Abstract
Background:
We propose a methodology to analyze complex real-life glucose data in insulin pump users.
Methods:
Patients with type 1 diabetes (T1D) on insulin pumps were recruited from an academic endocrinology practice. Glucose data, insulin bolus (IB) amounts, and self-reported alcohol consumption and exercise events were collected for 30 days. Rules were developed to retrospectively compare IB recommendations from the insulin pump bolus calculator (IPBC) against recommendations from a proposed decision aid (PDA) and for assessing the PDA’s recommendation for exercise and alcohol.
Results:
Data from 15 participants were analyzed. When considering instances where glucose was below target, the PDA recommended a smaller dose in 14%, but a larger dose in 13% and an equivalent IB in 73%. For glucose levels at target, the PDA suggested an equivalent IB in 58% compared to the subject’s IPBC, but higher doses in 20% and lower in 22%. In events where postprandial glucose was higher than target, the PDA suggested higher doses in 25%, lower doses in 13%, and equivalent doses in 62%. In 64% of all alcohol events the PDA would have provided appropriate advice. In 75% of exercise events, the PDA appropriately advised an IB, a carbohydrate snack, or neither.
Conclusions:
This study provides a methodology to systematically analyze real-life data generated by insulin pumps and allowed a preliminary analysis of the performance of the PDA for insulin dosing. Further testing of the methodological approach in a broader diabetes population and prospective testing of the PDA are needed.
Keywords: alcohol, bolus calculator, exercise, insulin pump, postprandial blood glucose, retrospective analysis
Current standards of care for patients with type 1 diabetes (T1D) advocate for tight control of blood glucose (BG).1 One treatment challenge for patients with T1D is optimization of postprandial glucose levels.2-4 To help patients achieve improved glucose regulation, continuous subcutaneous insulin infusion devices (CSII, aka “insulin pumps”) sometimes coupled with continuous glucose monitoring systems (CGMs), have been developed. Although devices can assist patients in making insulin dosing decisions through the use of bolus calculators, it is unknown how accurate the bolus recommendations are in real-life scenarios when complex lifestyle choices, such as exercise and alcohol intake, have to be considered in decision making. Recent data suggests that patients are often confused and inconsistent when trying to factor in these behaviors when deciding insulin doses.5,6
Models exist to study insulin delivery recommendations in controlled, simulated settings. Before undergoing clinical trials, a common practice to facilitate the design, development and testing of diabetes technology is to use in silico methods.7-12 Recently, Wong et al proposed a method to retrospectively compare insulin bolus (IB) recommendations using Intensive Care Unit (ICU) data.13 They concluded that in silico comparisons appear to be an efficient nonclinical method for allowing rapid and inexpensive identification of computer-based protocols that justify expensive and burdensome clinical trials.
Although models exist to study IB recommendations in controlled environments, there is a lack of methods capable of analyzing glucose data simultaneously with patient behaviors. The aims of this study were to (1) develop an analytic method to retrospectively compare prandial IB recommendations, (2) apply the proposed method in a real-life setting to test the performance of an evidence-based proposed decision aid (PDA) against the bolus calculator of an insulin pump, and (3) share lessons learned from collecting, aggregating and analyzing real-life data generated by insulin pumps and self-reported patient behaviors.
Methods
Description of the iDECIDE Evidence-Based Insulin Bolusing Dosing Decision Aid
iDECIDE, the PDA evaluated here, is an evidence-based decision aid to recommend IB doses, carbohydrate intake, or both, by taking into account carbohydrates and alcohol consumed, and/or exercise plans.14 The PDA was deployed as a smartphone app to help patients with T1D incorporate varied lifestyle choices simultaneously into decisions about prandial insulin dosing. The PDA is based on the formula proposed by Colin15 to include alcohol,16,17 exercise,18-21 and the absorption rate of rapid-acting insulin.22 The PDA corrects to the nearest target glucose setting when the blood glucose is out of range, but would not account for the CGMS trendline. Exercise is accounted for based on body weight and duration and intensity of exercise, while the alcoholic beverage type and volume consumed are necessary to adjust for alcoholic beverages.
