Scrub typhus point-of-care testing: A systematic review and meta-analysis

Kartika Saraswati 1,2,3, Nicholas P J Day 1,3, Mavuto Mukaka 1,3, Stuart D Blacksell 1,3,*
Editor: Ana LTO Nascimento4
PMCID: PMC5892940  PMID: 29579046

Abstract

Background

Diagnosing scrub typhus clinically is difficult, hence laboratory tests play a very important role in diagnosis. As performing sophisticated laboratory tests in resource-limited settings is not feasible, accurate point-of-care testing (POCT) for scrub typhus diagnosis would be invaluable for patient diagnosis and management. Here we summarise the existing evidence on the accuracy of scrub typhus POCTs to inform clinical practitioners in resource-limited settings of their diagnostic value.

Methodology/principal findings

Studies on POCTs which can be feasibly deployed in primary health care or outpatient settings were included. Thirty-one studies were identified through PubMed and manual searches of reference lists. The quality of the studies was assessed with the Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS-2). About half (n = 14/31) of the included studies were of moderate quality. Meta-analysis showed the pooled sensitivity and specificity of commercially available immunochromatographic tests (ICTs) were 66.0% (95% CI 0.37–0.86) and 92.0% (95% CI 0.83–0.97), respectively. There was a significant and high degree of heterogeneity between the studies (I2 value = 97.48%, 95% CI 96.71–98.24 for sensitivity and I2 value = 98.17%, 95% CI 97.67–98.67 for specificity). Significant heterogeneity was observed for total number of samples between studies (p = 0.01), study design (whether using case-control design or not, p = 0.01), blinding during index test interpretation (p = 0.02), and QUADAS-2 score (p = 0.01).

Conclusions/significance

There was significant heterogeneity between the scrub typhus POCT diagnostic accuracy studies examined. Overall, the commercially available scrub typhus ICTs demonstrated better performance when ‘ruling in’ the diagnosis. There is a need for standardised methods and reporting of diagnostic accuracy to decrease between-study heterogeneity and increase comparability among study results, as well as development of an affordable and accurate antigen-based POCT to tackle the inherent weaknesses associated with serological testing.

Author summary

Scrub typhus is a disease caused by a mite-transmitted bacterium, Orientia tsutsugamushi. Aside from fever and necrotic skin lesion at the site of mite inoculation (eschar), scrub typhus has virtually no distinctive clinical features, making diagnosis without laboratory tests difficult. However, the laboratory tests often have limited diagnostic accuracy and are generally in limited supply in resource-limited or outpatient settings. In those settings, point-of-care testing (POCT) can be useful in making a quick diagnosis. In this study, the authors reviewed the available evidence on the diagnostic value of scrub typhus POCT. A meta-analysis, focusing on commercially available immunochromatography tests (ICTs), was done. The pooled sensitivity and specificity were 66.0% (95% CI 0.37–0.86) and 92.0% (95% CI 0.83–0.97), respectively. This indicates better performance when ‘ruling in’ a positive diagnosis rather than ‘ruling out’ a negative diagnosis. As ICTs have low false positive rates, confidence in interpreting a positive result can be high. The studies reviewed displayed wide variation in both methodology and quality, indicating a need to develop standards for study methodology and reporting.

Introduction

Scrub typhus is a febrile illness caused by the obligate intracellular bacterium, Orientia tsutsugamushi. It is transmitted by the bite of infected larvae of a number of trombiculid mite species known to be prevalent in Asia, the Pacific Rim islands, pockets in the north of Australia, and some areas of Chile [14]. In 2010, a novel species from the same genus, Orientia chuto sp. nov., was identified in an acutely febrile patient infected in Dubai [5]. Scrub typhus responds to certain antibiotics (i.e. doxycycline, tetracycline, azithromycin, chloramphenicol), but if left untreated, the mortality rate may reach 70% [6]. One estimate, based on scant data, is that there are one billion people at risk of this disease; with one million clinical cases annually in Southeast Asia alone [1]. Although the exact prevalence of scrub typhus is not available, several studies showed that the disease burden in rural Asia is high–causing in some areas over 20.0% of febrile illness admitted to hospital [7,8].

Infected patients usually present with acute fever; lymphadenopathy (regional or generalised) and sensorineural hearing loss may occur, neither of which is sensitive or specific enough for establishing diagnosis [1]. With few distinguishing clinical characteristics, scrub typhus is difficult to differentiate from other tropical febrile illnesses, such as dengue, typhoid fever, leptospirosis, and murine typhus [1,9]. The presence of the pathognomonic eschar, the painless black crust at the site of mite inoculation, can help in establishing clinical diagnosis due to its high specificity (98.9%), however, its presence in patients varies widely (7.0%-97.0%) [1,1012].

Therefore, the role of laboratory tests in establishing diagnosis in scrub typhus cases is very important. Laboratory tests for scrub typhus often have limited diagnostic accuracy and are generally in limited supply in resource-limited or outpatient settings [9]. Failure in diagnosing scrub typhus may result in prolonged illness, complications including pneumonitis, acute respiratory distress syndrome, renal failure, meningoencephalitis, and unnecessary treatment with inappropriate antibiotics [1,9,13].

Serology remains the mainstay of diagnosis. The immunofluorescent assay (IFA) and immunoperoxidase test (IIP) are considered imperfect gold standards, in view of their limitations which include high expense, requirement for substantial training to perform, inter-operator variability in result interpretation, and the often-retrospective nature of diagnosis that does not help in directing treatment [1,9,12,14,15]. Another antibody detection method, the enzyme-linked immunosorbent assay (ELISA) has been developed and shown to have both sensitivity and specificity of greater than 90.0%; however this is highly dependent on endemicity and the application of a previously investigated and geographically-based cut-off [1,16]. Besides antibody-based diagnostics, molecular detection methods, including using the polymerase chain reaction (PCR) to detect various genes targets (e.g. 47 kDa, 56 kDa, groEL, 16S rRNA genes) have also been developed, however they have limitations in terms of diagnostic sensitivity due to the limited period of rickettsaemia [1,9,12,14]. PCR is still deemed impractical in resource-limited endemic areas because it requires considerable training and expense [1,9,12,14]. The bacteria can be isolated through in vitro and in vivo cultivation methods, such as cell culture and mouse inoculation, respectively [1,9]. These methods need considerable training, biosafety level 3 (BSL 3) laboratory containment facilities for large-scale propagation, and usually take several weeks which contributed to the retrospective nature of the diagnosis [1,9].

Therefore, there is clearly a need for affordable point-of-care testing (POCT) for scrub typhus diagnosis in endemic settings with resource constraints. There are varied definitions of POCT, but fundamentally POCT should provide quick results to inform patient management and be convenient enough to be performed close to the patient (i.e. not in a central laboratory) [17,18].

