Abstract
Background
Prospective trials have demonstrated the advantage of dose-escalated radiotherapy for the biochemical and clinical control of intermediate risk prostate cancer. Dose escalation improves outcomes but increases risks of urinary and bowel toxicity. Recently the contribution of “spacers” positioned in the septum between the rectum and the prostate could improve the functional results of intensity modulated radiation therapy (IMRT). To date most of the evaluated devices were polyethylen glycol (PEG) and hyaluronic acid (HA). Men on the Spacer arm had decreased bowel toxicity and less decline in both urinary and bowel quality of life as compared to Control men in a randomized trial.
Methods
This is an interventional, multi-center study to evaluate the use of biodegradable inflatable balloon for patients with intermediate risk prostate cancer treated by IMRT (74 to 80 Gy, 2 Gy/fraction) with daily image guided radiotherapy.
Discussion
This multicenter prospective study will yield new data regarding dosimetric gain and implantation stages of Bioprotect balloon. Acute and late toxicities and quality of life will be registered too.
Trial registration
NCT02478112, date of registration: 15/06/2015.
Keywords: Implantable biodegradable balloon, Rectal spacer, Intensity modulated radiotherapy, Prostate cancer
Background
Prospective trials have demonstrated the advantage of dose-escalated radiotherapy for the biochemical and clinical control of prostate cancer. This benefit observed with three-dimensional conformational irradiation is counterbalanced by an increase of the urinary and digestive toxicities [1–4]. The Medical Research Council (MRC) conducted the MRC 01 randomized multicenter trial [2] comparing a conformal RT (2 Gy / session) delivering either 64 Gy or 74 Gy, in combination with neoadjuvant hormone therapy during 3 to 6 months. The 5-year biochemical relapse–free survival was 71% versus 60% (p = 0,0007) with 74 and 64 Gy respectively. In France the French Group for the Study of Uro-Genital Tumors (GETUG) conducted the GETUG 06 multicenter trial [3]. Dose escalation from 70 to 80 Gy provided a better 5-year biochemical outcome with slightly greater toxicity. Peeters and al. [4] reported the Dutch trial evaluating dose-response for 664 randomized patients in radiotherapy for prostate cancer. Patients were randomly assigned to a tridimensional conformal radiation treatment of either 68 Gy or 78 Gy (in 2 Gy fractions). The 5-year biological relapse–free survival was 54 and 64% respectively (p = 0.02). In these randomized trials dose escalation improved biological relapse free survival but was associated with higher rate of rectal toxicity.
There are no randomized trials comparing conformational three-dimensional conformational irradiation with intensity modulated radiation therapy (IMRT), but experiments conducted by several teams, including historically that of the Memorial Sloan-Kettering Cancer Center (MSKCC) [5] showed that it was possible to deliver increased radiation doses to the prostate while decreasing frequency of urinary and digestive complications of this “high dose” RT. However this approach imposes a very strict control of the position of the target volume (prostate) under the accelerator in order to translate this dosimetric advantage into clinical benefit. Image Guided Radiotherapy (IGRT) guarantees this positioning accuracy. First clinical benefits of using IGRT in combination with IMRT were published in 2012 by the MSKCC team in a retrospective analysis of 180 “IGRT” patients (fiducial markers implanted in the prostate and daily kV imaging) treated with 86.4 Gy, whatever the initial risk group, between 2008 and 2009 compared with a cohort of patients treated without IGRT between 2006 and 2007 [6]. Patients in high risk group in this study showed a significant improvement in biochemical control (from 77 to 97%, p = 0.041). For all analyzed patients, IGRT would lead to a significant reduction in urinary late toxicity: grade ≥ 2 toxicity rate decreased from 20 to 10.4% at 3 years thanks to the IGRT [6]. Despite significant technical advances (IMRT, IGRT) rectum dose remains a limiting factor in dose escalation. Although the role of moderate doses has been recently shown, severe toxicity is strongly related to high doses. Patients with V70 below or above 26% had a risk of grade 2 rectal morbidity of 13 and 54%, respectively [7].
