Abstract
As part of the scientific community’s development of medical countermeasures against Ebola virus disease, optimization of standardized assays for product evaluation is paramount. The recent outbreak heightened awareness to the scarcity of available assays and limited information on performance and reproducibility. To evaluate the immunogenicity of vaccines entering Phase I–III trials and to identify survivors, two enzyme-linked immunosorbent assays, the Filovirus Animal Non-Clinical Group assay and the Alpha Diagnostics International assay, were evaluated for detection of immunoglobulin G against Ebola virus glycoprotein. We found that the Filovirus Animal Nonclinical Group assay produced a wider range of relative antibody concentrations, higher assay precision, larger relative accuracy range, and lower regional background. Additionally, to sufficiently power a vaccine trial, use of the Filovirus Animal Nonclinical Group assay would require one third the number of participants than the Alpha Diagnostics International assay. This reduction in needed study participants will require less money, fewer man hours, and much less time to evaluate vaccine immunogenicity.
Keywords: Ebola virus, ELISA, Serology, FANG, ADI, Antibodies, Immune response
1. Introduction
Over 11,000 individuals succumbed to the recent and largest Ebola virus disease (EVD) outbreak that occurred in Western Africa (World Health Organization, 2016c; World Health Organization, 2016b). Though many individuals recovered from disease, clear evidence demonstrates that Ebola virus (EBOV) can persist in EVD survivors long after signs and symptoms have subsided (MacDermott and Bausch, 2016; World Health Organization, 2016a). Persistent genomic virus RNA signals have been detected in EVD survivors over a year after resolution of both clinical disease and viremia(Crozier, 2016; Deen et al., 2017; Uyeki et al., 2016). As a result, sexual and vertical transmission of the virus may occur, as evidenced by at least 3 EVD clusters in Liberia, Sierra Leone, and Guinea (Arias et al., 2016; Blackley et al., 2016;Fischer et al., 2016; Fischer and Wohl, 2016; Mate et al., 2015; Sow et al., 2016; Yang et al., 2015). Though the outbreak has ended, countermeasures should be developed to combat any potential reintroductions of EVD into the population by these or other traditional transmission routes. Paramount to these countermeasures is an effective and long-lasting vaccine.
For effective and durable protection, a vaccine must elicit a strong and sustained adaptive immune response against a critical viral target and disrupt the viral replication cycle (e.g., infection, integration, replication). Since the discovery of EVD in 1976, a variety of assays have been developed to detect antibodies targeting filoviruses. While critical in the development of subsequent methods, widespread use of these assays has been limited by reagent quantity, batch-to-batch variability, and a lack of characterization and standardization. These shortfalls are expected as these assays were often developed by researchers for small scale and research-only use; they were not intended for advancement of countermeasures or developed with the intent of seeking regulatory approval. With a standardized assay, the scientific community could compare and replicate assay results across institutions.
Here we present a comparison between two enzyme-linked immunosorbent assays (ELISAs) measuring relative antibody concentrations (rel[Ab]) of immunoglobulin G against EBOV glycoprotein (GP), an immune response thought to be a correlate of protection from death related to EVD (Pushko et al., 2000; Wilson and Hart, 2001; Wong et al., 2012; Xu et al., 1998). The first ELISA was developed by the Filovirus Animal Nonclinical Group (FANG), a consortium of researchers and institutes attempting to standardize multiple aspects of filovirus research. This assay is currently submitted to the Food and Drug Administration for approval as a validated assay. The second ELISA is the commercially available Alpha Diagnostic International (ADI) ELISA. Rel [Ab] range, assay precision, baseline background, correlation between assays, study power requirements, and relative assay accuracy expressed as dilutional linearity were used as points of comparison using identical sample sets for each assay. For this investigation, rel[Ab] range is defined as the range of rel[Ab] observed using each assay for a specific set of samples. Preliminary data collected for the Partnership for Research on Ebola Virus in Liberia (PREVAIL) I vaccine trial suggested that the FANG assay provides less variable rel[Ab] due to plate format, analysis method, and stringent acceptance criteria. Therefore, we applied some aspects of the FANG assay, including a 4-parameter logistic (4-PL) calibrator fit, sample masking criteria, and a pooled serum positive control, to the ADI ELISA to create an optimized ADI assay (ADIO). Additionally, analysis methods from the FANG assay were modified and applied to previously collected ADI ELISA data retrospectively (ADI+) to increase precision and avoid reprocessing samples using the ADIO assay. As the FANG assay is currently in the process of FDA validation and is not commercially available, we present this ADI+ and ADIO assay as alternatives. Due to a lack of reagent and sample availability, the ADIO and ADI+ ELISAs were only compared to the FANG and ADI assays to establish assay precision and relative assay accuracy.