Immunochromatographic tests (ICTs), dot-blot, and loop-mediated isothermal amplification (LAMP) assays all have the principal qualities of POCT. ICTs and dot-blot tests have the same inherent problems of IFA as serology-based tests (e.g. the retrospective nature of diagnosis in cases where diagnosis relies on a convalescent sample, delicate cut-off setting), while offering more simplicity and speed [1,9]. LAMP is an alternative technique which involves amplification and detection of bacterial DNA. Similar in principle to conventional PCR assays, LAMP assay does not require intricate DNA extraction, a thermocycler, or special equipment to read the result [9,19,20].

This study aims to summarise the existing evidence on the accuracy of scrub typhus POCTs to inform clinical practitioners of their diagnostic value when providing care in resource-limited settings where scrub typhus is endemic.

Methods

Eligibility criteria

This review included articles on POCTs that would be feasible in primary health care provider or outpatient settings. Only articles published in English were included. To ensure feasibility in resource-limited settings, studies evaluating methods which were inherently more complicated, requiring relatively high levels of expertise and/or specialised equipment were excluded. Articles on POCTs not performed on human samples were excluded. Studies on the Weil-Felix test were excluded due to its established poor diagnostic accuracy and the lengthy time required to perform [9,21]. Meta-analysis and meta-regression were performed on studies of commercially available POCTs with an extractable diagnostic accuracy 2 by 2 table.

Search strategy

After a preliminary search, ICT, dot-blot, and LAMP were searched for specifically. The search was conducted on articles cited in PubMed up to 2 February 2017 combining the search terms ‘scrub typhus’, ‘immunochromatography’, ‘dot blot immunoassay’, and ‘loop mediated isothermal amplification’ without any other restrictions (i.e., "scrub typhus" AND (rapid diagnosis OR immunochromatograph* OR dot blot immunoassay OR loop mediated isothermal amplification). The titles and abstracts were screened and the full text of relevant articles were reviewed. Manual screening of the reference list of relevant articles was also performed.

Quality assessment

The quality of the studies was assessed with the Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS-2) [22]. QUADAS-2 was used as a quality scoring system to determine the risk of bias and the applicability of the paper [22]. It evaluates four main areas: ‘patient selection’, ‘index test’, ‘reference standard’, and ‘flow and timing’ [22]. These are assessed by using seven ‘signalling questions’ (e.g., “was a case-control design avoided?”) with ‘yes’, ‘no’, and ‘unclear’ answer [22]. The answers to these ‘signalling questions’ were then used to judge whether the risk of bias is low and if there is low concern for the applicability of the research [22]. If the response to the risk of bias and applicability questions were ‘low risk’ or ‘low concern’, the articles were given one point each. The articles were grouped based on their score into high (6–7 points), moderate (4–5 points), and low (0–3 points) quality categories.

Data extraction

Data was extracted primarily by one author (KS) and where the results were unclear a second author (SB) was consulted. The data was recorded on a form developed through an iterative process to ensure that all the required data could be collected for future reference. The parameters extracted include: citation information, methodology (i.e., study design, participant characteristics, index and reference test details), and the diagnostic accuracy results (including sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and if available, numbers required to construct a 2 by 2 contingency table.

Statistical data analysis and reporting

The extracted data were compiled into summary tables and analysed through narrative synthesis. Meta-analysis and meta-regression were performed on commercially available POCT diagnostic accuracy data, excluding studies with low quality (i.e. QUADAS-2 score of 3 or less). The ones in development stage/prototype were not included in the meta-analysis and meta-regression, but included in the narrative synthesis. If one study derived more than one 2 by 2 table, each table was extracted as separate data. However, if one study used more than one reference test cut-off titre, only data using one cut-off value above 1:3,200 were used to ensure accuracy [12]. In performing the meta-regression, relevant signalling questions with ‘unclear’ as the answers were entered as ‘no’ to turn these into dichotomous variables. Statistical analysis was done with STATA/IC 14.0 (College Station, TX) using MIDAS and METANDI commands. In the meta-analyses, heterogeneity was assessed using the Chi-square statistic, higher values of the Chi-square (and hence low p-value) being consistent with heterogeneity. In the summary statistics of the resulting forest plot, overall sensitivity and specificity were estimated and reported alongside the 95% confidence interval. In addition, study specific estimates were provided in the same plot to visualize how the estimates from each of the studies deviate from the overall estimate. Most of the results was presented graphically. The data was analysed, summarised, and presented following the Preferred Reporting Items for Systematic reviews and Meta-Analyses (PRISMA) statement as much as possible [23]. This review was registered in the International Prospective Register for Systematic Review (PROSPERO) with registration number CRD42017056727.

Results

Search results

There were 133 articles in total identified through database searching and reference list screening (Fig 1). After title and abstract screening and full text review, we included 31 relevant articles. There were six articles excluded after full text review. Since this study only focused on human diagnostics, one article using rabbit sera in its negative sera panel was excluded. One study involved DNA extraction, which is not applicable as a POCT in resource-limited settings. Four other papers were excluded due to language (not in English, n = 2) and study design (not experimental diagnostic accuracy studies, n = 2). There were 20 articles on ICTs, eight articles on dot-blot assays, four articles on LAMP assays, and one each on a passive hemagglutination assay, an IgM dot immunobinding assay, and a latex agglutination test. Four articles evaluated more than one type of diagnostic test. There were 11 studies that evaluated diagnostic tests still in development and 21 studies on prototype/commercial tests.

Fig 1. PRISMA flow diagram.

Fig 1

Characteristics of the included studies

In total, there were 6,772 samples analysed. The samples were taken from 12 countries, with most studies recruiting in Thailand (n = 15, 48.4% of included studies), India (n = 5, 16.1%), Laos (n = 3, 9.7%), and Korea (n = 3, 9.7%). There was one study each (3.2%) conducted on samples from Sri Lanka, Nepal, Malaysia, Peru, Indonesia, United States of America, and Australia. There were five studies (16.1%) with unclear sample collection location. Ten (32.2%) studies collected paired samples (acute and convalescent phase), while 15 studies (48.4%) did not provide sufficient details on sample collection timing (Table 1).

Table 1. Summary of included studies.