When the user launches the PDA application the first time he is prompted to set up a diabetes profile: weight, insulin-to-carbohydrate ratios, target BG levels, correction factors and active insulin time.23 Although participants did not set up their user profile for the study, those that did not use paper logs interacted with the self-reporting module to log (1) exercise, describing duration and intensity, (2) food intake, specifying food type, serving size and carbohydrate content, and (3) alcohol intake, indicating number of drinks, size, and type of drink (Figure 1). In addition, when self-reporting plans, the user is expected to enter the BG reading. The PDA subsequently recommends an IB or carbohydrate intake by incorporating current evidence on the way food and alcohol carbohydrates and exercise influence BG, but these recommendations were assessed retrospectively and were not provided to the participants.
Participant Recruitment
Following Institutional Review Board approval, 31 study participants were recruited from an outpatient academic endocrinology practice. Patients with T1D 18 years or older who had been under the care of the endocrinology team while on CSII therapy using a Medtronic (Minneapolis, MN) insulin pump for at least one year were eligible to participate.
Data Collection
Participants were asked to continue their usual fitness and nutrition routine. For 30 days, participants recorded their exercise activity and alcohol consumption via paper logs or the self-reporting module of the PDA, according to subject’s preferences. Exercise was recorded by start time, duration and intensity, and categorized as light, moderate or vigorous. Alcohol was recorded by tracking drink time, type, volume, and number (eg, 6PM, 1 pint of beer, no carbohydrates entered). Carbohydrate content was entered in the insulin pump. After 30 days, logs were manually encoded into tables or downloaded from a secure cloud-based server.
Self-reported data on exercise and alcohol was used as input for the PDA. For exercise, the PDA recommends an IB or carbohydrate intake by considering body weight and intensity and duration of exercise.18-21 For alcohol, the PDA accounts for the carbohydrates of the alcoholic drinks based on type, volume and count.
CSII data from the corresponding 30-day timeframe was downloaded in tabular format. CSII device data included carbohydrates recorded by the participant, glucose levels either from a CGMS or capillary BG monitor or both, amount of insulin delivered, pump settings, and the IB suggested by the insulin pump bolus calculator (IPBC).
Retrospective Comparison of Two Insulin Bolus Calculators
To evaluate the performance of the PDA against conventional approaches to prandial insulin dosing, the authors adapted methodology from Wong et al.13 For this study, the conventional approaches to insulin dosing were defined as either use of the IPBC or participant’s self-determined doses. The PDA’s recommendations were compared against those made by the participant’s IPBC, or against the participant when they either overrode or neglected to get advice from their IPBC (Figure 2).
The “appropriateness” of an IB was defined as one that brings the postprandial glucose to the desired target.13 The method assumes that a conventional insulin dosing calculator, BCa (ie, IPBC or the participant), has made an IB recommendation. The point in time when BCa made the IB suggestion and when the insulin was delivered is referred to as the initial time, ti. The method assumes that a proposed insulin dosing calculator, BCp (eg, PDA), is retrospectively executed at the same data point, ti, to compare at time ti+1 the effect on BG of the insulin suggestion from BCp against the actual suggestion that was made by BCa. We considered that one calculator “outperformed” another calculator if there was a major performance enhancement over the competitor. For instance, in the case of a low postprandial BG we consider that a lower insulin dose recommendation outperformed higher insulin dose advice, potentially avoiding a hypoglycemic event.
Applying this methodology requires that each preprandial BG at ti can be paired with a corresponding postprandial BG at ti+1. For meal events and BG corrections, we defined ti+1 to be the first BG reading obtained 3 hours ± 15 minutes, after ti. This time frame was chosen considering that the majority of the carbohydrate load and the rapid-acting insulin analog bolus would have been absorbed and BG levels would have stabilized.16 The BG readings at ti+1 were broken into three categories, based on predetermined individual target BG levels obtained from the insulin pump settings of each participant. The analysis determines which calculator provided at time ti an IB recommendation that would have placed the participant closer to their target BG based on the category of the actual BG reading at ti+1. In the case of a target postprandial BG reading, we considered that a smaller insulin recommendation outperforms a larger recommendation because it could have avoided a hypoglycemic event.
The method outlined in Figure 2 was used to compare the appropriateness of two calculators, BCa and BCp, and assumes that BCa (IPBC) has made IB recommendations that were delivered to the patient. A variation of that method is needed to assess the appropriateness of recommendations from BCp (PDA) when there is no available data from BCa (ie, no recommendation from the IPBC).
