Abstract
Background
Heparin therapy and prophylaxis may be accompanied by bleeding and thrombotic complications due to individual responses to treatment. Dosage control based on standard laboratory assays poorly reflects the effect of the therapy. The aim of our work was to compare the heparin sensitivity of new thrombodynamics (TD) assay with sensitivity of other standard and global coagulation tests available to date.
Study population and methods
A total of 296 patients with high risk of venous thromboembolism (deep vein thrombosis (DVT), early postoperative period, hemoblastosis) were enrolled in the study. We used a case-crossover design to evaluate the sensitivity of new thrombodynamics assay (TD) to the hemostatic state before and after unfractionated heparin (UFH) and low-molecular-weight heparin (LMWH) therapy/prophylaxis and to compare it with the activated partial thromboplastin time (APTT), anti-Xa activity test, thrombin generation test (TGT) and thromboelastography (TEG). A receiver operating characteristic (ROC) curve analysis was used to evaluate changes before and after heparin prophylaxis and therapy. Blood was sampled before heparin injection, at the time of maximal blood heparin concentration and before the next injection.
Results
Hypercoagulation before the start of heparin treatment was detected by TD, TGT and TEG but not by APTT. The area under the ROC curve (AUC) was maximal for TD and anti-Xa, intermediate for TGT and TEG and minimal for APTT.
Conclusions
These results indicate that TD has a high sensitivity to the effects of UFH and LMWH after both prophylactic and therapeutic regimes and may be used for heparin monitoring.
Introduction
Heparin treatment and prophylaxis for venous thrombosis and thromboembolism (VTE) are widely used in clinical practice. However, wide variability in the effects of unfractionated heparin (UFH) and low-molecular-weight heparin (LMWH) is observed due to different concentrations of platelet factor 4, coagulation factors (antithrombin, fibrinogen and factor VIII) and individual variations in pharmacokinetics [1,2]. Monitoring of LMWH is used in special cases, including patients with extremely low or high body weight, age >75 years, renal disease, critical status or following a change in anticoagulant drug therapy [3–5]. The frequency of bleeding and thrombotic complications during heparin treatment and prophylaxis is 5–10% [6–11]. Therefore, monitoring of heparin efficacy is essential in many clinical situations.
The activated partial prothrombin time (APTT) and anti-Xa activity assays are used in routine control of heparin treatment. APTT is sensitive only to high (therapeutic) doses of UFH [12], and some APTT methods are insensitive to LMWH treatment [13]. The anti-Xa activity assay may be used to monitor both prophylactic and therapeutic doses of UFH or LMWH. However, this assay provides only information about the blood heparin concentration and does not show the overall clotting status of the blood, as the anticoagulant effect of heparin is influenced by individual plasma characteristics (e.g., concentration of antithrombin and heparin-binding proteins and the ability of the system to generate thrombin). Furthermore, the mechanism of the antithrombotic action of LMWH probably does not only depend upon anti-Xa and anti-IIa activities. For example, UFH and LMWH release Tissue Factor Pathway Inhibitor (TFPI) from vascular sites and this could explain some of the antithrombotic effect of subcutaneously administered LMWH [14]. Anti-Xa level has not been demonstrated to be a good predictor of bleeding risk and antithrombotic efficacy in LMWH thromboprophylaxis and treatment [14]. Additionally, a significant percentage of patients demonstrate resistance to heparin treatment [1,2,15], which cannot be detected using conventional heparin control assays.
Poor efficacy of heparin therapy may also be related to inadequate individual dosing because of insufficiently sensitive methods of laboratory control. Global hemostasis assays superimpose the state of the patient’s clotting system and the effects of anticoagulant treatment. This fact makes global hemostasis assays essential instruments for control of heparin treatment efficacy. Thromboelastography (TEG) and the thrombin generation test (TGT) were shown to be sensitive to the effects of heparin [13,16–23]. Thrombodynamics (TD) is a new global hemostasis assay that was shown to be sensitive to both hypercoagulation and hypocoagulation [24–27]. The aim of our work was to compare the heparin sensitivity of thrombodynamics assay with sensitivity of APTT, TEG and TGT.
Materials and methods
Patients and study design
To evaluate assays sensitivity to hypercoagulation we used classical case-control design with use of control group consisting of healthy volunteers (aged 19–65 years) (n = 75). To evaluate assays’ sensitivity to heparin we used a case-crossover design, a variation of a case-control design [28] due to the fact that patients requiring heparin therapy or prophylaxis often have many additional confounding factors (such as presence of concomitant treatment, temporal immobility and baseline hypercoagulation due to the major surgery or malignancy), which makes the selection of appropriate controls rather challenging. To prevent bias from these potential confounders we decided that the best selection would be the cases themselves at the time point before the heparin administration. The number of cases in the area during the study period determined the sample size.
Group 1 consisted of 124 patients with lower extremity DVT (confirmed by ultrasonography and D-dimers level) who were admitted to the Surgical Departments of City Clinical Hospitals #15 and #61 (Moscow, Russia). Subcutaneous injections of UFH (initial dose was 150 IU/kg three times a day, here and below we used IU anti-Xa activity) or LMWH (initial dose was 6,000 IU two times a day) were used for treatment.
Group 2 consisted of 78 early postoperative patients who were admitted to the Surgical Department of the National Research Center for Hematology (Moscow, Russia), the Surgical Department of Altay Department of the National Research Center for Hematology (Barnaul, Russia) and City Clinical Hospital #24 (Moscow, Russia). Patients underwent orthopedic (total hip replacement, 48 patients) and surgical procedures (colon cancer surgery, 10 patients; splenectomy and cholecystectomy, 18 patients; lung biopsy, 2 patients). LMWH prophylaxis by subcutaneous injection (3,000–4,000 IU once a day) was initiated during the first 24 hours after surgery. All patients had VTE Caprini scores of 3 or greater.