First author Year Sample collection location Index test assay type Reference test assay type QUADAS a 2 score
Anitharaj et al [33] 2016 India ICT IgM ELISA 5
Blacksell et al [34] 2010 Laos ICT IgM IFA 6
Blacksell et al [28] 2010 Laos and Thailand ICT IgM; ICT IgG, IgM IFA and/or PCR and/or culture 5
Blacksell et al [19] 2012 Thailand ICT IgM; ICT IgG, IgM; LAMP STIC (cell culture isolation, PCR, IFA) 6
Cao et al [25] 2007 China ICT IgG, IgM; ICT IgG; ICT IgM Unclear 2
Ching et al [24] 2001 Unclear ICT IgG, IgM; ICT IgG; ICT IgM IFA 2
Chinprasatsak et al [35] 2001 Thailand Dot-blot IgG, IgM IIP 6
Coleman et al [36] 2002 Thailand Dot-blot IgG, dot-blot IgM, ICT IgG, ICT IgM IIP 7
Huber et al [37] 2012 Unclear LAMP PCR 2
Kim et al [38] 1993 Unclear Passive hemagglutination assay IFA 3
Kim et al [39] 2013 Korea ICT IgG, IgM; dot-blot IgG Unclear 2
Kim et al [40] 2016 Korea, Sri Lanka, India ICT IgG, IgM IFA 3
Kingston et al [41] 2015 Thailand, Nepal ICT IgM PAb, Mab IFA 6
Koay et al [42] 1995 Unclear IgM dot-immunobinding assay IIP 3
Lee et al [27] 2014 Korea ICT IgG, IgM, IgA IFA 4
Paris et al [43] 2008 Thailand, Laos LAMP In vitro isolates, IFA, PCR, ICT IgG & IgM 3
Paris et al [20] 2011 Thailand ICT IgM, LAMP STIC (Cell culture isolation, PCR, IFA) 6
Pradutkanchana et al [44] 1997 Thailand Dot-blot IgG, IgM IFA 4
Prakash et al [45] 2006 India Dot-blot IgG, IgM Dot EIA, ELISA, clinical response to antibioticsa 3
Ramyasree et al [46] 2015 Unclear ICT IgM ELISA 5
Rodkvamtook et al [47] 2015 Thailand Dot-blot IgG, IgM IFA 5
Silpasakorn et al [48] 2012 Thailand ICT IgM, ICT IgG PCR & IFA 4
Silpasakorn et al [49] 2012 Thailand ICT IgG, IgM, IgA PCR & IFA 3
Stephen et al [50] 2015 India ICT IgG, IgM, IgA ELISA IgM, IgG 5
Stephen et al [51] 2016 India ICT IgG, ICT IgM IFA 5
Watt et al [52] 1998 Thailand Dot-blot IgG, IgM IFA 5
Watthanaworawit et al [29] 2015 Thailand ICT IgM; IgG, IgM, IgA PCR & IFA 5
Weddle et al [53] 1995 Malaysia, Peru, Indonesia, USA Dot-blot immunoassay IgG, IgM IFA 4
Wilkinson et al [54] 2003 Thailand, Australia ICT IgG, ICT IgM Unclear 2
Wongchotigul et al [55] 2005 Thailand Latex agglutination test IFA 4
Zhang et al [26] 2011 China InBios ICT IgG, ICT IgM IgM and IgG antibodies detection (tests unclear) and PCR 4

aDoxycycline or chloramphenicol

Quality of articles

There are 7, 14, and 10 articles with good, moderate, and low quality, respectively (Table 1). Two articles fulfilled all of the QUADAS-2 main criteria (S1 Dataset). Most of the articles (n = 8, 25.8%) scored 5 points. Two articles (6.5%) scored full points for the risk of bias component (i.e. 4 points), although most of the articles scored only 1 point (n = 12, 38.7%). The majority of the articles scored 3 points in the applicability component (n = 22, 71.0%). There were 24.4% (n = 53/217) sub-questions with ‘unclear’ response. Among the ‘unclear’ response, most of those responses (n = 45, 84.9%) were observed in the index test and reference standard domain of the risk of bias assessment (Fig 2). The main reason for this is that most of the articles did not mention explicitly whether they performed blinding during the conduct and interpretation of both the index test (n = 23, 74.2%) and reference test (n = 25, 80.6%) (S2 Dataset). There were 20 studies (64.5%) with a case-control study design.

Fig 2. QUADAS-2 finding per domain.

Fig 2

Performance of POCT

IgM ICT

There were five manufacturers of IgM ICT identified (Table 2), namely: AccessBio, InBios, PanBio, Standard Diagnostic (SD), and ImmuneMed. There were two studies (6.6%) that assessed in-house tests [24,25]. The sensitivity ranged from 23.3% to 100.0%, and the specificity ranged from 73.0% to 100.0% (Fig 3, results from Zhang et al, 2011 and Lee et al, 2014 were not plotted since only sensitivity values were presented [26,27]). The accuracy across data points of the same manufacturers varied across the studies. The InBios IgM ICT tests reported >80.0% sensitivity and >90.0% specificity on average. The ImmuneMed IgM ICT demonstrated sensitivity of 99.0% and specificity of 98.0%, however, there was only one data point for this manufacturer. One of the studies reported sensitivity and specificity for the AccessBio ICT IgM of 97.0% and 93.0%, respectively. However, the other AccessBio studies did not demonstrate such a high degree of accuracy (Fig 3).

Table 2. Summary of ICT IgM.
Assay Study Year Location of sample collection Sample collection timing Reference test assay type Reference assay positivity cut-off Sensitivity (%)/ specificity (%)
CareStart, AccessBio Blacksell et al [28] 2010 Laos and Thailand Acute phase IFA and/or PCR and/or culture IFA: 4-fold rise in paired serum samples 96.8/93.3
Blacksell et al [19] 2012 Thailand Acute (median 5 days of fever, IQR 3 to 7 days) and convalescent phase STIC IFA: an admission IgM titre of 1:12,800, and/or a 4-fold rising IgM titre in paired serum samples 55.6/90.0
Watthanaworawit et al [29] 2015 Thailand Acute (median 2 days of fever, IQR: 2–3 days) and convalescent phase (median interval to convalescent sample collection was 14 days, range: 11–30 days). PCR & IFA ≥ 4-fold increase in IFA IgM titer, 1:25,600 Acute samples: 23.3/81.4
Paired samples: 32.6/79.1
ImmuneMed Stephen et al [51] 2016 India Acute phase with partly paired samples IFA 1:40 87.0/94.6
InBios Zhang et al [26] 2011 China Acute phase: 2 to 10 days after onset IgM and IgG antibodies detection (unclear) and PCR Unclear Sensitivity 93.9%
Silpasakorn et al [48] 2012 Thailand Acute and convalescent phase PCR & IFA IFA: IgM 1:400 or a 4-fold increase 90.2/85.5
InBios prototype Kingston et al [41] 2015 Thailand, Nepal Unclear IFA 1:400 PAb: 52.0/97.0
1:400 MAb: 52.0/96.0
1:1,600 PAb: 82.0/98.0
1:1,600 MAb: 82.0/96.0
1:6,400 PAb: 92.0/95.0
1:6,400 MAb: 92.0/94.0
1:25,600 PAb: 100.0/94.0
1:25,600 MAb: 100.0/93.0
PanBio Blacksell et al [34] 2010 Laos Admission phase: median 6 days of fever (IQR = 5–10 days); median interval to convalescent sampling: 7 days (IQR = 5–10 days) IFA ≥ 1:400 23.8/86.2
4-fold increase in paired samples 31.9/99.5
Combination of the above 34.7/99.7
Paris et al [20] 2011 Thailand Mostly paired samples. Median fever before admission = 5 days (IQR: 3–7) STIC IFA: an admission IgM titer ≥1:12,800 and/or a 4-fold increase 47.0/95.0
Blacksell et al [19] 2012 Thailand Acute (median 5 days of fever, IQR 3 to 7 days) and convalescent phase STIC IFA: an admission IgM titer ≥1:12,800 and/or a 4-fold increase 46.3/95.1
PanBio prototype Wilkinson et al [54] 2003 Thailand, Australia Unclear Unclear Unclear 95.8/86.0
SD Bioline Tsutsugamushi Blacksell et al [19] 2012 Thailand Acute (median 5 days of fever, IQR 3 to 7 days) and convalescent phase STIC IFA: an admission IgM titer ≥1:12,800 and/or a 4-fold increase 67.9/73.0
Ramyasree et al [46] 2015 India? Unclear ELISA Unclear Agreement 97%
In-house tests Ching et al [24] 2001 Unclear Unclear IFA >1:40 74.0/99.0
Cao et al [25] 2007 China Unclear Unclear Unclear 81.2%/100.0
Fig 3. IgM ICT sensitivity and specificity range*.

