Abstract
Patient: Female, 65
Final Diagnosis: Clavicle Kirschner wire migration into left lung
Symptoms: No symptoms
Medication: —
Clinical Procedure: Thoracotomy
Specialty: Surgery
Objective:
Diagnostic/therapeutic accidents
Background:
Kirschner wires are often used to perform osteosynthesis. Migration through tissue of these wires is a rare but well-known occurrence.
Case Report:
A 65-year-old female presented with light intensity pain complaints in the upper left chest area; personal history included left clavicle fracture 20 years ago that was treated surgically with fixation using a K-wire. Chest radiography showed the presence of metallic foreign body in the left pulmonary apex. An exploratory axillary thoracotomy was performed, and the foreign body was extracted by a pneumotomy.
Conclusions:
To obtain satisfactory results with a K-wire, some peculiarities in their application should be respected. The time from orthopedic surgery of the collarbone to migration into the chest of the metal rod used can vary from one day to nearly 20 years. Although the migration mechanism remains unclear, it is likely that it involves shoulder movements, breathing movements, negative intrathoracic pressure, gravitational force, or local bone resorption. Caution should be exercised when orthopedic pins and wires are used for the fixation of fractures and dislocations of the shoulder girdle. If there is migration of the wire, it should be removed immediately to avoid sudden and fatal complications.
MeSH Keywords: Clavicle, Device Removal, Lung, Risk Management, Thoracic Surgery
Background
Clavicle fractures and sternum-clavicular dislocations are common after traumatic chest injuries. In some cases, orthopedic surgeons need to use metal rods to fix and stabilize these fractures [1].
Kirschner wire (K-wire) was developed at the beginning of the nineteenth century [2]; now it is commonly used for stabilization of complex fractures [3] to perform osteosynthesis.
Migration through tissue of these wires is a rare but well-known occurrence [1,3,4]. Several significant complications of clavicle fixation have been previously reported such as migration of K-wires into the chest cavity that could cause spinal cord, esophagus, lung, trachea, heart, pericardium, or great vessels drilling [1,4–8].
Case Report
Twenty years ago, a healthy 45-year-old female suffered a traumatic left mid-shaft clavicle fracture when she had an accident with a motorcycle. At that time, she was treated in a rural hospital then referred to an institution with orthopedic services. The treatment of the clavicle fracture was surgical with fixation using a K-wire. The correct position of the wire was documented by a postoperative chest radiograph. After surgery, the patient continued with her normal activities and did not require a follow-up appointment as she had no complaints.
In May 2013, she started complaining of light intensity pain in the upper left chest area that was mainly associated with movements of the upper ipsilateral limb in the last 6 months with worsening in the last month. She had no other symptoms and no limitation of movement or impairment of daily activities.
Chest radiography showed the presence of a metallic foreign body in the left pulmonary apex (Figure 1). Radiography was thus requested, and completion of the imaging study with computed tomography revealed the presence of a metallic linear radiodense artifact tangential to the apex of the left upper lobe, measuring about 6.6 cm along its largest longitudinal axis (Figure 2).
Figure 1.
Anteroposterior (A) and lateral (B) chest radiograph revealing the presence of the foreign body in the left lung apex region.
Figure 2.

Computed tomography of the chest, revealing metallic artifact in the left upper lobe apex.
An exploratory axillary thoracotomy was performed. After selective left bronchial intubation, the patient was placed in the right lateral decubitus position. One incision was made, in the 5th intercostal space at the anterior axillary line intersection. With the aid of a ring clamp, the foreign body was extracted by a pneumotomy (Figure 3). Hemostasis and an air leak test were performed. A chest drain was connected to a pleural drainage system. The patient had no complications after surgery and was discharged on the 4th day after surgery. She remained in outpatient follow-up for 2 years.
Figure 3.

The K-wire removed from the upper left lobe.