Thanks to innovative techniques, rectal side effects could be reduced by moving the prostate away from the rectal wall through an injection of a biodegradable substance that creates a space in anterior perirectal fat. To date most of the evaluated devices were polyethylen glycol (PEG) and hyaluronic acid (HA). In Mok et al. review, a total of 11 studies involving human prostate cancer patients were identified in 6 studies using implants in patients treated with external beam radiotherapy and 5 studies treating patients with brachytherapy (BT). Four studies used PEG spacers, 5 used HA spacers, 1 study used implanted biodegradable balloons, and 1 study used collagen implants [8]. Prostate rectum (PR) separation created by the different PR spacers varied between 7 and 20 mm and was largely dependent on implantation protocol. The increased PR separation was associated with improved dosimetric rectal profiles. Relative reduction of V70 Gy ranged from 46 to 61%; V40 and V60 Gy were decreased too, from 40 to 65%. The use of prophylactic antibiotic therapy is estimated to reduce the risk of infection to less than 5% [8].
Outcomes following PEG spacer implantation was assessed by a prospective multicenter randomized controlled trial [9]. Computed tomography (CT) and magnetic resonance imaging (MRI) scans for treatment planning were used for 222 patients with prostate cancer with clinical stage T1 or T2. They were randomized to receive spacer implantation or no implantation (control). Image guided IMRT (79.2 Gy in 1.8-Gy fractions) was used. In this trial, spacer implantation was rated as “easy” or “very easy” in 98.7% of the patients. The hydrogel placement success rate was 99%. Overall acute rectal adverse event rates were the same between groups, but fewer spacer patients presented with rectal pain (p = 0.02). A significant decrease in late (3 to 15 months) rectal toxicity in the spacer group was noted (p = 0.04), with a 2.0 and 7.0% late rectal toxicity incidence in the spacer and control arms, respectively. At 6, 12, and 15 months, a lower ratio of spacer patients presented with bowel quality of life (QOL) decrease. 11.6% of spacer patients and 21.4% of control patients experienced 10-point decrease at 15 months (p = 0.087). Furthermore, at 6 months, 8.8% of spacer patients and 22.2% of control patients had 10-point urinary decreases (p = 0.003). At 3-years patients on the spacer group had less bowel toxicity and less decrease in both urinary and bowel QOL in comparison to control patients. On the control arm, 41% of patients presented with a detectable decline in bowel QOL (5-points) by patient reported outcomes, and 21% had a more serious decline (10-points). These rates were both reduced by 70% on the spacer arm (14 and 5%, respectively) [10].
The use of HA spacers in hypofractionated RT regimens were evaluated by Chapet et al. This phase II study aims to assess the rates of late rectal toxicities of grade ≥ 2 after hypofractionated radiotherapy of prostate cancer of 62 Gy in 20 fractions of 3.1 Gy with an HA spacer. Thirty-six patients with low- to intermediate-risk prostate cancer according to the D’Amico classification are included in the present protocol. As part of this phase 2 study, the patients received a 10 cm3 transperineal HA injection. HA spacer significantly reduced rectal wall dose and could allow a dose escalation from 6.5 Gy to 8.5 Gy per fraction without increasing the dose to the rectum. A phase 2 study is under way to assess the rate of acute and late rectal toxicities when SBRT (5 × 7.5 Gy) is combined with an injection of HA [11]. Other trials are currently evaluating rectal spacer in patients treated by stereotactic radiotherapy [https://clinicaltrials.gov/ct2/show/NCT02353832, https://clinicaltrials.gov/ct2/show/NCT02911922].
A biodegradable balloon can also be used: Bioprotect has designed an adapted device for this implantation procedure. Animal studies have confirmed its efficacy and also its good tolerance [12]. ProSpace® system is a deflated balloon made of a biodegradable polymer which is inserted perineally after hydrodissection thanks to an introducer.
The implantation procedure is performed under general anesthesia through a small perineal incision [13, 14]. A multi-institutional phase II study has been carried out in 6 centers using IMRT or 3D conformal RT [13]. Twenty three patients were analyzable and balloon was biodegraded within 6 months. The space between the prostate and rectum created by balloon implantation was about 2 cm, rising from 0.22 ± 0.2 cm to 2.47 ± 0.47 cm. This gap lasted during all the RT. In this first study three patients experienced acute urinary retention which resolved quickly following bladder drainage. In Melchert et al. [14] the prostate rectal wall separation resulted in an average reduction of the rectal V70% by 55.3% (±16.8%), V80% by 64.0% (±17.7%), V90% by 72.0% (±17.1%) in 26 patients.
Methods
This is an interventional, multi-center study to evaluate the use of the BioProtect Balloon Implant System for patients with a prostate cancer of intermediate risk treated by intensity modulated radiotherapy.
