Comparison of TG‐43 dosimetric parameters of brachytherapy sources obtained by three different versions of MCNP codes

Neda Zaker 1, Mehdi Zehtabian 1,, Sedigheh Sina 2, Craig Koontz 3, Ali S Meigooni1 4
PMCID: PMC5874963  PMID: 27074460

Abstract

Monte Carlo simulations are widely used for calculation of the dosimetric parameters of brachytherapy sources. MCNP4C2, MCNP5, MCNPX, EGS4, EGSnrc, PTRAN, and GEANT4 are among the most commonly used codes in this field. Each of these codes utilizes a cross‐sectional library for the purpose of simulating different elements and materials with complex chemical compositions. The accuracies of the final outcomes of these simulations are very sensitive to the accuracies of the cross‐sectional libraries. Several investigators have shown that inaccuracies of some of the cross section files have led to errors in 125I and 103Pd parameters. The purpose of this study is to compare the dosimetric parameters of sample brachytherapy sources, calculated with three different versions of the MCNP code — MCNP4C, MCNP5, and MCNPX. In these simulations for each source type, the source and phantom geometries, as well as the number of the photons, were kept identical, thus eliminating the possible uncertainties. The results of these investigations indicate that for low‐energy sources such as 125I and 103Pd there are discrepancies in gL(r) values. Discrepancies up to 21.7% and 28% are observed between MCNP4C and other codes at a distance of 6 cm for 103Pd and 10 cm for 125I from the source, respectively. However, for higher energy sources, the discrepancies in gL(r) values are less than 1.1% for 192Ir and less than 1.2% for 137Cs between the three codes.

PACS number(s): 87.56.bg

Keywords: brachytherapy, MCNP, dosimetry, TG‐43

I. INTRODUCTION

Brachytherapy has been a cornerstone treatment technique for the management of various malignancies such as gynecological, prostate, and breast tumors. In low‐dose‐rate (LDR) brachytherapy treatments, 125I,125Pd,125Ir, and 137Cs sources have been utilized for many years. Task Group 43 (TG‐43) of the American Associations of Physicists in Medicine (AAPM) 1 , 2 has recommended the dosimetric evaluation of low energy brachytherapy sources using experimental technique and/or Monte Carlo simulations. 3 , 4 , 5 , 6 , 7 , 8 , 9 , 10 , 11 , 12 , 13 , 14 , 15 A subsequent Working Group from AAPM has recommended the dosimetric evaluation of high‐energy sources (16) using a similar method. Several different Monte Carlo codes have been utilized for the dosimetric evaluation of brachytherapy sources including MCNP4C2, 17 , 18 MCNP5, (19) MCNPX, (20) EGS4, (21) EGSnrc, (22) PTRAN, (5) and GEANT4. (23) Each of these codes uses a different cross‐sectional library for both photons and electrons for all the chemical elements within the energy range of 1 eV to 1 GeV. The accuracies of the simulated data have been demonstrated by comparison with the experimental data. 9 , 11 However, the large statistical uncertainties of the experimental data masked the true uncertainties of the Monte Carlo simulations due to the cross‐sectional errors or selection of insufficient number of photons in the simulations. In 2002, DeMarco et al. (7) performed an extensive study regarding the accuracies of cross‐sectional data utilized in different Monte Carlo simulation codes and their impacts on the absorbed doses and collisional kerma values. They reported up to 10% differences in the photoelectric cross sections for water at 30 keV between the DLC‐200 from RSICC at Oak Ridge National Laboratory and XCOM cross‐sectional database maintained at the National Institute of Standards and Technology. 24 , 25 Furthermore, they demonstrated that differences of up to 10% are observed in the photoelectric cross section for water at 30 keV between the standard MCNP cross‐sectional dataset (DLC‐200) and the most recent XCOM/NIST tabulation. More specifically, they had shown that at energies of 20 keV and 30 keV, the absolute dose rates in water at 1.0 cm depth increased by 3.5% and 7.1%, respectively, when the cross‐sectional tabulated data were changed from DLC‐200 to XCOM/NIST data. However, these changes were found to be insignificant when using DLC‐146. (26)

The exact cause of the differences between various Monte Carlo reported data generated by different investigators for the same source type was left unresolved. For example, the dose rate constant of 125I, Model 6733 was reported to be 0.965cGyh1U1 by Mosleh‐Shirazi et al. (19) using the MCNP5 code, while it is has been reported to be 0.993 cGyh1U1 by the same authors using the MCNP4C2 code and 0.970 cGyh1U1 by Sowards and Meigooni (27) using the PTRAN code. There is a need to understand whether the differences between these values are due to the errors in the cross‐sectional data which has been used or some other cause. These causes could include the differences in the geometrical designs of the sources and phantoms used in each project, the differences in number of histories used for simulations, or size of the tally cells. This project is designed to resolve these issues.

