Abstract
Intentional and unintentional contamination incidents, such as terrorist attacks, natural disasters, and accidental spills, can result in large volumes of contaminated water. These waters may require pre-treatment before disposal and assurances that treated waters will not adversely impact biological processes at wastewater treatment facilities, or receiving waters. Based on recommendations of an industrial workgroup, this study addresses such concerns by studying electrochemical advanced oxidation process (EAOP) pre-treatment for contaminated waters, using a boron-doped diamond (BDD) anode, prior to discharge to wastewater treatment facilities. Reaction conditions were investigated, and microbial toxicity was assessed using the Microtox® toxicity assay and the Nitrification Inhibition test. A range of contaminants were studied including herbicides, pesticides, pharmaceuticals and flame retardants. Resulting toxicities varied with supporting electrolyte from 5%−92%, often increasing, indicating that microbial toxicity, in addition to parent compound degradation, should be monitored during treatment. These toxicity results are particularly novel because they systematically compare the microbial toxicity effects of a variety of supporting electrolytes, indicating some electrolytes may not be appropriate in certain applications. Further, these results are the first known report of the use of the Nitrification Inhibition test for this application. Overall, these results systematically demonstrate that anodic oxidation using the BDD anode is useful for addressing water contaminated with refractory organic contaminants, while minimizing impacts to wastewater plants or receiving waters accepting EAOP-treated effluent. The results of this study indicate nitrate can be a suitable electrolyte for incident response and, more importantly, serve as a baseline for site specific EAOP usage.
Keywords: Boron-Doped Diamond Anode, Nitrification Inhibition, Microtox® Toxicity, Electrochemical Advanced Oxidation Process
1. Introduction
Effectively managing water contamination, resulting from intentional/unintentional events such as terrorist attacks, natural disasters, or accidental spills, continues to be a primary objective for the water industry. Water contamination also arises from firefighting or decontamination activities, including washing of contaminated infrastructure and equipment following an indoor or outdoor contamination incident. These incidents can produce large volumes of contaminated water, which can pose pre-treatment, treatment, and disposal concerns for wastewater treatment works (WWTWs) and receiving waters.
Advanced oxidation processes (AOPs) can be effective at pre-treating contaminated water because they produce hydroxyl radicals, highly reactive and non-selective oxidants that exhibit higher oxidation potentials than chlorine or ozone. This chemical oxidation allows AOPs to degrade, and potentially mineralize, contaminants without the use of reagents like chlorine, reducing the potential reagent handling concerns and the formation of chlorinated byproducts (Sirés et al., 2014). These advantages make AOPs appropriate for pre-treatment of contaminated waters, especially where onsite treatment is necessary. However, the use of AOPs in this application has knowledge gaps related to the potential to produce oxidation byproducts and effluents that may stress the biological treatment processes at WWTWs or result in the discharge of toxic contaminants into the receiving waters. Wastewater utilities and related stakeholders discussed these knowledge gaps at a workshop hosted by the Water Environment Research Foundation (currently known as Water Environment & Reuse Foundation), where “participants agreed that various bench-scale studies to develop information on the treatability of contaminants by AOP (such as rate constants) and toxicity testing are needed to demonstrate an acceptable quality of AOP-treated water” (Striano et al., 2011).
Electrochemical advanced oxidation processes (EAOPs) pose advantages over some other types of AOP treatment. One main advantage associated with the EAOP approach is that it does not require reagents with handling concerns, thereby reducing the design, operation, and maintenance considerations. EAOPs may require an electrolyte, which may not be present in sufficient quantities for the types of contaminated waters generated from contamination incident response scenarios. Many electrolytes have been proposed and the literature demonstrates that each may participate in reactions to produce a variety of secondary oxidants ( Martínez-Huitle and Brillas, 2009; Anglada et al., 2011; Kornienko et al., 2011; Sirés et al., 2014; Brillas and Martínez-Huitle, 2015; Jalife-Jacobo et al., 2016). These electrolyte and source-water specific oxidants and their reaction products may pose toxicity concerns. To the authors’ knowledge, there are presently no systematic studies assessing the microbial toxicity produced across a range of contaminants and electrolytes upon EAOP treatment using BDD anodes.
The boron-doped diamond (BDD) anode, in particular, is ideal for EAOP treatment because it is able to attain a higher range of working potentials and is more stable to corrosion than other electrodes, resulting in higher overall performance (Kornienko et al., 2011). Moreover, the BDD anode is non-active, meaning it has an inert surface, reducing concerns about adsorption and deactivation due to fouling (Saylor et al., 2012). In electro-oxidation with BDD anodes, contaminants can be degraded either through direct anodic oxidation, or, more likely, through mediated reactions with adsorbed hydroxyl radicals produced from the electrolysis of water (Eq. 1). A variety of other side reactions are possible, some of which consume hydroxyl radicals, leading to contaminant degradation via other reactive oxygen species and secondary oxidants (Sirés et al., 2014; Uranga-Flores et al., 2015; Jalife-Jacobo et al., 2016). The BDD literature covers a wide range of reaction conditions such as pH values, electrolytes, current densities, and contaminant/electrolyte combinations and concentrations. To the authors’ knowledge, few studies explore contaminant degradation in un-buffered solutions at natural solution pH, as are often found in real-world contaminated waters.
