Stem Fixation in Revision Total Knee Arthroplasty: Indications, Stem Dimensions, and Fixation Methods

Se Gu Kang 1, Cheol Hee Park 2,, Sang Jun Song 1
PMCID: PMC6122937  PMID: 30157588

Abstract

Although stems improve initial mechanical stability in revision total knee arthroplasty (TKA), ideal indications, proper lengths and diameters, and appropriate fixation methods remain controversial. The topics of the present article include the indications, selection of lengths and diameters, and fixation methods of stems in revision TKA. The use of a stem in revision TKA can protect the juxta-articular bone. A stem cannot be a substitute for optimal component fixation; it plays an adjunctive role in transferring the loads from the compromised metaphysis to the stronger diaphysis. Proper bone surface preparation and appropriate use of the stem based on a great store of knowledge are required to support the stemmed components effectively in revision TKA. The balance between overshielding and overloading the juxta-articular bone would provide excellent structural protection. The stem length and diameter should be tailored according to patients’ anatomical characteristics and determined fixation strategy. There are two traditional methods of stem fixation including the total cementation technique and the hybrid technique with a cementless press-fit stem. Selection of a cementation technique should be based on thorough consideration of advantages and disadvantages of each technique.

Keywords: Knee, Arthroplasty, Revision, Stem

Introduction

One of the primary goals of total knee arthroplasty (TKA) is to recreate a stable joint with accurate position and orientation of prostheses, mimicking the normal knee kinematics. However, it would be a more difficult task in the revision setting due to combined severe bone defects and soft tissue insufficiency. Stems are required in most revision TKAs to help transfer loads from the compromised articular and metaphyseal bone to the remaining tibial cortical rim and to widely distribute the increased stress of a constrained articulation.

They improve the mechanical stability at the cost of stress shielding. Mechanical stability can be improved by resistance to shear, reduced lift-off, and decreased micromotion1). However, long stems may have disadvantages including end-of-stem pain and stress shielding along their length with associated reduction in bone density, a theoretical risk of subsidence, loosening, and periprosthetic fracture1,2). It is known that the axial load can be reduced by 23% to 39% when a stem length reaches 70 mm3). With a stem up to 150 mm in length, marked stress shielding of the proximal tibial cortex and doubling of the strain are noted at the stem tip3).

Although the need for stem to improve initial mechanical stability and ultimate component survival is well accepted, the ideal indications, proper lengths and diameters, and fixation methods remain controversial. No evidence-based guidelines are available to help determine when to use a stem, which length and diameter are ideal, and whether or not to use cement fixation4). Such decisions should be based on a great store of knowledge rather than on presumption. In this review article, we will discuss the indications, selection of stem lengths and diameters, and fixation methods in revision TKA.

1. Indications

Stems should be used in revision TKA when the remaining bone stock is insufficient to support the prosthesis. Bone grafting for large volume defects may require the use of a stem to protect the graft from excessive load. The knee joint bears loads that are several times the body weight. If a stem fails to transfer the load, then the remaining cancellous bone will experience load beyond its ultimate strength, which will lead to a loss in component fixation in the initial phase5). Modular metal augments (blocks or wedges), cones, or sleeves can be used when the cortical rim or either the distal femur or proximal tibia is breached. The selection of metal augments typically mandates the use of a stem. Brand et al.6) reported the results of 22 TKAs with modular metal wedges and small tibial cemented stems: no tibial tray was considered loose at an average of 37 months of follow-up.

Wolff ’s Law states that bone remodels in response to applied stress and strain in the chronic phase. The strain for normal remodeling of the proximal tibia should be within a physiological range of 50 to 1,500 μɛ1). Below 50 μɛ, stress shielding is likely to occur and bone resorption takes place1). On the contrary, the risk of damage within the cancellous bone with microfractures increase above 1,500 μɛ, and the risk of collapse with pathological overloading increases above 3,000 μɛ1). The stiffer the material, the greater the stress shielding effect. Also, the greater the difference in Young’s modulus of elasticity between the material and surrounding bone, the greater the stress shielding. Greater shielding is therefore expected in cobalt-chrome components than in titanium components. Accordingly, factors affecting the selection of a stem include not only bone stock and the type of bone defect but also component material. However, recent studies with finite element analysis have shown no difference in stress shielding between long stems made of titanium and cobalt-chrome1). The development of metaphyseal cones or sleeves and porous trabecular metals have provided an alternative to stems although biomechanical studies are warranted7). More research on various stem lengths, surface finishes, and different stem fixation techniques are needed under different circumstances to optimize the strain environment.

