Abstract
Purpose
This study aims to assess the accuracy of source position verification during high‐dose rate (HDR) prostate brachytherapy using a novel, in‐house developed two‐dimensional (2D) diode array (the Magic Plate), embedded exactly below the patient within a carbon fiber couch. The effect of tissue inhomogeneities on source localization accuracy is examined.
Method
Monte Carlo (MC) simulations of 12 source positions from a HDR prostate brachytherapy treatment were performed using the Geant4 toolkit. An Ir‐192 Flexisource (Isodose Control, Veenendaal, the Netherlands) was simulated inside a voxelized patient geometry, and the dose deposited in each detector of the Magic Plate evaluated. The dose deposited in each detector was then used to localize the source position using a proprietary reconstruction algorithm.
Results
The accuracy of source position verification using the Magic Plate embedded in the patient couch was found to be affected by the tissue inhomogeneities within the patient, with an average difference of 2.1 ± 0.8 mm (k = 1) between the Magic Plate predicted and known source positions. Recalculation of the simulations with all voxels assigned a density of water improved this verification accuracy to within 1 mm.
Conclusion
Source position verification using the Magic Plate during a HDR prostate brachytherapy treatment was examined using MC simulations. In a homogenous geometry (water), the Magic Plate was able to localize the source to within 1 mm, however, the verification accuracy was negatively affected by inhomogeneities; this can be corrected for by using density information obtained from CT, making the proposed tool attractive for use as a real‐time in vivo quality assurance (QA) device in HDR brachytherapy for prostate cancer.
Keywords: brachytherapy, diode, in vivo, Magic Plate, source tracking
1. INTRODUCTION
When used in combination with external beam radiotherapy (EBRT) in the form of a boost, high‐dose rate (HDR) brachytherapy has been shown to be a safe and effective treatment modality for intermediate‐ and high‐risk prostate cancer.1, 2 Despite recent technological developments in the field of brachytherapy, such as image‐guided brachytherapy,3, 4 treatment planning,5 electromagnetic tracking,6 and applicator development,7, 8, 9 poor execution of HDR prostate brachytherapy can still have a significant effect on patient outcomes. Furthermore, the incidences of errors that may occur in the practice of HDR prostate brachytherapy is relatively unknown, and limited options exist for independent routine monitoring of treatment delivery. There are a number of published documents by the International Commission on Radiological Protection (ICRP)10, 11 as well as the International Atomic Energy Association (IAEA)12 describing errors that have occurred in HDR brachytherapy. Many of these errors are related to human miscalculations, and less often, due to machine or computational malfunction. The likelihood of remote afterloader malfunction is generally considered extremely low; however, small deviations from the plan in source dwell position and time can result in significant errors in the dose delivered to the patient.13
The American Association of Physicists in Medicine (AAPM) Radiation Therapy Task Group (TG) No. 5914 recommends that institutions employ a quality assurance (QA) program that exploits redundancy and review the entire treatment planning and delivery process to isolate any actions susceptible to errors. The report suggests that incidence of these errors may be reduced by the introduction of pretreatment QA in the time between treatment planning and delivery. Further to this, another AAPM Report from TG 5615 recommends that the source position, source dwell time, and transit time be quantified by the medical physicist on a regular basis. A combination of these regular and pretreatment QA checks, along with a well‐documented treatment planning and delivery protocol will go a long way to ensuring safe and successful delivery of HDR brachytherapy treatment plans. However, this type of QA program will not safeguard the HDR brachytherapy delivery from all types of errors. An ideal system for HDR brachytherapy treatment verification should be able to provide real‐time identification of the dwell positions, measure the dwell and transit times, and compare these parameters with the prescribed treatment plan both before and during treatment.16
Real‐time source identification of dwell positions during HDR prostate brachytherapy treatments based on electronic portal imaging devices (EPIDs) have been performed previously.17, 18, 19 In one study, the authors retrospectively compared the planned vs measured source positions using an EPID embedded into the couch for eight treatment fractions, and the mean linear distance between the planned and measured dwell positions was found to be 1.8 mm (range 0.7–3.9 mm).17 However, these EPID‐based devices suffer from low frame rates and slow readout electronics, resulting in loss of data for short dwell times.18
This study aims to investigate the feasibility of using a two‐dimensional (2D) diode array, the “Magic Plate” (MP), developed at the Centre for Medical Radiation Physics (CMRP), University of Wollongong (UoW), Australia, for in vivo source position verification of a HDR prostate brachytherapy treatment. Previous studies performed with the MP in homogeneous phantom media and have reported source localization accuracy of less than 1 mm16, 20, 21 and temporal resolution of 1 ms,20 making it an ideal device for real‐time source position verification. It is hypothesized that the introduction of heterogeneous media associated with the patient geometry may compromise the accuracy of source localization using this device.
