Abstract
Estimating the depth of anaesthesia (DoA) in operations has always been a challenging issue due to the underlying complexity of the brain mechanisms. Electroencephalogram (EEG) signals are undoubtedly the most widely used signals for measuring DoA. In this paper, a novel EEG-based index is proposed to evaluate DoA for 24 patients receiving general anaesthesia with different levels of unconsciousness. Sample Entropy (SampEn) algorithm was utilised in order to acquire the chaotic features of the signals. After calculating the SampEn from the EEG signals, Random Forest was utilised for developing learning regression models with Bispectral index (BIS) as the target. Correlation coefficient, mean absolute error, and area under the curve (AUC) were used to verify the perioperative performance of the proposed method. Validation comparisons with typical nonstationary signal analysis methods (i.e., recurrence analysis and permutation entropy) and regression methods (i.e., neural network and support vector machine) were conducted. To further verify the accuracy and validity of the proposed methodology, the data is divided into four unconsciousness-level groups on the basis of BIS levels. Subsequently, analysis of variance (ANOVA) was applied to the corresponding index (i.e., regression output). Results indicate that the correlation coefficient improved to 0.72 ± 0.09 after filtering and to 0.90 ± 0.05 after regression from the initial values of 0.51 ± 0.17. Similarly, the final mean absolute error dramatically declined to 5.22 ± 2.12. In addition, the ultimate AUC increased to 0.98 ± 0.02, and the ANOVA analysis indicates that each of the four groups of different anaesthetic levels demonstrated significant difference from the nearest levels. Furthermore, the Random Forest output was extensively linear in relation to BIS, thus with better DoA prediction accuracy. In conclusion, the proposed method provides a concrete basis for monitoring patients’ anaesthetic level during surgeries.
Keywords: Electroencephalogram, Sample entropy, ANOVA, Depth of anaesthesia, Random Forest
Introduction
General anaesthesia (GA) is a temporary drug-induced unconsciousness state, which is reversible under human manipulation (Purdon et al., 2015). Modern GA is widely used in clinical operations because it guarantees safe surgical operations for millions of patients every year (Vutskits & Xie, 2016). However, numerous published articles have provided strong evidence of postoperative side effects, such as delirium (Lepouse et al., 2006) and cognitive dysfunction (Monk et al., 2008). Limited solutions are available to deal with these problems effectively because the underlying mechanism of anaesthesia is still not entirely understood (Uhrig, Dehaene & Jarraya, 2014). Therefore, optimal titration of anaesthetic dose is crucial to achieve the ideally effective analgesia, unconsciousness, and immobility for reducing the potential negative influences of either underdosing or overdosing. Thus, a precise evaluation of the depth of anaesthesia (DoA) is crucial due to the emerging critical requirements of patients’ surgery safety and experience.
Anaesthetic agents mainly rely on enhancing the activity of inhibitory cells or suppressing the activity of excitatory cells. Gamma-aminobutyric acid (GABA) and N-methyl-D-aspartate (NMDA), pervasive receptors, are usually induced by drugs including propofol, sevoflurane, and ketamine (Brown, Lydic & Schiff, 2010; Uhrig, Dehaene & Jarraya, 2014). In general, some disruption of neural communications occurs during the loss of consciousness (LoC). Anaesthetics alter the brain’s normal information processing. Thus, discovery of how to assess the degree of the brain state might reveal a feasible method for determining DoA; researchers have regarded this line of inquiry as promising (Shalbaf et al., 2013). Nevertheless, the monitoring of the brain state is not routinely preferred in practical anaesthesia care because markers that reliably reflect consciousness level variations under GA have yet to be identified (Palanca, Mashour & Avidan, 2009). The current standards for indirect DoA measurements, by which anaesthesiologists judge whether patients are adequately anaesthetised, include: (1) observing changes in the heart rate, blood pressure, and muscle tone and (2) monitoring the drug pharmacokinetics, pharmacodynamics, and the level of exhaled anaesthetic gas of the inhaled anaesthetics (Purdon et al., 2013). Anaesthesiologists primarily rely on autonomic and behavioral responses to optimize levels of anaesthesia and provide appropriate analgesia. However, individual anaesthesiologists apply these traditional methods idiosyncratically. Moreover, analysis of these signs can be difficult and unreliable due to the use of some medications such as muscle relaxants (Shalbaf et al., 2013). Accumulating evidence shows that electrical potential in the frontal cortex, namely electroencephalogram (EEG), is highly associated with GA (Kiersey, Bickford & Faulcone Jr, 1951; Martin, Faulconer & Bickford, 1959). Brain information processing is based on transmission and reception of the spike or action potential by neurons, which is one of the fundamental mechanisms underlying central nervous system interactions (functions) (Kandel et al., 2000). Information processing is altered by drugs (Vizuete et al., 2014). They accordingly change neural behaviour, such as the rate of firing, which can definitely be reflected by the neurons’ electrical potential. However, subcortical areas produce weak and difficult-to-detect electrical potentials because the electric field strength decreases as the distance of a potential from its source increases (Hämäläinen et al., 1993). Scalp EEG patterns reflect the states of both cortical and subcortical structures due to their strong connections (Ching et al., 2010). Thus, EEG can be used to explore the interactions between different structures under anaesthetics. EEG behaviour has been proved to be strongly correlated with different DoA levels (Brown, Lydic & Schiff, 2010; Purdon et al., 2013). One typical phenomenon is the burst-suppression pattern. Therefore, the EEG is the most pervasive noninvasive signal that reflects the neural communications and states. Therefore, numerous EEG-based DoA-level-monitoring machines have been produced and are described in following passages.
Fortunately, various electrophysiological measurements of brain activity have been investigated to quantify DoA (McKeever, Johnston & Davidson, 2014; Poorun et al., 2016). Spectrum-power analysis involves the evaluation of specific frequency component decompositions, compressed spectral arrays, two-dimensional plots of the spectrogram, and so on (Billard et al., 1997). Moreover, several spectral indices have been proposed, such as the 95% spectral edge frequency (SEF) (Katoh, Suzuki & Ikeda, 1998), median frequency (MF) and spectral entropy (SE) (Höcker et al., 2010). Among them, the SE has already been applied to the commercial monitor Datex–Ohmeda S/5 (GE Healthcare, Helsinki, Finland). However, these indices are used for studying spectral characteristics which are based on traditional fast Fourier transform (FFT) techniques. Raw EEG data spectrogram is often investigated to identify the difference between the awake state and unconsciousness state (Akeju et al., 2014; Purdon et al., 2013) for overcoming some drawbacks of index currently applied though they just compared the unique features of spectrogram under different conditions. However, it is difficult to practically distinguish the states. Most frequency analysis methods, which are based on Fourier theory, assume the data is stationary in a short period. Because an EEG presents non-stationary and nonlinear features in human biological system, those frequency based methods may ignore valuable information. Consequently, it is reasonable to apply the nonlinear dynamics and information theory to EEG for the prediction of DoA. Besides, various other studies only focus on the segment comparisons (Akeju et al., 2014). Therefore, it is crucial to conduct the whole perioperative studies. Moreover, few researchers use other physiological methods such as electrocardiography (ECG) (Liu et al., 2017b) or blood pressure combined with the heart rate (Shalbaf, Behnam & Moghadam, 2015). However, it is obvious that anaesthetics affect the brain more directly, which results in the variations of EEG signals.