2. Materials and methods
2.1. Origin of samples
Serum samples were collected for a double-blind, randomized, placebo-controlled Phase II efficacy trial of the candidate EBOV vaccines: recombinant chimpanzee adenovirus expressing EBOV GP (rChAd3-EBO-Z) vaccine and recombinant vesicular stomatitis virus modified to replace gene encoding G glycoprotein with the gene encoding EBOV GP (rVSVΔG-ZEBOV-GP) vaccine (Partnership for Research on Ebola Virus in Liberia I trial [PREVAIL I] ClinicalTrials.gov registration no. NCT02344407). Collection of samples from 1500 healthy adult Liberians with no reported history of EVD was approved by the Ministry of Health, Liberia (Kennedy et al., 2016). The protocol used in this clinical vaccine trial was approved by the National Research Ethics Board of Liberia, National Cancer Institute institutional review board, the Liberian Medicines and Health Products Regulatory Authority, and the Food and Drug Administration. Written informed consent was obtained from all the participants.
2.2. Experimental design
Adults enrolled in this trial were randomized 1:1:1 to three groups to receive an intramuscular injection of: recombinant chimpanzee adenovirus expressing rChAd3-EBO-Z vaccine (1×1011 particle units/ mL; 2 mL), rVSVΔG-ZEBOV-GP vaccine (2×107 particle units/mL; 1 mL), or placebo (saline) (Kennedy et al., 2017). Sera were separated from blood samples collected during the treatment visit (baseline) and 1-month post-vaccination. These samples were processed on the FANG and ADI ELISAs to quantify the antibody response to EBOV GP and to determine the correlation between the two assays.
A positive reference sample was established by equally pooling sera from five EVD survivors enrolled in the PREVAIL III Ebola Natural History Study (NCT02431923) with previously detected high GP-IgG titers as determined by both the FANG and ADI ELISAs. Pooled sera were diluted to form three samples representing high, medium, and low titers (semi-quantitative). EBOV-naïve serum was collected from one healthy, American volunteer. EBOV-naïve sera were also collected from 100 participants in a Mali human immunodeficiency virus study from 2004–2011.
2.3. Filovirus animal non-clinical group enzyme-linked immunosorbent assay
The methods used to perform the FANG ELISA were described in the PREVAIL I vaccine trial (Kennedy et al., 2017; Sabourin, 2014). Prior to test sample addition, Immulon 2 HB microplates (Thermo Fisher Scientific, Walkersville, MD, USA) were coated with 100 µL of recombinant EBOV GP and incubated at 4 °C for at least 14 h in the absence of light. Following this incubation, test sample dilutions were prepared in a dilution block in duplicate as a 6-point, 1:2 dilution series on a single plate. The starting dilution of 1:62.5 was prepared using an ELISA diluent containing 5% weight-by-volume dry milk (LabScientific, Highlands, NJ, USA) and 0.1% volume-by-volume Tween-20 (Sigma Aldrich, St. Louis, MO, USA) in phosphate buffered saline (Gibco, Gaithersburg, MD, USA). The reference standard (calibrator) was also prepared in a dilution block as an 11-point, 1:2 dilution series using the ELISA diluent described above. Additionally, a negative control was prepared as a 1:50 dilution using the ELISA diluent, and a high concentration quality control and a low concentration quality control were diluted as a sample processed in singlet. Following this preparation, EBOV GP-coated plates were washed 3 times with 300 µL phosphate buffered saline with 0.1% volume-by-volume Tween-20 (FANG wash buffer) and 100 µL of the sample dilutions, reference standard dilutions, and controls were transferred to the plate. Plates were then incubated at 37 °C for 1 h and washed 3 times with FANG wash buffer, and 100 µL of the horseradish peroxidase-conjugated, anti-human IgG (Thermo Fisher Scientific, Walkersville, MD, USA) was added. Plates were then incubated for an additional hour at 37 °C before washing an additional 3 times with 300 µL of FANG wash buffer and adding TMB Substrate (Thermo Fisher Scientific, Walkersville, MD, USA). Following a 30- minute incubation at room temperature in the absence of light, 100 µL of TMB stop solution (Thermo Fisher Scientific, Walkersville, MD, USA) was added to each well and the plates were read on a SpectraMax Plus 384 plate reader (Molecular Devices, Sunnyvale, CA, USA) at 450 nm. Individual well background optical densities were measured at 650 nm and subtracted from the 450-nm optical density readings prior to performing any calculations. A 4-PL curve were fit to the reference standard using Softmax Pro software provided by the manufacturer. Sample replicate dilutions were masked if optical density readings were outside the bounds of the 4-PL curve. Back-calculated rel[Ab] for each sample replicate dilution were calculated based on the 4-PL calibrator fit and the dilution factor in Softmax Pro. An average and %CV for all sample replicate dilutions were then calculated. If the %CV was greater than a 20% threshold, the sample dilution with a back calculated value furthest from the mean was masked until the %CV fell below that 20% threshold. If the %CV did not fall below 20% or too many dilution points were masked, the sample would fail, and the process was repeated. A sample replicate found to have 0–1, 2–4, or 5 dilution points with optical density readings below the bounds of the reference standard curve required at least 3, 2, or 1 unmasked dilution points to pass acceptance criteria, respectively. If both sample replicates passed, an average rel[Ab] and %CV would be calculated between replicates. If that %CV was greater than 30%, then the sample would fail and be repeated. When all acceptance criteria were met, rel[Ab] were reported.