Group 3 consisted of 48 patients with hematological malignancies (multiple myeloma, 37 patients; diffuse large B-cell lymphoma, 11 patients) who were admitted to the Chemotherapy Hematologic Diseases Department and the Bone Marrow Transplantation Department of the National Research Center for Hematology (Moscow, Russia). UFH prophylaxis was administered by continuous intravenous infusion (12,000 IU per day depending on the individual risk of VTE). All patients had VTE Caprini scores of 3 or greater.
No thrombotic or bleeding episodes occurred during the investigation.
All the recruitment and analyzing was performed in 2012–2014. The research protocol was approved by the Ethics Committee of the Center for Theoretical Problems of Physicochemical Pharmacology (Protocol #1, September 5, 2011).
The lyophilized human plasma (Hemosil Normal Control Assayed, Instrumentation Laboratory, MA, USA) was used for the in vitro characterization of thrombodynamics. Plasma was prepared according to the manufacturer instructions and was supplemented with human factor XIa (FXIa, 1 pM, Enzyme Research Laboratories, IN, USA) and LMWH (0.16 IU/ml of calcium nodraparin, Aspen Health, France).
To assess the in vitro dose-dependent effect, UFH (5,000 IU/mL, Heparin-Ferein, CJSC Brinsalov-A, Russia) or LMWH (10,000 IU/mL, Enoxaparin sodium, Sanofi Winthrop Industrie, France) were diluted with normal saline and then were added to the whole blood samples of healthy volunteers (150 μL of UFH per 9 mL of citrated blood) to achieve blood concentrations of 0.04, 0.08, 0.17 and 0.33 IU/mL (for UFH) and 0.1, 0.2, 0.3 and 0.4 IU/mL (for LMWH). Normal saline was added to the control samples.
Blood sampling
Blood samples for coagulation checking were collected at the following thee time points: before the heparin treatment start (Point 0); 2h (for UFH) or 4h (for LMWH) after the heparin injection or 24h after continuous heparin infusion (Point 1, corresponds the maximal heparin concentration [29,30]); and 12h (for group 1) or 24h (for group 2) after heparin injection, immediately before the next injection (Point 2, corresponds the minimal heparin concentration).
Blood was drawn into 10 ml vacuum tubes (Monovette, Sarstedt, Germany) with 106 mM sodium citrate buffer (pH 5.5) at a 9:1 blood:anticoagulant volume ratio through the membrane screw cap needle 21G/8 mm (Multifly, Sarstedt, Germany). The samples were transported to the laboratory within 30 minutes from the blood draw. During transportation, blood samples were stored at 37°C. Whole blood was used for thromboelastography. The remaining blood was processed by centrifugation at 1600g for 15 min to obtain platelet-poor plasma (PPP) and analyzed for APTT and anti-Xa activity assay. A part of PPP was repeatedly processed by centrifugation at 10,000g for 5 min to obtain platelet-free plasma (PFP), which was used for a thrombodynamics assays. The remaining PPP was frozen in liquid nitrogen and stored at −80°C for the thrombin generation assay.
Laboratory methods
APTT and anti-Xa assay was performed using the Sysmex CA-1500 (Sysmex Corporation, Japan) automated analyzer with Pathromtin SL (Dade Behring, Marburg, Germany) and Berichrom Heparin (Siemens Healthcare Diagnostics, Erlangen, Germany).
TGT was performed in our modification as previously described [24]. Frozen PPP was thawed and incubated in a water bath at 37°C for 1 hour before the experiment. The following reagents were used: 7-amino-4-methylcoumarin (AMC, Sigma Aldrich, St. Louis, MO, USA), phospholipid vesicles prepared by extrusion from phosphatidylcholine and phosphatidylserine (Avanti, Alabaster, AL) at a molar ratio of 7:3 and and stored in a nitrogen atmosphere at 4°C, fluorogenic substrate Z-Gly-Gly-Arg-AMC (Bachem, Bubendorf, Switzerland) and relipidated rabbit tissue factor (Renam, Moscow, Russia). The final concentrations in plasma were: 4 μM of phospholipid; 5 pM of tissue factor and 400 μM of substrate. We used ACTICHROME® TF activity assay (American Diagnostica, GA, USA) for concentration of TF measurement as it was described in [31]. The endogenous thrombin potential (ETP), peak thrombin concentration (Amax), time of peak thrombin concentration (Tmax), and lag time until the appearance of thrombin (LagT) parameters were calculated.
TEG was performed using the TEG 5000 Hemostasis Analyzer System (Haemonetics Corporation, USA) and reagents. For Ca2+ level restoration and to start clotting, 0.2 M calcium chloride solution was used. The reaction time (R), clot formation time (k), alpha angle (α) and maximal amplitude (MA) parameters were calculated.
TD was performed using a Thrombodynamics Analyzer and Thrombodynamics kit (LLC HemaCore, Moscow, Russia) consisting of corn trypsin inhibitor (CTI), calcium acetate and a plastic insert with immobilized relipidated recombinant TF from Instrumentation Laboratory (Bedford, MA, USA). Manufacturer (LLC Hemacore) provided the information about density of the TF in the Thrombodynamics kits: 113±26 pmoles/m2; ACTICHROME® TF activity assay was used. Plasma preparation, experimentation and image processing were performed as previously described [24]. This method is based on registering spatial fibrin clot growth after activation of clotting in a thin layer of plasma after contact with an immobilized tissue factor bearing surface. The process of clot growth was registered by serial photography during the test (Fig 1A). Based on the photos, a plot of clot growth versus time was obtained (Fig 1B). The lag time (Tlag, the delay between the test start and the onset of clot formation), the initial velocity of clot growth (Vi, calculated as the mean clot growth velocity over a Tlag+2 to Tlag+6 interval), the stationary velocity of clot growth (Vst, calculated as the mean clot growth velocity over a Tlag+15 min to Tlag+25 min interval), the velocity of clot growth (V, the parameter, calculated as the mean clot growth velocity over the 10-min interval before spontaneous clotting occurs and equal to Vst in cases without spontaneous clot formation) were calculated.