Study objectives
The main objective is to assess the dosimetric gain from the contribution of the implantable BioProtect balloon on organs at risk.
Secondary objectives are: evaluation of implantation stages of the balloon and its technical feasibility, evaluation of acute and late toxicities, correlation between the delay (between the implantation of the balloon and the radiotherapy) and the complications due to the balloon implantation, benefit from the Bioprotect Balloon Implant System compared to usual treatments for acute proctitis and measurement of the quality of life.
Main inclusion and exclusion criteria are specified in Table 1
Table 1.
Main inclusion criteria All the following must be met at the time of screening. – Patient over 18 years old – With a localized adenocarcinoma of the prostate: ○ of intermediate risk of D’AMICO ○ and of stage MRI < T3 – Requiring a treatment with Intensity Modulated Radiotherapy – PSA (Prostate-Specific Antigen) levels ≤20 ng/mL before external beam radiotherapy – Prostate volume > 15 cm3 – Short hormone therapy possibly associated (4-6 months) – Patient without clinical signs of progressive disease – Performance status ECOG (Eastern Cooperative Oncology Group) ≤ 1 – Life expectancy ≥10 years – Informed consent signed – Affiliation to a social security system | |
Main exclusion criteria – Incompatibility to the implantation of a Bioprotect balloon: ○ ongoing anticoagulant by vitamin K antagonist (VKA) or heparintherapy ○ patient with immunosuppression or with serious chronic diseases such as heart failure, cirrhosis, chronic kidney failure, colic or rectal digestive inflammatory disease ○ history of prostatitis or of repeated prostatic resections ○ history of recto-colic inflammatory disease or of lower gastrointestinal infection ○ untreated perineal wound – Prior treatment with hormone therapy (> 6 months) – History of another invasive cancer within 5 years prior to study entry (excepted a treated basal cell skin carcinoma) – History of pelvic radiotherapy – Severe hypertension non controlled by an adapted treatment (≥ 160 mmHg in systole and/or ≥ 90 mmHg in diastole) – Ongoing antineoplastic therapy – Person deprived of liberty or under tutorship – Inability to submit to the medical monitoring of the study for geographical, social or psychological reasons. – Conformal radiotherapy without intensity modulation |
Evaluation criteria
Dosimetric gain from the implantable BioProtect balloon on organs at risk (OAR)
A dosimetric computed tomographic (CT) scan is systematically performed before (CT1) and after (CT2) the implantation. Several dosimetric criteria will be evaluated including near maximum rectal dose, rectum volume receiving 90 to 50% of the prescribed dose at 74-80 Gy. A relative reduction of 50% of the rectal volume receiving at least 70 Gy is the primary endpoint. Bladder wall doses will also be reported at V50 and V60. Patients will be treated with IMRT and daily IGRT, 2 Gy/fraction, total dose 74-80 Gy.
Implantation stages of the balloon and technical feasibility
Balloon implantation, difficulties for implantation, implant procedure time (mn) and radiotherapy delivery will be evaluated.
Tolerance evaluation
Implant-related toxicities and urinary and rectal toxicities due to irradiation will be evaluated according to the NCI-CTCAE v4.0 scale. Acute toxicities will be defined as occurring within the first 6 months after radiotherapy, and late toxicity when occurring beyond that period (Table 2).
Table 2.
Within 60 days before the first day of radiation therapy | Within 30 days before the first day of radiation therapy | Within 3 weeks before the first day of radiation therapy | During radiotherapy | At the end of treatment | Monitoring period | Study exit | ||
---|---|---|---|---|---|---|---|---|
Before balloon implantation | After balloon implantation | |||||||
Clinical exams | ||||||||
History of the disease, main health history, previous treatments | X | |||||||
Complete clinical examination with assessment of the general condition (Performance Status) | X | X 1 / week |
X | X | X | |||
IPSS score | X | X | X At mid-treatment |
X | X At 3, 6, 12 et 24 months |
X | ||
Laboratory tests | ||||||||
PSA level | X | X At 3 months then every 6 months |
X | |||||
Quality of life | ||||||||
EORTC-QLQ-C30 Questionnaire | X | X | X At mid-treatment |
X | X At 3, 6, 12 et 24 months |
X | ||
EORTC-QLQ-PR25 Questionnaire | X | X | X At mid-treatment |
X | X At 3, 6, 12 et 24 months |
X | ||
Paraclinical exams | ||||||||
Tumor evaluation by computed tomography scanner | X | |||||||
Full description and measurements according to RECIST | X | |||||||
Dosimetric computed tomography scanner | X | X | ||||||
Treatments | ||||||||
Side effects related to radiotherapy | X 1 / week |
X | X At 3, 6, 12 et 24 months |
X |
The evaluation of the correlation between the delay (between positioning of the device and radiotherapy treatment) and the occurrence of complications due to the device implantation will be done by using a Wilcoxon or Student test (depending on the nature of the variables).