Fig 3

Total antibody ICT

The ICTs with IgG, combination of IgG and IgM, and combination of IgG, IgM, and IgA as the detection target were grouped together under ‘total antibody ICT’ (Table 3). There were five manufacturers identified, namely: AccessBio, ImmuneMed, InBios, PanBio, and SD. The remaining studies (n = 3, 6.6%) assessed the diagnostic performance of in-house tests in development. The sensitivity and specificity ranged from 20.9% to 99.1% and 67.9% to 100.0%, respectively (Fig 4, results from and Zhang et al, 2011 and Lee et al, 2014 were not plotted since only sensitivity values were presented [26,27]). As in the case of IgM ICT, the accuracy across data points of total antibody ICT of the same manufacturer varies. The ImmuneMed total antibody ICT demonstrated >95.0% specificity and >80.0% sensitivity.

Table 3. Summary table of total antibody ICT.
Assay Study Year Location of sample collection Sample collection timing Reference test assay type Reference assay positivity cut-off Sensitivity (%)/ specificity (%)
CareStart, AccessBio Blacksell et ala [28] 2010 Laos and Thailand Acute phase IFA and/or PCR and/or culture IFA: 4-fold rise in paired serum samples 97.6/71.4
Blacksell et ala [19] 2012 Thailand Acute (median 5 days of fever, IQR 3 to 7 days) and convalescent phase STIC IFA: an admission IgM titre of 1:12,800, and/or a 4-fold rising IgM titre in paired serum samples 60.5/67.9
ImmuneMed Kim et ala [39] 2013 Korea Unclear Unclear Unclear 99.1/98.9
Kim et alb [40] 2016 Koread Unclear IFA IgM 1:10 98.6/98.2
IFA IgG 1:40 97.1/97.7
Koreae Unclear IFA IgM 1:10 98.6/97.6
IFA IgG 1:40 97.1/97.0
Sri Lanka Unclear IFA Unclear 92.1/96.1
India Unclear IFA IgM 1:10 Sensitivity 86.0%
IFA IgG 1:40 Sensitivity 92.0%
Stephen et alb [51] 2016 India Acute and convalescent phase IFA 1:80 77.3/86.4
InBios Zhang et alb [26] 2011 China Acute phase: 2 to 10 days after onset of illness IgM and IgG antibodies detection (unclear) and PCR Unclear Sensitivity 90.9%
Silpasakorn et alc [48] 2012 Thailand Acute and convalescent phase PCR & IFA IFA: IgG 1:800 or a 4-fold increase 66.7%/98.4
PanBio Coleman et alb [36] 2002 Thailand Acute phase IIP 1:400 90.4/95.8
Panbio prototype Wilkinson et alb [54] 2003 Thailand, Australia Unclear Unclear Unclear 96.1/93.0
SD Bioline Tsutsugamushi Silpasakorn et alc [49] 2012 Thailand Acute phase: median 6 days of fever (range 1–47 days); median interval between obtaining admission and convalescence phase: 13 days (range 3–32 days) PCR & IFA IFA: IgM or IgG IFA assay titre > 1:400 or a 4-fold increase 66.7/98.4
Lee et alc [27] 2014 Korea Unclear IFA 4-fold rise or single titre ≥ 1:160 Sensitivity 72.6%
Stephen et alc [50] 2015 India Acute and convalescence phase ELISA IgM Unclear 91.7/90.5
ELISA IgG Unclear 85.7/100.0
Watthanaworawit et alc [29] 2015 Thailand Median fever = 2 days (IQR: 2–3 days), and the median interval between obtaining initial acute-phase specimens and convalescent specimens was 14 days (range: 11–30 days). PCR & IFA ≥ 4-fold increase in IFA IgM titre, 1:25,600 Acute samples: 20.9/74.4
Paired samples: 76.7/76.7
In-house tests Ching et ala [24] 2001 Unclear Unclear IFA >1:40 IgG: 86.0/97.0
IgG, IgM: 89.0/97.0
Cao et ala [25] 2007 China Unclear Unclear Unclear IgG: 94.6/98.9
IgG, IgM: 98.2/98.1

aTotal antibody: IgG & IgM

bIgG

cIgG, IgM, IgA

dHealthy control sera

eDiseases non-scrub typhus control sera

Fig 4. Total antibody ICT sensitivity and specificity range*.

Fig 4

Dot-blot

Aside from the in-house tests assessed by four studies, there were two dot-blot assay manufacturers, namely Integrated Diagnostics and PanBio. The range was 59.6% to 100.0% and 83.0% to 98.7% for sensitivity and specificity, respectively (Table 4).