Assessing the Appropriateness of an Insulin Bolus Recommendation for Alcohol and Exercise
Conventional IPBCs do not provide IB recommendations for alcohol. For these cases the method explained in Figure 3 was adopted. The postprandial time frame of interest, ti+1, was defined as the first BG reading obtained within 3 hours ± 15 minutes. This time frame neglects to consider any delayed effects from alcohol induced hypoglycemia and primarily focuses on the carbohydrates associated with alcoholic beverages.
As with alcohol ingestion, when participants exercised there were no recommendations made by the IPBC. For those cases, we used the method in Figure 4. We modified the window of ti+1 to be the first BG reading within 15 minutes of finishing exercise as recorded by the participant to detect any immediate effects of exercise-induced hypoglycemia. For example, if the participant finished exercising at 8:30 am, we used the first available BG between 8:30 and 8:45 am. In the case of exercise, the PDA’s recommendations could be a carbohydrate snack in addition to an IB dose. For exercise scenarios, the appropriateness of the IB and/or carbohydrate was defined as in Figure 4.
Data Analysis
Computer programs were written to automate the process of collating and analyzing the data generated by the insulin pumps with the self-reported patient behaviors. In addition, assessing the performance of the PDA at ti+1 against the IPBC, or against participants’ self-dosing choices when the IPBC was not used as anticipated was automated. Comparisons were made according to participant glucose targets (below, at, or above target).
Results
Participant Characteristics and Data
There were 31 participants recruited for the study, with 4 withdrawals. Of the remaining 27 participants, a subset of 15 participants (Table 1) had preprandial glucose readings paired with ti+1 BG readings, with 13 of them on CGMS (9 on Minimed 530G-551, 3 on Minimed 530G-751, and 1 on Paradigm Revel-723).
Table 1.
Characteristic | Value |
---|---|
Age (years) | 48.7 (13.9) |
% women | 73.3 |
% white | 93.3 |
Hemoglobin A1C | 7.5 (1.2) |
Diabetes duration (years) | 26.9 (11.8) |
Duration on insulin pump (years) | 11.5 (5.3) |
Daytime low/high target BG | 89.9 (8.6) / 112.3 (10.8) |
# Analyzable exercise events/day | 1.1 (0.34) |
# Analyzable alcohol events/day | 0.2 (0.18) |
Data are mean (SD) or %.
A total of 2104 events had postprandial glucose readings that allowed for a comparison between the IPBC and the PDA, and there were 419 events where the PDA was compared against cases where the participants did not use their IPBC, they overrode the IPBC recommendations, or they did not provide a prandial BG. There were 235 exercise and 105 alcohol events that had sufficient data for analysis. Most (56%) exercise events were of moderate intensity. There were few (14%) alcohol events where participants accounted for the carbohydrates associated with the beverage.
IPBCs allow different settings (BG target, insulin-to-carbohydrate ratio, and correction factor) throughout the day and the PDA accounted for these different settings for each participant at each time of day. While participants used different Medtronic insulin pumps, all used the same formula for computing IB recommendations, and an adaptation of Mudaliar et al24 for computing active insulin. The Medtronic 530G includes a threshold suspend feature, that is designed to automatically stop insulin delivery when the CGMS value falls below a patient-specific preset threshold. There were 5 insulin suspension events that occurred in close temporal proximity to events of interest; such low frequency did not warrant removing data from the analysis.
Comparison of iDECIDE Against the IPBC or Patient
We used the method described in Figure 2 to compare the appropriateness of the PDA’s recommendations against events when the patient followed the IPBC recommendations for BG correction doses and/or carbohydrate loads that included a prandial and postprandial BG.
