Skip to main content
EPA Author Manuscripts logoLink to EPA Author Manuscripts
. Author manuscript; available in PMC: 2019 Aug 1.
Published in final edited form as: Environ Toxicol Chem. 2018 Jul 13;37(8):2089–2097. doi: 10.1002/etc.4167

Using performance reference compounds to compare mass transfer calibration methodologies in passive samplers deployed in the water column

Abigail S Joyce a,*, Robert M Burgess b
PMCID: PMC6122610  NIHMSID: NIHMS1500792  PMID: 29744915

Abstract

Performance reference compounds (PRCs) are often added to passive samplers prior to field deployments to provide information about mass transfer kinetics between the sampled environment and the passive sampler. Their popularity has resulted in different methods of varying complexity to estimate mass transfer and better estimate freely dissolved concentrations (Cfree) of targeted compounds. Three methods for describing a mass transfer model are commonly used: a first order (FO) kinetic method, a non-linear least squares (NLS) fitting of sampling rate, and a diffusion (DIF) method. PRC-loaded, low-density polyethylene (PE) strips of four different thicknesses were used as passive samplers to create an array of PRC results to assess the comparability and reproducibility of each of the methods. Samplers were deployed in the water column at three stations in New Bedford Harbor (MA, USA). Collected data allowed Cfree comparisons to be performed in two ways: (1) comparing Cfree derived from one thickness using different methods and (2) comparing Cfree derived by the same method using different thicknesses of PE. Overall, the NLS and DIF methods demonstrated the most precise results for all of the PCBs measured and generated Cfree values that were often statistically indistinguishable. Relative standard deviations (RSDs) for total PCB measurements using the same thickness and varying model types range from 0.04–12% and increased with sampler thickness and RSDs for estimates using the same method and varying thickness ranged from 8 – 18%. Environmental scientists and managers are encouraged to use these methods when estimating Cfree from passive sampling and PRC data.

Keywords: passive sampling, polychlorinated biphenyls (PCBs), persistent organic pollutants (POPs), performance reference compounds (PRCs)

Graphical Abstract

graphic file with name nihms-1500792-f0004.jpg

Mean freely dissolved water concentration (Cfree) per PCB conger as measured without PRC correction and by each examined PRC modeling approach for a 76 ¼m thick LDPE passive sampler deployed in the water column of New Bedford Harbor.

INTRODUCTION

Passive sampling in environmental applications has evolved from a scientific concept to a tool being promoted for use in regulatory decision making (Greenberg et al., 2014, Booij et al., 2016, Mayer et al., 2003). During this evolution, several review articles have summarized the scientific foundations and approaches for passive sampling including its ability to estimate a chemical’s freely dissolved concentration (Cfree)(Lohmann, 2012, Lydy et al., 2014, Mayer et al., 2014, Ghosh et al., 2014). The importance of Cfree as a surrogate for a chemical’s actual bioavailable concentrations was described by (Di Toro et al., 1991) but measuring or estimating Cfree in the water column or sediment interstitial waters can be challenging because of experimental artifacts (e.g., presence of colloids and dissolved organic carbon) and modeling uncertainties. Several studies have demonstrated the advantages of using passive sampling compared to traditional sampling and modelling approaches for measuring Cfree (Gschwend et al., 2011, Burgess et al., 2015, Maruya et al., 2009). Using passive sampling to estimate Cfree, environmental managers can make better informed decisions with greater certainty about the magnitude of risk associated with target contaminants at contaminated sites.

Ideally, when using passive sampling to measure Cfree in the water column or sediment interstitial waters a thermodynamic equilibrium between the sampling environment and passive sampler is attained for all targeted contaminants. This equilibrium allows for the use of Equation 1 for calculating Cfree (ng/L):

Cfree=CPSKPW (1)

where, CPS(ng/kgpolymer)is the concentration of a target contaminant in the passive sampler at equilibrium and KPW (Lwater/kgpolymer) is its polymer-water partition coefficient. In some instances, the application of the passive samplers requires ‘real world’ scenarios to be assessed and exposures must be performed in the field where environmental conditions may control the rate at which equilibrium is attained (Fernandez et al., 2014). In this respect, highly hydrophobic organic contaminants (HOCs) of interest may require months or even years to reach equilibrium. Assuming equilibrium has been attained and using Equation 1 to calculate Cfree may result in under-estimating Cfree and ultimately under-estimating exposure and risk for many hydrophobic compounds that have not reached equilibrium. In these cases, the use of performance reference compounds (PRCs) allows Cfree to be estimated from the measured passive sampler concentration at a given time, CPSt(ng/kgpolymer), even when a compound has not attained equilibrium (Huckins et al., 2002). PRCs are dosed into a passive sampler prior to deployment and their release from the sampler can then be used to determine mass transfer kinetics to more accurately estimate Cfree. Several methods, of varying complexity, describing this mass transfer have been developed to generate practical information about equilibrium conditions using PRC data.

The availability of multiple methods for estimating Cfree using PRC data allows environmental scientists and managers options for assessing risk. The objective of this investigation was to generate a data set of target contaminant concentrations, estimate Cfree using three different methods to model mass transfer, and compare the results. Low-density polyethylene (PE) in four thicknesses was used as a passive sampler and the target contaminants were 27 polychlorinated biphenyl (PCB) congeners. The deployment was performed in the water column of three stations at the New Bedford Harbor Superfund site in New Bedford, Massachusetts (USA).

From an environmental management perspective, the goal of this research is to provide guidance to environmental managers and scientists at contaminated sediment sites for selecting which method(s) to use when analyzing PRC data from passive sampling deployments.

