Skip to main content
Biomicrofluidics logoLink to Biomicrofluidics
. 2018 Jun 26;12(4):042218. doi: 10.1063/1.5023041

Barriers-on-chips: Measurement of barrier function of tissues in organs-on-chips

Yusuf B Arık 1,2,1,2,, Marinke W van der Helm 2, Mathieu Odijk 2, Loes I Segerink 2, Robert Passier 1,3,1,3, Albert van den Berg 2, Andries D van der Meer 1
PMCID: PMC6019329  PMID: 30018697

Abstract

Disruption of tissue barriers formed by cells is an integral part of the pathophysiology of many diseases. Therefore, a thorough understanding of tissue barrier function is essential when studying the causes and mechanisms of disease as well as when developing novel treatments. In vitro methods play an integral role in understanding tissue barrier function, and several techniques have been developed in order to evaluate barrier integrity of cultured cell layers, from microscopy imaging of cell-cell adhesion proteins to measuring ionic currents, to flux of water or transport of molecules across cellular barriers. Unfortunately, many of the current in vitro methods suffer from not fully recapitulating the microenvironment of tissues and organs. Recently, organ-on-chip devices have emerged to overcome this challenge. Organs-on-chips are microfluidic cell culture devices with continuously perfused microchannels inhabited by living cells. Freedom of changing the design of device architecture offers the opportunity of recapitulating the in vivo physiological environment while measuring barrier function. Assessment of barriers in organs-on-chips can be challenging as they may require dedicated setups and have smaller volumes that are more sensitive to environmental conditions. But they do provide the option of continuous, non-invasive sensing of barrier quality, which enables better investigation of important aspects of pathophysiology, biological processes, and development of therapies that target barrier tissues. Here, we discuss several techniques to assess barrier function of tissues in organs-on-chips, highlighting advantages and technical challenges.

INTRODUCTION

The human body contains numerous barriers, some of which separate the internal environment from the external environment and others that separate different compartments inside the body. These barriers are found in, for example, skin, airways, brain, eye, and blood vessels, and they maintain homeostasis by regulating the interactions between the compartments that they separate. Moreover, barriers such as the blood-brain barrier (BBB), blood retinal barrier (BRB), and the pulmonary air-liquid interface (ALI) are highly selective to prevent toxins from affecting vital organs. Disruption and dysfunction of such tissues are of major importance in the pathophysiology of many human diseases (e.g., BBB disruption in multiple sclerosis, meningitis, encephalitis,1 BRB disruption in diabetic retinopathy, macular degeneration,2 ALI disruption in pulmonary edema3).

It is well known that the biophysical and biochemical tissue microenvironment in terms of blood flow, interstitial flow, tissue shape and curvature, mechanical strain, paracrine signaling, and the local interaction between various cell types all play important roles in maintaining or altering the barrier function of tissues.4–7 Current in vitro methods fail to provide this dynamic physicochemical microenvironment. Therefore, there is a strong need for advanced in vitro systems that allow the controlled and routine inclusion of a realistic tissue microenvironment when studying the barrier function of cultured cells.

Organs-on-chips are a new class of microphysiological in vitro models of human organs and tissues that rely on culturing cells in a well-controlled microenvironment that has been engineered to include key physical and biochemical parameters.5,8–14 Organs-on-chips show great promise in mimicking human tissues and organs and are being used in both fundamental and translational biomedical research. For organs-on-chips to be valuable as research tools, it is essential that the state of the cells in an organ-on-a-chip can be probed and quantified in various ways. Some of the most often measured physiological parameters in the current generation of organs-on-chips are related to tissue barrier function. Importantly, measuring permeability in organs-on-chips is associated with unique challenges that are related to their small size, low volumes, and dynamic nature. It is essential to understand these challenges and to analytically characterize the organ-on-a-chip system that is being used.

In this review, we give examples of organ-on-chip systems in which various parameters related to barrier function were routinely measured. We discuss the advantages and challenges of measuring barrier function in organ-on-a-chip systems, and we give practical pointers for avoiding the most common measurement errors. Although active receptor-mediated transport is very important in physiology and drug discovery, and organ-on-chip systems show great promise in realistically mimicking physiological expression profiles of receptors,15–17 active transportation of molecules will not be discussed in this review. The assessment of cellular active transport in vitro has been discussed elsewhere,18 and the same is true for the potential role of organs-on-chips in drug discovery.19

CELL CULTURING PLATFORMS FOR BARRIER ASSESSMENT

Prior to giving examples and information about the methods to quantify barrier integrity in organs-on-chips, the section on “Conventional cell culturing systems” gives an overview of methods in conventional in vitro models which commonly use Transwell systems. Since fundamental principles of these techniques are similar in different platforms, basics discussed below will help to understand the techniques in organ-on-a-chip platforms.

Conventional cell culturing systems

Because barriers are so important in health and disease, experimental in vitro tools that can be used to quantify and characterize the barrier function of cells and tissues are currently widely used. Most conventional techniques typically make use of a Transwell cell culture system, which relies on a tissue-culture plate with two culture compartments—the well and the insert—that are separated by a synthetic porous membrane (Fig. 1). When cells are grown on the synthetic membrane, their barrier function can be assessed by measuring various parameters. In addition to assessing barriers by imaging cell-cell junction proteins using fluorescent and electron microscopy, there are various parameters that can be measured to evaluate the barrier function of cultured cell layers: electrical resistance, mass transport, and hydraulic conductivity; all three parameters will be discussed briefly below.

FIG. 1.

FIG. 1.

Conventional barrier assessment methods. (A) TEER measurements in Transwell systems uses two electrode pairs that are submerged in different compartments, measuring resistance of the cellular monolayer seeded onto the membrane. (B) Evaluating barrier by means of transport of fluorescently labeled dextran starts with cells cultured to a monolayer (B-I), then treated with a disease stimulus to change their permeability (B-II). After that FITC-dextran is added to the insert (B-III), and samples can be collected from the bottom compartment to measure the integrity of the cell layer (B-IV). Schematics of experimental setup for measuring hydraulic conductivity (C) starts with Transwell insert sealed in a chamber. For water transport, the reservoir is lowered to create a pressure gradient across the cell layer, then the water flux across the cell monolayer is measured using bubble tracker, and the hydraulic conductivity is determined accordingly [Reproduced with permission from Li et al., Ann. Biomed. Eng. 38, 2499 (2010). Copyright 2010 Springer Nature.]

Transepithelial/endothelial electrical resistance (TEER) is one of the widely used methods for evaluating barriers; it gives an indication of the tightness of cell-cell junctions in the paracellular space by means of electrical resistance across a monolayer. For measurements, a commercially available Epithelial Voltohmmeter (EVOM) is often used which consists of a pair of legs with two pairs of electrodes. One of these legs is placed in the upper compartment whereas the other is submerged into the culture medium in the lower compartment [Fig. 1(A)]. Each of these legs contains an electrode to apply a current to the barrier, while the other electrode of the pair is used to measure the resulting voltage over the barrier. Since direct current can be damaging to the cells and electrodes, an alternating current (AC) of 10 μA amplitude with a square waveform and a relatively low frequency (typically 12.5 Hz) is applied. For analyzing TEER, first the resistance of the permeable membrane only (without cells, Rmembrane) is measured, followed by a measurement of the resistance across the cell layer on the membrane (Rtotal). The specific resistance of the cell layer (Rcells) is then calculated by subtracting the blank resistance from the total resistance [Eq. (1)] and normalizing for cell culture area (Amembrane) [Eq. (2)]20

Rcells=RtotalRmembrane, (1)
TEERcells=Rcells×Amembrane. (2)

Readings of TEER are highly dependent on the electrode positions, and careful handling while placing the electrodes is important as it might disturb the cell monolayer. In addition, having a uniform current density generated by the electrodes has an impact on TEER values. To accurately provide that, correct type of electrode systems should be chosen. For instance, classical STX2/Chopstick electrode cannot be used for a relatively large membrane (i.e., 24 mm diameter) in tissue culture inserts.21 This may result in overestimation of the TEER value, and as an alternative in this case, a better suited EndOhm chambers can be used to cover larger areas.22 TEER is a sensitive, non-invasive method and with dedicated measurement systems, it is possible to monitor live cells during various stages of growth, differentiation, or experimental treatment.

Barrier integrity of cells can also be assessed by measuring the paracellular diffusive transport of tracer compounds of various molecular weights. While TEER measures the ion flux through the barrier, studies of paracellular transport can give more detailed information about the paracellular spacing when using different tracers of defined molecular weights. Tracers are typically added to the insert, and their diffusion over the cell layer into the well is tracked over time to determine the molecular flux [Fig. 1(B)]. Tracers can be radioactively, fluorescently, or enzymatically labeled. Radiolabeling is capable of detecting subtle changes in a barrier; however, they require special handling and safety measures, and their short half-life means that they cannot be stored for long periods. Therefore, this type of labeling is not usually preferred for barrier assessment.23 On the other hand, usage of enzymatic markers (e.g., horseradish peroxidase) has been reported for macromolecular diffusion. Low amounts of enzymes can still be sensitively quantified with the addition of sufficient amount of substrate and spectroscopically measuring the product of the catalyzed reaction, but the activity of enzymes can be affected by factors such as pH, temperature, and serum constituents thereby limiting its application.24 Due to ease of handling, non-radioactive, fluorescently labeled marker polysaccharides [e.g., fluorescein isothiocyanate (FITC)-labeled dextran] or proteins are widely used for permeability assays.25 Depending on the biological application, the size of tracer compounds can vary widely [e.g., inulin (5 kDa), mannitol (182 Da), albumin (67 kDa)].20 Despite ease-of-use, fluorescent tracers sometimes lack the required sensitivity to detect subtle changes in barriers due to poor specific activity (fluorescence/mg protein) or fluorophore instability.24 In general, it should be noted that the use of any tracer compounds may interfere with the transport process under study and may affect the barrier integrity as well as rendering the tested cells unusable for further experiments.20

Quantification of paracellular diffusive transport of tracer molecules typically starts with cell seeding to a Transwell membrane [Fig. 1(B-I)] followed by treatment of the cellular monolayer with a molecule of interest which would induce a change in permeability [Fig. 1(B-II)]. After treatment, a known concentration of a labeled tracer, such as FITC-dextran, is added to the insert [Fig. 1(B-III)] and over time, its diffusion over the cell layer is measured by taking repeated samples from the bottom well [Fig. 1(B-IV)]. The concentration of labeled dextran in individual samples can be calculated by measuring fluorescence intensity with a plate reader and standardizing against a calibration curve. If the increase in concentration is linear over time (which is typically only true in the initial stage of the experiment, when the concentration gradient between insert and well is still constant), the permeability coefficient of a solute can be calculated using the following equation:

P=1CidCwdt0VwA, (3)

where the permeability coefficient P is a function of Ci, the initial concentration in the insert; (dCw/dt)0, the linear fit for the rate of increase in concentration at the start of the experiment; Vw, the volume of the well; and A, the culture area.

In order to isolate the permeability coefficient of the cell layer (Pcell) [Eq. (4)], a blank measurement in which permeability of the membrane (P0) was established according to the method above should be subtracted from the measured permeability of cells grown on the membrane (Ptotal)

1Pcell=1Ptotal1P0. (4)

In addition to aforementioned techniques, measuring the flux of water across a cellular monolayer, also known as the hydraulic conductivity of a tissue, is another method to assess the barrier function of cultured cells. When performing measurements of hydraulic conductivity, water flux is facilitated by a defined pressure gradient. In addition, hydraulic conductivity can also be used to determine the optimal transmural pressure required to prevent delamination of endothelial cells from scaffold walls, which is a common challenge in micro-vessel engineering.26–29 Hydraulic conductivity can vary similar to permeability across cellular monolayers found in different locations in the human body (e.g., endothelial cells). In vitro endothelial monolayers may be optimized to produce tighter or leakier vessels to water by using different cell material and exposing the cells to different shear stresses and pressures to model different tissues, such as the tight blood-brain barrier or the permeable kidney glomerulus.29–32 In vitro measurements of hydraulic conductivity were demonstrated by Li et al. using a Transwell system [Fig. 1(C)].33 First, cells were grown on the Transwell insert to a monolayer. The filter was then sealed within a chamber. This chamber was connected to a water reservoir by a Tygon and borosilicate glass tube [Fig. 1(C)]. A difference in hydrostatic pressure was created across the filters by adjusting the height difference between the reservoir and the fluid covering the cell layers. Flow of water across the cell layers was then measured by tracking the position of a bubble pre-inserted into the glass tube. Using the volumetric flow rate derived from the displacement of the air bubble, hydraulic conductivity (Lp) can be calculated by [Eq. (5)]

Lp=JvA×Δp, (5)

where Jv is the volumetric flow rate, A is the surface area of the Transwell insert, and Δp is the hydraulic pressure difference across the cell layer.

Organ-on-a-chip systems

When using conventional in vitro systems, the barrier function of tissues is often found to be decreased compared to the physiological in vivo situation. For example, in vivo values of barrier tightness of the blood-brain barrier have been reported to be larger than what can be achieved with simple in vitro systems.34 Since inclusion of biochemical and mechanical stimuli that the cells would normally experience in their in vivo microenvironment has such an impact on their barrier function, there is a need for advanced conventional models that incorporate such factors. Therefore, in order to meet the shortcomings of the conventional models, microfluidic organ-on-a-chip systems have been developed. These systems provide a clearly defined, well-controlled physicochemical microenvironment for cell and tissue organization. Cells are exposed not only to fluid shear stresses by perfused microchannels but also forces such as mechanical cyclic strain similar to what they would normally experience in living organs during processes such as breathing as in the case of lung-on-a-chip device reported by Huh et al.5 In another example, electric fields can be incorporated into these systems to pace contractile cells.35 As a result, organs-on-chips demonstrate functional realism that is normally not found in other in vitro systems. Despite their improved functional realism, organ-on-chips devices typically require dedicated measurement setups and present specific challenges for assessing tissue barrier function. The section on “Overview of assessing barrier integrity in organs-on-chips” provides an overview of how barrier function is typically assessed in organs-on-chip systems along with their unique challenges.

Conventional TEER measurement setups (i.e., EVOM) are mostly confined to static and macroscopic cellular environments. Therefore, they are not suitable to be used in microfluidic systems due to the small scale of the devices which makes electrode placement in close proximity to the cells impossible. This leads to variations between measurements when electrodes are not firmly secured in the same positions. Integrating the electrodes directly into an organ-on-a chip model and placing the electrodes closer to the cellular monolayer can reduce the influence of electrical resistance from the cell culture medium and the noise generated by any electrode motion. Moreover, electrodes can be scaled relative to the size of the microchannel dimensions within the system; thus, compared to conventional systems, TEER can be measured with much smaller surface areas in organ-on-chip systems. However, one needs to ensure a uniform current density across a cellular monolayer.20

As mentioned before, organs-on-chips can incorporate physiologically relevant fluid flow to study cells in conditions that resemble the in vivo situation more closely. Thus, TEER has been commonly used to evaluate functionality of several barriers including BBB, gastrointestinal tract, and pulmonary tissues.5,6,36–47 An example of such a system is the BBB-chip reported by Van der Helm et al., a multi-layered microfluidic device comprising two polydimethylsiloxane (PDMS) parts with defined microchannels, separated by a membrane made of polycarbonate, and containing 4 integrated platinum electrodes (200 μm in diameter) that are inserted into the culture channels through side channels in the PDMS (Fig. 2).44,48 For TEER measurements, a lock-in amplifier with a probe cable circuit is coupled with two of the four electrodes. A series of six resistance values is recorded using all the possible pairs of electrodes. Subsequently, Gaussian elimination is used to determine the resistance of the cellular barrier and membrane from these six resistance values. The four-electrode system enables direct isolation of barrier resistance regardless of variations in the system (e.g., temperature fluctuations and changes in medium solute concentrations) affecting the inherent resistance of the system. Resulting TEER values obtained using this method were comparable to the values obtained by conventional Transwell systems.44

FIG. 2.

FIG. 2.

Integrated electrodes for measuring TEER in a BBB-Chip. (A) Exploded view of the device with top channel (TC), membrane (M), bottom channel (BC), and platinum wire electrodes (E1, E2, E3, E4). (B) Assembled device. (C) Schematic top view of the device. (D) Cross section schematic of the device showing endothelial cells (EC) cultured on membrane M in the TC. (E) Simplified equivalent circuit of the device, showing electrodes E1-E4, resistors representing the TC (R1 and R3), resistors representing the BC (R2 and R4), and resistor Rm representing the membrane and EC barrier. [Reproduced with permission from van der Helm et al., Biosens. Bioelectron. 85, 924 (2016). Copyright 2016 Elsevier.]

Assessment of barrier quality of cells in organs-on-chips with TEER has been challenging due to various factors. In these systems, temperature and physical support for cell culture as well as the characteristics of the electrodes such as material, quality, and surface state have an influence on the TEER values.40 Non-uniform current densities are a well-established source of measurement error for TEER in any system,40,49,50 but are particularly important in organs-on-chips, due to the relatively low volume of medium in microfluidic channels resulting in high electrical resistance comparable to the cell layer resistance.40,49 In order to ensure a uniform current density and thus an equal potential drop over the entire cell culture area, one could integrate electrodes along the entire channel even though this might not be compatible with devices where mechanical deformations (e.g., stretching) are applied.40 Alternatively, correction factors can be applied when calculating TEER from raw measurements to account for non-uniform current densities.40,49 Other potential sources of measurement errors are chip-to-chip variation in positioning of the electrodes and air bubbles present in microchannels (different cross-sectional areas). In addition to these physical sources of measurement errors, variations can be caused by incomplete cell coverage, even though cells of interest in the monolayer express cell-cell junction proteins. A slight gap (0.4%) in cell coverage can potentially reduce the TEER measured by 80%.40 It is essential to control all these sources of variation to enable comparison of TEER values between different microfluidic systems.

Assessment of a cellular barrier by means of resistance provides label-free, real time information, but it does require dedicated measurement setups as well as device designs. Therefore, assessment of barriers by measuring transport of tracer molecules is used more often in organ-on-chip systems. An example of such a system is the BBB-on-chip model of Achyuta et al., which involves two cell types cultured on microchannels assembled into a chip (Fig. 3-I). In this device, barrier integrity was measured using diffusion of fluorescently labeled dextran (3 kDa), which was perfused in the vascular channel and collected at the neural layer. The amount of diffused dextran was measured using a plate reader.51 Another example of using fluorescently labeled molecules to assess barrier integrity is the lung-on-a-chip device used by Huh et al. (Fig. 3-II).5 This device is a three-layered sandwich where two adjacent channels were separated by a porous membrane. Added to the upper alveolar layer, FITC-conjugated albumin transport was measured by sampling liquid flowing through the lower channel. Similarly, Kim et al. demonstrated the diffusion of fluorescently labeled dextran added to the upper channel of a gut-on-chip device, by taking hourly samples from the bottom layer (Fig. 3-III).6 Contrary to the traditional Transwell systems, these devices are dynamic. That means culture medium is often continuously perfused through the top and bottom channels, which makes it impossible to measure an increase in concentration in the collecting channel by repeated sampling. The concentration of tracer molecules in all repeatedly collected samples will be typically be constant, and will only depend on the barrier tightness and the residence time—the time for molecules to accumulate in the fluid of the collecting compartment as it flows through the chip. The residence time is dictated by the volumetric flow rate and the volume of the collecting channel. Typically, due to continuous perfusion and the low volume of compartments, residence times are short, which means that the effective “sampling time” that can be used to estimate (dCw/dt)0 for Eq. (3) is also short. This may lead to low effective concentrations of tracers to reliably estimate the rate of increase. To overcome this issue, multiple measures can be taken: flow rates in the channels can be decreased, concentrations of tracer molecules in the source channel can be increased, and smaller tracer molecules with higher permeability coefficients can be used. Another complication of using this method in organs-on-chips is that differences in pressures or hydraulic resistances between channels, although small, might cause transport of tracer molecules through advection instead of diffusion. Therefore, one should be careful with the fluid levels in different inlets to be equal to prevent any pressure differences.

FIG. 3.

FIG. 3.

Typical organ-on-a-chip devices containing two adjacent channels separated by a semi-permeable membrane. (I) Device of Achyuta et al. that consists of 2 parts which are assembled following the cell seeding. [Reproduced with permission from Achyuta et al., Lab Chip 13, 542 (2012). Copyright 2012 The Royal Society of Chemistry.] (II) Device of Huh et al. that consists of two adjacent channels separated by a porous PDMS membrane. [Reproduced with permission from Huh et al., Science 328, 1662 (2010). Copyright 2010 American Association for the Advancement of Science.] (III) Design of Kim et al. that contains two channels, one of which seeded with gut epithelial cells, other containing the interstitial fluid. [Reproduced with permission from Kim et al., Lab Chip 12, 2165 (2012). Copyright 2012 The Royal Society of Chemistry.]

Organs-on-chips are becoming progressively more three-dimensional to mimic in vivo tissue structure and function. This means that many organs-on-chips now contain 3D vessels or networks of vessels. Assessment of barrier function in such devices with 3D culture area geometries typically relies on imaging of fluorescent tracers because measuring TEER is currently not possible due to challenges related to integration of electrodes as well as ensuring a uniform electric field along the culture area.52 An example of a device with a 3D vascular architecture is provided by Moya et al. They reported a microfluidic device with individual cell culture chambers which were filled with endothelial cells and fibrin matrix. Cells in these chambers self-assembled into a capillary network in the presence of cell media supplemented with growth factors (Fig. 4).53 Permeability of vessels could be assessed by injecting fluorescent dextran (70 and 150 kDa) to the channels followed by imaging with fluorescence microscopy. Typically, this type of microscopy data is reported to make a qualitative or semi-quantitative statement about barrier function.

FIG. 4.

FIG. 4.

Microfluidic system reported by Moya et al. for 3D vasculature modelling. (a) PDMS based microfluidic device contains outer microfluidic channels that connect to a series of central micro-tissue chambers through a communication pore on each side. (b) Central chamber is inhabited by endothelial cells and stromal cells embedded in a fibrin matrix [cross-section of panel (a) indicated by a black dotted line]. (c) Hydrostatic pressure is necessary for media flow and is enabled by large media reservoirs. (d) and (e) Microfluidic system enables robust interconnected vessel network formation within 14–21 days (scale bar= 200 μm). [Reproduced with permission from Moya et al., Tissue Eng. Part C 19, 730 (2013). Copyright 2013 Mary Ann Liebert Inc.]

Still, microscopic tracking of tracer molecules can in principle be used to make quantitative statements about barrier function in organs-on-chips with 3D geometries. For instance, Herland et al. constructed a 3D blood vessel-on-a-chip inside with lumens created by viscous finger patterning in a collagen I matrix.52 Cells seeded inside these lumens created a 3D vessel. Barrier quality in these systems was evaluated by infusing a fluorescently labeled dextran (3 kDa) followed by continuous recording of fluorescent images. Using these images, the apparent permeability coefficient (Papp) can be calculated by analyzing the total fluorescence intensity in an area and applying

Papp=1ΔIdIdt0r2, (6)

where ΔI is the step increase in the total fluorescence intensity upon adding dextran, (dI/dt)0 is the initial rate of increase in intensity as dextran diffuses out of the vessels into the surrounding gel, and r is the radius of the tube.54 For this type of measurements, it is essential to ensure a stable monolayer of cells in the beginning of the dye addition, otherwise if diffusion of fluorescently labeled molecules is too fast to reliably establish the intensity step ΔI, and quantification will not be possible.

Overview of assessing barrier integrity in organs-on-chips

Since their inception, a wide variety of organs-on-chip designs have been optimized to model various tissues of the human body. By the controlled incorporation of physiologically relevant forces, flows, and geometries that are also found in their native in vivo environment, one can better recapitulate tissue and organ level physiology. Below is an overview of different organ-on-chip systems used to assess cellular barriers of different tissues. The list includes only a small fraction of a vast number of models as the focus of this review was restricted to models investigating cellular barriers (Table I).

TABLE I.

Overview of barrier assessment techniques in organ-on-a-chip systems.

Organ Assessment Co-culture Culturing type References
Blood-brain Barrier TEER No 2D 41 and 46
Blood-brain Barrier PTT No 2D 60 and 55
Blood-brain Barrier PTT Yes 2D 56 and 67
Blood-brain Barrier PTT Yes 3D 57 and 56
Blood-brain Barrier PTT No 3D 57
Blood-brain Barrier TEER Yes 2D 58
Blood-brain Barrier TEER, PTT Yes 2D 36, 46, 59, and 60
Blood-brain Barrier TEER, PTT Yes 3D 41, 61, and 62
Cornea PTT No 2D 63
Gastrointestinal Tract PTT Yes 2D 64 and 65
Gastrointestinal Tract TEER Yes 2D 66
Gut TEER, PTT No 2D 6
Kidney TEER, PTT Yes 2D 38
Liver PTT No 2D 67
Lung TEER, PTT Yes 2D 5
Lung PTT Yes 2D 68
Retina PTT Yes 2D 69
Retina PTT Yes 3D 70
Vasculature PTT No 3D 53 and 71
Multiple organs TEER Yes 2D 72
Multiple organs TEER, PTT Yes 2D 46

Following abbreviations were used to describe each of the techniques: TEER for trans-endothelial/epithelial electrical resistance and PTT for paracellular transport of tracer molecules. Each row of Table I has been classified based on the organ of interest, barrier assessment method (whether TEER, PTT, or combination), co-culture (presence of multiple cell types in the same model), and culturing type (whether cells have been cultured in hydrogels to provide a 3D microenvironment).

As it is challenging to setup TEER for 3D culture areas due to previously stated factors, it has not been preferred in many models, and instead barrier function is more often evaluated using PTT method.56,61 Moreover, it has been reported that the cells exposed to mechanical forces exhibit increased paracellular permeability even though TEER values for the cell layer remains stable, possibly due to increased transcytosis.6 On the other hand, the assessment of barrier function by TEER has unique advantages, as it can be performed continuously, non-invasively and in a controlled atmosphere. Moreover, technical proof-of-concept studies in impedance spectroscopy have demonstrated that also in 3D cell culture, electrical signals from integrated electrodes can still provide information about, e.g., cell numbers and barrier properties in chips.73,74 Therefore, whenever possible, a combination of both methods should be applied to ensure a reliable readout on the barriers.

Future technical development of mimicking barriers in organs-on-chips

As is clear from Table I, many different barrier tissues have already been modeled with organ-on-chip technology. As organ-on-chip technology is developed further, the 3D culture configurations will become increasingly complex. This can already be observed in recent studies that focus on advanced 3D scaffolds for alveolus-on-chip75 and colon-on-chip systems.76 Obviously, this increasing complexity will also lead to challenges in measuring barrier function, e.g., with interpreting signals from electrical sensors. It will therefore be important to keep developing innovative read-outs and more sophisticated sensor technology, such as 3D biocompatible electrodes that can directly integrate in the cultured tissues.77

Most organs-on-chips are currently fabricated from PDMS, but the high gas permeability of this material makes it challenging to control specific gas pressures in an organ-on-chip system. Local control over gas concentrations is important in studying the transport of gases over barrier tissues in, for example, lung-on-a-chip or the vessel-on-a-chip systems. In addition, the barrier function of many tissues is affected by local oxygen concentrations. For example, the permeability of blood vessels changes dynamically in episodes of ischemia and reperfusion,78 and the barrier function of intestinal epithelium is affected by interactions with anaerobic bacteria that only survive in low-oxygen conditions.79 Another challenge when using PDMS-based devices is selective adsorption and absorption of molecules from the culture media, which in turn reduces their effective concentrations and ability to affect cells. This is especially important in drug efficacy or toxicology studies where compound availability to the cells is required to determine the dosage and efficiency of the drug.80 To reduce absorption, PDMS channel surfaces are often coated to block the passage of compounds.81 Alternatively, organ-on-chip systems are increasingly being manufactured with materials other than PDMS, such as polystyrene, glass, and cyclic olefin copolymer to allow control over gas pressures as well as to prevent absorption of compounds.82,83

Combined with the engineering of organ-on-chip systems that contain ever more physiologically relevant cues, in vitro barrier models will also integrate progressively more cell types. By doing so, it will be possible to mimic cellular dynamics as well as crosstalk between cell types, such as barrier tissue cells and immune cells. However, one needs to be cautious about the media compatibility when incorporating several cell types. Cells may not survive in each other's respective medium and continuous fluid flow may be needed to allow locally stable culture conditions for various cell types. Together with an increasing 3D complexity of organ-on-chip systems, this will require innovative solutions for microfluidic actuation, for example, by 3D-printing parts of organ-on-chip systems.

CONCLUSION

Barriers exist in our bodies to maintain homeostasis and protect vital organs. Disruption of tissue barriers leads to various diseases. It is undeniable that investigation of these barriers in diseases might reveal new mechanisms and treatments. Therefore, proper in vitro tools are required to evaluate the integrity and characteristics of barriers. Conventional Transwell systems suffer from not fully recapitulating the complexity of the microenvironment as well as inclusion of physical forces that have an impact on the development and differentiation of cells. Microfluidic organ-on-chip systems are great tools to overcome these challenges. In addition to integrating physical stimuli, they consist of additional co-cultures with immune cells and microbes to mimic the physiological tissue environment more realistically, thus giving more accurate information about underlying organ physiology and disease mechanisms.8 Due to their complexity and a wide plethora of designs, it can be challenging to compare measurements performed in a specific organ-on-chip system with data from other organs-on-chips or conventional assays. To overcome this challenge, it is essential to implement assays that quantitatively measure barrier function, independent of exact system design. Assessing barrier function will be of importance especially in cases where multiple organ models are combined to create body-on-chip systems, where barrier functions influenced by other organ models (e.g., via inflammation) can be studied. Incorporating such measurements in organs-on-chips may require adjustments or corrections to avoid common measurements errors, as discussed in this review.

Finally, advances in stem cell technology and access to patient-derived cells will improve the physiological relevance of current organ-on-chip models and will contribute to more realistic and patient-specific disease models. Continuous monitoring of barriers without disrupting viability of the cells in such organ-on-a-chip models will yield unique insights into mechanisms of disease, thus contributing to the development of patient-specific treatments in the context of precision medicine.

ACKNOWLEDGMENTS

We acknowledge Stichting Toegepast Wetenschappelijk Instituut voor Neuromodulatie (TWIN) project “Inflammation and Edema in an Organ-on-a-Chip Model of Wet Age-Related Macular Degeneration” for financial support of this research.

References

  • 1. Rosenberg G. A., “Neurological diseases in relation to the blood–brain barrier,” J. Cerebral Blood Flow Metabolism 32(7), 1139–1151 (2012). 10.1038/jcbfm.2011.197 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Cunha-Vaz J. et al. , “Blood-retinal barrier,” Eur. J. Ophthalmol. 21(6_suppl), 3–9 (2011) 10.5301/EJO.2010.6049. [DOI] [PubMed] [Google Scholar]
  • 3. Hurley J., “Current views on the mechanisms of pulmonary oedema,” J. Pathol. 125(2), 59–79 (1978). 10.1002/path.1711250202 [DOI] [PubMed] [Google Scholar]
  • 4. Cucullo L. et al. , “The role of shear stress in Blood-Brain Barrier endothelial physiology,” BMC Neurosci. 12(1), 40 (2011). 10.1186/1471-2202-12-40 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Huh D. et al. , “Reconstituting organ-level lung functions on a chip,” Science 328(5986), 1662–1668 (2010). 10.1126/science.1188302 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Kim H. J. et al. , “Human gut-on-a-chip inhabited by microbial flora that experiences intestinal peristalsis-like motions and flow,” Lab Chip 12(12), 2165–2174 (2012). 10.1039/c2lc40074j [DOI] [PubMed] [Google Scholar]
  • 7. Wang Y. et al. , “Formation of human colonic crypt array by application of chemical gradients across a shaped epithelial monolayer,” Cell. Mol. Gastroenterol. Hepatol. 5(2), 113–130 (2018) 10.1016/j.jcmgh.2017.10.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Bhatia S. N. and Ingber D. E., “Microfluidic organs-on-chips,” Nat. Biotechnol. 32(8), 760–772 (2014). 10.1038/nbt.2989 [DOI] [PubMed] [Google Scholar]
  • 9. Bergers L. et al. , “Skin-on-chip: Integrating skin-tissue and microsystems engineering,” Tissue Eng. Part A 21, S337 (2015). [Google Scholar]
  • 10. Du G. et al. , “Microfluidics for cell-based high throughput screening platforms—A review,” Anal. Chim. Acta 903, 36–50 (2016). 10.1016/j.aca.2015.11.023 [DOI] [PubMed] [Google Scholar]
  • 11. Hutson M. S. et al. , “Organs-on-chips as bridges for predictive toxicology,” Appl. Vitro Toxicol. 2(2), 97–102 (2016). 10.1089/aivt.2016.0003 [DOI] [Google Scholar]
  • 12. Loskill P. et al. , “WAT-on-a-chip: A physiologically relevant microfluidic system incorporating white adipose tissue,” Lab Chip 17(9), 1645–1654 (2017). 10.1039/C6LC01590E [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Marx U. et al. , “Human-on-a-chip'developments: A translational cutting-edge alternative to systemic safety assessment and efficiency evaluation of substances in laboratory animals and man?,” ATLA Altern. Lab. Anim. 40(5), 235 (2012). [DOI] [PubMed] [Google Scholar]
  • 14. van der Meer A. D. and van den Berg A., “Organs-on-chips: Breaking the in vitro impasse,” Integr. Biol. 4(5), 461–470 (2012). 10.1039/c2ib00176d [DOI] [PubMed] [Google Scholar]
  • 15. Gao D. et al. , “Characterization of drug permeability in Caco-2 monolayers by mass spectrometry on a membrane-based microfluidic device,” Lab Chip 13(5), 978–985 (2013). 10.1039/c2lc41215b [DOI] [PubMed] [Google Scholar]
  • 16. Imura Y. et al. , “Micro total bioassay system for ingested substances: Assessment of intestinal absorption, hepatic metabolism, and bioactivity,” Anal. Chem. 82(24), 9983–9988 (2010). 10.1021/ac100806x [DOI] [PubMed] [Google Scholar]
  • 17. Imura Y. et al. , “Micro total bioassay system for oral drugs: Evaluation of gastrointestinal degradation, intestinal absorption, hepatic metabolism, and bioactivity,” Anal. Sci. 28(3), 197 (2012). 10.2116/analsci.28.197 [DOI] [PubMed] [Google Scholar]
  • 18. Yang N. J. and Hinner M. J., “Getting across the cell membrane: An overview for small molecules, peptides, and proteins,” in Site-Specific Protein Labeling: Methods Protocols ( Springer, 2015), pp. 29–53. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Esch E. W. et al. , “Organs-on-chips at the frontiers of drug discovery,” Nat. Rev. Drug Discovery 14(4), 248–260 (2015). 10.1038/nrd4539 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Srinivasan B. et al. , “TEER measurement techniques for in vitro barrier model systems,” J. Lab. Autom. 20(2), 107–126 (2015). 10.1177/2211068214561025 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.See http://www.wpi-europe.com/products/cell-and-tissue/teer-measurement/evom2.aspx for Epithelial Voltohmmeter.
  • 22.See http://www.wpi-europe.com/products/cell-and-tissue/teer-measurement/endohm-24snap.aspx for EndOhm Chamber.
  • 23. Bowman P. D. et al. , “Brain microvessel endothelial cells in tissue culture: A model for study of blood‐brain barrier permeability,” Ann. Neurol. 14(4), 396–402 (1983). 10.1002/ana.410140403 [DOI] [PubMed] [Google Scholar]
  • 24. Duffy S. L. and Murphy J. T., “Colorimetric assay to quantify macromolecule diffusion across endothelial monolayers,” BioTechniques 31(3), 495–500 (2001). [DOI] [PubMed] [Google Scholar]
  • 25. Horibe Y. et al. , “Polar solute transport across the pigmented rabbit conjunctiva: Size dependence and the influence of 8-bromo cyclic adenosine monophosphate,” Pharm. Res. 14(9), 1246–1251 (1997). 10.1023/A:1012123411343 [DOI] [PubMed] [Google Scholar]
  • 26. Price G. M. et al. , “Effect of mechanical factors on the function of engineered human blood microvessels in microfluidic collagen gels,” Biomaterials 31(24), 6182–6189 (2010). 10.1016/j.biomaterials.2010.04.041 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Wong K. H. et al. , “Artificial lymphatic drainage systems for vascularized microfluidic scaffolds,” J. Biomed. Mater. Res. Part A 101A(8), 2181–2190 (2013). 10.1002/jbm.a.34524 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Zheng Y. et al. , “In vitro microvessels for the study of angiogenesis and thrombosis,” Proc. Natl. Acad. Sci. 109(24), 9342–9347 (2012). 10.1073/pnas.1201240109 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Bogorad M. I. et al. , “in vitro microvessel models,” Lab Chip 15(22), 4242–4255 (2015). 10.1039/C5LC00832H [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Clough G. and Michel C., “Quantitative comparisons of hydraulic permeability and endothelial intercellular cleft dimensions in single frog capillaries,” J. Physiol. 405(1), 563–576 (1988). 10.1113/jphysiol.1988.sp017348 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Fraser P. and Dallas A., “Permeability of disrupted cerebral microvessels in the frog,” J. Physiol. 461(1), 619–632 (1993). 10.1113/jphysiol.1993.sp019532 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Sill H. W. et al. , “Shear stress increases hydraulic conductivity of cultured endothelial monolayers,” Am. J. Physiol. -Heart Circ. Physiol. 268(2), H535–H543 (1995). 10.1152/ajpheart.1995.268.2.H535 [DOI] [PubMed] [Google Scholar]
  • 33. Li G. et al. , “Permeability of endothelial and astrocyte cocultures: In vitro blood–brain barrier models for drug delivery studies,” Ann. Biomed. Eng. 38(8), 2499–2511 (2010). 10.1007/s10439-010-0023-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Butt A. M. et al. , “Electrical resistance across the blood‐brain barrier in anaesthetized rats: A developmental study,” J. physiology 429(1), 47–62 (1990). 10.1113/jphysiol.1990.sp018243 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Agarwal A. et al. , “Microfluidic heart on a chip for higher throughput pharmacological studies,” Lab a Chip 13(18), 3599–3608 (2013). 10.1039/c3lc50350j [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Booth R. and Kim H., “Characterization of a microfluidic in vitro model of the blood-brain barrier (μBBB),” Lab a chip 12(10), 1784–1792 (2012). 10.1039/c2lc40094d [DOI] [PubMed] [Google Scholar]
  • 37. Douville N. J. et al. , “Fabrication of two-layered channel system with embedded electrodes to measure resistance across epithelial and endothelial barriers,” Anal. Chem. 82(6), 2505–2511 (2010). 10.1021/ac9029345 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Ferrell N. et al. , “A microfluidic bioreactor with integrated transepithelial electrical resistance (TEER) measurement electrodes for evaluation of renal epithelial cells,” Biotechnol. Bioeng. 107(4), 707–716 (2010). 10.1002/bit.22835 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Griep L. et al. , “BBB on chip: Microfluidic platform to mechanically and biochemically modulate blood-brain barrier function,” Biomed. Microdevices 15(1), 145–150 (2013). 10.1007/s10544-012-9699-7 [DOI] [PubMed] [Google Scholar]
  • 40. Odijk M. et al. , “Measuring direct current trans-epithelial electrical resistance in organ-on-a-chip microsystems,” Lab Chip 15(3), 745–752 (2015). 10.1039/C4LC01219D [DOI] [PubMed] [Google Scholar]
  • 41. Partyka P. P. et al. , “Mechanical stress regulates transport in a compliant 3D model of the blood-brain barrier,” Biomaterials 115, 30–39 (2017). 10.1016/j.biomaterials.2016.11.012 [DOI] [PubMed] [Google Scholar]
  • 42. Sbrana T. et al. , “Dual flow bioreactor with ultrathin microporous TEER sensing membrane for evaluation of nanoparticle toxicity,” Sens. Actuators, B 223, 440–446 (2016). 10.1016/j.snb.2015.09.078 [DOI] [Google Scholar]
  • 43. Shah P. et al. , “A microfluidics-based in vitro model of the gastrointestinal human–microbe interface,” Nat. Commun. 7, 11535 (2016). 10.1038/ncomms11535 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. van der Helm M. W. et al. , “Direct quantification of transendothelial electrical resistance in organs-on-chips,” Biosens. Bioelectron. 85, 924–929 (2016). 10.1016/j.bios.2016.06.014 [DOI] [PubMed] [Google Scholar]
  • 45. Vogel P. A. et al. , “Microfluidic transendothelial electrical resistance measurement device that enables blood flow and postgrowth experiments,” Anal. Chem. 83(11), 4296–4301 (2011). 10.1021/ac2004746 [DOI] [PubMed] [Google Scholar]
  • 46. Walter F. R. et al. , “A versatile lab-on-a-chip tool for modeling biological barriers,” Sens. Actuators, B 222, 1209–1219 (2016). 10.1016/j.snb.2015.07.110 [DOI] [Google Scholar]
  • 47. Wang Y. I. et al. , “Microfluidic blood–brain barrier model provides in vivo‐like barrier properties for drug permeability screening,” Biotechnol. Bioeng. 114(1), 184–194 (2017). 10.1002/bit.26045 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. van der Helm M. et al. , “Fabrication and validation of an organ-on-chip system with integrated electrodes to directly quantify transendothelial electrical resistance,” J. Visualized Exp. 127, e56334 (2017) 10.3791/56334. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49. Yeste J. et al. , “Geometric correction factor for transepithelial electrical resistance measurements in transwell and microfluidic cell cultures,” J. Phys. D: Appl. Phys. 49(37), 375401 (2016). 10.1088/0022-3727/49/37/375401 [DOI] [Google Scholar]
  • 50. Jovov B. et al. , “A spectroscopic method for assessing confluence of epithelial cell cultures,” Am. J. Physiol. Cell Physiol. 261(6), C1196–C1203 (1991). 10.1152/ajpcell.1991.261.6.C1196 [DOI] [PubMed] [Google Scholar]
  • 51. Achyuta A. K. H. et al. , “A modular approach to create a neurovascular unit-on-a-chip,” Lab Chip 13(4), 542–553 (2013). 10.1039/C2LC41033H [DOI] [PubMed] [Google Scholar]
  • 52. Herland A. et al. , “Distinct contributions of astrocytes and pericytes to neuroinflammation identified in a 3D human blood-brain barrier on a chip,” PLoS One 11(3), e0150360 (2016). 10.1371/journal.pone.0150360 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53. Moya M. L. et al. , “In vitro perfused human capillary networks,” Tissue Eng. Part C 19(9), 730–737 (2013). 10.1089/ten.tec.2012.0430 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Huxley V. et al. , “Quantitative fluorescence microscopy on single capillaries: Alpha-lactalbumin transport,” Am. J. Physiol. -Heart Circ. Physiol. 252(1), H188–H197 (1987). 10.1152/ajpheart.1987.252.1.H188 [DOI] [PubMed] [Google Scholar]
  • 55. Prabhakarpandian B. et al. , “SyM-BBB: A microfluidic blood brain barrier model,” Lab Chip 13(6), 1093–1101 (2013). 10.1039/c2lc41208j [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56. Adriani G. et al. , “A 3D neurovascular microfluidic model consisting of neurons, astrocytes and cerebral endothelial cells as a blood–brain barrier,” Lab Chip 17(3), 448–459 (2017). 10.1039/C6LC00638H [DOI] [PubMed] [Google Scholar]
  • 57. Cho H. et al. , “Three-dimensional blood-brain barrier model for in vitro studies of neurovascular pathology,” Sci. Rep. 5, 15222 (2015) 10.1038/srep15222. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58. Booth R. and Kim H., “Permeability analysis of neuroactive drugs through a dynamic microfluidic in vitro blood–brain barrier model,” Ann. Biomed. Eng. 42(12), 2379–2391 (2014). 10.1007/s10439-014-1086-5 [DOI] [PubMed] [Google Scholar]
  • 59. Deosarkar S. P. et al. , “A novel dynamic neonatal blood-brain barrier on a chip,” PloS One 10(11), e0142725 (2015). 10.1371/journal.pone.0142725 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60. Wang J. D. et al. , “Organization of endothelial cells, pericytes, and astrocytes into a 3D microfluidic in vitro model of the blood–brain Barrier,” Mol. Pharm. 13(3), 895–906 (2016). 10.1021/acs.molpharmaceut.5b00805 [DOI] [PubMed] [Google Scholar]
  • 61. Brown J. A. et al. , “Recreating blood-brain barrier physiology and structure on chip: A novel neurovascular microfluidic bioreactor,” Biomicrofluidics 9(5), 054124 (2015). 10.1063/1.4934713 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62. Xu H. et al. , “A dynamic in vivo-like organotypic blood-brain barrier model to probe metastatic brain tumors,” Sci. Rep. 6, 36670 (2016). 10.1038/srep36670 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63. Puleo C. M. et al. , “Integration and application of vitrified collagen in multilayered microfluidic devices for corneal microtissue culture,” Lab Chip 9(22), 3221–3227 (2009). 10.1039/b908332d [DOI] [PubMed] [Google Scholar]
  • 64. Esch M. B. et al. , “Body-on-a-chip simulation with gastrointestinal tract and liver tissues suggests that ingested nanoparticles have the potential to cause liver injury,” Lab Chip 14(16), 3081–3092 (2014). 10.1039/C4LC00371C [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65. Ramadan Q. et al. , “NutriChip: Nutrition analysis meets microfluidics,” Lab Chip 13(2), 196–203 (2013). 10.1039/C2LC40845G [DOI] [PubMed] [Google Scholar]
  • 66. Mahler G. J. et al. , “Characterization of a gastrointestinal tract microscale cell culture analog used to predict drug toxicity,” Biotechnol. Bioeng. 104(1), 193–205 (2009). 10.1002/bit.22366 [DOI] [PubMed] [Google Scholar]
  • 67. Lee P. J. et al. , “An artificial liver sinusoid with a microfluidic endothelial‐like barrier for primary hepatocyte culture,” Biotechnol. Bioeng. 97(5), 1340–1346 (2007). 10.1002/bit.21360 [DOI] [PubMed] [Google Scholar]
  • 68. Huh D. et al. , “A human disease model of drug toxicity–induced pulmonary edema in a lung-on-a-chip microdevice,” Sci. Transl. Med. 4(159), 159ra147–159ra147 (2012). 10.1126/scitranslmed.3004249 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69. Chen L.-J. et al. , “Microfluidic co-cultures of retinal pigment epithelial cells and vascular endothelial cells to investigate choroidal angiogenesis,” Sci. Rep. 7, 3538 (2017) 10.1038/s41598-017-03788-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70. Chung M. et al. , “Wet‐AMD on a chip: Modeling outer blood‐retinal barrier in vitro,” Adv. Healthcare Mater. 7, 1700028 (2017) 10.1002/adhm.201700028. [DOI] [PubMed] [Google Scholar]
  • 71. Kim J. A. et al. , “Collagen-based brain microvasculature model in vitro using three-dimensional printed template,” Biomicrofluidics 9(2), 024115 (2015). 10.1063/1.4917508 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72. Maschmeyer I. et al. , “A four-organ-chip for interconnected long-term co-culture of human intestine, liver, skin and kidney equivalents,” Lab Chip 15(12), 2688–2699 (2015). 10.1039/C5LC00392J [DOI] [PubMed] [Google Scholar]
  • 73. Henry O. Y. et al. , “Organs-on-chips with integrated electrodes for trans-epithelial electrical resistance (TEER) measurements of human epithelial barrier function,” Lab Chip 17(13), 2264–2271 (2017). 10.1039/C7LC00155J [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74. Lei K. F., “Review on impedance detection of cellular responses in micro/nano environment,” Micromachines 5(1), 1–12 (2014). 10.3390/mi5010001 [DOI] [Google Scholar]
  • 75. Jain A. et al. , “Primary human lung alveolus‐on‐a‐chip model of intravascular thrombosis for assessment of therapeutics,” Clin. Pharmacol. Ther. 103(2), 332–340 (2018). 10.1002/cpt.742 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76. Wang Y. et al. , “Capture and 3D culture of colonic crypts and colonoids in a microarray platform,” Lab Chip 13(23), 4625–4634 (2013). 10.1039/c3lc50813g [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77. Inal S. et al. , “Conducting polymer scaffolds for hosting and monitoring 3D cell culture,” Adv. Biosyst. 1(6), 1700052 (2017) 10.1002/adbi.201770038. [DOI] [Google Scholar]
  • 78. Carden D. L. and Granger D. N., “Pathophysiology of ischaemia–reperfusion injury,” J. Pathol. 190(3), 255–266 (2000). [DOI] [PubMed] [Google Scholar]
  • 79. Ulluwishewa D. et al. , “Live Faecalibacterium prausnitzii in an apical anaerobic model of the intestinal epithelial barrier,” Cell. Microbiol. 17(2), 226–240 (2015). 10.1111/cmi.12360 [DOI] [PubMed] [Google Scholar]
  • 80. van Meer B. et al. , “Small molecule absorption by PDMS in the context of drug response bioassays,” Biochem. Biophys. Res. Commun. 482(2), 323–328 (2017). 10.1016/j.bbrc.2016.11.062 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 81. Sasaki H. et al. , “Parylene-coating in PDMS microfluidic channels prevents the absorption of fluorescent dyes,” Sens. Actuators, B 150(1), 478–482 (2010). 10.1016/j.snb.2010.07.021 [DOI] [Google Scholar]
  • 82. Fredrickson C. K. et al. , “Effects of fabrication process parameters on the properties of cyclic olefin copolymer microfluidic devices,” J. Microelectromech. Syst. 15(5), 1060–1068 (2006). 10.1109/JMEMS.2006.880352 [DOI] [Google Scholar]
  • 83. Ren K. et al. , “Materials for microfluidic chip fabrication,” Acc. chem. Res. 46(11), 2396–2406 (2013). 10.1021/ar300314s [DOI] [PubMed] [Google Scholar]

Articles from Biomicrofluidics are provided here courtesy of American Institute of Physics

RESOURCES