Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2019 Jun 1.
Published in final edited form as: Proc Inst Mech Eng H. 2018 Apr 15;232(6):545–552. doi: 10.1177/0954411918770700

Finite element evaluation of the newest ISO testing standard for polyethylene TKR liners

Steven P Mell 1, Spencer Fullam 1, Markus A Wimmer 1, Hannah J Lundberg 1
PMCID: PMC6022396  NIHMSID: NIHMS962963  PMID: 29658386

Abstract

Current treatment for end-stage osteoarthritis is a total knee replacement. Given that the number of total knee replacement surgeries is expected to approach 3.48 million by 2030, understanding long term failure is important. One of the preclinical tests for total knee replacements is carried out using mechanical wear testing under generic walking conditions. Used for this purpose is the International Standards Organization’s generic walking profile. Recently this standard was updated by reversing the direction of anterior/posterior translation and internal/external rotation. The effects of this change have not been investigated and therefore it is unknown if comparisons between wear tests utilizing the old and new version of the standard are valid. In this study, we used a finite element model along with a frictional energy based wear model to compare the kinematic inputs, contact conditions, and wear from the older and newer versions of the ISO standard. Simulator tested components were used to validate the computational model. We found that there were no visible similarities in the contact conditions between the old and new versions of the standard. The new version of the standard had a lower wear rate, but covered a larger portion of the articular surface. Locations of wear also varied considerably. The results of the study suggest that major differences between the old and new standard exist and therefore historical wear results should be compared with caution to newly obtained results.

Keywords: Total Knee Arthroplasty, Standardized Testing, Finite Element Modelling/Analysis [Medical], Wear Analysis/Testing [Biomechanics], Knee Prostheses, Orthopaedic Materials

Introduction

Osteoarthritis of the knee is one of the leading causes of disability globally [1]. End-stage treatment for knee arthritis is a total knee replacement (TKR). The number of TKR surgeries is expected to approach 3.48 million by 2030 [2], making study of how and why these implants fail important. One of the leading causes of long term failure of TKRs is osteolysis caused by polyethylene wear particles [37]. This can lead to implant loosening, pain, and the eventual failure of the implant. Once an implant has failed, there is a need to perform revision surgery, which can be both costly and risky for the patient, and challenging for the surgeon [8]. To estimate how long TKRs will last before they are implanted, TKRs undergo preclinical wear testing. The gold standard for TKR preclinical wear testing is mechanical wear testing using generic standard walking profiles [9]. Currently used for this purpose is the ISO standard for TKR wear testing (ISO 14243-3) [10]. Wear testing using standardized protocols is necessary for FDA approval, and for comparing older designs to new ones.

Recently, there was a change in the ISO standard for knee testing (ISO 14243-3): the direction of motion in the anterior/posterior (AP) and internal/external (IE) rotation directions were reversed. The International Organization for Standardization does not provide any reasoning for the change. However, these modifications followed publications that compared motion analysis studies of TKR subjects during walking to the force and displacement control ISO standards [11, 12]. Studies had shown that contrary to ISO(2004), patients exhibited external rotation of the tibia during swing phase [11, 12]. A study examining how the natural knee responded to force control ISO standard at the time showed anterior tibial translation during swing phase [13]. These modifications may affect volumetric wear, wear area, and contact conditions, all of which may lead to different damage mechanisms. The effects of this change have not been investigated and therefore it is unknown if comparisons between wear tests utilizing the old and new standard are valid.

In the present study, we used a finite element analysis (FEA) model of a TKR, along with a novel frictional energy based wear model [14], to investigate differences between ISO 14243-3(2004)[15] and ISO 14243-3(2014)[10]. Kinematic inputs, contact conditions, and wear from the older and newer versions of the ISO standard were compared, and wear predicted by the FEA model was validated against simulator tested components.