Theory

The dominating environmental factor controlling sampling kinetics in the water column is the water boundary layer (WBL). The WBL is the region separating the passive sampler from the bulk water where molecular diffusion dictates mass transfer and resistance to mass transfer. HOCs are often rate limited by diffusion through the WBL when using PE as a sampling medium (Tomaszewski and Luthy, 2008, Belles et al., 2016, Vrana and Schüürmann, 2002). When kinetics are WBL-controlled there are several methods to estimate Cfree using PRCs pre-loaded into a passive sampler. The simplest method for estimating Cfree is using the release rate of PRCs based on first order kinetics which assumes that uptake into the passive sampler follows a first order (FO) kinetic regime as described by:

Cfree=CPStKPW(1eket) (2)

where, t (days) is the deployment time and ke(days−1) is the overall exchange rate constant. The value for is determined from the fraction of PRC retained by the sampler, fret, over the course of the deployment:

ke=(ln(cPRCtcPRC0)t) (3)

where, CPRC0and CPRCt (ng/kgpolymer) are the PRC concentrations in the PE prior to deployment and after deployment, respectively, this ratio defines fret. Often there is a specific PRC for each targeted HOC in the study (Joyce et al., 2015), though has been interpolated for target HOCs by correlation with KOW, molecular volume, or molecular weight when a dedicated PRC was not used (Friedman et al., 2012, Perron et al., 2013a, Perron et al., 2013b). Because of analytical issues associated with low fret (i.e., quantification limit issues for low KOW PRCs) and high fret (i.e., difficulty differentiating CPRC0 and CPRCt of high KOW PRCs), the literature suggests limiting the use of the fret to 0.2 < fret< 0.8, sometimes referred to as the 80/20 method (Smedes, 2007, Booij and Smedes, 2010).

Taking the mass transfer model one step further and calculating a sampling rate, RS (L/day), introduces physiochemical characteristics of the PRCs and target compounds such that predictive relationships between all PRC data can be used to estimate compound specific sampling rates for targeted compounds. (Booij and Smedes, 2010) reported a method for better estimating RS using a non-linear least squares (NLS) fit:

fret=eRst/mpKPW (4)

where mp is the mass (kg) of the passive sampler and RS is the product of mp, KPW, and ke. It allows for using all acquired PRC data – even data for which the fraction of PRC retained by the passive sampler is less than 0.20 or greater than 0.8. This formula gives a site-specific RS, guidance in calculating compound specific can be found in the literature(Booij and Smedes, 2010, Rusina et al., 2010). Once a compound specific RS has been determined, Cfree can be estimated from:

Cfree=CPStmpRSt (5)

A third method for modelling mass transfer has also been developed for both the sediments and the water column and accounts for compound-specific diffusivities within both the sampler and the WBL or interstitial water (Fernandez et al., 2009, Fernandez et al., 2012, Apell and Gschwend, 2014, Thompson et al., 2015). The diffusion method (DIF) is useful when uptake is controlled or partially controlled by the polymer, or if it is unclear how a compound’s diffusion through the polymer will affect transfer kinetics. The model, derivation, and code used in this study were originally reported by (Thompson et al., 2015). It is based on Fickian diffusion in a symmetrical system with PE thickness (cm) of 2l and WBL of thickness b where x = 0 is the center of the PE film. Diffusion is assumed to be unidirectional through half the sheet thickness. The resulting governing equations for the system become:

CPSt=DPE2CPSx2−for0<x<l (6)
CWBLt=DW2CWBLx2 for l<x<l+b (7)

where, DPE and DW are the diffusivities (¼m2/s) of each HOC through the PE and water, respectively, x is the distance from the center of the PE, l is half the thickness of the PE (¼m), and b is the thickness of the WBL (¼m). Solving this system of partial differential equations allows for a calculation of the WBL thickness using PRC results. Assuming that compounds with varying physio-chemical properties experience the same WBL, data from several PRCs can be used to estimate an average WBL thickness though similar fret restrictions to the FO method must be implemented into the input. Once WBL thickness is known, fractional equilibriums (feq), can be estimated for target HOCs. For clarification, the feq refers to the fraction of equilibrium that had been attained by a target contaminant over the deployment period. It is related to the fret in that the sum of feq and fret should be one for a target contaminant and its corresponding PRC. Once feq are calculated for the target compounds, Cfree can then be calculated as:

Cfree=(cPStfeq)/KPW (8)

The error associated with Cfree can be attributed to different factors depending on the method used to estimate equilibrium conditions. In general, the error is mainly dependent on KPW. Log KPW values usually have an error of ±0.2 which can affect the calculated Cfree by a factor of two. This work used an estimated set of log KPW based on a log KOW values (Ghosh et al., 2014, Hawker and Connell, 1988), as such KPW would contribute to the systematic error of Cfree estimates. For compounds near equilibrium, KPW will be the dominant contributor to error and as the CPRCt/CPRC0 ratio gets smaller, analytical errors in these measurements will have a larger contribution to total error. When compounds are not at equilibrium, the KPW and analytical error can amplify the error in Cfree. For instance, error is introduced to the FO method because this method only uses one PRC for each targeted compound, whereas the NLS and DIF methods use data from all PRCs which should average-out analytical errors introduced by calculating CPRCtCPRC0. In this way, errors introduced by KPW, CPRCt, and CPRC0contribute to total error differently depending on how close to equilibrium a compound is. As such, errors contributed by KPW in Cfree for Equations 1,2, and 8 would only contribute a bias between the Cfree values from the different thicknesses tested. However, because KPW is used to determine RS, errors in KPW will contribute to the final Cfree generated by the NLS method, similar errors associated with DW and DPS will contribute to Cfree error using the DIF method. These factors will likely contribute more toward bias and not variability. Much of the variability from the equilibrium corrections will be introduced through analytical error associated withCPRCt, and CPRC0.