Discussion
Late migration of K-wine after surgeries has been reported in the literature [3,9–13] and the time from orthopedic surgery of the collarbone to migration into the chest of the metal rod used can vary from one day to nearly 30 years [1,3,4].
The migration mechanism remains unclear [14]. The possible reasons include muscular activity, movement of the shoulder, breathing movements, negative intrathoracic pressure with respiratory excursion, gravitational force, local bone resorption, [4] capillary action, [1] and even peripheral intravascular embolization to the heart [3].
The successful removal of intrathoracic K-wires using different approaches has been reported. [3,15]. The location of the K-wire and the clinical condition of the patient are the criteria used to choose the most appropriate technique; thoracoscopy is the preferred approach [3].
To obtain satisfactory results with a K-wire, some peculiarities in their application should be respected. The surgeon should accommodate local anatomy, and as a precaution, the insert should be made from the anatomical site of greatest risk and in the opposite direction [16]. When these locations cannot be avoided, it is essential to palpate the artery and insert the wire at a minimum distance of 2 cm from it. Note also that extreme care must be taken to avoid sliding the thread on the bone and to avoid damage to the surrounding structures [16,17].
Therefore, caution should be exercised when orthopedic pins and wires are used for the fixation of fractures and dislocations of the shoulder girdle. Care should also be taken to perform a fold at the end of the wire or use a restraint system.
The most important step in the prevention of this potentially lethal complication is to double the exposed portion of the wire or pin after fixation through use of holding devices, to perform rigorous clinical and radiography follow-up every 2–4 weeks [18], K-wire removal as soon as bone healing is achieved, and bending the external tip of each implanted wire [14]. Regardless, once the migration has been diagnosed, the wire should be removed immediately [4].
Nevertheless, according to Tsai et al. [12], physicians should be aware of the possibility of late migration of threaded wires and carefully instruct patients about the risks and the importance of returning for follow-up evaluations, even years later. Risk management allows early detection of incidents and contributes for patient safety.
Conclusions
Exploratory axillary thoracotomy is an effective treatment to remove a K-wire which migrated into left lung when the clinical condition of the patient is stable. Although the mechanism of late and silent migration is unclear, physicians should advice the patient about the risk and the need of clinical and radiographic follow-up, even years later after the insertion of the device, in order to allow for early detection of possible harms. If there is migration of the wire, it should be removed immediately to avoid sudden and fatal complications.
Acknowledgments
The authors would like to thank the Foundation for teaching support, research and assistance (Fundação De Apoio Ao Ensino, Pesquisa E Assistência – FAEPA) for the financial grant toward fees incurred for the publication of the paper.
Footnotes
Conflict of interest
None.
References:
- 1.Lyons FA, Rockwood CA. Migration of pins used in operations on the shoulder. J Bone Joint Surg Am. 1990;72(8):1262–67. [PubMed] [Google Scholar]
- 2.Franssen BBGM, Schuurman AH, Van der Molen AM, Kon M. One century of Kirschner wires and Kirschner wire insertion techniques: A historical review. Acta Orthop Belg. 2010;76(1):1–6. [PubMed] [Google Scholar]
