Abstract
Ventilation distribution calculation using 4D CT has shown promising potential in several clinical applications. This study evaluated the direct geometric ventilation calculation method, namely the ΔV method, with xenon‐enhanced CT (XeCT) ventilation data from four sheep, and compared it with two other published methods, the Jacobian and the Hounsfield unit (HU) methods. Spearman correlation coefficient (SCC) and Dice similarity coefficient (DSC) were used for the evaluation and comparison. The average SCC with one standard deviation was with a range between 0.29 and 0.61 between the XeCT and DLV ventilation distributions. The average DSC value for lower 30% ventilation volumes between the XeCT and ΔV ventilation distributions was with a range between 0.48 and 0.63. Ventilation difference introduced by deformable image registration errors improved with smoothing. In conclusion, ventilation distributions generated using ΔV‐4D CT and deformable image registration are in reasonably agreement with the in vivo XeCT measured ventilation distribution.
PACS number(s): 87.57.N‐, 87.57.nj, 87.57.Q‐, 87.85.Pq
Keywords: ventilation distribution, 4D CT, deformable image registration, evaluation, XeCT
I. INTRODUCTION
Pulmonary ventilation and perfusion are not uniform throughout the lung and this heterogeneity may be increased if pulmonary diseases are involved. (1) Different imaging modalities are currently used clinically for pulmonary ventilation evaluation. Nuclear medicine modalities, including nuclear scintigraphy, 2 , 3 single photon emission computed tomography (SPECT), 4 , 5 and positron emission tomography (PET), 6 , 7 are most commonly used. Magnetic resonance imaging (MRI) 8 , 9 and computed tomography (CT) 10 , 11 are also capable of pulmonary functional imaging.
Recently, four‐dimensional (or respiratory gated) CT (4D CT) has been proposed for ventilation calculation using deformable image registration (DIR). 12 , 13 , 14 In the so‐called HU algorithm, deformable image registration is applied between respiratory phases of a 4D CT image set. The deformation transformation calculated from the DIR process maps corresponding voxels between two different respiratory phases. The Hounsfield unit (HU) differences between the corresponding voxels of the two phases are used to generate the differences in local densities which are related to ventilation. (12) Ventilation can also be calculated using the Jacobian of the deformation transformation (13) or using the transformation directly (geometrically). (14) Recently, promising results have been reported on the agreement between ventilation results obtained using the Jacobian method and xenon‐enhanced CT (XeCT) in a large animal model. (15) The latter measures regional ventilation by observing the contrast gas, xenon (Xe), wash‐in or wash‐out rate on serial CT images and is considered a gold standard for regional ventilation imaging. (16) Patient‐based comparative studies have also shown promising results between 4D CT vs. 4D‐PET using ‐labeled nanoparticles, (17) and 4D CT vs. SPECT. 17 , 18 Ventilation data derived from 4D CT also demonstrated good reproducibility in both animal and human studies. 19 , 20 , 21 This new ventilation calculation method using 4D CT has also been applied in lung disease detection, (22) radiotherapy treatment planning studies, (23) and assessment of radiotherapy response. (24)
The advantages of ventilation imaging using 4D CT include: 1) 4D CT is a mature, commercial and widespread technology; 2) no additional procedures, such as inhalation of a contrast, are needed, which makes the clinical implementation straightforward; 3) high spatial resolution of functional lung imaging can be achieved, which is one of the major advantages over nuclear medicine; and 4) 4D CT is a much less expensive procedure than other imaging modalities, which is an important consideration for clinical implementation.
In the current clinical practice of thoracic cancer radiotherapy, the differences in pulmonary ventilation and/or perfusion are mostly not considered when generating treatment plans. All regions of the lung are usually deemed equal. Clinicians and researchers have proposed to advance normal lung sparing in thoracic cancer radiotherapy by introducing ventilation and/or perfusion imaging into radiotherapy treatment planning. 25 , 26 If clinically implemented, it could lead to reduced radiation toxicity to lung regions with high perfusion and/or ventilation while adequate dose coverage of tumors is maintained. Since 4D CT has become widely accepted in thoracic cancer radiation therapy, logistically it is an excellent choice for ventilation calculation for the purpose of sparing functional lung during radiation dose treatment planning. However, as the ventilation data derived from 4D CT are introduced in clinical usage, further proper validation studies are needed.
