. 2018 Feb 25;23(2):121–125. doi: 10.1016/j.rpor.2018.02.001

Assessment of the Monitor Unit Objective tool for VMAT in the Eclipse treatment planning system

Sara Jiménez-Puertas a,, David Sánchez-Artuñedo b, Marcelino Hermida-López b
PMCID: PMC5856680  PMID: 29556140

Abstract

Aim

This work aims to achieve the highest possible monitor units (MU) reduction using the MU Objective tool included in the Eclipse treatment planning system, while preserving the plan quality.

Background

The treatment planning system Eclipse (Varian Medical Systems, Palo Alto, CA) includes a control mechanism for the number of monitor units of volumetric modulated arc therapy (VMAT) plans, named the MU Objective tool.

Material and methods

Forty prostate plans, 20 gynecological plans and 20 head and neck plans designed with VMAT were retrospectively studied. Each plan (base plan) was optimized without using the MU Objective tool, and it was re-optimized with different values of the Maximum MU (MaxMU) parameter of the MU Objective tool. MU differences were analyzed with a paired samples t-test and changes in plan quality were assessed with a set of parameters for OARs and PTVs.

Results

The average relative MU difference (ΔMU¯) considering all treatment sites, was the highest when MaxMU = 400 (−4.2%, p < 0.001). For prostate plans, the lowest ΔMU¯ was obtained (−3.7%, p < 0.001). For head and neck plans ΔMU¯ was −7.3% (p < 0.001) and for gynecological plans ΔMU¯ was 7.0% (p = 0.002). Although similar MU reductions were observed for both sites, for some gynecological plans maximum differences were greater than 10%. All the assessed parameters for PTVs and OARs sparing showed average differences below 2%.

Conclusion

For the three studied clinical sites, establishing MaxMU = 400 led to the optimum MU reduction, maintaining the original dose distribution and dosimetric parameters practically unaltered.

Keywords: VMAT, RapidArc, Eclipse, Monitor units

1. Background

The incorporation of intensity-modulated techniques to radiotherapy is supposed to improve dose sparing and conformity compared to conventional 3DCRT techniques, but also implies the delivery of a higher number of monitor units (MU).1 The increase in MU entails more secondary radiation dose and, therefore, a higher risk of developing secondary cancers.2, 3

Volumetric modulated arc therapy (VMAT) is a technique based on a simultaneous variation of MLC position, gantry angle and dose rate to improve dose sparing and to shorten the treatment time.4, 5 In modulated techniques, the total MU are not linearly related to prescribed dose, and factors as leaf or gantry speed may substantially affect the total MU of a plan.

In the Eclipse treatment planning system v.11.0.13 (Varian Medical Systems, Palo Alto, CA), the Progressive Resolution Optimizer (PRO3) algorithm is included for VMAT planning, in which three different tools can be differentiated: Dose Volume Objective, Normal Tissue Objective and MU Objective (see Fig. 1). The first two affect the dose homogeneity and conformity in the tumor and dose sparing in healthy tissue. As dose homogeneity and conformity demands rise, the leaf movement often produces many small area segments resulting in increased MU. Conversely, the MU Objective aims to decrease the total MU of the plan. MU Objective and Normal Tissue Objective tools are not mandatory to optimize a plan, so it is a choice of the users to include them in the optimization process.

Fig. 1.

Fig. 1

MU Objective tool in Eclipse Progressive Resolution Optimizer.

Several parameters may be adjusted in the MU Objective tool, namely, Maximum MU (MaxMU), Minimum MU (MinMU) and Strength (S). The MU goal is defined by MinMU and MaxMU, and the priority enforced by the MU reduction goal is defined by the S value. In PRO3 algorithm the plan goals are defined by relative priorities assigned to dose–volume histogram (DVH) constraints that the user is asked to specify in the Dose Volume Objective section. The greater the value of a priority of a constraint, the more probable it is that the constraint will be reached, which is useful in conflicting constraints. In the MU Objective tool, the S parameter ranges from 0 to 100 and it is not in the same scale as the previous priorities.