The increased constraint of polyethylene insert in patients with soft tissue insufficiency can result in increased stress between the components and cement or between cement and host bone. Therefore, more stable fixation of components would be required in patients with severe soft tissue insufficiency and constrained prostheses8). Anderson et al.9) reported good results of 55 TKAs using a constrained condylar prosthesis without stem extension: there were one case of femoral loosening and one case of tibial fracture at 44.5 months of follow-up. Sabatini et al.10) reported favorable clinical and radiographic results with use of the second generation semi-constrained prosthesis with stem extension. Although the use of the stem remains controversial in constrained TKA, the stem should be used routinely if there is inadequate bony surface.

The stem can also be used for the treatment of periprosthetic fractures after TKA. Fractures around a loose prosthesis are classified as Rorabeck type III11); the exchange for a prosthesis with a longer stem for diaphyseal fixation is the treatment of choice in this type of periprosthetic fractures. As in the case of fractures around the femoral component, the stability of the prosthesis is important in the treatment of periprosthetic tibial fractures. In fractures with subtype B according to Felix classification, a loose prosthesis has to be exchanged for a revision prosthesis with diaphyseal stem fixation12).

Stems can be used depending on the extent of offset. Hicks et al.13) noted significant variability in the location of the tibial canal relative to the tibial plateau. They found that the intramedullary canal center was usually anterior and medial to the tibial plateau. Tang et al.14) described that the axis of the tibial shaft is located anterolateral to the center of the tibial plateau in Asian population. These studies highlighted the feasibility of offset stems of the tibial component in revision TKA.

2. Dimensions (Length and Diameter)

Micromotion of knee prostheses can be significantly reduced by increasing the stem length, diameter, and canal filling ratio (CFR) in revision TKA using press-fit stems1). Biomechanical studies also demonstrated significant improvement in joint stability with use of a press-fit stem with an increased length and diameter15,16).

When a short stem (25 to 30 mm in length) is used, it should be cemented for stability. However, the use of long modular stem extensions, which are canal filling and diaphyseal engaging, can provide the stability in a press-fit method without the use of cement. Wood et al.17) reported the results of 135 revision TKAs using press-fit diaphyseal fixation and cemented metaphyseal fixation: the 12-year survival analysis showed 98% probability of survival free of revision for aseptic loosening.

However, the literature review failed to identify the “ideal” length18). It is unlikely that an ideal length can be determined because of the heterogeneity of patients’ anatomical characteristics and revision circumstances. Patient-specific tailoring of reconstruction strategy will be needed to match the patient’s anatomy, bone loss, and constraint requirements.

Although some authors have recommended at least 4 cm of diaphyseal engagement for press-fit stems19), longer stems that approach the confined diaphyseal isthmus may need to be narrower20). The length of a diaphyseal engaging stem must be determined concurrently with the stem diameter20). No clinical evidence has effectively reconciled this relative influence of stem length and diameter. The ideal composite of stem length and diameter may be better expressed as the CFR, which is defined as the width of the stem divided by the width of the intramedullary canal. The canal filling stems permit considerable diaphyseal cortical contact, which may enhance component stability and improve mechanical alignment20,21). Fleischman et al.22) reported that the risk for mechanical failure was reduced by 41.2% for every 10% increase in CFR. Parsley et al.23) recommended a CFR of >0.85 to achieve stable intramedullary fit.

A long stem can force articular components into suboptimal position (Table 1). Gobba et al.24) demonstrated that a 120 mm tibial stem will force the tibial component into an excessively valgus position, and a 200 mm tibial stem will often force the tibial component into a posteromedial position above the tibial surface. Therefore, proper surgical strategies are required to prepare a cut surface that facilitates the entry of the long stem. We will describe such strategies in the next section.

Table 1.

