Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2018 Aug 1.
Published in final edited form as: Nucl Med Commun. 2015 Aug;36(8):782–789. doi: 10.1097/MNM.0000000000000317

Quantification of metabolic tumor activity and burden in patients with NSCLC: Is manual adjustment of semi-automatic gradient based measurements necessary?

Piotr Obara 1,#, Haiping Liu 2,#, Kristen Wroblewski 3, Chen-Peng Zhang 4, Peng Hou 2, Yulei Jiang 1, Chen Ping 2,#, Yonglin Pu 1,#
PMCID: PMC6069606  NIHMSID: NIHMS673835  PMID: 25888358

Abstract

Purpose:

Metabolic tumor burden (MTB) measurements including metabolic tumor volume (MTV) and total lesion glycolysis (TLG) have been shown to have prognostic value in non-small cell lung cancer (NSCLC). Calculating MTB typically utilizes software to semi-automatically draw volumes of interest (VOIs) around the tumor, which are subsequently manually adjusted by the radiologist to include the entire tumor. The manual adjustment step can be time-consuming and observer-dependent. We compared the agreement of MTB values obtained from the semi-automatic method with and without manual adjustment in NSCLC patients.

Methods:

This IRB approved prospective study included 134 patients with histologically proven NSCLC who underwent 18F-FDG-PET/CT. The MTB of the primary tumor was measured with a semi-automatic gradient-based method without manual adjustment (semi-automatic gradient method) and with manual adjustment (manually adjusted semi-automatic gradient method) by two radiologists by using the MIM PETedge tool. The paired t-test, Wilcoxon signed-rank test, and concordance correlation coefficient (CCC) were calculated to evaluate agreement between MTB measures obtained with these two methods, as well as agreement between the two radiologists for each method.

Results:

SUVmax values were identical between the two methods. No statistically significant difference was present for SUVpeak, MTV and TLG values between the two methods (p=0.23, 0.45 and 0.37, respectively). Excellent agreement between the two methods was found in terms of CCC (CCC > 0.98 for all measures). Inter-observer reliability was excellent for all measures (CCC > 0.90).

Conclusion:

The semi-automatic gradient-based tumor-segmentation method can be used without the additional manual adjustment step for metabolic tumor burden quantification of primary NSCLC tumors.

Keywords: non-small cell lung cancer, metabolic tumor burden, PET/CT, metabolic tumor volume, total lesion glycolysis

Introduction

18F-FDG positron emission tomography (18F-FDG-PET) exams are commonly used in patients with non-small cell lung cancer (NSCLC) for initial evaluation, staging, radiation therapy planning, and treatment response monitoring [14]. In addition to qualitative evaluation by visual inspection, 18F-FDG -PET imaging provides several semi-quantitative or quantitative measurements of radioactivity concentration, such as with the standardized uptake value (SUV) [5]. SUVmax and SUVpeak are routinely used in clinical and research settings [6]. More recently, measures of metabolic tumor burden (MTB) have been explored, including metabolic tumor volume (MTV) and total lesion glycolysis (TLG). MTV indicates the volume of metabolically active tumor. Total lesion glycolysis (TLG) is the product of SUVmean and MTV; it combines the volumetric and metabolic information from 18F-FDG-PET [7]. Metabolic tumor burden measurements of MTV and TLG as well as SUVmax of the primary tumor in patients with NSCLC have been shown to have prognostic value and utility in monitoring treatment response [816].

A critical step in calculating MTB values is the delineation of tumor margin in PET images; accuracy and efficiency of this step are necessary for MTB measures to be used in routine clinical practice. Completely manual tumor margin delineation by the radiologist involves hand drawing a region of interest around the tumor margin; this is not only time-intensive but also tends to have poor inter-observer agreement [17]. Manual adjustment following a semi-automatic gradient method has been used to accomplish this goal and MTB measurements with this method have been found to be associated with patient’s overall survival [1214, 16]. With this method, the initial semi-automatic tumor contouring requires the radiologist to identify the tumor center and its major and minor axes on a single 2-D slice of a 3-D PET image set. By utilizing a gradient method based on spatial derivatives, the software automatically draws a volume of interest (VOI) to delineate tumor margin, and subsequently calculate various tumor metabolic measures such as SUVmax, MTV, or TLG from the VOI[18]. The second step requires the radiologist to adjust the software-generated VOI to make sure that it includes the entire tumor and that no tissue outside the tumor is included. This manual adjustment step can also be time-consuming and observer-dependent.

