Skip to main content
Sensors (Basel, Switzerland) logoLink to Sensors (Basel, Switzerland)
. 2018 Mar 16;18(3):887. doi: 10.3390/s18030887

Accurate Determination of the Q Quality Factor in Magnetoelastic Resonant Platforms for Advanced Biological Detection

Ana Catarina Lopes 1, Ariane Sagasti 1,*, Andoni Lasheras 2, Virginia Muto 3, Jon Gutiérrez 1,4, Dimitris Kouzoudis 5, José Manuel Barandiarán 1
PMCID: PMC5876627  PMID: 29547578

Abstract

The main parameters of magnetoelastic resonators in the detection of chemical (i.e., salts, gases, etc.) or biological (i.e., bacteria, phages, etc.) agents are the sensitivity S (or external agent change magnitude per Hz change in the resonance frequency) and the quality factor Q of the resonance. We present an extensive study on the experimental determination of the Q factor in such magnetoelastic resonant platforms, using three different strategies: (a) analyzing the real and imaginary components of the susceptibility at resonance; (b) numerical fitting of the modulus of the susceptibility; (c) using an exact mathematical expression for the real part of the susceptibility. Q values obtained by the three methods are analyzed and discussed, aiming to establish the most adequate one to accurately determine the quality factor of the magnetoelastic resonance.

Keywords: magnetic biosensors, quality factor, magnetoelastic resonance

1. Introduction

Magnetoelastic resonators used as sensing devices present advantages like allowing remote “query and answer” [1,2] as well as low cost and low power consumption [3]. Due to these reasons, chemical and many other parameters can be detected: aqueous chemicals including pH [1], salt, and glucose concentrations [3], as well as inorganic salt depositions [4], gas humidity [5], gases such as carbon dioxide [6], or toxic volatile organic compounds (VOCs) such as benzene or hexane, among others [7]. In recent years, they have become a hot topic as novel wireless biosensors for bacteria, potentially lethal for humans, such as Salmonella [8,9], Bacillus anthracis [9], or Escherichia coli [10]. Such detection will be achieved if the surface of the magnetoelastic resonator is coated with an appropriate smart functionalized film that interacts with the target of interest.

The detection process in such sensors is based on the shift of the magnetoelastic resonance (MER) frequency under the action of an external agent, easily seen when measuring the magnetic susceptibility versus the frequency of the applied magnetic field. In the case of biological agents, the adhesion of different bacteria to the resonators causes an increase in the total mass, which leads to a decrease in the MER (see Figure 1).

Figure 1.

Figure 1

Principle of detection of biological targets using magnetoelastic resonators. The adhesion of the bacteria to the materials leads to an increase in the total mass of the system, which is detected as a shift (always decreasing) in the measured magnetoelastic resonance frequency.

Among the magnetoelastic materials to be used as biological sensors, Fe-based amorphous ferromagnetic alloys in the form of a ribbon are among the most suitable, mainly due to their high magnetoelastic coupling coefficient (k), high saturation magnetostriction (λS), and high saturation magnetization (MS) [11].

The good performance of a magnetoelastic sensing device is mainly determined by two parameters: sensitivity and quality factor. The sensitivity is related to the lowest detectable frequency change. This depends on the experimental system, but also on the sharpness of the resonance, which in turns depends on the Q factor. In some cases, the determination of Q can be more sensitive than that of the resonance frequency for detecting small changes in mass. With m0 and f0 the unloaded mass and corresponding MER frequency of a magnetoelastic resonator, respectively, its sensitivity to a change in mass due to an external target is given by the relationship:

S=δfδm=f02m0, (1)

where δf represents the resonance frequency shift caused by the presence of an external agent that causes a change of mass δm, mass change of the MER film. Thus, a high sensitivity value means a large δf shift for a given mass change. The linearity expressed by Equation (1) is valid for small mass changes compared to the initial MER film mass. Nevertheless, it is just an approximation of a more general expression [12] and is still subject to revision and discussion by the authors [13,14].