| (1) |
The effluent toxicity was a major concern to the WERF expert workgroup for ultimate disposal of contaminated waters treated in emergency response scenarios(Striano et al., 2011). Limited studies have reported that AOP treatments yield degradation intermediates that can be more toxic than the parent compound (Saylor et al., 2012; Haidar et al., 2013). A variety of microbial toxicity assays demonstrate potential eco-toxicity, as well as toxicity to activated sludge from WWTWs. These assays indicate that various waters are unlikely to harm WWTW processes or receiving waters (Striano et al., 2011). The Microtox® toxicity assay, a widely used and standardized test, uses Vibrio fischeri, luminescent marine bacteria, to indicate toxicity and is often utilized as a receiving water toxicity surrogate (Montalbán et al., 2016). This assay is sensitive, time-efficient, and low cost, making it a good screening test, especially if coupled with another toxicity assay (Dalzell et al., 2002). The Nitrification Inhibition (NI) toxicity assay is ideal to couple with the Microtox® toxicity assay for disposal of contaminated waters because, in spite of its high cost/sample, the NI test provides data directly applicable for determining toxicity to WWTW biological processes (Dalzell et al., 2002). The NI test utilizes activated sludge from local WWTWs to determine the response of nitrifying bacteria to contaminants/sample waters. Nitrifying bacteria are often considered to be the most sensitive bacteria in the WWTW processes due to their slow specific growth rates and sensitivity to toxic compounds. Thus, the nitrifying bacteria serve as an indicator of the worst case scenario for WWTW biological processes (Fox et al., 2006; Pagga et al., 2006).
The goal of this study is to systematically investigate factors affecting microbial toxicity arising from contaminant degradation using a BDD anode for electro-oxidation treatment following response activities for a variety of contamination scenarios. This study examines the efficacy of electrochemical oxidation using a BDD anode, while varying conditions including current densities, reaction times, and electrolyte composition and concentration. Further, it includes a novel and diverse range of contaminants of emerging interest with chemical structures spanning a variety of reaction mechanisms and reported hydroxyl radical reaction rates. To address treatment and toxicity concerns raised by wastewater utilities and stakeholders at a Water Environment Research Foundation workshop (Striano et al., 2011), electrochemical oxidation is assessed using both contaminant destruction and resulting microbial toxicity, indicated by the Microtox® toxicity assay and the NI test. Microbial toxicity is a parameter not often reported in conjunction with the EAOP literature utilizing BDD anodes, with only a few studies reporting Microtox® toxicity. To the authors’ knowledge, this is the first study to utilize the NI toxicity assay to assess toxic effects of waters treated with electro-oxidation using BDD anodes on WWTW biological processes. Further, this study uniquely reports directly comparable toxicities resulting not only from a variety of contaminants, but also from a variety of background electrolyte matrices for use in electro-oxidation with BDD anodes.
2. Materials and Methods
2.1. Electrochemical Oxidation System Using a BDD Anode
This electrochemical oxidation setup utilized a 10 L reservoir with a BDD anode in an electrode flow cell designed and manufactured by ADT (Diamonox™ Model 40 Monopolar Electrochemical Cell, ADT, Romeoville, IL). Local tap water, representing a realistic water that might be contaminated by emergency situations, was dechlorinated by passage through granulated activated carbon, and then was supplemented with approximately 10 mg L−1 of contaminants (listed in Table S1) and experimentally determined concentrations of Na2SO4, NaH2PO4, NaNO3, Na2CO3, NaCl, and Na2B4O7.12H2O electrolytes. After mixing for 10 minutes, this solution was recirculated over the 40 cm2 electrode with an applied current of 4.4 amperes (~13 volts) at 8–10 liters per minute, except for a few samples treated at 8.8 amperes for comparison. Most experiments were carried out for two hours, with samples collected for LC-MS/MS analysis at 0, 3, 10, 20, 40, 60, 90, and 120 minutes. A few longer time period experiments were performed (24 and 72 hours) with samples collected throughout the experimental period. Toxicity tests were performed on a pre-selected subset of these samples (see below).
2.2. Analytical Methods
The N,N-diethyl-p-phenylenediamine (DPD) method for free chlorine, based on Standard Methods 4500-Cl G (2014), and an adapted DPD method with 0.25 μL of 1 g L−1 horseradish peroxidase added to allow for hydrogen peroxide measurement were utilized to monitor oxidant presence. Samples were quenched with sodium thiosulfate where necessary and adjusted to neutral pH prior to analysis.
A direct-inject liquid chromatograph with tandem mass spectrometer (LC-MS/MS) was used to quantify concentrations of target contaminants using an AB SCIEX 5500 QTrap (SCIEX, Framingham, MA) or Thermo TSQ quantum MS/MS (Thermo Fisher Scientific, Waltham, MA) with a Dionex Ultimate LC system (Dionex, Sunnyvale, CA). The method included monitoring two multiple reaction monitoring (MRM) transitions for each analyte (see Table S2). Perfluorooctanoic acid (PFOA) was analyzed using a direct inject LC-MS/MS method using a Quatro Premier mass spectrometer (Waters Corporation, Milford, MA) monitoring the ions noted in Table S2.
2.3. Toxicity Measurements
Samples (175 mL, in triplicate) of non-contaminated drinking water (control), untreated contaminated drinking water (Time 0), and EAOP-treated contaminated drinking water (120 minutes) were placed in respirometer (Model WB1000-SB, N-CON, Crawford, GA) reaction vessels used for aeration throughout the test. Ammonia concentrations in these samples were adjusted to approximately 45 mg L−1. Mixed liquor was collected daily from a local WWTW and allowed to settle. Settled sludge flocs (175 mL) were added to each sample vessel, and the ammonia nitrogen concentration was measured both before and after two hours of aeration. Ammonia concentrations were determined by the Salicylate Method(2015) using Amver™ Diluent Reagent High Range Test ‘N Tube™ vials, and a DR/2000 spectrophotometer all from HACH® (Loveland, CO). The NI test was considered complete when the ammonia concentration of the control decreased by 25%.
The Microtox® toxicity assay used the Microtox® Model 500 (Modern Water Inc., New Castle, DE). Differences in light output above 10% are attributed to toxic effects of the sample according to ISO 11248–2007 (Sánchez et al., 2014). Samples were analyzed in duplicate after 0, 10, 40, 90, and 120 minutes of EAOP treatment for two-hour trials and at extended intervals for longer experiments.
3. Results and Discussion
3.1. Selection of BDD Electrolytes
3.1.1. Effect of Electrolyte Concentration and Toxicity
Initial batch experiments examined the suitability of six electrolytes for electro-oxidation using a BDD anode and microbial toxicity testing. The Microtox® test was utilized as a screening tool because it is the more sensitive of the two toxicity assays utilized (Dalzell et al., 2002). Microtox® toxicities of electrolyte solutions of 0.05 N, 0.1 N and 1.0 N Na2SO4, NaH2PO4, NaNO3, Na2CO3 and NaCl were tested to determine initial background toxicities of these potential electrolyte matrices. Solutions of 0.05 N, 0.1 N, and 0.2 N Na2B4O7 were utilized due to solubility constraints.
Higher normality electrolyte solutions tended to yield higher Microtox® toxicities. The NaH2PO4, Na2CO3 and Na2B4O7 electrolytes exhibited Microtox® inhibitions of 86% or greater at all normalities tested. The elevated observed pH values (pH 8 for Na2B4O7 and pH 11–12 for Na2CO3) and the buffering capacity of NaH2PO4 may have contributed to this observed toxicity. The other electrolytes exhibited less than 3% inhibition at both 0.01 N and 0.1 N concentrations. Background Microtox® toxicities were investigated using full blank trials with only NaCl, NaNO3, Na2SO4, or NaH2PO4 with pH correction prior to toxicity testing to investigate toxicity, solubility constraints, and pH considerations.
3.1.2. Impacts of Electrolytes on Solution pH
Changes in pH affect electrochemical oxidation via contaminant and oxidant speciation, so pH changes during treatment are an important electrolyte selection criterion. Increases up to approximately pH 12 in NaNO3, NaCl, and Na2SO4 blank experiments suggest electrolyte interactions within the electrochemical system. Na2SO4 trials exhibited the highest pH and greater increases in pH for higher electrolyte normalities. The NaH2PO4 was the only electrolyte that yielded fairly stable and lower (pH 4.3 to 5.7) pH values throughout trials, as expected given its buffering capacity. The buffering capacity of the NaH2PO4, as well as the high pH values of the other electrolytes, can pose operational challenges with pH neutralization, which will be necessary to eliminate toxicity to sludge and/or receiving water.
Changes in pH are expected during electro-oxidation using a BDD anode, because the electrochemical production of hydroxyl radicals (Eq. 1) produces H+, thereby decreasing the pH of treated waters (Uranga-Flores et al., 2015). Further, hydroxyl radicals may then combine to form hydrogen peroxide (Eq. 2), which may subsequently decay (Eq. 3), yielding oxygen and more H+ (Uranga-Flores et al., 2015).
| (2) |
| (3) |
However, because the NaNO3, NaCl, and Na2SO4 electrolytes all exhibited different amounts of pH increase, these reactions are not sufficient to explain the observed pH differences, indicating that more specific interactions with each electrolyte are likely occurring. Indeed, the production of secondary oxidants via reactions between electrolytes, the electrodes, and/or hydroxyl radicals have been reported to affect H+ and OH− concentrations, thereby affecting the solution pH. For example, chloride from the NaCl electrolyte may react to form chlorine (Eq. 4) and then, in an alkaline medium, consume OH- to produce hypochlorite (Eq. 5), thereby lowering the pH (Sirés et al., 2014).
| (4) |
| (5) |
Further, the sulfate from the Na2SO4 electrolyte has been reported to react through two mechanisms to ultimately produce peroxodisulfate. The first involves direct oxidation of sulfate at the electrode (Eq. 6) (Uranga-Flores et al., 2015; Jalife-Jacobo et al., 2016). If the solution pH is such that the sulfate is protonated, the HSO4− can also react, yielding H+ and decreasing the pH (Eq. 7) (Uranga-Flores et al., 2015). However, the second mechanism produces OH− through indirect hydroxyl radical mediated reactions (Eqs. 8 and 9), which would be expected to increase the pH (Uranga-Flores et al., 2015; Jalife-Jacobo et al., 2016).
| (6) |
| (7) |
| (8) |
| (9) |
Likewise, phosphate is also reported to undergo direct (Eq. 10) (Brillas and MartínezHuitle, 2015) and indirect hydroxyl radical mediated (Eq. 11) reactions to produce peroxodiphosphate, as well as other peroxophosphates, which are also affected by the pH and whether or not they are protonated (Sirés et al., 2014).
| (10) |
| (11) |
Unlike the other electrolytes, the nitrate electrolyte is not expected to undergo oxidation reactions. Rather, under some conditions, nitrate reduction may occur at the cathode (Eq. 12), yielding OH− formation and increases in pH, as well as transformation and removal of nitrogen species through off-gassing (Eq. 13) (Lévy-Clément et al., 2003; Martin de Vidales et al., 2016).
| (12) |
| (13) |
Another possible explanation for the observed increases in pH may result from off gassing of hydrogen gas produced via reduction of water at the cathode (Eq. 14) (Brillas and Martíne-zHuitle, 2015).
| (14) |
Some combination of these reactions, favoring those that lead to the observed pH increases, are likely occurring in the system presented in this study. Complex interactions have been reported with the utilization of BDD anodes for electro-oxidation (Cui et al., 2009; Faouzi Elahmadi et al., 2009; Ratiu et al., 2010) impacting the pH. In this study, usage of 0.05N NaNO3 electrolyte impacted pH less than the other electrolytes.
3.1.3. Minimizing Impacts of Secondary Oxidants
In addition to pH concerns, another potential microbial toxicity concern is that electrolyte interactions can form secondary oxidants, which may exhibit toxicity and require oxidant quenching. Secondary oxidants, including persulfates, percarbonates, perphosphates, chlorine, and even chlorate and bromate can be produced on many anodes, (Panizza and Cerisola, 2009; Sirés et al., 2014) complicating electrolyte selection in a given application by confounding data interpretation during system implementation. For example, the NaCl electrolyte produced oxidant(s) measuring up to 330 mg L−1 as active chlorine during the two-hour treatment time of the 1.0 N NaCl blank trial. These DPD-measurable oxidants were readily quenched with sodium thiosulfate. Toxic effects were observed at the highest electrolyte concentrations of 0.5–1.0 N NaCl, but Microtox® toxicity remained below 20% for the lower electrolyte concentrations.
Because chloride salts are widely available and might be used despite toxicity concerns, further investigation of the oxidant(s) detected at very high levels using the NaCl electrolyte may be necessary. Two test compounds were chosen, one that reacts readily with chlorine and another that does not. Experiments were performed to degrade either BPA or tris(2-chloroethyl) phosphate (TCEP) utilizing an electro-oxidation with a BDD anode, comparing the NaCl or the NaNO3 electrolytes. The results are summarized in Figure 1. TCEP, which is not expected to react readily with active chlorine, was degraded at a similar rate regardless of which electrolyte was used. BPA, which is expected to react readily with active chlorine, was degraded in less than three minutes of treatment using the NaCl electrolyte, even with very little time for the chlorine concentration to build. The reaction progressed much more slowly in the presence of NaNO3, which is not expected to participate in the reaction or produce reactive chlorine. Therefore, this electrochemical oxidation using a BDD anode, in the presence of the NaCl electrolyte, likely produces large concentrations of active chlorine, which may contribute to the degradation of contaminants, similar to reports with other electrodes (Murugananthan et al., 2011; Saylor et al., 2012). Although not directly measured in this study, the formation of organochlorinated intermediates, as well as chlorate, may present further concerns when using the NaCl electrolyte (Sirés et al., 2014).
Figure 1:
Degradation of tris(2-chloroethyl) phosphate(TCEP) and Bisphenol A(BPA) in an unbuffered electro-oxidation using a BDD anode with 0.05 N NaCl or NaNO3 electrolytes and 4.4 A applied current.
Prior to electro-oxidative treatment using the BDD anode, the DPD test did not indicate the presence of oxidants for the Na2SO4 electrolyte solutions. However, after treatment, the presence of persistent oxidants was observed. The 0.5 N solution produced oxidants equivalent to up to 103 mg L−1 as Cl2 almost two months after sample generation. Further trials with 0.05 N Na2SO4 indicated that the DPD color change continued for 20 minutes or more after DPD was mixed with the samples, as well as after quenching with sodium thiosulfate. This indicates that some portion of the oxidants formed react more slowly with DPD than chlorine, hydrogen peroxide, or chloramines and that they are not readily quenched by sodium thiosulfate. Indeed, persulfate reactions with sodium thiosulfate have been reported, but long reaction times may be necessary to achieve oxidant removal (Olmez-Hanci et al., 2014). Even after quenching, all concentrations of the Na2SO4 electrolyte exhibited variable toxicities. Together, these factors render this electrolyte less applicable.
The NaNO3 and NaH2PO4 electrolytes only measured up to 1.5 mg L−1 and 2.4 mg L−1 oxidant as Cl2, respectively. The NaNO3 electrolyte yielded greater Microtox® toxicities for higher normalities. However, no discernible trend was found between NaH2PO4 electrolyte concentration and toxicity. These results indicate that NaNO3 might be more inert than NaCl or Na2SO4 during electro-oxidation using a BDD anode and that, from a microbial toxicity standpoint, it could be advantageous over the other electrolytes at concentrations of approximately 0.05N.
3.2. Investigation of effect of electrolyte on toxicity and degradation
In light of the general greater toxicity and pH changes at higher normality, the 0.05 N electrolyte concentration was chosen for further trials with propanil as a model contaminant. The NaCl, NaNO3, Na2SO4, and NaH2PO4 electrolytes were studied to determine their simultaneous effects on contaminant degradation, as well as on microbial toxicity. Propanil degradation and the corresponding Microtox® toxicities are presented in Figure 2 (note that analytical variability contributes to large fluctuation in C C0−1 when both values are large). After two hours of treatment, the Na2SO4 electrolyte exhibited 100% propanil degradation, followed by NaH2PO4, (73% degradation), and NaNO3 (56% degradation). The NaCl electrolyte, frequently utilized in the literature and sometimes recommended by manufacturers, exhibited the slowest propanil degradation, with only 36% degradation after two hours of treatment. The propanil degradation was sensitive to electrolyte and followed the overall trend: Na2SO4 > NaH2PO4 > NaNO3 > NaCl.
Figure 2:
Propanil degradation (a) and Microtox® toxicity (b) using unbuffered electro-oxidation using a BDD anode with 0.05 N electrolytes and 4.4 A or 8.8 A (higher current) applied current.
The Microtox® toxicity (Figure 2b), however, did not follow this trend with the final Microtox® toxicities reading: Na2SO4 < NaCl < NaH2PO4 (quenched with sodium thiosulfate) < NaNO3 < NaH2PO4 (unquenched). The final Na2SO4 and NaCl samples did not exhibit Microtox® toxicities above the 10% toxicity threshold (Sánchez et al., 2014), with final toxicities of 5% ± 1% and 11% ± 2%, respectively. The solution treated with the NaH2PO4 electrolyte exhibited 92% ± 1% toxicity after two hours of treatment, though once quenched, the final toxicity decreased to 23% ± 1%. This toxicity was not observed in the blank and may indicate some secondary oxidant formation/reactions contributing to microbial toxicity. The NaNO3 exhibited a slightly greater toxicity to that of the quenched NaH2PO4 with 32 ± 1% inhibition after the two-hour treatment time. Interestingly, in Figure 2b, the NaNO3 samples showed increased toxicities within the first 10 minutes of treatment, then a subsequent decrease in toxicity.
In spite of exhibiting lower contaminant degradation compared to several other electrolytes, 0.05 N NaNO3 electrolyte was chosen for the majority of further trials due to lack of demonstrated toxicity and other process concerns such as inertness and buffering capacity. The NaNO3 electrolyte has been reported to be inert during electro-oxidation with BDD anodes, meaning that the electrolyte does not participate in reactions nor produce secondary oxidants, so the majority of the resulting contaminant degradation would likely result only from reactions with hydroxyl radicals (Murugananthan et al., 2011). Further, the NaNO3 avoids several pitfalls of other electrolytes such as the potential for producing chlorinated byproducts as with NaCl, along with the toxicity and pH adjustment issues related to the buffering capacity of NaH2PO4. In addition, the nitrate electrolyte itself is not expected to adversely impact biological processes and many utilities already have processes in place for its removal because it is a common nutrient. Since electrolytes are not expected to be present in many of the contaminated waters arising from incident response situations, waters may need to be amended with the nitrate electrolyte prior to electrochemical treatment with BDD anodes.
3.3. Treatment Performance and Toxicity for a Variety of Classes of Contaminants
A concentration of 0.05 N NaNO3 electrolyte was used to investigate the ability of electro-oxidation using a BDD anode to degrade a variety of contaminants. As detailed in Table S1, these contaminants include herbicides and pesticides such as propanil, carbofuran, aldicarb, atrazine and cyanazine; flame retardants such as PFOA, diethyl methyl phosphonate (DEMP) and TCEP; and pharmaceuticals/contaminants of interest such as BPA, carbamazepine, and phenylephrine. These contaminants were selected to cover a range of reported hydroxyl radical reaction rates and chemical structures (listed in Table S1), as well as potential contamination incident concerns. Contamination concentrations ranged up to 10 mg L−1, to represent concentrations that might be generated during some contamination scenarios.
Contaminant degradation and the corresponding Microtox® toxicity for these contaminants are presented in Figure 3. Contaminant degradation profiles for phenylephrine and aldicarb are not presented in Figure 3a due to analytical detection issues. Contaminants that consistently fell below the 10% Microtox® toxicity threshold, including TCEP, cyanazine, DEMP and PFOA, are not shown in Figure 3b (to yield a more readable figure). Data in Figure 3 demonstrate that the extent of electro-oxidation using a BDD anode is very contaminant-dependent, under the conditions tested. For example, carbofuran was nearly completely degraded, with 95% degradation in the two-hour treatment time, whereas PFOA and DEMP exhibited very little degradation with 13% and 15% degradation, respectively.
Figure 3:
Degradation (a) and Microtox® toxicity (b) of various contaminants using unbuffered anodic oxidation using a BDD anode with 0.05 N NaNO3 electrolyte and 4.4 A applied current.
Figure 3b demonstrates a variety of microbial toxicity behaviors that do not necessarily follow the corresponding contaminant degradation data. In particular, carbofuran, the contaminant with near complete degradation, also exhibited the highest Microtox® toxicity by the end of the two-hour treatment time. These results for carbofuran suggest that degradation intermediates are more toxic than the parent compound and, coupled with its near complete degradation, these intermediates may have been produced in quantities large enough to exhibit Microtox® toxicity. Accordingly, it is not unreasonable that some other compounds have toxic degradation intermediates as well, but either the parent compounds were refractory enough during the two-hour treatment time or that the intermediates were degraded quickly enough that the intermediates were not present in sufficient quantity to exhibit toxic effects.
Several compounds in this study, like BPA and propanil, showed measurable degradation that did not correspond to similar reductions in Microtox® toxicity. In contrast, other contaminants like DEMP and PFOA exhibited very little degradation but were consistently below the toxicity threshold. Microtox® toxicities for carbofuran, carbamazepine, atrazine, propanil increased during treatment. These multi-contaminant degradation and toxicity results demonstrate that parent contaminant degradation alone is not an adequate measure of treatment, and that Microtox® toxicity could be an important tool for understanding the effectiveness and implications of electro-oxidation of contaminated waters using a BDD anode.
While the Microtox® assay may be a useful indicator of ecotoxicity, it has been reported to be a poor indicator of microbial toxicity for activated sludge microbes due to its sensitivity (Dalzell et al., 2002). NI results, presented in Table 1, better indicate potential effects of treatment on activated sludge organisms for this study. A 50% inhibition benchmark for NI is used by some utilities to indicate a level of concern warranting further testing. This benchmark is based on research studying inhibition by heavy metals (Juliastuti et al., 2003; Hartmann et al., 2013). These toxicity thresholds are operational and can be defined at various levels based on the toxicity test and the individual situation. These inhibition results are generally categorized from 20% to 50% inhibition for similar activated sludge test methods (Juliastuti et al., 2003). Results show that only the pre-EAOP treatment sample containing propanil exhibited toxicity within its error of the 50% inhibition benchmark. Many of the pre- and post-treatment NI toxicity results are not significantly different from each other at the 95% confidence interval. Further, observed negative inhibition results indicate that, in some cases, EAOP-treated contaminated waters promoted nitrification in activated sludge. These results suggest that activated sludge is fairly resilient to contaminant loading, as well as to impact by EAOP-treated effluent. These results generally demonstrate low NI toxicity and should bolster the confidence—but not eliminate all concerns—of wastewater treatment works in accepting a variety of contaminated waters.
Table 1:
Nitrification Inhibition Percentages upon Electro-Oxidation with a BDD Anode. Inhibitions above or within the error of the 50% threshold are bolded.
| Compound | Pre-EAOP NI | % Error | Post-EAOP NI | % Error |
|---|---|---|---|---|
| Carbamazepine | −6% | 2% | 17% | 8% |
| Aldicarb | 3% | 32% | 7% | 5% |
| Phenylephrine | 4% | 5% | 12% | 5% |
| BPA | 0% | 12% | 2% | 4% |
| Carbofuran | 6% | 12% | 23% | 5% |
| Atrazine | −23% | 30% | 8% | 25% |
| Propanil | 42% | 13% | 29% | 18% |
| Cyanazine | 6% | 7% | 27% | 9% |
| DEMP | 8% | 20% | 17% | 4% |
| TCEP | 3% | 6% | −9% | 14% |
| PFOA | −25% | 10% | 22% | 22% |
4. Conclusions
This study provides systematic information about the impacts of background electrolytes on pH, potential secondary oxidant formation, contaminant degradation, and microbial toxicity upon electro-oxidation using BDD anodes. This study is particularly valuable for practical applications of EAOPs due to its systematic evaluation of toxicity resulting from various electrolytes and contaminant classes. The optimal electrolyte for this study, nitrate, balances competing factors and is compatible with current nutrient removal treatment processes.
The extent of contaminant degradation alone is not enough to monitor reaction progress nor infer that effluents are less toxic to receiving waters, highlighting the importance of testing microbial toxicity. Further, although the Microtox® toxicity test may be a good screening tool for some contaminants, its results may not be indicative of activated sludge behavior. Overall, in spite of incomplete contaminant degradation, observed NI results may bolster the confidence of wastewater utilities in the resiliency of their activated sludge processes to a variety of contaminants of concern, including effluents from EAOP treatment using BDD anodes.