We conducted this study to determine whether the semi-automatic gradient method without manual adjustment can be used to replace the method with manual adjustment by evaluating the agreement in metabolic and volumetric measurements of the primary tumor obtained with these two methods in 134 NSCLC patients.

Methods

Patient selection/demographics

This study was done under IRB approval. All patients with pathologically confirmed diagnosis of NSCLC and a baseline PET/CT scan between 09/26/2013 and 01/31/2014 at Guangzhou Medical University Hospital, which showed a measurable 18F-FDG-avid primary tumor, were enrolled in this prospective study.

PET/CT technique

All patients underwent a whole-body 18F-FDG PET/CT scan in accordance with the National Cancer Institute guidelines [3]. The 18F-FDG PET images were obtained from a PET/CT scanner (GE Discovery ST, GE, Milwaukee, WI, USA) equipped with high-resolution bismuth germanate (BGO) detector blocks and an 8-slice CT scanner. Patients fasted for at least 4–6 hours before intravenous administration of 3.70–5.55 MBq/kg of 18F-FDG. Serum glucose levels were tested via finger-stick sampling before 18F-FDG injection and a value less than 7.0 mmol/L was required for the PET/CT study to be performed. A whole-body CT scan without IV contrast was performed for PET attenuation correction, by using a standard protocol: 140 kVp, 150 mA, slice thickness 3.75 mm, pitch 1.675, and imaging matrix 512 × 512. About sixty minutes following injection of 18F-FDG, a whole-body static PET scan was acquired for about 21–25 min, starting at the thighs and proceeding to the head, with an acquisition time of 3.5 min per cradle position, by using the 2D acquisition mode with image matrix of 128 × 128. PET images were reconstructed by using the ordered subsets expectation maximization (OSEM) iterative algorithm. CT imaging matrices were converted to 128 × 128 and were used for attenuation correction for PET images.

Primary tumor segmentation and metabolic evaluation

All PET image analyses were done with the PETedge tool of the MIMvista software (MIMvista Corp, Cleveland, OH, USA) at the University of Chicago. After a radiologist indicated the approximate tumor center and drew the major and minor axes of tumor on a single 2-D axial image of a 3-D image set, which were shown automatically on sagittal and coronal planes, the software used a gradient-based (spatial derivatives) method to automatically generate volumes of interest (VOIs) and segment the tumor [18]. The tumor segmentation generated by the software in this manner was what we defined as the “semi-automatic gradient method.” The radiologist can also manually adjust the software-generated VOIs by visual inspection to ensure that the VOI contour matched the metabolically active tumor boundary exactly; we defined this as the “manually adjusted semi-automatic gradient method.” The manual adjustment step used the software’s 3-D ball tool, which is a manual adjustment tool for the tumor contour.

Two radiologists (YP and CZ with more than 7 years of clinical nuclear medicine and FDG PET experience) independently segmented each patient’s primary tumor twice with both the semi-automatic gradient method and manually adjusted semi-automatic gradient method (Figure 1). Subsequently, the following MTB and SUV measurements were obtained based on the segmentation results from each radiologist with each method: SUVmax, SUVpeak, MTV, and TLG. A 1.2-cm-diameter sphere within the tumor was used to calculate SUVpeak.

Figure 1:

Figure 1:

Figure 1:

PET images of a 63-year-old woman with a new diagnosis of NSCLC (adenocarcinoma), showing the measurements of MTV, TLG and maximum and peak SUVs of the primary tumor obtained with the MIMvista PETedge tool by using both the semi-automatic gradient methods without (A, semi-automatic gradient method) and with manual adjustment (B, manually adjusted semi-automatic gradient method).

Statistical analysis

The differences between values obtained from the semi-automatic gradient method versus manually adjusted semi-automatic gradient method were evaluated by using the paired t-test for normally distributed SUV measures, and the Wilcoxon signed-rank test was used for the non-normally distributed MTV and TLG. A natural logarithmic transformation was also applied to MTV and TLG in order to obtain more normally distributed data for these measures, and differences in log(MTV) and log(TLG) measured with the two methods were evaluated by using the paired t-test. Subsequently, measurement agreement between the two methods, and inter-radiologist agreement from each method, were assessed with concordance correlation coefficients (CCCs) [19].

Results

A total of 134 patients met the inclusion criteria and were included in the study. Information regarding patient demographics, diagnosis method, histology type, and stage are summarized in Table 1. Of these, only 121 patients were evaluated for SUVpeak because in 13 patients the small tumor size (less than 1.2 cm) did not meet the criteria for determining this measure.

Table 1.

