Abstract
The optical pickup unit (OPU) within a CD/DVD/Blu-ray drive integrates 780, 650, and 405 nm wavelength lasers, diffraction-limited optics, a high-bandwidth optoelectronic transducer up to 400 MHz, and a nanoresolution x-, z-axis, and tilt actuator in a compact size. In addition, the OPU is a remarkable piece of engineering and could enable different scientific applications such as sub-angstrom displacement sensing, micro- and nanoimaging, and nanolithography. Although off-the-shelf OPUs can be easily obtained, manufacturers protect their datasheets under nondisclosure agreements to impede their availability to the public. Thus, OPUs are black boxes that few people can use for research, and only experienced researchers can access all their functions. This review details the OPU mechanism and components. In addition, we explain how to utilize three commercially available triple-wavelength OPUs from scratch and optimize sensing quality. Then, we discuss scientific research using OPUs, from standard optical drive-based turnkey-biomarker array reading and OPU direct bioapplications (cytometry, optical tweezing, bioimaging) to modified OPU-based biosensing (DNA chip fluorescence scanning, biomolecular diagnostics). We conclude by presenting future trends on optical storage devices and potential applications. Hacking low-cost and high-performance OPUs may spread micro- and nanoscale biosensing research from research laboratories to citizen scientists around the globe.
Keywords: compact disc (CD), digital versatile disc (DVD), Blu-ray, pptical pickup-unit (OPU), nanobio imaging, cytometer, optical tweezer, DNA chip, fluorescence, medical diagnostics
Some years ago, the global market of digitized music, video, and data storage pushed the capacity and speed of optical drives to their limits, with a huge amount of R&D investment. From compact discs (CDs), digital versatile discs (DVDs), to the most recent Blu-ray discs, the data capacity evolved from the megabyte (MB) to the gigabyte (GB). Likewise, mass production and outstanding sales dramatically reduced the cost of optical storage drives while maintaining high quality and performance. Nowadays, Blu-ray players (12× speed, 432 Mb/s) are affordable devices that reliably read 150 nm data pits from a spinning disc at 10,000 rpm.
Optical drive components have been used for research over the past decades, as they adopt an accessible standard of CD/DVD/Blu-ray disc, and the spindle motor can be used for centrifugal lab-on-disc applications.1,2 Most of this research relies on either commercially available optoelectrical sensing systems or customized sensing mechanisms.3−8 A seldom considered component inside optical drives is the optical pickup unit (OPU), which is essential for converting physical data pits into electric signals. A CD/DVD/Blu-ray drive OPU equips at least one triple-wavelength laser and optimized optical components to achieve the diffraction limit of light for high-density data-pit reading. To focus the laser spot perfectly on a spinning and wobbling disc, an objective lens is actuated by a high-bandwidth triple-axis precision actuator. The light intensity signal reflected from the data pits is then transduced by a specialized high-bandwidth optoelectrical component. These functions are compactly integrated within an eraser-sized box, representing a low-cost and high-performance engineering achievement.
Although off-the-shelf OPUs can be unrestrictedly purchased, accessing their full functions can be much more challenging. In fact, every manufacturer protects the OPU datasheets and hardware pin assignment under secrecy, and hence any researcher needs to either sign a nondisclosure agreement or reverse engineer the OPU to access its functions. This technical barrier has prevented researchers, especially those without an engineering background, from utilizing OPUs with research purposes, and only few groups have managed to access all the OPU functions for its use in applications such as sub-angstrom displacement sensing,9 micro- and nanoimaging,10 and nanolithography.11 Interestingly, an OPU-based low-cost atomic force microscope (AFM) has already democratized nanoimaging up to the point that even children can perform nanoscale measurements by themselves.12,13
Before exploiting the full potential of the OPU in a CD/DVD/Blu-ray drive, we must understand its mechanism and main components. We distinguish the types of OPUs that are flexible for scientific research and explain how to increase the signal-to-noise ratio and couple external optical sensors to enhance sensing quality. Then, we further describe how to control the OPU in practice, by using either specific software tools or customized circuits. We dedicate another part of the review to surveying biosensing studies using OPUs, including standard optical drive-based turnkey biosensing, OPU direct bioimaging, and modified OPU-based systems for advanced bioapplications.
Opening the Black Box: The OPU Mechanism
From CDs and DVDs to Blu-ray discs, higher data capacity requires smaller data pits on the same 12-cm-diameter disc, as illustrated in Figure 1.14 Likewise, higher data density demands shorter laser wavelength λ (CD: 780 nm, DVD: 650 nm, Blu-ray disc: 405 nm) and higher numerical apertures (NAs) of objective lenses (CD: ∼0.5, DVD: ∼0.63, Blu-ray disc: 0.85) to focus the laser on nanoscale data pits (CD: 800 nm, DVD: 400 nm, Blu-ray disc: 150 nm). The OPU focuses a polarized laser to the diffraction limit of light, with full width at half-maximum being approximately 800 nm for CD, 530 nm for DVD, and 250 nm for Blu-ray disc.15 In addition, each data pit has a depth of λ/4, and when the laser hits the pit, the reflection is destructively interfered and the OPU receives a low reflection corresponding to the digital signal “0”, whereas if no pit is hit, the higher intensity reflection is translated as the digital signal “1”.
A transparent hard-coat polycarbonate cover layer16 (CD: 1.1 mm, DVD: 0.6 mm, Blu-ray: 0.1 mm in thickness) protects the data pits from being destroyed by scratching and has a refractive index of 1.6. This layer is an essential optical component of the storage system, as its absence could cause spherical aberration, creating an optical path difference which could exceed 0.2 of the light wavelength. Consequently, the optical path difference could reduce in more than 20% the reflected laser intensity.17
Figure 2 illustrates the structure of a typical CD/DVD/Blu-ray OPU.18 A triple-wavelength semiconductor laser diode, its most expensive component, emits a laser beam that is collimated, linearly polarized, and focused on a disc by a collimator, a polarized beam splitter, and an objective lens, respectively. The reflected laser from the disc passes through the same optical path back to an astigmatic lens and impinges onto a photodiode integrated circuit (PDIC).