Concerning the quality factor Q, it has been already experimentally observed that damping strongly affects both resonant frequency and magnetoelastic resonance curve shape (see, for example, [15,16,17]). A high Q value means a sharp resonance frequency and, consequently, a well-defined fr resonance frequency. From the measured susceptibility curve around the magnetoelastic resonance, the Q quality factor can be estimated as the resonance curve full bandwidth Δf signal (or full width at half maximum power) relative to its susceptibility maximum frequency fr, that is:

Q0=frΔf, (2)

which is a dimensionless number [18,19]. This classical empirical first approximation can lead to errors as high as 20% in the correct Q value determination, as previously noted by Kaczkowski [20]. Therefore, in biological detection based on the magnetoelastic resonance frequency shift, accurate determination of the Q quality factor beyond the empirical expression Equation (2) turns out to be a key parameter.

In the present work, we present an extensive study of the determination of the Q factor in magnetoelastic resonant platforms. To do this, strips (L=4 cm) of Fe-rich Fe64Co17Si6.6B12.4 composition homemade metallic glass have been used. Determination of the Q quality factor value has been performed in three different ways: (a) analyzing the full susceptibility curve around the resonance (real and imaginary components); (b) numerical fitting of the magnitude (modulus) of the susceptibility; and (c) using an exact mathematical expression for the Q value arising from analysis of the real part of the susceptibility curve at the resonance.

2. Experimental

2.1. Material: Magnetic and Magnetostrictive Characterization

In the present study, Fe-based Fe64Co17Si6.6B12.4 composition homemade metallic glass ribbons were used. They were prepared by the single roller quenching method in the form of a long ribbon. Equal length strips (L=4 cm) were cut to perform all the magnetic and magnetoelastic characterizations. Room-temperature hysteresis loops were measured by a classical induction method, obtaining a saturation magnetization (given as internal magnetic induction in Tesla) of μoMS1.6 T and a magnetic susceptibility χ15,000. A magnetostriction value of λS22 ppm was measured using strain gauges connected to an electronic Wheatstone bridge. Figure 2 shows the obtained hysteresis loop and magnetostriction curves.

Figure 2.

Figure 2

Fe64Co17Si6.6B12.4 composition metallic glass magnetic characterization: (a) hysteresis loop and (b) magnetostriction curve.

2.2. Magnetoelastic Characterization

The metallic glasses of the present study show excellent coupling between magnetic and elastic properties, that is, the applied mechanical stress and the magnetic field generating equivalent effects in the magnetization and deformation of the materials. A direct consequence of the magnetoelastic coupling is the dependence of the elastic constants of magnetostrictive materials on the external magnetic field H, in particular the dependence of the longitudinal Young’s modulus on H, known as the ΔE effect (ΔE=1E(H)/ES, with ES being the Young’s modulus measured at magnetic saturation (a detailed mathematical formula can be found in [21]).

This ΔE effect is easy to measure experimentally through the change in mechanical resonance as a function of the field. The resonance can be excited by an alternating field and detected by the changes in magnetic susceptibility. For this purpose, we used a home-mounted, computer-controlled magnetoelastic resonance detection apparatus [22,23] that automatically changes the DC external applied magnetic field Hdc, also known as bias field, and sweeps the frequency of the AC magnetic field Hac in order to drive the sample to magnetoelastic resonance at a given bias. This is the so-called resonance–antiresonance detection method. We use an HP 3589A Spectrum Analyzer in order to quickly measure the magnitude (χ) of the susceptibility curve at the magnetoelastic resonance and store the resonant frequency fr at the maximum and antiresonance frequency fa at the minimum signals, together with the signal amplitude at the resonance. Besides the susceptibility χ, we also measure its real (χ') and imaginary (χ'') components separately with the help of a Signal Recovery 7280 Lock-in Amplifier. In these measurements, we extract the frequencies f'M and fm', at the maximum and the minimum signal of the χ', respectively, and to the frequency f''M at maximum χ''.

All these measured frequencies and in particular the resonance (fr) vary with the bias field Hdc, and so does the Young’s modulus, determined as E(H)=[2Lfr2(H)]2ρ [24], where L and ρ are the length and density, respectively, of the sample. Other useful magnetoelastic parameters that can be determined from these measurements are the magnetoelastic coupling coefficient (k=(π2/8)(1(fr/fa)2)) [25] and the quality factor of the resonance Q. All such quantities are a function of the applied external magnetic field.