Patient demographics and tumor characteristics.

Variable Number Percentage (%)
Number of patients 134 100
Gender
 Female 50 37.3
 Male 84 62.7
Age
 Mean±SD 60.8±10.8
 Median 62
 Range 23–83
Diagnosis Methods
 Surgery 70 52.2
 Bronchoscope biopsy 41 30.6
 Transthoracic needle biopsy 21 15.7
 Pleural fluid cytopathology 2 1.5
Histology types
 Adenocarcinoma 89 66.4
 Squamous cell carcinoma 32 23.9
 Other type of NSCLC 13 9.7
TNM Stage
 Stage I A/I B 15/22 11.2/16.4
 Stage II A/II B 9/6 6.7/4.5
 Stage III A/III B 30/13 22.4/9.7
 Stage IV 39 29.1

Note: NSCLC =non-small cell lung cancer.

There was no statistically significant difference observed between values obtained using the semi-automatic gradient method versus manually adjusted semi-automatic gradient method (Table 2). The median (range) MTV values, in ml, was 14.50 (1.22–529.69) and 12.99 (1.42–529.39) from the semi-automatic gradient method and from manually adjusted semi-automatic gradient method, respectively (p=0.45, Wilcoxon signed-rank test). Similarly, the mean (SD) MTV were similar for the two methods, with values in ml of 37.62 (67.36) and 37.34 (67.47), respectively. The median (range) TLG value, in ml, was 82.47 (4.14–4228.87) and 85.17 (3.83–4197.61) from the semi-automatic gradient method and from manually adjusted semi-automatic gradient method, respectively (p=0.37, Wilcoxon signed-rank test). Similar results were found when comparing logarithmically transformed MTV and TLG values between the two methods (p=0.52 and p=0.72, respectively). The mean (standard deviation) of SUVpeak was 10.07 (4.78) and 10.12 (4.80) from the semi-automatic gradient method and from manually adjusted semi-automatic gradient method, respectively (p=0.23, from paired t-test). SUVmax was identical between the two methods in every case, with a mean (SD) of 12.10 (5.83); therefore, further statistical analysis was not performed for this measure.

Table 2.

Comparison of PET measurements obtained with the semi-automatic gradient method and manually adjusted semi-automatic gradient method.

Parameter Number of patients Semi-automatic Manually adjusted semi-automatic Difference^ p-value
SUVmax 134 NA
 Mean (SD) 12.10 (5.83) 12.10 (5.83) 0 (0)
 Median 11.78 11.78
 Range 2.25–32.86 2.25–32.86
SUVpeak 121 0.23*
 Mean (SD) 10.07 (4.78) 10.12 (4.80) −0.05 (0.44)
 Median 9.79 9.80
 Range 1.73–27.27 1.73–27.27
MTV 134 0.45$
 Mean (SD) 37.62 (67.36) 37.34 (67.47) 0.28 (5.67)
 Median 14.50 12.99
 Range 1.22–529.69 1.42–529.39
Log(MTV) 134 0.52*
 Mean (SD) 2.81 (1.23) 2.80 (1.22) 0.01 (0.21)
 Range 0.20–6.27 0.35–6.27
TLG 134 0.37$
 Mean (SD) 276.51 (606.71) 278.59 (607.58) −2.07 (21.38)
 Median 82.47 85.17
 Range 4.14–4228.87 3.83–4197.61
Log(TLG) 0.72*
 Mean (SD) 4.47 (1.49) 4.46 (1.51) 0.01 (0.16)
 Range 1.42–8.35 1.34–8.34

Note: NA = Not Applicable, Log = natural logarithmic transformation, MTV= metabolic tumor volume, SD = standard deviation, SUVmax = maximum standardized uptake value, SUVpeak =peak standardized uptake value, TLG= total lesion glycolysis

^

Semi-automatic – Manually Adjusted

*

Differences tested with the paired t-test.

$

Differences tested with the Wilcoxon signed-rank test.

SUVmax was identical between the two methods in every case; therefore, further statistical analysis was not performed for this measure.

Figure 2 shows scatterplots of the differences in MTV and TLG measured with the semi-automatic gradient method and the manually adjusted semi-automatic gradient method, versus SUVmax of the primary tumor. There was no evidence in these scatterplots to suggest that the differences in MTV and TLG measured with the two methods were dependent on SUVmax of the primary tumor.

Figure 2:

Figure 2:

Figure 2:

Scatterplots of (A) the differences in MTV, and (B) the differences in TLG, measured with the semi-automatic gradient method and with the semi-automatic gradient method followed by manual adjustment, versus SUVmax of the primary tumor. These scatterplots show that the differences in MTV and TLG measured with the two methods are not dependent on SUVmax of the primary tumor.

Excellent agreement between the semi-automatic gradient method and manually adjusted semi-automatic gradient method was found for all measures in terms of CCC, with the lower bound of 95% confidence interval being 0.98 or greater for all measures (Table 3).

Table 3.

Concordance correlation coefficients for PET measurements obtained with the semi-automatic gradient method and manually adjusted gradient method.

Parameter Number of patients CCC 95% CI
SUVmax 134 1.000 NA
SUVpeak 121 0.996 0.994 to 0.997
log (MTV) 134 0.986 0.980 to 0.990
log (TLG) 134 0.994 0.992 to 0.996

Notes: CCC = Concordance correlation coefficients, CI = confidence interval. Other abbreviations are the same as in Table 2.

Excellent agreement was also seen between the two radiologists for both the semi-automatic gradient and manually adjusted semi-automatic gradient method methods in terms of CCC, with the lower bound of 95% confidence interval being 0.86 or greater for all measures (Table 4).

Table 4.

Inter-observer agreement between two radiologists for PET measurements obtained with the semi-automatic gradient method and manually adjusted gradient method.

Parameter Method Number of patients CCC 95% CI
SUVpeak Semi-automatic 109 0.980 0.971 to 0.986
log (MTV) Semi-automatic 134 0.902 0.867 to 0.928
log (TLG) Semi-automatic 134 0.952 0.933 to 0.965
SUVpeak Manual 111 0.989 0.984 to 0.992
log (MTV) Manual 134 0.903 0.868 to 0.930
log (TLG) Manual 134 0.952 0.934 to 0.966

Notes: CCC = Concordance correlation coefficients, CI = confidence interval. Other abbreviations are the same as in Table 2.

Discussion

We demonstrate in this study excellent agreement between MTB measurements of the primary tumor in NSCLC obtained with the semi-automatic gradient method and manually adjusted semi-automatic gradient method, as well as between MTB measurements obtained independently by two radiologists. These results suggest that the additional time-consuming step of manually adjusting semi-automatically selected VOIs may be avoided without significantly affecting MTB values.

MTV and TLG can provide an estimate of the true metabolically active tumor volume. MTV and TLG of the primary tumor and of the whole body have been shown to have prognostic value in patients with NSCLC independent of stage, and better prognostic value than SUVmax or SUVmean [916]. The efficiency and objectivity of the semi-automatic method for calculating MTV and TLG values may help facilitate the use of MTB as a prognostic factor.

PET tumor SUV measurement is also important for evaluating tumor response to therapy [8]. PET Response Criteria in Solid Tumors (PERCIST) based on tumor SUVpeak measurements has been adopted as the new standard method for PET-based evaluation of treatment effectiveness [20]. The close similarity of SUVpeak obtained from the two methods in our study suggests that the semi-automatic method without manual adjustment can be more efficient for PERCIST assessment without sacrificing accuracy. Whole-body MTB quantification may also be valuable for standardized assessment of treatment response because of its demonstrated prognostic value, and the faster semi-automatic method can make MTB more practical to use clinically.

Semi-automatic tumor segmentation and contouring rely on the software’s ability to detect the true margin of the tumor. Two common algorithms used for tumor delineation are the gradient- and threshold-based methods [21,22]. Our study used the gradient method, which has been shown to be more accurate than the threshold method for tumor segmentation [18, 23,24]. MTV estimates from the gradient method are also insensitive to the tumor to background ratio and, for this reason, are expected to be less sensitive to the length of 18F-FDG uptake before PET imaging [18]. More sophisticated segmentation techniques have also been developed, including locally-adaptive region-growing, an iterative technique combining local background correction and adaptive thresholding, as well as other techniques employing fuzzy logic and spatial context to help identify the tumor boundary more accurately [17, 2530]. Although tumor location in the lung may affect accuracy, several studies employing these algorithms demonstrated good correlation of semi-automatically generated tumor volume with pathological specimens [24, 3036]. While semi-automatically derived measurements are more reproducible overall than those obtained from completely manual tumor delineation, the reproducibility of each semi-automatic algorithm varies and further testing must be performed before application of any of these methods to patient care [17, 25, 29, 37].

All of the methods above calculate tumor volume by segmenting the metabolically active region of a tumor seen on PET images. Theoretically, CT and MR images can also be used for estimation of whole-body tumor volume. However, the boundary of the metabolically active tumor can be difficult to differentiate from the surrounding anatomy based on CT or MR, and thus PET/CT is more accurate for this purpose. A study comparing tumor-volume measurement from CT, MRI, and PET for head and neck cancer showed that PET-derived MTV is the closest to the reference tumor volume from surgical specimens [38]. Another study of NSCLC showed that PET/CT images produce more accurate tumor-size measurement than PET alone or CT alone [39]. Other studies have also shown that PET/CT is more accurate than CT or MRI for characterization of various cancers, including NSCLC [4046]. Another advantage of PET/CT is that oncologic PET/CT scans are already routinely performed from “eyes to thighs” [47], thus facilitating whole-body metabolic tumor burden measurements for staging without incurring an additional imaging study.

Our study has several limitations. First, all the measurements were performed with one particular software tool and one particular initial semi-automatic segmentation method. It is possible that the agreement between the derived MTB values from the semi-automatic gradient method versus from manually adjusted semi-automatic gradient method that we have observed may not hold for software tools from other vendors. However, continued development of segmentation methods should lead to even more accurate and efficient calculation of metabolic tumor values. Second, although there is a possibility of bias toward agreement between the two segmentation methods, because the agreement would have been greatest if the observers simply did not manually adjust the VOIs, this study was not done for the purpose of comparing the two segmentation methods, but to compare VOIs measured from PET scans at two time points (results not reported here). Therefore, the observers simply tried to measure the VOIs as accurately as possible, and we expect the possibility of this bias to be minimal. Third, the spatial resolution of PET in our study was determined by a 128 × 128 matrix size, which is no longer considered state-of-the-art, compared with 192 or 256 matrices; it is unknown whether the software will perform better or worse in generating accurate tumor contours with larger matrices. Fourth, we compared MTB measurements of only the primary tumor in a limited number of patients with untreated NSCLC and we did not exclude the central necrosis in the tumor. Finally, given the high contrast between bright metabolically active tumor and dark background lung parenchyma, it is possible that the software performs better in delineating primary lung tumor contours than it will for tumor in other parts of the body where the tumor to background contrast is less pronounced. Further studies are needed to confirm that the semi-automatic method is equally accurate as the method with manual adjustment for a more heterogeneous patient population with metastasis in other locations.

Conclusion

We have demonstrated excellent agreement for metabolic PET measurement of the primary tumor obtained from the semi-automatic gradient method and manually adjusted semi-automatic gradient method in NSCLC patients. Given the close similarity in MTB values obtained with these two methods, the semi-automatic method can be used to provide better efficiency in metabolic tumor burden evaluation, and improve the utility of MTB in both research and clinical practice. However, we recommend that a final review of the VOI boundary by a radiologist should always be performed when using semi-automatic method.

Acknowledgments

Funding information: This work was supported in part by a grant (R21 CA181885) from the National Cancer Institute of the National Institutes of Health

Footnotes

Compliance with Ethical Standards

Disclosure of potential conflicts of interest:

Piotr Obara, Haiping Liu, Kristen Wroblewski, Chen-Peng Zhang, Peng Hou, Yulei Jiang, Chen Ping and Yonglin Pu declare that they have no conflict of interest.

Research involving Human Participants:

The study was done under IRB approval of the First Affiliated Hospital of Guangzhou Medical University.

Informed consent:

Informed consent was obtained from the research subjects.