The aim of this project is to verify the differences among three versions of the MCNP Monte Carlo codes (MCNP4C2, MCNP5, and MCNPX), which are widely utilized for dosimetry of brachytherapy sources. The TG‐43 parameters of four different brachytherapy sources — 192Ir (model Vari‐Source VS2000), 125I (model IAI‐125A), 103Pd (model Best 2335) and 137Cs (Selectron pellet sources from Nucletron) — have been determined using these codes. In the simulations for each source type, the phantom and source geometries, the number of particle histories/photons, the size of the tally cells, and dose points were kept identical for all the Monte Carlo codes. The cross‐sectional files used in the MCNP5, MCNP4C2, and MCNPX are ENDF/B‐VI release 8, (28) ENDF/B‐VI version 5, (29) and ENDF/B‐VI release 8, (28) respectively.

II. MATERIALS AND METHODS

A. Source characteristics

Four different commercially available brachytherapy sources were selected for this project. These sources are the Varian VariSource 192Ir, Model VS2000 (Varian, Varian Medical Systems, Palo Alto, CA), IsoAid Advantage 125I Model IAI‐125A (IsoAid LLC, Port Richey, FL), the Best Industries 103Pd Model Best 2335 (Best International, Best Medical International, Inc., Springfield, VA), and Nucletron 137Cs Selectron pellet sources (Nucletron BV, Veenedaal, The Netherlands). These sources were selected as representative of each source type and the choices do not reflect endorsement of the vendors. The physical characteristics of these sources are briefly described below.

A.1 125I

The IsoAid Advantage 125I (model IAI‐125A) source was introduced in the North American market in 2002. (9) Dimensions of this source for the Monte Carlo simulations are taken from the study by Meigooni et al. (9) The IsoAid Advantage seed contains a 3 mm long silver rod X‐ray marker with a diameter of 0.50 mm. The silver rod is coated with a 1 μm thick layer of AgI containing 125I. The active silver rod is encapsulated in 0.05 mm thick titanium (Ti) capsule with an outside diameter of 0.80 mm. The laser‐welded end‐caps have a maximum thickness of 0.10 mm. Two hemispheres are used for modeling the end‐caps, one with 0.40 mm radius Ti hemisphere overlapped with a 0.35 mm radius air. The center of the air hemisphere was shifted by 0.05 mm relative to the Ti hemisphere. The overall source length is 4.50 mm and the active length is 3.0 mm. The cylindrical source element is free to move approximately 0.150 mm along the longitudinal direction of the source and 0.100 mm in the radial direction. The schematic diagram of the simulated source is shown in Fig. 1(a).

Figure 1.

Figure 1

The Schematic diagram of the simulated sources: (a) the Best Industries 103Pd Model Best 2335, (b) IsoAid Advantage 125I Model IAI‐125A, (c) the Varian VariSource 192Ir, and (d) Nucletron 137Cs Selectron pellet sources.

A.2 103Pd

The 103Pd source (Model Best 2335) was obtained from Best Industries. The dimensions of the source were taken from previous investigations. (11) The Best 2335 source consists of a cylindrical tungsten X‐ray marker with a diameter of 0.5 mm and a length of 1.2 mm. There are three polymer spheres with 0.5 mm diameter located on either side of the X‐ray marker. The centers of these pellets are located at distances of 0.9 mm, 1.5 mm, and 2.1 mm relative to the center of the seed. The relative chemical compositions of these spheres by weight are: 89.73% C, 7.85% H, 1.68% O, and 0.740% N. In addition, the mass density of the polymer is assumed to be 1.00 g/cm3. The polymer spheres are uniformly coated in 103Pd. The spheres and the X‐ray marker are double encapsulated in titanium shell with total thickness of 0.080 mm and outer diameter of 0.800 mm. The overall source length is 5.00 mm and the active length is 4.76 mm. The maximum possible displacement of a source sphere is 0.570 mm along the seed axis and up to 0.070 mm in the radial direction. The schematic diagram of the simulated source is shown in Fig. 1(b).

A.3 192Ir

The source geometry and dimensions of the VariSource VS2000 were taken from the investigation of Angelopoulos et al., (13) Papagiannis et al., (14) and Llisp et al. (15) The VS2000 consists of two 2.50 mm long sources pellets. Each source pellet is made up of a 2.16 mm long cylindrical section with a 0.34 mm diameter and hemispherical ends with the same diameter. The two sources are placed laser drilled hole made at the end of a solid Ni/Ti wire (55.6%/44.4%) with a diameter of 0.59 mm. The end of the wire was sealed with a 1 mm thick plaque. This source end was modeled as a hemisphere with its center shifted 3.205 mm relative to the center of the source. During the simulation, the length of the wire was extended by about 5.0 cm from the center of the source to include the impact of the source wire on the dose distribution. The active length of this source is 5.0 mm. The schematic diagram of the simulated source is shown in Fig. 1(c).

A.4 137Cs

Spherical 137Cs pellet sources of the LDR Selectron remote afterloading system, with 2.50 mm in diameter encapsulated in 0.5 mm stainless steel material, was simulated for this source. (8) The active cores of the sources contain spherical borosilicate glass with 1.50 mm diameter. For treatment of gynecological diseases, different combinations of active pellet sources and dummy pellets are moved through a plastic catheter and stopped at the end of the applicator by a stainless stopping screw.

In this study, a source train containing five active Amersham pellets was considered at the end of the cylindrical applicator to simulate a linear 137Cs source, with 1 cm active length. The schematic diagram of the simulated source is shown in Fig. 1(d).