Reference values were determined by testing plasma obtained from healthy volunteers (aged 19–65 years): n = 75, 33, 25 and 75 for APTT, TGT, TEG and TD, respectively.
Statistical analysis
Statistical analysis was performed using Origin Pro 8 (Origin-Lab Corp., USA) and MedCalc (Belgium) software. The mean value, median value, standard deviation (SD) and coefficient of variation (CV) were used to estimate the assay results. The nonparametric pair sample Wilcoxon signed rank test and Mann-Whitney U-test were used for analysis, the significance level was set at P<0.05.
The relative effect (Q) of FXIa or FXIa+LMWH on parameters Vi, Vst, V, CS and D in vitro was calculated as:
where Pn and SDn—mean value and standard deviation of parameter in normal plasma and Pmod and SDmod—mean value and standard deviation of parameter in plasma supplemented with FXIa or FXIa+LMWH. Q was reversed for Tlag.
To estimate the coagulation tests’ sensitivity during heparin therapy a receiving operating curve (ROC) curve analysis was used. The presence or absence of anticoagulant therapy was selected as an “outcome”. A standard binomial exact statistics and Bootstrap estimation (analysis of 3000 alternative ROC-curves) were used for the calculation of the parameters: area under the curve (AUC) and cut-off value for each parameter. We considered the sensitivity as low when AUC was less than 0.7, moderate when AUC was from 0.7 to 0.8 and high when AUC was more than 0.8. We compared the sensitivity to heparin of all the parameters within each assay and presented data of only one parameter from each assay with the better sensitivity. After obtaining the cut-off, sensitivity and specificity, positive and negative predictive values of each test parameter were calculated.
Potential sources of bias
The main potential source of bias is our inability to measure all tests in all of the participants. To deal with missing data we restricted analyses to individuals with complete data on all variables required for a particular analysis.
Results
Thrombodynamics variability and sensitivity in the in vitro system
We used the lyophilized human plasma to characterize TD parameters. FXIa was added to imitate the hypercoagulant plasma state [32]. LMWH (0.16 IU/ml) was additionally added to imitate the heparin prophylaxis effect. The characteristic parameters were presented in S1 Table. Clot growth velocity (V) showed the highest relative effect (Q) of the factor XIa and FXIa+LMWH so in the following sections of the paper we demonstrated only this parameter of TD.
We examined in vitro dose-dependent effect of UFH and LMWH on V in TD (Fig 1C). Heparin concentration in blood was varied within the range of 0–0.33 IU/ml for UFH and 0–0.4 IU/ml for LMWH that approximately corresponded to plasma concentration of 0–0.57 IU/ml and 0–0.69 IU/ml, respectively (based on mean hematocrit value of 42%). These heparin concentrations are characteristic for treated patients [29,30].
Thrombodynamics vs APTT
We included patients from three high-VTE-risk groups: 1) DVT, 2) early postoperative period and 3) hemoblastosis. A total of 296 participants were enrolled in the study; 250 patients were included in the analysis, and 46 were excluded because a blood sample was not collected at Point 0. Patients’ characteristics are shown in Table 1.
Table 1. Characteristics of patients.
Group | n | m/f | Age, years median (min-max) | Heparin type | Heparin dosage | Point 0/Point 1/Point 2 | |||||
---|---|---|---|---|---|---|---|---|---|---|---|
Time* | APTT n |
TD n |
Anti-Xa n |
TGT n |
TEG n |
||||||
1 DVT |
24 | 14/10 | 59 (24–86) | LMWH | 6000 IU 2x a day |
Bef/3h/12h | 24/21/23 | 23/21/22 | 24/21/24 | - | - |
71 | 46/25 | 57 (21–84) | UFH | 150 IU/kg 3x a day |
Bef/2h/8h | 64/67/64 | 64/58/64 | - | 54/50/50 | - | |
29 | 15/14 | 67 (28–87) | UFH | 150 IU/kg 3x a day |
Bef/-/8h | 29/-/25 | 29/-/28 | - | - | - | |
2 SURGERY |
42 | 21/21 | 60 (37–76) | LMWH | 4000 IU 1x a day |
Bef/-/24h | 29/-/39 | 13/-/22 | - | 18/-/32 | 27/-/38 |
26 | 11/15 | 51 (20–75) | LMWH | 3000 IU 1x a day |
Bef/3h/24h | 26/25/26 | 26/26/26 | 24/26/25 | 10/10/16 | 24/23/22 | |
10 | 3/7 | 54 (26–65) | LMWH | 4000 IU 1x a day |
Bef/3h/24h | 5/10/9 | 5/10/9 | 5/10/9 | - | - | |
3 ONCO |
34 | 21/13 | 56 (27–65) | UFH | 1200 IU/d | Bef/24h†/NA | 30/27/ NA | 30/27/ NA | - | 27/23/ NA | 29/27/ NA |
14 | 4/10 | 43 (23–80) | UFH | 1200 IU/d | Bef/24h†/NA | 14/14/ NA | 14/14/ NA | - | - | 13/12/ NA |
DVT—patients with lower extremity DVT; SURGERY—early postoperative patients; ONCO—patients with hematological malignancies; APTT—activated partial thromboplastin time; TD—thrombodynamics; TGT—thrombin generation test; TEG—thromboelastography; m/f = male/female; Bef—before administration of heparin; NA—not applicable; n—number of participants.
* Time between heparin injection and blood sampling (at maximal blood heparin concentration).
† 24h after continuous heparin infusion.
To characterize the initial state of patients before anticoagulation (Point 0), we compared the results of APTT and TD of patients with those of healthy volunteers (Fig 2). We only presented data for V in TD because it turned out to be the most sensitive to heparin treatment.