MATERIALS AND METHODS

Materials

Four different thicknesses of low-density PE sheeting (Ace Hardware, Oak Brook, IL, USA; 12.5 ¼m, 25.4 μm, 50.8¼m, and 76.2 ¼m thick) were cut into strips (8cm x 50cm, ~ 0.5 to 3 g) and pre-cleaned in dichloromethane (DCM; 1 day), methanol (MeOH; 2 days), and stored in ultrapure water (18.3 MΩ) prior to loading PRCs. Passive samplers were separated according to thickness. PRCs were loaded into cleaned PE in 2 L of a 4:1 MeOH:water spiking solution containing 13C-labeled PCB congeners: 8, 28, 52, 101, 138, and 180 (99%, Cambridge Isotope Laboratories, Inc., Andover, MA, USA) for 36 days while gently shaking on an orbital shaker. The nominal concentration of each PRC in the spiking solution is given in Table S1. PRC-loaded PE strips were then dried and woven on stainless steel wire (Malin Co., Brookpark, OH, USA; 20 gauge) which was then shaped into a ring. Prior to weaving on the wire, a piece of PE (~5 cm) was cut from each sampler to determineCPE0. Samplers were stored at −20 °C or on ice in the dark until deployment.

The target compounds of interest included 27 chlorinated biphenyls (CBs): 8, 18, 28, 44, 52, 66, 70, 77, 81, 99, 101, 105, 110, 114, 118, 123, 126, 138, 153, 156, 157, 167, 169, 170, 180, 189, and 206. Standards for non-labeled target PCBs (Ultra Scientific, North Kingston, RI, USA) were purchased individually, combined, and diluted in hexane to make a 3000 ng/mL stock standard solution. Internal standard, CB198, was also purchased from Ultra Scientific and recovery standards, 13C-PCB: 9, 118, and 188, were acquired from Cambridge Isotope Laboratories, Inc. ACS/Pesticide grade DCM and MeOH, were purchased from Honeywell (Muskegon, MI, USA) and used without further purification. Ultra-resi analysed grade hexane (95% n-hexane) was purchased from J.T. Baker (Center Valley, PA, USA). Concentrated sulfuric acid (Fisher Scientific, Hampton, NH, USA) was used to clean-up extracts when necessary.

Field studies

In September of 2014, all thicknesses of PE were deployed for 30 days, in triplicate, at two stations in New Bedford Harbor (NBH 2 and NBH4) and one station in Buzzards Bay (NBH5) (Figure S1). One non-PRC-loaded PE of each thickness was exposed to air during deployment and retrieval operations at each station to serve as a station-specific travel/field blank. One PRC-loaded PE blank of each thickness was also exposed but not deployed at all three stations and served as a travel/field blank. Upon retrieval, PE samplers were placed in aluminum foil and stored on ice during transit. Once at the laboratory, samplers were wiped clean of visible residue/biological growth using a lab tissue and rinsed with de-ionized water. Samplers were then dried and individually stored in pre-cleaned glass vials at −20°C in the dark until extraction.

Sample processing and analysis

PE samplers and blanks were spiked with recovery standards: 13Clabeled PCB 9, 118, and 188 (12.5 ¼L; 20 ¼g/mL) and extracted three times by gently mixing on an orbital shaker in DCM (~90mL, overnight). The combined DCM extract was concentrated using a TurboVap (Zymark, Hopkinton, MA, USA; 5–10 psi, 35°C), exchanged to hexane, and further concentrated to 0.5 mL. The TurboVap vial was then rinsed with an additional 0.5 mL of hexane and combined with the concentrated extract for 1 mL total volume. Internal standard, CB198 (12.5 ¼L; 20 ¼g/mL), was then added to the extract. Concentrated extracts from Stations NBH4 and NBH5 were dark green and therefore were acid washed with concentrated H2SO4 (200 ¼L) and allowed to separate overnight. The organic solvent layer was then pipetted off and stored. All sampler extracts were stored in amber glass, crimp cap vials at −20 °C in the dark until analysis by GC-MS (details in Supporting Information).

Data analysis and quality control

Chromatograms were integrated using MSD ChemStation (E.02.00.493; Santa Clara, CA, USA) and quantified by the internal standard method using a seven-point calibration curve. Analysis of pre-cleaned, virgin PE showed that no target contaminants were detectable prior to deployment. Six of the 5-cm, pre-cut pieces per thickness were extracted to determine CPRC0 for all PRCs, CPRC0values for each PRC agreed within 15%. One 76 ¼m thick PE replicate was lost during analysis, thus only duplicates were available. Throughout this paper, data are presented as the mean ± one standard deviation, unless otherwise stated. Recoveries of 13C-PCB 9, 118, and 188 were 51±9, 83±7, and 76± 6, respectively, for all PE at all three stations. All target contaminants were corrected for surrogate recovery according to Table S2. Cfree was calculated using a compound specific KPW (Table S2) and corrected for nonequilibrium conditions using the correction methods described in the Theory section. To compare PCB congener and total PCB Cfree values generated by different thicknesses and method, analysis of variance (ANOVA) was performed followed by a protected Fisher’s test to identify any statistically-significant differences between method estimates of Cfree. An α = 0.05 was used to indicate significant differences between mean Cfree values. Analyses were performed using SAS 9.4 (Cary, NC, USA).