B. TG‐43 dose calculations

The AAPM Task Group 43 (TG‐43)(1‐2) introduced a formalism for calculation of dose distribution around a low‐energy sealed brachytherapy source. Also, there is a report from the AAPM and ESTRO for dose calculation around photon‐emitting brachytherapy sources with average energy higher than 50 keV. (16) We have used this report for dosimetric evaluation of 137Cs and 192Ir. This new report explains the differences of low‐energy sources and high‐energy sources for distances very close to the sources. At larger distances, there are not significant differences between the dosimetry of the sources using the two recommendations. Since the goal of this project is toward the dosimetry of the source for distances greater than 1 cm, the updated TG‐43 protocol (2) (i.e., TG‐43U1) will be used for all the sources.

According to the TG‐43U1 formalism, (2) the dose rates around brachytherapy sources are obtained from:

D˙=SKΛGl(r,θ)Gl(r0,θ0)gL(r)F(r,θ) (1)

where SK is the air‐kerma strength of the source, Λ is the dose rate constant, GL(r,θ) is the geometry function, gL(r) is the radial dose function, and F(r,θ) is the 2D anisotropy function. The (r0,θ0) is the reference point relative to the source with r0=1cm and θ0=π/2. The subscript “L” has been added in TG‐43U1 (2) to denote the linear source approximation. The air‐kerma strength, SK is the air‐kerma rate K˙δ(d) multiplied by the square of the distance for energy levels greater than δ:

SK=K˙δ(d).d2 (2)

Δ is the ratio of the dose rate at the reference point, P(r0,θ0) and SK. The unit of is cGyh1U1:

D˙=SKΛGl(r,θ)Gl(r0,θ0)gL(r)F(r,θ) (1)

The geometry function accounts for the activity distribution within the source and the distance between the point of interest and the source. TG‐43U1 defines the geometry function as

GL(r,θ)={βLrsinθifθ0(r2L24)1ifθ=0GL(r)=D˙(r,π2)GL(r0,π2)D˙(r0,π2)GL(r,π2) (4)

where β is the angle in radians, subtended by the point of interest, P(r,θ), to the tips of the active length of a hypothetical line; L is the active length of the source. The radial dose function, gL(r), takes into account the effects of photon scattering and attenuation and is defined as follows:

GL(r)=D˙(r,π2)GL(r0,π2)D˙(r0,π2)GL(r,π2) (5)

The 2D anisotropy function describes the variations in dose as a function of polar angle relative to the transverse plane and is defined as:

F(r,θ)=D˙(r,θ)D˙(r,θo)G(r,θo)G(r,θ) (6)

C. MCNP code

MCNP is a general purpose Monte Carlo radiation transport code that can be used for neutron, photon, electron, or coupled neutron/photon/electron transport. (30) Three versions of the MCNP code have been used in this study: MCNP4C2, MCNP5, and MCNPX. When a version of MCNP has been released, newly introduced features may be based on those from previous versions or these features may be novel to MCNP. However, the sources of these new features are not always explicit.

C.1 MCNP4C2

MCNP4C2 is the first major release of MCNP since version MCNP4B (February 4, 1997). (31) One of the major new features of MCNP4C2 relative to version MCNP4B is the use of ENDF/B‐VI improvements. Monte Carlo simulations are, in general, only as accurate as the underlying cross‐sectional libraries. (31) Several investigators 7 , 32 have discussed the problems with the photon cross‐sectional data that were included in MCNP (31) simulations. The effect of these problems is especially visible in dose calculations involving low‐energy photon emitting seeds. In 2002, the photon cross‐sectional data included with the standard distribution of MCNP was completely revised (X‐5 Monte Carlo Team, 2003). (31)

For MCNP4C2 there are two photon interaction data libraries Labeled as ZZZ000.nnP with nn=01 and nn=02. (29) For the ZAID = ZZZ000.01P library, the photon interaction tables were created for elements from z=1 through z=94 except z=84, 85, 87, 88, 89, 91, and 93. ZAID is Nuclide Identification number. This number is used to identify the element or nuclide desired. The form of the number is ZZZAAA.nnX, where ZZZ is the atomic number of the element or nuclide, AAA is the mass. These tables were based on evaluated data from ENDF (33) from 1 keV to 100 MeV. However, in the ZAID=ZZZ000.02P library, the values were a subset of the ZAID=ZZZ000.01P library with pair production thresholds added for the Storm‐Israel (34) data. Data above 15 MeV for the Storm‐Israel data and above 100 MeV for the ENDF data come from adaptation of the Livermore Evaluated Photon data library (EPDL (35) ) and are valid for energies up to 100 GeV. (29) For this project we used data library for nn=02. (34)

C.2 MCNP5

The MCNP5 code incorporates an arbitrary three‐dimensional configuration for materials in geometric cells bounded by first‐ and second‐degree surfaces and fourth‐degree elliptical tori. For photons, the code accounts for incoherent and coherent scattering, the possibility of fluorescent emission after photoelectric absorption, and the absorption in electron‐positron pair‐production. (28) There are four photon transport libraries maintained by X‐5 and distributed with MCNP: MCPLIB, MCPLIB02, MCPLIB03, and MCPLIB04. (35) MCPLIB04 was officially released in 2002. (29) The cross section, form factor, and fluorescence data are all derived from the ENDF/B‐VI.8 data library that are derived from EPDL97. (34) Cross‐sectional data are given for incident photon energies from 1 keV to 100 GeV. Fluorescence data are derived from the atomic relaxation data available in ENDF/B‐VI.8, but use the storage and sampling scheme defined by Everett and Cashwell. (33)