First assessed was how the PDA (ie, iDECIDE) compared against the IPBC (Table 2). The IPBC brought the participants to target glucose levels in 13% (278/2104) events, below target in 10% (207/2104) and above target in 77% (1619/2104). When considering very low and very high postprandial BG, the BG was below 70 mg/dl in 3% (55/2104) and over 180 mg/dl in 35% (737/2104). When considering instances where glucose was below target, the PDA would have recommended an appropriately smaller dose in 14% (28/207), but a larger dose in 13% (27/207) and an equivalent IB in 73% (152/207). For glucose levels at target, the PDA would have suggested an equivalent IB in 58% (162/278) compared to the subject’s IPBC, but a higher dose in 20% (56/278) and lower in 22% (60/278). In events where postprandial glucose was higher than target, the PDA would have suggested a higher dose in 25% (406/1619), a lower dose in 13% (212/1619), and an equivalent dose in 62% (1001/1619). Overall, the PDA would have recommended an equivalent dose compared to the IPBC in 63% (1315/2104) of IB decisions.
Table 2.
Event type | Postprandial BG (mg/dl) | PDA insulin recommendations |
Total | ||
---|---|---|---|---|---|
Larger dose | Smaller dose | Equivalent dose | |||
IPBC | Low (<target) | 27a | 28b | 152c | 207 |
Target (participant target) | 56a | 60b | 162d | 278 | |
High (>target) | 406b | 212a | 1001c | 1619 | |
Total | 489 | 300 | 1315 | 2104 | |
Participant | Low (<target) | 10a | 23b | 21c | 54 |
Target (participant target) | 18a | 24b | 4d | 46 | |
High (>target) | 107b | 86a | 126c | 319 | |
Total | 135 | 133 | 151 | 419 |
IPBC (or participant) was appropriate and the PDA recommendation was not appropriate, IPBC (or participant) outperformed the PDA.
PDA recommendation was appropriate and IPBC (or participant) was not appropriate, PDA outperformed the bolus calculator (or patient). When the PDA recommends a lower insulin dose recommendation than the bolus calculator (or participant) and the postprandial BG is on target, the PDA could potentially avoid a hypoglycemic event and therefore outperformed the bolus calculator (or participant).
Events where the PDA and IPBC (or participant) recommendations were not appropriate.
Events where the PDA and IPBC (or participant) recommendations were appropriate.
We used the method described in Figure 2 to compare the appropriateness of the PDA against decisions made by the participant (Table 2). The participants self-dosing led to above target postprandial glucose in 76% (319/419), below target in 13% (54/419) while participants only achieved target glucose levels in 11% (46/419). There were 3% (14/419) of the events with a postprandial BG below 70 mg/dl and 37% (154/419) over 180 mg/dl. When considering instances where glucose was below target, the PDA would have recommended an appropriately smaller dose in 43% (23/54), a larger dose in 19% (10/54), and an equivalent IB dose in 38% (21/54). For glucose levels at target, the PDA would have suggested an equivalent IB amount in 9% (4/46) compared to the subject’s own decision, but a higher dose 39% (18/46) and lower in 52% (24/46). In situations where postprandial glucose was greater than target, the PDA would have suggested a higher dose in 34% (107/319), a lower dose in 27% (86/319), and an equivalent dose in 39% (126/319). Overall, the PDA would have recommended an equivalent IB in only 36% (151/419) of instances compared to when the participants made their own decisions.
Assessment of the Appropriateness of iDECIDE’s Recommendations for Exercise and Alcohol
In cases of exercise and alcohol the pump does not suggest insulin. In these cases, the PDA is only assessed based on the BG outcomes since it could not be compared against the IPBC. We used the method described in Figure 3 to assess the appropriateness of the PDA’s recommendations when alcohol consumption was recorded. As reported earlier, patients self-reported accounting for the carbohydrate content of the beverage in 15 of the 105 events. As indicated in Table 3, in 64% (67/105) of overall alcohol events the PDA would have provided appropriate advice. The PDA performed well when the postprandial BG was high with 78% (64/82) appropriate IB recommendations, but had poor performance when the postprandial BG was at target with only 5% (1/19) recommendations deemed appropriate.
Table 3.
Postprandial BG | PDA recommendations |
Total | |
---|---|---|---|
Appropriate | Not appropriate | ||
Low (<target) | 2 | 2 | 4 |
Target (participant target) | 1 | 18 | 19 |
High (>target) | 64 | 18 | 82 |
Total | 67 | 38 | 105 |
We used the method described in Figure 4 to assess the appropriateness of the PDA’s recommendation before exercise (Table 4). The PDA appropriately suggested insulin or to ingest carbohydrates in 75% (176/235). Similar to the alcohol results, the PDA performed well when post exercise BG was high 87% (154/178), but only made appropriate suggestions in 37% (10/27) and 40% (12/30) when the post exercise BG was low or target, respectively. There were 26 exercise events that had a duration of 90 minutes or longer and the PDA made appropriate recommendations in only 27%.