Figure 3 shows the typical external applied magnetic field dependence of Young’s modulus E(H) and magnetoelastic coupling coefficient k(H) for our magnetostrictive material. It can be seen that there is a minimum in the E(H) curve that happens at a value of the applied external magnetic corresponding to Hk or effective anisotropy field of the sample. In the same field, the maximum of k(H) and minimum of Q(H) occur. This is an expected behavior since in fact the k value is high when the difference between fr and fa is also high, that is, the resonance curve is broad, and so its corresponding Q value (quality of the resonance curve) is poor. While the maximum of the coupling value k guarantees the best sensitivity S of a magnetoelastic resonator working as a biological or chemical sensor [13], the simultaneous occurrence of the worst Q value jeopardizes the accurate determination of the magnetoelastic resonance frequency. Bearing this in mind, we will study the Q factor under external bias field conditions for poor (k=0.065 at 16 Oe), medium (k=0.176 at 0.6 Oe), and good (k=0.282 at 2 Oe) magnetoelastic coupling, aiming to establish the most adequate method (analysis of the real and imaginary part of the susceptibility, numerical fitting of modulus of the susceptibility, and analytical calculations) to determine the quality factor of a magnetoelastic resonance.

Figure 3.

Figure 3

Magnetoelastic characterization of Fe64Co17Si6.6B12.4 metallic glass: magnetic field dependence of (a) Young’s modulus E(H); (b) magnetoelastic coupling coefficient k(H).

3. Results: Determination of the Q Quality Factor Value

3.1. Full Susceptibility Curve Analysis at Resonance

Figure 4 shows the magnetic susceptibility modulus (χ) curves for our Fe64Co17Si6.6B12.4 (L=4 cm) composition metallic glass, measured around the magnetoelastic resonance frequency for all the applied magnetic field cases under study. In these measurements, the frequency step between consecutive points was 10 Hz.

Figure 4.

Figure 4

Susceptibility modulus (χ) measured for the Fe64Co17Si6.6B12.4 (L=4 cm) composition metallic glass, for all the bias magnetic field cases.

From these susceptibility modulus (χ) curves, we can directly extract the resonance (fr) and antiresonance (fa) frequencies and so estimate the magnetoelastic coupling coefficient k and the quality factor Q, from Equation (2), as Table 1 summarizes. As mentioned before, however, the values of Q0 are quite inaccurate.

Table 1.

Resonance and antiresonance frequencies for Fe64Co17Si6.6B12.4 and calculated k and Q0 values determined directly from the experimental data (this last one obtained using Equation (2)).

H (Oe) fr (Hz) fa (Hz) Δf k Q0
0.6 53,260 53,940 160 0.176 333
2 52,496 54,280 184 0.282 285
16 56,380 56,476 32 0.065 1762

However, a careful measurement of those magnetoelastic resonance curves, using a Lock-in Amplifier, allows us to record the susceptibility real and imaginary parts (χ and χ, respectively ) as Figure 5 shows.

Figure 5.

Figure 5

(a) Real, χ', and (b) imaginary, χ'', parts of the magnetic susceptibility for the Fe64Co17Si6.6B12.4 (L=4 cm) metallic glass ribbon, for the three bias magnetic field cases under study.

With the measured frequencies corresponding to maximum and minimum values of χ' (fM' and fm', respectively, see Figure 5a), we can use the first approximated expression often used to give an accurate value of this Q quality factor [20]:

Q1frfmfM . (3)

Table 2 summarizes the experimentally obtained frequency data for maximum and minimum values of χ' (fM' and fm', respectively) and calculated Q1 values from Equation (3), as well as the relative difference between Q0 and Q1.

Table 2.

Frequencies for the maximum and minimum of χ' (fM' and fm', respectively), calculated Q1 values (using Equation (3)) and relative error between Q0 and Q1.