To assess the feasibility of source localization using the MP in the presence of patient‐related heterogeneities, MC simulations were performed using the Geant4 toolkit (v4.10.01).22, 23 During the HDR prostate brachytherapy treatment simulations, the Flexisource Ir‐192 source (Isodose Control, Veenendaal, the Netherlands) was simulated inside a voxelized patient geometry, and the dose deposited within the sensitive volume of each detector in the couch embedded 11 × 11 diode array was evaluated. The simulated detector dose was then used to determine the distance of all detectors in the array to each of the simulated source positions. Finally, the source position was determined using an iterative procedure where the source position is first estimated, and then repeatedly refined based upon the agreement of the predicted geometric distance from the source to the detectors against those determined by the detectors in the array.
2. MATERIALS AND METHODS
2.A. Ir‐192 Flexisource
The geometric design of the Flexisource model was obtained from the study performed by Granero et al.24 and is shown in Fig. 1. A detailed description of the Flexisource model used in the simulations is included in Appendix I.
2.B. Magic Plate diode array
The MP is a 2D silicon diode array developed at the CMRP, UoW, Australia, originally as a tool for intensity modulated radiation therapy (IMRT) QA.25, 26 The MP has since been validated as a tool for Ir‐192 source position verification with focus on pretreatment quality assurance.16, 21
The MP consists of an 11 × 11 array of epitaxial diodes mounted on a 0.6 mm Kapton substrate using the “drop‐in” technique. The structure of the MP is modeled in the Geant4 MC simulations in this study using the description by Wong, et al.25 and is shown in Figs. 2(a) and 2(b).
2.C. TG‐43 simulations
Before using the Flexisource model in MC studies, the physics modeling within the prospective MC simulations was validated against benchmark data via TG‐43 simulations. The AAPM TG43‐U1 report27 recommends that MC simulations used to obtain TG‐43 parameters be performed with the source placed in the center of a spherical 400 mm radius water phantom, so as to simulate an unbounded phantom.28 The phantom geometry used in this simulation was a 400 mm radius spherical phantom comprised of liquid water with physical density 0.998 g/cm3. The density of 0.998 g/cm3 was chosen so as to simulate the density of liquid water at 22°C as is recommended in TG43‐U1.27
The spectrum of gamma energies emitted from the Ir‐192 source was obtained from the NuDat database.29 The β spectrum was not considered in the study since its contribution to the dose delivered beyond the stainless steel shell is negligible.24, 30 A total of 109 primary photons were generated for each simulation run, and a total of 10 simulation runs were performed. Results from each run were averaged and the standard deviation (k = 1) calculated.
The interaction processes for photons (the photoelectric effect, Compton scattering and Rayleigh scattering) are modeled using the Geant4 Livermore Low Energy Package. The interactions cross‐sections tabulation was taken from the EPDL97 database.31 In order to improve the efficiency of the simulations, the linear track‐length kerma estimator32 was utilized with a photon cutoff energy of 250 eV.
Interactions for electrons (multiple scattering, ionization, and bremsstrahlung) are also modeled using the Geant4 Livermore Low Energy Package. Secondary particles with range less than 1 μm are assumed to deposit the dose locally in the interaction voxel.30
To obtain the dose rate in polar coordinates and calculate the TG‐43 parameters, the dose was scored in spherical sections with thickness of 0.5 mm (from 0 to 200 mm) and angular resolution of 1° (from 0° to 180°) concentric to the longitudinal axis of the source. The thickness and resolution of the voxels were chosen so as to ensure the effects of volume averaging was less than 0.1% for distances greater than 5 mm from the source.33 To calculate the absorbed dose in each of the spherical sections, the total energy deposited in each section was obtained and divided by the total mass of the section.
The Ir‐192 source is located at the origin of the calculation volume with its longitudinal axis placed along the y axis of the coordinate system, as shown in Fig. 1. To calculate the radial dose function, the absorbed dose along the z axis was normalized to the absorbed dose at 10 mm from the center of the source, before being divided by the normalized (at z = 10 mm) line source geometry function, as per eq. (A6) of the AAPM TG43‐U1 report.27
To calculate the 2D anisotropy function, for a given radial distance from the center of the source (r) the absorbed dose was plotted as a function of the angle from the longitudinal axis (θ), normalized to θ = 90°, and then divided by the normalized (again at θ = 90°) line source geometry function, as shown in Equation 8 of the AAPM TG43‐U1 report.27
2.D. Source position verification simulations
Once the Flexisource model had been validated through TG‐43 simulations, the same source was simulated inside a voxelized patient model, and the dose deposited in each of the diodes in the carbon couch embedded MP was tallied. The MP was modeled in the source position verification simulations as described in Section 2.B, embedded inside a 120‐mm thick carbon couch, offset 5 mm from its anterior surface.