Table 4.
Post exercise BG | PDA insulin dose and carbohydrate recommendations |
Total | |
---|---|---|---|
Appropriate | Not appropriate | ||
Low (<target) | 10 | 17 | 27 |
Target (participant target) | 12 | 18 | 30 |
High (>target) | 154 | 24 | 178 |
Total | 176 | 59 | 235 |
Discussion
Although advances in in silico models technology have allowed for incorporation of new features into existing technologies to improve BG control, these often do not account for variables that affect BG (eg, exercise, stress, sleep, and illness). Decision aids that assist patients with T1D to make better prandial insulin dosing decisions are needed, particularly when patients must account for multiple simultaneous lifestyle variables that may impact BG levels.
One of the main differences between this study and others that retrospectively evaluated the performance of prandial insulin dosing recommendations is the source of the clinical data. For instance, previous studies have compared the effectiveness of insulin dosing recommendations in controlled environments such as in the ICU,13,25 where glucose control is closely monitored and tracked and lifestyle behaviors are not a factor. In contrast, this study focused on free-living outpatients who made their own choices about insulin therapy, and where individual lifestyle choices have the potential to impact treatment decisions and outcomes.
One of the analytic challenges we encountered when developing, testing, and comparing the effectiveness of insulin dosing recommendations is the complex nature of data generated by free-living participants. In our study, many of the self-management and daily living activities recorded by the participants occurred in tight temporal succession and could not be assessed as isolated events. This required development of a new analytic approach to evaluating the data. An unexpected positive outcome of this study was gaining a better understanding of patients’ self-management behaviors as they interact with insulin pumps.5,6
The methodology outlined here permitted an assessment of how our PDA would perform when used in different scenarios. When compared to the IPBC embedded in the subject’s insulin pump, the PDA in general was noninferior, recommending IB doses equivalent to the IPBC standard in 63% of decisions overall and nearly equivalent number of smaller doses when glucose levels were below or at target. There were some instances where the PDA was superior to the IPBC, such as when it would have recommended larger doses in cases when postprandial glucose levels were above target. Initial analysis of the PDA in cases where the doses were too large or small, provided insights which were used to improve performance with continuing analysis necessary for further refinement of the recommendations.26,27 For instance, we used an initial setting of 3 hours of active insulin time to calculate IOB. To improve performance, this was later adjusted to 4 hours which reduced the number of inappropriate recommendations that could have led to hypoglycemia. In the future, the PDA will adapt to the insulin action time specified for each patient.
Employing the analytic paradigms developed here, we also assessed the performance of the PDA when there was a lack of recommendations from the IPBC with exercise and alcohol events. In these analyses the postprandial glucose was used as the outcome measure. For cases involving alcohol consumption, the PDA may have offered an advantage when deciding a compensatory insulin bolus. The PDA could have improved postexercise BG when the duration was 90 minutes or less and the PDA should be restricted to such events until further study.
There are limitations to the study. This study incorporated self-reported data for exercise, meal and alcohol behaviors. It is possible participants did not record all these events, or may have recorded them inaccurately. Also, participants’ insulin pump settings were not adjusted for the study. Inappropriate insulin pump settings, such as basal rates, could have influenced the results. Sample sizes for alcohol and exercise events were small with respect to the larger comparisons involving the IPBC. The study also did not consider late-onset hypoglycemia that can arise from engaging in exercise, and possibly when consuming alcohol. To automate the analysis, we opted against determining an appropriate postexercise timeframe on a case-by-case basis and instead focused on the immediate effects of exercise by employing a standard 15-minute postexercise timeframe. Considering BG levels outside of the time-frames used for analysis in this study is another important factor to consider in the future when assessing and calibrating IB calculators.