Quality of life
Patients will complete quality of life questionnaires prior to the start of treatment, at mid-treatment, at the end of the radiotherapy and at 3, 6, 12 and 24 months after the end of the radiotherapy (Table 2).
QLQ-C30 and QLQ-PR25 questionnaires will be completed by the patient. The first questionnaire (before implantation) will be completed at the first consultation. During the first consultation, the patient will be asked to complete these on-line questionnaires on the secured platform AQUILAB Share Place. He will receive personalized access codes to connect on line and fill out these questionnaires (Table 2). The results will be accessible to the physician on the platform.
Evaluation of urinary symptoms
The IPSS (International Prostate Score Symptom) was designed to be self-administered by the patient. Its French version has been validated. It is based on the answers to seven questions concerning urinary symptoms. The total score can therefore range from 0 to 35 (asymptomatic to very symptomatic). Symptoms are categorized as follows: mild (symptom score less than or equal to 7), moderate (symptom score range 8-19) or severe (symptom score range 20-35).
Statistics
Number of patients
In literature rectal spacers allows a relative reduction of about 50% of the rectal volume receiving at least 70Gy. The sample size of 50 (in order to obtain 44 evaluable patients) was calculated to have a 90% power at 5% statistical significance level to detect an irradiated volume difference V70 of 5 (decrease from 10 to 5%) with a standard deviation of 10. V70 with and without the balloon implantation will be compared using equality test on matched data: student test if data are normal or Wilcoxon test if data are non-normal.
Description of the device
BioProtect balloon implant
The BioProtect biodegradable Balloon Implant consists of a biodegradable inflatable balloon (mounted on a deployer) and an installation kit (echogenic needles, dilator and introducer sheath). It is provided as a sterile device. Once the balloon is in situ, it is inflated with sterile saline thanks to a plastic Luer-Lok syringe (20 ml or 50 ml). Users of this system must have been trained and certified by a Bioprotetc’s physician trainer before using the balloon. The implantation is performed under general anesthesia or neuroleptanalgesia. In case of difficulty during penetration of the perineum skin, a small incision (3-5 mm) using a scalpel at the point of insertion can be made. A minimum delay of 7 days is required between the implantation and the second computed tomography scanner.
Patient preparation
Two days before the implantation and during 5 consecutive days, the patient must be administered a broad-spectrum antibiotic (oral fluroquinolone). Prior to the implantation, patients must also use a bowel preparation (laxative).
It is recommended to introduce a urethral catheter into the bladder at the beginning of the session to empty the bladder and to help balloon positioning.
Discussion
Literature suggests that rectal spacer may reduce rectal side effects after prostate cancer radiotherapy. Several reports show important decrease of rectal dose [8]. Increased perirectal space using HA spacer reduced rectal irradiation, rectal toxicity severity, and decreased rates of patients experiencing declines in bowel quality of life in a randomized trial. Most of the series evaluated HA and PEG spacers. The aim of this study is to evaluate dosimetric gain, implantation procedure, acute and late toxicities with biodegradable BioProtect Balloon. The procedure seems a little more invasive than HA and PEG spacers, with the necessity of a short perineal incision. On the other hand, an advantage could be a better stability of this device, due to its inflatable balloon concept. The envelope prevents lateral and cranio-caudal dispersion of the product as observed sometimes with others spacers and could allow a more reproducible implantation.
Acknowledgements
S. Marchant for writing assistance.
List of investigators:
Principal Investigators:
David PASQUIER, Centre Oscar Lambret, Lille, France.
Franck DARLOY, Centre Léonard de Vinci, Dechy, France.
Alain TOLEDANO Clinique Hartmann, Levallois-Perret, France.
Denis FOSTER, Centre de Cancérologie Paris Nord, Sarcelles, France.
Co-coordonnator: Igor LATORZEFF.
Co-investigators:
François BONODEAU, Centre Oscar Lambret, Lille, France.
Xavier MIRABEL, Centre Oscar Lambret, Lille, France.
