3D confocal laser-scanning microscopy for large-area imaging of the corneal subbasal nerve plexus

Stephan Allgeier 1,, Andreas Bartschat 1, Sebastian Bohn 2, Sabine Peschel 3, Klaus-Martin Reichert 1, Karsten Sperlich 2, Marcus Walckling 2, Veit Hagenmeyer 1, Ralf Mikut 1, Oliver Stachs 2, Bernd Köhler 1
PMCID: PMC5945773  PMID: 29749384

Abstract

The capability of corneal confocal microscopy (CCM) to acquire high-resolution in vivo images of the densely innervated human cornea has gained considerable interest in using this non-invasive technique as an objective diagnostic tool for staging peripheral neuropathies. Morphological alterations of the corneal subbasal nerve plexus (SNP) assessed by CCM have been shown to correlate well with the progression of neuropathic diseases and even predict future-incident neuropathy. Since the field of view of single CCM images is insufficient for reliable characterisation of nerve morphology, several image mosaicking techniques have been developed to facilitate the assessment of the SNP in large-area visualisations. Due to the limited depth of field of confocal microscopy, these approaches are highly sensitive to small deviations of the focus plane from the SNP layer. Our contribution proposes a new automated solution, combining guided eye movements for rapid expansion of the acquired SNP area and axial focus plane oscillations to guarantee complete imaging of the SNP. We present results of a feasibility study using the proposed setup to evaluate different oscillation settings. By comparing different image selection approaches, we show that automatic tissue classification algorithms are essential to create high-quality mosaic images from the acquired 3D datasets.

Introduction

Diagnosis of peripheral neuropathies currently relies on assessments of nerve function that commonly detect pathologic alterations only after the manifestation of clinical symptoms. In this context, novel objective diagnostic methods with increased sensitivity are urgently needed to facilitate early diagnosis and to evaluate novel therapeutic approaches. Confocal laser scanning microscopy of the cornea is a promising novel approach which allows non-invasive in vivo imaging of the corneal subbasal nerve plexus (SNP) and therefore enables direct assessment of its morphology13. In recent years, corneal confocal microscopy (CCM) has been used to reveal significant morphological changes of the SNP in diverse diseases such as diabetic peripheral neuropathy (DPN), chronic migraine4, multiple sclerosis5,6 and familial amyloid neuropathy7. DPN in particular has been the main focus of various studies8. Since characteristic morphological alterations of the SNP already occur at an early stage of DPN9,10, CCM has the potential to provide reliable biomarkers for the early assessment of DPN. A major limitation of this imaging technology arises from the inhomogeneous distribution of nerve fibres across the area of the cornea11. The field of view of a single CCM image (typically 0.16 mm²) is therefore insufficient for the reliable morphometric characterisation of the SNP. Recommendations in the literature to analyse samples of multiple non-overlapping CCM images to effectively expand the examined corneal area range from three12 to eight13 images, which have to be manually selected from a larger set of acquired images according to predefined quality criteria.

An alternative approach to increase the examined area is to compose a mosaic image from several overlapping CCM images. Various algorithms for the creation of mosaic images have been proposed1420: Some rely on the operator to acquire appropriate image sequences manually, while others incorporate more automated methods for capturing an extended area in the recorded image sequence. The so-called EyeGuidance system that we described in preceding articles21,22 implements a fully automated image acquisition process. A computer-controlled fixation target, presented to the contralateral eye and moving in an outward-spiralling pattern, guides the eye movements of the patient during the entire image acquisition process. Using this approach, an average expansion rate of the acquired mosaic image area of 0.16 mm²/s can be achieved. Based on the examination of the progression of morphological SNP parameters in continuously expanding mosaic images, we recommended a minimum size of the mosaic images of 1.5 mm² for a reliable morphological characterisation of the SNP23, which requires a mean examination time of less than 10 seconds21 with the EyeGuidance system.

Due to the limited depth of field, which is characteristic to confocal microscopy, it cannot be guaranteed that all the recorded CCM images are well focused on the SNP layer. Even slight displacements perpendicular to the focus plane during the image acquisition process or small deviations from the surface-parallel arrangement of the SNP lead to images that partially or entirely contain extraneous tissue (epithelium or stroma)21. Consequently, the mosaic image generated from such an image sequence does not show SNP in the complete field of view (see Fig. 1), which ultimately leads to incorrect morphological SNP parameter measurements. This issue is also common to almost all of the previously proposed mosaicking methods1418,21 as they are inherently 2D processes without the ability of (controlled) focus adjustments. Only Lagali et al.19,20 describe a mosaicking process with manual focus variation during the image acquisition process; images that do not show SNP are sorted out manually afterwards.

Figure 1.

Figure 1

SNP mosaic image created using the EyeGuidance system without focus adjustments, as described by Allgeier et al.21. Several image areas show epithelium (black arrows) or stroma tissue (white arrows). The bottom left inset represents the size of a single CCM image (0.4 × 0.4 mm²) from the recorded dataset.