Table 4. Summary of articles on dot-blot, LAMP and other assays.
Assay type Assay Study Year Location of sample collection Sample collection timing Reference test assay type Reference assay positivity cut-off Sensitivity (%)/ specificity (%)
Immunoblot Dip-S-Ticks, Integrated Diagnostics Pradutkanchana et al [44] 1997 Thailand Unclear IFA ≥1:400 or a 4-fold or greater rise in IFA titre to at least 1:200 94.0/98.7
Chinprasatsak et al [35] 2001 Thailand Acute and convalescent phase IIP “Four-fold or greater rise in IIP titre to at least 1:200, or a single IgM antibody titre of ≥ 1:400 and/or IgG antibody titre of ≥ 1:1,600.” 86.7/94.3
Coleman et al [36] 2002 Thailand Acute phase IIP 1:400 IgM: 60.3/97.4
IgG: 59.6/95.3
Rickettsia Screen, PanBio Prakash et al [45] 2006 India Acute phase Dot EIA, ELISA, clinical response to antibiotics* ELISA: ≥16 PanBio units 100.0/93.5
In-house Weddle et al [53] 1995 Malaysia, Peru, Indonesia, USA Acute phase, unclear for negative panel IFA <1:64, >1:128, if in between depends on clinical condition 90.0/83.0
Watt et al [52] 1998 Thailand High-antigen dipstick: median = 4 days of fever (range = 2–14) IFA IgG 1:1600 and/or IgM 1:400 67.0/98.0
Low-antigen dipstick: median = 4 days of fever (range = 2–30) 100.0/98.0
Kim et al^ [39] 2013 Korea Unclear Unclear Unclear 97.7/98.6
Rodkvamtook et al [47] 2015 Thailand Acute phase: fever no more than 2 weeks; IFA 1:400 or 4-fold increase for IgG & IgM 98.5/96.3
Paired sample. Convalescent phase: collected after 3–14 days after the acute sample. 98.3/97.0
LAMP Loopamp kit, Eiken Chemical LAMP Paris et al [43] 2008 Thailand, Laos Acute phase for positive panel IFA 4-fold rise or single titre ≥ 1:160
Sensitivity 72.6%
Paris et al [20] 2011 Thailand Mostly paired samples. Median fever before admission = 5 days (IQR: 3–7) In vitro isolates, IFA, PCR, ICT IgG & IgM 1:400 and four-fold rise of paired serum for IgM IFA
100.0/100.0
In-house LAMP Huber et al [37] 2012 Not stated Not stated in detail PCR N/A 90.0/80.0
Combined Panbio ICT IgM and in-house LAMP Blacksell et al [19] 2012 Thailand Acute (median 5 days of fever, IQR 3 to 7 days) and convalescent phase STIC (cell culture isolation, PCR, IFA) IFA: an admission IgM titre of 1:12,800, and/or a 4-fold rising IgM titre in paired samples
66.7/90.6
Combined SD ICT IgM and in-house LAMP Blacksell et al [19] 2012 Thailand Acute (median 5 days of fever, IQR 3 to 7 days) and convalescent phase STIC (cell culture isolation, PCR, IFA) IFA: an admission IgM titre of 1:12,800, and/or a 4-fold rising IgM titre in paired samples
77.2/68.2
Combined AccessBio ICT IgG, IgM and in-house LAMP Blacksell et al [19] 2012 Thailand Acute (median 5 days of fever, IQR 3 to 7 days) and convalescent phase STIC (cell culture isolation, PCR, IFA) IFA: an admission IgM titre of 1:12,800, and/or a 4-fold rising IgM titre in paired samples
68.5/84.9
Combined AccessBio ICT IgM and in-house LAMP Blacksell et al [19] 2012 Thailand Acute (median 5 days of fever, IQR 3 to 7 days) and convalescent phase STIC (cell culture isolation, PCR, IFA) IFA: an admission IgM titre of 1:12,800, and/or a 4-fold rising IgM titre in paired samples
71.6/63.2
Others Passive hemagglutination assay Kim et al [38] 1993 Unclear Unclear IFA Seroconversion or 4-fold rise
99.0/98.9
IgM dot-immunobinding assay Koay et al [42] 1995 Unclear Unclear IIP 1:50
90.4/81.4
Latex agglutination assay Wongchotigul et al [55] 2005 Thailand Unclear IFA 1:400
89.1/98.2

*Doxycycline or chloramphenicol

^IgG only

LAMP

The Loopamp LAMP kit (Eiken Chemical, Japan) was assessed in two studies (Table 4). The other two studies evaluated in-house tests. The sensitivity ranged from 66.7% to 100.0% and specificity ranged from 63.2% to 100.0%.

Other methods

There was one study each on a passive hemagglutination assay, an IgM dot-immunobinding assay, and a latex agglutination assay (Table 4). These tests were all in the development phase.

Meta-analysis results

There were 11 data points extracted from four studies included in the meta-analysis (Fig 5). In the resulting forest plot (Fig 5), the top three data were extracted from studies assessing total antibody ICT (Blacksell et al, 2010 and Watthanaworawit et al, 2015) [28,29]. The rest of the 2 by 2 table data were extracted from studies assessing IgM ICT diagnostic performance. The pooled sensitivity and specificity were 66.0% (95% CI 0.37–0.86) and 92.0% (95% CI 0.83–0.97), respectively. The overall Chi-square heterogeneity statistics showed significant heterogeneity (p < 0.001). There is a high degree of heterogeneity present (I2 value = 97.48%, 95% CI 96.71–98.24 for sensitivity and I2 value = 98.17%, 95% CI 97.67–98.67 for specificity). Meta-regression on several covariates was performed in an attempt to explain this heterogeneity. Significant heterogeneity was observed for total number of samples (p = 0.01), study design (whether using case-control design or not, p = 0.01), blinding during index test interpretation (p = 0.02), and QUADAS-2 score (p = 0.01). No significant heterogeneity was observed for the blinding during reference test interpretation (p = 0.21) and antibody target detection of the tests (p = 0.22). All of these studies used IFA as their reference standard, except Blacksell et al, 2010 which used IFA with the addition of PCR and culture [28]. None of the meta-analysed studies used an IFA cut off lower than 1:400 as the reference comparator.

Fig 5. Forest plot of sensitivity and specificity of current commercially available POCT.

Fig 5

Discussion

There were 31 relevant articles included in this review. Almost half of the included articles were of moderate quality. The meta-analysis showed moderately low pooled sensitivity and good specificity of the current commercially available scrub typhus POCT. However, the studies were heterogeneous with the I2 value indicating a high degree of heterogeneity. Hence, the pooled sensitivity and specificity value needs to be interpreted with caution.

The systematic review and meta-analysis highlighted the methodological and clinical heterogeneity across scrub typhus POCT diagnostic accuracy studies. These differences made it difficult to pool results and compare studies. Meta-regression for other covariates of interest (e.g. sample collection timing) could not be performed because of limited information presented in the original articles.

Almost a quarter of the responses gathered in the main seven questions on QUADAS-2 quality assessment were ‘unclear’. Although we did not assess the quality of reporting in this review, this finding indicates that the quality of reporting in the included studies is still arguably poor. Poorly conducted and controlled diagnostic accuracy studies are a waste of time, resources, and effort; moreover, if research is not accurately reported, it can hinder critical appraisal, replication, and meta-analysis of studies [30]. The launch of reporting guidelines such as Standards for Reporting Diagnostic Accuracy (STARD) and PRISMA is a starting point in improving the quality of reporting, although they are not applied as much as they should be [30,31]. Besides adhering to reporting guidelines, regulations should be adapted to incentivise better and more complete reporting [31]. Creating a reporting infrastructure and building the capability of both authors and reviewers are also necessary to encourage better reporting [31].

Approximately two thirds of the included studies used case-control design. Compared with studies that recruit patients consecutively, case-control study design (evaluating index test in separate diseased population and control group) overestimates diagnostic performance [32]. Therefore, the results presented need to be interpreted with caution.

Aside from LAMP assay, all of the other POCTs reviewed here relied on antibody detection. Serological diagnosis is problematic in several ways. First, the primary serum collection may not contain sufficient antibodies since it takes time for antibodies to increase to a detectable level creating a “false negative” result. Second, in endemic populations with significant background immunity, an appropriate cut-off needs to be established to ensure accurate diagnosis otherwise there is the possibility of “false positives”. Furthermore the issues regarding the selection of IgM as opposed to whole antibody is very much dependent on the situation. Conventional thought is that whole antibodies may give higher number of “false positive” results in endemic situations due to the presence of residual IgG from previous scrub typhus infections as was the case with the AccessBio ICTs tested in Thailand and Laos [19,28]. These shortcomings of serology highlight the need to develop alternative diagnostic strategies.