Electron interaction data tables are required both for problems in which electrons are actually transported, and for photon problems in which the thick‐target bremsstrahlung model is used. Electron data tables are identified by ZAIDs of the form ZZZ000.nnE, and are selected by default when the problem mode requires them. There are two electron interaction data libraries: el01 (ZAID endings of .01e) and el03 (ZAID endings of .03e). (30) The electron libraries contain data on an element‐by‐element basis for atomic numbers from Z equal 1 to 94. The library data contain energies for tabulation, radiative stopping power parameters, bremsstrahlung production cross sections bremsstrahlung energy distributions, K‐edge energies, and Auger electron production energies. It also contains parameters for the evaluation of the Goudsmit‐Saunderson theory (29) for angular deflection based on the Riley cross‐sectional calculation. This file includes Mott correction factors to the Rutherford cross sections used in the Goudsmit‐Saunderson theory. The el03 library includes the atomic data of Carlson that was used in the density effect calculation. (30)

C.3 MCNPX

MCNPX is a general purpose Monte Carlo radiation transport code that tracks nearly all particles, as well as photons and electrons, at almost all energies. It is a superset of MCNP4C3 (36) and has many capabilities beyond MCNP4C3 such as heavy ion transport and long file names. It was first released to the public in 1999 as version 2.1.5. Many new tally sources and variance reduction options have been developed and new tallies have been created specific to the intermediate and high‐energy physics ranges. (36) The MCNPX photon transport libraries are identical to the MCNP5 libraries that were defined in the previous section. The major difference of MCNP5 and MCNPX is their energy ranges. The energy ranges for MCNP5 are 1 keV to 1 GeV for electrons, and 1 keV to 100 GeV for photons.

C.4 Revised MCNP4C

Another code that has been generated in this project (referred here after as MCNP4C‐revised) is the original MCNP4C2 code with the cross‐sectional library taken from the MCNP5 and MCNPX library. Therefore, the cross‐sectional library is changed from ENDF/B‐VI version 5 to ENDF/B‐VI. 8. This adjustment is made in order to identify if the differences between MCNP4C2 and other codes are simply the differences in their cross‐sectional library or if changes in the code are responsible for these differences.

D. Phantom geometry

To obtain the dose rate constant, Δ, radial dose function, gL(r), and the anisotropy function, F(r,θ), the simulations were performed in a spherical water phantom with 30 cm radius. Each source type was placed at the center of the phantom, and spherical tally cells were defined at different distances from the source center. The radius of the tally cells was 0.02 cm for distances less than 2 cm and 0.05 cm for distances greater than 3 cm. For each simulation, 109 particle histories were considered to ensure that the relative errors (R(%)=100Sx¯X¯ that represents the estimated relative error (%) at the 1 σ level) of MCNP simulations at all distances are less than 1%. To obtain air‐kerma strength, each source type was placed at the center of an air phantom, the kerma rates were obtained in spherical tally cells, using F6‐tally. The air‐kerma rates, obtained for distances 1 to 10 cm, were multiplied by the square of the distance to obtain the air‐kerma strength. The average values of the air‐kerma strength were considered as the SK value for each source. In all simulations *F4 and F6 tallies were used. The results of the

*F4 tally were multiplied by the values of mass absorption coefficients, to convert them to dose values. The TG‐43 parameters (i.e., dose rate constant, gL(r), and F(r,θ)) were obtained according to Eq. (2) to (6). The uncertainty analysis of each parameter was obtained using the error propagation method. F(r,θ) is obtained at distances from 1 cm to 10 cm and the angles of 0° to 180° by 10° spacing. Radial dose functions have been obtained at distances of 1 cm to 6 cm of the source. The photon energy spectrum for 125I and 103Pd were taken from the TG43U1 report, (2) and the spectrum for 192Ir was taken from national nuclear data center. (37) For 137Cs, 662 keV monoenergetic gamma rays were considered.

III. RESULTS

A. Dose rate constant

Table 1 shows the dose rate constant of 103Pd,125I,192Ir,137Cs sources obtained by the four different MCNP codes and also the published values by different investigators. 9 , 11 , 13 , 38 , 39 For a better visibility of the differences between different Monte Carlo codes, these results were compared with the values obtained from the MCNP5 code (the most recently used code). Differences between the data from all the codes are excellent (0.1%) except MCNP4C2 that is 4.4% and 1.1% for 103Pd and 125I, respectively. It should be noted that the results of the simulations by the MCNP4C‐revised code was similar to those of MCNP5 and MCNPX codes. Therefore, the main source of error for the dose rate constant of the 103Pd and 125I from MCNP4C2 code was the use of the cross‐sectional file.

Table 1.