The patient model was created by converting a DICOM CT study set from a prostate HDR brachytherapy treatment into a voxelized model that can be used in Geant4, as shown in Fig. 3. This was achieved by first converting the Hounsfield unit (HU) numbers to a mass density value using a CT–density curve, and then converting from mass density to a material using a look up table.34, 35, 36 Once imported into the simulation, a geometrical phantom is created, within which is an array of voxels containing the materials (and their compositions) determined from the HU numbers.36 The compositions and the densities of materials used in the simulations were obtained from the AAPM TG 186 Report.37 The voxel size was set to 3 × 3 × 3 mm3 in this study, to model an accurate geometrical definition of patient‐related inhomogeneities and prevent prohibitively long simulation times.
A selection of 12 source positions from a HDR prostate brachytherapy treatment plan created in the Oncentra Brachytherapy treatment planning system (Elekta Brachytherapy, Veenendaal, the Netherlands) were used in the simulations. The source position coordinates were selected as three consecutive source positions from four catheters. The catheters were selected such that they spanned the extent of the prostate (L–R and A–P), to determine if consecutive source positions along a catheter could be localized by the MP at the maximum and minimum distances expected in a clinical HDR prostate brachytherapy treatment. The step size of the source used in the treatment planning system was 3 mm.
The same geometrical source model as described in Section 2.A was used in the simulations, along with the same gamma spectrum and interaction processes described in Section 2.C. To prevent overlapping volumes, which causes tracking errors in simulations, a parallel geometry was used to place the source within the patient geometrical phantom at the planned dwell positions.38 To calculate the absorbed dose in each of the sensitive silicon volumes, the total energy deposited in each volume was obtained and divided by the total mass of the volume. Each source position was simulated with 109 primary photons for each simulation run. A total of 20 simulation runs were performed for each source position; results from each run were averaged and the standard deviation (k = 1) was less than 1%.
Each of the source localization simulations was then repeated with each voxel in the patient geometry overridden to the density of water, to compare the source localization accuracy of the MP with and without the presence of patient‐related inhomogeneities.
To determine the distance of each of the 12 source positions to all detectors in the array (αi), a separate group of simulations were first performed to determine the dose deposited in a single detector placed at 10 mm from the source (D10), along the z axis (as shown in Fig. 1) in a water phantom. A total of 10 simulations of 109 primary photons were performed for this configuration, and the dose deposited in the single detector averaged across the 10 simulations. This average dose was then used to normalize the dose from each detector in the patient geometry simulations (Di), before converting the relative dose to radial distance.
(1) |
The radial distance from each detector to each source position can then be determined by converting the relative diode dose to distance via a fit of the TG‐43 parameters calculated in Section 2.C. This approach assumes that the diodes are present within a homogeneous water phantom, when in fact the diode dose was calculated within the heterogeneous patient voxelized phantom. The source positions were determined using an iterative procedure (Appendix 2).39, 40 Once an initial estimation of the source position is found, a correction factor is then applied to the response of each of the MP detectors to take into account the angular dependence of the detectors. The source position is then re‐estimated using the above method but uses the initial estimated source position of the previous calculation. Finally, the calculated source position is compared to known source positions obtained from the Oncentra Brachytherapy treatment planning system.
3. RESULTS
3.A. TG‐43 simulations
The radial dose function and anisotropy function (at a radial distance of 10 mm) calculated in this study, using an active length of 3.5 mm for the calculation of GL(r,θ), are compared to Granero et al.24 and Taylor & Rogers41 in Figs. 4 and 5, respectively.
The calculated radial dose function from this study was shown to agree (discrepancies < 1%) with both the Granero24 and Taylor & Rogers41 benchmark datasets within the calculated uncertainty (1.2%) in the radial distance range of 1–200 mm. The 2D anisotropy was also found to agree with the two benchmark datasets to within 1% across the investigated range, verifying the source model and simulation physics were adequate to be used in the study. Larger discrepancies can be observed at polar angles between 0–15° and 165–180° due to the minor variations in source capsule modeling between the studies. In this study, the non‐cable end weld of the stainless steel shell is modeled as a cylinder of length 0.65 mm and diameter 0.85 mm, whereas in the studies by Granero et al.24 and Taylor & Rogers41 the end weld is modeled as a 0.108 mm thick conical section with a half angle of 23° and the radius of the face being 0.17 mm. This conical section was then attached to a 0.49 mm long solid cylindrical section to complete the end weld. This minor difference in the end weld modeling was not observed to have a significant effect on the calculated anisotropy. There is also a substantial decrease in magnitude of the dose scored in the voxels close to the ends of the source encapsulation due to the significant attenuation through the end welds, this leads to an increased statistical uncertainty and may also contribute to the larger discrepancies in calculated anisotropy between the studies.
