Abstract
Mandibular distraction is a surgical process that progressively lengthens bone. To improve the distraction procedure and devices, the load of distraction and the mechanical strain of soft tissues during the process must be determined. We tested the assumption that it could be the periosteum primarily opposing distraction. Therefore we assessed the mechanical properties of the human mandibular periosteum and compared the stress-strain data with the torque measured on the activator during a cadaveric mandibular distraction. A 20 mm horizontal mandibular distraction was performed in 7 cadavers using standard distractors. Torque was measured with a torquemeter placed on the activation rods of the devices, providing a load (Lt) for each millimeter of distraction. In parallel, 18 periosteum samples were harvested from 9 cadaver mandibles. Uniaxial tensile tests were performed on the specimens and an estimated load (Lc) was calculated using periosteal stress-strain data and mandibular dimensions. During the distraction process, we observed an increase of the load Lt from 11.6 to 50.6 N. The periosteum exhibited a nonlinear viscoelastic stress-strain relationship, typical of biological tissues composed of collagen and elastin. The median Lc and Lt were not significantly different for the first millimeter of distraction. We demonstrated the periosteum is primarily responsible for opposing the distraction load.
Introduction
Distraction osteogenesis (DO) is a surgical procedure consisting of the progressive lengthening of a bone segment (1 mm/day). The technique requires the implantation of a distractor device, and daily activation is responsible for the lengthening of the bone. The use of distraction is widespread in the craniofacial area for the treatment of congenital malformations or acquired large bone defects [1]. Despite the technical improvements that lead to miniaturized intraoral distractors [2, 3], their activation still requires a transmucosal or transcutaneous rod to rotate the endless screw. This activation rod may be responsible for multiple adverse events and discomfort [4]. To overcome these issues, new generations of distractor devices such as automated distractors or distractors with distant activation have been proposed, but are currently not suitable for human use [5, 6]. In a previous study, we demonstrated the feasibility of distant activation with a magnetically activated device for mandibular distraction [7]. Despite the increasing use of DO and the development of new mechanisms in the field of DO, the role of the surrounding soft tissues, mainly the periosteum, and their participation in the mechanical load opposing the distraction vector are not understood. Forces involved in the process are also not precisely determined and they are essential to develop new generation distractor devices [5, 7].
The periosteum is a bilayered soft tissue surrounding bone [8]. The outer fibrous layer contains primarily collagen and elastin fibres, whereas the inner layer has a higher cell density, primarily of osteoblasts and periosteum-derived mesenchymal stem cells (PDCs). The periosteum is anchored to the bone by Sharpey’s fibres, strong fibres with a high collagen content [8]. It exhibits the properties of a viscoelastic anisotropic material [9–11]. Its behaviour during a uniaxial tensile test has been characterized by a nonlinear and partially reversible stress-strain relationship, which varies according to the orientation of its fibres. The periosteum stress-strain curve has the same aspect as other connective tissues composed of collagen and elastin with an initial toe region with a low elastic modulus, followed by a heel (transition region) and a stiffer linear phase (high elastic modulus) before rupture. Moreover, detachment of the periosteum from its bone support results in natural shrinkage along the long and short axes. Hence, in vivo, the periosteum is pre-stressed on the bone [9, 10, 12, 13]. These properties are consistently observed but vary across species and according to the anatomical location [9–15]. Despite its known role in many bone regeneration aspects [8, 14], to our knowledge, no study of the biomechanics of the human mandibular periosteum has been conducted, and none have incorporated it into a mandibular distraction process (distraction may be compared to a uniaxial traction test).
We hypothesized that the load opposing the distraction process during its firsts millimeters—before the callus formation—is provided by the periosteum. In the first part of the study, we measured the torque at the endless screw of the devices during a mandibular distraction to determine the global load of distraction. As we assumed that distraction is similar to a uniaxial traction test, the second part explored the mechanical properties of the human mandibular periosteum, especially the stress generated through uniaxial elongation to determine the periosteal load of distraction. Finally, we compared the torque-based load of distraction to the periosteal stress-strain-based load to test our main hypothesis and to provide the first elements for a distraction analytic model.
Materials and methods
From January to June 2017, 9 human cadavers (3 males and 6 females numbered from I to IX) were dissected. These cadavers were provided by the Ecole de Chirurgie, Agence Générale des Equipements et Produits de Santé-AGEPS, Assistance Publique Hopitaux de Paris-APHP, APHP. Permission to perform cadaveric study on cadaveric specimens was obtained from the institutional review board (Ecole de Chirurgie, AGEPS, APHP). All the cadaveric subjects had given their consent for the use of their body for medical research. Ages at death ranged from 75 to 97 years. Lengths of cold preservation (at -18°C) extended from 13 to 460 days. A complete 20 mm horizontal distraction with torque assessment was performed on 7 cadavers (III to IX). For each distraction a different distractor device was used. Eighteen periosteum samples (2 per left mandibular corpus) were harvested.