Methods

Development of finite element wear model

A FEA model was created from CAD models of a left sided NexGen Cruciate Retaining TKR (Zimmer, Warsaw, IN). 78,108 linear hexahedral elements (C3D8R) were used to model the ultra-high molecular weight polyethylene (UHMWPE) component. The cobalt chromium molybdenum femoral component was modeled as a rigid surface, utilizing a mix of 2nd order quadrilaterals and tetrahedrons (4080 SFM3D8, 547 SFM3D6) (Figure 1). The UHMWPE component was modeled using the J-2 Plasticity model [16] with a density of 9.4E-7 kg/mm3 and an elastic modulus of 1,051 MPa. Parameters used in the J-2 Plasticity model can be found in Table 1. Contact between the UHMWPE insert and femoral component was modeled as penalty contact, with the friction coefficient set to 0.04 on the articular surface [1719]. IE rotation, and AP translation were applied to the UHMWPE insert as velocity boundary conditions at a reference point located at the center of rotation as defined during simulator testing. Medial/lateral (ML) translation and abduction/adduction were left free, as was the case in our experimental testing. Flexion/extension was applied to the femoral component which was otherwise fixed. Axial force was applied through the UHMWPE reference point and offset 5 mm medially per ISO 14243-3. A summary of the boundary conditions applied to the model can be found in (Figure 1). FEA models were run using ABAQUS v2016/Standard (Dassault Systèmes, Waltham, MA). The loads and motions specified by simulator standard ISO 14243-3(2014) were compared to ISO 14243-3(2004) (Figure 2). Terms “ISO(2014)” and “ISO(2004)” will be used throughout to indicate ISO 14243-3(2014) and ISO 14243-3(2004), respectively.

Figure 1.

Figure 1

Summary of the model and associated boundary conditions.

Table 1.

Parameters for the J-2 Plasticity Model

Stress (MPa) Strain
12.1 0.00
21.4 0.03
23.8 0.11
44.0 0.55
92.4 0.98
135.0 1.09
515.0 1.34

Figure 2.

Figure 2

Model inputs for ISO(2014) and ISO(2004). Blue is ISO(2014) and orange is ISO(2004). Red is inputs shared by both versions of the standard.

Wear was predicted following the concept of a previously published wear model based on energy loss due to friction [14] (Equation 1). Newly, the model was applied to the FEA mesh using the Fortran user subroutine UMESHMOTION available from within ABAQUS. The wear model utilizes a parameter, unit work, defined as the frictional energy required to remove a unit volume of material both parallel to and perpendicular to the primary fibril orientation. Values for unit work parameters (in J/mm3) were set as: δWorkx = 3.547E07, δWorky = 3.855E08. All symbols defined in Equation 1 are listed in Table 2.

Table 2.

List of symbols for the wear model (Equations 1 and 2).

SD Sliding Direction
V Sliding Velocity
Ff Tangential Force
θ Primary Fiber Orientation
A Area
t Time
δWorkx Unit work in X
δWorky Unit work in Y
WearDepthcycle=StarttimeEndtimeFf(t)V(t)cos2(|θ-SD|)δWorkx+Ff(t)V(t)sin2(|θ-SD|)δWorkyΔtΔA 1)

As the wear model does not consider UHMWPE fibril mobility, we initialize each node as having an initial fibril direction. This means that UHMWPE modeled by a particular node is assumed to be aligned in a particular direction at the start of the wear simulation. We calculate this assumed direction by running the FEA model through an ISO cycle without modeling wear, and calculating the total frictional work at the UHMWPE contact surface in both the x and y directions. The initial direction is assigned as the angle between the total work done in each direction, or the resultant direction of total frictional work (Equation 2, symbols in Table 2).

θinitial=tan-1Ff,yVyFf,xVx 2)

To simulate wear in the FEA model (Figure 3), first, for each contact node, ABAQUS outputs the nodal quantities for tangential contact forces (Ff) and sliding velocity (V). Angle (θ) is calculated as the difference between the tangential contact force direction, and the primary fibril direction defined in the model. Using an ABAQUS user subroutine, for every time increment, wear is calculated at each node using Equation 1. Wear is scaled by 500,000 cycles. The wear step is then repeated for a total of 1 million cycles. The number of mesh updates necessary was determined through a convergence study and is comparable to the values utilized in similar models [20].

Figure 3.

Figure 3

Process for wear prediction. t is the total time within a single gait cycle and n is the number of gait cycles simulated.

Mesh convergence was determined by comparing the contact pressure to the contact pressure error indicators available in ABAQUS/Standard. Local contact pressure error of an order of magnitude less than reported contact pressure was deemed acceptable as per guidelines by the developers [21].