3.B. Source position verification simulations
The average difference between MP predicted and actual source positions was found to be 2.1 ± 0.8 mm (k = 1) when all detectors in the array were used in the localization algorithm. Table 1 summarizes the localization results in three dimensions, along with the calculated three‐dimensional difference vector, for different number of detectors used in the source localization algorithm. When not all detectors in the array were used in the source position verification algorithm, the detectors with the highest deposited dose were chosen. As can be seen from Table 1, the MP could localize the source to within 1 mm in the X and Y directions (left/right and superior/inferior directions, respectively). However, it consistently overestimated the distance in the Z direction (anterior/posterior direction), with an average error of 1.9 mm.
Table 1.
Number of detectors used | 9 | 25 | 49 | 81 | 121 |
---|---|---|---|---|---|
X | 0.5 ± 1.0 | 0.2 ± 0.4 | 0.0 ± 0.2 | 0.0 ± 0.2 | 0.2 ± 0.2 |
Y | 0.5 ± 1.1 | 0.0 ± 0.5 | 0.0 ± 0.3 | 0.2 ± 0.3 | 0.5 ± 0.5 |
Z | 1.8 ± 0.5 | 1.8 ± 0.5 | 1.9 ± 0.5 | 2.0 ± 0.5 | 2.0 ± 0.5 |
3D | 2.4 ± 1.0 | 1.9 ± 0.5 | 1.9 ± 0.5 | 2.0 ± 0.5 | 2.1 ± 0.6 |
The source position verification simulations were then repeated using the exact same methods, but with each voxel in the patient geometry assigned a density of water [Figs. 6(a)–6(d)]. The heterogeneous and water only results were compared by means of a Student's t test (P < 0.05) in each of the X, Y, and Z directions. Only the Z direction differences were found to be statistically significant (P < 0.001).
The heterogeneous results were used for a one‐way ANOVA42 analysis (P < 0.05) to determine if there was a statistically significant difference in the localization accuracy depending on the number of detectors used. The only significant difference was found for the three‐dimensional vector (P < 0.001). Subsequent Student's t tests were performed to compare the datasets for the three‐dimensional vector. From this, it was found that only the dataset with nine detectors had statistically significant differences to the other groups.
4. DISCUSSION
The uncertainties quoted in this study have been evaluated using the combination of both type A and type B uncertainties combined in quadrature, as recommended in AAPM TG43‐U127 and AAPM TG138.43 A summary of the uncertainty budget for both the TG‐43 simulations and the source tracking simulations is presented in Table 2.
Table 2.
Type A | Type B | |
---|---|---|
Statistical variation in absorbed dose determination from repeated MC simulation runs | 1% | |
Variations of the source geometry from one source to another in manufacturing process24 | 0.5% | |
Uncertainty in cross‐section library data for Ir‐19244 | 0.5% | |
Effect of volume averaging on absorbed dose calculation in sensitive volumes45 | 0.1% | |
Uncertainty in composition of tissues used during source tracking simulations46, a | 2% | |
Total TG‐43 simulation uncertainty | 1.2% | |
Total source tracking simulation uncertainty | 2.3% |
Applies to source tracking simulations only.
In a water only geometry, the Magic Plate was able to localize the source to within 1 mm. The source localization accuracy, however, was found to decrease with the introduction of inhomogeneities. This decrease in accuracy of source position localization due to the presence of inhomogeneities was found to be primarily in the direction perpendicular to the diode array (z direction). This is due to the source localization algorithm39, 40 z direction estimate being more sensitive to the changing ratio of primary to secondary photons due to the presence inhomogeneities and increased source to detector distance.
The distance estimate in the z direction is depends directly on the absolute dose deposited in the detector array, whereas source localization in the x and y directions depends only on the relative difference between the distance estimate for each detector and therefore is less sensitive to inhomogeneities. Small inhomogeneities, however, can affect the x and y estimate if a smaller number of detectors are used in the localization algorithm.39, 40
This indicates that to track the Ir‐192 source with the desired accuracy during HDR prostate brachytherapy treatments a correction may be required, based on density information obtained from the patient CT scan performed prior to treatment. This information could be used, along with a model‐based dose calculation algorithm37 built into the localization algorithm to more accurately predict the source to detector distances.