Apart from the manufacturer definition, there are no further explanations in literature about how the MU Objective tool works or how it reduces the MU and its utilization may envelop a compromise between the MU reduction and the plan quality.

Ahamed et al.6 studied the MU Objective tool for ten hypopharinx cancer patients to quantify differences between MU deprived plans and those freely optimized. Clemente et al.7 investigated the maximum MU reduction for prostate plans in seven patients and reported the dose differences to organs at risk (OARs). Bao-Tiang et al.8 followed a similar procedure in stereotactic ablative VMAT lung plans obtaining a MU reduction higher than 200 MU/Gy without compromising the target dose coverage and the OARs dose sparing. These studies evaluated the effect of the MU Objective tool on the total MU and reported the changes obtained in dosimetric parameters.

2. Aim

This work aims to achieve the highest possible MU reduction using the MU Objective tool while preserving the plan quality as unaltered as possible. As the MU Objective tool acts as a ‘black box’, we have evaluated this tool empirically. To improve the statistics and to find recommended values of the MU Objective parameters without losing the PTV dose coverage and OARs dose sparing, we analyzed a higher number of patients for the prostate, head and neck, and gynecological sites than in previous works. These sites represent 93% of the patients treated with VMAT in our institution.

3. Materials and methods

We retrospectively studied the last 40 prostate plans, 20 gynecological plans and 20 head and neck plans treated with RapidArc (Varian implementation of VMAT) in 2016 at our center.

Prescribed doses for prostate plans were 57 Gy or 59.5 Gy for the prostate plus seminal vesicles PTV and 72 Gy or 77 Gy for the simultaneous integrated boost to the prostate PTV, in 30 or 35 fractions. For gynecological plans, prescribed doses were 45 Gy or 50.4 Gy in 25 or 28 fractions. For head and neck plans, prescribed doses were 60 Gy or 70 Gy for the tumoral region plus affected lymph nodes, and 54 Gy for non-affected lymph nodes in 30 or 35 fractions.

We used one full arc for the prostate and two full arcs for the head and neck and gynecological sites. Plans were optimized with the PRO3 algorithm and calculated with the Acuros XB algorithm with 2.5 mm grid resolution in the Eclipse treatment planning system v.11.0.31 using 6 MV photon beams with a maximum dose rate of 600 MU/min.

Fig. 1 displays the interface of PRO3 in Eclipse during the optimization of a prostate plan. At the top-left corner of the screen, the Normal Tissue Objective, the Dose Volume Objective and priorities of the OARs and PTVs are specified. To the right, the DVH and the evolution of the optimization process are displayed. At the bottom-left corner, the MU Objective tool and its parameters are available to use.

Clemente et al.7 studied the effect of the MU Objective tool on the MU variation for S = 50 and S = 100. The highest mean MU reduction was obtained for S = 100 (ΔMU = −28%) at the expense of a high variation in the target mean homogeneity index value (ΔHI = −23%). Following the manufacturer suggestion to set the S parameter from 50 to 100, we set S = 70 for all the tests reported. As in previous studies, we kept MinMU = 0 and fixed values of other optimization parameters, as dose constraints, arc-length or collimator angle, to separately assess the influence of the MaxMU parameter on the calculated MU.

For each patient, a plan was optimized without using the MU Objective tool, called base plan, with the corresponding MU called MUbase. This plan was re-optimized with the following MaxMU values: 400, 500, 600, 700, 800, 1000 and 1200; thus generating seven additional plans, called re-optimized plans. In each re-optimization, the dose constraints and their priorities remained unchanged to evaluate only the MU Objective tool influence on the final MU. Once the whole set of plans was re-optimized and re-calculated, the differences obtained in the cumulative DVH and in MU were analyzed with a paired samples t-test (α = 0.05). Plan quality was assessed with the dosimetric parameters indicated in Table 1 for OARs, and with the average dose, dose coverage (D2%, D98%), dose conformity (V95%, V107%) and dose homogeneity (D5%–D95%) for the PTVs.

Table 1.

Dosimetric parameters evaluated for OARs.