In the past decade, the concept of entropy has been widely utilised in medical and neurological techniques such as the approximate entropy (ApEn) (Bruhn et al., 2001) and Shannon entropy (Bruhn et al., 2001; Shannon, 2001). The concept of entropy is derived from the time series. Sample Entropy (SampEn), an improved form of the previous entropy methods (Richman & Moorman, 2000), estimates the probability that data series are closely correlated in a dataset within a given tolerance. SampEn calculates the irregularity of a time series and reflects the randomness and complexity in time domain. In general, regular systems have lower values of entropy and vice versa. In particular, SampEn applies to non-stationary data and is more noise resistant, which overcomes the drawbacks of many spectral methods that are mostly based on linear Fourier transform. Therefore, SampEn has been applied to many biomedical dynamics such as heart rate variability and EEG analysis (Chu et al., 2017; Huang et al., 2013; Shalbaf et al., 2012). Previous research efforts, including ours, ignored the variability in different anaesthetic levels to some extent. Unconsciousness is not only an unresponsive state but also is notably associated with the concentration level of anaesthetic agents, which affect both neuronal activity and brain networks. Moreover, any disruption in the neuronal activity and brain networks may lead to conditions such as burst suppression or postoperative cognition dysfunction (Purdon et al., 2015). These adverse effects may be avoided if DoA can be precisely controlled using an appropriate drug dosage. Therefore, further studies should be conducted for assessing DoA at different levels.
Currently, many machine learning methods receive much attention for classification and regression in data mining, such as Support Vector Machine (SVM), Artificial Neural Network (ANN) and Random Forest. SVM, a type of supervised learning method, depends on one or multiple hyperplanes to be used in classification and regression (Drucker et al., 1997). The larger the margin is from the training data to the hyperplane, the better performance is. ANN is an advanced modelling tool in statistics, machine learning, and cognitive science (Alpaydin, 2014; Kriegeskorte, 2015). Neural network models are widely applied to theoretical neuroscience and computational neuroscience. In general ANN implementations, artificial neurons and connections typically adapt their connection weights on the basis of a learning procedure that maps input signals to outputs. An ANN is organized in layers that exhibit different types of transformations on their inputs. Random Forest is an ensemble learning method and can be considered as an ensemble of decision trees used for classification, regression and other tasks (Liaw & Wiener, 2002). By a random selection of features at each tree node, the correlation between the trees in the forest can be decreased thus decreasing the forest error rate and reducing over fitting.
In this paper, EEG analysis was conducted to estimate DoA using Bispectral index (BIS) as the reference value (Aspect Medical Systems, Newton, MA, USA). BIS is the most commonly used commercial index, it is considered as the gold standard to evaluate our method performance due to its relatively high accuracy in the GABA receptor anaesthetic situation and the BIS has been approved by the US Food and Drug Administration (FDA). The range of the recorded BIS is from 0 (EEG suppression state) to 100 (awake state). The BIS index has significant advantages such as comprehensive clinical validation (Johansen & Sebel, 2000; Luginbühl et al., 2003). However, the algorithm remains unreleased and has some specific weak points which are comprehensively discussed in section ‘Discussion’, thus posing the need for a superior open source methodology. In this study, the focus is on patients receiving GABA-receptor dependent anaesthetic agents by employing BIS as the gold standard for preliminary investigation due to its precision when using specific anaesthetic drugs. In total, 24 datasets of EEG data were analyzed to evaluate DoA. First, Multivariate Empirical Mode Decomposition (MEMD), a non-linear, non-stationary method, was applied to filter the noise. Subsequently, SampEn features were extracted from specific combinations of Intrinsic Mode Function (IMF), generated from MEMD. To analyse the performance of SampEn, it was compared with recurrence quantification analysis (RQA) measurement and permutation entropy (PEn). Next, three regression models were chosen to train the features for regression for optimising the performance. Finally, the capability of the models was evaluated by comparing the outputs of these models with the BIS gold standard. In order to analyze this method capability in more detail, analysis of variance (ANOVA) analysis was undertaken for the data pairs of four subgroups.
Materials and Methods
Subjects
Ethical approval was granted by the Research Ethics Committee of the National Taiwan University Hospital (NTUH) (No: 201302078RINC). Written informed consent forms were obtained from all the patients who were involved. In this study, patients who received regular surgery under GA were recruited from NTUH (i.e., no patients underwent high risk operations related to brain, heart, lung, etc.). Those who had alcohol or smoking habits, or had other medical illness that affected the data recording, were excluded. Finally, 24 patients (ASA I or II, age (yr): 44.5 ± 12.9, height (cm): 164.2 ± 7.1, weight (kg): 63.4 ± 14.8, and BMI (kg/m2): 23.4 ± 4.2, gender: 14 females/10 males) were eligible for this study. Operations last 126.4 ± 72.9 min from anaesthetic medical recording forms.
Anaesthetic procedure manipulation
The anaesthetic procedures, as shown in Fig. 1, can be illustrated as follows: first, patients’ clinical information, including height, weight, age, gender, and type of operation performed, were provided to anaesthesiologists for preparing clinical assessment of the anaesthetic plan. Prior to the beginning of this study, patients were required to eat nothing for at least 8 h. All the mandatory procedures were conducted such as routine and emergency drug check. Then each patient received an appropriate volume of anaesthetic agents for the regular operations. In this study, intravenous propofol was used to induce unconsciousness along with some other compounds such as muscle relaxants and analgesic. Propofol is a GABA-receptor dependent drug, which provides a rapid anaesthetic effect within a few minutes. When patients lost consciousness (i.e., no response to verbal commands or stimulus), the next stage was attained. For the maintenance period, either propofol was titrated for short-term surgeries or gas anaesthetics (desflurane or sevoflurane) together with air and oxygen in order to maintain unconsciousness. Practically, the concentration levels were adjusted within 1–1.5 minimal alveolar concentration. The anaesthetic gases were delivered by a medical mask. When approaching the end of a surgery, additional drugs are administrated to comfort the patients like vagostin, atropine, morphine, etc. All the anaesthetic agents’ information is detailed in Table 1. To maximize the universality of the data, drugs were not specified. Thus, patients were often injected with a combination of drugs used in ordinary operations. General anaesthesia was performed safely through all stages by monitoring the physiological signals such as EEG, ECG, photoplethysmography (PPG) and by examining the intermittent vital signs of blood pressure (BP), heart rate (HR), pulse rate (PR), pulse oximeter oxygen saturation (SpO2), and so on. These measurements were shown on the physiological monitor MP60 (Intellivue; Philips, Foster City, CA, USA). If any observed signal changed irregularly, doctors will adjusted the intraoperative standard anaesthesia machine accordingly. The surgical procedure and anaesthetic management details were acquired from the anaesthetic recording sheets by the hospital staff.
Table 1. Properties of drugs administrated in the anaesthetic treatment procedures.
Anaesthetic management | Mean (std) |
---|---|
Induction combinations | |
Propofol (mg) | 119.6(25.1), n = 23 |
Fentanyl (mg) | 116.7(20.9), n = 24 |
Atropine (mg) | 0.6(0.19), n = 9 |
Nimbex (mg) | 9.6(2.1), n = 10 |
Xylocaine (mg) | 42.0(12.3), n = 20 |
Maintenance | |
Sevoflurane (%, min) | 2∼3, 146.8(75.3), n = 6 |
Desflurane (%, min) | 6∼9, 123.8(80.4), n = 16 |
Propofol ( µg, min) | 1.5∼4, (126,100)a, n = 2 |
Additional drugs administered close to the end | |
Morphine (mg) | 6.7(2.9), n = 3 |
Ketamine (mg) | 18.3(7.3), n = 3 |
Atropine (mg) | 1(0), n = 21 |
Vagostin (mg) | 2.4(0.2), n = 23 |
Notes.