Moreover, it was found that source localization accuracy can be improved with an increased number of detectors used in the localization algorithm. By increasing the number of detectors used in the localization algorithm, a redundancy is built in to reduce the uncertainty introduced due to small heterogeneous media in the patient geometry.47 The increased number of detectors is also beneficial due to the relatively isotropic dose profile at large source to detector distances.40
Previous studies have shown that source position verification using EPIDs is achievable. However, these studies have also shown that source position verification using EPIDs is restricted by the limited frame rate and readout electronics of the devices and can result in a significant number of dwell positions not being captured by the EPID when performing source position verification.18 Source position verification using EPIDs also requires large and expensive systems that have limited availability. This study has shown that similar source position verification accuracy to EPIDs17, 18, 19 may be achieved with the MP system. Furthermore, the MP system has been shown to have a superior timing resolution of less than 1 ms.20 As such, the MP delivers a dedicated, inexpensive HDR brachytherapy in vivo source position verification system with superior timing resolution that can easily be mass produced and is practical for routine clinical use.
The results of this study, along with previously published experimental results,16, 40 indicate that the MP will have sufficient sensitivity to detect errors in the order of 1–2 mm during the delivery of HDR prostate brachytherapy treatments when embedded in a carbon fiber couch beneath the patient. Such errors may be due to incorrect catheter connection or incorrect source strength. However, discrepancies less than 1–2 mm in catheter reconstruction, and small movements of catheters in the time between simulation and treatment may not be detectable unless patient‐related heterogeneities can be taken into account during source position verification. Furthermore, one of the most significant challenges for clinical implementation of source position verification using the MP system is the registration of MP and patient coordinate systems. Previous studies have overcome this challenge by using stereoscopic imaging,17, 48 and electromagnetic tracking technology.49, 50 Overcoming this challenge will be a focus in future publications.
With the move toward real‐time prostate brachytherapy treatment planning based on transrectal ultrasound imaging, future studies will also aim to examine the effect of the ultrasound probe on source localization accuracy and attempt to optimize the MP position to minimize the effects of heterogeneities.
5. CONCLUSION
Source localization using the Magic Plate during a HDR prostate brachytherapy treatment was examined using MC simulations. In a homogenous geometry, the Magic Plate was able to localize the source to within 1 mm. The effect of tissue inhomogeneities in the patient geometry on source localization accuracy was also examined and was found to increase the difference between Magic Plate predicted and known source positions from the brachytherapy treatment planning system to 2.1 ± 0.81 mm (k = 1). However, this accuracy can be improved using density information obtained from CT with the MP accurately registered to the patient geometry, making the proposed tool attractive for use as a real‐time in vivo QA device in HDR brachytherapy for prostate cancer.
CONFLICTS OF INTEREST
The authors have no conflicts of interest to report.
ACKNOWLEDGMENTS
The authors acknowledge the University of Wollongong Information Management & Technology Services (IMTS) for computing time on the UoW High Performance Computing Cluster. This research was supported by the National Computer Merit Allocation Scheme (NCMAS) project zi60, as well as the Australian Government Research Training Program (RTP).
APPENDIX 1. DESCRIPTION OF IR‐192 FLEXISOURCE MODEL
1.1.
The Ir‐192 core is modeled as a pure Iridium cylinder of length 3.5 mm and diameter 0.6 mm; it has a physical density of 22.42 g/cm3.
The active Ir‐192 core is surrounded by a stainless steel shell of length 4.6 mm, 0.85 mm outer diameter, and inner diameter of 0.67 mm. This results in a shell thickness of 0.09 mm. The composition by weight of the stainless steel shell is modeled as follows: Fe 67.92%, Cr 19%, Ni 10%, Mn 2%, Si 1%, and C 0.08%, and the physical density is 7.999 g/cm3.
The non‐cable end weld of the stainless steel shell is modeled as a cylinder of length 0.65 mm and diameter 0.85 mm. The cable end weld of the stainless steel shell is modeled as a partial cone of maximum diameter 0.85 mm, minimum diameter of 0.5 mm, and length 0.4 mm.
Finally, the stainless steel cable is modeled as a cylinder of 5 mm length and 0.5 mm diameter, as recommended by Granero et al.24 The space between the outer stainless steel shell and the inner Iridium core was modeled as dry air with a physical density of 1.20 mg/cm3.
APPENDIX 2. SOURCE POSITION VERIFICATION ALGORITHM
2.1.