References

  • 1.Black QC, Grills IS, Kestin LL, Wong CY, Wong JW, Martinez AA et al. Defining a radiotherapy target with positron emission tomography. International journal of radiation oncology, biology, physics 2004; 60: 1272–1282. [DOI] [PubMed] [Google Scholar]
  • 2.Bradley J, Thorstad WL, Mutic S, Miller TR, Dehdashti F, Siegel BA et al. Impact of FDG-PET on radiation therapy volume delineation in non–small-cell lung cancer. International journal of radiation oncology, biology, physics 2004; 59: 78–86. [DOI] [PubMed] [Google Scholar]
  • 3.Shankar LK, Hoffman JM, Bacharach S, Graham MM, Karp J, Lammertsma AA et al. Consensus recommendations for the use of 18F-FDG PET as an indicator of therapeutic response in patients in National Cancer Institute Trials. J Nucl Med 2006;47: 1059–1066. [PubMed] [Google Scholar]
  • 4.Hoekstra CJ, Paglianiti I, Hoekstra OS, Smit EF, Posmus PE, Teule GJ, et al. Monitoring Response to Therapy in Cancer Using 18F-2-fluoro-2-deoxy-D-glucose and Positron Emission Tomography: an Overview of Deferent Analytical Methods. European Journal of Nuclear Medicine and Molecular Imaging 2000; 27: 731–743. [DOI] [PubMed] [Google Scholar]
  • 5.Strauss LG, Conti PS. The applications of PET in clinical oncology. J Nucl Med 1991;32: 623–648. [PubMed] [Google Scholar]
  • 6.Bai B, Bading J, Conti P. Tumor quantification in clinical positron emission tomography. Theranostics 2013;3: 787–801. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Larson SM, Erdi Y, Akhurst T, Mazumdar M, Macapinlac HA, Finn RD et al. Tumor treatment response based on visual and quantitative changes in global tumor glycolysis using PET-FDG imaging: the visual response score and the change in total lesion glycolysis. Clin Positron Imaging 1999;2: 159–171. [DOI] [PubMed] [Google Scholar]
  • 8.Bral S, Duchateau M Volumetric response analysis during chemoradiation as predictive tool for optimizing treatment strategy in locally advanced unresectable NSCLC. Radiotherapy and Oncology 2009;91: 438–442. [DOI] [PubMed] [Google Scholar]
  • 9.Mehta G1, Chander A, Huang C, Kelly M, Fielding Feasibility study of FDG PET/CT-derived primary tumour glycolysis as a prognostic indicator of survival in patients with non-small-cell lung cancer. Clin Radiol 2014;69: 268–74. [DOI] [PubMed] [Google Scholar]
  • 10.Tong A, Han S, Yan P, Gong H, Zhao H, Yao H et al. Prognostic value of FDG uptake in primary inoperable non-small cell lung cancer. Med Oncol 2014;31:780. [DOI] [PubMed] [Google Scholar]
  • 11.Horne ZD, Clump DA, Vargo JA, Shah S, Beriwal S, Burton SA, Quinn AE et al. Pretreatment SUVmax predicts progression-free survival in early-stage non-small cell lung cancer treated with stereotactic body radiation therapy. Radiat Oncol 2014;9: 41. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Liao S, Penney BC, Wroblewski K, Zhang H, Simon CA, Kampalath R et al. Prognostic value of metabolic tumor burden on 18F-FDG PET in nonsurgical patients with non-small cell lung cancer. Eur J Nucl Med Mol Imaging 2012;39: 27–38. [DOI] [PubMed] [Google Scholar]
  • 13.Liao S, Penney BC, Zhang H, Suzuki K, Pu Y Prognostic value of the quantitative metabolic volumetric measurement on 18F-FDG PET/CT in Stage IV nonsurgical small-cell lung cancer. Acad Radiol 2012;19: 69–77. [DOI] [PubMed] [Google Scholar]
  • 14.Zhang H, Wroblewski K, Liao S, Kampalath R, Penney BC, Zhang Y et al. Prognostic value of metabolic tumor burden from (18)F-FDG PET in surgical patients with non-small-cell lung cancer. Acad Radiol 2013;20: 32–40. [DOI] [PubMed] [Google Scholar]
  • 15.Zhang H, Wroblewski K, Liao S, Kampalath R, Penney BC, Zhang Y et al. Prognostic value of metabolic tumor burden from (18)F-FDG PET in surgical patients with non-small-cell lung cancer. Acad Radiol 2013;20: 32–40. [DOI] [PubMed] [Google Scholar]
  • 16.Davison J, Mercier G, Russo G, Subramaniam RM. PET-based primary tumor volumetric parameters and survival of patients with non-small cell lung carcinoma. AJR Am J Roentgenol 2013;200: 635–40. [DOI] [PubMed] [Google Scholar]
  • 17.Hofheinz F, Pötzsch C, Ohme L, Beuthien-Baumann B, Steinbach J, Kotzerke J et al. Automatic volume delineation in oncological PET. Evaluation of a dedicated software tool and comparison with manual delineation in clinical data sets. Nuklearmedizin 2012;51: 9–16. [DOI] [PubMed] [Google Scholar]
  • 18.Werner-Wasik M, Nelson AD, Choi W, Arai Y, Faulhaber PF, Kang P et al. What is the best way to contour lung tumors on PET scans? Multiobserver validation of a gradient-based method using a NSCLC digital PET phantom. Int J Radiat Oncol Biol Phys 2012;82: 1164–1171. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Lin LI.A concordance correlation coefficient to evaluate reproducibility. Biometrics 1989;45: 255–268. [PubMed] [Google Scholar]
  • 20.Wahl RL, Jacene H, Kasamon Y, Lodge MA. From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med 2009;50: 122S–150S. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Schinagl DA, Vogel WV, Hoffmann AL, van Dalen JA, Oyen WJ, Kaanders JH. Comparison of five segmentation tools for 18F-Fluoro-deoxy-glucose–positron emission tomography–based target volume definition in head and neck cancer. International Journal of Radiation Oncology Biology Physics 2007;69: 1282–1289. [DOI] [PubMed] [Google Scholar]
  • 22.Nestle U, Schaefer-Schuler A, Kremp S, Groeschel A, Hellwig D, Rube C et al. Target volume definition for (18)F-FDG PET-positive lymph nodes in radiotherapy of patients with non-small cell lung cancer. Eur J Nucl Med Mol Imaging 2007;34: 453–462. [DOI] [PubMed] [Google Scholar]
  • 23.Wu K, Ung YC, Hwang D, Tsao MS, Darling G, Maziak DE et al. Autocontouring and manual contouring: which is the better method for target delineation using 18F-FDG PET/CT in non-small cell lung cancer? J Nucl Med 2010;51: 1517–23. [DOI] [PubMed] [Google Scholar]
  • 24.Wanet M, Lee JA, Weynand B, De Bast M, Poncelet A, Lacroix V et al. Gradient-based delineation of the primary GTV on FDG-PET in non-small cell lung cancer: a comparison with threshold-based approaches, CT and surgical specimens. Radiother Oncol 2011;98: 117–125. [DOI] [PubMed] [Google Scholar]
  • 25.Hatt M1, Cheze-Le Rest C, Aboagye EO, Kenny LM, Rosso L, Turkheimer FE et al. Reproducibility of 18F-FDG and 3’-deoxy-3’−18F-fluorothymidine PET tumor volume measurements. J Nucl Med 2010;51: 1368–1376. [DOI] [PubMed] [Google Scholar]
  • 26.Li H, Thorstad WL, Biehl KJ, Laforest R, Su Y, Shoghi KI et al. A novel PET tumor delineation method based on adaptive region-growing and dual-front active contours. Med Phys 2008; 35: 3711–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Hatt M, Cheze le Rest C, Turzo A, Roux C, Visvikis D.A fuzzy locally adaptive Bayesian segmentation approach for volume determination in PET. IEEE Trans Med Imag 2009; 28: 881–893. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Belhassen S, Zaidi H. A novel fuzzy C-means algorithm for unsupervised heterogeneous tumor quantification in PET. Med Phys 2010;37: 1309–1324. [DOI] [PubMed] [Google Scholar]
  • 29.Hatt M, Cheze Le Rest C, Albarghach N, Pradier O, Visvikis D. PET functional volume delineation: a robustness and repeatability study. Eur J Nucl Med Mol Imag 2011;38: 663–72. [DOI] [PubMed] [Google Scholar]
  • 30.Hatt M, Cheze-le Rest C, van Baardwijk A, Lambin P, Pradier O, Visvikis D. Impact of tumor size and tracer uptake heterogeneity in 18F-FDG PET and CT non-small cell lung cancer tumor delineation. J Nucl Med 2011;52: 1690–1697. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Schaefer A, Kim YJ, Kremp S, Mai S, Fleckenstein J, Bohnenberger H et al. PET-based delineation of tumour volumes in lung cancer: comparison with pathological findings. Eur J Nucl Med Mol Imaging 2013;40: 1233–1244. [DOI] [PubMed] [Google Scholar]
  • 32.Cheebsumon P1, Boellaard R, de Ruysscher D, van Elmpt W, van Baardwijk A, Yaqub M et al. Assessment of tumour size in PET/CT lung cancer studies: PET- and CT-based methods compared to pathology. EJNMMI Res 2012; 2: 56. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.van Baardwijk A, Bosmans G, Boersma L, Buijsen J, Wanders S, Hochstenbag M et al. PET-CT-based auto-contouring in non-small-cell lung cancer correlates with pathology and reduces interobserver variability in the delineation of the primary tumor and involved nodal volumes. Int J Radiat Oncol Biol Phys 2007;68: 771–778. [DOI] [PubMed] [Google Scholar]
  • 34.Hatt M, le Cheze RC, Descourt P, Dekker A, De RD, Oellers M et al. Accurate automatic delineation of heterogeneous functional volumes in positron emission tomography for oncology applications. Int J Radiat Oncol Biol Phys 2010;77: 301–308. [DOI] [PubMed] [Google Scholar]
  • 35.Yu J, Li X, Xing L, Mu D, Fu Z, Sun X et al. Comparison of tumor volumes as determined by pathologic examination and FDG-PET/CT images of non-small-cell lung cancer: a pilot study. Int J Radiat Oncol Biol Phys 2009;75: 1468–1474. [DOI] [PubMed] [Google Scholar]
  • 36.Wu K, Ung YC, Hornby J, Freeman M, Hwang D, Tsao MS, Dahele M et al. PET CT thresholds for radiotherapy target definition in non-small-cell lung cancer: how close are we to the pathologic findings? Int J Radiat Oncol Biol Phys 2010;77: 699–706. [DOI] [PubMed] [Google Scholar]
  • 37.Jacene HA, Leboulleux S, Baba S, Chatzifotiadis D, Goudarzi B, Teytelbaum O, Horton KM et al. Assessment of interobserver reproducibility in quantitative 18F-FDG PET and CT measurements of tumor response to therapy. J Nucl Med 2009;50: 1760–9. [DOI] [PubMed] [Google Scholar]
  • 38.Daisne JF, Duprez T, Weynand B, Lonneux M, Hamoir M, Reychler H et al. Tumor volume in pharyngolaryngeal squamous cell carcinoma: Comparison at CT, MR imaging, and FDG PET and validation with surgical specimen. Radiology 2004;233: 93–100. [DOI] [PubMed] [Google Scholar]
  • 39.Yu HM, Liu YF, Hou M, Liu J, Li XN, Yu JM.Evaluation of gross tumor size using CT, 18F-FDG PET, integrated 18F-FDG PET/CT and pathological analysis in non-small cell lung cancer. Eur J Radiol 2009;72: 104–113. [DOI] [PubMed] [Google Scholar]
  • 40.Beggs AD, Hain SF, Curran KM, O’Doherty MJ. FDG-PET as a “metabolic biopsy” tool in non-lung lesions with indeterminate biopsy. Eur J Nucl Med Mol Imaging 2002;29: 542–546. [DOI] [PubMed] [Google Scholar]
  • 41.Antoch G, Saoudi N, Kuehl H, Dahmen G, Mueller SP, Beyer T et al. Accuracy of whole-body dual-modality fluorine-18–2-fluoro-2-deoxy-D-glucose positron emission tomography and computed tomography (FDG-PET/CT) for tumor staging in solid tumors: Comparison with CT and PET. J Clin Oncol 2004;22: 4357–4368. [DOI] [PubMed] [Google Scholar]
  • 42.Fischer BM, Mortensen J, Langer SW, Loft A, Berthelsen AK, Petersen BI, et al. A prospective study of PET/CT in initial staging of small-cell lung cancer: comparison with CT, bone scintigraphy and bone marrow analysis. Ann Oncol 2007;18: 338–345. [DOI] [PubMed] [Google Scholar]
  • 43.Ohno Y, Koyama H, Onishi Y, Takenaka D, Nogami M, Yoshikawa T et alNon-small cell lung cancer: whole-body MR examination for M-stage assessment-utility for whole-body diffusion weighted imaging compared with integrated FDG PET/CT. Radiology 2008;248: 643–654. [DOI] [PubMed] [Google Scholar]
  • 44.Yi CA, Shin KM, Lee KS, Kim BT, Kim H, Kwon OJ et al. Non-small cell lung cancer staging: efficacy comparison of integrated PET/CT versus 3.0-T whole-body MR imaging. Radiology 2008;248: 632–642. [DOI] [PubMed] [Google Scholar]
  • 45.Plathow C, Aschoff P, Lichy MP, Eschmann S, Hehr T, Brink I et al. Positron emission tomography/computed tomography and whole-body magnetic resonance imaging in staging of advanced non-small cell lung cancer: initial results. Invest Radiol 2008;43: 290–297. [DOI] [PubMed] [Google Scholar]
  • 46.Xu G, Zhao L, He Z. Performance of whole-body PET/CT for the detection of distant malignancies in various cancers: A systematic review and meta-analysis. J Nucl Med 2012;53: 1847–1854. [DOI] [PubMed] [Google Scholar]
  • 47.Coleman RE, Delbeke D, Guiberteau MJ, Conti PS, Royal HD, Weinreb JC et al. Concurrent PET/CT with an integrated imaging system: intersociety dialogue from the joint working group of American College of Radiology, the Society of Nuclear Medicine, and the Society of Computed Body Tomography and Magnetic Resonance. J Nucl Med 2005;46: 1225–39. [PubMed] [Google Scholar]

RESOURCES