After the publication of the manuscript titled “Evaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: a preliminary study”1, an error was discovered in measured delay time (Δtgate‐on) and Table 1 (Δtbeam‐on).
Table 1.
Comparison of delay times of various respiratory gating systems
| Author | modality | ▵tbeam‐on (ms) | ▵tbeam‐off (ms) |
|---|---|---|---|
| Chang et al.(19) | Novalis Tx + ExacTrac | 200 ± 30 | |
| Novalis Tx + RPM | 90 ± 10 | ||
| Smith et al.(20) | Trilogy + Calypso | 75.0 ± 12.7 | 65.1 ± 12.9 |
| Woods et al.(21) | TrueBeam + RPM | 139 ± 10 | |
| AAPM TG142(22) | 100 | 100 | |
| This study | TrueBeam + SyncTraX | ||
| 6 MV‐FF | 109.5 ± 11.8 | 71.5 ± 12.1 | |
| 6 MV‐FFF | 115.9 ± 12.1 | 62.7 ± 14.1 | |
| 10 MV‐FF | 103.5 ± 8.6 | 68.1 ± 7.2 | |
| 10 MV‐FFF | 112.1 ± 15.3 | 69.0 ± 10.7 |
RPM, real‐time positioning management; FF, flattened; FFF = flattening‐filter‐free.
The incorrected data have been removed from the results. Additionally, correct data have been added to generate a new Table 1 (shown below). The specific values for Δtgate‐on and Δtbeam‐on have got smaller, but all of the original findings and conclusions remain valid.
The new corrected table 1 values have resulted in the following changes (bold italics) to the text of the manuscript.
Abstract
The total time delay between TrueBeam and SyncTraX was <200 ms for each photon beam.
Results
B. Evaluation of time delay using oscilloscope
The mean ± SD values of Δtgate‐on were 64.2 ± 11.6 ms, 70.3 ± 12.2 ms, 58.3 ± 8.6 ms, and 66.4 ± 15.4 ms in the case of 6 MV‐FF, 6 MV‐FFF, 10 MV‐FF, and 10 MV‐FFF photon beams, respectively.
For each photon beam, Δtbeam‐on + Δtbeam‐off <200 ms.
Discussion
In this study, the time delay was measured using an oscilloscope, and the total time delay was <200 ms for each photon beam.
Reference
- 1. Shiinoki T, Kawamura S, Uehara T, et al. Evaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: a preliminary study. J Appl Clin Med Phys. 2016;17(4):202–213. [DOI] [PMC free article] [PubMed] [Google Scholar]