RESULTS AND DISCUSSION

Comparisons of the Cfree generated by the different methods were performed in two ways: comparing Cfree from different thicknesses of PE using the same correction method and comparing Cfree from the same thickness using a different correction method. Similar concentration trends were observed at each sampling station. Comparisons from NBH2 will be shown in the main text and corresponding tables and figures for NBH4 and NBH5 are provided in the Supporting Information. All Cfree values are given in Supporting Information (Tables S3 - S5). Cfree results using equation 1 are reported, and compared, labeled as “no model” to illustrate the differences when equilibrium is assumed to be achieved.

PRC results

As expected, fret increased with increasing PRC log KOW as well as with increasing PE thickness (Figure 1). PRC fraction retained was lowest at NBH5, followed by NBH2 and highest at NBH4 (Table S6). In most cases, small variations were observed between a station’s triplicates. However, in some instances, substantial variability was observed, so calculations of ke, Rs, and WBL thickness (Table S7) were performed for each replicate as opposed to taking a mean for each station. The retention of the PRCs, 13C-PCB 8 and 13C-PCB 28, on the 25.4 ¼m thick PE was higher than expected at all stations and on all replicates. The 13C-PCB 8 results had a small influence on the resulting Cfree calculations which can be observed in the elevated Cfree concentration of CB8 using the FO method. The 13C-PCB 28 results had more variation as evidenced in both the FO and DIF model. The origins of this behavior are unclear but this behavior was consistent for all deployed PE. There was no indication of contamination in either the laboratory blanks or non-PRC loaded field blanks, and full scan data was not available due to their low concentrations. There was no evidence that these results were incorrect and they were used when applicable. All estimated feq values are given in Supporting Information Table S8.

Figure 1.

Figure 1.

Performance reference compound (PRC) fraction retained (fret) from the deployment of four thicknesses of polyethylene passive samplers at station NBH2. PRCs are 13C-labelled PCB 8, 28, 52, 101, 138 and 180. The fret values are plotted against log KOW of the PRCs.

Intra-method comparison of congener Cfree

PRCs suggested that CB8 had reached equilibrium for all PE deployed. Calculating Cfree using Equation 1 for all PE yielded approximately 20% relative standard deviation (RSD; Figure 2). This RSD was also observed at the NBH4 station for CB8 (Figure S4) suggesting that a 20% RSD would be representative of the precision for passive sampling in this investigation and served as a standard for good agreement for congeners where an equilibrium correction was appropriate. (Apell and Gschwend, 2016) came to a similar conclusion finding an estimated PE sampler RSD precision of 27%.

Figure 2.

Figure 2.

Relative standard deviation (RSD; n = 12) of Cfree values for all four thicknesses at station NBH2 measured with a given equilibrium model: (1) No model, (2) first order kinetic (FO) method, (3) non-linear least squared (NLS) sampling rate method, and (4) Diffusion (DIF) method. The designations (e.g., ‘Di’, ‘Tri’, ‘Tetra’) indicate the chlorination level of congeners along the x-axis.

The RSDs increased (Figure 2) as the congeners became more hydrophobic suggesting that equilibrium had not been attained on all samplers during deployment. Referencing a congener from each chlorination level, specifically, CB28, CB52, CB101, CB138, CB180, and CB206, the corresponding RSDs were: 20%, 15%, 34%, 46%, 50%, and 53%, respectively, suggesting that CB28 and CB52 had reached equilibrium based on the previously defined 20% RSD standard. Both the NLS and DIF methods suggested that CB8 and CB28 had a feq of 0.9 or greater for most of the PE deployed, CB52 predicted feq ≥ 0.9 in half of the PE deployed.

Estimating Cfree using the FO method required set stipulations due to instances where the fret was greater than one resulting in a negative ke. In these cases, a ke correlating to a fret of 0.99 was used. The 80/20 rule was not used because these correction restrictions were not applied for the other methods. A minimum feq of 0.01 equivalent was used while no upper bound for feq was implemented. Applying the FO method decreased Cfree RSD values for several of the congeners measured through the pentachlorinated homolog (Figures 2, S4 and S5). Of these, less than a quarter of the congeners had RSDs greater than 20% at NBH2. Hexachlorinated congeners had RSD values greater than 100%. These large RSDs were due to the 76 ¼m measurements. PRC, 13C-PCB138, had a measured fret greater than one, so a feq equivalent of 0.01 was used (NLS and DIF methods predicted feqs 0.06), resulting in an estimated Cfree that was about five times larger than the other thicknesses measured. This example illustrates the risk of over-correction using the FO method when CPEt and CPE0 do not show a significant difference.

The NLS method allowed for an individual feq value to be estimated for each target PCB. Cfree values calculated using this model resulted in a decrease in measured RSDs throughout the suite of PCBs investigated (Figures 2, S4, S5), relative to other models or no correction at all. Only congener CB126 had a RSD value above 20%. Calculated RSD values ranged from 10 to 23% with a mean of 15% for the suite of PCBs. The 76.2 ¼m PE generally demonstrated the highest Cfree estimates followed by the 25.4 ¼m sampler. The Cfrees for the 12.7 ¼m and 50.8 ¼m samplers often agreed within one standard deviation of each other (Tables S3a,b).

WBL measurements for the passive samplers ranged from 66 to 268 ¼m (Table S7) and were in agreement with those reported or estimated (Fernandez et al., 2012, Lohmann, 2012, Apell et al., 2016). Only fret values between 0.15 and 0.85 were used in the calculation of the WBL for each sampler. Using fraction retained values where an equilibrium may have been achieved (e.g., fret < 0.15) resulted in very large WBL thickness estimates. Anomalous WBL values were also observed (both large and very small ranging from 0.6 ¼m to 660 ¼m) when fret values were larger than 0.85. Due to analytical uncertainties associated with these endmember frets, they were not used, which reduced the estimated WBL RSD by about 30% and, at least, two values were used for each WBL estimation. The triplicate feq variability is most notable with the 25.4 ¼m polymer thicknesses measurements. Closer inspection of this dataset from NBH2 revealed the fret for 13CPCB 28 was acceptable for only two of the replicates. The average WBL for the two replicates that used 13C-PCB 28 data are nearly identical and almost double that of the 13C-PCB 28 replicate that was not included (fret = 0.11). However, the Cfree results from the replicate that did not use 13CPCB 28 data were closer to the results generated from the other thicknesses. This range in WBL values likely contributed to feq variability and ultimately increased Cfree RSDs for this method.

Variations observed in the DIF-estimated feq resulted in larger Cfree RSDs for the four thicknesses tested (Figures 2, S4, S5) than the NLS method. An RSD of greater than 20% was observed for many congeners with a log KOW ≥ 6.2, though the RSDs did decrease compared to the non-corrected RSD values. Only two congeners, CB126 (35%) and CB189 (30%) had RSDs above 30%, whereas the majority of Cfree measurements demonstrated relatively favorable agreement it was not as good as measurements where equilibrium had been attained and over 70% of the PCBs measured had a RSD of 27% or less. Similar to the NLS method, the 25.4 and 76.2¼m PE thicknesses estimated the highest Cfree concentrations and in the majority of cases were not significantly different. (Tables S3a and b).

When statistically comparing the Cfree by thickness, congeners CB8, CB18 and CB28 demonstrated no significant differences between thicknesses for any of the equilibrium correction methods (Table S9). The occurrence of significantly different Cfree estimates between the thicknesses increased starting with the tetrachlorinated PCBs and continued for higher chlorinated congeners. Cfree values determined for each thickness using the DIF method were not statistically different 70% of the time. These results may be due, in part, to the relatively large standard deviations between replicates for each thickness resulting in the different thicknesses being difficult to distinguish statistically. Results of similar statistical analyses for NBH4 and NBH5 are provided in Tables S11 and S13. Overall, this data suggests in the cases of the DIF and NLS method using multiple PRCs to deduce WBL or RS, resulted in similar Cfree results between data sets and reduced variation in the results.

Inter-method comparison of congener Cfree

Estimates of Cfree were also compared by the three equilibrium correction methods used to generate them. Methods were compared by plotting the Cfree ratios of the different combinations of correction methods with the target compound’s KPW (Figure 3). Plotting the data this way illustrates how well the different equilibrium methods agree for compounds as correction factors increased (1/feq increased with log KPW) . Cfree values below the quantitation limits were not included, and the data was separated by sampler thickness. Naturally, compounds that reached or nearly reached equilibrium (e.g., CB8, CB18, CB28, CB 44, and CB52) during the deployment showed near perfect agreement for all equilibrium models as little to no corrections were made.

Figure 3.

Figure 3.

Inter-method variability of PRC-equilibrium model estimates of Cfree, comparing three methods: diffusion based (DIF), sampling rate fit to a nonlinear least square relationship (NLS), and first order kinetic fit (FO) as a function of the polymer partitioning coefficient (KPW).

The DIF and NLS methods showed the greatest and most consistent agreement of the three approaches tested (Figure 3a). In fact, the Cfree values calculated agree within a factor of two for all estimated Cfree. On average, the Cfree as estimated by the DIF method were 1.1 times larger than those estimated by NLS. The largest differences in Cfree were those estimated on the 25.4 ¼m thickness where, on average, the Cfree ratio between the two methods was 1.25. Cfree ratios ranged from 0.93 to 1.19 for the other thicknesses. This 25.4 ¼m thickness discrepancy is likely due to the unexpectedly high and variable fret for PRC 13C-PCB28 in this sampler, which resulted in variable and relatively high Cfree estimates of the targeted congeners, as discussed in the Intra-Method Comparison of Congener Cfree section.

The consistent agreement between these two methods resulted in similar comparisons with the FO method (Figures 3b and c). Overall, the FO-predicted Cfree values were, on average, higher than the other two methods. In these comparisons variation increased as the KPW increased, and as the feq decreased, and correction factor increased. Differences larger than a factor of 2, began with PCBs with a log KPW greater than six. These values correlate to estimated feq as high as 0.5. Unlike the DIF and NLS comparison, there was no one thickness of PE that illustrated any observable patterns in variance in the FO comparisons. There were however some observable patterns. These varying behaviors are likely due to some congeners having different physicochemical characteristics than the PRC used to predict their ke in the FO method. This trend is most observable for all PE thicknesses with log KPW between 6.80 and 8.25. In both @AB CD⁄ and 364 3EC CD⁄ the Cfree ratios are different by a factor of about ten. As KPW increases to about 7.5 (CB180 has KPW of 7.42 L/kg) the ratio approaches one. As KPW exceeds 7.5, the ratios continue to increase indicating the NLS and DIF methods are estimating a larger Cfree than the FO method. Much of the variation in these results is a direct result of the large fret of the PRC, both 13C-CB138 and 13C-CB180, had a different default fret been defined the patterns may have been different. Many of the large and small ratios observed here incorporated the default feq of 0.01 used in the FO model. This result gives further evidence of the risks associated with using the FO model beyond the 80:20 rule in that it can both over- and under-predict Cfree estimates.