H (Oe) fM (Hz) fm (Hz) Q1 Relative Difference (%) (Q0,Q1)
0.6 53,212 53,396 290 13
2 52,456 52,672 243 15
16 56,372 56,412 1410 20

It must be noted that to separately obtain the real and imaginary parts of the susceptibility is a quite difficult and time-consuming experimental task. Therefore, a method to obtain Q based only on the magnitude or modulus of the susceptibility is highly desirable, though it demands more complex numerical treatment afterwards.

3.2. Numerical Fitting of the Magnitude of the Susceptibility Curve

In 1978 Savage et al. [25] derived the following expression for the susceptibility around the magnetoelastic resonance in a free-standing cylinder-shaped sample:

χ(ω)=χ0[18k2π2n1n2×11ωn2ω2+iQ1ωnω], (4)

where k is the magnetoelastic coupling coefficient, ωn=2πfn is the frequency of the nth harmonic of the excited fundamental mode (n=1), Q1 is a phenomenological damping coefficient, and χ0 is the magnetic susceptibility measured at a frequency far below the resonance [26]. Equation (4) also applies to rectangular section ribbons with a proper choice of the k factor. Figure 6 shows an example of the calculated magnetic susceptibility, up to the fifth harmonic, by using Equation (4).

Figure 6.

Figure 6

Calculated magnetic susceptibility vs. frequency behavior for a magnetoelastic ribbon, up to the fifth harmonic, using Equation (4).

A different approach to estimate Q of a magnetoelastic resonance curve is the numerical fitting of the modulus or magnitude of the experimental susceptibility around its first resonant mode (n=1) by using Equation (4). Thus, we proceed to perform numerical fittings using Mathematica© software (v.11.0), following two different strategies: (a) by using the measured fr, fa, χ0 values as fixed parameters and (b) by leaving these parameters to vary around the experimentally obtained ones. In both cases, our goal is to search for the optimum Q value that minimizes the L2 norm between the fit and experimental data. We define such a norm (also called residual) as: ℛ =1Ni=1,N(χexp,iχfit,iχmax)2, where χmax=max(χmax,exp,χmax,fit) and N is the number of experimental points. In our measurements, N=397, 691 and 639 for H=0.6, 2 and 16 Oe, respectively. With such a definition, 0R 1, and a value of R close to 0 means very good fitting.

3.2.1. Numerical Fitting of the Susceptibility Curve Using Fixed Parameters

Figure 7 shows the measured magnetoelastic resonance curve at H=2 Oe and the fitted one when procedure a) is used. The only free parameter was the quality factor Q with an initial value of Q0=285. The sweep range for Q was 50380, and the optimum fitting was found for Qfit1=178 (see Figure 7 inset). For all the fits performed, the frequency step between consecutive points will be 1 Hz. The fit of Figure 7 is the best one obtained by following the a procedure. It looks satisfactory, but can still be improved.

Figure 7.

Figure 7

Measured resonance curve at H=2 Oe (black dots) and fitted one (magenta marks). The inset shows the residual ℛ change versus Q values, all obtained in calculations using procedure a.

Table 3 summarizes the results obtained for the three different magnetic field cases. While all obtained Qfit1 values are lower than the previous Q0 ones, it is noticeable that the worst fit corresponds to the applied magnetic field, where magnetoelastic coupling is maximum but quality factor is minimum.

Table 3.

Q values from the fitting of the χ susceptibility modulus, using fixed experimental parameters (procedure a).

H (Oe) fr (Hz) fa (Hz) Qfit1
0.6 53,260 53,940 208 0.0041
2 52,496 54,280 178 0.0039
16 56,380 56,476 1067 0.00031

3.2.2. Numerical Fitting of the Susceptibility by Leaving All Parameters Free

Figure 8 shows the measured magnetoelastic resonance curve at H=2 Oe and the fitted one when this second procedure is used: all parameters were left free in a range around the starting guess given by the experimental values appearing in Table 1. In this case we swept the Q value in the range 50440, finding the best fit for Qfit2=229, for the case of applied magnetic field H=2 Oe (see Figure 8 inset). Now the fit has greatly improved, as the norm values are much lower, especially for the curve with the highest Q value curve (at H=16 Oe) (see Table 4).

Figure 8.