2.4. Alpha Diagnostics International enzyme-linked immunosorbent assay
The ADI ELISA plates were processed according to manufacturer’s instructions (Alpha Diagnostics International, TX, USA). All reagents and plates were provided by the manufacturer. Samples were diluted in dilution blocks as a 6-point, 1:2 dilution series with a 1:250 starting dilution in ADI sample diluent. Then, 100 µL of each sample dilution and ADI plate calibrators, and controls were transferred to ADI microtiter plates pre-coated with recombinant EBOV GP and incubated for 60 min at room temperature. Plates were then washed four times with 300 µL of ADI wash buffer and 100 µL of ADI Anti-Human IgG horseradish peroxidase was added to each well. Following a 30-minute incubation at room temperature, plates were washed an additional 5 times with 300 µL of ADI wash buffer before 100 µL of 3, 3′, 5, 5′-tetramethylbenzidineTMB substrate was added to each well and incubated at room temperature in complete darkness for 15 min. ADI stop solution (100 µL) was then added to each well, and plates were read on a BioTek ELx808 plate reader (Winooski, VT) at 450 nm within 30 min of stop solution addition. Individual well background optical densities were recorded at 630 nm and subtracted from the 450 nm reading prior to any calculations. Rel[Ab] were calculated for each sample dilution based on the line of best fit through the optical densities recorded for the 4 pre-made assay calibrators supplied by the manufacturer using the Gen5 software provided by BioTek. A rel[Ab] of 0 ELISA units per milliliter (EU/mL) was assigned if all dilution points were found to have an optical density below the optical density of the 1 EU/mL calibrator.
2.5. Alpha Diagnostics International+ enzyme-linked immunosorbent assay
Plates were prepared and processed as described in the ADI method. Unlike the ADI ELISA, rel[Ab] were retrospectively calculated based on a 4-PL model fit (Findlay and Dillard, 2007) to the averaged optical densities for each sample recorded for the 4 pre-made calibrators and the assay blank. Back-calculated rel[Ab] for each average sample dilution were then calculated based on the dilution factor. The %CV threshold of 20% and dilution point masking used for the FANG ELISA were used as acceptance criteria for the ADI+ ELISA.
2.6. Alpha Diagnostics InternationalO enzyme-linked immunosorbent assay
Serum samples were prepared as described in the ADI method. The 10 unit calibrator was serially diluted 1:2 in ADI sample diluent for a total of six dilution points. Rel[Ab] were calculated for each sample dilution based on the calibrator 4-PL curve of best fit. Additionally, the positive reference (described previously) was processed on every plate as an additional positive control. Back-calculated rel[Ab] for each sample replicate dilution were calculated based on 4-PL calibrator fit and the dilution factor. An average and %CV for all sample replicate dilutions were then calculated. The %CV threshold of 20% and dilution point masking used for the FANG ELISA were used as acceptance criteria for the ADIO ELISA. If both sample replicates passed, an average rel[Ab] and %CV would be calculated between replicates. If that %CV was greater than 30%, then the sample would fail, and the process would be repeated.
2.7. Relative antibody concentration range and correlation
The rel[Ab] range following PREVAIL I trial vaccinations were compared using the FANG and ADI ELISAs. The difference in median rel [Ab] between the FANG and ADI assays were compared within each vaccine group at 1-month using the Wilcoxon Signed Rank test. Additionally, Spearman correlations between the rel[Ab] for FANG and ADI assays were calculated for all baseline and 1-month samples by vaccine group to measure agreement between assays.
2.8. Vaccine study power comparison
The change in rel[Ab] between baseline and 1-month post-vaccination was compared between the combined active vaccine groups and the placebo group for each assay using Welch’s t-test (non-equal variances). Rel[Ab] were adjusted by adding 1 to every value (to avoid zeros) and log10 transforming the data. All rel[Ab] changes are listed in log10 EU/mL. The ratio (ADI versus FANG) of sample sizes required to power a future vaccine study using these assays was calculated as the squared ratio of the z-score for each ELISA. A 95% confidence interval for the ratio was calculated using the bootstrap procedure.
2.9. Dilutional linearity (relative accuracy)
A high titer, EVD survivor serum pool (positive reference) was serially diluted 1:2 in EBOV-naïve serum to create a total of 9 individual samples encompassing dilutions from 0 to 1:256. Each sample was processed in duplicate on the FANG, ADI, ADI+, and ADIO ELISAs by two independent operators. The sample set was then plotted as log10 rel [Ab] vs. dilution factor. A linear regression (18 points) model was fit to the data, and a slope and 90% confidence interval were calculated. If the confidence interval was found not to be between 0.7 and −1.2, the most dilute sample was removed until the 90% confidence interval fell within that bound.
2.10. Assay precision
Serum from an EVD survivor with a rel[Ab] greater than both seroconversion cutoffs (FANG and ADI cutoffs) described previously was diluted in EBOV-naïve serum to create a total of three samples with a semi-quantitative high, medium, and low GP-IgG rel[Ab] for each assay. Individual sample sets (3) were created for each assay and processed on either the FANG, ADI, ADI+, or ADIO ELISA by two distinct operators for a total of 24 replicates for each assay and sample. Assay precision was then established at these three semi-quantitative rel[Ab] by calculating the coefficient of variation.
