Esophageal laceration related to mechanical trauma from a General Purpose (esophageal/rectal) temperature probe introducer sheath during atrial fibrillation ablation

Tawseef Dar 1, Bharath Yarlagadda 1, Cheen Alkhatib 2, Dhanunjaya Lakkireddy 1
PMCID: PMC6006969  PMID: 29988266

Abstract

Catheter Radiofrequency ablation (RFA) for the management of atrial fibrillation (AF) can be associated with serious thermal injuries of the esophagus due to the close proximity of later. Use of Esophageal/Rectal temperature monitoring probes have become a standard practice now during these procedures in order to prevent such complications. However these probes need small introducer sheaths in order to guide them into the esophagus without coiling. Due to the small size of these sheaths, they can easily get dislodged into the trachea or esophagus and cause serious complications including mucosal lacerations.

Keywords: Ablation, Esophageal Injury, Laceration, Probes

Introduction

Pulmonary vein isolation (PVI) using trans-catheter Radiofrequency ablation (RFA) is the standard of care for the management of atrial fibrillation (AF), especially in symptomatic patients with high AF burden[1]-[3]. Although relatively safe, AF ablation is associated with some serious esophageal complications due to the proximity to the left posterior atrial wall and PV ostia. Such complications include atrioesophageal fistulae, esophageal erythema, esophageal ulcerations and periesohageal nerve injury, collectively referred to as esophageal thermal lesions/Injuries (ETL/ETI)[4]-[7]. Incidence of esophageal thermal injuries (ETI) is reported around 20% [8]. In an effort to achieve effective and long lasting electrical isolation of pulmonary veins (PV), high temperatures are usually reached during the ablation process which can conduct to the esophageal wall, thereby leading to different injuries. Interestingly, atrioesophageal fistula (AEF) usually develops after a period of 2-3 weeks after the procedure as the lesions mature, andatrioesophageal fistula is a life threatening complication[4]. Thus, strategies to prevent such complications are always actively sought by electrophysiologists (EP). Two commonly applied strategies to avert these undesirable complications during RFA are-accurate localization of the esophagus and continuous monitoring of the effects of ablation on the esophagus. The latter is achieved by use of an esophageal temperature monitoring (ETM)probe during the procedure. However, there has been a recent trend towards the use of General Purpose (GP) probes(esophageal/rectal probes) instead of conventional Esophageal Stethoscope (ES)to monitor esophageal luminal temperature (ELT) due to a notion that these probes are more sensitive. Such substitution can be wrought with complications as we describe in the case below.

Case:

68 year old male with past medical history of coronary artery disease, hypertension (HTN), hyperlipidemia, obstructive sleep apnea (OSA), and recurrent AF status post ablation presented to the hospital for repeat RFAprocedure. As part of the procedure, patient hada GP (esophageal/rectal)probe placed in his esophagus to monitor ELT. But given the flimsiness of GP(esophageal/rectal probes) probe, a 30F nasal trumpet was inserted into the mouth as a sheath to guide the temperature probe into the esophagus. After unsuccessful placement due to coiling of the probe in the oropharynx, a small 5-0 F endotracheal tube (ET)was subsequently used due to its more rigid structure and length. Successful placement was confirmed by fluoroscopy and the procedure continued. During the procedure, the temperature probe is advanced and withdrawn for precise monitoring. RFA procedure was completed successfully but it was discovered that the ET had migrated into the esophagus when the surgical drapes were removed. The tube could not be visualized with direct and video laryngoscopy. The views were also obstructed due to the significant amount of barium in the oropharynx (used for the procedure).

Fluoroscopy localized the endotracheal tube within the thoracic region. Upper gastrointestinal (UGI) endoscopy was done to retrieve the ET. The proximal end of the ET was visualized just below the cricopharygeus and had caused a mucosal injury with a partial disruption underneath the upper esophageal sphincter (Figure 1).

Figure 1. Endoscopic evidence of partial mucosal injury of the Esophagus.

Figure 1.

In the process of retrieving the ET, further mucosal injury of the esophagus incurred. Post procedure endoscopy demonstrated a deep, partial-thickness laceration of the proximal esophagus with no signs of active bleeding. Patient was kept in the hospital for an additional 24 hours and was discharged in stable condition.

Discussion:

ELT monitoring has become a standard practice during the RFA procedure for AF in an effort to avoid ETI, especially life threatening AEF. Modern RFAtechniques for AFhave achieved higher success ratesalong with a reduction in pulmonary vein stenosis, albeit with an increased risk for AEF [3].Using the esophageal temperature monitoring system, ablation is controlled by setting an upper limit to the esophageal intraluminal temperature. Ablation is stopped anytime temperature monitor hits that point and is resumed at a lower energy. The set upper limit temperature varies (39-41 C) between different institutions, clinicians and with patient characteristics [9].Off note, the utility of these temperature monitoring probes in preventing esophageal injuries is still debated with some studies even showing that the use of these probes, in the setting of RFA, are associated with esophageal injuries [10]. General consensus, however, still favors their use during catheter ablation procedures for AF. Some clinicians and institutions have taken a step further and are using GP(esophageal/rectal probes) probes, originally designed for rectal/nasopharyngeal temperature monitoring, in place of conventional ES for ELT monitoring during AF ablation procedures. The rationale provided is that these GP(esophageal/rectal probes)probes are more sensitive to temperature changes than ES. However,due to their inherent flimsiness, GP(esophageal/rectal probes)probes often need a sheath to guide them into the esophagus. Usually a 30F nasal trumpet or a small 5-0 F ET is used for such purposes (Figure 2A and 2B).Such practice can result in serious adverse events if these devices (sheaths) migrate toward the trachea or esophagus while manipulating the temperature probe, as happened in our case.

Figure 2A. Showing different types of Temperature sensors.

Figure 2A.

Figure 2B. Showing different types of Devices used as Sheaths for General Purpose (GP) probes ((esophageal/rectal probes).

Figure 2B

Conclusion:

The trend of using GP(esophageal/rectal probes)probes for esophageal monitoring should be reviewed in order to find a reasonable rationale, backed with substantial evidence, for their use. If their use is rational,then efforts should be made to design sheaths to allow safe and easy manipulation of such probes in and out of the esophagus during the procedure. Meanwhile, great caution should be exercised at present while we continue to use small ET tubes as sheaths for such probes to avoid serious adverse consequences.

References

  • 1.Haïssaguerre M, Jaïs P, Shah D C, Takahashi A, Hocini M, Quiniou G, Garrigue S, Le Mouroux A, Le Métayer P, Clémenty J. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N. Engl. J. Med. 1998 Sep 03;339 (10):659–66. doi: 10.1056/NEJM199809033391003. [DOI] [PubMed] [Google Scholar]
  • 2.Pappone C, Rosanio S, Oreto G, Tocchi M, Gugliotta F, Vicedomini G, Salvati A, Dicandia C, Mazzone P, Santinelli V, Gulletta S, Chierchia S. Circumferential radiofrequency ablation of pulmonary vein ostia: A new anatomic approach for curing atrial fibrillation. Circulation. 2000 Nov 21;102 (21):2619–28. doi: 10.1161/01.cir.102.21.2619. [DOI] [PubMed] [Google Scholar]
  • 3.Oral Hakan, Scharf Christoph, Chugh Aman, Hall Burr, Cheung Peter, Good Eric, Veerareddy Srikar, Pelosi Frank, Morady Fred. Catheter ablation for paroxysmal atrial fibrillation: segmental pulmonary vein ostial ablation versus left atrial ablation. Circulation. 2003 Nov 11;108 (19):2355–60. doi: 10.1161/01.CIR.0000095796.45180.88. [DOI] [PubMed] [Google Scholar]
  • 4.Pappone Carlo, Oral Hakan, Santinelli Vincenzo, Vicedomini Gabriele, Lang Christopher C, Manguso Francesco, Torracca Lucia, Benussi Stefano, Alfieri Ottavio, Hong Robert, Lau William, Hirata Kirk, Shikuma Neil, Hall Burr, Morady Fred. Atrio-esophageal fistula as a complication of percutaneous transcatheter ablation of atrial fibrillation. Circulation. 2004 Jun 08;109 (22):2724–6. doi: 10.1161/01.CIR.0000131866.44650.46. [DOI] [PubMed] [Google Scholar]
  • 5.Kumar Saurabh, Ling Liang-Han, Halloran Karen, Morton Joseph B, Spence Steven J, Joseph Stephen, Kistler Peter M, Sparks Paul B, Kalman Jonathan M. Esophageal hematoma after atrial fibrillation ablation: incidence, clinical features, and sequelae of esophageal injury of a different sort. Circ Arrhythm Electrophysiol. 2012 Aug 01;5 (4):701–5. doi: 10.1161/CIRCEP.112.970624. [DOI] [PubMed] [Google Scholar]
  • 6.Yamasaki Hiro, Tada Hiroshi, Sekiguchi Yukio, Igarashi Miyako, Arimoto Takanori, Machino Takeshi, Ozawa Mahito, Naruse Yoshihisa, Kuroki Kenji, Tsuneoka Hidekazu, Ito Yoko, Murakoshi Nobuyuki, Kuga Keisuke, Hyodo Ichinosuke, Aonuma Kazutaka. Prevalence and characteristics of asymptomatic excessive transmural injury after radiofrequency catheter ablation of atrial fibrillation. Heart Rhythm. 2011 Jun;8 (6):826–32. doi: 10.1016/j.hrthm.2011.01.045. [DOI] [PubMed] [Google Scholar]
  • 7.Knopp H, Halm U, Lamberts R, Knigge I, Zachäus M, Sommer P, Richter S, Bollmann A, Hindricks G, Husser D. Incidental and ablation-induced findings during upper gastrointestinal endoscopy in patients after ablation of atrial fibrillation: a retrospective study of 425 patients. Heart Rhythm. 2014 Apr;11 (4):574–8. doi: 10.1016/j.hrthm.2014.01.010. [DOI] [PubMed] [Google Scholar]
  • 8.Schmidt Martin, Nölker Georg, Marschang Harald, Gutleben Klaus-Jürgen, Schibgilla Volker, Rittger Harald, Sinha Anil-Martin, Ritscher Guido, Mayer Dirk, Brachmann Johannes, Marrouche Nassir F. Incidence of oesophageal wall injury post-pulmonary vein antrum isolation for treatment of patients with atrial fibrillation. Europace. 2008 Feb;10 (2):205–9. doi: 10.1093/europace/eun001. [DOI] [PubMed] [Google Scholar]
  • 9.Halm Ulrich, Gaspar Thomas, Zachäus Markus, Sack Stephan, Arya Arash, Piorkowski Christopher, Knigge Ingrid, Hindricks Gerhard, Husser Daniela. Thermal esophageal lesions after radiofrequency catheter ablation of left atrial arrhythmias. Am. J. Gastroenterol. 2010 Mar;105 (3):551–6. doi: 10.1038/ajg.2009.625. [DOI] [PubMed] [Google Scholar]
  • 10.Müller Patrick, Dietrich Johannes-Wolfgang, Halbfass Philipp, Abouarab Aly, Fochler Franziska, Szöllösi Atilla, Nentwich Karin, Roos Markus, Krug Joachim, Schade Anja, Mügge Andreas, Deneke Thomas. Higher incidence of esophageal lesions after ablation of atrial fibrillation related to the use of esophageal temperature probes. Heart Rhythm. 2015 Jul;12 (7):1464–9. doi: 10.1016/j.hrthm.2015.04.005. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Atrial Fibrillation are provided here courtesy of CardioFront, LLC

RESOURCES