The dose rate constant (Δ:cGyh1U1) of 103Pd,125I,192Ir, and 137Cs sources

Source Method Dose Rate Constant
(cGy1U1)
Differences Relative to the MCNP5 Value
103Pd
MCNP4C2
0.726±0.6%
4.6%
MCNPX
0.695±0.6%
0.1%
MCNP5
0.694±0.6%
MCNP4C‐Revised
0.695±0.6%
0.1%
Meigooni et al. (11) 0.69
0.6%
TG‐43U1S1 (38) 0.685
1.3%
Taylor & Rogers (39) 0.650
6.3%
125I
MCNP4C2
0.935±0.6%
1.1%
MCNPX
0.924±0.6%
0.1%
MCNP5
0.925±0.6%
MCNP4C‐Revised
0.928±0.6%
0.3%
Meigooni et al (9) 0.99 7%
TG‐43U1S1 (38) 0.981 6%
Taylor & Rogers (39) 0.925 0.0%
192Ir
MCNP4C2
1.119±0.5%
0.1%
MCNPX
1.119±0.5%
0.1%
MCNP5
1.118±0.5%
MCNP4C‐Revised
1.119±0.5%
0.1%
Angelopoulos et al. (13) 1.101
1.5%
Taylor & Rogers (22) 1.099
1.7%
137Cs
MCNP4C2
0.993±0.5%
0.2%
MCNPX
0.995±0.5%
0.0%
MCNP5
0.995±0.5%
0.0%
MCNP4C‐Revised
1.007±0.5%
1.2%

The uncertainty of dose rate constant, originates from two parameters: SK and D˙(r0,θ0). The error of SK is calculated by the quadratic summation of the uncertainties from a point dose and the values calculated by deviations from the fitted line source. The uncertainty of D˙(r0,θ0) is calculated statistical fluctuations by the MCNP codes. Then, the uncertainty of the dose rate constant is calculated following the error propagation formula. (32)

B. Radial dose function

Table 2 shows the simulated radial dose functions of the 103Pd,125I,192Ir,137Cs sources as a function of radial distance, obtained by the four different Monte Carlo codes, using *F4 and F6 tally options. The percentage difference between the results of these codes relative to the MCNP5 data sets for 103Pd and 192Ir are shown in Fig. 2. These results indicate that for the F6 tally option, the percentage differences between the values obtained by the MCNP5 and MCNPX Monte Carlo codes are less than 6.7% for all distances. However, the percentage differences between the data obtained by the MCNP4C2 and MCNP5 codes increases with increasing distance from the source. The percentage differences between MCNP4C2 and MCNP5 codes at 6 cm distance from the 103Pd source reach approximately 27% and 28% for F6 and *F4 tallies, respectively. However, the percentage differences between the MCNP4C2 and MCNP5 for 125I,192Ir, and 137Cs sources at a distance of 10 cm are less than 10.5%, 0.5%, and 1%, respectively. Interestingly, the results of the simulations by the MCNP4C‐revised code are similar to those of MCNP5 and MCNPX codes. Therefore, the main source of error for the radial dose function of the low‐energy sources from MCNP4C2 code was its cross‐sectional

Table 2.

Comparison of the radial dose function of 103Pd source obtained by F6 and *F4 tally options

Source r (cm) MCNP4C2 *F4 tally MCNP4C2 F6 tally MCNP5 *F4 tally MCNP5 F6 tally MCNPX *F4 tally MCNPX F6 tally MCNP4C‐Revised *F4 tally MCNP4C‐Revised F6 tally
 103Pd
1
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+00
2
6.04E01
6.03E01
5.89E01
5.89E01
5.83E01
5.82E01
5.82E01
5.82E01
3
3.52E01
3.52E01
3.18E01
3.18E01
3.23E01
3.23E01
3.23E01
3.22E01
4
1.96E01
1.95E01
1.68E01
1.68E01
1.69E01
1.69E01
1.69E01
1.69E01
5
1.12E01
1.11E01
9.73E02
9.73E02
9.61E02
9.61E02
9.61E02
9.61E02
6
6.92E02
6.90E02
5.42E02
5.42E02
5.41E02
5.41E02
5.41E02
5.41E02
 125I
1 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000
2 0.8483 0.8484 0.8165 0.8137 0.8180 0.8159 0.8171 0.8143
3 0.6797 0.6801 0.6358 0.6357 0.6422 0.6385 0.6368 0.6368
4 0.5369 0.5372 0.4913 0.4913 0.4936 0.4907 0.4904 0.4904
5 0.4141 0.4146 0.3721 0.3722 0.3731 0.3709 0.3711 0.3710
6 0.3119 0.3121 0.2779 0.2778 0.2762 0.2746 0.2735 0.2735
7 0.2402 0.2400 0.2081 0.2080 0.2053 0.2065 0.2024 0.2023
8 0.1752 0.1750 0.1519 0.1516 0.1544 0.1553 0.1539 0.1538
9 0.1365 0.1365 0.1123 0.1123 0.1105 0.1111 0.1110 0.1109
10 0.0961 0.0961 0.0873 0.0873 0.0810 0.0814 0.0796 0.0796
 192Ir
1 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000
2 0.9932 0.9933 0.9958 0.9958 0.9934 0.9936 0.9935 0.9936
3 1.0036 1.0049 1.0040 1.0036 1.0031 1.0038 1.0042 1.0036
4 0.9990 1.0008 1.0044 1.0033 0.9981 0.9987 0.9997 0.9990
5 0.9986 0.9960 1.0018 0.9989 1.0001 0.9948 0.9950 0.9999
6 0.9928 0.9971 0.9965 0.9952 0.9923 0.9939 0.9938 0.9926
7 0.9729 0.9681 0.9732 0.9715 0.9672 0.9656 0.9675 0.9664
8 0.9555 0.9432 0.9459 0.9395 0.9524 0.9444 0.9446 0.9426
9 0.9338 0.9313 0.9391 0.9407 0.9320 0.9335 0.9340 0.9321
10 0.9243 0.9243 0.9122 0.9141 0.9277 0.9238 0.9265 0.9240
 137Cs
1 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000
2 0.9909 0.9938 0.9978 0.9948 0.9908 0.9938 0.9938 0.9908
3 0.9883 0.9942 0.9909 0.9852 0.9879 0.9941 0.9941 0.9879
4 0.9853 0.9935 0.9891 0.9810 0.9856 0.9939 0.9939 0.9856
5 0.9706 0.9805 0.9763 0.9661 0.9693 0.9794 0.9794 0.9693
6 0.9401 0.9527 0.9538 0.9421 0.9400 0.9526 0.9526 0.9400
7 0.9526 0.9386 0.9496 0.9360 0.9520 0.9378 0.9520 0.9378
8 0.9234 0.9053 0.9188 0.9045 0.9227 0.9047 0.9196 0.9047
9 0.8838 0.8680 0.8840 0.8695 0.8842 0.8686 0.8835 0.8686
10 0.8852 0.8540 0.9000 0.8612 0.8850 0.8516 0.8680 0.8516