2.11. Regional baseline determination
One hundred presumably EBOV-naïve serum samples from Mali were run on the FANG and ADI ELISAs. A suggested qualitative seroconversion cut-off value for each assay was assigned based on the average rel[Ab] plus three standard deviations using the log10 transformed data.
3. Results
3.1. Relative antibody range and correlation of FANG and ADI ELISAs
The median and interquartile ranges (IQR) of rel[Ab] at baseline and 1-month for the 1500 Prevail 1 participants using the FANG and ADI assays are presented in Fig. 1 by treatment group. The median and IQR rel[Ab] humoral vaccine responses for both active vaccine groups observed at 1 month were greater in the samples tested by the FANG ELISA than when these samples were tested by the ADI ELISA (p < 0.001 between FANG and ADI for both vaccine groups individually).
Additionally, Spearman correlations between the rel[Ab] for the FANG and ADI ELISAs were calculated to measure agreement between assays. Spearman correlation coefficients ranged from 0.213 and 0.274 at baseline among the three groups and ranged from 0.303 and 0.674 at 1-month (Table 1). Correlations were greater at 1-month for the active vaccine groups than for the placebo group.
Table 1.
Time of evaluation | rChAd3 EBO Z | rVSVΔG-ZEBOV GP | Placebo |
---|---|---|---|
Baseline | 0.213 | 0.269 | 0.274 |
1 mo post-vaccination | 0.584 | 0.674 | 0.303 |
Total no. tested | 487 | 484 | 491 |
ChAd3 EBO Z, chimpanzee adenovirus type 3 expressing Ebola virus glycoprotein Yambuku variant; rVSVΔG-ZEBOV, recombinant vesicular stomatitis virus with substitution of gene for Ebola virus glycoprotein for the VSV G glycoprotein.
3.2. Vaccine study power comparison of ADI and FANG ELISAs
The change in rel[Ab] (log scale) for the vaccine and placebo groups as measured on the FANG and ADI ELISAs at 1-month post-vaccination compared to baseline was calculated and the distribution graphed on a histogram (Fig. 2). The average and standard deviation of the changes (log10 EU/mL) are given in Table 2. The variation in change within the active groups was smaller with the FANG assay then with the ADI assay, resulting in a larger Z-score when comparing active to placebo groups. The ratio (FANG versus ADI) of sample sizes required to power a vaccine study using these assays was calculated as the squared ratio of the resultant z-scores. Based on the calculated squared z-score ratio of 3.1 (95% confidence interval 2.5, 3.9), we estimate that the ADI ELISA would require about three times as many participants than if the FANG assay was used.
Table 2.
ELISA type | Avg rel[Ab] difference log10 EU/ml (SD), Vaccine groups |
Avg rel[Ab] difference log10 EU/ml (SD), Placebo group |
Z score |
---|---|---|---|
FANG | 0.964 (0.613) | −0.043 (0.326) | 41.15 |
ADI | 1.314 (1.469) | −0.221 (1.024) | 23.33 |
Total tested | 978 | 494 |
ADI, Alpha Diagnostic International; ELISA, enzyme-linked immunosorbent assay; FANG, Filovirus Animal Nonclinical Group; rel[Ab], relative antibody concentration.
3.3. Dilution linearity (relative accuracy) of the ADI, ADI°, ADI+, and FANG ELISAs
Serum pooled from EVD survivors (positive reference) was serially diluted 1:2 in EBOV-naïve serum to create a total of 9 individual samples encompassing dilutions from 0 to 1:256. These samples were processed by the ADI, ADI+, ADIO, and FANG ELISAs for rel[Ab] and plotted separately for each assay with a regression line of best fit determined (Fig. 3, blue line). If the slope and 90% confidence interval were not found to be between −0.78 and −1.20, the most dilute sample was removed until an acceptable linear fit was found (red line, see methods) The slopes [90% confidence interval] for each full dilution regression fit are: −0.893 [−0.916, −0.871], −1.573 [−1.812, −1.335], −0.620 [−0.675, −0.564], and −0.631 [−0.702, −0.561] for the FANG, ADI, ADIO, and ADI+ assays, respectively. By the above definition, the full dilution regression fit on the FANG ELISA was found to be dilutionally linear, whereas the ADI assays were not dilutionally linear. The slope [90% confidence interval] for each dilutionally linear regression fit are: −0.980 [−1.102, −0.859], −0.788 [−0.827, −0.748], and −0.829 [−0.898, −0.760] for ADI, ADIO, and ADI+, respectively. The lower limit of quantification was found to be greater than a 1:256 dilution for the FANG ELISA and a 1:32 dilution for all ADI ELISA methods ().
3.4. Assay precision from the ADI, ADI°, ADI+, and FANG ELISAs
Coefficients of variation (100×standard deviation / mean; %CV) are shown from repeated measurements (24 replicates per sample) of three samples semi-quantitatively representing low, medium, and high rel[Ab] processed on the ADI, ADIO, ADI+, and FANG ELISAs (Table 3). Percent CVs were smaller on the FANG assay than for the ADI, ADIO, and ADI+ assays; this was true for high, medium, and low titer samples.
Table 3.
Sample titer | FANG ELISA CV % |
ADI ELISA CV % |
ADIO ELISA Optimized CV% |
Retrospective ADI+ ELISA data reanalyzed CV% |
---|---|---|---|---|
High titer | 15.8 | 23.3 | 21.8 | 23.9 |
Medium titer | 13.5 | 21.9 | 15.0 | 17.5 |
Low titer | 8.8 | 71.1 | 40.8 | 36.0 |
ADIO, optimized ADI ELISA, ADI+, reanalysis of data obtained from commercially available ADI ELISA, CV, coefficient of variation; ELISA, enzyme-linked immunosorbent assay; FANG, Filovirus Animal Nonclinical Group; rel[Ab], relative antibody concentration. N=24 replicates for each sample.
3.5. Regional assay background assessment for the ADI and FANG ELISAs
The average rel[Ab] and standard deviation were calculated for all presumably EBOV-naïve Mali samples (N=100) processed on either the ADI or FANG ELISA (Table 4). A seroconversion cutoff, defined as antibody titers against EBOV larger than 3 standard deviations above the mean (log10 transformed), was determined to be 607 and 840 EU/mL (back-transformed), respectively for the FANG and ADI ELISAs. Using these cutoffs, the percentage of participants estimated to seroconvert at 1 month were higher with the FANG assay than with the ADI assay in both vaccine groups (Table 5).
Table 4.
ELISA type | Avg rel[Ab] (EU/ml) |
SD of rel[Ab] (EU/ml) |
Seroconversion cutoffa (EU/ ml) |
---|---|---|---|
FANG | 104.1 | 156.9 | 607 |
ADI | 199.4 | 224.3 | 840 |
Seroconversion cutoff is defined as antibody titers against EBOV of 3 standard deviations above the mean (log10 transformed).
ADI, Alpha Diagnostic International; EBOV, Ebola virus; ELISA, enzyme-linked immunosorbent assay; FANG, Filovirus Animal Nonclinical Group; rel[Ab], relative antibody concentration; SD, standard deviation.
Table 5.
ELISA type | rChAd3 EBO Z responders (%) N=487 |
rVSVΔG-ZEBOV responders (%) N=484 |
Placebo N=491 |
---|---|---|---|
FANG | 250 (51.3%) | 358 (74.0%) | 19 (3.9%) |
ADI | 150 (30.8) | 167 (34.5%) | 29 (5.9%) |
ELISA, enzyme-linked immunosorbent assay, FANG, Filovirus Animal Nonclinical Group. Seroconversion cutoffs for the FANG and ADI ELISAs were 607 and 840 EU/ml, respectively.
4. Discussion
4.1. Comparing the FANG and ADI ELISAs
Our extensive processing of samples from the PREVAIL I trial on the FANG and ADI ELISAs has shown both assays to be acceptable ELISA platforms. Both assays showed large and significant changes in rel[Ab] in participants that were vaccinated with active vaccine. These assays can be used in a variety of laboratory spaces and locations, including field locations. While the correlations shown in section 3.1 highlight that both assays trend similarly when measuring patient serum samples for IgG against EBOV GP, the FANG assay had some notable advantages.
First, as made evident by the PREVAIL I vaccine data set, the rel[Ab] range of the FANG ELISA exceeds the range of the ADI assay, allowing an investigator to better compare vaccine responses. Though not tested in this study, rel[Ab] range differences could be due to a number of factors, including antigen concentrations on the plate, secondary antibody concentrations, or differences introduced by antigen manufacturing or coating processes. However, as the ADI ELISA is proprietary, differences between antigen manufacturing and plate coating processes cannot be properly compared. Additionally, the different instruments used to obtain optical density readings for the FANG and ADI ELISAs in this work could have an effect on the ranges reported here. Yet, as both instruments were routinely maintained, any potential differences in values would most likely be negligible. Interestingly, two distinct groups of PREVAIL I vaccine responders were observed using the ADI ELISA as opposed to the single bell-curve distribution observed for the FANG ELISA (Fig. 2). This bimodal distribution seems to suggest that the ADI ELISA was not able to detect the full response to the vaccine in all participants. As similar proportions of individuals receiving each of the two vaccines appear to seroconvert using the ADI ELISA (Table 5), it is unlikely that the vaccine type contributed to this effect. However, further study will be needed to fully investigate why this distribution is observed using the ADI ELISA.