This paper presents the in vivo evaluation of the 4D CT‐based geometric method of ventilation calculation, also called the ΔV method, and compares it with other two methods, the Jacobian and the HU methods, using previously published XeCT‐based ventilation data from studies on sheep in vivo. (15)
II. MATERIALS AND METHODS
A. Image data
Appropriate animal ethics approval was obtained for imaging acquisition protocols from the University of Iowa Animal Care and Use Committee and the study adhered to NIH guidelines for animal experimentation. Four adult male sheep were anesthetized and under positive ventilation during the experiments. 4D CT and XeCT scans were acquired with animal in the supine position. 4D CT images were reconstructed at different respiration phases. Zero% and 100% of the near end of expiration and near end of inspiration phases were used to calculate ventilation. XeCT scans were acquired with 12 axial slices over 45 breaths under respiratory gating at near end expiration. The XeCT based estimated regional lung ventilation was computed using Pulmonary Analysis Software Suite 11.0 by finding the constant of the exponential rise of the density from xenon gas wash‐in over multiple breaths. After resample, all images are with voxel dimension . More details of the image acquisition protocols can be found in the previous publication. (15)
B. Deformable image registration (DIR)
The Diffeomorphic Morphons (DM) method is based on matching edges and lines. (27) The morphon iteratively deforms a moving image into a target image by morphing the former. The process can be divided into three parts: estimation of displacement, accumulation of the deformation field, and deformation.
Estimation of displacement needed to deform the moving image into the target image is based on quadrature phase difference. (27) The accumulation of the deformation field uses the estimate of the displacement to update the deformation field. This is a two‐step process. The first step is to update the deformation field and to regularize the estimates of the accumulated field in order to fit the observed deformation to a deformation model. The second step is to morph the moving image to the target image according to the accumulated deformation field. These two steps are done iteratively as long as the displacement estimates are larger than a specified minimum displacement indicating acceptable convergence.
The quadrature phase difference method is used to estimate local displacement between two images. The advantage of this method over others, such as those based on gradient and polynomial expansion, is its invariance to image intensity and weak gradients. (28) Quadrature phase is a measure of local structure. Edges between bright and dark areas, dark lines, lines on dark background, and bright patches, all have different phases. The transition from one phase to another is continuous. Therefore, the difference in local phase between the moving and target images is a good measure of how much the moving image has to move to fit the target image. The local displacement is a function of the local phase along its associated direction. To estimate the local displacement a least squares estimator is used as follows:
(1) |
where v is the displacement field estimate, is the displacement field associated with direction i, is a measure of certainty derived from the magnitude of the phase difference, and is the unit normal vector for the i direction.
The displacement field of the current iteration is given by Eq. (1), which is used to interpolate a deformed version of the moving image, which is deformed based on the accumulated field and then compared to the target image in order to estimate a displacement field for the current iteration. The updated field () is formed by combining the accumulated field () and the displacement field () from the current iteration.
(2) |
where and are the uncertainties in the accumulated field and the current field, respectively. After acquiring the updated field in Eq. (2), as well as the uncertainty from the field, the following equation is used to determine the accumulated displacement field:
(3) |
where is the uncertainty in the updated field.
The morphons method is optimized to become diffeomorphic. As a result, Eq. (3) becomes:
(4) |
The deformation field produced by the DM is smoothed and iteratively used to transform the moving image and register it on to the target image. Smoothing is achieved using Gaussian regularization of the displacement field thereby reducing the influence of extreme values in a deformation field.
DM and diffeomorphic demons (DD) (29) DIR methods were evaluated using the 4D CT data of the sheep with manually delineated landmarks in the end‐expiration and end‐inspiration phases in this study. The deformation transformations were used to map the landmarks from the expiration phase to the inspiration phase. The mapped landmarks were then compared with the manually delineated landmarks in the inspiration phase and the difference was defined as the target registration error (TRE). There were 220 landmark pairs in each case. The registrations were performed with the multiscale feature. (30) Four levels of resolution were used in the registration, namely, , , , and . They were validated previously with average TRE (end‐expiration to end‐inspiration) was equal to for DM and for DD. (30)
Based on the evaluation results alluded to above, the DM DIR method was selected for the ventilation calculation in this study. The DM DIR computer program was developed by a group in the Netherlands. (27) It runs within the MATLAB platform (MathWorks, Natick, MA). Each registration took about 3 hrs on a personal computer using a 3 GHz CPU. The deformation transformations from the 4D CT data were used for ventilation matrices calculation.
C. Ventilation calculation
C.1 ΔV ventilation calculation algorithm
This study used the ΔV ventilation calculation algorithm 14 , 31 to derive ventilation from 4D CT data for four sheep. In the expiration phase of a 4D CT dataset, each voxel is a cuboid defined by eight vertices. In the inspiration phase, this cuboid is changed into a 12‐face polyhedron which is still comprised of the corresponding eight vertices. Any hexahedron or 12‐face polyhedron can be divided into six tetrahedrons. The volumes of the cuboid and the 12‐face polyhedron (deformed cuboid) are the sums of the volumes of their six corresponding tetrahedrons. In the inspiration phase, the DIR algorithm calculated the corresponding locations of the eight vertices that define the cuboid in the expiration phase. The volumes of each cuboid and the corresponding deformed cuboid are calculated using the corresponding vertices of each respective polyhedron.
The fundamental volume calculation is based on the volume calculation for each tetrahedron. The coordinates of the four vertices of a tetrahedron are used to determine its volume:
(5) |
where a, b, c, d are the vertices’ coordinates expressed as vectors. The volumes of the six tetrahedrons are summed up to generate the volume of the given polyhedron. The ventilation distribution was calculated as the distribution of,
(6) |
where ΔV is the volume change between expiration and inspiration, and is the initial volume for all voxels at expiration, which is constant across the whole exhale image. (11)
C.2 Jacobian algorithm
The Jacobian is a mathematical method where volume change is estimated using the first derivative of the deformation field. 13 , 24 , 32 , 33 Local volume change of the lung is calculated using the Jacobian of the transformation that maps the end expiration phase of 4D CT image to the end inspiration phase. Consider a vector displacement D(x,y,z) that transforms a voxel from its end expiration image to its corresponding location in the end inspiration image. The Jacobian operator J of this transformation is given by,
(7) |
where I is the identity matrix, (x,y,z) is the x component of D(x,y,z), (x,y,z) is the y component of D(x,y,z), and (x,y,z) is the z component of D(x,y,z). The determinant of the Jacobian is calculated at each voxel position according to Eq. (7). If the determinant of the Jacobian is unity, then no expansion or contraction of the voxel occurs. If the determinant is greater than one, there is local voxel (lung tissue) expansion; if less than one, there is local voxel contraction.
C.3 Hounsfield units (HU) method
The HU method uses deformable image registration to correlate voxels from the expiration image set to the anatomically corresponding voxels in the inspiration image. The change in density for each voxel is calculated by direct comparison of HUs. 11 , 12 The fraction of air in a CT lung volume may be estimated as,
(8) |
where HU represents the average Hounsfield unit value within the volume. Let be the fraction of air in the exhale CT volume, and the fraction of air in the inhale CT volume, then the change in volume is given by:
(9) |
Substitute Eq. (8) into Eq. (9), we get:
(10) |
C.4 Ventilation comparison
Due to the limits in dynamic scans, the volumes of the ventilation distributions by Xe‐enhanced CT imaging are usually limited to partial lungs. These partial lung distributions served as the masks in the ventilation distribution calculations with the 4D CT datasets. Thus the volumes with calculated ventilation matched with the XeCT data and the calculated ventilation distributions can be directly compared with the Xe ventilation images.