Although being pooled from heterogeneous studies, the cumulative specificity confidence interval was above 80.0%. This indicates that commercially available ICTs have value in “ruling in” for the diagnosis of scrub typhus. However, it is difficult to draw conclusions with confidence based on the currently available evidence given the high degree of heterogeneity amongst the studies.

Another important consideration when performing a diagnostic accuracy study is the choice of reference comparator. Often the IFA, the serological “gold standard”, is selected. However, this test is far from perfect. IFA result interpretation is subjective and there remains a lack of consensus on the optimum positivity cut-off. A cut-off is often adopted without sufficient local evidence, potentially resulting in incorrect diagnostic accuracy measures for the diagnostic under test [15]. To evaluate the true accuracy of IFA, Bayesian latent class models (LCM) have been used. The models showed that the IFA IgM has sensitivity and specificity of 70.0% and 83.8%, respectively; therefore suggesting it to be unfit as a reference standard [12]. An alternative reference comparator is the composite Scrub Typhus Infection Criteria (STIC), which have been proposed as a more robust method to diagnose scrub typhus with more confidence, by including a panel of parameters with high specificity [20]. The panel includes: (i) isolation of O. tsutsugamushi, (ii) at least two positive out of the three PCR assays targeting the 56kDa, 47kDa, and groEL genes, (iii) admission IFA IgM titre of ≥1:12,800, (iv) 4-fold rise in IgM titre from paired samples [12,19,20]. At least one of these criteria needs to be fulfilled for a positive scrub typhus diagnosis. However the Bayesian LCM also showed that STIC’s sensitivity and specificity are less than optimal for a reference test (90.5% and 82.5%, respectively).

This study has several limitations. First, the search was performed in one database, and in English only which might have resulted in non-inclusion of relevant articles. Second, the article inclusion and quality assessment were completed primarily by one person, though we attempted to decrease the risk of bias by routine discussion of contentious studies amongst the authors. Third, several studies did not present all of the parameters that we wished to extract, hence limiting the meta-analysis, as we could not include all relevant papers due to the limited information presented in the papers.

In reality, there is only a small number of scrub typhus POCTs that are commercially available. The Panbio POCTs are no longer available and therefore the market is dominated by a few companies including InBios, Standard Diagnostics and AccessBio. Selecting the most appropriate POCT is very much dependant on a few key factors such as availability of product locally, the price and the shelflife. In this study we have not examined these local practical aspects that should be considered when selecting POCTs.

In the absence of robust POCTs, the presence of an eschar can be a valuable clinical sign strongly suggesting a diagnosis of scrub typhus. Although its reported presence is very variable, it is still regarded as pathognomonic. An eschar may go unnoticed by the patient since it is painless, often does not itch, can be small, and looks similar to post trauma scab. In addition, it may be located in a concealed area, such as the perineum or under the breasts. This emphasises the importance of performing a thorough physical examination.

Treating patients empirically based on the pathogen pattern in an area is common practice, but bears the risk of unnecessary or inappropriate treatment (with the attendant risks of side effects) and promotion of antimicrobial resistance. POCTs can play an important role in reducing the number of patients treated empirically and increasing the proportion of patients treated appropriately.

There is an urgent need to develop an affordable and accurate POCT. However, if POCTs still rely on serological measures only, they might not be able to provide diagnosis in time to inform treatment. Future research should also be directed towards developing new antigen-based tests to improve diagnostic accuracy in the early period of disease.

Supporting information

S1 Dataset. QUADAS-2 scores.

(XLSX)

S2 Dataset. QUADAS-2 sub-question.

(XLSX)

S3 Dataset. Extracted data.

(XLSX)

S1 Checklist. PRISMA Checklist.

(DOC)

Acknowledgments

We would like to thank Mahidol-Oxford Tropical Medicine Research Unit in which Kartika Saraswati was doing her internship during the writing of this paper.