This study demonstrates the applicability of electro-oxidation using the BDD anode toward treatment of refractory organic compounds in waters generated by intentional and unintentional incidents. These results address immediate knowledge gaps and help provide decision makers with an additional treatment tool in a toolbox approach to address portable and incident/situation-specific goals. Studies comparing this electro-oxidation system using a BDD anode with O3/H2O2 and an UV/H2O2 AOPs will be presented in a future publication.
Supplementary Material
Acknowledgments
The U.S. Environmental Protection Agency through its Office of Research and Development funded and collaborated in the research described here under Contract EP-C-10–001. It has been subjected to the Agency’s review and has been approved for public release. EPA does not endorse the purchase or sale of any commercial products or services. This project was supported in part by an appointment to the Research Participation Program at US EPA, administered by the Oak Ridge Institute for Science and Education through an interagency agreement between the U.S. Department of Energy and EPA. This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Footnotes
Supplementary Material
Selected physicochemical properties and LC-MS/MS conditions for target contaminants
References
- 2014. Method 8021: Chlorine, Free, USEPA DPD Method (2.00 mg/L). Hach Company. [Google Scholar]
- 2015. Method 10031: Nitrogen, Ammonia, Salicylate Method (50.0 mg/L). in: Company H. (Ed.). [Google Scholar]
- Anglada Á, Urtiaga A, Ortiz I, Mantzavinos D, Diamadopoulos E, 2011. Boron-doped diamond 452 anodic treatment of landfill leachate: Evaluation of operating variables and formation of oxidation by-453 products. Water Research 45, 828–838. [DOI] [PubMed] [Google Scholar]
- Brillas E, Martínez-Huitle CA, 2015. Decontamination of wastewaters containing synthetic organic dyes by electrochemical methods. An updated review. Applied Catalysis B: Environmental 166–167, 603–643. [Google Scholar]
- Cui Y. h., Li X. y., Chen G, 2009. Electrochemical degradation of bisphenol A on different anodes. Water Research 43, 1968–1976. [DOI] [PubMed] [Google Scholar]
- Dalzell D, Alte S, Aspichueta E, De la Sota A, Etxebarria J, Gutierrez M, Hoffmann C, Sales D, Obst U, Christofi N, 2002. A comparison of five rapid direct toxicity assessment methods to determine toxicity of pollutants to activated sludge. Chemosphere 47, 535–545. [DOI] [PubMed] [Google Scholar]
- Faouzi Elahmadi M, Bensalah N, Gadri A, 2009. Treatment of aqueous wastes contaminated with Congo Red dye by electrochemical oxidation and ozonation processes. Journal of Hazardous Materials 168, 1163–1169. [DOI] [PubMed] [Google Scholar]
- Fox JT, Brandriff CJ, Bott CB, 2006. Assessing the potential for nitrification inhibition at wastewater treatment facilities as a result of zinc orthophosphate addition to potable water distribution systems. Proceedings of the Water Environment Federation 2006, 6593–6622. [Google Scholar]
- Haidar M, Dirany A, Sirés I, Oturan N, Oturan MA, 2013. Electrochemical degradation of the antibiotic sulfachloropyridazine by hydroxyl radicals generated at a BDD anode. Chemosphere 91, 1304–1309. [DOI] [PubMed] [Google Scholar]
- Hammami S, Bellakhal N, Oturan N, Oturan MA, Dachraoui M, 2008. Degradation of Acid Orange 7 by electrochemically generated •OH radicals in acidic aqueous medium using a boron-doped diamond or platinum anode: A mechanistic study. Chemosphere 73, 678–684. [DOI] [PubMed] [Google Scholar]
- Hartmann S, Skrobankova H, Drozdova J, 2013. Inhibition of activated sludge respiration by heavy metals. Proceedings of the 2013 International Conference on Environment, Energy, Ecosystem and Development, pp. 28–30. [Google Scholar]
- Jalife-Jacobo H, Feria-Reyes R, Serrano-Torres O, Gutiérrez-Granados S, Peralta-Hernández JM, 2016. Diazo dye Congo Red degradation using a Boron-doped diamond anode: An experimental study on the effect of supporting electrolytes. Journal of Hazardous Materials 319, 78–83. [DOI] [PubMed] [Google Scholar]
- Juliastuti S, Baeyens J, Creemers C, 2003. Inhibition of nitrification by heavy metals and organic compounds: the ISO 9509 test. Environmental Engineering Science 20, 79–90. [Google Scholar]
- Koparal AS, Yavuz Y, Gürel C, Öğütveren ÜB, 2007. Electrochemical degradation and toxicity reduction of CI Basic Red 29 solution and textile wastewater by using diamond anode. Journal of Hazardous Materials 145, 100–108. [DOI] [PubMed] [Google Scholar]
- Kornienko GV, Chaenko NV, Maksimov NG, Kornienko VL, Varnin VP, 2011. Electrochemical oxidation of phenol on boron-doped diamond electrode. Russ J Electrochem 47, 225–229. [Google Scholar]
- Lévy-Clément C, Ndao N, Katty A, Bernard M, Deneuville A, Comninellis C, Fujishima A, 2003. Boron doped diamond electrodes for nitrate elimination in concentrated wastewater. Diamond and Related Materials 12, 606–612. [Google Scholar]
- Martin de Vidales MJ, Millán M, Sáez C, Cañizares P, Rodrigo MA, 2016. What happens to inorganic nitrogen species during conductive diamond electrochemical oxidation of real wastewater? Electrochemistry Communications 67, 65–68. [Google Scholar]
- Martínez-Huitle CA, Brillas E, 2009. Decontamination of wastewaters containing synthetic organic dyes by electrochemical methods: a general review. Applied Catalysis B: Environmental 87, 105–145. [Google Scholar]
- Montalbán MG, Hidalgo JM, Collado-González M, Díaz Baños FG, Víllora G, 2016. Assessing chemical toxicity of ionic liquids on Vibrio fischeri: Correlation with structure and composition. Chemosphere 155, 405–414. [DOI] [PubMed] [Google Scholar]
- Murugananthan M, Latha SS, Bhaskar Raju G, Yoshihara S, 2011. Role of electrolyte on anodic mineralization of atenolol at boron doped diamond and Pt electrodes. Separation and Purification Technology 79, 56–62. [Google Scholar]
- Murugananthan M, Yoshihara S, Rakuma T, Shirakashi T, 2008. Mineralization of bisphenol A (BPA) by anodic oxidation with boron-doped diamond (BDD) electrode. Journal of Hazardous Materials 154, 213–220. [DOI] [PubMed] [Google Scholar]
- Olmez-Hanci T, Arslan-Alaton I, Dursun D, 2014. Investigation of the toxicity of common oxidants used in advanced oxidation processes and their quenching agents. Journal of Hazardous Materials 278, 330–335. [DOI] [PubMed] [Google Scholar]
- Özcan A, Şahin Y, Koparal AS, Oturan MA, 2008. Propham mineralization in aqueous medium by anodic oxidation using boron-doped diamond anode: Influence of experimental parameters on degradation kinetics and mineralization efficiency. Water Research 42, 2889–2898. [DOI] [PubMed] [Google Scholar]
- Pagga U, Bachner J, Strotmann U, 2006. Inhibition of nitrification in laboratory tests and model wastewater treatment plants. Chemosphere 65, 1–8. [DOI] [PubMed] [Google Scholar]
- Panizza M, Cerisola G, 2009. Direct And Mediated Anodic Oxidation of Organic Pollutants. Chemical Reviews 109, 6541–6569. [DOI] [PubMed] [Google Scholar]
- Ratiu C, Manea F, Lazau C, Grozescu I, Radovan C, Schoonman J, 2010. Electrochemical oxidation of p-aminophenol from water with boron-doped diamond anodes and assisted photocatalytically by TiO2-supported zeolite. Desalination 260, 51–56. [Google Scholar]
- Sánchez OG, Arana J, Díaz OG, Melián JH, Rodríguez JD, Peña JP, 2014. Detoxification of the herbicide propanil by means of Fenton process and TiO2 photocatalysis. Journal of Photochemistry and Photobiology A: Chemistry 291, 34–43. [Google Scholar]
- Saylor GL, Chen L, Kupferle MJ, 2012. Using toxicity testing to evaluate electrochemical reactor operations. Environmental Toxicology and Chemistry 31, 494–500. [DOI] [PubMed] [Google Scholar]
- Sirés I, Brillas E, Oturan MA, Rodrigo MA, Panizza M, 2014. Electrochemical advanced oxidation processes: today and tomorrow. A review. Environmental Science and Pollution Research 21, 8336–8367. [DOI] [PubMed] [Google Scholar]
- Striano E, Woltering DM, Pramanik A, 2011. Expert Workshop on Toxicity Testing of Water Undergoing Advanced Oxidation Processes Prior to Discharge. Water Environment Research Foundation, Alexandria, VA. [Google Scholar]
- Uranga-Flores A, de la Rosa-Júarez C, Gutierrez-Granados S, de Moura DC, Martínez-Huitle CA, Peralta Hernández JM, 2015. Electrochemical promotion of strong oxidants to degrade Acid Red 211: Effect of supporting electrolytes. J. Electroanal. Chem 738, 84–91. [Google Scholar]
- Velegraki T, Balayiannis G, Diamadopoulos E, Katsaounis A, Mantzavinos D, 2010. Electrochemical oxidation of benzoic acid in water over boron-doped diamond electrodes: Statistical analysis of key operating parameters, kinetic modeling, reaction by-products and ecotoxicity. Chemical Engineering Journal 160, 538–548. [Google Scholar]
- Yavuz Y, Savaş Koparal A, Öğütveren ÜB, 2011. Electrochemical oxidation of Basic Blue 3 dye using a diamond anode: evaluation of colour, COD and toxicity removal. Journal of Chemical Technology & Biotechnology 86, 261–265. [Google Scholar]
- Yavuz Y, Shahbazi R, 2012. Anodic oxidation of Reactive Black 5 dye using boron doped diamond anodes in a bipolar trickle tower reactor. Separation and Purification Technology 85, 130–136. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.