A. Assessment of the load of distraction: Lt (n = 7 cadavers)
The torque-based load Lt was obtained following a two-stage procedure: all the distractors were calibrated prior to the cadaveric experiments in order to determine their specific torque-load relationship, and then the torque was measured for each millimeter at the activation rod during standard 20 mm right horizontal distractions.
Distractor calibration and Lt acquisition
For a given load exerted on the two plates of the distractor, the torque required to activate the distraction (i.e. to depart the two plates) depends on the screw-thread size and the friction inside the endless screw. To discard the influence of friction, all distractors were calibrated before cadaveric experiments for loads ranging from 0 to 60 N from the first to the twentieth millimeters of distraction, using a custom-made apparatus (See S1 Text). Each device required four clockwise turns of the endless screw to achieve a 1 mm displacement. For each mm, the torque (in mN.m) on the activation rod during 5 s acquisitions was measured with a torque meter (Center Easy TT, Andilog Technologies, Marseille Vitrolles, France) at a sampling rate of 200 Hz. The torque values were averaged, providing a mean torque for each millimeter and for each load tested. This device-specific torque-load relationship provided an estimation of the load on the plate of the distractor for a given torque.
Cadaveric torque measurements
In this section, cadaveric distraction followed typical patient workflow [16]. A mandibular distraction was performed on the right horizontal branch of the mandible. Through an intra-oral approach, the mucosa and periosteum were incised longitudinally at the upper part of the mandible. Then, the periosteum was lifted 40 mm on the vestibular side and 10 mm on the lingual side of the mandible, then a complete vertical osteotomy of the mandibular corpus was performed. A 20 mm mandibular distractor device (Medicon eG, Tuttlingen, Germany) was subperiosteally implanted and fixed with four 5 mm titanium screws. The mucosa and periosteum were sutured and the endless screws were inserted through the mucosa (S2A Fig). The torque was measured on the activation rods during the entire distraction (20mm) with the same torquemeter, and the mean torque values for each millimeter were recorded using the same protocol as described above. Finally, the load of distraction Lt was determined using a linear interpolation between each point of the device-specific torque-load curves.
B. Periosteal mechanical behaviour: Lc
The primary objective of this study was to evaluate the mechanical role of the periosteum involved during distraction. As we assumed that a mandibular corpus distraction would stretch the periosteum and the other surrounding soft tissues along a single, anteroposterior axis, the next phase of our study was to characterize the periosteum behavior during a uniaxial tensile test.
Periosteum sample harvest and histological assessment
For each subject, the vestibular and lingual periosteum were harvested from the left mandibular corpus using a transmandibular approach (S2B Fig). Templates of 40 mm per 10 mm were drawn on the surface of the mandibular corpus (both lingual and vestibular sides) in an anteroposterior direction consistent with the horizontal mandibular distraction direction. Dimensions were set before harvesting the tissue to compensate for natural shrinkage. The periosteum is first detached from the overlying soft tissue. Then, the periosteum was cut following the drawings; slices were collected using a sharp elevator and immediately stored in saline solution. Specimens were removed from the saline solution a few hours after harvest and suspended 10 minutes in ambient air in a vertical position to drain the excess water. The objectives of this procedure were to allow complete relaxation, to avoid desiccation, and to limit the artefact due to tissue swelling.
Separate samples (5 by 10 mm, see S2B Fig) of each periosteal specimen were harvested at the posterior end of the tested specimens, fixed in formol and embedded in paraffin. Sections 4 μm thick were cut from each paraffin block. Staining with haematoxylin/eosin/saffron (HES) was automated using a Leica Autostainer (Leica Biosystems GmbH, Nussloch, Germany). The sections were mounted in synthetic resin (WVR International, Radnor, PA, USA). The mean thickness (h) of the periosteum was measured on cross-section slides (averaged on 5 random thickness measures for each samples) using Image J version 1.48 (public domain software, National Institutes of Health, Bethesda, MD, USA).