Length and Diameter of Stem according to Providers in Korea

Variable Depuy PFC Sigma Zimmer NexGen LCCK Biomet Vanguard 360 Stryker Triathlon TS Smith & Nephew Legion revision
Cemented (Tibia) 30 or 60×13 30×15 No 50×9, 12, 15 20×10, 12, 14
(Femur) 90 or 130×13 100×10–16 100×9, 12, 15 60×10, 12, 14
155×10, 12, 14, 16 150×9,12,15 100×10, 12, 14
Cementless (Tibia) 75×10, 12, 14, 16, 18 No 40×10–16 No 60×9–16, 18
115×10, 12, 14, 16, 18 80×10–16 100×9–16, 18
(Femur) 95×10, 12, 14, 16, 18 120×10–16
135×10, 12, 14, 16, 18

Values are presented as length×diameter (mm).

A preoperative estimate along with the intraoperative assessment of stem length and diameter is required to determine the extent of press-fit. The preoperatively templated stem diameter has to be confirmed intraoperatively to the point at which cortical chatter is felt or heard with a reamer at the preset stem length. A greater degree of press-fit is required when there are severe bone defect and soft tissue insufficiency3). Reaming to a greater diameter and removal of more endosteal bone induce more robust support for the stemmed component; it facilitates the insertion of a stem of larger diameter and provides an enlarged stem-to-bone contact associated with positive effects on stability25). The press-fit stem length and diameter have to be tailored to optimize the canal filling with the consideration of patients’ anatomical characteristics and degree of bone defect and prosthesis constraint. Manufacturers provide various press-fit or cemented stems with different lengths and diameters (Table 1), which need to be prepared to match various patients’ anatomical characteristics (Fig. 1).

Fig. 1.

Fig. 1

Preoperative planning on the diameter and length of a stem. (A) Tibial component loosening required stable fixation of the tibial component using a stem in revision total knee arthroplasty, but the narrow intramedullary canal made the use of a stem difficult (double-headed arrow). (B) The 75×10 mm press-fit stem was altered to be 45×10 mm to prevent an insertion failure. (C) Fortunately, the ordinary 75×10 mm stem could be inserted with modification of the stem insertion point.

3. Fixation Methods

Two traditional methods of stem fixation have been used: total cementation technique and hybrid technique with a cementless press-fit stem. Both techniques have their own advantages and disadvantages (Table 2)20,26). Based on the available literature, no superiority of any type of stem fixation has been found26). Due to the inhomogeneity of failure mechanism, bone stock, implant design, and comorbidities, a strictly randomized controlled trial (RCT) for revision TKA is nearly impossible to compare the outcomes according to the cementation technique26). A recent RCT27) performed radiostereometric analysis with (1) constructs of the same length, (2) the same Anderson Orthopaedic Research Institute type of bone defect, (3) implants with the same level of constraint, and (4) exclusion of re-revision TKAs. No difference was reported in median migration or the number of migrating components between total and hybrid cementation techniques in revision TKAs. However, another study28) showed that cemented stems were significantly more likely to have radiographic loosening compared to uncemented stems (4.9% vs. 1.6%, p=0.02). It suggested that the reamed diaphysis was a poor surface for cement interdigitation, leading to higher rates of radiolucency (32% vs. 17%, p=0.006).

Table 2.

Advantages and Disadvantages of Total Cementation versus Hybrid Cementation Techniques in Revision Total Knee Arthroplasty

Variable Total cementation with fully cemented stem Hybrid cementation with press fit stem
Advantage Easy to implant Potential for bone stock preservation
Better initial stability Less stress shielding
Potential benefit of intramedullary elution of antibiotics Optimizing mechanical alignment by diaphyseal referencing
More variability in component positioning and sizing Easy to remove
Useful especially in diaphyseal bowing Applicable in patients with previous DFO or HTO
Disadvantage Difficult to remove Potential for end-of-stem pain
Decreased references for proper alignment Potential forcing of component into suboptimal position without offset stem
Risk of periprosthetic fractures with canal preparation

DFO: distal femoral osteotomy, HTO: high tibial osteotomy.

Some surgeons favor the use of a specific technique, but others select them on an ad hoc basis. Most surgeons prefer the hybrid technique with a press-fit stem in ordinary revision TKA (Table 3)20). However, the total cementation technique is occasionally required when the metaphyseal bone is severely destructed and the inner cortex is too poor in patients with severe osteoporosis. We will describe the hybrid cementation technique first and then introduce the use of cement plugs in the total cementation technique.