In addition, the analysis was done retrospectively. A prospective analysis, where the PDA makes suggestions in real time, would help further delineate its capabilities, improve performance and assess user acceptance. A recent analysis suggests that mobile apps can offer advantages in diabetes management, but more rigorous studies are needed.28 Finally, the analytic methods tested here were for a very specialized group of patients (T1D on insulin pumps) and we did not conduct an analysis of the outcomes in relation to A1c scores. Testing these methodologies in a wider selection and more diverse population of patients (eg, T1D patients on multiple daily insulin injections or patients with type 2 diabetes) would be needed to test the generalizability of the approach.
Conclusion
We introduced an analytic method to use prospective real-life data to retrospectively compare insulin dosing recommendations. This method was used to assess the recommendations of an evidence-based decision aid. Additional prospective testing of the proposed decision aid with a bigger patient cohort is being planned to further validate the proposed method.
Footnotes
Abbreviations: BG, blood glucose; CGMS, continuous glucose monitoring system; CSII, continuous subcutaneous insulin infusion devices; IB, insulin bolus; ICU, intensive care unit; IPBC, insulin pump bolus calculator; PDA, proposed decision aid; T1D, type 1 diabetes.
Authors’ Note: Danielle Groat and Maria A. Grando share first authorship of this article.
Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: iDECIDE: Patient-centered decision support based on device data (1U54HL108460), funded by NLM.
References
- 1. American Diabetes Association. Standards of medical care in diabetes—2016. Diabetes Care. 2016;39(suppl 1):S52-S59.26696682 [Google Scholar]
- 2. Aiello LP, DCCT/EDIC Research Group. Diabetic retinopathy and other ocular findings in the diabetes control and complications trial/epidemiology of diabetes interventions and complications study. Diabetes Care. 2014;37(1):17-23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. De Boer IH, DCCT/EDIC Research Group. Kidney disease and related findings in the diabetes control and complications trial/epidemiology of diabetes interventions and complications study. Diabetes Care. 2014;37(1):24-30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Martin CL, Albers JW, Pop-Busui R, DCCT/EDIC Research Group. Neuropathy and related findings in the diabetes control and complications trial/epidemiology of diabetes interventions and complications study. Diabetes Care. 2014;37(1):31-38. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Groat D, Grando MA, Soni H, et al. Self-management behaviors in adults on insulin pump therapy what are patients really doing? J Diabetes Sci Technol. 2017;11(2):233-239. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Grando MA, Groat D, Soni H, et al. Characterization of exercise and alcohol self-management behaviors of type 1 diabetes patients on insulin pump therapy. J Diabetes Sci Technol. 2017; 11(2):240-246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Colmegna P, Peña RS. Analysis of three T1DM simulation models for evaluating robust closed-loop controllers. Comput Methods Programs Biomed. 2014;113(1):371-382. [DOI] [PubMed] [Google Scholar]
- 8. Dalla Man C, Micheletto F, Lv D, Breton M, Kovatchev B, Cobelli C. The UVA/PADOVA type 1 diabetes simulator new features. J Diabetes Sci Technol. 2014;8(1):26-34. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Percival MW, Wang Y, Grosman B, et al. Development of a multi-parametric model predictive control algorithm for insulin delivery in type 1 diabetes mellitus using clinical parameters. J Process Control. 2011;21(3):391-404. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Schiavon M, Man CD, Kudva YC, Basu A, Cobelli C. The artificial pancreas on the threshold of ambulatory use: setting the stage for a critical transition: in silico optimization of basal insulin infusion rate during exercise: implication for artificial pancreas. J Diabetes Sci Technol. 2013;7(6):1461. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. van Heusden K, Dassau E, Zisser HC, Seborg DE, Doyle FJ. Control-relevant models for glucose control using a priori patient characteristics. IEEE Trans Biomed Eng. 2012;59(7):1839-1849. [DOI] [PubMed] [Google Scholar]
