Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2019 Jul 1.
Published in final edited form as: Otol Neurotol. 2018 Jul;39(6):755–764. doi: 10.1097/MAO.0000000000001807

Lateral semi-circular canal pressures during cochlear implant electrode insertion: A possible mechanism for postoperative vestibular loss

Renee M Banakis Hartl 1, Nathaniel T Greene 1, Herman A Jenkins 1, Stephen P Cass 1, Daniel J Tollin 1,2
PMCID: PMC6002841  NIHMSID: NIHMS951213  PMID: 29889786

Abstract

Hypothesis

Insertion of cochlear implant electrodes generates transient pressure spikes within the vestibular labyrinth equivalent to high intensity acoustic stimuli.

Background

Though cochlear implant (CI) surgery is regarded as having low-risk of impacting the vestibular system, several studies have documented changes in vestibular function after implantation. The mechanism of these changes is not understood. We have previously established that large, potentially-damaging pressure transients can be generated in the cochlea during electrode insertion, but whether pressure transients occur within the vestibular labyrinth has yet to be determined. Here, we quantify the exposure of the vestibular system to potentially-damaging pressure transients during CI surgery.

Methods

Five human cadaveric heads were prepared with an extended facial recess and implanted sequentially with eight different CI electrode styles via a round window approach. Fiber-optic sensors measured intralabyrinthine pressures in scala vestibuli (SV), scala tympani (ST), and the lateral semicircular canal (LSCC) during insertions.

Results

Electrode insertion produced a range of high-intensity pressure spikes simultaneously in the cochlea and LSCC with all electrodes tested. Pressure transients recorded were found to be significantly higher in the vestibular labyrinth than the cochlea and occurred at peak levels known to cause acoustic trauma.

Conclusion

Insertion of CI electrodes can produce transients in intralabyrinthine fluid pressure levels equivalent to high-intensity, impulsive acoustic stimuli. Results from this investigation affirm the importance of atraumatic surgical techniques and suggest that in addition to the cochlea, the vestibular system is potentially exposed to damaging fluid pressure waves during cochlear implantation.

Keywords: Cochlear implant, intracochlear pressures, vestibular loss, semicircular canals

Introduction

Use of cochlear implants (CI) to treat severe to profound sensorineural hearing loss has increased as excellent communication outcomes have made it the modality of choice. Though cochlear implantation is generally considered as a low-complication procedure1, recent reports have described a wide range of incidence of adverse outcomes affecting balance, with estimates as low as 0.3%14 and as high as 75% of patients5,6.

Changes in vestibular function following cochlear implantation can be evaluated with both subjective and objective measures that do not always correlate with vestibular symptoms5,711. The clinical implications of changes can be difficult to assess and the mechanism of implantation-related loss of vestibular function remains poorly understood.

A series of recent investigations1216 reported pressure changes within the cochlea during the placement of CI electrodes. These studies hypothesized that insertion-induced pressure changes can stimulate the cochlea in an “acoustic” manner, with levels at times equivalent to those known to cause acoustic trauma. Given the interconnected nature of the fluid spaces in the inner ear, it would follow that these same high level pressure changes may also propagate through the vestibular labyrinth. Though we have previously demonstrated that high-level acoustic stimuli can cause pressure changes in the vestibular system in the absence of pathologic canal dehiscence in cadaveric specimen17, it is unknown if CI insertion-related pressure changes are transmitted to the vestibular end organs. To the best of our knowledge, no intralabyrinthine measurements specifically examining electrode insertion-related pressure changes within the vestibular system have been reported previously. Here, we aim to overcome this knowledge gap and characterize intralabyrinthine pressure changes associated with CI insertions in both the cochlear and vestibular systems of cadaveric specimen.

Methods

Data was collected and analyzed from five ears in fresh-frozen whole cadaveric heads. All specimens were confirmed to have intact temporal bones and verified to have no history of middle ear disease (MD Global, Aurora, CO, USA). The use of cadaveric human tissue was in compliance with the University of Colorado Anschutz Medical Campus Institutional Biosafety Committee and Review Board (COMIRB EXEMPT #14-1464).

Temporal Bone Preparation

Temporal bone specimens were prepared as previously described by our laboratory12,1724. After overnight thawing in warm water, intact whole cadaveric heads were inspected to rule out injury or disease. A canal-wall-up mastoidectomy with an extended facial recess approach was performed. The cochlear promontory was blue-lined near the oval and round windows, as was the bone overlying the lateral semicircular canal. The specimens were suspended with a Mayfield Clamp (Integra Lifesciences Corp., Plainsboro, NJ, U.S.A.) attached to a stainless-steel baseplate. Cochleostomies into the scala tympani (ST), scala vestibuli (SV), and the lateral semicircular canal (LSCC) were created using a fine pick under a droplet of water, and fiber-optic pressure sensors (FOP-M260-ENCAP, FISO Inc., Quebec, QC, Canada) were inserted (see Fig. 1) using micromanipulators (David Kopf Instruments, Trujunga, CA, U.S.A.) rigidly mounted onto the head clamp. Probes were sealed in place with alginate dental impression material (Jeltrate; Dentsply International Inc., York, PA).

Figure 1.

Figure 1

Photomicrograph of the right ear in a single specimen during data collection. Through the extended facial recess, the anatomical landmarks (stapes and round window niche) can be visualized. Pressure probes (PSV, PST, PLSCC) were placed in cochleostomies in scala vestibuli, scala tympani, and the lateral semicircular canal. All cochleostomies were sealed with dental impression material. Implant electrode (CI) was inserted into the round window.

Baseline measurements of stapes velocity (VStap) and intralabyrinthine pressures (PIL) were made to acoustic stimuli. Out-of-plane velocities were measured with a single-axis laser Doppler vibrometer (LDV) (OFV-534 & OFV-5000; Polytec Inc., Irvine, CA) mounted to a dissecting microscope (M400; Leica Microsystems, Buffalo Grove, IL, United States). Microscopic retro-reflective glass beads (P-RETRO 45–63 μm dia., Polytec Inc., Irvine, CA) were placed on the stapes to enhance measurement of the LDV signal [41, 42].