Gaelle JIMENEZ, Clinique Pasteur, Toulouse, France.
Louis GRAS, Centre Léonard de Vinci, Dechy, France.
Damien CARLIER, Centre Léonard de Vinci, Dechy, France.
Denis FOSTER, Clinique Hartmann, Levallois-Perret, France.
Hanah LAMALLEM-ALGHAZIRI, Clinique Hartmann, Levallois-Perret, France.
Marc BOLLET, Clinique Hartmann, Levallois-Perret, France.
Muriel BOTTI, Centre de Cancérologie Paris Nord, Sarcelles, France.
Cyril LAPORTE, Centre de Cancérologie Paris Nord, Sarcelles, France.
Guillaume SERGENT, Centre de Cancérologie Paris Nord, Sarcelles, France.
Funding
Centre Oscar Lambret and Aquilab.
Centre Oscar Lambret and Aquilab finance a part of biostatistics and administrative costs. Aquilab finances the whole Datamanagement part.
The funding body has no role in study design and collection, data analysis and interpretation, or manuscript writing.
Availability of data and materials
The data set used and/or analysed during the current study are available from the corresponding author on reasonable request. Not all data are obtained yet since the study is still ongoing.
Abbreviations
- BT
Brachytherapy
- CT
Computed tomography
- ECOG
Eastern Cooperative Oncology Group
- GETUG
French Group for the Study of Uro-Genital Tumors
- HA
Hyaluronic acid
- IGRT
Image guided radiotherapy
- IMRT
Intensity modulated radiation therapy
- IPSS
International Prostate Score Symptom
- MRC
Medical Research Council
- MRI
Magnetic resonance imaging
- MSKCC
Memorial Sloan-Kettering Cancer Center
- PEG
Polyethylen glycol
- PR
Prostate rectum
- PSA
Prostate-specific antigen
- RTP
Radiation treatment planning
- VKA
Vitamin K antagonist
Authors’ contributions
DP, TL, FB, EL, IL participated in the design of the study and ED designed the statistical analysis. IL, DP, EL conceived the study, and participated in its design and coordination. DP, IL, SM helped to draft the manuscript. All authors read and approved the final manuscript.
Ethics approval and consent to participate
The study has been submitted and approved by regulatory authorities (ANSM; date of approval: 13/10/2016) and ethics committee (Centre de Protection des Personnes; date of approval: 13/10/2016). The study opened in September 2015.
A written informed consent will be obtained from the study participants.
There is an agreement between each participating center and the Centre Oscar Lambret. Each protocol version is signed by the principal investigator. We have a copy of each signed document.
In France, according to the current law, a protocol can be subjected to any regional Ethics Committee, even if no hospital of this region takes part to the trial. The choice is made according to the workload of every committee. The opinion of this Ethics Committee applies to all the national centers.
Competing interests
A part of the trial cost is financially supported by Aquilab. The study protocol has undergone peer-review by Aquilab.
The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
David Pasquier, Phone: +33 3 20 29 59 11, Email: d-pasquier@o-lambret.fr.
Emilie Bogart, Email: e-bogart@o-lambret.fr.
François Bonodeau, Email: f-bonodeau@o-lambret.fr.
Thomas Lacornerie, Email: t-lacornerie@o-lambret.fr.
Eric Lartigau, Email: e-lartigau@o-lambret.fr.
Igor Latorzeff, Email: ilatorzeff@clinique-pasteur.com.