Only two cases maintained with propofol, therefore, we showed both dosage values respectively.
Data recording
Prior to EEG recording, a conductive paste was used to improve the contact between the frontal scalp skin and EEG BIS™ Quatro Sensor (Aspect Medical Systems, Newton, MA, US) in order to reduce the impedance under 5 kΩ. The EEG sensor was connected to the MP60 machine via the standard factory connection cable. Raw continuous EEG waveform data; including the aforementioned discrete routine vital signals; were recorded on a laptop through the serial RS-232 port utilising software developed in Borland C++ Builder 6 (C++ version 6; Borland Company, Austin, TX, USA). The EEG data were acquired at a sampling rate of 128 Hz. The intermittent BIS value was obtained every 5 s. Other intermittent vital signs such as, HR, PR, BP, and SPO2 were also recorded every 5 s using specific sensors with MP60 (e.g., pulse oximetry, ECG sensor and blood pressure cuff). The recording began 5 min ahead of the onset of induction, at which time patients were fully awake, and it terminated when patients began to respond to doctors either by voice or movement.
Data preprocessing
All the clinical administration and physiological data were sorted in order. Subsequently, tasks such as conversion of the EEG data format and labelling of the event time points were conducted. All cases’ data were then visually inspected roughly to remove the specific segments of artefacts resulting in waveform saturation (e.g., electrical artefacts caused by medical equipment or body movement), thus considerably reducing the outlier point segments. A notch filter was used to remove the 60-Hz line noise. Subsequently, a 5 s data epoch was used to conduct the MEMD analysis, and a 30-s segment of the reconstructed signal which was used for the SampEn calculation. The window step size was set to 5 s to be consistent with the BIS value for further comparisons.
Data analysis
Filtering by MEMD
EEG signal can be easily contaminated by artefacts such as noise from electromyography (EMG), electrooculography (EOG), and electric circuits (Huang et al., 2013). Thus, it is necessary to filter the signals to reduce the computation error. Regarding the nonlinear and nonstationary characteristics of EEG, a method known as the empirical mode decomposition (EMD) was proposed by Huang et al. (1998). The EMD method decomposes signal into a series of IMFs. The extraction procedure of an IMF from a signal is known as sifting. In this study, all the local extrema were found, and then, the upper envelope was generated by drawing a cubic spline line through all the local maxima. Similar steps were conducted for the local minima to produce the lower envelope.
All the data should fall in the range between the upper and lower envelopes. The first component h1 is obtained by computing the difference between the data and mean:
(1) |
where m1 denotes the mean, and h1 can only be considered as a proto-IMF. In the next step, h1 is treated as data:
(2) |
After sifting up to k times, h1 becomes an IMF as follows:
(3) |
Then, the first IMF is acquired as h1k:
(4) |
One can repeat these steps for the residual generated by previous loops to obtain successive IMFs.
Finally, the signal can be decomposed as:
(5) |
where N is the number of IMFs, ci(t) is the ith IMF and rN(t) is the residual. A combination of different IMFs can reconstruct a signal to eliminate the artefacts. However, EMD induces the mode mixing problem. An improved EMD known as MEMD was presented by Rehman & Mandic (2009) and noise assisted MEMD (N-A-MEMD) was also introduced by (Ur Rehman & Mandic, 2011). The new N-A-MEMD solves the mode mixing problems, and also can be applied to single channel data by adding Gaussian noise together to constitute multichannel data. The definition is as in Eq. (6).
(6) |
where eθi(t) is the multivariate envelope curves of the whole set of direction vector, K is the length of the vectors, and m(t) is calculated by means of the multivariate envelope curves. Unlike the traditional method that has constant amplitude and frequency in the harmonic component, an IMF provides a simple oscillatory mode as a counterpart and has a variable amplitude and frequency over time. Because decomposition is based on the local characteristics of a data series, it is applicable to nonlinear and nonstationary processes, thus overcoming both pseudo-linearity and stationarity. In our previous study (Huang et al., 2013), the EEG signals were reconstructed by summing IMF2 and IMF3 after decomposition to obtain the filtered signals due to their superior discrimination ability of different states of anaesthesia.
SampEn analysis
Due to the non-stationary characteristic of EEG data, SampEn is appropriate to quantify the degree of irregularity of EEG data. A highly complicated series generates high SampEn values. SampEn, proposed by Richman & Moorman (2000), improves ApEn to more precisely measure the complexity of a physical time series by following steps:
Step 1: Obtain a time series with N points . Set parameters r and m, which represent tolerance and embedding dimension, respectively.
Step 2: The m-dimension template vector is defined as:
(7) |
Step 3: Calculate the distance between two vectors:
(8) |
Step 4: Let Bi be the number of vectors within r of , then:
(9) |
(10) |
Step 5: If we set m = m + 1 and repeat Step 1 to Step 4, we obtain the following:
(11) |
Finally, SampEn is obtained as:
(12) |
Before computing the SampEn, three crucial parameters must be set: the length of the time series N, tolerance r, and dimension m. N was chosen to be 3,750 points (30 s). Various theoretical and clinical applications have described the validity of m = 1 or 2 and r in the range of 0.1 × std − 0.2 × std, where std is the standard deviation. The algorithm was initialized with the following parameters: m = 1 and r = 0.1 × std.
In addition, permutation entropy and recurrence analysis are utilised to analyse the nonstationary EEG data. Permutation entropy is an ordinal analysis method and is quantified by dividing time series into ordinal patterns for describing the order of relations between the present values and a fixed number of equidistant past values. Researchers claim it offers simplicity, robustness, and low computational complexity (Zanin et al., 2012); these advantages are mostly required in nonstationary data analysis, including studies related to anaesthesia (Li, Cui & Voss, 2008). In this study, we chose dimension m = 6 and lag τ = 1, as suggested in previous studies (Li, Cui & Voss, 2008; Liu et al., 2016). Recurrence analysis is a kind of statistical analysis method used to quantify the frequency of the phase space trajectory of a dynamical system visiting approximately the same area in some phase space (Huang, Wang & Singare, 2006; Marwan et al., 2007). We used the RQA tool to conduct the analysis (Marwan et al., 2007). The parameters were chosen: embedding dimension equal to 3, delay equal to 5, and threshold equal to 0.5, which were decided by the global false nearest neighbour and mutual information. In this study, both recurrence rate (RR) and determinism (DET) were used to measure the recurrence analysis. The data segment window size and step were the same as in the SampEn processing part.
Regression models
In order to symbolise DoA more accurately, three typical regression models were applied to train our datasetdash SVM, ANN and Random Forest. These models serve vital purposes in machine learning and have strong self-learning capabilities. Libsvm was employed to conduct SVM regression (Chang & Lin, 2011; Drucker et al., 1997). The exponential function was chosen as the kernel. From engineering perspective, three to four layers were mostly used for ANN (Kourentzes, Barrow & Crone, 2014; Ripley & Ripley, 2001). The backpropagation neural network (BPNN), a structure of 1–9–18–1, was chosen as the model after trials. The commonly recognised BIS values were regarded as the target whereas the entropy features were employed as input data. The input sample data was randomly divided into training (70%), validation (15%), and testing datasets (15%) to adjust the weights and bias for generating the eventual model. Random Forest is a type of ensemble of trees. We used 1,000 trees to train the SampEn extracted from filtered EEG. For all the training processes, the corresponding BIS acted as a target as well.