C. Parameters and tensile protocol
Tensile tests and protocol
A tensile test protocol was determined in order to simulate the periosteum condition during distraction. Tensile tests were performed with a uniaxial elongation machine (3342 Single Column, Instron Corp., Illinois Tool Works Inc., Glenview, IL, USA) at room temperature (18-20°C). Each periosteal sample was moistened every 3 min during the experiment using an atomizer (see Fig 1A). Periosteum slice ends were secured in two grips connected to the elongation device. The distance between the grips is controlled at a precision of 0.01 mm. The bottom jaw was fixed on the machines base; the upper jaw could move at a specified speed in the vertical direction and was fixed to a static load cell with a 100 N capacity (2519 series, Instron Corp., Illinois Tool Works Inc.).
Stress is defined as a load normalized on a surface. Hence, we have:
(1) |
where σ is the axial stress in megapascals (MPa), L is the load in newtons (N) recorded by the load cell (see above for the load cell characteristics), and w × h is the initial cross-sectional area of the specimen (in m2) perpendicular to the direction of stretch; h is the thickness of the periosteum measured on histology (see above); w is the initial width of periosteum slice measured on the bone.
Strain is defined as:
(2) |
where Δl is the variation of distance between the two grips measured during the tensile test and l is the initial distance between the grips.
Periosteal test specimens were loaded in tension at a rate of 0.25 mm/s until 15% deformation, εg = 0.15 (Phase 1), and relaxed 300 s (phase 2). The traction rate was set at 0.25 mm/s, which approximately corresponds to the distraction speed in vivo. The duration of relaxation was arbitrarily set to 300 s, as a 24-hr relaxation (as in a surgical distraction) was impossible to conduct under experimental conditions.
The tensile test method was elaborated using Bluehill 3 software (Instron, Illinois Tool Works Inc.).
Strain homogeneity
To demonstrate the homogeneity of the deformation during a uniaxial tensile test, strain in the central region of the 5 samples was measured using digital image correlation (DIC). Samples were speckled using India ink, and periosteal elongation was recorded using a 5-megapixel resolution video camera (GO-5000M-PMCL, JAI, Copenhagen, Denmark). Images were processed using Vic-2D version 6.0.6 digital image correlation software (Correlated Solutions, Inc., Irmo, SC, USA) to compute the Lagrange deformation fields of the samples (e.g. Fig 1B). The measured strain in the central region is linear with ε/εg = k. Therefore, we corrected the strain εg by this coefficient to determine the strain ε of the sample homogeneous region. For the samples where the DIC was not available, we have corrected εg by the average coefficient calculated on the other experiments. In the following, we only consider the strain of the homogeneous region ε ().
Stress-strain and relaxation data
The periosteum stress-stain curve presents two linear regions with a low and high modulus (toe and steep regions, respectively). The toe and steep regions of the stress-strain curves of phase 1 were fitted using two linear functions. The elastic moduli, Etoe and Esteep (in MPa), are the slopes of the fits:
(3) |
where ε0 and are respectively the strain and stress values of the intersection point of the two linear fits.
For phase 2, the stress over time relation σ(ε, t), normalized by the stress at the beginning of the relaxation σmax, was fitted with the bi-exponential function
(4) |
providing two characteristic times, τ1 and τ2, specific to each periosteal sample [17]. σ∞ is the residual stress over σmax.
Stress-strain-based load estimation
The first step was to determine the periosteum stress and strain on the bone (σ0 and ε0, respectively). According to Bertram et al., this physiological state corresponds to the transition point between the toe and steep regions of the stress-strain curve [12]. This observation is supported by studies conducted on other collagen tissues such as tendons [18]. Hence, we determined ε0 as the intersection point of the two linear curves fitting the toe and steep regions. σ0 is the value of the experimental stress-strain curve at ε0.
During experimental cadaveric distraction, the periosteum was lifted approximately 40 mm on the vestibular side and 10 mm on the lingual side. Therefore, assuming that the sharpey fibers clamped the periosteum on the bone everywhere except where the operator lifted it, we assessed that 1 mm of distraction corresponded to 2.5% and 10% strain for the vestibular () and lingual () sides. Accordingly the strain on each side of the mandibule is:
(5) |
for the vestibular and lingual sides, respectively.
Using the fit of the steep region of phase 1, we evaluated stresses , on the vestibular and lingual side of the mandibule as:
(6) |
To approximate the load generated by the periosteum during a distraction, we calculated a stress-strain-based load Lc for each cadaver.
(7) |
where and are the stress measured at the εf strain for periosteal samples harvested on the lingual and vestibular sides of the mandible, respectively. We assumed a 60 mm perimeter for a complete mandibular corpus, in which the lingual and vestibular sides accounted for half of this perimeter (30 mm). Then, we defined wl and wv as the lingual and vestibular parts of the mandibular perimeter, respectively. hl and hv are the thickness of the lingual and vestibular periosteal slices (histological measurements), respectively.