Figure 2.

Figure 2

The percentage differences between the radial dose functions of  103Pd (a),  125I (b),  192Ir (c), and  137Cs (d) sources calculated by different MCNP codes and the values obtained by the MCNP5 code using the F6 tally.

file. A comparison of the simulated radial dose function of the sources in this project using F6 tally with the published data is shown in Fig. 3.

Figure 3.

Figure 3

Comparison of the Monte Carlo simulated radial dose functions of  103Pd (a),  125I (b),  192Ir (c), and  137Cs (d) sources obtained by the four MCNP codes in this project, using F6 tally. These results are also compared with the published data by Meigooni et al. (11) for  103Pd, Meigooni et al. (9) for  125I, and Angelopoulos et al. (13) for  192Ir.

C. Anisotropy function

The values of 2D anisotropy function, F(r,θ), for the  103Pd, 125I, 192Ir, 137Cs sources were obtained by MCNP4C2, MCNP5, MCNPX, and MCNP4C‐revised Monte Carlo codes at distances of 1 cm to 10 cm. Figure 4 shows the comparison of the 2D anisotropy functions of these sources by the four codes using *F4 tally and also published data by other investigators. 9 , 11 , 13 , 38 Similar results were obtained by F6 tally. No significant differences between different codes had been observed.

Figure 4.

Figure 4

Comparison of the Monte Carlo simulated 2D anisotropy functions of  125I (a),  103Pd (b),  192Ir (c), and  137Cs (d) sources obtained by the three MCNP codes and the revised MCNP4C2 code using F6 tally. These results are also compared with the published data by Meigooni et al. (11) for  103Pd, Meigooni et al. (9) for  125I, and Angelopoulos et al. (13) for  192Ir.

IV. DISCUSSION & CONCLUSION

Four versions of Monte Carlo codes (MCNP4C2, MCNPX, MCNP5, and MCNP4C‐revised) were used to obtain the TG‐43 dosimetry parameters for low‐ and high‐energy brachytherapy sources. To demonstrate the variation of the results between the codes, all the simulated parameters were compared with the data from the MCNP5 code. The results of these investigations for the high energy brachytherapy sources (i.e.,  192Ir and  137Cs) indicated that the differences among the data from the four codes are less than 5.6%. However, in the case of low‐energy brachytherapy sources such as  103Pd and  125I, the difference between the radial dose functions obtained by MCNP4C2 and those obtained by two other versions of the code (MCNPX and MCNP5) changes as a function of distance. The differences were found to be more than 27% for  103Pd at r=6 cm, and more than 10% for  125I at r=10 cm. The values of the dose rate constants for MCNP4C2 were approximately 4.4% and 1.1% larger than MCNP5 value for  103Pd and  125I, respectively. No significant differences were found on 2D anisotropy functions.

Such differences show that the cross‐sectional library used by the MCNP4C2 code for low‐energy gamma rays were inaccurate. Therefore, the simulations were performed using the MCNP4C‐revised code, which means that the MCNP4C2 cross section was changed to that used in the MCNPX and MCNP5 (ENDF/B‐VI.8) codes. A comparison of results obtained with the MCNP4C‐revised and MCNP5 codes show that the average differences in the dose rate constant are 1.2%, 0.1%, 0.2%, and 0.1% for  137Cs, 192Ir, 125I, and  103Pd respectively.

The maximum difference for *F4 and F6 tallies obtained by MCNPX and MCNP4C‐revised in all points around the source were not more than 0.01%, 0.21%, 0.67%, and 0.41% for  137Cs, 192Ir, 125I, and  103Pd respectively.