In order to overcome the issue described above and to realise a reliable, highly automated SNP mosaicking process, we propose a novel approach for automatic online SNP focus tracking in the present article. To this end, the confocal microscope is equipped with a computer-controlled piezo drive unit that performs an oscillating movement of the focus plane during the image acquisition process. We present results from a feasibility test of the proposed setup in conjunction with the EyeGuidance system22. The aim of the present study is to examine the potential benefit in mosaic image quality and to evaluate the main process parameter of the focus oscillation, i.e. the oscillation velocity, and to compare three different approaches to create mosaic images from the recorded image sequences.

Methods

CCM imaging was performed using a Heidelberg Retina Tomograph II (HRT II, Heidelberg Engineering GmbH, Heidelberg, Germany) in conjunction with a modified Rostock Cornea Module (RCM) and the EyeGuidance system21,22 to guide the eye movements of the examined person. The study received appropriate ethics committee approval from the institutional review board (Rostock University Medical Centre Ethics committee) in accordance with the Declaration of Helsinki. All examined volunteers gave their informed consent and – in accordance with the ethics committee approval – did not have any known neuropathic or ocular diseases.

Modified rostock cornea module

Compared to the original RCM1 (Heidelberg Engineering GmbH, Heidelberg, Germany), the modified RCM is additionally equipped with a piezo driven actuator (MIPOS 600 SG DIG, Piezosystem Jena GmbH, Jena, Germany) to move the internal objective lens. Due to the changed local arrangement of the lenses of the RCM objective, the square field of view decreases from 0.160 mm² to 0.123 mm².

The piezo drive has a setting range of 500 µm in closed loop control and can be controlled by a computer via an RS232 interface. An in-house developed adapter device integrates the piezo actuator into the RCM (see Fig. 2); due to the compact design, the piezo drive unit does not interfere with the CCM examinations.

Figure 2.

Figure 2

The modified Rostock Cornea Module (RCM). (a) Photograph of the modified RCM attached to the HRT II and next to the eyepiece of the EyeGuidance system. (b) Schematic drawing of the modified RCM; the piezo drive translates the objective (highlighted by red border) to effect focus changes; the contact TomoCap remains fixed in the process.

EyeGuidance

To control the movements of the examined eye during the CCM image acquisition process, the EyeGuidance system presents a computer controlled fixation target to the contralateral eye. A smartphone (Motorola Moto G, Motorola Mobility Germany GmbH, Idstein, Germany) with an Android operating system is used to display the fixation target. The 4.5″ display has a pixel resolution of 1280 × 720. A special optical setup composed of four convex lenses and a tilted mirror realises a compact design suitable for CCM. Three linear axes allow the precise alignment of the eyepiece of the EyeGuidance system. A more detailed technical description of the EyeGuidance hardware setup has been published22.

The EyeGuidance smartphone is connected to the controlling PC via USB. The parameters for the fixation target (e.g. path layout, size, colour, speed, acceleration) are defined in an XML file. During the CCM image acquisition process, the EyeGuidance software continuously calculates the position of the fixation target and sends it to the EyeGuidance App running on the smartphone. The update frequency depends on the PC performance; for the setup used in the present study, we measured an update frequency of the fixation target position of approximately 923 Hz, which reliably guarantees a smooth motion of the fixation target even at high velocities.

CCM setup with modified RCM and EyeGuidance

The schematic illustration of the CCM setup is shown in Fig. 3. Three dedicated PCs, connected via a TCP/IP network, are used to control the HRT II, the EyeGuidance system, and the modified RCM. The complete CCM image acquisition process is controlled via the HRT computer (HRT-PC, see Fig. 3) as described below.

Figure 3.

Figure 3

Schematic illustration of the CCM setup with focus plane control and guided eye movements.

CCM image acquisition process

As a preparation step the cornea of the examined eye was locally anaesthetised by instilling Proparakain 0.5% eye drops (Ursapharm, Saarbrücken, Germany) and the Vidisic gel (Bausch & Lomb/Dr. Mann Pharma, Berlin, Germany) is used as a coupling medium between objective and TomoCap as well as between TomoCap and cornea. After positioning the head of the patient in the forehead support, the operator aligns the EyeGuidance system in front of the not examined eye. The operator adjusts the HRT in the conventional way to achieve frontal SNP images of the central cornea. A joystick connected to the piezo drive offers a fine tuning of the focus.

Following these manual initialisation steps, the operator starts the fully automated image acquisition process with the start button, which triggers a signal to EG-PC and RCM-PC. The movement of the EyeGuidance fixation target and the piezo oscillation start simultaneously. The fixation target moves with the predefined speed along an outward spiralling pattern with constantly spaced spiral windings. The piezo drive constantly oscillates with an amplitude aosc = 20 µm and a constant speed vosc along the slopes, describing a periodic triangular function (see Fig. 4). The period length is then given as tperiod = 4aosc/vosc. The operator terminates the image acquisition process using the stop button after texam = 40 seconds, or earlier, if visual assessment of the live image quality is deemed unsatisfying. In that case, the examination is repeated.

