Glucose meters differ from many other analytes since for some time, glucose meters have had accuracy performance standards for patient samples. The ISO 15197 standard of 2003, while for home glucose meters, was adopted by the Clinical Laboratory Standards Institute (CLSI) C30-A2, to specify glucose meter performance for point-of-care testing (POCT) meters. I critiqued these standards1-3 since, among other things, they left 5% of the results as unspecified.
In 2013, CLSI issued an updated standard which had more narrow limits but still left 2% of the results as unspecified.4 This prompted another critique,5 and in 2014 the Food and Drug Administration (FDA) issued a draft guidance that had a remarkable statement: “FDA believes that the criteria set forth in the ISO 15197 standard do not adequately protect patients using BGMS devices in professional settings, and does not recommend using these criteria.” FDA also issued performance limits for 100% of the data.
But now, in 2016, FDA has released an updated guidance,6 which is no longer a draft, which again leaves 2% of the results as unspecified!
A summary of the performance limits is listed in Table 1.
Table 1.
Year | Source | Percentage of data specified | Details |
---|---|---|---|
2003 | ISO | 95 | 95% of results must be < ±15 mg/dL below 75 mg/dL and < ±20% above 75 mg/dL |
2013 | CLSI | 98 | 95% of results must be < ±12 mg/dL below 100 mg/dL and < ±12.5% above 100 mg/dL No more than 2% > 15 mg/dL below 75 mg/dL and > 20% above 75 mg/dL |
2014 | FDA | 100 | 99% of results must be < ±7 mg/dL below 70 mg/dL and < ±10% above 70 mg/dL 0% > 15 mg/dL below 70 mg/dL and > 20% above 70 mg/dL |
2016 | FDA | 98 | 95% of results must be < ±12 mg/dL below 75 mg/dL and < ±12% above 75 mg/dL 98% of results must be < ±15 mg/dL below 75 mg/dL and < ±15% above 75 mg/dL |
One argument that has surfaced is that one can never prove that 100% of the results will fall within limits. While this is true, it is illogical to use this fact as a reason to exclude a percentage of results from being specified. Nor does one see this in other medical arenas. Thus, one does not hear that 98% of surgeries should be performed on the correct site.
A simple remedy is to use an error grid for 100% of the data.
Footnotes
Abbreviations: CLSI, Clinical Laboratory Standards Institute; FDA, Food and Drug Administration; POCT, point-of-care testing.
Declaration of Conflicting Interests: The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: JSK is an employee of Krouwer Consulting.
Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This work was funded by Krouwer Consulting.
References
- 1. Krouwer JS. Wrong thinking about glucose standards. Clin Chem. 2010;56:874-875. [DOI] [PubMed] [Google Scholar]
- 2. Krouwer JS, Cembrowski GS. A review of standards and statistics used to describe blood glucose monitor performance. J Diabetes Sci Technol. 2010;4:75-83. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Krouwer JS. Why specifications for allowable glucose meter errors should include 100% of the data. Clin Chem Lab Med. 2013;51:1543-1544. [DOI] [PubMed] [Google Scholar]
- 4. Clinical and Laboratory Standards Institute. POCT12-A3 Point-of-care blood glucose testing in acute and chronic care facilities; approved guideline—third edition. [Google Scholar]
- 5. Krouwer JS. The new glucose standard, POCT12-A3 misses the mark. J Diabetes Sci Technol. 2013;7:1400-1402. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Food and Drug Administration. Self-monitoring blood glucose test systems for over-the-counter use. October 11, 2016. http://www.fda.gov/downloads/medicaldevices/deviceregulationandguidance/guidancedocuments/ucm380327.pdf. Accessed October 19, 2016.