To achieve better signal integrity and lower relative intensity noise, the laser diode is modulated within a frequency from 300 to 500 MHz to mitigate laser coherence.19 Next to the laser diode, a diffraction grating splits the laser into three beams to generate data tracking feedback control signals.20,21 Then, the collimator lens, which is linearly driven by a miniature stepper motor along the laser beam direction, adjusts its position to accurately vary the OPU focal distance. This fine-tuning compensates the effect from both the different disc cover layer thicknesses and the switching target data layers (for single-sided dual-layer DVDs and Blu-ray discs) while reading the disc. The objective lens integrates aspherical and diffractive optical design to focus 780, 650, and 405 nm laser beams to the diffraction limit of light.22 A dichroic filter reflects the laser light to the objective lens, which has reflective coatings for the OPU operation wavelengths. Some OPUs have a clear aperture in the back of the dichroic filter to allow external optical access, despite being preferable to avoid the OPU operation wavelengths. This aperture enables the external coupling of optical components for applications such as optical imaging23 and fluorescence light sensing.24
While reading data pits on a fast-spinning disc, an x-, z-axis and tilt electromagnetic actuator, known as voice coil motor (VCM), allows focusing the laser on the data layer. Three pairs of suspension wires mechanically suspend an objective lens holder and electrically conduct driving signals to the x-, z-axis, and tilt coils inside the holder. The VCM has an operation bandwidth of 20 kHz for moving the objective lens along the z-axis (±1000 μm) for data layer focusing and the x-axis (±350 μm) for following spiral data tracks, whereas tilting (±1°) compensates the data layer angular variation due to disc wobbling. The x- and z-axes usually have a sensitivity of 1 μm/mV to achieve nanoscale resolution with precise diving signals that can be used in various applications.25−27
The PDIC consists of current preamplifiers for photodiodes A to H in Figure 2 (CXA2875GA, Sony Co., Tokyo, Japan), which have an operation bandwidth up to 400 MHz. Given the laser splitting grating, there are three laser spots reflected onto the PDIC, and photodiodes A to H monitor the laser spots and provide signals SA to SH, respectively.
Figure 3 shows the center laser spot shape according to the laser focusing, which is described as an astigmatic method.28 When the laser is focused on an object (z = 0), the laser spot on the PDIC projects a circular shape. When the focal point is off-center, the laser spot projects an elliptical shape with the major axis along photodiodes A–C (SFE > 0) or B–D (SFE < 0). This changing shape can be expressed as focus error signal SFE = (SA + SC) – (SB + SD), which is used to control the VCM for focusing the laser precisely on a target and has been proven to exhibit subatomic resolution for AFM applications.28,29
OPUs can feature different functions (reading and writing) and sizes (half height, slim type). From our experience, CD/DVD/Blu-ray OPUs can provide flexibility for scientific research given the characteristics shown in Table 1, which summarizes the core components from ten different types of OPUs.
Table 1. Characteristics of Main Components from Different CD/DVD/Blu-ray OPUs.
component | characteristic | type | value | unit |
---|---|---|---|---|
Objective lens | Lens NA | CD | 0.47–0.53 | - |
DVD | 0.6–0.66 | |||
Blu-ray | 0.85 | |||
Laser spot size (full width at half-maximum) | CD | ∼800 | nm | |
DVD | ∼530 | |||
Blu-ray | ∼250 | |||
S-curve linear region | CD | ∼15 | μm | |
DVD | ∼6 | |||
Blu-ray | ∼0.3 | |||
Working distance (lens to disc) | CD | 0.55–0.86 | mm | |
DVD | 0.63–1.25 | |||
Blu-ray | 0.27–0.61 | |||
Semiconductor laser diode | Wavelength | CD | 770–790 | nm |
DVD | 645–660 | |||
Blu-ray | 400–410 | |||
Power (Average) | CD | 160–1130 | mW | |
DVD | 170–830 | |||
Blu-ray | 340–450 | |||
VCM | Working distance | x-axis | ±350 | μm |
z-axis | ±1000 | μm | ||
Tilt | ±1 | ° | ||
PDIC | Operation bandwidth | CD | 25–90 | MHz |
DVD | 50–130 | |||
Blu-ray | 110–400 |
Driving OPUs from Scratch
The easiest way to gain access to the OPU for biosensing is by using an optical disc diagnostic software, which can be downloaded online for free. There are at least four diagnostic software tools: PlexUtilities v 1.3.3 (Plextor, USA), K-Probe2, QpxTool by Gennady Kozlov, and Opti Drive Control by Erik Deppe. PlexUtilities offers more diagnostic tools, whereas Opti Drive Control has enhanced hardware compatibility.30 Furthermore, a Linux-based software allows one to burn and analyze data sectors on the disc.31 This type of software can read logical error correction codes (ECCs) on the disc including detect parity inner errors and parity inner failures, where the former indicate correctable reading/data errors of a logical ECC block on the disc, and the latter determines whether an ECC block contains uncorrectable errors. The Blu-ray utilizes more powerful ECCs which contains long distance code and burst-indicating subcode to protect the data on the disc.32 The ECCs can indicate scratches or polluted parts on the disc surface. We can exploit these features to drive almost every kind of OPU inside optical drives and perform certain biosensing applications as described later.
Inside the CD/DVD optical drives, it is relatively easy to gain access to the OPU. A very efficient method to determine the input/output pin assignment and operating parameters of the OPU laser, PDIC, and VCM is by measuring the voltage and current of each component during operation. Then, the retrieved parameters can be used to drive the OPU for different applications. Unfortunately, it is much more difficult to reverse engineer triple-wavelength OPUs and determine their pin assignment using this method, because most of them have built-in microcontrollers that receive and deliver digital commands for operation.
Supporting Information of this review provides three controller circuit designs with their parameters for the widely available OPUs PHR-803T (Toshiba Co., Tokyo, Japan) inside the XBOX 360 (Microsoft Co., Redmond, WA, USA), KEM 410 with dual objective lens (Play Station 3; Sony Co., Tokyo, Japan), and SF-BC620L (Sanyo Electric Co., Ltd., Osaka, Japan). The customized controllers grant access to all the functions of these OPUs, including CD/DVD/Blu-ray laser switching and power adjustment, modulation frequency control, VCM x-, z-axis and tilt actuation, PDIC filter, gain and modes switching, and miniature stepper motor control.
Recommendations for OPU-Based Applications
It is recommended to keep the cover layer, whose thickness depends on the operation wavelength, in front of the OPU objective lens to guarantee optimal laser focusing. Microscopy cover glasses provide a similar refractive index (1.47 to 1.5) as the cover layer or one can simply use the disc hard-coat polycarbonate cover layer. Furthermore, the cover layer can be used for sealing microfluidic channels. OPU-based imaging or sensing through different media, such as liquid or gas, demands the optimization of the distance between the cover layer and measurement target.33 Moreover, removing the laser splitting grating can increase the laser intensity up to 25% for improved signal-to-noise ratio in sensing applications.
Compared with CD/DVD laser, the Blu-ray 405 nm laser has a focal point of approximately 250 nm, thus being suitable for high-resolution fluorescence imaging. However, the Blu-ray laser may destroy live-cell samples while imaging depending on the dose. Likewise, the Blu-ray phototoxicity might cause plasma membrane permeabilization, cytoskeleton destruction,34 and DNA damage.35 These aspects should be thoroughly considered when using Blu-ray technology for biosensing and imaging.