- 3.Cortés-Julián G, Mier JM, Briseño C. Thoracoscopic excision of migrated Kirschner wire to right pulmonary hilum. Ann Thorac Surg. 2015;100(4):1461–63. doi: 10.1016/j.athoracsur.2014.12.102. [DOI] [PubMed] [Google Scholar]
- 4.Ballas R, Bonnel F. Endopelvic migration of a sternoclavicular K-wire. Case report and review of literature. Orthop Traumatol Surg Res. 2012;98(1):118–21. doi: 10.1016/j.otsr.2011.09.015. [DOI] [PubMed] [Google Scholar]
- 5.Bennis S, Scarone P, Lepeintre J-F, et al. Asymptomatic spinal canal migration of clavicular K-wire at the cervicothoracic junction. Orthopedics. 2008;31(12) doi: 10.3928/01477447-20081201-03. pii: orthosupersite.com/view.asp?rID=32939. [DOI] [PubMed] [Google Scholar]
- 6.Cameliere L, Rosat P, Heyndrickx M, et al. Migration of a Kirschner pin from the shoulder to the lung, requiring surgery. Asian Cardiovasc Thorac Ann. 2013;21(2):222–23. doi: 10.1177/0218492312450021. [DOI] [PubMed] [Google Scholar]
- 7.Nakayama M, Gika M, Fukuda H, et al. Migration of a Kirschner wire from the clavicle into the intrathoracic trachea. Ann Thorac Surg. 2009;88(2):653–54. doi: 10.1016/j.athoracsur.2008.12.093. [DOI] [PubMed] [Google Scholar]
- 8.Kumar P, Godbole R, Rees GM, Sarkar P. Intrathoracic migration of a Kirschner wire. J R Soc Med. 2002;95(4):198–99. doi: 10.1177/014107680209500409. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Suzuki T, Matsumura N, Iwamoto T, Sato K. Migration of a Kirschner wire into the lung with shoulder dislocation. BMJ Case Rep. 2017;2017 doi: 10.1136/bcr-2017-221850. pii: bcr-2017-221850. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Mankowski B, Polchlopek T, Strojny M, et al. Intraspinal migration of a Kirschner wire as a late complication of acromioclavicular joint repair: A case report. J Med Case Rep. 2016;10:66. doi: 10.1186/s13256-016-0844-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Hédon C, Khoueiry Z, Verges M, Pasquié J-L. Late intracardiac orthopedic wire migration presenting as tamponade and stroke. Eur Heart J. 2015;36(24):1546. doi: 10.1093/eurheartj/ehu416. [DOI] [PubMed] [Google Scholar]
- 12.Tsai C-H, Hsu H-C, Huan C-Y, et al. Late migration of threaded wire (schanz screw) from right distal clavicle to the cervical spine. J Chin Med Assoc. 2009;72(1):48–51. doi: 10.1016/S1726-4901(09)70021-8. [DOI] [PubMed] [Google Scholar]
- 13.Stevenson KL, Wetzel M, Pollack IF. Delayed intracranial migration of cervical sublaminar and interspinous wires and subsequent cerebellar abscess. Case report. J Neurosurg. 2002;97(1 Suppl.):113–17. doi: 10.3171/spi.2002.97.1.0113. [DOI] [PubMed] [Google Scholar]
- 14.Mamane W, Breitel D, Lenoir T, Guigui P. [Spinal migration of a Kirschner wire after surgery for clavicular nonunion. A case report and review of the literature] Chir Main. 2009;28(6):367–69. doi: 10.1016/j.main.2009.08.007. [in French] [DOI] [PubMed] [Google Scholar]
- 15.Park S-Y, Kang J-W, Yang DH, Lim T-H. Intracardiac migration of a Kirschner wire: Case report and literature review. Int J Cardiovasc Imaging. 2011;27(Suppl. 1):85–88. doi: 10.1007/s10554-011-9977-0. [DOI] [PubMed] [Google Scholar]
- 16.Zhou S, Hao Y, Shi X, et al. [Design and clinical application of the drilling guide in the treatment of acromioclavicular joint dislocation with closed reduction and Kirschner fixation] Zhongguo Gu Shang. 2011;24(3):186–88. [in Chinese] [PubMed] [Google Scholar]
- 17.Bragin VB, Bezgodkov IA. [Comparative evaluation of the methods of treatment of dislocations of the clavicle] Vestn Khir Im I I Grek. 2002;161(4):33–36. [in Russian] [PubMed] [Google Scholar]
- 18.Hegemann S, Kleining R, Schindler HG, Holthusen H. [Kirschner wire migration in the contralateral lung after osteosynthesis of a clavicular fracture] Unfallchirurg. 2005;108(11):991–93. doi: 10.1007/s00113-005-0946-8. [in German. [DOI] [PubMed] [Google Scholar]