Lc-Lt comparison
The stress-strain-based estimated load Lc was compared to the torque-based load Lt for the first millimeter of distraction.
Data processing and statistical analysis
All curves were fitted using Matlab (MathWorks, Inc., Natick, MA, USA). Statistical analyses were conducted using SPSS Statistics for Windows, Version 23.0 (IBM Corp., Armonk, NY, USA). Significance was set at p < 0.05. Continuous variables were tested for normality using a Wilks-Shapiro test. The ones with normal distribution were displayed as the mean and standard error of the mean (SEM) and compared using a t-test. Non-parametric variables were displayed as the median and interquartile range (IQR) and compared using a Wilcoxon ranks test.
Results
The mean thickness was h = 0.23 mm (the interquartile range, IQR = 0.11 mm, Table 1). The coefficient k are constant along the whole tensile test (for all strains). The values are for vestibular samples , , and the coefficients are for lingual samples , , .
Table 1. Thickness value of the periosteum measured on histological cross-section slides (averaged on 5 random thickness measures for each samples).
Cad. | I | II | III | IV | V | VI | VII | VIII | IX |
---|---|---|---|---|---|---|---|---|---|
hv | 0.281 | 0.317 | 0.215 | 0.296 | 0.208 | 0.416 | 0.277 | 0.290 | 0.308 |
hl | 0.167 | 0.179 | 0.140 | 0.243 | 0.136 | 0.187 | 0.185 | 0.139 | 0.176 |
A. Load of distraction assessment: Lt
Torque-based load estimates Lt
To cope with outliers in the series of torque measurements, and to compare these values with stress-strain-based loads (Lc), Lt was displayed as the median and IQR of each cadaver.
The torque-based load Lt was calculated for each millimeter of distraction in each cadaveric experiment (Fig 2). We observed a steady increase of Lt through the distraction process, with median loads ranging from 11.6 N to 50.6 N.
B. Periosteal mechanical behaviour: Lc
Stress-strain relationship and relaxation
The digital image correlation showed homogeneous strains in the central region of the periosteum samples during tensile tests (Fig 1B). A typical non-linear stress-strain relationship was observed, with an initial toe region followed by a heel and a linear steep region (Fig 3). For phase 1, the median Etoe = 0.3 MPa (IQR = 0.6 MPa), while the median Esteep = 10.8 MPa (IQR = 10.0 MPa).
Regarding the transition region of the stress-strain curve, we measured a median ε0 = 0.032 (IQR = 0.01). The corresponding median stress is σ0 = 0.043 MPa (IQR = 0.031 MPa).
Parameters of phases 1, 2 for all the specimens are displayed in Table 2.
Table 2. Moduli of the steep linear region of the stress-strain curves in phase 1 (Esteep), ε0 strains, σ0 pre-stresses, τ1 and τ2 characteristic times, stress-strain and torque-based loads (Lc and Lt) are displayed for each periosteal sample and the corresponding cadaver.
Cad. | loc | Esteep | ε0 | σ0 | τ1 | τ2 | Lc | Lt |
---|---|---|---|---|---|---|---|---|
I | v | 8.1 | 0.033 | 0.018 | 127.8 | 9.6 | 7.3 | n/a |
l | 9.6 | 0.034 | 0.033 | 122.1 | 10.1 | |||
II | v | 7.7 | 0.030 | 0.035 | 122.8 | 8.4 | 7.0 | n/a |
l | 9.2 | 0.033 | 0.031 | 99.4 | 8.4 | |||
III | v | 29.7 | 0.037 | 0.244 | 388.5 | 68.1 | 13.3 | 34.9 |
l | 10.2 | 0.052 | 0.065 | 107.9 | 8.7 | |||
IV | v | 107.1 | 0.012 | 0.216 | 74.0 | 6.1 | 42.4 | 37.9 |
l | 24.9 | 0.036 | 0.049 | 99.3 | 6.3 | |||
V | v | 12.9 | 0.031 | 0.089 | 75.4 | 5.5 | 6.3 | 16.8 |
l | 9.6 | 0.021 | 0.040 | 107.2 | 9.3 | |||
VI | v | 27.1 | 0.035 | 0.119 | 126.5 | 7.5 | 20.4 | 18.0 |
l | 20.6 | 0.027 | 0.046 | 134.8 | 11.6 | |||
VII | v | 13.7 | 0.024 | 0.036 | 62.8 | 6.6 | 11.6 | 4.7 |
l | 15.6 | 0.029 | 0.038 | 92.4 | 7.7 | |||
VIII | v | 10.3 | 0.043 | 0.030 | 100.6 | 5.7 | 4.8 | 12.3 |
l | 4.2 | 0.025 | 0.025 | 143.8 | 9.7 | |||
IX | v | 8.0 | 0.021 | 0.052 | 105.1 | 8.9 | 8.4 | 24.2 |
l | 11.2 | 0.041 | 0.061 | 128.7 | 10.5 | |||
Mean | 123.3 | 11.6 | ||||||
SEM | 21.2 | 16.1 | ||||||
Median | 10.8 | 0.032 | 0.043 | 11.6⋆ | 18.0 | |||
IQR | 10.0 | 0.010 | 0.031 | 11.8⋆ | 15.0 | |||
p | 0.22† |
We fitted the relaxation curves with the bi-exponential function with R2 superior to 0.98, giving a mean τ1 of 123.3 s (2SEM = 21.2 s), a mean τ2 of 11.6 s (2SEM = 16.1 s), and an average σ∞ of 0.4 (2SEM = 0.01). Relaxation curves and their parameters are displayed in Fig 4.