OPU-Based Biosensing
Turnkey Biosensing
The above-mentioned software tools can turn standard CD/DVD/Blu-ray optical drives into turnkey biosensing instruments.30 Besides custom-made bio-CD for multiplexing DNA microarray detection,36 biomolecular complex assay spots can be inject-printed onto the cover layer of a standard CD.31 Then, a diagnostic analysis of ECCs can be performed while the OPU reads data through the bioassays to quantitatively extract colorimetric characteristics of different assay spots on the cover layer. Figure 4 illustrates the typical ECC-based method for biotin–streptavidin binding, DNA hybridization, and protein–protein interaction sensing.37 In addition, this method can be employed to detect DNAzyme assay at the part-per-billion level.38
The ECC-based method has been extended for applications such as onsite pregnancy test,39 multiplexed drug abuse diagnostics,40 heavy metal detection,41 and acute myocardial infarction monitoring.41 Furthermore, Blu-ray optical drives have been shown to increase the resolution and sensitivity for biosensing,42,43 reaching a sensitivity and selectivity comparable to standard enzyme-linked immunosorbent assays.
A professional CD/DVD/Blu-ray disc testing or quality control platform, which is normally used during optical disc production, can be repurposed to monitor on-disc biotin–streptavidin binding at a linear speed of 4.0 m/s. Unlike ECC-based biosensing, such platforms provide an analog reflection intensity signal from the disc.44 Following that approach, a commercially available optical drive can connect the OPU PDIC analog output to a data acquisition device, thus enabling more flexibility for biosensing applications, such as reading on-disc biochemical films to determine Ca2+ concentrations with a detection limit of ±5 ppm,45,46 measuring on-disc multiplexed microimmunoassays (e.g., pesticides, antibiotics),47 detecting Haemagglutinin of influenza virus,48 and sensing RNA aptamers generated against reverse transcriptase interaction.49 Furthermore, the OPU can be extracted from the drive for C-reactive protein measurement with a detection limit of 1 pM.50
Cytometer and Optical Trap
Blood cell count and sizing can provide insightful information during AIDS, sepsis, anemia, and leukemia diagnoses. A conventional CD modified with a polydimethylsiloxane microfluidic channel can be read by a standard optical drive for counting microparticles and living cells and determining concentrations based on ECCs.51Figure 5 illustrates a conventional and an OPU-based cytometer, which integrates a lab-on-chip device. The OPU-based cytometer has been used to count individual polystyrene beads between yeast cells,52 erythrocytes,53 Chinese hamster ovary,54 and cattle erythrocytes55 by analyzing focus error signal SFE as depicted in Figure 6.
A DVD burner OPU equips a laser with approximate power of 200 mW and can generate tens of piconewtons of force at the focal point. This force is enough to trap colloid or red blood cells.56 Furthermore, the VCM can precisely steer the cells to different channels. This setup has been used to isolate microparticles and red blood cells nondestructively by controlling the OPU with an Arduino board (Arduino LLC, USA) in a gravity-driven microfluid device,57 as shown in Figure 7.
Direct Bioimaging
The OPU laser focus to the diffraction limit of light has been used as principle for a high-resolution laser scanning microscope to monitor morphological changes in astrocytes and investigate apoptosis triggered by Toxocara canis larval excretory-secretory antigens.58 The OPU-based direct bioimaging system setup is similar to that shown in Figure 5B, except for the microchannel. The OPU laser focuses on astrocytes incubated on a reflective substrate. Then, the substrate is raster scanned by a piezoelectric scanner while the OPU reads out the focus error signal. The astrocytes can be represented by mapping the focus error array into a grayscale image. Figure 8 shows the images obtained using a conventional high-end optical microscope (Figure 8A; magnification 1000×, phase contrast mode) and a laser imaging system based on a DVD OPU (Figure 8B; λ: 650 nm, NA: 0.6). The OPU-based system reveals detailed filament structures of the astrocytes and retrieves a higher contrast than the conventional optical microscope.
Biosensing Using Transducers
Microelectromechanical systems (MEMS) cantilever-based biosensors are traditionally monitored using optical beam deflection,59 which implies a complicated configuration. In contrast, OPUs provide submicron laser spots and subatomic sensing resolution9 to monitor MEMS28,29 and even nanoelectromechanical systems.60 These cantilevers can be functionalized with receptor molecules as label-free biomolecular transducers that are immersed in an analyte and monitored by the OPU61 using static bending and frequency changes,62 as shown in Figure 9.
In addition, a Blu-ray OPU-based vibrometer combining MEMS resonators inside microfluidic chips has been used to characterize biopolymer degradation under the action of enzymes in a controlled flow condition. An algorithm enables the OPU to measure 12 resonators within 4 min, thus dramatically reducing the degradation measurement time from 6 weeks to 8 h.63
Furthermore, the OPU has a high sensing tolerance of ±5° to the cantilever initial angular tilt, allowing to scan cantilever-based biosensors on a rotating disc. This scanning system has a theoretical throughput of 500,000 cantilevers per second.64 Moreover, the OPU scanning data has been used to reconstruct 3D topography and surface roughness of each cantilever to provide extra physical information for detection of pesticide derivative 2,6-dichlorobenzamide,65 as shown in Figure 10. OPU cantilever-based biosensing has also been used in applications such as detection of vapor and liquid phase of 2,4-dinitrotoluene,66 platelet derived growth factor proteins,67 and soluble urokinase plasminogen activator receptor inflammatory biomarker.68
Interestingly, an OPU can be used for nanoscale biomolecule imaging beyond the diffraction limit by monitoring a MEMS AFM probe (Figure 11A),9,10,28,32 which has a tip with a typical radius of 10 nm. This OPU-based AFM is capable of imaging DNA in air or solution environments,33 as shown in Figure 11B.
Modified OPU for Biosensing
A conventional DNA microarray scanner requires microscale precision for xy positioning, laser excitation, precise optics for focusing, and optical sensing. Consequently, the resulting system is expensive and bulky. By replacing the OPU components with a single-mode optical fiber and attaching an external sensing setup, the OPU can be used as the head of a scanning confocal microscope to enable fluorescent-based biosensing.69
Figure 12 shows an OPU-based DNA microarray scanner. The OPU is coupled to a photomultiplier tube (PMT) detector through a dichroic filter, such as that shown in Figure 2. The OPU–PMT setup can acquire signals from fluorescent dyes excited by the OPU laser of 650 nm.24 The OPU VCM provides dynamic autofocusing that enables higher detection performance than conventional microarray scanners, as shown in Figure 13. The OPU–PMT-based DNA microarray scanner has been successfully used to measure a commercial bacterial artificial chromosome oligonucleotide DNA chip and a 32 12 Cy5 fluorescent dye calibration slide (DS01).70 Combined with microfluidic chips, OPU–PMT fluorescence detection can measure 2.5 μm fluorescent beads inside a microchannel,71,72 as shown in Figure 14.