Statistical analysis
To verify the proposed index capability for decimating the anaesthesia stages, parametric paired Student t-test was used. For the comparisons of intact surgery assessment of DoA, the receiver operating characteristic curve (ROC) was calculated to obtain area under the curve (AUC), which is often used when diagnosing diseases. The BIS binary threshold for distinguishing between anaesthesia condition and awake state is set to 65 (Johansen, Sebel & Sigl, 2000). Pearson’s correlation coefficient and mean absolute error (MAE) were computed between the index and BIS value. To pursue the most appropriate regression model, the Pearson correlation coefficient, MAE, and AUC were calculated for each model by comparing output to the gold standard BIS. Moreover, DoA states were classified into four groups based on the BIS (Johansen, Sebel & Sigl, 2000). ANOVA analysis was conducted for the corresponding stage period SampEn to study the performance in detail. The statistical analysis was performed using SPSS Statistics (IBM v22; Armonk, NY, USA) and MATLAB (Mathworks R2014b; Natick, MA, USA).
The entire workflow diagram is shown in Fig. 2. First, the valid subjects were selected, and then, the preprocessing was conducted. Afterwards, MEMD was applied to filter the data, thus constructing new data by combining IMF2 and IMF3. Then, SampEn features were extracted from data segments for all cases followed by ANN modeling. Finally, the results were statistically compared to BIS to evaluate the performance, including the perioperative analysis evaluation and detailed sub-level group ANOVA analysis.
Results
MEMD filter performance
As discussed previously, the non-stationary and non-linear characteristics of physiological signals (i.e., EEG in this study) require a robust and appropriate method to eliminate the weakness of traditional FFT methods. MEMD was applied to reconstruct the signal. A representative patient’s SampEn derived by summing both the combination signal (i.e., IMF2 + IMF3) and raw EEG data were presented in Fig. 3 with the gold standard BIS. By a rough visual evaluation, it is obvious that the filtered outcome behaved better than the raw one. Statistically, Pearson correlation coefficient is calculated between the perioperative SampEn and BIS for both the filtered signal and raw signal, as displayed in Fig. 4. The coefficient displayed a significant increase from 0.51 ± 0.17 to 0.72 ± 0.09 (p < 0.05). Thus, MEMD filtering improved the purity of a signal although we only used SampEn as an example to demonstrate this filtering effect.
Comparison with recurrence analysis and permutation entropy
Discrimination of the different anaesthetic stages
The awake stage denotes the period before the anaesthetic injection. Patients are notably aware during this stage. Then LoC is slowly induced by drugs, eventually followed by a stable unconsciousness state. By the end of the operation, the patients gradually recover consciousness the as the drug volume decreases. Therefore, it is worthwhile to distinguish the stages prior to tracking DoA for the entire surgery. A 2-min noise-free segment was extracted from the three stages for each patient. By applying a window size of 30 s and a window step of 5 s, 24 data points were calculated for each stage, and then an average of the points of all cases were taken. Figure 5 displayed four indexes percentile distribution of the three stages—RQA-RR, RQA-DET, PEn, and SampEn. All the indexes percentile distributions displayed a significant difference between the unconsciousness and conscious state (p < 0.05). Among them (Figs. 5A and 5D), the RR and SampEn were more robust and compatible to the awake and recovery period. Therefore, these distributions were more capable of processing the consciousness stage data and reducing the risk of classifying the EEG into a wrong stage. Moreover, both distributions were more sensitive to the recovery from anaesthesia. Both RQA results increased during anaesthesia, whereas the entropy ones had a low value at LoC (Figs. 5C and 5D). This finding is consistent with the stable and less complex brain activities during unconsciousness.
Perioperative performance of evaluating DoA
To find the optimal index to symbolise DoA, a further statistical analysis was conducted. By regarding the corresponding perioperative BIS value as the gold standard, the correlation coefficient, MAE, and AUC under ROC were obtained, as listed in Table 2. From the comparisons, we observed that the entropy method had a much better capability than the RQA group. Moreover, SampEn had the highest correlation with BIS (0.72 ± 0.09), and the corresponding AUC was the highest as well (0.95 ± 0.04). The AUC usually ranged from 0.5 to 1. The more effectively a method classified input, the larger the value was. This provided strong evidence for using SampEn to assess DoA in this study.
Table 2. Comparisons between SampEn and other typical nonstationary and nonlinear analysis methods.
CC | AUC | MAE | |
---|---|---|---|
RR (RQA) | −0.16 ± 0.16 | 0.24 ± 0.13 | 48.60 ± 7.95 |
DET (RQA) | −0.39 ± 0.16 | 0.16 ± 0.10 | 58.33 ± 7.98 |
PEn | 0.55 ± 0.24 | 0.91 ± 0.09 | 42.81 ± 7.92 |
SampEn | 0.72 ± 0.09 | 0.95 ± 0.04 | 47.23 ± 7.84 |
Notes.
MAE is not an absolute metric because it can decrease through regression. Values are presented with mean ± STD. Markers symbolize the statistical significant Student t-test.
- PEn
- Permutation
- CC
- correlation coefficient
- AUC
- area under curve
- MAE
- mean absolute error
- RQA
- recurrence quantification analysis
- RR
- recurrence rate
- DET
- determinism
- SampEn
- sample entropy
Regression model performance
To improve the capability of the selected SampEn index for predicting DoA, non-linear regression models including SVM, ANN and Random Forest, were used. Table 3 provides statistical result of output performance of the three models. It can be clearly confirmed that Random Forest generated the best values in terms of the correlation coefficient (0.90 ± 0.05), AUC (0.98 ± 0.02) and MAE (5.22 ± 1.12) among all three models. In addition, it is found that SVM had a statistical result that is similar to ANN, which implied their approximate effect on the data. Moreover, this can be observed in Fig. 6, in which the Random Forest (Fig. 6C) showed a strong linear relationship with BIS, of which the R2 is the highest. Clearly, Figs. 6A and 6B displayed a similar scatter distribution, which was consistent with the results displayed in the Table 3.
Table 3. Comparisons of performance of the three different regression methods.
CC | AUC | MAE | |
---|---|---|---|
SVM | 0.76 ± 0.09 | 0.95 ± 0.04 | 9.08 ± 2.72 |
ANN | 0.79 ± 0.08 | 0.96 ± 0.04 | 8.88 ± 2.46 |
Random Forest | 0.90 ± 0.05 | 0.98 ± 0.02 | 5.22 ± 1.12 |
Notes.
Values are presented with mean ± STD. Markers symbolize the statistical significant Student t-test.
- SVM
- support vector machine
- ANN
- artificial neural network
- CC
- correlation coefficient
- AUC
- area under curve
- MAE
- mean absolute error
ANOVA analysis of four data-pair subgroups
By considering BIS as the gold standard, four groups were obtained–Phase 1 (0–40, deep anaesthesia), Phase 2 (40–65, general anaesthesia), Phase 3 (65–85, light anaesthesia), and Phase 4 (85–100, awake). The corresponding model output was automatically divided into these phases. Thus, 33,895 data pairs (i.e., models’ output and BIS) existed for the whole procedure over 24 patients. Statistical analysis was performed using one-way ANOVA on ranks and Student-Newman-Keuls test for pairwise comparisons. The results showed a general ascending trend and a significant difference when compared with its previous subgroups (p < 0.05) for each model, as shown in Fig. 7. The variation of Random Forest (Fig. 7C) in each stage was the smallest, and the gap between different phases was the largest; this implied a superior DoA tracking ability. Moreover, in Tables 4–6, the values distribution in each phase was presented. For all the phase groups, the majority values fell into the corresponding phase definition range except Phase 4 in SVM and Phase 3 in ANN (detailed in the captions of Tables 4–6). This behavior can be more clearly observed in Figs. 6A and 6B of the scatter plot. Note that the Random Forest had the optimal subgroup value distribution, which was consistent with Fig. 6C. This demonstrated a general linear relationship between the two variables. The Random Forest regression performed the best regression effect.