Figure 8

Measured resonance curve at H=2 Oe (black dots) and fitted one (magenta marks). The inset shows the residual ℛ change versus Q values, all obtained in calculations when using procedure b.

Table 4.

Obtained parameters (resonance and antiresonance frequencies and Q) from the fitting leaving all parameters free (procedure b).

H (Oe) fr (Hz) fa (Hz) Qfit2
0.6 53,301 53,927 283 0.00027
2 52,566 54,296 229 0.00060
16 56,390 56,466 1321 0.000011

Table 4 summarizes the results obtained for the three applied magnetic field cases. Again, as with the first simulation procedure, all obtained Qfit2 values are lower than the previous Q0 ones and the worst fit corresponds to the applied magnetic field where magnetoelastic coupling is maximal but the quality factor is minimal.

3.3. An Exact Expression for the Q Factor

Starting with Equation (4) and taking into account the shape of the χ(ω) curve, one of us (J.G.) has derive an exact analytical expression for the Q factor value calculation. Since magnetic susceptibility χ can be described by Equation (4) as a complex number, χ=χ'+iχ'', we can separate the real and imaginary parts of this complex expression (taking into account only the first harmonic, n=1) as follows:

χ'χ0=18k2π2ω2(ω2ω12)(ω2ω12)2+(ωω1Q)2 , (5)

and

χ''χ0=8k2π2ω3ω1Q(ω2ω12)2+(ωω1Q)2 . (6)

The real part shows resonance at its maximum: (ωM') and antiresonance at its minimum: (ωm) (see Figure 5a). It is well known that at the local maxima and minima, the derivative vanishes so we can apply this criteria for the frequencies ω=ωM' and =ωm:

ddω(χ'χσ)=0=>ddωω2(ω2ω12)(ω2ω12)2+(ωω1Q)2=0 . (7)

After a short calculation, and taking into account that Q>0 and single-valued (which means ωM<ωr<ωm), the Q factor value can be written in the following two ways:

Q=ωm2ωm2ω12 and Q=ωM2ω12ωM2 . (8)

Solving these two expressions, first for ωM=ωr and afterwards for Q, we finally get an exact analytical expression for the Q factor value:

Qcal=ωm2+ωM2ωm2ωM2=1+u21u2 , (9)

where u=ω'M/ω'm=f'M/f'm. Table 5 summarizes the experimentally measured values for those frequencies and the subsequently obtained Qcal values.

Table 5.

Fe64Co17Si6.6B12.4 experimentally obtained data for resonance and antiresonance frequencies of the real part of the magnetic susceptibility, and calculated Q values using Equation (9).

H (Oe) fM' (Hz) fm' (Hz) Qcal
0.6 53,212 53,396 290
2 52,456 52,672 243
16 56,372 56,412 1410

4. Discussion

All the Q values obtained by the different calculation procedures explained in this study are summarized in Table 6, while Table 7 shows the estimated errors (in %, calculated as Error (%)=|Q1QQ1|·100) among those obtained Q values. The first thing to notice is that in all cases (or calculation procedures) the higher the estimated Q value, the higher the error in its determination, and this always corresponds to the highest applied field. This is a direct consequence of the sharpness of the measured resonance curves, as well as of its low amplitude (magnetic susceptibility), as can be seen in Figure 5a. As predicted previously by other authors [20], this estimated error (if the classical Q0 definition is used) can be as high as 20%. Surprisingly, the obtained norm of the fitting for this high applied field case is the lowest, this is, the numerical fits are the best for this case.

Table 6.

Q values for all the applied magnetic field cases, obtained by the different procedures shown in this study.

H (Oe) Q0=frΔf Q1frfm'fM' Qfit1 Qfit2 Qcal=fm2+fM2fm2fM2
0.6 333 290 208 283 290
2 285 243 178 229 243
16 1762 1410 1067 1321 1410

Table 7.

Comparison and estimated errors of the Q values obtained with the different calculation procedures.