Second, the FANG ELISA exceeds the ability of the ADI assay to accurately and precisely determine the rel[Ab] of diluted samples as expressed by dilutional linearity and assay precision. Due to sample availability, a serum pool from EBOV survivors as opposed to vaccinated individuals was used to assess the precision and relative accuracy of these two assays. Another important distinction between assays is the level of background rel[Ab] associated with a presumably naïve population (100 naïve persons from Mali). In this group, the standard deviation and average rel[Ab] were smaller on the FANG ELISA than the ADI ELISA. Thus, a higher seroconversion cutoff is associated with the ADI assay, and discerning whether an individual seroconverts is hampered as compared to the FANG ELISA, as evidenced by the lower percentage of individuals who appeared to seroconvert in the PREVAIL I clinical vaccine trial (Table 5).
Most importantly, however, based on the PREVAIL I vaccine responses and the z-scores comparing active vaccine groups to placebo groups (Table 2), we estimate that the use of the ADI ELISA would require about three times as many participants to power a future vaccine study than if the FANG assay was used. Due to the reduction in needed study participants associated with the FANG assay, vaccine efficacy will be determined more quickly and will likely require less labor and cost. For these reasons, though both assays are shown to be versatile for use in multiple locations, we would suggest using the FANG assay to quantify rel[Ab] of IgG against the EBOV GP. Though this work does not necessarily translate to immune response quantification in other species, we suspect similar advantages will be seen using the FANG ELISA in animal models of EBOV infections. However, work comparing these ELISAs in non-human primate and other animal research has yet to be completed.
4.2. ADIO and ADI+ ELISA comparison
As the FANG assay is not yet widely available, optimization methods were developed for the commercially available ADI assay for use in the field. Specifically, a 4-PL calibrator curve fit (Findlay and Dillard, 2007) and stringent acceptance criteria were applied to both the ADIO and ADI+ optimization methods. A positive serum control was also added to the ADIO ELISA. As mentioned previously, assay precision and relative accuracy were the only aspects of these optimization methods assessed due to reagent and sample availability restrictions. Therefore, conclusions regarding rel[Ab] range and study participant requirements using the optimization methods could not be properly assessed through this work. Though no improvements in relative accuracy were observed between the ADI assay and the ADI+ and ADIO methods, precision increased for both optimization methods at the low and medium titer samples. Given these data and that the FANG assay is not yet commercially available, the ADI assay can be a good alternative when necessary, especially when the optimization and reanalysis steps (i.e., ADIO and ADI+) described above are applied. The comparison presented demonstrates the importance of certain aspects of ELISA platforms in generating a more precise result, including a 4-PL curve fit, stringent acceptance criteria and the use of well qualified matrix controls.
Acknowledgments
Funding
The authors acknowledge funding by the National Institutes of Allergy and Infectious Diseases; ClinicalTrials.gov number, NCT02344407. The content of this publication does not necessarily reflect the views or policies of the US Department of Health and Human Services (DHHS) or of the institutions and companies affiliated with the authors. This work was funded in part through Battelle Memorial Institute’s prime contract with the US National Institute of Allergy and Infectious Diseases (NIAID) under Contract No. HHSN272200700016I. J.L., K.T., J.P., E.J.S. and J.M. performed this work as employees of Battelle Memorial Institute. A subcontractor to Battelle Memorial Institute who performed this work is: G.G.O, an employee of MRI Global.
The authors thank the trial participants and their families and the PREVAIL I trial study team, including the Integrated Research Facility’s Response Team for helping to generate the ELISA data. The authors thank the FANG ELISA colleagues (C. Dorsey, C. Bounds, and C. Badorrek) within the US Department of Defense (DoD) Joint Program Executive Office for Chemical-Biological Defense (JPEO-CBD), Medical Countermeasure Systems' Joint Vaccine Acquisition Program (MCSJVAP) for reagent access and their critical assessment and comments on this manuscript. The authors would also like to thank Laura Bollinger and Jiro Wada, Battelle employees, for critically editing the manuscript and figure preparation, respectively. Finally, the authors would like to acknowledge all Mali, SEREFO/UCRC staff, laboratory and clinical, who contributed to patient recruitment, sample processing and data collection.
Footnotes
Competing Interests
Battelle Memorial Institute was involved in the development and attempted validation of the FANG ELISA.
Ethics
The protocol used in this clinical vaccine trial was approved by the National Research Ethics Board of Liberia, National Cancer Institute institutional review board, the Liberian Medicines and Health Products Regulatory Authority, and the Food and Drug Administration. Written informed consent was obtained from all the participants.
Author Contributions
J.L., K.T., E.J.S., G.G.O., and L.E.H. designed the study. J.L. and K.T. performed the experiments. J.L., J.M. and K.T. analyzed the data. M.F. and F.K.B. coordinated sample collection and laboratory space in Liberia. D.F., G.G., and C.R. provided statistical analysis. J.L. and K.T. wrote the manuscript. All authors reviewed and edited the manuscript.
References
- Arias A, Watson SJ, Asogun D, Tobin EA, Lu J, Phan MVT, Jah U, Wadoum REG, Meredith L, Thorne L, Caddy S, Tarawalie A, Langat P, Dudas G, Faria NR, Dellicour S, Kamara A, Kargbo B, Kamara BO, Gevao S, Cooper D, Newport M, Horby P, Dunning J, Sahr F, Brooks T, Simpson AJH, Groppelli E, Liu G, Mulakken N, Rhodes K, Akpablie J, Yoti Z, Lamunu M, Vitto E, Otim P, Owilli C, Boateng I, Okoror L, Omomoh E, Oyakhilome J, Omiunu R, Yemisis I, Adomeh D, Ehikhiametalor S, Akhilomen P, Aire C, Kurth A, Cook N, Baumann J, Gabriel M, Wolfel R, Di Caro A, Carroll MW, Gunther S, Redd J, Naidoo D, Pybus OG, Rambaut A, Kellam P, Goodfellow I, Cotten M. Rapid outbreak sequencing of Ebola virus in Sierra Leone identifies transmission chains linked to sporadic cases. Virus Evol. 2016;2 doi: 10.1093/ve/vew016. vew016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Blackley DJ, Wiley MR, Ladner JT, Fallah M, Lo T, Gilbert ML, Gregory C, D’Ambrozio J, Coulter S, Mate S, Balogun Z, Kugelman J, Nwachukwu W, Prieto K, Yeiah A, Amegashie F, Kearney B, Wisniewski M, Saindon J, Schroth G, Fakoli L, Diclaro JW, 2nd, Kuhn JH, Hensley LE, Jahrling PB, Stroher U, Nichol ST, Massaquoi M, Kateh F, Clement P, Gasasira A, Bolay F, Monroe SS, Rambaut A, Sanchez-Lockhart M, Scott Laney A, Nyenswah T, Christie A, Palacios G. Reduced evolutionary rate in reemerged Ebola virus transmission chains. Sci. Adv. 2016;2:e1600378. doi: 10.1126/sciadv.1600378. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Crozier I. Ebola virus RNA in the semen of male survivors of Ebola virus disease: the uncertain gravitas of a privileged persistence. J. Infect. Dis. 2016;214:1467–1469. doi: 10.1093/infdis/jiw079. [DOI] [PubMed] [Google Scholar]
- Deen GF, Broutet N, Xu W, Knust B, Sesay FR, McDonald SLR, Ervin E, Marrinan JE, Gaillard P, Habib N, Liu H, Liu W, Thorson AE, Yamba F, Massaquoi TA, James F, Ariyarajah A, Ross C, Bernstein K, Coursier A, Klena J, Carino M, Wurie AH, Zhang Y, Dumbuya MS, Abad N, Idriss B, Wi T, Bennett SD, Davies T, Ebrahim FK, Meites E, Naidoo D, Smith SJ, Ongpin P, Malik T, Banerjee A, Erickson BR, Liu Y, Liu Y, Xu K, Brault A, Durski KN, Winter J, Sealy T, Nichol ST, Lamunu M, Bangura J, Landoulsi S, Jambai A, Morgan O, Wu G, Liang M, Su Q, Lan Y, Hao Y, Formenty P, Stroher U, Sahr F. Ebola RNA persistence in semen of Ebola virus disease survivors - final report. N. Engl. J. Med. 2017;377:1428–1437. doi: 10.1056/NEJMoa1511410. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Findlay JW, Dillard RF. Appropriate calibration curve fitting in ligand binding assays. AAPS J. 2007;9:E260–E267. doi: 10.1208/aapsj0902029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fischer RJ, Judson S, Miazgowicz K, Bushmaker T, Munster VJ. Ebola virus persistence in semen ex vivo. Emerg. Infect. Dis. 2016;22:289–291. doi: 10.3201/eid2202.151278. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fischer WA, 2nd, Wohl DA. Confronting Ebola as a sexually transmitted infection. Clin. Infect. Dis. 2016;62:1272–1276. doi: 10.1093/cid/ciw123. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kennedy SB, Bolay F, Kieh M, Grandits G, Badio M, Ballou R, Eckes R, Feinberg M, Follmann D, Grund B, Gupta S, Hensley L, Higgs E, Janosko K, Johnson M, Kateh F, Logue J, Marchand J, Monath T, Nason M, Nyenswah T, Roman F, Stavale E, Wolfson J, Neaton JD, Lane HC Prevail l Study Group. Phase 2 placebo-controlled trial of two vaccines to prevent Ebola in Liberia. N. Engl. J. Med. 2017;377:1438–1447. doi: 10.1056/NEJMoa1614067. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kennedy SB, Neaton JD, Lane HC, Kieh MW, Massaquoi MB, Touchette NA, Nason MC, Follmann DA, Boley FK, Johnson MP, Larson G, Kateh FN, Nyenswah TG. Implementation of an Ebola virus disease vaccine clinical trial during the Ebola epidemic in Liberia: design, procedures, and challenges. Clin. Trials. 2016;13:49–56. doi: 10.1177/1740774515621037. [DOI] [PubMed] [Google Scholar]
- MacDermott NE, Bausch DG. Virus persistence and recrudescence after Ebola virus disease: what are the risks to healthcare workers? J. Hosp. Infect. 2016;94:113–115. doi: 10.1016/j.jhin.2016.07.004. [DOI] [PubMed] [Google Scholar]
- Mate SE, Kugelman JR, Nyenswah TG, Ladner JT, Wiley MR, Cordier-Lassalle T, Christie A, Schroth GP, Gross SM, Davies-Wayne GJ, Shinde SA, Murugan R, Sieh SB, Badio M, Fakoli L, Taweh F, de Wit E, van Doremalen N, Munster VJ, Pettitt J, Prieto K, Humrighouse BW, Stroher U, DiClaro JW, Hensley LE, Schoepp RJ, Safronetz D, Fair J, Kuhn JH, Blackley DJ, Laney AS, Williams DE, Lo T, Gasasira A, Nichol ST, Formenty P, Kateh FN, De Cock KM, Bolay F, Sanchez-Lockhart M, Palacios G. Molecular evidence of sexual transmission of Ebola virus. N. Engl. J. Med. 2015;373:2448–2454. doi: 10.1056/NEJMoa1509773. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pushko P, Bray M, Ludwig GV, Parker M, Schmaljohn A, Sanchez A, Jahrling PB, Smith JF. Recombinant RNA replicons derived from attenuated Venezuelan equine encephalitis virus protect guinea pigs and mice from Ebola hemorrhagic fever virus. Vaccine. 2000;19:142–153. doi: 10.1016/s0264-410x(00)00113-4. [DOI] [PubMed] [Google Scholar]
- Sabourin C. [Accessed 12 Apr 2017];Assays to measure Ebola humoral and cellular responses in humans. Immunology of Protection from Ebola virus Infection Workshop, Dec 12, 2014, Fishers Lane Conference Center, Rockville, MD, Immunology of protection from Ebola virus infection Workshop, Dec 12, 2014, Fishers Lane Conference Center, Rockville, MD, National Institute of Allergy and Infectious Diseases. Aired Dec 12, 2014. Run time: 4:17:45-. Talk initiated at 4:17.55. 2014 Available at: http://videocast.nih.gov/launch.asp?18779.
- Sow MS, Etard JF, Baize S, Magassouba N, Faye O, Msellati P, Toure A, 2nd, Savane I, Barry M, Delaporte E Postebogui Study Group. New evidence of long-lasting persistence of Ebola virus genetic material in semen of survivors. J. Infect. Dis. 2016;214:1475–1476. doi: 10.1093/infdis/jiw078. [DOI] [PubMed] [Google Scholar]
- Uyeki TM, Erickson BR, Brown S, McElroy AK, Cannon D, Gibbons A, Sealy T, Kainulainen MH, Schuh AJ, Kraft CS, Mehta AK, Lyon GM, 3rd, Varkey JB, Ribner BS, Ellison RT, 3rd, Carmody E, Nau GJ, Spiropoulou C, Nichol ST, Stroher U. Ebola virus persistence in semen of male survivors. Clin. Infect. Dis. 2016;62:1552–1555. doi: 10.1093/cid/ciw202. [DOI] [PubMed] [Google Scholar]
- Wilson JA, Hart MK. Protection from Ebola virus mediated by cytotoxic T lymphocytes specific for the viral nucleoprotein. J. Virol. 2001;75:2660–2664. doi: 10.1128/JVI.75.6.2660-2664.2001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wong G, Richardson JS, Pillet S, Patel A, Qiu X, Alimonti J, Hogan J, Zhang Y, Takada A, Feldmann H, Kobinger GP. Immune parameters correlate with protection against ebola virus infection in rodents and nonhuman primates. Sci. Transl. Med. 2012;4:158ra146. doi: 10.1126/scitranslmed.3004582. [DOI] [PMC free article] [PubMed] [Google Scholar]
- World Health Organization. [Accessed 24 November 2017];Ebola Situation Report – 30 March 2016. 2016a http://apps.who.int/ebola/current-situation/ebola-situationreport-30-march-2016.
- World Health Organization. Interim Guidance: Clinical Care for Survivors of Ebola Virus Disease. World Health Organization; Geneva, CH: 2016b. [Accessed 24 November 2017]. http://apps.who.int/iris/bitstream/10665/204235/1/WHO_EVD_OHE_PED_16.1_eng.pdf?ua=1. [Google Scholar]
- World Health Organization. [Accessed 5 July 2016];Ebola Virus Disease Situation Report 10 June 2016. 2016c Available at: http://apps.who.int/iris/bitstream/10665/208883/1/ebolasitrep_10Jun2016_eng.pdf?ua=1.
- Xu L, Sanchez A, Yang Z, Zaki SR, Nabel EG, Nichol ST, Nabel GJ. Immunization for Ebola virus infection. Nat. Med. 1998;4:37–42. doi: 10.1038/nm0198-037. [DOI] [PubMed] [Google Scholar]
- Yang W, Zhang W, Kargbo D, Yang R, Chen Y, Chen Z, Kamara A, Kargbo B, Kandula S, Karspeck A, Liu C, Shaman J. Transmission network of the 2014–2015 Ebola epidemic in Sierra Leone. J. R. Soc. Interface. 2015;12 doi: 10.1098/rsif.2015.0536. . 20150536. [DOI] [PMC free article] [PubMed] [Google Scholar]