Electrode insertions were performed primarily by resident physicians under the guidance of Neurotology faculty. Implant electrodes were inserted into the cochlea as described in previous experiments12. Cochleostomies for CI electrode insertion were made in the round window with a fine pick. Electrodes were inserted by hand underwater using an insertion tool, if provided by the manufacturer, and with a pair of fine forceps otherwise. CI electrodes used in these experiments were: Nucleus CI24RE Contour Advance (NCA; Cochlear Ltd, Sydney, Australia), Nucleus CI422 Slim Straight inserted to 20 mm (CI422; Cochlear Ltd, Sydney, Australia), Nucleus Hybrid L24 (HL24; Cochlear Ltd, Sydney, Australia), Nucleus CI532 Slim Modiolar (CI532; Cochlear Ltd, Sydney, Australia), HiFocus Mid-Scala (MS; Advanced Bionics AG, Stäfa, Switzerland), FLEX 24 (F24; Med-El, Innsbruck, Austria), FLEX 28 (F28; Med-El, Innsbruck, Austria), and Digisonic SP EVO (EVO; Oticon Medical, Vallauris, France). Electrodes were inserted one after another, in random order.

Sound Presentation, Data Acquisition, and Data Analysis

All experiments were performed in a double-walled, sound-attenuating chamber (IAC Inc., Bronx, NY). Acoustic stimuli were presented as described previously12,1724 in order to generate baseline transfer functions used to calculate estimated equivalent ear canal pressures for each insertion. Briefly, digitally generated stimuli were presented via a closed-field magnetic speaker (MF1; Tucker-Davis Technologies Inc., Alachua, FL) coupled directly to the ear with a foam insert altered to accommodate flexible speaker tubing. The speaker was driven by an external sound card (Hammerfall Multiface II, RME, Haimhausen, Germany) amplified with one channel of a stereo amplifier (TDT SA1). Sound intensity in the ear canal was measured with a probe-tube microphone (type 4182; Bruel & Kjær, Nærum, Denmark), also placed through the modified foam earplug. Baseline acoustic transfer functions were generated from presentation of short (1s duration) tone pips between 100 and 12000 Hz ramped on and off with one half (5 ms) of a Hanning window. Input from the microphone, LDV, and pressure sensors were simultaneously captured via the sound card analog inputs. The fiber-optic sensors were factory calibrated and sensitivity was verified by normalizing the sensor response to LDV displacement measurements made while generating a known motion in a cup of fluid controlled by a B&K mini-shaker (Bruel & Kjær Type 4810, Nærum, Denmark). The magnitude of the LDV signal was adjusted using a cosine correction (1/cos(θ)) based on an estimate of the difference in angle between the primary axis of the stapes and the orientation of the LDV laser (~45°). Signals acquired were band-pass filtered between 15–15000 Hz with a second order Butterworth filter.

Baseline temporal bone measurements are shown as acoustic transfer functions, which are generated by measuring the response of a system to a known stimulus. In this case, we use acoustic stimuli and characterize the response of the external, middle, and inner ears utilizing intralabyrinthine pressures. We are then able to use the resulting transfer function to generate a filter and calculate an estimate of the intensity in decibels sound pressure level (dB SPL) of acoustic stimulus in the external ear canal (EAC) that would be required to generate a given pressure measurement12. These equivalent sound pressure levels in the ear canal (dB SPL Eq) are provided as peak pressure (in Pa) observed, and in dB SPL Peak (20 * log10(peak pressure/20μPa). Differential pressure (PDiff), which is thought to be the input to the cochlear partition and the driving force for auditory transduction2529, is calculated as the difference in pressures between scala vestibuli and scala tympani. Responses were only analyzed for recordings with a signal to noise ratio greater than 6 dB, and SNR was higher than 10 dB for the majority of the recordings.

Intralabyrinthine pressure changes with CI electrode insertion were determined according to methods previously detailed12,19. Briefly, to investigate maximum pressure exposure, the absolute peak pressures were found in each recording using the findpeaks function in Matlab (R2014b; The Mathworks, Inc., Natick, MA, U.S.A.). Filter criteria required peaks to a minimum peak amplitude of the greater of either four times the standard deviation within each recording or 100 Pa. Peaks were only assessed if they occurred within 10 ms of one another across recording locations to avoid inclusion of artifacts.

Results

Acoustic closed-field transfer functions

Baseline responses of VStap and PIL were assessed after placement of intralabyrinthine pressure probes and prior to making the RW cochleostomy for CI insertion in order to verify the condition of each temporal bone. Mean (±SEM) closed-field acoustic transfer function magnitudes for specimens are shown in Figure 2. Responses are overlain onto the 95% confidence interval for stapes velocities30, and the mean and the range of responses observed for intracochlear pressures reported previously by Nakajima et al26. Responses collected were consistent with previous reports12,18,2024.

Figure 2.

Figure 2

Mean baseline stapes velocity (VStap), scala vestibuli pressure (PSV), scala tympani pressure (PST), and differential pressure (PDiff; PSV-PST) transfer function magnitudes to air-conducted stimuli. Responses recorded in specimens are shown normalized to the SPL recorded in the ear canal (PEAC) and are superimposed onto the 95% CI and range of responses (gray bands) observed previously26,30. Colored bands indicate +/− standard error of the mean.

Closed-field acoustic transfer functions were also determined for the lateral semicircular canal. Given the relatively smaller number of specimens in this study (n=5), we elected to include lateral canal measurements from previous investigation of our lab17 in the calculation of the mean lateral canal transfer function. Figure 3 shows both the individual lateral canal pressure responses for all specimens as well the mean +/− standard error of the mean.

Figure 3.

