Abstract
Objective
The approximate life span of a silicone maxillofacial prosthesis is as short as 1.5–2 years of clinical service, then a new prosthesis should be fabricated. The most common reason for re-making the prosthesis is silicone mechanical properties degradation. The aim of this study was to assess some mechanical properties of VST-30 silicone for maxillofacial prostheses after addition of intrinsic pigments.
Methods
Two types of intrinsic pigments (rayon flocking and burnt sienna); each of them was incorporated into silicone. One hundred and twenty samples were prepared and split into 4 groups according to the conducted tests (tear strength, hardness, surface roughness, and tensile strength and elongation percentage) with 30 samples for each test. Each group was equally split into three subgroups. Group (A) was without pigment (control group), group (B) was with rayon flocking and group (C) was with burnt sienna.
Results
Samples with rayon flocking showed a highly significant decrease in hardness and there was a significant increase in tear strength, while there were non-significant differences in surface roughness, tensile strength and elongation percentage. Samples with burnt sienna showed a highly significant increase in tear strength and a highly significant decrease in hardness, but surface roughness, tensile strength and elongation percentage showed non-significant differences. However, there were non-significant differences between experimental groups in all tests.
Conclusions
The addition of each of rayon flocking and burnt sienna changed the mechanical properties of the VST-30 silicone, while no superior pigment-silicone combination was revealed in all the conducted tests.
Keywords: Maxillofacial prostheses, Pigments, VST-30 silicone, Mechanical properties, Rayon flocking, Burnt sienna
1. Introduction
“Every human has the divine right to look human.”. This is a quote given by a psychiatrist Dr. William J. Mayo concerning the patient with facial deformities. Keeping in mind the importance of this philosophy, a maxillofacial prosthodontist should work to ensure the return back of the affected individual to society (Rajni et al., 2006).
Increasing numbers of maxillofacial deformity cases were reported in Iraq that are related to non-fatal injuries caused by explosive devices, which now comprise 30% of all battlefield injuries (Lew et al., 2010, Owens et al., 2008). Plastic surgery is the first choice of treatment, but when surgery is inadvisable, rehabilitation with maxillofacial prostheses provides a means of improving patient aesthetics and self-esteem and facilitating their return to society (Guiotti et al., 2010).
Results of the prosthetic treatment are influenced by the properties of the prosthetic material. Mechanical properties and color degradation are the most common reason for maxillofacial prostheses re-making (Begum et al., 2011). Concentration of fillers, additives and pigments used and their types determine the required physical and mechanical properties of the silicone prosthesis, thus it should be tailored to fulfill the requirements of a strong yet elastic material having mechanical properties that fulfill the clinical requirements (Chi, 2014, Eleni et al., 2009).
Pigments used for prostheses coloration are classified as intrinsic and extrinsic pigments. In comparison with extrinsic pigment, intrinsic pigment which forms the requisite color and translucency is less susceptible for handling and ecological status, while it is more probable to influence the mixture characteristics (Han et al., 2010).
However, the effects of commonly used pigments by maxillofacial prosthodontists on the mechanical properties of popular materials used for maxillofacial prostheses have not been investigated (Montgomery and Kiat-Amnuay, 2010).
2. Materials and methods
Two types of intrinsic pigments, rayon flocking and burnt sienna (FI (functional intrinsic)) (Factor II Inc., Lakeside, USA) were separately incorporated into a room temperature vulcanized VST-30 silicone (Factor II Inc. Lakeside, USA). About 120 samples were prepared and split into 4 groups according to the conducted tests with 30 samples for each test. Each group was further more split into three subgroups: (A, B and C) with 10 samples for every subdivision. Group (A) represents control group without pigment, group (B) with rayon flocking and group (C) with burnt sienna (FI).
2.1. Pilot study
A pilot study was carried out to determine the optimum concentration of each pigment to be used in the main study by testing its effect on tear strength and hardness. For rayon flocking, the optimum concentration was 0.1 wt.% concentration, while it was 0.2 wt.% for burnt sienna (FI).
2.2. Mold making
Samples dimensions were designed utilizing AutoCAD 2013 (Autodesk Inc., San Rafael, CA, USA) then processed using computer numerical control machine to form the matrix part of the mold into which the material was poured (Chi, 2014).
2.3. Mixing of the silicone base with the pigments
TheVST-30 silicone mixing was in a proportion of 10:1 for the base and catalyst as recommended by manufacturer’s instructions.
Each pigment was hand-mixed with the base for 5 min ± 5 s by the same operator with a clean stiff flat-ended metal spatula followed by 5 min ± 5 s mixing by a mechanical mixer in a glass beaker (Guiotti et al., 2015, Han et al., 2013, Nguyen et al., 2013).
2.4. Deaeration of the mixture
The mixture of experimental groups was deaerated in a vacuum chamber and allowed to reach its maximum capacity and fall to the bottom of the mixing beaker. Afterward, the vacuum was held for another 5 ± 1 min to eliminate all the air bubbles (Zayed et al., 2014).
The vacuum pressure was 28 inches Hg as recommended by the manufacturer. Deaeration at this stage would decrease the time necessary to have a pore-free mixture after addition of the catalyst (according to the pilot study).
2.5. Refrigeration
The glass beaker containing the material was placed in a zip-locked plastic bag and placed in a refrigerator. Improvement of handling properties of the material was achieved by refrigeration of the base prior to use as suggested by the manufacturer.
2.6. Adding the catalyst
Mixing of the base and catalyst was at 50 ± 10% RH (relative humidity) and 23 ± 2 °C (controlled temperature). A flat-ended metal spatula was used for mixing of the base and catalyst as recommended by the manufacturer. The base and catalyst were hand-mixed by stirring with the spatula for 1 min by the same operator and in one direction (Guiotti et al., 2015, Hulterstrӧm, 2012, Willett and Beatty, 2015).
2.7. Pouring of the mixture into the deaeration and pouring syringe
After mixing of the base for the control group or the pigmented base for the experimental groups with the catalyst, the mixture was loaded into a custom-made syringe for deaeration and pouring.
In order to remove all the air bubbles entrapped during mixing with the catalyst, a vacuum pump was attached to the deaeration and pouring syringe for (3 ± 1) minutes as recommended by the manufacturer’s instructions(Fig. 1).
2.8. Pouring the mixed material into the mold
The matrix and the bottom parts of the mold were previously securely attached and placed on the vibrator with the mold spaces coated with petroleum jelly (Hatamleh and Watts, 2010, Zayed et al., 2014).
The material was injected from the deaeration and pouring syringe into the mold and a glass slab, which was previously coated with petroleum jelly, was laid onto the matrix filled with the material.
Lying of the glass slab was started from one side by resting the bottom edge of the slab and holding the top edge, while the glass slab was carefully and slowly lowered onto matrix to force excess material and air out ahead of it (Fig. 2). Finally, the cover of mold was placed on the glass slab, a mass of 1 kg was applied on the center and the cover was tightened.
2.9. Demolding and storage of samples
The material sets in about 30 min according to manufacturer’s product description, then the samples were removed carefully from the mold (Pinheiro et al., 2014). Samples that had visible defects (to the same operator) were discarded before testing (Al-Harbi et al., 2015).
The samples were stored in a custom-made lightproof box in an air-conditioned room. During storage, the temperature was 10–30 °C and RH did not excess 80% (Brown, 2006).
2.10. Conditioning of samples
Samples were conditioned for 24 h prior to testing and an ultrasonic humidifier was used to increase humidity if RH was below 50% (ASTMD624, 2012). Then, samples were conditioned at a standard laboratory temperature of 23 ± 2 °C for a minimum of 3 h after removal of flash (ASTM International, 2010). The flash was removed with a scalpel and sharp surgical blade # 10 (Zayed et al., 2014).
2.11. Mechanical testing procedures
2.11.1. Tear strength test
Samples preparation and testing were done according to ASTM, D624 (Standard and ISO, 2010) (American Society for Testing and Materials). Type C sample was used to measure tear initiation strength. The following equation was used to determine the tear strength:
where
F: The maximum force required for sample breaking in kilonewtons.
D: The median thickness of each sample in meter (ASTMD624, 2012).
2.11.2. Hardness test
The test was performed according to ASTM D2240 on 25 × 25 × 6 mm3 samples and type A shore hardness digital tester was used (Standard, 2010).
The mean value of five readings from five different points apart from each other by 6 mm while keeping 6 mm away from the border was considered as the hardness of the sample (Hatamleh and Watts, 2010).
2.11.3. Surface roughness test
In the surface roughness average (Ra) test, the sample dimensions were (10 mm × 10 mm × 2 mm) (Al-Askari et al., 2014, Khalaf, 2014). Profilometer tester was used and for each sample, three measurements were taken, then the mean value of them was calculated and considered as the surface roughness of the sample (Goiato et al., 2009).
2.11.4. Tensile strength testing
The test was managed depending on ISO 37 (International Organization for Standardization) and dumbbell-shaped samples (type 2) were prepared (Standard and ISO, 2010). The ultimate tensile was calculated from the maximum stretching force at break divided by the sample original cross-sectional area using the equation below:
where
: The force recorded at break in N.
: The original cross-sectional area of the sample in mm2 (Standard and ISO, 2010).
2.11.5. Elongation percentage
In accordance with ISO 37, elongation before break was executed at the time of tensile strength measuring. The break elongation was measured from the original length of tensile sample and the length of the sample at break using the equation:
where
L°: The original length in mm.
Lb: Extension at break in mm (Standard and ISO, 2010).
2.12. Statistical analyses
The statistical package for the social sciences software (version 23) was used for analyzing the data of this study. The following statistics were performed:
-
(a)
Descriptive statistics: Graphical display by bar charts.
-
(b)Inferential statistics: One-way ANOVA (One-way analysis of variance) and LSD (least significant difference) as a post hoc were used with the following significance levels:
- P > 0.05 NS Non-significant
- 0.05 ≥ P > 0.01 S Significant
- P ≤ 0.01 HS Highly significant
3. Results
3.1. Tear strength test
The highest mean value of tear strength test appeared in group (C), then followed by group (B), while group (A) mean value was the lowest among the groups (Fig. 3). One-way ANOVA for tear strength results showed a highly significant difference (Table 1).
Table 1.
Sum of squares | df | Mean square | F | Sig. | |
---|---|---|---|---|---|
Between groups | 78.961 | 2 | 39.480 | 9.318 | 0.001 HS |
Within groups | 114.401 | 27 | 4.237 | ||
Total | 193.362 | 29 |
To compare the mean values of each two groups of all the three groups, Post-hoc LSD test was performed. There was a significant difference between (A and B) groups and a highly significant difference between (A and C) groups while there was a non-significant difference between experimental groups (Table 2).
Table 2.
Study groups | Mean difference | SE | Sig. | |
---|---|---|---|---|
Group (A) | Group (B) | −2.268 | 0.921 | 0.020 S |
Group (B) | Group (C) | −1.692 | 0.921 | 0.077 NS |
Group (C) | Group (A) | 3.960 | 0.921 | 0.000 HS |
3.2. Hardness test
The highest mean value appeared in group (A), then followed by group (C), while the lowest mean value among the groups was in group (B) (Fig. 4). One-way ANOVA for shore A hardness results showed highly significant difference among groups (Table 3).
Table 3.
Sum of squares | df | Mean square | F | Sig. | |
---|---|---|---|---|---|
Between groups | 11.861 | 2 | 5.931 | 11.856 | 0.000 HS |
Within groups | 13.506 | 27 | 0.500 | ||
Total | 25.367 | 29 |
Post-hoc LSD test was performed to compare mean values of each two groups for all the three groups. Comparison between the control group and each of the experimental groups showed highly significant differences. The difference between the experimental groups was non-significant (Table 4).
Table 4.
Study groups | Mean difference | SE | Sig. | |
---|---|---|---|---|
Group (A) | Group (B) | 1.455 | 0.316 | 0.000 HS |
Group (B) | Group (C) | −0.290 | 0.316 | 0.367 NS |
Group (C) | Group (A) | −1.165 | 0.316 | 0.001 HS |
3.3. Surface roughness test
The highest mean value appeared in group (B), and then followed by group (C), while group (A) mean value was the lowest among the groups (Fig. 5). One-way ANOVA for roughness test results showed non-significant difference among groups (Table 5).
Table 5.
Sum of squares | df | Mean square | F | Sig. | |
---|---|---|---|---|---|
Between groups | 0.025 | 2 | 0.012 | 3.223 | 0.056 NS |
Within groups | 0.104 | 27 | 0.004 | ||
Total | 0.129 | 29 |
3.4. Tensile strength test
The highest mean value of tensile strength test appeared in group (B), and then followed by group (A), while group (C) mean value was the lowest among the groups (Fig. 6). One-way ANOVA for tensile strength test results showed a non-significant difference among groups (Table 6).
Table 6.
Sum of squares | df | Mean square | F | Sig. | |
---|---|---|---|---|---|
Between groups | 9.885 | 2 | 4.943 | 1.148 | 0.332 NS |
Within groups | 116.216 | 27 | 4.304 | ||
Total | 126.101 | 29 |
3.5. Elongation percentage test
In group (A), the highest mean value of elongation percentage test was found, then followed by group (B), while group (C) mean value was the lowest among the groups (Fig. 7). One-way ANOVA for the results of elongation percentage test revealed a non-significant difference among groups (Table 7).
Table 7.
Sum of squares | df | Mean square | F | Sig. | |
---|---|---|---|---|---|
Between groups | 22708.267 | 2 | 11354.133 | 0.946 | 0.401 NS |
Within groups | 323977.600 | 27 | 11999.170 | ||
Total | 346685.867 | 29 |
4. Discussion
The mechanical value of silicones is only reported by manufacturers without pigments, fillers or additives and this is not a real representation of silicones clinical performance when used for extra-oral prosthesis. For this reason, maxillofacial prosthodontists and anaplastologists should deal cautiously with the manufacturers’ values when using a material for making a facial prosthesis (Nguyen et al., 2013).
The results of tear strength test indicated that the tear strength was increased significantly when rayon flocking was added and highly significantly when burnt sienna (FI) was added.
The significant increase in tear strength when rayon flocking was incorporated may be attributed to the rayon flocking being fibers and these fibers bridged the tear and hindered or obstructed the propagating tear (Kumar and Thomas, 1995, Murty and De, 1984, Sreeja, 2012).
The highly significant increase in tear strength when burnt sienna (FI) was added may be due to liquid colorant’s action as a plasticizer which could enhance the tear strength as one of the main functions of plasticizer is improvement of tear resistance (Guiotti et al., 2015, Haug et al., 1999, Wypych, 2004).
The result of hardness test revealed a significantly high decrease in the mean value of hardness when each of rayon flocking and burnt sienna (FI) was added in comparison with the control group.
The highly significant decrease in hardness results may be because the intrinsic pigment incorporation affected the process of silicone polymerization, resulting in decreased hardness (Guiotti et al., 2015).
The results of surface roughness test indicated that both types of pigments increased the roughness mean value as compared with the control group.
Rayon flocking has short whiskers protruding from its surface (Leny and Narayanankutty, 2009). It is assumed that the rayon flocking fibers were randomly arranged during sample preparation. These different orientations across the surface along with these protruding whiskers that were distributed on the surface of the silicone might cause the increase in mean value after addition of rayon flocking.
Burnt sienna (FI), which is a liquid pigment, uses a vehicle for the pigment that facilitates handling of the pigment in liquid form. When the vehicle was absorbed or evaporated the material hardens (Haug et al., 1999). Eventually, the distributed pigment particles on the silicone surface may increase surface roughness (Yu et al., 1980).
To the knowledge of the researcher, to date, no researchers have attempted to assess the effect of rayon flocking on surface roughness of maxillofacial silicones (both types RTV and HTV) and that was the uniqueness of this study.
The results of tensile strength test indicated that rayon flocking addition to silicone increased the tensile strength mean value, while the addition of burnt sienna (FI) reduced the mean value of tensile property in comparison with the control group.
The increase in tensile strength mean value after rayon flocking incorporation may be due to stress transmit from the weaker resin matrix to the much stronger fibers and the effective restraining of the matrix by fibers, which led to hindering of the growing crack (Leny and Narayanankutty, 2009, Rosato and Rosato, 2004, Sreeja, 2012).
According to the manufacturer’s product description, burnt sienna (FI) is a combination of crushed cosmetic pigments in cross-linking fluid of silicone making a liquid but viscous pigment. The reduction in tensile strength mean value after addition of burnt sienna (FI) may be explained by the increased degree of cross-linking, which was caused by the silicone cross-linking fluid, that interferes with the redistribution of strain energy resulting in greater localized stress then early fracture(Polyzois et al., 1992).
The results of elongation percentage showed that both types of pigments decreased the elongation percentage mean value in comparison with the control group.
The reduction in elongation percentage mean value after addition of rayon flocking may be due to that the fibers prevent the flow and orientation of molecular chains making the matrix more restrained leading to initiation of failure at multiple points and hence causing considerably lower elongation percentage values(Leny and Narayanankutty, 2009, Sreeja and Kutty, 2002, Sreeja, 2012).
The decrease in mean value after adding burnt sienna (FI), most likely due to the gradually increased crystallization facilitated by the increased chain mobility and the interactions in the presence of plasticizer as stated previously in tear strength that liquid pigment may act as a plasticizer (Wypych, 2004).
5. Conclusions
As a conclusion, a change in the mechanical properties resulted when intrinsic pigments were incorporated into the RTV silicone VST-30, while both intrinsic pigments exhibited different results. No superior pigment-silicone combination was revealed in all the conducted tests.
Conflict of interest
The author of this article declares no conflict of interest.
Footnotes
Peer review under responsibility of King Saud University.
References
- Al-Askari S.K., Ariffin Z., Husein A., Reza F. Comparison of microbial adherence to silicone elastomers for maxillo-facial prostheses. World J. Med. Sci. 2014;11(2):161–165. [Google Scholar]
- Al-Harbi F.A., Ayad N.M., Saber M.A., ArRejaie A.S., Morgano S.M. Mechanical behavior and color change of facial prosthetic elastomers after outdoor weathering in a hot and humid climate. J. Prosthet. Dent. 2015;113(2):146–151. doi: 10.1016/j.prosdent.2014.09.008. [DOI] [PubMed] [Google Scholar]
- ASTM D2240-05 . ASTM International; West Conshohocken, PA: 2010. Standard Test Method for Rubber Property-Durometer Hardness. [Google Scholar]
- ASTM D624-00 . ASTM International; West Conshohocken, PA: 2012. Standard Test Method for Tear Strength of Conventional Vulcanized Rubber and Thermoplastic Elastomers. [Google Scholar]
- Begum Z., Kola M.Z., Joshi P. Analysis of the properties of commercially available silicone elastomers for maxillofacial prostheses. IJCD. 2011;2(4) [Google Scholar]
- Brown, R., 2006. Physical Testing of Rubber, fourth ed. Springer Science & Business Media.
- Chi, Y.H., 2014. Effect of silica filler on the mechanical properties of silicone maxillofacial prosthesis. M.Sc. Thesis, School of Dentistry, Indiana University.
- Eleni P.N., Katsavou I., Krokida M.K., Polyzois G.L. Color stability of facial silicone prosthetic elastomers after artificial weathering. Dent. Mater. J. 2009;5(2):71–79. doi: 10.1016/j.dental.2009.06.018. [DOI] [PubMed] [Google Scholar]
- Goiato M.C., Pesqueira A.A., dos Santos D.M., Dekon S.F.de C. Evaluation of hardness and surface roughness of two maxillofacial silicones following disinfection. Braz. Oral Res. 2009;23(1):49–53. doi: 10.1590/s1806-83242009000100009. [DOI] [PubMed] [Google Scholar]
- Guiotti A.M., Goiato M.C., dos Santos D.M. Evaluation of the shore a hardness of silicone for facial prosthesis as to the effect of storage period and chemical disinfection. J. Craniofac. Surg. 2010;21(2):323–327. doi: 10.1097/SCS.0b013e3181cf5fa4. [DOI] [PubMed] [Google Scholar]
- Guiotti, A.M., Goiato, M.C., dos Santos, D.M., Vechiato-Filho, A.J., Cunha, B.G., Paulini, M.B., et al., 2015. Comparison of conventional and plant-extract disinfectant solutions on the hardness and color stability of a maxillofacial elastomer after artificial aging. J. Prosthet. Dent. [DOI] [PubMed]
- Han Y., Zhao Y., Xie C., Powers J.M., Kiat-amnuay S. Color stability of pigmented maxillofacial silicone elastomer: effects of nano-oxides as opacifiers. J. Dent. 2010;38(Suppl 2):e100–e105. doi: 10.1016/j.jdent.2010.05.009. [DOI] [PubMed] [Google Scholar]
- Han Y., Powers J.M., Kiat-amnuay S. Effect of opacifiers and UV absorbers on pigmented maxillofacial silicone elastomer, part 1: Color stability after artificial aging. J. Prosthet. Dent. 2013;109(6):397–401. doi: 10.1016/S0022-3913(13)60327-3. [DOI] [PubMed] [Google Scholar]
- Hatamleh M.M., Watts D.C. Mechanical properties and bonding of maxillofacial silicone elastomers. Dent. Mater. J. 2010;26(2):185–191. doi: 10.1016/j.dental.2009.10.001. [DOI] [PubMed] [Google Scholar]
- Haug S.P., Andres C.J., Moore B.K. Color stability and colorant effect on maxillofacial elastomers. Part I: colorant effect on physical properties. J. Prosthet. Dent. 1999;81(4):418–422. doi: 10.1016/s0022-3913(99)80008-0. [DOI] [PubMed] [Google Scholar]
- Hulterstrӧm, A.K., 2012. Silicone obturators and the bacterial flora in symptomatic nasal septal perforations. Umeå Universitet, Ph.D. Thesis, Faculty of Medicine, Department of Odontology, Umea University.
- ISO 37, 2011. Rubber, vulcanized or thermoplastic—Determination of tensile stress-strain properties.
- Khalaf, S., 2014. Ariffin, Z., Husein, A., Reza, F. (Eds.), Surface Coating of Gypsum-Based Molds for Maxillofacial Prosthetic Silicone Elastomeric Material: The Surface Topography. J. Prosthodont. [DOI] [PubMed]
- Kumar R.P., Thomas S. Short fibre elastomer composites: effect of fibre length, orientation, loading and bonding agent. Bull. Mater. Sci. 1995;18(8):1021–1029. [Google Scholar]
- Leny, M., Narayanankutty, S.K., 2009. Development of Elastomeric Hybrid Composite Based on Synthesised Manosilica and Short Nylon Fiber. Ph.D. Thesis, Cochin University of Science and Technology.
- Lew T.A., Walker J.A., Wenke J.C., Blackbourne L.H., Hale R.G. Characterization of craniomaxillofacial battle injuries sustained by United States service members in the current conflicts of Iraq and Afghanistan. J. Oral. Maxillofac. Surg. 2010;68(1):3–7. doi: 10.1016/j.joms.2009.06.006. [DOI] [PubMed] [Google Scholar]
- Montgomery P.C., Kiat-Amnuay S. Survey of currently used materials for fabrication of extraoral maxillofacial prostheses in North America, Europe, Asia, and Australia. J. Prosthodont. 2010;19(6):482–490. doi: 10.1111/j.1532-849X.2009.00538.x. [DOI] [PubMed] [Google Scholar]
- Murty V., De S. Short-fiber-reinforced styrene-butadiene rubber composites. J. Appl. Polym. Sci. 1984;29(4):1355–1368. [Google Scholar]
- Nguyen C.T., Chambers M.S., Powers J.M., Kiat-amnuay S. Effect of opacifiers and UV absorbers on pigmented maxillofacial silicone elastomer, part 2: mechanical properties after artificial aging. J. Prosthet. Dent. 2013;109(6):402–410. doi: 10.1016/S0022-3913(13)60328-5. [DOI] [PubMed] [Google Scholar]
- Owens B.D., Kragh J.F., Jr, Wenke J.C., Macaitis J., Wade C.E., Holcomb J.B. Combat wounds in operation Iraqi Freedom and operation Enduring Freedom. J. Trauma Acute Care Surg. 2008;64(2):295–299. doi: 10.1097/TA.0b013e318163b875. [DOI] [PubMed] [Google Scholar]
- Pinheiro J.B., Reis A.C., Pisani M.X., Leite V.M.F., Souza R.F., Paranhos H.F.O. Microstructural characterization and evaluation of the properties of polymeric materials for maxillofacial prosthetics. J. Med. Eng. Technol. 2014;38(2):67–75. doi: 10.3109/03091902.2013.864715. [DOI] [PubMed] [Google Scholar]
- Polyzois G., Stafford G.D., Winter R. A study of some mechanical properties of an RTV polydimethylsiloxane for extraoral maxillofacial prostheses. Clin. Mater. 1992;9(1):21–29. doi: 10.1016/0267-6605(92)90006-f. [DOI] [PubMed] [Google Scholar]
- Rajni, D., Balaji, P., Smita, A., 2006. Prosthetic rehabilitation of hemi-mandibulectomy patient. J. Indian Prosthodont. Soc. 6(1), 47–51.
- Rosato, D.V., Rosato, D.V., 2004. Reinforced Plastics Handbook, third ed. Elsevier.
- Sreeja T., Kutty S. Studies on acrylonitrile butadiene rubber-short nylon fiber composites. J. Elastomers Plast. 2002;34(2):157–169. [Google Scholar]
- Sreeja, T., 2012. Studies on Short Nylon Fiber-Reclaimed Rubber/Elastomer Composites. Ph.D Thesis, Cochin University of Science and Technology.
- Willett E.S., Beatty M.W. Outdoor weathering of facial prosthetic elastomers differing in Durometer hardness. J. Prosthet. Dent. 2015;113(3):228–235. doi: 10.1016/j.prosdent.2014.09.009. [DOI] [PubMed] [Google Scholar]
- Wypych G. Chem Tec Publishing; Toronto, Canada: 2004. Plasticizers Use and Selection for Specific Polymers. [Google Scholar]
- Yu R., Koran A., Craig R.G. Physical properties of a pigmented silicone maxillofacial material as a function of accelerated aging. J. Dent. Res. 1980;59(7):1141–1148. doi: 10.1177/00220345800590070801. [DOI] [PubMed] [Google Scholar]
- Zayed, S.M., Alshimy, A.M., Fahmy, A.E., 2014. Effect of surface treated silicon dioxide nanoparticles on some mechanical properties of maxillofacial silicone elastomer. Int. J. Biomater. [DOI] [PMC free article] [PubMed]
Further reading
- ASTM D1349 . ASTM International; West Conshohocken, PA: 2014. Standard Temperature for Testing. [Google Scholar]