To reduce the cost of using triple-wavelength laser diodes, some OPUs are endowed with dual-optics 780 and 650 nm lasers for CD/DVD, and a 405 nm laser for Blu-ray disc reading. These dual-optics OPUs have two objective lenses driven by the same VCM and can perform two separate optical sensing tasks while the objective lenses move synchronously. Figure 15 shows a diagram of a Blu-ray scanning microscope (BSM) comprising a dual-optics OPU coupled with a PMT sensor. The BSM uses the Blu-ray optical path to excite fluorescent light and capture a cell fluorescence image. Simultaneously, the DVD optical path is focused on an address-patterned area to monitor the xy scanning area and z-axis focusing.73 This BSM setup has been successfully used to measure monkey-derived kidney epithelial cells and fibroblast cells stained with fluorophore phalloidin CF405 (Biotum, Inc., Fremont, CA, USA),74 as shown in Figure 16. Moreover, the compact size of the BSM allows embedding it into a cell culture chamber.
The Blu-ray objective lens can also be replaced by a 488 nm optimized lens for imaging HA22T/VGH and VERO cell stained with phalloidin CF405 and Alexa Fluor 488 (Thermo Fisher Scientific, Waltham, MA, USA). The imaging depth can be adjusted in a range of ±20 μm through collimator lens B4 (Figure 15) inside the OPU.75 Besides the PMT, the OPU can also equip an avalanche photodiode for Cy5 detection of stained cell fluorescence on a spinning disc.76
Nanoparticle labeling can enhance the biomolecule signals up to 100 times.77 Using this labeling technique, OPUs can scan the DNA microarray with an external photodiode, which is much less expensive than a PMT. Figure 17 shows an OPU–photodiode-based DNA microarray scanner to measure human papillomavirus with Ag-deposited Au nanoparticle labels.78
Besides labeling, functionalized magnetic nanoparticle (MNP)-based antigen and antibody assay reactions can be opto-magnetically79 monitored using a Blu-ray OPU (objective lens removed) instead of the expensive and bulky superconducting quantum interference devices.80 Specifically, the Blu-ray OPU–MNP sensing shines a 405 nm parallel laser through a solution contains the MNPs, then acquiring reflection via a mirror. In addition, the Blu-ray OPU PDIC monitors the reflected laser frequency, phase, and amplitude while two coils apply an oscillating magnetic field to the MNPs. Combining Blu-ray OPU–MNP sensing with the lab-on-disc technique can carry out microliter-scale whole blood separation and antibody sensing81 in few minutes (Figures 18 and 19).
Blu-ray OPU–MNP biosensing has also been applied on Escherichia coli, Proteus mirabilis, Pseudomonas aeruginosa,82,83 NS1 antigen of dengue,84 and adenosine triphosphate.85 This system features a compact size and can be integrated with commercial bioimaging systems to study the aptamer-conjugated MNPs86 and the action mechanism of type 2 diabetes drugs.87
Conclusions and Future Perspectives
CD/DVD/Blu-ray OPUs provide outstanding characteristics such as light weight, low cost, compact size, and high performance, which could be exploited to democratize and spread micro- and nanoscale biotechnology research for institutes in developing countries and citizen scientists. This review shows the possibility to open the black box of commercial OPUs and describes its mechanism, component characteristics, and several OPU-based bioapplications. We place special attention on gaining access to the OPU functions for various biosensing applications, and describe this access using circuits for three commercially available triple-wavelength OPUs in the Supporting Information.
High-speed Internet and solid-state drives have overcome optical drives for digital data transference and storage, with optical disc drives being discontinued from devices such as notebooks. Still, operative optical storage drives can be found in some music players and gaming consoles such as triple-wavelength Blu-ray OPUs are equipped in the Sony’s Play Station 4 (Sony Co., Tokyo, Japan) and XBOX One X (Microsoft Co., Redmond, WA, USA), which are devices that can operate for over 10 years. Moreover, repair assistance for OPUs can guarantee sustained scientific research for years and maybe decades to come.
Optical storage devices have a low cost, high durability, zero-energy storage, and compactness, which helps end users keep their information safe and private from enterprises (e.g., social networks), which can misuse sensitive information, and avoid big data cloud storage. Moreover, the femtosecond laser-based 360 TB with 13.8-billion-year storage time data recording technique proposed in ref (88) suggests an exciting future for optical storage technology. Of course, successful engineering, commercialization, and mass production processes will determine the future of this technology and reduce the costs for potential applications in further scientific research.
Acknowledgments
The authors would like to thank Mr. Chung-Hsiang Cheng and Mr. Christian Werner for the OPU driving circuits design. The authors also acknowledge the support from the European Research Council under the European Union’s Seventh Framework Program (FP7/2007-2013), Grant no. 320535-HERMES and Center for Intelligent Drug Delivery and Sensing Using Microcontainers and Nanomechanics (IDUN) (grant no. DNRF122) funded by the Danish National Research Foundation and the Velux Foundations.
Glossary
Vocabulary
- numerical aperture
a range of angles over which a lens can emit or collect light
- refractive index
a dimensionless number of a material which describes speed of light propagates through the material
- voice coil motor
an actuator driven by an electric current energized copper coil that operates inside a magnetic field
- optical path difference
rays emitted from a point traveling through media with different ray path length
- dichroic filter
color filter that selectively reflect certain light wavelengths
- photodiode
a semiconductor-based component to convert light into an electrical signal
- error correction code
sequence of numbers that corrects errors during data transmission
Supporting Information Available
The Supporting Information is available free of charge on the ACS Publications website at DOI: 10.1021/acssensors.8b00340.
Triple-wavelength OPU controller circuit designs (PDF)
Author Contributions
The manuscript was written through contributions of all authors. All authors have given approval to the final version of the manuscript.
The authors declare no competing financial interest.