Validation of wear model

Preliminary validation of the FEA model was carried out by comparing gravimetric wear from the simulator tested components to wear predicted by the FEA model for the ISO(2004) profile. A four-station knee wear simulator (Endolab, Rosenheim, Germany) was used to test tibial liners (NexGen CR, Zimmer, Inc., Warsaw, IN). In three stations, tibial liners were tested under ISO(2004) conditions in displacement control mode. In one station, the tibial liner acted as a loaded soak control and did not undergo displacement. The applied degrees of freedom were the same as described for the FEA model (Figure 1). In each of the four stations, the lubricant, tibial liner, tibial baseplate, and femoral component were contained in a sealed, temperature controlled (37° C) chamber. Lubricant consisted of bovine serum (Life Technologies Corporation, Carlsbad, California) diluted with saline to a protein content of 30 g/L and pH of 7.6. Ethylenediaminetetraacetic acid (EDTA) was added at 200 mg/L. Tests were conducted at 1.0 Hz for 4 million cycles. Testing was stopped every 0.5 million cycles for cleaning and weighing of the tibial liners. Before and after test completion, the articular surfaces of the tibial liners were digitized using a coordinate measuring machine (SmartScope, OGP, NY). A previously described autonomous mathematical reconstruction method [22] was used to construct wear contours on the surfaces of the tibial components from the pre- and post-test digitized surfaces for comparison to the FEA model. The simulator-produced gravimetric wear was compared to FEA wear by examining the change in volume of the entire mesh over the course of the FEA simulation and converted to gravimetric wear using the density for UHMWPE used in the model.

Results

Finite element predicted contact conditions and wear

There were no visible similarities between ISO(2004) and ISO(2014) contact pressure contours at most time points during the gait cycle. For example, at 15% and 50% of the gait cycle, the location of contact pressure contours for ISO(2004) are farther anterior than ISO(2014) on both the medial and lateral sides, and the lateral side contact pressure contours are more posterior than the medial side (Figure 4). The contact pressure contours for ISO(2014) during the two time points are on the posterior side of the tibial plateau, with both the medial and lateral compartment contact pressure contours parallel. At other points in the gait cycle, however, the contours are almost the same (for example 0%, 45%, and 100%). Overall, contact for ISO(2014) spends more time on the posterior side of the plateau, which would be expected from the input waveforms.

Figure 4.

Figure 4

Contact pressure comparison between ISO(2004) and ISO(2014) at key points of the gait cycle. There were no obvious correlations between the old and new standard.

FEA predicted wear scars for ISO(2014) inputs covered a larger area of the articular surface and were longer than for ISO(2004) inputs (Figure 5, Table 3). Locations of higher wear predicted by the wear model varied considerably between ISO(2004) and ISO(2014), with high wear in the ISO(2004) predicted wear scar located more anterior on the tibial component, but more posterior for ISO(2014). The angle between the medial and lateral wear scar centroids (counter-clockwise is external rotation) was also different, with the ISO(2014) wear scars internally rotated compared to the ISO(2004) wear scars. FEA predicted wear rates showed the older standard having a higher wear rate (8.66 mg/million cycles) while the newer standard had a lower wear rate (4.45 mg/million cycles).

Figure 5.

Figure 5

Contact pressure comparison between ISO(2004) and ISO(2014) at key points of the gait cycle. There were no obvious correlations between the old and new standard.

Table 3.

Wear scar attributes for ISO(2004) and ISO(2014) from FEA. Coordinate system for centroids is located at the bottom edge centered medially/laterally. The angle between the centroids has counter clockwise as positive.

Area (mm2) ML Centroid (mm) AP Centroid (mm) Angle (°) Perimeter (mm) AP Length (mm)
ISO(2004) Medial 344.8 22.91 9.99 NA 75.61 14.49
Lateral 300.5 −23.56 3.45 NA 78.87 22.66

Total 645.3 8.02

ISO(2014) Medial 386.7 22.95 0.43 NA 84.63 22.64
Lateral 356.3 −23.80 5.25 NA 91.11 26.18

Total 742.9 −5.88

% Difference from ISO(2004) Medial 12.13 0.17 −95.70 NA 11.93 56.25
Lateral 18.59 1.02 52.17 NA 15.52 15.53

Total 15.13 −173.3

Model validation

FEA wear scar predictions compared well with the simulator (Figure 6). There is a large variation present in the simulator tested components which the FEA wear scar predictions fall within. The location of highest wear in both the simulator tested components and the FEA tested components is on the medial side of the medial compartment, while the lateral compartment wear scar is smaller with less intense wear visible. Both simulator components and FEA simulations had medial wear scars located farther anterior than the lateral wear scars (externally rotated). Examining gravimetric wear, the wear of the simulator components was 5.14±2.18 mg/million cycles. FEA predicted wear for the 4 million cycle test was 8.66 mg/million cycles.

