Initial clinical observations of intra- and interfractional motion variation in MR-guided lung SBRT

David H Thomas 1,, Anand Santhanam 2, Amar U Kishan 2, Minsong Cao 2, James Lamb 2, Yugang Min 2, Dylan O'Connell 2, Yingli Yang 2, Nzhde Agazaryan 2, Percy Lee 2, Daniel Low 2
PMCID: PMC5965474  PMID: 29166129

Abstract

Objective:

To evaluate variations in intra- and interfractional tumour motion, and the effect on internal target volume (ITV) contour accuracy, using deformable image registration of real-time two-dimensional-sagittal cine-mode MRI acquired during lung stereotactic body radiation therapy (SBRT) treatments.

Methods:

Five lung tumour patients underwent free-breathing SBRT treatments on the ViewRay system, with dose prescribed to a planning target volume (defined as a 3–6 mm expansion of the 4DCT-ITV). Sagittal slice cine-MR images (3.5 × 3.5 mm2 pixels) were acquired through the centre of the tumour at 4 frames per second throughout the treatments (3–4 fractions of 21–32 min). Tumour gross tumour volumes (GTVs) were contoured on the first frame of the MR cine and tracked for the first 20 min of each treatment using offline optical-flow based deformable registration implemented on a GPU cluster. A ground truth ITV (MR-ITV20 min) was formed by taking the union of tracked GTV contours. Pseudo-ITVs were generated from unions of the GTV contours tracked over 10 s segments of image data (MR-ITV10 s).

Results:

Differences were observed in the magnitude of median tumour displacement between days of treatments. MR-ITV10 s areas were as small as 46% of the MR-ITV20 min.

Conclusion:

An ITV offers a “snapshot” of breathing motion for the brief period of time the tumour is imaged on a specific day. Real-time MRI over prolonged periods of time and over multiple treatment fractions shows that ITV size varies. Further work is required to investigate the dosimetric effect of these results.

Advances in knowledge:

Five lung tumour patients underwent free-breathing MRI-guided SBRT treatments, and their tumours tracked using deformable registration of cine-mode MRI. The results indicate that variability of both intra- and interfractional breathing amplitude should be taken into account during planning of lung radiotherapy.

Introduction

Intra- and interfractional motion remains one of the biggest challenges in radiation therapy. In thoracic and abdominal sites, four-dimensional CT (4DCT) techniques that allow both the range of motion as well as the mean tumour position to be determined are regularly employed. Using respiratory gating, 4DCT methods assume that respiratory motion is approximately periodic, and that the depth of breathing can be assumed to remain constant from cycle-to-cycle with no interfractional variation. Respiratory-gated 4DCT techniques also enable the calculation of an averaged CT image or maximum intensity projection (MIP).1 Often, the internal target volume (ITV) is the union of all tumour positions during the 4DCT.2 For many patients, breathing patterns are stable and regular, with naturally repeated cycles that are reasonably periodic, but various studies have shown that irregularities in breathing patterns can lead to the ITV underestimating the true range of motion.3, 4

The introduction of MRI-guided radiotherapy systems allows real-time soft-tissue imaging during delivery of radiotherapy. The only clinical system currently in use is the ViewRay® MRIdian, a commercially available MRI-guided Co-60 radiotherapy system that uses three Co-60 sources combined with a 0.35 T MRI system for patient setup and real-time tumour motion tracking during treatment. Real-time cine imaging is currently limited to two-dimensional (2D) sagittal planes.

MRI offers specific advantages for the real-time delineation of certain tumours owing to improved soft-tissue contrast, but image quality is generally degraded by motion.5 Various studies have employed single-slice 2D dynamic MRI to measure respiratory motion. Several studies,3,613 have all employed fast-sequence cine-mode MRI of free breathing patients to investigate the reproducibility of respiratory motion, but limitations of these studies include either short imaging sessions or only a single study session. Various imaging parameters and study size/durations are summarized in Table 1.

Table 1.

Literature summary of studies evaluating MRI-based lung tumour motion

Author Year Number of subjects Frames per second Session time (s) Number of sessions
Koch14 2004 7 (3 healthy volunteers, 4 lung cancer patients) 2.2 45 2
Lui14 2004 7 (3 healthy volunteers, 4 lung cancer patients) 2.2 45 2
Plathow15 2004 20 lung patients 0.3 30 1
Blackall7 2006 10 volunteers, 5 lung cancer patients. 3 Several minutes 1
Cai16 2007 3 healthy volunteers 10 300 2
Cai16 2007 14 (8 healthy volunteers, 6 lung cancer patients) 3 300 1
Cai12 2008 17 (9 healthy volunteers, 8 lung cancer patients) 10 300 2
Cai13 2008 15 (7 healthy volunteers, 8 lung cancer patients) 3 300 1
Cai3 2010 18 (10 healthy volunteers, 8 lung cancer patients) 3 300 1
Seregni8 2016 6 lung cancer patients 4 256 1
Thomasa 2016 5 lung cancer patients 4 1200 3–4

aThe current study is indicated.

This study reports tumour tracking results from five lung tumour patients who underwent free-breathing stereotactic body radiation therapy (SBRT) treatment on the ViewRay MRIdian at the UCLA Department of Radiation Oncology. Lung tumours were tracked for 20 mins per fraction for 3 or 4 fractions of treatment, and the inter- and intrafractional motion analysed with regards to the variation in ITV margin.

Methods and materials

Five patients (three male, two female) were treated under free-breathing with lung SBRT on the ViewRay system. Patient ages were 59, 68, 69, 72, 73. Simulation took place on both CT, including 4DCT and MRI (ViewRay), on the same day. Free-breathing MRI scans of 172 s duration were acquired (Siemens TrueFISP; repetition time = 3.37 ms, echo time = 1.45 ms) with 1.5 × 1.5 × 1.5 mm3 resolution. Table 2 shows the tumour locations. Gross tumour volumes (GTVs) were contoured by a physician, and ITV’s were contoured on both 4DCT and MRI. It was assumed that the 172 s free-breathing acquisitions captured sufficient motion, i.e. similar to the 4DCT, to be considered an ITV.

Table 2.

Tumour size (GTV) and motion (ITV), and associated motion of GTV centroid as measured during simulation ; data from 4DCT and free-breathin g (172 s) MRI (bottom row)

Patient 1 Patient 2 Patient 3 Patient 4 Patient 5
Simulation Contour size (mm) Motion (mm) Contour size (mm) Motion (mm) Contour size (mm) Motion (mm) Contour size (mm) Motion (mm) Contour size (mm) Motion (mm)
CT GTVa 28.7 43.2 27 14.6 14.8
4DCT ITVa 41.7 13 49.8 6.6 31.6 4.6 21.1 6.5 21.2 6.4
FB MRI ITVa 38.9 10.2 59.3 16.1 35.6 8.6 21.9 7.3 23.7 8.9
Location Right lower lung Right lower lung Right upper lung Right medial lung Right medial lung
aMeasured along dimension of maximum motion, through the central slice
Cine mode Mean motion (mm) σ (mm) Mean motion (mm) σ (mm) Mean motion (mm) σ (mm) Mean motion (mm) σ (mm) Mean motion (mm) σ (mm)
Fraction 1 11.1 2.0 9.4 2.0 8.4 1.7 1.9 1.2
Fraction 2 13.7 1.3 11.0 2.2 3.0 1.2 3.5 0.9
Fraction 3 18.0 3.0 6.0 0.8 4.8 1.5 2.0 0.4
Fraction 4 15.2 3.3 5.1 1.1

Cine mode measured motion (mean and standard deviation) of GTV centroid.

MRI-guided treatments were performed with real-time 2D cine-mode MR imaging performed using the 0.35 T whole body MRI scanner included with the ViewRay system. A balanced steady-state free precession sequence (Siemens TrueFISP; repetition time = 2.1 ms, echo time = 0.91 ms) was used to acquire images of a single 5 mm thick sagittal slice through the centre of the tumour at 4 fps, with a 35 × 35 cm2 field of view with 3.5 × 3.5 mm2 spatial resolution. Five patients received three to four fractions of treatment with cine-mode imaging of duration greater than 20 min acquired each fraction. Patients with central lung tumours (N = 2) received 50 Gy in four fractions, and patients with peripheral lung tumours (N = 3) received 54 Gy in three fractions. Cine-mode images were exported from the ViewRay system after each treatment, and converted to DICOM (Digital Imaging and Communications in Medicine) format using Matlab® R2016 (The MathWorks, Inc., Natick, MA) and subsequently exported to MIM v. 6 (MIM Software, Cleveland, OH) for contouring. The GTV was contoured on the first cine frame of each fraction by a physician. Images were then exported to a  graphics processing unit (GPU) cluster for deformable image registration (DIR) in order to propagate the GTV onto each frame of the cine. A well-validated anatomy-based multilevel multiresolution optical flow and thin-plate splines17, 18 method was used. This method characterizes a three-dimensional (3D) object in 2D images as features and correlates the features from one image to another in such a way that the cumulative intensity displacement gradient is minimized. The optical flow registration method starts with an initial estimated displacement (typically set to zero for each voxel) and then refines it according to partial derivative of voxel intensity and weighted sum of displacement from all nearby voxels. Displacement estimation is iteratively updated until the displacement in the whole field converges.19 One advantage of this method is that it tracks the overall motion of the field even if some feature points cannot be tracked, although this is not exclusive to this method. Each frame of the cine was registered to the first frame, and the subsequent deformation vectors used to warp the GTV contour. The accuracy of the DIR generated contours was checked visual in all images, and spot-checked in comparison to manually generated contours in 10 frames per fraction. For each time point t, the centroid of the GTV was recorded, as well as a cumulative ITV generated from the union of all GTVs from times time = 0 to time = t. Figure 1 shows a final cine frame with the GTV tracked across four fractions of treatment for Patient 1, with the ITV generated from the union of all GTV contours tracked for 20 min in each case (MR-ITV20 min).