Based on the derived distances (ri) of each diode (i) in the array to the source, as described in Section 2.D, the estimated source position, Ses(a,b,c) can be calculated. The geometrical distance, di, between Ses and the coordinate of the i‐th detector Di(xi,yi,zi) is calculated by
(A1) |
To determine the true source position, the geometrical distance, di, is fitted to the derived distance, ri, by adjusting the estimated source position. Employing a nonlinear least squares fit method to determine the estimated source position. In least squares fitting, the estimate of error assessment can be expressed as the sum of squares of the relative error, Χ2
(A2) |
and assumes that the derived distance, ri, is correct. As there is an uncertainty associated with deriving ri, if the estimated source position were equal to the true source position, then calculating the square of the sums of the percentage difference of the value di and ri would result in a minimum value.
To determine a source position that gives the minimal value to the estimate Χ2 a multivariable Newtown's method approach is adopted. The Newton's method is used in this case to determine the roots of a function by finding successively better approximations. In this analysis, it is necessary to determine the minimum values of Χ2 for all three dimensions of the estimated source position, and can be expressed as
(A3) |
The Newton's method for the three source coordinates can be expressed for the k‐th iteration as
(A4) |
where δa, δb, and δc are the changes made to the source position to produce the improved approximation. These changes can be determined by solving a set of linear equations, expressed in matrix form as
(A5) |
This process is repeated until all δ's are sufficiently small, or until further estimations of the source coordinates fail to reduce Χ2. This approach can converge rapidly to a minimum when close, as all three source coordinates are modified in a single iteration.
The initial guess is determined by the coordinates of the detector with the highest response, Dmax(xmax,ymax,z = 0), as the source is assumed to be closest to this position. The sum of the squares is calculated using the first estimation of the source position,
(A6) |
The source position is then re‐estimated using the above method, but uses the initial estimated source position of the previous calculation.
REFERENCES
- 1. Hoskin PJ, Rojas AM, Bownes PJ, et al. Randomised trial of external beam radiotherapy alone or combined with high‐dose‐rate brachytherapy boost for localised prostate cancer. Radiother Oncol. 2012;103:217–222. [DOI] [PubMed] [Google Scholar]
- 2. Michalski JM, Moughan J, Purdy J, et al. A randomized trial of 79.2 Gy versus 70.2 Gy radiation therapy (RT) for localized prostate cancer. J Clin Oncol (Meeting Abstracts). 2015;33:4. [Google Scholar]
- 3. Tanderup K, El Naqa I, Carlson DJ, Klein EE. Advances in Image‐Guided Brachytherapy. Int J Radiat Oncol Biol Phys. 2017;97:873–875. [Google Scholar]
- 4. Tanderup K, Fokdal LU, Sturdza A, et al. Effect of tumor dose, volume and overall treatment time on local control after radiochemotherapy including MRI guided brachytherapy of locally advanced cervical cancer. Radiother Oncol. 2016;120:441–446. [DOI] [PubMed] [Google Scholar]
- 5. Nkiwane KS, Pötter R, Tanderup K, et al. Single line source with and without vaginal loading and the impact on target coverage and organ at risk doses for cervix cancer Stages IB, II, and IIIB: treatment planning simulation in patients treated with MRI‐guided adaptive brachytherapy in a multicentre study (EMBRACE). Brachytherapy. 2013;12:317–323. [DOI] [PubMed] [Google Scholar]
- 6. Poulin E, Racine E, Binnekamp D, Beaulieu L. Fast, automatic, and accurate catheter reconstruction in HDR brachytherapy using an electromagnetic 3D tracking system. Med Phys. 2015;42:1227–1232. [DOI] [PubMed] [Google Scholar]
- 7. Dadkhah H, Hopfensperger KM, Kim Y, Wu X, Flynn RT. Multisource rotating shield brachytherapy apparatus for prostate cancer. Int J Radiat Oncol Biol Phys. 2017;99:719–728. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Lindegaard JC, Madsen ML, Traberg A, et al. Individualised 3D printed vaginal template for MRI guided brachytherapy in locally advanced cervical cancer. Radiother Oncol. 2016;118:173–175. [DOI] [PubMed] [Google Scholar]
- 9. Safigholi H, Han DY, Mashouf S, et al. Direction modulated brachytherapy (DMBT) for treatment of cervical cancer: a planning study with 192Ir, 60Co, and 169Yb HDR sources. Med Phys. 2017;44:6538–6547. [DOI] [PubMed] [Google Scholar]
- 10. Lopez PO, Cosset JM, Dutreix A, Landberg T. Prevention of accidental exposures to patients undergoing radiation therapy. ICRP Publication 86, Annals of the ICRP; 2000.
- 11. Ashton LP, Cosset J, Levin V, Martinez A, Nag S. Prevention of high‐dose‐rate brachytherapy accidents. ICRP Publication 97, Annals of the ICRP; 2004. [DOI] [PubMed]
- 12. IAEA . Lessons Learned from Accidental Exposures in Radiotherapy. Vienna: International Atomic Energy Agency; 2000. [Google Scholar]
- 13. Okamoto H, Aikawa A, Wakita A, et al. Dose error from deviation of dwell time and source position for high dose‐rate 192Ir in remote afterloading system. J Radiat Res. 2014;55:780–787. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Kubo HD, Glasgow GP, Pethel TD, Thomadsen BR, Williamson JF. High dose‐rate brachytherapy treatment delivery: Report of the AAPM radiation therapy committee task group No. 59. Med Phys. 1998;25:375–403. [DOI] [PubMed] [Google Scholar]
- 15. Nath R, Anderson LL, Meli JA, et al. Code of practice for brachytherapy physics: Report of the AAPM radiation therapy committee task group No. 56. Med Phys. 1997;24:1557–1598. [DOI] [PubMed] [Google Scholar]
- 16. Espinoza A, Petasecca M, Cutajar D, et al. Pretreatment verification of high dose rate brachytherapy plans using the ‘magic phantom’ system. Biomed Phys Eng Expr. 2015;1:025201. [Google Scholar]
- 17. Smith R, Haworth A, Millar J, et al. OC‐0177: clinical implementation of in vivo source position verification in high dose rate prostate brachytherapy. Radiother Oncol. 2015;115:S87–S88. [Google Scholar]
- 18. Smith RL, Haworth A, Hanlon M, et al. Clinical application of pre‐treatment image verification of catheter positions for HDR prostate brachytherapy. Brachytherapy. 2017;16:S114–S115. [Google Scholar]
- 19. Fonseca GP, Podesta M, Bellezzo M, et al. Online pretreatment verification of high‐dose rate brachytherapy using an imaging panel. Phys Med Biol. 2017;62:5440–5461. [DOI] [PubMed] [Google Scholar]
- 20. Espinoza A, Petasecca M, Fuduli I, et al. The evaluation of a 2D diode array in “magic phantom” for use in high dose rate brachytherapy pretreatment quality assurance. Med Phys. 2015;42:663–673. [DOI] [PubMed] [Google Scholar]
- 21. Espinoza A, Beeksma B, Petasecca M, et al. The feasibility study and characterization of a two‐dimensional diode array in “magic phantom” for high dose rate brachytherapy quality assurance. Med Phys. 2013;40:111702. [DOI] [PubMed] [Google Scholar]
- 22. Agostinelli S, Allison J, Amako K, et al. Geant4—a simulation toolkit. Nucl Instrum Methods Phys Res, Sect A. 2003;506:250–303. [Google Scholar]
- 23. Allison J, Amako K, Apostolakis J, et al. Geant4 developments and applications. IEEE Trans Nucl Sci. 2006;53:270–278. [Google Scholar]
- 24. Granero D, Pérez‐Calatayud J, Casal E, Ballester F, Venselaar J. A dosimetric study on the Ir‐192 high dose rate Flexisource. Med Phys. 2006;33:4578–4582. [DOI] [PubMed] [Google Scholar]
- 25. Wong JHD, Fuduli I, Carolan M, et al. Characterization of a novel two dimensional diode array the “magic plate” as a radiation detector for radiation therapy treatment. Med Phys. 2012;39:2544–2558. [DOI] [PubMed] [Google Scholar]
- 26. Alrowaili ZA, Lerch ML, Petasecca M, et al. Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate. J Appl Clin Med Phys. 2016;17:85–98. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Rivard MJ, Coursey BM, DeWerd LA, et al. Update of AAPM Task Group No. 43 Report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31:633–674. [DOI] [PubMed] [Google Scholar]
- 28. Pérez‐Calatayud J, Granero D, Ballester F. Phantom size in brachytherapy source dosimetric studies. Med Phys. 2004;31:2075–2081. [DOI] [PubMed] [Google Scholar]
- 29. N.N.D. Center . Nuclear data from NuDat, a web‐based database maintained by the Nuclear Data Center. Upton, NY: Brookhaven National Laboratory. [Google Scholar]
- 30. Safigholi H, Meigooni AS, Song WY. Comparison of 192Ir, 169Yb, and 60Co high‐dose rate brachytherapy sources for skin cancer treatment. Med Phys. 2017;44:4426–4436. [DOI] [PubMed] [Google Scholar]
- 31. Cullen DE, Hubbell JH, Kissel L. EPDL97: The evaluated photon data library, 97 version. UCRL‐50400. 1997;6:1‐28.