Figure 6.

Figure 6

Wear scar comparison of 3 simulator tested components to a FEA predicted wear scar after 4 million simulated cycles.

Discussion

The objective of this study was to compare the kinematic inputs, contact conditions, and wear from the ISO(2014) against ISO(2004). In ISO(2014), both the IE and AP inputs were reversed. This did not result in wear being mirrored medially/laterally or anteriorly/posteriorly as might be expected. Instead, there was no discernable relationship between wear scars or contact pressure generated using ISO(2004) or ISO(2014) inputs. ISO(2014) produced wear scars that were both longer and covered a larger area as compared ISO(2004). ISO(2004) inputs created wear scars with more localized wear and smaller wear scars. ISO(2014) had lower wear rates despite the larger wear scar. The wear rates produced during the simulator test for ISO(2004) were reasonably close to the wear rates for the ISO(2004) FEA predictions. Wear scars between the simulator tests and the FEA predictions were similar, with several features in common: (1) areas of high wear on the medial edge of the medial side were present on both the simulator components and the predictions; (2) the medial side wear scar was more anterior than the lateral side wear scar in both the FEA predictions and the simulator tests.

A limitation of the current study is that the volumetric wear calculation of the wear model has not been fully validated. While the model duplicated the distribution of wear (wear scars) seen in the simulator tested components well, the ISO(2004) wear rate was overpredicted. The range of errors in published computational models of wear of TKRs can be quite large, with the predictive power (R2) ranging from 0.12 (Archard) to 0.60 [23]. Another more recently published study [24] had errors as large as approximately 3.5 mm3/MC in both directions, very similar to the difference noted in our study. The error in the model can be accounted for by several factors. First, there is a large amount of variability in wear rates during mechanical testing. This can be seen in both the large standard deviation of the wear rate for our experimental test (±2.18mg/MC), and by examining the wear scars produced. Other factors could include differences in the lubricant, differences in the UHMWPE used to calibrate the model versus that used in the TKR samples (including the degree of crosslinking), the positioning of the components in the simulator, and the lack of UHMWPE chain mobility in the computational wear model. The original mechanical testing used to generate the unit work parameters [14] was also carried out at room temperature, while our simulator tests were run at 37° C. In addition, the comparison between ISO(2014) and ISO(2004) relies less on precise wear rates and more on the magnitude of the difference, as well as the distribution of wear. This study examined whether there were differences between the old and new versions of the ISO standard and our study predicted that there are indeed differences in wear volume and wear distribution. Attempts were made to find literature utilizing either version of the standard under displacement control and using NexGen components for comparison purposes however none were found. There are studies that utilize NexGen components under force control [9, 23, 25], and the wear rates in these studies vary from 5.12 mg/MC to 17.35 mg/MC. One study using a modified version of the ISO(2004) standard but with higher loading during swing phase showed a wear rate for NexGen tibial liners at 14.4 ± 2.8 mg/Mc [26, 27]. Another study using ISO-derived inputs reported wear rates of 22.0±5.8 mg/MC [28]. A study using a similar component under displacement ISO(2004) was shown to have a wear rate of 5.2 ± 1.9 mm3/MC [24]. The large variation in wear rates supports the decision to accept the results of the wear model. Another limitation of the study was that the tibial tray was excluded, and no attempt was made to consider backside wear. Given however that backside wear is a small portion of the total wear [20, 28], it was ignored for the sake of computational efficiency. Finally, while our model considers plastic deformation, viscoelastic effects are not included in our material model. Including viscoelasticity would increase CPU time and would likely only effect linear penetration, and not the comparison between the two versions of the standard. As such, we have not included values of linear penetration in our comparison of wear simulator test samples and our FEA predicted wear scars.