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

KS, NPJD, MM, and SDB are funded by the Wellcome Trust of the United Kingdom. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1.Paris D, Day NP. Tropical rickettsial infections In: Farrar J, Hotez P, Junghanss T, Kang G, Lalloo D, White N, editors. Manson’s Tropical Diseases. Philadelphia: Elsevier Saunders; 2014:273–291. [Google Scholar]
  • 2.Balcells ME, Rabagliati R, García P, Poggi H, Oddó D, Concha M, et al. Endemic scrub typhus-like illness, Chile. Emerg Infect Dis. 2011;17(9): 1659–1663. doi: 10.3201/eid1709.100960 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Weitzel T, Dittrich S, Lopez J, Phuklia W, Martinez-Valdebenito C, Velasquez K, et al. Endemic scrub typhus in South America. N Engl J Med. 2016;375(10): 954–961. doi: 10.1056/NEJMoa1603657 [DOI] [PubMed] [Google Scholar]
  • 4.Kelly DJ, Fuerst PA, Ching W, Richards AL. Scrub typhus: the geographic distribution of phenotypic and genotypic variants of Orientia tsutsugamushi. Clin Infect Dis. 2009;48(s3): S203–S230. [DOI] [PubMed] [Google Scholar]
  • 5.Izzard L, Fuller A, Blacksell SD, Paris DH, Richards AL, Aukkanit N, et al. Isolation of a novel Orientia species (O. chuto sp. nov.) from a patient infected in Dubai. J Clin Microbiol. 2010;48(12): 4404–4409. doi: 10.1128/JCM.01526-10 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Taylor AJ, Paris DH, Newton PN. A systematic review of mortality from untreated scrub typhus (Orientia tsutsugamushi). PLoS Negl Trop Dis. 2015;9(8): 1–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Phongmany S, Rolain JM, Phetsouvanh R, Blacksell SD, Soukkhaseum V, Rasachack B, et al. Rickettsial infections and fever, Vientiane, Laos. Emerg Infect Dis. 2006;12(2): 256–262. doi: 10.3201/eid1202.050900 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Brown GW, Robinson DM, Huxsoll DL, Ng TS, Lim KJ, Sannasey G. Scrub typhus: a common cause of illness in indigenous populations. Trans R Soc Trop Med Hyg. 1976;70(5–6): 444–448. [DOI] [PubMed] [Google Scholar]
  • 9.Koh GCKW, Maude RJ, Paris DH, Newton PN, Blacksell SD. Diagnosis of scrub typhus. Am J Trop Med Hyg. 2010;82(3): 368–370. doi: 10.4269/ajtmh.2010.09-0233 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Silpapojakul K, Varachit B, Silpapojakul K. Paediatric scrub typhus in Thailand: a study of 73 confirmed cases. Trans R Soc Trop Med Hyg. 2004;98(6): 354–359. doi: 10.1016/j.trstmh.2003.10.011 [DOI] [PubMed] [Google Scholar]
  • 11.Ogawa M, Hagiwara T, Kishimoto T, Shiga S, Yoshida Y, Furuya Y, et al. Scrub typhus in Japan: epidemiology and clinical features of cases reported in 1998. Am J Trop Med Hyg. 2002;67(2): 162–165. [DOI] [PubMed] [Google Scholar]
  • 12.Lim C, Paris DH, Blacksell SD, Laongnualpanich A, Kantipong P, Chierakul W, et al. How to determine the accuracy of an alternative diagnostic test when it is actually better than the reference tests: a re-evaluation of diagnostic tests for scrub typhus using bayesian LCMs. PLoS One. 2015;10(5):e0114930 doi: 10.1371/journal.pone.0114930 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Tsay RW, Chang FY. Serious complications in scrub typhus. J Microbiol Immunol Infect. 1998;31(4): 240–244. [PubMed] [Google Scholar]
  • 14.Paris DH, Shelite TR, Day NP, Walker DH. Unresolved problems related to scrub typhus: a seriously neglected life-threatening disease. Am J Trop Med Hyg. 2013;89(2):301–307. doi: 10.4269/ajtmh.13-0064 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Blacksell SD, Bryant NJ, Paris DH, Doust JA, Sakoda Y, Day NPJ. Scrub typhus serologic testing with the indirect immunofluorescence method as a diagnostic gold standard: a lack of consensus leads to a lot of confusion. Clin Infect Dis. 2007;44(1): 391–401. [DOI] [PubMed] [Google Scholar]
  • 16.Blacksell SD, Tanganuchitcharnchai A, Nawtaisong P, Laongnualpanich A, Basnyat B, Day NPJ, et al. Diagnostic accuracy of the InBios scrub typhus detect enzyme-linked immunoassay for the detection of IgM antibodies in Northern Thailand. Clin Vaccine Immunol. 2015;23(2): 148–154. doi: 10.1128/CVI.00553-15 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Price CP. Point of care testing. BMJ Clin Res. 2001;322: 1285–1288. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Shaw JLV. Practical challenges related to point of care testing. Pract Lab Med. 2016;4: 22–29. doi: 10.1016/j.plabm.2015.12.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Blacksell SD, Paris DH, Chierakul W, Wuthiekanun V, Teeratakul A, Kantipong P, et al. Prospective evaluation of commercial antibody-based rapid tests in combination with a loop-mediated isothermal amplification PCR assay for detection of Orientia tsutsugamushi during the acute phase. Clin Vaccine Immunol. 2012;19(3): 391–395. doi: 10.1128/CVI.05478-11 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Paris DH, Blacksell SD, Nawtaisong P, Kemajittra J, Chierakul W, Wuthiekanun V, et al. Diagnostic accuracy of a loop-mediated isothermal PCR assay for detection of Orientia tsutsugamushi during acute scrub typhus infection. PLoS Negl Trop Dis 2011. 5(9): e1307 doi: 10.1371/journal.pntd.0001307 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Kularatne SAM, Gawarammana IB. Validity of the Weil-Felix test in the diagnosis of acute rickettsial infections in Sri Lanka. Trans R Soc Trop Med Hyg. 2009;103(4): 423–424. doi: 10.1016/j.trstmh.2008.11.020 [DOI] [PubMed] [Google Scholar]
  • 22.Whiting PF, Rutjes AWS, Westwood ME, Mallet S, Deeks JJ, Reitsma JB, et al. QUADAS-2: A revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155(4): 529–536. [DOI] [PubMed] [Google Scholar]
  • 23.Moher D, Liberati A, Tetzlaff J, Altman DG, The PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: The PRISMA statement. PLoS Med. 2009;89(9):e1000097 doi: 10.1371/journal.pmed.1000097 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Ching W, Rowland D, Zhang Z, Bourgeois AL, Kelly D. Early diagnosis of scrub typhus with a rapid flow assay using recombinant major outer membrane protein antigen (r56) of Orientia tsutsugamushi. 2001;8(2):409–414. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Cao M, Guo H, Tang T, Wang C, Li X, Pan X, et al. Preparation of recombinant antigen of O. tsutsugamushi Ptan strain and development of rapid diagnostic reagent for scrub typhus. Am J Trop Med Hyg. 2007;76(3): 553–558. [PubMed] [Google Scholar]
  • 26.Zhang L, He S, Wang S, Yu H, Li X, Zhang D, et al. Comparison of a rapid diagnostic test and microimmunofluorescence assay for detecting antibody to Orientia tsutsugamushi in scrub typhus patients in China. Asian Pac J Trop Med. 2011: 666–668. doi: 10.1016/S1995-7645(11)60169-7 [DOI] [PubMed] [Google Scholar]
  • 27.Lee K, Moon C, Oh WS, Sohn KM, Kim B. Diagnosis of scrub typhus: Introduction of the immunochromatographic test in Korea. Korean J Intern Med. 2014;29: 253–255. doi: 10.3904/kjim.2014.29.2.253 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Blacksell SD, Jenjaroen K, Phetsouvanh R, Wuthiekanun V, Day NPJ, Newton PN, et al. Accuracy of AccessBio Immunoglobulin M and total antibody rapid immunochromatographic assays for the diagnosis of acute scrub typhus infection. Clin Vaccine Immunol. 2010;17(2): 263–266. doi: 10.1128/CVI.00448-08 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Watthanaworawit W, Turner P, Turner C, Tanganuchitcharnchai A, Jintaworn S, Hanboonkunupakarn B, et al. Diagnostic accuracy assessment of immunochromatographic tests for the rapid detection of antibodies against Orientia tsutsugamushi using paired acute and convalescent specimens. Am J Trop Med. Hyg. 2015;93(6): 1168–1171. doi: 10.4269/ajtmh.15-0435 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Paul P, Irwig L, et al. STARD 2015: An updated list of essential items for reporting diagnostic accuracy studies. BMJ. 2015;351:h5527 doi: 10.1136/bmj.h5527 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Glasziou P, Altman DG, Bossuyt P, Boutron I, Clarke M, Julious S, et al. Reducing waste from incomplete or unusable reports of biomedical research. Lancet. 2014;383(9913): 267–276. doi: 10.1016/S0140-6736(13)62228-X [DOI] [PubMed] [Google Scholar]