The number of patients in each group for which APTT was measured is reported in Table 1. APTT was not significantly different from healthy volunteers in group 1 and was prolonged in groups 2 and 3. Parameter V in TD was increased in all three groups of patients thus demonstrated hypercoagulant state. Therefore, the parameter V was able to detect hypercoagulation for patients in all high VTE risk groups while APTT remained in the normal range or even prolonged, which is generally interpreted as hypocoagulation.
Changes in APTT and V after heparin treatment were presented in S2 Table. APTT was significantly prolonged only for the UFH treatment while V was significantly decreased for both UFH and LMWH treatment.
We compared the sensitivity of APTT and V to heparin treatment using ROC analysis. The data set of patient’s samples before treatment was termed “controls” (there was no heparin in the blood samples), and the data set of patient’s samples after treatment was termed “cases” (there was heparin in the blood samples). To perform this analysis, we used only samples with both APTT and TD data present. We performed ROC analysis separately for Point 1 and Point 2 to obtain data about heparin effect at maximal and minimal blood concentration, respectively. The ROC curves for the APTT and V are presented in Fig 3, and the parameters of ROC-analysis are shown in S3 Table. APTT was moderately sensitive to LMWH therapy in group 1 at Point 1 (AUC = 0.720) and insensitive at Point 2 (AUC = 0.623; not significantly different from the 0.5 level) whereas V had a high sensitivity at both Point 1 and Point 2 (AUC = 1.000 and 0.864; significantly higher compared with the AUC for APTT, P<0.05). APTT was moderately sensitive to UFH therapy in group 1 at Point 1 (AUC = 0.755) and had a low sensitivity at Point 2 (AUC = 0.683) whereas V had a high sensitivity at both Point 1 (AUC = 0.849, P>0.05) and Point 2 (AUC = 0.850; significantly higher compared with the AUC for APTT, P<0.05). APTT was insensitive to LMWH prophylaxis in group 2 at both Point 1 and Point 2 (AUC = 0.611 and 0.545; not significantly different from the 0.5 level) whereas V had a high sensitivity at Point 1 (AUC = 0.897; significantly higher compared with the AUC for APTT, P<0.05) and had a low sensitivity at Point 2 (AUC = 0.616, P = 0.42). APTT had a low sensitivity to UFH prophylaxis in group 3 at Point 1 (AUC = 0.644), V had a high sensitivity at Point 1 (AUC = 0.851; significantly higher compared with the AUC for APTT, P<0.05).
Thrombodynamics vs anti-Xa assay
The number of patients in each group for which anti-Xa was measured is reported in Table 1. Anti-Xa was measured only in LMWH therapy and prophylaxis groups, so the comparison of anti-Xa and TD was performed only in these patients. The ROC curves for the anti-Xa and V are presented in Fig 4, and the parameters of ROC-analysis are shown in S4 Table. To perform this analysis, we used only samples with both anti-Xa and TD data present.
Both assays demonstrated equal sensitivity to treatment with therapeutic and prophylactic LMWH doses. Anti-Xa assay had a high sensitivity to LMWH therapy both at Point 1 (AUC = 1.000) and at Point 2 (AUC = 0.896), V in TD showed same results both at Point 1 (AUC = 1.000) and at Point 2 (AUC = 0.871). For LMWH prophylaxis anti-Xa assay had a high sensitivity at Point 1 (AUC = 0.984) but low sensitivity at Point 2 (AUC = 0.522; not significantly different from 0.5 level). V in TD showed the same tendency both at Point 1 (AUC = 0.986) and at Point 2 (AUC = 0.629; not significantly different from 0.5 level). No significant differences between AUC of anti-Xa and V was shown in any of the treatment regimes.
Thrombodynamics vs thrombin generation test
Number of patients in each group for which TGT was measured is reported in Table 1. To characterize the initial state of patients before anticoagulation, we compared the results of TGT and TD of patients with those of healthy volunteers (Fig 5). We only presented data for ETP in TGT and V in TD because these parameters turned out to be the most sensitive to heparin treatment.
ETP in TGT was significantly higher than in healthy volunteers in group 3 and was not significantly different from healthy volunteers in groups 1 and 2. V in TD was increased in groups 2 and 3 but was not different from healthy volunteers in group 1. The difference in results demonstrated in Figs 2B and 5B was due to decrease the patient’s number for correct comparison with TGT. Both TGT and TD were able to detect higher VTE risk for patients in group 3.
The ROC curves for the ETP and V are presented in Fig 6, and the parameters of ROC-analysis are shown in S5 Table. To perform this analysis, we used only samples with both TGT and TD data present. ETP had a high sensitivity to UFH therapy in group 1 at Point 1 (AUC = 0.826) and low sensitivity at Point 2 (AUC = 0.641) whereas V had a high sensitivity at both Point 1 (AUC = 0.802) and Point 2 (AUC = 0.800; significantly higher compared with the AUC for ETP, P<0.05). ETP had a high sensitivity to LMWH prophylaxis in group 2 at Point 1 (AUC = 0.853) and low sensitivity at Point 2 (AUC = 0.504; not significantly different from the 0.5 level), V had a high sensitivity at Point 1 (AUC = 1.00; significantly higher compared with the AUC for ETP, P<0.05) and low sensitivity at Point 2 (AUC = 0.616; not significantly different from the 0.5 level). ETP had a high sensitivity to UFH prophylaxis in group 3 at Point 1 (AUC = 0.813), V had a high sensitivity too (AUC = 0.859).
Thrombodynamics vs thromboelastography
Number of patients in each group for which TEG was measured is reported in Table 1. To characterize the patients’ state before anticoagulation, we compared the results of TEG and TD of patients with those of healthy volunteers (Fig 7). We only presented data for alpha in TEG and V in TD because it turned out to be the most sensitive to heparin treatment.