H (Oe) Error (%) Respect to Q1or Qcal
Q0 Qfit1 Qfit2
0.6 13 28 2.4
2 15 27 5.7
16 20 24 6.3

We also found that the values obtained for Q1 (the value given by Kaczkowski [20]) and Qcal (exact analytical expression) are almost equal. While Kaczkowski’s expression was an approximation obtained graphically from the impedance circle of an electrical circuit, our exact formula comes from the analysis of the real part of the magnetic susceptibility around the magnetoelastic resonance.

So, if we only have the possibility of measuring or working with the magnetic susceptibility modulus, numerical fitting of the measured magnetoelastic resonance curve is needed. From the results of the numerical fittings used (as can be observed in Figure 7 and Figure 8 and Table 7), it is clear that the second procedure (leaving all parameters fr, fa and χ0 free) leads to a much better result than the first one, as deduced from the obtained lowest norm values for procedure b. That is, we can affirm that the Qfit2 value obtained by using fitting procedure b can be considered the best approximation of the true Q quality factor of the magnetoelastic resonance curve.

On the other hand, if we compare the Q values obtained from numerical fits with the approximated Q1 value given by Kaczkowski or with the Qcal value, the estimated error when using fitting procedure a is always higher than 20%, while for fitting (procedure b) the range is only approximately 26%. Thus, Q1 or Qcal can be taken as approximated initial values when performing a numerical fit in order to get the most accurate Q value of a susceptibility magnetoelastic resonance curve.

It is also noticeable that Q values obtained with fitting (procedure b) are systematically lower than Q1 or Qcal, but this is a fact that should be expected: Equation (9) gives us the exact quality factor Q of the real part of the magnetic susceptibility around the frequency at which a magnetoelastic resonance happens, with this real part being a sharper curve than the corresponding measured susceptibility modulus. As is already well known, the sharper the curve, the higher the quality factor value. Finally, from the obtained error values, we can affirm that when using a magnetoelastic resonant platform for biological or chemical detection purposes, it is convenient to apply a bias field in the range 0<Hbias<Hk, searching for a compromise between moderate magnetoelastic coupling and low enough error in Q value determination.

5. Conclusions

We have presented an extensive study of the determination of the Q factor of a magnetoelastic resonance curve. This type of resonance is of great interest in order to fabricate devices for biological or chemical detection purposes. The use of the numerical fitting of the magnetic susceptibility modulus around that magnetoelastic resonance turns out to be a useful tool to give accurate Q quality factor values. These differ by up to 20% compared with the Q values determined by following the classical definition. Comparison with approximated Q value given by Kaczkowski and by the exact analytical solution obtained from the real part of the measured susceptibility shows, as expected, that in these two last cases the quality factor value obtained is always slightly higher than that estimated from the numerical fitting. This is a direct consequence of the fact that, while the numerical fit is performed over the magnetic susceptibility modulus, Kaczkowski’s and the exact expression for have been obtained from the real part of that susceptibility curve, which is always sharper than the susceptibility modulus one.

Future work should aim to obtain an analytical expression for the Q quality factor directly obtained from the magnetic susceptibility modulus measured around the magnetoelastic resonance.

Acknowledgments

Ana Catarina Lopes thanks MSCA-IF-2015 (Marie Skłodowska Curie Actions) of the European Union’s Horizon 2020 Programme for the received funds under grant agreement no. [701852]. Ariane Sagasti wishes to thank BCMaterials Centre for financial support. Jon Gutierrez, Andoni Lasheras, and José Manuel Barandiarán would like to acknowledge the financial support from the Basque Government under the ACTIMAT project (Etortek 2018 program) and Research Groups IT711-13 project. Dimitris Kouzoudis is thankful for financial support under Erasmus + Mobility Agreement between the University of Patras and the University of the Basque Country (UPV/EHU). Technical and human support provided by SGIker (UPV/EHU, MICINN, GV/EJ, ESF) is gratefully acknowledged.

Author Contributions

A.C. Lopes, A. Lasheras, J. Gutiérrez and J.M. Barandiarán conceived and designed the work and needed measurements; A.C. Lopes, A. Sagasti and A. Lasheras performed the experiments; A.C. Lopes and V. Muto performed the numerical calculations; A. Sagasti, J. Gutiérrez, D. Kouzoudis and J.M. Barandiarán analysed the data; A. Sagasti, A. Lasheras, J. Gutiérrez, D. Kouzoudis and J.M. Barandiarán wrote the manuscript. All authors discussed the results and implications, and commented on the manuscript at all stages. All authors read and approved the final manuscript.