Supplementary Material
References
- Mark D.; Haeberle S.; Roth G.; von Stetten F.; Zengerle R. Microfluidic Lab-on-a-Chip Platforms: Requirements, Characteristics and Applications. Chem. Soc. Rev. 2010, 39, 1153–1182. 10.1039/b820557b. [DOI] [PubMed] [Google Scholar]
- Strohmeier O.; Keller M.; Schwemmer F.; Zehnle S.; Mark D.; von Stetten F.; Zengerle R.; Paust N. Centrifugal Microfluidic Platforms: Advanced Unit Operations and Applications. Chem. Soc. Rev. 2015, 44, 6187–6229. 10.1039/C4CS00371C. [DOI] [PubMed] [Google Scholar]
- Kido H.; Maquieira A.; Hammock B. D. Disc-Based Immunoassay Microarrays. Anal. Chim. Acta 2000, 411, 1–11. 10.1016/S0003-2670(00)00776-5. [DOI] [Google Scholar]
- Morais S.; Carrascosa J.; Mira D.; Puchades R.; Maquieira A. Microimmunoanalysis on Standard Compact Discs to Determine Low Abundant Compounds. Anal. Chem. 2007, 79, 7628–7635. 10.1021/ac070328b. [DOI] [PubMed] [Google Scholar]
- Bañuls M.-J.; Gonzalez-Pedro V.; Puchades R.; Maquieira A. PMMA Isocyanate-Modified Digital Discs as a Support of Oligonucleotide-Based Assays. Bioconjugate Chem. 2007, 18, 1408–1414. 10.1021/bc700194n. [DOI] [PubMed] [Google Scholar]
- Morais S.; Tamarit-López J.; Carrascosa J.; Puchades R.; Maquieira A. Analytical Prospect of Compact Disc Technology in Immunosensing. Anal. Bioanal. Chem. 2008, 391, 2837–2844. 10.1007/s00216-008-2224-4. [DOI] [PubMed] [Google Scholar]
- Ramachandraiah H.; Amasia M.; Cole J.; Sheard P.; Pickhaver S.; Walker C.; Wirta V.; Lexow P.; Lione R.; Russom A. Lab-on-DVD: Standard DVD Drives as a Novel Laser Scanning Microscope for Image Based Point of Care Diagnostics. Lab Chip 2013, 13, 1578–1585. 10.1039/c3lc41360h. [DOI] [PubMed] [Google Scholar]
- Weng S.; Li X.; Li Y.; Yu H.-Z. Optical Disc Technology-Enable Analytical Devices: From Hardware Modification to Digitized Molecular Detection. Analyst 2016, 141, 6190–6201. 10.1039/C6AN01781A. [DOI] [PubMed] [Google Scholar]
- Hwu E.-T.; Huang K.-Y.; Hung S.-K.; Hwang I.-S. Measurement of Cantilever Displacement Using a Compact Disk/Digital Verstaile Disk Pickup Head. Jpn. J. Appl. Phys. 2006, 45, 2368–2371. 10.1143/JJAP.45.2368. [DOI] [Google Scholar]
- Wang W.-M.; Huang K.-Y.; Huang H.-F.; Hwang I.-S.; Hwu E.-T. Low-Voltage and High-Performance Buzzer-Scanner Based Streamlined Atomic Force Microscope System. Nanotechnology 2013, 24, 455503. 10.1088/0957-4484/24/45/455503. [DOI] [PubMed] [Google Scholar]
- Rothenbach C. A.; Gupta M. C. High Resolution, Low Cost Laser Lithography Using a Blu-ray Optical Head Assembly. Opt. Lasers Eng. 2012, 50, 900–904. 10.1016/j.optlaseng.2011.12.004. [DOI] [Google Scholar]
- Grey F. Creativity Unleashed. Nat. Nanotechnol. 2015, 10, 480. 10.1038/nnano.2015.95. [DOI] [PubMed] [Google Scholar]
- Lopez-Martinez F.; Lombraña D. L.; Grey F.; Hwu E.-T. A Crowdsourcing-Based Air Pollution Measurement System Using DIY Atomic Force Microscopes. Hum. Comput. 2016, 3, 235–241. 10.15346/hc.v3i1.14. [DOI] [Google Scholar]
- Blu-ray Disc Association, White Paper Blu-Ray DiscTM Format-Physical Format Specifications for BD-ROM, 2010; pp 1–48.
- Wu Q.; Feke G. D.; Grober R. D. Realization of Numerical Aperture 2.0 Using a Gallium Phosphide Solid Immersion Lens. Appl. Phys. Lett. 1999, 75, 4064–4066. 10.1063/1.125537. [DOI] [Google Scholar]
- Hayashida N.; Hirata H.; Komaki T.; Usami M.; Ushida T.; Itoh H.; Yoneyama K.; Utsunomiya H. High-Performance Hard Coat for Cartrige-Free Blu-ray Disc. Jpn. J. Appl. Phys. 2003, 42, 750–753. 10.1143/JJAP.42.750. [DOI] [Google Scholar]
- Hwu E.-T.Fundamentals of an Atomic Force Microscope Based on a Digital Versatile Disk Optical Pick-up Unit, 2014; pp 31–36; https://books.google.dk/books?id=2TlzAwAAQBAJ. [Google Scholar]
- Blu-ray Disc Association, White Paper Blu-Ray DiscTM Format-Key Technologies, 2004; pp 1–48.
- Hsu H.-S.; Tsai P. J. A High Frequency and Amplitude Modulated CMOS Laser Diode Driver for Low Noise Readout Applications. Analog Integr. Circ. Sig. Process. 2015, 85, 461–472. 10.1007/s10470-015-0622-z. [DOI] [Google Scholar]
- Uchiyama M.; Ebihara T.; Omi K.; Kitano H.; Hoshino I.; Mori K. Development of Optical Pickup for Digital Versatile Disc Using Two-Wavelength-Integrated Laser Diode. Jpn. J. Appl. Phys. 2000, 39, 1549–1553. 10.1143/JJAP.39.1549. [DOI] [Google Scholar]
- Wu C.-H.; Chen W.-S. Injection Molding and Injection Compression Molding of Three-beam Grating of DVD Pickup Lens. Sens. Actuators, A 2006, 125, 367–375. 10.1016/j.sna.2005.07.025. [DOI] [Google Scholar]
- Tanaka Y.; Komma Y.; Shimizu Y.; Shimazaki T.; Murata J.; Mizuno S. Lens Design of Compatible Objective Lens for Blu-ray Disc and Digital Versatile Disk with Diffractive Optical Element and Phase Steps. Jpn. J. Appl. Phys. 2004, 43, 4742–4745. 10.1143/JJAP.43.4742. [DOI] [Google Scholar]
- Wang W.-M.; Cheng C.-H.; Molnar G.; Hwang I.-S.; Huang K.-Y.; Danzebrink H.-U.; Hwu E.-T. Optical Imaging Module for Astigmatic Detection System. Rev. Sci. Instrum. 2016, 87, 053706. 10.1063/1.4952438. [DOI] [PubMed] [Google Scholar]
- Kim K.-H.; Lee S.-Y.; Kim S.; Lee S.-H.; Jeong S.-G. A new DNA Chip Detection Mechanism Using Optical Pick-up Actuators. Microsyst. Technol. 2007, 13, 1359–1369. 10.1007/s00542-006-0367-9. [DOI] [Google Scholar]
- Perraut F.; Lagrange A.; Pouteau P.; Peyssonneaux O.; Puget P.; McGall G.; Menou L.; Gonzalez R.; Labeye P.; Ginot F. A New Generation of Scanners for DNA Chips. Biosens. Bioelectron. 2002, 17, 803–813. 10.1016/S0956-5663(02)00073-8. [DOI] [PubMed] [Google Scholar]
- Matsui T.; Shimomura T.. DVD Optical-head with Fluorescent Detection. In Proceedings of Optical Data Storage Topical Meeting of IEEE Lasers & Electro-Optics Society, April 23–26, 2006; pp 103–105.