Stress-strain-based load estimation Lc
For this part, we assumed that a non-surgically detached periosteum is still attached to the bone through the Sharpey fibers. We observed a median load Lc = 11.6 MPa (IQR = 11.8 MPa).
C. Lt-Lc comparison
The median Lc accounted for more than 64% of the median Lt(s) for s = 1 mm. No statistical difference was found between Lc and Lt (p = 0.22). The results are summarized in Table 2.
Discussion
In this study, we assumed that a uniaxial tensile test and distraction would generate similar periosteal mechanical stress. Our results showed that the periosteum is the primary tissue contributing to the load-opposing distraction at the beginning of distraction activation.
Material behaviour (σ(ε, t))
The stress-strain relationship of the human mandibular periosteum is typical of biological connective tissues. Indeed, we observed nonlinear J-shaped stress-strain curves, stress relaxation at constant strain. [18]. Most of these features have already been described in the literature on the periosteum in other anatomical locations or species [8]. For instance, Bertram et al. [12] found a mean modulus of 229.5 MPa (±89.6, n = 72), when testing leghorn chicks tibiotarsal periosteum in an axial direction. They also reported a change in periosteal tensile properties over the course of growth, namely an increase in periosteal stiffness in the early period, followed by a decrease in the latest stages of growth. Popowics et al. [10], tested the pig periosteum in three anatomical regions, in both the long-axis and transverse orientation. Although directional effect seemed mild (except for a significantly higher longitudinal peak strain in the metacarpal area), substantial differences were found in periosteal stiffness among the three bones: moduli of the zygomatic arch, metacarpal, and mandibular periosteum were 91.7 MPa, 84.7 MPa and 63.0 MPa, respectively. Mc Bride et al. [9] tested the ovine femoral diaphyseal periosteum both in an axial and circumferential orientation. The modulus of the stiff region of the stress-strain curve was 25.67 MPa, in the axial direction and 4.41 MPa in the circumferential direction. These studies also suggest that periosteal properties are influenced by life cycles [12], muscle or tendon attachments [9, 10, 12], and fibers orientation [9]. In our study, the median moduli of the steep region of the stress-strain curve was 10.75 MPa. This result was 3 times as low as in the pig mandible for a tensile test along the same axis [10]. In our study, periosteal samples were harvested in old, partially or totally edentulous subjects, in the mandibular corpus region. These facts may explain the low moduli we measured.
The results of displayed in Fig 4 confirmed that modeling the relaxation with a bi-exponential function is reasonable. This assumption was reinforced by the fact that the periosteum is composed by two materials (elastin and collagen). The two characteristic times that were measured, τ1 and τ2, were significantly less than that the duration of 24 hours between each activation during in vivo distraction (123.3, 11.6 s). Therefore, if we neglected bone remodelling, it is probable that the periosteum is in a relaxed state during a conventional distraction before each activation (one activation per 12 or 24 hours). However, it must be noted that a longer relaxation test (unavailable here) would probably show the existence of a longer characteristic time [17]. It has to be noticed that these relaxation tests were performed on cadaveric subjects aged at death of at least 75 years old, which could lead to different relaxation times than the ones in children distraction.