Figure 4.

Figure 4

Oscillation of the focus plane during image acquisition. The focus starts in the SNP layer, and then describes a periodic triangular function.

In order to assess the influence of the focus oscillation parameters on the acquired image data and the resulting mosaic images, we performed repeated examinations with different settings for vosc and the velocity of the fixation target vtarg. Table 1 lists the absolute values used for the examinations E1 (slow focus oscillation and fixation target movement) through E4 (fast focus oscillation and fixation target movement).

Table 1.

Process parameters for examinations E1 through E4.

E1 E2 E3 E4
Focus oscillation velocity vosc [µm/s] 60.6 121.2 181.8 242.4
Focus oscillation amplitude aosc [µm] 20 20 20 20
Focus oscillation period tperiod [s] 1.32 0.66 0.44 0.33
Fixation target velocity vtarg [pixel/s] 25.5 51 76.5 102

Mosaicking process

In a post-processing step, mosaic images are calculated from the acquired image sequences. We use in-house developed algorithms to align the single images and to correct motion artefacts2426. In order to create mosaic images from datasets with guided eye movements we developed a special image alignment process suitable for datasets recorded with a fixation target movement on an outward spiralling pattern21.

As a first step, all images of a sequence were used for image alignment and correction of motion artefacts. Then, three different approaches for the montage of the mosaic image were implemented and applied to each dataset. The first approach (M1) considers all images of a dataset for the mosaic montage. A second mosaic image (M2) was created using only images from the oscillating focus sequence that had been recorded at a reference focus level. By definition of the process protocol, the imaging process starts in the SNP layer, thus this level was used as reference. Only one image was included from each slope of the oscillation process. In effect, this approach selects a 2D plane from the 3D image data, emulating the EyeGuidance process without focus oscillation21 and therefore facilitating a direct comparison of both approaches in terms of mosaic image quality. The third image selection approach (M3) used an automated tissue classification algorithm that had been trained to classify CCM images as SNP or other corneal tissue27. Only images classified as SNP were incorporated into the M3 mosaic images.

Data Availability

The datasets generated during the current study are available from the corresponding author on reasonable request.

Results

Data Acquisition

The present study included nine volunteers that were recruited from the clinic staff. All volunteers underwent the four examinations described in the Methods section. The imaging duration was 40 seconds for all examinations. Because of a very limited time slot of one volunteer, we decided to reduce the imaging duration to 30 seconds for all examinations for that person. According to our examination protocol, an examination was immediately repeated if the quality of the acquired image dataset was deemed insufficient based on the visual assessment of the live image during image acquisition. This was not the case for E1, E2 or E3, but for two volunteers in E4. In one of these two cases, we used the second of the two acquired datasets. In the other case, we stopped examination E4 after three unsuccessful attempts. Of the remaining eight successfully acquired datasets from E4 examinations, we had to exclude an additional one since it proved to be unsuitable for automated mosaic image generation due to insufficient image quality that originated from inhomogeneity in the contact gel, which was not recognised during image acquisition. The following results are therefore based on nine datasets for examinations E1 through E3 and the seven datasets from examination E4 that are suitable for mosaic image generation.

Mosaic Images

Three mosaic images M1, M2 and M3 were created from each of the datasets. Including all images, process M1 of course generates the maximum mosaic image size for all examinations (see Table 2). The number of images used for M2 is defined by the number of oscillation periods, texam/tperiod, and therefore increases from 72.0 (5.7% of the recorded images) for E1 nearly proportionally to 240.1 (20.3% of the recorded images) for E4. Compared to M1, the average size of the M2 mosaic images is reduced about 16.6%, 11.0%, 8.1% and 6.4% for examinations E1, E2, E3 and E4, respectively. For M3, the percentage of images used is similar for all examinations (between 36.3% and 39.7%) and the ratio of the average image size between M3 and M1 lies in the range of 91% to 95%.

Table 2.

Mosaic image sizes and area expansion rates for examinations E1 through E4.