Alpha in TEG in groups 2 and 3 was significantly higher than in healthy volunteers as well as V in TD. We decreased the patient’s number in TD for correct comparison with TEG. Both TEG and TD were able to detect high VTE risk for patients in groups 2 and 3.
The ROC curves for the angle alpha and V are presented in Fig 8, and the parameters of ROC-analysis are shown in S6 Table. To perform this analysis, we used only samples with both TEG and TD data present. Alpha had a high sensitivity to LMWH prophylaxis in group 2 at Point 1 (AUC = 0.861) and low sensitivity at Point 2 (AUC = 0.561; not significantly different from the 0.5 level), V had a high sensitivity at Point 1 (AUC = 0.999; significantly higher compared with the AUC for alpha, P<0.05) and low sensitivity at Point 2 (AUC = 0.613; not significantly different from the 0.5 level). Alpha was moderately sensitive to UFH prophylaxis in group 3 at Point 1 (AUC = 0.777), V had a high sensitivity at Point 1 (AUC = 0.864). Therefore, sensitivity to heparin treatment of TD was at least comparable to sensitivity of TEG.
Discussion
Thrombodynamics assay is a new global hemostasis assay; the first publications on this technology date back to the early 2000s [33–35]. Heparins effect on thrombodynamics was published for small cohorts of patients or in case-reports [24,25,27]. The earlier study presenting the comparison of thrombodynamics with other standard and global coagulation assays using in vitro setting has already shown that thrombodynamics is a stable, reproducible and sensitive test with narrow parameter distributions which accurately describes pharmacodynamics of any heparin [36]. To confirm the utility of this method for clinical use, we studied the effects of heparin for the APTT, TGT, TEG and TD assays ex vivo in blood samples of patients undergoing heparin treatment. We evaluated both UFH and LMWH effects in therapeutic and prophylactic regimes of treatment.
The intra-laboratory variability of TD’s parameters was 2.4% for the clot growth velocity V, which is comparable with the APTT‘s variability (about 2%, [37]) and better than variability of TGT (13% for ETP [38]) and TEG (10% for alpha [39]). V was the most sensitive to heparin effect compared with the other TD parameters; V continuously decreased with increase of heparin concentration within all range of possible heparin concentrations in the patients’ blood. The UFH effect on V was more pronounced compared with LMWH. The reason may be that we built the dependence based only on their anti-Xa activity, whereas UFH also has equal anti-Xa and anti-thrombin activity [40].
We assessed APTT, TEG, TGT and TD to reveal the hypercoagulation in patients at high risk of VTE. APTT did not reflect changes in any of the patient groups; moreover, APTT was prolonged in groups 2 and 3 that can be wrongly regarded as an increased risk of bleeding. TD demonstrated significant hypercoagulant changes in all patients groups or at least in groups 2 and 3, TGT–only in group 3, TEG–in groups 2 and 3 (for group 1 TEG was not measured).
APTT prolongation in groups 2 and 3 could be due to decrease of coagulation factors concentration after blood loss/blood dilution during surgery or coagulation factors consumption. This is indirectly confirmed by our previous study. We have already observed prolonged APTT (38±7 sec while normal range was 25.1–36.6 sec) and normal to hypercoagulant V in TD (V>20 μm/min) in patients receiving warfarin therapy after cardiac surgery (S1 Fig) [41]. Concentrations of vitamin-K dependent clotting factors were significantly reduced compared with the normal value (100%): 32±14% for prothrombin, 22±14% for factor VII, 69±27% for factor IX and 24±15% for factor X.
We speculate that TEG, TGT and TD reflected the real coagulation state of these patients. Thus, the hypercoagulation observed in these assays may be a sign of the higher VTE risk, while the normalization or even shift to hypocoagulation during effective heparin therapy may be a sign of decreased VTE risk. Surprisingly, TGT showed the lowest hypercoagulation sensitivity among global assays tested. The lack of sensitivity of TGT to hypercoagulation preceding heparin injection in some of the groups can be, at least partially, explained by insufficient sample size. To deal with missing data we restricted analyses to individuals with complete data on all variables required for a particular analysis (as stated in Materials and methods section). However, in thrombodynamics even after reduction of the sample, required for correct comparison with TGT, all the tendencies remained visible and significant, except for the difference between control group and group 1. So, if there were any evidence that TGT registers hypercoagulation in groups 1 and 2, we expected to observe it even on this sample size.
In order to perform the comparison of the sensitivity of tests to heparins we have identify three grades of sensitivity: low (AUC<0.7), medium (AUC Є [0.7–0.8]) and high (AUC>0.8). The gradations for all time points, heparin type and dosages for each assay are presented in S7 Table. For each assay, we determined the proportion of situations where TD have a better sensitivity to heparin treatment. We found that TD is more sensitive than APTT in 100% of cases (n = 7), it is comparable to anti-Xa in 100% of cases (n = 4), it is more sensitive than Amax in TGT in 60% of cases (n = 5) and it is more sensitive than alpha in TEG in 67% of cases (n = 3). There were no cases when TD had a sensitivity to heparin lower than any other assay. All assays had a low sensitivity to heparin at Point 2 during prophylactic treatment. Thus, TD showed a better heparin sensitivity than APTT, comparable to Anti-Xa and better or comparable to TGT and TEG.
Separately, we would like to discuss the mechanisms by which thrombodynamics is so highly sensitive to heparins. Heparin is a cofactor of antithrombin and enhances its affinity for clotting factors: thrombin, Xa and IXa. The thrombodynamics assay allows us to identify the phase of the propagation, for which the complexes of internal tenase and protrombinase are responsible [42]. Both complexes contain factors that are inhibited by antithrombin: IXa and Xa, respectively. The activation phase, localized in the thrombodynamics assay at the surface with tissue factor, includes external tenase and prothrombinase. Antithrombin does not affect the external tenase. Thus, it turns out that the effect of heparin is more pronounced in the phase of propagation. This is consistent with the data obtained in our work and a comparison of the sensitivity of thrombodynamics to heparin in vitro [36]. Thus, the clot growth velocity is most sensitive to the effect of heparin.