Physiological state (ε0)
Like many connective tissues composed of collagen and elastin (such as skin and tendons), the periosteum is stretched on its support [8]. Therefore, to compare the stress-strain data of a tensile test with the load measured during distraction, it was necessary to determine the ε0 strain of the periosteum (its physiological state on the bone).
Despite several attempts during preliminary experiments, it was difficult to provide reliable outcomes. Indeed, the periosteum naturally shrinks when detached from its bone support, and returning the samples to their on-the-bone state required stretching along the long and short axes. Moreover, we were not able to find the exploitable retraction rates, because saline storage was responsible for excessive retraction due to tissue swelling [19].
To overcome this issue, we relied on the literature to place ε0 in the transition region of the stress-strain relationship [12, 18]. In this study, the median ε0 was 3.2%, with a corresponding median stress σ0 = 0.043 MPa. Although these values were significantly lower than the findings of other animal periosteum studies [12], but they were comparable to those of other collagen tissues [18]. Describing and modeling the behavior of connective tissues based on their microstructural organization is not new [18, 20–24]. Nevertheless, a real-time microstructural analysis in parallel with an optimized tensile protocol would allow for correlation of the organization of the ground substance (primarily collagen) with the stress and strain levels, along the elongation, and during relaxation.
Lt measures
We showed an increasing load through the 20 mm distractions in this study, with a median Lt per millimetre ranging from 11.6 to 50.6 N. These values were of the same order of magnitude as the mean load measured in vivo by Robinson et al. (35.6 N) [25] and Burstein et al. (20 N) [26], whereas, in vivo, there is the formation of a callus, which is absent in cadavers.
Lc calculation versus Lt measures
Using the stress-strain data of the tensile tests, we evaluated the median load applied on the distractor during a distraction to be Lc = 11.6 MPa. We obtained comparable data when comparing the estimated load opposing distraction generated from the tensile test Lc to the measured load in the first millimeter of distraction Lt. This result confirmed that the periosteum is the primary tissue opposing distraction, which had been suggested previously without biomechanical proof.
Additionally, it is difficult to compare the aspect of the load mm curves with the J-shaped stress-strain curve of the periosteum. Indeed, considering the dimensions of an adult mandible with a 40 mm stretched periosteal segment, a 5% axial strain would correspond to only 2 mm of distraction. Moreover, provided that the mean relaxation characteristic time τ2 was 11.6 s, and that there was a mandatory latency period varying from 10 to 30 s between each activation (to save the measures and tare the torque meter), the periosteum significantly relaxed during these periods.
Although a viscoelastic model could be applied to tensile tests, it is difficult to compare beyond the first millimeter of distraction. Indeed, beyond this value the strain of the periosteum is above the maximum strain we explored in the tensile tests (see Fig 3). According to our assumptions, an activation of 2 mm would imply a strain of 20% of the periosteum in the lingual side of the mandibule. At this strain value, based on preliminary tensile tests (data not shown), some periosteum specimens ruptured. To modelize cadaveric distraction during the whole distraction process, we should assume a periosteum strain up to 200% (20 mm-distraction). Then, three issues appeared: (1) behavior of sharpey fibers out of the lifted area; (2) periosteum partial rupture; (3) relaxation between each activation. (1) In distraction, sharpey fibers are altered during the periosteum lifting and we considered that out of this area, these fibers are attached to the bone. However during 20 mm distraction, shear forces exerted on sharpey fibers could lead to their rupture. Thus, with this protocol, the periosteum slides on the bone out of the lifted area. Then the periosteum stretched area is bigger than the area considered in our calculation. (2) During preliminary experimentations, we showed periosteum rupture around 25% strain. A partial rupture of the periosteum should be considered during 200% strain. A total rupture is unlikely, since we showed a stiffening of the load during cadaver distraction. (3) To modelize distraction, the relaxation of the periosteum should be taken into account to determine the new initial state before each activation of 1 mm.
Conclusions
This study provided the first features to understand the biomechanical role of the periosteum during distraction. We showed that the periosteum is directly stretch by the distraction activation and that the periosteum is the principal contributor to the stress involved in a distraction at the beginning of activation. Further studies should be performed to understand the origin of the stress involve in a distraction beyond the first millimeters. While the conclusion of our study seems clear, these results could be empowered and deepen by increasing the number of cadavers. The loads of distraction measured on cadaver will be useful for the development of new types of distractor devices. Sizing of activators (motorized or magnetic) depends directly on the stress involved in the distraction. Further studies with an increased number of samples, an optimized experimental method such as microstructure recording during distraction, and a microstructure-based model are required to propose a viscoelastic model of the periosteum based on this microstructure. In fine, the objective will be to predict the stress involved during in vivo from data obtained from the microstructural analysis of periosteum biopsies.