There has not been any attempt in the literature yet to compare the differences between the old and new versions of the standard. One study was found that compared the effect of reversing the polarity of the AP displacement, and similar to our results, they found that a switch to anterior translation of the tibia resulted in less wear (4.8 mm3/MC with anterior translation, 6.0 mm3/MC with posterior translation) [25]. The effect of this change was not investigated using the ISO Standard however, and only investigated changes in AP translation. The current direction of the field has shifted towards multi-activity and patient specific gait data [29], and while this is an important step, the standard should not be ignored as it is still relevant for regulatory approval, preclinical testing, and for comparison to other studies. The large differences between the old and new versions of the ISO standard indicate that care should be taken when comparing preclinical tests and wear studies that utilized different versions of the standard. It should be noted that wear scars seen in retrievals have been shown to cover a larger portion of the articular surface than was found in both the ISO(2014) and ISO(2004) predictions [30]. In future studies, we intend to investigate multi-activity and patient specific gait tests using our wear model and compare these to the ISO standard.

In conclusion, the results of the study suggest that major differences between ISO(2004) and ISO(2014) standard exist and therefore historical wear results should be compared with caution to newly obtained results. In addition, this study demonstrates the utility of FEA in wear analysis, though the wear model needs further work and validation before it can be used as a supplement to simulator testing.

Acknowledgments

This work was supported by the National Institutes of Health (R01 AR059843, MAW). Zimmer-Biomet (Warsaw, IN) provided CAD models for FEA analysis.

Footnotes

Conflict of Interest Statement

The authors have no conflicts to disclose.