Since we used related data we performed an additional calculation using the Bootstrap estimation [43]. We did not find any significant difference from the parameters calculated via binomial exact method (S8 Table).
Very different individual patients’ responses to heparin treatment were detected with the global hemostasis assays. We found that up to 21% of patients were in a state of hypercoagulation during heparin treatment at Point 1, while 5–70% of patients were in a state of hypercoagulation at Point 2 (S9 Table). This finding may indicate the inefficiency of heparin treatment and the higher thrombotic risk for these patients. This is an area for further clinical studies.
The main limitation of our study is that we measured anti-Xa assay, TGT and TEG only for a part of patients, so we obtained small groups and relatively large variability. Thus, TEG was not performed for patients on heparin therapy; TGT was not performed for patients on LMWH therapy. This reduces our ability to compare the sensitivity of the tests in these cases. However, comparison of the sensitivity of these assays with TD was performed only for paired data, so the comparison itself is objective and trustworthy.
In our research we used only one APTT reagent for comparison with TD. Various APTT reagents do have different sensitivity to heparin. However, from the literature we know that not only the reagent type do influence APTT sensitivity to heparins, but also the model of coagulometer does influence it [44]. For most of the reagents mentioned in literature the correlation of APTT with anti-Xa activity results varies in the range of r = 0.7–0.8 [44,45]. We decided not to focus on comparison of various APTT reagents for the sake of standardization of protocol in all clinical settings of various specialty, which took part in the study. We chose Pathromtin SL mainly for two reasons: 1) it is widely used; 2) it may be not the most heparin-responsive reagent, but it cannot also be considered low-responsive (as are, for example, Actin or IL Test [44]). We do hope that this coagulometer-reagent pairing (Sysmex CA-1500+Pathromtin SL) may, to some extent, represent a kind of "average" APTT, enough to illustrate, at least, qualitative difference between the assays mentioned in the article.
We compared assays that used different sample preparation (regarding either whole blood in TEG or platelet free plasma in TD). The decreased heparin effect in TEG may be due to the platelet factor 4 from alpha granules binding to heparin molecules [46]. However, our task was to compare clinical efficacy of TEG and TD, so we consider our comparison to be correct.
Our study showed that the thrombodynamics assay is effective for monitoring heparin. TD is more sensitive to heparin than APTT and, apparently, its sensitivity is comparable to anti-Xa sensitivity and comparable or higher than that of TGT and TEG. Further clinical studies are needed to determine whether thrombodynamics can detect cases of ineffective therapy leading to thrombosis or bleeding.
Supporting information
Data Availability
All dataset files are available from the Harvard Dataverse (doi:10.7910/DVN/LUPIM3).
Funding Statement
This work was supported by the Russian Federation President Grant for Young Scientists МК-913.2017.4 to ANB (https://grants.extech.ru/); Thrombodynamics analyzing was supported by grant from Russian Science Foundation (16-14-00-224) to FIA (Fig. 2 B-C, 3 A-D, 4 B) and by grant from Russian Science Foundation (17-74-10-224) to ANB (Fig. 5 A-C, 6 B and 7 A-B) (rscf.ru). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1.Ranucci M, Isgrò G, Cazzaniga A, Soro G, Ranucci M, Centre C, et al. Predictors for heparin resistance in patients undergoing coronary artery bypass grafting. Perfusion. 1999;14: 437–442. doi: 10.1177/026765919901400606 [DOI] [PubMed] [Google Scholar]
- 2.Muedra V, Bonanad S, Gomez M, Villalonga V, Sanchez F, Llopis JE. Relationships between antithrombin activity, anticoagulant efficacy of heparin therapy and perioperative variables in patients undergoing cardiac surgery requiring cardiopulmonary bypass. Perfusion. 2011;26: 487–495. doi: 10.1177/0267659111412999 [DOI] [PubMed] [Google Scholar]
- 3.Bounameaux H, de Moerloose P. Is laboratory monitoring of low-molecular-weight heparin therapy necessary? No. J Thromb Haemost. 2004;2: 551–4. doi: 10.1111/j.1538-7933.2004.00648.x [DOI] [PubMed] [Google Scholar]
- 4.Gouin-Thibault I, Pautas E, Siguret V. Safety profile of different low-molecular weight heparins used at therapeutic dose. Drug Saf. 2005;28: 333–349. [DOI] [PubMed] [Google Scholar]
- 5.Fareed J, Hoppensteadt D, Walenga J, Iqbal O, Ma Q, Jeske W, et al. Pharmacodynamic and pharmacokinetic properties of enoxaparin: Implications for clinical practice. Clin Pharmacokinet. 2003;42: 1043–1057. doi: 10.2165/00003088-200342120-00003 [DOI] [PubMed] [Google Scholar]
- 6.Decousus H, Tapson VF, Bergmann JF, Chong BH, Froehlich JB, Kakkar AK, et al. Factors at admission associated with bleeding risk in medical patients: Findings from the improve investigators. Chest. 2011;139: 69–79. [DOI] [PubMed] [Google Scholar]
