COMP report: CPQR technical quality control guidelines for brachytherapy remote afterloaders

Normand Frenière 1,
PMCID: PMC5859323  PMID: 29417727

Abstract

The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested in a variety of Canadian radiation treatment facilities. The development process enables rapid review and update to keep the guidelines current with changes in technology (the most updated version of this guideline can be found on the CPQR website). This particular TQC details recommended quality control testing of brachytherapy remote afterloaders.

Keywords: brachytherapy remote afterloaders, quality control guidelines, radiation treatment therapy equipment

1. INTRODUCTION

The Canadian Partnership for Quality Radiotherapy (CPQR) is an alliance amongst the three key national professional organizations involved in the delivery of radiation treatment in Canada: the Canadian Association of Radiation Oncology (CARO), the Canadian Organization of Medical Physicists (COMP), and the Canadian Association of Medical Radiation Technologists (CAMRT). Financial and strategic backing is provided by the federal government through the Canadian Partnership Against Cancer (CPAC), a national resource for advancing cancer prevention and treatment. The mandate of the CPQR is to support the universal availability of high quality and safe radiotherapy for all Canadians through system performance improvement and the development of consensus‐based guidelines and indicators to aid in radiation treatment program development and evaluation.

This document contains detailed performance objectives and safety criteria for Brachytherapy Remote Afterloaders. Please refer to the overarching document Technical Quality Control Guidelines for Canadian Radiation Treatment Centres 1 for a programmatic overview of technical quality control, and a description of how the performance objectives and criteria listed in this document should be interpreted. The development of the individual TQC guidelines is spearheaded by expert reviewers and involves broad stakeholder input from the medical physics and radiation oncology community.2

All information contained in this document is intended to be used at the discretion of each individual center to help guide quality and safety program improvement. There are no legal standards supporting this document; specific federal or provincial regulations and licence conditions take precedence over the content of this document.

2. SYSTEM DESCRIPTION

Brachytherapy is the placement of encapsulated radionuclides or a miniaturized x‐ray tube in, or adjacent to, tissue which has been prescribed a radiation dose.3, 4, 5, 6, 7, 8, 9, 10 This practice offers unique advantages to the management of several treatment sites and has been used to complement or replace external beam radiation therapy since the onset of radiation oncology.

Remote afterloading equipment was developed to reduce, and in many cases eliminate, the radiation exposure to members of the staff. With remote afterloading systems the user does not directly handle the radioactive source and the patient is irradiated in a shielded room with staff operating and monitoring the process remotely.

High dose rate (HDR) refers to treatment dose rates larger than 20 cGy/min. For all HDR remote afterloaders, a single and small (< 1 mm × 5 mm) radioactive source (mostly iridium‐192, rarely cobalt60), laser‐welded to a metallic cable, is moved out of the safe by a motor‐drive mechanism to step along the prescribed positions (dwell positions) with different irradiation times (dwell times). The user can preselect dwell positions and dwell times at selected positions in a number of applicator lines. The remote afterloader could receive up to two such sources with two independent cables permitting dose delivery in two applicator lines simultaneously. The source strength is approximately 40,000 cGy cm² h−1 (activity of ~370 GBq) on installation of a new iridium‐192 source, while it is of 23,000 cGy cm² h−1 (activity of ~74 GBq) for a new cobalt‐60 source. Because iridium‐192 has a relatively short half‐life (73.8 days), the sources are usually replaced about every 3 months. Cobalt‐60 has a longer half‐life (1 925 days or 5.27 yr), offering less frequent source replacement to every 5 yr. Typical HDR irradiation times are 5 to 30 min, and a treatment course may consist of several fractions.

Another form of treatment is pulsed dose rate (PDR) treatment. With a PDR device, irradiations are given in short “pulses” with the total treatment being given in 48 to 72 h. The mechanism for PDR units is very similar to that used in HDR units. PDR remote afterloaders also use a single iridium‐192 source attached to a cable. However, the source strength for these units is typically only 10% of the source strength of an iridium‐192 HDR unit.

High‐dose rate miniaturized x‐ray tube remote devices, along with intravascular and cardiovascular brachytherapy remote afterloaders using beta emitting radioactive sources, are beyond the scope of this document.

Various recommendations for brachytherapy quality assurance have been reported in the literature.11, 12, 13, 14, 15, 16, 17, 18 “Per treatment” tests must be executed prior to each treatment. “Treatment day” tests must be scheduled before treating the first patient of the day. For PDR remote afterloaders where treatments may last several days, “treatment day” tests should be performed prior to the initiation of the treatment.

Table 1.

Per treatment quality control tests

Designator Test Performance
Action
Per treatment (executed prior to each treatment)
T1 Plan data transfer from treatment planning computer Data integrity
T2 Plan dwell times adjustment See note
T3 Minimum dwell times See note
T4 Plan catheters’ connection to remote afterloader Reproducible
T5 Complete source retraction Functional
Treatment day (or per treatment for PDR)
D1 Treatment interrupt Functional
D2 Console displays (treatment status indicator) and key switch Functional
D3 Date, time, and source strength in treatment unit See note
D4 Source (and dummy) positional accuracy 2 mm
D5 Dwell time accuracy 2%

Table 2.

Quarterly quality control tests

Designator Test Performance
Action
Quarterly (or at source replacement)
Q1 Mechanical integrity of applicators, guide tubes, connectors Functional
Q2 Internal battery power supply (power failure recovery) Functional
Q3 Source/dummy interlocks Functional
Q4 Dummy wire positional accuracy 3 mm (1 mm see note)
Q5 Radiological source positional accuracy 1 mm
Q6 Source strength calibration 5%
Q7 Source homogeneity Reproducible
Q8 Records Complete

Table 3.

Annual quality control tests

Designator Test Performance
Action
Annually
A1 Hand crank operation Functional
A2 Leakage radiation Reproducible
A3 Multi‐channel indexer function Functional
A4 Dwell time accuracy 1%
A5 Timer linearity 1%
A6 Transit time/transit dose reproducibility Reproducible
A7 Dosimetric length of applicators and guide tubes 1 mm
A8 Applicators and templates dimensions Reproducible
A9 Shield integrity of shielded applicators Reproducible
A10 X ray marker positional accuracy 1 mm
A11 Document staff review of emergency response procedures Complete
A12 Independent quality control review Complete
A13 PDR sequencing (for PDR only) Functional

3. RELATED TECHNICAL QUALITY CONTROL GUIDELINES

In order to comprehensively assess brachytherapy remote afterloader performance, additional guideline tests, as outlined in related CPQR Technical Quality Control (TQC) guidelines, must also be completed and documented, as applicable. Related TQC guidelines, available at http://www.cpqr.ca/programs/technical-quality-control/, include:

  • Safety Systems19

  • Major Dosimetry Equipment20

  • Treatment Planning Systems21

4. TEST TABLES

4.A. Notes on per treatment tests

T1 Plan data imported from a treatment planning system into the treatment console should be verified for source strength, dwell positions, and dwell times. In case of many dwell positions, the verification of a sub‐set of positions is acceptable. For a plan already present in the treatment console, the same verification should be made to assure proper plan selection
T2 Plan dwell times adjustment by the treatment console for the treatment date should be verified (by an independent calculation: hand calculation, decay factor chart, or software calculation). Action level will depend on the treatment console decay frequency and time resolution; express in percentage of difference and/or in second
T3 Minimum dwell time should be verified against the device driven limits. Those limits should take into account both the effect of transit dose and positioning reproducibility. Some remote afterloader systems might have a positioning reproducibility dependence on dwell time. All dwell times should be equal or greater than set limit
T4 Catheters/applicators connections to remote afterloader indexer channels must match plan
T5 Survey the treatment room and patient to ensure that source has been completely retracted

4.B. Notes on treatment day (or per treatment for PDR) tests

D1 During source exposure, verify that the non‐emergency interrupt button (if equipped) retracts the source to its safe and shielded position
D2 On the treatment console, displays should be verified. At minimum, treatment status indicators should be verified by exposing a source. Indicators could be visual and audible. When a treatment key is available, its deactivation should prevent source exposition
D3 Remote afterloader console date and time are properly set. Decayed source strength is accurate compared to an independent calculation (hand calculation, decay factor chart or software calculation), taking into account treatment console decay frequency
D4 Verify accuracy of source drive mechanism positioning. A visual inspection with a camera is acceptable. Apply also to dummy drive mechanism if used to measure catheter length
D5 Comparison of dwell time accuracy with external standard such as a stopwatch. The dwell time used should be sufficiently long such that errors in the measurement of the time (e.g., reaction time of the observer) are less than 1%

4.C. Notes on quarterly tests

Q1 Verify the applicators, guide tubes, and connectors are exempt of damage (excessive wear, kinks, etc.)
Q2 The configuration of this test will depend on the design of the facility and equipment. Safety is the concern and tests should be designed accordingly. The first objective is to verify that the equipment safely retracts the source wire after a power failure. The second objective is to verify that the equipment properly records treatment delivered before power failure and permits to resume the treatment after power recovery
Q3 Verify functionality of remote afterloader interlocks related to source and dummy wires. This includes incorrect connection of applicator to transfer guide tube, incorrect connection of transfer guide tube to remote afterloader, and obstruction
Q4 Verify accuracy of dummy drive mechanism positioning. The purpose is to assure proper obstruction detection by assuring no false positive or false negative obstruction. If dummy drive mechanism is used to measure catheter length for treatment planning, then the action level should be lowered to 1 mm
Q5 Accuracy of source drive mechanism to be verified. Autoradiographs or ion‐chamber measurements could be used. If visual checks with in‐room cameras are to be used, source positioning in the cable construction should be verified independently
Q6 Comparison of measured source strength with manufacturer supplied value. On installation of a new source, source strength must be measured using calibrated re–entrant chamber and electrometer traceable to a national standards laboratory. The re‐entrant chamber and electrometer should have been calibrated within the last 2 yr. Measured source strength should be used for planning and treatment purposes. Discrepancies greater than 5% between the measured and the manufacturer's supplied source strengths must be investigated. This action level could be lowered to 3%22 if the manufacturer supplied source strength offers such precision. Stability of re‐entrant chamber should be verified prior to use. A second qualified medical physicist should perform a check of the calibration
Q7 Visual check on film that the radioactive material is evenly distributed in the encapsulated source. Most important for sources composed of multiple source pellets
Q8 Documentation relating to the daily quality control checks, preventive maintenance, service calls, and subsequent checks must be complete, legible, and the operator identified

4.D. Notes on annual tests

A1 Manual emergency hand crank functionality should be verified with manufacturer service engineer. It is desirable that each person responsible to operate the hand crank, in an emergency situation, practices its operation annually when a dummy wire replaces the source wire
A2 Monitor leakage radiation to check afterloader's safe integrity. Intensity of leakage radiation must be lower than the value set by manufacturer and local regulations
A3 Verify multichannel indexer functions properly. The wire must be sent to the proper programmed channel
A4 Comparison of dwell time accuracy with an external standard performed more rigorously than the treatment day test
A5 Verification of the linearity of the timer over a clinically relevant range. The action level represents deviations of measured values from those calculated using a linear fit to the measured data
A6 Reproducibility of transit time effect or transit dose effect or source speed between dwell positions. Can be verified using autoradiographs, ion‐chamber measurements, or visual checks with in‐room cameras. A fixed and reproducible applicator geometry is required to assure the same wire drive speed
A7 Reusable applicators and transfer guide tubes length should be measured to verify dosimetric lengths if used clinically as nominal values. This measurement could also verify that no debris has come into the lumen
A8 Verify physical dimensions of reusable applicators and templates (e.g., diameters, angles, shields). They must match dimensions used in the planning process
A9 Verify shielding integrity of shielded applicators. Visual and radiographic inspections should be performed
A10 Check x‐ray markers positional accuracy if used clinically for source positioning. If markers are only used to draw the applicator path, then only the integrity should be checked
A11 The configuration of this test will depend on the design of the facility and equipment, and local regulations. All staff should review the emergency procedures when a source fails to retract properly and remains outside the safe
A12 To ensure redundancy and adequate monitoring, a second qualified medical physicist must independently verify the implementation, analysis, and interpretation of the quality control tests at least annually
A13 For PDR only, verify pulse sequencing functionality according to manufacturer's recommendations

CONFLICT OF INTEREST

No conflicts of interest.

DISCLAIMER

All information contained in this document is intended to be used at the discretion of each individual centre to help guide quality and safety program improvement. There are no legal standards supporting this document; specific federal or provincial regulations and licence conditions take precedence over the content of this document. As a living document, the information contained within this document is subject to change at any time without notice. In no event shall the Canadian Partnership for Quality Radiotherapy (CPQR) or its partner associations, the Canadian Association of Radiation Oncology (CARO), the Canadian Organization of Medical Physicists (COMP), and the Canadian Association of Medical Radiation Technologists (CAMRT), be liable for any damages, losses, expenses, or costs whatsoever arising in connection with the use of this document.

ACKNOWLEDGMENTS

We thank the many people who participated in the production of this guideline. These include: Michelle Nielsen and Marie‐Joelle Bertrand (associate editors); the Quality Assurance and Radiation Safety Advisory Committee; the COMP Board of Directors, Erika Brown and the CPQR Steering Committee, and all individuals that submitted comments during the community review of this guideline.

REFERENCES

  • 1. Canadian Partnership for Quality Radiotherapy . Technical quality control guidelines for Canadian radiation treatment centres. 2016. May 1 [cited 2017 Jul 18]. Available from: http://www.cpqr.ca/programs/technical-quality-control
  • 2. Nielsen MK, Malkoske KE, Brown E, et al. Production, review, and impact of technical quality control guidelines in a national context. J Appl Clin Med Phys. 2016;17:2–15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Erickson BA, Bittner NHJ, Chadha M, Mourtada F, Demanes DJ. The American College of Radiology and the American Brachytherapy Society practice parameter for the performance of radionuclide‐based high‐dose‐rate brachytherapy. Brachytherapy. 2017;16:75–84. Available from: http://www.sciencedirect.com/science/article/pii/S1538472116304810 [DOI] [PubMed] [Google Scholar]
  • 4. Glasgow GP. Brachytherapy In: Van Dyk J, ed. The Modern Technology of Radiation Oncology: A Compendium for Medical Physicists and Radiation Oncologists. Madison, WI: Medical Physics Publishing; 1999:695–752. ISBN 978‐0944838228. [Google Scholar]
  • 5. Glasgow GP, Bourland JP, Grigsby PW, Meli JA, Weaver KA. Report of Task Group 41: Remote Afterloading Technology. New York, NY: American Institute of Physics; 1993. Available from: https://www.aapm.org/pubs/reports/rpt_41.pdf [Google Scholar]
  • 6. Kubo HD, Glasgow GP, Pethel TD, Thomadsen BR, Williamson BF. High dose‐rate brachytherapy treatment delivery: report of the AAPM Radiation Therapy Committee Task Group No. 59. Med Phys. 1998;25:375–403. [DOI] [PubMed] [Google Scholar]
  • 7. Nath R, Anderson LL, Meli JA, Olch AJ, Stitt JA, Williamson JF. Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine. Med Phys. 1997;24:1557–1598. [DOI] [PubMed] [Google Scholar]
  • 8. Podgorsak EB, ed. Radiation Oncology Physics: A Handbook for Teachers and Students. Vienna, Austria: International Atomic Energy Agency; 2005. Jul [cited 2016 Jul 19]. Available from: http://www-pub.iaea.org/mtcd/publications/pdf/pub1196_web.pdf [Google Scholar]
  • 9. Purdy JA, editor. Medical physics monograph In: Advances in Radiation Oncology Physics: Dosimetry, Treatment Planning, and Brachytherapy. Vol. 19 Woodbury, NY: American Institute of Physics; 1992. ISBN 978‐1563960543. [Google Scholar]
  • 10. Rivard MJ, Beaulieu L, Thomadsen BR, editors. Clinical Brachytherapy Physics. AAPM Monograph No 38. Madison, WI, USA: Medical Physics Publishing; 2017. ISBN 978‐1‐936366‐57‐6 (hardcover book), ISBN 978‐1‐936366‐58‐3 (eBook). [Google Scholar]
  • 11. American College of Radiology . ACR‐AAPM technical standard for the performance of high‐dose‐rate brachytherapy physics. Resolution 50. Reston, VA: American College of Radiology; 2015. Available from: https://www.acr.org/-/media/ACR/Files/Practice-Parameters/HDR-BrachyTS.pdf [Google Scholar]
  • 12. Bastin KT, Podgorsak MB, Thomadsen BR. The transit dose component of high dose‐rate brachytherapy: direct measurements and clinical implications. Int J Radiat Oncol Biol Phys. 1993;26:695–702. Available from: https://doi.org/10.1016/0360-3016(93)90291-3 [DOI] [PubMed] [Google Scholar]
  • 13. DeWerd LA, Jursinic P, Kitchen R, Thomadsen BR. Quality assurance tool for high dose rate brachytherapy. Med Phys. 1995;22:435–440. Available from: https://doi.org/10.1118/1.597469 [DOI] [PubMed] [Google Scholar]
  • 14. Kutcher GJ, Coia L, Gillin M, et al. Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. Med Phys. 1994;21:581–618. Available from: https://doi.org/10.1118/1.597316 [DOI] [PubMed] [Google Scholar]
  • 15. Nag S, Dobelbower R, Glasgow G, et al. Inter‐society standards for the performance of brachytherapy: a joint report from ABS, ACMP and ACRO. Crit Rev Oncol Hematol. 2003;48:1–17. Available from: https://doi.org/10.1016/s1040-8428(03)00026-x [DOI] [PubMed] [Google Scholar]
  • 16. Rickey DW, Sasaki D, Bews J. A quality assurance tool for high dose rate brachytherapy. Med Phys. 2010;37:2525–2532. Available from: https://doi.org/10.1118/1.3425786 [DOI] [PubMed] [Google Scholar]
  • 17. Venselaar J, Pérez‐Calatayud J, editors. A practical guide to quality control of brachytherapy equipment, European Guidelines for Quality Assurance in Radiotherapy booklet no. 8. Brussels, Belgium: ESTRO; 2004. ISBN 90‐804532‐8. [Google Scholar]
  • 18. Zuofeng L, Mitchell TP, Palta JR, Liu C. A quality assurance test tool for high dose‐rate remote afterloading brachytherapy units. Med Phys. 1998;25:232–235. Available from: https://doi.org/10.1118/1.598185 [DOI] [PubMed] [Google Scholar]
  • 19. Canadian Partnership for Quality Radiotherapy . Technical Quality Control Guidelines for Safety Systems at Radiation Treatment Centres. 2016. Jul 12 [cited 2017 Jul 18]. Available from: http://www.cpqr.ca/programs/technical-quality-control/
  • 20. Canadian Partnership for Quality Radiotherapy . Technical Quality Control Guidelines for Major Dosimetry Equipment. 2015. Feb 28 [cited 2017 Jul 18]. Available from: http://www.cpqr.ca/programs/technical-quality-control/
  • 21. Canadian Partnership for Quality Radiotherapy . Technical Quality Control Guidelines for Treatment Planning Systems. 2015. Jun 29 [cited 2017 Jul 18]. Available from: http://www.cpqr.ca/programs/technical-quality-control/
  • 22. DeWerd LA, Ibbott GS, Meigooni AS, et al. A dosimetric uncertainty analysis for photon‐emitting brachytherapy sources: Report of AAPM Task Group No. 138 and GEC‐ESTRO. Med Phys. 2011;38:782−801. Available from: https://doi.org/10.1118/1.3533720 [DOI] [PMC free article] [PubMed] [Google Scholar]