Supporting information
Acknowledgments
This cadaveric experiment was supported by l’Ecole de chirurgie de l’APHP.
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
This cadaveric experiment was supported by l’Ecole de chirurgie de l’APHP, Idex Paris Saclay grant to JB, a Strasbourg Osteosynthesis Research Group research grant to NK, the Coline Association to NK, and Association Françaises des Familles ayant un enfant atteint du Syndrome Silver Russell (SSR) ou né Petit pour l’Age Gestationnel (PAG) et leurs amis to NK. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Adolphs N, Ernst N, Menneking H, Hoffmeister B. Significance of distraction osteogenesis of the craniomaxillofacial skeleton—A clinical review after 10 years of experience with the technique. Journal of Cranio-Maxillofacial Surgery. 2014;42(6):966–975. doi: 10.1016/j.jcms.2014.01.018 [DOI] [PubMed] [Google Scholar]
- 2. Diner PA, Kollar EM, Martinez H, Vazquez MP. Intraoral distraction for mandibular lengthening: a technical innovation. Journal of Cranio-Maxillo-Facial Surgery: Official Publication of the European Association for Cranio-Maxillo-Facial Surgery. 1996;24(2):92–95. doi: 10.1016/S1010-5182(96)80019-9 [DOI] [PubMed] [Google Scholar]
- 3. McCarthy JG, Stelnicki EJ, Mehrara BJ, Longaker MT. Distraction osteogenesis of the craniofacial skeleton. Plastic and Reconstructive Surgery. 2001;107(7):1812–1827. doi: 10.1097/00006534-200106000-00029 [DOI] [PubMed] [Google Scholar]
- 4. Verlinden CRA, van de Vijfeijken SECM, Jansma EP, Becking AG, Swennen GRJ. Complications of mandibular distraction osteogenesis for congenital deformities: a systematic review of the literature and proposal of a new classification for complications. International Journal of Oral and Maxillofacial Surgery. 2015;44(1):37–43. doi: 10.1016/j.ijom.2014.07.009 [DOI] [PubMed] [Google Scholar]
- 5. Goldwaser BR, Papadaki ME, Kaban LB, Troulis MJ. Automated Continuous Mandibular Distraction Osteogenesis: Review of the Literature. Journal of Oral and Maxillofacial Surgery. 2012;70(2):407–416. doi: 10.1016/j.joms.2011.01.042 [DOI] [PubMed] [Google Scholar]
- 6. Saman M, Abramowitz JM, Buchbinder D. Mandibular osteotomies and distraction osteogenesis: evolution and current advances. JAMA facial plastic surgery. 2013;15(3):167–173. doi: 10.1001/jamafacial.2013.44 [DOI] [PubMed] [Google Scholar]
- 7. Boisson J, Strozyk H, Diner P, Picard A, Kadlub N. Feasibility of magnetic activation of a maxillofacial distraction osteogenesis, design of a new device. Journal of Cranio-Maxillofacial Surgery. 2016;44(6):684–688. doi: 10.1016/j.jcms.2016.03.009 [DOI] [PubMed] [Google Scholar]
- 8. Evans SF, Chang H, Knothe Tate ML. Elucidating Multiscale Periosteal Mechanobiology: A Key to Unlocking the Smart Properties and Regenerative Capacity of the Periosteum? Tissue Engineering Part B: Reviews. 2013;19(2):147–159. doi: 10.1089/ten.teb.2012.0216 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. McBride SH, Evans SF, Knothe Tate ML. Anisotropic mechanical properties of ovine femoral periosteum and the effects of cryopreservation. Journal of Biomechanics. 2011;44(10):1954–1959. doi: 10.1016/j.jbiomech.2011.04.036 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Popowics TE, Zhu Z, Herring SW. Mechanical properties of the periosteum in the pig, Sus scrofa. Archives of oral biology. 2002;47(10):733–741. doi: 10.1016/S0003-9969(02)00065-1 [DOI] [PubMed] [Google Scholar]
- 11. Yiannakopoulos CK, Kanellopoulos AD, Trovas GP, Dontas IA, Lyritis GP. The biomechanical capacity of the periosteum in intact long bones. Archives of Orthopaedic and Trauma Surgery. 2007;128(1):117–120. doi: 10.1007/s00402-007-0433-5 [DOI] [PubMed] [Google Scholar]