Table 3.

Comparison of Indication and Concept between Total Cementation and Hybrid Cementation Techniques in Revision Total Knee Arthroplasty

Parameter Total cementation with fully cemented stem Hybrid cementation with press fit stem
Indication Huge bone defect Ordinary revision setting
Poor bone quality
Concept Metaphyseal engaging Diaphyseal engaging
Length Short Long
Characteristic Smooth, small range of diameter option Fluted, wide range of diameter option
Material Co-Cr Titanium

Preoperative radiographic planning includes assessment of the intramedullary canal to ensure that stem insertion is possible without ipsilateral hardware and deformity or that the intramedullary canal axis conforms to the mechanical axis orientation. Eccentric joint surfaces may require the use of offset stems29). The stem length and diameter are estimated to obtain adequate endosteal press-fit. Preparation of at least two possible stem lengths with correspondingly different stem diameters would allow optimal application and prevent cortical impingement.

Intraoperatively, technical efforts should be focused on exposing the good inner cortex to achieve a sufficient press-fit and matching the cut surface and stem orientation, when using an extramedullary alignment guide. Alternatively, preparation of the cut surface can be easily performed using a broad intramedullary guide rod or a reamer. The initial reaming removes ectatic and sclerotic inner cortex which can lever the reamers away from the true intramedullary canal axis3). Then, the intramedullary canal is sequentially reamed to the appropriate length and diameter to accept the press-fit stems. Traditionally, reaming is performed approximately 1 cm past the tip of the stem to ensure that there is no tip impingement with the possibility of cortical erosion. It is continued in millimeter increments until minimal endosteal cortical contact is felt at the preoperatively determined stem length. For structural-bulk allografts, reaming should be continued at least 1mm wider beyond the point at which cortical chatter is felt and heard to improve initial stability.

The position of components is dictated by the position of the press-fit stem. If the canal is eccentric, an offset stem should be considered. In particular, the femoral component position in the sagittal plane with a straight stem is also dictated by the stem and therefore determines the flexion gap. Because posterior shifting of the femoral component’s position or addition of posterior femoral augments to alter the flexion gap cannot be accomplished with a straight stem, a posterior offset stem or a femoral component with a different housing junction point can be a good viable option3).

When components are easily seated without a stem but failure of seating occurs with the stem, the surgeon should consider the following causes: (1) improper stem size; (2) occurrence of endosteal impingement; and (3) mismatch between the prepared bone surface and the intramedullary position of the stem. The use of an offset stem and adjustment of stem position with intramedullary widening should be considered as a solution.

After sufficient reaming and confirmation of the match between the canal and prepared bone surface, cement is mixed and packed along the cut surfaces with component undersurface and metaphyseal keel, but not around the stem tip.

For the total cementation technique with a fully-cemented stem, impingement of stem or mismatch between the prepared bone surface and stem orientation is uncommon (Table 2). However, optimization of mechanical alignment by diaphyseal referencing cannot be expected. Care should be taken to achieve accurate positioning of components and alignment during the bone preparation and cementation. With a cement plug placed above the expected level of the stem tip, cement is pressure injected into the intramedullary canal. Additional cement is mixed and packed along the prepared bone surfaces with undersurface of the articular components, surfaces of the metaphyseal keel, and around the stem.

Conclusions

Use of a stem in revision TKA can protect the juxta-articular bone and transfer the load to the stronger diaphyseal bone. The balance between overshielding and overloading of the juxta-articular bone is crucial for excellent structural protection. Proper bone surface preparation and appropriate use of stems based on a great store of knowledge are required to support the stemmed components effectively in revision TKA. The stem length and diameter should be tailored according to the patients’ anatomical characteristics and selected fixation strategy. Two traditional methods of stem fixation (total cementation technique and hybrid technique with cementless press-fit stem) have their own advantages and disadvantages, which should be carefully taken into consideration in the selection of a cementation technique.

Footnotes

Conflict of Interest

No potential conflict of interest relevant to this article was reported.