Figure 1.

Figure 1.

Example sagittal 2D cine-mode imaging during four fractions of treatment for Patient 1. GTV tracking from the off-line deformable image registration is shown for each fraction (inner contour), with the cumulative ITV for the entire fraction also plotted (outer contour ). 2D, two-dimensional; GTV, gross tumour volume; ITV, internal target volume.

A breathing trace for each fraction was generated by calculating the vector displacement length of the GTV centroid from the 5th percentile of motion in the anterioposterior (AP) and craniocaudal (CC) directions. The time points at which local peaks of inhalation occurred were extracted from a version of the trace smoothed using a median filter, and the amplitude of each peak subsequently identified from the original data. The mean breathing period was determined from the mean duration between peaks, and any peak occurring less than an arbitrary threshold of 10% of the mean period, chosen from visual inspection of the data, was determined to be noise in the data and rejected from the analysis. The variation in the magnitude of tumour centroid displacement was analysed using the boxplot function in Matlab Statistics Toolbox (R2016).

A simulation was performed to determine the variation in the dimensions of ITV contours as although the patient had been scanned using a commercial low-pitch helical CT protocol and retrospective phase-based sorting technique. For this simulation, the scanner parameters were taken from the Phillips Big Bore 4DCT protocol: pitch of 0.06, rotation time of 0.5 s and detector width of 2.4 cm. This translated to an imaging time of approximately 10 s for stationary tissues. For each 1 s of the GTV tracking data per fraction, an ITV was generated from the union of the corresponding GTV contours from the following 10 s of image data (MR-ITV10 s). The variation in the MR-ITV10 s contours (a total of 1190 per fraction) was compared with the ITV generated from the entire treatment MIP (MR-ITV20 min).

Results

The DIR-generated contours were found to be visually acceptable in all fractions and within 7 mm, measured along the direction of maximum motion, of manually generated contours (10 frames per fraction). The motion of the GTV measured during CT- and MR-simulation is listed in Table 2 (top row, motion in red). Figure 2 shows the location of the GTV centroid for the first 500 s of two fractions for Patient 1, plotted as AP vs CC directions normalized to the fifth percentile of each axis. The motion of the GTV centroid shows a distinct pattern in each fraction. Fraction 3 showed the largest motion (not shown for figure clarity), with mean GTV centroid displacement of 18.0 ± 3.0 mm (± one standard deviation). Fraction 1 shows the smallest mean centroid displacement (11.1 ± 2.0 mm) and Fraction 2 shows an increase in the amount of hysteresis. Figure 2 shows an example of 1 min of breathing data from the middle of Fraction 1 for the same patient, showing the detected inhalation peaks (red circles) overlaid on the motion amplitude (black line). Table 2 (bottom row) shows the mean (red) and standard deviation (black) of the GTV centroid motion for each fraction and patient. The agreement between the 4DCT, 172 s MR and cine mode MR data sets vary.

Figure 2.

Figure 2.

GTV tracking data for Patient 1. (a) Shows AP vs CC motion of the GTV centroid for Patient 1, Fractions 1–4 (first 500 s of each fraction shown), showing interfractional variation in the amplitude of breathing and amount of hysteresis. (b) Shows breathing amplitude as calculated from the vector length displacement of GTV centroid relative to fifth percentile reference position. One minute of data is shown here with the detected inhalation peaks indicated with red circles. (