Conflicts of Interest

The authors declare no conflict of interest.

References

  • 1.Grimes C.A., Seitz W.R., Horn J., Doherty S.A., Rooney M.T. A remotely interrogatable magnetochemical pH sensor. IEEE Trans. Magn. 1997;33:3412–3414. doi: 10.1109/20.617961. [DOI] [Google Scholar]
  • 2.Stoyanov P.G., Doherty S.A., Grimes C.A., Seitz W.R. A remotely interrogatable sensor for chemical monitoring. IEEE Trans. Magn. 1998;34:1315–1317. doi: 10.1109/20.706533. [DOI] [PubMed] [Google Scholar]
  • 3.Grimes C.A., Mungle C.S., Zeng K., Jain M.K., Dreschel W.R., Paulose M., Ong G.K. Wireless magnetoelastic resonance sensors: A critical review. Sensors. 2002;2:294–313. doi: 10.3390/s20700294. [DOI] [Google Scholar]
  • 4.Bouropoulos N., Kouzoudis D., Grimes C.A. The real-time, in situ monitoring of calcium oxalate and brushite precipitation using magnetoelastic sensors. Sens. Actuators B. 2005;109:227–232. doi: 10.1016/j.snb.2004.12.054. [DOI] [Google Scholar]
  • 5.Grimes C.A., Kouzoudis D., Dickey E.C., Kiang D., Anderson M.A., Shahidain R., Lindsey M., Green L. Magnetoelastic sensors in combination with nanometer-scale honeycombed thin film ceramic TiO2 for remote query measurement of humidity. J. Appl. Phys. 2000;87:5341–5343. doi: 10.1063/1.373341. [DOI] [PubMed] [Google Scholar]
  • 6.Cai Q.Y., Cammers-Goodwin A., Grimes C.A. A wireless, remote query magnetoelastic CO2 sensor. J. Environ. Monit. 2000;2:556–560. doi: 10.1039/b004929h. [DOI] [PubMed] [Google Scholar]
  • 7.Baimpos T., Gora L., Nikolakis V., Kouzoudis D. Selective detection of hazardous VOCs using zeolite/Metglas composite sensors. Sens. Actuators A. 2012;186:21–31. doi: 10.1016/j.sna.2011.12.030. [DOI] [Google Scholar]
  • 8.Lakshmanan R.S., Guntupalli R., Hu J., Kim D.J., Petrenko V.A., Barbaree J.M., Chin B.A. Phage immobilized magnetoelastic sensor for the detection of Salmonella typhimurium. J. Microbiol. Methods. 2007;71:55–60. doi: 10.1016/j.mimet.2007.07.012. [DOI] [PubMed] [Google Scholar]
  • 9.Huang S., Yang H., Lakshmanan R.S., Johnson M.L., Wan J., Chen I.H., Wikle H.C., III, Petrenko V.A., Barbaree J.M., Chin B.A. Sequential detection of Salmonella typhimurium and Bacillus anthracis spores using magnetoelastic biosensors. Biosens. Bioelectron. 2009;24:1730–1736. doi: 10.1016/j.bios.2008.09.006. [DOI] [PubMed] [Google Scholar]
  • 10.Ruan C., Zeng K., Varghese O.K., Grimes C.A. Magnetoelastic immunosensors: Amplified mass immunosorbent assay for detection of Escherichia coli O157:H7. Anal. Chem. 2003;75:6494–6498. doi: 10.1021/ac034562n. [DOI] [PubMed] [Google Scholar]
  • 11.Luborsky F.E. Chapter 6: Amorphous ferromagnets in Ferromagnetic Materials. In: Wohlfart E.P., editor. Handbook of Ferromagnetic Materials. Volume 1. Elsevier; Amsterdam, The Netherlands: 1980. [Google Scholar]
  • 12.Stoyanov P.G., Grimes C.A. A remote query magnetostrictive viscosity sensor. Sens. Actuators A. 2000;80:8–14. doi: 10.1016/S0924-4247(99)00288-5. [DOI] [PubMed] [Google Scholar]