- 7.Hirsh J, Warkentin TE, Raschke R, Granger C, Ohman EM, Dalen J. Heparin and low-molecular-weight heparin: mechanisms of action, pharmacokinetics, dosing considerations, monitoring, efficacy, and safety. Chest. 1999;115: 1760. [DOI] [PubMed] [Google Scholar]
- 8.Kahn SR, Lim W, Dunn AS, Cushman M, Dentali F, Akl EA, et al. Prevention of VTE in nonsurgical patients: Antithrombotic Therapy and Prevention of Thrombosis, 9th ed: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines. Chest. 2012;141: e195S–e226S. doi: 10.1378/chest.11-2296 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kearon C, Akl EA, Ornelas J, Blaivas A, Jimenez D, Bounameaux H, et al. Antithrombotic Therapy for VTE Disease: CHEST Guideline and Expert Panel Report. Chest. 2016;149: 315–352. doi: 10.1016/j.chest.2015.11.026 [DOI] [PubMed] [Google Scholar]
- 10.Przybysz T, Huang D. Does dalteparin PROTECT better than heparin? Crit Care. 2011;15: 315 doi: 10.1186/cc10581 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.van Es N, Bleker SM, Wilts IT, Porreca E, Di Nisio M. Prevention and Treatment of Venous Thromboembolism in Patients with Cancer: Focus on Drug Therapy. Drugs. 2016;76: 331–41. doi: 10.1007/s40265-015-0526-3 [DOI] [PubMed] [Google Scholar]
- 12.Price EA, Jin J, Nguyen HM, Krishnan G, Bowen R, Zehnder JL. Discordant aPTT and anti-Xa values and outcomes in hospitalized patients treated with intravenous unfractionated heparin. Ann Pharmacother. 2013;47: 151–8. doi: 10.1345/aph.1R635 [DOI] [PubMed] [Google Scholar]
- 13.Thomas O, Lybeck E, Strandberg K, Tynngåd N, Schött U. Monitoring low molecular weight heparins at therapeutic levels: Dose-responses of, and correlations and differences between aptt, anti-factor xa and thrombin generation assays. PLoS One. 2015;10: 1–16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Greaves M. Limitations of the laboratory monitoring of heparin therapy. Scientific and Standardization Committee Communications: on behalf of the Control of Anticoagulation Subcommittee of the Scientific and Standardization Committee of the International Society of. Thromb Haemost. 2002;87: 163–4. [PubMed] [Google Scholar]
- 15.Srinivasan C, Kuppuswamy B. Rhabdomyolysis complicating non-invasive blood pressure measurement. Indian J Anaesth. 2012;56: 428–430. doi: 10.4103/0019-5049.100847 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Artang R, Frandsen NJ, Nielsen JD. Application of basic and composite thrombelastography parameters in monitoring of the antithrombotic effect of the low molecular weight heparin dalteparin: an in vivo study. Thromb J. 2009;7: 14 doi: 10.1186/1477-9560-7-14 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Gibbs NM, Sear JW. Effect of ketorolac, bupivacaine and low-dose heparin on thrombelastographic variables in vitro. Br J Anaesth. 1995;75: 27–30. [DOI] [PubMed] [Google Scholar]
- 18.Schaden E, Jilch S, Hacker S, Schober A, Kozek-Langenecker S. Monitoring of unfractionated heparin with rotational thrombelastometry using the prothrombinase-induced clotting time reagent (PiCT®). Clin Chim Acta. Elsevier B.V.; 2012;414: 202–5. [DOI] [PubMed] [Google Scholar]
- 19.Tekkesin N, Tekkesin M, Kaso G. Thromboelastography for the monitoring of the antithrombotic effect of low-molecular-weight heparin after major orthopedic surgery. Anadolu Kardiyol Derg. 2015;15: 932–937. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Zmuda K, Neofotistos D, Ts’ao CH. Effects of unfractionated heparin, low-molecular-weight heparin, and heparinoid on thromboelastographic assay of blood coagulation. Am J Clin Pathol. 2000;113: 725–31. doi: 10.1309/Q4AE-BMCW-CQ7J-NUVT [DOI] [PubMed] [Google Scholar]
- 21.Chowdary P, Adamidou D, Riddell A, Aghighi S, Griffioen A, Priest P, et al. Thrombin generation assay identifies individual variability in responses to low molecular weight heparin in pregnancy: implications for anticoagulant monitoring. Br J Haematol. 2015;168: 719–727. doi: 10.1111/bjh.13193 [DOI] [PubMed] [Google Scholar]
- 22.al Dieri R, Alban S, Beguin S, Hemker HC. Thrombin generation for the control of heparin treatment, comparison with the activated partial thromboplastin time. J Thromb Haemost. 2004;2: 1395–1401. doi: 10.1111/j.1538-7836.2004.00798.x [DOI] [PubMed] [Google Scholar]
- 23.Hacquard M, Perrin J, Lelievre N, Vigneron C, Lecompte T. Inter-individual variability of effect of 7 low molecular weight antithrombin-dependent anticoagulants studied in vitro with Calibrated Automated Thrombography. Thromb Res. Elsevier Ltd; 2011;127: 29–34. [DOI] [PubMed] [Google Scholar]
- 24.Gracheva MA, Urnova ES, Sinauridze EI, Tarandovskiy ID, Orel EB, Poletaev A V, et al. Thromboelastography, thrombin generation test and thrombodynamics reveal hypercoagulability in patients with multiple myeloma. Leuk Lymphoma. 2015;56: 3418–25. doi: 10.3109/10428194.2015.1041385 [DOI] [PubMed] [Google Scholar]
- 25.Seregina EA, Nikulina OF, Tsvetaeva N V., Rodionova MN, Gribkova I V., Orel EB, et al. Laboratory tests for coagulation system monitoring in a patient with β-thalassemia. Int J Hematol. 2014;99: 588–596. doi: 10.1007/s12185-014-1559-1 [DOI] [PubMed] [Google Scholar]
- 26.Soshitova NP, Karamzin SS, Balandina AN, Fadeeva OA, Kretchetova A V, Galstian GM, et al. Predicting prothrombotic tendencies in sepsis using spatial clot growth dynamics. Blood Coagul Fibrinolysis. 2012;23: 498–507. doi: 10.1097/MBC.0b013e328352e90e [DOI] [PubMed] [Google Scholar]
