Abstract
Malignant pleural mesothelioma (MPM) is an extremely difficult disease to treat. This pilot study investigates the feasibility of using volumetric‐modulated arc therapy (VMAT) for malignant pleural mesothelioma (MPM), and compares VMAT to static field intensity‐modulated radiotherapy (IMRT) for five patients. To identify the best treatment technique for MPM, in five patients, we made a representative comparative analysis of two kinds of techniques for radiation therapy planning: IMRT and VMAT. The plans were created for an Elekta Synergy linear accelerator with 6 MV photons using Oncentra version 4.3 treatment planning system. Dose prescription was 50 Gy to the average of the planning target volume (PTV). PTV coverage and homogeneity, dose of organs at risk, numbers of segments, MUs, and delivery time were evaluated for all techniques. VMAT allowed better homogeneous and conformity indices compared with IMRT ( vs. 0.12, vs. 0.77, respectively, ). VMAT plan had a significantly shorter delivery time (326 s) compared with in IMRT plans (510 s), (). In the dose verification, an average of 93.16% of the detector points passed the 3%/3 mmγ criterion for VMAT plans, while in IMRT plans the dose verification was 95.12%.().
PACS number(s): 87.55.D, 87.55.km, 87.56.Fc
Keywords: pleural mesothelioma, intensity‐modulated radiotherapy/volumetric‐modulated arc therapy, plan compare; dose verification, Delta4 detector array
I. INTRODUCTION
Malignant pleural mesothelioma (MPM) is a deadly disease to treat worldwide, with the median overall survival ranging between 9 and 17 months, regardless of stage. (1) In the past, surgery and chemotherapy were more commonly used for treatment of MPM than radiotherapy, due to the limitations of technique. A number of studies have investigated whether more advanced radiotherapy techniques, such as IMRT, would lead to better local control and lower doses to organs at risk (OARs) than standard RT. VMAT is a next generation of IMRT technique that can decrease treatment delivery times with similar or better plan quality for different treatment sites. 2 , 3 , 4 , 5
The aim of this study was to compare the two kinds of radiotherapy techniques and analyze the feasibility of using VMAT for MPM.
II. MATERIALS AND METHODS
All patients underwent computed tomography (CT) simulation before radiation therapy. Simulation took place while the patients were supine and immobilized in an upper‐body cradle, with both arms overhead. The hemithorax was contoured as the clinical target volume, which included the pleural space, scars, drains sites, and involved nodal stations. This volume was then expanded to include a margin for internal motion, and an additional 0.5 to 1.0 cm planning target volume (PTV) margin was added. All PTVs and OARs were delineated by an experienced clinician. Both VMAT and IMRT plans were generated for a treatment in 25 fractions, to deliver a total dose of 50 Gy to the PTV. Both the treatment plans were optimized to cover at least 95% of the volume of PTV, and the maximal allowed point dose for spinal cord was 45 Gy, the mean dose allowed for heart was 30 Gy. All the treatment planning objectives of the plans were listed as in Table 1. VMAT and IMRT plans were generated using Oncentra version 4.3 treatment planning system (Elekta AB, Stockholm, Sweden) and commissioned for Elekta's Synergy linear accelerator with 6 MV photons. Step‐and‐shoot IMRT plans were generated using typically seven coplanar beams with a total of 50 segments. The lower limit for the segment size was . VMAT plans were generated using dual arcs with an arc length close to the range of IMRT plan's beams. The control points within the arc were set to 4° and collimator angle was set to 0°. All IMRT and VMAT plans were generated using 6 MV photons. The conformity index ) and homogeneity index (HI, ) were also compared (6) where DT indicates total dose of the target. We used Delta4 Discover detector array (ScandiDos AB, Uppsala, Sweden) to compare the dose verification, and all the data from these patients was analyzed in SPSS version 15.0 (IBM Corp., Armonk, NY).
Table 1.
Target | Dose—Volume Constraints | Weight | |
---|---|---|---|
PTV | Min dose 50 Gy, to 98% volume | 100 | |
Max dose 53 Gy | 100 | ||
Uniform dose 50.5 Gy | 100 | ||
Normal Tissue | Dose—Volume Constraints | Weight | |
Spinal cord | No portion may receive 45 Gy | 75 | |
Contralateral lung |
|
10 | |
Whole lung |
|
15 | |
|
13 | ||
|
10 | ||
Heart |
|
10 | |
Ring | Max dose | 10 |
III. RESULTS
Generally, both techniques can achieve satisfying clinical plans. VMAT plans offered tighter isodoses surfaces and encompassed the PTV with similar or better sparing of OARs. The dose distributions of IMRT and VMAT plans are shown in Fig. 1. The high dose of VMAT plans was less than IMRT plans, while the low dose was nearly the same. In addition, with VMAT plan the isodose surfaces encompassed the PTVs more smoothly and fewer hot spots outside the PTVs were observed. This means VMAT plan succeeded in producing a better dose of PTV without increasing the dose to OARs. The differences in dose and volume histogram between them are shown in Fig. 2: both and of PTV in VMAT plan decreased compared to IMRT plan, while the , and in VMAT plan increased compared to IMRT plan; (i.e., VMAT plans improved the conformity index and homogeneity index significantly, as shown in Table 2).