Figure 3 shows the intra- and interfractional variation in the displacement magnitude of the GTV centroid. Individual inhalations are shown (grey dots), with the number of inhalations (N) measured per fraction listed on the x-axis. Boxplots per fraction showing the median (black line), interquartile range (coloured boxes) and min–max range (whiskers) of the displacements are displayed over the data. The interval endpoints of the medians are displayed as notches in the boxes, representing the 95% confidence interval on the median. Figure 4 shows an example of the ITV generation (Patient 3, Fraction 1 shown, with GTV area = 12.3 mm2). Figure 4 shows MIP image from a 20-min fraction of SBRT, and the corresponding ITV contour, resulting in an ITV area = 26.6 mm2. Figure 4 shows the “best-case” scenario (i.e. closest to the entire treatment) for a MR-ITV10 s, generated from 10 s snapshot which includes a deep inhalation, and generates an ITV with ITV area = 26.2 mm2. Figure 4 shows the “worst-case” scenario, where no motion is captured during the 10 s snapshot, and the ITV is similar in size to the GTV (ITV area = 12.6 mm2). ITV area was chosen as a metric, as opposed to maximum diameter, as area will reflect changes to both depth of breathing and any changes in hysteresis, as observed in Figure 2.

Figure 3.

Figure 3.

Variation in amplitude of breathing, showing amplitude of individual inhalations (grey dots) and boxplots per fraction showing the median (black line), interquartile range (coloured boxes) and min–max range (whiskers) of the amplitudes. The interval endpoints of the medians are displayed as notches in the boxes, representing the 95% confidence interval on the median.

It is worth noting that MR-ITV20 min includes all time points, and as such may be susceptible to outliers, such as involuntary patient motions (coughs etc.) which would have potentially little dosimetric consequence. Cumulative histograms showing the distribution of ITV area with respect to the entire treatment MIP for each fraction, each patient, are shown in Figure 4. MR-ITV10 s with area as small as 40% of the MR-ITV20 min are observed. The median (50th percentile) values range from 45 to 90%.

Figure 4.

Figure 4.

(a) Maximum intensity projection image of the entire treatment–from a single fraction– showing the MR-ITV20 min contour, and of the corresponding “best-case” and “worst case” MR-ITV10 s. (d) Cumulative distributions for the all possible snapshot MR-ITV10 s generated per fraction.

Discussion

This study presents analyses of a large amount of tumour tracking data, and is the first study of its kind to analyse patients in the treatment position during treatment over multiple fractions. Many of the previous studies were performed with arms-down, and all were done without the immobilization devices used for treatment being present. While many studies investigating breathing motion have studied the effect of intrafractional breathing variation, the data presented here show that interfractional variation in breathing amplitude may be larger than previously considered.

Only patients with tumours with sufficient contrast for tumour tracking on MRI were chosen for this study, which limited the size of the patient cohort. The choice of patient inclusion in the study was made by physician, based on the ability to easily identify the GTV on the simulation images. Comparison between contours drawn on different modalities is problematic, owing to inherent differences in contrast and acquisition times, especially when acquired during free breathing. The comparison between the 4DCT and MR (both the high resolution 172 s acquisition and the cine-mode data) ITV dimensions shown in Table 2 has a large amount of intrinsic uncertainty. Of note is that even with a 172 s acquisition acquired at simulation, the ITV can be smaller than that acquired during subsequent day’s treatment (e.g. Patient 1, Fraction 3). Further work is merited to investigate if patient specific margins can be calculated from a single session of breathing data which will account for interfraction motion changes. For more advanced approaches to treatment delivery, such as tumour tracking with prediction modelling, we show here that a single 4DCT is not sufficient to estimate the variation in tumour motion throughout the course of treatment, and more robust motion modelling is warranted. While the acquisition of multiple 4DCTs20 has the ability to offer increased information to validate the ITV, the unwanted increase in patient imaging dose typically prevents such studies.

As improvements to hardware and acquisition sequences continue to increase image quality and temporal resolution, MRI will offer more accurate measurement of breathing variability owing to its non-ionizing nature and the ability to image for an extended period of time.21 Currently the ViewRay is the only clinically available MR-guided system, and uses a gated approach to irradiate the tumour at a specific breathing phase.22 These initial five patients were treated free breathing; we have since then transitioned to a breath-hold (deep inspiratory) treatment protocol with only a 3–5 mm planning target volume margin added to the GTV acquired at breathold (no ITV). Gated treatments suffer from lower efficiency; various publications suggest improved accuracy and sparing of normal tissues can be achieved by adjusting the beam delivery to track the tumour,2327 as well as shorter treatment times. Dynamic MRI currently has poor spatial resolution compared with 4DCT, and suffers from low lung signal intensity and magnetic susceptibility effects.28, 29 The limitations of real-time MRI, as well as the unavoidable latencies of tracking, such as data processing time and hardware reaction time (up to 500 ms in the case of the ViewRay system)22 suggest that motion modelling30, 31 and breathing prediction methods3234 will still have an important role to play in MRI-guided treatments, as well as X-ray-guided treatments. The data presented here show that interfractional motion is an important consideration for such approaches.