- Shimomura T.; Izawa C.; Matsui T. Development of a Compact Optical System for Microarray Scanning Using a DVD Pickup Head. Rev. Sci. Instrum. 2008, 79, 035101. 10.1063/1.2885609. [DOI] [PubMed] [Google Scholar]
- Hwu E.-T.; Hung S.-K.; Yang C.-W.; Huang K.-Y.; Hwang I.-S. Simultaneous Detection of Translational and Angular Displacements of Micromachined Elements. Appl. Phys. Lett. 2007, 91, 221908. 10.1063/1.2817750. [DOI] [Google Scholar]
- Hwu E.-T.; Hung S.-K.; Yang C.-Y.; Huang K.-Y.; Hwang I.-S. Real-Time Detection of Linear and Angular Displacements with a Modified DVD Optical Head. Nanotechnology 2008, 19, 115501. 10.1088/0957-4484/19/11/115501. [DOI] [PubMed] [Google Scholar]
- Zhao X.; Li X.; Cui C.; Yu H.-Z. DVD Diagnostic Software for reading Disc-Based Bioassays, a Comparative Study. Sens. Actuators, B 2014, 195, 116–122. 10.1016/j.snb.2013.12.087. [DOI] [Google Scholar]
- La Clair J. J.; Burkart M. D. Molecular Screening on a Compact Disc. Org. Biomol. Chem. 2003, 1, 3244–3249. 10.1039/b306391g. [DOI] [PubMed] [Google Scholar]
- Weng S.; Li X.; Niu M.; Ge B.; Yu H.-Z. Blu-Ray Technology-Based Quantitative Assays for Cardiac Markers: From Disc Activation to Multiplex Detection. Anal. Chem. 2016, 88, 6889–6896. 10.1021/acs.analchem.6b01604. [DOI] [PubMed] [Google Scholar]
- Liao H.-S.; Huang K.-Y.; Hwang I.-S.; Chang T.-J.; Hsiao W. W.; Lin H.-H.; Hwu E.-T.; Chang C.-S. Operation of Astigmatic-Detection Atomic Force Microscopy in Liquid Environments. Rev. Sci. Instrum. 2013, 84, 103709. 10.1063/1.4826494. [DOI] [PubMed] [Google Scholar]
- Wäldchen S.; Lehmann J.; Klein T.; van de Linde S.; Sauer M. Light-Induced Cell Damage in Live-Cell Super-Resolution Microscopy. Sci. Rep. 2015, 5, 15348. 10.1038/srep15348. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kong X.; Mohanty S. K.; Stephens J.; Heale J. T.; Gomez-Godinez V.; Shi L. Z.; Kim J.-S.; Yokomori K.; Berns M. W. Comparative Analysis of Different Laser Systems to Study Cellular Responses to DNA Damage in Mammalian Cells. Nucleic Acids Res. 2009, 37, e68. 10.1093/nar/gkp221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barathur R.; Bookout J.; Sreevatsan S.; Gordon J.; Werner M.; Thor G.; Worthington M. New Disc-Based Technologies for Diagnostic and Research Applications. Psychiatr. Genet. 2002, 12, 193–206. 10.1097/00041444-200212000-00002. [DOI] [PubMed] [Google Scholar]
- Yu H.-Z.; Li Y.; Ou L. M.-L. Reading Disc Based Bioassays with Standard Computer Drives. Acc. Chem. Res. 2013, 46, 258–268. 10.1021/ar300104b. [DOI] [PubMed] [Google Scholar]
- Wang H.; Ou L. M. L.; Suo Y.; Yu H.-Z. Computer-Readable DNAzyme Assay on Disc for ppb-Level Lead Detection. Anal. Chem. 2011, 83, 1557–1563. 10.1021/ac103177w. [DOI] [PubMed] [Google Scholar]
- Li X.; Weng S.; Ge B.; Yao Z.; Yu H.-Z. DVD Technology-Based Molecular Diagnosis Platform: Quantitative Pregnancy Test on a Disc. Lab Chip 2014, 14, 1686–1694. 10.1039/c3lc51411k. [DOI] [PubMed] [Google Scholar]
- Zhang L.; Li X.; Li Y.; Shi X.; Yu H.-Z. Indirect Competitive Assays on DVD for Direct Multiplex Detection of Drugs of Abuse in Oral Fluids. Anal. Chem. 2015, 87, 1896–1902. 10.1021/ac5040715. [DOI] [PubMed] [Google Scholar]
- Zhang L.; Wong J. X. H.; Li X.; Li Y.; Yu H.-Z. Detection and Quantitation of Heavy Metal Ions on Bona Fide DVDs Using DNA Molecular Beacon Probes. Anal. Chem. 2015, 87, 5062–5067. 10.1021/acs.analchem.5b00899. [DOI] [PubMed] [Google Scholar]
- Li X.; Shi M.; Cui C.; Yu H.-Z. Inkjet-Printed Bioassays for Direct Reading with a Multimode DVD/Blu-Ray Optical Drive. Anal. Chem. 2014, 86, 8922–8926. 10.1021/ac501870w. [DOI] [PubMed] [Google Scholar]
- Shi M.; Weng S. S. H.; Li X.; Yu H.-Z. Digitized Molecular Detection on Off-the-shelf Blu-Ray Discs: Upgraded Resolution and Enhanced Sensitivity. Sens. Actuators, B 2017, 242, 79–86. 10.1016/j.snb.2016.11.042. [DOI] [Google Scholar]
- Arai T.; Gopinath S. C. B.; Mizuno H.; Kumar P. K. R.; Rockstuhl C.; Awazu K.; Tominaga J. Toward Biological Diagnosis System Based on Digital Versatile Disc Technology. Jpn. J. Appl. Phys. 2007, 46, 4003–4006. 10.1143/JJAP.46.4003. [DOI] [Google Scholar]
- Potyrailo R. A.; Morris W. G.; Leach A. M.; Sivavec T. M.; Wisnudel M. B.; Boyette S. Analog Signal Acquisition fom Computer Optical Disk Drives for Quantitative Chemical Sensing. Anal. Chem. 2006, 78, 5893–5899. 10.1021/ac060684e. [DOI] [PubMed] [Google Scholar]
- Potyrailo R. A.; Morris W. G.; Leach A. M.; Hassib L.; Krishnan K.; Surman C.; Wroczynski R.; Boyette S.; Xiao C.; Shrikhande P.; Agree A.; Cecconie T. Theory and Practice of Ubiquitous Quantitative Chemical Analysis Using Conventional Computer Optical Disk Drives. Appl. Opt. 2007, 46, 7007–7017. 10.1364/AO.46.007007. [DOI] [PubMed] [Google Scholar]
- Morais S.; Tortajada-Genaro A. A.; Arnandis-Chover T.; Puchades R.; Maquieira A. Multiplexed Microimmunoassays on a Digital Versatile Disk. Anal. Chem. 2009, 81, 5646–5654. 10.1021/ac900359d. [DOI] [PubMed] [Google Scholar]
- Gopinath S. C. B.; Kumar P. K. R.; Tominaga J. A BioDVD Media with Multilayered Structure is Suitable for Analyzing Biomolecular Interactions. J. Nanosci. Nanotechnol. 2011, 11, 5682. 10.1166/jnn.2011.4500. [DOI] [PubMed] [Google Scholar]
- Gopinath S. C. B.; Suenaga E.; Nagumo O.; Kumar P. K. R.; Tominaga J.. BioDVD Platform for Bio-Recognition; International Conference on Biological, Biomedical and Pharmaceutical Sciences of the International Association of Chemical, Biological & Medical Sciences Researchers, July 28–29, 2012.