Table 2.
|
|
|
|
|
||||||
---|---|---|---|---|---|---|---|---|---|---|
IMRT |
|
|
|
|
|
|||||
VMAT |
|
|
|
|
|
|||||
t value | 3.47 | 4.69 | 1.25 | 1.81 | 0.92 | |||||
p value |
|
|
|
|
|
|||||
(Gy) | (Gy) | (Gy) | HI | CI | ||||||
IMRT |
|
|
|
|
|
|||||
VMAT |
|
|
|
|
|
|||||
t value | 2.93 | 4.36 | 2.67 | 5.32 | 8.40 | |||||
p value |
|
|
|
|
|
Comparison of the plan parameters for VMAT and IMRT plans generated showed significantly better sparing for most OARs (see Table 3). The maximum point dose and mean dose to spinal cord were higher with VMAT than that with IMRT. With VMAT we observed a significantly lower (the percentage of volume receiving more than 10 Gy and 20 Gy), and average mean dose for heart and the lung. As for the of the lung, IMRT plans achieved an extremely lower dose.
Table 3.
Cord
|
Heart
|
Heart
|
Heart
|
Heart
|
||||||
---|---|---|---|---|---|---|---|---|---|---|
IMRT |
|
|
|
|
|
|||||
VMAT |
|
|
|
|
|
|||||
t value | 3.27 | 4.90 | 10.80 | 0.08 | 4.54 | |||||
p value |
|
|
|
|
|
|||||
Whole lung
|
Whole Lung
|
Whole Lung
|
Whole Lung
|
Whole Lung
|
||||||
IMRT |
|
|
|
|
|
|||||
VMAT |
|
|
|
|
|
|||||
t value | 3.52 | 7.88 | 4.33 | 0.17 | 3.42 | |||||
p value |
|
|
|
|
|
|||||
Contralateral Lung
|
Contralateral Lung
|
Contralateral Lung
|
Contralateral Lung
|
Contralateral Lung
|
||||||
IMRT |
|
|
|
‐ |
|
|||||
VMAT |
|
|
|
‐ |
|
|||||
t value | 4.26 | 6.52 | 3.21 | ‐ | 3.13 | |||||
p value |
|
|
|
‐ |
|
Although there were more control points of VMAT plan than IMRT plan, the numbers of MUs and delivery time were decreased significantly. Dose verifications were similar between VMAT and IMRT plans (see Table 4).
Table 4.
Total Segments | Total MUs | Delivery Time (s) | 3 mm/3% γ | |||||
---|---|---|---|---|---|---|---|---|
IMRT |
|
|
|
|
||||
VMAT |
|
|
|
|
||||
t value | 6.81 | 4.76 | 6.42 | 2.99 | ||||
p value |
|
|
|
|
IV. DISCUSSION
MPM is one of the most challenging tumor entities in oncology that requires a multidisciplinary approach. 7 , 8 , 9 , 10 The large target volume that comprises the complete operated hemithorax with its complex shape and the proximity of many sensitive risk structures forces the radiation oncologist to use very sophisticated methods. Modern conformal radiation technologies, such as IMRT, open new possibilities in the treatment of complex‐shaped targets like MPM. However, Allen et al. (11) point out how essential the reduction of mean lung dose in the radiotherapy of MPM is with severe pneumonitis seen after a mean dose of 15 Gy. Sterzing et al. (12) compared tomotherapy and IMRT plans, and reported that, regarding the PTV, the biggest differences in plan comparison were seen in target coverage homogeneity. Tomotherapy was capable of delivering dose in excellent homogeneity. In their data, the maximum dose especially could be lowered by approximately 8 Gy on average when using tomotherapy.
Other studies have shown that the VMAT technique could achieve better plans. 13 , 14 , 15 , 16 Analysis of the data resulting from this study of VMAT for MPM shows that VMAT plans have an improved plan quality compared to IMRT plans.
IMRT can provide both dosimetric superiority and good clinical outcomes when appropriate dose constraints are used. Other studies from M.D. Anderson Cancer Center and other institutions have shown low rates of high‐grade pneumonitis and median survival times of 16 months or more with the use of IMRT. The large volume to irradiate requires a long irradiation time with IMRT techniques. Long delivery time results in patient motion during the daily treatment, with the consequence of dose delivery different from the plan that was designed in the TPS. In addition, IMRT plans need high MUs to achieve a good dose sparing, but increasing MUs means increasing probability of secondary tumors due to radiation leakage. The decreased treatment delivery time obtained with VMAT will improve patient comfort and result in a smaller impact of intrafraction movements.