Sites OARs Parameters analyzed apart from average dose
Prostate Rectum V50Gy, V60Gy, V70Gy
Bladder D2%, D67%, V30Gy
Femoral heads D2%, V45Gy



Gynecological Rectum V30Gy, V40Gy, V45Gy, D50%
Bladder V30Gy, V40Gy, V45Gy
Bowel V40Gy, D30%
Femoral heads V30Gy, D2%



Head and neck Mandible V70Gy
Spinal cord and brainstem D2%
Parotids glands D33%, D50%, D66%

4. Results

Fig. 2 reports the average relative MU difference (ΔMU¯) as a function of the MaxMU value considering all treatment sites. Three results are observed:

  • MU reduction was the highest when MaxMU = 400 (−4.2%, p < 0.001).

  • A trend of decreasing the ΔMU¯ is observed when using a MaxMU value greater than 400.

  • The resulting MU tend to increase for MaxMU values higher than MUbase, as 1000 and 1200.

Fig. 2.

Fig. 2

Global MU variation according to MaxMU values.

Fig. 3 displays the relative ΔMU¯ according to MUbase for different clinical sites. Each point represents one patient where the ΔMU¯ was obtained considering the whole different MaxMU values. Only for prostate plans, a differentiated behavior was found according to MUbase. For MUbase greater than 650, the ΔMU¯ is −2.2% (p < 0.001), being the maximum for MaxMU = 500 (−3.7%, p < 0.001). On the other hand, for prostate base plans with MUbase lower than 650, ΔMU¯=+1.7% (p < 0.001).

Fig. 3.

Fig. 3

MU reduction according to MUbase for each treatment site.

Table 2 summarizes the value of MaxMU that involves the maximum ΔMU¯ obtained for each treatment site.

Table 2.

MaxMU values with the largest average MU reduction for each clinical site.

Sites MaxMU ΔMU¯(%) p
Prostate (MUbase > 650 MU) 500 −3.7% <0.001
Gynecological 400 −7.0% 0.002
Head and neck 400 −7.3% <0.001

For all OARs in prostate plans, the average differences of each parameter were lower than 0.2%, except for the femoral heads, with an average dose difference of +1.3% (p = 0.002), and in hot–spots (ΔD2%¯=+1.4%, p = 0.005). As for the coverage, conformity and homogeneity in PTVs, differences were lower than 0.4%.

For gynecological plans, the only statistically significant difference was found in the bladder for V40Gy(%), becoming of +0.9% (p = 0.007). In a small set of patients, maximum differences greater than 10% were found, corresponding to plans with the highest MU reductions, up to 333 MU.

Head and neck results showed the maximum decrease in ΔMU¯ of all analyzed sites without compromising the OARs sparing. The average differences for all dosimetric parameters were below 0.7%, this value being the maximum average difference obtained in D2% and D5%–D95% for the high-dose PTV and in D50% for the parotid glands. Differences in the rest of dosimetric parameters for OARs and PTVs were lower than 0.3%. Unlike gynecological plans, all particular differences were lower than 5%, which confirms that the MU reduction maintaining the plan quality is a global trend for this site.

5. Discussion

Second malignancies risk in radiotherapy associated with secondary radiation dose is an issue of concern when developing new treatment techniques with high modulation,2, 3 due to the increase in MU. With the Eclipse MU Objective optimization tool, we investigated a practical method to reduce the MU keeping the plan quality unaltered. With fixed values of S = 70 and MinMU = 0, an optimal value of the MaxMU parameter to reach the maximum ΔMU¯ was obtained for different treatment sites.

For prostate plans, we obtained the lowest ΔMU¯ in comparison to other clinical sites being statistically significant only for MUbase greater than 650. Although the plan quality stayed unaltered, the variability in ΔMU¯ made it difficult to ensure satisfactory results in all cases. When MUbase was lower than 650, the trend is to increase the final MU. This trend becomes more pronounced when using a MaxMU value higher than MUbase (see Fig. 2), which is an unexpected result, as a higher MaxMU value is not supposed to incorporate an additional restriction that affects MUbase.