Table 4. SVM outputs distribution across each phase.
BIS Level | Model output | ||||
---|---|---|---|---|---|
n | 0-40 | 40-65 | 65-85 | 85-100 | |
BIS 0–40 | 15,467 | 11,293 (73.0%) | 4,090 (26.4%) | 84 (0.6%) | 0 (0%) |
BIS 40–65 | 13,952 | 6,302 (45.2%) | 7,348 (52.7%) | 287 (2.1%) | 15 (0%) |
BIS 65–85 | 2,274 | 83 (3.6%) | 899 (39.5%) | 1,008 (44.4%) | 284 (12.5%) |
BIS 85–100 | 2,202 | 47 (2.1%) | 508 (23.1%) | 872 (39.6%) | 775 (35.2%) |
Table 6. Random Forest outputs distribution across each phase.
BIS level | Model output | ||||
---|---|---|---|---|---|
n | 0–40 | 40–65 | 65–85 | 85–100 | |
BIS 0–40 | 15,467 | 12,820 (82.9%) | 2,647 (17.1%) | 0 (0%) | 0 (0%) |
BIS 40–65 | 13,952 | 2,622 (18.8%) | 11,091 (79.5%) | 239 (1.7%) | 0 (0%) |
BIS 65–85 | 2,274 | 0 (0%) | 761 (33.5%) | 1,349 (59.3%) | 164 (7.2%) |
BIS 85–100 | 2,202 | 0 (0%) | 13 (0.6%) | 836 (38.0%) | 1352 (61.4%) |
Table 5. ANN outputs distribution across each phase.
BIS level | Model output | ||||
---|---|---|---|---|---|
n | 0–40 | 40–65 | 65–85 | 85–100 | |
BIS 0–40 | 15,467 | 9171 (59.3%) | 6,219 (40.2%) | 76 (0.5%) | 1 (0%) |
BIS 40–65 | 13,952 | 4218 (30.2%) | 9,466 (67.8%) | 248 (1.7%) | 20 (0.1%) |
BIS 65–85 | 2,274 | 47 (2.1%) | 958 (42.1%) | 851 (37.4%) | 418 (18.4%) |
BIS 85–100 | 2,202 | 24 (1.1%) | 547 (24.8%) | 623 (28.3%) | 1008 (45.8%) |
Discussion
In this paper, a novel method for estimating DoA was proposed. By extracting EEG signal features, the anaesthesia-related EEG patterns were acquired with different anaesthetic levels. After selecting the SampEn as the input, three regression models were established and compared in terms of the regression effect by taking BIS as a DoA reference. The distribution of the proposed output index in four different subgroups (phases) demonstrated a linear relationship with BIS, of which the Random Forest performed the best. Results illustrated the capability of the proposed method to signify DoA over 24 patients. The analysis is focused on the perspective of the time domain features, which could facilitate DoA research method exploration. From the time domain, instead of only the spectral angle, this methodology might provide a universal solution to those situations where the commercial index is not applicable, such as BIS deficiency in ketamine or nitrous oxide (Avidan et al., 2008).
EEG serves a vital role for brain-related disease studies such as epilepsy, and Parkinson’s diseases due to its strong connection with brain activity (Le Van Quyen et al., 2001; Ly et al., 2016; Vlisides & Mashour, 2017)). Similar to anaesthesia, EEG traces show dominant low-frequency and high-amplitude oscillations. However, the high frequency components decrease from common medication, and thus provide the opportunity to develop EEG-based methods to track DoA automatically (Purdon et al., 2015). However, EEG is easy to be contaminated by EOG, EMG baseline drift and nonlinear distortion of the amplitude and so on (Liu et al., 2017a). Moreover, due to its non-stationary and non-linear properties, the traditional filtering method may not work appropriately. In our procedure, MEMD produced an optimal artefact cancellation capability. By comparing the correlation coefficients in Fig. 4, it is observed that the filtered effect is significantly improved from 0.51 ± 0.17 to 0.72 ± 0.09 (p < 0.05). Although MEMD is an improved EMD method (Rehman & Mandic, 2009), it still requires a long computation time. Therefore, MEMD cannot be applied to real-time situations currently. This drawback of MEMD might be overcome due to hardware development. Another point of concern is the principle of selecting the combination of IMFs (Huang, 2014; Komaty et al., 2014). The decomposition result varies on an individual basis because EEG signals might vary from patient to patient, although IMF2 + IMF3 were selected based on previous study (Huang et al., 2013). Further exploration might reveal superior combinations. These properties should be analysed more precisely, in terms of different anaesthetics, age group, and so on. Another solution is that ICA (Long & Sheng, 2014), or wavelet (Mamun, Al-Kadi & Marufuzzaman, 2013) can be employed for obtaining superior artefact rejection (Khatwani & Tiwari, 2013).
Entropy is an optimal concept for the series randomness analysis (Fast, 1962), and is becoming a powerful analysis tool for information process of the EEG activity currently (Al-Kadi, Reaz & Ali, 2013; Chu et al., 2017). Entropy implies the degree of irregularity of a system. Therefore, we observed significant decrease in the entropy values during anaesthesia, as shown in Fig. 4. However, the factors are chosen empirically. It is worthwhile to study the rationale of proper factors to improve the measurement accuracy further. Results should be compared with a range of these parameters. Moreover, an EEG signal naturally exhibits complicated variations, which occur from complex self-regulating systems over multiple temporal scales. Therefore, the advanced multiscale entropy family, such as multiscale SampEn or multiscale PEn, can be considered as alternative methods although PEn was inferior to SampEn in our data analysis (Shalbaf et al., 2013). Moreover, brains spatial difference suggest that multichannel analysis should be employed, which may present a different level relationship related to anaesthesia.
As observed in the regression step, Random Forest proves to be best. The correlation coefficient, AUC, and MAE of the Random Forest were the highest compared to those of the other two models. SVM and ANN demonstrated similar capability of regression. Both models largely addressed the gap between SampEn and BIS; however, their correlation coefficient and AUC remained approximate. Random Forest improved the metrics significantly. Because we chose the typical and most widely used parameters for each model based on literature (Chang & Lin, 2011; Li, Cui & Voss, 2008; Liu et al., 2016; Ripley & Ripley, 2001; Shalbaf, Behnam & Moghadam, 2015), more research on the appropriate parameters may be required. Moreover, these learning models rely on large datasets to improve themselves. Therefore, a higher amount of data should be provided for training. Moreover, some other machine learning methods, such as deep neural network or boosting, could be applied to increase the accuracy of DoA assessment.