- 32. Williamson JF. Monte Carlo evaluation of kerma at a point for photon transport problems. Med Phys. 1987;14:567–576. [DOI] [PubMed] [Google Scholar]
- 33. Ballester F, Hernández C, Pérez‐Calatayud J, Lliso F. Monte Carlo calculation of dose rate distributions around Ir192 wires. Med Phys. 1997;24:1221–1228. [DOI] [PubMed] [Google Scholar]
- 34. Kimura A, Aso T, Yoshida H, et al. Presented at the Nuclear Science Symposium Conference Record, 2004. IEEE; 2004. (unpublished).
- 35. Aso T, Kimura A, Yamashita T, Sasaki T. Presented at the 2007 IEEE Nuclear Science Symposium Conference Record; 2007. (unpublished).
- 36. Giacometti V, Guatelli S, Bazalova‐Carter M, Rosenfeld AB, Schulte RW. Development of a high resolution voxelised head phantom for medical physics applications. Phys Med. 2017;33:182–188. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Beaulieu L, Carlsson Tedgren Å, Carrier JF, et al. Report of the Task Group 186 on model‐based dose calculation methods in brachytherapy beyond the TG‐43 formalism: current status and recommendations for clinical implementation. Med Phys. 2012;39:6208–6236. [DOI] [PubMed] [Google Scholar]
- 38. Apostolakis J, Asai M, Cosmo G, et al. Presented at the Proceedings of the 2008 IEEE NSS/MIC/RTSD Conference; 2008. (unpublished).
- 39. Rozenfeld A, Zaider M. Method and apparatus for real time dosimetry. Google Patents; 2008.
- 40. Espinoza AA. University of Wollongong; 2016.
- 41. Taylor REP, Rogers DWO. EGSnrc Monte Carlo calculated dosimetry parameters for Ir192 and Yb169 brachytherapy sources. Med Phys. 2008;35:4933–4944. [DOI] [PubMed] [Google Scholar]
- 42. Hoaglin DC, Welsch RE. The hat matrix in regression and ANOVA. Am Stat. 1978;32:17–22. [Google Scholar]
- 43. DeWerd LA, Ibbott GS, Meigooni AS, et al. A dosimetric uncertainty analysis for photon‐emitting brachytherapy sources: Report of AAPM Task Group No. 138 and GEC‐ESTRO. Med Phys. 2011;38:782–801. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Ballester F, Granero D, Pérez‐Calatayud J, Casal E, Puchades V. Monte Carlo dosimetric study of best industries and alpha omega Ir‐192 brachytherapy seeds. Med Phys. 2004;31:3298–3305. [DOI] [PubMed] [Google Scholar]
- 45. Jashni HK, Safigholi H, Meigooni AS. Influences of spherical phantom heterogeneities on dosimetric characteristics of miniature electronic brachytherapy X‐ray sources: Monte Carlo study. Appl Radiat Isot. 2015;95:108–113. [DOI] [PubMed] [Google Scholar]
- 46. Ghorbani M, Salahshour F, Haghparast A, Moghaddas TA, Knaup C. Effect of tissue composition on dose distribution in brachytherapy with various photon emitting sources. J Contemp Brachyther. 2014;6:54. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47. Nakano T, Suchowerska N, Bilek M, et al. High dose‐rate brachytherapy source localization: positional resolution using a diamond detector. Phys Med Biol. 2003;48:2133. [DOI] [PubMed] [Google Scholar]
- 48. Smith RL, Haworth A, Panettieri V, Millar JL, Franich RD. A method for verification of treatment delivery in HDR prostate brachytherapy using a flat panel detector for both imaging and source tracking. Med Phys. 2016;43:2435–2442. [DOI] [PubMed] [Google Scholar]
- 49. Beaulieu L, Racine E, Han DY, et al. Real‐time electromagnetic tracking–based treatment platform for high‐dose‐rate prostate brachytherapy: clinical workflows and end‐to‐end validation. Brachytherapy. 2017;17:103–110. [DOI] [PubMed] [Google Scholar]
- 50. Kellermeier M, Herbolzheimer J, Kreppner S, et al. Electromagnetic Tracking (EMT) technology for improved treatment quality assurance in interstitial brachytherapy. J Appl Clin Med Phys. 2017;18:211–222. [DOI] [PMC free article] [PubMed] [Google Scholar]