Image smoothing techniques are often used in image processing to reduce noise (34) and an averaging filter is a simple smoothing technique. (35) Different averaging filter sizes were applied to the 4D CT ventilation distributions before the comparison with the Xe ventilation distributions in order to reduce registration errors; these were , , and filters. Blood vessels and other high‐density regions were excluded in the ventilation calculation by all three methods. The ventilation value of the central voxel was the average value inside the box defined by the averaging filter. When the averaging filter mask was at the edge of the calculated ventilation region or where some ventilation values were not available due to the presence of blood vessels, the actual number of valid voxels inside the filter mask was used for the averaging calculation.
Because of the different metrics between the two ventilation imaging modalities, namely density difference for the XeCT versus volume change for the ΔV method, the two ventilation distributions were converted to the relative percentile distribution (32) first and then the Dice similarity coefficient (DSC) (36) was applied to calculate the similarity between the two ventilation volumes. When volumes A and B are compared, DSC is calculated as,
(11) |
The values of DSC index range between 1.0 and 0.0. ADSC of 1.0 indicates a complete similarity of the two volumes whereas a DSC of 0.0 indicates no similarity at all, with intermediate values describing proportional amounts of similarity. Similar to the cumulative dose‐volume histogram, if a certain percentage of lung volume is covered by no more than a certain ventilation value, this ventilation value is associated with the corresponding percentage value of the lung volume in the percentile distribution. The two aligned ventilation datasets were compared for similarity for the higher 30% and 50% of ventilation volumes (or 70 and 50 percentile).
Another way to compare the ventilation distributions is to calculate the Spearman correlation coefficient (SCC) between the ventilation data sets. (17) The voxel‐wise correlation was calculated between the distribution by XeCT and the other three 4D CT methods for all four cases. The absolute ventilation distributions without conversion to percentile ventilation distributions were used for the SCC calculations.
III. RESULTS
The average TRE with one standard deviation (1 SD) in the lungs for the four cases was for DM and for DD. The majority of the errors were within 3 mm, which is comparable to other studies 15 , 37 and consistent with our previous study. (30) Based on the evaluation results, DM DIR was selected for the ventilation calculations in this study.
The average Spearman coefficient r with 1 SD was (range 0.29–0.61) between the XeCT and ΔV ventilation distributions over the four cases (sheep). It was (range 0.31–0.61) between the XeCT and the Jacobian ventilation distributions, and (range 0.17–0.42) between the XeCT and the HU ventilation distributions. Based on the Spearman coefficient, Figs. 1 and 2 show the ventilation distributions for the best () and worst () cases, respectively, between XeCT, and XXDV, Jacobian, and HU methods. The qualitative agreements are patent by visual inspection.
The average DSC value for the upper 50% ventilation volumes between the XeCT and ΔV ventilation distributions was (range 0.61–0.74), while for the upper 70% ventilation volumes, it was (range 0.78–0.84). The average DSC results between XeCT and Jacobian ventilation distributions were the same (the individual values were different), while between XeCT and HU they were for the upper 50% ventilation volumes (range 0.58–0.68) and for the upper 70% ventilation volumes (range 0.74–0.81). The similarity of the lower 30% ventilation volumes was between the XeCT and ΔV and between the XeCT and HU, respectively.
It was found that the similarity between the ventilations calculated using 4D CT and XeCT depends on the averaging filter size applied to the 4D CT ventilation data. Figure 3(a) shows the mean Dice similarity coefficient versus averaging filter size and Fig. 3(b) the mean Spearman coefficient versus averaging filter size for the four cases. The dependence of the HU method on filter size was flat with the size larger than 3. At the filter size of 3 and smaller, HU method demonstrated better similarity to the ventilation distribution from XeCT than the other two methods, while with the larger filter size, the other two methods showed better similarity than HU method.
Figure 4 shows the mean DSC versus averaging filter size and ventilation volume between the ventilation calculated using ΔV and the ones calculated using the Jacobian and HU methods. DSC was high between the XXDV and Jacobian methods, while DSC was strongly dependent on ventilation volume and averaging filter size between the ΔV and HU methods; more smoothing (larger filter size) resulted in better similarity.
IV. DISCUSSION
No matter how accurate the deformable image registration algorithm is, registration errors are inevitable due to the noise inherent in CT images. (38) The registration errors in DIR are the reason for the similarity dependence on averaging filter size. The averaging is applied to smooth the ventilation distributions. The ventilation difference introduced by the registration error is reduced by smoothing, and the resulting ventilation distribution is thus closer to that obtained with XeCT.
The other major source of errors in DIR is artifacts in CT images. The sheep 4D CT data used in this study did not show obvious artifacts because they were mechanically ventilated and therefore breathed regularly. For clinical human 4D CT, motion artifacts exist when irregular motion is involved, in which mushroom artifacts due to the irregular motion of the large surface of the diaphragm are most obvious. (39) Such artifacts introduce DIR errors and images with such artifacts should not be used in ventilation calculations. (40)
The HU method needs less smoothing compared to the other two methods because during the ventilation calculation it excludes voxels in which large density difference occurs after registration. The registration errors often yield nonperfect matches between the two phases of a 4D CT dataset, such as chest wall‐lung tissue and blood vessels‐lung tissue interfaces, which could introduce large density differences at such interfaces. (14) These edge artifacts are easily removed during the ventilation calculation using the HU method by checking if any of the involved two sets of CT data has a higher HU number than normal lung tissue for any given voxel. The ventilation values for the artifact voxels were set to the values of the closest valid voxels. Because of the removal of the large errors during the ventilation calculation, HU method does not need much smoothing of the resulted ventilation distribution. For the other two methods, the deformation transformation from the registration is used to calculate the volume change, and the volume change errors introduced by the registration errors are not easily removed in the ventilation calculation. Additional smoothing of the ventilation distribution is an effective way to reduce the ventilation errors.
Both ΔV and Jacobian methods calculate ventilation based on the local volume change. One performs a direct geometric calculation and the other approximates volume change using the Jacobian determinant of a transformation. The volume change calculations are performed based on the same deformation transformation in each case. Thus the ventilation distributions generated by these two methods are similar to each other. HU method calculates ventilation based on the local HU change. Although the same deformation transformation is used, different target properties from the transformation are used in the calculations. The resulted ventilation distributions could be different due to registration errors and thus can be reduced by smoothing, which is seen in Fig. 4.
Because the accuracy of the ventilation distribution generated from 4D CT data and deformable image registration depends on the accuracy of the registration, which in turn depends on the registration algorithm and the quality of the 4D CT data, high‐quality 4D CT data are essential for producing accurate ventilation calculation (38) — registration algorithms need to be validated for such applications. (41)
Validation of the ΔV‐4D CT ventilation calculation method is important because 4D CT‐based techniques are finding medical applications given that these scans are faster and do not require injection of radioactive materials or contrast agents. For example, 4D CT‐based ventilation has been used in radiation therapy treatment planning to avoid functional lung and to estimate loss of lung function or normal tissue toxicity after radiotherapy in a dose‐dependent manner. 40 , 42 , 43 , 44 , 45 Additional medical applications, such as in diagnostic imaging, may be possible.
V. CONCLUSIONS
Ventilation distributions calculated using 4D CT and deformable image registration were compared to the ventilation distribution derived from Xe‐enhanced CT data, which was considered the gold standard, for four sheep. The 4D CT‐based ventilation distributions agreed with the ones from XeCT reasonably well. Ventilation differences introduced by registration errors can be reduced by smoothing. This evaluation study and previous reports 15 , 18 support the use of ventilation calculated using 4D CT in clinical applications such as radiation treatment planning, 12 , 23 , 24 , 40 , 42 , 43 , 44 , 45 pulmonary function, 15 , 18 and lung disease. (22)
COPYRIGHT
This work is licensed under a Creative Commons Attribution 4.0 International License.
REFERENCES
- 1. Nakagawa T, Sakuma H, Murashima S, Ishida N, Matsumura K, Takeda K. Pulmonary ventilation‐perfusion MR imaging in clinical patients. J Magn Reson Imaging. 2001;14(4):419–24. [DOI] [PubMed] [Google Scholar]
- 2. Bunow B, Line BA, Horton MR, Weiss GH. Regional ventilatory clearance by xenon scintigraphy: a critical evaluation of two estimation procedures. J Nucl Med. 1979;20(7):703–10. [PubMed] [Google Scholar]
- 3. Gottschalk A, Sostman HD, Coleman RE, et al. Ventilation‐perfusion scintigraphy in the PIOPED study. Part II. Evaluation of the scintigraphic criteria and interpretations. J Nucl Med. 1993;34(7):1119–26. [PubMed] [Google Scholar]
- 4. Harris B, Bailey D, Miles S, et al. Objective analysis of tomographic ventilation‐perfusion scintigraphy in pulmonary embolism. Am J Resp Crit Care Med. 2007;175(11):1173–80. [DOI] [PubMed] [Google Scholar]
- 5. Petersson J, Sánchez‐Crespo A, Rohdin M, et al. Physiological evaluation of a new quantitative SPECT method measuring regional ventilation and perfusion. J Appl Physiol. 2004;96(3):1127–36. [DOI] [PubMed] [Google Scholar]
- 6. Vidal Melo FV, Layfield D, Harris RS, et al. Quantification of regional ventilation‐perfusion ratios with PET. J Nucl Med. 2003;44(12):1982–91. [PubMed] [Google Scholar]
- 7. Willey‐Courand DB, Harris RS, Galletti GG, Hales CA, Fischman A, Venegas JG. Alterations in regional ventilation, perfusion, and shunt after smoke inhalation measured by PET. J Appl Physiol. 2002;93(3):1115–22. [DOI] [PubMed] [Google Scholar]
- 8. Levin DL, Chen Q, Zhang M, Edelman RR, Hatabu H. Evaluation of regional pulmonary perfusion using ultrafast magnetic resonance imaging. Magn Reson Med. 2001;46(1):166–71. [DOI] [PubMed] [Google Scholar]
- 9. Suga K, Yuan Y, Ogasawara N, Tsukuda T, Matsunaga N. Altered clearance of gadolinium diethylenetriaminepentaacetic acid aerosol from bleomycin‐injured dog lungs. Am J Respir Crit Care Med. 2003;167(12):1704–10. [DOI] [PubMed] [Google Scholar]
- 10. Marcucci C, Nyhan D, Simon BA. Distribution of pulmonary ventilation using Xe‐enhanced computed tomography in prone and supine dogs. J Appl Physiol. 2001;90(2):421–30. [DOI] [PubMed] [Google Scholar]
- 11. Simon BA. Non‐invasive imaging of regional lung function using X‐ray computed tomography. J Clin Monit Comput. 2000;16(5‐6):433–42. [DOI] [PubMed] [Google Scholar]
- 12. Guerrero T, Sanders K, Noyola‐Martinez J, et al. Quantification of regional ventilation from treatment planning CT. Int J Radiat Oncol Biol Phys. 2005;62(3):630–34. [DOI] [PubMed] [Google Scholar]
- 13. Reinhardt JM, Ding K, Cao K, Christensen GE, Hoffman EA, Bodas SV. Registration‐based estimates of local lung tissue expansion compared to xenon CT measures of specific ventilation. Med Image Anal. 2008;12(6):752–63. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Zhang GG, Huang TC, Dilling T, Stevens C, Forster KM. Derivation of high‐resolution pulmonary ventilation using local volume change in four‐dimensional CT data. In: Dössel O, Schlegel WC, editors. World Congress on Medical Physics and Biomedical Engineering. Vol 25 Munich, Germany: Springer; 2009. [Google Scholar]
- 15. Ding K, Cao K, Fuld MK, et al. Comparison of image registration based measures of regional lung ventilation from dynamic spiral CT with Xe‐CT. Med Phys. 2012;39(8):5084–98. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Chon D, Simon BA, Beck KC, et al. Differences in regional wash‐in and wash‐out time constants for xenon‐CT ventilation studies. Respir Physiol Neurobiol. 2005;148(1‐2):65–83. [DOI] [PubMed] [Google Scholar]
- 17. Kipritidis J, Siva S, Hofman MS, Callahan J, Hicks RJ, Keall PJ. Validating and improving CT ventilation imaging by correlating with ventilation 4D‐PET/CT using 68Ga‐labeled nanoparticles. Med Phys. 2014;41(1):011910. [DOI] [PubMed] [Google Scholar]
- 18. Yamamoto T, Kabus S, Lorenz C, et al. Pulmonary ventilation imaging based on 4‐dimensional computed tomography: comparison with pulmonary function tests and SPECT ventilation images. Int J Radiat Oncol Biol Phys. 2014;90(2):414–22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Du K, Bayouth JE, Cao K, Christensen GE, Ding K, Reinhardt JM. Reproducibility of registration‐based measures of lung tissue expansion. Med Phys. 2012;39(3):1595–608. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Du K, Bayouth JE, Ding K, Christensen GE, Cao K, Reinhardt JM. Reproducibility of intensity‐based estimates of lung ventilation. Med Phys. 2013;40(6):063504. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Du K, Reinhardt JM, Christensen GE, Ding K, Bayouth JE. Respiratory effort correction strategies to improve the reproducibility of lung expansion measurements. Med Phys. 2013;40(12):123504. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Castillo R, Castillo E, McCurdy M, et al. Spatial correspondence of 4D CT ventilation and SPECT pulmonary perfusion defects in patients with malignant airway stenosis. Phys Med Biol. 2012;57(7):1855–71. [DOI] [PubMed] [Google Scholar]
- 23. Huang T‐C, Hsiao C‐Y, Chien C‐R, Liang J‐A, Shih T‐C, Zhang G. IMRT treatment plans and functional planning with functional lung imaging from 4D‐CT for thoracic cancer patients. Radiat Oncol. 2013;8(1):3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Ding K, Bayouth JE, Buatti JM, Christensen GE, Reinhardt JM. 4DCT‐based measurement of changes in pulmonary function following a course of radiation therapy. Med Phys. 2010;37(3):1261–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Lavrenkov K, Christian JA, Partridge M, et al. A potential to reduce pulmonary toxicity: the use of perfusion SPECT with IMRT for functional lung avoidance in radiotherapy of non‐small cell lung cancer. Radiother Oncol. 2007;83(2):156–62. [DOI] [PubMed] [Google Scholar]
- 26. Shioyama Y, Jang SY, Liu HH, et al. Preserving functional lung using perfusion imaging and intensity‐modulated radiation therapy for advanced‐stage non‐small cell lung cancer. Int J Radiat Oncol Biol Phys. 2007;68(5):1349–58. [DOI] [PubMed] [Google Scholar]
- 27. Janssens G, Jacques L, Orban de Xivry J, Geets X, Macq B. Diffeomorphic registration of images with variable contrast enhancement. International Journal of Biomedical Imaging. 2011;2011:891585. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Wrangsjo A, Pettersson J, Knutsson H. Non‐rigid registration using morphons. Image Analysis, Proceedings. 2005;3540:501–10. [Google Scholar]
- 29. Vercauteren T, Pennec X, Perchant A, Ayache N. Diffeomorphic demons: Efficient non‐parametric image registration. Neuroimage. 2009;45(1, Suppl 1):S61–S72. [DOI] [PubMed] [Google Scholar]
- 30. Latifi K, Zhang G, Stawicki M, van Elmpt W, Dekker A, Forster K. Validation of three deformable image registration algorithms for the thorax. J Appl Clin Med Phys. 2013;14(1):19–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Zhang G, Huang T‐C, Dilling T, Stevens C, Forster K. Comments on ‘Ventilation from four‐dimensional computed tomography: density versus Jacobian methods’. Phys Med Biol. 2011;56(11):3445–46. [DOI] [PubMed] [Google Scholar]
- 32. Castillo R, Castillo E, Martinez J, Guerrero T. Ventilation from four‐dimensional computed tomography: density versus Jacobian methods. Phys Med Biol. 2010;55(16):4661–85. [DOI] [PubMed] [Google Scholar]
- 33. Reinhardt J, Christensen G, Hoffman E, Ding K, Cao K. Registration‐derived estimates of local lung expansion as surrogates for regional ventilation. In: Karssemeijer N, Lelieveldt B, editors. Information Process Med Imaging. Vol 4584 Berlin Heidelberg: Springer; 2007. [DOI] [PubMed] [Google Scholar]
- 34. Lee J‐S. Digital image smoothing and the sigma filter. Comput Vis Graph Image Process. 1983;24(2):255–69. [Google Scholar]
- 35. Mastin GA. Adaptive filters for digital image noise smoothing: an evaluation. Comput Vis Graph Image Process. 1985;31(1):103–21. [Google Scholar]
- 36. Dice LR. Measures of the amount of ecologic association between species. Ecology. 1945;26(3):297–302. [Google Scholar]
- 37. Yamamoto T, Kabus S, Klinder T, et al. Four‐dimensional computed tomography pulmonary ventilation images vary with deformable image registration algorithms and metrics. Med Phys. 2011;38(3):1348–58. [DOI] [PubMed] [Google Scholar]
- 38. Latifi K, Huang T‐C, Feygelman V, et al. Effects of quantum noise in 4D‐CT on deformable image registration and derived ventilation data. Phys Med Biol. 2013;58(21):7661–72. [DOI] [PubMed] [Google Scholar]
- 39. Goerres G, Kamel E, Heidelberg T‐N, Schwitter M, Burger C, von Schulthess G. PET‐CT image co‐registration in the thorax: influence of respiration. Eur J Nucl Med Mol Imaging. 2002;29(3):351–60. [DOI] [PubMed] [Google Scholar]
- 40. Latifi K, Dilling T, Feygelman V, et al. Impact of dose on lung ventilation change calculated from 4D‐CT using deformable image registration in lung cancer patients treated with SBRT. J Radiat Oncol. 2015;4(3):265–70. [Google Scholar]
- 41. Latifi K, Forster KM, Hoffe SE, et al. Dependence of ventilation image derived from 4D CT on deformable image registration and ventilation algorithms. J Appl Clin Med Phys. 2013;14(4):150–62. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Kadoya N, Cho SY, Kanai T, et al. Dosimetric impact of 4‐dimensional computed tomography ventilation imaging‐based functional treatment planning for stereotactic body radiation therapy with 3‐dimensional conformal radiation therapy. Pract Radiat Oncol. 2015;5(5):e505–e512. [DOI] [PubMed] [Google Scholar]
- 43. Kimura T, Doi Y, Nakashima T, et al. Combined ventilation and perfusion imaging correlates with the dosimetric parameters of radiation pneumonitis in radiation therapy planning for lung cancer. Int J Radiat Oncol Biol Phys. 2015;93(4):778–87. [DOI] [PubMed] [Google Scholar]
- 44. King MT, Maxim PG, Diehn M, Loo BW Jr., Xing L. Analysis of long‐term 4‐dimensional computed tomography regional ventilation after radiation therapy. Int J Radiat Oncol Biol Phys. 2015;92(3):683–90. [DOI] [PubMed] [Google Scholar]
- 45. Kipritidis J, Hugo G, Weiss E, Williamson J, Keall PJ. Measuring interfraction and intrafraction lung function changes during radiation therapy using four‐dimensional cone beam CT ventilation imaging. Med Phys. 2015;42(3):1255–67. [DOI] [PMC free article] [PubMed] [Google Scholar]