- 12. Bertram JE, Polevoy Y, Cullinane DM. Mechanics of avian fibrous periosteum: tensile and adhesion properties during growth. Bone. 1998;22(6):669–675. doi: 10.1016/S8756-3282(98)00035-0 [DOI] [PubMed] [Google Scholar]
- 13. McBride SH, Dolejs S, Brianza S, Knothe U, Knothe Tate ML. Net Change in Periosteal Strain During Stance Shift Loading After Surgery Correlates to Rapid De Novo Bone Generation in Critically Sized Defects. Annals of Biomedical Engineering. 2011;39(5):1570–1581. doi: 10.1007/s10439-010-0242-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Saunders MM, Lee JS. The Influence of Mechanical Environment on Bone Healing and Distraction Osteogenesis. Atlas of the Oral and Maxillofacial Surgery Clinics. 2008;16(2):147–158. doi: 10.1016/j.cxom.2008.04.006 [DOI] [PubMed] [Google Scholar]
- 15. Zeng YJ, Sun Xp, Yang J, Wu Wh, Xu Xh, Yan Yp. Mechanical properties of nasal fascia and periosteum. Clinical Biomechanics. 2003;18(8):760–764. doi: 10.1016/S0268-0033(03)00136-0 [DOI] [PubMed] [Google Scholar]
- 16. Bell WH, Guerrero CA. Mandibular lengthening by Distraction Osteogenesis In: Distraction Ostegenesis of the facial skeleton. PMPH-USA; 2007. p. 327–340. [Google Scholar]
- 17. Lynch B, Bancelin S, Bonod-Bidaud C, Gueusquin JB, Ruggiero F, Schanne-Klein MC, et al. A novel microstructural interpretation for the biomechanics of mouse skin derived from multiscale characterization. Acta biomaterialia. 2017;50:302–311. doi: 10.1016/j.actbio.2016.12.051 [DOI] [PubMed] [Google Scholar]
- 18. Fung YC. Bioviscoelastic solids In: Biomechanics. Springer; 1993. p. 242–320. Available from: http://link.springer.com/chapter/10.1007/978-1-4757-2257-47. [Google Scholar]
- 19. Foolen J, Van Donkelaar CC, Soekhradj-Soechit S, Ito K. European Society of Biomechanics SM Perren Award 2010: an adaptation mechanism for fibrous tissue to sustained shortening. Journal of biomechanics. 2010;43(16):3168–3176. doi: 10.1016/j.jbiomech.2010.07.040 [DOI] [PubMed] [Google Scholar]
- 20. Bancelin S, Lynch B, Bonod-Bidaud C, Ducourthial G, Psilodimitrakopoulos S, Dokládal P, et al. Ex vivo multiscale quantitation of skin biomechanics in wild-type and genetically-modified mice using multiphoton microscopy. Scientific Reports. 2015;5(1). doi: 10.1038/srep17635 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Comninou M, Yannas IV. Dependence of stress-strain nonlinearity of connective tissues on the geometry of collagen fibers. Journal of Biomechanics. 1976;9(7):427–433. doi: 10.1016/0021-9290(76)90084-1 [DOI] [PubMed] [Google Scholar]
- 22. Gupta HS, Seto J, Krauss S, Boesecke P, Screen HRC. In situ multi-level analysis of viscoelastic deformation mechanisms in tendon collagen. Journal of Structural Biology. 2010;169(2):183–191. doi: 10.1016/j.jsb.2009.10.002 [DOI] [PubMed] [Google Scholar]
- 23. Manschot JF, Brakkee AJ. The measurement and modelling of the mechanical properties of humanskin in vivo–II. The model. Journal of Biomechanics. 1986;19(7):517–521. doi: 10.1016/0021-9290(86)90125-9 [DOI] [PubMed] [Google Scholar]
- 24. Puxkandl R, Zizak I, Paris O, Keckes J, Tesch W, Bernstorff S, et al. Viscoelastic properties of collagen: synchrotron radiation investigations and structural model. Philosophical Transactions of the Royal Society B: Biological Sciences. 2002;357(1418):191–197. doi: 10.1098/rstb.2001.1033 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Robinson RC, O’Neal PJ, Robinson GH. Mandibular distraction force: Laboratory data and clinical correlation. Journal of Oral and Maxillofacial Surgery. 2001;59(5):539–544. doi: 10.1053/joms.2001.22688 [DOI] [PubMed] [Google Scholar]
- 26. Burstein FD, Lukas S, Forsthoffer D. Measurement of torque during mandibular distraction. Journal of Craniofacial Surgery. 2008;19(3):644–647. doi: 10.1097/SCS.0b013e31816ae1a8 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the paper and its Supporting Information files.