- Lange S. A.; Roth G.; Wittemann S.; Lacoste T.; Vetter A.; Grässle J.; Kopta S.; Kolleck M.; Breitinger B.; Wick M.; Horber J. K. H.; Dübel S.; Bernard A. Measuring Biomolecular Binding Events with a Compact Disc Player Device. Angew. Chem., Int. Ed. 2006, 45, 270–273. 10.1002/anie.200501243. [DOI] [PubMed] [Google Scholar]
- Imaad S. M.; Kulsharova G.; Liu G. L. Microparticle and cell counting with digital microfluidic compact disc using standard CD drive. Lab Chip 2011, 11, 1448–1456. 10.1039/c0lc00451k. [DOI] [PubMed] [Google Scholar]
- Kostner S.; Vellekoop M. J.. Detection of Single Biological Cells Using a DVD Pickup Head; The 14th International Conference on Solid-State Sensors, Actuators and Microsystems of IEEE Transducers, June 10–14, 2007.
- Kostner S.; Vellekoop M. J.. Low Cost Cytometer Based on a DVD Pickup Head; 11th International Conference on Miniaturized Systems for Chemistry and Life Sciences of The Chemical and Biological Microsystems Society, October 7–11, 2007; pp 739–741.
- Kostner S.; Vellekoop M. J. Cell Analysis in a microfluidic cytometer applying a DVD pickup head. Sens. Actuators, B 2008, 132, 512–517. 10.1016/j.snb.2007.11.038. [DOI] [Google Scholar]
- Kostner S.; Vellekoop M. J. Microsystems for Optical Cell Detection: Near versus Far Field. Part. Part. Syst. Charact. 2008, 25, 92–98. 10.1002/ppsc.200700013. [DOI] [Google Scholar]
- Kasukurti A.; Potcoava M.; Desai S. A.; Eggleton C.; Marr D. W. M. Single-cell isolation using a DVD optical pickup. Opt. Express 2011, 19, 10377–10386. 10.1364/OE.19.010377. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kasukurti A.; Eggleton C. D.; Desai S. A.; Disharoon D. I.; Marr D. W. M. A simple Microfluidic Dispenser for Single-Microparticle and Cell Samples. Lab Chip 2014, 14, 4673–4679. 10.1039/C4LC00863D. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hsiao W.-W.; Liao H.-S.; Lin H.-H.; Lee Y.-L.; Fan C.-K.; Liao C.-W.; Lin P.-Y.; Hwu E.-T.; Chang C.-S. Biophysical Analysis of Astrocytes Apoptosis Triggered by Larval E/S Antigen from Cerebral Toxocarosis-Causing Pathogen Toxocara Canis. Anal. Sci. 2013, 29, 885–892. 10.2116/analsci.29.885. [DOI] [PubMed] [Google Scholar]
- Fritz J.; Baller M. K.; Lang H. P.; Rothuizen H.; Vettiger P.; Meyer E.; Güntherodt H.-J.; Gerber Ch.; Gimzewski J. K. Translating Biomolecular Recognition into Nanomechanics. Science 2000, 288, 316–318. 10.1126/science.288.5464.316. [DOI] [PubMed] [Google Scholar]
- Chen C.-H.; Liao H.-S.; Hwang I.-S.; Hwu E.-T.; Huang K.-Y.; Larsen P. E.; Schmid S.; Boisen A. Compact MEMS/NEMS Characterization Platform Using a DVD Optical Pick-Up Unit with Optical Imaging Function. TechConnect Briefs 2013, 2, 485–488. [Google Scholar]
- Bosco F. G.; Hwu E.-T.; Keller S.; Greve A.; Boisen A. Self-aligned Cantilever Positioning for n-substrate Measurements Using DVD Pickup Head. Microelectron. Eng. 2010, 87, 708–711. 10.1016/j.mee.2009.12.064. [DOI] [Google Scholar]
- Hwu E.-T.; Liao H.-S.; Bosco F. G.; Chen C.-H.; Keller S. S.; Boisen A.; Huang K.-Y. An Astigmatic Detection System for Polymeric Cantilever-Based Sensors. J. Sens. 2012, 2012, 580939. 10.1155/2012/580939. [DOI] [Google Scholar]
- Ceccacci A. C.; Chen C.-H.; Hwu E.-T.; Morelli L.; Bose S.; Bosco F. G.; Schmid S.; Boisen A. Blu-Ray-Based Micromechanical Characterization Platform for Biopolymer degradation Assessment. Sens. Actuators, B 2017, 241, 1303–1309. 10.1016/j.snb.2016.09.190. [DOI] [Google Scholar]
- Hwu E.-T.; Chen C.-H.; Bosco F. G.; Wang W.-M.; Ko H.-C.; Hwang I.-S.; Boisen A. High-performance Spinning Device for DVD-based Micromechanical Signal Transduction. J. Micromech. Microeng. 2013, 23, 045016. 10.1088/0960-1317/23/4/045016. [DOI] [Google Scholar]
- Bosco F. G.; Hwu E.-T.; Chen C.-H.; Keller S.; Bache M.; Jakobsen M. H.; Hwang I.-S.; Boisen A. High Throughput Label-free Platform for Statistical Bio-molecular Sensing. Lab Chip 2011, 11, 2411–2146. 10.1039/c1lc20116f. [DOI] [PubMed] [Google Scholar]
- Bosco F. G.; Bache M.; Hwu E.-T.; Chen C.-H.; Andersen S. S.; Nielsen K. A.; Keller S. S.; Jeppesen J. O.; Hwang I.-S.; Boisen A. Statistical Analysis of DNT Detection Using Chemically Functionalized Microcantilever Arrays. Sens. Actuators, B 2012, 171–172, 1054–1059. 10.1016/j.snb.2012.06.033. [DOI] [Google Scholar]
- Bosco F. G.; Bache M.; Yang J.; Chen C.-H.; Hwu E.-T.; Lin Q.; Boisen A. Micromechanical PDGF Recognition via Lab-on-a-disc Aptasensor Arrays. Sens. Actuators, A 2013, 195, 154–159. 10.1016/j.sna.2012.06.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bache M.; Bosco F. G.; Brogger A. L.; Frohling K. B.; Alstrom T. S.; Hwu E.-T.; Chen C.-H.; Eugen-Olsen J.; Hwang I.-S.; Boisen A. Nanomechanical Recognition of Prognostic Biomarker suPAR with DVD-ROM Optical Technology. Nanotechnology 2013, 24, 444011. 10.1088/0957-4484/24/44/444011. [DOI] [PubMed] [Google Scholar]
- Benschop J.; van Rosmalen G. Confocal Compact Scanning Optical Microscope Based on Compact Disc Technology. Appl. Opt. 1991, 30, 1179–1184. 10.1364/AO.30.001179. [DOI] [PubMed] [Google Scholar]
- Kim K.-H.; Lee S.-Y.; Kim S.; Jeong S.-G. DNA Microarray Scanner with a DVD Pick-up Head. Curr. Appl. Phys. 2008, 8, 687–691. 10.1016/j.cap.2007.04.047. [DOI] [Google Scholar]
- Segerink L. I.; Koster M. J.; Sprenkels A. J.; Vermes I.; Berg A.. A Cheap 2D Fluorescence Detection System for μm-sized Beads on-chip; 15th International Conference on Miniaturized Systems for Chemistry and Life Sciences of The Chemical and Biological Microsystems Society, October 2–6, 2011.