- 27.Seregina EA, Tsvetaeva N V, Nikulina OF, Zapariy AP, Erasov A V, Gribkova I V, et al. Eculizumab effect on the hemostatic state in patients with paroxysmal nocturnal hemoglobinuria. Blood Cells Mol Dis. 2015;54: 144–50. doi: 10.1016/j.bcmd.2014.11.021 [DOI] [PubMed] [Google Scholar]
- 28.Lombardi DA. The case-crossover study: A novel design in evaluating transient fatigue as a risk factor for road traffic accidents. Sleep. 2010;33: 283–284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Bara L, Billaud E, Ph D, Kher A, Samama M, Ph D. Increased Anti-Xa Bioavailability for a Low Molecular Weight Heparin (PK 10169) Compared with Unfractionated Heparin. Semin Thromb Hemost. 1985;11: 0–1. [DOI] [PubMed] [Google Scholar]
- 30.Sanderink GJ, Le Liboux A, Jariwala N, Harding N, Ozoux ML, Shukla U, et al. The pharmacokinetics and pharmacodynamics of enoxaparin in obese volunteers. Clin Pharmacol Ther. 2002;72: 308–318. doi: 10.1067/mcp.2002.127114 [DOI] [PubMed] [Google Scholar]
- 31.Balandina AN, Shibeko AM, Kireev DA, Novikova AA, Shmirev II, Panteleev MA, et al. Positive feedback loops for factor V and factor VII activation supply sensitivity to local surface tissue factor density during blood coagulation. Biophys J. 2011;101: 1816–24. doi: 10.1016/j.bpj.2011.08.034 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Lipets E, Vlasova O, Urnova E, Margolin O, Soloveva A, Ostapushchenko O, et al. Circulating contact-pathway-activating microparticles together with factors IXa and XIa induce spontaneous clotting in plasma of hematology and cardiologic patients. PLoS One. 2014;9: e87692 doi: 10.1371/journal.pone.0087692 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Ovanesov M V., Krasotkina J V., Ul’yanova LI, Abushinova K V., Plyushch OP, Domogatskii SP, et al. Hemophilia A and B are associated with abnormal spatial dynamics of clot growth. Biochim Biophys Acta—Gen Subj. 2002;1572: 45–57. [DOI] [PubMed] [Google Scholar]
- 34.Ovanesov M V., Lopatina EG, Saenko EL, Ananyeva NM, Ul’yanova LI, Plyushch OP, et al. Effect of factor VIII on tissue factor-initiated spatial clot growth. Thromb Haemost. 2003;89: 235–242. [PubMed] [Google Scholar]
- 35.Ovanesov M V., Ananyeva NM, Panteleev MA, Ataullakhanov FI, Saenko EL. Initiation and propagation of coagulation from tissue factor-bearing cell monolayers to plasma: Initiator cells do not regulate spatial growth rate. J Thromb Haemost. 2005;3: 321–331. doi: 10.1111/j.1538-7836.2005.01128.x [DOI] [PubMed] [Google Scholar]
- 36.Sinauridze EI, Vuimo TA, Tarandovskiy ID, Ovsepyan RA, Surov SS, Korotina NG, et al. Thrombodynamics, a new global coagulation test: Measurement of heparin efficiency. Talanta. 2018;180: 282–291. doi: 10.1016/j.talanta.2017.12.055 [DOI] [PubMed] [Google Scholar]
- 37.Doherty TM, Shavelle RM, French WJ. Reproducibility and variability of activated clotting time measurements in the cardiac catheterization laboratory. Catheter Cardiovasc Interv. 2005;65: 330–337. doi: 10.1002/ccd.20355 [DOI] [PubMed] [Google Scholar]
- 38.Dargaud Y, Wolberg AS, Luddington R, Regnault V, Spronk H, Baglin T, et al. Evaluation of a standardized protocol for thrombin generation measurement using the calibrated automated thrombogram: An international multicentre study. Thromb Res. Elsevier Ltd; 2012;130: 929–934. [DOI] [PubMed] [Google Scholar]
- 39.Thane K, Bedenice D, Pacheco A. Operator-based variability of equine thromboelastography. J Vet Emerg Crit Care. 2017;27: 419–424. [DOI] [PubMed] [Google Scholar]
- 40.Garcia DA. Parenteral Anticoagulants. Chest. 2012;141: e24S doi: 10.1378/chest.11-2291 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Goncharova EI, Spiridonova EA, Balandina AN, Poletaev A V., Sepoyan AM, Ataullakhanov FI, et al. Diagnosis of Impairments in the Hemostatic System in the Use of Warfarin in Cardiac Surgical Patients. Gen Reanimatol. 2015;11: 51. [Google Scholar]
- 42.Panteleev MA, Ovanesov M V., Kireev DA, Shibeko AM, Sinauridze EI, Ananyeva NM, et al. Spatial propagation and localization of blood coagulation are regulated by intrinsic and protein C pathways, respectively. Biophys J. 2006;90: 1489–1500. doi: 10.1529/biophysj.105.069062 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Skalská H, Freylich V. Web-Bootstrap Estimate of Area Under ROC Curve. Austrian J Stat. 2006;35: 325–330. [Google Scholar]
- 44.Bates SM, Weitz JI, Johnston M, Hirsh J, Ginsberg JS. Use of a fixed activated partial thromboplastin time ratio to establish a therapeutic range for unfractionated heparin. Arch Intern Med. 2001;161: 385–91. [DOI] [PubMed] [Google Scholar]
- 45.Manzato F, Mengoni A. Evaluation of the Activated Partial Thromboplastin Time (APTT) Sensitivity to Heparin Using Five Commercial Reagents: Implications for Therapeutic Monitoring. 1998;36: 975–980. [DOI] [PubMed] [Google Scholar]
- 46.Mixon TA, Dehmer GJ. Recombinant platelet factor 4 for heparin neutralization. Semin Thromb Hemost. 2004;30: 369–377. doi: 10.1055/s-2004-831050 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All dataset files are available from the Harvard Dataverse (doi:10.7910/DVN/LUPIM3).