Our present work has several limitations. First, the BIS value has some time delay (Shalbaf et al., 2013) and drawbacks for paediatric application somehow. Therefore, the Expert Assessment of Consciousness Level, which is obtained from experienced anaesthesiologists’ comprehensive evaluation through considering multiple clinical indicators used in practice, would be referred to be as the gold standard for our method aimed at minimising drawbacks and maximising universality. Second, the BIS has been a subject of some controversy on the measurement of DoA over different drugs (Avidan et al., 2008), because the BIS may not correctly reflect patients status when NMDA-dependent anaesthetic agents such as ketamine and nitrous oxide are used; NMDA-dependent anaesthetic agents induce anaesthesia by enhancing inhibitory receptors. This fact explains why BIS (i.e., when using EEG) is not considered as a compulsory standard in practical operation. However, the BIS correctly reflects patients status when GABA-dependent anaesthetic agents such as propofol and desflurane, which induce anaesthesia by suppressing the excitatory nervous circuits, are employed (Uhrig, Dehaene & Jarraya, 2014). Therefore, 24 patients received the GABA receptor dependent drugs in this study. BIS could be selected as the gold standard for our methodology. However, this implies that our method might not be applied to other agent with different anaesthesia mechanisms such as ketamine, nitrous oxide mentioned previously. This needs to be confirmed whether it works as well taking new gold standard instead of BIS, which could exemplify rationale of our methodology. Furthermore, the sample size could be improved further in order to decrease the sample variations, although 24 cases are initially required to obtain statistical results. Fourth, only adults were selected for our study. There is a considerable difference between adults and infants from the clinical viewpoint (Schechter, Allen & Hanson, 1986). Thus, our method may not be applicable to children. Fifth, our definition of the four groups of different anaesthetic levels may be questionable. Further appropriate categorisation of the groups might be required. Analysis with respect to this should be more comprehensive. Finally, the estimation of DoA with a single parameter such as the EEG measure does not work well over time. Accurate DoA monitoring might be improved by taking activity of the autonomic nervous system into consideration (Cohen, Cameron & Duncan, 1990). Therefore, even though numerous trials have sought to integrate several modes of observations for increasing the accuracy of DoA evaluation, there is no single perfect index. Overall, these challenges provide many possibilities for researches to explore a more advanced DoA assessment in future.
Conclusions
In this paper, a novel method protocol based on the SampEn concept and Random Forest learning method for estimating DoA through EEG signals was proposed. The method provides optimal performance compared with other typical methods. First, the SampEn feature derived from EEG signals distinguished the three selected traditional stages of anaesthesia. Second, the feasibility of MEMD to filter EEG data and regression model effect was demonstrated. By considering the BIS as the gold standard, optimal DoA assessment performance for the entire perioperative duration was verified with ROC. More importantly, ANOVA analysis was conducted for 24 patients induced with GABA dependent anaesthetic agents under different anaesthetic levels, which demonstrates a clear linear relationship with the gold standard. The results exhibited high accuracy of DoA estimation. This work provides a basis for DoA evaluation that may be especially helpful for optimising anaesthetic titration to avoid unwarranted effects when deep anaesthesia is performed inappropriately and for promoting fast postoperative recovery.
Funding Statement
This work was supported by the Innovation Center for Big Data and Digital Convergence, Yuan Ze University, Taiwan, which are sponsored by Ministry of Education, and the National Chung-Shan Institute of Science & Technology in Taiwan. It is also supported by the Wuhan University of Technology international exchange program (No. 2015-JL-012) and the National Natural Science Foundation of China (No. 51475342,51675389). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Additional Information and Declarations
Competing Interests
The authors declare there are no competing interests.
Author Contributions
Quan Liu analyzed the data, prepared figures and/or tables, approved the final draft.
Li Ma performed the experiments, analyzed the data, prepared figures and/or tables, approved the final draft.
Shou-Zen Fan performed the experiments, contributed reagents/materials/analysis tools, approved the final draft.
Maysam F. Abbod authored or reviewed drafts of the paper, approved the final draft.
Jiann-Shing Shieh conceived and designed the experiments, prepared figures and/or tables, authored or reviewed drafts of the paper, approved the final draft.
Human Ethics
The following information was supplied relating to ethical approvals (i.e., approving body and any reference numbers):
All studies were approved by the Research Ethics Committee, National Taiwan University Hospital (NTUH), Taiwan, and written informed consent was obtained from patients (No: 201302078RINC).
Data Availability
The following information was supplied regarding data availability:
Ma, Li (2017): EEG and BIS raw data. figshare. Fileset. https://doi.org/10.6084/m9.figshare.5589841.v1.
References
- Akeju et al. (2014).Akeju O, Westover MB, Pavone KJ, Sampson AL, Hartnack KE, Brown EN, Purdon PL. Effects of sevoflurane and propofol on frontal electroencephalogram power and coherence. Anesthesiology. 2014;121:990–998. doi: 10.1097/ALN.0000000000000436. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Al-Kadi, Reaz & Ali (2013).Al-Kadi MI, Reaz MBI, Ali MAM. Evolution of electroencephalogram signal analysis techniques during anesthesia. Sensors. 2013;13:6605–6635. doi: 10.3390/s130506605. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Alpaydin (2014).Alpaydin E. Introduction to machine learning. Cambridge: MIT press; 2014. [Google Scholar]
- Avidan et al. (2008).Avidan MS, Zhang L, Burnside BA, Finkel KJ, Searleman AC, Selvidge JA, Saager L, Turner MS, Rao S, Bottros M. Anesthesia awareness and the bispectral index. New England Journal of Medicine. 2008;358:1097–1108. doi: 10.1056/NEJMoa0707361. [DOI] [PubMed] [Google Scholar]
- Billard et al. (1997).Billard V, Gambus PL, Chamoun N, Stanski DR, Shafer SL. A comparison of spectral edge, delta power, and bispectral index as EEG measures of alfentanil, propofol, and midazolam drug effect. Clinical Pharmacology & Therapeutics. 1997;61:45–58. doi: 10.1016/S0009-9236(97)90181-8. [DOI] [PubMed] [Google Scholar]
- Brown, Lydic & Schiff (2010).Brown EN, Lydic R, Schiff ND. General anesthesia, sleep, and coma. New England Journal of Medicine. 2010;363:2638–2650. doi: 10.1056/NEJMra0808281. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bruhn et al. (2001).Bruhn J, Lehmann LE, Röpcke H, Bouillon TW, Hoeft A. Shannon entropy applied to the measurement of the electroencephalographic effects of desflurane. Anesthesiology: The Journal of the American Society of Anesthesiologists. 2001;95:30–35. doi: 10.1097/00000542-200107000-00010. [DOI] [PubMed] [Google Scholar]
- Chang & Lin (2011).Chang C-C, Lin C-J. LIBSVM: a library for support vector machines. ACM Transactions on Intelligent Systems and Technology. 2011;2(3) Article 27. [Google Scholar]
- Ching et al. (2010).Ching S, Cimenser A, Purdon PL, Brown EN, Kopell NJ. Thalamocortical model for a propofol-induced α-rhythm associated with loss of consciousness. Proceedings of the National Academy of Sciences of the United States of America. 2010;107:22665–22670. doi: 10.1073/pnas.1017069108. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chu et al. (2017).Chu Y-J, Chang C-F, Shieh J-S, Lee W-T. The potential application of multiscale entropy analysis of electroencephalography in children with neurological and neuropsychiatric disorders. Entropy. 2017;19(8):428. doi: 10.3390/e19080428. [DOI] [PubMed] [Google Scholar]
- Cohen, Cameron & Duncan (1990).Cohen MM, Cameron CB, Duncan PG. Pediatric anesthesia morbidity and mortality in the perioperative period. Anesthesia & Analgesia. 1990;70:160–167. doi: 10.1213/00000539-199002001-00160. [DOI] [PubMed] [Google Scholar]
- Drucker et al. (1997).Drucker H, Burges CJ, Kaufman L, Smola AJ, Vapnik V. Advances in neural information processing systems. MIT Press; Cambridge: 1997. Support vector regression machines; pp. 155–161. [Google Scholar]
- Fast (1962).Fast JD. Entropy. New York: McGraw Hill Education; 1962. [Google Scholar]
- Hämäläinen et al. (1993).Hämäläinen M, Hari R, Ilmoniemi RJ, Knuutila J, Lounasmaa OV. Magnetoencephalography—theory, instrumentation, and applications to noninvasive studies of the working human brain. Reviews of Modern Physics. 1993;65(2):413–497. doi: 10.1103/RevModPhys.65.413. [DOI] [Google Scholar]
- Höcker et al. (2010).Höcker J, Raitschew B, Meybohm P, Broch O, Stapelfeldt C, Gruenewald M, Cavus E, Steinfath M, Bein B. Differences between bispectral index and spectral entropy during xenon anaesthesia: a comparison with propofol anaesthesia. Anaesthesia. 2010;65:595–600. doi: 10.1111/j.1365-2044.2010.06344.x. [DOI] [PubMed] [Google Scholar]
- Huang (2014).Huang NE. Hilbert-Huang transform and its applications. World Scientific; Singapore: 2014. [Google Scholar]
- Huang et al. (2013).Huang J-R, Fan S-Z, Abbod MF, Jen K-K, Wu J-F, Shieh J-S. Application of multivariate empirical mode decomposition and sample entropy in EEG signals via artificial neural networks for interpreting depth of anesthesia. Entropy. 2013;15:3325–3339. doi: 10.3390/e15093325. [DOI] [Google Scholar]
- Huang et al. (1998).Huang NE, Shen Z, Long SR, Wu MC, Shih HH, Zheng Q, Yen N-C, Tung CC, Liu HH. The empirical mode decomposition and the Hilbert spectrum for nonlinear and non-stationary time series analysis. Proceedings of the Royal Society of London A: Mathematical, Physical and Engineering Sciences. 1998;454:903–995. [Google Scholar]
- Huang, Wang & Singare (2006).Huang L, Wang W, Singare S. International conference on natural computation. Berlin: Springer; 2006. Using back propagation feedback neural networks and recurrence quantification analysis of EEGs predict responses to incision during anesthesia; pp. 364–373. [Google Scholar]
- Johansen & Sebel (2000).Johansen JW, Sebel PS. Development and clinical application of electroencephalographic bispectrum monitoring. Anesthesiology. 2000;93:1336–1344. doi: 10.1097/00000542-200011000-00029. [DOI] [PubMed] [Google Scholar]
- Johansen, Sebel & Sigl (2000).Johansen JW, Sebel PS, Sigl JC. Clinical impact of hypnotic-titration guidelines based on EEG bispectral index (BIS) monitoring during routine anesthetic care. Journal of Clinical Anesthesia. 2000;12:433–443. doi: 10.1016/S0952-8180(00)00187-2. [DOI] [PubMed] [Google Scholar]
- Kandel et al. (2000).Kandel ER, Schwartz JH, Jessell TM, Siegelbaum SA, Hudspeth AJ. Principles of neural science. McGraw-hill; New York: 2000. [Google Scholar]
- Katoh, Suzuki & Ikeda (1998).Katoh T, Suzuki A, Ikeda K. Electroencephalographic derivatives as a tool for predicting the depth of sedation and anesthesia induced by sevoflurane. Anesthesiology. 1998;88:642–650. doi: 10.1097/00000542-199803000-00014. [DOI] [PubMed] [Google Scholar]
- Khatwani & Tiwari (2013).Khatwani P, Tiwari A. A survey on different noise removal techniques of eeg signals. International Journal of Advanced Research in Computer and Communication Engineering. 2013;2:1091–1095. [Google Scholar]
- Kiersey, Bickford & Faulcone Jr (1951).Kiersey DY, Bickford RG, Faulcone Jr A. Electro-encephalographic patterns produced by thiopental sodium during surgical operations: description and classification. Oxford: Oxford University Press; 1951. [DOI] [PubMed] [Google Scholar]
- Komaty et al. (2014).Komaty A, Boudraa A-O, Augier B, Dare-Emzivat D. EMD-based filtering using similarity measure between probability density functions of IMFs. IEEE Transactions on Instrumentation and Measurement. 2014;63:27–34. doi: 10.1109/TIM.2013.2275243. [DOI] [Google Scholar]
- Kourentzes, Barrow & Crone (2014).Kourentzes N, Barrow DK, Crone SF. Neural network ensemble operators for time series forecasting. Expert Systems with Applications. 2014;41:4235–4244. doi: 10.1016/j.eswa.2013.12.011. [DOI] [Google Scholar]
- Kriegeskorte (2015).Kriegeskorte N. Deep neural networks: a new framework for modeling biological vision and brain information processing. Annual Review of Vision Science. 2015;1:417–446. doi: 10.1146/annurev-vision-082114-035447. [DOI] [PubMed] [Google Scholar]
- Lepouse et al. (2006).Lepouse C, Lautner C, Liu L, Gomis P, Leon A. Emergence delirium in adults in the post-anaesthesia care unit. British Journal of Anaesthesia. 2006;96:747–753. doi: 10.1093/bja/ael094. [DOI] [PubMed] [Google Scholar]
- Le Van Quyen et al. (2001).Le Van Quyen M, Martinerie J, Navarro V, Boon P, D’Havé M, Adam C, Renault B, Varela F, Baulac M. Anticipation of epileptic seizures from standard EEG recordings. The Lancet. 2001;357:183–188. doi: 10.1016/S0140-6736(00)03591-1. [DOI] [PubMed] [Google Scholar]
- Li, Cui & Voss (2008).Li X, Cui S, Voss LJ. Using permutation entropy to measure the electroencephalographic effects of sevoflurane. Anesthesiology: The Journal of the American Society of Anesthesiologists. 2008;109:448–456. doi: 10.1097/ALN.0b013e318182a91b. [DOI] [PubMed] [Google Scholar]
- Liaw & Wiener (2002).Liaw A, Wiener M. Classification and regression by randomForest. R News. 2002;2:18–22. [Google Scholar]
- Liu et al. (2016).Liu Q, Chen Y-F, Fan S-Z, Abbod MF, Shieh J-S. A comparison of five different algorithms for EEG signal analysis in artifacts rejection for monitoring depth of anesthesia. Biomedical Signal Processing and Control. 2016;25:24–34. doi: 10.1016/j.bspc.2015.10.010. [DOI] [Google Scholar]
- Liu et al. (2017a).Liu Q, Chen Y-F, Fan S-Z, Abbod MF, Shieh J-S. EEG artifacts reduction by multivariate empirical mode decomposition and multiscale entropy for monitoring depth of anaesthesia during surgery. Medical & Biological Engineering & Computing. 2017a;55:1435–1450. doi: 10.1007/s11517-016-1598-2. [DOI] [PubMed] [Google Scholar]