- 12. Turksoy K, Samadi S, Feng J, Littlejohn E, Quinn L, Cinar A. Meal detection in patients with type 1 diabetes: a new module for the multivariable adaptive artificial pancreas control system. IEEE J Biomed Health Inform. 2016;20(1):47-54. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Wong AF, Pielmeier U, Haug PJ, Andreassen S, Morris AH. An in-silico method to identify computer-based protocols worthy of clinical study: an insulin infusion protocol use case. J Am Med Inform Assoc. 2016;23(2):283-288. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Lloyd B, Groat D, Cook CB, Kaufman D, Grando MA. iDECIDE: a mobile application for insulin dosing using an evidence based equation to account for patient preferences. Paper presented at: MEDINFO Conference; August 19-23, 2015; São Paulo, Brazil. [PMC free article] [PubMed] [Google Scholar]
- 15. Colin IM, Paris I. Glucose meters with built-in automated bolus calculator: gadget or real value for insulin-treated diabetic patients? Diabetes Ther. 2013;4(1):1-11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Richardson T, Weiss M, Thomas P, Kerr D. Day after the night before: influence of evening alcohol on risk of hypoglycemia in patients with type 1 diabetes. Diabetes Care. 2005;28(7):1801-1802. [DOI] [PubMed] [Google Scholar]
- 17. Turner BC, Jenkins E, Kerr D, Sherwin RS, Cavan DA. The effect of evening alcohol consumption on next-morning glucose control in type 1 diabetes. Diabetes Care. 2001;24(11):1888-1893. [DOI] [PubMed] [Google Scholar]
- 18. Bolderman KM. Putting Your Patients on the Pump. 2nd ed. Alexandria, VA: American Diabetes Association; 2013. [Google Scholar]
- 19. Kourtoglou GI. Insulin therapy and exercise. Diabetes Res Clin Pract. 2011;93(suppl 1):S73-S77. [DOI] [PubMed] [Google Scholar]
- 20. Toni S, Reali MF, Barni F, Lenzi L, Festini F. Managing insulin therapy during exercise in type 1 diabetes mellitus. Acta Bio-Medica Atenei Parm. 2006;77(suppl 1):34-40. [PubMed] [Google Scholar]
- 21. Rabasa-Lhoret R, Bourque J, Ducros F, Chiasson JL. Guidelines for premeal insulin dose reduction for postprandial exercise of different intensities and durations in type 1 diabetic subjects treated intensively with a basal-bolus insulin regimen (ultralente-lispro). Diabetes Care. 2001;24(4):625-630. [DOI] [PubMed] [Google Scholar]
- 22. Lindholm A, McEwen J, Riis AP. Improved postprandial glycemic control with insulin aspart. A randomized double-blind cross-over trial in type 1 diabetes. Diabetes Care. 1999;22(5):801-805. [DOI] [PubMed] [Google Scholar]
- 23. Diabetes Teaching Center at the University of California, San Francisco. Programming your pump. Diabetes Education Online. 2016. Available at: https://dtc.ucsf.edu/types-of-diabetes/type1/treatment-of-type-1-diabetes/medications-and-therapies/type-1-insulin-pump-therapy/how-to-use-your-pump/programming-your-pump/. Accessed December 15, 2016.
- 24. Mudaliar SR, Lindberg FA, Joyce M, et al. Insulin aspart (B28 asp-insulin): a fast-acting analog of human insulin: absorption kinetics and action profile compared with regular human insulin in healthy nondiabetic subjects. Diabetes Care. 1999;22(9):1501-1506. [DOI] [PubMed] [Google Scholar]
- 25. Allart L, Vilhelm C, Mehdaoui H, et al. An architecture for online comparison and validation of processing methods and computerized guidelines in intensive care units. Comput Methods Programs Biomed. 2009;93(1):93-103. [DOI] [PubMed] [Google Scholar]
- 26. Groat D, Soni H, Thompson B, Cook CB, Grando A. Introducing a method to retrospectively compare insulin dosing recommendations. Paper presented at: Diabetes Technology Meeting; October 2015; Bethesda, MD. [Google Scholar]
- 27. Groat D, Grando A, Wallstrom G, et al. Retrospective evaluation of an evidence-based equation for insulin dosing accounting for exercise and alcohol. Paper presented at: American Diabetes Association 76th Scientific Session; June 2016; New Orleans, LA. [Google Scholar]
- 28. Hou C, Carter B, Hewitt J, Francisa T, Mayor S. Do mobile phone applications improve glycemic control (HbA1c) in the self-management of diabetes? A systematic review, meta-analysis, and GRADE of 14 randomized trials. Diabetes Care. 2016;39(11):2089-2095. [DOI] [PubMed] [Google Scholar]