  • 13.Sagasti A., Gutiérrez J., Sebastián M.S., Barandiarán J.M. Magnetoelastic resonators for highly specific chemical and biological detection: A critical study. IEEE Trans. Magn. 2017;53:4000604. doi: 10.1109/TMAG.2016.2622359. [DOI] [Google Scholar]
  • 14.Sagasti A. Ph.D. Thesis. University of the Basque Country (UPV/EHU); Leioa, Spain: 2018. Functionalized Magnetoelastic Resonant Platforms for Chemical and Biological Detection Purposes. unpublished. [Google Scholar]
  • 15.Grimes C.A., Kouzoudis D., Ong K.G., Crump R. Thin-film magnetoelastic microsensors for remote query biomedical monitoring. Biomed. Microdevices. 1999;2:51–60. doi: 10.1023/A:1009907316867. [DOI] [Google Scholar]
  • 16.Bravo-Ímaz I., García-Arribas A., Gorritxategi E., Arnaiz A., Barandiarán J.M. Magnetoelastic viscosity sensor for on-line status assessment of lubricant oils. IEEE Trans. Magn. 2013;49:113–116. doi: 10.1109/TMAG.2012.2217117. [DOI] [Google Scholar]
  • 17.Sagasti A., Bouropoulos N., Kouzoudis D., Panagiotopoulos A., Topoglidis E., Gutierrez J. Nanostructured ZnO in Metglas/ZnO/Hemoglobin modified electrode to detect the oxidation of the hemoglobin simultaneously by cyclic voltammetry and magnetoelastic resonance. Materials. 2017;10:849. doi: 10.3390/ma10080849. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Peterson P.J., Anlage S.M. Measurement of resonant frequency and quality factor of microwave resonators: Comparison of methods. J. Appl. Phys. 1998;84:3392–3402. doi: 10.1063/1.368498. [DOI] [Google Scholar]
  • 19.Cory D., Hutchinson I., Chaniotakis M. Introduction to Electronics, Signals and Measurements. Springer; Berlin, Germany: 2006. Chapter 17. [Google Scholar]
  • 20.Kaczkowski Z. Piezomagnetic dynamics as a new parameter of magnetostrictive materials and transducers. Bull. Pol. Acad. Sci. 1997;45:19–42. [Google Scholar]
  • 21.Du Tremolet de Laichesserie E. Magnetostriction: Theory and Application of Magnetoelasticity. CRC Press; Boca Raton, FL, USA: 1993. [Google Scholar]
  • 22.Gutiérrez J., Barandiarán J.M., Nielsen O.V. Magnetoelastic properties of some Fe-rich Fe-Co-Si-B metallic glasses. Phys. Status Solidi A. 1989;111:279–283. doi: 10.1002/pssa.2211110129. [DOI] [Google Scholar]
  • 23.Gutiérrez J. Ph.D. Thesis. University of the Basque Country (UPV/EHU); Leioa, Spain: 1992. Propiedades Magnéticas y Magnetoelásticas de Nuevas Aleaciones Amorfas de Interés Tecnológico. [Google Scholar]
  • 24.Landau L.D., Lifshitz E.M. Theory of Elasticity. Oxford Pergamon Press; Oxford, UK: 1975. p. 116. [Google Scholar]
  • 25.Savage H., Abbundi R. Perpendicular susceptibility, magnetomechanical coupling and shear modulus in Tb0.27Dy0.73Fe2. IEEE Trans. Magn. 1978;14:545–547. doi: 10.1109/TMAG.1978.1059991. [DOI] [Google Scholar]
  • 26.Hernando A., Madurga V., Barandiarán J.M., Liniers M. Anomalous eddy currents in magnetostrictive amorphous ferromagnets: A large contribution from magnetoelastic effects. J. Magn. Magn. Mater. 1982;28:109–116. doi: 10.1016/0304-8853(82)90034-8. [DOI] [Google Scholar]

Articles from Sensors (Basel, Switzerland) are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES