A 36-year-old man underwent ERCP for a common bile duct stone in another hospital. To prevent post-ERCP pancreatitis, a 5F and 3-cm single-flanged pancreatic duct (PD) straight stent had been placed. However, the stent was inadvertently inserted upside down and migrated up to the comparatively distal PD. Attempts to remove the stent with use of a balloon catheter and a Soehendra stent retriever were unsuccessful. For further endoscopic treatment, he was referred to our hospital. We used the SpyBite biopsy forceps (outer diameter 1 mm; Boston Scientific, Natick, Mass) through the Soehendra biliary dilation catheter (4F-7F; Cook Medical, Bloomington, Ind) to remove the migrated stent (Fig. 1A). First, a 0.025-inch guidewire was placed in the PD beyond the migrated stent, and the catheter was advanced over the guidewire in front of the migrated stent. After removal of the guidewire, the biopsy forceps was inserted through the catheter to the distal end of the stent (Fig. 1B). Then, the PD stent could be grasped by the biopsy forceps. Finally, the stent was completely removed (Fig. 1C; Video 1, available online at www.VideoGIE.org) without any procedure-related adverse event.
Figure 1.
A, We used the SpyBite biopsy forceps (outer diameter 1 mm) through the Soehendra biliary dilation catheter (4F-7F, taper length 3 cm). B, The biopsy forceps was inserted through the catheter and grasped the distal end of the stent directly. C, The stent was completely removed.
Disclosure
All authors disclosed no financial relationships relevant to this publication.
Footnotes
Written transcript of the video audio is available online at www.VideoGIE.org.
Supplementary data
The Soehendra biliary dilation catheter was advanced over the guidewire in front of the migrated stent. After removal of the guidewire, the SpyBite biopsy forceps was inserted through the catheter to the distal end of the stent. Then, the biopsy forceps grasped the stent directly. Finally, the stent was completely removed.
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
The Soehendra biliary dilation catheter was advanced over the guidewire in front of the migrated stent. After removal of the guidewire, the SpyBite biopsy forceps was inserted through the catheter to the distal end of the stent. Then, the biopsy forceps grasped the stent directly. Finally, the stent was completely removed.