The results indicate that the only difference between MCNP4C‐revised and MCNPX is the difference in their cross‐sectional libraries and, as such, the results obtained with the two codes are similar. However, the difference between the results obtained with the MCNP5 and MCNP4C‐revised codes is greater than the difference between the MCNPX and MCNP4C‐revised codes.

ACKNOWLEDGMENTS

The authors would like to thank Ms. Michele A. Klimke, the senior dosimetrist at Comprehensive Cancer Centers of Nevada, for her valuable comments and suggestions for editorial aspects of the manuscript.

COPYRIGHT

This work is licensed under a Creative Commons Attribution 4.0 International License.

REFERENCES

  • 1. Nath R, Anderson LL, Luxton G, Weaver KA, Williamson JF, Meigooni AS. Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No.43. Med Phys. 1995;22(2):209–34. [DOI] [PubMed] [Google Scholar]
  • 2. Rivard MJ, Coursey BM, DeWerd LA, et al. Update of AAPM Task Group No. 43 Report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31(3):663–74. [DOI] [PubMed] [Google Scholar]
  • 3. Meigooni AS, Gearheart DM, Sowards K. Experimental determination of dosimetric characteristics of Best 125I brachytherapy source. Med Phys. 2000;27(9):2168–73. [DOI] [PubMed] [Google Scholar]
  • 4. Nath R and Yue N. Dosimetric characterization of an encapsulated interstitial brachytherapy source of 125I on a tungsten substrate. Brachytherapy. 2002;1(2):102–09. [DOI] [PubMed] [Google Scholar]
  • 5. Sowards K and Meigooni AS. Monte Carlo evaluation of the dosimetric characteristics of the Best Model 2301 125I brachytherapy source. Appl Radiat Isot. 2002;57(3):327–333. [DOI] [PubMed] [Google Scholar]
  • 6. Li Z, Fan JJ, Palta JR. Experimental measurements of dosimetric parameters on the transverse axis of a new 125I source. Med Phys. 2000;27(6):1275–80. [DOI] [PubMed] [Google Scholar]
  • 7. DeMarco JJ, Wallace RE, Boedeker K. An analysis of MCNP cross‐sections and tally methods for low‐energy photon emitters. Phys Med Biol. 2002;47(8):1321–32. [DOI] [PubMed] [Google Scholar]
  • 8. Perez‐Calatayud J, Granero D, Ballester F, Puchades V, Casal E. Monte Carlo dosimetric characterization of the Cs‐137 selectron/LDR source: evaluation of applicator attenuation and superposition approximation effects. Med Phys. 2004;31(3):493–99. [DOI] [PubMed] [Google Scholar]
  • 9. Meigooni AS, Hayes JL, Zhang H, Sowards K. Experimental and theoretical determination of dosimetric characteristics of IsoAid ADVANTAGE 125I brachytherapy source. Med Phys. 2002;29(9):2152–58. [DOI] [PubMed] [Google Scholar]
  • 10. Solberg TD, DeMarco JJ, Hugo G, Wallace RE. Dosimetric parameters of three new solid core I‐125 brachytherapy sources. J Appl Clin Med Phys. 2002;3(2):119–34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Meigooni AS, Bharucha Z, Yoe‐Sein M, Sowards K. Dosimetric characteristics of the best double‐wall 103Pd brachytherapy source. Med Phys. 2001;28(12):2567–75. [DOI] [PubMed] [Google Scholar]
  • 12. Peterson SW and Thomadsen B. Measurements of the dosimetric constants for a new 103Pd brachytherapy source. Brachytherapy. 2002;1(2):110–19. [DOI] [PubMed] [Google Scholar]
  • 13. Angelopoulos A, Baras P, Sakelliou L, Karaiskos P, Sandilos P. Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source. Med Phys. 2000;27(11):2521–27. [DOI] [PubMed] [Google Scholar]
  • 14. Papagiannis P, Angelopoulos A, Pantelis E, et al. Dosimetry comparison of 192Ir sources. Med Phys. 2002;29:2239–46. [DOI] [PubMed] [Google Scholar]
  • 15. Lliso F, Perez‐Calatayud J, Carmona V, Ballester F, Puchades V, Granero D. Technical note: Fitted dosimetric parameters of high dose‐rate 192Ir sources according to the AAPM TG43 formalism. Med Phys. 2003;30(4):651–54. [DOI] [PubMed] [Google Scholar]
  • 16. Perez‐Calatayud J, Ballester F, Das RK, et al. Dose calculation for photon‐emitting brachytherapy sources with average energy higher than 50 KeV: report of the AAPM and ESTRO. Med Phys. 2012;39(5):2904–29. [DOI] [PubMed] [Google Scholar]
  • 17. Zehtabian M, Faghihi R, Sina S. Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations. Iran J Radiat Res. 2010;7(4):217–22. [Google Scholar]
  • 18. Mowlavi AA and Yazdani M. Dosimetry Parameters Determination of 169Yb Brachytherapy Source Model X1267. Appl Phys Res. 2010;2(1):118–24. [Google Scholar]