E1 E2 E3 E4
M1 Average mosaic image size [mm²] 1.34 2.44 3.39 4.62
Average expansion rate [mm²/s] 0.033 0.062 0.089 0.117
Average number of images used 1245.8 1187.3 1119.9 1178.7
M2 Average mosaic image size [mm²] 1.12 2.18 3.12 4.31
Average expansion rate [mm²/s] 0.028 0.055 0.082 0.109
Average number of images used 72.0 127.1 175.2 240.1
M3 Average mosaic image size [mm²] 1.25 2.30 3.09 4.38
Average expansion rate [mm²/s] 0.031 0.059 0.081 0.111
Average number of images used 482.4 469.2 406.3 468.7

Figure 5 depicts representative examples for the mosaic images created from a single dataset with the processes M1, M2 and M3 (high-resolution versions of these images as well three additional examples from other datasets are available as Supplementary Figures S1 through S4). The mosaic images were evaluated qualitatively with regard to the visibility of the nerve structures of the SNP. Due to the oscillating movement of the focus plane, mosaic images created from all images in a dataset (M1, see Fig. 5a) always show a superimposition of epithelial tissue, nerve structures in the SNP layer, structures from Bowman’s layer (so-called K-structures) and the keratocyte nuclei of the stroma. Most nerve structures are visible but provide only low contrast because of the superimposition of several tissues at most image locations. Where SNP nerves are visible in mosaic images created from the subset of images acquired on the reference focus level defined by the start of the imaging process (M2, see Fig. 5b and red circles in Fig. 6), they commonly possess very good contrast. However, the proportion of the mosaic image area that is optimally focused on the SNP layer varies widely. Some M2 mosaic images are focused well over almost the entire image field; others contain large regions of non-basal epithelial or stroma tissue that encompass more image area than SNP regions. The mosaic images created from the subset of images classified as SNP (M3, see Fig. 5c and solid circles in Fig. 6) show at least the nerves structures that are discernible in the M1 mosaic image. In contrast to the M1 montaging process, the majority of images containing non-SNP tissue is excluded prior to montaging the M3 mosaic images; according to the numbers of images used for the respective mosaicking processes (see Table 2), the proportion of excluded images ranges between 60.3% and 63.7% of all images. Consequently, the nerve structures manifest in M3 with good contrast against the surrounding tissue. The exclusion of images leads to the above-mentioned reduction of the field of view of the M2 and M3 mosaics.

Figure 5.

Figure 5

Mosaic images generated from a dataset (subject 8, examination E2); (a) mosaic image M1 using all 1245 images; (b) mosaic image M2 using the subset of 130 images acquired on the initial focus level; (c) mosaic image M3 using the subset of 610 images classified as SNP tissue. The arrows denote nerves that are visible in mosaic image M3 but not in M2.

Figure 6.

Figure 6

Plot of the focus levels of the CCM images (circles) of the dataset shown in Fig. 5. Red circles mark images acquired on the initial focus level. Solid circles mark CCM images classified as containing SNP tissue. These tend to lie below the initial focus level, and consequentially the M2 mosaic image in Fig. 5b tends towards showing epithelium tissue.

Discussion

Although the majority of CCM studies of the SNP published in literature are based on the examination of sets of single CCM image frames, the number of publications on wide-field CCM image creation by different research groups has increased in recent years. Both the CCM-frame approach and the wide-field approach have their unique properties12,19. The most important advantages of the CCM-frame approach are the relatively straightforward and well-established image acquisition process and the good support for morphometric analysis of the standard-sized and -shaped CCM frames12 by established software. Its major disadvantage is the necessity for manual selection of non-overlapping CCM frames from a larger pool of up to several hundreds of CCM images12,28 that are recorded during the image acquisition phase; see Fig. 7 for a sample of eight SNP images manually selected from the same dataset used for Fig. 5. This step is time-consuming, requires expert knowledge and is prone to subjective bias effects12,19. By contrast, the wide-field approach is commonly more complex in terms of the software12,1421 and hardware17,21 setup. Semi-automatic morphometric analysis of large mosaic images has been described as more laborious compared to standard CCM frames12, although further developments towards fully-automatic software tools can provide a solution19,20. A major advantage of the wide-field approach, on the other hand, is its capability to provide image data within a larger field of view, potentially allowing the clinician to repeatedly assess identical tissue regions19. Regarding the diagnostic value of both general approaches, Kheirkhah et al. find no significant differences12, whereas Lagali et al. point out potential benefits in morphometric analysis of wide-field mosaic images19.

Figure 7.

Figure 7

Exemplary sample of eight CCM images of the SNP manually selected from the dataset used for Fig. 5 (subject 8, examination E2, 1246 images total). Repeated presence of the same SNP region in multiple images (as for example in subfigures (b,c,e) see black arrows) can potentially introduce bias effects to morphometric parameter measurements and should therefore be avoided; identification of such repeatedly represented SNP regions in a manual image selection procedure is not trivial.

Our aim is to develop a reliable, easy-to-use and fast tool for imaging and visualising an expanded area of the SNP. The 2D-only EyeGuidance process (2D-EG) without focus adjustment, originally published in 201421, achieved an average area expansion rate of approximately 0.16 mm²/s. Those examinations also established that a high velocity of the fixation target movement does not negatively influence the resulting mosaic image quality but has positive effects regarding the patient compliance, by decreasing the required examination time and because fast target movement is perceived as more comfortable than slow movement. Although already yielding promising results, the primary limitation of the 2D-EG process is its inability to compensate deviations of the focus plane from the SNP layer, which are encountered in practice for multiple reasons and are difficult or even impossible to avoid21.

The focus of the present study lies on the evaluation of an automated oscillation of the focus plane during image acquisition, effectively extending the process into the third dimension (3D-EG). The main process parameter that changes between the examinations E1 through E4 is the focus oscillation velocity vosc. The fixation target velocity was adjusted such that the fixation target translation throughout an oscillation period always remained constant, i.e. the ratio between vosc and fixation target velocity was identical for all examinations (see Table 1). The relation between the two velocities was chosen such that after one focus oscillation period (corresponding to two passes through the SNP layer) the resulting lateral displacement between the CCM images was approximately 200 µm.

The first objective was to analyse the examination duration, as this is essential for patient comfort and compliance. As expected, the measured average area expansion rates (see Table 2) are directly proportional to the speed of the fixation target respectively the oscillation velocity vosc (see Table 1). An average area expansion rate of 0.11 to 0.12 mm²/s (depending on the mosaicking approach used) was measured for the fastest oscillation velocity, which is a decrease of only 26–31% compared to the 0.16 mm²/s of the 2D-EG process21. This is an interesting and unexpected finding, considering that the field of view of the modified RCM is reduced about 23% and the new 3D-EG process essentially covers a volume of tissue extending 40 µm in depth, instead of just a single plane. Regarding the question, whether vosc can be increased further in order to accelerate the imaging process, it is important to look at the translation of the focus plane between two adjacent images. In examination E4, the focus plane moves about 8 µm during the recording of a single image. This distance is already in the order of the depth of field of the microscope29. Increasing the vosc beyond the value used for E4, although technically possible, is not practicable because it would no longer be guaranteed that at least a single image of the SNP layer is acquired along each slope of the focus oscillation process. Further acceleration of the imaging process can therefore only be achieved by decreasing the oscillation amplitude.

Regarding the reliability of the imaging process, we did not find any qualitative difference between examinations E1 through E3. All of those examinations were successfully completed at the first attempt. In contrast, examination E4 had to be repeated for two volunteers due to unsatisfactory quality of the dataset, and we finally only obtained seven mosaic images from the nine volunteers, as described in the Results section. For one of the two missing datasets, the exclusion could be attributed to inhomogeneity in the contact gel, which happens during preparation of the device. This exclusion is therefore unrelated to the process parameters of the oscillation process. The noticeable increase in the number of repetitions in E4 might be related to the process parameters, i.e. the higher oscillation and fixation target velocities. However, as the order of the examinations was not randomised and the other three examinations always preceded E4, it may also be an effect of diminishing participant compliance over time. Finally, considering the relatively low number of participants, no correlation can be established between vosc and the reliability of the imaging process based on the analysed data.

Based on the proportional relationship between oscillation velocity and average area expansion rate stated above, we conclude that faster oscillation velocities possess the advantage of reducing the required examination time and allow faster fixation target movement, which is perceived as more comfortable by the examined volunteers. Further investigation is required to find whether a very high vosc in the order of the velocity used for E4 reduces the reliability of the imaging process. However and in any case, even the process parameters of examination E3 allow an average expansion rate of covered SNP area of more than 0.08 mm²/s and facilitate an examination duration of less than 20 seconds for a target area of 1.5 mm².

The second objective of the present study was to assess the mosaic image quality. No systematic differences were found between examinations E1 through E4. The differences in mosaic image quality between the mosaicking approaches M1 through M3, however, are obvious (see Fig. 5 and Supplementary Figures S1 through S4). The comparative qualitative evaluation of the three presented mosaic montaging processes has to be guided by the question of how the characteristics of the resulting mosaic images affect the reliable quantitative morphometric characterisation of the SNP. Two potential detrimental factors play a major role in this regard:

  • incorrect nerve fibre segmentation, encompassing both unsegmented nerve structures (false negatives) and segmented structures that are not nerve fibres (false positives) and

  • incorrect tissue regions, i.e. mosaic image regions showing tissue from above or below the SNP layer.

The second point is of particular interest as, even assuming a perfect segmentation procedure, such incorrect tissue regions lead to an underestimation bias of the morphometric parameter measurements, which are commonly normalised to the assessed area (cf. corneal nerve fibre length, corneal nerve fibre density, corneal nerve branch density13,24,30,31). Incorrect tissue regions represent image area without adding any nerve fibres or branches.

These two points clearly reveal the deficits of the mosaicking approaches M1 and M2. The M1 mosaics contain practically no regions with no SNP tissue visible, which of course is the intent of the focus oscillation during the imaging process. However, the low contrast of the nerve structures in the mosaics is counterproductive for correct nerve fibre segmentation and therefore carries the risk for increased deviations of the measured SNP parameters compared to the M2 and M3 mosaics. In some cases, the contrast can even be reduced to a level where the nerve fibres almost disappear, although nerves have been imaged at the respective location, as evidenced by the corresponding M3 mosaic (see Fig. 5).

