. 2018 Jan 11;12(5):1061–1063. doi: 10.1177/1932296817753629

Blood Glucose Monitoring Data Should Be Reported in Detail When Studies About Efficacy of Continuous Glucose Monitoring Systems Are Published

Stefan Pleus 1,, Lutz Heinemann 2, Guido Freckmann 1
PMCID: PMC6134630  PMID: 29322825

Abstract

Recently, two clinical trials about a “sensor-based flash glucose monitoring system” and its efficacy in reducing time in hypoglycemia were published. Interestingly, patients spent more time at low glucose concentrations in these studies than in other studies related to the efficacy of real-time continuous glucose monitoring (rtCGM). Although it is possible that the study populations differed from those in other studies, another potential explanation is that the CGM system used in these two studies had a negative glucose measurement bias. Such a negative bias was reported in recent literature, suggesting that the CGM system may inaccurately indicate hypoglycemia. Reporting blood glucose monitoring data would help to interpret the CGM data at least in the context of time spent in various glucose ranges as a parameter with which quality of diabetes therapy is measured.

Keywords: blood glucose monitoring, continuous glucose monitoring, clinical trial, hypoglycemia


Data from two clinical trials with a “sensor-based flash glucose monitoring system” and its efficacy in reducing time in hypoglycemia (TiH) were published.1,2 In the trial with patients with type 1 diabetes conducted by Bolinder et al.1 TiH was the trial’s primary endpoint. Blood glucose monitoring (BGM) data were not reported in these publications; however, as outlined subsequently these would be of help to interpret the study results.

A literature search for trials regarding the effect of real-time continuous glucose monitoring (rtCGM) or intermittent-scanning CGM (isCGM) on diabetes therapy outcomes that were published within the last 5 years yielded 50 results. A timeframe of the last 5 years was chosen to reflect the increasing analytical accuracy of newer rtCGM system generations that became commercially available, because accuracy may affect time spent in hypoglycemia as measured by the rtCGM system. Of these 50 publications, 12 were excluded because the number of participants was comparably small, and another 29 publications were excluded because they did not provide data for HbA1c or time spent <70 mg/dL (3.9 mmol/L) or both. The remaining 9 publications were analyzed for TiH, defined as time spent <70 mg/dL (3.9 mmol/L) (see Figure 1).1-9 Figure 1 shows only baseline data from these studies: intervention group results are affected by the intervention, and even control group results might be influenced by a potential study effect. Weighted mean values were calculated for TiH and HbA1c based on the individual results presented in the publications and the number of patients in the intervention and control groups.

Figure 1.

Figure 1.

Time spent <70 mg/dL (3.9 mmol/L) measured by rtCGM (blue) or isCGM (orange) glucose levels in various studies. Measurements were performed in patients with type 1 or type 2 diabetes and different baseline HbA1c values.

*Ly et al3 and Heinemann et al4 actively recruited participants with hypoglycemia unawareness.

For Lind et al5 time spent <70 mg/dL (3.9 mmol/L) and HbA1c were reported for different numbers of patients.

We would like to focus on two specific studies with rtCGM systems, published by Aleppo et al7 and by Beck et al.9 The second study was the only suitable study with patients with type 2 diabetes. The first study was published recently and had a large number of participants whose HbA1c was comparable to that of the participants studied by Bolinder et al.1

Patients in both studies seemed to spend considerably less TiH than in the isCGM studies: For the “well-controlled” adults (average HbA1c 6.8% [51 mmol/mol]) with type 1 diabetes studied by Bolinder et al,1 TiH was approximately 205 min per day at baseline. This is well above the approximately 45 min for adults with comparable HbA1c values (7.1% [53 mmol/mol]) studied by Aleppo et al.7 For the adults with type 2 diabetes (HbA1c 8.7% [72 mmol/mol]) who participated in the study by Haak et al,2 TiH was 74 min, whereas Beck et al9 reported 12 min for adults with type 2 diabetes, who had similar HbA1c (8.5% [69 mmol/mol]).

Clearly it is possible that the populations in the two studies with the isCGM system differ from those in the other studies, because inclusion and exclusion criteria were not identical in these studies. For example, Aleppo et al7 did exclude subjects who experienced severe hypoglycemia, whereas Bolinder et al1 did not. Haak et al2 included people using continuous subcutaneous insulin infusion, whereas Beck et al9 did not.

However, another potential explanation is that the isCGM system had a negative systematic measurement difference (bias) with respect to BGM. If such a bias existed, it would (artificially) increase TiH at baseline. Such a bias could also be a possible explanation why in the study by Haak et al2 HbA1c decreased slightly less (difference not statistically significant) in the intervention group using nonblinded isCGM than in the control group using regular therapy or blinded isCGM.

A negative bias was reported in a recent publication for the isCGM system when BGM were ≤150 mg/dL (≤8.3 mmol/L).10 Food and Drug Administration (FDA) premarket approval data for this isCGM system also indicate a negative bias.11 For the professional-use version of this isCGM system, the FDA explicitly noted in its premarket approval that this system may inaccurately indicate hypoglycemia, because in 40% of cases in which sensor glucose readings were <60 mg/dL (<3.3 mmol/L), BGM were in the range of 81 to 160 mg/dL (4.5 to 8.9 mmol/L).12

Therefore, we suggest that neither rtCGM nor isCGM data are reported without accompanying BGM data; at least in the context of time spent in various glucose ranges as a parameter with which quality of diabetes therapy is measured. Reporting BGM data would help to interpret rtCGM/isCGM data. If a considerable systematic measurement difference (bias) between CGM and BGM data were found, acknowledging this bias in publications would also be helpful. Quantitative analysis of bias, if one exists, could possibly provide additional information. It should not be forgotten that laboratory analyzers and even most of the hand-held BGM systems for self-use still provide more accurate glucose measurement results than the current generation of rtCGM/isCGM systems.

Recommendation

  • If rtCGM or isCGM data are reported in the context of a CGM system’s impact on time spent in various glucose ranges, BGM data should also be provided.

Footnotes

Abbreviations: BGM, blood glucose monitoring; CGM, continuous glucose monitoring; FDA, Food and Drug Administration; isCGM, intermittent-scanning continuous glucose monitoring; rtCGM, real-time continuous glucose monitoring; TiH, time in hypoglycemia.

Declaration of Conflicting Interests: The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: SP is employee of the IDT (Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany). LH is member of advisory boards/consultant for a number of companies that are developing diagnostic and therapeutic options. He is shareholder of the Profil Institut für Stoffwechselforschung, Neuss, Germany, and ProSciento, San Diego, CA, USA. GF is general manager of IDT, which carries out clinical studies on the evaluation of BG meters and medical devices for diabetes therapy on its own initiative and on behalf of various companies. GF/IDT have received speakers’ honoraria or consulting fees from Abbott, Ascensia, Bayer, Berlin-Chemie, Becton-Dickinson, Dexcom, LifeScan, Menarini Diagnostics, Novo Nordisk, Roche, Sanofi, Sensile, and Ypsomed.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

References

  • 1. Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kroger J, Weitgasser R. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet. 2016;388(10057):2254-2263. [DOI] [PubMed] [Google Scholar]
  • 2. Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, Rayman G. Flash glucose-sensing technology as a replacement for blood glucose monitoring for the management of insulin-treated type 2 diabetes: a multicenter, open-label randomized controlled trial. Diabetes Ther. 2017;8(1):55-73. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Ly TT, Nicholas JA, Retterath A, Lim EM, Davis EA, Jones TW. Effect of sensor-augmented insulin pump therapy and automated insulin suspension vs standard insulin pump therapy on hypoglycemia in patients with type 1 diabetes: a randomized clinical trial. JAMA. 2013;310(12):1240-1247. [DOI] [PubMed] [Google Scholar]
  • 4. Heinemann L, Freckmann G, Faber-Heinemann G, Ehrmann D, Hermanns N. Can CGM discriminate between type 1 diabetes patients with and without severe hypoglycaemia? Results of the baseline assessment from the HypoDE study. Available at: https://www.easd.org/virtualmeeting/home.html#!resources/can-cgm-discriminate-between-type-1-diabetes-patients-with-and-without-severe-hypoglycaemia-results-of-the-baseline-assessment-from-the-hypode-study-f8071a79-3608-4791-8263-8bba22480c05. Accessed November 9, 2017.
  • 5. Lind M, Polonsky W, Hirsch IB, et al. Continuous glucose monitoring vs conventional therapy for glycemic control in adults with type 1 diabetes treated with multiple daily insulin injections: the GOLD randomized clinical trial. JAMA. 2017;317(4):379-387. [DOI] [PubMed] [Google Scholar]
  • 6. Beck RW, Riddlesworth T, Ruedy K, et al. Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial. JAMA. 2017;317(4):371-378. [DOI] [PubMed] [Google Scholar]
  • 7. Aleppo G, Ruedy KJ, Riddlesworth TD, et al. REPLACE-BG: a randomized trial comparing continuous glucose monitoring with and without routine blood glucose monitoring in adults with well-controlled type 1 diabetes. Diabetes Care. 2017;40(4):538-545. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Mauras N, Beck R, Xing D, et al. A randomized clinical trial to assess the efficacy and safety of real-time continuous glucose monitoring in the management of type 1 diabetes in young children aged 4 to <10 years. Diabetes Care. 2012;35(2):204-210. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Beck RW, Riddlesworth TD, Ruedy K, et al. Continuous glucose monitoring versus usual care in patients with type 2 diabetes receiving multiple daily insulin injections: a randomized trial. Ann Intern Med. 2017;167(6):365-374. [DOI] [PubMed] [Google Scholar]
  • 10. Fokkert MJ, van Dijk PR, Edens MA, et al. Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. BMJ Open Diabetes Res Care. 2017;5(1):e000320. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Food and Drug Administration. Premarket approval (PMA): FreeStyle Libre Flash glucose monitoring system (P160030). Available at: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P160030. Accessed November 13, 2017.
  • 12. Food and Drug Administration. Premarket approval (PMA): FreeStyle Libre Pro Flash glucose monitoring system (P150021). Available at: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P150021. Accessed November 13, 2017.