“Evaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: A preliminary study” [JACMP, 17(4), 2016]

T Shiinoki, S Kawamura, T Uehara, Y Yuasa, K Fujimoto, M Koike, T Sera, Y Emoto, H Hanazawa, K Shibuya
PMCID: PMC5874949  PMID: 28681447

After the publication of the manuscript titled “Evaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: a preliminary study”1, an error was discovered in measured delay time (Δtgate‐on) and Table 1 (Δtbeam‐on).

Table 1.

Comparison of delay times of various respiratory gating systems

Author modality ▵tbeam‐on (ms) ▵tbeam‐off (ms)
Chang et al.(19) Novalis Tx + ExacTrac 200 ± 30
Novalis Tx + RPM 90 ± 10
Smith et al.(20) Trilogy + Calypso 75.0 ± 12.7 65.1 ± 12.9
Woods et al.(21) TrueBeam + RPM 139 ± 10
AAPM TG142(22) 100 100
This study TrueBeam + SyncTraX
6 MV‐FF 109.5 ± 11.8 71.5 ± 12.1
6 MV‐FFF 115.9 ± 12.1 62.7 ± 14.1
10 MV‐FF 103.5 ± 8.6 68.1 ± 7.2
10 MV‐FFF 112.1 ± 15.3 69.0 ± 10.7

RPM, real‐time positioning management; FF, flattened; FFF = flattening‐filter‐free.

The incorrected data have been removed from the results. Additionally, correct data have been added to generate a new Table 1 (shown below). The specific values for Δtgate‐on and Δtbeam‐on have got smaller, but all of the original findings and conclusions remain valid.

The new corrected table 1 values have resulted in the following changes (bold italics) to the text of the manuscript.

Abstract

The total time delay between TrueBeam and SyncTraX was <200 ms for each photon beam.

Results

B. Evaluation of time delay using oscilloscope

The mean ± SD values of Δtgate‐on were 64.2 ± 11.6 ms, 70.3 ± 12.2 ms, 58.3 ± 8.6 ms, and 66.4 ± 15.4 ms in the case of 6 MV‐FF, 6 MV‐FFF, 10 MV‐FF, and 10 MV‐FFF photon beams, respectively.

For each photon beam, Δtbeam‐on + Δtbeam‐off <200 ms.

Discussion

In this study, the time delay was measured using an oscilloscope, and the total time delay was <200 ms for each photon beam.

Reference

  • 1. Shiinoki T, Kawamura S, Uehara T, et al. Evaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: a preliminary study. J Appl Clin Med Phys. 2016;17(4):202–213. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Applied Clinical Medical Physics are provided here courtesy of Wiley