- Segerink L. I.; Koster M. J.; Sprenkels A. J.; van den Berg A. A Low-cost 2D Fluorescence Detection System for μ-sized Beads on-chip. Lab Chip 2012, 12, 1780–1783. 10.1039/c2lc21187d. [DOI] [PubMed] [Google Scholar]
- Yu H.-C.; Lyu H.-C.; Lee Y.-C.; Ju J.-J.; Kao F.-J.. A Novel Optical Scanner Based on Blue-ray DVD Pick-up Head for Analysis of Multifunctional Bio-information; Focus on Microscopy conference, April 17–20, 2011.
- Tsai R.-Y.; Chen J.-P.; Lee Y.-C.; Huang C.-C.; Huang T.-T.; Chiang H.-C.; Cheng C.-M.; Lo F.-H.; Chang S.-L.; Weng K.-Y.; Chung L.-P.; Chen J.-C.; Tiao G. Position-addressable Digital Laser Scanning Point Fluorescence Microscopy with a Blu-ray Disk Pickup Head. Biomed. Opt. Express 2014, 5, 427–438. 10.1364/BOE.5.000427. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tsai R.-Y.; Chen J.-P.; Lee Y.-C.; Chiang H.-C.; Cheng C.-M.; Huang C.-C.; Huang T.-T.; Cheng C.-T.; Tiao G. Cell Depth Imaging by Point Laser Scanning Fluorescence Microscopy with an Optical Disk Pickup Head. Jpn. J. Appl. Phys. 2015, 54, 09MD01. 10.7567/JJAP.54.09MD01. [DOI] [Google Scholar]
- Huang D-R.; Ju J.-J.; Lee Y.-C.; Chen J.-S.; Lo F.-H.; Chang S.-Li. New Disc Format for Biosensing by Using Optical Pick-Up Head System. IEEE Trans. Magn. 2014, 50, 3500704. 10.1109/TMAG.2014.2300340. [DOI] [Google Scholar]
- Taton T. A.; Mirkin C. A.; Letsinger R. L. Scanometric DNA Array Detection with Nanoparticle Probes. Science 2000, 289, 1757–1760. 10.1126/science.289.5485.1757. [DOI] [PubMed] [Google Scholar]
- Li X. Z.; Kim S.; Cho W.; Lee S.-Y. Optical Detection of Nanoparticle-enhanced Human Papillomavirus Genotyping Micoarrays. Biomed. Opt. Express 2013, 4, 187–192. 10.1364/BOE.4.000187. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Donolato M.; Antunes P.; Bejhed R. S.; Zardan Gomez de la Torre T.; Osterberg F. W.; Stromberg M.; Nilsson M.; Stromme M.; Svedlindh P.; Hansen M. F.; Vavassori P. Novel Readout Method for Molecular Diagnostic Assays Based on Optical Measurement of Magnetic Nanobead Dynamics. Anal. Chem. 2015, 87, 1622–1629. 10.1021/ac503191v. [DOI] [PubMed] [Google Scholar]
- Koh I.; Josephson L. Magnetic Nanoparticle Sensors. Sensors 2009, 9, 8130–8145. 10.3390/s91008130. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Donolato M.; Antunes P.; Burger R.; Bosco F.; Olsson M.; Yang J.; Chen C.-H.; Lin Q.; Hwu E.-T.; Boisen A.; Hansen M. F.. Lab-On-Blu-ray: Low-Cost Analyte Detection on a Disk; 18th International Conference on Miniaturized Systems for Chemistry and Life Sciences of The Chemical and Biological Microsystems Society, October 26–30, 2014; pp 2044–2046.
- Mezger A.; Fock J.; Antunes P.; Osterberg F. W.; Boisen A.; Nilsson M.; Hansen M. F.; Ahlford A.; Donolato M. Scalable DNA-Based Magnetic Nanoparticle Agglutination Assay for Bacterial Detection in Patient Samples. ACS Nano 2015, 9, 7374–7382. 10.1021/acsnano.5b02379. [DOI] [PubMed] [Google Scholar]
- Donolato M.; Antunes P.; Zardan Gomez de la Torre T.; Hwu E.-T.; Chen C.-H.; Burger R.; Rizzi G.; Bosco F. G.; Strømme M.; Boisen A.; Hansen M. F. Quantification of Rolling Cycle Amplified DNA Using Magnetic Nanobeads and a Blu-ray Optical Pick-p Unit. Biosens. Bioelectron. 2015, 67, 649–655. 10.1016/j.bios.2014.09.097. [DOI] [PubMed] [Google Scholar]
- Antunes P.; Watterson D.; Parmvi M.; Burger R.; Boisen A.; Young P.; Cooper M. A.; Hansen M. F.; Ranzoni A.; Donolato M. Quantification of NS1 Dengue Biomarker in Serum via Optomagnetic Nanocluster Detection. Sci. Rep. 2015, 5, 16145. 10.1038/srep16145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yang J.; Donolato M.; Pinto A.; Bosco F. G.; Hwu T.-T.; Chen C.-H.; Alstrom T. S.; Lee G.-H.; Schafer T.; Vavassori P.; Boisen A.; Lin Q.; Hansen M. F. Blu-Ray Based Optomagnetic Aptasensor for Detection of Small Molecules. Biosens. Bioelectron. 2016, 75, 396–403. 10.1016/j.bios.2015.08.062. [DOI] [PubMed] [Google Scholar]
- Uddin R.; Burger R.; Donolato M.; Fock J.; Creagh M.; Hansen F. M.; Boisen A. Lab-on-a-disc Agglutination Assay for Protein Detection by Optomagnetic Readout and Optical Imaging Using Nano-and Micro-sized Magnetic Beads. Biosens. Bioelectron. 2016, 85, 351–357. 10.1016/j.bios.2016.05.023. [DOI] [PubMed] [Google Scholar]
- Uddin R.; Nur-E-Habiba; Rena G.; Hwu E.-T.; Boisen A. New Evidence for the Mechanism of Action of a Type-2 Diabetes Drug Using a Magnetic Bead-Based Automated Biosensing Platform. ACS Sens. 2017, 2, 1329–1336. 10.1021/acssensors.7b00384. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhang J.; Gecevicius M.; Beresna M.; Kazansky P. Seemingly Unlimited Lifetime Data Storage in Nanostructured Glass. Phys. Rev. Lett. 2014, 112, 033901. 10.1103/PhysRevLett.112.033901. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.