References

  • 1.Scott CE, Biant LC. The role of the design of tibial components and stems in knee replacement. J Bone Joint Surg Br. 2012;94:1009–15. doi: 10.1302/0301-620X.94B8.28289. [DOI] [PubMed] [Google Scholar]
  • 2.Murray PB, Rand JA, Hanssen AD. Cemented long-stem revision total knee arthroplasty. Clin Orthop Relat Res. 1994;(309):116–23. [PubMed] [Google Scholar]
  • 3.Weeden SH, Poprosky WG, Sekundiak TD. Use of stems in revision total knee arthroplasty. In: Callaghan JJ, Rosenberg AG, Rubash HE, Simonian PT, Wickiewicz TL, editors. The adult knee. Philadelphia, PA: Lippincott Williams and Wilkins; 2003. pp. p1439–45. [Google Scholar]
  • 4.Bottner F, Laskin R, Windsor RE, Haas SB. Hybrid component fixation in revision total knee arthroplasty. Clin Orthop Relat Res. 2006;446:127–31. doi: 10.1097/01.blo.0000214418.36959.c5. [DOI] [PubMed] [Google Scholar]
  • 5.Crawford DA, Berend KR, Morris MJ, Adams JB, Lombardi AV., Jr Results of a modular revision system in total knee arthroplasty. J Arthroplasty. 2017;32:2792–8. doi: 10.1016/j.arth.2017.03.076. [DOI] [PubMed] [Google Scholar]
  • 6.Brand MG, Daley RJ, Ewald FC, Scott RD. Tibial tray augmentation with modular metal wedges for tibial bone stock deficiency. Clin Orthop Relat Res. 1989;(248):71–9. [PubMed] [Google Scholar]
  • 7.Daines BK, Dennis DA. Management of bone defects in revision total knee arthroplasty. Instr Course Lect. 2013;62:341–8. [PubMed] [Google Scholar]
  • 8.Engh GA, Ammeen DJ. Bone loss with revision total knee arthroplasty: defect classification and alternatives for reconstruction. Instr Course Lect. 1999;48:167–75. [PubMed] [Google Scholar]
  • 9.Anderson JA, Baldini A, MacDonald JH, Tomek I, Pellicci PM, Sculco TP. Constrained condylar knee without stem extensions for difficult primary total knee arthroplasty. J Knee Surg. 2007;20:195–8. doi: 10.1055/s-0030-1248042. [DOI] [PubMed] [Google Scholar]
  • 10.Sabatini L, Risitano S, Rissolio L, Bonani A, Atzori F, Masse A. Condylar constrained system in primary total knee replacement: our experience and literature review. Ann Transl Med. 2017;5:135. doi: 10.21037/atm.2017.03.29. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Rorabeck CH, Taylor JW. Classification of periprosthetic fractures complicating total knee arthroplasty. Orthop Clin North Am. 1999;30:209–14. doi: 10.1016/S0030-5898(05)70075-4. [DOI] [PubMed] [Google Scholar]
  • 12.Felix NA, Stuart MJ, Hanssen AD. Periprosthetic fractures of the tibia associated with total knee arthroplasty. Clin Orthop Relat Res. 1997;(345):113–24. [PubMed] [Google Scholar]
  • 13.Hicks CA, Noble P, Tullos H. The anatomy of the tibial intramedullary canal. Clin Orthop Relat Res. 1995;(321):111–6. [PubMed] [Google Scholar]
  • 14.Tang Q, Zhou Y, Yang D, Xu H, Liu Q. The offset of the tibial shaft from the tibial plateau in Chinese people. J Bone Joint Surg Am. 2010;92:1981–7. doi: 10.2106/JBJS.I.00969. [DOI] [PubMed] [Google Scholar]
  • 15.Conlisk N, Gray H, Pankaj P, Howie CR. The influence of stem length and fixation on initial femoral component stability in revision total knee replacement. Bone Joint Res. 2012;1:281–8. doi: 10.1302/2046-3758.111.2000107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Jazrawi LM, Bai B, Kummer FJ, Hiebert R, Stuchin SA. The effect of stem modularity and mode of fixation on tibial component stability in revision total knee arthroplasty. J Arthroplasty. 2001;16:759–67. doi: 10.1054/arth.2001.25507. [DOI] [PubMed] [Google Scholar]
