Abstract
Background
To evaluate the effect of pretreatment megavoltage computed tomographic (MVCT) scan methodology on setup verification and adaptive dose calculation in helical TomoTherapy.
Methods
Both anthropomorphic heterogeneous chest and pelvic phantoms were planned with virtual targets by TomoTherapy Physicist Station and were scanned with TomoTherapy megavoltage image-guided radiotherapy (IGRT) system consisted of six groups of options: three different acquisition pitches (APs) of ‘fine’, ‘normal’ and ‘coarse’ were implemented by multiplying 2 different corresponding reconstruction intervals (RIs). In order to mimic patient setup variations, each phantom was shifted 5 mm away manually in three orthogonal directions respectively. The effect of MVCT scan options was analyzed in image quality (CT number and noise), adaptive dose calculation deviations and positional correction variations.
Results
MVCT scanning time with pitch of ‘fine’ was approximately twice of ‘normal’ and 3 times more than ‘coarse’ setting, all which will not be affected by different RIs. MVCT with different APs delivered almost identical CT numbers and image noise inside 7 selected regions with various densities. DVH curves from adaptive dose calculation with serial MVCT images acquired by varied pitches overlapped together, where as there are no significant difference in all p values of intercept & slope of emulational spinal cord (p = 0.761 & 0.277), heart (p = 0.984 & 0.978), lungs (p = 0.992 & 0.980), soft tissue (p = 0.319 & 0.951) and bony structures (p = 0.960 & 0.929) between the most elaborated and the roughest serials of MVCT. Furthermore, gamma index analysis shown that, compared to the dose distribution calculated on MVCT of ‘fine’, only 0.2% or 1.1% of the points analyzed on MVCT of ‘normal’ or ‘coarse’ do not meet the defined gamma criterion. On chest phantom, all registration errors larger than 1 mm appeared at superior-inferior axis, which cannot be avoided with the smallest AP and RI. On pelvic phantom, craniocaudal errors are much smaller than chest, however, AP of ‘coarse’ presents larger registration errors which can be reduced from 2.90 mm to 0.22 mm by registration technique of ‘full image’.
Conclusions
AP of ‘coarse’ with RI of 6 mm is recommended in adaptive radiotherapy (ART) planning to provide craniocaudal longer and faster MVCT scan, while registration technique of ‘full image’ should be used to avoid large residual error. Considering the trade-off between IGRT and ART, AP of ‘normal’ with RI of 2 mm was highly recommended in daily practice.
Electronic supplementary material
The online version of this article (10.1186/s13014-018-0989-y) contains supplementary material, which is available to authorized users.
Keywords: TomoTherapy, Megavoltage computed tomographic, Image-guided radiotherapy, Adaptive radiotherapy
Background
Daily repeatability of patient positioning is vital to the successful treatment outcome in radiotherapy. The megavoltage computed tomographic (MVCT) image quality from helical TomoTherapy (HT) has been proved sufficient for image-guided radiotherapy (IGRT) including tumor identification and setup verification [1]. Moreover, considering that MVCT has a reliable CT number to electron density calibration curve, MVCT has also been proved accurate to be used for calculating adaptive daily dose distribution which is an assurance of adaptive radiotherapy (ART) [2]. HT MVCT consisted of advantages in lower absorption dose and a larger imaging capacity (theoretically 40 cm reconstruction field of view (FOV) × 160 cm longitudinal scanning length) than C-arm based cone beam CT (CBCT) technology [3, 4]; thus making HT daily MVCT imaging safe in assessing patient treatment location and make on-line adaptive re-planning possible using the same imaging data sets.
In the process of MVCT image acquisition on TomoTherapy Operator Station (version 5.0, Accuray, Sunnyvale, CA), there are options for ‘acquisition pitch’ (AP) and ‘reconstruction interval’ (RI). AP is directly proportional to couch speed, scan duration and especially correlated to the patient absorption dose [5]. In the meanwhile, RI is much related to 3D reconstruction with varied slice thickness which will impact MVCT image qualities especially in the craniocaudal resolution.
There are still debates regarding to the effect of MVCT scan options on setup verification. Studies argued that there is no noticeable difference in image quality among different APs [6] or the issue could be ignored [7], while some studies concluded that pitch variation and longitudinal MVCT image resolution make statistically significant contribution to the MVCT- planning kilovoltage CT (KVCT) registration process [8, 9]. To the best of our knowledge, there is no study published on the effect of MVCT scan options on adaptive dose calculation. In order to investigate the quantitative effect of MVCT scan and reconstruction options on IGRT and ART, an experiment with two different phantoms was performed and the clinical effects were analyzed and presented in terms of image qualities, the accuracy for setup verification and dosimetry discrepancies.