It is clear that VMAT, the newest technique, is well positioned with respect to the alternative approaches from IMRT and could offer significant improvement from the logistic viewpoint. In our study, VMAT extremely reduced the delivery time and MUs, while maintaining adequate target coverage and dose sparing to the OARs. VMAT and IMRT showed both good index of PTV coverage and homogeneity. VMAT is better than IMRT.
We use the Delta4 detector array for dose verification. It is very important to determine the error between the calculations and effective delivery. The results show that both VMAT and IMRT plans could used for treatment with the error within the clinical limit.
V. CONCLUSIONS
VMAT is another promising radiotherapy option for MPM. It allows reducing dose to most OARs without compromising target coverage, meanwhile keeping a shortest treatment time.
COPYRIGHT
This work is licensed under a Creative Commons Attribution 4.0 International License.
Supporting information
REFERENCES
- 1. Tsao AS, Wistuba I, Roth JA, Kindler HL. Malignant pleural mesothelioma. J Clin Oncol. 2009;27(12):2081–90. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Bedford JL. Treatment planning for volumetric modulated arc therapy. Med Phys. 2009;36(11):5128–38. [DOI] [PubMed] [Google Scholar]
- 3. Verbakel WF, Cuijpers JP, Hoffmans D, Bieker M, Slotman BJ, Senan S. Volumetric intensity‐modulated arc therapy vs. conventional IMRT in head‐and‐neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys. 2009;74(1):252–59. [DOI] [PubMed] [Google Scholar]
- 4. Rao M, Yang W, Chen F, et al. Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: plan quality, delivery efficiency and accuracy. Med Phys. 2010;37(3):1350–59. [DOI] [PubMed] [Google Scholar]
- 5. Lu SH, Cheng JC, Kuo SH, et al. Volumetric modulated arc therapy for nasopharyngeal carcinoma: a dosimetric comparison with TomoTherapy and step‐and‐shoot IMRT. Radiother Oncol. 2012;104(3):324–30. [DOI] [PubMed] [Google Scholar]
- 6. van't Riet A, Mak AC, Moerland MA, Elders LH, van der Zee W. A conformation number to quantify the degree of conformality in brachytherapy and external beam irradiation: application to the prostate. Int J Radiat Oncol Biol Phys. 1997;37(3):731–36. [DOI] [PubMed] [Google Scholar]
- 7. Neragi‐Miandoab S. Multimodality approach in management of malignant pleural mesothelioma. Eur J Cardio‐Thorac. 2006;29(1):14–19. [DOI] [PubMed] [Google Scholar]
- 8. Sterman DH and Albelda SM. Advances in the diagnosis, evaluation, and management of malignant pleural mesothelioma. Respirology. 2005;10(3):266–83. [DOI] [PubMed] [Google Scholar]
- 9. Ung YC, Yu E, Falkson C, et al. Therole of radiation therapy in malignant pleural mesothelioma: a systematic review. Radiother Oncol. 2006;80(1):13–18. [DOI] [PubMed] [Google Scholar]
- 10. Yajnik S, Rosenzweig KE, Mychalczak B, et al. Hemithoracic radiation after extrapleural pneumonectomy for malignant pleural mesothelioma. Int J Radiat Oncol Biol Phys. 2003;56(1):1319–26. [DOI] [PubMed] [Google Scholar]
- 11. Allen AM, Czerminska M, Jänne PA, et al. Fatal pneumonitis associated with intensity‐modulated radiation therapy for mesothelioma. Int J Radiat Oncol Biol Phys. 2006;65(3):640–65. [DOI] [PubMed] [Google Scholar]
- 12. Sterzing F, Sroka‐Perez G, Schubert K, et al. Evaluating target coverage and normal tissue sparing in the adjuvant radiotherapy of malignant pleural mesothelioma: helical tomotherapy compared with step‐and‐shoot IMRT. Radiother Oncol. 2008;86(2):251–57. [DOI] [PubMed] [Google Scholar]
- 13. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35(1):310–17. [DOI] [PubMed] [Google Scholar]
- 14. Yoo S, Wu QJ, Lee WR, Yin FF. Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vesicles and lymph nodes. Int J Radiat Oncol Biol Phys. 2010;76(3):935–42. [DOI] [PubMed] [Google Scholar]
- 15. Yu CX. Intensity‐modulated arc therapy with dynamic multileaf collimation: alternative to tomotherapy. Phys Med Biol. 1995;40(9):1435–49. [DOI] [PubMed] [Google Scholar]
- 16. Guckenberger M, Richter A, Krieger T, Wilbert J, Baier K, Flentje M. Is a single arc sufficient in volumetric‐modulated arc therapy (VMAT) for complex‐shaped target volumes? Radiother Oncol. 2009;93(2):259–65. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.