  • 32.Lijmer JG, Mol B, Heisterkamp S, Bonsel GJ, Prins MH, van der Meulen JHP, et al. Empirical evidence of design-related bias in studies of diagnostic tests. JAMA. 1999;282(11): 1061–1066. [DOI] [PubMed] [Google Scholar]
  • 33.Anitharaj V, Stephen S, Pradeep J, Park S, Kim SH, Kim YJ, et al. Serological diagnosis of acute scrub typhus in Southern India: Evaluation of InBios Scrub Typhus Detect IgM Rapid Test and comparison with other serological tests. J Clin Diagn Res. 2016;10(11):DC07–DC10. doi: 10.7860/JCDR/2016/24051.8861 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Blacksell SD, Jenjaroen K, Phetsouvanh R, Tanganuchitcharnchai A, Phouminh P, Phongmany S, et al. Accuracy of rapid IgM-based immunochromatographic and immunoblot assays for diagnosis of acute scrub typhus and murine typhus infections in Laos. Am J Trop Med Hyg. 2010;83(2): 365–369. doi: 10.4269/ajtmh.2010.09-0534 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Chinprasatsak S, Wilairatana P, Looareesuwan S, Chappuis F, Akkhavong K, Laferl H, et al. Evaluation of a newly developed dipstick test for the rapid diagnosis of scrub typhus in febrile patients. Southeast Asian J Trop Med Public Heal. 2001;32(1): 132–136. [PubMed] [Google Scholar]
  • 36.Coleman R, Sangkasuwan V, Suwanabum N, Eamsila C, Mungviriya S, Devine P, et al. Comparative evaluation of selected diagnostic assays for the detection of IgG and IgM antibody to Orientia tsutsugamushi. Am J Trop Med Hyg. 2002;67(5): 497–503. [DOI] [PubMed] [Google Scholar]
  • 37.Huber E, Ji D, Howell L, Zhang Z, Chen HW, Ching W, et al. Loop-mediated isothermal amplification assay targeting the 47-kDa gene of Orientia tsutsugamushi: A rapid and sensitive alternative to real-time PCR. J Med Microb Diagn. 2012;1(4). doi: 10.4172/2161-0703.1000112 [Google Scholar]
  • 38.Kim I, Seong S, Woo S, Choi M, Kang J, Chang W. Rapid diagnosis of scrub typhus by a passive hemagglutination assay using recombinant 56 kilodalton polypeptides. J Clin Microbiol. 1993;31(8): 2057–2060. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Kim Y, Yeo S, Park S, Woo Y, Kim M, Kim S, et al. Improvement of the diagnostic sensitivity of scrub typhus using a mixture of recombinant antigens derived from Orientia tsutsugamushi serotypes. J Korean Med Sci 2013;28: 672–679. doi: 10.3346/jkms.2013.28.5.672 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Kim Y, Park S, Premaratna R, Selvaraj S, Park S, Kim S, et al. Clinical evaluation of rapid diagnostic test kit for scrub typhus with improved performance. 2016;(12): 1190–1196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Kingston H, Blacksell SD, Tanganuchitcharnchai A, Laongnualpanich A, Basnyat B, Day NPJ, et al. Comparative accuracy of the InBios scrub typhus detect IgM rapid test for the detection of IgM antibodies by using conventional serology. Clin Vaccine Immunol. 2015;22(10): 1130–1132. doi: 10.1128/CVI.00390-15 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Koay AS, Tay ST, Cheong YM, Yasin RM. Development and preliminary evaluation of an IgM dot-immunobinding assay (IgM-DIA) for rapid serodiagnosis of scrub typhus infection. Southeast Asian J Trop Med Public Health. 1995;26(2): 350–353. [PubMed] [Google Scholar]
  • 43.Paris DH, Blacksell SD, Newton PN, Day NP. Simple, rapid and sensitive detection of Orientia tsutsugamushi by loop-isothermal DNA amplification. Trans R Soc Trop Med Hyg. 2008;102(12):1239–1246. doi: 10.1016/j.trstmh.2008.04.040 [DOI] [PubMed] [Google Scholar]
  • 44.Pradutkanchana J, Silpapojakul K, Paxton H, Pradutkanchana S, Kelly D, Strickman D. Comparative evaluation of four serodiagnostic tests for scrub typhus in Thailand. Trans R Soc Trop Med Hyg. 1997;91: 425–428. [DOI] [PubMed] [Google Scholar]
  • 45.Prakash J, Abraham O, Mathai E. Evaluation of tests for serological diagnosis of scrub typhus. Trop Doct. 2006;36: 212–213. doi: 10.1258/004947506778604715 [DOI] [PubMed] [Google Scholar]
  • 46.Ramyasree A, Kalawat U, Rani N, Chaudhury A. Seroprevalence of scrub typhus at a tertiary care hospital in Andhra Pradesh. Indian J Med Microbiol. 2015;33(1): 68–72. doi: 10.4103/0255-0857.148381 [DOI] [PubMed] [Google Scholar]
  • 47.Rodkvamtook W, Zhang Z, Chao C, Huber E, Bodhidatta D, Gaywee J. Dot-ELISA rapid test using recombinant 56-kDa protein antigens for serodiagnosis of scrub typhus. 2015;92(5): 967–971. doi: 10.4269/ajtmh.14-0627 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Silpasakorn S, Srisamut N, Ekpo P, Zhang Z, Chao C. Development of new, broadly reactive, rapid IgG and IgM lateral flow assays for diagnosis of scrub typhus. Am J Trop Med Hyg. 2012;87(1): 148–152. doi: 10.4269/ajtmh.2012.11-0583 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Silpasakorn S, Waywa D, Hoontrakul S, Suttinont C, Losuwanaluk K, Suputtamongkol Y. Performance of SD Bioline Tsutsugamushi Assays for the Diagnosis of Scrub Typhus in Thailand. J Med Assoc Thai. 2012;95(2): 18–22. [PubMed] [Google Scholar]
  • 50.Stephen S, Sangeetha B, Ambroise S, Sarangapani K, Gunasekaran D, Hanifah M, et al. Outbreak of scrub typhus in Puducherry & Tamil Nadu during cooler months. Indian J Med Res. 2016;142(5): 591–597. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Stephen S, Kim S, Pradeep J, Kim YJ, Kim EY, Kim MW, et al. Evaluation of ImmuneMed scrub typhus rapid test kit, for diagnosis of scrub typhus. J Vector Borne Dis. 2016(53): 283–287. [PubMed] [Google Scholar]
  • 52.Watt G, Strickman D, Kantipong P, Jongsakul K, Paxton H. Performance of a dot blot immunoassay for the rapid diagnosis of scrub typhus in a longitudinal case series. JID. 1998;177: 800–802. [DOI] [PubMed] [Google Scholar]
  • 53.Weddle J, Chan T, Thompson K, Paxton H, Kelly DJ, Dasch G, et al. Effectiveness of a dot-blot immunoassay of anti-Rickettsia tsutsugamushi antibodies for serologic analysis of scrub typhus. Am J Trop Med Hyg. 1995;53(1): 43–46. [PubMed] [Google Scholar]
  • 54.Wilkinson R, Rowland D, Ching W. Development of an improved rapid lateral flow assay for the detection of Orientia tsutsugamushi–specific IgG / IgM antibodies. Ann NY Acad Sci. 2003;990: 386–390. [DOI] [PubMed] [Google Scholar]
  • 55.Wongchotigul V, Waicharoen S, Riengrod S, Chimsumang S, Rattanadakul J, Chinprasatsak S, et al. Development and evaluation of a latex agglutination test for the rapid diagnosis of scrub typhus. Southeast Asian J Trop Med Public Health Med Public Health. 2005;36(1): 108–112. [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

S1 Dataset. QUADAS-2 scores.

(XLSX)

S2 Dataset. QUADAS-2 sub-question.

(XLSX)

S3 Dataset. Extracted data.

(XLSX)

S1 Checklist. PRISMA Checklist.

(DOC)

Data Availability Statement

All relevant data are within the paper and its Supporting Information files.