As the M2 mosaics are composed from images from a specific, constant focus depth, they manifest a very similar characteristic as the ones that result from the originally proposed 2D-EG process21. Their good contrast supports the correct segmentation of nerve structures. Contrary to the other two mosaicking approaches, however, they can potentially contain ill-focused image regions and consequentially carry the risk of reducing the accuracy of the measured SNP parameters.

The M3 mosaics combine the advantages of both of the other approaches while at the same time avoiding the disadvantages: They provide satisfactory contrast of the nerve fibres in order to support reliable nerve segmentation and they do not contain ill-focused tissue regions. Such regions can only be avoided by continually controlling the focus, which cannot be adjusted manually while at the same time minimising the duration of examination21. The advantageous effects of small focus variations on the wide-field visualisation of the SNP are in general accordance with the conclusions drawn by Lagali et al.19.

The primary purpose of the present article is to provide a detailed description of the highly automated 3D-EG process and to evaluate its technical parameters. The results demonstrate the feasibility of the process and the qualitative advantages with regard to SNP visualisation, as compared to 2D-only mosaicking approaches. Further evaluation of the diagnostic utility of the 3D-EG process was not within the scope of the presented feasibility study, therefore no morphometric parameters were calculated in the mosaic images. However, two previously published results from larger studies have to be taken into account: the reported absence of significant quantitative differences between SNP morphometry measurements in sets of single CCM frames and in wide-field images12, and the reported benefits of systematic focus variations on the resulting wide-field images of the SNP19. Given these previous results, the diagnostic utility of the presented 3D-EG process can be assumed to be at least equal to other published CCM techniques.

A further limitation of the presented system is the complex setup with three connected PCs (see Fig. 3), which is not suitable in clinical practice. This setup originates from the distributed development of the software modules. For a potential future system intended to be used for wide-field CCM in a clinical environment, it is necessary to integrate all software modules into a single, dedicated PC system.

In the current implementation of the described process, the mosaicking step can only start after the entire image dataset has been recorded. The quality of the imaging process has to be estimated based on the rapidly changing live image (see Supplementary Videos S5 through S8), which requires a lot of expert knowledge and is prone to misjudgement. With a potential real-time mosaic image visualisation, the operator could directly assess the quality of the growing mosaic image. This could allow earlier termination and repetition of the examination and reduce overall examination time for the patient. Future development work will therefore be dedicated towards realising an online mosaicking process with tissue classification27,32 and a real-time control of the focus oscillation parameters. The purpose of this control scheme is to reduce the oscillation amplitude and continuously adjust the zero-position of the oscillation in order to reduce the examination duration further while at the same time reliably keeping the SNP layer inside the oscillation range (see Supplementary Video S9 for a simulated animation of online mosaicking with tissue classification).

As the presented process captures image data extending into the third dimension, reconstructing a volume representation of the acquired tissue region could provide valuable additional information, not only for imaging the SNP but also potentially for other tissues. This will also be an interesting topic for future research work.

In conclusion: We proposed a novel concept for automated wide-field CCM imaging of the SNP and provided results of the first feasibility study. Using an externally mounted, piezo-driven focus control unit, we realised a fast 3D imaging process that facilitates the reliable acquisition of the SNP and adjacent tissue layers. A tissue classification algorithm filters the CCM datasets, keeping only images of the SNP layer. The direct comparison of mosaic images composed of these SNP-filtered 3D datasets against mosaic images montaged from a 2D plane clearly reveals the advantages of the proposed concept over 2D-only mosaicking approaches.

Electronic supplementary material

Supplementary Video S5 (10.1MB, mp4)
Supplementary Video S6 (10.1MB, mp4)
Supplementary Video S7 (7.6MB, mp4)
Supplementary Video S8 (2.5MB, mp4)
Supplementary Video S9 (10.3MB, mp4)

Acknowledgements

The work was supported by the DFG (German Research Foundation, grant numbers KO 5003/1-1, MI 1315/5-1 and STA 543/6-1).

Author Contributions

S.A., R.M., O.S. and B.K. designed the experiment. S.B., S.P., K.M.R., K.S., M.W., O.S. and B.K. conducted the experiment. S.B., K.S., O.S. and B.K. developed and operated the modified RCM module. S.A., K.M.R. and B.K. developed and applied the mosaicking algorithm. A.B. and R.M. developed and applied the tissue classification algorithm. S.A. and B.K. analysed and interpreted the results. S.A. and B.K. wrote the manuscript. S.A., A.B., S.B., S.P., K.M.R., K.S., M.W., V.H., R.M., O.S. and B.K. critically reviewed the manuscript.

Competing Interests

The authors declare no competing interests.

Footnotes

Electronic supplementary material

Supplementary information accompanies this paper at 10.1038/s41598-018-25915-6.

Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Guthoff, R. F., Baudouin, C. & Stave, J. Atlas of Confocal Laser Scanning In-vivo Microscopy in Ophthalmology. Principles and Applications in Diagnostic and Therapeutic Ophthalmology (Springer, 2006).
  • 2.Kim J, Markoulli M. Automatic analysis of corneal nerves imaged using in vivo confocal microscopy. Clin. Exp. Optom. 2018;101:147–161. doi: 10.1111/cxo.12640. [DOI] [PubMed] [Google Scholar]
  • 3.Cruzat A, Qazi Y, Hamrah P. In Vivo Confocal Microscopy of Corneal Nerves in Health and Disease. Ocul. Surf. 2017;15:15–47. doi: 10.1016/j.jtos.2016.09.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Kinard KI, et al. Chronic migraine is associated with reduced corneal nerve fiber density and symptoms of dry eye. Headache. 2015;55:543–549. doi: 10.1111/head.12547. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Bitirgen G, Akpinar Z, Malik RA, Ozkagnici A. Use of Corneal Confocal Microscopy to Detect Corneal Nerve Loss and Increased Dendritic Cells in Patients With Multiple Sclerosis. JAMA Ophthalmol. 2017;135:777–782. doi: 10.1001/jamaophthalmol.2017.1590. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Petropoulos IN, et al. Corneal Confocal Microscopy. An Imaging Endpoint for Axonal Degeneration in Multiple Sclerosis. Invest. Ophthalmol. Vis. Sci. 2017;58:3677–3681. doi: 10.1167/iovs.17-22050. [DOI] [PubMed] [Google Scholar]
  • 7.Rousseau A, et al. Potential Role of In Vivo Confocal Microscopy for Imaging Corneal Nerves in Transthyretin Familial Amyloid Polyneuropathy. JAMA Ophthalmol. 2016;134:983–989. doi: 10.1001/jamaophthalmol.2016.1889. [DOI] [PubMed] [Google Scholar]
  • 8.De Clerck EEB, et al. New ophthalmologic imaging techniques for detection and monitoring of neurodegenerative changes in diabetes: a systematic review. Lancet Diabetes Endocrinol. 2015;3:653–663. doi: 10.1016/S2213-8587(15)00136-9. [DOI] [PubMed] [Google Scholar]
  • 9.Tavakoli M, et al. Corneal confocal microscopy: a novel noninvasive test to diagnose and stratify the severity of human diabetic neuropathy. Diabetes Care. 2010;33:1792–1797. doi: 10.2337/dc10-0253. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Lovblom LE, et al. In vivo corneal confocal microscopy and prediction of future-incident neuropathy in type 1 diabetes. A preliminary longitudinal analysis. Can. J. Diabetes. 2015;39:390–397. doi: 10.1016/j.jcjd.2015.02.006. [DOI] [PubMed] [Google Scholar]
  • 11.Patel DV, McGhee CNJ. Mapping of the normal human corneal sub-basal nerve plexus by in vivo laser scanning confocal microscopy. Invest. Ophthalmol. Vis. Sci. 2005;46:4485–4488. doi: 10.1167/iovs.05-0794. [DOI] [PubMed] [Google Scholar]
  • 12.Kheirkhah A, et al. Comparison of standard versus wide-field composite images of the corneal subbasal layer by in vivo confocal microscopy. Invest. Ophthalmol. Vis. Sci. 2015;56:5801–5807. doi: 10.1167/iovs.15-17434. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Vagenas D, et al. Optimal image sample size for corneal nerve morphometry. Optom. Vis. Sci. 2012;89:812–817. doi: 10.1097/OPX.0b013e31824ee8c9. [DOI] [PubMed] [Google Scholar]
  • 14.Zhivov A, Blum M, Guthoff RF, Stachs O. Real-time mapping of the subepithelial nerve plexus by in vivo confocal laser scanning microscopy. Br. J. Ophthalmol. 2010;94:1133–1135. doi: 10.1136/bjo.2009.175489. [DOI] [PubMed] [Google Scholar]
  • 15.Poletti, E., Wigdahl, J., Guimarães, P. & Ruggeri, A. Automatic montaging of corneal sub-basal nerve images for the composition of a wide-range mosaic in Proceedings of the 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 5426–5429 (IEEE, 2014). [DOI] [PubMed]
  • 16.Turuwhenua JT, Patel DV, McGhee CNJ. Fully automated montaging of laser scanning in vivo confocal microscopy images of the human corneal subbasal nerve plexus. Invest. Ophthalmol. Vis. Sci. 2012;53:2235–2242. doi: 10.1167/iovs.11-8454. [DOI] [PubMed] [Google Scholar]
  • 17.Edwards K, et al. Wide-field assessment of the human corneal subbasal nerve plexus in diabetic neuropathy using a novel mapping technique. Cornea. 2012;31:1078–1082. doi: 10.1097/ICO.0b013e318245c012. [DOI] [PubMed] [Google Scholar]
  • 18.Vaishnav YJ, Rucker SA, Saharia K, McNamara NA. Rapid, automated mosaicking of the human corneal subbasal nerve plexus. Biomed. Tech. (Berl.) 2017;62:609–613. doi: 10.1515/bmt-2016-0148. [DOI] [PubMed] [Google Scholar]
  • 19.Lagali NS, et al. Reduced Corneal Nerve Fiber Density in Type 2 Diabetes by Wide-Area Mosaic Analysis. Invest. Ophthalmol. Vis. Sci. 2017;58:6318–6327. doi: 10.1167/iovs.17-22257. [DOI] [PubMed] [Google Scholar]
  • 20.Lagali, N. S. et al. Wide-field corneal subbasal nerve plexus mosaics in age-controlled healthy and type 2 diabetes populations. Sci. Data (accepted, 2018). [DOI] [PMC free article] [PubMed]
  • 21.Allgeier S, et al. Mosaicking the subbasal nerve plexus by guided eye movements. Invest. Ophthalmol. Vis. Sci. 2014;55:6082–6089. doi: 10.1167/iovs.14-14698. [DOI] [PubMed] [Google Scholar]
  • 22.Köhler B, et al. EyeGuidance – a computer controlled system to guide eye movements. Curr. Dir. Biomed. Eng. 2016;2:433–436. [Google Scholar]
  • 23.Allgeier S, et al. A novel approach to analyze the progression of measured corneal sub-basal nerve fiber length in continuously expanding mosaic images. Curr. Eye. Res. 2017;42:549–556. doi: 10.1080/02713683.2016.1221977. [DOI] [PubMed] [Google Scholar]
  • 24.Ziegler D, et al. Early detection of nerve fiber loss by corneal confocal microscopy and skin biopsy in recently diagnosed type 2 diabetes. Diabetes. 2014;63:2454–2463. doi: 10.2337/db13-1819. [DOI] [PubMed] [Google Scholar]
  • 25.Allgeier S, et al. Image reconstruction of the subbasal nerve plexus with in vivo confocal microscopy. Invest. Ophthalmol. Vis. Sci. 2011;52:5022–5028. doi: 10.1167/iovs.10-6065. [DOI] [PubMed] [Google Scholar]
  • 26.Allgeier, S. et al. Elastische Registrierung von in-vivo-CLSM-Aufnahmen der Kornea in Bildverarbeitung für die Medizin 2011. Algorithmen - Systeme - Anwendungen (eds Handels, H., Ehrhardt, J., Deserno, T. M., Meinzer, H.-P. & Tolxdorff, T.), 149–153 (Springer, 2011).
  • 27.Bartschat, A. et al. Automatic corneal tissue classification using bag-of-visual-words approaches in Forum Bildverarbeitung 2016 (eds Puente León, F. & Heizmann, M.), 245–256 (KIT Scientific Publishing, 2016).
  • 28.Petropoulos IN, et al. Rapid automated diagnosis of diabetic peripheral neuropathy with in vivo corneal confocal microscopy. Invest. Ophthalmol. Vis. Sci. 2014;55:2071–2078. doi: 10.1167/iovs.13-13787. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Zhivov A, Stachs O, Stave J, Guthoff RF. In vivo three-dimensional confocal laser scanning microscopy of corneal surface and epithelium. Br. J. Ophthalmol. 2009;93:667–672. doi: 10.1136/bjo.2008.137430. [DOI] [PubMed] [Google Scholar]
  • 30.Messmer EM, Schmid-Tannwald C, Zapp D, Kampik A. In vivo confocal microscopy of corneal small fiber damage in diabetes mellitus. Graefes Arch. Clin. Exp. Ophthalmol. 2010;248:1307–1312. doi: 10.1007/s00417-010-1396-8. [DOI] [PubMed] [Google Scholar]
  • 31.Pritchard N, et al. Longitudinal assessment of neuropathy in type 1 diabetes using novel ophthalmic markers (LANDMark). Study design and baseline characteristics. Diabetes Res. Clin. Pract. 2014;104:248–256. doi: 10.1016/j.diabres.2014.02.011. [DOI] [PubMed] [Google Scholar]
  • 32.Bartschat, A. et al. Augmentations of the bag of visual words approach for real-time fuzzy and partial image classification in Proceedings. 27. Workshop Computational Intelligence. Dortmund, 23. - 24. November 2017 (eds Hoffmann, F., Hüllermeier, E. & Mikut, R.), 227–242 (KIT Scientific Publishing, 2017).

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Video S5 (10.1MB, mp4)
Supplementary Video S6 (10.1MB, mp4)
Supplementary Video S7 (7.6MB, mp4)
Supplementary Video S8 (2.5MB, mp4)
Supplementary Video S9 (10.3MB, mp4)

Data Availability Statement

The datasets generated during the current study are available from the corresponding author on reasonable request.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group