References

  • 1.Cross M, Smith E, Hoy D, et al. The global burden of hip and knee osteoarthritis: estimates from the global burden of disease 2010 study. Ann Rheum Dis. 2014;73:1323–1330. doi: 10.1136/annrheumdis-2013-204763. [DOI] [PubMed] [Google Scholar]
  • 2.Kurtz S, Ong K, Lau E, et al. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89:780–785. doi: 10.2106/JBJS.F.00222. [DOI] [PubMed] [Google Scholar]
  • 3.Naudie DDR, Ammeen DJ, Engh GA, et al. Wear and osteolysis around total knee arthroplasty. J Am Acad Orthop Surg. 2007;15:53–64. doi: 10.5435/00124635-200701000-00006. [DOI] [PubMed] [Google Scholar]
  • 4.Sharkey PF, Hozack WJ, Rothman RH, et al. Insall Award paper. Why are total knee arthroplasties failing today? Clin Orthop Relat Res. 2002:7–13. doi: 10.1097/00003086-200211000-00003. [DOI] [PubMed] [Google Scholar]
  • 5.Gupta SK, Chu A, Ranawat AS, et al. Osteolysis after total knee arthroplasty. J Arthroplasty. 2007;22:787–799. doi: 10.1016/j.arth.2007.05.041. [DOI] [PubMed] [Google Scholar]
  • 6.Lachiewicz MP, Lachiewicz PF. Are the relative indications for revision total knee arthroplasty changing? J Surg Orthop Adv. 2009;18:74–76. [PubMed] [Google Scholar]
  • 7.Gallo J, Goodman SB, Konttinen YT, et al. Osteolysis around total knee arthroplasty: a review of pathogenetic mechanisms. Acta Biomater. 2013;9:8046–8058. doi: 10.1016/j.actbio.2013.05.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Bozic KJ, Kurtz SM, Lau E, et al. The epidemiology of revision total knee arthroplasty in the United States. Clin Orthop Relat Res. 2010;468:45–51. doi: 10.1007/s11999-009-0945-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Knight LA, Pal S, Coleman JC, et al. Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading. J Biomech. 2007;40:1550–1558. doi: 10.1016/j.jbiomech.2006.07.027. [DOI] [PubMed] [Google Scholar]
  • 10.ISO 14243-3:2014. Implants for surgery -wear of total knee-joint prosthesis - part 3: Loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test.
  • 11.Ngai V, Wimmer MA. Kinematic evaluation of cruciate-retaining total knee replacement patients during level walking: a comparison with the displacement-controlled ISO standard. J Biomech. 2009;42:2363–2368. doi: 10.1016/j.jbiomech.2009.06.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Ngai V, Schwenke T, Wimmer MA. In-vivo kinematics of knee prostheses patients during level walking compared with the ISO force-controlled simulator standard. Proc Inst Mech Eng H. 2009;223:889–896. doi: 10.1243/09544070JEIM549. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Sutton LG, Werner FW, Haider H, et al. In vitro response of the natural cadaver knee to the loading profiles specified in a standard for knee implant wear testing. JBiomech. 2010;43:2203–2207. doi: 10.1016/j.jbiomech.2010.03.042. [DOI] [PubMed] [Google Scholar]
  • 14.Schwenke T, Wimmer MA. Cross-Shear in Metal-on-Polyethylene Articulation of Orthopaedic Implants and its Relationship to Wear. Wear. 2013;301:168–174. doi: 10.1016/j.wear.2013.01.069. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.ISO 14243-3:2004. Implants for surgery - wear of total knee-joint prosthesis - part 3: Loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test.
  • 16.Bergström JS, Kurtz SM, Rimnac CM, et al. Constitutive modeling of ultra-high molecular weight polyethylene under large-deformation and cyclic loading conditions. Biomaterials. 2002;23:2329–2343. doi: 10.1016/s0142-9612(01)00367-2. [DOI] [PubMed] [Google Scholar]
  • 17.O’Brien ST, Bohm ER, Petrak MJ, et al. An energy dissipation and cross shear time dependent computational wear model for the analysis of polyethylene wear in total knee replacements. J Biomech. 2014;47:1127–1133. doi: 10.1016/j.jbiomech.2013.12.017. [DOI] [PubMed] [Google Scholar]
  • 18.Godest AC, Beaugonin M, Haug E, et al. Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J Biomech. 2002;35:267–275. doi: 10.1016/s0021-9290(01)00179-8. [DOI] [PubMed] [Google Scholar]
  • 19.Halloran JP, Petrella AJ, Rullkoetter PJ. Explicit finite element modeling of total knee replacement mechanics. J Biomech. 2005;38:323–331. doi: 10.1016/j.jbiomech.2004.02.046. [DOI] [PubMed] [Google Scholar]
  • 20.O’Brien S, Luo Y, Wu C, et al. Computational development of a polyethylene wear model for the articular and backside surfaces in modular total knee replacements. Tribol Int. 2013;59:284–291. [Google Scholar]
  • 21.Modeling Contact with Abaqus/Standard 6.14.
  • 22.Knowlton CB, Wimmer MA. An autonomous mathematical reconstruction to effectively measure volume loss on retrieved polyethylene tibial inserts. J Biomed Mater Res Part B Appl Biomater. 2013;101:449–457. doi: 10.1002/jbm.b.32782. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Strickland MA, Taylor M. In-silico wear prediction for knee replacements--methodology and corroboration. J Biomech. 2009;42:1469–1474. doi: 10.1016/j.jbiomech.2009.04.022. [DOI] [PubMed] [Google Scholar]
  • 24.Brockett CL, Abdelgaied A, Haythornthwaite T, et al. The influence of simulator input conditions on the wear of total knee replacements: An experimental and computational study. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2016;230:429–439. doi: 10.1177/0954411916645134. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Cottrell JM, Babalola O, Furman BS, et al. Stair ascent kinematics affect UHMWPE wear and damage in total knee replacements. J Biomed Mater Res Part B Appl Biomater. 2006;78:15–19. doi: 10.1002/jbm.b.30451. [DOI] [PubMed] [Google Scholar]
  • 26.Popoola OO, Yao JQ, Johnson TS, et al. Wear, delamination, and fatigue resistance of melt-annealed highly crosslinked UHMWPE cruciate-retaining knee inserts under activities of daily living. J Orthop Res. 2010;28:1120–1126. doi: 10.1002/jor.21104. [DOI] [PubMed] [Google Scholar]
  • 27.Johnson TS, Laurent MP, Yao JQ, et al. Comparison of wear of mobile and fixed bearing knees tested in a knee simulator. Wear. 2003;255:1107–1112. [Google Scholar]
  • 28.Muratoglu OK, Rubash HE, Bragdon CR, et al. Simulated Normal Gait Wear Testing of a Highly Cross-Linked Polyethylene Tibial Insert. J Arthroplasty. 2007;22:435–444. doi: 10.1016/j.arth.2006.07.014. [DOI] [PubMed] [Google Scholar]
  • 29.Freed RD, Simon JC, Knowlton CB, et al. Are Instrumented Knee Forces Representative of a Larger Population of Cruciate-Retaining Total Knee Arthroplasties? J Arthroplasty. doi: 10.1016/j.arth.2017.01.054. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Moslemi Rad E. Master’s Thesis. University of Illinois; Chicago: 2015. Volumetric Wear Assessment and Characterization of Striated Pattern of Retrieved UHMWPE Tibial Inserts. [Google Scholar]

RESOURCES