrations in clinical settings. However, depending on the stability of the utilized calibration methodology, this may not be necessary. The authors strongly encourage other researchers in the field to familiarize themselves with and apply the proposed or similar alternative error metrics in their research that can identify undesired parallel components and objectively assess the validity of the source parameters. However, it cannot be precluded that the proposed error metrics may not be able to reveal other unknown types of calibration errors.

The authors would like to emphasize that the objective of the present study is not to compare the performance of the ER-10X and Titan ear probes, but rather to explore underlying calibration methodologies. Ideally, methods for measuring impedance and reflectance should be independent of the utilized probe. The beveling of the probe tube is only possible for the ER-10X probe due to the axisymmetrical construction of the probe tube. In addition, an appropriate coupling mechanism of the ER-10X probe to the waveguides proposed by Nørgaard et al. (2017b) does not currently exist, and thus the Titan probe was utilized instead. However, this study does point toward the need for a calibration methodology that can be used in combination with the built-in ER-10X waveguide that takes evanescent modes into account. While a set of causal source parameters should generally be desired, it has the effect that subsequent measurements of impedance can be interpreted as a transfer impedance, the ratio of sound pressure on the probe microphone to volume flow injected by the probe speaker. These measurements thus remain affected by evanescent modes, and additional compensation, such as proposed by Fletcher et al. (2005) or Nørgaard et al. (2017a) must be applied to obtain an estimate of the plane-wave impedance or reflectance.

VI. CONCLUSION

This paper has proposed a set of error metrics that can be used to assess the physical validity of the source parameters resulting from an acoustic Thévenin calibration procedure. As opposed to existing metrics, the proposed metrics are capable of identifying undesired physical or phantom parallel components in the source parameters. Four different calibrations using two existing calibration methodologies were compared to assess the effects of different physical and phantom components on the error metrics. The results showed that the proposed error metrics were accurately able to identify and quantify both the physical and phantom parallel components. Furthermore, the calibration methodology proposed by Nørgaard et al. (2017b) yielded a set of causal source parameters. The proposed additional error metrics could potentially pave the way for more reliable and accurate measurements of ear-canal impedance and reflectance, and increase the precision in in situ calibration of stimulus levels.

ACKNOWLEDGMENTS

The authors would like to thank Douglas H. Keefe for pointing out the similarity of the L-C coupling with an elemental transmission-line section. This research was funded by Interacoustics A/S, the Innovation Fund Denmark under File No. 5189-00100B, and by a grant from the National Institutes of Health: Grant No. R01 DC008318 (S.T.N.).

Footnotes

1

An alternative interpretation is that the sign change occurs when a series component is represented as an equivalent inverse parallel component prior to being incorporated into the source parameters.

2

www.playrec.co.uk (Last viewed 12 February 2018).

References

  • 1. Allen, J. B. (1985). “ Measurement of eardrum acoustic impedance,” in Peripheral Auditory Mechanisms, edited by Allen J., Hall J., Hubbard A., Neely S., and Tubis A. ( Springer, New York: ), pp. 44–51. [Google Scholar]
  • 2. Brass, D. , and Locke, A. (1997). “ The effect of the evanescent wave upon acoustic measurements in the human ear canal,” J. Acoust. Soc. Am. 101, 2164–2175. 10.1121/1.418244 [DOI] [PubMed] [Google Scholar]
  • 3. Ellison, J. C. , Gorga, M. , Cohn, E. , Fitzpatrick, D. , Sanford, C. A. , and Keefe, D. H. (2012). “ Wideband acoustic transfer functions predict middle-ear effusion,” Laryngoscope 122, 887–894. 10.1002/lary.23182 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Feeney, M. P. , and Keefe, D. H. (2001). “ Estimating the acoustic reflex threshold from wideband measures of reflectance, admittance, and power,” Ear Hear. 22, 316–332. 10.1097/00003446-200108000-00006 [DOI] [PubMed] [Google Scholar]
  • 5. Fletcher, N. H. , Smith, J. , Tarnopolsky, A. Z. , and Wolfe, J. (2005). “ Acoustic impedance measurements-correction for probe geometry mismatch,” J. Acoust. Soc. Am. 117, 2889–2895. 10.1121/1.1879192 [DOI] [PubMed] [Google Scholar]
  • 6. Hsu, H. P. (1967). “ Causality and the Hilbert transform,” in Fourier Analysis, revised edition ( Simon and Schuster, New York: ), pp. 239–243. [Google Scholar]
  • 7. Keefe, D. H. (1984). “ Acoustical wave propagation in cylindrical ducts: Transmission line parameter approximations for isothermal and nonisothermal boundary conditions,” J. Acoust. Soc. Am. 75, 58–62. 10.1121/1.390300 [DOI] [Google Scholar]
  • 8. Keefe, D. H. , and Benade, A. H. (1981). “ Impedance measurement source and microphone proximity effects,” J. Acoust. Soc. Am. 69, 1489–1495. 10.1121/1.385783 [DOI] [Google Scholar]
  • 9. Keefe, D. H. , Folsom, R. C. , Gorga, M. P. , Vohr, B. R. , Bulen, J. C. , and Norton, S. J. (2000). “ Identification of neonatal hearing impairment: Ear-canal measurements of acoustic admittance and reflectance in neonates,” Ear Hear. 21, 443–461. 10.1097/00003446-200010000-00009 [DOI] [PubMed] [Google Scholar]
  • 10. Keefe, D. H. , Ling, R. , and Bulen, J. C. (1992). “ Method to measure acoustic impedance and reflection coefficient,” J. Acoust. Soc. Am. 91, 470–485. 10.1121/1.402733 [DOI] [PubMed] [Google Scholar]
  • 11. Keefe, D. H. , Sanford, C. A. , Ellison, J. C. , Fitzpatrick, D. F. , and Gorga, M. P. (2012). “ Wideband aural acoustic absorbance predicts conductive hearing loss in children,” Int. J. Audiol. 51, 880–891. 10.3109/14992027.2012.721936 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Lewis, J. D. , McCreery, R. W. , Neely, S. T. , and Stelmachowicz, P. G. (2009). “ Comparison of in-situ calibration methods for quantifying input to the middle ear,” J. Acoust. Soc. Am. 126, 3114–3124. 10.1121/1.3243310 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. McCreery, R. W. , Pittman, A. , Lewis, J. , Neely, S. T. , and Stelmachowicz, P. G. (2009). “ Use of forward pressure level to minimize the influence of acoustic standing waves during probe-microphone hearing-aid verification,” J. Acoust. Soc. Am. 126, 15–24. 10.1121/1.3143142 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Merchant, G. R. , Röösli, C. , Niesten, M. E. F. , Hamade, M. A. , Lee, D. J. , McKinnon, M. L. , Ulku, C. H. , Rosowski, J. J. , Merchant, S. N. , and Nakajima, H. H. (2014). “ Power reflectance as a screening tool for the diagnosis of superior semicircular canal dehiscence,” Otol. Neurotol. 36, 172–177 10.1097/MAO.0000000000000294. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Nørgaard, K. R. , Fernandez-Grande, E. , and Laugesen, S. (2017a). “ Compensating for evanescent modes and estimating characteristic impedance in waveguide acoustic impedance measurements,” J. Acoust. Soc. Am. 142, 3497–3509. 10.1121/1.5016808 [DOI] [PubMed] [Google Scholar]
  • 16. Nørgaard, K. R. , Fernandez-Grande, E. , and Laugesen, S. (2017b). “ Incorporating evanescent modes and flow losses into reference impedances in acoustic Thévenin calibration,” J. Acoust. Soc. Am. 142, 3013–3024. 10.1121/1.5010891 [DOI] [PubMed] [Google Scholar]
  • 17. Oppenheim, A. V. , and Schafer, R. W. (1975). “ Discrete Hilbert transforms,” in Digital Signal Processing ( Prentice-Hall, Englewood Cliffs, NJ: ), pp. 337–375. [Google Scholar]
  • 18. Papoulis, A. (1962). “ Hilbert transforms,” in The Fourier Integral and Its Applications ( McGraw-Hill, New York: ), pp. 198–201. [Google Scholar]
  • 19. Piskorski, P. , Keefe, D. H. , Simmons, J. L. , and Gorga, M. P. (1999). “ Prediction of conductive hearing loss based on acoustic ear-canal response using a multivariate clinical decision theory,” J. Acoust. Soc. Am. 105, 1749–1764. 10.1121/1.426713 [DOI] [PubMed] [Google Scholar]
  • 20. Rasetshwane, D. M. , and Neely, S. T. (2011). “ Inverse solution of ear-canal area function from reflectance,” J. Acoust. Soc. Am. 130, 3873–3881. 10.1121/1.3654019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Rasetshwane, D. M. , Neely, S. T. , Allen, J. B. , and Shera, C. A. (2012). “ Reflectance of acoustic horns and solution of the inverse problem,” J. Acoust. Soc. Am. 131, 1863–1873. 10.1121/1.3681923 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Scheperle, R. A. , Goodman, S. S. , and Neely, S. T. (2011). “ Further assessment of forward pressure level for in situ calibration,” J. Acoust. Soc. Am. 130, 3882–3892. 10.1121/1.3655878 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Scheperle, R. A. , Neely, S. T. , Kopun, J. G. , and Gorga, M. P. (2008). “ Influence of in situ, sound-level calibration on distortion-product otoacoustic emission variability,” J. Acoust. Soc. Am. 124, 288–300. 10.1121/1.2931953 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Siegel, J. H. , and Neely, S. T. (2017). “ Ear tip modification greatly reduces evanescent waves,” in Proc. 40th MidWinter Meet., Association for Research in Otolaryngology, pp. 55–56. [Google Scholar]
  • 25. Souza, N. N. , Dhar, S. , Neely, S. T. , and Siegel, J. H. (2014). “ Comparison of nine methods to estimate ear-canal stimulus levels,” J. Acoust. Soc. Am. 136, 1768–1787. 10.1121/1.4894787 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Voss, S. E. , and Allen, J. B. (1994). “ Measurement of acoustic impedance and reflectance in the human ear canal,” J. Acoust. Soc. Am. 95, 372–384. 10.1121/1.408329 [DOI] [PubMed] [Google Scholar]
  • 27. Withnell, R. H. , Jeng, P. S. , Waldvogel, K. , Morgenstein, K. , and Allen, J. B. (2009). “ An in situ calibration for hearing thresholds,” J. Acoust. Soc. Am. 125, 1605–1611. 10.1121/1.3075551 [DOI] [PubMed] [Google Scholar]

Articles from The Journal of the Acoustical Society of America are provided here courtesy of Acoustical Society of America

RESOURCES