References
- 1.Zelefsky MJ, Pei X, Chou JF, Schechter M, Kollmeier M, Cox B, et al. Dose escalation for prostate cancer radiotherapy: predictors of long-term biochemical tumor control and distant metastases-free survival outcomes. Eur Urol. 2011;60(6):1133–1139. doi: 10.1016/j.eururo.2011.08.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Dearnaley DP, Khoo VS, Norman AR, Meyer L, Nahum A, Tait D, et al. Comparison of radiation side-effects of conformal and conventional radiotherapy in prostate cancer: a randomised trial. Lancet. 1999;353:267–272. doi: 10.1016/S0140-6736(98)05180-0. [DOI] [PubMed] [Google Scholar]
- 3.Beckendorf V, Guerif S, Le Prisé E, Cosset J-M, Bougnoux A, Chauvet B, et al. 70 Gy versus 80 Gy in localized prostate cancer: 5-year results of GETUG 06 randomized trial. Int J Radiat Oncol Biol Phys. 2011;80:1056–1063. doi: 10.1016/j.ijrobp.2010.03.049. [DOI] [PubMed] [Google Scholar]
- 4.Peeters ST, Heemsbergen WD, Koper PC, van Putten WL, Slot A, Dielwart MF, et al. Dose-response in radiotherapy for localized prostate cancer: results of the Dutch multicenter randomized phase III trial comparing 68 Gy of radiotherapy with 78 Gy. J Clin Oncol. 2006;24:1990–1996. doi: 10.1200/JCO.2005.05.2530. [DOI] [PubMed] [Google Scholar]
- 5.Cahlon O, Hunt M, Zelefsky MJ. Intensity-modulated radiation therapy: supportive data for prostate cancer. Semin Radiat Oncol. 2008;18:48–57. doi: 10.1016/j.semradonc.2007.09.007. [DOI] [PubMed] [Google Scholar]
- 6.Zelefsky MJ, Kollmeier M, Cox B, Fidaleo A, Sperling D, Pei X, et al. Improved clinical outcomes with high-dose image guided radiotherapy compared with non-IGRT for the treatment of clinically localized prostate cancer. Int J Radiat Oncol Biol Phys. 2012;84:125–129. doi: 10.1016/j.ijrobp.2011.11.047. [DOI] [PubMed] [Google Scholar]
- 7.Huang EH, Pollack A, Levy L, Starkschall G, Dong L, Rosen I, et al. Late rectal toxicity: dose-volume effects of conformal radiotherapy for prostate cancer. Int J Radiat Oncol Biol Phys. 2002;54:1314–1321. doi: 10.1016/S0360-3016(02)03742-2. [DOI] [PubMed] [Google Scholar]
- 8.Mok G, Benz E, Vallee J-P, Miralbell R, Zilli T. Optimization of radiation therapy techniques for prostate cancer with prostate-rectum spacers: a systematic review. Int J Radiat Oncol Biol Phys. 2014;90:278–288. doi: 10.1016/j.ijrobp.2014.06.044. [DOI] [PubMed] [Google Scholar]
- 9.Mariados N, Sylvester J, Shah D, Karsh L, Hudes R, Beyer D, et al. Hydrogel spacer prospective multicenter randomized controlled pivotal trial: dosimetric and clinical effects of perirectal spacer application in men undergoing prostate image guided intensity modulated radiation therapy. Int J Radiat Oncol Biol Phys. 2015;92:971–977. doi: 10.1016/j.ijrobp.2015.04.030. [DOI] [PubMed] [Google Scholar]
- 10.Hamstra DA, Mariados N, Sylvester J, Shah D, Karsh L, Hudes R, et al. Continued benefit to rectal separation for prostate RT: final results of a phase III trial. Int J Radiat Oncol Biol Phys. 2017;97:976–985. doi: 10.1016/j.ijrobp.2016.12.024. [DOI] [PubMed] [Google Scholar]
- 11.Chapet O, Udrescu C, Tanguy R, Ruffion A, Fenoglietto P, Sotton MP, et al. Dosimetric implications of an injection of hyaluronic acid for preserving the rectal wall in prostate stereotactic body radiation therapy. Int J Radiat Oncol Biol Phys. 2014;88:425–432. doi: 10.1016/j.ijrobp.2013.10.039. [DOI] [PubMed] [Google Scholar]
- 12.Ben-Yosef R, Paz A, Levy Y, Alani S, Muncher Y, Shohat S, et al. A novel device for protecting rectum during prostate cancer irradiation: in vivo data on a large mammal model. J Urol. 2009;181:1401–1406. doi: 10.1016/j.juro.2008.11.010. [DOI] [PubMed] [Google Scholar]
- 13.Gez E, Cytron S, Yosef RB, London D, Corn BW, Alani S, et al. Application of an interstitial and biodegradable balloon system for prostate-rectum separation during prostate cancer radiotherapy: a prospective multi-center study. Radiat Oncol. 2013;8:96. doi: 10.1186/1748-717X-8-96. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Melchert C, Gez E, Bohlen G, Scarzello G, Koziol I, Anscher M, et al. Interstitial biodegradable balloon for reduced rectal dose during prostate radiotherapy: results of a virtual planning investigation based on the pre- and post-implant imaging data of an international multicenter study. Radiother Oncol. 2013;106:210–214. doi: 10.1016/j.radonc.2013.01.007. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data set used and/or analysed during the current study are available from the corresponding author on reasonable request. Not all data are obtained yet since the study is still ongoing.