- Liu et al. (2017b).Liu Q, Ma L, Chiu R-C, Fan S-Z, Abbod MF, Shieh J-S. HRV-derived data similarity and distribution index based on ensemble neural network for measuring depth of anaesthesia. PeerJ. 2017b;5:e4067. doi: 10.7717/peerj.4067. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Long & Sheng (2014).Long L, Sheng L. EEG signal denoising based on fast independent component analysis. Computer Measurement & Control. 2014;11 Article 77. [Google Scholar]
- Luginbühl et al. (2003).Luginbühl M, Wüthrich S, Petersen-Felix S, Zbinden A, Schnider T. Different benefit of bispectal index (BIS™) in desflurane and propofol anesthesia. Acta Anaesthesiologica Scandinavica. 2003;47:165–173. doi: 10.1034/j.1399-6576.2003.00041.x. [DOI] [PubMed] [Google Scholar]
- Ly et al. (2016).Ly QT, Handojoseno AA, Gilat M, Nguyen N, Chai R, Tran Y, Lewis SJ, Nguyen HT. Detection of Gait Initiation Failure in Parkinson’s disease patients using EEG signals. Engineering in medicine and biology society (EMBC), 2016 IEEE 38th annual international conference of the; Piscataway. 2016. pp. 1599–1602. [DOI] [PubMed] [Google Scholar]
- Mamun, Al-Kadi & Marufuzzaman (2013).Mamun M, Al-Kadi M, Marufuzzaman M. Effectiveness of wavelet denoising on electroencephalogram signals. Journal of Applied Research and Technology. 2013;11:156–160. doi: 10.1016/S1665-6423(13)71524-4. [DOI] [Google Scholar]
- Martin, Faulconer & Bickford (1959).Martin JT, Faulconer A, Bickford RG. Electroencephalography in anesthesiology. The Journal of the American Society of Anesthesiologists. 1959;20:359–376. doi: 10.1097/00000542-195905000-00017. [DOI] [PubMed] [Google Scholar]
- Marwan et al. (2007).Marwan N, Romano MC, Thiel M, Kurths J. Recurrence plots for the analysis of complex systems. Physics Reports. 2007;438:237–329. doi: 10.1016/j.physrep.2006.11.001. [DOI] [Google Scholar]
- McKeever, Johnston & Davidson (2014).McKeever S, Johnston L, Davidson AJ. Sevoflurane-induced changes in infants’ quantifiable electroencephalogram parameters. Pediatric Anesthesia. 2014;24:766–773. doi: 10.1111/pan.12366. [DOI] [PubMed] [Google Scholar]
- Monk et al. (2008).Monk TG, Weldon BC, Garvan CW, Dede DE, Van Der Aa MT, Heilman KM, Gravenstein JS. Predictors of cognitive dysfunction after major noncardiac surgery. Anesthesiology: The Journal of the American Society of Anesthesiologists. 2008;108:18–30. doi: 10.1097/01.anes.0000296071.19434.1e. [DOI] [PubMed] [Google Scholar]
- Palanca, Mashour & Avidan (2009).Palanca BJA, Mashour GA, Avidan MS. Processed electroencephalogram in depth of anesthesia monitoring. Current Opinion in Anesthesiology. 2009;22:553–559. doi: 10.1097/ACO.0b013e3283304032. [DOI] [PubMed] [Google Scholar]
- Poorun et al. (2016).Poorun R, Hartley C, Goksan S, Worley A, Boyd S, Cornelissen L, Berde C, Rogers R, Ali T, Slater R. Electroencephalography during general anaesthesia differs between term-born and premature-born children. Clinical Neurophysiology. 2016;127:1216–1222. doi: 10.1016/j.clinph.2015.10.041. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Purdon et al. (2013).Purdon PL, Pierce ET, Mukamel EA, Prerau MJ, Walsh JL, Wong KFK, Salazar-Gomez AF, Harrell PG, Sampson AL, Cimenser A. Electroencephalogram signatures of loss and recovery of consciousness from propofol. Proceedings of the National Academy of Sciences of the United States of America. 2013;110:E1142–E1151. doi: 10.1073/pnas.1221180110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Purdon et al. (2015).Purdon PL, Sampson A, Pavone KJ, Brown EN. Clinical electroencephalography for anesthesiologistspart I: background and basic signatures. The Journal of the American Society of Anesthesiologists. 2015;123:937–960. doi: 10.1097/ALN.0000000000000841. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rehman & Mandic (2009).Rehman N, Mandic DP. Multivariate empirical mode decomposition. Proceedings of The Royal Society of London A: Mathematical, Physical and Engineering Sciences. 2009;466(2117):1291–1302. doi: 10.1098/rspa.2009.0502. [DOI] [Google Scholar]
- Richman & Moorman (2000).Richman JS, Moorman JR. Physiological time-series analysis using approximate entropy and sample entropy. American Journal of Physiology-Heart and Circulatory Physiology. 2000;278:H2039–H2049. doi: 10.1152/ajpheart.2000.278.6.H2039. [DOI] [PubMed] [Google Scholar]
- Ripley & Ripley (2001).Ripley BD, Ripley RM. Clinical applications of artificial neural networks. Cambridge: Cambridge University Press; 2001. Neural networks as statistical methods in survival analysis; pp. 237–255. [Google Scholar]
- Schechter, Allen & Hanson (1986).Schechter NL, Allen D, Hanson K. Status of pediatric pain control: a comparison of hospital analgesic usage in children and adults. Pediatrics. 1986;77:11–15. [PubMed] [Google Scholar]
- Shalbaf, Behnam & Moghadam (2015).Shalbaf R, Behnam H, Moghadam HJ. Monitoring depth of anesthesia using combination of EEG measure and hemodynamic variables. Cognitive Neurodynamics. 2015;9:41–51. doi: 10.1007/s11571-014-9295-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shalbaf et al. (2013).Shalbaf R, Behnam H, Sleigh JW, Steyn-Ross A, Voss LJ. Monitoring the depth of anesthesia using entropy features and an artificial neural network. Journal of Neuroscience Methods. 2013;218:17–24. doi: 10.1016/j.jneumeth.2013.03.008. [DOI] [PubMed] [Google Scholar]
- Shalbaf et al. (2012).Shalbaf R, Behnam H, Sleigh J, Voss L. Measuring the effects of sevoflurane on electroencephalogram using sample entropy. Acta Anaesthesiologica Scandinavica. 2012;56:880–889. doi: 10.1111/j.1399-6576.2012.02676.x. [DOI] [PubMed] [Google Scholar]
- Shannon (2001).Shannon CE. A mathematical theory of communication. ACM SIGMOBILE Mobile Computing and Communications Review. 2001;5(1):3–55. [Google Scholar]
- Uhrig, Dehaene & Jarraya (2014).Uhrig L, Dehaene S, Jarraya B. Cerebral mechanisms of general anesthesia. Annales Francaises d’anesthesie et de Reanimation. 2014;33(2):72–82. doi: 10.1016/j.annfar.2013.11.005. [DOI] [PubMed] [Google Scholar]
- Ur Rehman & Mandic (2011).Ur Rehman N, Mandic DP. Filter bank property of multivariate empirical mode decomposition. IEEE Transactions on Signal Processing. 2011;59:2421–2426. doi: 10.1109/TSP.2011.2106779. [DOI] [Google Scholar]
- Vizuete et al. (2014).Vizuete JA, Pillay S, Ropella KM, Hudetz AG. Graded defragmentation of cortical neuronal firing during recovery of consciousness in rats. Neuroscience. 2014;275:340–351. doi: 10.1016/j.neuroscience.2014.06.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vlisides & Mashour (2017).Vlisides PE, Mashour GA. Monitoring the nervous system for anesthesiologists and other health care professionals. Cham: Springer; 2017. Clinical application of raw and processed EEG; pp. 193–204. [Google Scholar]
- Vutskits & Xie (2016).Vutskits L, Xie Z. Lasting impact of general anaesthesia on the brain: mechanisms and relevance. Nature Reviews Neuroscience. 2016;17:705–717. doi: 10.1038/nrn.2016.128. [DOI] [PubMed] [Google Scholar]
- Zanin et al. (2012).Zanin M, Zunino L, Rosso OA, Papo D. Permutation entropy and its main biomedical and econophysics applications: a review. Entropy. 2012;14:1553–1577. doi: 10.3390/e14081553. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The following information was supplied regarding data availability:
Ma, Li (2017): EEG and BIS raw data. figshare. Fileset. https://doi.org/10.6084/m9.figshare.5589841.v1.