Abstract
Magnetic resonance fingerprinting (MRF) is a promising method for fast simultaneous quantification of multiple tissue parameters. The objective of this study is to improve the coverage of MRF based on echo-planar imaging (MRF-EPI) by using a slice-interleaved acquisition scheme. For this, the MRF-EPI is modified to acquire several slices in a randomized interleaved manner, increasing the effective repetition time of the spoiled gradient echo readout acquisition in each slice. Per-slice matching of the signal-trace to a precomputed dictionary allows the generation of T1 and T2* maps with integrated B1+ correction. Subsequent compensation for the coil sensitivity profile and normalization to the cerebrospinal fluid additionally allows for quantitative proton density (PD) mapping. Numerical simulations are performed to optimize the number of interleaved slices. Quantification accuracy is validated in phantom scans and feasibility is demonstrated in-vivo. Numerical simulations suggest the acquisition of four slices as a trade-off between quantification precision and scan-time. Phantom results indicate good agreement with reference measurements (Difference T1: −2.4 ± 1.1%, T2*: −0.5 ± 2.5%, PD: −0.5 ± 7.2%). In-vivo whole-brain coverage of T1, T2* and PD with 32 slices was acquired within 3:36 minutes, resulting in parameter maps of high visual quality and comparable performance with single-slice MRF-EPI at 4-fold scan-time reduction.
Introduction
Quantification of tissue properties has long been an overarching goal in Magnetic Resonance Imaging (MRI) research, allowing for inter-patient and inter-scan comparability1. Recently, signal quantification has achieved major clinical impact in multiple fields of MRI2–6. Neurological applications of quantitative MRI have gained interest with the introduction of magnetic resonance fingerprinting (MRF)7, due to the premise of fast simultaneous multi-parameter quantification. MRF is based on generating unique signal signatures, termed “fingerprints,” for different tissue types based on their underlying MRI properties. This is achieved by the rapid acquisition of numerous images with varying contrast weightings induced by the variation of sequence parameters including flip angle and echo time (TE). Matching these fingerprints to a precomputed dictionary allows parameter mapping of relaxation parameters including T1, T2 and T2* 8,9, tissue properties such as perfusion10 and system parameters such as B1+ 11. MRF has been used in clinical studies to evaluate the range and progression of MRF-derived relaxometry values in the brain as a function of the age of healthy volunteers12. Recently, a study demonstrated that MRF can differentiate common types of adult brain tumors, providing initial evidence for its clinical utility13.
The original MRF method was based on a balanced steady state free precession7 sequence design with highly undersampled spiral readout allowing for joint T1 and T2 mapping. Unbalanced fast imaging with steady state precession (FISP)8,14–16 was subsequently introduced to overcome sensitivity to B0 field inhomogeneities at the expense of reduced signal-to-noise ratio (SNR). Recently, we introduced an alternative MRF sequence for simultaneous T1 and T2* mapping based on spoiled gradient echo imaging with Cartesian echo-planar imaging k-space readout (MRF-EPI)9, potentially fostering robustness towards gradient deviations and trajectory inaccuracies compared with unspoiled spiral readouts17.
While MRF has successfully enabled efficient multi-parameter quantification in a single-slice, its applicability with improved coverage has been limited. Conventionally, full magnetization relaxation needs to be ensured prior to the acquisition of the next slice in order to use the same signal model for each measurement. A recent study has proposed shortened relaxation intervals, maintaining similar quantification accuracy at the cost of compromised precision due to lower SNR in the baseline images18. Further, simultaneous multi-slice (SMS) imaging19 has been incorporated into MRF20,21 by creating time-varying phase modulation between the acquired slices, in order to alleviate the problem of confined coverage and to increase scan-time efficiency. While obtaining an acceleration factor up to 3, computationally complex kernel fitting is needed for complete slice separation and additional training data must be acquired prior to the measurement, increasing overall measurement time. Further, quantification precision is compromised due to overlapping coil-geometries in the SMS reconstruction20. Most recently, 3D MRF methods were also proposed for improved spatial coverage22,23. In these studies highly regularized image reconstructions23 or repeated acquisition from the steady-state22 were used to enable reconstruct of a continuous imaging volume from a 3D stack-of-spirals k-space.
Slice-interleaved acquisition is a complimentary approach for volumetric imaging and clinical standard in numerous applications24, including diffusion MRI, fMRI and gradient echo sequences. Slice-interleaved schemes achieve similar SNR compared to 3D25 acquisitions, while allowing arbitrary slice spacing. Compared to single-slice imaging scan-efficiency is substantially improved due to increased effective TR, leading to higher SNR for each slice. However, the need for coherent signal-paths limits the effective TR in balanced sequence designs26, so far preventing the use of slice-interleaved acquisitions in MRF.
In this study, we sought to increase scan-time efficiency of volumetric coverage in MRF parameter mapping by integrating a slice-interleaved acquisition scheme in MRF-EPI. Spoiled gradient echo readouts enable increased effective TRs, ultimately enabling whole-brain coverage in clinically acceptable scan-times. The number of interleaved slices is numerically optimized to provide a trade-off between scan-time and quantification precision. Phantom experiments are performed to validate quantification accuracy of joint T1, T2* and proton density (PD) mapping. In-vivo images in healthy subjects and patients suffering from multiple sclerosis are obtained in order to study feasibility of whole-brain quantification for clinical usage and compare image quality to single-slice acquisitions.
Materials and Methods
Pulse sequence design
MRF-EPI9 is modified to allow for slice-interleaved acquisition of multiple slices (Fig. 1A). Following a global inversion pulse, numerous single-slice EPI readout modules are acquired in rapid succession. The slice position is varied in a pseudo-random fashion. The randomization is performed block-wise for groups containing the acquisition of each slice once. Slice order within these groups is randomly permuted. This ensures that each slice is sufficiently sampled throughout the measurement, while creating pseudo-random signal traces. Especially in the initial stages of the acquisition this guarantees that the sampling frequency of each slice is similar during the early part of the inversion recovery, ensuring comparable sensitivity. The pseudo-randomization influences the effective slice-TR (Fig. 1B), which includes the acquisition time of the other slices, leading to higher signal dissimilarities of the resulting fingerprints. As shown in Fig. 1C,D, TE and flip-angles are also varied to obtain sensitive tissue fingerprints as previously proposed8. Fat-suppression was included to minimize EPI imaging artifacts; gradient spoiling was incorporated by using crusher gradients of equal polarity performed before and after the fat-suppression pulse. For whole-brain coverage multiple interleaved slice-group acquisitions were performed, each simultaneously measuring four slices (Fig. 1A), separated by a 10 second pause to guarantee full magnetization recovery due to the global inversion pulse.
Dictionary
The dictionary was generated off-line using MATLAB (The MathWorks; Natick, MA) by simulating the evolution of the magnetization based on Bloch-equation simulations on a per-slice basis, as detailed by Rieger et al.9. B1+ compensation was integrated by simulating a scaling factor to the flip-angle excitation pulse14. Dictionary matching was performed by choosing the entry with the highest inner product between the magnitude of the dictionary entry and the magnitude of the measured signal. Due to the varying TR pattern of each slice within a slice group, a separate dictionary was precomputed for each unique TR pattern. Each per-slice MRF dictionary consisted of 157,938 entries with the following parameter range: B1+ amplitude correction14 using a linear scaling factor in the range of 0.6–1.4 in steps of 0.1, T1 = 20–2000 ms in steps of 10 ms, and 2000–6000ms in steps of 500 ms; T2* = 10–100 ms in steps of 2 ms, and 100–300 ms in steps of 5 ms. Unrealistic entries with T1 < T2* were discarded.
Proton Density (PD) Mapping
Numerous factors influence the voxel intensity in an MRI measurement, and need to be accounted for when quantifying the tissue PD1. In the present sequence, relaxation induced signal changes are incorporated as T1 and T2* in the dictionary model. Transmit radiofrequency field B1+ inhomogeneities are compensated using scaling of the effective flip-angle as described above. Semi-quantitative M0 maps, can be calculated from the matching dictionary entry f and the measured signal k, using a least-square fit with the closed form solution of
1 |
Here, the semi-quantitative M0 is characterized as
2 |
depending on PD ρ, coil sensitivities S and a constant scaling factor C that includes spatial-invariant scaling such as receiver gain and DICOM export window-leveling. The coil sensitivity maps S were obtained with the method described by Volz et al., based on the idea, that the proton density map is a combination of the data set M0 with full spatial resolution and a bias field comprising of low spatial frequencies27. It was shown that the coil sensitivity S can be calculated with a probabilistic framework including optimised parameters28, by which in-vivo images are registered, segmented and bias-corrected. The model incorporates a smoothness intensity variation estimation termed field bias, which is proportional to S27. The field bias map was calculated from the M0 map using the segmentation toolbox of the SPM12 software package (http://www.fil.ion.ucl.ac.uk/spm), using the default toolbox parameters: Regularization = 0.001, FWHM = 60 mm cutoff. To quantify ρ, C is calculated using cerebrospinal fluid (CSF) as a reference point, such that 100 percentage units (pu) corresponds to the known PD of 110.3 mol/l at 37° 29. The scaling factor was determined by manually placing regions of interests (ROIs) in CSF and normalizing these areas to 100pu.
Numerical Simulations
For a given scan-time the proposed sequence requires a trade-off between number of slices per slice group and number of baseline images per slice. However, an increased number of slices also increases the average effective slice TR, leading to higher baseline SNR. To determine the optimal number of slices for a given measurement time of 17 seconds per slice group, the stability of the parameter quantification from the MRF data was evaluated in dependence of the number of baseline images by estimating the noise amplification in the linearized MRF system for T1 and T2* parameters as described by Rieger et al.9. The average normalized noise amplification was calculated for fingerprints generated from 18 pairs of relaxation times in the in-vivo range (T1: 1000–2500 ms T2*: 50–70 ms) for 1 to 160 baseline images, each. The TE and flip angle patterns were interpolated according to the number of measurements and TR was maximized to reach a measurement time of 17 seconds.
To verify the analysis based on linearization, Monte-Carlo simulations were performed to determine the quantification accuracy. The same set of fingerprints, TE, TR and flip angles patterns were used as in the noise amplification simulation. Noise was added (n = 1000) to the simulated patterns and matched with the dictionary. The mean normalized quantification accuracy was calculated depending on the number of baseline images.
The randomization of the slice order was performed to achieve more homogenous quantification characterization across the slices, as compared with sequential order. Numerical simulations comparing the quantification precision between these slice ordering schemes are provided in Supplementary Information.
Acquisition
To test the performance of the sequence, phantom and in-vivo data were acquired on a 3T whole-body scanner (Magnetom Trio; Siemens Healthcare, Erlangen, Germany) using a standard 32-channel head array coil for in-vivo measurements and 30-channels of a body and spine array for phantom scans. This study was approved by the local institutional review board (Institutional Review Board II, Medical Faculty Mannheim, Germany), all subjects provided written informed consent prior to examination and all methods were performed in accordance with the relevant guidelines and regulations. For the proposed slice-interleaved MRF acquisition, following parameters were used: 4 slices per slice group, TE/TR = 17–78 ms/80–755 ms, flip angle = 4–58°, FOV = 220 × 220 × 140 mm3, slice-gap = 0.9 mm, voxel size = 1.0 × 1.0 × 3.0 mm3, band-width = 1136 Hz/pixel, partial-Fourier = 5/8, parallel imaging factor 3 with GRAPPA reconstruction, reference lines = 48 acquired in-place for each baseline image, acquisition time per slice group = 17 s, total number of baseline images = 160 (4 slices with 40 images each). The single-slice MRF-EPI used the same parameters, though only acquiring one slice with 160 baseline images and an acquisition time per slice = 17 s.
Phantom Experiments
The accuracy and precision of the sequence were evaluated in phantom experiments and compared to reference measurements, inversion-recovery turbo spin-echo for T1 (IR-TSE, 6 images, TI = 50–4000 ms, TE/TR = 6 ms/15 s, turbo factor = 16, matrix/FOV = 64 × 128/110 × 220 mm2, slice thickness = 3 mm, BW = 399 Hz/pixel, scan-time = 4 min 30 sec, two-parameter fit) and spoiled gradient echo for T2* (GRE, 6 images, TE = 5–300 ms, TR = 1000 ms, alpha = 15°, matrix/FOV = 64 × 128/110 × 220 mm2, slice thickness = 3 mm, BW = 390 Hz/pixel, scan-time = 6 min 18 sec, three parameter fit). Each phantom was acquired 10 times in separate measurements with the slice-interleaved MRF-EPI and single-slice MRF-EPI. Reference measurements were acquired once per phantom. The nine phantoms consisted of a single gadoterate-meglumine (Dotarem; Guerbet, Villepinte, France, concentration: 0.075–0.15 µmol/ml) doped agarose compartments each (concentration: 0.5–1.5%). The mean relaxation times were determined by manually drawing ROIs in the phantoms. The accuracy of the slice-interleaved MRF-EPI was determined by comparing the average deviation and the normalized root-mean-square error (NRMSE) between the method and the reference measurements. A two sample Student’s t-test was used to conclude if the single-slice and slice-interleaved MRF-EPI have significantly different means. P values less than 0.05 were considered to be significant.
Consistency within a slice group was tested to study quantification differences among the slices caused by different acquisition parameters. For each phantom four separate measurements (A, B, C, D) were performed with the slice-interleaved MRF sequence, each acquiring four slices (A1, …, A4, and B1, …B4, …). The slice group location was shifted among the four measurements in such a way that the center of the phantom was covered by a different one of the four slices in each measurement (i.e. the center of the phantom was covered by A1, B2, C3 and D4). Consistency among the four slices was defined as the difference between the measurement in A1, B2, C3 and D4, using the same manually drawn ROI.
PD mapping was evaluated in a phantom consisting of gadoterate-meglumine doped water. Reference PD maps were acquired using multiple GRE images with long TR to avoid T1 weighting, and varying echo-time to compensate for T2* decay (GRE, 5 images, TE = 3–60 ms, TR = 1500 ms, alpha = 90°, matrix/FOV = 64 × 64/220 × 220 mm2, BW = 390 Hz/pixel, scan-time = 7 min 50 sec, three parameter fit). Imaging was performed with the body coil, which was used both for transmit and receive. B1+ maps were acquired using a double-angle method (GRE, 2 images, TE = 10 ms, TR = 1500 ms, alpha = 45°/90°). Given the reciprocity assumption, as transmit and receive were performed with the same body coil, B1+ maps were also used for receive coil profile correction. Accordingly, PD maps were calculated with . The reference and slice-interleaved measurements were performed with a slice thickness of 10 mm with varying amounts of water and air within the slice, thus changing the PD depending on the water-air ratio. PD values of the slice-interleaved EPI and the reference measurements were compared by placing ROIs within the PD maps and acquiring average values in each. The constant scaling factor was chosen such that 100% water is normalized to 100pu in manually drawn ROIs in a full water compartment. The B1+ maps acquired using a double-angle method were also used in the correction of the MRF-EPI PD instead of bias field correction, as the latter probabilistic method is known to be unsuitable for phantom measurements28.
In-vivo experiments
Whole-brain in-vivo MRF quantifications were acquired with the proposed method in 6 healthy volunteers (4 men, 31 ± 6 years old) and 4 multiple sclerosis (MS) patients (2 men, 42 ± 5 years old). T1, T2* and PD values were obtained for white matter and grey matter by segmenting a slice of each healthy volunteer with the segmentation toolbox of SPM12. To avoid partial voluming effects in small structures, the masks were eroded with MATLAB image erosion algorithm (disk radius = 1 pixel). T1, T2* and PD for MS patients were determined by manually placing ROIs in the lesions, as identified on separate clinical measurements.
Data availability statement
The datasets generated during and the current study are not publicly available as data clean up and further analysis is currently underway but are available from the corresponding author on reasonable request.
Results
Numerical Simulations
Figure 2A shows the average noise amplification in the parameter recovery from MRF data as a function of the number of baseline images. The combined average noise amplification of T1 and T2* (thick blue curve) has a similar amplitude for 40 to 160 baseline images. The noise amplification increases rapidly, when reducing the number of images below 40. The separate T1 and T2* curves have different characteristics, while the T2* noise amplification is lowest with 29 baseline images, it increases slightly with more images, implying that for T2* quantification fewer higher SNR baseline images have higher noise resilience than many low SNR images. T1 has lowest noise amplification with a high number of baseline images with a slight increase until 20 baseline images. For less than 20 images, the amplification is rapidly rising, as a too low number of baseline images causes the inversion recovery curve to be sampled sparsely, decreasing noise resilience. Figure 2B shows the normalized quantification precision in the Monte-Carlo simulations as a function of the number of baseline images. The results show the same characteristics as the noise amplification using the linearized system (Fig. 2A) with similar average precision for 40 to 160 baseline images. Reducing the number of baseline images increases the effective TR, thereby enabling the acquisition of other slices during these pauses. As the average noise amplification has a similar value for the baseline images between 40 and 160, acquiring 4 slices for each slice group in the proposed acquisition results in the highest acceleration factor with only marginal loss of precision.
Phantom Experiments
Figure 3A shows the T1 and T2* maps of the phantom experiments using MRF-EPI and reference measurements. The proposed method yields homogeneous T1 estimates within the phantoms. T2* maps show a higher degree of variation than T1 maps, due to magnetic susceptibilities and large field inhomogeneities influencing the T2* values, both in the reference and MRF measurements. Figure 3B depicts the T1, T2* and PD quantification using the slice-interleaved MRF-EPI as compared to the reference method. Proposed MRF-EPI shows slight underestimation of T1 (deviation: −2.4 ± 1.1% [min: −4.5%, max −0.8%], NRMSE: 3.0%), T2* (−0.5 ± 1.5% [min: −2.7%, max 1.6%], NRMSE: 1.5%) and PD (−0.5 ± 7.2 pu [min: −11.6 pu, max 6.3 pu], NRMSE: 6.5%) values compared to the reference measurements.
The Bland-Altman plots in Fig. 3C compare the quantification accuracy of the single-slice MRF-EPI to the slice-interleaved MRF-EPI for T1 and T2*. No significant difference was found between the single and slice-interleaved MRF (p ≤ 0.05). The slice-interleaved implementation results on average in slightly higher values for T1 (24 ± 9 ms [min: 12 ms, max: 38 ms])) and T2* (2 ± 1 ms [min: 0 ms, max 3 ms]) maps.
Precision and accuracy were not affected by different TR patterns among the 4 slices within a slice group, resulting in a deviation of less than 1% for all parameters when quantifying the exact same slice location in a phantom as a test for inter-slice consistency.
In-vivo Experiments
Full brain MRF data with 32 slices was successfully acquired in all healthy volunteers and patients. Figure 4 shows representative T1, T2* and PD maps from a healthy volunteer. The mean T1/T2*/PD values of all healthy volunteers of an exemplary slice are: white matter: 746 ± 57 ms/57 ± 6 ms/72 ± 7 pu, grey matter 1200 ± 100 ms/53 ± 6 ms/92 ± 18 pu. While T2 values are generally higher in grey than white matter, similar values for T2* in grey and white matter are obtained. This is well in line with previous studies and attributed mostly to magnetic susceptibility30. Figure 5 shows two example slices of a healthy volunteer acquired with the slice-interleaved and single-slice MRF-EPI. Both methods yield visually comparable T1 and T2* maps. The volunteer has intracranial calcifications in the frontal lobe of the brain (white arrow), leading to signal dropouts in the region in T1 as well as T2* maps, which is visible in both methods. Results from a patient scan are depicted in Fig. 6 (female, 40 years). The lesion of the MS-patient is clearly visible in T1, T2* and PD maps (Fig. 6a, further images in Supplementary Figure S2). Exemplary T1, T2* and PD values of lesions from three MS patients are higher compared to surrounding tissue (1285 ± 200 ms/95 ± 17 ms/62 ± 0 pu, 1528 ± 113 ms/121 ± 21 ms/62 ± 0 pu, 1270 ± 66 ms/81 ± 12 ms/60 ± 0 pu), allowing clear quantitative discrimination especially in the T2* maps. Sample fingerprints of the MS patient retrieved from manually placed ROIs in grey matter, white matter and a lesion display clearly differentiable signal paths (Fig. 6B). The same trend and good differentiability is observed across all subjects, despite minor subject specific variations.
Discussion
In this work, we integrated slice-interleaved scanning in the MRF-EPI method to increase scan-time efficiency, enabling quantification with increased volumetric coverage with an acceleration factor of 4 compared to single-slice MRF. Acquisition of slice groups with four slices provided high quantification accuracy in phantom experiments, comparable to the single-slice implementation and in agreement with reference measurements. Whole-brain in-vivo scans with 32 slices were acquired within 3:36 minutes in multiple volunteers, resulting in robust and artifact-free T1, T2* and PD parameter maps.
In comparison to the original MRF-EPI with 160 baseline images, a reduced number of 40 baseline images were acquired per slice. Our numerical simulations of noise amplification and phantom experiments showed that a reduced number of baseline images with longer TR and therefore higher SNR can lead to comparable quantification precision compared with a high number of low SNR baseline images. The results indicate that by reducing the number of baseline images while increasing TR, T2* accuracy is enhanced due to the higher SNR images. T1 accuracy declines slightly, as less data is acquired during the inversion recovery period at the start of the measurement. Our simulations indicate that acquiring four interleaved slices sharing a global inversion pulse leads to the highest scan-time efficiency gain for values in the in-vivo range, without compromising accuracy. The T1 precision could possibly be improved by including further inversion pulses during the acquisition of the slices, facilitating even higher acceleration. Alternative sequence designs incorporating this idea warrant investigation in future studies.
Simultaneous multi-slice imaging has recently received increasing interest19, including quantitative applications31,32, as it provides means for scan-time acceleration where the only loss in SNR is due to coil geometries. SMS has recently been applied to trueFISP-based MRF in two studies20,33. While gaining a 3-fold acceleration, higher factors are currently limited by noise amplification and signal leaking. Further, the sequence needs to acquire additional training data per slice and the computationally intensive regridding algorithms pose a challenge, while the quantification precision is compromised. Combining SMS with EPI by applying a cyclic phase-shift among the k-space lines allows for an intuitive interpretation of the CAIPIRINHA approach as FOV shifts induced in separate bands. SMS-EPI is commercially available from a major vendor and is successfully integrated into several clinical and large scale cohort studies with previously reported acceleration rate of up to 1634. Therefore, the combination of SMS with the proposed slice-interleaved scheme bears great promise for ultra-rapid whole-brain quantification of multiple parameters.
Several methods have been recently proposed to shorten reconstruction times by altering the process of dictionary matching35–37. These can be readily incorporated in the quantification step of the proposed method, replacing conventional dictionary matching, in order to speed up post-processing. However, due to a lower number of baseline images, reconstruction times are typically less of a concern, as compared with other MRF methods. In the proposed scheme, as well as in other MRF techniques, spatial resolution is limited by the readout duration per image. Advanced undersampling and reconstruction or denoising techniques have been recently proposed in the context of MRF37–44 to facilitate increased undersampling rates or to improve noise performance, by exploiting structure or inter-dependencies mainly along the temporal dimension. As these approaches are applicable to a variety of sampling schemes or have specifically been demonstrated for EPI45, they are fully compatible with the proposed slice-interleaved MRF method. Due to the increased baseline SNR in the proposed approach, these reconstruction methods promise large gains in the feasible undersampling factor at minor loss in quantification quality. This may ultimately facilitate high-resolution quantification without scan-time penalty and warrants investigation in future studies.
Volumetric MRF sequences based on highly undersampled stack of spirals trajectories have demonstrated the feasibility of whole-brain quantitative T1 and T2 imaging. Liao et al.22 minimized scan time by including Cartesian GRAPPA with a factor of 3 in kz direction and simultaneously reducing the baseline images to 420 per slice, compared to 1000 in the original MRF-FISP sequence. However, to allow for sufficient k-space data to reconstruct a 3D volume per imaging contrast, repeated acquisitions of the signal train are performed from a steady-state. This enabled whole-brain quantification with 1 mm isotropic resolution within 7.5 min, although with extensive reconstruction complexity amounting to ~20 h computation time. Ma et al.23 used an interleaved sampling pattern, acquiring 4 interleaved slices per group and sequentially measuring multiple groups, while also reducing the number of baseline images per slice to 480 to decrease scan time. Further, the relaxation time between the slice groups was set to 3 seconds to reduce scan time, therefore preventing full relaxation before the acquisition of the next slice group, as previously proposed18. Within 5 min T1 and T2 maps of 48 slices with a resolution of 1.2 × 1.2 × 3.0 mm were acquired, though an additional B1 measurement was needed to correct for B1 inhomogeneity effects and improve the accuracy of T1 and T2 estimates. To achieve reconstruction of an imaging volume without the need for repeated acquisitions per contrast, image regularization was integrated with a sliding window reconstruction across multiple images. Several recent works have enabled 3D EPI or echo volumar imaging (EVI), based on highly, accelerated imaging Cartesian readouts46,47. Readout-times could be further shortened by exploiting dependencies between the acquisition of multiple images46, or by employing image regularization48. These methods offer interesting potential to enable the acquisition of multiple interleaved 3D volumes in the proposed sequence scheme, similar to the method of Ma et al. The combination of slice-interleaving and volumetric coverage allows to synergistically benefit of the SNR gain of both methods and will be subject of future studies.
Slice-interleaved 2D and 3D acquisitions are complimentary techniques for volumetric coverage, each offering a distinct profile of advantages. The SNR gain is reported to be very similar in many practical applications25. Comparing the proposed slice-interleaved MRF-EPI to a hypothetical 3D acquisition with varying coverage but constant scan time, the 3D acquisition would achieve constant SNR, due to 49. For 1 to 4 slices, no drawback in terms of SNR is observed for the slice-interleaved implementation as shown by the simulations (Fig. 2). Hence, in this regime the slice-interleaved MRF-EPI provides comparable to SNR to an idealized 3D implementation. While 3D sequences commonly allow for lower minimal slice thickness and improved slice profiles, they often result in higher undersampling of the k-space and might require more elaborate reconstructions schemes, while conventional reconstructions are applicable to the proposed scheme. Further, 3D sequences are limited to the acquisition of a continuous volume. Multislice 2D allow for arbitrary slice spacing, enabling time efficient coverage of larger volumes without compromising resolution, by using slice gaps. Thus slice-interleaved acquisitions are often preferred in clinical applications, including scout and overview scans, which have the highest demand in terms of volumetric coverage.
In the present sequence, relaxation periods are required between different slice-groups due to the application of a non-selective adiabatic inversion-pulse. This can be circumvented by the application of slice-selective inversion. However, this requires a multi-band inversion pulse covering all spatially separated slices within the slice group and introduces substantial contrast weighting on the B1+ profile of the inversion-pulse. B1+ correction of the inversion similar to the one in proposed by Buonincontri et al.14 can be integrated, and warrants further studies. Recently, it was shown that the relaxation periods can be significantly shortened between MRF pulse train repetitions by starting the repetition using a non-relaxed initial spin state18. While quantification accuracy is maintained, increased computational complexity is needed as the shortened relaxation times must be accounted for within the Bloch simulations. This approach might be used in combination with slice-interleaved MRF in future studies to minimize wait times and reduce acquisition time, at the cost of reduced precision due to lower baseline SNR.
Integrating B1+ compensation in MRF has been shown to improve quantification precision. This has been done by either using a Bloch-Siegert reference scan prior to the MRF measurement50 or by integration a scaling factor to the flip-angle excitation pulse in the dictionary simulation. As previously evaluated in MRF-EPI, the latter method was chosen as to increase quantification accuracy in the presence of imperfect excitation slice profiles and inhomogeneities in the transmit field without the need for additional scan time. However, this scheme assumes a single flip-angle representative of the slice-profile. This has been shown to be a valid approximation for small flip-angles in steady-state conditions51. To further overcome residual inaccuracies which might be observed with high flip-angles or pseudo-random acquisitions, actual slice-profile simulations can also be integrated in the dictionary, albeit at increased computational complexity.
The present slice-interleaved implementation has a similar computational complexity for the dictionary matching process to the single-slice MRF-EPI. On the one hand, four dictionaries need to be calculated, as each slice in a slice group has a unique TR pattern, while on the other hand the dictionary size per slice is reduced. Including slice-interleaved acquisitions in MRF favors a non-balanced spoiled sequence design, as the acquisition of interleaved slices hereby does not affect the signal paths of the other slices. MRF methods based on a balanced sequence design require coherent signal-paths limiting the effective TR, prohibiting the application in slice-interleaving without major adaptation.
PD mapping has a number of clinical applications, including multiple sclerosis52 and brain tumors53. However, PD mapping is challenging, as traditionally a number of separate measurements need to be performed to compensate for contrast induced intensity variation. These variations include relaxation processes, inhomogeneous transmit and receive fields, as well as potentially other contrast mechanisms1. MRF has been proposed as a promising method for fast, joint quantification of a number of parameters. Previous studies have included a first-step towards MRF PD mapping by providing semi-quantitative M0 maps7. However, to achieve PD quantification additional bias correction for B1+ and coil sensitivity maps is required. Furthermore, in a balanced sequence design as previously proposed the M0 measurement is confounded by residual contrast sensitivity towards molecular diffusion54 and magnetization-transfer55, potentially necessitating further corrections to obtain reliable PD maps. Due to the spoiled gradient echo contrast in MRF-EPI all necessary bias corrections can be performed in a two-step process without the need of additional measurements. A B1+ correction scheme was suggested by Buonincontri et al.14, including flip angle correction in the dictionary matching process, which has been integrated in MRF-EPI9. Coil sensitivity correction can be performed by calculating the field bias maps27 based on a probabilistic per-image framework28. The bias field maps have been shown to have a high in-vivo accuracy compared to separately measured coil sensitivity maps56. As spoiled gradient-echo imaging is commonly not associated to other contrast sensitivities, the proposed approach compensates for the portfolio of confounders commonly considered in previous PD mapping studies27. However, residual inaccuracies can be induced by deviations from the assumed signal model, including non-monoexponential transverse signal decay, as previously reported for complex tissue structures such as lung alveolus57.
Patient motion is one of the main causes of artefacts in clinical MRI. While MRF has been shown to be partially resilient to certain kinds of patient movement, initial results demonstrate that generic motion can induce significant intra-image variance to the MRF signal trace, resulting in motion artefacts in parameter maps58–60. This effect can be exacerbated if increased scan time is necessary for volumetric coverage. In the proposed scheme, the total duration between first and last data acquisition for any given slice, is comparable to previously proposed single slice sequences. Hence, the sensitivity to patient motion can be expected to be comparable to other MRF techniques. Furthermore, MRF seems well suited for correction of residual in-plane motion using co-registration of the baseline data, due to the rapid image acquisition. Motion correction schemes that are insensitive to contrast variation among the baseline images have been previously proposed61,62. Hence, further mitigation of in-plane motion effects using contrast invariant motion correction of the baseline images prior to dictionary matching warrants future investigation.
While the proposed method is used for joint T1 and T2* quantification, it could be extended for T1 and T2 quantification by incorporating refocusing pulses prior to the readout. This is commonly performed in multiple applications of EPI63, benefiting from increased SNR at the expense of increased minimal TE and longer scan times. However, this would also allow for an integrated assessment of diffusion biomarkers by including randomized diffusion gradients. Extending the portfolio of simultaneous quantification of biomarkers with the proposed method is subject of further research, and may facilitate the usage of these methods for a wider range of diseases.
The study and the proposed method have limitations regarding comparability with other methods and the number of patients. The proposed method uses the same TE and flip angle scheme as the single-slice MRF-EPI method, subsampled by a factor of four. To further increase accuracy or accelerate the sequence, optimization of sequence parameters may be necessary, which will be subject of further research. Due to the lack of the original MRF sequence at our center, no direct comparison could be performed regarding accuracy and precision. Further, in this study only a small number of subjects were scanned to prove the in-vivo feasibility. Larger cohorts with specific diseases remain to be evaluated.
Conclusion
In the study, we have demonstrated the feasibility to accelerate the MRF-EPI for volumetric coverage by a factor of four while maintaining quantification accuracy using a slice-interleaved acquisition scheme. Within 17 seconds four slices with a resolution of 1 × 1 × 3 mm3 are acquired, resulting in artifact free T1, T2* and PD maps.
Electronic supplementary material
Acknowledgements
NIH; Grant numbers: R00HL111410, NSF; Grant number: CCF-1651825. Partially funded by a Proyecto de Investigacion en Salud PI15/00587, S.L., A.S., integrated in the Plan Estatal I+D+I and co-funded by ISCIII-Subdirección General de Evaluación and European Regional Development Fund (ERDF). Red Española de Esclerosis Múltiple (REEM) (RD16/0015/0002, RD16/0015/0003, RD12/0032/0002, RD12/0060/01-02). S.L. is recipient of a Juan Rodes grant from the Instituto de Salud Carlos III (JR14/00015). We acknowledge the financial support of the Deutsche Forschungsgemeinschaft and Ruprecht-Karls-Universität Heidelberg within the funding programme Open Access Publishing.
Author Contributions
B.R. and S.W. conceived the study, evaluated the measurements and wrote the main manuscript text, B.R. and J.P. performed the measurements, S.L. and E.M.H. undertook the medical interpretation of the images and handled the volunteers and patients, M.A. and L.S. supported in the development and improvement of the method and experimental design of the study. All authors reviewed the manuscript.
Competing Interests
The authors declare no competing interests.
Footnotes
Sebastian Weingärtner and Lothar R. Schad jointly supervised this work.
Electronic supplementary material
Supplementary information accompanies this paper at 10.1038/s41598-018-24920-z.
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Tofts, P. Quantitative MRI of the Brain: Measuring Changes Caused by Disease (John Wiley & Sons, 2005).
- 2.Baksi AJ, Pennell DJ. T1 Mapping in Heart Failure. Circ. Cardiovasc. Imaging. 2013;6:861–863. doi: 10.1161/CIRCIMAGING.113.001178. [DOI] [PubMed] [Google Scholar]
- 3.Bulluck H, et al. Myocardial T1 Mapping. Circ. J. 2015;79:487–94. doi: 10.1253/circj.CJ-15-0054. [DOI] [PubMed] [Google Scholar]
- 4.Feng X, Deistung A, Reichenbach JR. Quantitative susceptibility mapping (QSM) and R2* in the human brain at 3T. Z. Für Med. Phys. 2017 doi: 10.1016/j.zemedi.2017.05.003. [DOI] [PubMed] [Google Scholar]
- 5.Radenkovic D, Weingärtner S, Ricketts L, Moon JC, Captur G. T1 mapping in cardiac MRI. Heart Fail. Rev. 2017;22:415–430. doi: 10.1007/s10741-017-9627-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Ross KA, et al. Comparison of Three Methods to Quantify Repair Cartilage Collagen Orientation. Cartilage. 2013;4:111–120. doi: 10.1177/1947603512461440. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Ma D, et al. Magnetic resonance fingerprinting. Nature. 2013;495:187–192. doi: 10.1038/nature11971. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Jiang Y, Ma D, Seiberlich N, Gulani V, Griswold MA. MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. Magn. Reson. Med. 2015;74:1621–1631. doi: 10.1002/mrm.25559. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Rieger, B., Zimmer, F., Zapp, J., Weingärtner, S. & Schad, L. Magnetic resonance fingerprinting using echo‐planar imaging: Joint quantification of T1 and relaxation times. Magn. Reson. Med.78, 1724–1733 (2016). [DOI] [PubMed]
- 10.Christen T, et al. MR vascular fingerprinting: A new approach to compute cerebral blood volume, mean vessel radius, and oxygenation maps in the human brain. NeuroImage. 2014;89:262–270. doi: 10.1016/j.neuroimage.2013.11.052. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Buonincontri, G., Schulte, R. F., Cosottini, M. & Tosetti, M. Spiral MR fingerprinting at 7T with simultaneous B1 estimation. Magn. Reson. Imaging41, 1–6 (2017). [DOI] [PubMed]
- 12.Badve C, et al. Simultaneous T1 and T2 Brain Relaxometry in Asymptomatic Volunteers using Magnetic Resonance Fingerprinting. Tomogr. J. Imaging Res. 2015;1:136–144. doi: 10.18383/j.tom.2015.00166. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Badve C, et al. MR Fingerprinting of Adult Brain Tumors: Initial Experience. Am. J. Neuroradiol. 2017;38:492–499. doi: 10.3174/ajnr.A5035. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Buonincontri G, Sawiak SJ. MR fingerprinting with simultaneous B1 estimation. Magn. Reson. Med. 2015;76:1127–1135. doi: 10.1002/mrm.26009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Assländer, J., Glaser, S. J. & Hennig, J. Pseudo Steady-State Free Precession for MR-Fingerprinting. Magn. Reson. Med.77, 1151–1161 (2017). [DOI] [PubMed]
- 16.Jiang, Y. et al. MR fingerprinting using the quick echo splitting NMR imaging technique. Magn. Reson. Med.77, 979–988 (2017). [DOI] [PMC free article] [PubMed]
- 17.Block KT, Frahm J. Spiral imaging: A critical appraisal. J. Magn. Reson. Imaging. 2005;21:657–668. doi: 10.1002/jmri.20320. [DOI] [PubMed] [Google Scholar]
- 18.Amthor, T. et al. Magnetic Resonance Fingerprinting with short relaxation intervals. Magn. Reson. Imaging41, 22–28 (2017). [DOI] [PubMed]
- 19.Barth M, Breuer F, Koopmans PJ, Norris DG, Poser BA. Simultaneous multislice (SMS) imaging techniques. Magn. Reson. Med. 2016;75:63–81. doi: 10.1002/mrm.25897. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ye, H. et al. Simultaneous multislice magnetic resonance fingerprinting (SMS-MRF) with direct-spiral slice-GRAPPA (ds-SG) reconstruction. Magn. Reson. Med.77, 1966–1974 (2017). [DOI] [PMC free article] [PubMed]
- 21.Jiang, Y. et al. Use of pattern recognition for unaliasing simultaneously acquired slices in simultaneous multislice MR fingerprinting. Magn. Reson. Med.78, 1870–1876 (2017). [DOI] [PMC free article] [PubMed]
- 22.Liao C, et al. 3D MR fingerprinting with accelerated stack-of-spirals and hybrid sliding-window and GRAPPA reconstruction. NeuroImage. 2017;162:13–22. doi: 10.1016/j.neuroimage.2017.08.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Ma, D. et al. Fast 3D magnetic resonance fingerprinting for a whole-brain coverage. Magn. Reson. Med.79, 2190–2197 (2018). [DOI] [PMC free article] [PubMed]
- 24.Fautz H-P, Paul D, Scheffler K, Hennig J. TRIM: TR independent multislice imaging. Magn. Reson. Med. 2004;51:1239–1246. doi: 10.1002/mrm.20093. [DOI] [PubMed] [Google Scholar]
- 25.Johnson G, Wadghiri YZ, Turnbull DH. 2D multislice and 3D MRI sequences are often equally sensitive. Magn. Reson. Med. 1999;41:824–828. doi: 10.1002/(SICI)1522-2594(199904)41:4<824::AID-MRM23>3.0.CO;2-1. [DOI] [PubMed] [Google Scholar]
- 26.Nielsen J-F, Nayak KS. Interleaved balanced SSFP imaging: artifact reduction using gradient waveform grouping. J. Magn. Reson. Imaging JMRI. 2009;29:745–750. doi: 10.1002/jmri.21628. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Volz S, Nöth U, Deichmann R. Correction of systematic errors in quantitative proton density mapping. Magn. Reson. Med. 2012;68:74–85. doi: 10.1002/mrm.23206. [DOI] [PubMed] [Google Scholar]
- 28.Ashburner J, Friston KJ. Unified segmentation. NeuroImage. 2005;26:839–851. doi: 10.1016/j.neuroimage.2005.02.018. [DOI] [PubMed] [Google Scholar]
- 29.Warntjes Jbm, Dahlqvist O, Lundberg P. Novel method for rapid, simultaneous T1, T*2, and proton density quantification. Magn. Reson. Med. 2007;57:528–537. doi: 10.1002/mrm.21165. [DOI] [PubMed] [Google Scholar]
- 30.Wansapura JP, Holland SK, Dunn RS, Ball WS. NMR relaxation times in the human brain at 3.0 tesla. J. Magn. Reson. Imaging. 1999;9:531–538. doi: 10.1002/(SICI)1522-2586(199904)9:4<531::AID-JMRI4>3.0.CO;2-L. [DOI] [PubMed] [Google Scholar]
- 31.Setsompop K, et al. Improving diffusion MRI using simultaneous multi-slice echo planar imaging. NeuroImage. 2012;63:569–580. doi: 10.1016/j.neuroimage.2012.06.033. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Weingärtner S, et al. Simultaneous multislice imaging for native myocardial T1 mapping: Improved spatial coverage in a single breath-hold. Magn. Reson. Med. 2017;78:462–471. doi: 10.1002/mrm.26770. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Ye H, et al. Accelerating magnetic resonance fingerprinting (MRF) using t-blipped simultaneous multislice (SMS) acquisition. Magn. Reson. Med. 2015;75:2078–85. doi: 10.1002/mrm.25799. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Moeller S, et al. Multiband multislice GE-EPI at 7 tesla, with 16-fold acceleration using partial parallel imaging with application to high spatial and temporal whole-brain fMRI. Magn. Reson. Med. 2010;63:1144–1153. doi: 10.1002/mrm.22361. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.McGivney DF, et al. SVD Compression for Magnetic Resonance Fingerprinting in the Time Domain. IEEE Trans. Med. Imaging. 2014;33:2311–2322. doi: 10.1109/TMI.2014.2337321. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Cauley SF, et al. Fast group matching for MR fingerprinting reconstruction. Magn. Reson. Med. 2015;74:523–528. doi: 10.1002/mrm.25439. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Yang M, et al. Low rank approximation methods for MR fingerprinting with large scale dictionaries. Magn. Reson. Med. 2018;79:2392–2400. doi: 10.1002/mrm.26867. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Wang, Z., Li, H., Zhang, Q., Yuan, J. & Wang, X. Magnetic Resonance Fingerprinting with compressed sensing and distance metric learning. Neurocomputing174, Part B, 560–570 (2016).
- 39.Pierre, E. Y., Ma, D., Chen, Y., Badve, C. & Griswold, M. A. Multiscale reconstruction for MR fingerprinting. Magn. Reson. Med.75, 2481–2492 (2016). [DOI] [PMC free article] [PubMed]
- 40.Assländer, J. et al. Low rank alternating direction method of multipliers reconstruction for MR fingerprinting. Magn. Reson. Med79, 83–96 (2018). [DOI] [PMC free article] [PubMed]
- 41.Zhao B, Setsompop K, Ye H, Cauley SF, Wald LL. Maximum Likelihood Reconstruction for Magnetic Resonance Fingerprinting. IEEE Trans. Med. Imaging. 2016;35:1812–1823. doi: 10.1109/TMI.2016.2531640. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Doneva M, Amthor T, Koken P, Sommer K, Börnert P. Matrix completion-based reconstruction for undersampled magnetic resonance fingerprinting data. Magn. Reson. Imaging. 2017;41:41–52. doi: 10.1016/j.mri.2017.02.007. [DOI] [PubMed] [Google Scholar]
- 43.Zhao B, et al. Improved magnetic resonance fingerprinting reconstruction with low-rank and subspace modeling. Magn. Reson. Med. 2018;79:933–942. doi: 10.1002/mrm.26701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Liao C, et al. Acceleration of MR Fingerprinting with Low Rank and Sparsity Constraint. In Proceedings of the 24th Annual Meeting ISMRM. 2016;2016:4227. [Google Scholar]
- 45.Davies M, Puy G, Vandergheynst P, Wiaux Y. A Compressed Sensing Framework for Magnetic Resonance Fingerprinting. SIAM J. Imaging Sci. 2014;7:2623–2656. doi: 10.1137/130947246. [DOI] [Google Scholar]
- 46.Afacan O, Hoge WS, Janoos F, Brooks DH, Morocz IA. Rapid full-brain fMRI with an accelerated multi shot 3D EPI sequence using both UNFOLD and GRAPPA. Magn. Reson. Med. 2012;67:1266–1274. doi: 10.1002/mrm.23106. [DOI] [PubMed] [Google Scholar]
- 47.Posse, S. et al. High-Speed Real-Time Resting-State fMRI Using Multi-Slab Echo-Volumar Imaging. Front. Hum. Neurosci. 7 (2013). [DOI] [PMC free article] [PubMed]
- 48.Jung H, Sung K, Nayak KS, Kim EY, Ye JC. k-t FOCUSS: A general compressed sensing framework for high resolution dynamic MRI. Magn. Reson. Med. 2009;61:103–116. doi: 10.1002/mrm.21757. [DOI] [PubMed] [Google Scholar]
- 49.Bernstein, M., King, K. & Zhou, X. J. Handbook of MRI Pulse Sequences (Academic Pr Inc, 2004).
- 50.Ma D, et al. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn. Reson. Med. 2017;78:1781–1789. doi: 10.1002/mrm.26580. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Weingärtner S, Roujol S, Akçakaya M, Basha TA, Nezafat R. Free-breathing multislice native myocardial T1 mapping using the slice-interleaved T1 (STONE) sequence. Magn. Reson. Med. 2015;74:115–124. doi: 10.1002/mrm.25387. [DOI] [PubMed] [Google Scholar]
- 52.Laule C, et al. Water content and myelin water fraction in multiple sclerosis. J. Neurol. 2004;251:284–293. doi: 10.1007/s00415-004-0306-6. [DOI] [PubMed] [Google Scholar]
- 53.Neeb H, Zilles K, Shah NJ. A new method for fast quantitative mapping of absolute water content in vivo. NeuroImage. 2006;31:1156–1168. doi: 10.1016/j.neuroimage.2005.12.063. [DOI] [PubMed] [Google Scholar]
- 54.Buxton RB. The diffusion sensitivity of fast steady-state free precession imaging. Magn. Reson. Med. 1993;29:235–243. doi: 10.1002/mrm.1910290212. [DOI] [PubMed] [Google Scholar]
- 55.Gloor M, Scheffler K, Bieri O. Quantitative magnetization transfer imaging using balanced SSFP. Magn. Reson. Med. 2008;60:691–700. doi: 10.1002/mrm.21705. [DOI] [PubMed] [Google Scholar]
- 56.Weiskopf N, et al. Unified segmentation based correction of R1 brain maps for RF transmit field inhomogeneities (UNICORT) NeuroImage. 2011;54:2116–2124. doi: 10.1016/j.neuroimage.2010.10.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Zapp J, Domsch S, Weingärtner S, Schad LR. Gaussian signal relaxation around spin echoes: Implications for precise reversible transverse relaxation quantification of pulmonary tissue at 1.5 and 3 Tesla. Magn. Reson. Med. 2017;77:1938–1945. doi: 10.1002/mrm.26280. [DOI] [PubMed] [Google Scholar]
- 58.Mehta, B. B., Ma, D., Coppo, S. & Griswold, M. A. Image Reconstruction Algorithm for Motion Insensitive Magnetic Resonance Fingerprinting (MRF). in Proceedings of the 25th Annual Meeting of the International Society forMagnetic Resonance in Medicine (ISMRM), Honolulu, Hawaii, USA, 0302 (2017).
- 59.Cruz, G., Botnar, R. & Prieto, C. Motion corrected Magnetic Resonance Fingerprinting using Soft-weighted key-Hole (MRF-McSOHO). In Proceedings of the 25th Annual Meeting of the International Society forMagnetic Resonance in Medicine (ISMRM), Honolulu, Hawaii, USA, 0935 (2017).
- 60.Rieger, B., Wenning, M., Akçakaya, M., Schad, L. R. & Weingärtner, S. Improved motion robustness in Magnetic Resonance Fingerprinting using Intensity based Motion Correction. In Proceedings of the ISMRM Workshop on Motion Correction in MRI & MRS, Cape Town, South Africa (2017).
- 61.Roujol S, Foppa M, Weingärtner S, Manning WJ, Nezafat R. Adaptive registration of varying contrast-weighted images for improved tissue characterization (ARCTIC): application to T1 mapping. Magn. Reson. Med. 2015;73:1469–1482. doi: 10.1002/mrm.25270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Xue H, et al. Motion correction for myocardial T1 mapping using image registration with synthetic image estimation. Magn. Reson. Med. 2012;67:1644–1655. doi: 10.1002/mrm.23153. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Tyler DJ, Moore RJ, Marciani L, Gowland PA. Rapid and accurate measurement of transverse relaxation times using a single shot multi-echo echo-planar imaging sequence. Magn. Reson. Imaging. 2004;22:1031–1037. doi: 10.1016/j.mri.2004.01.069. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets generated during and the current study are not publicly available as data clean up and further analysis is currently underway but are available from the corresponding author on reasonable request.