One obvious limitation to this study is the small patient sample size, which limits the power to detect significant differences in breathing variation between tumours of different size and location. However, the large number of breaths and days tracked in each patient gives large statistical power to the analysis, and all five patients in the cohort demonstrated significant variation in between fractions. Similar simulations of the variability of ITV accuracy have been presented in the literature, e.g. Dou et al35 have performed a statistical study of one-dimensional simulations of 4DCT tumour motion measurement error using 50 patient’s external breathing surrogates. They show that 61% of patients exhibit a 5% chance of tumour motion underestimation, depending on the snapshot of breathing taken during 4DCT, but data are only presented for a single session. The data presented here represent a continuation of that approach, and although limited in the number of patients, has the advantage of tumour tracking over multiple days and not relying on surrogate data.

A proprietary free-form deformable registration (similar to the work of Lu et al36) is available on the ViewRay clinical treatment delivery system which allows a user specified contour to be tracked in real time, but this real time tracking data are currently not available to be exported from the system for further analysis, so was not used in this study. In addition, tracking low contrast tumours can be difficult using real time approaches owing to limited signal-to-noise in the low field images. Various approaches have been documented in the literature for improved real-time tracking, including snakes-based tracking, template matching and manifold learning,3741 including a recent approach using scale-invariant feature transform descriptors applied directly to ViewRay cine images.41 As real-time tracking is not required for this study, we used an offline approach based on anatomy-based multilevel multiresolution optical flow analysis and thin-plate splines for registering the frames. Although the frame-to-frame registration is approximately 12 times slower than real time, even when running on a GPU cluster, this approach provides improved tracking of tumours that lack well-defined boundaries in the presence of limited signal-to-noise owing to the low field strength. Target registration error was typically on the order of one voxel.

The amplitude of motion in this study is measured as the vector length of the distance between the current GTV centroid and its fifth-AP and CC percentile position. This approach effectively removes any effect of a baseline drift from the analysis. As image guidance (either CBCT for patient positioning or continual MRI for MR-guided treatments) is the clinical standard, we chose to remove the baseline shifts from the analysis in this study, and focus on the accuracy of the area of the ITV contour generated for planning. Corrective action such as patient repositioning either before or during treatment to account for a baseline shift is possible as long as the treatment is image-guided. However, replanning to account for an inaccurate ITV is not achievable without real-time adaptive treatment planning.

The dosimetric effect of under covering the true ITV is not considered here and should be investigated further. Another limitation of this study was the acquisition of 2D data, which prohibited tracking of motion in 3D. We assume here that out of plane motion is small in comparison to the sagittal plane. Fast 3D sequences to allow true 4D-MRI imaging is an active area of development, but the resultant images have poor resolution in the axial slice direction, on the order of 10 mm,14 often suffer from low signal-to-noise ratio, and are not yet clinically available.

A recent retrospective study to evaluate whether gated radiotherapy, whereby radiation is only delivered when the tumour is within the GTV contour, provides dosimetric improvements to organs-at-risk in lung SBRT showed that only patient’s with the largest tumour motion (>2 cm) would receive a clinically relevant benefit.42 Given the success of current ITV-based lung SBRT treatments in achieving local control without significant toxicity,4345 current ITV approaches may be adequate, but still leave room for improvement in cases for which minimizing treatment margins is critical. MR’s improved ability to accurately delineate organs at risk, combined with gated- or tracked-delivery of lung SBRT, may lead to the expansion of SBRT to more central lung tumours owing to the associated reduction in toxicity in normal tissue.

Conclusions

The results presented here indicate that variability of both intra- and interfractional tumour motion and variation in the amount of hysteresis should be taken into account during both motion modelling of breathing motion and planning of lung radiotherapy.

Contributor Information

David H Thomas, Email: david.h.thomas@ucdenver.edu.

Anand Santhanam, Email: ASanthanam@mednet.ucla.edu.

Amar U Kishan, Email: aukishan@mednet.ucla.edu.

Minsong Cao, Email: MinsongCao@mednet.ucla.edu.

James Lamb, Email: jlamb@mednet.ucla.edu.

Yugang Min, Email: ymin@mednet.ucla.edu.

Dylan O'Connell, Email: DOConnell@mednet.ucla.edu.

Yingli Yang, Email: yyang@mednet.ucla.edu.

Nzhde Agazaryan, Email: nagazaryan@mednet.ucla.edu.

Percy Lee, Email: percylee@mednet.ucla.edu.

Daniel Low, Email: dlow@mednet.ucla.edu.