  • 17.Wood GC, Naudie DD, MacDonald SJ, McCalden RW, Bourne RB. Results of press-fit stems in revision knee arthroplasties. Clin Orthop Relat Res. 2009;467:810–7. doi: 10.1007/s11999-008-0621-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.El-Zayat BF, Heyse TJ, Fanciullacci N, Labey L, Fuchs-Winkelmann S, Innocenti B. Fixation techniques and stem dimensions in hinged total knee arthroplasty: a finite element study. Arch Orthop Trauma Surg. 2016;136:1741–52. doi: 10.1007/s00402-016-2571-0. [DOI] [PubMed] [Google Scholar]
  • 19.Gililland JM, Gaffney CJ, Odum SM, Fehring TK, Peters CL, Beaver WB. Clinical & radiographic outcomes of cemented vs. diaphyseal engaging cementless stems in aseptic revision TKA. J Arthroplasty. 2014;29(9 Suppl):224–8. doi: 10.1016/j.arth.2014.03.049. [DOI] [PubMed] [Google Scholar]
  • 20.Patel AR, Barlow B, Ranawat AS. Stem length in revision total knee arthroplasty. Curr Rev Musculoskelet Med. 2015;8:407–12. doi: 10.1007/s12178-015-9297-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Fleischman AN, Azboy I, Restrepo C, Maltenfort MG, Parvizi J. Optimizing mechanical alignment with modular stems in revision TKA. J Arthroplasty. 2017;32(9S):S209–13. doi: 10.1016/j.arth.2017.05.039. [DOI] [PubMed] [Google Scholar]
  • 22.Fleischman AN, Azboy I, Fuery M, Restrepo C, Shao H, Parvizi J. Effect of stem size and fixation method on mechanical failure after revision total knee arthroplasty. J Arthroplasty. 2017;32(9S):S202–8. doi: 10.1016/j.arth.2017.04.055. [DOI] [PubMed] [Google Scholar]
  • 23.Parsley BS, Sugano N, Bertolusso R, Conditt MA. Mechanical alignment of tibial stems in revision total knee arthroplasty. J Arthroplasty. 2003;18(7 Suppl 1):33–6. doi: 10.1016/S0883-5403(03)00302-4. [DOI] [PubMed] [Google Scholar]
  • 24.Gobba MS, Chan N, Patel R, Noble PC, Incavo SJ. Tibial stems in revision total knee arthroplasty: is there an anatomic conflict? J Arthroplasty. 2015;30(9 Suppl):86–9. doi: 10.1016/j.arth.2014.12.039. [DOI] [PubMed] [Google Scholar]
  • 25.Pfeifer R, Sellei R, Pape HC. The biology of intramedullary reaming. Injury. 2010;41( Suppl 2):S4–8. doi: 10.1016/S0020-1383(10)70002-4. [DOI] [PubMed] [Google Scholar]
  • 26.Wang C, Pfitzner T, von Roth P, Mayr HO, Sostheim M, Hube R. Fixation of stem in revision of total knee arthroplasty: cemented versus cementless-a meta-analysis. Knee Surg Sports Traumatol Arthrosc. 2016;24:3200–11. doi: 10.1007/s00167-015-3820-4. [DOI] [PubMed] [Google Scholar]
  • 27.Heesterbeek PJ, Wymenga AB, van Hellemondt GG. No Difference in implant micromotion between hybrid fixation and fully cemented revision total knee arthroplasty: a randomized controlled trial with radiostereometric analysis of patients with mild-to-moderate bone loss. J Bone Joint Surg Am. 2016;98:1359–69. doi: 10.2106/JBJS.15.00909. [DOI] [PubMed] [Google Scholar]
  • 28.Edwards PK, Fehring TK, Hamilton WG, Perricelli B, Beaver WB, Odum SM. Are cementless stems more durable than cemented stems in two-stage revisions of infected total knee arthroplasties? Clin Orthop Relat Res. 2014;472:206–11. doi: 10.1007/s11999-013-3139-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Foruria X, Schmidt-Braekling T, Nabarte DA, Faschingbauer M, Kasparek M, Boettner F. Does the tibia component design affect the need for offset stems in revision total knee arthroplasty? Arch Orthop Trauma Surg. 2017;137:853–60. doi: 10.1007/s00402-017-2677-z. [DOI] [PubMed] [Google Scholar]