For gynecological plans, particular cases showed the greatest dosimetric differences related to the greatest ΔMU¯ values, demanding an additional evaluation of these re-optimized plans.

For head and neck plans, the maximum ΔMU¯ was achieved with the lowest dosimetric differences, making the optimization tool a useful choice for this site.

Clemente et al.7 achieved a ΔMU¯=28% for prostate plans with S = 100, MaxMU = 0.5 × MUbase and MinMU = 0, although differences in mean dose for the bladder and femoral heads were −12.4% and +9.6%, respectively. We did not achieve such a high MU average reduction but all dosimetric differences were below 2%. This implies a MU reduction keeping the plan practically unaltered, especially for plans with MUbase greater than 650.

Ahamed et al.6 investigated three values of MaxMU: 0.8 × MUbase, 0.65 × MUbase and 0.5 × MUbase; and reported superior quality scores for base plans of hypopharinx plans. They reported a MU decrease between 8.6% and 34.7% for MaxMU = 0.8 × MUbase. In our study, the average dosimetric differences were lower than 0.7%, but with a lower ΔMU¯ (−7.3%) for MaxMU = 400.

6. Conclusion

For the prostate, gynecological, and head and neck sites, a plan of similar quality to the base plan was obtained with lower MU. To improve the statistical results, we assessed a bigger set of VMAT plans than in previous publications. Establishing MaxMU = 400 leads to the optimum MU reduction in most cases, keeping the original dose distribution and dosimetric parameters practically unaltered.

Conflict of interest

None declared.

Financial disclosure

None declared.

Acknowledgements

The authors thank Marc Sancho and Diego Carrasco for their help in the data collection.

Contributor Information

Sara Jiménez-Puertas, Email: jpfcsara@gmail.com.

David Sánchez-Artuñedo, Email: d.sanchez.artunedo@gmail.com.

Marcelino Hermida-López, Email: mhermida@vhebron.net.

References

  • 1.Wolff D., Stieler F., Welzel G. Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer. Radiother Oncol. 2009;93(2):226–233. doi: 10.1016/j.radonc.2009.08.011. [DOI] [PubMed] [Google Scholar]
  • 2.Hall E.J. Intensity-modulated radiation therapy, protons, and the risk of second cancers. Int J Radiat Oncol Biol Phys. 2006;65(1):1–7. doi: 10.1016/j.ijrobp.2006.01.027. [DOI] [PubMed] [Google Scholar]
  • 3.Kry S.F., Salehpour M., Followill D.S. The calculated risk of fatal secondary malignancies from intensity-modulated radiation therapy. Int J Radiat Oncol Biol Phys. 2005;62(4):1195–1203. doi: 10.1016/j.ijrobp.2005.03.053. [DOI] [PubMed] [Google Scholar]
  • 4.Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35(1):310–317. doi: 10.1118/1.2818738. [DOI] [PubMed] [Google Scholar]
  • 5.Teoh M., Clark C.H., Wood K., Whitaker S., Nisbet A. Volumetric modulated arc therapy: a review of current literature and clinical use in practice. Br J Radiol. 2011;84(1007):967–996. doi: 10.1259/bjr/22373346. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Ahamed S., Singh N., Gudipudi D. Assessment of monitor unit limiting strategy using volumetric modulated arc therapy for cancer of hypopharynx. Phys Med. 2017;35:73–80. doi: 10.1016/j.ejmp.2017.01.016. [DOI] [PubMed] [Google Scholar]
  • 7.Clemente S., Cozzolino M., Chiumento C., Fiorentino A., Caivano R., Fusco V. Monitor unit optimization in RapidArc plans for prostate cancer. J Appl Clin Med Phys. 2013;14(3):4114. doi: 10.1120/jacmp.v14i3.4114. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Huang B.T., Lin Z., Lin P.-X., Lu J.-Y., Chen C.-Z. Monitor unit optimization in stereotactic body radiotherapy for small peripheral non-small cell lung cancer patients. Sci Rep. 2015;5:18453. doi: 10.1038/srep18453. [DOI] [PMC free article] [PubMed] [Google Scholar]