  • 19. Mosleh‐Shirazi MA, Hadad K, Faghihi R, Baradarn‐Ghahfarokhi M, Naghshnezhad Z, Meigooni AS. EchoSeed Model 6733 Iodine‐125 brachytherapy source: improved dosimetric characterization using the MCNP5 Monte Carlo code. Med Phys. 2012;39(8):4653–59. [DOI] [PubMed] [Google Scholar]
  • 20. Wu CH, Liao YJ, Hsueh Liu YW, Hung SK, Lee MS, Hsu SM. Dose distributions of an 192Ir brachytherapy source in different media. BioMed Res Int. 2014;Volume(2014):article ID 946213. Retrieved November 1, 2015 from http://www.hindawi.com/journals/bmri/2014/946213/abs/ [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Wang R and Sloboda RS. EGS4 dosimetry for cylindrical symmetric brachytherapy sources. Med Phys. 1996;23(8):1459–65. [DOI] [PubMed] [Google Scholar]
  • 22. Taylor RE and Rogers DW. EGSnrc Monte Carlo calculated dosimetry parameters for 192Ir and 169Yb brachytherapy sources. Med Phys. 2008;35(11):4933–44. [DOI] [PubMed] [Google Scholar]
  • 23. Perez‐Calatayud J, Granero D, Ballester F. Phantom size in brachytherapy source dosimetric studies. Med Phys. 2004;31(7):2075–81. [DOI] [PubMed] [Google Scholar]
  • 24. Berger MJ and Hubbell JH. XCOM: Photon cross‐sections on a personal computer. NBSIR (Washington, DC: NBS) 87‐3597. Gaithersburg, MD: NBS; 1987. [Google Scholar]
  • 25. Berger MJ and Hubbell JH. Photon attenuation coefficients. In: Lide DR, editor. CRC handbook of chemistry and physics, 79th edition Boca Raton, FL: CRC Press; 1998/1999. [Google Scholar]
  • 26. Trubey DK. HUGI VI notes documentation for the DLC‐146 code package. Radiation Safety Information Computational Center. Oak Ridge, TN: OakRidge National Laboratory; 1989. [Google Scholar]
  • 27. Sowards K and Meigooni AS. A Monte Carlo evaluation of the dosimetric characteristics of the Echo Seed Model 6733 125I brachytherapy source. Brachytherapy. 2002;1(4):227–32. [DOI] [PubMed] [Google Scholar]
  • 28. Cullen DE, Hubbel JH, Kissel LD. EPDL97: the evaluated photon data library,'97 version. Lawrence Livermore National Laboratory report, UCRL‐50400, Volume 6,Rev 5. Livermore, CA: Lawrence Livermore National Laboratory; 1997. [Google Scholar]
  • 29. Briesmeister J. MCNP— a general Monte Carlo N‐particle transport code, version 4C. LA‐13709‐M. Los Alamos, NM: Los Alamos National Laboratory; 2000. [Google Scholar]
  • 30. MCNP‐5 Monte Carlo Team . MCNP — a general Monte Carlo N‐Particle transport code, version 5. Los Alamos, NM: Los Alamos National Laboratory; 2003. [Google Scholar]
  • 31. Duggan DM. Improved radial dose function estimation using current version MCNP Monte‐Carlo simulation: Model 6711 and ISC3500 125I brachytherapy sources. Appl Radiat Isot. 2004;61(6):1443–50. [DOI] [PubMed] [Google Scholar]
  • 32. Knoll GF. Radiation detection and measurement, 4th edition Hoboken, NJ: John Wiley & Sons, Inc; 2010. [Google Scholar]
  • 33. Everett CJ and Cashwell ED. MCNP code fluorescence‐routine. Discussion. Los Alamos National Laboratory report LA‐5240‐MS. Los Alamos, NM: Los Alamos National Laboratory; 1973. [Google Scholar]
  • 34. Storm E and Israel HI. Photon cross sections from 0.001 to 100 Mev for elements 1 through 100. Los Alamos Scientific Laboratory report LA‐3753. Los Alamos, NM: Los Alamos Scientific Laboratory; 1967. [Google Scholar]
  • 35. Cullen DE, Chen MH, Hubbell JH, et al. Tables and graphs of photon‐interaction cross sections from 10eV to 100 GeV derived from the LLNL evaluated photon data library (EPDL). Lawrence Livermore National Laboratory report UCRL‐50400. Livermore, CA: Lawrence Livermore National Laboratory; 1989. [Google Scholar]
  • 36. Monte Carlo N‐Particle Transport Code System Including MCNP6.1, MCNP5‐1.60, MCNPX‐2.7.0 and Data Libraries. Available from: https://rsicc.ornl.gov/codes/ccc/ccc8/ccc-810.html.
  • 37. NUDAT 2.6. Upton, NY: National Nuclear Data Center, Brookhaven National Laboratory; 2012. [Google Scholar]
  • 38. Rivard MJ, Butler WM, DeWerd LA, et al. Supplement to the 2004 Update of the AAPM Task Group No. 43 Report. Med Phys. 2007;34(6):2187–205. [DOI] [PubMed] [Google Scholar]
  • 39. Taylor RE and Rogers DW. An EGSnrc Monte Carlo‐calculated database of TG‐43 parameters. Med Phys. 2008;35(9):4228–41. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Applied Clinical Medical Physics are provided here courtesy of Wiley