ANOVA analysis showed that the estimates of Cfree demonstrated no statistical differences between results by correction method and Cfree estimates when no model was used (Equation 1) for all thicknesses for di- and tri-chlorinated congeners (Table S10). At NBH2, the NLS and DIF Cfree estimates were indistinguishable, in 96% of the comparisons tested, and results from the FO method were not statistically different from the NLS and DIF methods in 40% and 34% of the measurements, respectively. Results of similar statistical analyses for NBH4 and NBH5 are reported in Tables S12 and S14, respectively.

Comparison of total PCB Cfree

Table 1 reports total PCB Cfree values calculated using each of the methods as the sum of up to 27 individual congener Cfrees by thickness at NBH2, NBH4 and NBH5. For NBH2, mean Cfrees ranged across models and thicknesses with values of 387 to 505 ng/L. For each thickness, the largest total PCB values were generated by the FO method followed by the DIF method, and then the NLS method, where the NLS and DIF methods were always within 5% of each other, and no one thickness of PE consistently demonstrated the largest nor the smallest mean total PCB measurement. No significant differences in total PCB Cfree values between methods were determined at NBH2 (Table 1). The same trends were observed at NBH4 although the total PCB Cfree range was an order of magnitude lower: 32.2 to 55.6 ng/L. At NBH5, the total PCB Cfree were two orders of magnitude below the NBH4 levels with a range of 0.13 to 0.27 ng/L. At NBH4 and NBH5, statistical analyses of total PCB Cfrees by thickness detected several significant differences in total PCB Cfree estimates (Table 1). However, at all three stations, statistically significant differences between Cfrees based on the NLS and DIF methods were never detected. In about 55% of the total PCB Cfree estimates a significant difference existed between the FO and NLS or DIF modeled results. These results also illustrate how some congeners can statistically dominate the total PCB estimates. For instance, at NBH2 several of the lighter, faster equilibrating congeners contribute to a large fraction of the total PCB, and as a result, the PRC-modeled results are agreeable with the results to which no PRC corrections were made. However, when these congeners make up a smaller percentage of the mixture significant differences in the results occur. These statistically different results again highlight the importance of using PRCs to estimate accurate Cfree. A similar statistical analysis as discussed above but focusing on differences between passive sampler thicknesses by model is reported in Table S15.

Table 1.

Mean total PCB Cfree values (ng/L) estimated as the sum of up to 27 individual PCB congeners using each of the correction methods by thickness at NBH2, NBH4 and NBH5

Station Thickness (¼m) Model ANOVA p-value
No Model FO NLS DIF
NBH2 12.5 385 ± 67.6 396 ± 65.4 395 ± 68.1 396 ± 68.1 0.9963
25.4 427 ± 65.0 505 ± 91.0 444 ± 68.0 463 ± 83.9 0.6533
50.8 397 ± 56.2 390 ± 58.4 408 ± 61.2 388 ± 58.5 0.9735
76.2 305 ± 23.5 456 ± 50.2 387 ± 40.3 401 ± 39.4 0.0772
NBH4 12.5 36.0 ± 0.83B 40.4 ± 2.01A 38.1 ± 1.28A,B 38.9 ± 0.87A 0.0215
25.4 39.9 ± 1.05C 55.6 ± 1.75A 43.7 ± 1.01B,C 47.8 ± 4.45B 0.0003
50.8 27.1 ± 1.60C 43.8 ± 1.10A 32.2 ± 1.26B 34.3 ± 1.16B <0.0001
76.2 26.8 ± 1.46C 47.1 ± 1.26A 39.7 ± 1.13B 39.3 ± 0.64B <0.0001
NBH5 12.5 0.18 ± 0.01C 0.25 ± 0.01B 0.26 ± 0.01A,B 0.27 ± 0.02A <0.0001
25.4 0.09 ± 0.03A 0.13 ± 0.04A 0.17 ± 0.06A 0.18 ± 0.06A 0.1853
50.8 0.07 ± 0.01B 0.13 ± 0.01A 0.14 ± 0.01A 0.14 ± 0.01A <0.0001
76.2 0.06 ± 0.01C 0.13 ± 0.01B 0.16 ± 0.01A 0.17 ± 0.01A <0.0001

This comparison of mass transfer-based correction methods for passive samplers found that using multiple PRCs to estimate sampling rates or fractional equilibria for target PCBs (i.e., NLS and DIF methods, respectively) resulted in more consistent and precise estimates of Cfree than using results from one PRC (i.e., the FO method). In addition, the NLS and DIF methods generated Cfree values that were statistically similar when expressed as total PCB or on a single congener basis. These findings suggest that contaminated site managers and scientists can apply either mass transfer method to use PRC results to estimate Cfree for their target contaminants. Assistance running DIF method is available on-line (https://www.serdp-estcp.org/Tools-andTraining/Tools/PRC-Correction-Calculator and https://www.epa.gov/superfund/superfundcontaminated-sediments-guidance-documents-fact-sheets-and-policies). Assuming equilibrium is discouraged as it will likely under-estimate Cfree for contaminants that are slow to equilibrate or under in situ conditions with slow kinetics (i.e., quiescent environmental conditions). For all methods, it is important to consider using PRCs that span the physicochemical characteristics (e.g., log KOW) of the target HOCs and that will result in useful fret, guidance on estimating PRC results given environmental conditions can be found in the literature (Apell et al., 2016).

Supplementary Material

Supplement1

ACKNOWLEGMENT

The authors thank the AED reviewers for their comments: Adeyemi Adeleye, David Katz, and Jonathan Serbst. This research was performed while A.S. Joyce held an NRC Research Associateship award at U.S. Environmental Protection Agency, ORD/NHEERL Atlantic Ecology Division. The authors would like to acknowledge Donald Cobb and Barbara Bergen for their assistance in deploying and retrieving these passive samplers.

Footnotes

DATA AVAILABILITY STATEMENT

Readers can access much of the data in the Supporting Information. Additional data and supporting calculation tools are available upon request.

This is NHEERL Contribution ORD-021785.

Mention of trade names or commercial products does not constitute endorsement or recommendation for use. The views expressed in this article are those of the authors and do not necessarily represent the views or policies of the U.S. Environmental Protection Agency.

REFERENCE LIST

  1. APELL JN & GSCHWEND PM 2014. Validating the Use of Performance Reference Compounds in Passive Samplers to Assess Porewater Concentrations in Sediment Beds. Environmental Science & Technology, 48, 10301–10307. [DOI] [PubMed] [Google Scholar]
  2. APELL JN & GSCHWEND PM 2016. In situ passive sampling of sediments in the Lower Duwamish Waterway Superfund site: Replicability, comparison with ex situ measurements, and use of data. Environmental Pollution, 218, 95–101. [DOI] [PubMed] [Google Scholar]
  3. APELL JN, TCACIUC AP & GSCHWEND PM 2016. Understanding the rates of nonpolar organic chemical accumulation into passive samplers deployed in the environment: Guidance for passive sampler deployments. Integrated Environmental Assessment and Management, 12, 486–492. [DOI] [PubMed] [Google Scholar]
  4. BELLES A, ALARY C & MAMINDY-PAJANY Y 2016. Thickness and material selection of polymeric passive samplers for polycyclic aromatic hydrocarbons in water: Which more strongly affects sampler properties? Environmental Toxicology and Chemistry, 35, 1708–1717. [DOI] [PubMed] [Google Scholar]
  5. BOOIJ K, ROBINSON CD, BURGESS RM, MAYER P, ROBERTS CA, AHRENS L, ALLAN IJ, BRANT J, JONES L, KRAUS UR, LARSEN MM, LEPOM P, PETERSEN J, PRÖFROCK D, ROOSE P, SCHÄFER S, SMEDES F, TIXIER C, VORKAMP K & WHITEHOUSE P 2016. Passive Sampling in Regulatory Chemical Monitoring of Nonpolar Organic Compounds in the Aquatic Environment. Environmental Science & Technology, 50, 3–17. [DOI] [PubMed] [Google Scholar]
  6. BOOIJ K & SMEDES F 2010. An Improved Method for Estimating in Situ Sampling Rates of Nonpolar Passive Samplers. Environmental Science & Technology, 44, 6789–6794. [DOI] [PubMed] [Google Scholar]
  7. BURGESS RM, LOHMANN R, SCHUBAUER-BERIGAN JP, REITSMA P, PERRON MM, LEFKOVITZ L & CANTWELL MG 2015. Application of passive sampling for measuring dissolved concentrations of organic contaminants in the water column at three marine superfund sites. Environmental Toxicology and Chemistry, 34, 1720–1733. [DOI] [PubMed] [Google Scholar]
  8. DI TORO DM, ZARBA CS, HANSEN DJ, BERRY WJ, SWARTZ RC, COWAN CE, PAVLOU SP, ALLEN HE, THOMAS NA & PAQUIN PR 1991. Technical basis for establishing sediment quality criteria for nonionic organic chemicals using equilibrium partitioning. Environmental Toxicology and Chemistry, 10, 1541–1583. [Google Scholar]
  9. FERNANDEZ LA, HARVEY CF & GSCHWEND PM 2009. Using Performance Reference Compounds in Polyethylene Passive Samplers to Deduce Sediment Porewater Concentrations for Numerous Target Chemicals. Environmental Science & Technology, 43, 8888–8894. [DOI] [PubMed] [Google Scholar]
  10. FERNANDEZ LA, LAO W, MARUYA KA & BURGESS RM 2014. Calculating the Diffusive Flux of Persistent Organic Pollutants between Sediments and the Water Column on the Palos Verdes Shelf Superfund Site Using Polymeric Passive Samplers. Environmental Science & Technology, 48, 3925–3934. [DOI] [PubMed] [Google Scholar]
  11. FERNANDEZ LA, LAO W, MARUYA KA, WHITE C & BURGESS RM 2012. Passive Sampling to Measure Baseline Dissolved Persistent Organic Pollutant Concentrations in the Water Column of the Palos Verdes Shelf Superfund Site. Environmental Science & Technology, 46, 11937–11947. [DOI] [PubMed] [Google Scholar]
  12. FRIEDMAN CL, CANTWELL MG & LOHMANN R 2012. Passive sampling provides evidence for Newark Bay as a source of polychlorinated dibenzo-p-dioxins and furans to the New York/New Jersey, USA, atmosphere. Environmental Toxicology and Chemistry, 31, 253–261. [DOI] [PubMed] [Google Scholar]
  13. GHOSH U, KANE DRISCOLL S, BURGESS RM, JONKER MTO, REIBLE D, GOBAS F, CHOI Y, APITZ SE, MARUYA KA, GALA WR, MORTIMER M & BEEGAN C 2014. Passive sampling methods for contaminated sediments: Practical guidance for selection, calibration, and implementation. Integrated Environmental Assessment and Management, 10, 210–223. [DOI] [PMC free article] [PubMed] [Google Scholar]
  14. GREENBERG MS, CHAPMAN PM, ALLAN IJ, ANDERSON KA, APITZ SE, BEEGAN C, BRIDGES TS, BROWN SS, CARGILL JG, MCCULLOCH MC, MENZIE CA, SHINE JP & PARKERTON TF 2014. Passive sampling methods for contaminated sediments: Risk assessment and management. Integrated Environmental Assessment and Management, 10, 224–236. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. GSCHWEND PM, MACFARLANE JK, REIBLE DD, LU X, HAWTHORNE SB, NAKLES DV & THOMPSON T 2011. Comparison of polymeric samplers for accurately assessing PCBs in pore waters. Environmental Toxicology and Chemistry, 30, 1288–1296. [DOI] [PubMed] [Google Scholar]
  16. HAWKER DW & CONNELL DW 1988. Octanol-water partition coefficients of polychlorinated biphenyl congeners. Environmental Science & Technology, 22, 382–387. [DOI] [PubMed] [Google Scholar]
  17. HUCKINS JN, PETTY JD, LEBO JA, ALMEIDA FV, BOOIJ K, ALVAREZ DA, CRANOR WL, CLARK RC & MOGENSEN BB 2002. Development of the Permeability/Performance Reference Compound Approach for In Situ Calibration of Semipermeable Membrane Devices. Environmental Science & Technology, 36, 85–91. [DOI] [PubMed] [Google Scholar]
  18. JOYCE AS, PIROGOVSKY MS, ADAMS RG, LAO W, TSUKADA D, CASH CL, HAW JF & MARUYA KA 2015. Using performance reference compound-corrected polyethylene passive samplers and caged bivalves to measure hydrophobic contaminants of concern in urban coastal seawaters. Chemosphere, 127, 10–17. [DOI] [PubMed] [Google Scholar]
  19. LOHMANN R 2012. Critical Review of Low-Density Polyethylene’s Partitioning and Diffusion Coefficients for Trace Organic Contaminants and Implications for Its Use As a Passive Sampler. Environmental Science & Technology, 46, 606–618. [DOI] [PubMed] [Google Scholar]
  20. LYDY MJ, LANDRUM PF, OEN AMP, ALLINSON M, SMEDES F, HARWOOD AD, LI H, MARUYA KA & LIU J 2014. Passive sampling methods for contaminated sediments: State of the science for organic contaminants. Integrated Environmental Assessment and Management, 10, 167–178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. MARUYA KA, ZENG EY, TSUKADA D & BAY SM 2009. A passive sampler based on solid-phase microextraction for quantifying hydrophobic organic contaminants in sediment pore water. Environmental Toxicology and Chemistry, 28, 733–740. [DOI] [PubMed] [Google Scholar]
  22. MAYER P, PARKERTON TF, ADAMS RG, CARGILL JG, GAN J, GOUIN T, GSCHWEND PM, HAWTHORNE SB, HELM P, WITT G, YOU J & ESCHER BI 2014. Passive sampling methods for contaminated sediments: Scientific rationale supporting use of freely dissolved concentrations. Integrated Environmental Assessment and Management, 10, 197–209. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. MAYER P, TOLLS J, HERMENS JLM & MACKAY D 2003. Peer Reviewed: Equilibrium Sampling Devices. Environmental Science & Technology, 37, 184A–191A. [DOI] [PubMed] [Google Scholar]
  24. PERRON MM, BURGESS RM, SUUBERG EM, CANTWELL MG & PENNELL KG 2013a. Performance of passive samplers for monitoring estuarine water column concentrations: 1. Contaminants of concern. Environmental Toxicology and Chemistry, 32, 2182–2189. [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. PERRON MM, BURGESS RM, SUUBERG EM, CANTWELL MG & PENNELL KG 2013b. Performance of passive samplers for monitoring estuarine water column concentrations: 2. Emerging contaminants. Environmental Toxicology and Chemistry, 32, 2190–2196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. RUSINA TP, SMEDES F, KOBLIZKOVA M & KLANOVA J 2010. Calibration of Silicone Rubber Passive Samplers: Experimental and Modeled Relations between Sampling Rate and Compound Properties. Environmental Science & Technology, 44, 362–367. [DOI] [PubMed] [Google Scholar]
  27. SMEDES F 2007. Monitoring of chlorinated biphenyls and polycyclic aromatic hydrocarbons by passive sampling in concert with deployed mussels In: GREENWOOD R, MILLS GA & VRANA B (eds.) Passive Sampling Techniques in Environmental Monitoring. Amsterdan: Elsevier. [Google Scholar]
  28. THOMPSON JM, HSIEH C-H & LUTHY RG 2015. Modeling Uptake of Hydrophobic Organic Contaminants into Polyethylene Passive Samplers. Environmental Science & Technology, 49, 2270–2277. [DOI] [PubMed] [Google Scholar]
  29. TOMASZEWSKI JE & LUTHY RG 2008. Field Deployment of Polyethylene Devices to Measure PCB Concentrations in Pore Water of Contaminated Sediment. Environmental Science & Technology, 42, 6086–6091. [DOI] [PubMed] [Google Scholar]
  30. VRANA B & SCHÜÜRMANN G 2002. Calibrating the Uptake Kinetics of Semipermeable Membrane Devices in Water:  Impact of Hydrodynamics. Environmental Science & Technology, 36, 290–296. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplement1

RESOURCES