Methods
CT simulation and planning on phantoms
Anthropomorphic heterogeneous chest (Model 002LFC, CIRS, Norfolk, VA, USA) and pelvic phantoms (Model 002PRA) were used in this study (Fig. 1a,b). Three metal particles were placed on the surface of each phantom to serve as reference points for red laser alignment during planning and treatment setup. Spiral KVCT scans were performed for these phantoms on Philips Brilliance Big Bore CT (Philips, Cleveland, OH, USA) using a 3 × 3 mm stacked axial slice technique with a pitch of 0.938. The KVCT image resolution was 512 × 512 pixels. KVCT serials were transferred to Pinnacle (Philips Medical System, Eindhoven, Netherlands) treatment planning system (TPS) for contouring. Once the synthetic targets and normal tissues were all delineated on transverse sections, they were sent to TomoTherapy Physicist Station and the image resolution was downgraded to 256 × 256 pixels during this transfer process. To assess the dosimetry discrepancies among different MVCT serials, a simulated TomoTherapy treatment plan was mimicked on each phantom as shown in Fig. 1d, e. On KVCT with chest phantom (Fig. 1a), the spinal cord was contoured as a target to simulate a craniospinal irradiation (Fig. 1d); while with pelvic phantom (Fig. 1b), a region of soft tissue was contoured as a target to simulate a prostate irradiation (Fig. 1e). A prescription dose of 3Gy with one fraction was assigned to the simulated targets. The TomoTherapy planning parameters were as follow: dose calculate grid = fine, field width = 2.51 cm, modulation factor = 2.00 and pitch = 0.287.
Phantoms setup and MVCT scan
First, the phantoms were set up in position without shift. For chest and pelvic phantoms, each was placed on TomoTherapy treatment couch respectively by aligning the metallic marker correlated to the red lasers. Then six series of MVCT images were obtained on helical TomoTherapy system with 6 groups of scan settings (Technical setting were listed in the left column of Additional file 1: Table S1). The groups consisted of 3 AP options of ‘fine’ (4 mm/rotation), ‘normal’ (8 mm/rotation) and ‘coarse’ (12 mm/rotation) multiplied by 2 with corresponding RI (fine: 1 and 2 mm; normal: 2 and 4 mm; coarse: 3 and 6 mm). The craniocaudal scan length and scan time were recorded for each MVCT scan. All MVCTs were transferred to TomoTherapy Physicist Station and Planned Adaptive module for adaptive dose calculation.
Secondly, setup with shift. After each chest and pelvic phantom was placed according to the red lasers and metal particle positions, the couch was manually shifted to X, Y, Z axial positive directions for 5.0 mm respectively by Positioning Control Panels which is located on the gantry of the machine with the precision of 0.1 mm in a cardinal IEC scale. Wherein positive direction denotes deviation in left (+X), superior (+Y) and anterior (+Z); while negative direction denotes deviation in right (−X), inferior (-Y) and posterior (−Z). Then transverse MVCT images were obtained again for each phantom with the above scan parameters respectively for the 6 serials. All the 6 serials of MVCT were also transferred to TomoTherapy Planned Adaptive module for offline image registration analysis.
In order to evaluate MVCT scan options on image quality, the cylindrical Virtual Water™ phantom (named ‘cheese phantom’) from TomoTherapy with 20 inserts of 7 various densities (Fig. 1c,f) was also scanned by all MVCT scan options for 6 serials. These series of MVCT images (Fig. 1f) were exported and the 7 different density plugs were contoured on each slice in Varian Eclipse treatment planning system (Varian Medical Systems, Inc., Walnut Creek, CA). The average CT numbers and image noises in the delineated regions of interest (ROIs) on the MVCT images were recorded.
The other MVCT acquisition factors during the TomoTherapy scan include jaw width projected to isocenter ~ 4 mm, gantry period 10 s/rotation, reconstructed FOV ~ 39 cm, irradiation energy of 3.5 MV and dose rate of 45 MU/min.
Registration schemes
To investigate the accuracy of MVCT-KVCT registration, MVCT images were automatically registered by a modified mutual information algorithm called ‘extracted feature fusion’ with planning KVCT in the module of TomoTherapy Planned Adaptive for 6 serials, respectively. For any registration manipulation, four combinations of settings and their corresponding parameters, see Additional file 1, were available to choose from. The helical TomoTherapy system has 3 modes of rigid image registration: one based on bony anatomy, the other based on soft tissue, and the third based on bony anatomy as well as soft tissue, depending on the threshold applied to the Hounsfield numbers [10]. A previous study found that the bony anatomy-based auto registration is most useful [6]. While study from Suh argued that if MVCT provided better soft-tissue contrast, the soft-tissue image based registration would potentially improve setup accuracy [9]. Therefore, for the option of ‘selecting down sampling’ (three parameters), each was used respectively to test the involved pixels in the registration. For the option of ‘Uniform down sampling’, considering that down sampling will obviously affect the registration results [8, 11, 12] without significant time saving, the parameter of ‘Superfine’ was recommended to be applied. For the option of ‘Incomplete FOV’, it is not evaluated since both phantoms in this study will not extend outside of MVCT FOV. For the option of ‘rotational degrees of freedom’, the parameter of ‘translations only’ was selected since 96.6% of the rotational corrections were less than 4° and TomoTherapy treatment couch can only support translational movement in three dimensions [6].
Evaluation methods and factors
Considering treatment efficiency, the scan time, craniocaudal scan length and reconstruction slice numbers of each MVCT series were recorded. Scatterplot and bar histogram were used to present the deviation in average CT number and image noise among 3 metrics of APs. Average CT number was calculated as the mean value of all pixels in each different density plug on the same transversal slice of MVCT image. Image noise was identified as the standard deviation of CT number divided by the mean value of all the pixels in the same density measurement area.
The simulated adaptive patient treatment plan of each phantom was recalculated on 6 MVCT data sets using TomoTherapy Planned Adaptive module. The corresponding dosimetric results were presented and compared with 6 groups of dose volume histogram (DVH) outcomes. Covariance statistical analysis was also used to evaluate the difference on intercept and slope of the DVH curves that were calculated on the most system elaborated (AP of ‘fine’ + RI of ‘1 mm’) and the coarsest (AP of ‘coarse’ + RI of ‘6 mm’) combinations MVCT data sets. Two curves are identified as different when covariance statistical analysis presents significant with p < 0.05. In order to encompass the whole phantom besides of certain ROIs, gamma index analysis method [13] was also used to compare ART dose distributions point by point on MVCTs from different APs by SNC Patient™ software (Sun Nuclear Corporation, Melbourne, FL). Parameters were set as follow: Threshold (TH %) 0.0%, which is the minimum dose percent value that must be met in either dose data from AP of ‘fine’ or dose data from AP of ‘normal’ and ‘coarse’ for the point to be included (0.0% means the whole dose distribution data in the entire image space) in the analysis; Diff(%) 1.0 and Dist 1.0 mm, that are dose difference tolerance and distance tolerance equal to or below which the gamma value of a point is considered ‘passing’ and beyond which a point is considered ‘failing’.
To mathematically evaluate the deviation between the calculated shift distance given by CT-MVCT rigid registration and the expected 5 mm shift distance, an equation of distance-to-agreement (DTA, unit millimeter) was defined as follow:
where Dx, Dy, and Dz are the deviations between the registration result and the expected 5 mm shift distance on X (left-right), Y (superior-inferior) and Z (anterior-posterior) axis direction respectively. Theoretically, the registration result should ideally calculate as − 5.0 mm in each direction in order to meet the artificial 5.0 mm shift originally. Therefore, the smaller DTA value (close to 0) means the better registration result that indicates the lower effect of scan options on setup verification.
Results
MVCT scan analysis
Table 1 shows the phantom scanning time and reconstruction slice numbers in 6 groups of scanning settings in TomoTherapy MVCT IGRT system. For the same phantom, the scan time was basically inversely proportional to the AP, and the reconstruction slice number was inversely proportional to the RI. The scanning time of AP of ‘fine’ was approximately 2 times of ‘normal’ and 3 times of ‘coarse’, which is not affected by the RI options. The reconstruction procedure is independent of RI, which is fast enough to be taken as real-time.
Table 1.
Scan Options | Chest Phantom | Pelvic Phantom | |||||
---|---|---|---|---|---|---|---|
Acquisition Pitch | Reconstruction Interval | Scan Length (mm) | Scan Time (s) | Reconstruction Slices | Scan Length (mm) | Scan Time (s) | Reconstruction Slices |
Fine (4 mm/ra) |
1 mm | 135 | 356 | 135 | 106 | 283 | 106 |
2 mm | 137 | 361 | 69 | 106 | 281 | 53 | |
Normal (8 mm/r) |
2 mm | 136 | 191 | 69 | 106 | 151 | 53 |
4 mm | 134 | 186 | 34 | 107 | 151 | 27 | |
Coarse (12 mm/r) |
3 mm | 139 | 136 | 47 | 108 | 108 | 36 |
6 mm | 137 | 131 | 23 | 114 | 111 | 19 |
amm/r: the movement distance of treatment couch during one rotation of the beam gantry
Effects on MVCT image quality
Figure 2 shows the relationships between CT number, image noise and physical densities on 3 series of MVCT images. Since reconstruction interval is independent of transversal image performances, thus, only acquisition pith (AP) selections (fine, normal and coarse) were considered. Three sets of CT density curves were nearly overlapped in Fig. 2a indicating that MVCT acquisition with different APs have very similar CT numbers. In Fig. 2b, the discrepancy of MVCT image noise from three APs presents minimum significance in 7 selected density regions.
Effects on dose calculation
DVHs with 5 allocated ROIs in the chest (Fig. 3a) and pelvic (Fig. 3b) phantoms were plotted in Fig. 3 with adaptive dose calculation in these 6 series of MVCT image sets. It was concluded that for all ROIs, each DVH curve calculated on different MVCT images overlapped together with no difference by objective observation. Analytical analysis indicated that the covariance statistical analysis result which aims to compare DVH curves were as follows: on both chest and pelvic phantoms, all p values of simulated spinal cord (p = 0.761 and 0.277), heart (p = 0.984 and 0.978), lungs (p = 0.992 and 0.980), soft tissue (p = 0.319 and 0.951) and bony structures (p = 0.960 and 0.929) were definitely present no major significance between the most elaborated and the coarsest MVCT imaging techniques.
Regarding to the gamma index analysis, a total 129,762 check points were included in the whole dose distribution space calculated on MVCT images of AP ‘fine’, ‘normal’ and ‘coarse’ with smaller RIs. After gamma values were checked point by point, passed ratio were 99.8% (129,491 points with gamma< 1) between MVCT AP of ‘fine’ and ‘normal’, and 98.9% (128,391 points with gamma< 1) between ‘fine’ and ‘coarse’. It indicated that only 0.2% or 1.1% of the points analyzed do not meet the defined gamma criterion. A 2-D plot of the gamma values that failed the comparison criteria (Threshold 0.0%, dose difference tolerance 1.0% and distance tolerance 1 mm) on the coronal center slice were colored for ‘fine vs. normal’ (Fig. 4a) and ‘fine vs. coarse’ (Fig. 4b) respectively. It also send the same message with above that, if MVCT ‘fine’ is considered as reference, fewer points from MVCT ‘normal’ failed the comparison than MVCT ‘coarse’. Fig. 4 also shows that MVCT image with larger AP may underestimate the low dose at the superior-inferior edge of the MVCT image space. However, most of the failed points appears on the low dose area (< 20% prescription dose), which is usually not considered and not presented during the planning process. Considering the standard y-jaw setting for MVCT imaging mode is fixed to 4 mm regardless of Aps option [14], the larger difference on dose calculation on the MVCT craniocaudal edge may attribute to faster couch moving during the imaging process. More experiments are expected to prove this assumption.
So far, both visual inspections as well as statistical factors indicated that the choices of AP and RI would not significantly impact the adaptive dose calculation, no matter which parameter was selected during the MVCT scan on TomoTherapy IGRT system.
Effects on registration
Figure 5 shows the registration errors (on axis of X, Y, Z respectively) and the DTAs distribution, which was generated from the discrepancy between the calculated shift distance given by CT-MVCT rigid registration and the pre-defined 5.0 mm translational manually shift.
Figure 5a present the results with chest phantom showing that: first, all registration errors > 1 mm appear on axis of Y (superior-inferior); secondly, even using the smallest AP (fine) and RI (1 mm) could not avoid relatively large craniocaudal errors (> 1 mm); thirdly, in term of DTA estimation, the larger AP would bring larger registration errors. Since each transversal MVCT slice of the chest phantom was almost identical (shown in Fig. 1d), it was assumed that the low and limited craniocaudal spatial resolution effect caused the major errors on axis of Y, which were larger than axis of X, and Z, inherently.
Figure 5b shows the results from pelvic phantom: first, as assumed above, craniocaudal errors were much smaller than that from chest phantom because the craniocaudal resolution of pelvic phantom is higher with its simulated bony structures; second, larger errors and DTA (about 3 mm) were obvious with AP of ‘coarse’; however, the ‘full image’ registration technique will reduce the large errors caused by AP of ‘coarse’ from 2.90 mm to 0.22 mm.
Comparing the results from both chest and pelvic phantoms, the AP of ‘normal’ option present good performance with less scan time and comparable registration accuracy to pitch of ‘fine’. In all cases, there is no significant difference on image registration accuracy for RI option, no matter it is along with AP of small (fine) or large (coarse).
Discussion
IGRT is an efficient tool to improve the accuracy of targeting and is necessary to ART [15]. In head and neck and prostate cancer patients, setup error > 5 mm can occur in 10% [16] and 20% [4] fractions respectively when volume imaging was performed every second day. Li found that the interfractional variations in patient setup and in shapes, sizes, and positions of both targets and normal structures are site specific and may be used to determine the site-specific margins, which emphasized the importance of ART [17]. Duma [18] recommends daily MVCT scan for patients treated with doses close to the tolerance dose of the critical ROI (such as spinal cord). Furthermore, random setup errors can only be eliminated through the use of daily image guidance, whilst the dose from MVCT scan is critical to evaluate the treatment outcome. Quantitatively, Shah [4] proved that the typical HT MVCT imaging dose is approximately 1.5 cGy per image and the uniform MVCT dose delivered using HT is greatest when the anatomic thickness is the smallest and the pitch is set to the lowest value. Since reduplicative MVCT scans on human body is unethical, an anthropomorphic phantom based experiment was created for this study, which aims to give a reasonable MVCT scan protocol by investigating the effect of scanning setting and technique during setup verification and adaptive dose calculation in TomoTherapy system.
Since option of AP is directly proportional to scan time and creating additional irradiation dose to patients [5], it is recommended to use pitch as large as possible if there is no impact on IGRT imaging quality and ART accuracy. In term of adaptive dose calculation, our study shown that MVCT from any AP and RI combinations has no significant effect on these MVCT calculated dose distributions. Langen [2] has indicated that the MVCT pitch ratio has little effect on dose recalculation. The robustness of the recalculations is impressed, given the coarse sagittal reconstructions that were used [2]. With the conclusion, then the use of AP of ‘coarse’ may be tempting in an effort to reduce the imaging dose, which a patient has to receive due to the MVCT acquisition. Yue [19] also argued that different APs have no significant difference on CT number (p = 1.000) and image noise (p = 0.667) among MVCT imaging technical settings. But there is no study, to the best of our knowledge, has ever reported the effect of MVCT RIs. According to our study, we have concluded and recommended the fastest and harmless scan option in case of adaptive planning: AP of ‘coarse’ with RI of 6 mm. In term of ART, the image stability is another critical issue. As Langen [20] resulted that, during a 9-month period, an average variation of 20 Hounsfield units was seen (by a CT-to-electron density phantom) for the fan-beam MVCT imaging system. However, Langen [2] also argued that, this variation does not affect the dose calculation result significantly. An agreement of 0.5% was found between the planned and recalculated dose received by 95% of the target volume. Therefore, it was concluded that the fan-beam MVCT dose calculation accuracy is similar to that of the initial diagnostic KVCT dose calculation. And in another perspective, to approve the stability, MVCT image quality is one of the most important content in the Tomotherapy monthly QA protocol according to report TG-148, where image noise, uniformity, spatial resolution, contrast, and the MVCT dose are monitored. Furthermore, ‘CT number calibration’ is a mandatory weekly QA procedure on TomoTherapy operation station, which will compare the average measured HU to expected HU for both homogeneity water equivalent cheese phantom and air. If the absolute difference is smaller than 70 HU, TomoTherapy system will automatically correct the CT numbers in order to ensure consistency of the CT number values. So far, based on the arguments above, we can be confident to the MVCT image’s stability for accurate dose calculation in adaptive radiotherapy.
However, in term of IGRT, our study shows that large AP (coarse) at KVCT-MVCT registration would lead to large error, which was coincident to a previous study which found that 4 and 2 mm inter-slice spacings generally have lower residual error than 6 mm spacing [21] Therefore, to tradeoff between IGRT and ART requirement, AP of ‘normal’ with RI of 2 mm is recommended in daily IGRT and ART clinical application based on our research results. However, in case of stereotactic body radiation therapy (SBRT) with early stage lung tumor (< 4 cm) [22], since AP of ‘normal’ and ‘coarse’ have larger step size per rotation (8 mm/rotation and 16 mm/rotation, respectively) which may lose critical target anatomy information, tumor density [23] and texture features [24] which may be used to predict therapeutical effect, AP of ‘fine’ (4 mm/rotation) with RI 1 mm setting is recommended.
Smaller RI may improve the longitudinal display resolution, with neither impact on scan time, patient dose nor impact on ART dose recalculation as well as IGRT registration accuracy. Furthermore, larger reconstruction resolution may miss smaller tumor or lymph nodes in adaptive delineation and replanning process. Levegrun [8] also reminded that even with the finest pitch, the MVCT images of a thin object are blurred longitudinally to limit the visibility of small anatomic structures. Therefore, no matter which AP is selected, the smaller RI is recommended to provide relative higher resolution in daily IGRT and ART applications. However, since the maximum number of slices allowed for an MVCT scan is 300, thus the ‘coarse’ setting and larger RI is a compromise if the scan is necessary to cover as long as possible. In this case, ‘full image’ registration technique should be used to remedy the disadvantage of large AP that might generate large registration errors. A reasonable explanation is that registration with ‘full image’ technique enrolled more voxels from MVCT, which provided closer axial resolution to planning CT. Our recent studies also proved that, if two series of images have a closer axial resolution, the registration will be matched better and will deliver higher registration accuracy [12, 25].
Compared to lateral and ventrodorsal directions, our study shows that registration error in craniocaudal direction is significantly larger, which is consistent with previous published data and our previous study from both KVCT-MVCT and KVCT-CBCT registration techniques [8, 26–29]. This phenomenon may be caused by the trade-off against imaging speed and increase with the pitch of the MVCT scan [30]. Regarding to the craniocaudal registration errors, we found that it decreased in the pelvic phantom than in the chest phantom and we assumed it may come from the lower craniocaudal resolution in the chest but rather higher one in pelvis because of the complicated pelvic bony structures. Considering that the bony structures of the chest are similar in between transversal slices, the CT-MVCT registration protocol of craniospinal irradiation should probably be different from the other anatomic site such as head and neck and pelvic etc. Therefore, we suggest the craniocaudal scanning length should encompass at least one entire vertebra (including thoracic spine or lumbar spine) in craniospinal irradiation in order to correct the body roll direction and this claim should be substantiated by further research. Furthermore, the automatic registration algorithm could potentially eliminate inter-observer’s deviation altogether but also can impair local matches highly weighted by individual observers and may introduce a procedure-dependent new error [8]. Therefore, manually perform a double check and adjustment to the automatic registration result by radiation oncologist should be part of the strict CT-MVCT registration protocol in craniospinal irradiation, lest low craniocaudal resolution of MVCT may bring large registration error just as one vertebra is incorrectly matched to the adjacent one [30, 31].
MVCT scan takes approximate 10 s for one gantry rotation regardless of AP options, which will include 2~ 5 respiratory cycles for chest imaging. With the highest pitch, AP of ‘coarse’ will cover craniocaudal12mm per rotation. Considering that each fraction of scan should encompass the motion envelope of the target for accurate IGRT and ART, inspiratory motion and resultant imaging artifacts cannot be avoided by breath holding in TomoTherapy MVCT serials. A phantom based experimental study has proved that motion artifacts in TomoTherapy MVCT or planning CT studies changed the accuracy of the automatic registration process by less than 2.0% [21]. Based on this result, effects of AP and RI on IGRT and ART accuracy were investigated by static phantoms in this study, and we propose that it is possible to apply present findings to the moving targets or organs. However, this conclusion relies on two assumptions: first, moving target is much smaller than the total volume of the thorax imaged; second, respiratory movement of the organs like ribcage, diaphragm does not affect the registration process. If they cannot be met, gross motion may bring unexpected impact to IGRT and ART accuracy, which should be investigated by further research.
One of the probable resolution of gross respiratory motion and target shape change is deformable registration. As mentioned above, since manually performed double check is critical to accurate IGRT, a clinician may modify the automatic registration result based on their judgement on the alignment of the target. This may bring a new tradeoff between alignment of target and surrounded organs. Deformable registration can enroll more voxels and multi scales of mutual information during the match, which may meet the requirements from both target and organs, and bring more accurate registration in real situations. If so, we assume that AP of ‘fine’ with smaller RI may bring better deformable registration result and fewer residual errors because of its more sufficient imaging information. However, this assumption should also be substantiated by further research.
Beside of the translational corrections, TomoTherapy IGRT system can use the gantry angle to correct for the roll (along with the craniocaudal axis) adjustment automatically. TomoTherapy software allows a roll correction of any angle bigger than 0.1° (0.1° is the limiting value of gantry’s mechanical precision) prior to treatment. Rotational adjustment in one direction results in gantry rotation in the opposite direction to achieve the correct angular relationship between the radiation source and patient. However, roll correction was not enrolled in this study. We considered that if the image volume is rotated clockwise or counterclockwise, lateral and vertical adjustments become linked, and future lateral or vertical adjustments will proportionally affect each other. Since this study predominantly aimed to reveal the residual errors on 3 translational directions, rotational disagreement was eliminated to avoid its translational impacts in the experiment. However, if deformable registration method can be used in TomoTherapy IGRT module, rotational disagreement value and the corresponding correction may bring higher accuracy, which should also be substantiated by further research.
Conclusions
From our study, AP of ‘coarse’ with RI of 6 mm is recommended in ART planning to provide craniocaudal longer and faster MVCT scan, while registration technique of ‘full image’ should be used to avoid large residual error. Considering the trade-off between IGRT and ART, AP of ‘normal’ with RI of 2 mm was highly recommended in daily practice.
Additional files
Acknowledgements
We are grateful to Dr. Yang Ching-Chong Jack (Department of Radiation Oncology, Monmouth Medical Center, New Jersey, USA) and Ms. Chen Jiuhong (Accuray Medical Equipment, Shanghai, China) for reviewing the manuscript in English writing.
Funding
This work was supported by the National Natural Science Foundation of China (grant numbers 81671785, 81530060 and 81301298), the National Key Research and Develop Program of China (grant number 2016YFC0105106), the Taishan Scholar Foundation (grant number ts20120505), the China Postdoctoral Science Foundation (grant number 2016 T90412) and the Natural Science Foundation of Shandong Province (grant number ZR2014YL034).
Availability of data and materials
The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.
Abbreviations
- AP
Acquisition pitch
- ART
Adaptive radiotherapy
- CBCT
Cone beam CT
- DTA
Distance-to-agreement
- DVH
Dose volume histogram
- FOV
Field of view
- HT
Helical TomoTherapy
- IGRT
Image-guided radiotherapy
- KVCT
Kilovoltage computed tomographic
- MVCT
Megavoltage computed tomographic
- RI
Reconstruction interval
- ROI
Regions of interest
- TPS
Treatment planning system
Authors’ contributions
JZ designed the experiment, analyzed the experimental raw data, and was a major contributor in writing the manuscript. TB and JG executed the experiment process, recorded the data and revised the manuscript for important intellectual contents. ZS provided the two phantoms used in this experiment and revised the manuscript critically for important intellectual contents. YW executed the CT simulation and plan designing work on both phantoms used in this experiment. BL checked the experimental raw data and were secondly major contributors in writing the manuscript. YY made substantial contributions to conception and design the whole experiment. All authors read and approved the final manuscript.
Ethics approval and consent to participate
Not applicable. No human participants, human data or human tissue were involved in this phantom based experimental work of this paper.
Consent for publication
Not applicable
Competing interests
The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Footnotes
Electronic supplementary material
The online version of this article (10.1186/s13014-018-0989-y) contains supplementary material, which is available to authorized users.
Contributor Information
Jian Zhu, Phone: +86-531-67626427/907, Email: zhujian.cn@163.com.
Tong Bai, Email: baitong0411@163.com.
Jiabing Gu, Email: 1536919133@qq.com.
Ziwen Sun, Email: 1874603740@qq.com.
Yumei Wei, Email: wymlzjljy@163.com.
Baosheng Li, Email: baoshli1963@163.com.
Yong Yin, Phone: +86-531-67626427/907, Email: yinyongsd@126.com.
References
- 1.De Los SJ, Popple R, Agazaryan N, Bayouth JE, Bissonnette J-P, Bucci MK, Dieterich S, Dong L, Forster KM, Indelicato D. Image guided radiation therapy (IGRT) technologies for radiation therapy localization and delivery. Int J Radiat Oncol Biol Phys. 2013;87:33–45. doi: 10.1016/j.ijrobp.2013.02.021. [DOI] [PubMed] [Google Scholar]
- 2.Langen K, Meeks S, Poole D, Wagner T, Willoughby T, Kupelian P, Ruchala K, Haimerl J, Olivera G. The use of megavoltage CT (MVCT) images for dose recomputations. Phys Med Biol. 2005;50:4259. doi: 10.1088/0031-9155/50/18/002. [DOI] [PubMed] [Google Scholar]
- 3.Murphy MJ, Balter J, Balter S, BenComo JA, Jr, Das IJ, Jiang SB, Ma CM, Olivera GH, Rodebaugh RF, Ruchala KJ, et al. The management of imaging dose during image-guided radiotherapy: report of the AAPM task group 75. Med Phys. 2007;34:4041–4063. doi: 10.1118/1.2775667. [DOI] [PubMed] [Google Scholar]
- 4.Shah AP, Langen KM, Ruchala KJ, Cox A, Kupelian PA, Meeks SL. Patient dose from megavoltage computed tomography imaging. Int J Radiat Oncol Biol Phys. 2008;70:1579–1587. doi: 10.1016/j.ijrobp.2007.11.048. [DOI] [PubMed] [Google Scholar]
- 5.Chen M, Chao E, Lu W. Quantitative characterization of tomotherapy MVCT dosimetry. Med Dosim. 2013;38:280–286. doi: 10.1016/j.meddos.2013.02.009. [DOI] [PubMed] [Google Scholar]
- 6.Kaiser A, Schultheiss TE, Wong JY, Smith DD, Han C, Vora NL, Pezner RD, Chen Y-J, Radany EH. Pitch, roll, and yaw variations in patient positioning. Int J Radiat Oncol Biol Phys. 2006;66:949–955. doi: 10.1016/j.ijrobp.2006.05.055. [DOI] [PubMed] [Google Scholar]
- 7.Chen YJ, Han C, Liu A, Schultheiss TE, Kernstine KH, Shibata S, Vora NL, Pezner RD, Wong JY. Setup variations in radiotherapy of esophageal cancer: evaluation by daily megavoltage computed tomographic localization. Int J Radiat Oncol Biol Phys. 2007;68:1537–1545. doi: 10.1016/j.ijrobp.2007.04.023. [DOI] [PubMed] [Google Scholar]
- 8.Levegrun S, Pottgen C, Jawad JA, Berkovic K, Hepp R, Stuschke M. Megavoltage computed tomography image guidance with helical tomotherapy in patients with vertebral tumors: analysis of factors influencing interobserver variability. Int J Radiat Oncol Biol Phys. 2013;85:561–569. doi: 10.1016/j.ijrobp.2012.04.010. [DOI] [PubMed] [Google Scholar]
- 9.Suh S, Schultheiss TE. Distance-to-agreement investigation of Tomotherapy's bony anatomy-based autoregistration and planning target volume contour-based optimization. Int J Radiat Oncol Biol Phys. 2013;85:862–865. doi: 10.1016/j.ijrobp.2012.06.015. [DOI] [PubMed] [Google Scholar]
- 10.Ruchala KJ, Olivera GH, Kapatoes JM. Limited-data image registration for radiotherapy positioning and verification. Int J Radiat Oncol Biol Phys. 2002;54:592–605. doi: 10.1016/S0360-3016(02)02895-X. [DOI] [PubMed] [Google Scholar]
- 11.Xu L, Wan JW, Bian T. A continuous method for reducing interpolation artifacts in mutual information-based rigid image registration. IEEE Trans Image Process. 2013;22:2995–3007. doi: 10.1109/TIP.2013.2265880. [DOI] [PubMed] [Google Scholar]
- 12.Yu G, Liang Y, Yang G, Shu H, Li B, Yin Y, Li D. Accelerated gradient-based free form deformable registration for online adaptive radiotherapy. Phys Med Biol. 2015;60:2765–2783. doi: 10.1088/0031-9155/60/7/2765. [DOI] [PubMed] [Google Scholar]
- 13.Low DA, Dempsey JF. Evaluation of the gamma dose distribution comparison method. Med Phys. 2003;30:2455–2464. doi: 10.1118/1.1598711. [DOI] [PubMed] [Google Scholar]
- 14.Langen KM, Papanikolaou N, Balog J, Crilly R, Followill D, Goddu SM, Grant W, 3rd, Olivera G, Ramsey CR, Shi C. QA for helical tomotherapy: report of the AAPM task group 148. Med Phys. 2010;37:4817–4853. doi: 10.1118/1.3462971. [DOI] [PubMed] [Google Scholar]
- 15.Yoo S, Yin FF. Dosimetric feasibility of cone-beam CT-based treatment planning compared to CT-based treatment planning. Int J Radiat Oncol Biol Phys. 2006;66:1553–1561. doi: 10.1016/j.ijrobp.2006.08.031. [DOI] [PubMed] [Google Scholar]
- 16.Zeidan OA, Langen KM, Meeks SL, Manon RR, Wagner TH, Willoughby TR, Jenkins DW, Kupelian PA. Evaluation of image-guidance protocols in the treatment of head and neck cancers. Int J Radiat Oncol Biol Phys. 2007;67:670–677. doi: 10.1016/j.ijrobp.2006.09.040. [DOI] [PubMed] [Google Scholar]
- 17.Li XA, Qi XS, Pitterle M, Kalakota K, Mueller K, Erickson BA, Wang D, Schultz CJ, Firat SY, Wilson JF. Interfractional variations in patient setup and anatomic change assessed by daily computed tomography. Int J Radiat Oncol Biol Phys. 2007;68:581–591. doi: 10.1016/j.ijrobp.2006.12.024. [DOI] [PubMed] [Google Scholar]
- 18.Duma MN, Kampfer S, Schuster T, Aswathanarayana N, Fromm L-S, Molls M, Andratschke N, Geinitz H. Do we need daily image-guided radiotherapy by megavoltage computed tomography in head and neck helical tomotherapy? The actual delivered dose to the spinal cord. Int J Radiat Oncol Biol Phys. 2012;84:283–288. doi: 10.1016/j.ijrobp.2011.10.073. [DOI] [PubMed] [Google Scholar]
- 19.Yue Q, Duan J, Wang Z, Gu D, Yang X, Li R. The research on the factors of effecting the CT number and noise of TomoTherapy MVCT. Chin J Radiat Oncol. 2014;23:527–529. [Google Scholar]
- 20.Langen KM, Meeks SL, Pouliot J. Quality assurance of onboard megavoltage computed tomography imaging and target localization systems for on-and off-line image-guided radiotherapy. Int J Radiat Oncol Biol Phys. 2008;71:S62–S65. doi: 10.1016/j.ijrobp.2007.04.094. [DOI] [PubMed] [Google Scholar]
- 21.Woodford C, Yartsev S, Van Dyk J. Image registration of a moving target phantom with helical tomotherapy: effect of the CT acquisition technique and action level proposal. Phys Med Biol. 2008;53:5093–5106. doi: 10.1088/0031-9155/53/18/016. [DOI] [PubMed] [Google Scholar]
- 22.Ettinger DS, Akerley W, Borghaei H, Chang AC, Cheney RT, Chirieac LR, D'Amico TA, Demmy TL, Govindan R, Grannis FW, Jr, et al. Non-small cell lung cancer, version 2.2013. J Natl Compr Cancer Netw. 2013;11:645–653. doi: 10.6004/jnccn.2013.0084. [DOI] [PubMed] [Google Scholar]
- 23.Feng M, Yang C, Chen X, Xu S, Moraru I, Lang J, Schultz C, Li XA. Computed tomography number changes observed during computed tomography-guided radiation therapy for head and neck cancer. Int J Radiat Oncol Biol Phys. 2015;91:1041–1047. doi: 10.1016/j.ijrobp.2014.12.057. [DOI] [PubMed] [Google Scholar]
- 24.Li Z, Mao Y, Li H, Yu G, Wan H, Li B. Differentiating brain metastases from different pathological types of lung cancers using texture analysis of T1 postcontrast MR. Magn Reson Med. 2016;76:1410–1419. doi: 10.1002/mrm.26029. [DOI] [PubMed] [Google Scholar]
- 25.Chen Y, Shi LY, Feng QJ, Yang J, Shu HZ, Luo LM, Coatrieux JL, Chen WF. Artifact suppressed dictionary learning for low-dose CT image processing. Ieee T Med Imaging. 2014;33:2271–2292. doi: 10.1109/TMI.2014.2336860. [DOI] [PubMed] [Google Scholar]
- 26.Al-Wassia R, Bahig H, Poon E, Parker W, Freeman C. Daily setup uncertainty analysis for craniospinal irradiation using helical tomotherapy. Pract Radiat Oncol. 2013;3:349–355. doi: 10.1016/j.prro.2012.07.005. [DOI] [PubMed] [Google Scholar]
- 27.Han C, Schiffner DC, Schultheiss TE, Chen YJ, Liu A, Wong JY. Residual setup errors and dose variations with less-than-daily image guided patient setup in external beam radiotherapy for esophageal cancer. Radiother Oncol. 2012;102:309–314. doi: 10.1016/j.radonc.2011.07.027. [DOI] [PubMed] [Google Scholar]
- 28.Li H, Li B, Zhu J, Li X. Primary study of an offline dose compensation technique in a repeatable cyclic way to compensate for uncertainties after online correction with CBCT for NSCLC. Chin J Cancer Prev Treat. 2009;16:210–212. [Google Scholar]
- 29.Ryu S, Fang Yin F, Rock J, Zhu J, Chu A, Kagan E, Rogers L, Ajlouni M, Rosenblum M, Kim JH. Image-guided and intensity-modulated radiosurgery for patients with spinal metastasis. Cancer. 2003;97:2013–2018. doi: 10.1002/cncr.11296. [DOI] [PubMed] [Google Scholar]
- 30.Chao E, Bichay T, Lucas D, Ruchala K, Olivera G. SU-GG-J-120: longitudinal resolution of the TomoTherapy® MVCT image and potential improvements. Med Phys. 2008;35:2706. doi: 10.1118/1.2961669. [DOI] [Google Scholar]
- 31.Zhou J, Uhl B, Dewit K, Young M, Taylor B, Fei DY, Lo YC. Analysis of daily setup variation with tomotherapy megavoltage computed tomography. Med Dosim. 2010;35:31–37. doi: 10.1016/j.meddos.2009.01.005. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.