Figure 3

Individual and mean baseline and lateral semicircular canal pressure (PLSCC) transfer function magnitude to air-conducted stimuli. Responses recorded in specimens are shown normalized to the SPL recorded in the ear canal (PEAC). Data collected from individual specimens are shown as colored lines, the black line represents the mean, and gray bands indicate +/− standard error of the mean.

Peak Intralabyrinthine Pressures During CI Electrode Insertion

Figure 4 shows a summary of the absolute maximum peak pressures observed in all recordings in all specimens for all electrodes tested. Figure 4A shows the unfiltered recorded raw pressures while the maximum peak pressures in estimated EAC SPL are illustrated in Figure 4B. Estimated EAC pressures recorded during electrode insertion ranged up to169 dB SPL Peak; these levels are consistent with data previously observed in our lab12 and comparable in magnitude to high intensity acoustic stimuli.

Figure 4.

Figure 4

Summary of peak sound pressure levels observed in all specimens during all electrode insertions. Unfiltered peak intracochlear pressure measurements (A) and estimated EAC pressures (B) are shown for each pressure recording as a function of recording location. Box plots represent the median +/−25% of the range of pressures observed, whiskers show the full range of the estimated distribution, and +’s mark outliers. Significant differences between groups are indicated with asterisks (* p<0.05).

Results of a one-way ANOVA, with peak pressures serving as the dependent variable and recording location as the independent variable, revealed a significant effect of electrode location on pressure for both raw pressures and estimated EAC SPL (F=8.18, p<0.0001 and F=9.29, p<0.0001, respectively). Post hoc Tukey honest significant difference (hsd) pairwise comparisons showed significantly higher pressure in the lateral canal than scala tympani and differential pressure, as well as significantly higher pressure in scala vestibuli than tympani in unfiltered data (see Table 1). Estimated equivalent EAC pressures were significantly higher in the lateral semicircular canal than the three other recording locations (see Table 2).

Table 1.

Posthoc Tukey honest-significant difference (hsd) pairwaise comparisons for raw pressure by location

Electrode Electrode Lower limit 95% CI for true mean difference Difference between estimated group means Upper limit 95% CI for true mean difference p-value
1 (Diff) 2 (LSCC) −1415.16 −773.52 −131.89 0.0105*
1 (Diff) 3 (ST) −119.13 497.41 1113.95 0.1621
1 (Diff) 4 (SV) −815.40 −198.86 417.68 0.8409
2 (LSCC) 3 (ST) 602.28 1270.93 1939.58 <0.0001**
2 (LSCC) 4 (SV) −93.99 574.66 1243.32 0.1210
3 (ST) 4 (SV) −1340.88 −696.27 −51.65 0.0283*

Table 2.

Posthoc Tukey honest-significant difference (hsd) pairwaise comparisons for SPL by location

Electrode Electrode Lower limit 95% CI for true mean difference Difference between estimated group means Upper limit 95% CI for true mean difference p-value
1 (Diff) 2 (LSCC) −1041.61 −684.10 −326.60 <0.0001**
1 (Diff) 3 (ST) −581.22 −254.98 71.27 0.1850
1 (Diff) 4 (SV) −380.59 −54.34 271.91 0.9737
2 (LSCC) 3 (ST) 61.61 429.13 796.64 0.0144*
2 (LSCC) 4 (SV) 262.25 629.76 997.28 0.0001**
3 (ST) 4 (SV) −136.55 200.64 537.82 0.4202

A breakdown of the absolute maximum peak pressures by electrode type in shown in Figure 5. Figure 5A shows the unfiltered recorded raw pressures while the maximum peak pressures in estimated EAC SPL are illustrated in Figure 5B. In general, the presence of high peak pressure transients was reasonably consistent across electrode styles (peak EAC SPL Eq exceeded 164 dB SPL Peak in all cases); however, differences were noted in the prevalence of large transients.

Figure 5.

Figure 5

Peak sound pressure levels observed in all specimens during all electrode insertions as a function of recording location and electrode type. Unfiltered peak intracochlear pressure measurements (A) and estimated EAC pressures (B) are shown in the upper and lower panels, respectively. Box plots represent the median +/−25% of the range of pressures observed, whiskers show the full range of the estimated distribution, and +’s mark outliers. Significant differences between groups are indicated with asterisks (* p<0.05).

Peak pressures served as the dependent variable and electrode style acted as the independent variables for a one-way ANOVA for both unfiltered and estimated EAC SPL conditions. Significant differences were noted in both raw pressures and estimated EAC pressures for electrode type (F=8.31, p<0.0001 and F=13.81, p<0.0001, respectively). Post hoc testing also defined the effect of electrode type, with significant differences between electrode types illustrated in Figure 5 and Tables 3 and 4.

Table 3.

Posthoc Tukey honest-significant difference (hsd) pairwaise comparisons for unfiltered pressure by electrode pairs

Electrode Electrode Lower limit 95% CI for true mean difference Difference between estimated group means Upper limit 95% CI for true mean difference p-value
1 (NCA) 2 (CI532) −599.74 654.89 1909.51 0.7613
1 (NCA) 3 (EVO) 825.19 1847.57 2869.94 <0.0001**
1 (NCA) 4 (F24) 672.48 1705.24 2738.00 <0.0001**
1 (NCA) 5 (F28) 462.66 1690.46 2918.25 0.0008**
1 (NCA) 6 (HL24) 874.26 2195.24 3516.22 <0.0001**
1 (NCA) 7 (MS) 202.05 1487.66 2773.27 0.0107*
1 (NCA) 8 (CI422) 1165.32 2230.98 3296.64 <0.0001**
2 (CI532) 3 (EVO) 237.57 1192.68 2147.79 0.0038**
2 (CI532) 4 (F24) 84.13 1050.35 2016.57 0.0221*
2 (CI532) 5 (F28) −136.80 1035.57 2207.95 0.1295
2 (CI532) 6 (HL24) 270.72 1540.36 2809.99 0.0058**
2 (CI532) 7 (MS) −400.02 832.77 2065.57 0.4496
2 (CI532) 8 (CI422) 574.78 1576.09 2577.40 0.0001**
3 (EVO) 4 (F24) −778.53 −142.33 493.87 0.9976
3 (EVO) 5 (F28) −1076.69 −157.11 762.47 0.9996
3 (EVO) 6 (HL24) −693.06 347.67 1388.41 0.9727
3 (EVO) 7 (MS) −1355.37 −359.91 635.55 0.9579
3 (EVO) 8 (CI422) −304.91 383.41 1071.72 0.6950
4 (F24) 5 (F28) −945.89 −14.78 916.34 1.0000
4 (F24) 6 (HL24) −560.94 490.00 1540.94 0.8516
4 (F24) 7 (MS) −1223.71 −217.58 788.55 0.9980
4 (F24) 8 (CI422) −177.92 525.74 1229.39 0.3134
5 (F28) 6 (HL24) −738.34 504.78 1747.91 0.9230
5 (F28) 7 (MS) −1408.28 −202.80 1002.68 0.9996
5 (F28) 8 (CI422) −426.96 540.52 1507.99 0.6917
6 (HL24) 7 (MS) −2007.85 −707.58 592.68 0.7200

Table 4.

Posthoc Tukey honest-significant difference (hsd) pairwaise comparisons for SPL by electrode pairs

Electrode Electrode Lower limit 95% CI for true mean difference Difference between estimated group means Upper limit 95% CI for true mean difference p-value
1 (NCA) 2 (CI532) −583.16 21.02 625.19 1.0000
1 (NCA) 3 (EVO) 506.23 1015.30 1524.37 <0.0001**
1 (NCA) 4 (F24) 417.56 923.97 1430.37 <0.0001**
1 (NCA) 5 (F28) −55.62 599.79 1255.19 0.1016
1 (NCA) 6 (HL24) 344.28 1078.28 1812.28 0.0002**
1 (NCA) 7 (MS) 12.63 678.23 1343.84 0.0422*
1 (NCA) 8 (CI422) 661.53 1185.78 1710.03 <0.0001**
2 (CI532) 3 (EVO) 528.71 994.29 1459.87 <0.0001**
2 (CI532) 4 (F24) 440.28 902.95 1365.62 <0.0001**
2 (CI532) 5 (F28) −43.46 578.77 1201.00 0.0902
2 (CI532) 6 (HL24) 352.72 1057.26 1761.80 0.0001**
2 (CI532) 7 (MS) 24.25 657.22 1290.18 0.0353*
2 (CI532) 8 (CI422) 682.63 1164.77 1646.90 <0.0001**
3 (EVO) 4 (F24) −420.25 −91.34 237.58 0.9907
3 (EVO) 5 (F28) −945.89 −415.52 114.85 0.2539
3 (EVO) 6 (HL24) −561.92 62.98 687.87 1.0000
3 (EVO) 7 (MS) −880.00 −337.07 205.86 0.5636
3 (EVO) 8 (CI422) −185.29 170.48 526.25 0.8326
4 (F24) 5 (F28) −852.00 −324.18 203.64 0.5776
4 (F24) 6 (HL24) −468.42 154.31 777.04 0.9954
4 (F24) 7 (MS) −786.16 −245.73 294.70 0.8675
4 (F24) 8 (CI422) −90.14 261.81 613.77 0.3191
5 (F28) 6 (HL24) −270.44 478.49 1227.43 0.5256
5 (F28) 7 (MS) −603.59 78.45 760.49 1.0000
5 (F28) 8 (CI422) 41.04 586.00 1130.96 0.0248*
6 (HL24) 7 (MS) −1157.92 −400.05 357.83 0.7506

Discussion

Our results suggest that placement of CI electrodes into scala tympani via a round window approach can generate intralabyrinthine pressures in both the cochlea and vestibular organs comparable to high-intensity acoustic stimulation. Here, as in our previous study investigating pressures in the semicircular canals17, we confirm that pressure changes do indeed propagate from the cochlea to the vestibular labyrinth in normal temporal bones that are free of pathology, such as canal dehiscence. Estimated peak equivalent ear canal pressure was as high as 169 dB SPL, with significantly higher exposures associated with measurements made from the lateral semicircular canal. Additionally, two perimodiolar electrode styles (the NCA and the CI532) showed significantly higher pressures when compared with the remaining electrodes in both unfiltered pressures and the estimated equivalent SPL measurements. This trend for higher pressures with perimodiolar electrodes was present in our previous study12, but the difference was not significant for all perimodiolar electrodes as it is in results presented here. As previously noted, not every insertion generated a pressure spike above the equipment noise floor.

Vestibular Loss After Cochlear Implantation

Transient postoperative dizziness or imbalance is a commonly reported side effect of cochlear implant surgery1, while prolonged or persistent symptoms are more rarely described. The reported rates of postoperative changes in vestibular function varies widely, with some studies reporting little to no changes13 and others seeing effects in the majority of patients5,6. Methods for evaluation of postoperative dizziness generally fall into one of two categories: objective assessments of vestibular function, such as the videonystagmography (VNG), head impulse test (HIT), vestibular evoked myogenic potentials (VEMP), or caloric testing, and subjective testing, which includes both informal questionnaires and validated balance surveys, such as the Dizziness Handicap Index (DHI).

Postoperative loss of vestibular function as measured by postoperative changes in HIT ranges from 3.6–30%10,31, though one study showed no significant change in postoperative HIT testing3. The range of postoperative change in caloric weakness was even greater, from estimates as low as 6.3% ranging upward to 60%5,7,8,10,3235. Systematic review of caloric changes found 34% of patients demonstrated new-onset unilateral canal weakness36. Changes in VEMP, which include reduction in response amplitudes, development of unilateral weakness, or loss of response postoperatively was found to be most common, occurring in 31–60% of patients5,9,10,37. Age of patient at implantation may be an important factor as one review saw significant changes in objective testing only in patients over 60 years of age38. Another study saw no change between pre- and post-operative VEMP, caloric, or HIT, though the studied patients were all under the age of 5 at the time of implantation2.

Subjective assessment of patient dizziness with the validated DHI can confirm worsening dizziness and resultant quality of life31; however, degree of handicap is frequently low4,11 and in some cases demonstrates no postoperative change35. Direct questioning regarding the presence of postoperative dizziness and vertigo or the use of specifically-designed questionnaires places estimates of new symptoms between 24–53%5,79,32,3943, though some estimates are as low as 3%44 and as high as 75%6.

In addition to examining objective and subjective vestibular changes independently, several studies have attempted to correlate changes on balance function tests with patient-reported symptoms. Mismatch between reported symptoms of balance changes and objective testing has been seen in several investigations, where patients complaining of new dizziness show no new abnormalities on vestibular function testing and those with objective findings do not necessarily report vertigo or imbalance5,711,44. One study demonstrated a significant change in DHI associated with changes in vestibular function testing, though this was for a relatively small number of patients (n=10)31.

Possible Mechanisms for Vestibular Loss

Though many clinical studies have reviewed both subjective and objective changes in balance and vestibular function following cochlear implantation, the exact mechanism underlying new onset vestibular remains unclear. Here, we propose that pressure changes generated from implant insertion can cause a stimulus similar to acoustic trauma, which may potentially cause disruption of the delicate sensory epithelium of the inner ear. Results from this study, as well as previous investigation17, have confirmed that pressure energy is propagated from the cochlear to the vestibular labyrinths in the absence of a third window. Others have suggested that direct trauma to the epithelium from contact with the electrode or translocation through the basilar membrane and subsequent mixing of endolymph and perilymph may explain changes in vestibular neurophysiology36,45. Though not a direct measure of trauma, no association between insertion depth and vestibular function has been found46.

Several studies have examined histopathological changes of the cochlea following cochlear implantation, but few have looked at the vestibular system. When the vestibular end organs of implanted patients are examined post-mortem, patterns that can be attributed to implantation include saccular membrane distortion47,48, as well as fibrosis, calcification, and ossification48. These changes are most often seen in the saccule, which is not surprising given the anatomic proximity to the cochlear implant, but have also been noted with less frequency in the utricle and semicircular canal48. No deafferentation in the peripheral vestibular system has been seen on histologic examination47.

Multiple theoretical mechanisms have been proposed to explain immediate or acute-onset vestibular symptoms after cochlear implantation. Benign positional vertigo has also been reported in the postoperative period in a small number of patients42,49, and has been proposed to be due either to inadvertent entry of bone dust into the labyrinth or dislodging of otoconia50. Iatrogenic perilymphatic fistula has also been proposed as a potential mechanism for postoperative vertigo5,43,51, and a single case study has demonstrated this etiology52. Spreading of electric current from the implant to stimulate the otolithic organs has been described to cause dizziness is some patients who did not report preoperative sound-induced vertigo32,33,39,53.

In addition to immediate onset changes in vestibular function, a delayed-onset vertigo has been reported in a subset of patients. These individuals have vertigo spells starting more than one month postoperatively, associated with Meniere’s-like symptoms of fluctuating hearing changes and tinnitus42,51. The presence of cochlear hydrops presenting with saccular membrane abnormalities on histology has led some to hypothesize that obstruction of endolymphatic flow can cause dizziness in implanted patients47.

Source for Pressure Transients

We have previously theorized multiple sources for pressure transient generation during electrode insertion12 and multiple potential mechanical mechanisms have been previously discussed for cochlear trauma during CI electrode insertion. Basilar membrane translocation of the electrode from scala tympani to scala vestibuli would likely generate pressure changes54, though this seems unlikely to be the sole mechanism as it would likely generate only a single transient and we frequently see multiple trains of spikes within each recording. Contact with the crista fenestra or lateral wall may also be a source for implantation trauma55, but this seems unlikely to be the only mechanism given the presence of spikes with both lateral wall and perimodiolar style electrodes. The size and shape of the round window cochleostomy may also contribute to fluid build-up and release as the non-compressible cochlear fluid may face increased impedance releasing through a smaller round window cochleostomy56. While data presented here is not able to determine to what degree these mechanisms may play a role, it is likely that some combination of the above-mentioned theories represent the true source for the pressure transients and additional study is needed in this area to determine the relevance of each proposed mechanism.

Limitations

Although this investigation represents the first description of pressure measurements made in both the cochlea and vestibular end organs resulting from insertion of CI electrodes, several limitations in our methodology limit the ability to generalize results from this study. Though full cephali represent a more complete model than isolated temporal bones, cadaveric specimens may be subject to post-mortem degradation. Regulation of intracochlear pressure via the cochlear aqueduct in living humans is not present in cadaveric specimens and pressure changes from physiologic baseline cannot be verified. Additionally, while it is possible that the very nature of creating openings into the otic capsule may alter pressure dynamics within the labyrinth, it does not appear to be affected by a greater number of probe insertions (up to 4 in previous investigation)17. Two cochlear probes not only provide important acoustic baseline measurements and are necessary for the exclusion of artifacts, we think that the scala vestibuli probe may roughly estimate exposures of the otolithic organs and may provide insight into the relative potential trauma to each subsite of the vestibular system.

Additional testing on the specimens after insertions, such as imaging or histological studies, may provide additional information regarding the implant trajectory and damage to anatomical structures leading to further insight into the mechanism of this potential source of trauma, but these tools are not available for this study. The mechanism proposed here is speculative, as directly linking observed pressure transients to cellular level changes remains elusive.

Our objective in this study was to investigate a possible mechanism for vestibular trauma and not evaluate the effect of surgeon experience; however, one important consideration is that we may be overestimating potential trauma exposure given that electrode insertions were performed primarily by resident physicians. Additionally, it is possible given the repeated nature of insertions within a single specimen, anatomical damage occurred to specimen during initial insertions that affected subsequent measurements. Though this is not possible to confirm without additional anatomical dissection, we have attempted to mitigate any effect by randomizing the order of electrodes inserted. And finally, while we propose a single mechanism for postoperative changes in vestibular function, we also appreciate that studies performed to date have shown that these changes are likely to be multifactorial.

Conclusions

Patients undergoing cochlear implantation for the treatment of severe to profound hearing loss risk the development of postoperative dizziness and vertigo. Data presented here have demonstrated that pressure changes in the cochlea and lateral semicircular canal resulting implant electrode insertion may be of sufficiently high amplitude to cause injury. Our data also suggest that these pressures may be higher in the vestibular system than the cochlea, though additional data is needed to further investigate this association. This study underscores the importance of atraumatic surgical techniques to minimize likelihood of generation of transient intralabyrinthine pressure spikes and confirms that the vestibular end organs, as well as the cochlea, are at risk for pressure-related trauma from cochlear implant insertion.

Acknowledgments

Funding: RMBH was funded by NIH NIDCD T32 DC-012280.

Footnotes

Conflict of Interest Statement:

SPC is a consultant for Cochlear Corporation.

References

  • 1.Terry B, Kelt RE, Jeyakumar A. Delayed Complications After Cochlear Implantation. JAMA Otolaryngol Head Neck Surg. 2015;141(11):1012–1017. doi: 10.1001/jamaoto.2015.2154. [DOI] [PubMed] [Google Scholar]
  • 2.Ajalloueyan M, Saeedi M, Sadeghi M, Zamiri Abdollahi F. The effects of cochlear implantation on vestibular function in 1–4 years old children. Int J Pediatr Otorhinolaryngol. 2017;94:100–103. doi: 10.1016/j.ijporl.2017.01.019. [DOI] [PubMed] [Google Scholar]
  • 3.Jutila T, Aalto H, Hirvonen TP. Cochlear implantation rarely alters horizontal vestibulo-ocular reflex in motorized head impulse test. Otol Neurotol. 2013;34(1):48–52. doi: 10.1097/MAO.0b013e318277a430. [DOI] [PubMed] [Google Scholar]
  • 4.Buchman CA, Joy J, Hodges A, Telischi FF, Balkany TJ. Vestibular effects of cochlear implantation. Laryngoscope. 2004;114(10 Pt 2) Suppl 103:1–22. doi: 10.1097/00005537-200410001-00001. [DOI] [PubMed] [Google Scholar]
  • 5.Katsiari E, Balatsouras DG, Sengas J, Riga M, Korres GS, Xenelis J. Influence of cochlear implantation on the vestibular function. Eur Arch Otorhinolaryngol. 2013;270(2):489–495. doi: 10.1007/s00405-012-1950-6. [DOI] [PubMed] [Google Scholar]
  • 6.Steenerson RL, Cronin GW, Gary LB. Vertigo after cochlear implantation. Otol Neurotol. 2001;22(6):842–843. doi: 10.1097/00129492-200111000-00021. [DOI] [PubMed] [Google Scholar]
  • 7.Holinski F, Elhajzein F, Scholz G, Sedlmaier B. Vestibular dysfunction after cochlear implant in adults. HNO. 2012;60(10):880–885. doi: 10.1007/s00106-012-2526-x. [DOI] [PubMed] [Google Scholar]
  • 8.Krause E, Louza JPR, Hempel J-M, Wechtenbruch J, Rader T, Gurkov R. Effect of cochlear implantation on horizontal semicircular canal function. Eur Arch Otorhinolaryngol. 2009;266(6):811–817. doi: 10.1007/s00405-008-0815-5. [DOI] [PubMed] [Google Scholar]
  • 9.Krause E, Wechtenbruch J, Rader T, Gurkov R. Influence of cochlear implantation on sacculus function. Otolaryngol Head Neck Surg. 2009;140(1):108–113. doi: 10.1016/j.otohns.2008.10.008. [DOI] [PubMed] [Google Scholar]
  • 10.Melvin T-AN, Santina Della CC, Carey JP, Migliaccio AA. The effects of cochlear implantation on vestibular function. Otol Neurotol. 2009;30(1):87–94. doi: 10.1097/mao.0b013e31818d1cba. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Parmar A, Savage J, Wilkinson A, Hajioff D, Nunez DA, Robinson P. The role of vestibular caloric tests in cochlear implantation. Otolaryngol Head Neck Surg. 2012;147(1):127–131. doi: 10.1177/0194599812442059. [DOI] [PubMed] [Google Scholar]
  • 12.Greene NT, Mattingly JK, Banakis Hartl RM, Tollin DJ, Cass SP. Intracochlear Pressure Transients During Cochlear Implant Electrode Insertion. Otol Neurotol. 2016;37(10):1541–1548. doi: 10.1097/MAO.0000000000001232. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Mittmann M, Ernst A, Mittmann P, Todt I. Insertional depth-dependent intracochlear pressure changes in a model of cochlear implantation. Acta Otolaryngol. 2017;137(2):113–118. doi: 10.1080/00016489.2016.1219918. [DOI] [PubMed] [Google Scholar]
  • 14.Mittmann P, Mittmann M, Ernst A, Todt I. Intracochlear Pressure Changes due to 2 Electrode Types: An Artificial Model Experiment. Otolaryngol Head Neck Surg. 2017;156(4):712–716. doi: 10.1177/0194599816684104. [DOI] [PubMed] [Google Scholar]
  • 15.Todt I, Mittmann M, Ernst A, Mittmann P. Comparison of the effects of four different cochlear implant electrodes on intra-cochlear pressure in a model. Acta Otolaryngol. 2016 Sep;:1–7. doi: 10.1080/00016489.2016.1232490. [DOI] [PubMed] [Google Scholar]
  • 16.Todt I, Mittmann P, Ernst A. Intracochlear fluid pressure changes related to the insertional speed of a CI electrode. Biomed Res Int. 2014;2014:507241. doi: 10.1155/2014/507241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Maxwell AK, Banakis Hartl RM, Greene NT, et al. Semicircular Canal Pressure Changes During High-intensity Acoustic Stimulation. Otol Neurotol. 2017;38(7):1043–1051. doi: 10.1097/MAO.0000000000001456. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Farrell NF, Banakis Hartl RM, Benichoux V, Brown AD, Cass SP, Tollin DJ. Intracochlear Measurements of Interaural Time and Level Differences Conveyed by Bilateral Bone Conduction Systems. Otol Neurotol. 2017 doi: 10.1097/MAO.0000000000001556. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Banakis Hartl RM, Mattingly JK, Greene NT, Farrell NF, Gubbels SP, Tollin DJ. Drill-induced Cochlear Injury During Otologic Surgery: Intracochlear Pressure Evidence of Acoustic Trauma. Otol Neurotol. 2017;38(7):938–947. doi: 10.1097/MAO.0000000000001474. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Banakis Hartl RM, Mattingly JK, Greene NT, Jenkins HA, Cass SP, Tollin DJ. A Preliminary Investigation of the Air-Bone Gap: Changes in Intracochlear Sound Pressure With Air- and Bone-conducted Stimuli After Cochlear Implantation. Otol Neurotol. 2016;37(9):1291–1299. doi: 10.1097/MAO.0000000000001184. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Mattingly JK, Greene NT, Jenkins HA, Tollin DJ, Easter JR, Cass SP. Effects of Skin Thickness on Cochlear Input Signal Using Transcutaneous Bone Conduction Implants. Otol Neurotol. 2015;36(8):1403–1411. doi: 10.1097/MAO.0000000000000814. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Deveze A, Koka K, Tringali S, Jenkins HA, Tollin DJ. Techniques to improve the efficiency of a middle ear implant: effect of different methods of coupling to the ossicular chain. Otol Neurotol. 2013;34(1):158–166. doi: 10.1097/MAO.0b013e3182785261. [DOI] [PubMed] [Google Scholar]
  • 23.Deveze A, Koka K, Tringali S, Jenkins HA, Tollin DJ. Active middle ear implant application in case of stapes fixation: a temporal bone study. Otol Neurotol. 2010;31(7):1027–1034. doi: 10.1097/MAO.0b013e3181edb6d1. [DOI] [PubMed] [Google Scholar]
  • 24.Tringali S, Koka K, Deveze A, Holland NJ, Jenkins HA, Tollin DJ. Round window membrane implantation with an active middle ear implant: a study of the effects on the performance of round window exposure and transducer tip diameter in human cadaveric temporal bones. Audiol Neurootol. 2010;15(5):291–302. doi: 10.1159/000283006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Ravicz ME, Rosowski JJ. Inner-ear sound pressures near the base of the cochlea in chinchilla: further investigation. J Acoust Soc Am. 2013;133(4):2208–2223. doi: 10.1121/1.4792139. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Nakajima HH, Dong W, Olson ES, Merchant SN, Ravicz ME, Rosowski JJ. Differential intracochlear sound pressure measurements in normal human temporal bones. J Assoc Res Otolaryngol. 2009;10(1):23–36. doi: 10.1007/s10162-008-0150-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Voss SE, Rosowski JJ, Peake WT. Is the pressure difference between the oval and round windows the effective acoustic stimulus for the cochlea? J Acoust Soc Am. 1996;100(3):1602–1616. doi: 10.1121/1.416062. [DOI] [PubMed] [Google Scholar]
  • 28.Lynch TJ, Nedzelnitsky V, Peake WT. Input impedance of the cochlea in cat. J Acoust Soc Am. 1982;72(1):108–130. doi: 10.1121/1.387995. [DOI] [PubMed] [Google Scholar]
  • 29.Dancer A, Franke R. Intracochlear sound pressure measurements in guinea pigs. Hear Res. 1980;2(3–4):191–205. doi: 10.1016/0378-5955(80)90057-x. [DOI] [PubMed] [Google Scholar]
  • 30.Rosowski JJ, Chien W, Ravicz ME, Merchant SN. Testing a method for quantifying the output of implantable middle ear hearing devices. Audiol Neurootol. 2007;12(4):265–276. doi: 10.1159/000101474. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Batuecas-Caletrio A, Klumpp M, Santacruz-Ruiz S, Benito Gonzalez F, Gonzalez Sánchez E, Arriaga M. Vestibular function in cochlear implantation: Correlating objectiveness and subjectiveness. Laryngoscope. 2015;125(10):2371–2375. doi: 10.1002/lary.25299. [DOI] [PubMed] [Google Scholar]
  • 32.Ito J. Influence of the multichannel cochlear implant on vestibular function. Otolaryngol Head Neck Surg. 1998;118(6):900–902. doi: 10.1016/S0194-5998(98)70295-5. [DOI] [PubMed] [Google Scholar]
  • 33.Jacot E, Van Den Abbeele T, Debre HR, Wiener-Vacher SR. Vestibular impairments pre- and post-cochlear implant in children. Int J Pediatr Otorhinolaryngol. 2009;73(2):209–217. doi: 10.1016/j.ijporl.2008.10.024. [DOI] [PubMed] [Google Scholar]
  • 34.Vibert D, Häusler R, Kompis M, Vischer M. Vestibular function in patients with cochlear implantation. Acta Otolaryngol Suppl. 2001;545:29–34. doi: 10.1080/000164801750388063. [DOI] [PubMed] [Google Scholar]
  • 35.Wagner JH, Basta D, Wagner F, Seidl RO, Ernst A, Todt I. Vestibular and taste disorders after bilateral cochlear implantation. Eur Arch Otorhinolaryngol. 2010;267(12):1849–1854. doi: 10.1007/s00405-010-1320-1. [DOI] [PubMed] [Google Scholar]
  • 36.Kuang H, Haversat HH, Michaelides EM. Impairment of Caloric Function After Cochlear Implantation. J Speech Lang Hear Res. 2015;58(4):1387–1395. doi: 10.1044/2015_JSLHR-H-15-0010. [DOI] [PubMed] [Google Scholar]
  • 37.Krause E, Louza JPR, Wechtenbruch J, Gurkov R. Influence of cochlear implantation on peripheral vestibular receptor function. Otolaryngol Head Neck Surg. 2010;142(6):809–813. doi: 10.1016/j.otohns.2010.01.017. [DOI] [PubMed] [Google Scholar]
  • 38.Brey RH, Facer GW, Trine MB, Lynn SG, Peterson AM, Suman VJ. Vestibular effects associated with implantation of a multiple channel cochlear prosthesis. Am J Otol. 1995;16(4):424–430. [PubMed] [Google Scholar]
  • 39.Fina M, Skinner M, Goebel JA, Piccirillo JF, Neely JG, Black O. Vestibular dysfunction after cochlear implantation. Otol Neurotol. 2003;24(2):234–242. doi: 10.1097/00129492-200303000-00018. [DOI] [PubMed] [Google Scholar]
  • 40.Filipo R, Patrizi M, La Gamma R, D’Elia C, La Rosa G, Barbara M. Vestibular impairment and cochlear implantation. Acta Otolaryngol. 2006;126(12):1266–1274. doi: 10.1080/00016480600678789. [DOI] [PubMed] [Google Scholar]
  • 41.Krause E, Louza JPR, Hempel J-M, Wechtenbruch J, Rader T, Gurkov R. Prevalence and characteristics of preoperative balance disorders in cochlear implant candidates. Ann Otol Rhinol Laryngol. 2008;117(10):764–768. doi: 10.1177/000348940811701011. [DOI] [PubMed] [Google Scholar]
  • 42.Rah YC, Park JH, Park JH, Choi BY, Koo J-W. Dizziness and vestibular function before and after cochlear implantation. Eur Arch Otorhinolaryngol. 2016;273(11):3615–3621. doi: 10.1007/s00405-016-3988-3. [DOI] [PubMed] [Google Scholar]
  • 43.Shoman N, Ngo R, Archibald J, Pijl S, Chan S, Westerberg BD. Prevalence of new-onset vestibular symptoms following cochlear implantation. J Otolaryngol Head Neck Surg. 2008;37(3):388–394. [PubMed] [Google Scholar]
  • 44.Bonucci AS, Costa Filho OA, Mariotto LDF, Amantini RCB, de Alvarenga KF. Vestibular function in cochlear implant users. Braz J Otorhinolaryngol. 2008;74(2):273–278. doi: 10.1016/S1808-8694(15)31100-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Ribári O, Küstel M, Szirmai A, Répássy G. Cochlear implantation influences contralateral hearing and vestibular responsiveness. Acta Otolaryngol. 1999;119(2):225–228. doi: 10.1080/00016489950181710. [DOI] [PubMed] [Google Scholar]
  • 46.Louza J, Mertes L, Braun T, Gurkov R, Krause E. Influence of insertion depth in cochlear implantation on vertigo symptoms and vestibular function. Am J Otolaryngol. 2015;36(2):254–258. doi: 10.1016/j.amjoto.2014.11.007. [DOI] [PubMed] [Google Scholar]
  • 47.Handzel O, Burgess BJ, Nadol JB. Histopathology of the peripheral vestibular system after cochlear implantation in the human. Otol Neurotol. 2006;27(1):57–64. doi: 10.1097/01.mao.0000188658.36327.8f. [DOI] [PubMed] [Google Scholar]
  • 48.Tien H-C, Linthicum FH. Histopathologic changes in the vestibule after cochlear implantation. Otolaryngol Head Neck Surg. 2002;127(4):260–264. doi: 10.1067/mhn.2002.128555. [DOI] [PubMed] [Google Scholar]
  • 49.Limb CJ, Carey JP, Srireddy S, Minor LB. Auditory function in patients with surgically treated superior semicircular canal dehiscence. Otol Neurotol. 2006;27(7):969–980. doi: 10.1097/01.mao.0000235376.70492.8e. [DOI] [PubMed] [Google Scholar]
  • 50.Limb CJ, Francis HF, Lustig LR, Niparko JK, JAMMAL H. Benign positional vertigo after cochlear implantation. Otolaryngol Head Neck Surg. 2005;132(5):741–745. doi: 10.1016/j.otohns.2005.01.004. [DOI] [PubMed] [Google Scholar]
  • 51.Kubo T, Yamamoto K, Iwaki T, Doi K, Tamura M. Different forms of dizziness occurring after cochlear implant. Eur Arch Otorhinolaryngol. 2001;258(1):9–12. doi: 10.1007/pl00007519. [DOI] [PubMed] [Google Scholar]
  • 52.Kusuma S, Liou S, Haynes DS. Disequilibrium after cochlear implantation caused by a perilymph fistula. Laryngoscope. 2005;115(1):25–26. doi: 10.1097/01.mlg.0000150680.68355.cc. [DOI] [PubMed] [Google Scholar]
  • 53.Coordes A, Basta D, Götze R, et al. Sound-induced vertigo after cochlear implantation. Otol Neurotol. 2012;33(3):335–342. doi: 10.1097/MAO.0b013e318245cee3. [DOI] [PubMed] [Google Scholar]
  • 54.Schuster D, Kratchman LB, Labadie RF. Characterization of intracochlear rupture forces in fresh human cadaveric cochleae. Otol Neurotol. 2015;36(4):657–661. doi: 10.1097/MAO.0000000000000573. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Adunka O, Unkelbach MH, MacK M, Hambek M, Gstoettner W, Kiefer J. Cochlear implantation via the round window membrane minimizes trauma to cochlear structures: a histologically controlled insertion study. Acta Otolaryngol. 2004;124(7):807–812. doi: 10.1080/00016480410018179. [DOI] [PubMed] [Google Scholar]
  • 56.Mittmann P, Ernst A, Todt I. Intracochlear pressure changes due to round window opening: a model experiment. ScientificWorldJournal. 2014;2014(7):341075–341077. doi: 10.1155/2014/341075. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES